This supplement contains information
necessary to ensure the accuracy of
the above manual.
On page 10, prior to Basic Operation, add the following:
Charging the Internal Battery
Before you use the 1735 Logger, charge the internal battery as
follows:
1. Switch the 1735 Battery Charger to either 115V or 230V as
appropriate.
2. With the Logger power off, attach the BC1735 battery
charger to a power outlet and then to the 1735.
3. Charge the 1735 for 5 hours before using it for the first time.
4. On subesquent use, turn the Logger power on before you
connect the BC1735 Battery Charger.
This helps insure that the fast charge mode is activated. If the
Logger does not turn on due to a discharged battery, recharge the
battery for 5 hours with the Logger power off as described in
steps 2 and 3 above.
Change #2
On page 52, under EMC replace the Emission, with the
following:
Emission:
IEC/EN 61326-1:1997+A1:1998+A2:2000 class B
Change #3
On page 2, add the following symbol to Table 1:
-
Do not apply around or remove from the Hazardous
Live conductors
On page 4, replace Table 3 with the following:
Table 3. Optional Accessories
Description
Accessory
I1A/10A CLAMP PQ3, 3-PHASE
1A/10A MINI CURRENT CLAMP
SET FOR PQ
I1A/10A CLAMP PQ3
I1A/10A CLAMP PQ4, 4-PHASE
1A/10A MINI CURRENT CLAMP
SET FOR PQ
I1A/10A CLAMP PQ4
I5A/50A CLAMP PQ3, 3-PHASE
5A/50A MINI CURRENT CLAMP
SET FOR PQ
I5A/50A CLAMP PQ3
I5A/50A CLAMP PQ4, 4-PHASE
5A/50A MINI CURRENT CLAMP
SET FOR PQ
I5A/50A CLAMP PQ4
I20/200A CLAMP PQ3, 3-PHASE
20A/200A MINI CURRENT
CLAMP SET FOR PQ
I20/200A CLAMP PQ3
I20/200A CLAMP PQ4, 4-PHASE
20A/200A MINI CURRENT
CLAMP SET FOR PQ
I20/200A CLAMP PQ4
3000/6000A FLEX
4,3000A/6000A 36-INCH FLEXI
PROBE 4 PHASE
3000/6000A FLEX 4
C435 HARDCASE
TRANSIT CASE
On page 21, add the following Note under the Warning:
Note
When using either flexi-probes or current clamp
sets, make sure the arrow on the current probe
points towards the load.
On page 21, under Connecting the Power Logger to the
network add the following text and Figure:
•
To avoid shock or personal injury, keep fingers
behind the tactile barrier, see the figure.
cab01f.eps
Using the Optional Microclamps
On page 22, replace the table with the following:
Network
Test Leads
L1
A (L1)
Not connected
B (L2)
Not connected
C (L3)
N
N
On page 26, change the figure title to:
Figure 8. Three-Phase Delta Δ Connections-Blondel
(Aron, Two-Element Delta)
And add the following figure and title:
CURRENT INPUT
CURSOR
SCOPE
OFF
HARMONICS
POWER
EVENTS
METER
V A Hz
ENTER
MAX 30V
POWER LOGGER
1735
SAVE
SCREEN
ESC
MENU
VOLTAGE INPUT
HOLD
RUN
RECORD
MEASURE
600V CAT
L1
L2
L3
N
L1
L1
L2
L3
L2 L3
N
edx045.eps
Figure 9. Three-Phase Delta Δ Connections-Blondel
(Aron, Three-Element Delta)
On page 49, under Power Measurements, delete the total
active power Blondel (Aron) formula and add:
3
1
P
P
Ptot
+
=
Total Active Power-Blondel (Aron)
2
2
tot
tot
tot
P
S
Q
−
=
Total Reactive Power-Blondel (Aron)
3
2
3
2
2
2
1
2
31
2
23
12 2
I
I
I
V
V
V
Stot
+
+
⋅
+
+
=
Total Apparent Power
)
(
3
1
2
I
I
I
+
=−
Blondel (Aron) circuit
On page 50, underTotal Harmonic Distortion, delete the
Total Harmonic Distortion formula and add:
100%
1
)
(
50
2
2
×
= ∑
=
V
V
THD
h
h
Total Harmonic Distortion
V1 to RMS of the fundamental
Vh to RMS of the h-th harmonic
∑
=
⎟⎟ ⋅
⎠
⎞
⎜⎜
⎝
⎛
=
−
50
1
2
2
h
RMS
h
h
I
I
factor
k
k-factor
Ih … h-th harmonic (amps)
IRMS … RMS amps
h … harmonic order