YSI EXO Quick Start Manual
Text of the manual
Sonde Quick Start Guide
Thank you for using an EXO sonde
for your environmental monitoring!
This document will assist with the
initial steps required to use an
EXO™ sonde.
Refer to the EXO User Manual for:
• Part numbers and accessories
• Maintenance guidelines
• Sensor calibration instructions
• Connecting EXO to a Data Logger
See reverse for
more information.
Items Needed
Included in Packaging
• Sonde
• Probes (Sensors)
• Batteries
• Battery cover wrench
• Sensor tool with magnet
(to activate Bluetooth)
• KOR™ software on USB flash drive
Not Included
• EXO Handheld or PC running
KOR™ software
• USB cable and adapter, field cable,
or Bluetooth dongle (sold separately)
Preparing Sonde for Use
To prepare the sonde for operation,
you will need to:
• Install probes into the connectors
on the sonde bulkhead
• Install batteries
• Load KOR™ software onto PC
• Connect sonde to EXO Handheld
device or PC
Registration
Register your product online to
activate your warranty and keep
informed of software updates:
YSI.com/warranty
TM
TM The EXO
Sonde Line
EXO1
EXO2
EXO3
EXO2s
Keep your serial number on file:
Date of purchase:
Unpacking and Inspection
Inspect the outside of the shipping
container(s) for damage. If you see any
damage, contact the shipping
carrier immediately.
Remove the equipment from the shipping
box. Some parts are loose in the box, so
check the packing material carefully.
item# 603794REF
dwg# A603794
Jan 2017 Rev. C
YSI.com
Install probes into the sonde bulkhead
Install Probes
1. Remove buffer bottles and hydration
caps on the sensors (if applicable).
2. Remove the calibration cup and guard
from the sonde.
3. Remove port plugs and insert sensors into
ports on the bulkhead. Finger-tighten
the locking rings until snug. Ports are
“universal,” which means any sensor can
be plugged into any port.
Exception – The wiper must be installed in
the correct port to function properly:
• EXO2 / EXO2s – Central Port 7
• EXO3 – Port 5
Interface with the sonde using KOR™
Connect sonde to computer with USB
cable and adapter
Connect sonde to handheld with
field cable
If not using cable, activate Bluetooth
wireless communication with magnet
Communication
Connect Sonde to Computer
via USB Cable
1. First, install the KOR™ software on your
computer. Follow the prompts to install
the new hardware and drivers.
2. Plug in the USB Signal Output Adapter
to the top of the sonde.
3. Plug in the small end of the USB cable to
the top of the adapter and the large end
in a standard USB port on the computer.
The USB adapter provides a hardwired
communications link to a PC as well as
providing power to the sonde.
via Bluetooth®
A computer must be equipped with an
internal Bluetooth radio or use a Bluetooth
dongle to establish a wireless connection to
the sonde.
Activate Bluetooth by holding the included
sensor tool with magnet at the magnetic
activation area on sonde. Once activated,
the blue light will be on. The software will
automatically pair the instrument to
your PC.
Troubleshooting
Establish Bluetooth link (Windows OS)
1. Check your version of Windows for the
proper method to pair a Bluetooth device
to your PC.
2. When prompted, locate the sonde from
the list. The sonde name should begin
with YSI.
3. Select “Enter the device’s pairing code,”
and click Next. Enter the code 9876.
Click “Next.”
5. Select Finish.
Discover the Sonde in KOR™
Upon startup of the software KOR™ will
search for and discover the activated
sonde and request to connect to it. Once
connected via Bluetooth, the blue light on
the sonde will blink.
Connect Sonde to EXO Handheld
via Field Cable
Press in the male 6-pin connector on the
cable to the top of the sonde. Press in the
female 6-pin connector on the cable to the
bottom of the handheld, then screw down
the retaining collars on each end.
Install Batteries
The EXO1 and EXO3 Sondes use two (2)
D-cell alkaline batteries and the EXO2 Sonde
uses four (4) D-cell alkaline batteries as the
recommended power source. For the EXO2s,
and at the discretion of the user for EXO1,
EXO2, and EXO3, external power sources
may be used.
Alternatively, the sonde may use
rechargeable NiMH D-cell batteries (not
provided). We recommend a NiMH D-cell
battery with a minimum rating of 10,000
milliamp hours. However, due to variations
in manufacturer capacity and discharge
curves, deployment life cannot be estimated
when using rechargeable batteries.
1. Remove the battery cover from
the sonde:
EXO1: Twist the blue battery cover
counter-clockwise to unscrew, lift up to
remove. Use included wrench to loosen,
if necessary.
EXO2 / EXO3: Unscrew and remove
battery cap. Use included wrench to
loosen, if necessary.
2. Insert the batteries with negative (-)
terminals facing down toward
the probes.
3. Replace the battery cover or cap and
tighten until snug. Do not overtighten.
LED Indicators
When power is supplied to the sonde, two
indicator lights will provide status updates.
Red LED – Sonde State
0.1 Hz (10s Blink): Asleep, logging enabled
1 Hz (1s Blink): Awake
On Solid: Awake with faults
Blue LED – Bluetooth
On Solid: On, not linked
2 Hz (0.5s Blink): On, successfully linked
Off: Bluetooth disabled
Remove battery cover/cap and
install batteries
EXO1
Side View
EXO2 / EXO3
Top View
item# 603794REF
dwg# A603794
Jan 2017 Rev. C
You are now ready to calibrate your
sensors and start collecting data!
YSI.com