354 Lighting System

Owner’s Product Identification

(information that you will need to provide for servicing - key information is highlighted) Date of Purchase Serial Number Name of Owner / Facility / Department Model Number Name of Authorized Dealer Telephone # of Authorized Dealer Address of Authorized Dealer MODEL AND SERIAL NUMBER LOCATION

Scope and Purpose of This Manual................................................................................2

Intended Use of Product.................................................................................................2 Safety Instructions ..........................................................................................................2 Explanation of Safety Symbols and Notes......................................................................2 Transportation and Storage Conditions ..........................................................................3 DESCRIPTION........................................................................................................................3 Introduction.....................................................................................................................3 Features .........................................................................................................................4 COMPONENTS OVERVIEW ..................................................................................................5 CONTROLS & INDICATORS .................................................................................................6 OPERATION ...........................................................................................................................7 Electromagnetic Interference..........................................................................................7 Operating Lighthead Assembly.......................................................................................7 OPERATOR MAINTENANCE ................................................................................................8 Preventive Maintenance Schedule .................................................................................8 Troubleshooting Guide ...................................................................................................8 Bulb Replacement Procedure.......................................................................................10 Fuse Replacement Procedure......................................................................................12 Ball Pivot Tension Adjustment Procedure.....................................................................13 Cleaning and Disinfecting.............................................................................................14 CALLING FOR SERVICE.....................................................................................................15 SPECIFICATIONS ................................................................................................................15 LIMITED WARRANTY..........................................................................................................16

Scope and Purpose of This Manual

This manual covers complete instructions for the operation of the 354 Lighting System and is intended to be used by medical personnel responsible for operat- ing the 354 Lighting System during medical procedures or performing operator level maintenance. The installation manual is a separate document. Intended Use of Product This product is intended for use in all medical environments where illumination is required for external examinations and procedures. Safety Instructions The primary concern of Midmark is that this equipment is operated and main- tained with the safety of the patient and staff in mind. To assure safer and more reliable operation: • Read and understand this manual before attempting to install or operate the ceiling light system. • Assure that appropriate personnel are informed on the contents of this man- ual; this is the responsibility of the purchaser. • Assure that this manual is located near the ceiling light system. Explanation of Safety Symbols and Notes DANGER Indicates an imminently hazardous situation which, if not avoided, will result in death or serious injury. The DANGER symbol is limited to the most extreme situations. WARNING Indicates a potentially hazardous situation which, if not avoid- ed, could result in death or serious injury. CAUTION Indicates a potentially hazardous situation which, if not avoided, may result in minor or moderate injury. It may also be used to alert against unsafe practices. Important Information

Transportation and Storage Conditions

• Ambient Temperature Range:..... -22°F to +140°F (-30°C to +60°C) • Relative Humidity........................ 10% to 90% (non-condensing) • Atmospheric Pressure ................ 0.5 bar to 1.06 bars (500hPa to 1060hPa) This product contains glass, so it should be transported and stored with care to limit vibration and shocks. DESCRIPTION Introduction Lighthead The 354 lighthead assembly is a compact spotlight. The spotlight design pro- vides excellent control over the diameter of the lighted area. Through the move- ment of two levers, the user has complete control over the light beam diameter between 3 in. (7.6 cm) to 10 in. (25.4 cm) and over the brightness of the lighted area. The peak illumination is 4,000 fc (43,000 lux) at a distance of 36 in. (91.4 cm). The plastic handle can be easily removed for sterilization or it accepts a Devon EZ Handle™ without requiring an awkward adapter. The arm assembly has been precisely designed, assembled, and balanced so that the lighthead EQUIPMENT ALERT Indicates an imminently or potentially hazardous situation which, if not avoided, will or may result in serious, moderate, or minor equipment damage. NOTE Amplifies an operating procedure, practice, or condition. Indicates that the unit is rated: Type B, Applied Part. Indicates that the operator’s man- ual should be consulted for impor- tant information. Indicates the presence of a dan- gerous voltage / shock hazard. Indicates a fuse rating specifi- cation Indicates a protective earth ground. Indicates that the product is fragile; do not handle roughly. Indicates the proper shipping ori- entation for the product. Indicates that the product must be kept dry. Indicates a hot surface. V A Description

can be positioned with minimal force and no drifting will occur. In addition, mul-

