idi Aspect100HT Operator's Manual
Read Online
Read the idi Aspect100HT Operator's Manual online using the built-in PDF viewer.
Text of the manual
Operator Manual Aspect 100HT
Made in the USA
Aspect100HT
High-Tilt
Mobile Imaging Table
Operator Manual
Operator Manual Aspect 100HT
L100-3064 REV A
www.imagediagnostics.com
ii
Table of Contents
PRODUCT DATA ............................................................................................................................................ 1
INTRODUCTION ............................................................................................................................................ 2
CUSTOMER SUPPORT ................................................................................................................................... 3
SYMBOLS AND DATA PLATE ......................................................................................................................... 4
SPECIFICATIONS ............................................................................................................................................ 6
ELECTROMAGNETIC COMPATIBILITY STATEMENT ...................................................................................... 7
SETUP ............................................................................................................................................................ 8
SAFETY HAZARDS.......................................................................................................................................... 9
SAFETY HAZARD ALERTS ............................................................................................................................ 10
SAFETY INSTRUCTIONS ............................................................................................................................... 10
OPERATOR CONTROLS ............................................................................................................................... 14
EMERGENCY STOP SWITCH ........................................................................................................................ 16
EQUIPOTENTIAL TERMINAL ....................................................................................................................... 16
ELECTRIC SYSTEM ....................................................................................................................................... 17
MAINTENANCE ........................................................................................................................................... 19
USER REPLACEMENT ITEMS ....................................................................................................................... 21
DISPOSAL OF COMPONENTS ...................................................................................................................... 22
WARRANTY ................................................................................................................................................. 23
Linak documentation can be found at their web site: www.linak-us.com
Select “Products | User manuals” and click on “Actuators & Electronics”.
The text of this manual was originally written, approved
and published by the manufacturer in English.
Operator Manual Aspect 100HT
L100-3064 REV A
www.imagediagnostics.com
1
PRODUCT DATA
• Low attenuation carbon fiber tabletop
• Low attenuation patient pad
• Tabletop: 61cm x 203cm (24in x 80in)
• Tilt: 0° (Horizontal) to 75°
• Height: 81cm (32in)
• Patient capacity: 238kg (525lbs)
• Locking swivel casters
• Foot operator control standard
• Emergency Stop Switch
• Equipotential Terminal
• Foot Rest
• Optional hand operator control
Operator Manual Aspect 100HT
L100-3064 REV A
www.imagediagnostics.com
2
INTRODUCTION
Overview
This manual pertains to the specified product only. It is intended for qualified medical
personnel who have been trained in the use of medical equipment. It is not designed to
replace or substitute for certified training in the application of this equipment.
Functional capabilities and operation of the equipment described herein which can be
employed in a variety of diagnostic, therapeutic, and surgical applications. It is designed
for use as either a fluoroscopic or radiographic table.
Owner Responsibilities
The owner is responsible to ensure system compatibility, the qualifications of the
operator and maintenance personnel. The operator must be properly trained and have
obtained credentials from the appropriate authorities.
This equipment should only be installed in an area provided with the proper electrical
power.
The owner is responsible for verifying continued compliance with all applicable
regulations and standards. Consult local, state, federal and/or international agencies
regarding specific requirements and regulations applicable to the use of this equipment.
Image Diagnostics, Inc. certifies this equipment. After-sale operating practices and safety
are the responsibility of the owner and operator. Image Diagnostics, Inc. assumes no
liability or responsibility for after-sale operating or safety practices; nor can it be
responsible for personal injury or damage resulting from misuse.
Never make modifications or adjustments to the equipment unless directed by a qualified
Image Diagnostics representative. This equipment, when properly assembled, meets US
federal regulations and international standards. Unauthorized modifications to the
equipment may impact adherence to these standards and make the equipment unsafe to
operate.
Operator Manual Aspect 100HT
L100-3064 REV A
www.imagediagnostics.com
3
CUSTOMER SUPPORT
Image Diagnostics will make available, on request, circuit diagrams,
component part lists, calibration instructions, or other information which
will assist the user’s appropriately qualified technical personnel to repair
those parts of equipment which are designated by the manufacturer as
repairable.
