GHM DeltaOHM HD2010UC Operating Manual
Read Online
Read the GHM DeltaOHM HD2010UC Operating Manual online using the built-in PDF viewer.
Text of the manual
Integrating Sound Level Meter
HD2010UC
www.deltaohm.com
English
Keep for future reference.
Companies / Brands of GHM
- 3 -
V4.2
1. Microphone.
2. Preamplifier.
3. Preamplifier or extension cable connector.
4. Symbol showing measurement status: RUN, STOP, PAUSE, RECORDING or HOLD.
5. Keypad LEFT key: in graphic mode, it moves the selected cursor towards lower values.
6. Keypad CURSOR key: in graphic mode, it allows to select one or both of the two cursors.
7. HOLD key: it temporarily stops display updating.
8. MENU key: it activates the different configuration menus of the instrument.
9. REC key (recording): combined with START/STOP/RESET, it activates data recording on
memory (with “Advanced Data Logger” option). When pressed for at least 2 seconds, the dis-
played data can be stored in memory as a single record; alternatively, the Auto-Store mode can
be activated.
10. PAUSE/CONTINUE key: pauses integrated measurements. From PAUSE mode, integrated
measurements can be resumed by pressing the same key. In PAUSE mode, press
START/STOP/RESET to reset measurements.
11. SELECT key: enables modification mode of displayed parameters by selecting them in se-
quence.
12. ENTER key. It confirms entered data or edited parameters.
13. LEFT key: in the menu, it is used when editing parameters with attribute. In graphic mode, it
reduces the vertical scale.
14. M12 connector for multi-standard serial port, RS232C and USB.
15. Auxiliary power supply connector.
16. DC output connector (∅ 2.5mm jack).
17. DOWN key: in the menu, it selects the next line or decreases the selected parameter. In graphic
mode, it increases the vertical scale levels; the graph is shifted downwards.
18. RIGHT key: in the menu, it is used when editing parameters with attribute. In graphic mode it
extends the vertical scale.
19. MODE key: if “Advanced Data Logger” is installed, selects in circular order the instrument’s
different view modes, from the display of 3 channels in numeric format, to the profile (“Data
Logger” option), to the statistics screens (“Advanced Data Logger” option”).
20. UP key: in the menu, it selects the previous line or increases the selected parameter. In graphic
mode, it decreases the vertical scale levels; the graph is shifted upwards.
21. START/STOP/RESET key: when pressed in STOP mode, it starts the measurements (RUN
mode). In RUN mode, it stops the measurements. When pressed in PAUSE mode, it resets the
integrated measurements, such as Leq, SEL, MAX/MIN levels, etc.
22. PROG key: activates the program selection mode.
23. PRINT key: transfers the displayed data to the serial port. When pressed for more than 3 sec-
onds, it enables the continuous printing (Monitor). Monitoring will be stopped by pressing the
key once more.
24. ON/OFF key: turns the instrument on and off.
25. Keypad RIGHT key: in graphic mode, it moves the selected cursor towards higher values.
26. Battery symbol: indicates the battery level. The more the symbol is empty, the more the battery
has run down.
27. LINE: un-weighted input or output connector (∅ 3.5mm jack).
- 4 -
V4.2
CONNECTOR FUNCTION
The instrument is equipped with five connectors: one in front, one to the side and three at the bot-
tom. The figure on page 2 shows:
n. 3 - 8-pole DIN connector for preamplifier or extension cable. The connector, located on the
instrument front face, has a positioning notch and a screw ring nut to ensure proper lock-
ing.
n.14 - M12 connector for RS232C multi-standard serial port and USB. For the connection to a
PC’s RS232 port you have to use the dedicated serial null-modem cable (code
HD2110/RS), fitted with a 9-pole female connector. As alternative the sound level meter
can be connected to a PC USB port by using the dedicated cable (HD2110/USB), fitted
with type A USB connector.
n.15 - Male connector for external power supply (∅ 5.5mm-2.1mm socket). It requires a
9…12Vdc/300mA power supply. The positive (pole) power supply must be connected to
the central pin.
n.16 - Jack (∅ 2.5mm) or the analogue (DC) A weighted output with FAST constant time updat-
ed 8 times/s.
n.27 - Jack (∅ 3.5 mm) for the analogue (LINE) un-weighted output placed on the right side of
the conical part/detail.
- 5 -
V4.2
INTRODUCTION
The HD2010UC is an integrating portable sound level meter performing statistical analyses. The in-
strument has been designed combining price competitiveness and simplicity of use. Attention has
been paid to the possibility of updating the instrument, and the HD2010UC can be integrated, at any
time, with other options like “Advanced Data Logger” to extend its application scope. The firmware
can be upgraded directly by the user using the NoiseStudio program supplied with the instrument.
The HD2010UC conforms to IEC 61672-1 of 2002 and IEC 61672-1 ed. 2.0 of 2013, IEC 60651
and IEC 60804 specifications with class 2 or class 1 tolerances. The acoustic calibrator complies
with IEC 60942 class 2 or class 1 specifications.
The HD2010UC is an integrating sound level meter and analyzer suitable for the following applica-
tions:
•
Assessment of the environmental noise level,
•
Occupational noise measurements at the workplace (according to ISO9612)
•
Measurement at the workplace,
•
Selection of personal protective equipment,
•
Production quality control,
•
Measurement of machine noise.
With the optional “Advanced Data logger” function the sound level meter is able to perform the
following task
•
Environmental noise monitoring,
•
Analysis of sound events,
•
Complete statistical analysis with percentiles calculation from L1 to L99.
Using a HD2010UC you can measure the sound level by programming 3 parameters and freely se-
lecting the frequency weightings and time constants. You can measure parameters such as Leq, SEL
and maximum and minimum sound levels with integration times from 1 second to 99 hours. If an
undesired sound event produces an over-load indication, or simply alter the result of an integration,
its contribution can be excluded using the versatile Back-Erase function.
The measured sound levels can be recorded in the large non-volatile memory in order to be trans-
ferred to a PC using the supplied NoiseStudio program.
For long term noise monitoring needs, it’s possible to increase the HD2010UC storing capacity with
the optional HD2010MC Secure Digital memory card reader. This device is supplied complete with
a 2GB memory card.
As a statistical analyzer, the HD2010UC samples the sound signal 8 times per second with A-
frequency weighting and FAST constant, and it analyzes it statistically in 0.5 dB classes. It is possi-
ble to display up to 3 percentile levels between L1 and L99.
For further analysis, the LINE un-weighted output allows recording the sound signal either on tape
or directly on a PC equipped with a audio acquisition card.
The high speed of the USB interface combined with the flexibility of the RS232 interface allows
quick data transfers from the sound level meter to the PC mass storage, but can also control a mo-
dem or printer. For example, in case of lengthy recordings, you can activate the “Monitor” function.
This function allows to send the displayed data to a PC via the RS232 serial interface, to be directly
stored on the PC mass storage.
The sound level meter can be completely controlled by a PC through the multistandard serial inter-
face (RS232 and USB) by using a special communication protocol. Through the RS232 interface,
the sound level meter can also be connected to a PC via modem.
- 6 -
V4.2
The calibration can be made either using the provided acoustic calibrator (type 1 according to IEC
60942) or the built-in reference generator. The electric calibration employs a special preamplifier
and checks the sensitivity of the measuring channel, microphone included. A protected area in the
non-volatile memory, reserved to factory calibration, is used as a reference in the user’s calibra-
tions, allowing keeping instrument drifts under control and preventing the instrument from wrong
calibrations. The check of the complete sound level meter functionality can be made directly by the
user, on site, thanks to a diagnostic program.
The HD2010UC sound level meter can perform the measurements required to evaluate workers’
noise exposure (ISO 9612/11 and D.Lgs.81/08). The selection of the personal protective equipment
can be carried out, according to UNI EN 458, through comparison of the A and C weighted equiva-
lent levels that can be measured simultaneously (SNR method). The noise source impulsivity is
easily evaluated (according to UNI9432) measuring the equivalent sound pressure level A
weighted with Impulse time constant (LAeqI).
The class 1 HD2010UC sound level meter with the “Advanced Data Logger” option is suitable for
sound level monitoring and acoustic mapping. Using the “Advanced Data Logger” option, it can al-
so perform assessments of the acoustic climate with capture and analysis of sound events. When
measuring traffic noise near airports, railways and roads, the sound level meter can be used as a
multi-parameter sound recorder, combining statistical analysis features. Remote electrical calibra-
tions and diagnostic tests can be executed using its remote control capabilities.
The HD2010UC can be configured in accordance with different customers’ needs: the available op-
tions can be activated on the new instrument, as well as, later on, when requested by the user. The
provided options are:
“Advanced Data Logger” option
Display and recording of the A-weighted sound level profile with FAST time constant, sampled
8 times per second.
Storage of the profiles of 3 programmable parameters, sampled twice per second. Simultaneous-
ly, the profiles of 3 programmable parameters are recorded at 2 samples/second. The HD2010
sound level meter changes into a sound level logger capable of storing 4 parameters for more than
23 hours. Impulsive events can be easily identified thanks to the possibility of analysing the profile
of the A-weighted sound level with FAST, SLOW, and IMPULSE constants.
A second recording mode allows storing 3 parameters programmed together with the average
spectra, at intervals programmable from 1 second up to 99 hours. In this recording mode, you can
store 3 parameters according to intervals of 1 minute for over 80 days using the supplied memory (4
MB expandable to 8 MB). The sound level meter automatically resets all integrated levels at the be-
ginning of each acquisition interval.
Recordings can later be searched in the memory and viewed on the graphic display using the
“replay” function, which reproduces the time trend of the sound trace.
The HD2010UC sound level meter with the “Advanced Data Logger” option has more sound
level analyzer functions.
• Statistical analysis is available graphically in the form of the sound level distribution of prob-
abilities and the percentile levels graph.
• Parallel to the normal sound level profiles recording mode, you can select a report recording,
programmable according to intervals from 1s to 1 hour, with 5 integrated parameters and the
sound level distribution of probabilities.
• It is possible to enable a trigger feature to capture sound events, and false events can be fil-
tered by requiring that the sound level variation be of a certain length. You can record 5 inte-
grated parameters per each event, and the sound level distribution of probabilities being sam-
pled during the event. The event parameters recording does not exclude logging of profiles
and reports. The event trigger feature can also be activated by pressing a key. Thanks to “Ad-
vanced Data Logger” option, the HD2010UC can acquire the noise profile of 3 acoustic de-
- 7 -
V4.2
scriptors in parallel with reports sequences with 5 dedicate parameters and complete statistical
analysis. Moreover, during data logging, the trigger function enables user to identify events
recording 5 acoustic parameters together with statistical descriptors of each event.
• During recording, up to 9 different markers are available in order to signal the occurrence of
particular situations to be considered during the traces analysis phase.
• A timer programs delayed start of capture.
“Memory module” option
This option doubles the memory capacity from 4MB to 8MB. “Advanced Data Logger” option re-
quired.
- 8 -
V4.2
HD2010UC BLOCK DIAGRAMS
The block diagram shows the main elements of the HD2010UC sound level meter.
- 9 -
V4.2
THE MICROPHONE
The UC52 condenser microphone is pre-polarized and has a ½” standard diameter. The fre-
quency response in free field is flat from 20Hz to 16kHz for type 1 sound level meter version
HD2010UC Kit1 (UC52/1). Version used for type 2 sound level meter version is flat from 25Hz to
10KHz.
Pre-polarized condenser microphones are available, optimized for free field measurements,
with ¼ diameter and 2 mV/Pa (MC24E) and 0.25 mV/Pa (MC24EH) sensitivity. With these micro-
phones the frequency response in free field is flat throughout the whole audio range and the maxi-
mum measurable sound levels are 160 dB and 180 dB respectively.
THE OUTDOOR MICROPHONE UNIT HD WME
The microphone unit HD WME is suitable for long term outdoor noise measurements, includ-
ing from fixed unattended station. The unit is adequately protected from rain and wind; the heated
preamplifier allows performing measurements within a wide weather condition range.
The special preamplifier is fitted with a calibration circuit that uses a charge partition technique in
order to calibrate the unit, inclusive of the microphone
capsule. The output driver allows guiding cables up to
10m in length, without important losses.
The “free field” response of the UC52 microphone fitted
with the HD WME, maintains the Class 1 specifications
of the microphone only. Its ease of assembly and disas-
sembly mean the electro-acoustic characteristics periodic
check can be made like using a normal measurement mi-
crophone.
HD WME unit must be positioned in vertical position
to fulfill correctly its function as weather protection.
Used in this way it can measure both community and
airplane noise (90° and 0°). Delta Ohm sound level me-
ters apply spectral corrections to measurements in order
to comply with IEC 61672 type 1 and type 2 tolerances.
Installation and uninstallation procedure is very easy; in
this way periodical acoustic verification (using a calibra-
tor) of the microphone is as simple as for a standard
measurement microphone.
- 10 -
V4.2
The unit is composed of a central body plus the following components:
• HD SAV3: wind-screen (3)
• HD WME1: birds spike (4)
• HD WME2: rain shield (2)
• HD WME3: stainless steel support (1)
• Microphonic capsule with optimized free field correction
• Microphone preamplifier:
HD2010PNE2W: heated preamplifier for UC52 microphone including CTC calibrator
and driver for up to 10m extension cable.
• Extension cable standard 5m length (10m on request).
For detailed information on the use of HD WME unit, please refer to the specific part of this
manual on page. 45; for information on assembling and disassembling the unit please refer to
appendix on page 115.
THE PREAMPLIFIER
The HD2010PNE2 preamplifier amplifies the weak signal provided by the microphone. The
preamplifier has a gain selectable between 0 and 10 dB and is supplied with a charge partition cali-
bration device which allows to measure the frequency response of the whole amplification chain,
microphone included (diagram described on page.49).
The output driver allows transmitting the microphone signal via a cable to a distance up to 10m.
HD2010PNE2 preamplifier together with UC52 microphone can measure signals up to 140 dB.
The following models are available:
• HD2010PNE2: connector for ½” UC52 microphone and driver for up to 10m extension cables.
This preamplifier, supplied with CTC electric calibration device, can be directly plugged into the
HD2010UC sound level meter or connected via extension cable (up to 10mt).
- 11 -
V4.2
• HD2010PNE2W: heated preamplifier with connector for ½” UC52 microphone and driver for
extension cables. This preamplifier, supplied with CTC calibration device, can be connected to
sound level meters using the 5mt supplied extension cable (10m as option).
• HD2010PNE4: preamplifier for MC24E ¼” microphone. Equipped with CTC calibration device
for electric calibration and driver for cable up to 100m. Requires the HDP079A02 microphone
adapter.
• HD2010PNE4H: preamplifier for MC24EH ¼” microphone. Equipped with CTC calibration
device for electric calibration and driver for cable up to 100m. Requires the HDP079A02 micro-
phone adapter.
THE INSTRUMENT
The signal of the preamplifier comes to the instrument receiver and its output is sent to the
LINE connector and to the A/D converter input. The instrument can be set to use the LINE channel
in place of the signal coming from the preamplifier.
The A/D converts the analogue signal into the numeric format at 20 bits. The measurement dynam-
ics, exceeding 140 dB, is divided into 5 ranges using a variable gain amplifier with steps of 10 dB,
between 0 dB and 20 dB, positioned at the input.
The signal of the preamplifier comes to the instrument receiver and its output is sent to the
LINE connector and to the A/D converter input. The instrument can be set to use the LINE channel
in place of the signal coming from the preamplifier.
The A/D converts the analogue signal into the numeric format at 20 bits. The measurement dynam-
ics, exceeding 140 dB, is divided into 5 ranges using a variable gain amplifier with steps of 10 dB,
between 0 dB and 20 dB, positioned at the input.
The levels with (A, C and Z) wideband frequency weighting are calculated in parallel in the
DSP). Peak (C and Z) levels are also calculated. The levels calculated by the DSP are transmitted to
the microprocessor for further processing, ready to be displayed, stored and printed.
The microprocessor controls all the instrument processes: management of the electrical calibrator,
Flash memory, display, keyboard and multi-standard serial interface (RS232 and USB).
The microprocessor also supplies the electrical signal corresponding to the A weighted instant level
with FAST time constant, sent to DC output.
- 12 -
V4.2
DESCRIPTION OF DISPLAY MODES
The HD2010UC measures simultaneously 3 selectable parameters (statistic ones too) and dis-
plays them at a fixed frequency corresponding to 2 measurements/s; simultaneously (with “Ad-
vanced Data Logger” option) measures the A weighted sound pressure level with FAST time con-
stant and displays it at 8 samples/s; as a statistic analyser , it calculates the probability distribution
and the percentile levels (cumulative distribution). To be able to display all these data, the
HD2010UC, provides 4 different display modes as shown in the figures below.
56.9
60.8
79.3
30
110
01:08:25
Tint= s
10
LFp
Leq
Lpk
dBA
dBC
dBC
00:02:05
LFmx
dBA
1/ 2s
10
90
20/
Fig. 1 - SLM
Fig. 2 - Time Profile
Fig. 3 – Probability distribution
Fig. 4 – Percentiles Levels
Press MODE at any time to jump from a screen page to the next one. The display will show:
first the SLM screen with 3 measuring parameters in numeric format and, if the “Advanced Data
Logger” option is enabled, the Profile screen with the A weighted sound pressure level with FAST
time constant (LAFp), the Probability distribution and the Percentile Levels screens. Upon power
on, the sound level meter displays the SLM screen.
Some indications are shown in all modes. They are
(see the figure on the right):
• Measurement status indicator
• Overload/Under-range indicator
• Battery level indicator
The first symbol in the left corner at the top shows
the measurement status of the sound level meter.
• RUN: the instrument is measuring.
• PAUSE: the calculation of integrated measure-
ments and the recording of measurement have
been suspended. Instantaneous parameters are
still being measured and displayed.
• REC: the instrument is measuring and recording.
• STOP: the instrument is not making any measurement.
- 13 -
V4.2
• HOLD: the calculation of integrated measurements has come to the end of the set integration
interval, or HOLD was pressed.
• P (Print): indicates that printing is in progress.
• M (Monitor): indicates (flashing) that continuous data printing has been started.
• R (Replay): appears (flashing) when the “Memory Navigator” program is in use (“Advanced
Data Logger” option required), to view a file saved in the instrument memory (see page.42).
Just on the right of the symbol indicating the logging mode, there is the symbol showing a
possible overload or under-range. An arrow directed upwards indicates that the input level has ex-
ceeded the maximum measurable level, while an arrow directed downwards indicates that the input
level is lower than the minimum measurable level.
The maximum measurable level corresponding to the selected measurement range is given in
the technical specifications (see page 74). The minimum measurable level is 80 dB lower than the
maximum one. The noise levels for each frequency weighting are listed in the technical specifica-
tions. Using an appropriate parameter (MENU >> Instrument >> Measurement >> Overload Level)
you can program the maximum measurable limit at lower levels (chapter on page 74).
An empty arrow indicates that the limit has been exceeded, while a full arrow indicates that
the overload is in progress.
The integration time Tint, programmable between 1s and 99h, is displayed to the right of the
overload indicator. When the integration mode is set on MULT, the “Tint” symbol on the SLM will
flash (see the “DESCRIPTION OF THE DIFFERENT INTEGRATION MODES” chapter on page
25).
In the right corner at the top, there is the battery symbol. The more the symbol is empty, the
more the battery has run down. When the instrument autonomy reaches 10%, corresponding to
about 30 minutes, the battery symbol will start flashing. A protection device prevents the instrument
from making measurements with insufficient battery levels and automatically switches off the in-
strument when the battery level is at the minimum.
The battery level, expressed in percentage, is visible in the menu main screen page and in the
program page; press MENU and PROG to access them. To jump back to the measurement screen,
press MENU and PROG again.
Pressing SELECT, you will select in sequence the parameters relevant to the displayed page.
While the selected parameter flashes, you can change it with the UP and DOWN keys. Press
ENTER to quit the selection mode (automatic exit after 10s).
In graphic display mode, use the UP, DOWN, LEFT and RIGHT keys to change the vertical
scale parameters. The LEFT and RIGHT keys reduce and expand the vertical scale, while the UP
and DOWN keys decrease and increase the levels of the vertical scale; the graph is so shifted up-
wards or downwards, respectively.
- 14 -
V4.2
SLM (SOUND LEVEL METER)
This is the display mode upon power on.
Three parameters (selectable among the following ones) can be displayed simultaneously:
• Instantaneous acoustic parameters such as Lp, Leq(Short) and Lpk. The instantaneous pressure
level is displayed as the maximum level reached every 0.5s.
•
Integrated acoustic parameters with wideband frequency weightings, such as Lpmax, Leq,
LIeq and Lpkmax, updated every 0.5s.
•
Percentile levels selectable between L1 ad L99.
•
Sound exposure level.
•
Daily personal exposure level.
•
Dose and daily Dose with programmable Exchange Rate, Criterion and Threshold Levels.
•
Overload Time (in %).
The display is updated every 0.5 seconds.
The data recording depends on the integration mode selected (single or multi) and on the activation
or not of the auto-store mode. (see the chapter “DESCRIPTION OF THE DIFFERENT
INTEGRATION MODES” on page 25).
