Condair Cube Operation Manual

Specifications

Brand: Condair
Category: Measuring Instruments
Model: Cube
File Type: PDF
File Size: —
Pages: 24

Read Online

Read the Condair Cube Operation Manual online using the built-in PDF viewer.

Text of the manual

This document and the information disclosed herein are proprietary data of Condair Group

AG. Neither this document, nor the information contained herein shall be reproduced, used, or disclosed to others without the written authorisation of Condair Group AG, except to the extent required for installation or maintenance of recipient's equipment. Liability Notice Condair Group AG does not accept any liability due to incorrect installation or operation of the equipment or due to the use of parts/components/equipment that are not authorised by Condair Group AG. Copyright Notice © Condair Group AG, All rights reserved. Technical modifications reserved.

Thank you for purchasing your Condair Cube

4 1.2 Notes on the operation manual 4 2 For your safety 5 3 Quickstart 6 4 Product overview 7 4.1 Components 7 4.2 Main screen view 8 4.2.1 Traffic light display of the measured values 9 4.3 Application overviews 10 4.3.1 View health chart 10 4.3.2 View historical readings 11 4.3.3 Settings 11 5 Calibration 12 5.1 Manual Calibration 12 5.2 Auto calibration 14 6 Time and date setting 15 7 Options 17 7.1 SD card 17 7.2 Power bank 17 7.3 Display on large screen 17 8 Scope of delivery 18 9 Disposal/Recycling 19 10� Product�specification� 20 10.1 Technical data Sensirion SCD40 20 10.1.1 CO2 Sensing Performance 20 10.1.2 Humidity Sensing Performance 20 10.1.3 Temperature Sensing Performance 20 11 Appendix 21 11.1 CE declaration of conformity 21

Thank you for purchasing your Condair Cube

The Condair Cube uses a sensor to detect CO2, relative humidity and temperature in the ambient environment and is primarily intended for indoor use. The Condair Cube incorporates the latest technical advances and meets all rec- ognized safety standards. Nevertheless, improper use of the Condair Cube may result in danger to the user or third parties. Further, Condair shall not be liable for any incorrect readings and any damage resulting therefrom. To ensure a safe, proper, and economical operation of the Condair Cube please observe and comply with all information contained in the present documentation as well as in the separate documentations of the components installed in the Condair Cube. If you have questions after reading this documentation, please contact your Condair representative. They will be glad to assist you. 1.2 Notes on the operation manual Limitation The subject of this operation manual is the Condair Cube. The various op- tions and accessories are only described as necessary for proper operation of the equipment. This operation manual is restricted to the operation of the Condair Cube and is meant for users of this device. Please note that some illustrations in this manual may include options and acces- sories that may not be standard or available in your country. Please check with your Condair representative for availability and specification details. Safekeeping Please keep this operation manual in a safe place, where it can be immediately accessed. If the equipment changes hands, the documentation must be passed on to the new owner If the documentation gets misplaced, please contact your Condair representative. Language versions This operation manual is available in various languages. Please contact your Condair representative for information.

For your safety

2 For your safety General Every person operating the Condair Cube must have read and understood these instructions before use. Knowledge of the contents of this manual for the Condair Cube is a basic require- ment for protecting against any kind of danger, to prevent faulty installations, and to operate the Condair Cube safely and correctly. All symbols, signs and markings applied to the components of the Condair Cube must be observed and kept in readable state. Intended use The Condair Cube is intended exclusively for indoor use. Any other type of ap- plication is considered as not conforming with the intended purpose and may lead to deviating results being displayed by the Condair Cube. Any warranty claim will be voided if the device is not used for its intended purpose. Intended use also includes observing all the information contained in this manual. Prohibited�modifications�to�the�unit No�modifications�must�be�undertaken on the Condair Cube without the express written consent of the manufacturer. For the replacement of defective components only use original accessories and spare parts available from your Condair representative.

