Installation and Operation

Manual 419 Power Examination Table 419 MA518602i

Owner’s Product Identification

(Information required when calling for service) Date of Purchase Serial Number Name of Owner / Facility / Department Model Number Name of Authorized Midmark Dealer Telephone # of Authorized Midmark Dealer Address of Authorized Midmark Dealer MA518702i 419 Model Number Serial Number

IMPORTANT INFORMATION ........................................................................................................ 2

Intended Use of Product........................................................................................................ 2 Patient Positioning................................................................................................................. 2 Authorized CE Representative .............................................................................................. 2 Safety Instructions................................................................................................................. 2 Transportation and Storage Conditions................................................................................. 2 Explanation of Safety Symbols and Notes ............................................................................ 3 INSTALLATION.............................................................................................................................. 4 Unpacking ............................................................................................................................. 4 Headrest Adjustment Tools ................................................................................................... 5 Electrical Requirements ........................................................................................................ 5 Leveling The Table ................................................................................................................ 5 Foot / Hand Control Installation............................................................................................. 6 Upholstery Installation........................................................................................................... 7 OPERATION.................................................................................................................................. 8 Standard Table Functions...................................................................................................... 9 Programming Table Positions (Programmable tables only)................................................. 10 Home Function (Programmable tables only)....................................................................... 12 Multiple Function Positioning............................................................................................... 13 Footrest Extension Lamp .................................................................................................... 13 Armrest Adjustment............................................................................................................. 14 Armrest Removal / Installation ............................................................................................ 14 Headrest Positioning / Removal / Orientation...................................................................... 15 Headrest Slide Adjustment.................................................................................................. 16 Headrest Pivot Tension Adjustment .................................................................................... 17 Accessory Rails................................................................................................................... 17 List of Authorized Accessories ............................................................................................ 18 OPERATOR MAINTENANCE ..................................................................................................... 20 Preventive Maintenance...................................................................................................... 20 Care of Table....................................................................................................................... 20 COMPONENTS OVERVIEW ....................................................................................................... 21 CONTROLS & INDICATORS ...................................................................................................... 22 CALLING FOR SERVICE............................................................................................................ 27 SPECIFICATIONS ....................................................................................................................... 27 RANGE OF MOTION................................................................................................................... 28 LIMITED WARRANTY................................................................................................................. 29

Intended Use of Product

This product is intended to be used as a chair / table to provide positioning and support of patients during general examinations and procedures (primarily fo- cusing around the shoulder, neck, and head area of the patient) conducted by qualified medical professionals. Patient Positioning The table is designed to accommodate the following examination positions: • Flat Table Position • Chair Position Authorized CE Representative Countries in the EEC should direct all questions, incidents, and complaints to Midmark’s Authorized CE representative listed below: Promotal 22, Rue De St. Denis 53500 Ernee, FRANCE Phone: 33-2-430-517-76 Fax: 33-2-430-569-10 Safety Instructions To assure safe and reliable operation: • Read and understand this manual before installing or operating the table. • Inform appropriate personnel of the contents of this manual • Store this manual near the table. Transportation and Storage Conditions • Ambient Temperature Range: ..... +41°F to 100°F (+5°C to +38°C) • Relative Humidity ........................ 10% to 90% (non-condensing) • Atmospheric Pressure................. 500hPa to 1060hPa (0.49atm to 1.05atm) Important Information

Explanation of Safety Symbols and Notes

DANGER Indicates an imminently hazardous situation which, if not avoided, will result in death or serious injury. The DANGER symbol is limited to the most extreme situations. WARNING Indicates a potentially hazardous situation which, if not avoid- ed, could result in death or serious injury. CAUTION Indicates a potentially hazardous situation which, if not avoided, may result in minor or moderate injury. It may also be used to alert against unsafe practices. EQUIPMENT ALERT Indicates an imminently or potentially hazardous situation which, if not avoided, will or may result in serious, moderate, or minor equipment damage. NOTE Amplifies an operating procedure, practice, or condition. Consult operator’s manual for important information Proper shipping orientation Fragile. Type B, Applied Part. Protective earth ground Duty cycle: (motor run time 15 seconds on, 5 minutes off). Keep dry. Minimum and maximum storage temperatures for the unit. Maximum stacking height of palleted units Fuse rating specification Dangerous voltage / shock hazard. 30°C 100°F 5°C 41°F 3 V A T

