Text of the manual

Art. No. 7001-0127

Operating Manual Translation of the Original Operating Manual APT.line™ FDL 115 Safety / paint drying ovens for drying of limited quantities of solvents with microprocessor program controller RD3 Model Art. No. FDL 115 (E2) 9010-0269, 9110-0269 BINDER GmbH Address Post office box 102 D-78502 Tuttlingen Tel. +49 7462 2005 0 Fax +49 7462 2005 100 Internet http://www.binder-world.com E-mail [email protected] Service Hotline +49 7462 2005 555 Service Fax +49 7462 2005 93 555 Service E-Mail [email protected] Service Hotline USA +1 866 885 9794 or +1 631 224 4340 Service Hotline Spain +34 9492 677 23 Service Hotline Asia Pacific +852 39070500 or +852 39070503 Service Hotline Russia and CIS +7 495 98815 17

FDL (E2) 03/2011

page 2/68 EG - KONFORMITÄTSERKLÄRUNG EC - DECLARATION OF CONFORMITY CE - DECLARATION DE CONFORMITE Anbieter / Supplier / Fournisseur: BINDER GmbH Anschrift / Address / Adresse: Im Mittleren Ösch 5, D-78532 Tuttlingen Bezeichnung der Maschine / Denomination of the machine / Dénomination de la machine: Sicherheits- / Lacktrockenschrank Safety / paint drying oven Armoire séchante de sécurité / pour peintures et vernis Typenbezeichnung / Type / Type: FDL 115 Die oben beschriebene Maschine ist konform mit folgenden EG-Richtlinien (gemäß Veröf- fentlichung im Amtsblatt der europäischen Kommission): The machine described above is in conformity with the following EC guidelines (as published in the Official Journal of the European Union): La machine décrite ci-dessus est conforme aux directives CE suivantes (selon leur publication dans le Journal officiel de l’Union européenne): Maschinenrichtlinie 2006/42/EG Machinery directive 2006/42/EC Directive Machines 2006/42/EC Richtlinie 2006/42/EG des Europäischen Parlaments und des Rates vom 17. Mai 2006 über Maschinen und zur Änderung der Richtlinie 95/16/EG (Neufassung) Directive 2006/42/EC of the European Parliament and of the Council of 17 May 2006 on machinery, and amending Directive 95/16/EC (recast) Directive 2006/42/CE du Parlement Européen et du Conseil du 17 mai 2006 relative aux machines et modifiant la directive 95/16/CE (refonte) EMV-Richtlinie 2004/108/EG EMC Directive 2004/108/EC Directive CEM 2004/108/CE Richtlinie 2004/108/EG des Europäischen Parlaments und des Rates vom 15. Dezember 2004 zur Angleichung der Rechtsvor- schriften der Mitgliedstaaten über die elektromagnetische Ver- träglichkeit und zur Aufhebung der Richtlinie 89/336/EWG. Directive 2004/108/EC of the European Parliament and of the Coun- cil of 15 December 2004 on the approximation of the laws of the Member States relating to electromagnetic compatibility and repeal- ing Directive 98/336/EEC. Directive 2004/108/CE du Parlement Européen et du Conseil du 15 décembre 2004 relative au rapprochement des législations des États membres concernant la compatibilité électromagnétique et abro- geant le directive 98/336/CEE. ATEX-Richtlinie 94/9/EG ATEX Directive 94/9/EC Directive ATEX 94/9/CE Richtlinie 94/9/EG des Europäischen Parlaments und des Rates zur Angleichung der Rechtsvorschriften der Mitgliedstaaten für Geräte und Schutzsysteme zur bestimmungsgemäßen Verwendung in ex- plosionsgefährdeten Bereichen. Directive 94/9/EC of the European parliament and the council of 23 March 1994 on the approximation of the laws of the Member States concerning equipment and protective systems intended for use in potentially explosive atmospheres Directive 94/9/CE du parlement européen et du conseil du 23 mars 1994 concernant le rapprochement des législations des États mem- bres pour les appareils et les systèmes de protection destinés à être utilisés en atmosphères explosibles. 1 / 3

