INNOVATEST DSAS Manual 

Specifications

Brand: INNOVATEST
Category: Test Equipment
Model: DSAS (Manual )
File Type: PDF
File Size: —
Pages: 12

Read Online

Read the INNOVATEST DSAS Manual  online using the built-in PDF viewer.

Text of the manual

Changes in products and/or product specifications

can emerge due to new technologies and continuous development. We reserve the right to change or modify specifications of products without prior notice. We recommend you to contact our sales office for up-to-date information. © All rights reserved

Using the supplied aluminium reference block

2 2. The display 3 2.1 Set mode 4 2.2 Reference mode 4 2.3 Freeze display 4 2.4 Preset mode 5 3. Battery changing 6 4. Data transmission 6 5. IP65 protection 6 6. Warranty 6 7. Bench stand 7 Manual DSAS/DSDS 1

The test specimen should be at least 6mm (0.25”) in thickness for the Shore A models and

at least 3mm (0.12”) in thickness for the Shore D models. b. The lateral dimensions of the specimen should be sufficient to permit measurements at least 12mm (0.5”) from any edge. c. The surface of the specimen should be flat and parallel over a sufficient area to permit thepresser foot to contact the specimen over an area having a radius of at least 6mm (0.25”)from the indentor point. d. A specimen may be composed of plyed pieces to obtain the necessary thickness, but determinations made on such specimens may not agree with those made on solid specimens because the surface faces between plyes may not be in complete contact 1.1 Testing procedure a. Ensure the tester is set to zero before commencing any tests. b. Change to REF I mode when using the durometer. c. Place the specimen on a hard, horizontal surface. d. Hold the durometer in a vertical position with the point of the indentor a least 12mm (0.5”) from any edge. e. Apply the presser foot to the specimen as rapidly as possible, without shock, keeping the foot parallel to the surface of the specimen. Apply just sufficient pressure to obtain firm contactbetween presser foot and specimen. f. The hardness should be read off 1 second after the presser foot of the durometer has made contact. g. In case of the specimen exhibiting a marked flow propensity, the reading may also be taken after 15 seconds. 1.2 Using the supplied aluminium reference block To check the tester has not been damaged simply place the indentor of the tester in the slot of the test block. Apply gentle pressure keeping the tester vertical, then simply read off the value obtained. This should be within ±2 units. Please note: INNOVATEST Europe BV recommend that the unit is returned annually for re-calibration. 2 Manual DSAS/DSDS Factory set calibration screw Important do not adjust

Button pressed less than 1 second

- Button pressed more than 1 second - To switch on press SET briefly. CAL is displayed. Press SET briefly to activate electronics. To switch off press SET and hold. Automatic sleep condition occurs after 4 hours (display off), press MODE or SET to wake up electronics. - There are three electronic functions available: SET, REF I or PRESET To change between each mode, press MODE and hold Manual DSAS/DSDS 3

To zero the display press SET (000.0 must be entered in the preset).

If a calibration value is entered in the preset, pressing SET will display this value. 2.2 Reference mode Press and hold MODE until REF I is displayed. 2.3 Freeze display To freeze the display press SET briefly and HOLD will appear. To unfreeze the display, press SET briefly again. 4 Manual DSAS/DSDS

Press and hold MODE until PRESET appears

Press MODE to move cursor along. Press SET to enter value or zero digit. Press MODE and hold to store value and return to SET. Manual DSAS/DSDS 5

If display shows a B, the battery requires changing.

Insert battery as shown. Ensure correct polarity. Lithium 3V battery, part no. 2032 (20mm diameter, 3.2mm thick) Note: Removing the battery will provoke a general reset of the instrument. 4. DATA TRANSMISSION Remove battery to connect Power Lead/Opto cable. Pressing SET, sends a data transmission. 5. IP65 PROTECTION Please note the IP65 protection rating for the instrument contained in these instructions applies to the electronics only. 6. WARRANTY Except for battery exchange only the manufacturer has the ability to dismount the electronics. Any tampering will invalidate warranty. 6 Manual DSAS/DSDS

Using the tester with the Shore bench stand.

The tester used in conjunction with the Shore Bench Stand will give better reproducibility and accuracy when being used by either experienced or inexperienced operator. The Shore Bench Stand comes with 2 different masses, namely a 1kg mass for Shore A and a 4kg mass which is added to the 1kg mass to produce 5kg for Shore D. Flip-off the plastic dust cover from the top of the tester and screw in the adaptor with an M8 thread into the tester. Place the test piece on the anvil of the bench stand and apply pressure to the operating lever, so as to lift the anvil and test piece into the indentor of the tester. Continue to apply pressure to the operating handle, lifting the mass upwards so as to apply the full mass to the tester. The reading can now be taken. Manual DSAS/DSDS 7

Kapcsolat, bővebb információ:

Sidex Trade Kft. 1224 Budapest, Dózsa György út 105. TT Ipartelep (volt Mechanikai Művek) Tel : (+36-1) 424-1705, Mobil: (+36-30) 9349-611 Fax :(+36-1) 424-1704 Web: http://www.sidex.hu – E-mail: [email protected] CORPORATE HEAD OFFICE MANUFACTURING, DISTRIBUTION & SERVICE INNOVATEST Europe BV Borgharenweg 140 6222 AA Maastricht (The Netherlands) Phone: +31 43 3520060 Fax: +31 43 3631168 Email: [email protected] www.innovatest-europe.com IN09-056 V090712