1. Switch on the Condair Cube: To do this, press the power button inside

the cavity on the lower, left side of the cube (see product overview). Note: When the Condair Cube is plugged in via the USB cable, it switches on automatically. Fully charge the device before using it for the first time. 2. Place the Condair Cube upright in the room where you want to check the room air. The Cube must be placed as shown in the figure below, so that the sensor on the back of the Cube is reached by the room air and the excess heat can escape through the ventilation slots.  3. Once the Condair Cube is in place, it requires an adjustment time of ap- prox. 20 minutes until all values are displayed correctly. In particular, the temperature may deviate during the first phase after switching on, since the sensor is not yet at operating temperature. 4. When the Condair Cube is put into operation for the first time, it must be calibrated. This procedure is described in chapter 5.

1 housing from 2 parts

– Display with rechargeable battery (M5core2) – Sensor (Sensirion SCD40) – USB charging cable – Optional Micro SD card (not included in delivery) Button1 Button 2 LED (green) Ventilation slots Slot for Micro SD card Sensor Housing Display Power button USB charging cable port

Main screen view

ppm %RH °C CO2 10:15 21.08.2021 85% ppm %RH °C 28 21 1092 CO2 CO2 value Time Date Battery status Menu Humidity value Temperature value Time/Date: Shows the current time and date. Battery status: Shows the current state of charge of the battery. The display ranges from 0-100 %. CO2 value: Displays the current CO2 content in the air. The value is given in ppm (parts per million). Temperature value: Displays the current temperature value in °C (Celsius). Humidity value: Displays the current relative humidity of the ambient air. Indication in %RH. Menu bar: Via the menu bar you can access the various submenus. – Health chart – Settings – Graphical view of the temporal courses of CO2, humidity and temperature Button 1: Turn on screen and increase brightness Button 2: Turn off screen and reduce brightness LED: When the display is darkened via button 2, the LED lights up green as an operating indicator.

The circles around the values show the current state of the value in the manner

of a traffic light system: Green – Healthy (Everything in the optimal range) Yellow – Acceptable (Caution: Need for action) Red – Critical (Critical readings) CO2 Green circle: CO2 content of the ambient air is optimal. (value is ≤ 800 ppm) Yellow circle: CO2 content in the ambient air is increased. Ventilation recom- mended. (value is between > 800 and ≤ 1200 ppm) Red circle: CO2 content in the ambient air is critical. Ventilation necessary. (value is > 1200 ppm) Relative humidity Green circle: Relative humidity of the ambient air is at an optimum. (value is between ≥ 40 and ≤ 60 % rH) Yellow circle: Relative humidity of the ambient air is rather low or rather high. (value is between ≥ 30 and < 40 % rH or > 60 and < 70 % rH) Red circle: Relative humidity of the ambient air is too low or too high. (value is < 30 % rH or ≥ 70 % rH) Temperature Green circle: Comfortable ambient temperature. (value is between ≥ 20 and < 23.5 °C) Yellow circle: Ambient temperature is rather low or high. (value is ≥ 17 and < 20 °C or ≥ 23.5 and < 27 °C) Red circle: The comfortable temperature range is undershot or overshot. (value is < 17 °C or ≥ 27 °C) Note: Please note that the sensor requires an adjustment time of up to 20 minutes after switching on, a reset or when moving to a different location, until the temperature reading is displayed correctly.

4.3.1 View health chart

18°C 22°C 26°C 20% 40% 60% 80% ppm CO2 800 1200 10:15 21.08.2021 85% CO2 display in bar format Currently measured value Range "healthy" Range "acceptable" Temperature Air humidity The range "healthy" represents the air quality, which provides increased health protection and is perceived as pleasant. The ranges "healthy, "acceptable" and "critical" are distinguished. The aim is that the current measured value (red point) of the room air is in the inner green diamond and thus in the "healthy" range. In the upper area of the view, the currently measured CO2 content of the room air is displayed in a bar view. If the current measured value is outside the green range, the room should be ventilated to reduce the CO2 content in the room. Note: If the currently measured value is displayed as a gray dot at the edge of the diagram, it indicates that the measured value is outside the displayed range.

