Extech Instruments EA33 Operation & User’s Manual

Brand: Extech Instruments
Category: Measuring Instruments
Model: EA33 (Operation & User’s Manual)
File Type: PDF
File Size: —
Pages: 10
Text of Extech Instruments EA33 Operation & User’s Manual

EasyView Light Meter with Memory

Model EA33 www. .com [email protected] 1.800.561.8187

Congratulations on your purchase of the Extech EA33 EasyView Light Meter which measures light

level (luminance). This device measures up to 99,990 Fc and 999,990 Lux and stores up to 50 readings for easy recall. Careful use of this meter will provide years of reliable service. Meter Description 1. Light measurement display 2. Timer display 3. TIME button 4. HOLD button 5. Unit of measure button (Fc/lux) 6. Enter button 7

Extech Instruments EA33 With the sensor cover in place, press the button to turn the meter on. The meter will perform

a short self-test. 2. Remove the and place the sensor in the area to be measured. 3. Press the lux / fc button to select the desired units of measure. 4. Read the light level on the display. If the measurement is out of range, ation will appear. 5. When the meter is turned on, the elapsed timer starts. The elapsed ti

Extech Instruments EA33 MODE 2 Elapsed Timer with Hold

In this mode, measurements are made while the countdown timer is running. When the timer reaches zero, the last measurement is held on the display (see diagram). 1. Press the SET key; SEt 01 will display. 2. Press the down arrow key to change to the SEt 02 parameter. - 3. Press the ENTER key to call up the timer. 4. Use the up/down arrow keys to set the desired seconds. 5. Press the ENTER key to prog

Extech Instruments EA33 MODE 5 - Comparator Mode

In this mode the meter compares its measurements to one HIGH and one LOW limit. When the high limit is exceeded, the plus sign appears on the LCD. When the measurement falls below the low limit, the negative sign appears on the LCD. 1. 2. parameter. 3. Press the ENTER key to enter the Comparator setting mode. 4. Press the down arrow key to select the desired unit of measure (Lux or Fc). 5. Press ENTER to begi

Luminous Intensity Feature

Luminous intensity is the measurement of light from a light source at a known distance from the sensor. The unit of measure is candela (cd). 1. Press the cd/COMP key to begin (for lux measurement) feet (for Fc measurement) will appear. 2. Use the up/down arrow keys to set the distance from the light source to the light sensor. The distance range is 0.01 to 9.99 meters or feet. 3. The measurement must be mad

AUTO POWER OFF Feature

To conserve battery life, the EA33 automatically turns off after several minutes. To defeat this feature: 1. Turn the meter OFF. 2. Press and hold the HOLD key while turning the meter ON. 3. The AUTO POWER OFF feature will now be disabled. 4. Note that AUTO POWER OFF is re-enabled each time the meter is turned on. 5. Also note that AUTO POWER OFF is always disabled in the Integral Illuminance mode. www. .com in

Cleaning and storage

1. The white plastic sensor dome should be cleaned with a damp cloth when necessary. 2. Store the meter in an area with moderate temperature and humidity (refer to the operating and storage range in the specifications chart earlier in this manual). Battery Replacement When the battery power falls low, the battery symbol will appear on the lower left-hand side of the battery compartment screw and acce

Extech Instruments EA33 Multi-function LCD with a maximum display of 999900

Ranging Autoranging over five ranges Over range indication Spectral response CIE photopic (CIE human eye response curve) Spectral accuracy V 1 Cosine response 2 2%; Cosine corrected for angular incidence of light Measurement Repeatability ±2% Temperature coefficient ±0.1% per oC Sampling rate 5 times per second Memory 50 manual data points Photo dete

Extech Instruments Measuring Instruments Owners Instructions: