hivolt.de GmbH & Co. KG

Oehleckerring 40 ∙ D-22419 Hamburg ∙ Germany ∙  +49 40 537122-0 ∙  +49 40 537122-99 ∙ [email protected] ∙ www.hivolt.de - HA2B5-S High Voltage Amplifier ±2000V HA3B3-S High Voltage Amplifier ±3000V HAR42-4 High Voltage Amplifier Mainframe for 4 Channels HA2B5-S / HA3B3-S Operating Manual

hivolt.de GmbH & Co. KG

Oehleckerring 40 ∙ D-22419 Hamburg ∙ Germany ∙  +49 40 537122-0 ∙  +49 40 537122-99 ∙ [email protected] ∙ www.hivolt.de 1 WARNINGS 3 2 OVERVIEW 5 3 DESCRIPTION 5 3.1 Block Diagram 5 3.2 Control Voltage Input 6 3.3 Monitor Outputs 6 3.4 Differential Output Mode 6 3.5 DISABLE 7 3.6 Loading Conditions 7 3.7 Protection 8 3.8 Interlock / HV Off 8 4 TECHNICAL DATA 9 4.1 Frequency Response of Current Monitor 10 4.2 Ambient Conditions 10 4.3 Mechanical Specifications 11 5 OPERATION 12 5.1 Initial Check 12 5.2 Warning Notices 12 5.3 Temperature Compensation 12 5.4 Power Supply 13 5.5 Ventilation 13 5.6 Functional Test 13 5.7 Connector X1 INTERLOCK 14 5.8 Output Connector X7 14 5.9 Grounding 14 5.10 DIP Switch S1 14 5.11 Internal Connectors 15 6 OPERATION / MAINTENANCE 16 6.1 Troubleshooting 16 6.2 Maintenance 16 6.3 Cleaning 16 7 DECLARATION OF CONFORMITY 18

hivolt.de GmbH & Co. KG

Oehleckerring 40 ∙ D-22419 Hamburg ∙ Germany ∙  +49 40 537122-0 ∙  +49 40 537122-99 ∙ [email protected] ∙ www.hivolt.de 1 Warnings Caution! This device produces dangerous voltage above 3000V. Due to capacitive charging, dangerous voltages can still be present in the set-up even after the amplifier has been switched off! Please respect the following rules before every activation of the high voltage amplifier:  The device should be operated only by skilled personnel, in accordance with the local regulations and the instructions given in this manual.  The device may only be operated as a component of an overall structure that fully complies with the regulations for working with high voltage systems.  Before switching the unit on, the experiment setup must be checked, and safety should be assured. High voltage areas have to be blocked and secured.  In case of suspected damage or malfunction, the device should immediately be put out of service, and it should be secured against unintentional or accidental operation.  The safety ground must always be connected! The grounding nut on the rear panel must be connected to the grounding point of the setup and to protective earth. Local regulations about grounding should be taken into account.  The output connectors must only be operated when the unit is switched off.  High voltages may still exist even after the switch-off of the device due to capacitive charge! Capacitances connected to the outputs of the device can possibly remain charged to dangerous voltages, even after switching off the device  Before removing any covers disconnect the unit from the power supply!  Before touching the output or working on the experimental setup, disconnect the unit from the power supply!  No wires or similar objects may project into the device through the ventilation slots  Fuses have to be replaced by types as rated on the nameplate on the rear of the unit. Personal safety must be given the highest priority!