Auto-Store: OFF
Auto-Store: ON
Integration
SINGLE
Recording twice a second of the 3 pa-
rameters and 8 times a second of LAF.
Automatic Stop at the end of the set inte-
gration interval.
Automatic recording of SLM page at the
end of the set integration interval.
MULTIPLE
Recording twice a second of the 3 pa-
rameters and 8 times a second of LAF.
Automatic reset of integrated levels at
every integration interval.
Automatic recording of the SLM screen at
intervals equals to the set integration time.
At the beginning of each period, integrated
levels and spectra are set to zero.
DISPLAY DESCRIPTION
Left at the top of the display there are the recording status symbol and the overload or under-
range indicator (described at the beginning of this chapter). In the midst there is the integration in-
terval and on the right the acquisition time (hours:minutes:seconds). When the integration mode is
set on MULT (MENU>>General>>Integr.Mode:MULT), the “Tint” symbol flashes. The battery
symbol is in the right corner, indicating battery level.
56.9
60.8
79.3
30
110
01:08:25
Tint= s
10
LFp
Leq
Lpk
dBA
dBC
dBC
Integration interval
Acquisition time
Maximum level
Minimum level
Displayed parameters
Bar showing
istantaneous level
- 15 -
V4.2
The “analogue bar” shows the unweighted sound pressure instantaneous level in an 80 dB in-
terval.
Three measuring parameters are displayed under the analogue bar. All displayed parameters can be
freely selectable among the available ones. There are no restrictions in the selection of frequency
weightings. Measuring parameters are displayed with a shortened label, followed by the numerical
value, by the unit of measurement, and, when necessary, by the frequency weighting. The corre-
spondence between the label and the effective parameter is to be found on page 99.
Integrated parameters like Leq (and Lmax or Lmin), which imply the accumulation of the sampled
sound levels, are displayed with a series of dashes (----) until the level remains lower than the min-
imum measurable level.
Before starting a new logging, the sound level meter automatically resets all measurements. If
the multiple integration mode is enabled (MENU >> Instrument >> Measurements >> Integration
Mode: MULT), integrated levels will be automatically set to zero at regular intervals equal to the
set integration time Tint.
If the “Advanced Data Logger” option is installed, with the Continuous Recording mode a se-
ries of values is stored every 0.5s together with the parameter displayed in the PROFILE screen,
which corresponds to the A weighted sound pressure level with FAST time constant, computed 8
times a second. Each sample corresponds to the highest sound level (LAFmx) calculated every 0.125s
on the level measured every 7.8ms.
SELECTING PARAMETERS
Some measuring parameters (integration interval, measuring range and the 3 parameters) can
be changed directly via the SLM screen.
Pressing SELECT you choose the different parameters in sequence. While the selected parameter
flashes, you can change it with the UP and DOWN keys.
If a parameter with attribute is selected, like, for example, LFp (FAST weighted pressure lev-
el), the relative frequency weighting will also flash (A in the example in the figure). In this case,
pressing UP and DOWN, you can modify the selected parameter without changing the attribute; for
example, if you press DOWN, you can go from LFp A weighted to LSp A weighted. Pressing
RIGHT you’ll jump to the attribute selection, which will be the only one to flash. Use then the UP
and DOWN keys to change the attribute. For example, if you press UP, you can go from A
weighted LSp to Z weighted LSp.
Pressing SELECT let you choose the next parameter; pressing ENTER, or automatically after
approximately 10s, will let you exit the selection mode.
Also the integration mode (see page 25) can be set using the LEFT and RIGHT keys: press
SELECT to choose the integration interval. When the integration interval numeric value flashes,
press RIGHT to set the multiple integration mode or LEFT to set the single integration mode. When
the integration mode is set on MULT, the “Tint” symbol flashes.
Parameters can be modified only when the instrument is in STOP mode: if you try to
make changes to any of the parameters while the instrument is in a status other than STOP, you will
be asked to stop the measurement in progress: pressing YES will stop recording and will allow you
to go on modifying parameters; pressing NO recording will continue without interruption.
The above settings can be made through the instrument configuration menus. See a detailed descrip-
tion on page.32.
BACK-ERASE FUNCTION (DATA EXCLUSION)
To stop a measurement in progress when recording, press the PAUSE/CONTINUE key.
All data logged until the moment key was pressed are used for calculation of integrated pa-
rameters. However, there are some cases when it is useful to clear the measurements recorded just
- 16 -
V4.2
before pressing PAUSE, for example, because they were caused by unexpected events and not char-
acterizing the sound being examined.
During measurement, press PAUSE/CONTINUE: integrated measurements update will be in-
terrupted. At this point, press the LEFT arrow to delete the last recorded data.
The integration time value will be temporarily replaced by the word ERASE followed by the
time interval (in seconds) to be deleted. Use the LEFT and RIGHT keys to increase or decrease the
erase interval. Displayed integrated parameters change accordingly, allowing choosing the erase
time depending on the effective need. When pressing PAUSE/CONTINUE again, measurement will
start again and the integrated parameters will have been removed from the selected interval.
The erase maximum time, divided into 5 steps, is set from menu: MENU >> Instrument >>
Measurement >> Max Back-Erase. Settable values are: 5, 10, 30 or 60 seconds, with 1s, 2s, 6s or
12s steps, respectively.
- 17 -
V4.2
TIME PROFILE MODE (REQUIRES THE “ADVANCED DATA LOGGER” OPTION)
This display mode, available if “Advanced Data Logger” is installed, presents the time profile
of the A weighted sound pressure level with FAST time constant (LAFp). The integration time is
equal to 1/8s and the last 100 measured samples are displayed.
The HD2010UC sound level meter calculates the sound level 128 times a second and displays the
maximum level at intervals equal to 125ms.
Pressing HOLD, the display update will be stopped; however, the instrument continues meas-
uring and pressing HOLD again will restart display updating.
The HOLD status does not affect neither Monitor (continuous printing) nor recording opera-
tions. When the continuous recording is activated with the single integration mode, the integration
time acts like a timer for data acquisition, stopping automatically the measurement when the time is
elapsed.
This screen-page is not recorded in the Auto-Store mode.
DISPLAY DESCRIPTION
Fig. 5 - Description of the Profile mode display
The Fig. 5 presents the time profile of the A weighted sound pressure level with FAST time
constant.
The integration interval is shown in the left corner at the bottom of the display. Always at the bot-
tom, in the centre, the display shows the measurement unit and the frequency weighting of the
measuring parameter
The amplitude of the vertical scale of the displayed graph corresponds to 5 divisions. The am-
plitude of each division is called “scale factor” of the graph and appears in the middle of the verti-
cal axis. Using the RIGHT (zoom +) and LEFT (zoom -) keys, this parameter is selectable in real
time among 20dB, 10dB or 5dB by division.
Use the UP and DOWN arrows to set the graph full scale with steps equal to the selected scale
factor, starting from the instrument full scale1. In this way, the graph can be shifted UP or DOWN,
depending on the key you have pressed.
An “analogue” bar indicator on the display right side provides the non-weighted instantaneous
level of the input sound pressure level, as for the SLM mode bar.
During Recording mode, 8 values per second of the LAFp level are stored together with 2 val-
ues per second of the SLM screen sound levels. Likewise, when the Monitor function is enabled, 4
values are sent to the interface every 0.5s.
1 The instrument full scale is determined by the selection of the input gain by choosing from the menu: MENU >> In-
strument >> Input gain.
- 18 -
V4.2
The integration mode do not influence this screen recording functioning.
The sound level displayed on this screen can be used as source for the event trigger (see para-
graph “EVENT TRIGGER FUNCTION” on page 23).
USING THE CURSORS
To activate cursors on the graph, press CURSOR on the keypad. If you press CURSOR re-
peatedly, either L1 or L2 cursor, or both ∆L cursors in “tracking” will be activated in succession: the
selected cursor will flash. Use the LEFT and RIGHT arrows on the keypad to move the selected
cursor on the graph.
The second line at the top of the display shows the level of the measuring parameter and the
time indicated by the active cursor or the time interval and the L1-L2 level difference between the
two cursors when they are both active.
The parameter level being lower than the minimum measurable level is indicated by a series
of dashes (- - - -).
Press CURSOR again to disable the cursors.
- 19 -
V4.2
STATISTICAL GRAPHS
The functioning mode with advanced statistical analyser, available with “Advanced Data
logger” installed, allows analyses on the sound pressure level with FAST time constant (sampled 8
times per second) or short equivalent level (integrated every 0.125s) or peak level (calculated twice
per second) with any frequency weighting (only C or Z for peak level).
The statistical analysis is done with 0.5dB classes for sound levels from 21dB to 140dB and
provides graphic display of the sound level distribution of probabilities and percentile levels. The
graphs can be enabled in Menu >> Statistical Analyzer >> Display Statistics. Disabling the displays
does not influence the programmable L1–L4 percentile level calculation.
The following figure shows the level distribution of probabilities on the 6-minute measure-
ment of the noise issued by a climatic room. During measurement an acoustic calibrator was
switched on for about 2 minutes near the microphone.
The distribution of probabilities shows the different “population” of the examined noise clear-
ly. From lower levels, the first peak (about 63dBA) reflects the room background noise caused by
the ventilation system. The second peak (about 65dB) concerns the cooling compressor activation.
The third peak (about 69dB) is the tone issued by the calibrator.
In the following figure the cumulative distribution for the same sample above can be seen.
The cumulative distribution is built from the 100% of the levels under the measured minimum, and
subtracting the probability of each you get 0 for the levels over the measured maximum.
The percentile levels are calculated interpolating the cumulative distribution.
The statistical analyzer resets the classes at the beginning of measurement and, if no report time has
been selected, it will continue accumulating the statistic until the end of the measurement. If a re-
port time is selected (menu Main >> Measurement >> Report time) or the “multi” mode is activat-
ed, the statistic is automatically reset at the beginning of each interval.
- 20 -
V4.2
The HOLD key does not affect this display mode.
When the continuous recording is activated with the single integration mode, the integration
time acts like a timer for data acquisition, stopping automatically the measurement when the time is
elapsed. When the level integration is performed in multiple mode, the statistical graphs are cleared
at the beginning of every interval. These screens are not recorded in the Auto-Store mode, and can
be recorded only as a single datum; therefore they are not available for continuous recording.
These display modes do not have a specific Monitor function. The statistical analysis is sent
to the serial interface, together with other measurements, when the monitor function is enabled in
MEASUREMENT (see paragraph “PRINT AND MONITOR FUNCTIONS” on page 29).
- 21 -
V4.2
Level distribution of probabilities
This function is available with “Advanced Data Logger” installed.
Fig. 6 - Description of the distribution of probabilities display
The figure shows the distribution of probabilities of the A weighted equivalent level with a
0.125s sampling interval. The vertical axis shows the sound levels in decibels and the probabilities
are on the horizontal axis.
The display shows the sampling interval in the left lower corner and the chosen measurement
parameter for the statistical analysis in the first line, to the left of the status indicator, and the possi-
ble overload indicator.
The amplitude of the vertical scale of the displayed graph corresponds to 5 divisions. The am-
plitude of each division is called “scale factor” of the graph and appears in the middle of the vertical
axis. This parameter is selectable in real time among 20dB, 10dB or 5dB by division. These corre-
spond to the 2dB, 1dB or 0.5dB classes in the graph. The scale factor can be set using the RIGHT
(zoom +) and LEFT (zoom -) keys.
Use the UP and DOWN arrows to set the graph full scale with steps equal to the selected scale fac-
tor. In this way, the graph can be shifted UP or DOWN according to the pressed key.
An “analogue” bar indicator on the display right side provides the non-weighted instantaneous
level of the input sound pressure level, as for the SLM mode bar.
The parameter chosen for statistical analysis can be changed without accessing the menus, but
simply using the SELECT keys, the four arrows UP, DOWN, LEFT and RIGHT and ENTER (for
more details, see "Selecting Parameters " on page 15).
Using the Cursors
To activate cursors on the graph, press CURSOR on the keypad. If you press CURSOR re-
peatedly, either L1 or L2 cursor, or both ∆L cursors in “tracking” will be activated in succession:
the selected cursor will flash. Use the LEFT and RIGHT arrows on the keypad to move the selected
cursor on the graph.
The second line at the top of the display shows the level and central frequency of the class and the
relevant probability indicated by the active cursor, or the probability for the levels in the interval be-
tween the two cursors when they are both active.
Press CURSOR again to disable the cursors.
- 22 -
V4.2
Percentile Levels Graph
The graphic display is available for the sound level distribution of probabilities and also for
the percentile levels (“Advanced Data Logger” required).
Fig. 7 - Description of the percentile level display
The figure shows the percentile levels graph corresponding to the distribution of probabilities
shown in the above paragraph.
From the sound level distribution of probabilities you can calculate the cumulative distribu-
tion of probabilities on the same classes. The cumulative distribution is equal to 100% for the clas-
ses with levels under the measured minimum, and 0% for the classes with levels over the measured
maximum. From the minimum measured level class, the cumulative distribution decreases for the
relevant probability of each class until the class corresponding to the maximum measured level,
where it becomes zero. The L1 – L99 percentile levels are calculated interpolating the cumulative
distribution of probabilities.
The vertical axis shows the sound levels in decibels and the percentile index is on the horizon-
tal axis. The display shows the sampling interval in the left lower corner and the chosen measure-
ment parameter for the statistical analysis in the first line, left of the status indicator and the possible
overload indicator.
The amplitude of the vertical scale corresponds to 5 divisions. The amplitude of each division
is called “scale factor” of the graph and appears in the middle of the vertical axis. This parameter is
selectable in real time among 20dB, 10dB or 5dB by division. The scale factor can be set using the
RIGHT (zoom +) and LEFT (zoom -) keys.
Use the UP and DOWN arrows to set the graph full scale with steps equal to the selected scale fac-
tor. In this way, the graph can be shifted UP or DOWN according to the pressed key.
An “analogue” bar indicator on the display right side provides the non-weighted instantaneous
level of the input sound pressure level, as for the SLM mode bar.
The parameter chosen for statistical analysis can be changed without accessing the menus, but
simply using the SELECT keys, the four arrows (UP, DOWN, LEFT and RIGHT) and ENTER (for
more details, see "Selecting Parameters " on page 15).
Using the Cursors
The CURSOR, LEFT and RIGHT keys on the keypad enable and move the cursor.
The second line at the top of the display shows the percentile level indicated by the cursor.
Press CURSOR again to disable the cursor.
- 23 -
V4.2
EVENT TRIGGER FUNCTION
The Event trigger function is available with the “Advanced Data Logger” option and can
be enabled only with the single integration mode. During measurement this function can be used to
isolate a sound event identifiable by sound level variation or by synchronization to an external sig-
nal or, manually, by pressing a key.
The acoustic descriptor used by the trigger function is selected in the PROFILE view (Menu
>> Trigger >> Source: LEV). The level variation that triggers the event can be positive or negative
(Menu >> Trigger >> Trigger Polarity) and the trigger threshold (Menu >> Trigger >> Trigger
Threshold and Menu >> Trigger >> Bottom Threshold) can be different from the deactivation
threshold (Menu >> Trigger >> Trigger Threshold and Menu >> Trigger >> Bottom Threshold).
The following figure shows an example of a positive polarity sound event capture. The sound level
(LAF) exceeds the trigger threshold for time T0 and, later, the bottom threshold for time T2.
Fig. 8 - Description of the event trigger parameters
To prevent short duration pulses being detected as sound events, a minimum trigger duration
can be set up to a maximum of 10s (Menu >> Trigger >> Minimum Duration). If the threshold is
exceeded for less than the set time, the event is neglected. Also a minimum deactivation duration
can be set: when the deactivation threshold is exceeded, the event close is delayed for the set time,
up to a maximum of 255s (Menu >> Trigger >> Stop Delay).
In the example, since the trigger conditions exceed the minimum duration, that is, they persist at
least for time T1, the event levels integration begins, including the 2 seconds before the threshold is
reached (pre-trigger). This pre-trigger time cannot be modified.
The event levels integration ends at time T3, that is, after the stop delay from the T2 time corre-
sponding to the bottom threshold being reached.
The event trigger feature can be activated also by pressing the ENTER key (Menu >> Trigger
>> Source: MAN). In this case the minimum duration parameter has no effect and the event begins
as soon as the trigger is detected.
- 24 -
V4.2
Per each event identified, 2010UC calculates the following:
• 5 programmable selectable parameters: maximum and minimum levels, peak level, equiva-
lent level and SEL.
• Full statistical analysis
These parameters are not displayed but can be stored, completely or partially, at the end of each
level. The menu Recording >> Event allows selection of the 5 parameters and memorization.
The event parameters integration begins 2 seconds before triggering. This pre-trigger time cannot be
modified.
A special printing function, synchronized with the trigger, is available for communication of the
event itself via RS232 (Menu >> Trigger >> Print).
If the trigger is enabled, the sound level profile is displayed with black background when the
trigger is not active so as to highlight the event’s portion.
Fig. 9 - Description of the “Event trigger - display”
- 25 -
V4.2
DESCRIPTION OF THE DIFFERENT INTEGRATION MODES
The 2010UC makes measurements according to two different acquisition modes either single
(standard for 2010UC) or multiple (“Advanced Data Logger” required).
The single integration begins by resetting the integrated levels (e.g. the Leq) and ends when the set
integration time Tint is elapsed or when the acquisition is manually interrupted by the RUN/STOP
key. The following figure shows the Leq Short profile calculated twice per second and the integrat-
ed Leq over a measurement time of 1 minute in single mode.
In the measurement time span, the Leq Short (LAeqS) shows three phases with high noise,
about 80dB, and a variable 52-60dB background noise.
The Leq profile shows that the integration of the three high noise phases gives an equivalent level
stabilizing at about 77dB at the end.
It is often necessary, or convenient, to divide the measurement time in equal duration intervals and
calculate the integrated levels, as Leq, maximum and minimum levels, etc. on each interval sepa-
rately instead of on the entire measurement. The multiple integration mode serves this purpose.
The multiple integration divides the measurement in equal intervals according to the set in-
tegration time (Tint). Each single interval begins by resetting the levels and ends when the set inte-
gration time is elapsed; the integration levels sequence ends when manually interrupted by the
RUN/STOP key.
The following figure shows the Leq Short profile calculated twice per second, and the integrated
Leq over intervals of 20seconds during a measurement time of 1 minute in multiple mode.
The Leq Short profile is similar to that of the previous figure. The Leq Short profile shows the
multiple integration mode divided the measurement time three intervals of 20 seconds, each isolat-
ing the three noise phases. Some markers point out the beginning of each interval, when the Leq
was reset. Three equivalent levels at about 75dB are obtained, corresponding to each interval.
To select the integration mode select the proper parameter (MENU >> Instrument >> Measurement
>> Integration Mode) as SING for single integration or MULT for multiple integration.
To perform the setting directly from the keyboard, press SELECT to choose the integration interval.
When the integration interval numeric value flashes, press RIGHT to set the multiple integration
mode or LEFT to set the single integration mode. When the integration mode is set on MULT, the
“Tint” symbol flashes.
- 26 -
V4.2
SINGLE INTEGRATION
When the integration mode is single, the sound level meter resets the integrated parameters
(i.e. Leq), starts measuring the instant sound levels (i.e. SPL) and calculates the integrated levels
(i.e. Leq, maximum and minimum levels or statistical levels), regularly until acquisition stops.
This mode provides, at the end of the measurement session, the integrated levels over the
whole acquisition period. The parameter “MENU >> Instrument >> Measurement >> Integration
Interval” allows to suspend the display update, when set time is over. Now, to store displayed data
press REC and hold it down for at least 2 seconds, then select the manual storage option. Press
PRINT if you want to send displayed data to the serial output.
While display update is paused (HOLD mode), the sound level meter continues measuring
and calculating the sound levels; press HOLD to let display update start again. If you do not wish to
continue beyond the set integration time, press STOP to interrupt the acquisition.
When the integrating mode is SINGLE and the sound level continuous recording is active,
the acquisition is automatically ended once the set integration time is elapsed.
The PAUSE/CONTINUE key can be used to suspend the calculation of integrated levels tem-
porarily, while instant levels are still being measured. During a pause, and as far as the integrated
levels displayed on the SLM page are concerned, you can delete the last integration seconds through
the “Back-Erase Function (data exclusion) ” see page. 15. The monitor function is not affected by
the acquisition pauses. The continuous recording function suspends data storage during the acquisi-
tion pauses and automatically stores a marker indicating the pause duration and the possible use of
the erase function.
The sound level meter has an additional timer for acquisition over time intervals (MENU >> Gen-
eral >> Measures >> Report). With this function it is possible to split the measurement time over
programmable duration intervals form 1 second up to 1 hour; for each interval it is then possible to
calculate a set of 5 integrated levels choosing from Leq, max, min levels, SEL and statistical levels.
These data are not directly available on the display but can be stored activating continuous record-
ing mode. The report levels can be evaluated directly on the sound level meter display using the
“Navigator” program, selecting the Report mode for the replay.
Interval acquisition can be activated only in “SINGLE” integration mode.
- 27 -
V4.2
MULTIPLE INTEGRATION (REQUIRES THE “ADVANCED DATA LOGGER” OPTION)
When the integration mode is at intervals, the sound level meter automatically makes a con-
tinuous sequence of acquisition intervals having a duration equal to the integration time set through
the parameter “MENU >> Instrument >> Measurements >> Integration interval”. Each acquisition
interval starts clearing the integrated levels and the “Tint” symbol on the SLM page flashes.