1. Remove packaging and inspect for shipping damage.

NOTE: Report any damage to shipping company and fill out a concealed damage report. 2. Cut cable ties securing headrest and footrest section. 3. Using 3/4” open end wrench or socket, remove two bolts (1, Figure 1) from underside of table base (one on each side). 4. Lift table at points (3 & 4) & remove from skid. Positon table as desired. WARNING Table weighs approximately 515.0 lbs (233.6 kgs). Use an assistant and proper lifting techniques when removing table from skid. Failure to do so could result in personal injury. EQUIPMENT ALERT Do not lift at points (2). These are unsupported areas. Failure to comply could result in damage to the table. 419 MA518802i 2 1 3 4 2 2 3 Figure 1. Installation

Headrest Adjustment Tools

The bag attached to the headrest assembly contains two allen wrenches [3/32” & 3/16”] to adjust the headrest tension. Remove this bag and store these tools with the Installation & Operation Manual. Electrical Requirements The electrical rating for the 115 V unit is 115 VAC, 50/60 Hz, 11 amps. The electrical rating for the 230 V unit is 230 VAC, 50/60 Hz, 5.5 amps. The three- pronged grounding plug on the table power cord must be plugged into a match- ing three-pronged, grounded, dedicated, correctly polarized receptacle. The ta- ble power cord provides power for all electrical functions of the table. Leveling The Table Adjust four leveling screws (1, Figure 2) [located under each corner of table] until table is level. WARNING Most exposed metal parts of the table are electrically ground- ed. When performing a cauterization or similar treatment, the patient must be insulated from the metal portions of the table by non- conductive material. Failure to do so could result in electrical shock or burns to patient. On 115 VAC units, use 115 VAC, 50/60 HZ alternating current only and on 230 VAC units, use 230 VAC, 50/60 HZ alternating current only. Fail- ure to do so could result in electrical shock to personnel and will result in damage to table. NOTE Grounding reliability can only be achieved if this equipment is connected to an equivalent receptacle marked Hospital Grade (NEMA 5-15R, HOSPITAL GRADE). Figure 2. Installation

Foot / Hand Control

Installation Foot Control 1. Insert foot control cord (1, Figure 3) into foot control receptacle (2). Hand Control (optional) 1. Insert hand control cord (3) into hand control receptacle (4). 2. Hang hand control on hanger (5) when not in use. NOTE The hand control hanger may be installed on opposite side of the table using the two predrilled holes behind the armrest. The foot control receptacle and hand control receptacle are interchangeable. 5 3 MIDMARK 1 2 4 3 P 4 MA524202i 2 1 Figure 3. Installation

1. Remove packaging and inspect for shipping damage.

NOTE: Report any damage to shipping company and fill out a concealed damage report. 2. Using foot / hand control, move table to horizontal position; then raise BACK section approximately 10°. Back 3. Position back upholstery (1, Figure 4) with two locking posts (A) in slots (B); then push upholstery toward foot end to engage posts. Secure with two large knobs (2). Seat 4. Position seat upholstery (3) and secure wih four large knobs (4). EQUIPMENT ALERT Do not use a sharp object to open upholstery packaging. This could result in accidental damage. MA524302i A B 1 2 3 5 6 7 10 4 9 9 8 Figure 4. Installation

