Sencon SI9000 plus Operation & User’s Manual
Read Online
Read the Sencon SI9000 plus Operation & User’s Manual online using the built-in PDF viewer.
Text of the manual
SI9000plus
Enamel Rater
User Manual
REF
OPERATION
MANUAL No.
092-00900-10
LANGUAGE
ENGLISH
REVISION
10
ENAMEL RATER MANUAL: V.10
3
1.
INTRODUCTION
5
1.1 THE ENAMEL RATING TEST
5
1.2 THE SI9000+
5
1.2.1
BODY CONTACT AND LEVEL CHECKING
6
1.2.2
SEMI-AUTOMATIC OPERATION
6
1.2.3
BATCH OPERATION
6
1.2.4
CONSTRUCTION
6
1.3 THE SI9005 CAN STAND
7
2.
GETTING STARTED WITH YOUR SI9000+
8
2.1 SAFETY
8
2.2 SETTING THE MAINS VOLTAGE
8
2.3 ADJUSTING AND CONNECTING THE CAN STAND
10
2.4 CONNECTING A NON-SENCON PROBE
12
2.5 SWITCHING ON AND THE SELF TEST
13
3.
OPERATION
14
3.1 GENERAL DESCRIPTION
14
3.2 CONFIGURATION CHECK
14
3.3 FREE RUN AND REVERSE POLARITY
15
3.4 BATCH MODE
16
3.4.1
RATING CANS
16
3.5 MASTER RESET
17
3.6 ERROR MESSAGES
18
4.
CONFIGURING YOUR SI9000
19
4.1 SETTING THE MAINS VOLTAGE
20
4.2 SETTING UP THE STATISTICAL OUTPUT
21
4.3 THE LEVEL INTERLOCK
21
4.4 THE CONTACT INTERLOCK
22
4.5 CHANGING THE RATING VOLTAGE
22
4.6 CHANGING THE RATING DURATION
22
4.7 AUTORANGING OPERATION
23
4.8 LINE NUMBER ENABLE
23
4.9 AUTOSTART
24
4.10 AUTOSAVE
24
5.
MAINTENANCE AND SERVICING
25
5.1 CARE OF YOUR SI9000+
25
5.2
ROUTINE MAINTENANCE
25
5.3 ROUTINE TESTING
25
5.5 FAULT FINDING
27
5.6 REPLACING THE FUSE
28
5.7 OBTAINING HELP FROM SENCON
28
6.
USING THE SI9000+ WITH A PRINTER
29
6.1 OVERVIEW
29
6.2 CHOOSING A PRINTER
29
6.3 CONNECTING THE PRINTER
30
6.3.1
THE EPSON FX OR EX PRINTER
30
6.3.2
THE TANDY DMP132 PRINTER
31
6.3.3
USING OTHER PRINTERS
31
6.4 SETTING UP THE RS232 DATA FORMAT
32
6.4.1
SETTING UP THE EPSON PRINTER
32
6.4.2
SETTING UP THE TANDY PRINTER
33
ENAMEL RATER MANUAL: V.10
6.4.3
SETTING UP OTHER PRINTERS
33
6.5 ENABLING THE PRINTOUT
33
6.6 SELECTING THE PRINTOUT FORMAT
34
6.7 PRINTER PROBLEMS
35
7.
USING THE SI9000+ WITH A COMPUTER
36
7.1 GENERAL
36
7.2 COMPUTER PORT CONNECTIONS
36
7.3 SETTING DATA FORMAT AND SPEED
37
7.4 UNI-DIRECTIONAL PROTOCOL
38
7.5 BI-DIRECTIONAL PROTOCOL
38
7.6 FORMAT OF MESSAGE TRANSMITTED
39
7.7 ERROR MESSAGES
39
ENAMEL RATER MANUAL: V.10
5
1.
INTRODUCTION
1.1
THE ENAMEL RATING TEST
The can industry has established a standard test for lacquer
integrity of food and beverage cans. The test involves applying a
voltage between a conductive liquid (electrolyte) within the can and
the metallic can body. As the internal lacquer coating is an
insulator, any current which flows is an indication of imperfections
in the lacquer . The larger the current the greater the area of the
imperfections and the poorer the quality of the can.
There is no definitive standard for the enamel rating test. Different
users have their own standards specifying the test voltage and test
duration, however a common standard requires a voltage of 6.3V to
be supplied for 4 seconds. In its factory supplied configuration the
SI9000+ uses this voltage and duration.
When a faulty can is detected the normal procedure is to reverse
the polarity of the applied voltage, which causes bubbles of gas to
be released from the area of the fault. This gives a visual indication
of the location of the fault and enables corrective action to be
taken.
1.2
THE SI9000+
The following description of operation assumes that the SI9000+ is
operated in its factory set configuration (see Appendix A Fig 1 for
settings).
ENAMEL RATER MANUAL: V.10
1.2.1
BODY CONTACT AND LEVEL CHECKING
One of the problems the enamel rating test has suffered in the past
is that if any part of the circuit has a poor, high resistance contact
only a small current will flow and the instrument will show a bad can
as good. This is a particular problem where electrical contact is
made with the can body.
In order to overcome this problem the SI9000+ has special circuitry
which checks the contact with the can body and will not allow the
test to proceed until a good contact has been detected. In addition
to this, a similar circuit checks that the electrolyte level within the
can is adequate and so avoids the problem of the upper zone of
the can not being fully tested due to insufficient electrolyte. These
two circuits together will also detect a fault anywhere in the
connector or wiring to the probe, making the SI9000+ fail-safe in its
operation.
1.2.2
SEMI-AUTOMATIC OPERATION
Another spin-off of these check-circuits is that the SI9000+ can be
configured to detect a change of can and will automatically start the
test once a new can is correctly inserted. An internal timer will
freeze the reading when a preset time (e.g. 4 seconds) has elapsed
and one key press then prepares the instrument for the next can.
1.2.3
BATCH OPERATION
As the operator rates cans the SI9000+ will store the readings (up
to 99) in its memory as a batch. At any time during the batch
statistical information about the batch can be displayed and at the
end of the batch the readings in tabular format together with the
statistical information can be output to a printer or computer if
required.