tiple pivot points make the positioning of the lighthead easy and flexible. Features The Model 354 lighting system . . . • has a power supply with three input taps, allowing the voltage output to the light bulb to be adjusted according to the particular input voltage available at a facility. This prevents premature failure of the light bulb and lighthead com- ponents as well as unsatisfactory performance. • has a peak illumination of 4,000 fc (43,000 lux) at 36 in. (91.4 cm), but can be adjusted to a lower intensity (brightness) if desired. • provides a round beam of light which can be adjusted from a diameter of 3 in. to 10 in. (7.6 to 25.4 cm) - based on a distance of 36 in. (91.4 cm) from surface being lit. • has two joint pivots with 580° of rotation, one pivot joint with 540° of rotation, and one pivot joint with 180° of rotation making positioning of the lighthead easy and flexible. • has an arm reach of 48 in. (122 cm) (from centerline of down tube to center of lighthead). • is precisely balanced so that the lighthead can be positioned with minimal force and no drifting will occur. • has a plastic handle which can be easily removed for sterilization or it accepts a Devon EZ Handle™. Description

The illustration below shows the location of the light system’s major compo-

nents and the chart below provides their descriptive name. DESCRIPTION OF COMPONENTS 1. Down Tube Assembly 6. Lamp Tube Assembly 2. Ceiling Plate 7. Lighthead Assembly 3. Junction Box (includes trans- former and fuse) 8. Positioning Handle (sterilizable) 4. Cross Tube Assembly 9. Ball Pivot Joint 5. Ballast Assembly Description Components Overview

CONTROLS & INDICATORS

The illustration below shows the location of the light system’s controls and indi- cators and the chart below describes their function; the on / standby control for the light system is located on the underside of the cross tube assembly and a replaceable fuse is located on the underside of the ceiling plate, under the ceil- ing cover (see fuse replacement procedure later in this manual). ON / STANDBY CONTROL ON STANDBY Controls & Indicators

This product is designed and built to minimize electromagnetic interference with

other devices. However, if interference is noticed between another device and this product, remove the interfering device from the room or plug this product in- to an isolated circuit. Operating Lighthead Assembly Turn the lighthead on by switching the ON / STANDBY switch (I / ) to ON “I” (See previous illustration). To adjust the position of the lighthead, grasp the sterile handle (3) and rotate the arm assemblies and lighthead as necessary to obtain correct illumination on the patient (See previous illustration for amount of rotation for an axis). To turn off the lighthead, switch the ON / STANDBY (I / ) to STANDBY “ “. Ref. Control Function 1 on / standby switch (indi- cated by international symbol for on / standby: I / turns the light system on or off. 2 fuse holder can be removed and inspected to determine if the fuse has been blown, indicating a current surge or problem with the light. 3 sterilizable positioning handle allows sterile personnel to move the lighthead without compromising the sterility of their hands. 4 brightness control lever used to adjust the brightness of the light beam. 5 beam size control lever used to adjust the size (diameter) of the light beam. EQUIPMENT ALERT Do not obstruct the airflow of the lighthead or damage to lighthead could result. If the 354 light system malfunctions, immediately turn the ON / STANDBY switch to STANDBY “ “. Controls & Indicators Operation

Preventive Maintenance Schedule

The following preventive maintenance schedule should be followed. If a prob- lem is detected, refer to the troubleshooting guide in this manual. Troubleshooting Guide If there is a malfunction with the light system, use the troubleshooting guide to correct the problem. Refer to the illustrations in the Components Overview and Controls & Indicators sections of this manual as necessary. FREQUENCY ACTIVITY semiannually Inspect rotation of lighthead. Make sure the lighthead rotates freely, without noise, and has 180° of rotation at lamp tube con- nection. Make sure the lighthead stays positioned at any point (without drifting) with respect to the lamp tube. semiannually Inspect the rotation of the lamp tube, cross tube, and down tube. Make sure the arm assemblies rotate freely and without noise. Make sure the arm assemblies don’t drift at any point in their range of motion. The lamp tube should have 540° of rotation at cross tube connection. The cross tube should have 580° of rota- tion at down tube connection and should also have a vertical range of motion from -35° to +35°. The down tube should have 580° of rotation at ceiling mount connection. semiannually Turn light ON “I”. Rotate lighthead, lamp tube, cross tube, and down tube while observing lighthead. The lighthead should not intermittently flicker or stop illuminating while joints are being ro- tated. semiannually Check for excessive joint rotation (more than 540° or 580° of ro- tation). The physical hard stops for an axis may be broken. semiannually Inspect inside of lighthead for indications of broken material or other signs of damage. If damage to any part of lighthead is evident, do not continue to use lighthead. semiannually Turn light ON “I”. Verify that fan is operating. If fan is not operat- ing, do not continue to use lighthead. Operating lighthead with- out a fan in operation may cause lenses to melt. PROBLEM POSSIBLE CAUSE SOLUTION no light from lighthead lighthead is on STAND- BY “ “. turn ON / STANDBY switch to ON “I” Operator Maintenance