5. Position footrest upholstery (5) and secure with four 3/4” knobs (6).

Headrest 6. Position headrest upholstery (7) and secure with two 1/2” knobs (8). Armrests 7. Position armrest upholstery (9) and secure with two female thread knobs (10). OPERATION DANGER If the table malfunctions, immediately release function button(s), unplug the table power cord, and assist the patient from the table. If table continues to malfunction, call for service.. Keep personnel and equipment clear of the table before initiating power movement to avoid personal injury or equipment damage. CAUTION To prevent foot control pedals and switches from being inadvert- ently activated, do not position foot control directly below foot sec- tion or headlock assembly of table. EQUIPMENT ALERT For optimal table performance, allow the table to reach room tem- perature before operating. Movement of the Tilt, Back, and Foot functions will automatically stop when the travel limit is reached, but the actuators will continue to run until button/ pedal is released. Release button/pedal when movement stops. Failure to comply may cause actuator motor to overheat and reduce the life of motor. NOTE This equipment is designed for intermittent operation (motor run time = 15 seconds on / 5 minutes off); each motor is protected by a thermal overload safety switch. If operation of a function ceases, immediately stop operation and allow the motor to cool for 10 to 15 minutes before next use. The thermal overload safety switches automatically reset when the motor cools to a safe operating temperature. Installation Operation

Standard Table Functions

To operate standard table functions [Base Up/Down, Back Up/Down, Tilt Up/Down, Foot Up/Down] follow the instructions below. Using Foot Control (standard): 1. Activate desired table function (Base, Back, Tilt, Foot) by pressing corresponding button. [Note: Each function has a light, which illuminates when that function is activated]. See Figure 5. 2. Within 10 seconds, press and hold desired side of Foot Pedal. [Left side of pedal = UP, Right side of pedal = DOWN] Release pedal to stop movement. Using Hand Control (optional): 1. Press and hold button for desired function. Release button to stop movement. See Figure 5. EQUIPMENT ALERT Do not immerse hand control in water / fluids. Failure to comply could result in damage to hand control and/or electric shock. MA597500i MIDMARK 1 2 4 3 P MIDMARK Base Down Base Up Back Up Tilt Down Back Down Tilt Up Foot Down Foot Up Base Foot Back Tilt Foot Pedal Figure 5. Operation

Programming Table Positions

(Programmable Tables only) Up to four table positions can be stored in memory. This allows the operator to recall frequently used positions quickly and easily. To change a programmed po- sition, simply repeat the programming procedure. The new position will replace the existing one. To program a table position: 1. Move table to desired position. 2. Press Program button (1, Figure 6); [Program Mode lamp will illuminate] Note: step 3 must be performed within 5 seconds of step 2. 3. Press the desired Program Position button (2). [Program Mode lamp will flash 3 times to indicate position was stored. If pro- gramming fails, repeat this procedure. ] To recall a programed position: 1. Using Hand Control: Press & hold the desired Program Position button (2). Using Foot Control: Press desired Program Position button (2). Note: w/foot control, step 2 must be performed within 10 seconds of step 1. [Lamp beside Program Position button will illuminate] 1. Press and hold either side of Foot Pedal (3). MA597600i MIDMARK 1 2 P MIDMARK 1 2 P 3 4 4 3 2 1 2 2 1 3 Footrest Extension Lamp Program Mode Lamp Footrest Extension Lamp Program Mode Lamp Figure 6. Operation

Listed below are solutions to common problems experienced during the

programming procedure. Solution: There is no table position saved for this Program Position button. Perform Programming Procedure to re-enter table position. Solution: Push footrest extension to its stowed position. Solution: Perform Error Clearing Procedure, then repeat Programming Procedure. Error Clearing Procedure: 1. Unplug table power cord. 2. Press and hold Program button (1, Figure 6) while plugging in power cord. [Program Mode lamp / Footrest Extension lamp will illuminate for 5 seconds] Release Program button when lamps go out. 3. After 10 - 20 seconds the table will sound two beeps to indicate memory was succesfully cleared. PROBLEM #1 When attempting to recall a programmed position, there is no movement and Program Mode lamp flashes. PROBLEM #2 When attempting to recall a programmed position, table stops and Footrest Extension Lamp flashes. PROBLEM #3 Programming procedure continually fails. Operation

Home Function (Programmable Tables only)

The Home button automatically moves the table’s Base function to its lowest height. The Back, Tilt, and Foot functions do not move. To activate Home function: 1. Press Home button (1, Figure 7). [Home function lamp will illuminate] If using hand control, table will move to its lowest height. 2. Press and release right side of foot pedal (2). [Home function lamp will go out and table will move to its lowest height] 3. To stop movement at any time, press Stop button (3). CAUTION Be sure personnel and equipment are clear of table before ac- tivating Home function. Failure to do so could result in per- sonal injury and / or equipment damage. NOTE If table stops before reaching its lowest height and Footrest Extension Lamp flashes, push footrest extension to stowed position and push Home button. MA597700i MIDMARK 1 2 P MIDMARK 1 2 P 3 4 4 3 2 3 1 1 3 Home Function Lamp Footrest Extension Lamp Footrest Extension Lamp Figure 7. Operation

