Binder FDL 115 Operating Manual
Read Online
Read the Binder FDL 115 Operating Manual online using the built-in PDF viewer.
Text of the manual
Art. No. 7001-0127
Operating Manual
Translation of the Original Operating Manual
APT.line™ FDL 115
Safety / paint drying ovens
for drying of limited quantities of solvents
with microprocessor program controller RD3
Model
Art. No.
FDL 115 (E2)
9010-0269, 9110-0269
BINDER GmbH
Address
Post office box 102
D-78502 Tuttlingen
Tel.
+49 7462 2005 0
Fax
+49 7462 2005 100
Internet
http://www.binder-world.com
E-mail
[email protected]
Service Hotline
+49 7462 2005 555
Service Fax
+49 7462 2005 93 555
Service E-Mail
[email protected]
Service Hotline USA
+1 866 885 9794 or +1 631 224 4340
Service Hotline Spain
+34 9492 677 23
Service Hotline Asia Pacific
+852 39070500 or +852 39070503
Service Hotline Russia and CIS +7 495 98815 17
FDL (E2) 03/2011
page 2/68
EG - KONFORMITÄTSERKLÄRUNG
EC - DECLARATION OF CONFORMITY
CE - DECLARATION DE CONFORMITE
Anbieter / Supplier / Fournisseur:
BINDER GmbH
Anschrift / Address / Adresse:
Im Mittleren Ösch 5, D-78532 Tuttlingen
Bezeichnung der Maschine /
Denomination of the machine /
Dénomination de la machine:
Sicherheits- / Lacktrockenschrank
Safety / paint drying oven
Armoire séchante de sécurité / pour peintures et vernis
Typenbezeichnung / Type / Type:
FDL 115
Die oben beschriebene Maschine ist konform mit folgenden EG-Richtlinien (gemäß Veröf-
fentlichung im Amtsblatt der europäischen Kommission):
The machine described above is in conformity with the following EC guidelines (as published in
the Official Journal of the European Union):
La machine décrite ci-dessus est conforme aux directives CE suivantes (selon leur publication
dans le Journal officiel de l’Union européenne):
Maschinenrichtlinie 2006/42/EG
Machinery directive 2006/42/EC
Directive Machines 2006/42/EC
Richtlinie 2006/42/EG des Europäischen Parlaments und des Rates
vom 17. Mai 2006 über Maschinen und zur Änderung der Richtlinie
95/16/EG (Neufassung)
Directive 2006/42/EC of the European Parliament and of the Council
of 17 May 2006 on machinery, and amending Directive 95/16/EC
(recast)
Directive 2006/42/CE du Parlement Européen et du Conseil du 17
mai 2006 relative aux machines et modifiant la directive 95/16/CE
(refonte)
EMV-Richtlinie 2004/108/EG
EMC Directive 2004/108/EC
Directive CEM 2004/108/CE
Richtlinie 2004/108/EG des Europäischen Parlaments und des
Rates vom 15. Dezember 2004 zur Angleichung der Rechtsvor-
schriften der Mitgliedstaaten über die elektromagnetische Ver-
träglichkeit und zur Aufhebung der Richtlinie 89/336/EWG.
Directive 2004/108/EC of the European Parliament and of the Coun-
cil of 15 December 2004 on the approximation of the laws of the
Member States relating to electromagnetic compatibility and repeal-
ing Directive 98/336/EEC.
Directive 2004/108/CE du Parlement Européen et du Conseil du 15
décembre 2004 relative au rapprochement des législations des États
membres concernant la compatibilité électromagnétique et abro-
geant le directive 98/336/CEE.
ATEX-Richtlinie 94/9/EG
ATEX Directive 94/9/EC
Directive ATEX 94/9/CE
Richtlinie 94/9/EG des Europäischen Parlaments und des Rates zur
Angleichung der Rechtsvorschriften der Mitgliedstaaten für Geräte
und Schutzsysteme zur bestimmungsgemäßen Verwendung in ex-
plosionsgefährdeten Bereichen.
Directive 94/9/EC of the European parliament and the council of 23
March 1994 on the approximation of the laws of the Member States
concerning equipment and protective systems intended for use in
potentially explosive atmospheres
Directive 94/9/CE du parlement européen et du conseil du 23 mars
1994 concernant le rapprochement des législations des États mem-
bres pour les appareils et les systèmes de protection destinés à être
utilisés en atmosphères explosibles.
1 / 3
FDL (E2) 03/2011
page 3/68
Die oben beschriebene Maschine entspricht aufgrund ihrer Konzipierung und Bauart sowie in der von uns
in Verkehr gebrachten Ausführung den einschlägigen grundlegenden Sicherheits- und Gesundheitsan-
forderungen der genannten EG-Richtlinien.
The machine described above is conform to the mentioned EC directives in regard to the relevant safety
and health demands due to its conception and its style of construction as well as to the version put onto
market by us.
La machine décrite ci-dessus correspond aux demandes de sécurité et de santé des directives citées de
la Communauté Européenne due à sa conception et construction et dans la réalisation mise sur le
marché par nous.
Die oben beschriebene Maschine trägt entsprechend die Kennzeichnung CE.
The machine described above, corresponding to this, bears the CE-mark.
La machine décrite ci-dessus, en correspondance, porte l’indication CE.
Die oben beschriebene Maschine ist konform mit folgenden harmonisierten Normen:
The machine described above is in conformity with the following harmonized standards:
La machine décrite ci-dessus est conforme aux normes harmonisées suivantes:
Sicherheit / safety / sécurité:
EN 61010-1:2010
Sicherheitsbestimmungen für elektrische Mess-, Steuer-, Regel- und
Laborgeräte – Teil 1: Allgemeine Anforderungen (DIN EN 61010-
1:2002 + Berichtigung 1:2002 + Berichtigung 2:2004)
Safety requirements for electrical equipment for measurement, con-
trol, and laboratory use – Part 1: General requirements (IEC 61010-
1:2010, BS EN 61010-1:2010)
Règles de sécurité pour appareils électriques de mesurage, de régu-
lation et de laboratoire – Partie 1: Prescriptions générales (CEI 61010-
1:2010, NF EN 61010:2011)
EN 61010-2-010:2003
Sicherheitsbestimmungen für elektrische Meß-, Steuer-, Regel- und
Laborgeräte – Teil 2-010: Besondere Anforderungen an Laborgeräte
für das Erhitzen von Stoffen (DIN EN 61010-2-010:2004)
Safety requirements for electrical equipment for measurement, con-
trol, and laboratory use – Part 2-010: Particular requirements for labo-
ratory equipment for the heating of materials (IEC 61010-2-10:2005,
BS EN 61010-2-10:2003)
Règles de sécurité pour appareils électriques de mesurage, de régu-
lation et de laboratoire – Partie 2-010 : Prescriptions particulières pour
appareils de laboratoire utilisés pour l’échauffement des matières
(CEI 61010-2-10:2003, NF EN 61010-2-10:2005)
EN ISO 12100:2010
Sicherheit von Maschinen - Allgemeine Gestaltungsleitsätze - Risi-
kobeurteilung und Risikominderung (DIN EN ISO 12100:2011)
Safety of machinery - General principles for design - Risk assessment
and risk reduction (BS EN ISO 12100:2010)
Sécurité des machines - Principes généraux de conception -
Appréciation du risque et réduction du risque (NF EN ISO
12100:2010)
EN ISO 13732-1:2008
Ergonomie der thermischen Umgebung - Bewertungsverfahren für
menschliche Reaktionen bei Kontakt mit Oberflächen. Teil 1: Heiße
Oberflächen (DIN EN ISO 13732-1:2008)
Ergonomics of the thermal environment - Methods for the assessment
of human responses to contact with surfaces. Part 1: Hot surfaces (BS
EN ISO 13732-1: 2008)
Ergonomie des ambiances thermiques - Méthodes d'évaluation de la
réponse humaine au contact avec des surfaces. Partie 1: Surfaces
chaudes (NF EN ISO 13732-1: 2008)
2 / 3
FDL (E2) 03/2011
page 4/68
Sicherheit / safety / sécurité:
EN 60204-1:2006 + A1:2009
Sicherheit von Maschinen. Elektrische Ausrüstung von Maschinen.
Teil 1: Allgemeine Anforderungen (DIN EN 60204-1:2007 + A1:2009)
Safety of machinery. Electrical equipment of machines. General re-
quirements (IEC 60204-1:2005 mod. + A1:2008, BS EN 60204-1:2006
+ A1:2009)
Sécurité des machines - Équipement électrique des machines - Partie
1 : règles générales (CEI 60204-1:2005 mod. + A1:2008, NF EN
60204-1:2006 + A1:2009)
EN 1539:2009
Trockner und Öfen, in denen brennbare Stoffe freigesetzt werden.
Sicherheitsanforderungen (DIN EN 1539: 2010)
Dryers and ovens, in which flammable substances are released.
Safety requirements (BS EN 1539:2009)
Séchoirs et fours dans lesquels se dégagent des substances inflam-
mables. Prescriptions de sécurité (NF EN 1539 : 2010)
EMV / EMC / CEM:
EN 61326-1:2006
+ Corr. 1:2008 + Corr. 2:2010
Elektrische Mess-, Steuer-, Regel- und Laborgeräte - EMV-
Anforderungen - Teil 1: Allgemeine Anforderungen (DIN EN 61326-
1:2006 + Berichtigung 1:2008 + Berichtigung 2:2011)
Electrical equipment for measurement, control and laboratory use -
EMC requirements - Part 1: General requirements (IEC 61326-1:2005
+ Corr. 1:2008 + Corr. 2:2010, BS EN 61326-1:2006+ A1:2008)
Matériel électrique de mesure, de commande et de laboratoire - Exi-
gences relatives à la CEM - Partie 1: Exigences générales (CEI
61326-1:2005 + AC1:2008, NF EN 61326-1:2006 mod.)
EN 61326-2-2:2006
Elektrische Mess-, Steuer-, Regel- und Laborgeräte – EMV-
Anforderungen. Teil 2-2: Besondere Anforderungen - Prüfanordnung,
Betriebsbedingungen und Leistungsmerkmale für ortsveränderliche
Prüf-, Mess- und Überwachungsgeräte in Niederspannungs-
Stromversorgungsnetzen. (DIN EN 61326-2-2:2006)
Electrical equipment for measurement, control and laboratory use –
EMC requirements. Part 2-2: Particular requirements - Test configura-
tions, operational conditions and performance criteria for portable test,
measuring and monitoring equipment used in low-voltage distribution
systems. (IEC 61326-2-2:2005, BS EN 61326-2-2:2006)
Matériel électrique de mesure, de commande et de laboratoire – Exi-
gences relatives à la CEM. Partie 2-2: Exigences particulières - Con-
figurations d’essai, conditions de fonctionnement et critères d’aptitude
à la fonction des matériels portatifs d’essai, de mesure et de surveil-
lance utilisés dans des systèmes de distribution basse tension. (CEI
61326-2-2:2005 + AC1:2007, NF EN 61326-2-2:2006)
D-78532 Tuttlingen, 30.03.2011
BINDER GmbH
P. M. Binder
Geschäftsführender Gesellschafter
Managing Director
Directeur général
Dr. H. von Both
Leiter F & E und Dokumentationsbevollmächtigter
Director R & D and documentation representative
Chef de service R&D et autorisé de documentation
3 / 3
FDL (E2) 03/2011
page 5/68
Certificate for the GS mark of conformity of the “Deutsche Gesetzliche Unfallversicherung e.V.”
(German Social Accident Insurance (DGUV)
FDL (E2) 03/2011
page 7/68
Contents
1.
SAFETY..................................................................................................................9
1.1
Legal considerations ...........................................................................................................................9
1.2
Structure of the safety instructions......................................................................................................9
1.2.1
Signal word panel......................................................................................................................9
1.2.2
Safety alert symbol..................................................................................................................10
1.2.3
Pictograms...............................................................................................................................10
1.2.4
Word message panel structure ...............................................................................................11
1.3
Localization / position of safety labels on the unit.............................................................................11
1.4
Type plate .........................................................................................................................................12
1.5
General safety instructions on installing and operating the safety / paint drying oven FDL.............13
1.6
Intended use .....................................................................................................................................14
1.7
Operating instructions .......................................................................................................................14
1.8
Measures to prevent accidents.........................................................................................................15
1.9
Important points to consider before commissioning .........................................................................16
1.9.1
Technical ventilation / permissible load...................................................................................16
1.9.2
Loading instructions ................................................................................................................16
1.9.3
Drying nitro-cellulose lacquers ................................................................................................16
1.9.4
Drying mould varnishes...........................................................................................................16
1.9.5
Drying impregnating resins......................................................................................................17
2.
UNIT DESCRIPTION............................................................................................ 17
2.1
Unit overview.....................................................................................................................................18
2.2
Control panel.....................................................................................................................................19
2.3
Solvent curve FDL 115 .....................................................................................................................19
3.
COMPLETENESS OF DELIVERY, TRANSPORTATION, STORAGE, AND
INSTALLATION....................................................................................................20
3.1
Unpacking, and checking equipment and completeness of delivery................................................20
3.2
Guidelines for safe lifting and transportation ....................................................................................21
3.3
Storage..............................................................................................................................................21
3.4
Location of installation and ambient conditions ................................................................................21
4.
INSTALLATION OF THE EQUIPMENT ............................................................... 22
4.1
Operating instructions .......................................................................................................................22
4.2
Electrical connection .........................................................................................................................23
4.3
Connection to a suction plant (optional) ...........................................................................................23
5.
START UP............................................................................................................ 24
5.1
Settings at the RD3 program controller ............................................................................................24
5.2
General indications ...........................................................................................................................25
6.
FIXED VALUE ENTRY MODE ............................................................................. 27
7.
WEEK PROGRAM EDITOR.................................................................................28
7.1
Program table template for Week program Editor ............................................................................30
8.
PROGRAM EDITOR.............................................................................................31
8.1
Selecting between set-point ramp and set-point step.......................................................................31
8.1.1
Programming with setting “Ramp” (default setting).................................................................31
8.1.2
Programming with setting “step” .............................................................................................33
8.1.3
General notes on programming temperature transitions ........................................................34
8.2
Set-point entry for program operation...............................................................................................34
8.3
Program table template.....................................................................................................................37
8.4
Deleting a program section...............................................................................................................38
FDL (E2) 03/2011
page 8/68
9.