no light from lighthead,

but ON / STANDBY switch is ON “I” bulb has burned out refer to bulb replacement procedure in this manual fuse in junction box is blown refer to the fuse replace- ment procedure in this man- ual no power to the junction box call your building mainte- nance electrician to check for facility power to the junc- tion box transformer is malfunc- tioning call an authorized service technician no light from lighthead al- though the bulb and fuse were checked; lighthead flashes intermittently when lighthead or arms are moved circuit or wiring problem within arm assemblies, junction box, or other part of lighting system call an authorized service technician down tube, lamp tube, or lighthead does not rotate freely or drifts when re- leased in desired position brakes need adjustment or improper installation of ceiling plate call an authorized service technician cross tube rotates (at ball pivot joint) too stiffly or drifts when released in de- sired position tension needs to be ad- justed or cross tube counterbalance needs to be adjusted refer to tension adjustment procedure in this manual. If proper tension cannot be achieved, call an authorized service technician Operator Maintenance

Bulb Replacement Procedure

1. If light is on, turn the ON / STANDBY switch to STANDBY “ “. Move light- head away from exam / surgical site. Lower the lighthead to gain access for bulb removal. 2. Using a phillips head screwdriver, loosen one screw (1) and then slide bulb carriage (2) out of outer housing (3). 3. Remove the old bulb (4) by first pulling the bulb ejector lever (5) to the rear and then gently pulling upward on the bulb with the other hand. 4. Push the bulb ejector lever (5) forward. CAUTION Turn the ON / STANDBY switch to STANDBY “ “ before replacing bulb. Otherwise, electrical shock or hand burns could result. Move the lighthead as far as possible from the exam / surgical site to prevent contaminants from falling onto the exam / surgical site. Do not try to remove the bulb until the unit is allowed to cool. Otherwise, burns to the hands and fingers could result. NOTE Screw (1) is a captive screw. Only loosen captive screw; do not try to remove it. The phillips head screw (1) is different from the other two screws on the top cap. Loosen this screw only to slide out the bulb carriage assembly (2); the two other screws secure the top cap to the fan mounting plate. WARNING The maximum allowable bulb wattage which can be used in this light is 150 Watts. There is a risk of fire if the 150 Watt limit is exceeded. Use Midmark Part Number: 015-1212-00. NOTE Halogen bulbs are sensitive to body oils. Be sure not to touch the inside sur- face of the reflector or glass capsule of the bulb during relamping or cleaning. Body oils create a hot spot on the bulb and may cause the bulb to burn out prematurely. If these portions of the bulb/reflector are handled, wipe with a clean, soft, lint free cloth. Wipe with alcohol and pat dry. Operator Maintenance

5. Grasp the new bulb (4) and insert the bulb into the bulb socket (6). Push the

bulb all the way down until the base of the bulb is firmly seated against the bulb socket. 6. Slide bulb carriage assembly (2) into outer housing (3) and secure by tight- ening screw (1). 2 3 4 5 6 MA457700 1 Operator Maintenance