Multiple Function Positioning

When using the foot control, only one function may be selected at a time; pressing any other function will deactivate the previously selected function. When using the hand control, multiple functions can be operated at the same time. However, if a function’s up and down buttons are pressed at the same time, neither function will move. Footrest Extension Lamp When the Footrest Extension Lamp (1, Figure 8) is illuminated, the following functions will not operate: Tilt Down, Base Down, & Foot Down. This prevents the footrest extension from colliding with the floor. To resume normal operation, push the footrest extension to its stowed position. MA597800i MIDMARK MIDMARK 1 2 4 3 P 1 1 Figure 8. Operation

1. Pull stop pin (1, Figure 9)

outward. 2. Rotate armrest (2) up or down until stop pin locks armrest in desired position. Armrest Removal / Installation Removal 1. Loosen knob (1, Figure 10) and pull armrest (2) out of drive slot (3). Installation 1. Insert drive rod of armrest (2) into drive slot (3). Be sure armrest is parallel to seat section (+10°). 2. Tighten knob (1). MA525601i 2 1 Figure 9. Figure 10. Operation

Headrest Positioning / Removal / Orientation

The headrest can be removed and exchanged with another approved headrest to facilitate specific procedures (See List of Authorized Accessories later in this manual). To position the headrest (1, Figure 11), slide in or out as desired. To remove the headrest (1), pull head- rest straight out of slide mechanism (2). To install headrest (1), slide headrest into slide mechanism (2). To orient the headrest: (Figure 12) Pivot Point (A) 1. Position headrest at pivot point (A) as desired. Pivot Points (B & C) 2. Pull handle(1) outward to unlock. 3. Position headrest at pivot pionts (B & C) as desired. 4. Push inward on handle (1) to lock in place. 5. Apply pressure to headrest to ensure headrest assem- bly is securely locked in position. CAUTION Be sure headrest is securely in position before use. Failure to secure headlock assembly could result in serious injury. MA525701i 1 2 Figure 11. C B A 1 MA598000i Figure 12. Operation

Headrest Slide Adjustment

The headrest slide mechanism holds the headrest in place using friction. The following procedure adjusts the friction holding the headrest in place, as well as the amount of force needed to slide it up or down.. To tighten headrest slide mechanism: 1. Tighten screw (1, Figure 13) until headrest bar (2) slides up/down as desired. To loosen headrest slide mechanism: 2. Loosen screw (1, Figure 13) until headrest bar (2) slides up/down as desired. CAUTION After making any adjustments, be sure headrest mechanism will hold desired position before performing next procedure. Failure to do so could result in serious injury to patient. NOTE Use the 3/16” allen wrench [provided] to perform the following adjustments. 2 MA526101i 1 Figure 13. Operation

Headrest Pivot Tension Adjustment

This adjustment affects the tension at pivot points (A & B, Figure 14). 1. Move handle to unlocked position. 2. Loosen set screw (1). 3. To increase tension: Tighten screw (2). To decrease tension: Loosen screw (2). 4. Tighten set screw (1). 5. Move handle to locked position & check for proper operation. Repeat procedure if necessary. Accessory Rails To mount authorized accessories, a 3/8 in. x 1 in. rail (1, Figure 15) is locat- ed along each side of the back sec- tion. (See List of Authorized Accessories in this manual). CAUTION After making any adjustments, be sure headrest mechanism will hold desired position before performing next procedure. Failure to do so could result in serious injury to patient. NOTE Use the two allen wrenches [provided] to perform the following adjustments. 1 2 Figure 14. A B Figure 15. Operation

