Midmark 354 Operation Manual
Read Online
Read the Midmark 354 Operation Manual online using the built-in PDF viewer.
Text of the manual
354 Lighting System
Owner’s Product Identification
(information that you will need to provide for servicing - key information is highlighted)
Date of Purchase
Serial Number
Name of Owner / Facility / Department
Model Number
Name of Authorized Dealer
Telephone # of Authorized Dealer
Address of Authorized Dealer
MODEL AND SERIAL
NUMBER LOCATION
Scope and Purpose of This Manual................................................................................2
Intended Use of Product.................................................................................................2
Safety Instructions ..........................................................................................................2
Explanation of Safety Symbols and Notes......................................................................2
Transportation and Storage Conditions ..........................................................................3
DESCRIPTION........................................................................................................................3
Introduction.....................................................................................................................3
Features .........................................................................................................................4
COMPONENTS OVERVIEW ..................................................................................................5
CONTROLS & INDICATORS .................................................................................................6
OPERATION ...........................................................................................................................7
Electromagnetic Interference..........................................................................................7
Operating Lighthead Assembly.......................................................................................7
OPERATOR MAINTENANCE ................................................................................................8
Preventive Maintenance Schedule .................................................................................8
Troubleshooting Guide ...................................................................................................8
Bulb Replacement Procedure.......................................................................................10
Fuse Replacement Procedure......................................................................................12
Ball Pivot Tension Adjustment Procedure.....................................................................13
Cleaning and Disinfecting.............................................................................................14
CALLING FOR SERVICE.....................................................................................................15
SPECIFICATIONS ................................................................................................................15
LIMITED WARRANTY..........................................................................................................16
Scope and Purpose of This Manual
This manual covers complete instructions for the operation of the 354 Lighting
System and is intended to be used by medical personnel responsible for operat-
ing the 354 Lighting System during medical procedures or performing operator
level maintenance. The installation manual is a separate document.
Intended Use of Product
This product is intended for use in all medical environments where illumination
is required for external examinations and procedures.
Safety Instructions
The primary concern of Midmark is that this equipment is operated and main-
tained with the safety of the patient and staff in mind. To assure safer and more
reliable operation:
•
Read and understand this manual before attempting to install or operate the
ceiling light system.
•
Assure that appropriate personnel are informed on the contents of this man-
ual; this is the responsibility of the purchaser.
•
Assure that this manual is located near the ceiling light system.
Explanation of Safety Symbols and Notes
DANGER
Indicates an imminently hazardous situation which, if not
avoided, will result in death or serious injury. The DANGER
symbol is limited to the most extreme situations.
WARNING
Indicates a potentially hazardous situation which, if not avoid-
ed, could result in death or serious injury.
CAUTION
Indicates a potentially hazardous situation which, if not avoided,
may result in minor or moderate injury. It may also be used to alert
against unsafe practices.
Important
Information
Transportation and Storage Conditions
•
Ambient Temperature Range:..... -22°F to +140°F (-30°C to +60°C)
•
Relative Humidity........................ 10% to 90% (non-condensing)
•
Atmospheric Pressure ................ 0.5 bar to 1.06 bars (500hPa to 1060hPa)
This product contains glass, so it should be transported and stored with care to
limit vibration and shocks.
DESCRIPTION
Introduction
Lighthead
The 354 lighthead assembly is a compact spotlight. The spotlight design pro-
vides excellent control over the diameter of the lighted area. Through the move-
ment of two levers, the user has complete control over the light beam diameter
between 3 in. (7.6 cm) to 10 in. (25.4 cm) and over the brightness of the lighted
area. The peak illumination is 4,000 fc (43,000 lux) at a distance of 36 in. (91.4
cm). The plastic handle can be easily removed for sterilization or it accepts a
Devon EZ Handle™ without requiring an awkward adapter. The arm assembly
has been precisely designed, assembled, and balanced so that the lighthead
EQUIPMENT ALERT
Indicates an imminently or potentially hazardous situation which, if
not avoided, will or may result in serious, moderate, or minor
equipment damage.
NOTE
Amplifies an operating procedure, practice, or condition.
Indicates that the unit is rated:
Type B, Applied Part.
Indicates that the operator’s man-
ual should be consulted for impor-
tant information.
Indicates the presence of a dan-
gerous voltage / shock hazard.
Indicates a fuse rating specifi-
cation
Indicates a protective earth
ground.
Indicates that the product is fragile;
do not handle roughly.
Indicates the proper shipping ori-
entation for the product.
Indicates that the product must be
kept dry.