hivolt.de GmbH & Co. KG

Oehleckerring 40 ∙ D-22419 Hamburg ∙ Germany ∙  +49 40 537122-0 ∙  +49 40 537122-99 ∙ [email protected] ∙ www.hivolt.de 1 Warnhinweise Achtung! Dieses Gerät erzeugt lebensgefährliche Spannungen über 3000V. Vor jeder Inbetriebnahme des Hochspannungsverstärkers sind u. a. folgende Regeln zu beachten:  Die Inbetriebnahme darf nur von einer Elektrofachkraft im Sinne der Berufsgenossenschaft der Feinmechanik und Elektrotechnik vorgenommen werden.  Vor der Inbetriebnahme muss die Bedienungsanleitung gelesen und verstanden worden sein.  Es sind die einschlägigen Bestimmungen und Vorschriften des Gesetzgebers, der Berufsgenossenschaft und des VDE zu beachten, insbesondere - DIN-VDE 104 "Errichten und Betreiben elektrischer Prüfanlagen" (s. auch BGI 891 "Errichten und Betreiben von elektrischen Prüfanlagen") - Unfallverhütungsvorschrift DGUV Vorschrift 3 / BGV A3 "Elektrische Anlagen und Betriebsmittel"  Das Gerät darf nur als Bestandteil eines im Ganzen den Vorschriften für den Umgang mit Hochspannung genügenden Gesamtaufbaus betrieben werden!  Vor jedem Einschalten des Geräts ist der Versuchsaufbau zu überprüfen und sicherzustellen, dass es zu keiner Gefährdung kommen kann. Die Hochspannung führenden Bereiche müssen vorschriftengerecht abgesperrt oder anderweitig gesichert sein!  Sollte der Verdacht bestehen, dass das Gerät beschädigt ist oder Fehlfunktionen zeigt, ist es umgehend außer Betrieb zu setzen und gegen beabsichtigten oder unbeabsichtigten Betrieb zu sichern.  Der Erdbolzen auf der Rückplatte des Geräts muss mit dem Zentralen Erdpunkt des Versuchsaufbaus und dem Schutzleiter verbunden werden. Die örtlichen Vorschriften über Erdung sind zu beachten.  Die Ausgangssteckverbinder dürfen nur bei spannungsfrei geschaltetem Gerät betätigt werden!  Bei Arbeiten am Versuchsaufbau oder bei Berühren der Ausgangsanschlüsse ist das Gerät zuvor von der Spannungsversorgung zu trennen.  Vor dem Öffnen des Geräts ist das Gerät von der Spannungsversorgung zu trennen.  Es dürfen keine Objekte – wie Drähte o. ä. durch die Lüftungsschlitze in das Gerät Hereinragen.  Im angeschlossenen Versuchsaufbau können evtl. vorhandene Kapazitäten auf Hochspannung aufgeladen werden. Diese können auch nach Abschalten des Geräts noch gefährliche Spannungen führen.  Sollte ein Ersatz der Netzsicherungen erforderlich sein, so ist sicherzustellen, dass nur Sicherungen der angegebenen Nennstromstärke und Nennspannung als Ersatz verwendet werden. Der Personensicherheit ist höchste Priorität einzuräumen!

hivolt.de GmbH & Co. KG

Oehleckerring 40 ∙ D-22419 Hamburg ∙ Germany ∙  +49 40 537122-0 ∙  +49 40 537122-99 ∙ [email protected] ∙ www.hivolt.de 2 Overview This manual covers the following models:  HA2B5-S single channel high voltage amplifier / 3U plug-in card  HA3B3-S single channel high voltage amplifier / 3U plug-in card  HAR42-4 Tabletop 4U subrack for up to 4 amplifier modules HA2B5-S / HA3B3-S 3 Description The single-channel high voltage amplifiers HA2B5-S / HA3B3-S with bipolar output voltage are designed to drive capacitive and resistive-capacitive loads. The amplifiers are characterized by high speed, good stability and low noise. Piezo elements, electroactive polymers, electrorheological fluids, electrostatic deflecting electrodes and many other loads can be driven by this amplifier easily. For HA2B5-S output voltages of -2000V to +2000V at load currents of up to ±5mA and >7mAP are provided. The signal gain is 200, the input voltage range is -10V…+10V. For HA3B3-S output voltages of -3000V to +3000V at load currents of up to ±3mA and >5mAP are provided. The signal gain is 300, the input voltage range is -10V…+10V. The output voltage can be monitored by a high-speed voltage monitor; a current monitor provides a representation of the output current. HA2B5-S and HA3B3-S are 3U plug-in cards intended to be mounted in a suitable subrack. The deviced incorporate the high voltage amplifier itself, high voltage sources for the output stage, measuring functions as well as monitoring and protective functions. The amplifier’s output is protected against overload, short circuit, overtemperature, transient overvoltage and high voltage flashover. LEDs on the front panel indicate operation and error statuses. The HAR42-4 is a tabletop subrack to accommodate up to 4 high voltage amplifier modules HA2B5-S or HA3B3-S. Both models can be combined in a single subrack. It features internal power supplies, cooling fans, a common Interlock input for all channels and a HV OFF button. Customized and full custom models are available on request. 3.1 Block Diagram