List of Authorized Accessories

Listed below are the accessories authorized for use with this table. Accessory Name Order Number Restrictions / Intended Use IV Pole 9A77001 Restrictions - Must be removed or rotated clear of moving parts before positioning table. Also, the IV Pole may interfere with an 8 ft. (20.3 cm) ceiling. Intended Use - To provide secure support for the fluid for an IV infusion. Used with a 9A272 Base Rail. Vision Block Screen 9A78001 Restrictions - Must be removed or rotated clear of moving parts before positioning table. Intended Use - To obscure the patient’s vision of a general procedure / examination. Used with a 9A272 Base Rail. Articulating Armboard 9A81001 Restrictions - Do not use with any other acces- sory on the same side of table. Use caution when positioning this accessory and positioning table. Intended Use - To hold an arm in a secure posi- tion. Special Procedures Armboard 9A82001 Restrictions - Do not operate table with this accessory installed. Intended Use - To provide positioning and restraint for an arm during a minor procedure on the hand. Fixed Armboard 9A179002 Restrictions - Do not use with any other acces- sory on the same side of table. Use caution when positioning this accessory and positioning table. Intended Use - To hold the arm in a secure posi- tion, while the table is in a flat position. Welch Allyn Bracket 9A182001 Intended Use - To hold a Welch Allyn light on the table. Used with a 9A272 Base Rail. Caster Base 9A213001 Restrictions - Do not use simultaneously with 9A273001, Knee Crutch. Intended Use - To attach to the examination table to allow for mobility of the examination table. Foot Control - Programmable 9A253001 Intended Use - To position the table before, during, or after an examination or general procedure by using the foot. Seat Rails 9A254001 9A254002 Restrictions - Do not use simultaneously on the same side of table with 9A272, Base Rail. Intended Use - To provide a secure support for the attachment of Midmark accessories to the seat section of the table. Non-articulating Headrest w/ Adjustment Pad 9A255xxx Intended Use - To provide non-articulated support of patient’s head Restraint Straps 9A265001 9A265002 Intended Use - To confine the patient to the table during an examination / general procedure. Operation

Intended Use - To provide a larger pillow support

area for patient’s head. Interchangeable with stan- dard headrest pillow on double & triple articulating headrests. Instrument Tray (double arm) 9A267001 Restrictions - Must be removed or rotated clear of moving parts before positioning table. Total weight allowed to be placed on tray is not to exceed 15 lbs (6.8 kg), centered on tray. Do not use more than one 9A267, Instrument Tray simultaneously on same side of table. Intended Use - To hold instru- ments used during an examination or general pro- cedure. Used with a 9A272 Base Rail. Instrument Tray (single arm) 9A267002 Restrictions - Must be removed or rotated clear of moving parts before positioning table. Total weight allowed to be placed on tray is not to exceed 30 lbs (13.6 kg), centered on tray. Do not use more than one 9A267, Instrument Tray simultaneously on same side of table. Intended Use - To hold instru- ments used during an examination or general pro- cedure. Used with a 9A272 Base Rail. Instrument Tray (only) 9A268001 Intended Use - To provide additional tray(s) for use with 9A267 Instrument Tray. Used with a 9A267 Instrument Tray. Base Rail Assembly 9A272001 9A272002 Restrictions - Do not use simultaneously on the same side of table with 9A254, Seat Rail Assembly. Intended Use - To provide a secure and stationary support for the attachment of other accessories (9A77 IV Pole, 9A78 Vision Block screen, 9A267 Instrument Tray, 9A182 Welch Allyn Bracket) of which the multiple attachment of these accesso- ries, at the same time, is not recommended. Knee Crutch Assembly 9A273001 Restrictions - Must be removed or rotated clear of moving parts before positioning table. Also, do not use simultaneously with 9A21300X, Caster Base Assembly. Intended Use - To provide support for the knee and legs. Hand Control - Programmable 9A277001 Intended Use - To position the table, before, dur- ing, or after an examination or general procedure by using the hand. Upholstery Kits 002-0607-xxx xxx - specifies color; call for avail- able colors. Intended Use - To allow a facility to install the type of upholstered top which best suits their patient’s needs. Wrist Rest Kit for Dbl. Articulating Headrest 9A278002 Restrictions - To be used with Dbl. Articulating Headrest 9A310xxx Intended Use - To provide support for wrists during general procedures and examinations. Dbl. Articulating Headrest 9A310xxx Intended Use - To provide double articulating sup- port for patients head. When ordering, specify upholstery color. Accessory Name Order Number Restrictions / Intended Use Operation