FDL (E2) 03/2011

page 3/68 Die oben beschriebene Maschine entspricht aufgrund ihrer Konzipierung und Bauart sowie in der von uns in Verkehr gebrachten Ausführung den einschlägigen grundlegenden Sicherheits- und Gesundheitsan- forderungen der genannten EG-Richtlinien. The machine described above is conform to the mentioned EC directives in regard to the relevant safety and health demands due to its conception and its style of construction as well as to the version put onto market by us. La machine décrite ci-dessus correspond aux demandes de sécurité et de santé des directives citées de la Communauté Européenne due à sa conception et construction et dans la réalisation mise sur le marché par nous. Die oben beschriebene Maschine trägt entsprechend die Kennzeichnung CE. The machine described above, corresponding to this, bears the CE-mark. La machine décrite ci-dessus, en correspondance, porte l’indication CE. Die oben beschriebene Maschine ist konform mit folgenden harmonisierten Normen: The machine described above is in conformity with the following harmonized standards: La machine décrite ci-dessus est conforme aux normes harmonisées suivantes: Sicherheit / safety / sécurité: EN 61010-1:2010 Sicherheitsbestimmungen für elektrische Mess-, Steuer-, Regel- und Laborgeräte – Teil 1: Allgemeine Anforderungen (DIN EN 61010- 1:2002 + Berichtigung 1:2002 + Berichtigung 2:2004) Safety requirements for electrical equipment for measurement, con- trol, and laboratory use – Part 1: General requirements (IEC 61010- 1:2010, BS EN 61010-1:2010) Règles de sécurité pour appareils électriques de mesurage, de régu- lation et de laboratoire – Partie 1: Prescriptions générales (CEI 61010- 1:2010, NF EN 61010:2011) EN 61010-2-010:2003 Sicherheitsbestimmungen für elektrische Meß-, Steuer-, Regel- und Laborgeräte – Teil 2-010: Besondere Anforderungen an Laborgeräte für das Erhitzen von Stoffen (DIN EN 61010-2-010:2004) Safety requirements for electrical equipment for measurement, con- trol, and laboratory use – Part 2-010: Particular requirements for labo- ratory equipment for the heating of materials (IEC 61010-2-10:2005, BS EN 61010-2-10:2003) Règles de sécurité pour appareils électriques de mesurage, de régu- lation et de laboratoire – Partie 2-010 : Prescriptions particulières pour appareils de laboratoire utilisés pour l’échauffement des matières (CEI 61010-2-10:2003, NF EN 61010-2-10:2005) EN ISO 12100:2010 Sicherheit von Maschinen - Allgemeine Gestaltungsleitsätze - Risi- kobeurteilung und Risikominderung (DIN EN ISO 12100:2011) Safety of machinery - General principles for design - Risk assessment and risk reduction (BS EN ISO 12100:2010) Sécurité des machines - Principes généraux de conception - Appréciation du risque et réduction du risque (NF EN ISO 12100:2010) EN ISO 13732-1:2008 Ergonomie der thermischen Umgebung - Bewertungsverfahren für menschliche Reaktionen bei Kontakt mit Oberflächen. Teil 1: Heiße Oberflächen (DIN EN ISO 13732-1:2008) Ergonomics of the thermal environment - Methods for the assessment of human responses to contact with surfaces. Part 1: Hot surfaces (BS EN ISO 13732-1: 2008) Ergonomie des ambiances thermiques - Méthodes d'évaluation de la réponse humaine au contact avec des surfaces. Partie 1: Surfaces chaudes (NF EN ISO 13732-1: 2008) 2 / 3