For technical assistance, call IDI at (978) 829-0009 or send fax to (978)
829-0027. Be prepared to give the complete model and serial number
found on the dataplate on the table base at the time of contact.
This X-ray component manufactured by Image Diagnostics, Inc. complies
with applicable FDA performance standards contained in 21CFR at date of
manufacture.
Operator Manual Aspect 100HT
L100-3064 REV A
www.imagediagnostics.com
4
SYMBOLS AND DATA PLATE
Intertek Testing Service
The system was tested and found to be in compliance with the
requirements of all relevant directives and standards in effect within
the European Union at the time of manufacture.
Attention! Consult accompanying documents.
Failure to follow these instructions can cause accidents resulting in
serious personal injury or damage to equipment. CD ROMs
containing file copies of all relevant drawings, BOMs, and
documentation are included with this manual, one for general
operators and one for the BIOMED department.
Type B equipment
Alternating current
Data Plate
Maximum Patient Weight
Equipotential Terminal (Featured on some models)
Indicates the Equipotential terminal of the table. Provides for a
connection between the table and the equipotential bus bar of the
facility.
Emergency Stop Button
Operator Manual Aspect 100HT
L100-3064 REV A
www.imagediagnostics.com
5
LATEX
Recycle
Some of the material can be recycled rather than discarded.
Not for Patient Transport. The table should not be relocated with a
patient on it nor should it be used to move a patient.
Do Not Sit
This label is found on parts that are not designed to support the
sitting weight of a patient.
Not made with natural rubber latex.
Single Use item.
This symbol, when used in this manual, indicates the potential of
exposure to harmful x-rays. Be sure to read and comply with all
warnings.
European Authorized
Representative
Warning! Information or instructions shown near this symbol must
be adhered to in order to prevent a potentially hazardous situation
which if not avoided, could result in death, personal injury or
damage to the equipment.
Protective Ground
This is the common tie point between the AC cord ground, frame
ground, and controller ground.
2
EC
REP
Advena Ltd. Pure Offices, Plato Close,
WARWICK CV34 6WE UK
Operator Manual Aspect 100HT
L100-3064 REV A
www.imagediagnostics.com
6
SPECIFICATIONS
Mode of Operation:
For continuous operation with short time loading
Duty Cycle: 10% (6 min/hr)
Type of Equipment:
Class I Type B (as defined by IEC 60601-1, UL 60601-1,
EN 60601-1, CSA 601.1, IEC 60601-2-32:1994 and IEC 60601-2-46)
Type B protection against electrical shock
Complies with FDA standards
Electrical:
Supply Voltage: 120±5% Vac 60Hz or 23010% Vac 50Hz
Duty Cycle: 10% (6 min/hr)
Current Rating: Less than 10 Amps
Internally Battery Powered
Environmental:
Operating Temperature Range: -10C to +40C
Operating Humidity Range: 30% to 75% relative humidity, noncondensing
Operating Pressure Range: 700hPa to 1060 hPa
Transport & Storage Temperature Range: -40C to +60C
Transport & Storage Humidity Range: 30% to 75% relative humidity, noncondensing
Transport & Storage Pressure Range: 500hPa to 1060 hPa
Rated IPX4 (Protected against splashing water)
Meets EMC requirements of EN 60601-2 (1993)
Tabletop:
The tabletop is made of carbon fiber and is certified to meet all the requirements of 21 CFR Subchapter J
Operator Manual Aspect 100HT
L100-3064 REV A
www.imagediagnostics.com
7
ELECTROMAGNETIC COMPATIBILITY STATEMENT
This equipment may generate and use radio frequency energy. The equipment must be installed and used
according to the manufacturer’s instructions in order to avoid radio frequency interference.
If this equipment generates or receives interference:
• Verify that the equipment is the cause by turning the system off and on.
• In the event of unintended motor actuation, immediately remove power to the equipment.
• Use only cables supplied by Image Diagnostics, Inc.
Operator Manual Aspect 100HT
L100-3064 REV A
www.imagediagnostics.com
8
SETUP
Unpack
No special handling is required for unpacking this equipment at the site. Conventional shipping materials
are used.
To remove the table from the shipping pallet, first remove the shipping brackets that secure the table to
the wooden pallet using a ½” wrench. Then roll the table down a temporary ramp or lift off pallet with a
forklift. When using a forklift, pad the forks to protect the paint. Lift from under the metal base. When
rolling the table off the pallet by hand, use the long metal rails attached to the tabletop to guide the table
after releasing the caster brakes.