PROGRAM START LEVEL.................................................................................. 39
10. USER LEVEL .......................................................................................................41
11. BEHAVIOR IN CASE OF FAILURES................................................................... 48
11.1 Behavior after power failure..............................................................................................................48
11.2 Alarm messages ...............................................................................................................................48
12. SAFETY DEVICES............................................................................................... 48
12.1 Temperature safety device class 2 (DIN 12880:2007) .....................................................................48
12.2 Exhaust air monitoring ......................................................................................................................50
13. OPTIONS..............................................................................................................50
13.1 Communication software APT-COM™ 3 DataControlSystem (option) ............................................50
13.1 Ethernet interface (option) ................................................................................................................50
13.2 Protocol printer (option) ....................................................................................................................51
13.3 Additional measuring channel for digital object temperature indicator with magnetic sensor or
pincer-type sensor (option) ...............................................................................................................52
14. MAINTENANCE, CLEANING, AND SERVICE .................................................... 53
14.1 Maintenance intervals, service..........................................................................................................53
14.2 Replacing the intake filter..................................................................................................................54
14.3 Cleaning and decontaminating the safety / paint drying oven..........................................................54
14.4 Sending the unit back to BINDER GmbH .........................................................................................55
15. DISPOSAL............................................................................................................56
15.1 Disposal of the transport packing .....................................................................................................56
15.2 Decommissioning..............................................................................................................................56
15.3 Disposal of the unit in the Federal Republic of Germany .................................................................56
15.4 Disposal of the unit in the member states of the EC except for the Federal Republic of Germany .57
15.5 Disposal of the unit in non-member states of the EC .......................................................................59
16. TROUBLESHOOTING ......................................................................................... 59
17. TECHNICAL DESCRIPTION................................................................................ 61
17.1 Factory calibration and adjustment...................................................................................................61
17.2 Definition of usable volume...............................................................................................................61
17.3 Technical data...................................................................................................................................61
17.4 Equipment and Options ....................................................................................................................63
17.5 Spare parts........................................................................................................................................64
17.6 Dimensions FDL 115 ........................................................................................................................65
18. CONTAMINATION CLEARANCE CERTIFICATE ............................................... 66
FDL (E2) 03/2011
page 9/68
Dear customer,
For the correct operation of the safety / paint drying ovens FDL, it is important that you read this operating
manual completely and carefully and observe all instructions as indicated. Failure to read, understand
and follow the instructions may result in personal injury. It can also lead to damage to the unit and/or poor
equipment performance.
1.
Safety
This operating manual is part of the components of delivery. Always keep it handy for reference. The
device should only be operated by laboratory personnel especially trained for this purpose and familiar
with all precautionary measures required for working in a laboratory. To avoid injuries and damage ob-
serve the safety instructions of the operating manual.
WARNING
Failure to observe the safety instructions.
Serious injuries and unit damage.
� Observe the safety instructions in this operating manual.
� Carefully read the complete operating instructions of the safety / paint drying oven FDL.
1.1
Legal considerations
This operating manual is for informational purposes only. It contains information for installing, start-up,
operation and maintenance of the product. Note: the contents and the product described are subject to
change without notice.
Understanding and observing the instructions in this operating manual are prerequisites for hazard-free
use and safety during operation and maintenance. In no event shall BINDER be held liable for any dam-
ages, direct or incidental arising out of or related to the use of this manual.
This operating manual cannot cover all conceivable applications. If you would like additional information,
or if special problems arise that are not sufficiently addressed in this manual, please ask your dealer or
contact us directly by phone at the number located on page one of this manual
Furthermore, we emphasize that the contents of this operating manual are not part of an earlier or exist-
ing agreement, description, or legal relationship, nor do they modify such a relationship. All obligations on
the part of BINDER derive from the respective purchase contract, which also contains the entire and ex-
clusively valid statement of warranty administration. The statements in this manual neither augment nor
restrict the contractual warranty provisions.
1.2
Structure of the safety instructions
In this operating manual, the following safety definitions and symbols indicate dangerous situations in
accordance with ISO 3864-2 and ANSI Z535.6.
1.2.1 Signal word panel
Depending on the probability of serious consequences, potential dangers are identified with a signal
word, the corresponding safety color, and if appropriate, the safety alert symbol.
DANGER
Indicates an imminently hazardous situation that, if not avoided, will result in death or serious (irreversi-
ble) injury.
WARNING
Indicates a potentially hazardous situation which, if not avoided, could result in death or serious (irre-
versible) injury.
FDL (E2) 03/2011
page 10/68
CAUTION
Indicates a potentially hazardous situation which, if not avoided, may result in moderate or minor (re-
versible) injury.
CAUTION
Indicates a potentially hazardous situation which, if not avoided, may result in damage to the product
and/or its functions or of a property in its proximity.
1.2.2 Safety alert symbol
Use of the safety alert symbol indicates a risk of injury.
Observe all measures that are marked with the safety alert symbol in order to avoid death or
injury.
1.2.3 Pictograms
Warning signs
Electrical hazard
Hot surface
Explosive atmosphere
Tipover hazard
Inhalation hazard
Pollution Hazard
Harmful substances
Biohazard
Mandatory action signs
Mandatory regulation
Read operating
instructions
Disconnect the power
plug
Lift with several persons
Environment protection
Prohibition signs
Do NOT touch
Do NOT spray with
water
Information to be observed in order to ensure optimum function of the product.
FDL (E2) 03/2011
page 11/68
1.2.4 Word message panel structure
Type / cause of hazard.
Possible consequences.
∅ Instruction how to avoid the hazard: prohibition
� Instruction how to avoid the hazard: mandatory action
Observe all other notes and information not necessarily emphasized in the same way, in order to avoid
disruptions that could result in direct or indirect injury or property damage.
1.3
Localization / position of safety labels on the unit
The following labels are located on the unit:
Pictograms (Warning signs)
Service label
Hot surface
Figure 1: Position of labels on the unit
Keep safety labels complete and legible.
Replace safety labels that are no longer legible. Contact BINDER Service for these replacements.
FDL (E2) 03/2011
page 12/68
1.4
Type plate
Figure 2: Position of type plate
Figure 3: Type plate (example of FDL 115 regular unit)
Indications of the type plate
Information
BINDER
Manufacturer: BINDER GmbH
FDL 115
Model FDL 115
Safety Drying Oven
Safety / paint drying oven
Serial No.
00-00000
Serial No. 00-00000
Nominal temperature
300 °C
572°F
Nominal temperature
Enclosure protection
IP 33
IP type of protection 33 acc. to EN 60529
Temp. safety device
DIN 12880
Temperature safety device acc. to standard DIN 12880
Class
2.0
Temperature safety device, class 2
Art. No.
9010-0269
Art. No. 9010-0269
Project No.
---
(Special application acc. to project no.)
Built
0000
Year of construction: 0000
2,90 kW
Nominal power 2.90 kW
230 V 1 N PE ~
Nominal voltage 230 V ± 10%, single-phase unit
13,0 A
Nominal current 13.0 Amp
50/60 Hz
Power frequency 50/60 Hz
Usable volume 0,115m³
Usable volume 0.115 m³
Steam space 0,156m³
Total steam space 0.156 m³
Max. solvent at nominal temp. 3,0g
Highest permissible solvent quantity at 300 °C / 572°F: 3.0g
Min. exhaust flow rate 24m³/h at +20 °C
Minimum exhaust flow rate at +20 °C: 24 m³/h
Max. temp. of heating surfaces +750 °C
Maximum temperature of heating surfaces +750 °C / 1382°F
Wiring diagram FDL 115 pages 1-3
Wiring diagram FDL 115 pages 1 to 3
Nominal temperature
300 °C
2,90 kW
Usable volume 0,115m³
572°F
230 V 1 N PE ~
Steam space 0,156m³
Enclosure protection
Temp. safety device
IP 33
DIN 12880
13,0 A
50/60 Hz
Max. solvent at nominal temp. 3,0g
Min. exhaust flow rate 24m³/h at +20 °C
Class
2.0
Max. temp. of heating surfaces +750 °C
Art. No.
9010-0269 US PATS 4585923 / 5222612 / 5309981
Wiring diagram FDL 115 pages 1-3
Project No.
Built
0000
5405194 / 5601143 / 5773287 / 6079403
Safety Drying Oven
D 78532 Tuttlingen / Germany
Tel. + 49 (0) 7462/ 2005-0
Internet: www.binder-world.com
FDL 115 Serial No. 00-00000
Made in Germany
FDL (E2) 03/2011
page 13/68
Symbol on the type plate
Information
CE conformity marking
Electrical and electronic equipment manufactured / placed on the
market in the EC after 13 August 2005 and to be disposed of in a
separate collection according to directive 2002/96/EC on waste elec-
trical and electronic equipment (WEEE).
The equipment is certified in the GOST R certification system of
GOSTSTANDARD Russia.
GS mark of conformity of the “Deutsche Gesetzliche Unfallversi-
cherung e.V. (DGUV), Fachausschuss Metall- und Oberflächenbe-
handlung, Prüf- und Zertifizierungsstelle im DGUV Test“ (German
Social Accident Insurance (DGUV), Expert Committee: Metal and
Surface Treatment, Testing and Certification Body in DGUV)
1.5
General safety instructions on installing and operating the safety / paint
drying oven FDL
With regard to operating the safety / paint drying oven FDL and to the installation location, please ob-
serve the guideline BGI/GUV-I 850-0 on safe working in laboratories (formerly BGR/GUV-R 120 or ZH
1/119 laboratory guidelines issued by the employers’ liability insurance association) (for Germany).
BINDER GmbH is only responsible for the safety features of the unit provided skilled electricians or quali-
fied personnel authorized by BINDER perform all maintenance and repair, and if components relating to
chamber safety are replaced in the event of failure with original spare parts.
To operate the unit, use only original BINDER accessories or accessories from third-party suppliers au-
thorized by BINDER. The user is responsible for any risk caused by using unauthorized accessories.
CAUTION
Danger of overheating.
Damage to the unit.
∅ Do NOT install the unit in unventilated recesses.
� Ensure sufficient ventilation for dispersal of the heat.
Do not operate the safety / paint drying oven FDL in hazardous locations.
DANGER
Explosion hazard.
Danger of death.
∅ Do NOT operate the unit in potentially explosive areas.
∅ KEEP explosive dust or air-solvent mixtures AWAY from the unit.
Familiarize yourself with any potential health risks caused by the charging material, the contained mois-
ture constituent or by reaction products which may arise during the temperature process. Take adequate
measures to exclude such risks prior to putting the safety / paint drying oven into operation.
FDL (E2) 03/2011
page 14/68
DANGER
Electrical hazard.
Danger of death.
∅ The unit must NOT become wet during operation or maintenance.
The safety / paint drying ovens were produced in accordance with VDE regulations and were routinely
tested in accordance to VDE 0411-1 (IEC 61010-1).
If a warning signal which indicates an of error condition, no further charging material must be introduced
into the unit.
DANGER
Explosion and ignition hazard.
Danger of death.
∅ Do NOT introduce any further charging material as long as there is a warning signal.
During and after the drying process, the inner surfaces have got a temperature close to the set-point.
CAUTION
The inner chamber, the outgoing air pipe, the door window (option), the door gasket,
and the access ports will become hot during operation.
Danger of burning.
∅ Do NOT touch the inner surfaces, the outgoing air pipe, the door window, the door gas-
ket, the access ports, or the charging material during operation.
1.6
Intended use
The BINDER safety / paint drying oven FDL 115 is suitable for drying and burn in of lacquers and similar
liquid coating materials which are containing solvents that can form explosive mixtures with air. The
maximally permitted drying temperature and the maximally permitted quantity of solvent are limited, see
chap. 1.9.
Other applications are not approved.
Do NOT use the unit to warm up coating materials in containers, vessels, etc. or for drying textiles
soaked in solvent.
Do NOT use the unit for drying purpose if greater quantities of steam or solvent gas leading to condensa-
tion will be set free.
Following the instructions in this operating manual and conducting regular maintenance work
(chap. 14.1) are part of the intended use.
1.7
Operating instructions
Depending on the application and location of the unit, the operator of the safety / paint drying oven must
provide the relevant information for safe operation of the unit in a set of operating instructions.
These operating instructions must be kept with the unit at all times in a place where they are
clearly visible. They must be comprehensible and written in the language of the employees.
FDL (E2) 03/2011
page 15/68
1.8
Measures to prevent accidents
During drying of fluid paints, flammable solvent/air mixtures can form and ignite.
The manufacturer took the following measures to prevent ignition and explosions:
• Indications on the type plate
See operating manual chap. 1.4
• Operating manual
An operating manual is available for each safety / paint drying oven.
A diagram in the operating manual (chap. 2.3) indicates the maximum permissible solvent quantities
for various operating states.
The operating manual asks the operator of the safety / paint drying oven to set up an instruction on
the permissible loading quantity.
• Maximum temperatures and maximum permissible steam concentration
By means of the diagram “Highest permissible solvent quantity”, which is included in the operating
manual chap. 2.3 and applied on the front of the safety / paint drying oven, the operator must adapt
the drying temperature to the maximum solvent quantity arising.
When using nitro-cellulose lacquers or nitro combination lacquers, the diagram ”Highest permissible
solvent quantity” specifies a limit temperature of max. 130 °C/ 266 °F, which must not be exceeded.
(In this context, all paints containing more than 5 % nitro-cellulose relative to the non-volatile contents
are regarded as nitro-cellulose lacquers and nitro-combination lacquers).
A gas-tight separation between the drying chamber and the heating chamber is not necessary since
there is an effective forced air motion in the entire steam room.
• Throttle valves
No throttle valves are used, i.e., the full air change occurs permanently.
• Protecting the heating surfaces against dripping
The BINDER APT.line™ preheating chamber protects all heating elements against lacquer dripping
and direct contact with lacquer coatings.
• Heat insulation
All heat insulation is sealed against penetration by lacquer vapors from the outside with high tempera-
ture-resistant and ageing-resistant sealant.
The insulation material consists of non-combustible mineral wool (class A1 according to DIN 4102-
1:1998).
• Temperature monitoring
The safety / paint drying oven is equipped with a temperature display, which can be read from outside.
A built-in additional temperature safety device can turn off the heating and is functionally independent
of the main controller. When turning off the unit, the forced convection and the control equipment’s
function are maintained. Visual (red indicator light) and audible (buzzer) signals indicate temperature
exceeding.
• Safety in case the technical ventilation fails
The heating only turns on if the air circulation is already in operation.
If the air circulation fails, the heating immediately turns off. In addition, there are audible (buzzer) and
visual (red indicator light) signals.
• Moving parts participating in the work process
It is impossible to touch the fan from outside the device or from the interior.
• Safety, measurement and control devices
The safety, measuring, and control equipment is easily accessible via the housing top cover.
FDL (E2) 03/2011
page 16/68
• Electrostatic charge
The interior parts are grounded.
• Protection against touchable surfaces
Tested according to EN ISO 13732-1:2006.
• Floors
See operating manual chap. 3.4 for the oven’s installation.
• Ventilation
Ventilation shall be realized by the operator according to GUV-R 500 chap. 2.29 “Verarbeiten von
Beschichtungsstoffen” (Processing of paints) (for Germany).
• Cleaning
See operating manual chap. 14.
• Examinations
The safety / paint drying oven has been inspected by the “Deutsche Gesetzliche Unfallversicherung
e.V. (DGUV), Fachausschuss Nahrungs- und Genussmittel, Prüf- und Zertifizierungsstelle im DGUV
Test“ (German Social Accident Insurance (DGUV), Expert Committee: “Metal and Surface Treatment”,
Testing and Certification Body in DGUV) and bears the GS mark.
1.9
Important points to consider before commissioning
1.9.1 Technical ventilation / permissible load
For safety reasons, the formation of a dangerous, explosive atmosphere must be avoided in all operating
modes (see GUV-R 500 chap. 2.28 “Betreiben von Trocknern für Beschichtungsstoffe” ("Dryers for coat-
ing materials")). Keeping the maximum permissible amount of solvent when loading the unit will meet this
requirement. The permissible quantity can be calculated according to the "principles for the calculation of
ventilation of chamber dryers and continuous dryers" (EN 1539:2009, Appendix B). According to this
standard, the safety / paint drying oven’s technical data (chap. 17.3) must be included in the calculation,
and it is required to set up a set of loading instructions (for Germany).
1.9.2 Loading instructions
The loading instructions should specify the amount of material, which can be loaded into the drying oven
for treatment without any risk of creating a dangerous, explosive atmosphere. GUV-R 500 chap. 2.28
“Betreiben von Trocknern für Beschichtungsstoffe” ("Dryers for coating materials") explicitly states that
the operator must set up a set of loading instructions (for Germany).