FDL (E2) 03/2011

page 4/68 Sicherheit / safety / sécurité: EN 60204-1:2006 + A1:2009 Sicherheit von Maschinen. Elektrische Ausrüstung von Maschinen. Teil 1: Allgemeine Anforderungen (DIN EN 60204-1:2007 + A1:2009) Safety of machinery. Electrical equipment of machines. General re- quirements (IEC 60204-1:2005 mod. + A1:2008, BS EN 60204-1:2006 + A1:2009) Sécurité des machines - Équipement électrique des machines - Partie 1 : règles générales (CEI 60204-1:2005 mod. + A1:2008, NF EN 60204-1:2006 + A1:2009) EN 1539:2009 Trockner und Öfen, in denen brennbare Stoffe freigesetzt werden. Sicherheitsanforderungen (DIN EN 1539: 2010) Dryers and ovens, in which flammable substances are released. Safety requirements (BS EN 1539:2009) Séchoirs et fours dans lesquels se dégagent des substances inflam- mables. Prescriptions de sécurité (NF EN 1539 : 2010) EMV / EMC / CEM: EN 61326-1:2006 + Corr. 1:2008 + Corr. 2:2010 Elektrische Mess-, Steuer-, Regel- und Laborgeräte - EMV- Anforderungen - Teil 1: Allgemeine Anforderungen (DIN EN 61326- 1:2006 + Berichtigung 1:2008 + Berichtigung 2:2011) Electrical equipment for measurement, control and laboratory use - EMC requirements - Part 1: General requirements (IEC 61326-1:2005 + Corr. 1:2008 + Corr. 2:2010, BS EN 61326-1:2006+ A1:2008) Matériel électrique de mesure, de commande et de laboratoire - Exi- gences relatives à la CEM - Partie 1: Exigences générales (CEI 61326-1:2005 + AC1:2008, NF EN 61326-1:2006 mod.) EN 61326-2-2:2006 Elektrische Mess-, Steuer-, Regel- und Laborgeräte – EMV- Anforderungen. Teil 2-2: Besondere Anforderungen - Prüfanordnung, Betriebsbedingungen und Leistungsmerkmale für ortsveränderliche Prüf-, Mess- und Überwachungsgeräte in Niederspannungs- Stromversorgungsnetzen. (DIN EN 61326-2-2:2006) Electrical equipment for measurement, control and laboratory use – EMC requirements. Part 2-2: Particular requirements - Test configura- tions, operational conditions and performance criteria for portable test, measuring and monitoring equipment used in low-voltage distribution systems. (IEC 61326-2-2:2005, BS EN 61326-2-2:2006) Matériel électrique de mesure, de commande et de laboratoire – Exi- gences relatives à la CEM. Partie 2-2: Exigences particulières - Con- figurations d’essai, conditions de fonctionnement et critères d’aptitude à la fonction des matériels portatifs d’essai, de mesure et de surveil- lance utilisés dans des systèmes de distribution basse tension. (CEI 61326-2-2:2005 + AC1:2007, NF EN 61326-2-2:2006) D-78532 Tuttlingen, 30.03.2011 BINDER GmbH P. M. Binder Geschäftsführender Gesellschafter Managing Director Directeur général Dr. H. von Both Leiter F & E und Dokumentationsbevollmächtigter Director R & D and documentation representative Chef de service R&D et autorisé de documentation 3 / 3

FDL (E2) 03/2011

page 5/68 Certificate for the GS mark of conformity of the “Deutsche Gesetzliche Unfallversicherung e.V.” (German Social Accident Insurance (DGUV)

FDL (E2) 03/2011

page 7/68 Contents 1. SAFETY..................................................................................................................9 1.1 Legal considerations ...........................................................................................................................9 1.2 Structure of the safety instructions......................................................................................................9 1.2.1 Signal word panel......................................................................................................................9 1.2.2 Safety alert symbol..................................................................................................................10 1.2.3 Pictograms...............................................................................................................................10 1.2.4 Word message panel structure ...............................................................................................11 1.3 Localization / position of safety labels on the unit.............................................................................11 1.4 Type plate .........................................................................................................................................12 1.5 General safety instructions on installing and operating the safety / paint drying oven FDL.............13 1.6 Intended use .....................................................................................................................................14 1.7 Operating instructions .......................................................................................................................14 1.8 Measures to prevent accidents.........................................................................................................15 1.9 Important points to consider before commissioning .........................................................................16 1.9.1 Technical ventilation / permissible load...................................................................................16 1.9.2 Loading instructions ................................................................................................................16 1.9.3 Drying nitro-cellulose lacquers ................................................................................................16 1.9.4 Drying mould varnishes...........................................................................................................16 1.9.5 Drying impregnating resins......................................................................................................17 2. UNIT DESCRIPTION............................................................................................ 17 2.1 Unit overview.....................................................................................................................................18 2.2 Control panel.....................................................................................................................................19 2.3 Solvent curve FDL 115 .....................................................................................................................19 3. COMPLETENESS OF DELIVERY, TRANSPORTATION, STORAGE, AND INSTALLATION....................................................................................................20 3.1 Unpacking, and checking equipment and completeness of delivery................................................20 3.2 Guidelines for safe lifting and transportation ....................................................................................21 3.3 Storage..............................................................................................................................................21 3.4 Location of installation and ambient conditions ................................................................................21 4. INSTALLATION OF THE EQUIPMENT ............................................................... 22 4.1 Operating instructions .......................................................................................................................22 4.2 Electrical connection .........................................................................................................................23 4.3 Connection to a suction plant (optional) ...........................................................................................23 5. START UP............................................................................................................ 24 5.1 Settings at the RD3 program controller ............................................................................................24 5.2 General indications ...........................................................................................................................25 6. FIXED VALUE ENTRY MODE ............................................................................. 27 7. WEEK PROGRAM EDITOR.................................................................................28 7.1 Program table template for Week program Editor ............................................................................30 8. PROGRAM EDITOR.............................................................................................31 8.1 Selecting between set-point ramp and set-point step.......................................................................31 8.1.1 Programming with setting “Ramp” (default setting).................................................................31 8.1.2 Programming with setting “step” .............................................................................................33 8.1.3 General notes on programming temperature transitions ........................................................34 8.2 Set-point entry for program operation...............................................................................................34 8.3 Program table template.....................................................................................................................37 8.4 Deleting a program section...............................................................................................................38