Periodically inspect the following areas:

• Power cord should be free of cuts or other visible damage. • All fasteners should be present and tightened securely. • All mechanical functions should operate properly. Periodically lubricate the following areas to manitain quiet, smooth operation: • Back, Tilt, & Foot hinges using a light machine oil • Accessory rails and foot extension slide using household furniture wax. Have your table inspected by an authorized service technician every six months. Care of Table Clean the table weekly, wiping the painted metal and plastic surfaces with a clean, soft cloth, and mild cleaners. Periodic applications of common furniture wax will ease cleaning and maintain the finished luster of the table. Wrist Rest Kit for Triple Articulat- ing Headrest 9A311001 Restrictions - To be used with Triple Articulating Headrest 9A312xxx Intended Use - To provide support for wrists during general procedures and examinations. Triple Articulat- ing Headrest 9A312xxx Intended Use - To provide triple articulating sup- port for patients head. When ordering, specify upholstery color. EQUIPMENT ALERT The table’s upholstery is resistant to most medicinal-type stains, but may be damaged by solvents and dyes. Immediately remove any fluids spilled on the upholstery. Accessory Name Order Number Restrictions / Intended Use Operator Maintenance

Figure 16 shows the location of the table’s major components. Refer to chart

below for description. DESCRIPTION OF COMPONENTS 1. Back Section (adjustable angle) 6. Leveling Feet 2. Seat Section (adjustable tilt) 7. Hand Control (optional) 3. Foot Section (adjustable angle) 8. Foot Control 4. Foot Rest Extension 9. Armrests 5. Headrest (adjustable) 10. Accessory Rails 1 2 3 4 5 6 9 10 7 8 MA524402i 9 Figure 16. Components Overview

CONTROLS & INDICATORS (Standard Functions)

Figures 17 & 18 show the controls & indicators for standard functions. The chart on the accompanying pages describes each function. [Controls & indicators for programmable functions (optional) are described later in this section]. MIDMARK 1 2 4 3 P MIDMARK MA535301i 14 4 6 7 8 1 5 3 2 9 10 13 12 5 11 Hand Control (Optional) Foot Control Figure 17. Controls & Indicators

Back Up button

raises back section 2 Back Down button lowers back section 3 Tilt Up button raises front of seat section, tilting entire table top 4 Tilt Down button lowers front of seat section, tilting entire table top. 5 Foot Rest Extension lamp illuminates when Table / Foot /Tilt Down, or Home button(s) are pressed while foot rest extension is extended. NOTE: a programmed position, with any of the movements listed above will cause lamp to illuminate when foot rest extension is extended. 6 Foot Up button raises foot section 7 Foot Down button lowers foot section 8 Table Up button raises table top 9 Table Down button lowers table top 10 Back Function Selector button & lamp when depressed, the Back function is acti- vated; the up / down foot pedal will position the back section. The lamp illuminates to indicate the Back function is activated. 11 Tilt Function Selector button & lamp when depressed, the Tilt function is activated; the up / down foot pedal will position the tilt section. The lamp illuminates to indicate the Tilt function is activated. 12 Foot Function Selector button & lamp when depressed, the Foot function is acti- vated; the up / down foot pedal will position the foot section. The lamp illuminates to indi- cate the Foot function is activated. 13 Table Function Selector button & lamp when depressed, the Table function is acti- vated; the up / down foot pedal will position the table top. The lamp illuminates to indicate the Table function is activated. 14 Up / Down Foot pedal when the foot pedal is tilted to the left, the active function raises. When the foot pedal is tilted to the right, the active function lowers. Also used to initiate the “Home” function after it has been activated. Controls & Indicators