1.9.3 Drying nitro-cellulose lacquers
When using the safety / paint drying oven FDL to dry material, which is coated with nitro-cellulose lac-
quers, the temperature safety device must be set to max. 130 °C / 266 °F, so that the surface tempera-
ture of the goods to be dried is ensured not to exceed 130 °C / 266 °F. Deviations are only permissible if
a report from a testing agency, which is recognized by the employer's liability insurance association, has
declared safe a higher surface temperature.
1.9.4 Drying mould varnishes
When using the safety / paint drying oven FDL to dry mould varnishes, the operator is allowed to increase
the highest permissible amount of solvent specified for surface drying (chap. 2.3) by up to 10 times (see
GUV-R 500 chap. 2.28 “Betreiben von Trocknern für Beschichtungsstoffe” ("Dryers for coating materials")
or EN 1539:2009, Appendix A.1.1.2).
FDL (E2) 03/2011
page 17/68
1.9.5 Drying impregnating resins
When using the safety / paint drying oven FDL to dry impregnating resins, the operator is allowed to in-
crease the highest permissible amount of solvent specified for surface drying (chap. 2.3) by up to 20
times (see GUV-R 500 chap. 2.28 „Betreiben von Trocknern für Beschichtungsstoffe“ ("Dryers for coating
materials") no. 3.7.4 or EN 1539:2009, Appendix A.1.1.2).
2.
Unit description
The safety / paint drying oven FDL 115 was built according to EN 1539:2009 (“Dryers and ovens, in
which flammable substances are released. Safety requirements”).
The fan in the rear of the safety / paint drying oven delivers a constant amount of fresh air through the
working area irrespective of the drying temperature. A large-area filter removes dust from the incoming
air (permeability up to approx. 1 micrometer).
A flow monitor in the upper part of the device (pressure differential switch) monitors the stream of exhaust
air. In the event of failure, the monitoring system turns the heater off immediately and registers this status
with a visual and an audible signal. The heating of the device is released with a time delay of 4 minutes
after you turn on the device by its main power switch.
The indicator light (3) in the operating panel only lights up as long as the air-flow monitor has not yet en-
abled the heating process. The drying temperature is also monitored constantly by the temperature limiter
(2). If the temperature would exceed the maximum permissible temperature, the heating is turned off
immediately and this state is indicated visually by indicator light (2a) and acoustically. In the event of fail-
ure, the safety / paint drying oven cannot be restarted until the RESET key (2b) or (4) has been reset
BINDER safety / paint drying ovens with forced convection FDL are equipped with the electronic program
controller RD3 with digital display. This permits programming of temperature cycles.
The APT.line™ preheating chamber system ensures high level of spatial and time-based temperature
precision, thanks to the direct and distributed air circulation into the interior. The air turbine supports ex-
act attainment and maintenance of the required temperature accuracy.
All functions of the multifunctional program control can be set simply and conveniently via the easy to
understand function keypad of the RD3 temperature program controller. This controller is equipped with
touch function keys and a digital display and permits exact temperature setting and programming tem-
perature cycles. The FDL provides almost unlimited possibilities of adapting to individual customer re-
quirements based upon extensive programming options and on the week program timer and real time
clock of the controller.
All unit functions are easy and comfortable to use thanks to their clear arrangement. Major features are
easy cleaning of all unit parts and avoidance of undesired contamination.
The inner chamber, the pre-heating chamber and the interior side of the doors are all made of stainless
steel (material no. 1.4301 (V2A) in Germany, US equivalent 304). When operating the chamber at tem-
peratures above 150 °C, the impact of the oxygen in the air may cause discoloration of the metallic sur-
faces (yellowish-brown or blue) by natural oxidation processes. These colorations are harmless and will
in no way impair the function or quality of the unit. The housing is RAL 7035 powder-coated. All corners
and edges are also completely coated.
The safety / paint drying oven FDL comes equipped with a serial interface RS 422 for computer commu-
nication, e.g. via the communication software APT-COM™ 3 DataControlSystem (option, chap. 13.1). For
further options, see chap. 17.4.
The units can be operated at an ambient temperature of 18 °C / 64.4 °F up to 40 °C / 104 °F in a tem-
perature control range by 5 °C above ambient temperature up to +300 °C / 572 °F.
FDL (E2) 03/2011
page 18/68
2.1
Unit overview
Figure 4: Front view FDL 115 with optional window
(A)
Control panel
(B)
Microprocessor program controller RD3
(C)
Temperature safety device class 2 according to DIN 12880:2007
(D)
Main power switch ON/OFF
(E)
Solvent curve
(F)
Door handle
(G)
Unit door
(H)
Window (optional)
RES
AI
(E)
(B)
(C)
(D)
(F)
(A)
(G)
(H)
FDL (E2) 03/2011
page 19/68
2.2
Control panel
(6) (5) (4) (3) (2) (2b) (2a) (1)
Figure 5: Control panel of standard unit
(1)
Main power switch ON/OFF
(2)
Temperature safety device class 2
(2a) Red pilot lamp for temperature safety device class 2
(2b) RESET key for temperature safety device
(3)
Red indicator light, loss of technical ventilation
(4)
RESET key for the heating cut-out due to failure of technical ventilation
(5)
Temperature program controller RD3
(6)
Solvent curve: Highest permissible amount of solvent Gtotal [g] as a function of the drying tempera-
ture
2.3
Solvent curve FDL 115
0,00
5,00
10,00
15,00
20,00
25,00
30,00
35,00
50
60
70
80
90
100
110
120
130
140
150
160
170
180
190
200
210
220
230
240
250
260
270
280
290
300
Temperature [°C]
Gtotal [g]
Figure 6: Solvent curve FDL 115
The diagram shows the highest permissible solvent quantity Gtotal [g] in the steam room in correspon-
dence to the drying temperature. This is based on the calculation acc. to EN 1539:2009 considering the
unit specific data, an assumed molecular weight of the solvent of 100g/Mol, and a lower explosion limit of
40g/m3 at 20 °C / 68 °F and at 760 Torr (1013 hPa) (assumptions for unknown solvents acc. to EN
1539:2009).
RESET
AIR
FDL (E2) 03/2011
page 20/68
In case of too high drying temperature and /or too high quantity of solvent in the steam room it could be
possible that the concentration of the solvent steam will lead to an explosion. The permitted maximum of
solvent brought into the drying oven and the maximal temperature must not be exceeded.
DANGER
Excessive drying temperature and/or excessive solvent quantity.
Danger of explosion.
Danger of death.
∅ Do NOT exceed the maximum solvent quantity.
∅ Do NOT exceed the maximum drying temperature for the solvent quantity.
Adjust the temperature safety device according to the selected set-point (chap. 12.1).
3.
Completeness of delivery, transportation, storage, and installa-
tion
3.1
Unpacking, and checking equipment and completeness of delivery
After unpacking, please check the unit and its optional accessories, if any, based on the delivery receipt
for completeness and for transportation damage. Inform the carrier immediately if transportation damage
has occurred.
The final tests of the manufacturer may have caused traces of the shelves on the inner surfaces. This
has no impact on the function and performance of the unit.
Please remove any transportation protection devices and adhesives in/on the unit and on the doors and
remove the operating manuals and accessory equipment.
CAUTION
Sliding or tilting of the unit.
Damage to the unit.
∅ Do NOT lift or transport the unit using the door or the handle.
� Lift the unit from the pallet at the four lower corners with the aid of four people.
If you need to return the unit, please use the original packing and observe the guidelines for safe lifting
and transportation (chap. 3.2).
For disposal of the transport packing, see chap. 15.1.
Note on second-hand units (Ex-Demo-Units):
Second-hand units are units that were used for a short time for tests or exhibitions. They are thoroughly
tested before resale. BINDER ensures that the chamber is technically sound and will work flawlessly.
Second-hand units are marked with a sticker on the unit door. Please remove the sticker before commis-
sioning the unit.
FDL (E2) 03/2011
page 21/68
3.2
Guidelines for safe lifting and transportation
After operation, please observe the guidelines for temporarily decommissioning the unit (chap. 15.2).
CAUTION
Sliding or tilting of the unit.
Damage to the unit.
∅ Do NOT lift or transport the unit using the door or the handle.
� Transport the unit in its original packaging only.
� For moving or shipping, secure the oven with transport straps.
� Lift the unit at the four lower corners with the aid of 4 people and place it on a transport
pallet with wheels. Push the pallet to the desired site and then lift the unit at its four lo-
wer corners from the pallet with the aid of four people.
• Permissible ambient temperature range during transport: -10 °C / 14 °F to +60 °C / 140 °F.
You can order transport packing and pallets for moving or shipping purposes from BINDER service.
3.3
Storage
Intermediate storage of the unit is possible in a closed and dry room. Observe the guidelines for tempo-
rary decommissioning (chap. 15.2).
• Permissible ambient temperature range during storage: -10 °C / 14 °F to +60 °C / 140 °F.
• Permissible ambient humidity: max. 70 % r.H., non-condensing
When after storage in a cold location you transfer the unit to its warmer installation site, condensation
may form. Before start-up, wait at least one hour until the chamber has attained ambient temperature and
is completely dry.
3.4
Location of installation and ambient conditions
Set up the safety / paint drying oven FDL on a flat, even and non-flammable surface, free from vibration
and in a well-ventilated, dry location and align it using a spirit level. The site of installation must be capa-
ble of supporting the unit’s weight (see technical data, chap. 17.3). The chambers are designed for set-
ting up inside a building (indoor use).
CAUTION
Danger of overheating.
Damage to the unit.
∅ Do NOT set up units in non-ventilated recesses.
� Ensure sufficient ventilation for dispersal of the heat.
• Permissible ambient temperature range during operation: +18 °C / 64.4 °F up to +40 °C / 104 °F. At
elevated ambient temperature values, fluctuations in temperature can occur.
The ambient temperature should not be substantially higher than the indicated ambient tem-
perature of +25 °C / 77°F to which the specified technical data relate. For other ambient condi-
tions, deviations from the indicated data are possible.
• Permissible ambient humidity: 70 % r.H. max., non-condensing.
• Installation height: max. 2000 m / 6.6 ft. above sea level.
When placing several units of the same size side by side, maintain a minimum distance of 250 mm / 9.84
in between each unit. Wall distances: rear 100 mm / 3.94 in, sides 160 mm / 6.29 in.
FDL (E2) 03/2011
page 22/68
CAUTION
Danger by stacking.
Damage to the units.
∅ Do NOT place safety / paint drying ovens on top of each other.
To completely separate the unit from the power supply, you must disconnect the power plug. Install the
unit in a way that the power plug is easily accessible and can be easily pulled in case of danger.
Do not install or operate the oven in potentially explosive areas.
DANGER
Explosion hazard.
Danger of death.
∅ Do NOT operate the unit in potentially explosive areas.
∅ KEEP explosive dust or air-solvent mixtures AWAY from the vicinity of the unit.
The standard equipped safety / paint drying oven FDL with housing protection IP 33 acc. to DIN 40050
(see type plate) must NOT be installed and operated in explosive and fire-endangered areas.
If the safety / paint drying oven is optionally built up with housing protection IP 54, it can be operated in
fire-endangered areas. Decisive is always the indication on the type plate.
When operating the safety / paint drying oven FDL with housing protection IP 54 in fire-
endangered areas, observe DIN VDE 0100-482:2003 (for Germany).
Operation of the cooling slots must on no account be impaired. The vapors produced when the heating
the content must be extracted from the safety / paint drying oven through non-combustible exhaust gas or
exhaust air ducts. The exhaust duct (nominal diameter 100 mm / 3.9 in) on the rear of the chamber
serves for this purpose, to which a suitable exhaust air duct can be connected, e.g., a corrugated alumi-
num hose. The exhaust connection must be made via a draught limiter; the exhaust air must not be
guided into ducts for combustible gas.
CAUTION
The rear exhaust duct will become hot during operation.
Danger of burning.
∅ Do NOT touch the exhaust duct during operation.
4.
Installation of the equipment
4.1
Operating instructions
Depending on the application and location of the unit, the operator of the safety / paint drying oven must
provide the relevant information for safe operation of the unit in a set of operating instructions.
Keep these operating instructions with the unit at all times in a place where they are clearly
visible. They must be comprehensible and written in the language of the employees.
FDL (E2) 03/2011
page 23/68
4.2
Electrical connection
The safety / paint drying oven FDL 115 is supplied ready for connection.
• Power connection: fixed power connection cable 1800 mm / 5.9 ft in length with a shockproof plug
• Power supply voltage 230 V (1N~) +/- 10 %, 50/60 Hz
• Housing protection type according to EN 60529:2000 : IP 33 (optional IP 54)
• Electrical protection: protection class I (with grounding conductor connection)
• Prior to connection and start-up, check the power supply voltage. Compare the values to the specified
data located on the unit’s type plate (unit front behind the door, bottom left-hand, see chap. 1.4).
• When connecting, please observe the regulations specified by the local electricity supply company as
well as the VDE directives (for Germany)
• Pollution degree (acc. to IEC 61010-1): 2
• Over-voltage category (acc. to IEC 61010-1): II
CAUTION
Danger of incorrect power supply voltage.
Damage to the equipment.
� Check the power supply voltage before connection and start-up.
� Compare the power supply voltage with the data indicated on the type plate.
See also electrical data (chap. 17.3).
To completely separate the unit from the power supply, you must disconnect the power plug.
Install the unit in a way that the power plug is easily accessible and can be easily pulled in
case of danger.
4.3
Connection to a suction plant (optional)
When directly connecting a suction plant the spatial temperature exactitude, the heating-up and the re-
covering times and the maximum temperature will be negatively influenced. So no suction plant should be
directly connected to the exhaust duct (nominal diameter 100 mm / 3.9 in) on the rear of the chamber.
Connect a suitable exhaust air duct, e.g., a corrugated aluminum hose, to the chamber exhaust duct. The
exhaust connection must be made via a draught limiter; the exhaust air must not be guided into ducts for
combustible gas.
Active suction from the oven must only be performed together with extraneous air. Perforate
the connecting piece to the suction device or place an exhaust funnel at some distance to the
outgoing air pipe.
CAUTION
The exhaust duct will become hot during operation.
Danger of burning.
∅ Do NOT touch the exhaust duct during operation.
FDL (E2) 03/2011
page 24/68
5.
Start up
After connecting the electrical supply (chap. 4.2), turn on the unit via the main power switch (1).
Set the main power switch (1) to position “I”. Indicator light (3) and an audible signal indicate that the
minimal volumetric flow rate of the technical ventilation has not yet been reached. Once the fan has
reached its operating speed, indicator light (3) goes off. Now press RESET key (4). The audible alarm
stops.
After turning on the main power switch (1), the heating is automatically enabled with a delay time of 4
minutes to ensure the required pre-ventilation acc. to EN1539:2009. Set the temperature set point (chap.
6). Detach the plastic cover over the temperature safety device class 2 (2) with a suitable Phillips screw-
driver, and then set the temperature safety device (2) to the maximum permissible drying temperature
(chap. 12.1) and restore the plastic cover to prevent misadjusting.
After reaching the set drying temperature, the heating keeps it constant by turning on and off regularly.
You can verify this on the controller display.
When loading the oven with solvent-containing material, do not exceed the maximally permitted solvent
quantity for the selected drying temperature. Refer to the solvent curve at the oven front, chap. 2.3.
Adjust the temperature safety device according to the selected set-point (chap. 12.1).