hivolt.de GmbH & Co. KG

Oehleckerring 40 ∙ D-22419 Hamburg ∙ Germany ∙  +49 40 537122-0 ∙  +49 40 537122-99 ∙ [email protected] ∙ www.hivolt.de 3.2 Control Voltage Input The BNC control voltage input In is connected to a differential amplifier to suppress common mode voltages between the external signal source and the amplifier (ground loops). Two jumpers (W2, W5) are provided on each amplifier module in order to adapt the control voltage input to the signal source. W5 connects the input signal reference via 100Ω to the amplifier signal ground. W2 adjusts the amplifier gain to the output impedance of the signal source. W5 1-2 Floating signal source Common mode voltage <1VRMS Default 2-3 Grounded signal source (connected to PE) Common mode voltage <5VRMS W2 1-2 Output impedance of signal source <1 Ω Default 2-3 Output impedance of signal source 50 Ω 3.3 Monitor Outputs Two BNC monitor outputs provide actual values of output voltage and output current, both of which are normalized to ±10 V. The voltage monitor output V Mon receives its signal via a compensated voltage divider and provides a scaled image of the output voltages. The frequency response is linear up to well above the upper cut-off frequency of the amplifiers. The frequency response of the I Mon current monitor output is limited to about 9kHz in order to filter out the common mode ripple of the internal high voltage sources. The monitor outputs are buffered (output impedance: 2kΩ) and short-circuit-proof. They are able to drive capacitive loads (coaxial cables), but are not designed to drive cables with a low-impedance termination. 3.4 Differential Output Mode Two amplifier modules can be configured to operate in a differential output mode. In this configuration, only amplifier 1 (Master) receives the input signal from the signal source. Amplifier 2 (Slave) receives the input signal through the backplane of the subrack. The signal to amplifier 2 is inverted (180° phase shift in respect to amplifier 1). This mode can be used e. g. to differentially drive two deflection plates or to drive a floating load with up to 8000VPP.

hivolt.de GmbH & Co. KG

Oehleckerring 40 ∙ D-22419 Hamburg ∙ Germany ∙  +49 40 537122-0 ∙  +49 40 537122-99 ∙ [email protected] ∙ www.hivolt.de Two jumpers (W3, W4) are provided on each amplifier module to configure differential output mode: W3 1-2 Master Default 2-3 Slave W4 1-2 Master Default 2-3 Slave 3.5 DISABLE The high voltage amplifier channels can be remotely switched off through the DISABLE input. The signal is available through connector X1 INTERLOCK on the HAR42-4 rear panel. It is TTL-compatible. A LOW-signal turns on all amplifier outputs. In case of open input or HIGH signal, the amplifier outputs are forced to 0V. During disable the internal high voltage sources are not turned off. The DISABLE input can be used to turn on and off the amplifier outputs relatively fast. If the DISABLE input is not needed, a wire jumper can be installed between DISABLE (X1.4) and GND (X1.5) to turn on the amplifiers permanently. The DISABLE input is not suitable to achieve a safe state at the amplifier output. It must not be used for safety relevant purposes. 3.6 Loading Conditions The amplifiers are designed to drive capacitive and resistive-capacitive loads. The output is stable even with large capacitive loads. In case of larger load capacitances overshoot of the output may occur. The slew rate that can be achieved on the output depends on the load capacitance. The effective load capacitance CL consists of the amplifier’s internal output capacitance (ca. 80pF), capacitance of the output cable (a typical coaxial cable: ca. 100pF/m) and capacitance of the connected load. The output stage can provide peak currents (IOP) of > ±7.5mAP for ca. 1s. The maximum static output current is +5mA and -5mA. Achievable slew rate: SR = IO / CL [V/s]. If the average positive or negative output current exceeds 5mA (HA2B5-S) or 3mA (HA3B3-S) the amplifier shuts down with Overload. If the maximum slew rate is exceeded, distortions occur in the output signals. In principle, the amplifier should thus only be controlled by input signals that satisfy the achievable slew rate for a given load. However, dynamic override of the inputs (e.g. control by square wave signals) is harmless. The edges of the output waveform are nearly linear then with a slight overshoot. Large capacitive loads driven by high frequency signals may result in a DC-offset superimposed on the output voltage. This occurs if the final value of the desired output voltage can not be reached periodically.