1.2.4
CONSTRUCTION
The acrylic coated steel case and waterproof polyester facia panel
are designed to ensure that the SI9000+ will withstand the rigours
of a factory environment. Further to this the case, mains-switch
and connectors are all sealed to I.P.65 rating, thus preventing any
ingress of corrosive electrolytes and allowing operation of the
SI9000+ in "hose-down" areas.
ENAMEL RATER MANUAL: V.10
7
1.3
THE SI9005 CAN STAND
Due to the corrosive nature of the electrolytes used in enamel
rating the life of probe attachments can be very poor. With this in
mind, the SI9005 has been designed using nylon parts wherever
possible, whilst all metal parts are made from specially selected
stainless steels, heat treated where appropriate. The wiring of the
can stand is completely sealed against corrosion from electrolytes
which not only improves the life but ensures accuracy and reliability
of the readings taken.
Electrical contact is made with the can by three hardened knife
edges, one of which is wired separately to allow electrical contact
checking, and the two probes (one for rating and a shorter probe
for level checking) are supported on an adjustable arm and
telescopic column. This arrangement allows the probes to be lifted
and rotated to one side while the can is being inserted.
ENAMEL RATER MANUAL: V.10
2.
GETTING STARTED WITH YOUR SI9000+
2.1
SAFETY
The test voltages used by the SI9000+ are sufficiently low so as to
present no hazard whatsoever under all circumstances. Although
potentially dangerous mains voltages are present inside the unit,
the construction of the case to I.P.65 rating will prevent any hazard
from this whilst the case remains intact. Should the facia become
perforated or the case become damaged in such a way as to
destroy the seal then the unit must be removed from service
immediately and returned to Sencon for repair. Operators should
be warned to keep electrolytes away from electrical socket outlets
and never handle plugs with wet hands.
2.2
SETTING THE MAINS VOLTAGE
The mains voltage of the SI9000+ is factory set as appropriate for
the particular country to which it is supplied and the setting will be
clearly marked on the back panel. Should you need to change the
voltage proceed as follows:-
i.
Remove the six button head screws in the base of the
case using the 2.5mm hex wrench supplied.
ii.
Remove the base panel taking care not to damage the
earthing cable.
iii.
Referring to Appendix A Fig 1 move the mains voltage
switch to the required position.
iv.
Check that the foam seal on the bottom of the case is
intact and replace the base panel and six screws.
v.
Change or remove the voltage label.
ENAMEL RATER MANUAL: V.10
9
Key to Enamel Rater Probe
1 -
Bushing
7 - Body Contact
2 -
Foot
8 - Locating Ring
3 -
Clamp Screw
9 - Probe Arm
4 -
Locking Screw
10 - Inner Column
-
Level Probe
11 - Outer Column
6 -
Main Probe
12 - Stand Base
FIGURE 2: ENAMEL RATER PROBE
ENAMEL RATER MANUAL: V.10
2.3
ADJUSTING AND CONNECTING THE CAN STAND
In order to make use of the electrolyte level checking facilities of
the SI9000+, it is important that the can stand is adjusted correctly
for the height of can to be tested. Referring to Fig.2 proceed as
follows:-
i.
Raise outer column (item 11) and place can through the
can locating ring (item 8) onto the body contacts (item 7).
ii.
Lower the outer column.
Note
If at any time the outer column will not return to
its normal position resting on the stand base
(item 12) due to the cable fouling inside, then
raise the outer column and rotate clockwise for
several turns until the column moves freely
again.
iii.
Slacken clamp screw (item 3) using the 3mm hex wrench
supplied, and slide the probe arm (item 9) so that the
bottom of the level probe (item 5) is at the minimum
permissible electrolyte level and the outer column is firmly
against the stand base.
iv.
Tighten the clamp screw making sure the main probe
(item 6) is central in the can. Take care not to overtighten
the clamp screw as this will cause the outer column to
bind on the inner column.
The can stand is now ready for use and should be connected to the
SI9000+ via the probe socket (refer to Fig.3). This connector is
polarised so it must be aligned before insertion. Always screw in
the locking ring and tighten finger-tight to ensure it is sealed against
electrolyte.
ENAMEL RATER MANUAL: V.10
2.4
CONNECTING A NON-SENCON PROBE
If you want to connect another probe to your SI9000+ for
specialised tests such as end, or side-seam enamel rating this is
always possible using the 5 pin plug supplied, but the following
points should be noted:-
i.
The cable diameter must be in the range 3mm to 5mm to
ensure correct sealing.
ii.
Other probes are unlikely to support body contact and
level checking and so these functions will be disabled.
iii.
If (ii) above applies, then the automatic restart facility after
each can will be disabled.
Soldering of the cable should be carried out by an electrician as
follows:-
Pin
Function
1
Main probe (electrolyte connection).
2
Level probe (if not used, connect to 1).
3
Body contact check (if not used connect to
4).
4
End or can body contacts.
5
Not connected.
The probe can now be connected to the SI9000+ via the probe
socket (refer to Fig.3). Ensure the connector is aligned before
insertion and screw in the locking ring finger-tight to ensure a good
seal.
ENAMEL RATER MANUAL: V.10
13
2.5
SWITCHING ON AND THE SELF TEST
Having satisfied yourself that the mains voltage setting is correct,
connect the mains plug to a suitable socket outlet or if you prefer
wire the mains cable directly into a fused outlet. Switch on the
SI9000+ by depressing the mains on/off switch on the back panel.
On power up, the system will perform a self-test. The numeric
displays will cycle through each digit from 0 through to 9 allowing
the operator to see that all the segments illuminate correctly. The
LED indicators will be illuminated in sequence.
Finally the SI9000+ will go to standby mode with the can number
and timer set to zero and the milliamp display set to dashes.
Enamel rating can then begin by pressing.
i.
The start key. Enamel rating will start when the level and
contact check requirements are satisfied (indicators off)
and continue for 4 seconds.
ii.
The Free Run key. The SI9000+ will enamel rate for up
to 99 seconds in free run.
iii.
The Reverse Polarity key. As ii) but with reversed
polarity (can body negative)
ENAMEL RATER MANUAL: V.10
3.
OPERATION
3.1
GENERAL DESCRIPTION
The SI9000+ will operate in two basic modes, "free run" and
"batch".
