Mat. 711540 Ver. A
This electrocardiograph is shipped with the battery disconnected to prevent discharge.
Connecting the Battery
1.
Unplug the electrocardiograph from AC power if connected.
2.
Turn the electrocardiograph upside-down.
3.
Unscrew and remove the battery door. See Figure 1.
4.
Lift out the battery. See Figure 2.
Your electrocardiograph was shipped with the battery (DC) fuse disconnected.
5.
Connect the wire terminal to the battery fuse as shown in the illustration. See Figure 3.
Caution To optimize battery operation and life, keep the electrocardiograph plugged in for at least 12 hours before using it on battery power.
WARNING Disconnect the power cord from AC power before servicing.
Figure 1. Removing the Battery Door
Figure 2. Lifting Out the Battery
2008 Welch Allyn Inc.
4341 State Street Road
PO Box 220, Skaneateles Falls, NY 13153-0220 USA
800 535 6663, 315 685 4560
www.welchallyn.com
©
CP100 & CP200
2
6.
Carefully tuck in all cables. Place the battery into the battery compartment.
Fold the strap so that it will not stick out when you replace the battery door.
See Figure 4.
7.
Replace the battery door and screws. Return the electrocardiograph to its
normal, upright position.
The electrocardiograph turns on automatically and displays a prompt to enter
date and time.
8.
Enter the date and time.
9.
Verify the local configuration settings. See “Reviewing the Device
Configuration Settings” in the electrocardiograph DFU for further information.
10. Keep the electrocardiograph plugged in for at least 12 hours before using
it on battery power.
Figure 3. Plugging In the Battery Fuse
Figure 4. Folding the Battery Strap
“Z” wire
The electrocardiograph is ready to
use after the battery is fully charged.