Install
Mobile tables are equipped with four locking caster brakes. Unlock brakes while positioning.
Foot controls are standard with this table with a hand control being an option. The operator should become
familiar with the controls before using.
The table operates on 110Vac (or 230Vac) or on internal battery backup. Plug the power cord into a properly
grounded hospital grade outlet.
The outlet used should be visible and accessible to the user. The cord should be routed where it will not be subject
to damage or be a hazard. Check that the power cord ground pin is in good condition. Check that the cord ground
wire is securely connected to the ground stud inside the unit.
During an AC failure, the table will automatically run on battery backup without any manual switching required.
It is recommended that power be applied at all times, even when not in use, to keep a proper charge on the
batteries and achieve maximum battery life. (Batteries are being charged during normal use).
Where the integrity of the external protective earth conductor is in doubt, this equipment shall be operated from its
internal electrical power source by unplugging the power cord from the mains.
Recycle or dispose of shipping
material per local regulations.
Operator Manual Aspect 100HT
L100-3064 REV A
www.imagediagnostics.com
9
Preparation for Patient Use
New Installation:
This equipment will need to be properly cleaned before patient use as it will inevitably come into contact with
contaminants during shipping, unpacking, storage, and installation.
After Installation:
This equipment will need to be properly cleaned between patient use as it will inevitably come into contact with
contaminants during procedures.
SAFETY HAZARDS
Operators using this equipment should understand the safety issues, emergency procedures, and
operating instructions provided.
Comments and questions regarding safety should be addressed to:
Customer Support
Image Diagnostics, Inc.
310 Authority Drive
Fitchburg, MA 01420
USA
Or call IDI at (978) 829-0009 or send fax to (978) 829-0027.
Or call toll-free at (877) 304-5434.
Review the SAFETY INSTRUCTIONS before
operating table
Operator Manual Aspect 100HT
L100-3064 REV A
www.imagediagnostics.com
10
SAFETY HAZARD ALERTS
Alert
Circumstances for Use
DANGER
Indicates an imminently hazardous situation which, if
not avoided, will result in death or serious injury.
WARNING
Indicates a potentially hazardous situation which, if not
avoided, could result in death or serious injury.
CAUTION
Indicates a potentially hazardous situation which, if not
avoided, may result in minor or moderate injury or
equipment damage.
WARNING!
This equipment has not been tested for use with high
frequency surgical equipment, cardiac defibrillators, or
cardiac defibrillator monitors.
WARNING!
If an antistatic path is required, use the equipment on an
antistatic floor. Use only the tabletop pad (patient
mattress) supplied with the table.
CAUTION
Secure power cord when in use so that it does not get
entangled with other equipment and/or compromise
the safety of the operator and/or staff.
CAUTION
Do not modify this equipment without authorization of
the manufacturer.
Operator Manual Aspect 100HT
L100-3064 REV A
www.imagediagnostics.com
11
SAFETY INSTRUCTIONS
This manual should be accessible to all personnel installing, operating, or servicing this equipment.
Only a qualified technician may install or maintain this equipment.
Only qualified persons may operate this equipment.
Failure to follow safety precautions may result in serious injury to patient or user or damage to equipment.
This symbol, when found marked on the equipment
means:
”Attention, consult accompanying documentation.”
WARNING!
Always safely position and secure patient on table
WARNING!
Do not exceed patient weight of 525 pounds (238kg)
CAUTION
Do not leave patient on table unattended
CAUTION
Patient must be restrained at all times. The restraining
straps are not intended to restrain an uncontrollable
patient.
CAUTION
The carbon fiber top is subject to damage or possible
damage from impact from other objects. Take caution
when using power driven diagnostic equipment around
table. Regular inspection of the tabletop is necessary for
the safety of patient and operator.
Operator Manual Aspect 100HT
L100-3064 REV A
www.imagediagnostics.com
12
Push down to lock
CAUTION
Ensure that the tabletop does not contact other
equipment as it moves.
CAUTION
During operation, listen for any unusual sounds and
watch for uneven operation.
CAUTION
Do not sit on the table top at the head of the table.