Indicates a hot surface.
V
A
Description
can be positioned with minimal force and no drifting will occur. In addition, mul-
tiple pivot points make the positioning of the lighthead easy and flexible.
Features
The Model 354 lighting system . . .
•
has a power supply with three input taps, allowing the voltage output to the
light bulb to be adjusted according to the particular input voltage available at
a facility. This prevents premature failure of the light bulb and lighthead com-
ponents as well as unsatisfactory performance.
•
has a peak illumination of 4,000 fc (43,000 lux) at 36 in. (91.4 cm), but can
be adjusted to a lower intensity (brightness) if desired.
•
provides a round beam of light which can be adjusted from a diameter of 3
in. to 10 in. (7.6 to 25.4 cm) - based on a distance of 36 in. (91.4 cm) from
surface being lit.
•
has two joint pivots with 580° of rotation, one pivot joint with 540° of rotation,
and one pivot joint with 180° of rotation making positioning of the lighthead
easy and flexible.
•
has an arm reach of 48 in. (122 cm) (from centerline of down tube to center
of lighthead).
•
is precisely balanced so that the lighthead can be positioned with minimal
force and no drifting will occur.
•
has a plastic handle which can be easily removed for sterilization or it
accepts a Devon EZ Handle™.
Description
The illustration below shows the location of the light system’s major compo-
nents and the chart below provides their descriptive name.
DESCRIPTION OF COMPONENTS
1. Down Tube Assembly
6. Lamp Tube Assembly
2. Ceiling Plate
7. Lighthead Assembly
3. Junction Box (includes trans-
former and fuse)
8. Positioning Handle (sterilizable)
4. Cross Tube Assembly
9. Ball Pivot Joint
5. Ballast Assembly
Description
Components
Overview
CONTROLS & INDICATORS
The illustration below shows the location of the light system’s controls and indi-
cators and the chart below describes their function; the on / standby control for
the light system is located on the underside of the cross tube assembly and a
replaceable fuse is located on the underside of the ceiling plate, under the ceil-
ing cover (see fuse replacement procedure later in this manual).
ON / STANDBY
CONTROL
ON
STANDBY
Controls &
Indicators
This product is designed and built to minimize electromagnetic interference with
other devices. However, if interference is noticed between another device and
this product, remove the interfering device from the room or plug this product in-
to an isolated circuit.
Operating Lighthead Assembly
Turn the lighthead on by switching the ON / STANDBY switch (I /
) to ON “I”
(See previous illustration). To adjust the position of the lighthead, grasp the
sterile handle (3) and rotate the arm assemblies and lighthead as necessary to
obtain correct illumination on the patient (See previous illustration for amount of
rotation for an axis). To turn off the lighthead, switch the ON / STANDBY (I /
) to STANDBY “
“.
Ref.
Control
Function
1
on / standby switch (indi-
cated by international
symbol for on / standby:
I /
turns the light system on or off.
2
fuse holder
can be removed and inspected to determine if
the fuse has been blown, indicating a current
surge or problem with the light.
3
sterilizable positioning
handle
allows sterile personnel to move the lighthead
without compromising the sterility of their
hands.
4
brightness control lever
used to adjust the brightness of the light
beam.
5
beam size control lever
used to adjust the size (diameter) of the light
beam.
EQUIPMENT ALERT
Do not obstruct the airflow of the lighthead or damage to lighthead
could result. If the 354 light system malfunctions, immediately
turn the ON / STANDBY switch to STANDBY “
“.
Controls &
Indicators
Operation
Preventive Maintenance Schedule
The following preventive maintenance schedule should be followed. If a prob-
lem is detected, refer to the troubleshooting guide in this manual.
Troubleshooting Guide
If there is a malfunction with the light system, use the troubleshooting guide to
correct the problem. Refer to the illustrations in the Components Overview and
Controls & Indicators sections of this manual as necessary.
FREQUENCY
ACTIVITY
semiannually
Inspect rotation of lighthead. Make sure the lighthead rotates
freely, without noise, and has 180° of rotation at lamp tube con-
nection. Make sure the lighthead stays positioned at any point
(without drifting) with respect to the lamp tube.
semiannually
Inspect the rotation of the lamp tube, cross tube, and down tube.
Make sure the arm assemblies rotate freely and without noise.