Warming chambers may release odors in the first few days after commissioning. This is not a quality
defect. To reduce odors quickly we recommend heating up the chamber to its nominal temperature for
one day and in a well ventilated location.
5.1
Settings at the RD3 program controller
After turning the unit on with the main power switch (1) the controller is in Normal Display / fixed value
operation mode.
Depending on the temperature value entered before LED (3a) is lit if the heating is active, or no LED if the
actual temperature is equal to or above the set-point.
In Display 1 of the controller the actual temperature Value is shown.
• With inactive week program timer:
In Display 2 of the controller the actual date and time are displayed. Example:
15.01.07 13:52
• With active week program timer:
In Display 2 of the controller the actual date and time and the states of the week program timer chan-
nels are displayed. Examples:
15.01.07 13:52 - -
15.01.07 13:52 - �
15.01.07 13:52 � -
15.01.07 13:52 � �
Channel 1 Off,
Channel 2: Off
Channel 1: Off,
Channel 2: On
Channel 1: On,
Channel 2: Off
Channel 1 On,
Channel 2: On
FDL (E2) 03/2011
page 25/68
Display 1
Display 2
Figure 7: RD3 program controller
LED function indications and their signification:
(3a)
(yellow)
Heating active
(3b)
(yellow)
No function
(3c)
(yellow)
No function
(3d)
(green)
Illuminates: program operation
Flashes: exceeding of the tolerance limits in “Fixed value entry mode” or in
“Program operation”. In “Program operation”: program interruption.
The program controller RD3 permits programming of temperature cycles.
Two programs with up to 10 sections each or one program with up to 20 sections can be entered (setting
in the user level, chap. 10).
When changing from 2 programs to 1 program or vice-versa, existing programs are deleted
The maximum length of an individual program section can be set to either 99 hs 59 min or to 999 hs 59
min (setting in the user level, chap. 10). This setting is then valid for all program sections.
Programming can be done directly via the controller keyboard or graphically at the computer using the
software APT-COM™ 3 DataControlSystem (option, chap. 13.1) specially developed by BINDER.
5.2
General indications
The program controller RD3 offers several functional levels:
Normal Display / fixed value operation:
• Display of the actual value of temperature (display 1) and of the actual date and time (display 2).
• The chamber is in fixed value operating mode, adjusting to the entered set-points.
Fixed value entry mode (chap. 6)
• Entry of the temperature set-point for fixed value operating mode
• Entry of temperature set-points 1 and 2 for week program operation
FDL (E2) 03/2011
page 26/68
Program editor (chap. 8)
• Two programs with up to 10 sections each or one program with up to 20 sections can be entered (se-
lection in the user level, chap. 10). Entry of temperature set-points in all program sections (chap. 8.2).
• Deleting a program section (chap. 8.4)
Program start level (chap. 9)
• Selection of an entered program
• Entry of settings affecting the program course, as “start delay time” or “number of program cycles”
• Program start
Week program editor (chap.7)
• Setting the shift points
User level (chap. 10)
• User specific controller settings
• Setting the real time clock
Normal Display / Fixed value operation
5 seconds
Fixed value entry mode
Program editor
Program start level
5 seconds
Week program editor
Menu visible only if week program
timer is activated in the user level
5 seconds
User level
If no button is touched within more than 120 sec. the controller returns from the current level to Normal
Display.
FDL (E2) 03/2011
page 27/68
6.
Fixed value entry mode
If you do not want to use the week program timer, deactivate it (factory setting, setting in the
user level, chap. 10) before entering any set-points.
Basic entry principle: Access the individual parameters with button X/W one after the other. Enter the
values with the arrow keys. A value flashing once after 2 seconds indicates that it has been applied by
the controller.
Normal Display
Display 1 shows
e.g. 39.8
(actual temperature value)
Display 2 shows e.g. 15.01.07 13:52 - -
(actual date and time)
(actual switching state of week program timer channel 1: Off,
channel 2: Off, visible only if week program timer is activated
in the user level, chap. 10)
Press key
Display 1 shows
e.g. 40.0
(actual temperature set-point 1)
Display 2 shows SP1 TEMPERATURE
(variable: temperature in °C)
Enter the temperature set-point in °C using
arrow keys
Value is shown in display 1.
Press key
Display 1 shows
e.g. 90.0
(actual temperature set-point 2)
(visible only if week program timer is activated
in the user level, chap. 10)
Display 2 shows SP2 TEMPERATURE
(variable: temperature in °C)
Enter temperature set-point in °C using the
arrow keys
Value is shown in display 1.
Press key
If no button is pressed within more than 120 sec, or if the EXIT button is pressed, the controller changes
to Normal Display.
When changing the set-point, check the setting of the temperature safety device (chap. 12.1).
The values entered in fixed-value entry mode remain valid after program run-off and are then
equilibrated.
If the week program timer is active, depending on the running week program another set-point (SP2) may
be targeted. Temperatures too high for the introduced solvent quantity can occur. Deactivate the week
program timer if you do not use it (default setting, setting in the User level, chap. 10).
DANGER
Too high temperature.
Danger of explosion.
Danger of death.
� Deactivate the week program timer if you do not use it.
FDL (E2) 03/2011
page 28/68
7.
Week program editor
The Week program editor permits defining up to 4 shift point for each week day. A shift point defines a
moment and the switching state ON or OFF of the channels that become active in this instance.
Channel function:
• Channel 1 On = Set-point 2 is equilibrated.
• Channel 1 Off = Set-point 1 is equilibrated
• Channel 2 = reserve
The week program timer is initially set to inactive (factory setting). Therefore, you need to
activate the week program timer in the user level (chap. 10).
Normal Display
Display 1 shows
e.g. 39.8
(actual temperature value)
Display 2 shows e.g. 15.01.07 13:52 - -
(actual date and time, actual state of week program timer
channel 1: Off, channel 2: Off)
Press down key
for 5 sec
Display 1 shows
e.g. 0000
Display 2 shows
PROGRAM EDITOR
(you are in the program editor)
Press down key
for 5 sec
Display 1 shows
0000
Menu visible only if week program timer is activated
in the user level (chap. 10)
Display 2 shows WEEK PROG. EDITOR
(you are in the week program editor)
Press program key
Display 1 shows
0000
Display 2 shows UserCod?
0000
(enter user code, display flashes)
Enter the user code using arrow keys
e.g. 0001 (basic setting, adjustable in the
user level, chap. 10).
Value is shown in both displays.
Automatically forward after 2 sec
Display 1 shows
0000
Display 2 shows
Monday
(selection of day of the week)
(actual selection: Monday)
Select the day of the week (Monday up to
Sunday) with key
Day of the week is shown in display 2.
Press program key
FDL (E2) 03/2011
page 29/68
Display 1 shows
0000
Display 2 shows
Shiftpt.
(no function)
Press program key
Display 1 shows
0000
Display 2 shows
Shiftpt.
1
(selection of the shift point)
(actual shift point: 1)
Select the shift point (1 up to 4) with key
Value is shown in display 2.
Press program key
Display 1 shows
e.g. --.--
(time of the selected shift point)
Display 2 shows
S1:
--:--
(actual selection of the shift point: S1)
(actual setting: shift point not programmed)
Press program key
Display 1 shows
--.--
(time of the selected shift point)
Display 2 shows Time
--:--
(entry of the time of the selected shift point)
(actual setting: shift point not programmed)
Enter the time (hh:mm) using arrow keys
Value is shown in display 2.
Press key
Display 1 shows
0000
Display 2 shows Ch1 = SP2:
Off
(entry of the state of channel 1)
(actual setting: Off)
Enter the state of channel 1
(On or Off) using arrow keys
Setting is shown in display 2.
Press key
Display 1 shows
0000
Display 2 shows Channel 2:
Off
(entry of the state of channel 2)
(actual setting: Off) (no function)
Enter the state of channel 2
(On or Off) using arrow keys
Setting is shown in display 2.
Press key
Press key EXIT
Display 1 shows
e.g., 08.30
(time of the selected shift point)
Display 2 shows S1:
08:30 - -
(actual selection of the shift point: S1)
(actual setting: time 08.30, channels Off)
Press program key
Press key EXIT twice
Select the next shift point
Select the next day of the week
To exit the menu, press several times key EXIT or wait for 120 seconds. Controller returns to normal
display.
FDL (E2) 03/2011
page 30/68
7.1
Program table template for Week program Editor
Program editor
Program title
Project
Date:
Time
Channel 1 (tem-
perature)
Channel 2*
Day of the week
hh:mm
AM
PM
ON = SP2
OFF = SP1
ON
OFF
S1
S2
S3
Monday
S4
S1
S2
S3
Tuesday
S4
S1
S2
S3
Wednesday
S4
S1
S2
S3
Thursday
S4
S1
S2
S3
Friday
S4
S1
S2
S3
Saturday
S4
S1
S2
S3
Sunday
S4
* Channel 2 is non-functional in the standard unit
FDL (E2) 03/2011
page 31/68
8.
Program editor
8.1
Selecting between set-point ramp and set-point step
You can program various kinds of temperature transitions. In the user level (chap. 10) you can select
between the settings “Ramp” (default setting) and “Step”.
Setting “Ramp” permits programming all kinds of temperature transitions.
With setting “Step” the controller will equilibrate only to constant temperatures; programming
ramps is no longer possible.
Switching between settings “Ramp” and “Step” will influence all programs. Please note that
this can cause the time courses of existing programs to change significantly.
8.1.1 Programming with setting “Ramp” (default setting)
Set-points always refer to the start of a program section, i.e., at the beginning of each program section,
the entered set-point will be reached. During program section operation, the temperature gradually
passes to the set-point entered for the subsequent program section.
You can program all kinds of temperature transitions by the appropriate design of the program section
timing:
• Gradual temperature changes “set-point ramp“
The set-point gradually moves from one set-point to the one of the following program section during a
given interval. The actual temperature value (X) follows the continually moving set-point (W) at any
time.
• Program sections with constant temperature
The initial values of two subsequent program sections are identical; therefore the temperature is kept
constant during the whole time of the first program section.
• Sudden temperature changes “set-point step”
Steps are temperature changes (ramps) that occur during a very short interval. Two program sections
with an identical set-point are followed by a section with a different set-point. If the duration of this
transitional program section is very short (minimum entry 1 min), the temperature change will proceed
rapidly in the minimum amount of time.
01
02
03
04
05
06
07
08
09
W
t
Figure 8: Possible temperature transitions
(with default setting “Ramp” in the user level (chap. 10)
FDL (E2) 03/2011
page 32/68
Program entry as set-point ramp (example):
W/°C
t/min.
S01
S02
S03
S04
0
20
40
60
80
100
30
120
180
380
S05
Program table corresponding to the diagram (with default setting “Ramp”):
Section
SEC
Temperature
set-point
[ °C]
TEMP
Section length
[hh.mm]
TIME
S01
40
00:30
S02
60
01:30
S03
90
01:00
S04
90
03:20
S05
110
00:01
The values of such a program table can now be entered to the RD3 program controller (chap. 8.2).
Program entry as set-point step (example):
W/°C
t/min.
S01 S02 S03 S04 S05 S06 S07 S08
0
20
40
60
80
100
30
120
180
380
FDL (E2) 03/2011
page 33/68
Program table corresponding to the diagram (with default setting “Ramp”):
Section
SEC
Temperature
set-point
[ °C]
TEMP
Section length
[hh.mm]
TIME
S01
40
00:30
S02
40
00:01
S03
60
01:30
S04
60
00:01
S05
80
01:00
S06
80
00:01
S07
110
03:20
S08
110
00:01
The values of such a program table can now be entered to the RD3 program controller (chap. 8.2).
The end point of the desired cycle must be programmed with an additional section (in our examples S05
for set-point ramp and S08 for set-point step) with a section time of at least one minute. Otherwise, the
program will stop one section too early because the program line is incomplete.
8.1.2 Programming with setting “step”
With setting “Step” selected, you don’t need to program the transition section in the Program Editor.
With setting “step” the controller will equilibrate only to constant temperatures; programming
ramps is no longer possible.
The set-points are maintained constant for the duration of a program section. At the start of each program
section, the unit heats up with the maximum speed in order to attain the entered set-point.
Program entry as set-point step (example):
W/°C
t/min.
S01 S02 S03 S04 S05 S06 S07 S08
0
20
40
60
80
100
30
120
180
380
FDL (E2) 03/2011
page 34/68
Program table corresponding to the diagram (with setting “Step”):
Section
SEC
Temperature
set-point
[ °C]
TEMP
Section length
[hh.mm]
TIME
S01
40
00:30
S02
60
01:30
S03
80
01:00
S04
110
03:20
The values of such a program table can now be entered to the RD3 program controller (chap. 8.2).
8.1.3 General notes on programming temperature transitions
If the tolerance limits set in the user level (chap. 10) are exceeded, the program is halted until the actual
temperature value returns to within the tolerance range. During this program interruption, the LED (3d)
flashes. Therefore, the duration of the program might be extended due to the programming of tolerances
The programming is saved even in case of a power failure or after turning off the unit.
After program rundown the controller returns to fixed value operation showing Normal Display and equili-
brates to the temperature value previously entered in fixed value entry mode.
Before starting the program, check the set-point value entered in fixed value entry mode.
After program rundown temperature will equilibrate to this value.
Deactivate the week program timer (factory setting, setting in the user level, chap. 10) before
starting a program.
8.2
Set-point entry for program operation
From Normal Display, press down button X/W for 5 sec to access the program editor. Then enter the set-
points one after the other in all program sections of a selected program.
You can enter two programs with up to 10 sections each or one program with up to 20 sections (setting in
the user level, chap. 10).
In order to avoid incorrect programming, we recommend entering the values of the program course into a
table (template in chap. 8.3).
Example of program table (with default setting “Ramp”):
Section
SEC
Temperature set-point
[ °C]
TEMP
Section length
[hh.mm]
TIME
S01
40
00:30
S02
60
01:30
S03
90
01:00
S04
90
03:20
S05
110
00:01
The values of the program table can now be entered to the RD3 program controller.
FDL (E2) 03/2011
page 35/68
Step 1 – Selecting the program and the program section:
Normal Display
Display 1 shows
e.g. 39.8
(actual temperature value)
Display 2 shows e.g. 15.01.07 13:52 - -
(actual date and time, actual state of week program timer
channel 1: Off, channel 2: Off)
Press down key
for 5 sec.
Display 1 shows
e.g. 0000
Display 2 shows
PROGRAM EDITOR
(you are in the program editor)
Press program key
Display 1 shows
0000
Display 2 shows
UserCod? 0000
(enter user code)
Enter user code using arrow keys
e.g. 0001
(basic setting, adjustable in the user level,
chap. 10).
Value is shown in both displays.
Automatically forward after 2 sec.
Display 1 shows
e.g. 01
(program P01 selected)
--- : --- PRG.
(program can be selected)
Display 2 shows
alternating
CONTINUE X/W
(information: to 1st program section with X/W)
Select program P01 or P02 using arrow
keys
Value is shown in display 1.
Press key
In the selected program P01 or P02, program sections can be selected:
Display 1 shows
e.g. 01
(section S01 selected)
P01: --- SEC.
Display 2 shows
alternating
CONTINUE X/W
section S01 has already been created.
enter new set-points for the individual variables with button
X/W
or:
Display 1 shows
e.g. 01
(section S01 selected)
P01: --- SEC.
Display 2 shows
alternating
NEW SEC. X/W
section S01 has not yet been created.
enter set-points for the individual variables with button X/W
Select sections S01 to S10 or to S20 using
arrow keys
As long as no program section has been entered, the display switches back to 01 in case of any entry
> 01, because all sections need to be entered one after the other, and each new section is created as
NEWSEC.