This integration mode, together with the Auto-Store function (MENU >> Instrument >> Re-
cording >> Auto-Store) allows to record the SLM parameters at intervals equal to the set integration
time (see “THE RECORD FUNCTION” on page 30). Automatic recording occurs at the end of
each integration interval.
When the Auto-Store function is active, the REC symbol will flash alternating with the indi-
cator of the sound level meter acquisition, up on the left side of display.
“MULTI” integration mode prevents the possibility to use the Report Interval parameter to
record at defined time intervals and over triggered events. Statistical analysis and integrated levels
are zeroed at the beginning of each integration interval.
- 28 -
V4.2
The following table gives the different measurement and storage modes of the HD2010UC.
Integration
Auto-Store
(Advanced Da-
ta Logger op-
tion)
Measurements
Continuous Recording
(Advanced Data Logger
option)
Single Recording
SINGLE
OFF
Press
to start.
The integration ends when
t=T.Int., enters in HOLD
mode, and it is possible to
continue by pressing HOLD
or to stop with
Press
+
to start.
Automatic stop when t =
T.Int
Press
to record the
displayed data.
ON
Press
to start.
Automatic
stop
when
t=T.Int. with recording of the
SLM,.
----
----
MULTIPLE
OFF
Press
to start.
The integration of the sound
levels occurs according to
intervals equal to T.Int. The
integrated levels are cleared
at the beginning of every
interval.
Press
+
to start.
Continuous recording with
LAST marker at the end of
each integration period. Stop
by pressing
.
Press
to record the
displayed data.
ON
Press
to start.
The integration of the sound
levels occurs according to
intervals equal to T.Int. The
integrated levels are cleared
at the beginning of every
interval and recorded at the
end.
----
----
- 29 -
V4.2
PRINT AND MONITOR FUNCTIONS
If you press and soon release the PRINT key, you can send to a PC or to a printer, via
the serial interface, the screen-page displayed when pressing the key, in ASCII format.
The suggested serial printer can be the HD40.1 (please refer to page 66) On the instru-
ment display, a letter “P” replacing the status indicator, highlights data transfer.
If the PRINT key is hold down until the letter M (Monitor function) and the recording status indica-
tor flash alternatively, the displayed screen will be continuously sent to the serial interface: press
PRINT again or STOP to end the operation.
After activating the Monitor function, even if you press the MODE key, the type of screen
sent to the serial interface will not change.
The PRINT function can also be selected starting from the STOP status. In this case the func-
tion will automatically activate as soon as the instrument switches into the RUN mode.
If the instrument gets into the PAUSE mode, the function will remain active, but sent data
will be combined with the “P” symbol indicating the suspension status of the integrated parameter
calculation.
The Monitor function does not interfere with data recording (which needs “advanced data
logger” option) on memory and can be activated simultaneously.
A series of values is sent every 0.5s.
- 30 -
V4.2
THE RECORD FUNCTION
The REC key supervises the function of recording data on the instrument memory.
Two recording modes are available: the single (manual or automatic) and continuous
recording. The automatic (Auto-Store) and continuous recording modes are available
with the “Advanced Data Logger” option”.
MANUAL AND AUTOMATIC SINGLE RECORDING
When only the REC key is pressed for at least 2 seconds, the displayed screen will be record-
ed as a single record. When you press REC and before the instrument stores the active screen, you
will be asked to confirm the recording title containing order date and number. This operation can be
performed in RUN, HOLD, PAUSE and STOP modes. With the “Advanced Data Logger” option,
when the single recording is activated while the instrument is in STOP mode, you will be first asked
to choose between automatic and manual storage.
If the manual recording has been selected, the instrument will behave as per the above description
(Single record recording).
If, on the contrary, the automatic recording mode “AUTO” (“Advanced Data Logger” required)
has been selected, the sound level meter is set for Auto-Store mode. The parameter “MENU >> Re-
cording >> Auto-Store” will be activated and the REC symbol blinks over the status indicator.
Press START to begin automatic recording: as soon as the measurement time reaches the set
integration time, the parameters shown on the SLM screen are recorded automatically. If the inte-
gration mode is single, the acquisition will be stopped. If it is multiple, on the SLM a new integra-
tion and recording cycle will start automatically after all parameters are reset.
Press REC while the instrument is in STOP mode, to disable the Auto-Store function and
jump back to the continuous integration mode.
The automatic recording can be also activated from the corresponding menu item (MENU >> Re-
cording >> Auto-Store).
If the Auto-Store function is enabled in single integration mode, the parameters displayed on
the SLM screen will be stored when the set integration time is over. Then the capture will be auto-
matically stopped.
When the parameter “MENU >> General >> Measurement >> Integr.Mode” is set to MULT
mode (the symbol “Tint” in the SLM screen starts to blink), then the acquisition is repeated over
time intervals corresponding to the set integration time; each time interval is automatically preced-
- 31 -
V4.2
ed by the reset to zero of integrated levels (ref. to chapter “DESCRIPTION OF THE DIFFERENT
INTEGRATION MODES” ).
This integration mode, together with Auto-Store function, enables to record, over a time peri-
od equal to the integration time, the parameters available on the SLM mode screen. The integration
time (equal to recording time period) is programmable on the SLM mode screen or, alternatively,
using the specific parameter on the menu (MENU >> General >> Measurements >> Integration In-
terval).
CONTINUOUS RECORDING
If the “Advanced Data Logger” option is installed and the keys REC and
START/STOP/RESET are pressed simultaneously, the continuous recording of data on memory
will be activated. The 3 parameters of SLM mode are stored twice a second, while the A weighted
sound level with FAST constant is stored 8 times a second.
Press PAUSE to stop recording temporarily; press CONTINUE to restart it. As soon as you
go back to the RUN status, a special record is stored, containing indications about the possible erase
(see the “Back-Erase Function” in SLM mode on page 15) besides date and time.
The HOLD key does not affect data recording.
When the continuous recording is active, if the integration mode is set to single, the integra-
tion time acts like a timer stopping automatically the acquisition as soon as the set time is elapsed.
Integration time is programmable from SLM display or using the relative menu item (Menu >> In-
strument >> Measurement >> Integration interval).
When integration mode is set to multiple, continuous recording do not stop when integration
time elapses: a special marker (“Last”) is stored together with the last data just before the integrated
levels clearing at the beginning of each integration period (see the chapter “DESCRIPTION OF
THE DIFFERENT INTEGRATION MODES” on page 25). The statistical levels and also the inte-
grated parameters displayed on the SLM screen are cleared.
- 32 -
V4.2
CONTINUOUS RECORDING OF REPORT AND EVENTS GROUPS
With the Advanced Data Logger option, by performing the measurements in single integra-
tion mode, you can also record reports and events. The parameters of the SLM and PROFILE
screens are included in the Measurement group. Together with the Measurement group recording,
you can also enable the Report and Event groups recording.
The Report and Event groups are composed of the following storable parameters:
•
5 integrated parameters
•
Statistics
The Report group is recorded at programmable intervals, using the parameter MENU >> In-
strument >> Measurement >> Report Time, from a minimum of 1s to a maximum of 1 hour. The 5
integrated parameters, the spectra and the statistics are automatically cleared at the beginning of
every report time.
The 5 reported parameters can include:
•
FAST, SLOW and IMPULSE weighted maximum and minimum levels
•
Peak level
•
Equivalent level
•
SEL
•
Preset percentile levels L1, L2, L3 and L4
The enabling of the report time is an alternative to the multiple integration mode. The report time
can be set only in single integration mode.
The Event group is recorded, per each detected event (see paragraph “EVENT TRIGGER
FUNCTION ” on page 23), at the end of the event itself. The 5 integrated parameters, the spectra
and the statistics are automatically cleared at the beginning, and are integrated for the entire dura-
tion of the event. The 5 event parameters can include:
•
FAST, SLOW and IMPULSE weighted maximum and minimum levels
•
Peak level
•
Equivalent level
•
SEL
The event trigger feature can be activated only in single integration mode. Therefore it is not
possible to enable event recording while multiple integration mode is selected.
When the Measurement group recording is activated together with the Report and Event
groups recording, the Measurement group continuous recording is enabled only with the events
sensed by the event trigger. This allows a lot of memory to be saved, minimizing information loss-
es: during the events the maximum quantity of logged information occurs, while with outside sound
events, the recording of the level is also carried out according to a reduced time resolution, as de-
fined in MENU >> Instrument >> Measurement >> Report time.
Fig. 10 shows the recording flow for the Report and Event groups.
The Measurement group recording interval is equal to 2 recordings a second.
In the following example, also the Report group items are memorized with a Report time amounting
to 10s.
- 33 -
V4.2
Fig. 10
When the trigger function detects an event, identified by the overcoming of the trigger thresh-
old, or when ENTER is pressed, a time marker is recorded.
Similarly, when the end of event conditions are sensed, as identified by the deactivation
threshold being reached, or by the external TRGIN signal, or when ENTER is pressed, and after the
set stop delay has elapsed, another time marker is recorded. After the time marker, when the event
is closed, the record containing the Event group information is logged.
When the event trigger uses the Profile view sound level as source (Menu >> Trigger >>
Source: LEV), the event data are recorded only when the trigger threshold (Menu >> Trigger >>
Trigger Threshold) exceeds the minimum duration time (Menu >> Trigger >> Minimum Duration).
The following figure shows the recording flow for the Measurement, Report and Event groups. The
Measurement group recording is enable only during the event; outside the events, only Reports are
recorded. In the following example a report is recorded every 10s.
Fig. 11 - Recording flow for the Measurement, Report and Event groups
When the trigger threshold is exceeded for a shorter period than the minimum duration time, the
record containing the Event group information is not logged.
- 34 -
V4.2
Delayed acquisition timer
A timer is available to enable data acquisition according to a programmable delay of up to 99
hours. To perform an acquisition with delayed start, the recording parameters need to be set first
and then the delayed acquisition timer programmed in the parameter Menu >> Sequencer >> Tim-
er.
After programming the sound level meter, you only need to press the REC and RUN keys together
(as if starting a logged measurement): confirm with the “YES” key. The instrument switches into
stand-by and turns off (see. Fig. 12).
Fig. 12 - Delayed acquisition timer warning screen
The instrument will turn back about one minute before the set delay, to allow execution of the
warm up time before starting the automatic acquisition. During this waiting minute, the “TIMER”
message blinks, showing the automatic acquisition feature has been activated.
Fig. 13 - Waiting for the timer star
The acquisition will end when the set integration time (Tint) has elapsed and the instrument will
automatically turn off after disabling the timer.
- 35 -
V4.2
DESCRIPTION OF THE MENU FUNCTIONS
The menu collects all of the parameters through which the instrument functions are set.
The menu can be accessed even when the instrument is measuring, but parameters can be modified
only if the instrument is in STOP mode. If this is not the case, a message will invite you to stop the
current measurement: “WARNING! Stop the measurement to continue".
SELECT PARAMETER
WARNING!
Stop measurement
to continue.
YES
NO
Press YES, and you will be allowed to change the selected parameter.
Some of the parameters listed in the menu can also be modified directly from the measurement dis-
plays: see the chapter concerning the different display modes (from page 12).
The menu is nest-structured, that is, it is organised in menus and submenus. To select a menu item,
use the UP and DOWN arrows until the selected item flashes. When the parameter next to a menu
item does not flash, it means that the item cannot be changed.
This parameter can be changed
This parameter CANNOT be changed
Press SELECT to access the selected submenu or to modify the selected parameter.
Use the UP and DOWN arrows to edit the flashing selected parameter: press ENTER to confirm the
new value, press MENU to cancel the input changes.
- 36 -
V4.2
Press MENU to exit a menu and return to the upper level until you get the measurement display
again.
When you access the menus, current date and time are displayed, as well as the battery charge level
and the available memory.
If you are in a submenu, the "SELECT MENU" item becomes "SELECT SUB-MENU".
The dots at the end of a list mean that there are other items following the visible ones: press the
DOWN arrow to display them.
INSTRUMENT
The Instrument menu includes all data relevant to the instrument identification, as well as
some general parameters of the instrument itself, input and output settings and measurement global
parameters. It consists of four submenus, as described below.
IDENTIFICATION
It includes the information that identifies the instrument and the microphone. These are all items
that cannot be changed by the user.
•
Instrument: instrument name.
•
Serial N.: instrument serial number.
•
Version: current firmware version loaded.
•
Microphone: supplied Free Filed microphone model is UC52.
•
Mic. S.N.: microphone serial number.
•
Mic. Response: type of microphone response. FF stands for Free Field, DF for Diffused
Field).
•
Class IEC 61672: Class of tolerance according to IEC 61672
•
Memory: memory available on the instrument. The standard is 2Mbyte.
•
Options: firmware options.
•
Ext. range: always OFF.
SYSTEM
It allows configuring some system parameters.
• Time: current time.
• Date: current date expressed as year/month/day.
•
Display contrast: allows adjusting the display contrast. When the ambient temperature
changes, the display contrast slightly changes: it can be adjusted entering a higher value to
increase the contrast or a lower one to decrease it. The value can be set between 3 (mini-
mum) and 9 (maximum).
- 37 -
V4.2
•
Auto-Power-Off: the instrument is provided with the auto-power-off function that auto-
matically switches it off after 5 minutes the instrument is in STOP mode and no key has
been pressed. Before switching off, the instrument will make a series of beep: press any
key to prevent the instrument shutting off. The function is active if this menu item is "ON".
If you set Auto-Power-Off = OFF, the instrument will not automatically shut off. In this
case the battery symbol will flash even if batteries are fully charged.
INPUT/OUTPUT
Submenu for the choice of parameters relevant to the instrument inputs and outputs.
•
Baud Rate: Baud Rate: this parameter and the next ones define the properties of the serial
connection. Selectable baud rate values go from a minimum of 300 to a maximum of
230400 baud. The higher is the value, the faster is the connection; therefore, when possible,
it is suggested to select the highest available value to speed up data transfer as much as pos-
sible. If the instrument is connected to a printer with RS232 serial input or with a seri-
al/parallel converter, the value to be set is the one provided by the printer manufacturer.
WARNING: When you use the serial interface, the communication between instru-
ment and computer (or printer with serial input) works only if the instrument and PC
(or device) baud rates are the same. Take care of this note when using data transfer
programs requiring a manual configuration of the serial port parameters, such as, i.e.
HyperTerminal. The NoiseStudio program, combined with the HD2010UC automati-
cally sets the serial port, so that no action by the operator is required.
•
RS232 Bits: (non-modifiable parameter) number of bits of which transferred data are made
up, the value is 8.
•
RS232 Stop Bits: (non-modifiable parameter) stop bit, the value is 2.
•
RS232 Parity: (non-modifiable parameter) parity bit, the value is: none (OFF).
•
Serial Device: identifies the device connected to the serial port.
The connection possibilities are:
PRINTER: connection to a printer with RS232 input
RS232:
connection to a PC equipped with RS232 port (physical)
MODEM: connection to a modem with RS232 input RS232 (see Connection to a mo-
dem on page 66)
USB:
connection to a PC by USB port (see “CONNECTION TO A PC WITH
USB INTERFACE” on page 68)
MC:
connection to optional module HD2010MC for SD memory card recording
(see paragraph on page 71)
MEASUREMENTS
The Measurement item includes the acquisition general parameters.
•
Input Gain: with Input Gain = 0 the measuring range upper limit equals 140dB, and as
much as the input gain increases the maximum measurable level decreases correspondingly
(see page 74). Select the proper gain according to the level of sound to be.
•
Integration interval: Once this time has been reached, the instrument automatically
switches to HOLD, interrupting display update. It can be set from a minimum of 1s to a
maximum of 99 hours. When continuous recording is activated, the integration time acts
like a timer for data acquisition, stopping automatically the measurement when the time is
elapsed. If set to 0s, the timer is disabled and the integration becomes continuous.
•
Report Time: report parameters are integrated over intervals corresponding to the set time.
At the beginning of every interval report parameters are automatically cleared. Report time
can be set to: 1, 2, 5, 10, 20 and 30 seconds, 1, 2, 5, 10, 20, 30 minutes and 1 hour. A re-
- 38 -
V4.2
port time can be set only for the single integration mode (see parameter “Integration
mode”). This parameter needs the “Advanced Data Logger” option”.
•
Max Back-Erase: maximum erase interval of data recorded in SLM mode. The available
values are: 5s, 10s, 30s and 60s: the erase interval is settable with 1s, 2s, 5s or 10s steps,
respectively. See description on page.15.
•
Exchange Rate: is used together with "DOSE Threshold" and "DOSE Criterion" in DOSE
calculation. It represents the variation of the sound pressure level corresponding to the
double or the half of the exposure maximum time with the same Criterion Level (indicated
as “DOSE Criterion”). Its value can be equal to 3dB, 4dB or 5dB.
•
DOSE Threshold: it is the noise level below which the DOSE is not increased. The value
can be set in the 0dB÷140dB interval, with 1dB steps.
•
DOSE Criterion: it is the noise level providing a DOSE equal to 100% after an 8 hour ex-
posure. The value can be set in the 60dB÷140dB interval, with 1dB steps.
•
Overload Level: according to the selected input gain, the instrument provides two measur-
ing ranges: the former with a 130dB upper limit, the latter with a 140dB one. If the sound
level exceeds the 131dB and 141dB limits, respectively, the display will highlight the over-
load indication (∆ and Λ). The indication can be activated also at lower input levels, pro-
gramming this parameter from 20dB, minimum, to 200dB maximum, with 1dB steps. The
shown level defines the overload threshold when the input gain corresponds to 0dB (Input
Gain). The overload threshold automatically scales with the input gain.
•
Integration mode: Two integration modes are available: the former is single (SING) and
the latter is multiple (MULT). For the description of the two modes see chapter
“DESCRIPTION OF THE DIFFERENT INTEGRATION MODES”. The multiple integra-
tion mode requires an Integration interval different from zero be set, and disables Report
Time and Event Trigger.
•
Lev. 1, 2, 3 and 4 Percentile: in the statistical analysis of sound events, LN percentile lev-
els are defined as the levels of noise exceeded for N time percentage throughout the whole
measurement interval. For example, L1 represents the sound level exceeded by 1% of the
measurement time. These items define 4 percentile levels selectable between 1% and 99%
with 1% steps. The corresponding variables are shown in the SLM view as L1, L2, L3 and
L4, combined with the respective percentage.
SOUND LEVEL METER
The Sound Level Meter menu features all parameters relevant to the SLM display mode. These
items can be changed directly from the respective display as described in "Selecting Parameters "
on page15
The first three items of the menu, from Par. 1 to Par. 3, define the three measuring parameters
and the respective frequency weightings relative to the SLM display mode.
Press RIGHT to change the time weighting of measuring parameters, when selected. When the
time weighting flashes, use the UP and DOWN arrows to change it.
The list of available parameters is shown on appendix A1. HD2010UC MEASURING
PARAMETERSon page 99
STATISTICAL ANALYZER
The Statistical Analyzer menu (needs the “Advanced Data Logger” option) collects the specific pa-
rameters related to the statistical graphs display modes. These items can be changed directly in their
screens.
• Param.: the parameter used for selected statistical calculations: LFp, Leq and Lpk with A, C
and Z (only C and Z for Lpk) weightings. The sampling frequency is equal to 8 samples/s (on-
ly 2 samples/s for Lpk).
- 39 -
V4.2
• Class width: statistical analysis is performed by 0.5dB classes.
• Display Stat.: enables (ON) or disables (OFF) displaying of the distribution of probabilities
and percentile levels graph.
TRIGGER
The Trigger menu (needs the “Advanced Data Logger” option) collects the specific parameters re-
lated to the event trigger.
• Source: the trigger source can be chosen among: profile view sound level (LEV), ENTER key
(MAN),.
• Trigger Threshold: the trigger threshold on the profile view sound level (LEV); it can by
programmed with 1dB steps.
• Trigger Threshold: the trigger threshold on the profile view sound level (LEV); it can by
programmed with 1dB steps.
• Bottom Threshold: the deactivation threshold, different than the trigger one, on the profile
view sound level (LEV); it can by programmed with 1dB steps.
• Trigger Polarity: it is possible to choose increasing (POS) or decreasing (NEG) levels for the
trigger on the profile view sound level (LEV). For the increasing levels trigger, the Trigger
Threshold will be higher than the Bottom Threshold, for the decreasing levels trigger the
Trigger Threshold will be lower than the Bottom Threshold.
• Minimum Duration: a duration filter is available to eliminate false triggers. It is activated
upon detection of the event, only if the trigger condition persists for a number of seconds at
least equal to this parameter. It is used only if the Source parameter is set to LEV.
• Stop Delay: when the trigger conditions are not present anymore, the event ends after a num-
ber of second equal to this parameter has elapsed.
• Printing: printing of a warning string (TAG) can be enabled through the serial interface for
each event.
RECORDING
In the Recording menu (“Advanced Data Logger” required) the parameters relating to the logging
of the measured data can be found. It collects the settings concerning the recording of the sound
levels measured in each screen, the report, and event parameters. If no parameter is enabled for re-
cording (all OFF), the instrument warns of the impossibility to record.
