Earth Ground Clamp

Calibration Adjustment 9 Table 4. Calibration Adjustment Steps (cont.) Step Action Meter Function Source Applied Value Adjust Expected Result/Instruction 13 Enter Cal Mode Harmonic displays 14 ~ A 1 A / 60 Hz Press Honce when reading is stable. [1] 15 ~ A 3 A / 60 Hz Press Honce when reading is stable. [1] 16 ~ A Calibrator and 6-in. wire loop 10 A / 60 Hz Press Honce when reading is stable. [1] 17 ~ A Calibrator and 3 turns copper magnet wire, 6-in. diameter 30 A / 60 Hz Press Honce when reading is stable. [1] Note [1] Disregard reading accuracy during these calibration ...

Earth Ground Clamp

10 IEC 6LR61 BATTERY IEC 6LR61 BATTERY Press tab and lift to remove battery holder from PCA. Do not remove shield during adjustment. VR1 VR2 Shield removed for demonstration purposes only. Do not remove during adjustment. VR10 VR6 is not used VR3 VR20 VR11 fcy02.eps Figure 2. Disassembly and Adjustment Locations ...

Earth Ground Clamp

Limited Warranty and Limitation of Liability 11 Limited Warranty and Limitation of Liability Each Fluke product is warranted to be free from defects in material and workmanship under normal use and service. The warranty period is one year and begins on the date of shipment. Parts, product repairs, and services are warranted for 90 days. This warranty extends only to the original buyer or end-user customer of a Fluke authorized reseller, and does not apply to fuses, disposable batteries, or to any product which, in Fluke's opinion, has been misused, altered, neglected, contam ...

Earth Ground Clamp

10 IEC 6LR61 BATTERY IEC 6LR61 BATTERY Press tab and lift to remove battery holder from PCA. Do not remove shield during adjustment. VR1 VR2 Shield removed for demonstration purposes only. Do not remove during adjustment. VR10 VR6 is not used VR3 VR20 VR11 fcy02.eps Figure 2. Disassembly and Adjustment Locations ...

Earth Ground Clamp

Performance Tests 7 Table 3. 1630 Accuracy Tests (cont.) Step UUT Function Source Applied Value Display Reading Limits/Instruction 22 Press F and set the HI alarm to 40 Ω 23 Press F to return to measurement function 24 50 Ω Clamp Beeps and Display shows HI__ 25 Press F and set the LO alarm to 30 Ω 26 Press F to return to measurement function 27 Decade Resistance Box and 6-in. diameter wire loop 10 Ω Clamp Beeps and Display shows LO__ 28 Continuity Press F until SEC is shown on display. Press the E and D to check if value can be incremented. 29 Decade Resistance Box and 6 ...

Earth Ground Clamp

2 Safety Information XW Safety Information To avoid possible electric shock or personal injury and ensure safe operation and service of the Clamp, follow these instructions: • Read the operating instructions before use and follow all safety instructions. • Use the Clamp only as specified in the operating instructions; otherwise, the Instrument’s safety features may be impaired. • Adhere to local and national safety codes. Individual protective equipment must be used to prevent shock and arc blast injury where hazardous live conductors are exposed. • Before eac ...

Fluke 1630 user guide recommended for: 87V Ex, 287, 8020A, Micro-Bath 7102, 79.

⚠️ Download Safety