Make sure the arm assemblies don’t drift at any point in their
range of motion. The lamp tube should have 540° of rotation at
cross tube connection. The cross tube should have 580° of rota-
tion at down tube connection and should also have a vertical
range of motion from -35° to +35°. The down tube should have
580° of rotation at ceiling mount connection.
semiannually
Turn light ON “I”. Rotate lighthead, lamp tube, cross tube, and
down tube while observing lighthead. The lighthead should not
intermittently flicker or stop illuminating while joints are being ro-
tated.
semiannually
Check for excessive joint rotation (more than 540° or 580° of ro-
tation). The physical hard stops for an axis may be broken.
semiannually
Inspect inside of lighthead for indications of broken material or
other signs of damage. If damage to any part of lighthead is
evident, do not continue to use lighthead.
semiannually
Turn light ON “I”. Verify that fan is operating. If fan is not operat-
ing, do not continue to use lighthead. Operating lighthead with-
out a fan in operation may cause lenses to melt.
PROBLEM
POSSIBLE CAUSE
SOLUTION
no light from lighthead
lighthead is on STAND-
BY “
“.
turn ON / STANDBY switch
to ON “I”
Operator
Maintenance
no light from lighthead,
but ON / STANDBY switch
is ON “I”
bulb has burned out
refer to bulb replacement
procedure in this manual
fuse in junction box is
blown
refer to the fuse replace-
ment procedure in this man-
ual
no power to the junction
box
call your building mainte-
nance electrician to check
for facility power to the junc-
tion box
transformer is malfunc-
tioning
call an authorized service
technician
no light from lighthead al-
though the bulb and fuse
were checked; lighthead
flashes intermittently
when lighthead or arms
are moved
circuit or wiring problem
within arm assemblies,
junction box, or other
part of lighting system
call an authorized service
technician
down tube, lamp tube, or
lighthead does not rotate
freely or drifts when re-
leased in desired position
brakes need adjustment
or improper installation of
ceiling plate
call an authorized service
technician
cross tube rotates (at ball
pivot joint) too stiffly or
drifts when released in de-
sired position
tension needs to be ad-
justed or cross tube
counterbalance needs to
be adjusted
refer to tension adjustment
procedure in this manual. If
proper tension cannot be
achieved, call an authorized
service technician
Operator
Maintenance
Bulb Replacement Procedure
1. If light is on, turn the ON / STANDBY switch to STANDBY “
“. Move light-
head away from exam / surgical site. Lower the lighthead to gain access for
bulb removal.
2. Using a phillips head screwdriver, loosen one screw (1) and then slide bulb
carriage (2) out of outer housing (3).
3. Remove the old bulb (4) by first pulling the bulb ejector lever (5) to the rear
and then gently pulling upward on the bulb with the other hand.
4. Push the bulb ejector lever (5) forward.
CAUTION
Turn the ON / STANDBY switch to STANDBY “
“ before replacing
bulb. Otherwise, electrical shock or hand burns could result.
Move the lighthead as far as possible from the exam / surgical site to prevent
contaminants from falling onto the exam / surgical site. Do not try to remove
the bulb until the unit is allowed to cool. Otherwise, burns to the hands and
fingers could result.
NOTE
Screw (1) is a captive screw. Only loosen captive screw; do not try to remove
it.
The phillips head screw (1) is different from the other two screws on the top
cap. Loosen this screw only to slide out the bulb carriage assembly (2); the
two other screws secure the top cap to the fan mounting plate.
WARNING
The maximum allowable bulb wattage which can be used in
this light is 150 Watts. There is a risk of fire if the 150 Watt
limit is exceeded. Use Midmark Part Number: 015-1212-00.
NOTE
Halogen bulbs are sensitive to body oils. Be sure not to touch the inside sur-
face of the reflector or glass capsule of the bulb during relamping or cleaning.
Body oils create a hot spot on the bulb and may cause the bulb to burn out
prematurely. If these portions of the bulb/reflector are handled, wipe with a
clean, soft, lint free cloth. Wipe with alcohol and pat dry.
Operator
Maintenance
5. Grasp the new bulb (4) and insert the bulb into the bulb socket (6). Push the
bulb all the way down until the base of the bulb is firmly seated against the
bulb socket.
6. Slide bulb carriage assembly (2) into outer housing (3) and secure by tight-
ening screw (1).
2
3
4
5
6
MA457700
1
Operator
Maintenance
Fuse Replacement Procedure
1. Turn the ON / STANDBY switch to STANDBY “
“.
2. Remove four screws (1) and lower ceiling cover (2) from ceiling plate (3).
3. Simultaneously press in on fuse cap (4) with a slotted screwdriver and rotate
it 1/4 turn in the counterclockwise direction; then pull fuse cap from fuse
holder (5).