The free run mode includes reverse polarity operation, and is a
totally manual operation not using any of the automatic features or
interlocks of the SI9000+. Although it can be used as an operating
mode on its own it is also available from within batch mode.
The normal operating mode is batch mode and this should be used
for rating cans even if a batch is not being tested.
All instructions assume the SI9000+ is initially in standby mode, i.e.
the milliamp display shows dashes and set to the factory
configuration given in Appendix A Fig 1.
3.2
CONFIGURATION CHECK
The Rating voltage and rating duration can be inspected whilst the
SI9000+ is in standby mode (i.e. the milliamp display shows
dashes) and before or after rating. To examine the rating voltage
and rating duration:
i.
Press the set key. The rating voltage is displayed on the
milliamp display. The ↑ and ↓ key are ignored.
ii.
Press the set key again. The rating duration is displayed
on the timer display. The ↑ and ↓ keys are ignored.
iii.
Press the End key. The SI9000+ drops back to the
condition it was in before the configuration check was
made.
ENAMEL RATER MANUAL: V.10
15
3.3
FREE RUN AND REVERSE POLARITY
i.
Place a can in the can stand, fill with electrolyte and lower
the probe.
ii.
Press FREE RUN.
The rating indicator will light, the timer display will
increment and the current will be displayed in milliamps.
The check circuits will be active but will not prevent
operation. If either of the red warning indicators is lit then
more electrolyte should be added or the can should be
rotated against the body contacts as necessary.
iii.
To freeze the reading at any point press FREE RUN.
Rating will stop and the timer will be halted. Rating can
be resumed by pressing FREE RUN once more.
iv.
To reverse the polarity press REVERSE POLARITY.
Note that the contact and level check circuits are disabled
when reverse polarity is selected.
v.
To return to normal polarity, press FREE RUN.
vi.
To return to standby mode, press END.
Reverse polarity operation is similar, but press REVERSE
POLARITY instead of FREE RUN.
If the timer should reach 100 seconds, the mode will be cleared
automatically and the SI9000+ will revert to standby mode.
Free run and reverse polarity are available from within batch mode
and operate as described above except that pressing END or the
timer reaching 100 seconds will return the SI9000+ to the stage
from where free run or reverse polarity had been selected.
ENAMEL RATER MANUAL: V.10
3.4
BATCH MODE
3.4.1 RATING CANS
i.
Place a can in the can stand, fill with electrolyte and lower
the probe.
If the body contact and level checks are clear, the can number will
change to 1 and rating will commence. If either of the red warning
indicators are lit then more electrolyte must be added or the can
should be rotated against the body contacts as necessary.
After the preset time has elapsed (usually four seconds), rating will
stop and the reading will be frozen. The operator has the following
options:-
i.
Save the reading in the internal memory by pressing
SAVE.
ii.
Abandon the reading by pressing CLEAR.
iii.
Enter free run mode by pressing FREE RUN.
iv.
Enter reverse polarity mode by pressing REVERSE
POLARITY.
v.
View the set rating time and rating voltage by pressing
SET.
If options iii or iv are chosen, operation will be as described in
section 3.3. Remember that if END is pressed, or timeout occurs,
the original timed reading will be redisplayed and the operator must
press SAVE or CLEAR to continue.
Having pressed SAVE or CLEAR the SI9000+ will be waiting for a
new can and it will use its check circuits to detect the can being
changed. Free run and reverse polarity are available at this point.
To test the next can proceed as follows:-
i.
Remove the can just tested.
ii.
Insert a new can, fill with electrolyte, and lower the probe.
If the SI9005 can stand is being used, the SI9000+ will detect the
new can and as soon as both checks are clear will start rating the
new can. If another probe that does not support body contact and
level checks is being used, then press START to tell the SI9000+ it
has a new can.
By repeating the above operations up to 99 cans can be tested and
their results stored. When 99 readings have been stored a "batch
end" is automatically invoked.
ENAMEL RATER MANUAL: V.10
17
3.5
MASTER RESET
In the event of a major change, error or fault it may become
necessary to completely reset the instrument. This is always
required when the operating mode is changed.
A reset will erase all stored settings and reading from the units
memory.
To perform a master reset proceed as follows;
A)
Switch off power to the unit.
B)
Simultaneously hold down the SET, ↑ and ↓
keys.
C)
Switch on the power.
D)
After 2 seconds release the keys and the
system will be reset.
When a master reset is performed the operator has the opportunity
to change the default times and voltage settings (factory set to 4
seconds and 6.3V) using the SET, ↑ and ↓ keys. The new default
values are stored in battery backed memory and are retained until
another master reset is performed. The rating voltage and duration
can be checked during operation as described in Section 3.2.
ENAMEL RATER MANUAL: V.10
3.6
ERROR MESSAGES
"OL"
The SI9000+ has tripped out on current
overload. Remove the cause of the overload
and press CLEAR.
"ERR"
The Err message is displayed if the Batch/Stats
key is depressed before any readings have been
taken. Under these conditions it is not possible
to calculate statistical information and there is
no information available to send to a printer.
"FAIL"
This message may flash on power up and
indicates that battery back-up of memory has
failed. A master reset should be performed (see
section 3.5). The clear key can be pressed to
continue operation, but the chance that
previously saved readings are corrupted is very
high.
"BUSY"
The SI9000+ is waiting for the printer (see
Section 6.7).
"BUFFER
Scrolling, computer buffer is 90% full
NEARLY FULL"
(see Section 7.7).
"FULL"
Flashing, computer buffer is full (see Section
7.7).
"PSU"
The power supply generating the rating voltage
has failed. The SI9000+ may reset when
CLEAR or END is pressed, alternatively switch
off the SI9000+ for a few seconds.
"CLR?"
If BATCH/STATS is pressed when no printer is
connected and the statistical output is disabled
(see section 4.2) the CLR? message is
displayed. Press CLEAR to clear the memory or
END to return to the condition before
BATCH/STATS was pressed.
"PR ?"
BATCH/STATS has been pressed when the
statistical output is off (see section 4.2). The
raw data may be printed by pressing SAVE.
ENAMEL RATER MANUAL: V.10
19
4.