Example: If three programs sections have been already entered, the next section to be entered is S04.
Before this, no section > S04 can be selected.
FDL (E2) 03/2011
page 36/68
Next step – set-point entry in the desired program sections:
Basic entry principle: Access the parameters of individual program sections with button X/W one after
the other. Enter the values of the individual parameters with the arrow keys. A value flashing once after 2
seconds indicates that it has been adopted by the controller. If several parameters are to be skipped (e.g.
in order to change a parameter in a posterior program section), the parameters can be rapidly jumped
over by holding down the X/W key. If no button is pressed for more than 120 sec the controller switches
back to Normal Display. The program entered to this point remains stored.
Press key
Display 1 shows
e.g. 40.0C
(actual temperature set-point)
S01: TEMP 40.0
(variable: temperature in °C)
Display 2 shows
alternating
CONTINUE X/W
(information: go on with X/W)
Enter temperature set-point of S01 in °C
using arrow keys
Press key
Display 1 shows
e.g. 00.30
(actual section length set-point)
S01: TIME 00:30
(variable: section length in hh:mm)
Display 2 shows
alternating
CONTINUE X/W
(information: go on with X/W)
Enter section length set-point of S01 in
hh.mm using arrow keys
Value is shown in both displays.
Press key
Selecting the next program sections to be entered
Display 1 shows
e.g. 02
(section S02 selected)
P01: --- SEC.
Display 2 shows
alternating
CONTINUE X/W
Section S02 has already been created.
enter new set-points for the individual parameters with X/W.
or:
Display 1 shows
e.g. 02
(section S02 selected)
P01: --- SEC.
Display 2 shows
alternating
NEW SEC. X/W
Section S02 has not yet been created.
enter set-points for the individual parameters with X/W
Select the next section to be entered using
arrow keys
Display 1 shows
e.g. 60.0C
(actual temperature set-point)
S02:TEMP 60.0
(variable: temperature in °C)
Display 2 shows
alternating
CONTINUE X/W
(information: go on with X/W)
Enter the temperature set-point of S02 in °C
using arrow keys
Etc.
If all sections up to S10 or up to S20 have been programmed, section S01 follows again. In order to quit
the entry mode, press the “EXIT” button several times or wait 120 sec → the controller will then return to
Normal Display.
When changing the set-point, check the setting of the safety device (chap. 12.1).
FDL (E2) 03/2011
page 37/68
8.3
Program table template
Program editor
Program title
Project
Program No.
Date:
Section
SEC
Temperature set-point
[ °C]
TEMP
Section length
[hh.mm]
TIME
S01
S02
S03
S04
S05
S06
S07
S08
S09
S10
S11
S12
S13
S14
S15
S16
S17
S18
S19
S20
FDL (E2) 03/2011
page 38/68
8.4
Deleting a program section
A program section is deleted from the program by setting the section duration to Zero.
Normal display
Press down key
for 5 sec.
Display 1 shows
e.g. 0000
Display 2 shows
PROGRAM EDITOR
(you are in the program editor)
Press program key
Display 1 shows
0000
Display 2 shows
UserCod? 0000
(enter user code)
Enter the user code using arrow keys
e.g. 0001
(basic setting, adjustable in the user level,
chap. 10).
Value is shown in both displays.
Automatically forward after 2 sec
Display 1 shows
e.g. 01
(program P01 selected)
--- : --- PRG.
(program can be selected)
Display 2 shows
alternating
CONTINUE X/W
(information: to 1st program section with X/W)
Select the desired program, e.g. P01, using
arrow keys
Value is shown in display1.
Press key
In the selected program P01 or P02, program sections can be selected:
Display 1 shows
e.g. 01
(actual selection of the section: S01)
P01: --- SEC.
(program section can be selected)
Display 2 shows
alternating
CONTINUE X/W
(information: set-point entry with X/W)
Select desired section, e.g. S03, using arrow
keys
(omitted if section S01 shall be deleted).
Press key
Display 1 shows
e.g. 90.0C
(actual temperature set-point)
S03:TEMP 90.0
(variable: temperature)
Display 2 shows
alternating
CONTINUE X/W
(information: continue with X/W)
No entry
Press key
Display 1 shows
e.g. 01.00
(actual section length)
S03:TIME 01:00
(variable: section length)
Display 2 shows
alternating
CONTINUE X/W
(information: continue with X/W)
Enter set-point Zero for section length in
hh:mm of S03 using arrow keys
Value is shown in display 2 or in both displays
(display depends on maximum time setting in
the user level chap. 10)
Display 1 shows
e.g. 00.00
(actual section length)
S03:TIME 00:00
(variable: section length)
Display 2 shows
alternating
DELETE SEC. X/W
(information: delete section with X/W)
Press key
FDL (E2) 03/2011
page 39/68
The following section (in our example now S03) is displayed:
Display 1 shows
e.g. 03
(actual selection of the section: S03)
P01:S03
(program section can be selected)
Display 2 shows
alternating
CONTINUE X/W
(information: set-point entry with X/W)
Press key
EXIT
or wait 120 sec
Controller returns to Normal Display
If you delete a program section which is followed by further sections, those following move up
in place of the deleted section.
In our example, section S03 has been deleted. If sections S04, S05, etc. have been programmed earlier,
they will now replace the preceding sections, i.e., S04 is now called S03 etc.
Deletion leads to overwriting the section by the following one. It is therefore not possible to temporarily
inactivate a program section. To enter a section later to a program, all the sections following the new one
must be entered again.
9.
Program start level
Before starting the program, check the set-point entered in Fixed value operation mode. After end of the
program, the temperature will equilibrate to this value. This value must not exceed the permitted drying
temperature for the used solvent quantity.
DANGER
Too high temperature after the program ends.
Danger of explosion.
Danger of death.
∅ Set-point of Fixed value operation must NOT exceed the maximum drying temperature
suitable for the solvent quantity.
� Check the set-point of Fixed value operation and if necessary adapt it.
After the program ends, the temperature will equilibrate to the set-point entered in Fixed value operation
mode. If the week program timer is active, another set-point (SP2) might be targeted according to pro-
gramming. Temperatures too high for the introduced solvent quantity can occur. Deactivate the week
program timer before starting the program (default setting, setting in the User level, chap. 10).
DANGER
Too high temperature after the program ends.
Danger of explosion.
Danger of death.
� Deactivate the week program timer before starting the program.
In the first step, select a program. This is on condition that a program has been entered previously (chap.
8.2) and that program type “2 programs with 10 sections each” has been selected in the user level (chap.
10).
FDL (E2) 03/2011
page 40/68
Then define the settings for the program course. Two parameters can be set:
• Program delay time, i.e. a defined time before a program starts. It can be entered with a precision of 1
minute, and its maximum value is 99.59 (99 hs 59 min). If the value is 00.00, the program will start
immediately.
• Number of program cycles, i.e. the desired number of program repeats. Values from 1 to 99 can be
entered. If the program is not going to be repeated, enter the value “0”. For infinite repeats enter the
value “–1”. The program is repeated as a whole; it is not possible to repeat individual sections.
In the last step start the selected program. These steps must be carried out subsequently.
Deactivate the week program timer (factory setting, setting in the user level, chap. 10) before
starting a program.
Step 1 – Program selection (only with program type “2 programs” set):
Normal Display
Press program key
Display 1 shows
e.g. 1
(actual selection of the program)
Display 2 shows
SEL.PRG.
(select program 1 or 2)
Enter program number 1 or 2 using arrow
keys
Value is shown in display 1.
Next step – entry of program course settings
Press program key
Display 1 shows
e.g. 00.00
(entered delay time hh.mm)
Display 2 shows
RUN TIME
(enter delay time of program start)
Set delay time in hh.mm using arrow keys
Value is shown in display 1.
Press program key
Display 1 shows
e.g. -1
(actual selection of the number of program cycles)
Display 2 shows
REPEAT
(enter number of program cycles)
Select number of cycles -1, 0, 1 etc. using
arrow keys
Value is shown in display 1.
Last step – program start:
Press program key
Display 1 shows
e.g. 1
(selected program)
Display 2 shows
RUN PRG.
(Question: start selected program?)
Press program key
Display 1 shows
e.g. 25.5C
actual temperature value
Display 2 shows
P01:S01 00:29:39
(time running backwards)
(actual program P01, actual section S01, and remaining
time of program section S01)
Program is running. The green LED (3d) lights up.
In addition to the green LED (3d) indicating a running program, the LED (3a) is lit if the heating is active,
or no LED if the actual temperature equals the set-point.
FDL (E2) 03/2011
page 41/68
During program course the arrow keys and the EXIT button are not functional.
By pressing the program key
for 3 seconds, you can terminate the program course.
If you press button
during program course, the entered set-point of the actually running program sec-
tion is shown for 5 seconds:
Display 1 shows
e.g. 65.5C
(actual temperature value)
Display 2 shows
P01:S03 00:47:12
(actual program P01, actual section S03, and remaining
time of program section S03)
Press key
Display 1 shows
e.g. 90
(actual temperature set-point 1)
Display 2 shows
SP1 TEMPERATURE
5 seconds
Display 1 shows
e.g. 30
(actual temperature set-point 2)
Display 2 shows
SP2 TEMPERATURE
(no function during program operation)
5 seconds
After program runoff (and, if appropriate, of the program repeats) the controller returns to fixed value op-
eration showing Normal Display and adjusting to the temperature value that has been previously entered
in the fixed value entry mode.
10.
User level
In this menu the following parameters can be se (in brackets the corresponding abbreviated information
given in display 2):
• Unit address (Adress)
Set the controller address (1 to 255) for operation with the communication software APT-COM™.
• User code (User-cod)
Modification of the user code setting (factory setting 0001) for access to the user level and the pro-
gram editor.
Keep in mind any modification of the user code. There is no access to these levels
without a valid user code.
• Decimal point position (Decimal)
Selection if integer values or one position after the decimal point can be entered. The integer repre-
sentation is shown in Display 2 (set-point entry) while the actual value in Display 1 is always shown
with one decimal point.
• Audio Alert (Buzzer)
Inactive: no audible signal (buzzer) in case of an alarm event.
Active: in case of an alarm event (see chap. 11.2) an audible signal (buzzer) will sound. It can be re-
set by pressing the “EXIT” button.
• Selection of controller menu language (Language)
German, English, or French can be selected.
FDL (E2) 03/2011
page 42/68
• Counter of operating hours (Oper.hs)
Information about the number of operating hours currently reached or since the last reset. (no setting,
display only).
• Max. number of operating hours (Op.limit)
Enter a limit number of operating hours, i.e., the maximum number of operating hours that can be run.
Maximum setting: 9999. Reaching the limit has no effect.
• Reset operating hours (Op.back)
Reset operating hours to zero.
• Interface protocol (Protocol)
“Modbus”: The chamber interface can be used as a communication interface to connect it to a com-
puter. This serves to control the chamber by the communication software APT-COM™. It is possible
to read and write the values of all parameters.
“Printer”: A protocol printer (option) for data printouts can be connected to the chamber interface. The
printer regularly protocols the actual temperature value with fixed formatting and with adjustable print
intervals (chap. 13.2).
In both cases an interface converter RS 422 / RS 232 is used.
• Print interval (Prt.-Inv.)
Set the print interval in minutes. Function is available only if setting “Printer” has been selected in the
previous menu point.
• Display illumination (Disp.LED)
Select between continuous display illumination and limited illumination, which will automatically go off
300 sec after the last entry.
• Program type selection (PrgSelec)
Select between entry of two programs with up to 10 sections each or of one program with up to 20
sections.
When changing from 2 programs to 1 program or vice-versa, existing programs are
deleted in the program editor.
• Maximum section duration (Prg.Time)
The maximum length of an individual program section can be set to either 99 hs 59 min or to 999 hs
59 min. This setting is then valid for all program sections.
When changing the maximum duration setting, pre-existing programs will be deleted in
the program editor.
• Set-point programming type (Setp.sim)
Select between “Ramp” and “Step”. With setting “Step” selected, you don’t need to program the transi-
tion section in the Program Editor.
If you select setting “Step”, the controller will equilibrate only to constant temperatures;
programming ramps becomes impossible.
A change between settings “ramp” and “step” will influence all programs. Note that signifi-
cant change in time courses may arise in existing programs.
• Tolerance limit range (Tol.band)
Entry of a tolerance limit value in °C. If the actual value of temperature exceeds the set-point of a pro-
gram section by more than the entered tolerance limit value, the program is halted (LED (3d) flashing)
until the actual temperature value is again within the tolerance range.
Entry of “0” means tolerance limits are off.
FDL (E2) 03/2011
page 43/68
• Activating or inactivating the week program timer (Prog.Clk)
„Inactive“: The week program timer is turned off (factory setting). The corresponding setting menu
(chap. 7) is not visible, nor is set-point 2 in the “Fixed value entry mode” (chap. 6).
„Active“: The week program timer is activated.
When deactivating the week program timer, any programming made in advance will re-
main in memory and take effect when the week program timer is activated again..
Deactivate the week program timer before staring a program (chap. 9).
• Display mode (12h/24h)
Select between 12 hours (display “AM” or “PM”) or 24 hours.
• Date of the real time clock (Date)
Main menu. Use the program key to access the settings of year, month, and day in the corresponding
submenus.
• Year of the real time clock (Year)
Enter the year (2006 up to 2050)
• Month of the real time clock (Month)
Enter the month (1 up to 12).
• Day of the real time clock (Day)
Enter the day (1 up to 31).
• Time of the real time clock (Time)
Main menu. Use the program key to access the settings of hour and minute in the corresponding sub-
menus.
There is no automatic switch between daylight saving time and regular time.
• Hour of the real time clock (Hour)
Enter the hour (0 up to 23).
• Minute of the real time clock (Minute)
Enter the minute (0 up to 59).
FDL (E2) 03/2011
page 44/68
Normal Display
Display 1 shows
e.g. 19.8
(actual temperature value)
Display 2 shows e.g. 15.05.06 13:52 - -
(actual date and time, actual switching state of week pro-
gram timer channel 1: Off, channel 2: Off)
Press down key
for 5 sec
Display 1 shows
e.g. 0000
Display 2 shows
PROGRAM EDITOR
(you are in the program editor)
Press down key
for 5 sec
Display 1 shows
0000
Menu visible only if week program timer is activated.
Display 2 shows WEEK PROG. EDITOR
(you are in the week program editor)
Press down key
for 5 sec
Display 1 shows
0000
Display 2 shows
USER – LEVEL
(you are in the user level)
Press program key
Display 1 shows
0000
Display 2 shows UserCod?
0000
(enter user code, display flashes)
Enter the user code using arrow keys
e.g. 0001 (basic setting,
or the valid code in case it has been previously
changed in this menu).
Value is shown in both displays.
Automatically forward after 2 sec
Display 1 shows
1
(actual address: 1)
Display 2 shows Adress
1
(entry of unit address)
(actual address: 1)
Enter the unit address (1 up to 254) using
arrow keys
Address is shown in both displays.
Press key
Display 1 shows
1
(actually valid user code: 1)
Display 2 shows User-cod
1
(change user code)
(actually set: 1)
Enter a new value using arrow keys
Value is shown in both displays.
Press key
Display 1 shows
0000
(no function)
Display 2 shows Saf.mode: Limit
(no function)
Press key
Display 1 shows
0
(no function)
Display 2 shows Saf.setp
0
(no function)
Press key
Display 1 shows
0000
(no function)
Display 2 shows Decimal:
XXX.X
(setting of decimal point position)
(actual setting: XXX.X)
Select decimal point position using arrow
keys
Decimal point position XXX.X or XXXX. is
shown in display 2.