hivolt.de GmbH & Co. KG

Oehleckerring 40 ∙ D-22419 Hamburg ∙ Germany ∙  +49 40 537122-0 ∙  +49 40 537122-99 ∙ [email protected] ∙ www.hivolt.de 3.7 Protection During operation, the internal auxiliary supplies, overload of the high voltage sources, overtempera- ture and interlock are monitored. In the event of a fault, the high voltage supplies are switched off; the shutdown state is latched and displayed on LEDs. A shutdown state can only be exited by means of a shutdown reset (push button SDN Reset on the HAR42-4 front panel or signal SDR (X1.1) on connector X1 INTERLOCK set to +24V). After power on the unit will be in shutdown state. LED Function Enable on: output stage is enabled by signal DISABLE=low HV ok on: internal HV sources on, no overload Interlock on: shutdown: interlock circuit open blink: shutdown after interlock Overload on: overload blink: shutdown due to overload Overtemp on: overtemperature blink: shutdown due to overtemperature 3.8 Interlock / HV Off The amplifier modules incorporate an interlock circuit through which the supply voltage of the high voltage sources is switched. The interlock signal is available via the rear connector X1 INTERLOCK on the subrack. The HV off palm button is looped into the interlock circuit. It is a closed circuit (break contacts). Its source voltage is 24VDC. To enable the high voltage generators +24V_IL (X1.2) and IL (X1.3) must be connected by a jumper or a contact. The contact can be part of an external interlock circuit. The quiescent current is 30mA typ. When the interlock circuit is closed and the HV off palm button is released, the palm button will be lit to indicate that the high voltage sources are switched on. The high voltage generation is not in operation until the Interlock shutdown state is cleared by shutdown reset. The device must be disconnected from the power supply before carrying out any operations on the test setup or before touching the output terminals.

hivolt.de GmbH & Co. KG

Oehleckerring 40 ∙ D-22419 Hamburg ∙ Germany ∙  +49 40 537122-0 ∙  +49 40 537122-99 ∙ [email protected] ∙ www.hivolt.de 4 Technical Data Mainframe HAR42-4 Parameter Conditions Mains voltage 95 – 265VAC Mains frequency 47 – 63Hz Input current VLine=115VAC, full load 1.0ARMS maximum VLine=230VAC, full load 0.5ARMS maximum Mains Fuses F1, F2 T6.3A, 250V, IEC127-2/V External Fusing 16A Protection Category I - Fuses F1, F2 are located within the mains connector unit on the rear. - Signal ground and high voltage ground are connected to chassis ground / protective earth and the earth connector (M4 nut). - The appliance conforms to protection category I. It must only be used on mains power sockets with a ground connection. Amplifier modules HA2B5-S / HA3B3-S Parameter Conditions Supply voltage, VS 24VDC ±10% Supply current, IS VS = 24V <1.6ADC Input voltage range Control input -10.0V – +10.0V Max. input voltage ±18V Input resistance 25kΩ typ. DC Gain HA2B5-S HA3B3-S 200 ±0.1% 300 ±0.1% Offset ≤ 100mV Output voltage range HA2B5-S HA3B3-S -2000V – +2000V -3000V – +3000V Load current range static HA2B5-S HA3B3-S -5mA – +5mA -3mA – +3mA Load current range dynamic, t < 1ms HA2B5-S HA3B3-S -7mA – +7mA -5mA – +5mA Power bandwidth CL = 0pF, THD ≤ 1% VO = 4000VPP HA2B5-S VO = 6000VPP HA3B3-S DC – ≥ 5kHz DC – ≥ 2.5kHz Power bandwidth CL = 100pF, THD ≤ 1% VO = 4000VPP HA2B5-S VO = 6000VPP HA3B3-S DC – ≥ 2.5kHz DC – ≥ 1.3kHz Small Signal Bandwidth CL = 100pF, VI = 2VPP, VO -3dB DC – ≥ 20kHz Slew-Rate CL = 100pF HA2B5-S HA3B3-S ≥ 40V/μs ≥ 25V/μs Internal output capacitance ca. 80pF Ripple, Noise CL = 100pF, HA2B5-S 1Hz – 20kHz HA3B3-S ≤ 20mVRMS ≤ 20mVRMS