CONFIGURING YOUR SI9000+
In order to change the configuration of the SI9000+ it is necessary
to remove the base panel. Instructions for doing this are given in
section 4.1 (i) and (ii). Before changing the configuration of the
SI9000+ it is recommended that users familiarise themselves with
the internal layout shown in Appendix A Figure 1. A list of the
switches and jumpers of interest follows:
SWITCH/
JUMPER
NAME
LOCATION
TYPE
OPERATION
SW1
Micro board (Top
Board) To right of
SW2
Block of 8 miniature
rocker switches
numbered 1 to 8 on the
device
A switch is off when its rocker
is pressed down adjacent to
the "open" legend.
SW2
Micro board (Top
Board) To left of SW1
Block of miniature rocker
switches numbered 1 to 8
on the device
A switch is off when its rocker
is pressed down adjacent to
the "open" legend.
SW11
Applications board
(Lower Board) Below
Fuses
Block of miniature rocker
switches numbered 1 to 8
on the device
A switch is off when its rocker
is pressed down adjacent to
the "open" legend.
Mains
Selector
Switch
Applications board
(Lower Board) To the
right of the fuses
Slide Switch
Slide to correct position for
supply voltage.
J1
Micro board (Upper
Board) R.H.S.
3 position jumper
Connect the pair of pins
adjacent to the required
legend A,B or C using the
jumper.
J2
Micro board (Upper
Board)
2 position jumper
Connect the upper pair of pins
for A or the lower pair for B,
using the jumper.
ENAMEL RATER MANUAL: V.10
SWITCH/
JUMPER
NAME
LOCATION
TYPE
OPERATION
J3
Micro board (Upper
Board)
2 position jumper
Connect the left hand pair
of pins for A or the right
hand pair for B, using the
jumper.
VR1VR5
Applications board (Lower
board) below the
transformer
Multiturn Pot
Adjust in accordance with
Section 5.4.1.
VR6VR7
Applications board (Lower
board) to the right of the
transformer
Multiturn Pot
Adjust in accordance with
Section 5.4.2.
The factory settings for the jumpers and switches are shown in Appendix A Fig 1.
4.1
SETTING THE MAINS VOLTAGE
The mains voltage of the SI9000+ is factory set as appropriate for
the particular country to which it is supplied and the setting will be
clearly marked on the back panel. Should you need to change the
voltage proceed as follows:-
i.
Remove the six button head screws in the base of the
case using the 2.5mm hex wrench supplied.
ii.
Remove the base panel taking care not to damage the
earthing cable.
iii.
Referring to Appendix A Fig 1 move the mains voltage
switch to the required position.
iv.
Check that the foam seal on the bottom of the case is
intact and replace the base panel and six screws.
v.
Change or remove the voltage label.
ENAMEL RATER MANUAL: V.10
21
4.2
SETTING UP THE STATISTICAL OUTPUT
When SW11/1 and SW11/2 on the applications board (see
Appendix A Fig 1) are set according to the following table statistical
information is available when the Batch/Stats key is pressed.
SW11/1
SW11/2
AVAILABLE STATISTICAL OUTPUT
Off
Off
None
Off
On
Average followed by Range
On
Off
Average followed by Standard Deviation
On
On
Average followed by 3 x Standard Deviation
The statistical information is available after the reading from any
can within the batch has been saved.
When the statistical information has been examined, 3 choices are
available:
i.
Press the END key to add further cans to the batch.
ii.
Press the SAVE key to download the batch data to a
printer (assuming the printer and SI9000+ are correctly
configured, see Section 6).
iii.
Press the CLEAR key to erase the batch information from
the SI9000+'s memory.
4.3
THE LEVEL INTERLOCK
When SW11/5 on the applications board (see Appendix A Fig 1) is
set off the operation of the SI9000+ changes as follows:
i
The SI9000+ can no longer detect when a can change
has occurred and will not auto start after the change.
Note that the level check indicator continues to operate
normally.
ii
Enamel rating with the level check indicator on is allowed
and commences when START is pressed.
ENAMEL RATER MANUAL: V.10
4.4
THE CONTACT INTERLOCK
When SW11/6 on the applications board (see Appendix A Fig 1) is
set off the operation of the SI9000+ changes as follows:
i
The SI9000+ can no longer detect when a can change
has occurred and will not autostart after the change. Note
that the contact check indicator continues to operate
normally.
ii
Enamel rating with the contact check indicator on is
allowed and commences when START is pressed.
4.5
CHANGING THE RATING VOLTAGE
When SW11/7 on the applications board (see Appendix A Fig 1) is
set on the operation of the SI9000+ changes as follows:
i
On completion of the power up self test the default
voltage setting of 6.3V is displayed on the mA display.
This can be changed to 4,10,12 or 15V using the ↑ and ↓
keys. When the required rating voltage is displayed,
press END to "drop back" to normal rating operation.
ii
The rating voltage may then be changed at the start or
end of a rating cycle, by pressing SET and using the ↑
and ↓ keys. When the required rating voltage is
displayed, press END to "drop back" to normal rating
operation.
4.6
CHANGING THE RATING DURATION
When SW11/8 on the applications board (see Appendix A Fig 1) is
set on the operation of the SI9000+ changes as follows:
i.
On completion of the power up self test the default rating
duration of 4 seconds is displayed on the timer display.
This can be altered over the range 2 to 99 seconds using
the ↑ and ↓ keys. When the required rating voltage is
displayed, press END to "drop back" to normal rating
operation.
ii.
The rating duration may then be changed at the start or
end of a rating cycle, by pressing SET and using the ↑
and ↓ keys. When the required rating duration is
displayed, press END to "drop back" to normal rating
operation.
ENAMEL RATER MANUAL: V.10
23
4.7
AUTORANGING OPERATION
When SW1/1 on the micro board (see Appendix A Fig 1) is set on
the operation of the SI9000+ changes as follows:
1.
All enamel rating currents up to 45mA are displayed and
saved to two decimal places (10μA). Note that 5mA of
hysteresis is incorporated in the design, so that with a
decreasing current the SI9000+ does not operate to two
decimal places until the current drops below 40mA.
Currents above the thresholds continue to be displayed to
one decimal place (100μA).
4.8
LINE NUMBER ENABLE
When SW1/2 on the micro board (see Appendix A Fig 1) is set on
the operation of the SI9000+ changes as follows:
1.