When the “Advanced Data Logger” option is enabled, the recording parameters are divided in three
sub-menus: Measurement, Report, and Event. Without this option only the Measurement sub-menu
can be accessed. For details please see chapters “THE RECORD FUNCTION” on page 30 and
“CONTINUOUS RECORDING OF REPORT AND EVENTS GROUPS” on page 32.
MEASUREMENTS
In this menu continuous recording is defined, (requires the “Advanced Data Logger” option
as described in “THE RECORD FUNCTION” on page 30).
• Auto-Store: activates the auto-recording mode of SLM screens, as described in “THE
RECORD FUNCTION” on page 30. By enabling this feature report time and event trigger are
disabled.
• SLM + PROFILE: activates the continuous recording of parameters of the SLM screen and
PROFILE.
- 40 -
V4.2
If the Auto-Store function is activated, SLM screens will be automatically stored when the set in-
tegration time is over. When this function is activated with the multiple integration mode, re-
cording is performed automatically according to the set integration time. Integration time is pro-
grammable from SLM display or using the relative menu item (Menu >> Instrument >> Meas-
urement >> Integration interval). The enabling of the recording mode is indicated by a flashing
REC over the status indicator. Recording is started by pressing RUN. To disable the Auto-Store
function press the REC key briefly.
REPORT
In this menu the recording of the reports is defined, (it requires the “Advanced Data Logger”
option). Each item can be enabled separately, as with the measurement recording. In order to
avoid occupying memory space uselessly, we recommend to enable only the needed items and
set the others to OFF. The integration interval (recording time) of the reports is programmable
using the relative menu item (Menu >> Instrument >> Measurement >> Report Time).
• Par.1 – Par.5: define five integrated parameters, with relevant frequency weightings.
• Parameters: enables recording of the 5 parameters Par.1 – Par.5, defined previously.
• Statistics: activates the recording of the statistics.
EVENT
In this menu (it requires the “Advanced Data Logger” option) the recording of the event re-
ports is defined, as described in the paragraph THE RECORD FUNCTION”.
Each item can be enabled separately, as with the measurement recording. In order to avoid occu-
pying memory space uselessly, we recommend to enable only the needed items and set the others
to OFF.
• Par. 1 Par. 5: define five integrated event parameters, with relevant frequency weightings.
• Parameters: enables recording of the 5 parameters Par.1 – Par.5, defined previously.
• Statistics: activates the recording of the event statistics
CALIBRATION
• Calibration Level: Sound level of the reference sound source used for the sound level meter
calibration. Allowed values vary from 90.0dB to 130.0dB with a 0.1dB resolution.
• Microphone Response: allows selecting the type of frequency response of the microphone
according to the sound field. The standard setting for the standard microphone (UC52) is
“Free Field” (FF), since it has a frequency response optimized for the “free field”. Set the pa-
rameter to “Random Incidence” (RI) to activate the correction for random incidence sound.
This setting is necessary to carry out measurements according to ANSI standards. The random
incidence correction is not available on the HD2010UC Class 2.
When HD WME microphone rain protection is used, for aircraft noise (0° incidence) meas-
urements the FF setting must be used while RI setting must be used for community noise (90°
incidence) measurements (e.g. road traffic noise).
• Wind Shield Correction: Allows correcting the frequency response of the sound level meter
when the HD SAV windshield provided with the sound level meter, or the HD WME outdoor
shield, is used. When this parameter is set to SAV or WME, the sound level meter frequency
response is corrected for the windshield or the outdoor shield presence, respectively. The ran-
dom incidence correction is not available on the HD2010UC Class 2. For additional details on
applied microphone correction, please see UC2 microphone manual.
- 41 -
V4.2
SEQUENCER
The Sequencer menu needs the “Advanced Data Logger” option.
• Timer: programmable acquisition delay in seconds, minutes, or hours up to a maximum of 99
hours (see paragraph “Delayed acquisition timer ” on page 34).
- 42 -
V4.2
PROGRAMS
The item PROGRAMS (PROG key) includes the following functions:
• Navigator: to display stored data (“Advance Data Logger” required),
• Electric and Acoustic Calibration
• Diagnostic Check: diagnostic test of instrument,
• Data copy on MC: this program allows to copy measurements stored on sound level meter’s
flash memory onto an external Secure Digital memory card (please refer to related chapter on
page 73).
Each program is here below described in details.
NAVIGATOR
This program, available with the “Advanced Data Logger” option, allows to access the data
stored on the instrument, display and print them, with no need to download them to a PC. It works
both with single session data and with multiple ones. Press: PROG >> Memory Navigator >>
SELECT to access it. The following screen will appear:
2003/01/01 10 00 00
:
:
Batt 95% Mem 92.5%
:
:
EXIT
CLEAR
SELECT
If you press CLEAR, the content of the instrument memory will be cleared. The user will be asked a
confirmation before clearing.
Press SEARCH to access the first session of stored data on memory.
- 43 -
V4.2
2003/01/01 10 00 00
:
:
Batt 95% Mem 92.5%
:
:
EXIT
LOAD
NEXT
The following data are indicated for each file: the type (single, multiple or automatic), the
progressive number given by the instrument upon logging and the date. Press NEXT to jump to the
next file, press LOAD to download the current file.
If you press LOAD, the instrument goes back to standard display and the STOP and battery sym-
bols alternate with the R (Replay) and P (Program) letters respectively).
R
P
56.9
60.8
79.3
30
110
01:08:25
Tint= s
10
LFp
Leq
Lpk
dBA
dBC
dBC
Four types of files are recorded:
• Sing. Single screen in manual recording
• Auto SLM screen in continuous recording (with “Advanced Data Logger” option”)
• Mult. Multiple screens in continuous recording (with “Advanced Data Logger” option”)
File type: “Sing”
Relative data are loaded and displayed automatically pressing the LOAD key
File type “Auto”
Press START to see the data session: the screens of the measurement session will be displayed in
the acquisition order. During the replay use the PAUSE/CONTINUE key to stop and restart the re-
play, or press STOP to finish. At the end, the instrument will position on STOP. A single screen
page can be sent at any moment to the serial interface.
While the replay is in pause, press START if you want to display the next data. If, while in pause,
the START key is hold down, the replay will be forwarded fast.
File type “Mult.”
Press START to see the data session: if no reports and/or events were recorded, the measurement
session screens will be displayed, as for the “Auto” files. If, in addition to measurements, reports
and/or events were also recorded (the “Advanced Data Logger” option is required), an intermediate
page is displayed allowing the display of measures, reports, or events to be chosen (see chapter
CONTINUOUS RECORDING OF REPORT AND EVENTS GROUPS on page 32).
- 44 -
V4.2
When replaying reports and events, in the SLM screen, reports and events parameters will be dis-
played respectively. When events are displayed they are shown one by one, including automatically
a pause between the event and the next one. During pause, pressing the START key, allows to load
the data of the following event and PAUSE to restart replay.
The replay of measurements associated with event recording, automatically enables the pauses at
the beginning and end of each event trigger. These pauses correspond to the time markers recorded
when the trigger spots the event and at the end when saving the relevant data.
If the recording contains measurements, reports, and events, the measurements are not registered
continuously but only in coincidence of the sound events detected (see EVENT TRIGGER
FUNCTION on page 23).
Disabling of measurements outside the events, in combination with event and report record-
ing, allows two different recording speeds to be maintained, slow and fast, associated with reports
and measurements respectively. The maximum recording resolution is used only during events by
enabling the Measurement group parameters recording (see CONTINUOUS RECORDING OF
REPORT AND EVENTS GROUPS on page 32) while for the other acquisition elements only the
Report group parameters are recorded, using a lower time resolution.
When the simultaneous recording of measurements, reports, and events is enabled, and the
event trigger uses the Profile view sound level as source (Menu >> Trigger >> Source: LEV), the
measurement recording begins as soon as the sound level exceeds the trigger threshold (Menu >>
Trigger >> Trigger Threshold) without waiting for the minimum duration time (Menu >> Trigger
>> Minimum Duration). The measurement recording ends as soon as the stop delay has elapsed
(Menu >> Trigger >> Stop Delay) after the level exceeded the deactivation threshold (Menu >>
Trigger >> Bottom Threshold).
After having examined a file size, press PROG to jump back to the Memory Navigator menu:
press LOAD to reload the current session, NEXT to display the properties of the next session or
EXIT to quit.
At the end of the list of stored files, “End of Dump” will be displayed. Press REWIND to go back
to the first file of the list.
- 45 -
V4.2
CALIBRATION
Calibration is periodically carried out to ensure the validity of measurements performed by
the sound level meter and to control possible long-term drifts of the measuring chain microphone-
preamplifier-instrument. To ensure that measurements performed by the sound level meter are made
in conditions of stability, the letter “W” will flash over the indicator of the instrument status
throughout the whole period needed to stabilize the microphone polarization, signalling the “warm-
up” period necessary every time the instrument is switched on.
The HD2010UC sound level meter stores all calibration typical parameters with respective date
and time in a reserved area. The calibration types can be:
• Acoustic calibration by means of a 1kHz sound level generator, like HD 9101
• Electric calibration (Capacitive Transducer Calibration) with the possibility to measure the fre-
quency response of the whole measurement chain, microphone included, using the built-in sig-
nal generator.
Calibration is necessary every time that the calibrator level, measured by the sound level meter, de-
viates from the nominal value for more than 0.5dB.
The acoustic calibration includes the capacitive transducer one and, before carrying it out, it is
suggested to ensure that the environment where you are operating is suitable: no sudden sounds, no
vibrations on the surface where the instrument is placed, instrument thermal stability. The electric
calibration allows a quick inspection of the electric parameters of microphone and instru-
ment. The calibration procedure includes the inspection of the microphone polarization.
Calibration programs are in the “PROGRAMS” menu, accessible through the PROG key.
Use the UP and DOWN arrows to select the calibration to be carried out:
Fig. 14
when you press SELECT, the calibration display will appear.
- 46 -
V4.2
The screen shows date and time of the last calibration and the calibrator sound level according to
the configuration of the respective parameter (MENU >> Calibration >> Calibr. Level). Answering
positively to the request to continue, the selected calibration program will start.
Calibration procedures are carried out in automatic mode and the operator is guided by the on-
screen instructions.
The calibration result is displayed when the procedure is over and you can choose whether to con-
firm it or not. If confirmed, the new calibration will be stored.
In order to grant the highest measuring accuracy when the wind shield mounted, select “Wind
Shield Correction” from the CALIBRATION menu to apply a correction to the sound level meter
frequency response and compensate the effects of the HD SAV shield provided with the instrument.
All measurement parameters with wideband frequency weightings are thus automatically adjusted.
In the 2010UC sound level meter, besides the windshield correction, the adjustment for the
acoustic field can be activated too.
With the standard microphone (UC52), which has a frequency response optimized for the “Free
Field”, no correction is applied when the correction is set on “Free Field”. When you set the correc-
tion on “Random Incidence” (RI), the correction for random incidence sound will be activated. This
setting is necessary to perform measurements according to ANSI standards (see the parameter of the
menu: CALIBRATION >> Microphone Response).
The HD2010UC sound level meter is suitable for measurements on site in a temperature range
between –10°C to +50°C, in a static pressure range between 65 kPa and 108 kPa and in a relative
humidity range between 25% and 90%. The microphone itself has defined drift coefficients of
acoustic sensitivity with temperature and static pressure that imply a measurable drift of the Micro-
phone-preamplifier-instrument chain, even though within the limits specified for class 1 according
to IEC 61672.
- 47 -
V4.2
PERIODIC CALIBRATION
The periodic calibration of the HD2010UC sound level meter is needed to ensure the traceability to
the laboratory standards and is carried out in accredited laboratories, for example, SIT, etc.
The HD2010UC sound level meter is calibrated at Delta Ohm Acoustic Laboratory (SIT laboratory
N. 124) before being supplied to the user.
“Factory” calibration, which is always made on new instruments and at every periodic calibration
(every 2 years, at least), includes the measurements of the acoustic response in pressure of the mi-
crophone-preamplifier-instrument chain, stored on the sound level meter non-volatile memory, to-
gether with the microphone acoustic sensitivity. Right after the measurement of the acoustic re-
sponse in pressure, also the Capacity Transducer Calibration (sound level meter electrical calibra-
tion including the microphone) is carried out, to be used as a reference for the calibrations made by
the user.
Every time a periodic calibration of the sound level meter is carried out at the factory, calibra-
tion constants are stored as a reference for following comparisons. The factory calibration can be
loaded onto the instrument to make a comparison or to correct a wrong calibration. This procedure
will also load the instrument default parameters so that, is there are any data stored on memory,
these will be deleted.
Follow this procedure to proceed with this operation:
Download any data stored on memory
Ensure that logging is on STOP mode
Remove the external power supply, if connected
Remove one of the batteries: the instrument will obviously switch off (this operation assures
that all the sound level meter internal circuits are discharged)
Press and hold down the ENTER key while inserting the missing battery
The instrument will automatically turn on and will show a warning message indicating that fac-
tory parameters have been loaded
Press CONTINUE to confirm and make an acoustic calibration to store calibration constants. If
this operation is not carried out, or in case the acoustic calibration should give a negative result,
upon the next power on, factory calibration data will be replaced by those stored in the last suc-
cessful calibration.
Electrical signals supplied by a generator connected to the HD2010UC preamplifier via a capac-
itive adapter (replacing the microphone) can be used for periodic tests. The capacitive adapter (spe-
cific for the model of microphone supplied with the instrument) is an accessory that can be supplied
by Delta Ohm S.r.l. Other capacitive adapters can be used, provided that the device equivalent ca-
pacity is included between 15 pF and 33 pF.
During calibration procedures, before performing electrical or acoustical test, it’s necessary to
disable sound level meter spectral corrections
• Menu >> Calibration >> Microphone Resp. Correction >> FF
• Menu >> Calibration >> Shield >> SAV
The frequency response check is made using an acoustic coupler under pressure field conditions
(for example using a B&K4226 multifrequency calibrator) . For additional information on correc-
tions to be applied to frequency response, see specific UC52 microphone user manual.
Microphone capsule maintenance
In order to avoid permanent deterioration of frequency response and subsequently a decay of
the specifications until the unit goes out of type 1 tolerances., it is necessary to prevent that dust
particles and dirt accumulate on the microphone membrane. It is suggested to calibrate the unit
every year at an accredited calibration laboratory.
- 48 -
V4.2
Avoid to use the unit in the presence of steam containing oil, conductive or corrosive substan-
ces. Also humidity condensation on the membrane must be avoided, as it modifies acoustic re-
sponse, causes corrosion and contributes to the formation of residues difficult to remove.
When HD WME weather protection is connected and sound level meter turned ON, power
supply and preamplifier heating avoid condensation of humidity on the microphone membrane.
When the sound level meter is OFF, heating of the preamplifier is disabled and it’s therefore possi-
ble that condensation is formed on the microphone. We advise to put the HD WME unit in a
dry environment when not in use.
- 49 -
V4.2
ELECTRIC CALIBRATION
The electric calibration, using the partition of charge injected into the microphone preamplifi-
er in “charge amplifier” configuration (Capacitive Transducer Calibration), even if it cannot com-
pletely replace the acoustic calibration,
provides however a useful means to keep un-
der control the instrument drifts, microphone
included. The figure on the right shows the
diagram of the CTC technique consisting in
sending an electrical signal to the preamplifi-
er through a high stability capacitor, so that
the output signal does not depend only on
amplification, but also on the microphone ca-
pacity. Many of the microphone malfunctions
reflect in a capacity drift identifiable by
means of this calibration technique. The elec-
tric calibration uses as a reference the result
of the last acoustic calibration, and according
to it corrects any possible drift of the instru-
ment. The electric calibration adjusts the acoustic response of the microphone-sound level meter
chain, both for wideband channels, and for those with a constant percentage bandwidth. In case of
continuous instrument drifts, it is suggested to carry out an acoustic calibration to verify that there
are no other problems both in instrument and microphone.
Operating Procedure
Switch on the sound level meter, if it is the case, press STOP to finish the measurement in progress
and make the following procedure:
1. Press PROG and use the DOWN arrow to select "Electric Calibration".
2. Press SELECT to start the function.
3. The inner signal generator will be turned on and the measured output signal will be com-
pared with the one detected in the last acoustic calibration.
The value that appears on the display (51.3 dBC in the example below), before starting cali-
bration, is the valued measured by the microphone upon pressure of the PROG key, and is
not relate to the calibration in progress.
ELECTRIC CALIBRATION
Last:2003/01/01 10.00
Calibr. level 94.0dB
Leq
dBC
51.3
YES
NO
Run calibration ?
4. Press YES to continue and NO to quit.
5. If you press YES, the calibration is run: wait till the procedure will be over.
Cm
Cc
Cf
V0
Calibration
Signal Source
Transducer
CTC
-A
Charge amplifier
Cb
Vi
V0
Vi
Cc
Cm + Cb+ Cc
= -
Cb
Cf
- 50 -
V4.2
6. At the end, the calibration result will be shown and you will be asked to confirm a new cali-
bration:
ELECTRIC CALIBRATION
Last:2003/01/01 10.00
Leq
dBC
Confirm calibration
∆
-0.1
Calibr. level 94.0dB
?
YES
NO
7. Press YES (keypad LEFT key) to confirm or NO (keypad RIGHT key) to refuse the calibra-
tion just finished. At the end, the instrument will switch to the SLM display, in STOP mode.
The stabilization on a value far from the reference one, indicated by a ΔLeq higher than some
tenths, means that one of the components of the microphone-preamplifier-instrument chain was af-
fected by a considerable drift and if this difference exceeds the maximum limit admissible by the
instrument, calibration will fail. In this case, refer to “Troubleshooting” (page 90), and if necessary
contact our service department.
- 51 -
V4.2
ACOUSTIC CALIBRATION
In order to keep the acoustic sensitivity of the microphone-sound level meter chain steady
over the time and in the different using conditions, a reference sound source is used, which gener-
ates a pure tone at a reference frequency with a given pressure level, stable over the time. For this
purpose acoustic calibrator class 1 or class 2 conforming to IEC 60942 and compatible with
the microphone capsule UC52 are used.
For the HD2010UC sound level meter are available Class 1 HD2020 and Class 2 HD2022
calibrators.
The control that the reference sound level by the acoustic calibrator is correctly measured by
the sound level meter (it is normally acceptable that the difference between the sound level meas-
ured by the sound level meter and the nominal level of the calibrator be lower than 0.5dB) has to be
carried out before and after a series of measurements, to ensure that measured values are correct.
When the difference between the sound calibrator level measured with the sound level meter and
the nominal value exceeds 0.5dB, it is necessary to perform an additional acoustic calibration.
Warning: to prevent damaging the sound level meter, it is important, during the calibration
procedure, to follow carefully the on-screen instructions and the indications provided by this
manual.
Operating Procedure
Switch on the sound level meter, if necessary, press STOP to finish the measurement on
progress, and perform the following procedure. The program will automatically control that the
warm-up time, indicated by a flashing W, is over.
1. Press PROG and, with the help of the DOWN arrow, select "Acoustic calibration". Press
SELECT to start calibration:
2003/01/01 10 00 00
:
:
Batt 95% Mem 92.5%
:
:
ACOUSTIC CALIBRATION
Last:2003/01/01 10.00
Calibr.level 94.0dB
Leq
dBC
Insert and turn on
the calibrator.
95.3
SELECT PROGRAM
Electric Calibration
Acoustic Calibration
Frequency Response
...
NEXT
EXIT
2. The first screen shows the date of the last calibration (Date:…) and, in the line below, the
calibrator sound level to be used in the calibration in progress (this value can be modified,
before starting the calibration program, by selecting “Calibration Level” from the MENU:
see page.40). Insert the microphone in the calibrator hole and switch it on.
3. Select the sound level indicated on the sound level meter display (94dB is the nominal val-
ue), then press NEXT to continue.
4. Now, the instrument measures the sound level applied and waits for it to become steady: the
measured level will be displayed. The indication "Waiting for level stabilization…" will be
viewed…".
When the sound level has stabilized, the measured value is compared to the reference one
and if the difference is acceptable, it is logged. In this case, the message “Turn off the Cal-
ibrator” will appear. Press NEXT to continue.
- 52 -
V4.2
ACOUSTIC CALIBRATION
Last:2003/01/01 10.00
Calibr.level 94.0dB
Leq
dBC
Turn off the calibra-
tor.
94.0
NEXT
5. After the acoustic calibration, the electric calibration will be automatically started. This
stage of the procedure generates the reference data for the following electrical calibrations.
ACOUSTIC CALIBRATION
Last:2003/01/01 10.00
Calibr.level 94.0dB
Leq
dBC
Confirm calibration?
∆
-0.1
ACOUSTIC CALIBRATION
Last:2003/01/01 10.00
Calibr.level 94.0dB
Waiting for electric
calibration ...
Leq
dBC
∆
12.5
YES
NO
6. At the end, if the values of the electrical calibration are acceptable, press YES to confirm the
new calibration (LEFT arrow of the keypad); at this point, press NO if you want to cancel
the whole calibration (RIGHT arrow of the keypad).
7. Finally, the microphone polarization check is carried out. Wait until "Take out the preampli-
fier from the calibrator" appears".