hivolt.de GmbH & Co. KG

Oehleckerring 40 ∙ D-22419 Hamburg ∙ Germany ∙  +49 40 537122-0 ∙  +49 40 537122-99 ∙ [email protected] ∙ www.hivolt.de Scaling monitor output V HA2B5-S HA3B3-S 10V  2000V ±0.2% 10V  3000V ±0.2% Bandwidth monitor output V DC – ≥ 50kHz Scaling monitor output I 10V  10mA ±1% Bandwidth monitor output I DC – ≥ 9kHz - Signal ground and high voltage ground are connected to the chassis ground / earth terminal. 4.1 Frequency Response of Current Monitor 4.2 Ambient Conditions Parameter Conditions Min. Max. Unit Ambient temperature - Operation - Storage and Transportation -20 -25 +50 +70 °C °C Relative humidity - Operation - Storage and Transportation Not condensing 5 5 80 95 % % - Depending on the ambient temperature, modulation amplitude and load capacitance, the maximum output power may need to be derated.

hivolt.de GmbH & Co. KG

Oehleckerring 40 ∙ D-22419 Hamburg ∙ Germany ∙  +49 40 537122-0 ∙  +49 40 537122-99 ∙ [email protected] ∙ www.hivolt.de 4.3 Mechanical Specifications Mainframe HAR42-4 Parameter Unit Depth overall 490 mm Depth case 420 mm Width overall 250 mm Width case 235 mm Height 185 mm Weight without amplifier modules 4.0 kg Amplifier module HA2B5-S / HA3B3-S Parameter Unit Depth overall 255 mm Depth module 235 mm Width 40 mm Height 128 mm Weight 0.51 kg

hivolt.de GmbH & Co. KG

Oehleckerring 40 ∙ D-22419 Hamburg ∙ Germany ∙  +49 40 537122-0 ∙  +49 40 537122-99 ∙ [email protected] ∙ www.hivolt.de 5 Operation 5.1 Initial Check Once the product is delivered, please check the packaging and the device for possible transport damage. Please check the device taken out of the packaging for any mechanical defects before the unit is put into operation. If the device has any signs of damage caused by transport, please immediately inform the shipping company so that damages can be claimed. 5.2 Warning Notices - For safe operation of this device it should be put into operation by a qualified electrician according to this Operating Manual. - The device may only be operated as a component of an overall setup that fully complies with the regulations for working with high voltage systems. - Output connectors may only be touched when the device is disconnected from the power supply! Otherwise, there is a risk of electric shock. - The test setup must be fully wired and protected against any contact before the device is put into operation. - The test setup must be checked each time before the device is put into operation to ensure that it is not potentially dangerous. It should be checked that the high voltage connections are faultless and the insulation of the wires is not damaged. - The high-voltage areas must be blocked in accordance with regulations or secured by other means. - Once the test setup is connected, any existing capacitances can be charged to high voltage. They may carry dangerous voltages even after the device is switched off. - The ground bolt on the rear panel of the device must be connected to the central grounding point of the test setup and to protective earth. Local regulations on grounding must be observed. - No wires or similar objects may project into the device through the ventilation slots - If it is suspected that safe operation is no longer possible, the device has to be taken out of operation and secured against unintentional operation. This symbol on the output terminals warns of the risk of electric shock. 5.3 Temperature Compensation To avoid condensation within the device, it should be allowed to reach the room temperature. Please unpack the product at least two hours prior to power-up.

hivolt.de GmbH & Co. KG

Oehleckerring 40 ∙ D-22419 Hamburg ∙ Germany ∙  +49 40 537122-0 ∙  +49 40 537122-99 ∙ [email protected] ∙ www.hivolt.de 5.4 Power Supply The HAR42-4 subrack provides the 24V supply voltage for the amplifier modules. The mainframe is fed by mains voltage. 5.5 Ventilation The HAR42-4 subrack is equipped with fans on the rear panel. Air intake is below the front panels. Sufficient space must be allowed for cooling air to reach the ventilation inputs and for the fan exhaust air to exit from the rear of the unit. The speed of the fans is controlled automatically by the temperature of the amplifier modules. 5.6 Functional Test Before the device is finally put into operation in a setup, a short functional test is to be carried out. It is necessary that the interlock circuit is closed. 1. Make sure that the supply voltage is disconnected. 2. Remove all input signal cables and the output cable from the device. 3. Connect the terminals +24V_IL (X1.2) and IL (X1.3) by means of a wire link. 4. Connect the terminals DISABLE (X1.4) and GND (X1.5) by means of a wire link. 5. Connect the mains supply voltage. 6. Switch on the Power switch. 7. Release the HV off palm button. 8. The Power switch and HV off button will light up, Enable LEDs will light up, IL, Overload and Overtemp LEDs will flash. 9. Push the SDN Reset push button. 10. The IL, Overload and Overtemp LEDs will turn off, the HV ok LEDs will light up. 11. Push in the HV off palm button. 12. The HV off button and HV ok LEDs will turn off, the IL LEDs will light up. 13. Open the interlock bridge at X1. 14. Release the HV off palm button. 15. The Interlock LEDs will light up continuously. 16. Press the SDN Reset push button. 17. The interlock status remains unchanged and the high voltage cannot be switched on. 18. Turn off the supply voltage.

