Fluke 1735 Manual Supplement

Text of the manual

© 2006-2008 Fluke Corporation. All rights reserved. Printed in the U.S.A.x

Manual Title: 1735 Users Supplement Issue: 4 Part Number: 2560330 Issue Date: 3/08 Print Date: March 2006 Page Count: 5 Revision/Date: 1, 4/06 This supplement contains information necessary to ensure the accuracy of the above manual.

On page 10, prior to Basic Operation, add the following:

Charging the Internal Battery Before you use the 1735 Logger, charge the internal battery as follows: 1. Switch the 1735 Battery Charger to either 115V or 230V as appropriate. 2. With the Logger power off, attach the BC1735 battery charger to a power outlet and then to the 1735. 3. Charge the 1735 for 5 hours before using it for the first time. 4. On subesquent use, turn the Logger power on before you connect the BC1735 Battery Charger. This helps insure that the fast charge mode is activated. If the Logger does not turn on due to a discharged battery, recharge the battery for 5 hours with the Logger power off as described in steps 2 and 3 above. Change #2 On page 52, under EMC replace the Emission, with the following: Emission: IEC/EN 61326-1:1997+A1:1998+A2:2000 class B Change #3 On page 2, add the following symbol to Table 1: - Do not apply around or remove from the Hazardous Live conductors

On page 4, replace Table 3 with the following:

Table 3. Optional Accessories Description Accessory I1A/10A CLAMP PQ3, 3-PHASE 1A/10A MINI CURRENT CLAMP SET FOR PQ I1A/10A CLAMP PQ3 I1A/10A CLAMP PQ4, 4-PHASE 1A/10A MINI CURRENT CLAMP SET FOR PQ I1A/10A CLAMP PQ4 I5A/50A CLAMP PQ3, 3-PHASE 5A/50A MINI CURRENT CLAMP SET FOR PQ I5A/50A CLAMP PQ3 I5A/50A CLAMP PQ4, 4-PHASE 5A/50A MINI CURRENT CLAMP SET FOR PQ I5A/50A CLAMP PQ4 I20/200A CLAMP PQ3, 3-PHASE 20A/200A MINI CURRENT CLAMP SET FOR PQ I20/200A CLAMP PQ3 I20/200A CLAMP PQ4, 4-PHASE 20A/200A MINI CURRENT CLAMP SET FOR PQ I20/200A CLAMP PQ4 3000/6000A FLEX 4,3000A/6000A 36-INCH FLEXI PROBE 4 PHASE 3000/6000A FLEX 4 C435 HARDCASE TRANSIT CASE On page 21, add the following Note under the Warning: Note When using either flexi-probes or current clamp sets, make sure the arrow on the current probe points towards the load.

On page 21, under Connecting the Power Logger to the

network add the following text and Figure: • To avoid shock or personal injury, keep fingers behind the tactile barrier, see the figure. cab01f.eps Using the Optional Microclamps

On page 22, replace the table with the following:

Network Test Leads L1 A (L1) Not connected B (L2) Not connected C (L3) N N On page 26, change the figure title to: Figure 8. Three-Phase Delta Δ Connections-Blondel (Aron, Two-Element Delta) And add the following figure and title: CURRENT INPUT CURSOR SCOPE OFF HARMONICS POWER EVENTS METER V A Hz ENTER MAX 30V POWER LOGGER 1735 SAVE SCREEN ESC MENU VOLTAGE INPUT HOLD RUN RECORD MEASURE 600V CAT L1 L2 L3 N L1 L1 L2 L3 L2 L3 N edx045.eps Figure 9. Three-Phase Delta Δ Connections-Blondel (Aron, Three-Element Delta)

On page 49, under Power Measurements, delete the total

active power Blondel (Aron) formula and add: 3 1 P P Ptot + = Total Active Power-Blondel (Aron) 2 2 tot tot tot P S Q − = Total Reactive Power-Blondel (Aron) 3 2 3 2 2 2 1 2 31 2 23 12 2 I I I V V V Stot + + ⋅ + + = Total Apparent Power ) ( 3 1 2 I I I + =− Blondel (Aron) circuit On page 50, underTotal Harmonic Distortion, delete the Total Harmonic Distortion formula and add: 100% 1 ) ( 50 2 2 × = ∑ = V V THD h h Total Harmonic Distortion V1 to RMS of the fundamental Vh to RMS of the h-th harmonic ∑ = ⎟⎟ ⋅ ⎠ ⎞ ⎜⎜ ⎝ ⎛ = − 50 1 2 2 h RMS h h I I factor k k-factor Ih … h-th harmonic (amps) IRMS … RMS amps h … harmonic order