4. Pull fuse (6) out of fuse cap (4).
5. Inspect fuse (6) for any indication that it has been blown (opened); i.e. burnt
look, discolored, fuse cord melted through, etc. Discard fuse.
6. Obtain a new fuse (6) of the same voltage rating, amperage rating, and type.
Refer to the Specifications section in this manual.
7. Insert one end of the new fuse (6) into the fuse cap (4).
8. Insert fuse cap (4) into fuse holder (5) and rotate it 1/4 turn in the clockwise
direction to secure it.
9. Position ceiling cover (2) on ceiling plate (3) and secure with four screws (1).
Operator
Maintenance
Ball Pivot Tension Adjustment Procedure
1. Remove screw (1) and slide ball pivot sleeve (2) up out of way.
2. Rotate cross tube (3) until adjustment hole (4) appears in the adjustment
window opening.
3. Insert a phillips screwdriver into adjustment hole (4). Then, using the screw-
driver, rotate ball pivot cam (5) to the desired tension setting.
4. Remove the screwdriver and move the cross tube (3) about the ball pivot
joint in a circular motion and up and down motion to ensure the setting is the
one desired. Move the cross tube (3) to a horizontal position and release it.
The cross tube should not drift in any direction. If it does, a higher tension
setting is required or the cross tube counterbalance needs adjusted (see
counterbalance adjustment in the Installation or Service and Parts Manual).
Repeat steps 3 and 4 until the desired tension adjustment is achieved with
no drifting of the cross tube.
5. Slide ball pivot sleeve (2) down into position and secure with screw (1).
NOTE
There are three settings which the ball pivot cam (5) can be set for: light ten-
sion, medium tension, or stiff tension. These settings may be changed ac-
cording to the operator’s preference.
ASSEMBLY /
DISASSEMBLY
POSITION
LIGHT TENSION
POSITION
MEDIUM TENSION
POSITION
STIFF TENSION
POSITION
Operator
Maintenance
Cleaning and Disinfecting
External Cleaning Procedures
For general cleaning, use a mild detergent and water solution. Wring excess
solution from sponge or cloth before wiping.
According to your facility’s procedure:
1. Clean lens with an antistatic acrylic cleaning solution using a soft, clean
cloth. Do not use alcohol or abrasive compounds on lens.
2. Wipe external surface of arm assemblies and lighthead with a mild detergent
and water solution.
3. Rinse all external surfaces with a soft cloth and clear water, wringing excess
from cloth before wiping.
4. Wipe all external surfaces dry.
External Disinfecting Procedures
According to your facility’s procedure:
1. Use only quaternary disinfectants to disinfect light. Staining, pitting, discol-
oration, or softening could occur if phenolic, iodophor, or glutaraldehyde-
based disinfectant is used on plastic surfaces of the lighthead. Also, use of
alcohol or aerosol spray cleaner / disinfectant containing substantial
amounts of alcohol in the formula can damage the lens.
2. Wring excess solution from cloth.
3. Using soft cloth, wipe all external surfaces of arm assemblies and lighthead.
EQUIPMENT ALERT
When cleaning or disinfecting the light, remove power from the
light, allow optical unit to cool. Clean EXTERNAL SURFACES
ONLY (arm assemblies and lighthead). Prevent fluids from leaking into interi-
or or onto electrical contacts. DO NOT ATTEMPT to clean or disinfect interi-
or; instead call an authorized dealer or service technician.
EQUIPMENT ALERT
Use only quaternary disinfectants to disinfect light. Staining, pit-
ting, discoloration, or softening could occur if phenolic, iodophor,
or glutaraldehyde-based disinfectant is used on plastic surfaces of the light-
head. Also, use of alcohol or aerosol spray cleaner / disinfectant containing
substantial amounts of alcohol in the formula can damage the lens.
Operator
Maintenance
4. Do not rinse or dry external surfaces. Allow disinfectant solution to air dry.
Handle Sterilization
•
Use only steam sterilization on the handle.
•
Follow steam sterilization instructions of sterilizer manufacturer.
CALLING FOR SERVICE
If you are having a problem or have a question, refer to the inside front cover of
this manual and call your dealer. Make sure that you have the information that
is highlighted on the inside front cover of this manual available. If you can’t re-
solve your question or problem with your dealer, call the following number:
1-800-Midmark (1-800-643-6275) or 937-526-3662
8:00 a.m until 5:00 p.m. (Eastern Standard Time in U.S.)
Monday through Friday, except for standard U.S. holidays.