CAUTION
Do not place or store any containers or large items
underneath the tabletop. As the tabletop is descending,
contact with an obstruction may cause permanent
damage to the table.
CAUTION
Do not use table for patient transport.
CAUTION
Secure wheel locks by depressing locks on all 4
wheels prior to patient usage and when table is not
in motion. Disengage to reposition.
Operator Manual Aspect 100HT
L100-3064 REV A
www.imagediagnostics.com
13
ACCESSORIES
• PATIENT RESTRAINT STRAPS
UPatient Restraint Strap SetU
Part Number C000-0571
(Two straps 3” wide)
•
PATIENT HANDGRIP
The patient handgrip can be quickly locked to the accessory rail in any position to suit the patient.
UPatient HandgripU
Part Number A100-1711
(Order 2 for a complete pair)
WARNING!
25lbs (11kg) Maximum Load
Check that handgrip is tight before each use.
Operator Manual Aspect 100HT
L100-3064 REV A
www.imagediagnostics.com
14
OPERATOR CONTROLS
HAND CONTROL (OPTIONAL)
• Buttons control the tilt motion
To move the tabletop, press the appropriate button and hold until the desired
position is reached.
UHand ControlU
Part Number: Handset K000-0041, Label D100-1824
Go to the Linak web site for disposal information.
The URL is listed on the table of contents page of this manual.
CAUTION
Always check that the patient is secure before
activating any table motion
Operator Manual Aspect 100HT
L100-3064 REV A
www.imagediagnostics.com
15
• FOOT CONTROL (OPTIONAL)
• The two pedals control the tilt motion
To move the tabletop, depress the appropriate pedal and hold until the
desired position is reached.
UFoot ControlU
Part Number A100-1823
Contact IDI Customer Service for individual components of foot control.
CAUTION
Always check that the patient is secure before
activating any table motion
Operator Manual Aspect 100HT
L100-3064 REV A
www.imagediagnostics.com
16
EMERGENCY STOP SWITCH
• The red emergency stop button is located on top of the control box
• Pressing this button down will stop all motor movement
• In the event that the hand or foot control malfunctions, the emergency stop control
prohibits motion by removing power from all system components
• ACTIVATE: Engage the emergency stop mode by pushing the button down
• RESET: Restore the electrical functions by rotating the button a quarter turn
Note: The battery will not charge if the emergency stop switch is pushed down.
EQUIPOTENTIAL TERMINAL
The Equipotential terminal provides a connection point between equipment and a
potential equalization busbar.
Symbol Terminal
Operator Manual Aspect 100HT
L100-3064 REV A
www.imagediagnostics.com
17
ELECTRIC SYSTEM
BATTERY FUNCTION (Motorized tables)
When AC power is applied, the battery will begin charging until the battery reaches full
capacity. When it does, the charging will taper off. If the power goes out, the table will
automatically switch to battery back-up mode.
Note: The battery will not charge if the emergency stop switch is pushed down.
An audible tone is sounded when using an operator control function if the battery
charge is low.
Local procedures may be in place for proper disposal or recycling. For more
information about batteries, such as battery life, production codes and
disposal go to the Linak web site for disposal information. Select
“Products|User Manuals” and click on “Actuators & Electronics”. Linak
documentation can be found at their web site: www.linak.com
Operator Manual Aspect 100HT
L100-3064 REV A
www.imagediagnostics.com
18
GENERAL CLEANING
The disinfectant cleaners and cleaning steps on this page apply to all the parts of the table.
• Clean tabletop, pads, accessories, and painted surfaces with a clean cloth dampened
with an approved cleaner (see list below).
Do not use abrasives, solvents, sprays, or corrosive agents.
APPROVED AND TESTED DISINFECTANT CLEANERS
• Sodium hypochlorite (generic household bleach) in solution of 5.25% sodium hypochlorite
diluted between 1:10 and 1:100 with water.
• Alcohol (generic)
• Envirocide ® Disinfectant and Cleaner
CAUTION
After each procedure, the table should be carefully
cleaned.
CAUTION
Avoid cleaning or getting cleaning solutions on
elevating tube if possible.