Fuse Replacement Procedure

1. Turn the ON / STANDBY switch to STANDBY “ “. 2. Remove four screws (1) and lower ceiling cover (2) from ceiling plate (3). 3. Simultaneously press in on fuse cap (4) with a slotted screwdriver and rotate it 1/4 turn in the counterclockwise direction; then pull fuse cap from fuse holder (5). 4. Pull fuse (6) out of fuse cap (4). 5. Inspect fuse (6) for any indication that it has been blown (opened); i.e. burnt look, discolored, fuse cord melted through, etc. Discard fuse. 6. Obtain a new fuse (6) of the same voltage rating, amperage rating, and type. Refer to the Specifications section in this manual. 7. Insert one end of the new fuse (6) into the fuse cap (4). 8. Insert fuse cap (4) into fuse holder (5) and rotate it 1/4 turn in the clockwise direction to secure it. 9. Position ceiling cover (2) on ceiling plate (3) and secure with four screws (1). Operator Maintenance

Ball Pivot Tension Adjustment Procedure

1. Remove screw (1) and slide ball pivot sleeve (2) up out of way. 2. Rotate cross tube (3) until adjustment hole (4) appears in the adjustment window opening. 3. Insert a phillips screwdriver into adjustment hole (4). Then, using the screw- driver, rotate ball pivot cam (5) to the desired tension setting. 4. Remove the screwdriver and move the cross tube (3) about the ball pivot joint in a circular motion and up and down motion to ensure the setting is the one desired. Move the cross tube (3) to a horizontal position and release it. The cross tube should not drift in any direction. If it does, a higher tension setting is required or the cross tube counterbalance needs adjusted (see counterbalance adjustment in the Installation or Service and Parts Manual). Repeat steps 3 and 4 until the desired tension adjustment is achieved with no drifting of the cross tube. 5. Slide ball pivot sleeve (2) down into position and secure with screw (1). NOTE There are three settings which the ball pivot cam (5) can be set for: light ten- sion, medium tension, or stiff tension. These settings may be changed ac- cording to the operator’s preference. ASSEMBLY / DISASSEMBLY POSITION LIGHT TENSION POSITION MEDIUM TENSION POSITION STIFF TENSION POSITION Operator Maintenance

Cleaning and Disinfecting

External Cleaning Procedures For general cleaning, use a mild detergent and water solution. Wring excess solution from sponge or cloth before wiping. According to your facility’s procedure: 1. Clean lens with an antistatic acrylic cleaning solution using a soft, clean cloth. Do not use alcohol or abrasive compounds on lens. 2. Wipe external surface of arm assemblies and lighthead with a mild detergent and water solution. 3. Rinse all external surfaces with a soft cloth and clear water, wringing excess from cloth before wiping. 4. Wipe all external surfaces dry. External Disinfecting Procedures According to your facility’s procedure: 1. Use only quaternary disinfectants to disinfect light. Staining, pitting, discol- oration, or softening could occur if phenolic, iodophor, or glutaraldehyde- based disinfectant is used on plastic surfaces of the lighthead. Also, use of alcohol or aerosol spray cleaner / disinfectant containing substantial amounts of alcohol in the formula can damage the lens. 2. Wring excess solution from cloth. 3. Using soft cloth, wipe all external surfaces of arm assemblies and lighthead. EQUIPMENT ALERT When cleaning or disinfecting the light, remove power from the light, allow optical unit to cool. Clean EXTERNAL SURFACES ONLY (arm assemblies and lighthead). Prevent fluids from leaking into interi- or or onto electrical contacts. DO NOT ATTEMPT to clean or disinfect interi- or; instead call an authorized dealer or service technician. EQUIPMENT ALERT Use only quaternary disinfectants to disinfect light. Staining, pit- ting, discoloration, or softening could occur if phenolic, iodophor, or glutaraldehyde-based disinfectant is used on plastic surfaces of the light- head. Also, use of alcohol or aerosol spray cleaner / disinfectant containing substantial amounts of alcohol in the formula can damage the lens. Operator Maintenance