4.3.2 View historical readings

18 22 26 h 1 2 3 4 °C °C21 10:15 21.08.2021 85% Currently measured value Historical measured values This view shows the course of the measured room temperature over a maximum of 4 hours. For every 10 minutes of measurement, a bar is displayed which shows the average value measured. This is colored according to the described traffic light system. See chapter chapter 4.2.1. If the bar is yellow or red, the temperature was too low or too high on average over the last 10 minutes, taking the health chart into account. A 4-hour chart is available for all measured values of CO2, temperature and humidity. These views are accessible via the arrow keys on the side of the circle. 4.3.3 Settings CO2 10:15 21.08.2021 85% Alarm settings Calibration CO2 Time and date settings Via the 3 icons you can call the different submenus: – Alarm settings – Calibration CO2 – Time and date settings Tap once on the respective icon to enter the submenus.

For accurate measurement results of the CO2 content, the Condair Cube must be

calibrated regularly. Two different calibration options are available: – Auto calibration – Manual calibration 5.1 Manual Calibration To calibrate the sensor manually, proceed as follows: 1. Menu → Settings → Calibration CO2 CO2 10:15 21.08.2021 85% CO2 CO2 Manual Auto 2. Place the sensor outdoors, for example outside the window in fresh air. Outside air is used as a reference, since it normally has a value of ap- prox. 400 ppm. Start 10:15 21.08.2021 85% 3. Press "Start" to start calibration.

4. The measurement takes approx. 3 minutes This is displayed graphically

with a bar graph and a percentage. The sensor resets its CO2 value several times during this time. ppm CO2 10:15 21.08.2021 85% 21 44% 5. After the calibration is completed, the Cube automatically switches to the main page. ppm %RH °C CO2 10:15 21.08.2021 85% ppm %RH °C 28 21 1092 CO2 6. Check the CO2 value. If it is within the range of 400 +/-25 ppm, the calibration was successful. If the value is above or below this, repeat the calibration. Important: The sensor must be calibrated during the first start-up. Also, if you place the Cube in a new geographical location, it must be recalibrated for accurate measurement results.

By default, the auto calibration is activated, whereby the sensor calibrates its

value independently. It is important here that the sensor is in operation as permanently as possible and is exposed to fresh outside air at least once a week (approx. 10 min). During auto-calibration, the lowest measured value over one week is taken as the reference value and recalibrated to 400 ppm. Note: This setting is also saved when the cube is switched off. 1. Menu → Settings → Calibration CO2 2. Auto calibration 3. Enable or disable auto calibration If the Condair Cube is not operated permanently, it is recommended to deactivate the auto calibration and to perform a manual calibration when using the Condair Cube. CO2 10:15 21.08.2021 85% CO2 CO2 Manual Auto OFF CO2 Auto

Time and date setting

6 Time and date setting The Condair Cube has an internal RTC (RealTimeClock) which is used for the time and date display. Time and date run for approx. 2 months when the unit is switched off and without charging the Cube. Afterwards, both must be reset. Follow the procedure below to reset the time and date: 1. Menu → Settings → Time and date settings CO2 10:15 21.08.2021 85% 2. Set the desired time using the up/down arrow keys. The time format is defined as follows: hh:mm:ss (Hours:Minutes:Seconds) Ok : : 10:15 21.08.2021 85% 10 15 33 3. Confirm the entry with "Ok".

Time and date setting

4. Enter the desired date in the next screen. The date format is defined as follows: dd:mm:yy (Day:Month:Year) Ok : : 10:15 21.08.2021 85% 21 08 21 5. Confirm the entry with "Ok" to complete the date entry.

An optional memory card can be integrated via the SD card slot on the right side of

the Cube. If a memory card is detected during the start-up process, the measured values are saved to the memory card every minute. With the help of Excel, this data can be displayed and analyzed. Recommended memory card: SANDISK microSDXC 16GB+AD memory card (16 GB, 98 MB/s, grey/red) or similar. 7.2 Power bank To extend the battery life of the Condair Cube, you can use any power bank with a 5V USB connection. 7.3 Display on large screen It is possible to connect the Condair Cube to an external monitor via a PC. To do this, connect the Condair Cube to the PC via the USB cable and install the necessary drivers. Then open the Condair Cube Tool to start the display on an external monitor. The values are updated at intervals of 30 seconds. Note: The minimum resolution for correct display of the values is 1024x768. Note that individual setting of zoom factors may have the negative effect on the display.