Operator Manual Aspect 100HT
L100-3064 REV A
www.imagediagnostics.com
19
APPROVED AND TESTED GENERAL PURPOSE CLEANERS
• Simple Green ™ cleaner
CLEANING STEPS
1. Move the tabletop into the horizontal position.
2. Lower the tabletop to the lowest position.
3. Power cord, hand control and foot control cords must be plugged in to protect the sockets.
4. Remove all pads and accessories.
5. Wipe off any access fluids with a water dampened cloth or sponge.
6. Clean the accessories, tabletop, and pads using approved cleaners.
7. Clean the table frame, castors, and base with Simple Green ™ cleaner.
8. Thoroughly rinse tabletop pad with water.
9. Gently rub with a soft clean cloth to dry.
If noise or uneven motion occurs when the table top is moving up or down,
extend and clean the lifting tube with a soft clean cloth.
MAINTENANCE
”Qualified Technician”
• All maintenance procedures should be done by an experienced technician with demonstrated
knowledge and skills (electrical and mechanical) relative to this type of equipment.
• This individual must have access to this manual and the proper tools.
ROUTINE MAINTENANCE
• Lubrication is not required however powdered graphite may be used on the lifting tube.
PERIODIC PERFORMANCE CHECKS
• Semiannually perform functional inspections of carbon fiber tabletop, support frame,
and the elevating tube.
• Daily inspect all external cables, controls, and the tabletop for wear and damage.
Replace damaged cables promptly.
• Weekly check battery operation by disconnecting the AC power and running the tabletop
up and down.
Check daily if table is routinely used without being attached to mains.
Operator Manual Aspect 100HT
L100-3064 REV A
www.imagediagnostics.com
21
USER REPLACEMENT ITEMS
These are easily replaced by the operator.
See also previous sections for operator hand and foot controls, armboard pads and protective covers.
• Standard Tabletop Pad
Pads are held in place by hook and loop fasteners. Simply pull pad up gently to remove. New
replacement pads are supplied with new self-adhesive fasteners installed and new mating pieces
for the tabletop. Peel off old fasteners from tabletop. Peel off the protective paper and carefully
place new pad into position. Apply pressure to complete installation.
Note: Pads comply to California Technical Bulletin 117 as well as Underwriters Laboratories 94 (UL94).
UTabletop PadU 2.5 cm x 61 cm x 203 cm (1” x 24” x 80”)
Part Number X100-1649
Optional Tabletop Pad with Hood 5cm x 61 cm x 203 cm (2” X 24” x 80”) with hood.
Part Number X100-2008
LATEX
Not made with natural rubber latex
Operator Manual Aspect 100HT
L100-3064 REV A
www.imagediagnostics.com
22
DISPOSAL OF COMPONENTS
IDI medical tables are made up of mostly steel, copper and aluminum parts which are
easily recycled. It is recommended that some components be disassembled before
disposal for recycling. The table below lists components typically found in IDI products
but varies with model and options.
COMPONENT
ITEM
RECYCLING GROUP
Actuator
Spindle and Motor
Housing
Cable
Metal (Steel and Copper)
Plastic
Copper
Control Box
PC Board
Plastic Housing
Cable
Transformer
Batteries
Electronic
Plastic
Copper
Copper
Lead Acid Batteries
Hand Control
PC Board
Housing
Cable
Electronic
Plastic
Copper
Foot Control
PC Board
Metal Housing
Cable
Electronic
Steel and Aluminum
Copper
Table Base
Frame
Casters
Metal (Steel)
Plastic and Steel
Operator Manual Aspect 100HT
L100-3064 REV A
www.imagediagnostics.com
23
WARRANTY
Image Diagnostics, Inc. Aspect Imaging Tables Official Product
Warranty for USA & Canada
1. Scope and Duration of Warranty: Image Diagnostics, Inc. (IDI) warrants, to the original Purchaser only,
that the Covered Products conform to the manufacturer's published specifications and are free from defects
in material or workmanship. The warranty period will commence on the date of delivery. The duration of
the warranty ("Warranty Period") is 36 months on parts and 12 months on labor, except for accessories,
hand controls, foot controls, casters, table top pads, batteries, power cords and electrical cords for hand
and/or foot controls, which have a warranty period of 12 months for both parts and labor[1]. If Purchaser
discovers within this Warranty Period a failure of the Covered Products to conform to specifications or a
defect in material or workmanship, Purchaser must promptly notify IDI by calling IDI Customer Service at
877-304-5434 during normal business hours: Monday through Friday, 8:00 a.m. through 5:00 p.m., Eastern
Time, excluding holidays. IDI's warranty obligations will apply only to such notifications made during the
warranty period and will not apply to notifications made after warranty expiration.