FDL (E2) 03/2011

page 8/68 9. PROGRAM START LEVEL.................................................................................. 39 10. USER LEVEL .......................................................................................................41 11. BEHAVIOR IN CASE OF FAILURES................................................................... 48 11.1 Behavior after power failure..............................................................................................................48 11.2 Alarm messages ...............................................................................................................................48 12. SAFETY DEVICES............................................................................................... 48 12.1 Temperature safety device class 2 (DIN 12880:2007) .....................................................................48 12.2 Exhaust air monitoring ......................................................................................................................50 13. OPTIONS..............................................................................................................50 13.1 Communication software APT-COM™ 3 DataControlSystem (option) ............................................50 13.1 Ethernet interface (option) ................................................................................................................50 13.2 Protocol printer (option) ....................................................................................................................51 13.3 Additional measuring channel for digital object temperature indicator with magnetic sensor or pincer-type sensor (option) ...............................................................................................................52 14. MAINTENANCE, CLEANING, AND SERVICE .................................................... 53 14.1 Maintenance intervals, service..........................................................................................................53 14.2 Replacing the intake filter..................................................................................................................54 14.3 Cleaning and decontaminating the safety / paint drying oven..........................................................54 14.4 Sending the unit back to BINDER GmbH .........................................................................................55 15. DISPOSAL............................................................................................................56 15.1 Disposal of the transport packing .....................................................................................................56 15.2 Decommissioning..............................................................................................................................56 15.3 Disposal of the unit in the Federal Republic of Germany .................................................................56 15.4 Disposal of the unit in the member states of the EC except for the Federal Republic of Germany .57 15.5 Disposal of the unit in non-member states of the EC .......................................................................59 16. TROUBLESHOOTING ......................................................................................... 59 17. TECHNICAL DESCRIPTION................................................................................ 61 17.1 Factory calibration and adjustment...................................................................................................61 17.2 Definition of usable volume...............................................................................................................61 17.3 Technical data...................................................................................................................................61 17.4 Equipment and Options ....................................................................................................................63 17.5 Spare parts........................................................................................................................................64 17.6 Dimensions FDL 115 ........................................................................................................................65 18. CONTAMINATION CLEARANCE CERTIFICATE ............................................... 66

FDL (E2) 03/2011

page 9/68 Dear customer, For the correct operation of the safety / paint drying ovens FDL, it is important that you read this operating manual completely and carefully and observe all instructions as indicated. Failure to read, understand and follow the instructions may result in personal injury. It can also lead to damage to the unit and/or poor equipment performance. 1. Safety This operating manual is part of the components of delivery. Always keep it handy for reference. The device should only be operated by laboratory personnel especially trained for this purpose and familiar with all precautionary measures required for working in a laboratory. To avoid injuries and damage ob- serve the safety instructions of the operating manual. WARNING Failure to observe the safety instructions. Serious injuries and unit damage. � Observe the safety instructions in this operating manual. � Carefully read the complete operating instructions of the safety / paint drying oven FDL. 1.1 Legal considerations This operating manual is for informational purposes only. It contains information for installing, start-up, operation and maintenance of the product. Note: the contents and the product described are subject to change without notice. Understanding and observing the instructions in this operating manual are prerequisites for hazard-free use and safety during operation and maintenance. In no event shall BINDER be held liable for any dam- ages, direct or incidental arising out of or related to the use of this manual. This operating manual cannot cover all conceivable applications. If you would like additional information, or if special problems arise that are not sufficiently addressed in this manual, please ask your dealer or contact us directly by phone at the number located on page one of this manual Furthermore, we emphasize that the contents of this operating manual are not part of an earlier or exist- ing agreement, description, or legal relationship, nor do they modify such a relationship. All obligations on the part of BINDER derive from the respective purchase contract, which also contains the entire and ex- clusively valid statement of warranty administration. The statements in this manual neither augment nor restrict the contractual warranty provisions. 1.2 Structure of the safety instructions In this operating manual, the following safety definitions and symbols indicate dangerous situations in accordance with ISO 3864-2 and ANSI Z535.6. 1.2.1 Signal word panel Depending on the probability of serious consequences, potential dangers are identified with a signal word, the corresponding safety color, and if appropriate, the safety alert symbol. DANGER Indicates an imminently hazardous situation that, if not avoided, will result in death or serious (irreversi- ble) injury. WARNING Indicates a potentially hazardous situation which, if not avoided, could result in death or serious (irre- versible) injury.