On power up the SI9000+ prompts for a line number.
This can be set to any value between 0 and 99 using the
↑ and ↓ keys. One or more cans can then be enamel
rated as a batch. When the data is sent to a printer or
host computer the line number is also sent.
2.
When a batch of measurements have been taken and
statistical information, a printout or data transfer
completed, pressing END will cause the SI9000+ to
prompt for a new line number. Note that once a new line
number is entered, all data from a previous line is erased.
The line number facility is provided primarily so that a
complete set of data can be sent to a printer or host
computer.
ENAMEL RATER MANUAL: V.10
4.9
AUTOSTART
When SW1/3 on the micro board (see Appendix A Fig 1) is set off
the operation of the SI9000+ changes as follows:
1.
The semi-automatic operation described in Para 1.2.2.
does not operate. After placing a can in the can stand the
operator must press start and ensure that the level and
contact check indicators are off before enamel rating will
commence.
Note that if either SW11/5 or SW11/6 is off then the
autostart function is disabled, and rating will commence
only when START is pressed.
4.10
AUTOSAVE
When SW1/4 on the micro board (see Appendix A Fig 1) is set on
the operation of the SI9000+ changes as follows:
1.
The SI9000+ operates in a "fully automatic" mode. After
each can has been rated the displayed reading is
automatically saved as the can is removed from the can
stand and the contact check and level check indicators
are illuminated. The combination of autosave and
autostart results in a "fully automatic" operation.
ENAMEL RATER MANUAL: V.10
25
5.
MAINTENANCE AND SERVICING
5.1
CARE OF YOUR SI9000+
The SI9000+ is a precision instrument and should be treated as
such. Avoid exposure to extremes of temperature, humidity,
mechanical vibration and shocks. Although the case is sealed to
I.P.65, electrolytes and other liquids should be kept away from the
unit as much as possible. These few simple precautions will ensure
your SI9000+ will hold its calibration closely and give years of
reliable service.
5.2
ROUTINE MAINTENANCE
The following procedures should be carried out weekly or more
frequently depending on usage:-
i.
Clean the enamel rater case and can stand with a cloth
dampened with water. Do not use solvents or abrasives
under any circumstances. If greasy or oily deposits are
present the use of dishwashing detergent is permissible.
ii.
Check the facia panel, case and seals of the SI9000+ for
damage.
iii.
Check the external wiring for damage.
iv.
Check all connectors for security.
v.
Check the operation of both the enamel rater and the can
stand.
5.3
ROUTINE TESTING
Once a day or more frequently depending on usage the following
test should be carried out using the test cable supplied with the
SI9000+:-
Referring to Fig.2 (Page 4)
i.
Remove the can locating ring (item 8).
ii.
Attach the large crocodile clip of the test cable to the body
contact (item 7) furthest from the inner column (item 10).
iii.
Attach the smaller crocodile clip to the main probe (item
6).
iv.
Press FREE RUN.
ENAMEL RATER MANUAL: V.10
The 'Rating', 'Low Level' and 'Contact Fault' indicators
should be lit and the milliamp display should show the
reading given in the table below for the rating voltage in
use:
RATING
VOLTAGE
MILLIAMP DISPLAY READING
4
4+0.2mA
6.3
6.3+0.2mA
10
10+0.3mA
12
12+0.4mA
15
15+0.5mA
If this is not the case first check that the crocodile clips
are making good contact. If this does not cure the fault
refer to section 5.5 'Fault Finding'.
v.
Repeat the test in reverse polarity.
Now the 'Low Level' and 'Contact Fault' indicators should
be off and the 'Reverse Polarity' and rating indicators
should be lit. The milliamp reading should be the same.
Note:
This test is far superior to internal calibration tests as it
will detect the more likely faults such as corroded or dirty
connectors and damaged probe cables. It will not,
however, show up equal and opposite drift of internal
voltage and current references. This test should therefore
be supplemented with calibration against a good quality
external meter as detailed in section 5.4.
ENAMEL RATER MANUAL: V.10
27
5.5
FAULT FINDING
None of the LED indicators or displays light when the on/off switch
is depressed.
i.
Check that the mains supply is connected.
ii.
Check that the fuse in the mains plug (if any) has not
blown.
iii.
Check that the internal fuse has not blown (Refer to
Section 5.6)
iv.
Contact Sencon.
The operation has "locked up" and/or the displays show
meaningless information.
i.
Perform a master reset (refer to Section 3.5). All readings
will be lost from memory. Ensure that the rating voltage
and duration is correctly reset.
ii.
Check that the mains voltage setting is correct.
iii.
Contact Sencon.
NOTE:
If this fault occurs frequently and operation 1 cures the
fault, it is likely that the mains supply to the unit is noisy or
interrupted. This must be corrected for reliable operation.
The milliamp reading is incorrect with the test cable connected.
i.
Check that the mains voltage setting is correct.
ii.
Check that the probe plug is secure in its socket.
iii.
Inspect the probe wiring and connectors for damage or
corrosion.
iv.
Proceed to Section 5.4 "Calibration".
ENAMEL RATER MANUAL: V.10
5.6
REPLACING THE FUSE
The internal fuse is 500mA anti-surge type, size 20mm x 5mm. A
spare fuse is contained inside the case. Proceed as follows :-
i.
Switch off the SI9000+ and isolate from the mains supply.
ii.
Remove the six button head screws using the 2.5mm hex
wrench supplied.
iii.
Remove the base panel taking care not to damage the
earthing wire.
iv.
Referring to Appendix A Fig.1 locate the working fuse
and the spare fuse.
v.
Remove the fuse carrier and fuse by lifting at one end.
NOTE:
The fuse type fitted may not show a break in the
conductor when it has blown. If in doubt test with an ohm-
meter.
vi.
Replace or exchange the fuse as necessary.
vii.
Refit the base panel, checking that the seal is intact.
viii.
Refit the six button head screws.
5.7
OBTAINING HELP FROM SENCON
If you have a problem with your SI9000+ and require help from
Sencon, the following information will greatly help us to assist you :-
i.
Serial number of the unit.
ii.
Version number of the software (This can be obtained by
pressing CLEAR and SAVE simultaneously whilst in
Standby mode).
iii.