2. Exclusive Product Warranty Remedies: If Purchaser promptly notifies IDI of Purchaser's warranty claim
and makes the Covered Product available for service, IDI will, at IDI's option, either repair, adjust or
replace (with new or exchange replacement parts) the non-conforming Covered Product or parts of the
Covered Product Replaced parts will become the property of IDI. The warranty period for any Covered
Product furnished to the customer as a warranty remedy will be the remaining portion of the warranty
period applicable to the repaired or replaced Covered Product All warranty service will be performed by an
IDI's authorized service representatives. During normal business hours, Warranty service will be performed
without charge. If Purchaser requests warranty service, the service visit will be scheduled at a mutually
acceptable time. If Purchaser refuses to make the Product available for service upon arrival of the IDI
service representative, the Purchaser will be responsible for payment of service travel time and expenses
and all time on site that the service representative is required to wait for access to the Product, whether or
not the service is completed. These charges will be billed at IDI's prevailing service rates. If Purchaser
requests Warranty service outside of normal business hours it will be provided at IDI's prevailing "after-
hours" service rates and will be subject to availability of service personnel.
3. What Is Not Covered Bv This Warranty: IDI does not warrant (i) any Product or part not sold by IDI or
its authorized representatives, (ii) defects caused by failure to provide a suitable installation environment
for the Covered Product (iii) damage caused by use of the Covered Product for purposes other than those
for which it was designed, (iv) damage caused by disasters such as fire, flood, wind, earthquake, lightning
or other natural disasters, (v) damage caused by unauthorized attachments or modification, (vi) abuse or
misuse by the Purchaser or its personnel, or (vii) other causes beyond IDI's control. Product damage or
failures not covered by this warranty may include, but are not limited to, failure to adhere to instructions
provided in the Product Operator Instructions.
4. Products not Covered bv This Warranty: The warranties set forth herein do not cover the following
Products: (i) consumable items, including but not limited to drapes, (ii) used or refurbished equipment, (iii)
Products serviced by anyone other than IDI or its authorized representatives during the Warranty Period.
5. Disclaimer of Warranty: THE FOREGOING WARRANTIES ARE IN LIEU OF ALL OTHER WARRANTIES,
EXPRESS OR IMPLIED, INCLUDING BUT NOT LIMITED TO THE IMPLIED WARRANTIES OF
MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE.
6. Limitation of Remedies: In no case shall IDI or its affiliates and representatives be liable for any special,
incidental or consequential damages based upon breach of warranty, breach of contract, negligence, strict
tort or any other legal theory. Such damages include, but are not limited to, loss of profits, loss of savings
or revenue, loss of use of the Covered Products or any associated equipment, cost of capital, cost of any
substitute equipment, facilities or services, downtime, the claims of third parties including customers and
injury to property. This limitation does not apply to damages caused by breach of the warranty of title and
patent or copyright infringement or to claims for personal injury.
Operator Manual Aspect 100HT
L100-3064 REV A
www.imagediagnostics.com
24
7. No Other Warranties: Unless modified in writing and signed by both parties, this Warranty is understood
to be the complete and exclusive product warranty agreement between the parties, superseding all prior
agreements, oral or written, and all other communications between the parties relating to the subject matter
of this Warranty. Except for an authorized IDI corporate officer, no IDI employee or IDI representative or
any other party is authorized to make any warranty in addition to those made in this Agreement.
8. Warranty Terms Subject to Change: IDI reserves the right to modify the terms and conditions of its
Official Warranty from time to time. The warranty terms and conditions, and IDI's obligations under such,
will be determined based on the prevailing version of IDI's Official Warranty in effect at the date of
purchase order.
[1] Certain parts subject to frequent wear and tear and misuse are limited to 12 months warranty as defined in Article 1
above.
Operator Manual Aspect 100HT
L100-3064 REV A
www.imagediagnostics.com
25
Image Diagnostics, Inc.