FDL (E2) 03/2011

page 10/68 CAUTION Indicates a potentially hazardous situation which, if not avoided, may result in moderate or minor (re- versible) injury. CAUTION Indicates a potentially hazardous situation which, if not avoided, may result in damage to the product and/or its functions or of a property in its proximity. 1.2.2 Safety alert symbol Use of the safety alert symbol indicates a risk of injury. Observe all measures that are marked with the safety alert symbol in order to avoid death or injury. 1.2.3 Pictograms Warning signs Electrical hazard Hot surface Explosive atmosphere Tipover hazard Inhalation hazard Pollution Hazard Harmful substances Biohazard Mandatory action signs Mandatory regulation Read operating instructions Disconnect the power plug Lift with several persons Environment protection Prohibition signs Do NOT touch Do NOT spray with water Information to be observed in order to ensure optimum function of the product.

FDL (E2) 03/2011

page 11/68 1.2.4 Word message panel structure Type / cause of hazard. Possible consequences. ∅ Instruction how to avoid the hazard: prohibition � Instruction how to avoid the hazard: mandatory action Observe all other notes and information not necessarily emphasized in the same way, in order to avoid disruptions that could result in direct or indirect injury or property damage. 1.3 Localization / position of safety labels on the unit The following labels are located on the unit: Pictograms (Warning signs) Service label Hot surface Figure 1: Position of labels on the unit Keep safety labels complete and legible. Replace safety labels that are no longer legible. Contact BINDER Service for these replacements.

FDL (E2) 03/2011

page 12/68 1.4 Type plate Figure 2: Position of type plate Figure 3: Type plate (example of FDL 115 regular unit) Indications of the type plate Information BINDER Manufacturer: BINDER GmbH FDL 115 Model FDL 115 Safety Drying Oven Safety / paint drying oven Serial No. 00-00000 Serial No. 00-00000 Nominal temperature 300 °C 572°F Nominal temperature Enclosure protection IP 33 IP type of protection 33 acc. to EN 60529 Temp. safety device DIN 12880 Temperature safety device acc. to standard DIN 12880 Class 2.0 Temperature safety device, class 2 Art. No. 9010-0269 Art. No. 9010-0269 Project No. --- (Special application acc. to project no.) Built 0000 Year of construction: 0000 2,90 kW Nominal power 2.90 kW 230 V 1 N PE ~ Nominal voltage 230 V ± 10%, single-phase unit 13,0 A Nominal current 13.0 Amp 50/60 Hz Power frequency 50/60 Hz Usable volume 0,115m³ Usable volume 0.115 m³ Steam space 0,156m³ Total steam space 0.156 m³ Max. solvent at nominal temp. 3,0g Highest permissible solvent quantity at 300 °C / 572°F: 3.0g Min. exhaust flow rate 24m³/h at +20 °C Minimum exhaust flow rate at +20 °C: 24 m³/h Max. temp. of heating surfaces +750 °C Maximum temperature of heating surfaces +750 °C / 1382°F Wiring diagram FDL 115 pages 1-3 Wiring diagram FDL 115 pages 1 to 3 Nominal temperature 300 °C 2,90 kW Usable volume 0,115m³ 572°F 230 V 1 N PE ~ Steam space 0,156m³ Enclosure protection Temp. safety device IP 33 DIN 12880 13,0 A 50/60 Hz Max. solvent at nominal temp. 3,0g Min. exhaust flow rate 24m³/h at +20 °C Class 2.0 Max. temp. of heating surfaces +750 °C Art. No. 9010-0269 US PATS 4585923 / 5222612 / 5309981 Wiring diagram FDL 115 pages 1-3 Project No. Built 0000 5405194 / 5601143 / 5773287 / 6079403 Safety Drying Oven D 78532 Tuttlingen / Germany Tel. + 49 (0) 7462/ 2005-0 Internet: www.binder-world.com FDL 115 Serial No. 00-00000 Made in Germany