Arm Rest Knob

Secures the arm rest to the back section. Unscrewing the knob allows the arm rest to be removed from the table. 16 Arm Rest Stop Pin used to position arm rest in normal or stowed position. Pulling on stop pin releases arm rest for positioning. Stop pin automatically returns to locked position when arm rest is moved into the normal or stowed positions. 17 Foot Rest Extension Release handle used to release foot rest extension so it may be extended to desired position. 18 Headlock Handle used to unlock the headrest assembly so it may be positioned, then relocked in the desired position. Ref. Control Function MA535401i 17 15 16 18 Figure 18. Controls & Indicators

CONTROLS & INDICATORS (Programmable Functions - optional)

Figure 19 shows the controls & indicators for programmable functions (optional). The chart on the accompanying page describes each function. [Controls & indi- cators for standard functions are described earlier in this section]. MIDMARK 1 2 4 3 P MA524501i 6 7 9 1 3 2 4 MIDMARK P 1 2 3 4 5 5 8 8 Hand Control (optional) Foot Control Figure 19. Controls & Indicators

when pressed, the Home Position mode or

Program mode is interrupted and stopped. All functions are deactivated. 2 Programmed Position buttons (1, 2, 3, and 4) when pressed, the table top automatically moves into the position which has been stored into memory for that particular button (movement continues only as long as the but- ton is depressed). 3 Home Position button when pressed, the table top automatically lowers to its lowest height for patient dis- mount. 4 Program button when pressed, the program mode is acti- vated. Pressing a programmed position (1, 2, 3, or 4) button now will store the table top’s present position in that button’s memory. 5 Program Mode indicator lamp illuminates to indicate that the Program Mode is activated (stays illuminated for five seconds after Program button is pressed). 6 Home Function Selec- tor button & indicator lamp when depressed, the Home function is acti- vated; depressing the up / down foot pedal to the right will cause the “Home” function to begin. The indicator lamp illuminates to indi- cate the Home function is activated. 7 Stop button when depressed, the Home mode or the Pro- gram mode is interrupted and stopped and all manual functions are temporarily disabled. 8 Programmed Position buttons (1, 2, 3, and 4) & indicator lamp when depressed, the Programmed Position function is activated for that button and its indicator lamp illuminates to indicate that the function is activated. Depressing the up / down foot pedal now will cause the table top to automatically move into the position which has been stored into memory for that particu- lar button (movement continues only as long as the foot pedal is depressed). 9 Program button when depressed, the Program mode is acti- vated. Pressing a Programmed Position (1, 2, 3, or 4) button now will store the table top’s present position in that button’s memory. Controls & Indicators

CALLING FOR SERVICE

Questions, comments, and service inquiries should be directed to your Midmark Authorized Dealer [refer to inside front cover of this manual]. Model and serial number information will be required when calling for service. If your dealer can- not resolve your problem, contact Midmark directly: 1-800-Midmark (1-800-643-6275) or 937-526-3662 8:00 a.m until 6:00 p.m. Monday through Thursday & 8:00 a.m until 5:00 p.m. Friday (Eastern Standard Time in the U.S.) [excluding standard U.S. holidays] SPECIFICATIONS Patient Weight (Maximum) ........... 325 lbs (147.4 kgs) maximum Weight of Table.............................. 515 lbs (233.6 kgs) Back Section Range...................... 0° (horizontal) to +85° ± 5° Table Top Height Range................ 22.4 ± 0.5 in. to 40.4 ± 1.0 in. (56.9 ± 1.3 cm to 102.6 ± 2.5 cm) Foot Section Range ...................... 0° (horizontal) to -90° ± 5° Table Top Tilt Range ..................... 0° (horizontal) to +30° ± 5° (foot end of table up) Dimensions: Upholstered top: w/o headrest extended.................. 23.0 in. wide x 65.0 in. long (w/back down & foot up) (54.4 cm x 165.1 cm) with headrest extended ................. adds 15.0 in. (38.1 cm) Footrest Extension (adjustable)....... adds 0.0 in. to 8.0 in (0 cm to 20.3 cm) Power Cord ..................................... 8 ft., 2 in. (248.9 cm) long. A black jacketed molded cord to fit IEC appliance inlet with destination specific attachment plug. Electrical Rating: 419(-003 & -004)............................. 115 VAC ±10%, 11 amps, 50/60 HZ 419(-005 & -006)............................. 230 VAC ±10%, 5.5 amps, 50/60 HZ Fuse Replacement (near power cord inlet): 419 (-003 & -004)............................ 12A, 250V, Time Lag 419 (-005 & -006)............................ 8A, 250V, Slo-Blo Duty Cycle.........................................Intermittent Operation [15 seconds on / 5 minutes off (motor run time)] Calling For Service Specifications