Aspect Imaging Tables Official Product Warranty
for Tables Installed Outside the USA & Canada
1. Scope and Duration of Warranty: Image Diagnostics, Inc. (IDI) warrants, to the original Purchaser only,
that the Covered Products conform to the manufacturer's published specifications and are free from defects
in material or workmanship. The duration of the warranty ("Warranty Period") is 36 months from date of
delivery, except for accessories, hand controls, foot controls, casters, table top pads, batteries, power cords
and electrical cords for hand and/or foot controls, which have a warranty period of 12 months[1] from date
of delivery. If Purchaser (Distributor) discovers within this Warranty Period a failure of the Covered
Products to conform to specifications or a defect in material or workmanship, Purchaser must promptly
notify IDI by calling IDI Customer Service at 978-829-0009 (or by email communication to
[email protected]) during normal business hours: Monday through Friday, 8:00 a.m. through
5:00 p.m., Eastern Time, excluding holidays. IDI's warranty obligations will apply only to such
notifications made during the warranty period and will not apply to notifications made after warranty
expiration.
2. Exclusive Product Warranty Remedies: If Purchaser promptly notifies IDI of Purchaser's warranty claim,
IDI will, at IDI's option, provide a replacement of the non-conforming Covered Product or parts of the
Covered Product with either new or exchange replacement parts. Replaced parts will become the property
of IDI. The warranty period for any Covered Product furnished to the customer as a warranty remedy will
be the remaining portion of the warranty period applicable to the repaired or replaced Covered Product. All
warranty service labor will be the responsibility of the Purchaser (Dealer).
3. What Is Not Covered By This Warranty: IDI does not warrant (i) any Product or part not sold by IDI or
its authorized Distributors or representatives, (ii) defects caused by failure to provide a suitable installation
environment for the Covered Product, (iii) damage caused by use of the Covered Product for purposes other
than those for which it was designed, (iv) damage caused by disasters such as fire, flood, wind, earthquake,
lightning or other natural disasters, (v) damage caused by unauthorized attachments or modification, (vi)
abuse or misuse by the Purchaser or its personnel, or (vii) other causes beyond IDI's control. Product
damage or failures not covered by this warranty may include, but are not limited to, failure to adhere to
instructions provided in the Product Operator Instructions.
4. Products not Covered bv This Warranty: The warranties set forth herein do not cover the following
Products: (i) consumable items, including but not limited to drapes, (ii) used or refurbished equipment, (iii)
Products serviced by anyone other than IDI or its authorized representatives during the Warranty Period.
5. Disclaimer of Warranty: THE FOREGOING WARRANTIES ARE IN LIEU OF ALL OTHER WARRANTIES,
EXPRESS OR IMPLIED, INCLUDING BUT NOT LIMITED TO THE IMPLIED WARRANTIES OF
MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE.
6. Limitation of Remedies: In no case shall IDI or its affiliates and representatives be liable for any special,
incidental or consequential damages based upon breach of warranty, breach of contract, negligence, strict
tort or any other legal theory. Such damages include, but are not limited to, loss of profits, loss of savings
or revenue, loss of use of the Covered Products or any associated equipment, cost of capital, cost of any
substitute equipment, facilities or services, downtime, the claims of third parties including customers and
injury to property. This limitation does not apply to damages caused by breach of the warranty of title and
patent or copyright infringement or to claims for personal injury.
7. No Other Warranties: Unless modified in writing and signed by both parties, this Warranty is understood
to be the complete and exclusive product warranty agreement between the parties, superseding all prior
agreements, oral or written, and all other communications between the parties relating to the subject matter
of this Warranty. Except for an authorized IDI corporate officer, no IDI employee or IDI representative or
any other party is authorized to make any warranty in addition to those made in this Agreement.
Operator Manual Aspect 100HT
L100-3064 REV A
www.imagediagnostics.com
26
8. Warranty Terms Subject to Change: IDI reserves the right to modify the terms and conditions of its
Official Warranty from time to time. The warranty terms and conditions, and IDI's obligations under such,
will be determined based on the prevailing version of IDI's Official Warranty in effect at the date of
purchase order.
[1] Certain parts subject to frequent wear and tear and misuse are limited to 12 months warranty as defined in Article
1 above.