FDL (E2) 03/2011

page 13/68 Symbol on the type plate Information CE conformity marking Electrical and electronic equipment manufactured / placed on the market in the EC after 13 August 2005 and to be disposed of in a separate collection according to directive 2002/96/EC on waste elec- trical and electronic equipment (WEEE). The equipment is certified in the GOST R certification system of GOSTSTANDARD Russia. GS mark of conformity of the “Deutsche Gesetzliche Unfallversi- cherung e.V. (DGUV), Fachausschuss Metall- und Oberflächenbe- handlung, Prüf- und Zertifizierungsstelle im DGUV Test“ (German Social Accident Insurance (DGUV), Expert Committee: Metal and Surface Treatment, Testing and Certification Body in DGUV) 1.5 General safety instructions on installing and operating the safety / paint drying oven FDL With regard to operating the safety / paint drying oven FDL and to the installation location, please ob- serve the guideline BGI/GUV-I 850-0 on safe working in laboratories (formerly BGR/GUV-R 120 or ZH 1/119 laboratory guidelines issued by the employers’ liability insurance association) (for Germany). BINDER GmbH is only responsible for the safety features of the unit provided skilled electricians or quali- fied personnel authorized by BINDER perform all maintenance and repair, and if components relating to chamber safety are replaced in the event of failure with original spare parts. To operate the unit, use only original BINDER accessories or accessories from third-party suppliers au- thorized by BINDER. The user is responsible for any risk caused by using unauthorized accessories. CAUTION Danger of overheating. Damage to the unit. ∅ Do NOT install the unit in unventilated recesses. � Ensure sufficient ventilation for dispersal of the heat. Do not operate the safety / paint drying oven FDL in hazardous locations. DANGER Explosion hazard. Danger of death. ∅ Do NOT operate the unit in potentially explosive areas. ∅ KEEP explosive dust or air-solvent mixtures AWAY from the unit. Familiarize yourself with any potential health risks caused by the charging material, the contained mois- ture constituent or by reaction products which may arise during the temperature process. Take adequate measures to exclude such risks prior to putting the safety / paint drying oven into operation.

FDL (E2) 03/2011

page 14/68 DANGER Electrical hazard. Danger of death. ∅ The unit must NOT become wet during operation or maintenance. The safety / paint drying ovens were produced in accordance with VDE regulations and were routinely tested in accordance to VDE 0411-1 (IEC 61010-1). If a warning signal which indicates an of error condition, no further charging material must be introduced into the unit. DANGER Explosion and ignition hazard. Danger of death. ∅ Do NOT introduce any further charging material as long as there is a warning signal. During and after the drying process, the inner surfaces have got a temperature close to the set-point. CAUTION The inner chamber, the outgoing air pipe, the door window (option), the door gasket, and the access ports will become hot during operation. Danger of burning. ∅ Do NOT touch the inner surfaces, the outgoing air pipe, the door window, the door gas- ket, the access ports, or the charging material during operation. 1.6 Intended use The BINDER safety / paint drying oven FDL 115 is suitable for drying and burn in of lacquers and similar liquid coating materials which are containing solvents that can form explosive mixtures with air. The maximally permitted drying temperature and the maximally permitted quantity of solvent are limited, see chap. 1.9. Other applications are not approved. Do NOT use the unit to warm up coating materials in containers, vessels, etc. or for drying textiles soaked in solvent. Do NOT use the unit for drying purpose if greater quantities of steam or solvent gas leading to condensa- tion will be set free. Following the instructions in this operating manual and conducting regular maintenance work (chap. 14.1) are part of the intended use. 1.7 Operating instructions Depending on the application and location of the unit, the operator of the safety / paint drying oven must provide the relevant information for safe operation of the unit in a set of operating instructions. These operating instructions must be kept with the unit at all times in a place where they are clearly visible. They must be comprehensible and written in the language of the employees.