FDL (E2) 03/2011
page 45/68
Press key
Display 1 shows
0000
(no function)
Display 2 shows Buzzer :
Active
(setting of the alarm buzzer)
(actual setting: „Active“)
Select between “Active” and “Inactiv” using
arrow keys
Setting is shown in display 2.
Press key
Display 1 shows
0000
(no function)
Display 2 shows Language : English
(selection of controller language)
(actual setting: English)
Select between languages German, English,
and French using arrow keys
Setting is shown in display 2.
Press key
Display 1 shows
e.g. 0004
(unit operating hours)
Display 2 shows
Oper.hs 0004:28
(operating hours up to now hhhh:mm)
(actually displayed: 4 hs 28 min)
Press key
Display 1 shows
1000
(actual setting: 1000 hs)
Display 2 shows Op.limit
1000:00
(maximum number of operating hours
hhhh:mm (actual setting: 1000 hs)
Set value using arrow keys
Setting is shown in both displays.
Press key
Display 1 shows
0000
(no function)
Display 2 shows Op.back :
No
(reset counter of operating hours ?)
(actual setting: No)
Select between “Yes” and “No” using arrow
keys
Setting is shown in display 2.
Press key
Display 1 shows
0000
(no function)
Display 2 shows Protocol:
MODBUS
(Selection of interface mode)
(actual setting: Modbus)
Select between protocols “MODBUS” and
“Printer” using arrow keys
Setting is shown in display 2.
Press key
Display 1 shows
e.g. 3
(actual setting: 3 min)
Display 2 shows Prt-Inv.
3
(print interval)
(actual setting: 3 min)
Select value between 0 and 255 minutes
using arrow keys
Setting is shown in displays 1 and 2.
Press key
Display 1 shows
0000
Display 2 shows Disp.LED:
No
(continuous display illumination?)
(actual setting: No)
Select between “Yes” and “No” using arrow
keys
Setting is shown in display 2.
FDL (E2) 03/2011
page 46/68
Press key
Display 1 shows
0000
Display 2 shows PrgSelec:
2Prg10S
(1 program with max. 20 sections or
2 programs with max. 10 sections each?)
(actual setting: 2Prg10S)
Select between “2Prg10S” and “1Prg20S”
using arrow keys
Setting is shown in display 2.
Press key
Display 1 shows
0000
Display 2 shows Prg.Time:
99:59
(max. section length 99:59 or 999:59?)
(actual setting: 99:59)
Select between 99:59 in hh:mm or 999:59 in
hhh:mm using arrow keys
Setting is shown in display 2.
Press key
Display 1 shows
0000
Display 2 shows Setp.sim
Ramp
(ramp or step?)
(actual setting: ramp)
Select between Ramp and Step using arrow
keys
Setting is shown in display 2.
Press key
Display 1 shows
0000
Display 2 shows Tol.band
0
(Tolerance limits in °C)
(actual setting: 0)
Set value in °C using arrow keys
Setting is shown in display 2.
Press key
Display 1 shows
0000
Display 2 shows Prog.Clk
Inactive
(Week program timer active or inactive?
(actual setting: Inactive)
Select between “Active” and “Inactive” using
arrow keys
Setting is shown in display 2.
Press key
Display 1 shows
0000
Display 2 shows 12h/24h
24h
(Display mode 12 hours or 24 hours?
(actual setting: 24h)
Select between 12 hours and 24 hours us-
ing arrow keys
Setting is shown in display 2.
FDL (E2) 03/2011
page 47/68
Press key
Display 1 shows
0000
Display 2 shows Date
(Main menu: Setting the date of the real time clock)
Press program key
Display 1 shows
e.g. 2006
(Actual setting: 2006)
Display 2 shows Year
2006
(Setting the year of the real time clock)
Set year (2006 up to 2050) using arrow keys
Setting is shown in display 2.
Press key
Display 1 shows
e.g. 5
(Actual setting: may)
Display 2 shows Month
5
(Setting the month of the real time clock)
Set month (1 up to 12) using arrow keys
Setting is shown in display 2.
Press key
Display 1 shows
e.g. 15
(Actual setting: 15)
Display 2 shows Day
15
(Setting the day of the real time clock)
Set day (1 up to 31) using arrow keys
Setting is shown in display 2.
Press key
EXIT
Press key
Display 1 shows
0000
Display 2 shows Time
(Main menu: Setting the time of the real time clock)
Press program key
Display 1 shows
e.g. 13
(Actual setting: 13, i.e. 1 p.m.)
Display 2 shows Hour
13
(Setting the hour of the real time clock)
Set hour (0 up to 23) using arrow keys
Setting is shown in display 2.
Press key
Display 1 shows
e.g. 30
(Actual setting: 30 minutes)
Display 2 shows Minute
30
(Setting the minute of the real time clock)
Set minute(0 up to 59) using arrow keys
Setting is shown in display 2.
Press several times key
EXIT
or wait for 120 seconds
Controller returns to normal display.
FDL (E2) 03/2011
page 48/68
11.
Behavior in case of failures
11.1 Behavior after power failure
Power failure during fixed-value operation (Normal Display): the entered parameters remain saved.
After power returns, operation continues with the set parameters. The heating starts operating after a
delay time of 4 minutes.
Power failure during program operation: After power return program course continues with the set-
points that have been reached previously during program operation. The heating starts operating after a
delay time of 4 minutes.
11.2 Alarm messages
Alarm messages, e g. “RANGE ERROR CH1” in case of sensor rupture, are shown in Display 2 only in
Normal Display.
A buzzer can be activated / deactivated in the user level (chap. 10). It can be reset by pressing the EXIT
button. The alarm text shown in Normal Display goes off only if the cause of the alarm does not exist any
longer.
12.
Safety devices
12.1 Temperature safety device class 2 (DIN 12880:2007)
The temperature safety device class 2 protects the unit, its environment and the charging material from
exceeding the maximum permissible temperature.
Please observe the guideline BGI/GUV-I 850-0 on safe working in laboratories (formerly BGR/GUV-R
120 or ZH 1/119 laboratory guidelines issued by the employers’ liability insurance association) (for Ger-
many).
In the event of a fault in the temperature controller, the safety device (2) permanently turns off the unit.
Only the fan continues running for safety reasons. This status is reported visually by the indicator lamp
(2a) and acoustically by the buzzer.
The operation of the safety device (2) is checked by moving it slowly counter-clockwise until it is turned
off. The safety device cut-off is reported visually by the indicator lamp (2a) and acoustically by the buzzer.
Then release the safety device by pressing the reset button (2b) and turn on the unit is as described pre-
viously.
Figure 9: Safety thermostat class 2
Function:
The safety thermostat class 2 is functionally and electrically independent of the temperature control de-
vice and turns off the unit at all poles.
When the control knob (2) is set to the end stop (position 10), the safety thermostat class 2 acts as a unit
protection device. If it is set somewhat higher than the set-point temperature selected on the controller, it
acts as a material protection device.
(2)
(2a)
(2b)
FDL (E2) 03/2011
page 49/68
When the safety device has turned off the unit, identifiable by the red alarm lamp (2a) lighting up, proceed
as follows:
• Disconnect the unit from the power supply.
• Have the cause of the fault examined and rectified by a technician.
• Release safety thermostat class 2 by pressing reset button (2b).
• Restart the unit as described in chap. 5.
Setting:
The diagram given in chap. 2.3 indicates the drying temperature to be set on the controller in relation to
the quantity of the introduced solvents. This must not be exceeded. For this reason, adjust the tempera-
ture safety device according to the selected set-point.
DANGER
Excessive drying temperature.
Danger of explosion.
Danger of death.
∅ Do NOT exceed the maximum drying temperature for the solvent quantity.
To check at which temperature the safety device activates, turn on the oven and set the required nominal
value on the temperature controller.
The scale division from 1 to 10 corresponds to the temperature range from 30 °C / 86°F up to 320 °C /
608°F and serves as a setting aid.
• Detach the plastic cover over the temperature safety device class
2 (2) with a suitable Phillips screwdriver.
• Turn the control knob (2) of the safety device using a coin to its
end-stop (position 10) (unit protection).
• When the set point is reached, turn back the control knob (2) until
its trip point (turn it counter-clockwise) is reached.
• The red alarm lamp (2a) lighting up identifies the trip point; and the
reset button (2b) pops out.
• The optimum setting of the safety device is obtained by turning the
knob clockwise by approx. one graduation mark on the scale.
• Push the reset button (2b) in again.
• Restore the plastic cover to prevent misadjusting
Figure 10: Setting the safety
device class 2
The unit is only active when the reset button (2b) is pushed in.
When the safety thermostat class 2 responds, the red alarm lamp (2a) illuminates, the reset button (2b)
pops out and the unit turns off permanently at all poles. Only the fan keeps turning for safety reasons.
Check the safety thermostat regularly and adjust it following changes of the set-point.
Fixed value operation: Adapt the temperature safety device every time the set-point for
temperature is changed. Set the set-point of temperature safety device by about 5 °C to 10
°C above the controller temperature set-point.
Program operation: Adapt the temperature safety device to the highest temperature set-
point value of the program actually used.
Functional check:
Test the functional capability of the safety thermostat at appropriate intervals. It is recommended having
this check performed by the authorized service personnel, e.g. before the beginning of a long working
process.
FDL (E2) 03/2011
page 50/68
12.2 Exhaust air monitoring
If the volumetric flow rate of the exhaust air is too low, the heating turns off immediately for safety rea-
sons. The indicator light (3) lights up and an audible signal is given. The indicator light and the buzzer
can only be turned off by pushing key (4). If the cause of the low volume of exhaust air is not resolved
when the key is released, the error messages will be given again.
RESET
(4) (3)
(3)
Red indicator light, loss of technical ventilation
(4)
RESET key for the heating cut-out due to failure of tech-
nical ventilation
Figure 11: Instrument panel (detail)
It is the operator’s responsibility to ensure that the doors of drying ovens are opened immediately on fail-
ure of the technical ventilation system (GUV-R 500 chap. 2.28).
On failure of the technical ventilation, open the door of the safety / paint drying oven.
13.
Options
13.1 Communication software APT-COM™ 3 DataControlSystem (option)
The unit is regularly equipped with a serial interface RS 422 that can connect the BINDER communica-
tion software APT-COM™ 3 DataControlSystem. The connection to a computer is established using the
FDL interface via an interface converter RS 422 / RS 232.
Confirm that the interface mode is correctly set to “Modbus” in the user level (chap. 10).
The actual temperature values are given at adjustable intervals. Programming can be performed graphi-
cally via PC. Up to 30 chambers with RS 422 interface can be cross linked. For further information,
please refer to the operating manual of the BINDER communication software APT-COM™ 3.
Pin allocation of the RS 422 interface at the rear of the incubator:
Pin 2:
RxD (+)
Pin 3:
TxD (+)
Pin 4:
RxD (-)
Pin 5:
TxD (-)
Pin 7:
Ground
13.1 Ethernet interface (option)
With this option, the chamber is equipped with an Ethernet interface that can connect the BINDER com-
munication software APT-COM™ 3 DataControlSystem. The actual temperature values are given at ad-
justable intervals. Programming can be performed graphically via PC. Up to 30 chambers with RS 422
interface can be cross-linked. The MAC Address is indicated below the Ethernet interface. For further
information, please refer to the operating manual of the BINDER communication software APT-COM™ 3.
The additional RS422 interface is only used for service purposes. Do NOT connect it to any network. The
interface is labeled accordingly.
FDL (E2) 03/2011
page 51/68
13.2 Protocol printer (option)
The protocol printer is connected using the FDL interface via an interface converter RS 422 / RS 232.
Confirm that the interface mode is correctly set to “Printer” in the user level (chap. 10).
Figure 12: Protocol printer
Interface configuration:
Baud rate:
9600
Stop bit:
1
Parity:
none
The actual temperature values are put out regularly with fixed formatting.
Printout: one printed line for each print interval with relative time stamp, temperature value with one deci-
mal point, curve representation (see “Manual for Setting Matrix Printer Epson LX-300+“, Art. No. 7001-
0041). In every 5th line the set print interval “Ptime” in minutes is noted. The printer interval is set in the
user level (chap. 10).
Example:
Temp.: 34.7 *
Temp.: 35.6 *
Temp.: 32.8 *
Temp.: 30.1 *
Ptime : 001. *
Temp.: 27.4 *
Temp.: 26.9 *
Temp.: 26.6 *
Temp.: 26.4 *
Ptime : 001. *
Temp.: 26.2 *
In this example the print interval is set to 1 min, i.e., every minute a new temperature value is printed.
In conjunction with the specified print interval, real-time reference can be achieved by noting
down the starting time of recording.
For operating instructions of the printer, please refer to the operating manual of the printer
manufacturer.
FDL (E2) 03/2011
page 52/68
13.3 Additional measuring channel for digital object temperature indicator with
magnetic sensor or pincer-type sensor (option)
The object temperature display enables the determination of the actual temperature of the specimen dur-
ing the whole process. The supplied pincer-type or magnetic sensor is used to display the object tem-
perature at Display 2 of the RD3 controller.
The object temperature data are put out together with the data of the temperature controller to the RS
422 interface as a second measuring channel and can be documented by the communication software
APT-COM™ developed by BINDER (option, chap. 13.1).
Socket
(inner chamber
Pincer-type sensor
Material sample
Magnetic sensor
Figure 13: Sensor probes for optional temperature display
With this option the chamber is equipped with an analog output 4-20 mA for temperature. This output
permits transmitting data to external data registration systems or devices.
The connection is realized as a DIN socket at the rear of the chamber as following:
ANALOG OUTPUT 4-20 mA DC
PIN 1: temperature –
PIN 2: temperature +
Temperature range:
0 °C / 32 °F to +300 °C / 572 °F
A suitable DIN plug is enclosed.
Figure 14: Pin allocation of DIN socket
FDL (E2) 03/2011
page 53/68
14.
Maintenance, cleaning, and service
14.1 Maintenance intervals, service
DANGER
Electrical hazard.
Danger of death.
∅ The unit must NOT become wet during operation or maintenance work.
∅ Do NOT remove the rear panel of the unit.
� Disconnect the unit before conducting maintenance work. Disconnect the power plug.
� Ensure all maintenance work is conducted by licensed electricians or experts author-
ized by BINDER.
Ensure regular maintenance work is performed at least once a year.
The warranty becomes void if maintenance work is conducted by non-authorized personnel.
It is particularly important to test the function of the flow monitor once a year.
Record the results of the maintenance tests in a service log
Replace the door gasket only when cold. Otherwise, the door gasket may become damaged.
We recommend taking out a maintenance agreement. Please consult BINDER Service.
BINDER telephone hotline:
+49 (0) 7462 2005 555
BINDER fax hotline:
+49 (0) 7462 2005 93555
BINDER e-mail hotline:
[email protected]
BINDER service hotline USA:
+1 866 885 9794 or +1 631 224 4340 (toll-free in the USA)
BINDER service hotline Spain
+34 9492 677 23
BINDER service hotline Asia Pacific:
+852 39070500 or +852 39070503
BINDER service hotline Russia and CIS
+7 495 98815 17
BINDER Internet website
http://www.binder-world.com
BINDER address
BINDER GmbH, post office box 102, D-78502 Tuttlingen
International customers, please contact your local BINDER distributor.