- 53 -
V4.2
ACOUSTIC CALIBRATION
Last:2003/01/01 10.00
Calibr.level 94.0dB
Take off the
calibrator
ACOUSTIC CALIBRATION
Last:2003/01/01 10.00
Calibr.level 94.0dB
...
Leq
dBC
Checking microphone
polarization
∆
-0.1
EXIT
Leq
dBC
96.1
8. Take out the preamplifier from the calibrator and press EXIT.
9. The procedure is over.
If calibration constants were incompatible with a correct working of the instrument, calibration
would fail and the message “Calibration failed! Consult the manual” would appear. In this case, re-
fer to “Troubleshooting” (page 90), and if necessary contact our service department.
- 54 -
V4.2
MICROPHONE REPLACEMENT
The 2010UC sound level meter is calibrated in the factory in conjunction with the micro-
phone. If the microphone capsule sensitivity deviates too much from factory calibration, the sound
level meter blocks acoustic calibrations and reports the microphone as damaged.
The acoustic calibration can therefore fail even if the capsule is replaced or repaired, or if us-
ing a capsule with different characteristics from the one supplied.
If you wish to replace the microphone capsule it is necessary to use wizard implemented in
Noise Studio under the menu >> Instrument management >> new microphone The procedure re-
quests the use of a certified acoustic calibrator; it is described in detail in Noise Studio’s online
Help under “Options >> New Microphone”.
DIAGNOSTIC CHECK
The diagnostic test is a program that verifies a series of electrical critical parameters. The fol-
lowing are checked: supply voltages, microphone polarization and sensitivity, and the type of pre-
amplifier. At the end of the procedure, should it fail, a table will be shown with the results of the in-
spection. If necessary, refer to “Troubleshooting” (page 90), and if needed contact our service de-
partment.
FIRMWARE UPGRADE
The firmware, that is the program managing all of the instrument functions, can be upgraded
by transferring the file from a PC to the HD2010UC through the serial/USB port. In this way, all
the instrument functions can be updated. Up-dating files are available at the authorized dealers.
To make the update, you need to use “firmware update” function of the NoiseStudio; Noise
Studio must be installed on your PC.
See “NoiseStudio Handbook” online help for details.
OPTIONS UPGRADE
HD2010UC options (HD2010.O0 “Memory Module” and HD2010.O2 “Advanced Data
Logger” can be purchased and installed after the purchase of the basic kit.
“Advance Data Logger” option, if purchased, can be activated directly by the user using
Noise Studio software package. Alternatively it’s possible to send the instrument to Delta Ohm
S.r.l. technical service for upgrade.
To activate the new option it can be used the “upgrade instrument options” function of Noise
Studio software using the activation code supplied after purchase. Please see online manual of
Noise Studio for details.
- 55 -
V4.2
BATTERY SYMBOL AND BATTERY REPLACEMENT
The battery symbol
in the upper right corner of the display constantly provides the charge
status of the instrument batteries. The more the batteries discharge, the more the symbol gets “emp-
ty " …
... when the battery voltage reaches the minimum value for a correct operation, the symbol flashes.
At this point, only 5 minutes of autonomy are left and batteries should be replaced as soon as possi-
ble.
If the instrument is still used, the battery voltage will decrease still further and the in-
strument cannot ensure anymore a correct measurement; data recording is automatically in-
terrupted, as well as data logging and the instrument goes into STOP mode. Under a given
level, the instrument will automatically shut off. Data stored on memory will remain. The in-
strument cannot be switched on again until the battery level remains low.
The battery charge level is available on the menu main screen and on the program screen, ex-
pressed as a percentage value. Press MENU or PROG to view it. When the level is indicated by 0%,
5 minutes of autonomy are left.
The battery symbol changes into a plug when the instrument is connected to an external power sup-
ply.
Note: the battery symbol also flashes when the auto power-off is disabled (AutoPowerOFF = OFF).
To replace the battery, switch the instrument off, then unscrew anticlockwise the two screws
of the cover of the battery space. After replacing batteries (4 1.5V alkaline batteries - type AA),
close again the cover and fasten the two screws clockwise. Check date and time. If this operation
takes less than two minutes, no clock adjustment should be needed.
- 56 -
V4.2
As an alternative to alkaline batteries, you can use rechargeable batteries. Batteries with lower
capacity usually show greater impedance, causing a worsening of the electrical noise generated by
the sound level meter, with repercussions on the measurement dynamic. Therefore zinc-carbon and
rechargeable NiCd batteries are not recommended.
WARNING ABOUT THE USE OF BATTERIES
• Batteries should be removed when the instrument is not used for an extended time.
• Flat batteries must be replaced immediately.
• Avoid loss of liquid from batteries.
• Use waterproof and good-quality batteries, if possible alkaline.
• If the instrument does not turn on after battery replacement:
Remove one of the batteries
Wait at least 5 minutes for the sound level meter internal circuits to discharge
Put the new battery in. If the batteries are charged the instrument should turn on au-
tomatically.
INSTRUMENT STORAGE
• Temperature: -25...+70°C.
• Humidity: less than 90% R.H. no condensation.
• To be avoided:
1. High humidity storage.
2. Direct sun irradiation over the instrument.
3. High temperature source near the instrument.
4. Presence of strong vibrations.
5. Steam, salt and/or corrosive environments.
The instrument case is in ABS plastics material and the protection belt in rubber: do not use any
solvent for cleaning.
- 57 -
V4.2
SERIAL INTERFACE
The instrument is provided with a versatile serial interface with double protocol: RS232C and
USB. The interface settings depend on the item “MENU >> General >> Input/Output >> Serial
Disp.”:
• PRINTER: connection with RS232 interface of the portable serial printer.
• MODEM: connection with RS232 interface to a modem.
• RS232: connection with RS232 interface to a PC equipped with COM type physical port.
• USB: connection with USB interface to a PC where the VCOM driver is installed.
The RS232 setting allows to connect the sound level meter to a COM type physical port of a
PC. This connection does not need any particular program to work as it is allowed by the common
architecture of PC equipped with RS232 (COM) port. The maximum speed of data transfer is up to
115200 baud.
In the last few years, in order to meet the needs of the new audio and video peripherals, the
USB standard has been used for information serial transfer. Recently many PC manufacturers do
not offer the COM type ports any longer, which are usually replaced by the USB type ports. There
is a 4 wire connection, two wires for information transfer, other two wires for the power supply. As
far as the data transfer is concerned, the main differences with respect to the RS232 interface RS232
are:
• The transfer occurs in simplex mode, i.e. it’s impossible to carry out simultaneously a
transfer in both the directions
• The data are transferred as package size
• The transfer time is defined by only one of the two units (the master)
• The transfer speed is fix at 1.5Mbit/s, 12Mbit/s or 480Mbit/s according to the USB stand-
ard and the kind of connected device.
The two devices connected through the USB interface are identified as master and slave. The
master supplies power to the slave and decide the sense and the transfer time scheduling.
The USB interface of sound level meter is a sort of slave and then it has to be connected to a
USB master able to supply with the necessary power and to manage the communication.
On the HD2010UC standard kit it is included a serial connection cable; the user can choose
between code HD2110RS for PC with COM ports or code HD2110USB for PC with USB ports.
The HD2110/RS cable is a null-modem cable with 9-way sub D female connector. The
HD2110/USB cable is fitted with an USB connector type A. On request, the connection cable for
modem or printer (DCE) with a 25-way sub D male connector (code HD2110/CSM) or with a 9-
way sub D male connector (code HD2110/CSP) can be supplied.
When the item “MENU >> Instrument >> Input/Output >> Serial Device” is set on “PRINTER”,
“MODEM” or “RS232”, the following signals are connected to the 8 pin male connector type M12
available on the instrument:
Pin
Direction Signal
Description
1
Input
CTS
Clear to send
2
Output
DTE
DTE ready
3
Input
DCE - CD
DCE ready – Carrier detect
4
Output
VDD
Power supply 3.3V
5
Input
RD
Receiving data channel
6
Output
RTS
Request to send
7
Output
TD
Transmitting data channel
8
-
GND
Reference ground
- 58 -
V4.2
The following signals are connected to the 9 pin sub D male connector of the HD2110/RS cable:
Pin
Direction
Signal
Description
1
DCE >> HD2010UC
DCE - CD DCE ready – Carrier detect
2
DCE >> HD2010UC
RD
Receiving data channel
3
HD2010UC >> DCE
TD
Transmitting data channel
4
HD2010UC >> DCE
DTE
DTE ready
5
-
GND
Reference ground
7
HD2010UC >> DCE
RTS
Request to send
8
DCE >> HD2010UC
CTS
Clear to send
9
HD2010UC >> DCE
VDD
Power supply 3.3V
When the item “MENU >> General >> Input/Output >> Serial Device” is set on “USB”, the follow-
ing signals are connected to the 8 pin male connector type M12 available on the instrument:
Pin
Direction Signal
Description
2
I/O
DP
Datum +
4
I/O
DM
Datum -
6
Input
VBUS
Power supply 5V
8
-
GND
Reference ground
While the sound level meter is connected to an active terminal (DCE active) via the RS232 in-
terface, the auto power off is disabled and the instrument cannot be switched off. If the instru-
ment is off, the connection to an active terminal (DCE active) will turn it on.
Standard parameters of the instrument serial transmission are:
• Baud rate
baud 38400
• Parity
None
• N. bit
8
• Stop bit
1
• Protocol
Hardware.
Data transmission speed can be changed through the "Baud rate" parameter inside the menu -
(MENU >> Instrument >> Input/Output >> Baud Rate – see page 37). The available baud rates are:
230400, 115200, 57600, 38400, 19200, 9600, 4800, 2400, 1200, 600, 300. The other transmission
parameters are fixed.
The 2010UC is provided with a complete set of commands to be sent via the serial port of a PC.
- 59 -
V4.2
COMMUNICATION PROTOCOL
The command consists of ASCII strings with a variable length, ending in CR-LF.
The instrument provides always a response, after a command has been received; if the command is
not accepted, the response string is always NAK-CR-LF. It is possible to disable the response, when
it is not expressly requested by the command, modifying the VERBOSE setup parameter (see the
SET paragraph).
Commands are divided into 5 groups, as shown in the following table.
Group
N. of Commands
Description
SET
60
SETUP: parameter configuration
KEY
21
KEY: keyboard simulation
STT
4
STATUS: instrument status
DMP
6
DUMP: memory dump
Each group contains a given number of commands. Each command is identified by a specific string.
The generic syntax of a command is the following:
< group>:::CR-LF
Ex.:
“SET:INPUT_GAIN:10\r\n”
sets the INPUT_GAIN parameter to 10dB (see SET paragraph).
Only capital characters are acknowledged. Each token can be shortened at the minimum number of
characters that univocally identify it. The example can be shortened as follows:
“SET:INP:10\r\n”
Here are the possible command formats.
A3 - SET:INTEGRATION_TIME:<{SS,MM,HH}>:< value >CRLF
A4 - SET:TIME::CRLF
A5 - SET:DATE:::CRLF
A6 - SET:x_SLM_PARAMETER:< Parameter abbreviation>:CRLF
A8 - SET::< value >CRLF
A10 - SET::?CRLF
C1 - KEY:CRLF
C2 - KEY::< value >CRLF
D1 - STT::< OPTION >CRLF
E1 - DMP:CRLF
If you enter “?” properly in the string, you can get either a help for the compilation of the desired
command or the current status of the instrument configuration parameters. Here are the formats of
the commands that use the “?” character “?”.
0
?CRLF
Provides the list of the groups of commands
A9 SET:?CRLF
Provides the command list of the SET group
A10 SET::?CRLF Provides the current status of the specified command
C3
KEY:?CRLF
Provides the command list of the KEY group
D2
STT:?CRLF
Provides the command list of the STT group
D3
STT::?CRLF Provides the current status of the specified command
E2
DMP:?CRLF
Provides the command list of the DMP group
- 60 -
V4.2
SET GROUP (SETUP)
The following table shows the list of the commands of the SET group (SETUP).
Command
Format
Description
INSTR_MODEL
A10
Instrument model - UNMODIFIABLE
INSTR_NUMBER
A10
Instrument serial number - UNMODIFIABLE
INSTR_VERSION
A10
Instrument version - UNMODIFIABLE
MIC_MODEL
A10
Microphone model – UNMODIFIABLE
MIC_NUMBER
A10
Microphone serial number – UNMODIFIABLE
MIC_TYPE
A10
Type of microphone – UNMODIFIABLE
CLASS
A10
Class of tolerance – UNMODIFIABLE
MEM_SIZE
A10
Memory size – UNMODIFIABLE
OPTIONS
A10
Firmware options – UNMODIFIABLE
TIME
A4
Time (hh:mm)
DATE
A5
Data (aaaa/mm/gg)
DISP_CONTRAST
A8
Display contrast (3÷9, default: 5)
AUTO_POWEROFF
A8
Instrument auto-power-off (ON/OFF, default: ON)
BAUD_RATE
A8
RS232 baud rate
DEVICE
A8
Serial device
INPUT_GAIN
A8
Input gain
INTEGRATION_TIME
A3
Integration time in s, m (1÷59) od h (1÷99)
REPORT_TIME
A8
Report Time
ERASE_TIME
A8
Erase interval
EXCHANGE_RATE
A8
Exchange rate in dB (3÷5)
DOSE_THRESHOLD
A8
Dose threshold in dB (0÷140)
CRITERION_LEVEL
A8
Criterion level in dB (60÷140)
VERBOSE
A8
Acknowledge (ON/OFF, default:ON). Always ON upon power
on.
OVERLOAD_LEVEL
A8
Overload level in dB (20÷200)
INT_MODE
A8
Integration Mode
1_PERC_LEVEL
A8
Percentile level 1 in % (1 ÷ 99, default: 1)
2_PERC_LEVEL
A8
Percentile level 2 in % (1 ÷ 99, default: 10)
3_PERC_LEVEL
A8
Percentile level 3 in % (1 ÷ 99, default: 50)
4_PERC_LEVEL
A8
Percentile level 4 in % (1 ÷ 99, default: 90)
1_SLM_PARAMETER
A6
SLM parameter 1 (see parameter list)
2_SLM_PARAMETER
A6
SLM parameter 2 (see parameter list)
3_SLM_PARAMETER
A6
SLM parameter 3 (see parameter list)
STAT_PARAMETER
A6
Parameter for statistical analysis
EVN_TRIGGER
A8
Event trigger source
EVN_ON_LEVEL
A8
Trigger activation level dB (10 ÷ 140, default: 90)
EVN_OFF_LEVEL
A8
Trigger deactivation level dB (10 ÷ 140, default: 60)
EVN_POLARITY
A8
Trigger level polarity (POS/NEG)
EVN_ON_TIME
A8
Trigger activation delay in seconds from 0 to 10
EVN_OFF_TIME
A8
Trigger deactivation delay in seconds from 0 to 255
EVN_PRINT
A8
Enabling event trigger printing
AUTO_STORE
A8
Enabling Auto-Store function (ON/OFF, default: OFF)
SLM+PROF_DLOGGE
R
A8
Continuous recording of SLM and PROFILE parameters
(ON/OFF, default: ON)
1_REP_PARAMETER
A6
REPORT parameter 1 (see parameter list)
2_REP_PARAMETER
A6
REPORT parameter 2 (see parameter list)
- 61 -
V4.2
Command
Format
Description
3_REP_PARAMETER
A6
REPORT parameter 3 (see parameter list)
4_REP_PARAMETER
A6
REPORT parameter 4 (see parameter list)
5_REP_PARAMETER
A6
REPORT parameter 5 (see parameter list)
REP_PARAMETERS
A8
Recording of REPORT parameters 1-5 (ON/OFF, default: OFF)
REP_STATISTICS
A8
Statistical recording (ON/OFF, default: OFF)
1_EVN_PARAMETER
A6
EVENT parameter 1 (see parameter list)
2_EVN_PARAMETER
A6
EVENT parameter 2 (see parameter list)
3_EVN_PARAMETER
A6
EVENT parameter 3 (see parameter list)
4_EVN_PARAMETER
A6
EVENT parameter 4 (see parameter list)
5_EVN_PARAMETER
A6
EVENT parameter 5 (see parameter list)
EVN_PARAMETERS
A8
Recording of EVENT parameters 1-5 (ON/OFF, default: OFF)
EVN_STATISTICS
A8
Statistical recording (ON/OFF, default: OFF)
CAL_LEVEL
A8
Acoustic calibrator level in dB (90.0 ÷ 130.0, default: 94.0)
MIC_CORR
A8
Acoustic range correction
WND_SHL_CORR
A10
Wind-shield correction (ON/OFF, default: OFF)
SEQ_TIMER
A3
Acquisition delay in s, m (1÷59) or h (1÷99)
The value that some parameters can take is listed in the following table. The default value is in bold
print.
Parameter
Value
BAUD_RATE
300
600
1.2k
2.4k
4.8k
9.6k
19.2k
38.4k
57.6k
115.2k
230.4k
DEVICE
RS232
MODEM
USB
PRINTER
INPUT_GAIN
0
10
20
30
40
ERASE_TIME
5s
10s
30s
60s
INT_MODE
SING
MULT
REPORT_TIME
1s
2s
5s
- 62 -
V4.2
Parameter
Value
10s
20s
30s
1m
2m
5m
10m
20m
30m
1h
EVN_TRIGGER
OFF
LEV
MAN
MIC_CORR
FF
RI
WND_SHL_CORR
OFF
SAV
WME
EVN_PRINT
OFF
TAG
The parameters that can be displayed in SLM mode are selectable among the following ones:
Parameter Attribute
Description
Lpk
Z or C
Instantaneous peak level, Z or C weighted
Lpkmx
Z or C
Peak maximum level
LeqS
Z, C or A
Short equivalent level, Z, C or A weighted
Leq
Z, C or A
Equivalent level
LFp
Z, C or A
FAST sound pressure level
LSp
Z, C or A
SLOW sound pressure level
LIp
Z, C or A
IMPULSE sound pressure level
LFmx
Z, C or A
FAST maximum sound pressure level
LSmx
Z, C or A
SLOW maximum sound pressure level
LImx
Z, C or A
IMPULSE maximum sound pressure level
LFmn
Z, C or A
FAST minimum sound pressure level
LSmn
Z, C or A
SLOW minimum sound pressure level
LImn
Z, C or A
IMPULSE minimum sound pressure level
LeqI
A
Equivalent sound pressure level with Impulse time constant
LE
A
A weighted exposure level (SEL)
Lep,d
A
Daily personal exposure level, A weighted (EEC/86/188)
Dose
A
A weighted dose
Dose,d
A
A weighted daily dose
L1
A
Percentile level (calculated on FAST pressure level, A weighted)
L2
A
Percentile level
L3
A
Percentile level
L4
A
Percentile level
OL
-
Time percentage in which an overload has occurred
The attribute of parameters that can be displayed in SLM mode indicates the respective frequency
weighting.
- 63 -
V4.2
KEY GROUP
The following table shows the command list of the KEY group.
Command
Format
Description
LEFT
C1
LEFT key
MENU
C1
MENU key
PRINT
C1
PRINT key
PROG
C1
PROG key
PAUSE
C1
PAUSE key
RUN
C1
RUN key
SELECT
C1
SELECT key
UP
C1
UP key
MODE
C1
MODE key
RIGHT
C1
RIGHT key
ENTER
C1
ENTER key
DOWN
C1
DOWN key
HOLD
C1
HOLD key
CURSOR
C1
CURSOR key
CLEFT
C1
CURSOR LEFT key
CRIGHT
C1
CURSOR RIGHT key
SER_MON
C1
Simulates pressure of the PRINT key for over 2 sec
STORE
C1
Simulates pressure of the REC key for over 2 sec
DATA_LOG
C1
REC+RUN key
PRN_VAL
C1
PRINT key without printing the heading
EXEC
C2
Program execution
GROUP (STATUS)
The following table shows the command list of the STT group (STATUS).
Command
Description
ACQUISITION
Acquisition control
DISPLAY
Display management
MONITOR
Monitor function via RS232
RECORDER
Recording management
- 64 -
V4.2
The STT:ACQUISITION commands are provided in the following table
Command
Format
Description
HOLD
D1
Interrupts display update
UPDATE
D1
Restarts display update
PAUSE
D1
Measurement in pause
RUN
D1
Starts measurements
STOP
D1
Ends measurements
CLEAR
D1
Clears measured levels
CONTINUE
D1
Restarts measuring
ERASE
D1
Erases the last x seconds of measure-
ments
RECORD
D1
Starts and records measurements
The STT:ACQUISITION:? command provides information on the acquisition status as shown in
the following example.
STT:ACQ:?
STT:ACQUISITION:STOP
BATTERY: 32%
MEMORY: 95.4%
DUMP TIME:00:00:01
LAST CALIBRATION: 2003/07/31 08:37
The STT:DISPLAY commands are listed in the table below.
Command
Format
Description
SLM
D1
Displays 5 selectable parameters in numeric format
PROFILE
D1
Displays the time profile of a selectable parameter
PROB_DISTR.
D1
Displays the level distribution of probabilities
CUMUL_DISTR.
D1
Displays the percentile levels graph
The STT:DISPLAY:? command provides information relevant to the sound level meter actual dis-
play as shown in this example.