FDL (E2) 03/2011

page 15/68 1.8 Measures to prevent accidents During drying of fluid paints, flammable solvent/air mixtures can form and ignite. The manufacturer took the following measures to prevent ignition and explosions: • Indications on the type plate See operating manual chap. 1.4 • Operating manual An operating manual is available for each safety / paint drying oven. A diagram in the operating manual (chap. 2.3) indicates the maximum permissible solvent quantities for various operating states. The operating manual asks the operator of the safety / paint drying oven to set up an instruction on the permissible loading quantity. • Maximum temperatures and maximum permissible steam concentration By means of the diagram “Highest permissible solvent quantity”, which is included in the operating manual chap. 2.3 and applied on the front of the safety / paint drying oven, the operator must adapt the drying temperature to the maximum solvent quantity arising. When using nitro-cellulose lacquers or nitro combination lacquers, the diagram ”Highest permissible solvent quantity” specifies a limit temperature of max. 130 °C/ 266 °F, which must not be exceeded. (In this context, all paints containing more than 5 % nitro-cellulose relative to the non-volatile contents are regarded as nitro-cellulose lacquers and nitro-combination lacquers). A gas-tight separation between the drying chamber and the heating chamber is not necessary since there is an effective forced air motion in the entire steam room. • Throttle valves No throttle valves are used, i.e., the full air change occurs permanently. • Protecting the heating surfaces against dripping The BINDER APT.line™ preheating chamber protects all heating elements against lacquer dripping and direct contact with lacquer coatings. • Heat insulation All heat insulation is sealed against penetration by lacquer vapors from the outside with high tempera- ture-resistant and ageing-resistant sealant. The insulation material consists of non-combustible mineral wool (class A1 according to DIN 4102- 1:1998). • Temperature monitoring The safety / paint drying oven is equipped with a temperature display, which can be read from outside. A built-in additional temperature safety device can turn off the heating and is functionally independent of the main controller. When turning off the unit, the forced convection and the control equipment’s function are maintained. Visual (red indicator light) and audible (buzzer) signals indicate temperature exceeding. • Safety in case the technical ventilation fails The heating only turns on if the air circulation is already in operation. If the air circulation fails, the heating immediately turns off. In addition, there are audible (buzzer) and visual (red indicator light) signals. • Moving parts participating in the work process It is impossible to touch the fan from outside the device or from the interior. • Safety, measurement and control devices The safety, measuring, and control equipment is easily accessible via the housing top cover.

FDL (E2) 03/2011

page 16/68 • Electrostatic charge The interior parts are grounded. • Protection against touchable surfaces Tested according to EN ISO 13732-1:2006. • Floors See operating manual chap. 3.4 for the oven’s installation. • Ventilation Ventilation shall be realized by the operator according to GUV-R 500 chap. 2.29 “Verarbeiten von Beschichtungsstoffen” (Processing of paints) (for Germany). • Cleaning See operating manual chap. 14. • Examinations The safety / paint drying oven has been inspected by the “Deutsche Gesetzliche Unfallversicherung e.V. (DGUV), Fachausschuss Nahrungs- und Genussmittel, Prüf- und Zertifizierungsstelle im DGUV Test“ (German Social Accident Insurance (DGUV), Expert Committee: “Metal and Surface Treatment”, Testing and Certification Body in DGUV) and bears the GS mark. 1.9 Important points to consider before commissioning 1.9.1 Technical ventilation / permissible load For safety reasons, the formation of a dangerous, explosive atmosphere must be avoided in all operating modes (see GUV-R 500 chap. 2.28 “Betreiben von Trocknern für Beschichtungsstoffe” ("Dryers for coat- ing materials")). Keeping the maximum permissible amount of solvent when loading the unit will meet this requirement. The permissible quantity can be calculated according to the "principles for the calculation of ventilation of chamber dryers and continuous dryers" (EN 1539:2009, Appendix B). According to this standard, the safety / paint drying oven’s technical data (chap. 17.3) must be included in the calculation, and it is required to set up a set of loading instructions (for Germany). 1.9.2 Loading instructions The loading instructions should specify the amount of material, which can be loaded into the drying oven for treatment without any risk of creating a dangerous, explosive atmosphere. GUV-R 500 chap. 2.28 “Betreiben von Trocknern für Beschichtungsstoffe” ("Dryers for coating materials") explicitly states that the operator must set up a set of loading instructions (for Germany). 1.9.3 Drying nitro-cellulose lacquers When using the safety / paint drying oven FDL to dry material, which is coated with nitro-cellulose lac- quers, the temperature safety device must be set to max. 130 °C / 266 °F, so that the surface tempera- ture of the goods to be dried is ensured not to exceed 130 °C / 266 °F. Deviations are only permissible if a report from a testing agency, which is recognized by the employer's liability insurance association, has declared safe a higher surface temperature. 1.9.4 Drying mould varnishes When using the safety / paint drying oven FDL to dry mould varnishes, the operator is allowed to increase the highest permissible amount of solvent specified for surface drying (chap. 2.3) by up to 10 times (see GUV-R 500 chap. 2.28 “Betreiben von Trocknern für Beschichtungsstoffe” ("Dryers for coating materials") or EN 1539:2009, Appendix A.1.1.2).