FDL (E2) 03/2011
page 54/68
14.2 Replacing the intake filter
The fresh air intake filter (fine particle filter for particles 1µm up to 10 µm, class F6/EU6 acc. to EN
779:2002) at the top right-hand side must be replaced from time to time, depending on the degree of
soiling. Pull out the slide and replace the filter cartridge.
Intake filter
Slide
Figure 15: Fresh air intake filter
14.3 Cleaning and decontaminating the safety / paint drying oven
DANGER
Electrical hazard.
Danger of death.
∅ Do NOT spill water or cleaning agents over the inner and outer surfaces.
� Disconnect the chamber before cleaning. Disconnect the power plug.
� Completely dry the appliance before turning it on again.
Cleaning
Always keep clean the inner parts of the safety / paint drying oven including the drip pans, drip
trays and exhaust-air ducts. Remove the remnants of coating material at suitable intervals.
Disconnect the chamber from the power supply before cleaning. Disconnect the power plug.
Wipe the surfaces with a moistened towel. In addition, you can use the following cleaning agents:
Exterior surfaces
inner chamber
racks
door gaskets
Standard commercial cleaning detergents free from acid or halogenides.
Alcohol based solutions.
We recommend using the neutral cleaning agent Art. No. 1002-0016.
Instrument panel
Standard commercial cleaning detergents free from acid or halogenides.
We recommend using the neutral cleaning agent Art. No. 1002-0016.
For surface protection, perform cleaning as quickly as possible.
After cleaning completely remove any cleaning agents from the surfaces with a moistened
towel.
Soapsuds may contain chlorides and must therefore NOT be used for cleaning.
FDL (E2) 03/2011
page 55/68
Decontamination
Disconnect the chamber from the power supply prior to decontamination. Pull the power plug.
You can use the following disinfectants:
Inner chamber
Standard commercial surface disinfectants free from acid or halogenides.
Alcohol based solutions.
We recommend using disinfectant Art. No. 1002-0022.
In case of contamination of the interior by biologically or chemically hazardous goods, there are three
possible procedures depending on the type of contamination and charging material.
(1) The safety / paint drying ovens FDL can be hot air sterilized at 190 °C / 374°F for at least 30 minutes.
All inflammable goods must be removed from the interior before.
(2) Spray the inner chamber with an appropriate disinfectant.
Before start-up, the unit must be absolutely dry and ventilated, as explosive gases may form during
the decontamination process.
(3) If necessary, have strongly contaminated inner chamber parts removed by an engineer for cleaning,
or have them exchanged. Sterilize the inner chamber parts in a sterilizer or autoclave.
With every decontamination method, always ensure adequate personal safety.
CAUTION
Danger of corrosion.
Damage to the unit.
∅ Do NOT use acidic or chlorine cleaning detergents.
We recommend using the neutral cleaning agent Art. No. Art. Nr. 1002-0016 for a thorough
and mild cleaning.
Any corrosive damage that may arise following use of other cleaning agents is excluded from
liability by BINDER GmbH.
14.4 Sending the unit back to BINDER GmbH
If you return a BINDER product to us for repair or any other reason, we will only accept the product upon
presentation of an authorization number that has previously been issued to you. An authorization number
will be issued after receiving your complaint either in writing or by telephone prior to your sending the
BINDER product back to us. The authorization number will be issued following receipt of the information
below:
• BINDER product type and serial number
• Date of purchase
• Name and address of the dealer from which you bought the BINDER product
• Exact description of the defect or fault
• Complete address, contact person and availability of that person
• Exact location of the BINDER product in your facility
• A contamination clearance certificate (chap. 18) must be faxed in advance
The authorization number must be applied to the packaging in such a way that it can be easily recog-
nized or be recorded clearly in the delivery documents.
For security reasons we cannot accept a unit delivery if it does not carry an authorization num-
ber.
FDL (E2) 03/2011
page 56/68
15.
Disposal
15.1 Disposal of the transport packing
Packing element
Material
Disposal
Straps to fix packing on pallet
Plastic
Plastic recycling
Non-wood (compressed match-
wood, IPPC standard)
Wood recycling
Wooden transport box (option)
with metal screws
Metal
Metal recycling
Solid wood (IPPC standard)
Wood recycling
Pallet
with foamed plastic stuffing
PE foam
Plastic recycling
Cardboard
Paper recycling
Shipping box
with metal clamps
Metal
Metal recycling
Edge protection
Styropor® or PE foam
Plastic recycling
Protection of doors and racks
PE foam
Plastic recycling
Bag for operating manual
PE foil
Plastic recycling
Insulating air cushion foil (packing of
optional accessories)
PE foil
Plastic recycling
If recycling is not possible, all packing parts can also be disposed of with normal waste.
15.2 Decommissioning
Turn off the main power switch (1). Disconnect the unit from the power supply.
When switching off the main power switch ON / OFF (1), the stored parameters remain saved.
• Temporal decommissioning: See indications for appropriate storage, chap. 3.3.
• Final decommissioning: Dispose of the unit as described in chap. 15.3 to 15.5.
15.3 Disposal of the unit in the Federal Republic of Germany
According to directive 2002/96/EC of the European Parliament and of the Council on waste electrical and
electronic equipment (WEEE), BINDER devices are classified as “monitoring and control instruments”
(category 9) only intended for professional use“. They must not be disposed of at public collecting points.
The safety / paint drying oven FDL bears the symbol for the marking of electrical and elec-
tronic equipment manufactured / placed on the market in the EC after 13 August 2005 and
be disposed of in separate collection according to the directive 2002/96/EC on waste elec-
trical and electronic equipment (WEEE) and German national law for electrical and elec-
tronic equipment (Elektro- und Elektronikgerätegesetz, ElektroG). WEEE marking: crossed-
out wheeled bin with solid bar under. A significant part of the materials must be recycled in
order to protect the environment.
At the end of the device’s service life, have the device disposed of according to the German national law
for electrical and electronic equipment (Elektro- und Elektronikgerätegesetz, ElektroG) from 23 March
2005, BGBl. I p. 762 or contact BINDER service who will organize taking back and disposal of the unit
according to the German national law for electrical and electronic equipment (Elektro- und Elektronik-
gerätegesetz, ElektroG) from 23 March 2005, BGBl. I p. 762.
FDL (E2) 03/2011
page 57/68
CAUTION
Violation against existing law.
∅ Do NOT dispose of BINDER devices at public collecting points.
� Have the device disposed of professionally at a recycling company that is certified ac-
cording to the German national law for electrical and electronic equipment (Elektro- und
Elektronikgerätegesetz, ElektroG) from 23 March 2005, BGBl. I p. 762.
or
� Instruct BINDER service to dispose of the device. The general terms of payment and
delivery of the BINDER GmbH apply, which were valid at the time of purchasing the
unit.
Certified companies disassemble waste (used) BINDER equipment in primary substances for recycling
according to directive 2002/96/EC. The devices must be free from toxic, infectious or radioactive sub-
stances in order to eliminate any health hazards to the employees of the recycling companies.
Prior to handing the unit over to a recycling company, it is the user’s responsibility that it is
free from toxic, infectious or radioactive substances.
− Prior to disposal clean all introduced or residual toxic substances from the unit.
− Prior to disposal disinfect the unit from all sources of infection. Be aware that sources of
infection may also be located outside the inner chamber.
− If you cannot safely remove all toxic substances and sources of infection from the unit,
dispose of it as special waste according to national law.
− Fill out the contamination clearance certificate (chap. 18) and enclose it with the unit.
WARNING
Contamination of the device with toxic, infectious or radioactive substances.
Danger of intoxication.
Danger of infection.
∅ NEVER take a unit contaminated with toxic substances or sources of infection for recy-
cling according to directive 2002/96/EC.
� Prior to disposal, remove all toxic substances and sources of infection from the unit.
� A unit from which all toxic substances or sources of infection cannot be safely removed
must be considered as “special” waste according to national law. Dispose of it accord-
ingly.
15.4 Disposal of the unit in the member states of the EC except for the Federal
Republic of Germany
According to directive 2002/96/EC of the European Parliament and of the Council on waste electrical and
electronic equipment (WEEE), BINDER devices are classified as “monitoring and control instruments”
(category 9) only intended for professional use“. They must not be disposed of at public collecting points.
The safety / paint drying oven FDL bears the symbol for the marking of electrical and elec-
tronic equipment manufactured / placed on the market in the EC after 13 August 2005 and
be disposed of in separate collection according to the directive 2002/96/EC on waste elec-
trical and electronic equipment (WEEE). WEEE marking: crossed-out wheeled bin with
solid bar under.
FDL (E2) 03/2011
page 58/68
At the end of the device’s service life, notify the distributor who sold you the device, who will take back
and dispose of the unit according to the directive 2002/96/EC of 27 January 2003 on waste electrical and
electronic equipment (WEEE).
CAUTION
Violation against existing law.
∅ Do NOT dispose of BINDER devices at public collecting points.
� Have the device disposed of professionally at a recycling company that is certified ac-
cording to conversion of the directive 2002/96/EC into national law.
or
� Instruct the distributor who sold you the device to dispose of it. The agreements apply
that were agreed with the distributor when purchasing the unit (e.g. his general terms of
payment and delivery).
� If your distributor is not able to take back and dispose of the unit, please contact
BINDER service.
Certified companies disassemble waste (used) BINDER equipment in primary substances for recycling
according to directive 2002/96/EC. The devices must be free from toxic, infectious or radioactive sub-
stances in order to eliminate any health hazards to the employees of the recycling companies.
Prior to handing the unit over to a recycling company, it is the user’s responsibility that it is
free from toxic, infectious or radioactive substances.
− Prior to disposal, clean all introduced or residual toxic substances from the unit.
− Prior to disposal, disinfect the unit from all sources of infection. Be aware that sources of
infection may also be located outside the inner chamber.
− If you cannot safely remove all sources of infection and toxic substances from the unit,
dispose of it as special waste according to national law.
− Fill out the contamination clearance certificate (chap. 18) and enclose it with the unit.
WARNING
Contamination of the device with toxic, infectious or radioactive substances.
Danger of intoxication.
Danger of infection.
∅ NEVER take a unit contaminated with toxic substances or sources of infection for recy-
cling according to directive 2002/96/EC.
� Prior to disposal, remove all toxic substances and sources of infection from the unit.
� A unit from which all toxic substances or sources of infection cannot be safely removed
must be considered as “special” waste according to national law. Dispose of it accord-
ingly.
FDL (E2) 03/2011
page 59/68
15.5 Disposal of the unit in non-member states of the EC
CAUTION
Alteration of the environment.
� For final decommissioning and disposal of the unit, please contact BINDER service.
� Follow the statutory regulations for appropriate, environmentally friendly disposal.
The main board of the safety / paint drying oven includes a lithium cell. Please dispose of it according to
national regulations.
16.
Troubleshooting
Fault description
Possible cause
Required measures
Heating
Unit door not properly closed.
Completely close unit door.
Door gasket defective.
Replace door gasket,
Set-point temperature is not
reached after specified time.
LED (3a) (loss of technical venti-
lation) not lit.
Controller not adjusted.
Calibrate and adjust controller.
Controller defective.
Pt 100 sensor defective.
Semiconductor relay defective
Contact BINDER service.
Chamber heating permanently,
set-point not maintained.
Controller not adjusted.
Calibrate and adjust controller.
Safety device has turned off
the oven.
Limit temperature reached.
Safety device class 2 (chap.
12.1) set too low.
Allow the oven to cool down the
oven and press the “RESET” button.
Check temperature set-point and
setting of safety device (chap. 12.1).
If appropriate, select suitable limit
value.
Semiconductor relay defective.
Controller defective.
Chamber doesn’t heat up.
LED (3a) (loss of technical venti-
lation) not lit.
LED (2a) of safety device lit.
Controller display off.
Fan turning.
Safety device defective.
Contact BINDER service.
Check connection to power supply.
No power supply.
Check if the main power switch (1)
is turned on.
Unit without any function.
Controller defective.
Contact BINDER service.
Deviations from the indicated
heating-up times.
Oven fully loaded.
Charge the oven less or consider
longer heating-up times.
Chamber just turned on, nec-
essary fan speed not yet
reached.
Wait a few minutes.
Fan defective.
Contact BINDER service.
Exhaust air channel blocked.
Flap of exhaust air channel
defective.
Check exhaust air system (customer
side).
Chamber doesn’t heat up.
LED (3a) (loss of technical venti-
lation) lit.
Intake opening blocked
(soiled filter).
Change filter.
FDL (E2) 03/2011
page 60/68
Fault description
Possible cause
Required measures
Ventilation
No volumetric flow rate for fresh
air and forced-air circulation.
LED (3a) (loss of technical venti-
lation) lit.
Fan defective.
Contact BINDER service.
No volumetric flow rate for fresh
air.
LED (3a) (loss of technical venti-
lation) lit.
Intake opening blocked
(soiled filter).
Change filter.
Exhaust air channel blocked.
No fresh air or exhaust air flow.
LED (3a) (loss of technical venti-
lation) lit.
Flap of exhaust air channel
defective.
Check exhaust air system (customer
side).
No pressure differential signal at
the pipe ends
The measuring pipe in the
exhaust passage blocked.
Contact BINDER service.
Pressure differential not sufficient
to switch the pressure switch
Rupture of pressure differential
measuring hose.
Contact BINDER service.
Controller
Program duration longer than
programmed.
Inappropriate tolerances have
been programmed.
For rapid transition phases, do NOT
program tolerance limits in order to
permit maximum heating speed.
Program stops one section too
early.
Program line is incomplete.
When programming, define the end
value of the desired cycle by adding
an additional section with a section
time of at least one minute (with
setting set-point ramp).
Programs have been deleted.
Change from 2 programs to 1
program or vice-versa
When changing, ensure that the
programs are no longer needed.
The controller returns to Normal
Display from any level.
No button was pressed for
more than 120 sec.
Repeat entries, enter the values
rapidly.
Message RANGE ERROR CH1
in Normal Display in Display 2
Sensor rupture between sen-
sor and controller
Contact BINDER service.
Ramp temperature transitions are
only realized as steps.
Set-point programming type
set to “Step” in the User level
(chap. 10).
Set the set-point programming type
ti setting “Ramp” in the User level
(chap. 10).
Only qualified service personnel authorized by BINDER must perform repair.
Repaired units must comply with the BINDER quality standards.
Every fire and explosion in relation to a lacquer dryer must be reported to the employer’s liabil-
ity insurance association (for Germany).
FDL (E2) 03/2011
page 61/68
17.
Technical description
17.1 Factory calibration and adjustment
This unit was calibrated and adjusted in the factory. Calibration and adjustment were performed using
standardized test instructions, according to the QM DIN EN ISO 9001 system applied by BINDER (certi-
fied since December 1996 by TÜV CERT). All test equipment used is subject to the administration of
measurement and test equipment that is also constituent part of the BINDER QM DIN EN ISO 9001 sys-
tems. They are controlled and calibrated to a DKD-Standard at regular intervals.
17.2 Definition of usable volume
The usable volume illustrated below is calculated as follows:
b
c
c
a
b
a
a
C
B
A
A, B, C = internal dimensions (W, H, D)
a, b, c = wall separation
a = 0.1*A
b = 0.1*B
c = 0.1*C
VUSE = (A - 2 * a) * (B - 2 * b) * (C - 2 * c)
Figure 16: Determination of the usable volume
The technical data refers to the so defined usable volume.
Do NOT place samples outside this usable volume.
Do NOT load this volume by more than half to enable sufficient airflow inside the chamber.