4. Do not rinse or dry external surfaces. Allow disinfectant solution to air dry.

Handle Sterilization • Use only steam sterilization on the handle. • Follow steam sterilization instructions of sterilizer manufacturer. CALLING FOR SERVICE If you are having a problem or have a question, refer to the inside front cover of this manual and call your dealer. Make sure that you have the information that is highlighted on the inside front cover of this manual available. If you can’t re- solve your question or problem with your dealer, call the following number: 1-800-Midmark (1-800-643-6275) or 937-526-3662 8:00 a.m until 5:00 p.m. (Eastern Standard Time in U.S.) Monday through Friday, except for standard U.S. holidays. SPECIFICATIONS Model 354 Ceiling Light Beam diameter @ 36 in.: .............. Variable from 3 to 10 in. (7.6 cm to 25.4 cm) (91.4 cm) Bulb:............................................... 21.0 VAC, 150 W halogen bulb Color temperature:........................ 4,200K Focal length:.................................. 36 in. (91.4 cm) Illumination @ 36 in. (91.4 cm): ... 4000 fc (43,000 lux) Electrical requirement: ................. 120 VAC, 60 HZ, 1.5 amps, single phase Reach of arm assemblies:............ 49.5 in. (126 cm) maximum from centerline of down tube to center of lighthead. Rotation of lamp tube:.................. 180° rotation at lamp tube connection Rotation of lamp tube:.................. 540° rotation at cross tube connection Rotation of cross tube:................. 580° rotation at down tube connection Vertical range of cross tube:........ -35° to +35° vertical movement Rotation of down tube:................. 580° rotation at ceiling plate connection Weight of 8 ft. (2.44 m) single light assembly:.............................. 59 lbs (26.8 kg) Weight of 9 ft. (2.74 m) single light assembly:.............................. 60 lbs (27.2 kg) Weight of 8 ft. (2.44 m) dual light assembly:.............................. 105 lbs (47.6 kg) Weight of 9 ft. (2.74 m) dual light assembly:.............................. 107 lbs (48.5 kg) Specifications Calling For Service Operator Maintenance

classified by Underwriters

Laboratories Inc. with respect to electric shock, fire, and mechanical hazards only in accordance with UL2601-1 and CAN/CSA C22.2, No.601.1. ISO-9001 Certified Classifications:..............................Class 1, Type B applied part, Ordinary Equipment, Continuous Operation Fuse Rating:...................................1.6 amp, 250 VAC, 5mm x 20mm, Type T, low breaking capacity, IEC 127-2/3 Equipment not suitable for use in the presence of a flammable anesthetic mixture with air, oxygen, or nitrous oxide. LIMITED WARRANTY SCOPE OF WARRANTY Midmark Corporation (“Midmark”) warrants to the original purchaser its new Alter- nate Care products and components (except for components not warranted under “Exclusions”) manufactured by Midmark to be free from defects in material and workmanship under normal use and service. Midmark’s obligation under this war- ranty is limited to the repair or replacement, at Midmark’s option, of the parts or the products the defects of which are reported to Midmark within the applicable warranty period and which, upon examination by Midmark, prove to be defective. APPLICABLE WARRANTY PERIOD The applicable warranty period, measured from the date of delivery to the original user, shall be one (1) year for all warranted products and components. EXCLUSIONS This warranty does not cover and Midmark shall not be liable for the following: (1) repairs and replacements because of misuse, abuse, negligence, alteration, acci- dent, freight damage, or tampering; (2) products which are not installed, used, and properly cleaned as required in the Midmark “Installation” and or “Installation / Oper- ation Manual for this applicable product. (3) products considered to be of a consum- able nature; (4) accessories or parts not manufactured by Midmark; (5) charges by anyone for adjustments, repairs, replacement parts, installation, or other work per- formed upon or in connection with such products which is not expressly authorized in writing in advance by Midmark. EXCLUSIVE REMEDY Midmark’s only obligation under this warranty is the repair or replacement of defec- tive parts. Midmark shall not be liable for any direct, special, indirect, incidental, exemplary, or consequential damages or delay, including, but not limited to, dam- ages for loss of profits or loss of use. Limited Warranty

No person or firm is authorized to create for Midmark any other obligation or liability

in connection with the products. THIS WARRANTY IS MIDMARK’S ONLY WARRANTY AND IS IN LIEU OF ALL OTHER WARRANTIES, EXPRESS OR IMPLIED. MIDMARK MAKES NO IMPLIED WARRANTIES OF ANY KIND INCLUDING ANY WARRANTIES OF MERCHANTABILITY OR FITNESS FOR ANY PARTICULAR PURPOSE. THIS WARRANTY IS LIMITED TO THE REPAIR OR REPLACEMENT OF DEFECTIVE PARTS. SF-1487 REV. A1 Limited Warranty

© Midmark Corporation - 1999

003-1071-00 Rev. B (7/00) Midmark Corporation, Versailles, Ohio 45380 U.S.A. 937-526-3662 FAX: 937-526-5542