hivolt.de GmbH & Co. KG

Oehleckerring 40 ∙ D-22419 Hamburg ∙ Germany ∙  +49 40 537122-0 ∙  +49 40 537122-99 ∙ [email protected] ∙ www.hivolt.de 5.7 Connector X1 INTERLOCK Connector type: Phoenix Mini-Combicon, 5-pin Mating connector: Phoenix Mini-Combicon, 5 circuits, FK-MCP1.5_ 5-ST-3.81 Pin Signal Direction Function 1 SDR I Control signal: 24V for 0.2s => Shutdown Reset; Caution: This reactivates the HV output after a shutdown! 2 +24V_IL O 24V output for interlock circuit and SDR signal; protected against overload 3 IL I Interlock / bridge to X1.+24V_IL 4 DISABLE I Control signal: TTL compatible; 2.5kΩ internal pull-up resistor; LOW == enable amplifier outputs 5 GND - Supply voltage GND 5.8 Output Connector X7 Connector type: SHV The high voltage connector may only be mated or unmated when the supply voltage is switched off! Otherwise, there is a risk of an electric shock. 5.9 Grounding The M4 ground nut on the back panel of the device must be connected to the central grounding point of the test setup and to protective earth. 5.10 DIP Switch S1 Switch Function Default S1-1 Disable Interlock shutdown latch OFF S1-2 Disable Overload shutdown latch OFF S1-3 Disable Overtemperature shutdown latch OFF S1-4 not used For safety reasons it is not recommended to disable the shutdown latches. In case of disabled shutdown latches high voltage will reappear at the outputs after a shutdown condition (Interlock, Overload, Overtemperature) is no longer pending. It is recommended to operate the unit with shutdown latches enabled and actively reset any shutdown condition via the SDN Reset push button or the SDN signal on X1.

hivolt.de GmbH & Co. KG

Oehleckerring 40 ∙ D-22419 Hamburg ∙ Germany ∙  +49 40 537122-0 ∙  +49 40 537122-99 ∙ [email protected] ∙ www.hivolt.de 5.11 Internal Connectors Internal connector X1 of HA2B5-S / HA3B3-S plug-in modules (DIN41612, Type F) Pins Signal Dir Function D2 B2 Z2 n.c. D4 B4 Z4 n.c. D6 B6 Z6 n.c. D8 B8 Z8 n.c. D10 B10 Z10 n.c. D12 B12 Z12 GND - GND D14 B14 Z14 n.c. D16 B16 Z16 n.c. D18 HVON_BP I HV enable input; used for interlock or general high voltage ON; approx. 10mA @ 24V B18 SDNRST_BP I Shutdown reset input; clear latched shutdown conditions; tP>200ms @ 24V; approx. 2.5mA @ 24V Z18 DISABLE_BP I Disable input; TTL compatible; 10kΩ internal pull-up resistor to +5V; LOW == enable amplifier outputs; tD≈50ms; see 3.5 D20 STAT_BP O Status output; 5V; HIGH == high voltage generation enabled B20 VFCON O Fan control output; 100mV/°C; 9.3V == 100°C; min. load: 10kΩ; DO NOT OVERLOAD OR SHORT TO GND Z20 CHA_SEL_BP I Channel select input; TTL compatible; 10kΩ input resistance; HIGH == activate backpanel monitor outputs IOAV_BP and VOAV_BP D22 VOAV_BP O Output voltage monitor output; ±10V; must be activated via CHA_SEL_BP; high impedance if disabled; see 3.3 B22 IOAV_BP O Output current monitor output; ±10V; must be activated via CHA_SEL_BP; high impedance if disabled; see 3.3 Z22 AV_RTN - Monitor output reference point; connected to analog GND D24 INP_S I Slave unit pos. differential input; ±10V; see 3.4 B24 INP_S I connected to D24 Z24 INN_S I Slave unit neg. differential input; ±10V; see 3.4 D26 VONV_BP O Inverted input signal output; ±10V; used to drive 2nd channel module in differential mode; see 3.4 B26 VONV_BP O connected to D26 Z26 INN_S I connected to Z24 D28 VONV_RTN - VONV_BP output reference point; connected to analog GND B28 VONV_RTN - connected to D28 Z28 VT_MAX O Temperature monitor output; 100mV/°C; 10V == 100°C; min. load: 10kΩ; DO NOT OVERLOAD OR SHORT TO GND D30 B30 Z30 +24V I +24VDC supply voltage input D32 B32 Z32 GND - GND