Do NOT divide the usable volume into separate parts with large area samples.
Do NOT place samples too close to each other in order to permit circulation between them
and thus obtain a homogenous distribution of temperature and humidity.
17.3 Technical data
Exterior Dimensions
Width
mm / inch
835 / 32.87
Height (including feet)
mm / inch
800 / 31.50
Depth
mm / inch
685 / 26.97
Additional depth of door handle
mm / inch
50 / 1.97
Wall clearance rear
mm / inch
100 / 3.94
Wall clearance side
mm / inch
160 / 6.30
Exhaust duct, outer diameter
mm / inch
100 / 3.94
Total steam space volume 1)
l / cu.ft.
156 / 5.51
Interior dimensions
Width
mm / inch
600 / 23.62
Height
mm / inch
435 / 17.13
Depth
mm / inch
435 / 17.13
Interior volume
l / cu.ft
115 / 4.1
Number of racks, chrome-plated (regular / max.)
2/5
Load per rack
Kg /lbs
20 / 44
Permitted total load
Kg /lbs
50 / 110
Weight of the unit (empty)
Kg /lbs
90 / 199
FDL (E2) 03/2011
page 62/68
Temperature data
Temperature range, 5 °C above ambient up to
°C / °F
300 / 572
Maximum temperature of the heating surface
°C / °F
750 / 1382
Maximum permitted oven temperature
°C / °F
300 / 572
Temperature fluctuation
± K
0.3
at 70 °C / 158°F
± K
1.5
at 150 °C / 302°F
± K
2.5
Temperature uniformity (variation)
at 300 °C / 572°F
± K
4
to 70 °C / 158°F
min
7
to 150 °C / 302°F
min
17
Heating-up time 2)
to 300 °C / 572°F
min
44
to 70 °C / 158°F
min
1
to 150 °C / 302°F
min
3
Recovery time after door was opened
for 30 sec. 2)
to 300 °C / 572°F
min
6
Ventilation data
Air change
approx. x/min.
3
Air circulation
approx. x/min.
40
approx. l/min
400
Minimum volumetric flow rate of exhaust air 3)
approx. m3/h
24.0
Air flow rate
m/sec
0.8 up to 1.2
Solvent data
Highest permitted solvent quantity 4)
(at T-180 °C, M-100g/mol, U-40g/m3, K=0,5)
g
6.65
Electrical data
Housing protection acc. to EN 60529
IP
33
Nominal voltage (+10 %) 50/60 Hz
V
230 1 N ~
Nominal current
Amp
13.0
Nominal power
kW
2.90
Sound-pressure level
dB(A)
< 70
at 70 °C / 158°F
Wh/h
388
at 150 °C / 302°F
Wh/h
1098
Energy consumption 5)
at 300 °C / 572°F
Wh/h
2023
Power plug
shock proof plug
Over-voltage category acc. to IEC 61010-1
II
Pollution degree acc. to IEC 61010-1
2
Legend:
1) If the volume of the dry goods inserted in the dryer exceeds 10% of the total steam area, it must be
deducted in the calculation of the total steam volume.
2) Up to 98 % of the set value
3) Procedure: "calculation of the volumetric flow rate through the exhaust duct based on the velocity of
flow measured there with a propeller anemometer "
4) At a drying temperature of 180 degrees Celsius and an average molecular weight of the solvent of
M=100g/mole. For a different drying temperature, the highest permissible solvent quantity must be re-
calculated. The calculation should be made on the basis of the "principles for the calculation of ventila-
tion of chamber dryers and continuous dryers" in accordance with industrial standard EN 1539:2009,
Appendix B.
5) Use this value for sizing air condition systems.
All technical data is specified for units with standard equipment at an ambient temperature of +25 °C /
77°F and a power supply voltage fluctuation of ±10. The temperature data is determined in accordance to
BINDER factory standard following DIN 12880, observing the recommended wall clearances of 10 % of
the height, width and depth of the inner chamber.
All indications are average values, typical for units produced in series. We reserve the right to
change technical specifications at any time.
If the cabinet is fully loaded, the specified heating up times may vary according to the load.
FDL (E2) 03/2011
page 63/68
17.4 Equipment and Options
To operate the safety / paint drying oven, use only original BINDER accessories or accesso-
ries / components from third-party suppliers authorized by BINDER. The user is responsible
for any risk arising from using unauthorized accessories.
Regular equipment
Electronically controlled APT.line™ preheating chamber technology
In compliance with all safety requirements according to EN 1539: 2009 and GUV-R 500 Kap. 2.28 „Be-
treiben von Trocknern für Beschichtungsstoffe““ ("Dryers for coating materials") (for Germany)
Multifunction program controller RD3 with digital display
Adjustable ramp functions via program editor
Temperature safety controller class 2 acc. to DIN 12880:2007
Communication and printer interface RS 422
Fresh-air monitoring with visual and audible alarms and automatic heating shut-off
Replaceable fresh-air filter cartridge (fine-particle filter for particle sizes 1 µm to 10 µm, class F6/EU6 acc.
to EN 779:2002)
FKM door gasket (for max. temp. 200 °C / 392°F)
2 chrome-plated racks
Rear exhaust duct, diameter 100 mm / 3.94 inch
Options / accessories
Access ports with various diameters, with silicone plug
Rack, chrome-plated or stainless steel
Perforated rack, stainless steel
Rack lockings (4 pieces)
Reinforced rack stainless steel, with 1 set rack lockings
Reinforced inner chamber with 2 reinforced racks
Door with window and touching protection
Lockable door
Silicone door gasket resistant to high temperatures > 200 °C / 392°F. Attention: Above 250 °C / 482°F
the gasket will age faster
High temperature resistance for inflammable areas (electrical area according to protection type IP 54)
Fresh air replacement filter (class F6/EU6 acc. to EN 779:2002) for particles 1µm to 10µm), with alumi-
num frame
Measurement of air change rate acc. to ASTM D 5374:1993/2005
Additional measuring channel for digital specimen temperature display (with magnet or clip sensor) with
analog output 4-20 mA at DIN socket (DIN plug included)
Serial printer with interface, for numerical and graphical temperature registration
Factory calibration certificate, measurement in the center
Extension of factory calibration certificate (additional value)
Measuring protocol acc. to DIN 12880:2007
Qualification folder
Neutral cleaning agent (liquid concentrate)
Stable table on wheels with castors and locking brakes
FDL (E2) 03/2011
page 64/68
17.5 Spare parts
BINDER GmbH is responsible for the safety features of the unit only, provided skilled electri-
cians or qualified personnel authorized by BINDER perform all maintenance and repair, and if
components relating to chamber safety are replaced in the event of failure with original spare
parts. The user is responsible for any risks arising from using unauthorized accesso-
ries/components.
Accessories and spare parts:
Description
Art. No.
Rack, chrome-plated
6004-0003
Rack, stainless steel
6004-0008
Perforated rack, stainless steel
6004-0030
Rack lockings (4 pieces)
8012-0531
Door gasket made of FKM (temperature resistant up to 200 °C / 392 °F)
6005-0030
Door gasket silicone (high temperature resistant)
8012-0499
Controller RD3 with week program timer
5014-0102
Thermostat class 2 30 °C/ 86 °F to 320 °C / 608 °F
5006-0008
Turning knob for thermostat class 2
8009-0004
Protective cover for thermostat class 2
6002-0077
Pilot lamp red
5008-0003
Temperature sensor Pt 100 bend-off
5002-0006
Temperature sensor magnetic head (option specimen temperature display)
5002-0001
Temperature sensor pincer-type head (option specimen temperature
display)
5002-0003
Unit foot, black
6002-0006
Protocol printer
8012-0418
Fresh air replacement filter, class F6/EU6
6014-0001
Measurement of air change rate acc. to ASTM D 5374:1993/2005
8012-0195
Calibration certificate
8012-0030
Extension for calibration certificate (additional value)
8012-0022
Measuring protocol acc. to DIN 12880:2007
8012-0156
Qualification folder
DL006031
Neutral cleaning agent, 1 kg
1002-0016
Stable table on wheels with castors and locking brakes
9051-0018
FDL (E2) 03/2011
page 65/68
17.6 Dimensions FDL 115
∅100
795
22
834
685
50
435
602
435
260
320
R
A
R
A
Figure 17: Dimensions FDL 115
FDL (E2) 03/2011
page 66/68
18.
Contamination clearance certificate
Unbedenklichkeitsbescheinigung
Declaration regarding safety and health
Erklärung zur Sicherheit and gesundheitlichen Unbedenklichkeit
The German Ordinance on Hazardous Substances (GefStofV), and the regulations regarding safety at
the workplace, require that this form be filled out for all products that are returned to us, so that the safety
and the health of our employees can be guaranteed.
Die Sicherheit und Gesundheit unserer Mitarbeiter, die Gefahrstoffverordnung GefStofV und die Vorschriften zur
Sicherheit am Arbeitsplatz machen es erforderlich, dass dieses Formblatt für alle Produkte, die an uns zurück-
geschickt wird.
Note: A repair is not possible without a completely filled out form.
Ohne Vorliegen des vollständig ausgefüllten Formblattes ist eine Reparatur nicht möglich.
• A completely filled out form must be transmitted via Fax (+49 (0) 7462 2005 93555) or by letter in
advance, so that this information is available before the equipment/component part arrives. A second
copy of this form must accompany the equipment/component part. In addition, the carrier should be
notified.
Eine vollständig ausgefüllte Kopie dieses Formblattes soll per Telefax (Nr. +49 (0) 7462 2005 93555) oder Brief
vorab an uns gesandt werden, so dass die Information vorliegt, bevor das Gerät/Bauteil eintrifft. Eine weitere Ko-
pie soll dem Gerät/Bauteil beigefügt sein. Ggf. ist auch die Spedition zu informieren.
• Incomplete information or non-conformity with this procedure will inevitably lead to substantial delays
in processing. Please understand the reason for this measure, which lies outside our area of influence
and will help us to speed up this procedure.
Unvollständige Angaben oder Nichteinhalten dieses Ablaufs führen zwangsläufig zu beträchtlichen Verzögerun-
gen in der Abwicklung. Bitte haben Sie Verständnis für Maßnahmen, die außerhalb unserer Einflussmöglichkeiten
liegen und helfen Sie mit, den Ablauf beschleunigen.
• Please print and fill out this form completely.
Bitte unbedingt vollständig ausfüllen!
1.
Unit/ component part / type: / Gerät / Bauteil / Typ:
2.
Serial No./ Serien-Nr.:
3.
Details about utilized substances / biological substances / Einzelheiten über die eingesetzten Sub-
stanzen/biologische Materialien:
3.1
Designations / Bezeichnungen:
a)
____________________________________________________________________________
b)
____________________________________________________________________________
c)
____________________________________________________________________________
3.2
Safety measures required for handling these substances / Vorsichtsmaßnahmen beim Umgang
mit diesen Stoffen:
a)
____________________________________________________________________________
b)
____________________________________________________________________________
c)
____________________________________________________________________________
FDL (E2) 03/2011
page 67/68
3.3
Measures to be taken in case of skin contact or release into the atmosphere / Maßnahmen
bei Personenkontakt oder Freisetzung:
a)
____________________________________________________________________________
b)
____________________________________________________________________________
c)
____________________________________________________________________________
d)
____________________________________________________________________________
3.4
Other important information that must be taken into account / Weitere zu beachtende und
wichtige Informationen:
a)
____________________________________________________________________________
b)
____________________________________________________________________________
c)
____________________________________________________________________________
4.
Declaration on the risk of these substances (please checkmark the applicable items) /
Erklärung zur Gefährlichkeit der Stoffe (bitte Zutreffendes ankreuzen) :
� 4.1 For non toxic, non radioactive, biologically harmless materials / für nicht giftige, nicht radioak-
tive, biologisch ungefährliche Stoffe:
We hereby guarantee that the above-mentioned unit / component part… / Wir versichern, dass o.g.
Gerät/Bauteil...
� Has not been exposed to or contains any toxic or otherwise hazardous substances / weder giftige noch
sonstige gefährliche Stoffe enthält oder solche anhaften.
� That eventually generated reaction products are non-toxic and also do not represent a hazard / auch
evtl. entstandene Reaktionsprodukte weder giftig sind noch sonst eine Gefährdung darstellen.
� Eventual residues of hazardous substances have been removed / evtl. Rückstände von Gefahrstoffen entfernt
wurden.
� 4.2 For toxic, radioactive, biologically harmful or hazardous substances, or any other hazard-
ous materials / für giftige, radioaktive, biologisch bedenkliche bzw. gefährliche Stoffe oder anderweitig ge-
fährliche Stoffe.
We hereby guarantee that … / Wir versichern, dass …
� The hazardous substances, which have come into contact with the above-mentioned equip-
ment/component part, have been completely listed under item 3.1 and that all information in this re-
gard is complete / die gefährlichen Stoffe, die mit dem o.g. Gerät/Bauteil in Kontakt kamen, in 3.1 aufgelistet sind und alle
Angaben vollständig sind.
� That the unit /component part has not been in contact with radioactivity / das Gerät/Bauteil nicht mit Radioak-
tivität in Berührung kam
5.
Kind of transport / transporter / Transportweg/Spediteur:
Transport by (means and name of transport company, etc.) Versendung durch (Name Spediteur o.ä.)
___________________________________________________________________________________
Date of dispatch to BINDER GmbH / Tag der Absendung an BINDER GmbH:
___________________________________________________________________________________
FDL (E2) 03/2011
page 68/68
We hereby declare that the following measures have been taken / Wir erklären, dass folgende Maßnah-
men getroffen wurden:
� Hazardous substances were removed from the unit including component parts, so that no hazard
exists for any person in the handling or repair of these items / das Gerät/Bauteil wurde von Gefahrstoffen be-
freit, so dass bei Handhabung/Reparaturen für die betreffenden Person keinerlei Gefährdung besteht
� The unit was securely packaged and properly identified / das Gerät wurde sicher verpackt und vollständig gek-
ennzeichnet.
� Information about the hazardousness of the shipment (if required) has been provided to the trans-
porter / der Spediteur wurde (falls vorgeschrieben) über die Gefährlichkeit der Sendung informiert.
We hereby commit ourselves and guarantee that we will indemnify BINDER GmbH for all damages that
are a consequence of incomplete or incorrect information provided by us, and that we will exempt
BINDER GmbH from eventual damage claims by third parties./ Wir versichern, dass wir gegenüber BINDER für
jeden Schaden, der durch unvollständige und unrichtige Angaben entsteht, haften und BINDER gegen eventuell entstehende
Schadenansprüche Dritter freistellen.
We are aware that, in accordance with Article 823 of the German Civil Code (BGB), we are directly liable
with regard to third parties, in this instance especially the employees of BINDER GmbH, who have been
entrusted with the handling / repair of the unit / component. / Es ist uns bekannt, dass wir gegenüber Dritten – hier
insbesondere mit der Handhabung/Reparatur des Geräts/des Bauteils betraute Mitarbeiter der Firma BINDER - gemäß §823 BGB
direkt haften
Name:
_______________________________________________________________________
Position/Title: _______________________________________________________________________
Date / Datum: _______________________________________________________________________
Signature / Unterschrift: _______________________________________________________________
Company stamp / Firmenstempel:
Equipment that is returned to the factory for repair must be accompanied by a completely filled
out contamination clearance certificate. For service and maintenance on site, such a contami-
nation clearance certificate must be submitted to the service technician before the start of any
work. No repair or maintenance of the equipment is possible, without a properly filled out con-
tamination clearance certificate.