SPECIFICATIONS
Model 354 Ceiling Light
Beam diameter @ 36 in.: .............. Variable from 3 to 10 in. (7.6 cm to 25.4 cm)
(91.4 cm)
Bulb:............................................... 21.0 VAC, 150 W halogen bulb
Color temperature:........................ 4,200K
Focal length:.................................. 36 in. (91.4 cm)
Illumination @ 36 in. (91.4 cm): ... 4000 fc (43,000 lux)
Electrical requirement: ................. 120 VAC, 60 HZ,
1.5 amps, single phase
Reach of arm assemblies:............ 49.5 in. (126 cm) maximum from centerline
of down tube to center of lighthead.
Rotation of lamp tube:.................. 180° rotation at lamp tube connection
Rotation of lamp tube:.................. 540° rotation at cross tube connection
Rotation of cross tube:................. 580° rotation at down tube connection
Vertical range of cross tube:........ -35° to +35° vertical movement
Rotation of down tube:................. 580° rotation at ceiling plate connection
Weight of 8 ft. (2.44 m) single
light assembly:.............................. 59 lbs (26.8 kg)
Weight of 9 ft. (2.74 m) single
light assembly:.............................. 60 lbs (27.2 kg)
Weight of 8 ft. (2.44 m) dual
light assembly:.............................. 105 lbs (47.6 kg)
Weight of 9 ft. (2.74 m) dual
light assembly:.............................. 107 lbs (48.5 kg)
Specifications
Calling For
Service
Operator
Maintenance
classified by Underwriters
Laboratories Inc. with respect to electric
shock, fire, and mechanical hazards only in
accordance with UL2601-1 and CAN/CSA
C22.2, No.601.1.
ISO-9001 Certified
Classifications:..............................Class 1, Type B applied part, Ordinary
Equipment, Continuous Operation
Fuse Rating:...................................1.6 amp, 250 VAC, 5mm x 20mm, Type T,
low breaking capacity, IEC 127-2/3
Equipment not suitable for use in the presence of a flammable anesthetic
mixture with air, oxygen, or nitrous oxide.
LIMITED WARRANTY
SCOPE OF WARRANTY
Midmark Corporation (“Midmark”) warrants to the original purchaser its new Alter-
nate Care products and components (except for components not warranted under
“Exclusions”) manufactured by Midmark to be free from defects in material and
workmanship under normal use and service. Midmark’s obligation under this war-
ranty is limited to the repair or replacement, at Midmark’s option, of the parts or the
products the defects of which are reported to Midmark within the applicable warranty
period and which, upon examination by Midmark, prove to be defective.
APPLICABLE WARRANTY PERIOD
The applicable warranty period, measured from the date of delivery to the original
user, shall be one (1) year for all warranted products and components.
EXCLUSIONS
This warranty does not cover and Midmark shall not be liable for the following: (1)
repairs and replacements because of misuse, abuse, negligence, alteration, acci-
dent, freight damage, or tampering; (2) products which are not installed, used, and
properly cleaned as required in the Midmark “Installation” and or “Installation / Oper-
ation Manual for this applicable product. (3) products considered to be of a consum-
able nature; (4) accessories or parts not manufactured by Midmark; (5) charges by
anyone for adjustments, repairs, replacement parts, installation, or other work per-
formed upon or in connection with such products which is not expressly authorized
in writing in advance by Midmark.
EXCLUSIVE REMEDY
Midmark’s only obligation under this warranty is the repair or replacement of defec-
tive parts. Midmark shall not be liable for any direct, special, indirect, incidental,
exemplary, or consequential damages or delay, including, but not limited to, dam-
ages for loss of profits or loss of use.
Limited
Warranty
No person or firm is authorized to create for Midmark any other obligation or liability
in connection with the products.
THIS WARRANTY IS MIDMARK’S ONLY WARRANTY AND IS IN LIEU OF ALL
OTHER WARRANTIES, EXPRESS OR IMPLIED. MIDMARK MAKES NO
IMPLIED WARRANTIES OF ANY KIND INCLUDING ANY WARRANTIES OF
MERCHANTABILITY OR FITNESS FOR ANY PARTICULAR PURPOSE. THIS
WARRANTY IS LIMITED TO THE REPAIR OR REPLACEMENT OF DEFECTIVE
PARTS.
SF-1487 REV. A1
Limited
Warranty
© Midmark Corporation - 1999
003-1071-00 Rev. B (7/00)
Midmark Corporation, Versailles, Ohio 45380 U.S.A.
937-526-3662
FAX: 937-526-5542