hivolt.de GmbH & Co. KG

Oehleckerring 40 ∙ D-22419 Hamburg ∙ Germany ∙  +49 40 537122-0 ∙  +49 40 537122-99 ∙ [email protected] ∙ www.hivolt.de 6 Operation / Maintenance 6.1 Troubleshooting If the device behaves unusual or erratic, please switch off the supply voltage and check the wiring of the load and that of the control and monitoring signals. Check the connected load and the signal source. Check the interlock circuit. During operation various parameters are monitored and analyzed. In case of a fault the error is reported via error LEDs on the front panel of the amplifier module. Press the SDN Reset push button to reset the error condition. Do not attempt to locate any faults within the device. This can be dangerous to life due to the high voltage used in the device. In such a case, please return the device to the manufacturer after consultation. Symptom Possible causes Device will not turn on; Power LED will not light up - No supply voltage present - Fuses F1, F2 in the IEC inlet defective HV off palm button will not light up - HV off palm button is in off position (pushed in) - Interlock circuit is open Output signal is distorted - Load capacitance too large for the desired slew rate (see section 3.6 for load conditions) Amplifier switches off, the OVERLOAD indicator lights up - Dynamic load current exceeded (see section 3.6 for load conditions) - Static load current exceeded (see section 3.6 for load conditions) Amplifier switches off, the OVERTEMP indicator lights up - Poor ventilation - Total load too large (see section 3.6 for load conditions) 6.2 Maintenance Depending on the cleanliness of the ambient air dust may accumulate within the unit possibly blocking the airflow. In that case the accumulated dust has to be removed by blowing out the unit cautiously. The manufacturer specified lifetime of the fans is >50000h. After about 6 years of continuous operation the fans might be replaced. Type: Papst 8314H (HAR42-4). Caution: disconnect the unit from the power supply before removing any modules or opening any covers. Further regular maintenance is not required. 6.3 Cleaning If necessary, wipe the device with a slightly damp cloth. Do not use abrasive detergents or solvents.

hivolt.de GmbH & Co. KG

Oehleckerring 40 ∙ D-22419 Hamburg ∙ Germany ∙  +49 40 537122-0 ∙  +49 40 537122-99 ∙ [email protected] ∙ www.hivolt.de © hivolt.de GmbH&Co. KG, Hamburg, Germany As of 03/2021 Document History: Version Date Name Changes 1.0 2019-09-03 WM Created 1.1 2020-10-19 WM extended description of internal connector X1 1.2 2021-03-09 WM HA3B3-S added 1.3 2021-07-16 WM pin-assignment X1 IOAV_BP, VOAV_BP corrected

hivolt.de GmbH & Co. KG

Oehleckerring 40 ∙ D-22419 Hamburg ∙ Germany ∙  +49 40 537122-0 ∙  +49 40 537122-99 ∙ [email protected] ∙ www.hivolt.de 7 Declaration of Conformity We declare under sole responsibility that the products Device: High Voltage Amplifiers Series: HA2B5-S, HA3B3-S Subrack HAR42-4 are in accordance with the following European directives: Low Voltage Directive 2014/35/EU EMC Directive 2014/30/EU and comply with the following European standards: EN 61010-1:2010 Safety requirements for electrical equipment for measurement, control, and laboratory use - Part 1: General requirements EN 61326-1:2013 Electrical equipment for measurement, control and laboratory use - EMC requirements - Part 1: General requirements Classification: Group 1, Class B Manufacturer: hivolt.de GmbH & Co. KG Oehleckerring 40 D-22419 Hamburg 2021-03-09 Wulf Müller Managing Director