STT:DIS:?
STT:DISPLAY:Mode:PROFILE
The following table lists the STT:MONITOR commands.
Command
Format
Description
ON
D1
Starts the Monitor function
OFF
D1
Ends the Monitor function
MEASUREMENT
D1
Monitor
SLM
D1
Monitor by 3 parameters
PROFILE
D1
Monitor by single parameter
REPORT
D1
Monitor of the reports
EVENT
D1
Monitor of the events
- 65 -
V4.2
The following table lists the STT:RECORDER commands.
Command
Format
Description
ON
D1
Starts the Record function
OFF
D1
Terminates the Record function
AUTO
D1
Enables the Auto-Store function
STT:MONITOR:? and STT:RECORDER:? commands provide information on the monitor and re-
cording status as shown in the following example.
STT:REC:?
STT:RECORDER:Measurement:SLM:OFF
DMP GROUP (DUMP)
The table below features the list of the DMP group commands (DUMP).
Comando
Formato
Descrizione
ON
E1
Starts memory dump
OFF
E1
Ends memory dump
NEXT_RECORD
E1
Requires the transmission of the next record
RECORD
E1
Requires the transmission of the current record
CLEAR
E1
Clears memory
Data download sequence is:
•
DMP:ON\r\n
The heading ending in the string “MEMORY DUMP\r\n” is printed if there are any data on
memory
•
DMP:RECORD\r\n
Prints the previous record in binary format
•
DMP:NEXT_RECORD\r\n
Prints the current record in binary format. If this is the last record, it prints the string “END
OF DUMP\r\n”
•
DMP:CLEAR\r\n (optional)
Clears memory
•
DMP:OFF\r\n
Ends data dump
Data dump can be interrupted through the sequences:
•
DMP:OFF\r\n
Ends data dump
- 66 -
V4.2
CONNECTION TO A MODEM
A modem connection allows a remote control of the HD2010UC sound level meter. The option-
al Noise Studio software module NS4 “Monitor” for Windows 98/ME/2000/Xp/Vista operating sys-
tems, can fully manage the sound level meter not only via a simple RS232 or USB serial connec-
tion, but also, via the telephone line by means of two modems. While the modem that connects the
PC to the telephone line must not meet any particular requirement, but being Hayes© compatible,
the modem connected to the HD2010UC sound level meter has to be configurable by the sound lev-
el meter itself and shall not interfere with improper messages during the delicate data transmission
phase from the sound level meter to the PC. Delta Ohm S.r.l. suggests three types of modems to be
used:
Multitech MT2834ZDX
Digicom SNM49
Digicom Botticelli
The connection with these modems has been tested. Other modems might be used but, due to
the great variety available on the market, we cannot provide service for the connection to mo-
dems other than those listed here.
The modem connected to the HD2010UC sound level meter must be configured before being used
for data transmission. The configuration is carried out automatically by the sound level meter itself,
according to the following steps.
1. Connect the modem to the HD2010UC by means of the HD2110/CSM special cable with M12
connector.
2. Connect the modem to the telephone line and the power supply.
3. Switch on the modem.
4. Switch on the HD2010UC sound level meter.
5. Set the baud rate at 38400 baud at least via the parameter: MENU >> Instrument >> In-
put/Output >> RS232 Baud Rate.
6. Select MODEM as serial connection through the parameter: MENU >> Instrument >> In-
put/Output >> Serial Device.
The instrument automatically gets into the modem configuration mode. The successful configu-
ration will be confirmed by the message “Modem Configured.” Should the configuration fail, the
sound level meter will automatically switch to PC mode and the message “Configuration failed!”
will be displayed.
When the modem is configured, it is possible to make the remote connection running NoiseStu-
dio >> Monitor program. Possible under voltages on the modem do not create problems as the con-
figuration is memorized and automatically loaded on turning it on.
The table shows the HD2110/CSM cable connections:
M12 connector (female 8-pole)
DB25 serial connector (male)
1
5
2
20
3
8
4
22
5
3
6
4
7
2
8
7
- 67 -
V4.2
Connectors - Front/External view
DB25
M12
CONNECTION TO A PRINTER
The HD2010UC sound level meter can print
the levels displayed in a format compatible with a
portable 24 column printer, such as the HD40.1.
Printer and sound level meter must be config-
ured properly.
Sound level meter configuration
1. Set the parameter MENU >> Instrument >> In-
put/Output >> RS232 Baud Rate to: 38.4k
2. Set the parameter MENU >> Instrument >> In-
put/Output >> Serial Device to: PRINTER.
Printer configuration
1. The printer Baud Rate must be the same of the
sound level meter (38400 baud).
2. Data bits: 8.
3. Parity: None
4. Stop bits: 1.
5. Handshaking: Xon/Xoff
6. Auto feed: Enabled.
Connect the HD2010UC sound level meter to the printer by means of the special HD2110/RS ca-
ble.
Follow the instructions in the documentation supplied with the printer.
- 68 -
V4.2
CONNECTION TO A PC WITH USB INTERFACE
The 2010UC sound level meter fitted with USB interface can be connected to a PC’s USB
port by using the HD2110/USB cable.
The USB port connection requires the previous installation of a driver contained in the Noise
Studio software. Before connecting the USB cable to the PC, install the NoiseStudio software.
With Windows 7 operating system it’s necessary to boot the PC taking care to disable the re-
quest for “driver signature”, as explained in the USB driver installation guide (included in the in-
stallation CD Rom of Noise Studio).
Proceed as follows:
1.
Set the instrument menu item “MENU >> Instrument >> Input/Output >> Serial Device” to
“USB”. Confirm and exit from the menu”.
2.
Do not connect the instrument to USB port until requested from the wizard
3.
Insert Noise Studio installation CDRom and wait for AutoPlay window to appear. For Win-
dows Vista and Windows 7 operating systems it’s shown the window below; select “auto-
run.exe execution” (see “note 1” in the next paragraph)
4.
The starting screen of Noise Studio will be opened: if the operating system prevents software
to start, please see “note 1” in the next paragraph.
5.
From starting window of Noise Studio click on “USB drive installation” to start the driver in-
stallation procedure.
6.
The program checks the operating system version and copies related drivers in a temporary
folder.
7.
At the end a message appears asking to plug the sound level meter connection cable to PC
USB port: press OK and close Noise Studio software pressing EXIT key.
8.
Connect the sound level meter to USB port and turn it ON: when Windows recognise the
instrument, it appears a message that a new device has been detected.
9.
Wait some seconds until the message “new hardware is installed and ready to work” appears.
10.
Driver installation procedure is ended: whenever the instrument will be connected it will be
automatically recognized.
- 69 -
V4.2
NOTE FOR INSTALLATION
Note 1: for Windows Vista and Windows 7 operating systems.
1. For software installation is requested Administrator permission
2. If the operating system prevents the software to open, boot the PC as administrator, insert in-
stallation CDRom and, when the following window appears, select “Open folder to view files”
3. In the following screen, click with right mouse key on file “Autorun” and select “Execute as ad-
ministrator”:
4. Continue from point 5 of previous paragraph.
Note 2: if the sound level meter has been connected to USB port before USB driver installation,
in Windows 2000 and XP operating systems it appears the window “new hardware installation”. In
Windows Vista and Windows 7 it appears an installation error under “device manager”: in both
cases, cancel operation, disconnect instrument and restart the complete procedure starting from the
beginning of this guide.
VERIFICATION OF PROPER DRIVER INSTALLATION
To verify the proper installation procedure, proceed as follows.
On Windows 2000 and XP operating systems:
Select “START >> Settings >> Control Panel”, double click on SYSTEM icon. Select
“Hardware >> Device Manager”
On Windows Vista and Windows 7 operating systems:
Select “START >> Control panel”, then click on “Device Manager” icon.
Connect instrument to USB port.
The following items should appear:
• “Port (COM and LPT) >> USB Serial Port (COM#)”. The value # is the number assigned to
virtual serial port.
• “Controller USB (Universal serial bus) >> USB serial converter”
- 70 -
V4.2
When USB cable is unplugged, the above items will disappear and they will appear again as soon as
the instrument is plugged to USB port.
In the documentation supplied with the Noise Studio CDRom, it’s available a detailed version (in-
cluding images) of the USB driver management guide.
- 71 -
V4.2
HD2010/MC MEMORY CARD READER
The memory card reader HD2010/MC can be used on the sound level meters provided with M12
serial connection. If they aren’t provided with M12 input, they can be modified in order to be able
to use the memory card reader HD2010/MC.
DESCRIPTION OF THE INTERFACE FOR MEMORY CARD HD2010/MC
The reader HD2010/MC allows to expand the storage capacity of the sound level meter. By em-
ploying SD type or MMC card type, it is possible to record, for example, a sound level profile con-
tinuously for several months. In addition, it is possible to download data from the internal memory
of the sound level meter to the card. This function can turn very useful when the effected measure-
ments need to be downloaded without moving the device or without using a notebook.
The maximum storage capacity of the card is 2GB, the formatting is FAT16 type.
The reader is powered by the sound level meter and doesn’t require any external power supply.
PREPARATION OF A NEW MEMORY CARD
Every new memory card needs to be formatted before use by means of the appropriate function of
the Noise Studio software.
The formatting function requires the PC, on which the Noise studio Software has been installed, to
be equipped with a PC Memory card reader (not included in supply). Those are typically included
in most recent laptops and desktop computers. Optionally it is possible to use an external memory
card, to be connected to an USB port.
How to format a memory card:
1. If the PC isn’t provided with a SD/MMC card reader, connect an external device to the USB
port of the PC.
2. Start up the Noise Studio software.
3. Press the key Instrument management in Noise Studio: press the key Memory Card Format-
ting.
4. Select the path of the card to be formatted and confirm with ENTER.
5. On the following screen, ensure that the parameter “File System = FAT” has been selected
and press START: the card will be formatted.
- 72 -
V4.2
6. When the message “Formatting completed” appears, press OK for confirmation and CLOSE
to exit.
7. Close the Noise Studio software.
8. The formatting has been accomplished and the card is ready for use.
CONNECTION OF HD2010/MC TO THE SOUND LEVEL METER AND USE OF THE MEMORY CARD
Connection of the HD2010/MC:
1. Switch off the sound level meter .
2. Connect HD2010/MC to the M12 connector of the sound level meter ensuring the correct fas-
tening of the connector’s ring nut. While inserting, please make sure that the arrow printed
on the reader is turned upwards. The reader has to be connected to the sound level meter in
vertical position.
3. Insert the memory card in the slot of HD2010/MC interface.
4. Switch on the sound level meter and set the menu item “MENU >> General >> Input/Output
>> Serial device ” on “MC”.
5. The sound level meter identifies the device. The display of the sound level meter shows “MC
connected” and the remaining dimension of memory. The LED on the card reader signalizes
the connection.
6. If the display visualizes “Connection failed”, check carefully if the memory card has been in-
troduced correctly into the slot and if the connector is plugged in properly.
When HD2010/MC interface is connected to the sound level meter, any recording is automati-
cally sent to the memory card rather than to the internal memory of the sound
level meter.
During recording, the LED on the reader flashes upon receipt of every data package.
When memory is used up entirely, the recording will be interrupted.
To disconnect HD2010/MC interface from the sound level meter:
1. Switch off the sound level meter.
2. Disconnect HD2010/MC interface.
NOTE: Slipping off the memory card while the sound level meter is running and interface is
connected causes the loss of all data.
In order to replace the memory card, switch off the sound level meter.
- 73 -
V4.2
For disabling HD2010/MC temporarily:
Set the menu item “MENU >> General >> Input/Output >> Serial device” on a different device
than MC.
In order to re-enable the reader, set the menu item on “MC”.
READING DATA DIRECTLY FROM PC
In order to read and copy files from the memory card to the PC, use a memory card reader for PC:
the card will be recognized as external peripheral mass storage device.
Do not write, cancel or modify the files through the PC.
In order to cancel the memory card, use the formatting function (see chapter “Preparation of a new
memory card”).
DATA TRANSFER FROM SOUND LEVEL METER TO MEMORY CARD
Data can be transferred from the internal memory of the sound level meter to the external memory
by executing the sound level meter program “PROG key >> Data download to MC”.
This program allows copying of the measurements recorded in the sound level meter into the exter-
nal memory card.
The program can be activated only after having connected the memory card interface HD2010/MC,
as described in the chapter “Connection of HD2010/MC to the sound level meter and use of the
memory card”.
During data download, which occurs automatically, the remaining memory space of the card and
the
estimation of program completion time are displayed on the screen.
The data transfer from the sound level meter to the memory card doesn’t cancel the internal
memory of the sound level meter. I order to cancel the content of the internal data memory of the
sound level meter, it is necessary to use the Navigator program.
To cancel the data of the card, use only the formatting function of Noise Studio: the erasure
of single files contained in the memory card is not foreseen.
The Navigator program is not able to manage registrations which have been effected with external
devices as interface for HD2010/MC memory card.
TECHNICAL SPECIFICATIONS
Type of card
MMC and SD
Maximum capacity
2GB
Power supply
provided by sound level meter
Function indicator
LED on the reader
Connector
8-pole female M12
- 74 -
V4.2
TECHNICAL SPECIFICATIONS
The HD2010UC sound level meter is a type 1 or 2 integrating sound level meter with statistical
analysis and optional logging.
HD2010UC complies with the following standards
IEC 61672:2002-5 and IEC 61672-1 Ed. 2.0 of 2013 Class 1 or Class 2 Group X
IEC 60651:2001-10 Class 1 or Class 2
IEC 60804:2000-10 Class 1 or Class 2
ANSI S1.4:1983 Type 1 or Type 2
Microphone Models
UC52, ½ inch pre-polarized microphone with 20 mV/Pa sensitivity; frequency response op-
timized for Free Field.
MC24E and MC24EH, ¼ inch with 2 mV/Pa (MC24E) or 0.25 mV/Pa (MC24EH) sensitivi-
ty, pre-polarized, optimized for free field measurements.
Preamplifier models
For ½ inch microphones, pre-polarized with 20 mV/Pa sensitivity:
HD2010PNE2: with connection for UC52, ½” microphone and driver for extension cable up
to 10mt. This preamplifier is provided with CTC device for electrical calibration, it can be di-
rectly plugged into the HD2010UC o using the extension cable..
HD2010PNE2W: heated preamplifier with connection for UC52, ½” microphone and driver
for extension cable. This preamplifier is provided with CTC device for electrical calibration,
it can be plugged into the sound level meter o using the 5mt extension cable provided (10mt
as an option).
Suitable for ¼ inch microphones
HD2010PNE4: preamplifier for MC24E ¼” microphone. Equipped with CTC calibration de-
vice for electric calibration and driver for cable up to 100m. Requires the HDP079A02 mi-
crophone adapter.
HD2010PNE4H: preamplifier for MC24EH ¼” microphone. Equipped with CTC calibration
device for electric calibration and driver for cable up to 100m. Requires the HDP079A02 mi-
crophone adapter.
Accessories
The use of the following accessories does not significantly modify the 2010UC technical specifi-
cations:
HD SAV Windshield (with spectral correction Menu >> Calibration >> Shield >> SAV for
type 1 version)
HDP079A02 microphone adapter for ¼ microphones, for the use of MC24E microphone with
HD2010PNE4 preamplifier and MC24EH microphone with HD2010PNE4H preamplifier.
Extension cable between the preamplifier and the body of the sound level meter up to a max-
imum of 10m.
SWD10 stabilized power supply.
HD40.1 portable thermal printer.
VTRAP tripod and HD2110/SA support for the preamplifier.
Outdoor shield HD WME: (with spectral correction Menu >> Calibration >> Shield >>
WME for type 1 version).
Memory card reader HD2110MC.
- 75 -
V4.2
METROLOGICAL CHARACTERISTICS
Frequency Weighting
A, C, Z for RMS measurements
C, Z for peak level measurements
Z weighting is flat along the whole sound spectrum with the following features:
Attenuation
[dB]
Frequency Range
[Hz]
< 0.1
100 ÷ 20 k
< 1
31.5 ÷ 21 k
< 5
16 ÷ 22.5 k
Frequency Response
Response/dB
Frequency / Hz
The selection of the UC52 microphones, on the basis of frequency response specifications,
and the spectral corrections calculated by the sound level meter DSP allows obtaining a Class 1
response according to IEC 61672. The HD2010UC is also available in the Class 2 version.
The frequency response of the microphone depends on the presence of devices like wind-
shield HD SAV or all-weather protection unit HD WME. In order to make measurements with
the maximum possible precision in different situations, the HD2010UC sound level meter can
automatically apply spectral corrections needed to measure in random field conditions, or with
the windshield or with the all-weather outdoor protection. Spectral correction for diffuse field
measurement is enabled setting parameter Menu >> Calibration >> Mic.Resp.Correction >> RI
while correction for HD SAV mounted, is enabled setting parameter Menu >> Calibration >>
Shield >> SAV.
For more information on the applied corrections please refer to the UC52 microphone manual.
HD WME outdoor protection
The windshield and rain protection units mounted on HD WME outdoor protection, modify
microphone frequency response: in order to apply the correction to sound level meter overall re-
1. Tolerance +
2. Sound level meter response
3. Microphone response
4. Tolerance –
5. Effect of the case (dotted line)
- 76 -
V4.2
sponse, it’s necessary to set the specific parameter Menu >> Calibration >> Shield >> WME.
With this setting applied, HD2010UC sound level meter with HD WME protection, fully com-
plies to type 1 specifications according to IEC 61672 standard; in such configuration, when used
in vertical position, the sound level meter is able to measure accurately environmental noise from
above (0° incidence) like for example the noise generated by aircraft overflights.
When parameter Menu >> Calibration >> Shield is set to WME and microphone acoustic re-
sponse ( menu Calibration >> Mic.Resp.Correction) is set to RI, it’s applied a frequency spectral
correction to obtain a flat frequency response in a diffuse field. Using this setting the
HD2010UC sound level meter with HD WME outdoor protection mounted in vertical position, is
suitable for measurements of environmental community noise due to noise sources on the ground
(road traffic noise, railway noise…).
For more information on spectral correction applied to HD2010UC sound level meter when
used with HD WME outdoor protection, please refer to UC52 microphone user manual.
Self-generated Noise
The intrinsic noise, measured replacing the microphone with the capacitive adapter,, ,is shown in
the following tables according to the different frequency weightings, either for RMS level or
peak level measurements:
Input Gain
[dB]
Intrinsic noise [dB]
LpA
LpC
LpZ
LpkC
LpkZ
0
50
49
53
61
65
10
40
39
43
51
55
20
31
32
36
45
48
30
26
30
33
43
45
40
24
29
33
43
45
Linearity range
The lower limit of the linearity range, was calculated as the maximum between the upper limit
decreased by 80dB and the self-generated noise increased by 7dB. The upper limit is equal to
141dB with a input gain equal to 0 dB and decrease by 10 dB for every 10 dB gain increment.
The linearity range is almost independent of frequency in the 31.5 Hz … 12.5 kHz interval, and
it is shown in the following table according to the input gain:
Input Gain [dB]
Parameter
Lower Limit
[dB]
Upper Limit
[dB]
0
LpA
60
141
LpC
60
LpZ
60
LpkC
68
144
LpkZ
72
10
LpA
50
131
LpC
50
LpZ
50
LpkC
58
134
LpkZ
62
20
LpA
40
121
LpC
40
LpZ
43
LpkC
52
124
LpkZ
55
- 77 -
V4.2
Input Gain [dB]
Parameter
Lower Limit
[dB]
Upper Limit
[dB]
30
LpA
33
111
LpC
37
LpZ
40
LpkC
50
114
LpkZ
52
40
LpA
31
101
LpC
36
LpZ
40
LpkC
50
104
LpkZ
52
The starting level for the linearity range detection matches the reference level (94 dB) at 1
kHz. With different frequencies, the starting level takes into account the attenuation of the frequen-
cy weighting being measured. In the secondary fields the starting level is subject to the same incre-
ment of the input gain.
Integration Time
It can be set from a minimum of 1s to a maximum of 99 hours.
Measurement Dynamics with Electromagnetic Fields
Minimum measurable level equal to 60dBA with carrier from 26 MHz to 1 GHz and amplitude
equal to 10V/m modulated 80% at 1 kHz.
Reference Conditions
The measuring range is that with input gain equal to 10 dB.
The level corresponds to 94 dB.
The acoustic calibration can be performed at a sound level included in the range 94dB ÷
124dB.
The reference direction of the acoustic signal is that of the preamplifier longitudinal axis.
The reference acoustic field is the free field
Operating conditions
Storage temperature: -25 ÷ 70°C.
Operating temperature: -10 ÷ 50°C.
Working relative humidity: 25 ÷ 90%RH, not condensing.
Static pressure: 65 ÷ 108kPa.
Protection degree: IP64.
In case of condensation growth, it must evaporate completely before using the sound level
meter.
Drifts
Temperature: ± 0.3 dB over the range -10 ÷ 50°C (with correction for microphone drift acti-
vated).
Relative humidity: ± 0.3 dB over the range 25 ÷ 90%RH, not condensing.
Static pressure: ± 0.3 dB over the range 65 ÷ 108kPa.
- 78 -
V4.2
ELECTRICAL CHARACTERISTICS
Pre-heating time
Lower than 10 seconds.