Exact description of the fault.
iv.
When it happens or what actions cause the fault to occur.
v.
Name of a contact (preferably technical) at the location.
ENAMEL RATER MANUAL: V.10
29
6.
USING THE SI9000+ WITH A PRINTER
6.1
OVERVIEW
The use of a printer with the SI9000+ relieves the operator from
having to record the results using pen and paper which is time
consuming and can be a source of errors. The unit is shipped with
the printer output off and this must be enabled before a printer is
connected.
6.2
CHOOSING A PRINTER
A large variety of printers could be used with the SI9000+, the most
obvious choice probably being a low cost 80 column dot matrix
printer intended for home or office use. In essence, the SI9000+
should work with any type of printer that is equipped with an RS232
serial interface, however, because of the large range available
Sencon cannot accept responsibility for the performance or settings
of a particular printer.
A word of caution is required regarding the RS232 interface.
Firstly, most printers intended for office use are fitted as standard
with a parallel or centronics interface, this is not suitable for
connection to the SI9000+ so the requirement for RS232 interface
must be clearly stated to your supplier. Secondly, it must be
accepted that the RS232 interface on the printer and possibly on
the SI9000+ will have to be set up, sometimes a confusing
operation for the uninitiated and thirdly, it may be necessary to
obtain and solder a different connector onto the printer lead
supplied by Sencon as standard.
Your printer supplier should be able to assist you with any of these
problems if they occur.
The following printers have been successfully tested with the
SI9000+;
1)
Any printer from the Epson FX or EX range provided it is
fitted with the Epson #8143 serial interface.
2)
The Tandy DMP132 printer. This model is fitted with an
RS232 interface as standard.
ENAMEL RATER MANUAL: V.10
6.3
CONNECTING THE PRINTER
The printer connection is the 3 pin socket at the side of the
instrument as shown. The pin out and cable colour code for
printer connections are given below;
FIG 3: SIDE PANEL
PIN
FUNCTION
SENCON
CABLE
COLOUR
1
GROUND
BLUE
2
TxD (TRANSMIT DATA)
RED
3
CTS (CLEAR TO SEND)
YELLOW
6.3.1
THE EPSON FX OR EX PRINTER
This printer will need to be fitted with the Epson #8143 serial
interface which offers a 25 way D type female connector. The wire
connection is as follows.
SI9000+
PIN
SENCON CABLE
COLOUR
EPSON PIN
1
BLUE
7
2
RED
3
3
YELLOW
11
It should be noted that a number of other printers share this type of
connector and pin connection.
PRINTER
CONNECTOR
ENAMEL RATER MANUAL: V.10
31
6.3.2
THE TANDY DMP132 PRINTER
This printer utilises a 4 pin DIN style connector for its serial
interface (Radio Shack Cat No. 274-007). Wire connection is;
SI9000+ PIN
SENCON CABLE
COLOUR
EPSON PIN
1
BLUE
3
2
RED
4
3
YELLOW
2
6.3.3
USING OTHER PRINTERS
If a different printer is used the connector type and pin numbers will
be given in the printer user manual. Having obtained the correct
connector it will be necessary to connect to it as follows;
SI900+
PIN
COLOUR
PRINTER CONNECTION
1
BLUE
Signal Ground or OV. Not to be confused with screen or
protective ground which may be a different pin.
2
RED
TxD or Transmit Data. This is the wire that carries the data
signal with respect to OV out of the SI9000+.
3
YELLOW
CTS or clear to send. This signal is generated by the printer
and is used to stop and start the flow of data from the SI9000+.
It must be connected otherwise it will stop data transmission.
ENAMEL RATER MANUAL: V.10
6.4
SETTING UP THE RS232 DATA FORMAT
The RS232 standard allows a large number of permutations in its
data transmission format. It is therefore necessary to ensure that
the printer is set to the same transmission format as the SI9000+.
The format of the data transmission on the printer port of the
SI9000+ is;
300/1200/9600 Baud - Selectable
8 DATA BITS
NO PARITY BIT
1 STOP BIT
It is possible to change the baud rate as described in section 6.5
below, however, Sencon recommend the 1200 baud is used.
6.4.1
SETTING UP THE EPSON PRINTER
The Epson printer fitted with the #8143 serial interface has a bank
of 8 small switches on the interface board that are used to select
the transmission format. This switch bank is marked SW1 and
should be set as below to achieve the correct format and 1200
baud.
SWITCH NO.
SWITCH NAME
SETTING
FUNCTION
1-1
JB3
OFF
BAUD RATE SELECTION
1-2
J8/7
OFF
WORD LENGTH
1-3
JB1
OFF
BAUD RATE SELECTION
1-4
JB2
OFF
BAUD RATE SELECTION
1-5
JO/E
OFF
PARITY ODD/EVEN
1-6
JPDS
OFF
PARITY ON/OFF
1-7
JB4
ON
BAUD RATE SELECTION
1-8
P/2
ON
SERIAL/PARALLEL
ENAMEL RATER MANUAL: V.10
33
6.4.2
SETTING UP THE TANDY PRINTER
The Tandy DMP132 has two sets of switches used to configure the
printer. These are accessed through an opening in the back of the
unit, near the serial port connector. The recommended settings for
these switches are;
SWITCH NO.
SETTING
FUNCTION
1-1
OFF
TANDY CODES
1-2
ON
CR=CR
1-3
OFF
LF=LF
1-4
OFF
11 INCH FORM
1-5
OFF
SKIP INVALID
1-6
OFF
1//6 INCH LF
1-7
OFF
TANDY CHARACTERS
1-8
ON
WP MODE
2-1
ON
1200 BAUD
2-2
OFF
SERIAL PORT
6.4.3
SETTING UP OTHER PRINTERS
For other printers it is essential that the serial port is set to match
the format as given in section 6.4 above. This may entail the
operation of switches or jumpers inside the printer as described in
the printer instruction manual.
The SI9000+ does allow the baud rate to be changed between 300,
1200 and 9600 baud. We recommend the use of 1200 baud as
this provides the optimum compromise between printout speed and
reliable operation.
6.5
ENABLING THE PRINTOUT
The SI9000+ is shipped with its printer port disabled. This is to
prevent the SI9000+ from attempting to send data to a non-existent
printer and subsequently reporting a communications failure.