Classifications ...............................Class 1, Type B Applied Part, Ordinary

Equipment, Intermittent Operation Certifications: Midmark Corporation.......................ISO 9001 Certified 419 (-003 & -004) comply to............UL2601-1, CAN/CSA 22.2, #601.1-M90 419 (-003, -004, -005, & -006) comply to ......................................MDD 93/42/EEC, Annex VII EN60601-1, EN60601-1-2, EN60601-1-4 Equipment not suitable for use in the presence of a flammable anesthetic mixture with air, oxygen, or nitrous oxide. RANGE OF MOTION MA598100i 40.4" (102.6 cm) 22.4" (56.9 cm) 30 85 90 Base Function Foot Function Back Function Tilt Function Range Of Motion

SCOPE OF WARRANTY

Midmark Corporation (“Midmark”) warrants to the original purchaser its new Alter- nate Care products and components (except for components not warranted under “Exclusions”) manufactured by Midmark to be free from defects in material and workmanship under normal use and service. Midmark’s obligation under this war- ranty is limited to the repair or replacement, at Midmark’s option, of the parts or the products the defects of which are reported to Midmark within the applicable warranty period and which, upon examination by Midmark, prove to be defective. APPLICABLE WARRANTY PERIOD The applicable warranty period, measured from the date of delivery to the original user, shall be one (1) year for all warranted products and components. EXCLUSIONS This warranty does not cover and Midmark shall not be liable for the following: (1) repairs and replacements because of misuse, abuse, negligence, alteration, acci- dent, freight damage, or tampering; (2) products which are not installed, used, and properly cleaned as required in the Midmark “Installation” and or “Installation / Oper- ation Manual for this applicable product. (3) products considered to be of a consum- able nature; (4) accessories or parts not manufactured by Midmark; (5) charges by anyone for adjustments, repairs, replacement parts, installation, or other work per- formed upon or in connection with such products which is not expressly authorized in writing in advance by Midmark. EXCLUSIVE REMEDY Midmark’s only obligation under this warranty is the repair or replacement of defec- tive parts. Midmark shall not be liable for any direct, special, indirect, incidental, exemplary, or consequential damages or delay, including, but not limited to, dam- ages for loss of profits or loss of use. NO AUTHORIZATION No person or firm is authorized to create for Midmark any other obligation or liability in connection with the products. THIS WARRANTY IS MIDMARK’S ONLY WARRANTY AND IS IN LIEU OF ALL OTHER WARRANTIES, EXPRESS OR IMPLIED. MIDMARK MAKES NO IMPLIED WARRANTIES OF ANY KIND INCLUDING ANY WARRANTIES OF MERCHANTABILITY OR FITNESS FOR ANY PARTICULAR PURPOSE. THIS WARRANTY IS LIMITED TO THE REPAIR OR REPLACEMENT OF DEFECTIVE PARTS. SF-1487 REV. A1 29 Limited Warranty

© Midmark Corporation - 2002

003-1290-00 Rev. C (2/03) Midmark Corporation declares that Model 419 (-003, -004, -005, -006) conform to the applicable requirements of the Medical Device Directive 93/42/EEC, Annex VII. Model 419-003 - Serial Numbers: LM001000 to present. Model 419-004 - Serial Numbers: LN001000 to present. Model 419-005 - Serial Numbers: LP001000 to present Model 419-006 - Serial Numbers: LR001000 to present. DECLARATION OF CONFORMITY MIDMARK CORPORATION Midmark Corporation, Versailles, Ohio 45380 U.S.A. 937-526-3662 FAX: 937-526-5542 Quality Assurance Manager Gene Harshbarger Midmark Corporation 60 Vista Drive, P O Box 286 Versailles, Ohio 45380 U.S.A. May 12, 1999