Power Supply
Batteries: 4 alkaline or rechargeable AA 1.5V internal batteries. The sound level meter does
not have a battery charger.
Autonomy: >12 hours in RUN mode with alkaline good-quality batteries. Amounting to 8
hours when using the outdoor microphone unit HD WME fitted with heated preamplifier.
External batteries: It is possible to connect external batteries to the sound level meter through
the male connector for external power supply (∅ 5.5mm socket). The positive (pole) power
supply must be connected to the central pin. The battery should supply 9 ÷ 12 V at about
200mA/h. The maximum limit for external power supply is 15V.
Mains: mains adapter with 9-12Vdc/300mA continuous voltage.
Switch-off: selectable auto-power-off.
When battery voltage falls below 3.9V, the sound level meter cannot perform any measurement.
However it is possible to access and download the data in the internal memory. Under 3.5V, the
instrument will automatically shut off. Logged data and configuration parameter are maintained
even without power supply.
Input maximum levels
The maximum tolerable sound level with UC52 microphone is 146 dB.
The level of the electrical signal applicable to the microphone input, despite the replacement of
the microphone capsule with the proper capacitive adapter, must not exceed 10Vrms.
LINE Output
3.5mm ∅ mono jack
Preamplifier output signal
Unweighted output protected against short-circuit
Gain: ~ .3 mV/Pa and ~30 mV/Pa respectively for an input gain equal to 0 dB and 20 dB, not
calibrated
Linearity: 100 dB with maximum output level equal to 1.2Vrms
Series impedance: 1kΩ
Typical Load: 100kΩ
DC Output
2.5mm ∅ mono jack
Short-circuit protected output
A weighted output with FAST time constant updated 8 times/s
Gain: 10 mV/dB
Linearity: 80dB
Series impedance: 1kΩ
Typical Load: 100kΩ
- 79 -
V4.2
Serial interface
Socket: 8-way M12
Type: RS232C (EIA/TIA574) or USB 1.1 or 2.0 not insulated
Baud rate: between 300 and 230400 baud
Data bits: 8
Parity: None
Stop bits: 1
Flow control: Hardware
Cable length: max 15m
Extension Cable for the Microphone
The microphone preamplifier can be connected to the sound level meter body through an exten-
sion cable up to 10m long (CPA). The sound level meter specifications are not significantly
modified by the cable presence.
STATISTICAL ANALYSIS
Acoustic descriptor LAFp or Leq Short or Lpk
Sampling 1/8 s.
Classes from 0.5 dB.
Measurement range: 30 dB ÷ 140 dB.
4 percentile levels, programmable between L1 to L99.
Statistical graphs calculation and display (“Advanced Data Logger” option).
Graph of distribution probability of the levels
Graph of the percentile levels from L1 to L99.
DISPLAY
Graphic display
128x64 pixels backlit on a 56x38mm surface.
Mode:
SLM (sound level meter) screen with 3 selectable parameters.
Time profile of the A weighted sound pressure level with FAST time constant according to
1/8s intervals (with “Advanced Data Logger” option”).
Level distribution of probabilities in 0.5dB, 1dB or 2dB classes (with “Advanced Data Log-
ger” option”).
Percentile levels graph from L1 to L99 (with “Advanced Data Logger” option”).
- 80 -
V4.2
STORAGE OF MEASUREMENTS
The basic version is fitted with a permanent 4 MB memory, corresponding to over 500 single
measurements. If “Advanced Data Logger” option is installed the memory capacity can be dou-
bled to 8 MB ( “memory module” option).
In this case it’s possible to store:
Continuous recording mode: almost 24 hours of recording (3 parameters twice per
second plus 1 parameter 8 times per second)
Auto-Store mode: over 80 days of recording (3 parameters on 1 minute intervals).
Report mode: over 5 days of recording on 1 minute interval of 5 user selectable pa-
rameters and complete statistical analysis.
Security of stored data
Unlimited, independent of battery charge
PROGRAMS
Calibrations and Diagnostic Programs
Acoustic calibration at 1kHz using a sound level calibrator included in the range: 94dB ÷
124dB
Electrical calibration with built-in generator.
“Diagnostic check” program”.
Interface and processing programs using a PC
NoiseStudio (basic version) for download and graphic display of stored data, as well as for
instrument configuration.
Noise Studio includes a number of optional modules for analysis and post-processing of noise
measurements captured using the Delta Ohm sound level meters. The functions in these mod-
ules are specifically developed for given applications (such as noise analysis in the workplace
or railway traffic noise analysis) and can be enabled on a licence protected hardware key
(CH20).
“Workers Protection” module – cod.NS1 module is used to get and process the sound meas-
urements performed in workplaces. Data are transferred from the sound level meter memory
to the PC using the Noise Studio program. Here, they can be processed in compliance with the
Italian and European Community rules concerning noise in workplaces.
“Acoustic Pollution” module – cod.NS2A – module for acoustic climate analysis on a daily,
weekly and annual basis including road, railway and airport noise. The software performs sta-
tistical and spectral analysis and automatically indentifies noisy events. The analyses are per-
formed in compliance with the national (D.L.194/2005 and D.M. 16/03/1998) and EU legis-
lation regarding the acoustic pollution and the mapping of the territory. (NS2A analysis re-
quest “advanced data logger” to be installed in the sound level meter).
“Monitor” module - cod.NS4 - for the acoustic monitoring and remote control via PC or via
modem. Programmed acquisition, event identification and synchronized audio recording.
For a detailed description of the currently available analysis functions, contact your authorized
dealer.
- 81 -
V4.2
Firmware
To be upgraded via the serial port with NoiseStudio software
OTHER CHARACTERISTICS
Printing
Direct printing of logged parameters (printing of a single event)
Continuous Printing (Monitor).
Case
Dimensions (Length x Width x Height): 445x100x50mm equipped with preamplifier
Weight: 740g (batteries included)
Materials: ABS, rubber
Time
Date and time: clock and date updated in real time
maximum deviation is less than 1min/month
- 82 -
V4.2
REFERENCE STANDARDS
• IEC 60651:2001, Class 1 or 2.
• IEC 60804:2000, Class 1 or 2.
• IEC 61672-1:2002 and IEC 61672-1 ed. 2.0 of 2013, Class 1 or 2 Group X
• ANSI S1.4-1983, Type 1 or 2.
ITALIAN LAWS
• Noise in working environment: D.L. 81/2008, UNI 9432/2011, ISO9612/11 and Directive
2008/46/CE.
• Acoustical climate assessment and monitoring (“Advanced Data Logger” option) with sound
events capture (“Advanced Data Logger” option”)
• Noise of entertainment dancing spaces: D.P.C.M. 215 del 16/04/99
• Noise emission from machines D.L 262 del 04/09/2002, Directive 2000/14/CE and 2005/88/CE
- 83 -
V4.2
ORDERING CODES
HD2010UC.Kit1: type 1 integrating sound level meter kit with 4MB memory
The kit includes the HD2010UC class 1 integrating sound level meter, HD2010PNE2 pream-
plifier preamplifier (or HD2010PNE2W with opt. HD2010.OE), UC52/1 pre polarized con-
denser microphone ½”, HD2110USB cable (alternatively, on request, HD2110RS serial cable
for RS232 connection), HDSAV windshield, Noise Studio PC software (basic version), IEC
61672 manufacturer conformity declaration, carrying case, instruction manual.
HD2010UC.Kit2: type 2 integrating sound level meter kit with 4MB memory
The kit includes the HD2010UC class 2 integrating sound level meter, HD2010PNE2 pream-
plifier preamplifier (or HD2010PNE2W with opt. HD2010.OE), UC52 pre polarized conden-
ser microphone ½”, HD2110USB cable (alternatively, on request, HD2110RS serial cable for
RS232 connection), HDSAV windshield, Noise Studio PC software (basic version), IEC
61672 manufacturer conformity declaration, carrying case, instruction manual.
Options, accessories and software
HD2010.O0 “Memory module”: 4 MB memory expansion. It requires the HD2010.O2 “Advanced
Data Logger” option.
HD2010.O2 “Advanced Data Logger”: sound level profiles automatic recording, full statistic
analysis, capture and analysis of sound events with trigger function, simultaneous data log-
ging of profiles, reports and events. “Navigator” program for reviewing stored data.
HD2010.O1/4 “Microphone chain for high levels measurements” (up to 160dB): replaces
HD2010PNE2 preamplifier and UC52/1 microphone with HD2010PNE4 preamplifier, com-
plete with HDP079A02 microphone adapter, and MC24E ¼” microphone with 2 mV/Pa sen-
sitivity. Available only for the kit HD2010UC.Kit1.
HD2010.O1/4H “Microphone chain for high levels measurements” (up to 180dB): replaces
HD2010PNE2 preamplifier and UC52/1 microphone with HD2010PNE4H preamplifier,
complete with HDP079A02 microphone adapter, and MC24EH ¼” microphone with 0.25
mV/Pa sensitivity. Available only for the kit HD2010UC.Kit1.
HD2010.OE “Microphone chain for outdoor measurements”: replaces the standard
HD2010PNE2 preamplifier with the heated version HD2010PNE2W, equipped with CTC de-
vice for electrical calibration and 5m integrated extension cable (10m on request); includes
the outdoor unit HD WME with windshield, rain shield and bird spikes. This option is availa-
ble only in combination with the standard microphone.
HD2020: class 1, according to IEC60942:2003, acoustic calibrator with LCD display. Suitable for
½” and, with HD2020AD4 adapter not included, ¼” standard microphones. Calibration fre-
quency 1000 Hz, levels 94 dB and 114 dB. ACCREDIA calibration certificate is included.
HD2022: class 2, according to IEC60942:2003, acoustic calibrator. Suitable for ½” and, with
HD2020AD4 adapter not included, ¼” standard microphones. Calibration frequency 1000 Hz,
level 114 dB. ACCREDIA calibration certificate is included.
HD2020AD4: ½”- ¼” adapter for ¼” microphones. It can be used with HD2020 and HD2022 cali-
brators.
CPA/5: 10m extension cable for HD2010PNE2 preamplifier.
CPA/10: 10m extension cable for HD2010PNE2 preamplifier.
SWD10: stabilized power supply with Vin =100÷240Vac / Vout=12Vdc/1000mA.
VTRAP: tripod, max. height 1550mm.
HD2110/SA: support to fix the preamplifier to the tripod.
- 84 -
V4.2
HD2110RS: RS232 serial cable null-modem with DB9 standard connector.
HD2110USB: USB cable with A type connector
HD40.1: the kit includes: 24-column portable thermal printer, serial interface, 57mm paper width,
four NiMH 1.2V rechargeable batteries, SWD10 power supply, instruction manual, 5 thermal
paper rolls.
HD2010MC: module for datalogging and data download to MMC and SD type memory cards. One
2 GB SD card is included.
HD WME: outdoor microphone protection. Includes: HD WME3 stainless steel support for the
preamplifier, HD SAV3 wind-screen, HD WME2 rain shield, HD WME1 birds spike.
Post processing software modules
CH20: Hardware key for PC with Windows operating system. When connected to USB port, it
enables the use of the Noise Studio software modules.
NS1 “Workers Protection” module. Analysis of the noise and vibrations exposure in the work-
place in compliance with D.L. 81/2008, UNI 9432:2011 and ISO 9612:2009 norms. Calcula-
tion of the exposure levels and related uncertainties and evaluation of the individual protec-
tion devices effectiveness, of the noise sources pulse index and of the presence of the DC-
shift phenomenon in the hand-arm transmitted vibrations analysis.
NS2A “Acoustic Pollution” module. Acoustic climate analysis on a daily, weekly and annual basis
including road, railway and airport noise. The software performs statistical and spectral anal-
yses and automatically identifies noisy events. The analyses are performed in compliance with
the national (D.L. 194/2005 and D.M. 16/03/1998) and EU legislation regarding the acoustic
pollution and the mapping of the territory.
NS4 “Monitor” module. Acoustic monitoring and PC remote control. Programmed data acquisi-
tion, events identification and synchronized audio recording.
Accessories and spare parts
UC52/1: pre-polarized microphone for free field with class 1 frequency response.
UC52: pre-polarized microphone for free field with class 2 frequency response.
MC24E: ¼” pre-polarized microphone for free field measurements up to 160dB. It can be com-
bined with HD2110PEL4 preamplifier by using the HDP079A02 adapter.
MC24EH: ¼” pre-polarized microphone for free field measurements up to 180dB. It can be com-
bined with HD2110PEL4H preamplifier by using the HDP079A02 adapter.
HD WME1: anti-bird spike for HD WME weather protection.
HD WME2: rain protection for HD WME.
HD SAV3: windshield for HD WME weather protection.
HD SAV: windshield for ½” microphones.
BAT-40: Spare battery pack for HD40.1 printer.
RCT: 4 rolls of thermal paper, 57 mm width and 32 mm diameter.
HD2010PNE2: Microphone preamplifier equipped with CTC device for electrical calibration and
driver for cable up to 10 m.
- 85 -
V4.2
HD2010PNE2W: Heated preamplifier for HD WME outdoor microphone unit. It is equipped with
CTC device for electrical calibration and driver for cable up to 10 m. 5 m connection cable
(10 m on request).
HD2010PNE4: preamplifier for MC24E ¼” microphone. Equipped with CTC calibration device for
electric calibration and driver for cable up to 100m. Requires the HDP079A02 microphone
adapter.
HD2010PNE4H: preamplifier for MC24EH ¼” microphone. Equipped with CTC calibration de-
vice for electric calibration and driver for cable up to 100m. Requires the HDP079A02 micro-
phone adapter.
- 86 -
V4.2
WHAT SHALL I DO IF …
This chapter deals with the step by step description on how to carry out the most recurrent
measurements in the acoustic field with the HD2010UC sound level meter.
If necessary, see the description of the key functions from page.92 and the different display modes
from page 12.
MEASUREMENT PROCEDURE
When the “Advanced Data Logger” option is active, the HD2010UC sound level meter can
acquire 3 parameters of SLM screen simultaneously (twice/second), as well as the A weighted
sound level with FAST time constant (8 times/second). The available parameters are those listed in
the tables of appendix A1, on page 99.
Sound Level Meter (SLM) – See also the description on page 14.
Use the MODE key to select the SLM display, where 3 measuring parameters are shown in numeric
format. Press SELECT repeatedly to set the integration time (Tint), the measuring range and the pa-
rameters to be displayed as described in the paragraph “Selecting Parameters” of chapter “SLM
(sound level meter) Mode” on page 14.
Alternatively, measuring parameters can be set from menu, as described in the chapter
“DESCRIPTION OF THE MENU FUNCTIONS on page” 35. Alternatively, measuring parameters
can be set from menu, as described in the chapter.
If the sound level meter is set to the single integration mode (MENU >> Instrument >> Meas-
urement >> Integration Mode: SINGLE), when the Tint time is over, the HOLD indication will ap-
pear and display update will stop. Now you can print or store the values. Meanwhile, the instrument
goes on measuring: to continue display updating, just press the HOLD key. When the continuous
recording is activated (with the “Advanced Data Logger” option) and the Tint time goes over, the
acquisition is stopped automatically.
If you press HOLD during the measuring phase, the display update will be temporarily inter-
rupted. When HOLD is pressed again, the updating will continue. Even if the display is not updated,
the instrument will go on measuring.
Pressing PAUSE, acquisition and calculation of the integrated parameters will be temporarily
interrupted. When in PAUSE, the calculation of integrated parameters, such as, for example, Leq
and the maximum levels, is suspended; in this phase the contribution of the last seconds of acquisi-
tion can be deleted using the “Back-Erase” function and the LEFT and RIGHT keys, as described in
the paragraph: “Back-Erase Function” on page 15. While in PAUSE, all integrated parameters can
be cleared pressing START/STOP/RESET. Press PAUSE once more to start measuring again.
Press PRINT at any time to print the on-screen data. To activate the continuous printing
(Monitor), press and hold down PRINT for at least 2 seconds. A flashing M overlapping the status
indicator, shows that the Monitor function is active. The Monitor function remains active even
jumping to other measuring views and can be disabled pressing PRINT again or
START/STOP/RESET to stop measurement.
If, on the contrary, the sound level meter is set with the multiple integration mode (MENU >>
Instrument >> Measurement >> Integration Mode: MULT), the integrated levels are cleared just be-
fore the beginning of each measurement cycle. The MONITOR function allows sending to the serial
interface the integrated spectrum at intervals equal to the Tint time. Recording in Auto-Store mode
allows to store on memory octave and third octave (if this option is installed) spectra together with
the parameters displayed on the SLM screen at a rate corresponding to the Tint time.
Press REC for 2 seconds at any time to store the displayed data. As soon as data are saved on
memory, a screen will allow you to enter the recording title. If single data recording is activated
- 87 -
V4.2
with the sound level meter in STOP mode and the “Advanced Data Logger” option is installed, a
screen allowing activating the Auto-Store will appear first.
History Mode (Time Profile) - See also the description on page 17.
With the MODE key you can open the PROFILE screen (requires the “Advanced Data Log-
ger” option) in which the time profile of the A weighted sound pressure level with FAST time con-
stant is shown graphically. The sampling interval is equal to 1/8s.
You can start the measurements by pressing the START/STOP/RESET key.
If the sound level meter is set to the single integration mode (MENU >> Instrument >> Measure-
ment >> Integration Mode: SING), once the Tint time (defined in the SLM page), is over, HOLD
will be displayed: display update will stop. Meanwhile, the instrument goes on measuring: to con-
tinue display updating, just press the HOLD key. When the continuous recording is activated and
the Tint time goes over, the acquisition is stopped automatically.
If you press HOLD during the measuring phase, the display update will be temporarily inter-
rupted. When HOLD is pressed again, the updating will continue. Even if the display is not updated,
the instrument will go on measuring.
If you press PAUSE, the acquisition will be temporarily suspended. While in pause, press
START/STOP/RESET to clear the graph. Press PAUSE once more to start measuring again.
If you press CURSOR at any moment, a cursor will be activated. If you press CURSOR once
more, a second cursor will activate, while, if you press it for the third time, both cursors will be ac-
tivated in “tracking”. Use the LEFT and RIGHT arrows of the keypad to shift the selected cursors
into the desired position to indicate the measured level and the corresponding acquisition time.
Press CURSOR again to disable cursors.
PRINT at any time to print the on-screen data. To activate the continuous printing (Monitor),
press and hold down PRINT for at least 2 seconds. The flashing letter M, over the status indicator,
indicates that the Monitor function has been activated. The Monitor function remains active even
when jumping to other measuring views and can be disabled pressing PRINT again or
START/STOP/RESET to stop measurement.
STORAGE OF MEASUREMENTS WITH THE “ADVANCED DATA LOGGER” OPTION
The HD2010UC, sound level meter, when provided with the “Advanced Data Logger” option, of-
fers two different storage modes:
1. The Continuous Recording (press REC and START keys simultaneously) implies the storage
of the SLM screen-page (2 samples a second) together with the time profile of the A weighted
sound pressure level with FAST constant time (8 samples a second). It’s also possible to store
the data of “Report and Events groups”, each consisting of 5 programmable parameters and full
statistical analysis. The data of the “Event group” are stored at the end of each event and the
data of “report group” are stored at programmable intervals from 1 sec to 1 hour. The REC
symbol as status indicator shows when the sound level meter is recording. If you press STOP,
recording will be stopped and you will be asked to confirm the title. While recording, you can
press PAUSE to suspend recording.
When integration mode is multiple (MENU >> Instrument >> Measurement >> Integration
Mode: MULT), the integrated levels are cleared just before the beginning of each measurement
cycle. A special marker identifies the last acquired data per each cycle.
2. The Auto-Store mode can be activated either through the parameter MENU >> Recording >>
Auto-Store or by pressing REC and holding it down for at least 2 seconds with the sound level
meter in STOP mode. Then select the AUTO option when you are requested to choose the re-
cording option. Activating the Auto-Store recording mode together with the multiple integra-
tion mode (MENU >> Instrument >> Measurement >> Integration Mode: MULT) allows to
- 88 -
V4.2
record at intervals equal to the set integration time (MENU >> Instrument >> Measurements
>> Integration interval) the data displayed on SLM screen.
Set in this way and once the integration time has been reached, the sound level meter will first
record data, then set all integrated levels to zero, and finally start a new integration period. The
flashing REC symbol, overlapping the status indicator, shows that the sound level meter is re-
cording. Press START to start recording and STOP to stop it. A flashing REC over the RUN
status indicator, shows when the sound level meter is recording. Press REC with the instrument
in STOP to disable the Auto-Store and the multiple integration modes. The Auto-Store record-
ing mode combined with the single integration mode (MENU >> Instrument >> Measurement
>> Integration Mode: SINGLE) allows to record automatically data displayed on SLM,
OCTAVE and T.OCTAVE (with option) screens at the end of the set integration interval
(MENU >> Instrument >> Measurements >> Integration Interval); the measurement will be au-
tomatically interrupted immediately after recording.
3. Single screen storage is obtained by pressing REC for at least 2 seconds when the instrument is
in RUN or STOP mode. If the instrument is in STOP mode, and the “Advanced Data Logger”
option is installed, you will be prompted to choose between automatic (AUTO) and manual
(MAN) storage; the current screen will be stored if choosing the latter.