The printout is enabled by SW1/7 and SW1/8 on the micro board
(Refer to Appendix A Fig 1). The settings of switches SW1/7 and
SW1/8 not only enable the printer but also set the baud rate of
transmission. We recommend the use of 1200 baud as the ideal
compromise between speed of operation and noise immunity and if
the settings detailed in sections 6.4.1 and 6.4.2 are used, 1200
baud must be selected.
Switch positions and corresponding selections are given below;
ENAMEL RATER MANUAL: V.10
SWITCH
SW1/7
SWITCH
SW1/8
SELECTION
OFF
OFF
PRINTOUT OFF
OFF
ON
PRINTOUT ON-9600 BAUD
ON
OFF
PRINTOUT ON-1200 BAUD
ON
ON
PRINTOUT ON-300 BAUD
Having enabled the SI9000+'s printer port and selected the printout
Format (see Section 6.6), the SI9000+ should be connected to a
printer configured as described in Section 6.4. The SI9000+ will
transmit data to the printer when SAVE is pressed after
BATCH/STATS. At this point the values of the cans rated in the
batch, the statistical conformation and the line number, if enabled,
will be printed.
6.6
SELECTING THE PRINTOUT FORMAT
The printed output is arranged in columns according to the settings
of SW1/5 and SW1/6 on the micro board (refer to Appendix A Fig
1). The relationship is given in the table below:
SW1:5
SW1:6
COLUMNS
OFF
OFF
1
OFF
ON
4
ON
OFF
6
ON
ON
8
The statistical information selected by SW11/1 and SW11/2 on the
applications board will also be printed (refer to Appendix A Fig 1
and Section 4.2).
If the prompt for line number option has been selected (See section
4.8) then the line number will be included on the printout.
ENAMEL RATER MANUAL: V.10
35
6.7
PRINTER PROBLEMS
If the printer option has been enabled (section 6.5) the instrument
will attempt to transmit the data when SAVE is pressed after
BATCH/STATS. Whilst the transmission takes place the word
"busy" will appear in the display. If the printer is disconnected, is
turned off or has some other problem the transmission will not take
place and the word busy will remain until the printer problem has
been corrected or the CLEAR key is pressed.
If the word "busy" appears check that the printer is O.K. (i.e. no
paper jam etc), check that it is on line, check the cable and
connections, and check the printer set up.
If the printout takes place but not correctly, check that both the
printer and SI9000+ are set to the same baud rate and that the
printer is set to the data format as given in section 6.4.
ENAMEL RATER MANUAL: V.10
7.
USING THE SI9000+ WITH A COMPUTER
7.1
GENERAL
The SI9000+ incorporates a second serial port intended for
connection to a host computer. This enables readings taken by the
SI9000+ to be transferred to the computer for statistical analysis
and archiving.
The data format and communications protocol are selectable and it
will require a suitably experienced person to organise and set up
such a connection. It will also be necessary to have software
produced for the computer that is capable of receiving and
understanding the data from the SI9000+.
The SI9000+ supports bi-directional or uni-directional
communication. Internal jumpers are used to select between
RS232C, RS422 and RS485 communication standards.
7.2
COMPUTER PORT CONNECTIONS
Serial Port 2 is the 4-way connector on the side of the SI9000+.
The pinout of this connector is shown below.
FIG 6: COMPUTER PORT CONNECTOR
ENAMEL RATER MANUAL: V.10
37
A suitable 4 way plug is supplied with each SI9000+.
7.3
SETTING DATA FORMAT AND SPEED
The computer port is set up using SW2, J1, J2 and J3 on the micro
board and SW11/3 and SW11/4 on the applications board (Refer to
Appendix A Fig 1. All switches must be set to match the data
transmission set up of the serial port on the host computer. SW2/1
on the micro board must be on to enable the computer port output.
The switch functions are as follows:
SWITCH BANK 2
SWITCH
POSITION
OPEN
CLOSED
S2/1
Computer Output Off
Computer Output On
S2/2
Software Handshake Off
Software handshake On
Computer Transmission Data Format as below:-
S2/3
S2/4
S2/5
FORMAT
OPEN
OPEN
OPEN
7 data bits-Even Parity-2 Stop Bits
OPEN
OPEN
CLOSED
7 data bits-Odd Parity-2 Stop Bits
OPEN
CLOSED
OPEN
7 data bits-Even Parity-1 Stop Bits
OPEN
CLOSED
CLOSED
7 data bits-Odd Parity-1 Stop Bits
CLOSED
OPEN
OPEN
8 data bits-No Parity-2 Stop Bits
CLOSED
OPEN
CLOSED
8 data bits-No Parity-1 Stop Bits
CLOSED
CLOSED
OPEN
8 data bits-Even Parity-1 Stop Bits
CLOSED
CLOSED
CLOSED
8 data bits-Odd Parity-1 Stop Bits
S2/6}
Computer Transmission Baud Rate as defined below:
S2/8
S2/7
S2/6
Baud Rate
OPEN
OPEN
OPEN
76800 *
OPEN
OPEN
CLOSED
38400
OPEN
CLOSED
OPEN
19200
OPEN
CLOSED
CLOSED
9600
CLOSED
OPEN
OPEN
4800
CLOSED
OPEN
CLOSED
2400
CLOSED
CLOSED
OPEN
1200
CLOSED
CLOSED
CLOSED
300
*NOTE: Sencon do not guarantee operation at this speed.
ENAMEL RATER MANUAL: V.10
The jumpers on the micro board must be changed according to the
communications standard used:
RS232
RS422
RS485
J1
C
A
B
J2
B
A
A
J3
B
A
A
SW11/4 on the applications board selects between RS422 and
RS485. When SW11/4 is off RS422 is selected. If RS422 or
RS485 is to be used the plug connecting the socket on the rear
panel to the microboard should be moved from P2 to P3. These
connectors are marked on the circuit board.
7.4
UNI-DIRECTIONAL PROTOCOL
If SW11/3 on the applications board is off (open), uni-directional
protocol is used. As each reading is saved it is automatically sent
down the serial port, if the hardware handshaking line is clear.
7.5
BI-DIRECTIONAL PROTOCOL
If SW11/3 on the applications board is on (closed), bi-directional
protocol is used. As each reading is saved it is stored in an internal
buffer, which is battery backed in case of power failure. When the
enamel rater receives a data request message from the host
computer the data stored in the buffer is sent.
Data Request message:
<40h><01h>
where:
<40h> (1 byte) ASCII character '@'
<01h> (1 byte)
When using bi-directional communications, software handshaking
is available as an option as well as hardware handshaking. If
selected, (SW2/2 closed on the micro board) the XON/XOFF
protocol is used. If XOFF is sent to the enamel rater it must be
followed by XON before any further transmission will occur.
ENAMEL RATER MANUAL: V.10
39
7.6
FORMAT OF MESSAGE TRANSMITTED
The format of the message sent is dependent on the settings of the
appropriate DIP switches, and the mode of communication
selected. If all options are selected the message is as follows:
<4CH><20H><52h><20h>
where:
<4CH>
(1 byte) ASCII 'L'
<20H>
(1 byte) ASCII ' '
(2 bytes) ASCII characters for digits,
a decimal point
e.g. 99 = '9' '9'
<52H>
(1 byte) ASCII 'R'
<20H>
(1 byte) ASCII ' '
(5 bytes) ASCII characters for digits, and a
decimal point
e.g. 126.4 = '1'' 2'' 6'' . ''4'
If the LINE NUMBER option is not selected then the appropriate
header and associated message are not transmitted and the two
byte header before the actual enamel reading is not sent.
When bi-directional communications are used an end-of-data byte
is transmitted when the whole buffer has been sent. This byte is
also sent if the buffer is empty when the data request message is
received.
End-of-data byte:
<04H>
NOTE:
RS485 communications facilities are not available. They
will be implemented at a later date.
7.7
ERROR MESSAGES
"Buffer nearly full" will scroll across the large display when the
internal buffer is 90% full.
The message is cleared automatically when the level drops below
90%. Alternatively the CLEAR key can be pressed.
FULL will flash on the large display when the internal buffer is full.
The message will clear automatically when space is available in the
buffer, i.e. after a message is sent. If CLEAR is pressed the oldest
data, not sent, in the buffer will be overwritten next time a reading is
saved.
ENAMEL RATER MANUAL: V.10
APPENDIX A
Voltage regulation switch
Working Fuse
Spare fuse
Voltage trim pots
Current trim pots
See note 1
See note 1
FIGURE 1: PCB LAYOUT
ENAMEL RATER MANUAL: V.10
41
1.
The factory switch and jumper settings are:
MICROBOARD
J1 = C
J2 = B
J3 = B
J4 = ON
SW1
SW2
1
OFF
OFF
2
OFF
OFF
3
ON
OFF
4
OFF
OFF
5
OFF
OFF
6
OFF
OFF
7
OFF
OFF
8
OFF
OFF
APPLICATIONS BOARD
SW11
1
OFF
2
OFF
3
OFF
4
OFF
5
ON
6
ON
7
OFF
8
OFF
Switches are off when the rockers are pressed down adjacent to
the open legend.
ENAMEL RATER MANUAL: V.10
2.
SWITCH AND JUMPER FUNCTIONS:
REF
SECTION
REF
SECTION
SW1/1 AUTORANGING OPERATION
4.7
SW2/1 COMPUTER OUTPUT
7.3
SW1/2 LINE NUMBER ENABLE
4.8
SW2/2 SOFTWARE
HANDSHAKING
7.3
SW1/3 AUTOSTART
4.9
SW2/3 COMPUTER
}
7.3
SW1/4 AUTOSAVE
4.10
SW2/4 DATA
SW1/5 PRINTOUT FORMAT
6.6
SW2/5 FORMAT
}
SW1/6 PRINTOUT FORMAT
6.6
SW2/6 AND
}
SW1/7 PRINTOUT ENABLE AND BAUD
6.5
SW2/7 BAUD
SW1/8 PRINTOUT ENABLE AND BAUD
6.5
SW2/8 RATE
SW11/1 STATISTICAL OUTPUT
4.2
SW11/2 STATISTICAL OUTPUT
4.2
SW11/3 BI-DIRECTIONAL COMMS
7.5
SW11/4 422/485
7.3
SW11/5 LEVEL INTERLOCK
4.3
SW11/6 CONTACT INTERLOCK
4.4
SW11/7 RATING VOLTAGE
4.5
SW11/8 RATING DURATION
4.6
JUMPERS J1, J2 AND J3 MUST BE CONFIGURED TO SUIT THE
COMMS STANDARD USED (REF SECTION 7.3).
ENAMEL RATER MANUAL: V.10
Specification
Operating voltage
220 setting
220-250 V a.c. 50-60
Hz
110 setting
100-125 V a.c. 50-60
Hz
Power consumption
30 watts
Rating Voltage
4.0 volts +/- 1%
for rating
6.3 volts +/- 1%
currents in
10.0 volts +/- 1%
the range
12.0 volts +/- 1%
0 to 100mA
15.0 volts +/- 1%
Accuracy of milliamp reading
+/- 1% of reading +/- 1 L.S.D.
Milliamp measurement range
0.00mA to 800.0mA
Operating Temperature
5°C to 40°C
Environmental Protection
I.P. 65
ENAMEL RATER MANUAL: V.10
45
REVISION AT V.01
i.
Para 3.6 "PSU"
-
"The SI9000+ may reset when
CLEAR or END is pressed " was
"The Si9000+ may reset
automatically".
ii.
Para 6.1
-
Paragraph revised for clarity.
iii.
Para 7.3
-
Switch S2/8 and S2/6 settings have
been swapped. Computer
Transmission Baud rate was
incorrectly defined as:
S2/6
S2/7
S2/8
Baud Rate
OPEN
OPEN
OPEN
76800
OPEN
OPEN
CLOSED
38400
OPEN
CLOSED
OPEN
19200
OPEN
CLOSED
CLOSED
9600
CLOSED
OPEN
OPEN
4800
CLOSED
OPEN
CLOSED
2400
CLOSED
CLOSED
OPEN
1200
CLOSED
CLOSED
CLOSED
300
iv.
P.40 Figure 7
-
The front Panel drawing added.
REVISION AT V.02
i.
Manual Format Change.