CONFIGURING YOUR ELI FOR ECG SAFE

ECG Safe Choose your device: • ELI 10 • ELI 150c and 250c • ELI 350 • ELI 280 ELI 10 Based on v1.13. 1. Delete all existing ECGs from memory. 1. From the main menu, press F6 More. 2. Press the number 3 for Set Time/Date. 3. Press and hold the Shift and Alt keys, and then press C. 4. Enter the ELI’s administrator password. 5. Press and hold the Shift and Alt keys, and then press F. 6. Press F1 OK. 7. Press and hold the Power button to reboot the ELI. 8. Wait about 10 seconds, and then press the Power button to turn the ELI on. 2. From the main menu, press F6 More. 3. Press the number 3 for Set Time/Date. 4. Press and hold the Shift and Alt keys, and then press C. 5. Enter the ELI’s administrator password. 6. If your ELI does not have Rx Multi-study support: 1. Press the F2 down arrow to move the cursor to Site Number. 2. Enter your ECG Safe’s Site Number. 3. Press the F2 down arrow to move the cursor to the Site Name. 4. Enter your organization’s name as you want it to appear in the footer of the ECGs. 7. Press the F4 Page button until you are on Page 5. 8. Press F3 to select WLAN or LAN, depending on which type of network connection you’ll be using. 9. Press F2 down arrow to move to the DHCP setting. 10. Select Yes if your network will assign the ELI an IP address. If you select No, you will need to configure a static IP Address, Default Gateway, and Subnet Mask for the ELI in the next 3 settings under DHCP. 11. Press F2 down arrow to move to the Host IP field. 12. Enter your ECG Safe’s Host IP. You must enter 3 digits in each field. Use leading 0’s as necessary. 13. Press F2 down arrow to move to the Port Number field. 14. Enter your ECG Safe’s Port Number.

CONFIGURING YOUR ELI FOR ECG SAFE

15. If using WLAN: 1. Press F2 down arrow to move to the SSID field. 2. Enter your WLAN’s SSID. 3. Press F2 down arrow to move to the Security field. 4. Press the F3 button until you have selected your network’s security protocol. 5. Press F2 down arrow to move to the Security field. 6. Enter the security information for your network, e.g. PSK Passphrase. 7. Press F2 down arrow to move to the WLAN Mode field. 8. Use F3 to select Fast if you want the ELI 10 to connect to your WLAN faster. Note, this mode uses a little more battery power to keep the WLAN radio powered. 16. Press F2 down arrow to move to the Time Sync field. 17. Press F3 to select ELI Link Xmt. This will cause the ELI to set its internal clock from ECG Safe’s every time it sends new ECGs to ECG Safe. Be sure the ELI’s time zone and DST settings are correct for your location. 18. Press F2 down arrow to move to the Comm. Protocol field. 19. Press F3 to select UNIPRO32. 20. If your ELI has Rx Multi-study support: 1. Press the F2 down arrow to move the cursor to one of the Site Number fields. 2. Enter your ECG Safe’s Site Number. 3. Press the F2 down arrow to move the cursor to the Site Name. 4. Enter your organization’s name as you want it to appear in the footer of the ECGs. 21. Press F6 Exit to save the settings. ELI 150c and 250c Based on v2.00. 1. Delete all existing ECGs from memory. 1. From the main menu, press F6 More. 2. Press F5 Set Time/Date. 3. Press and hold the Shift and Alt keys, and then press C. 4. Enter the ELI’s administrator password. 5. Press and hold the Shift and Alt keys, and then press F. 6. Press F6 Exit. 2. From the main menu, press F6 More. 3. Press F5 Set Time/Date. 4. Press and hold the Shift and Alt keys, and then press C. 5. Enter the ELI’s administrator password. 6. If your ELI does not have Rx Multi-study support: 1. Press the F2 down arrow to move the cursor to Site Number. 2. Enter your ECG Safe’s Site Number. 3. Press the F2 down arrow to move the cursor to the Site Name. 4. Enter your organization’s name as you want it to appear in the footer of the ECGs. 7. Press the F4 Page button until you are on Page 6 (for LAN), or Page 7 (for WLAN). 8. Press F2 down arrow to move to the DHCP setting. 9. Select Yes if your network will assign the ELI an IP address. If you select No, you will need to configure a static IP Address, Default Gateway, and Subnet Mask for the ELI in the next 3 settings under DHCP.