Components no longer in use cannot be disposed in domestic waste. Please

dispose of the individual components in accordance with local regulations at the authorised collecting point. If you have any questions, please contact the responsible authority or your local Condair representative. Thank you for your contribution to environmental protection.

10.1 Technical data Sensirion SCD40

10.1.1 CO2 Sensing Performance Default conditions of 25 °C, 50 % RH, ambient pressure 1013 mbar, default pe- riodic measurement and 3.3 V supply voltage apply to values in the table below, unless otherwise stated. Parameter Conditions Value CO2 output range – 0 – 40'000 ppm CO2 measurement accuracy 400 ppm – 2'000 ppm ± (50 ppm + 5 % of reading) Repeatability Typical ± 10 ppm Response time T63 %, typical 60 s Accuracy drift per year with au- tomatic selfcalibration algorithm enabled Typical ± (5 ppm + 0.5 % of reading) 10.1.2 Humidity Sensing Performance Parameter Conditions Value Humidity measurement range – 0 % RH – 100 % RH Accuracy (typ.) 15 °C – 35 °C, 20 % RH – 65 % RH ± 6 % RH -10 °C – 60 °C ± 9 % RH Repeatability Typical ± 0.4 % RH Response time T63 %, typical 90 s Accuracy drift – < 0.25 % RH / year 10.1.3 Temperature Sensing Performance Parameter Conditions Value Temperature measurement range – -10 °C – 60 °C Accuracy (typ.) 15 °C – 35 °C ± 0.8 °C -10 °C – 60 °C ± 1.5 °C Repeatability – ± 0.1 °C Response time T63 %, typical 120 s Accuracy drift – < 0.03 °C / year

11.1 CE declaration of conformity

Condair Group AG Gwattstrasse 17 8808 Pfäffikon, Switzerland Tel. +41 55 416 61 11, Fax +41 55 588 00 07 [email protected], www.condairgroup.com EC Konformitätserklärung Declaration of conformity Déclaration de conformité Wir, Condair Group AG CH-8808 Pfäffikon SZ erklären in alleiniger Verantwortung, dass das Produkt We, Condair Group AG CH-8808 Pfäffikon SZ declare under our sole responsibility, that the product Nous, Condair Group AG CH-8808 Pfäffikon SZ déclarons sous notre seule responsabilité, que le produit Condair Cube auf das sich diese Erklärung bezieht, mit den folgenden Normen oder normativen Dokumenten übereinstimmt to which this declaration relates is in conformity with the following standards or other normative standards auquel se réfère cette déclaration est conforme aux normes ou autres documents normatifs EN 55032 EN 61000-3-2 EN 61000-3-3 EN 301489-1 V2.2.3 EN 301489-17 V3.2.3 EN 300328 V2.2.2 und den Bestimmungen der folgenden Richtlinien entspricht and is corresponding to the following provisions of directives et est conforme aux dispositions des directives suivantes RED-Richtlinie 2014/53/EU EMC-Richtlinie 2014/30/EU ROHS Neufassung der Richtlinie 2011/65/EG RED directive 2014/53/EU EMC directive 2014/30/EU ROHS recast directive 2011/65/EC Directive RED 2014/53/EU Directive EMC 2014/30/EU Directive de refonte ROHS 2011/65/CE Pfäffikon , September 01, 2021 Condair Group AG Eric Roth Head of Engineering Adrian Spörri Corporate Systems Manager 2603018 DE/EN/FR 2109

Condair Group AG

Gwattstrasse 17, 8808 Pfäffikon SZ, Switzerland Phone +41 55 416 61 11, Fax +41 55 588 00 07 [email protected], www.condairgroup.com CONSULTING, SALES AND SERVICE: CH94/0002.00

Other Measuring Instruments Manuals