The Continuous Recording allows recording the time profile of instantaneous and integrated
levels. This means that you can record, for example, the sound pressure level with FAST time con-
stant 8 times a second and, at the same time, the sound pressure levels with SLOW time constant,
the peak level, and the Leq over 0.5s twice a second.
The Auto-Store recording mode, combined with the multiple integration mode, allows instead
to store sound levels at prefixed intervals, resetting them automatically at the beginning of each in-
terval. Thus you can record, for example, the Leq, the SEL, the maximum sound pressure level with
SLOW time constant, at intervals programmable between 1s and 99 hours.
MEASUREMENT OF NOISE DOSE
The Dose represents the percentage of a maximum value of noise exposure throughout a day.
dt
T
Q
D
T
q
L
L
c
c
∫
−
⋅
=
0
10
100
)
(
where:
D(Q) = exposure percentage for an Exchange Rate equal to Q.
Tc
= daily exposure time (usually 8 hours).
T
= measurement time.
L
= sound pressure level when it is higher than the Threshold Level, and -∞ otherwise.
Lc
= Criterion Level for a daily exposure corresponding to 100% of the dose.
Q
= Exchange Rate.
q
= parameter independent of the exchange rate and equal to:
• 10
for Q = 3dB
• 5/log2 for Q = 5dB
• 4/log2 for Q = 4dB
- 89 -
V4.2
The sound level meter calculates the following parameters: DOSE (A), the percentage of the daily
effective dose, and DOSE,d (A), the estimated daily DOSE according to the programmed parame-
ters.
The DOSE calculation is characterized by three parameters:
1. DOSE Criterion is the SPL constant value, which continuous exposure for 8 hours determines a
100% DOSE.
2. DOSE Threshold that represents the SPL level under which the DOSE is not increased.
3. Exchange rate is the variation of the SPL value that determines a double or half duration of ex-
posure with the same DOSE Criterion. Provided values are 3, 4 or 5dB.
The three configuration parameters are contained in the submenu Measurement (MENU >> Instru-
ment >> Measurement): once they have been set, select the submenu Sound Level Meter (MENU
>> Sound Level Meter) and choose, according to the type of measurements to be made, between the
parameters DOSE (A) or DOSE,d (A).
The integration time can be entered directly in the SLM measuring window. Now the instrument is
ready to carry out the measurement: press START. After Tint time has elapsed, the instrument turns
into the HOLD status and displays the DOSE calculated over the set time.
STATISTICAL ANALYSIS
Up to 3 percentile levels are selectable in the SLM display mode (MENU >> Instrument >>
Measurement >> Percentile Lev. 1-4), programmable between L1 and L99. The statistical analyzer
samples the sound pressure A weighted level with FAST time constant, 8 times a second; the levels
are accumulated in 0.5dB classes. The percentile levels are calculated interpolating the cumulative
distribution.
The “Advanced Data Logger” option can be used to choose which parameter the statistical analysis
is performed on: equivalent level, sound pressure level with FAST time constant, and peak level.
The complete statistical analysis is available with the distribution of probabilities graph and the L1
to L99 percentile level graph.
DATA PRINTING
The displayed values can be printed at any time in all display and acquisition modes.
Besides the Monitor function can be activated via the serial line when you press PRINT and hold it
down for 2 seconds. This function allows to continuously send displayed data in real time to the se-
rial interface. The transferred data are those of the active display mode at the time PRINT was
pressed. Data are continuously transferred until PRINT is pressed again, or until the acquisition is
stopped. The Monitor function can be activated even in the STOP acquisition mode; it will start as
soon as the instrument turns into RUN mode. The Monitor works independently from any recording
of data on memory.
Using the Monitor function it is possible, with the help of a PC, to make measurements limited
only by the PC storage capacity.
- 90 -
V4.2
TROUBLESHOOTING
The HD2010UC sound level meter is provided with a diagnostic program (DIAGNOSTIC
CHECK) that automatically checks the instrument main parameters. This program can be run at any
time to check the instrument operating conditions (see description on page 54).
One of the parameters being analyzed is the sensitivity of the amplification channel that includes,
through a charge partition circuit (CTC), the microphone capacity. Measurement is made at 1kHz.
DIAGNOSTIC CHECK
1. The DIAGNOSTIC CHECK program fails
Replace batteries and try again after waiting for the end of the stabilization time and, if the
problem remains, contact service.
CALIBRATION
1. The ELECTRIC CALIBRATION program fails
Ensure that the instrument is not subject to high noise and/or vibrations.
Try again after having waited for the end of the stabilization time and, if the problem re-
mains, run the ACOUSTIC CALIBRATION program.
2. The ACOUSTIC CALIBRATION program fails
Ensure that the instrument is not subject to high noise and/or vibrations and that acoustic
calibrator and sound level meter are steadily aligned and that the microphone is properly
plugged in the calibrator cavity. Check that the seal rubber ring is present and undamaged.
Try again after having waited for the end of the stabilization time and, if the problem re-
mains, load the factory calibration following these steps:
• Ensure that acquisition is on STOP.
• Remove one of the batteries while the instrument is switched on: this operation ensures
that all the instrument internal circuits are discharged.
• Press and hold ENTER down, and then insert the battery you had removed. The instru-
ment will switch on and will display a warning relevant to the load of factory calibration.
Release ENTER and press the key on the right near the CONTINUE key.
• After having waited for the stabilization time, run the program ACOUSTIC
CALIBRATION.
Should the program fail, contact the service department.
RESTORING FACTORY SETUP
The default configuration of the instrument parameters (factory setup) can be recalled at any
time by means of a combination of keys. This operation does not clear the content of data
memory.
While the instrument is off, press and hold down ENTER and power on the sound level meter. All
the menu items are simultaneously brought back to the default value.
RESTORING FACTORY CALIBRATION
The Factory Calibration can be recalled at any time by means of a combination of keys. This
operation does not clear the content of data memory.
While the instrument is off, remove one of the batteries and wait at least 5 minutes for the sound
level meter internal circuits to discharge.
- 91 -
V4.2
Press and hold down the ENTER key while reinserting the battery: the sound level meter will auto-
matically turn on. Confirm the load of factory calibration.
The sound level meter calibration parameters are restored to the last factory calibration; all menu
items are simultaneously brought back to the factory parameters (default).
MISCELLANEOUS PROBLEMS
1. The instrument does not turn on after battery replacement.
Remove one of the batteries and wait at least 5 minutes before reinserting it. The instru-
ment should turn on automatically when the battery is inserted.
2. The detected sound levels seem incorrect.
Ensure that no condensation is present on the capsule or preamplifier. Avoid turning on
the instrument in conditions of possible condensation growth. Use the outdoor micro-
phone unit HD WME to perform measurements in conditions of high humidity or when it
is raining.
Check that the warm-up time, indicated by a blinking letter “W” over the upper left status
indicator, has elapsed.
Check measurement accuracy using the acoustic calibrator.
Load factory calibration.
Check that the microphone protection grid is screwed down on the capsule securely.
3. Upon turning the sound level meter on, it turns off automatically right after the introduction
screen.
The batteries are flat.
4. The sound level meter does not communicate with the PC.
Check that the sound level meter and PC communication speed is the same (MENU >>
Instrument >> Input/Output >> Baud Rate).
Check that the connection cable is plugged correctly in the sound level meter and is con-
nected to a PC serial port or USB with the item MENU >> Instrument >> Input/Output
>> Serial device set up respectively on RS232 or USB respectively.
If you use a USB interface, check that the driver is correctly installed.
If you use a Noise Studio program, disable the AutoDetect function (Menu Option >>
Port Settings) and set the connection directly to the COM port to which the instrument is
connected with the same baud rate of the sound level meter (Menu >> Instrument >> In-
put/Output >> Baud Rate).
5. It is not possible to activate continuous recording. By pressing REC and RUN, the instrument
starts measurements without recording.
The instrument does not have available memory for additional data. Download the data
and/or erase the memory.
The “Advanced Data Logger” option is not installed.
- 92 -
V4.2
KEYBOARD DESCRIPTION
HOLD key
The HOLD key can be used to temporarily suspend the display update while the instrument contin-
ues making the requested measurements. A “H” in the left corner at the top shows that the display is
in this phase. Press the key again to go back to standard measurement. While the instrument is in
HOLD mode, you can jump from a screen-page to the other, activate the cursors on the graphic
pages, print and store data. Recording and the Monitor function are not affected by the HOLD sta-
tus.
ON/OFF key
To switch on or off the instrument, press the ON/OFF key for at least one second. Upon the power
on, the instrument shows for a few seconds the manufacturer’s logo and the program version. Then
it switches to the SLM working mode (Sound Level Meter) and displays 5 instantaneous or inte-
grated measurement parameters in numeric form.
HD2010
Sound Level Meter
Firmware: 303V2.7
56.9
60.8
79.3
30
110
01:08:25
Tint= s
10
LFp
Leq
Lpk
dBA
dBC
dBC
Before switching off the instrument, press STOP to end the ongoing measurement. If you do not do
it, a message will request you to stop the measurement in progress: “WARNING! Stop the meas-
urement to continue".
SELECT PARAMETER
WARNING!
Stop measurement
to continue.
YES
NO
Press YES and then ON/OFF to switch off the instrument.
- 93 -
V4.2
Auto Power Off Function
The instrument is provided with the Auto-PowerOff function that switches automatically off the in-
strument after 5 minutes when it is in STOP and no key is pressed in this time interval. Before
switching off, the instrument will make a series of beep: in this phase you can press a key to prevent
the instrument from switching off.
The function can be disabled from the MENU via the "Auto-Power-Off" item (MENU >> Instru-
ment >> System >> Auto-Power-Off = OFF). In this case, the battery symbol flashes to remind the
user that the instrument will not switch off automatically, but only if the key is pressed.
The Auto-Power-Off function is temporarily disabled when an external power supply is used or
when the instrument is recording or running a program.
MENU key
The HD2010UC sound level meter requires, according to its use, many parameters to be set. If you
press MENU, you can access all instrument parameters, for the following functions:
Main
Sound Level Meter (SLM)
statistical analyzer (“Advanced Data Logger” option)
Trigger (“Advanced Data Logger” option)
Recording (“Advanced Data Logger” option)
Calibration
Sequencer (“Advanced Data Logger” option)
Inside the menus you can:
Move from one item to the other within the same menu by means of the UP and DOWN ar-
rows,
Press SELECT to choose an item to be modified,
Use the UP and DOWN keys to modify the selected parameter,
Press ENTER to confirm a change, or press MENU to cancel it,
Press MENU to exit the menu or the submenu.
Some of the parameters available via menu can also be set directly during the measurement phase
(like, for example, integration interval, measuring range, etc.) by use of the SELECT, UP, DOWN,
LEFT, RIGHT and ENTER keys.
When you access the menus, the available memory will be displayed, as well as battery level, date
and time.
See a detailed description of the menu items on page 35 and following pages.
- 94 -
V4.2
PRINT key
Press PRINT to send the screen page to the RS232 serial interface, in a printable format.
Data can be downloaded to a PC or sent to a serial printer as for example the HD40.1 (see page 67),
directly connected to the sound level meter. In this last case, set the parameter MENU >> Instru-
ment >> Input/Output >> Serial Device to: PRINTER, to get a printing format compatible with a
portable 24 column printer.
Data downloading to a PC can be managed via a communication program, such as, i.e., Windows
HyperTerminal.
If the key is pressed and soon released, the single screen page is sent to the serial line; the let-
ter “P” will be displayed. If the key is hold down, the continuous printing (Monitor) will be
started and the letter “M” will be displayed: to stop it, press PRINT again or press
START/STOP/RESET to stop logging.
PROG key
The PROG key is used to access the menu of the instrument programs. Use the UP and DOWN ar-
rows to select the program; press SELECT to activate the selected program. These are the available
programs:
• Memory Navigator (with the “Advanced Data Logger” option): allows accessing stored data and
viewing them on the instrument display. It works both with single session data and with multi-
ple ones. (See details on pag.42).
• Electric Calibration: single-frequency (1kHz) with electrical signal coming from the built-in si-
nusoidal reference generator. (See details on page 49).
• Acoustic Calibration: is used for the sound level meter tuning at 1kHz with an acoustic calibra-
tor. (See details on page 51).
• Diagnostic Check: this program checks several instrument parameters: supply voltages, micro-
phone polarization and sensitivity, type of preamplifier (see details on page 54).
• Memory dump on MC: this program allows to copy measurement data from internal sound level
meter flash memory to external Secure Digital memory card (please see chapter on MC on page
71).
The selected program is performed upon pressing the SELECT key; some programs can be aborted
at any time by pressing RIGHT on the keypad. When you access the programs, the available
memory will be displayed, as well as battery level, date and time.
- 95 -
V4.2
PAUSE/CONTINUE key
The PAUSE key interrupts the calculation of integrated measurements (Leq, SEL, maximum or
minimum levels, spectra, etc.), as well as recording. The instantaneous levels are still measured and
displayed in the SLM screen. Press PAUSE/CONTINUE to start measuring again. The integrated
parameters are cleared if you press RUN/STOP/RESET while in PAUSE (during a measurement
session). The last seconds of integration can be excluded, from the calculation of integrated parame-
ters shown on the SLM screen (for example, to eliminate the effect of an undesired noise), using the
LEFT and RIGHT keys while in PAUSE. The maximum erasing interval is programmable from 5
seconds to 60 seconds via MENU >> Instrument >> Measurement.
If you press the RUN/STOP/RESET key while in pause during the replay of a recording (with the
“Advanced Data Logger” option), the next stored data will be displayed. If you press the
RUN/STOP/RESET key and hold it down, the replay will be performed in fast forward mode.
REC key
If REC is pressed and hold down for at least 2 seconds, the displayed data are stored as a single re-
port. You can also activate the automatic recording of the parameters displayed on the SLM screen,
(see THE RECORD FUNCTION on page 30).
When the “Advanced Data Logger” option is installed, the REC key, combined with
START/STOP/RESET, activates the data logging. Starting from the STOP condition, if you press
REC and hold it down, then press START/STOP/RESET, the data logging of measured values is
started. To stop logging, press START/STOP/RESET: registration number, date and time will be
displayed. Press ENTER to confirm.
RUN/STOP/RESET key
If you press RUN, while in stop, all the initial values of the integrated measurements like Leq, SEL,
MAX/MIN levels will be first cleared (RESET) and then measurement starts. Pressing the key
again (STOP), the measurements will be stopped. If pressed while in pause, all integrated pa-
rameters will be cleared.
During the replay of stored data (with the “Advanced Data Logger” option), if you press this key
while in pause, the next data will be displayed; if you press it and hold it down, the replay will be
executed in fast forward mode.
- 96 -
V4.2
SELECT key
The SELECT key activates the adjustment mode of displayed parameters by selecting them in se-
quence. Use the two arrows, UP and DOWN, to modify the values. When you have finished with
the adjustments, wait a few seconds or press ENTER to confirm and quit the selection mode.
UP key
The UP key selects a previous line in the menus or increases the selected parameter. Decreases the
scale limits of the time profile and of the frequency spectra moving the graph upwards.
MODE key
If “Advanced Data Logger” option is installed, the MODE key selects the different display modes
of the instrument from SLM to Time Profile, Probability Distribution, Cumulative Distribution.
All operating modes are active at the same time, even though not displayed: using the MODE key,
you can select the display mode without affecting measurement.
LEFT key
The LEFT key selects the previous character in the active line of the menu. It jumps to a previous
parameter during the selection of a measurement variable that needs the definition of more than one
parameter (see SELECT).
- 97 -
V4.2
ENTER key
The ENTER key confirms the selected parameter. When setting parameters from the menu, to quit
setting without saving a parameter, just press any key except SELECT, ENTER and the four ar-
rows, or press MENU. Factory setup is loaded into the sound level meter by turning it on while
holding the ENTER key.
RIGHT key
The RIGHT key selects the next character in the active line of the menu. It jumps to the next pa-
rameter during the selection of a measuring variable that needs the definition of more than one pa-
rameter (see SELECT).
DOWN key
The DOWN key selects the next line in the menus or decreases the selected parameter.
CURSOR (Keypad)
It activates the cursors in a graph. Press it repeatedly to activate in sequence: the first cursor L1, the
second one L2 or both of them in “tracking” (ΔL). To disable the cursors, press the key again.
Use the LEFT and RIGHT key of the keypad to move the flashing selected cursor over the graph.
Relevant values are shown at the top of the display.
LEFT (Keypad)
The LEFT key moves leftwards the cursor or the two active cursors (flashing).
- 98 -
V4.2
RIGHT (Keypad)
The RIGHT key moves rightwards the cursor or the two active cursors (flashing).
- 99 -
V4.2
ANNEX
A1. HD2010UC MEASURING PARAMETERS
In the next paragraphs, are shown the acoustic parameters that can be displayed in numeric or
graphic format, with the respective abbreviations used to identify them.
ACOUSTIC DESCRIPTORS (NUMERIC DISPLAY)
Instantaneous acoustic levels refreshed every 0.5s
Wideband
PARAMETER
ABBREV.
DEFINITION
WEIGH.
FREQ.
WEIGH.
TEMP.
LXeq(Short)
LeqS
dBX
Short equivalent level (0.5s)
X=Z, C, A
-
LXYp
LYp
dBX
Sound pressure level (SPL) 2
X=Z, C, A
Y=F, S, I
LXpk
Lpk
dBX
Instantaneous peak level
X=Z, C
-
Integrated acoustic levels
Wideband
PARAMETER
ABBREV.
DEFINITION
WEIGH.
FREQ.
WEIGH.
TEMP.
LXeq
Leq
dBX
Equivalent continuous level
X=Z, C, A
-
LXYmax
LYmx
dBX
Sound pressure maximum level (SPLmax)
X=Z, C, A
Y=F, S, I
LXYmin
LYmn
dBX
Sound pressure minimum level (SPLmin)
X=Z, C, A
Y=F, S, I
LXpkmax
Lpkmx
dBX
Peak maximum level
X=Z, C
-
Lnn
Li, i=1÷4
nn%
nn% percentile with nn=1÷99 3
A
F
A weighting
PARAMETER
ABBREV.
DEFINITION
WEIGH.
FREQ.
WEIGH.
TEMP.
LAE
LE
dBA
Exposure level throughout the measure-
ment time (SEL)
A
-
Dose %A
Dose
%
Dose percentage with programmable ex-
change rate, threshold level and criteria
A
-
Dose %A,d
Dose,d
%
Daily estimated dose with programmable
exchange rate, threshold level and criteria
A
-
LAIeq
LeqI
dBA
A weighted equivalent continuous level
with Impulse time constant
A
I
Other
PARAMETER
ABBREV.
DEFINITION
WEIGH.
FREQ.
WEIGH.
TEMP.
Overload %
OL%
Percentage of the measurement time in
which an overload occurs
-
-
2 The maximum reached level every 0.5s is displayed.
3 It is possible to program up to 4 different percentile levels.
- 100 -
V4.2
ACOUSTIC LEVELS (GRAPHIC DISPLAY)
Time Profile (with “Advanced Data Logger” option)
Wideband Levels
PARAMETER
ABBREV.
DEFINITION
WEIGH.
FREQ.
WEIGH.
TEMP.
LAFmax
LFmx
dBA
Sound pressure maximum level (SPLmax)
(SPLmax)
A
F
Statistical Analysis 4
PARAMETER
ABBREV.
DEFINITION
WEIGH.
FREQ.
WEIGH.
TEMP.
LXpk
LXpk
Peak level
X=Z, C
-
LXeq
LXeq
Equivalent level
X=Z, C, A
-
LXFp
LXFp
dBX
Sound pressure level with FAST time con-
stant (SPL)
X=Z, C, A
F
ACOUSTIC LEVELS THAT CAN BE STORED (WITH “ADVANCED DATA LOGGER” OPTION”)
Measurement group acoustic Levels
All the levels that can be displayed, described above, related to the SLM and PROFILE.
Report group acoustic levels
5 selected parameters
PARAMETER
ABBREV.
DEFINITION
WEIGH.
FREQ.
WEIGH.
TEMP.
LXeq
Leq
dBX
Equivalent continuous level
X=Z, C, A
LXYmax
LYmx
dBX
Sound pressure maximum level (SPLmax)
X=Z, C, A
Y=F, S, I
LXYmin
LYmn
dBX
Sound pressure minimum level (SPLmin)
X=Z, C, A
Y=F, S, I
LXpk
Lpk
dBX
Peak maximum level
X=Z, C
SEL
LE
dBA
Sound exposure level
A
Lnn
Li, i=1÷4
nn%
nn% percentile with nn=1÷995
LAIeq
LeqI
dBA
A weighted equivalent continuous level
with Impulse time constant
A
I
Statistical analysis on a selected parameter among the following
PARAMETER
ABBREV.
DEFINITION
WEIGH.
FREQ.
WEIGH.
TEMP.
LXeq
Leq
dBX
Equivalent continuous level
X=Z, C, A
LXF
LFp
dBX
Sound pressure maximum level (SPLmax)
X=Z, C, A
F
4 If the option Advanced Analyser is not activated, the statistic analysis is calculated on the A weighted sound pressure
with FAST time constant.
5 It is possible to program up to 4 different percentile levels.