CONFIGURING YOUR ELI FOR ECG SAFE

10. Press F2 down arrow to move to the Host IP field. 11. Enter your ECG Safe’s Host IP. You must enter 3 digits in each field. Use leading 0’s as necessary. 12. Press F2 down arrow to move to the Port Number field. 13. Enter your ECG Safe’s Port Number. 14. If using WLAN: 1. Press F2 down arrow to move to the SSID field. 2. Enter your WLAN’s SSID. 3. Press F2 down arrow to move to the Security field. 4. Press the F3 button until you have selected your network’s security protocol. 5. Press F2 down arrow to move to the Security field. 6. Enter the security information for your network, e.g. PSK Passphrase. 7. Press F2 down arrow to move to the WLAN Mode field. 15. Press F2 down arrow to move to the Comm. Protocol field. 16. Press F3 to select UNIPRO. 17. Press F2 down arrow to move to the Sync Mode field. 18. Press F3 to select Xmt. 19. Press F2 down arrow to move to the Sync Date Time field. 20. Press F3 to select Yes. This will cause the ELI to set its internal clock from ECG Safe’s every time it sends new ECGs to ECG Safe. Be sure the ELI’s time zone and DST settings are correct for your location. 21. If you want to make the patient’s last name, or the patient’s ID, or the technician’s ID mandatory, select Yes in the corresponding fields below. 22. If your ELI has Rx Multi-study support: 1. Press the F2 down arrow to move the cursor to one of the Site Number fields. 2. Enter your ECG Safe’s Site Number. 3. Press the F2 down arrow to move the cursor to the Site Name. 4. Enter your organization’s name as you want it to appear in the footer of the ECGs. 23. Press F6 Exit to save the settings. ELI 350 Based on v1.42. 1. Delete all existing ECGs from memory. 1. Select Settings from the main screen. 2. Enter clear data in the password field. This special phrase is all lower case and includes a space between clear and data. 3. Click the knob or press Enter to give the command. 4. You will see the message “Clearing Data…” in the System Settings button. 2. Select Settings from the main screen. 3. Enter the administrator password. 4. Select System Settings. 5. Move the cursor to Site Number. 6. Enter your ECG Safe’s Site Number. 7. Move the cursor to the Site Name. 8. Enter your organization’s name as you want it to appear in the footer of the ECGs. 9. Move the cursor to Comm. Protocol.

CONFIGURING YOUR ELI FOR ECG SAFE

10. Select UNIPRO32. 11. Move the curse to XMT (SYNC) Button. 12. Select Time Sync, Transmit ECG. 13. Move the cursor to XMT (SYNC) Media. 14. Select LAN or WLAN, depending on which type of network connection you’ll be using. 15. If using LAN: 1. Move the cursor to LAN DHCP. 2. Select Yes if your network will assign the ELI an IP address. If you select No, you will need to configure a static IP Address, Default Gateway, and Sub Net Mask for the ELI in the next 3 settings under DHCP. 3. Move the cursor to the LAN Host IP field. 4. Enter your ECG Safe’s Host IP. 5. Move the cursor to the LAN Port Number field. 6. Enter your ECG Safe’s Port Number. 16. If using WLAN: 1. Move the cursor to WLAN DHCP. 2. Select Yes if your network will assign the ELI an IP address. If you select No, you will need to configure a static IP Address, Default Gateway, and Sub Net Mask for the ELI in the next 3 settings under DHCP. 3. Move the cursor to the WLAN Host IP field. 4. Enter your ECG Safe’s Host IP. 5. Move the cursor to the WLAN Port Number field. 6. Enter your ECG Safe’s Port Number. 7. Move the cursor to the WLAN Connect field. 8. Select Fast if you want the ELI 350 to connect to your WLAN faster. Note, this mode uses a little more battery power to keep the WLAN radio powered. 9. Move the cursor to the SSID field. 10. Enter your WLAN’s SSID. 11. Move the cursor to the WLAN Security field. 12. Select your network’s security protocol. 13. Move the cursor to the appropriate Security field. 14. Enter the security information for your network. 17. Move the cursor to the Save button and select it to save the settings. ELI 280 Based on v1.10. 1. Delete all existing ECGs from memory. 1. Touch the Gears icon on the main screen. 2. Touch the Advanced button. 3. Enter the administrator password. 4. Touch the Service button. 5. Touch the Erase Records button. 6. Touch Yes to confirm. 7. Touch the Done button twice. 2. Touch the Gears icon on the main screen.

CONFIGURING YOUR ELI FOR ECG SAFE

3. Touch the Advanced button. 4. Enter the administrator password. 5. Touch the LAN or WLAN button, depending on which type of network connection you’ll be using. 6. If using LAN: 1. Touch DHCP. 2. Select Yes if your network will assign the ELI an IP address. If you select No, you will need to configure a static IP Address, Def. Gateway, and Subnet Mask for the ELI in the next 3 settings under DHCP. 3. Touch the Sync IP field. 4. Enter your ECG Safe’s Host IP. 5. Touch the Port Number field. 6. Enter your ECG Safe’s Port Number. 7. Touch the Done button to save the LAN settings. 7. If using WLAN: 1. Touch DHCP. 2. Select Yes if your network will assign the ELI an IP address. If you select No, you will need to configure a static IP Address, Def. Gateway, and Subnet Mask for the ELI in the next 3 settings under DHCP. 3. Touch the Sync IP field. 4. Enter your ECG Safe’s Host IP. 5. Touch the Port Number field. 6. Enter your ECG Safe’s Port Number. 7. Touch the SSID field. 8. Enter your WLAN’s SSID. 9. Touch the Security field. 10. Select your network’s security protocol. 11. If you need to enter a passphrase, touch the field and enter the passphrase. 12. Touch the Done button to save the WLAN settings. 8. Touch the System button. 9. Touch the Site Number field. 10. Enter your ECG Safe’s Site Number. 11. Touch the double right arrow button to move to the next page. 12. Touch the Site Name field. 13. Enter your organization’s name as you want it to appear in the footer of the ECGs. 14. Touch the Sync Mode field. 15. Select Transmit+Orders+Date/Time. 16. Touch the Comm. Protocol field. 17. Select UNIPRO32. 18. Touch the Done button to save the settings. Ref: 9508-201-03-ENG Rev 1 Copyright © 2014 Mortara Instrument. All Rights Reserved.