FDL (E2) 03/2011

page 17/68 1.9.5 Drying impregnating resins When using the safety / paint drying oven FDL to dry impregnating resins, the operator is allowed to in- crease the highest permissible amount of solvent specified for surface drying (chap. 2.3) by up to 20 times (see GUV-R 500 chap. 2.28 „Betreiben von Trocknern für Beschichtungsstoffe“ ("Dryers for coating materials") no. 3.7.4 or EN 1539:2009, Appendix A.1.1.2). 2. Unit description The safety / paint drying oven FDL 115 was built according to EN 1539:2009 (“Dryers and ovens, in which flammable substances are released. Safety requirements”). The fan in the rear of the safety / paint drying oven delivers a constant amount of fresh air through the working area irrespective of the drying temperature. A large-area filter removes dust from the incoming air (permeability up to approx. 1 micrometer). A flow monitor in the upper part of the device (pressure differential switch) monitors the stream of exhaust air. In the event of failure, the monitoring system turns the heater off immediately and registers this status with a visual and an audible signal. The heating of the device is released with a time delay of 4 minutes after you turn on the device by its main power switch. The indicator light (3) in the operating panel only lights up as long as the air-flow monitor has not yet en- abled the heating process. The drying temperature is also monitored constantly by the temperature limiter (2). If the temperature would exceed the maximum permissible temperature, the heating is turned off immediately and this state is indicated visually by indicator light (2a) and acoustically. In the event of fail- ure, the safety / paint drying oven cannot be restarted until the RESET key (2b) or (4) has been reset BINDER safety / paint drying ovens with forced convection FDL are equipped with the electronic program controller RD3 with digital display. This permits programming of temperature cycles. The APT.line™ preheating chamber system ensures high level of spatial and time-based temperature precision, thanks to the direct and distributed air circulation into the interior. The air turbine supports ex- act attainment and maintenance of the required temperature accuracy. All functions of the multifunctional program control can be set simply and conveniently via the easy to understand function keypad of the RD3 temperature program controller. This controller is equipped with touch function keys and a digital display and permits exact temperature setting and programming tem- perature cycles. The FDL provides almost unlimited possibilities of adapting to individual customer re- quirements based upon extensive programming options and on the week program timer and real time clock of the controller. All unit functions are easy and comfortable to use thanks to their clear arrangement. Major features are easy cleaning of all unit parts and avoidance of undesired contamination. The inner chamber, the pre-heating chamber and the interior side of the doors are all made of stainless steel (material no. 1.4301 (V2A) in Germany, US equivalent 304). When operating the chamber at tem- peratures above 150 °C, the impact of the oxygen in the air may cause discoloration of the metallic sur- faces (yellowish-brown or blue) by natural oxidation processes. These colorations are harmless and will in no way impair the function or quality of the unit. The housing is RAL 7035 powder-coated. All corners and edges are also completely coated. The safety / paint drying oven FDL comes equipped with a serial interface RS 422 for computer commu- nication, e.g. via the communication software APT-COM™ 3 DataControlSystem (option, chap. 13.1). For further options, see chap. 17.4. The units can be operated at an ambient temperature of 18 °C / 64.4 °F up to 40 °C / 104 °F in a tem- perature control range by 5 °C above ambient temperature up to +300 °C / 572 °F.

FDL (E2) 03/2011

page 18/68 2.1 Unit overview Figure 4: Front view FDL 115 with optional window (A) Control panel (B) Microprocessor program controller RD3 (C) Temperature safety device class 2 according to DIN 12880:2007 (D) Main power switch ON/OFF (E) Solvent curve (F) Door handle (G) Unit door (H) Window (optional) RES AI (E) (B) (C) (D) (F) (A) (G) (H)

FDL (E2) 03/2011

page 19/68 2.2 Control panel (6) (5) (4) (3) (2) (2b) (2a) (1) Figure 5: Control panel of standard unit (1) Main power switch ON/OFF (2) Temperature safety device class 2 (2a) Red pilot lamp for temperature safety device class 2 (2b) RESET key for temperature safety device (3) Red indicator light, loss of technical ventilation (4) RESET key for the heating cut-out due to failure of technical ventilation (5) Temperature program controller RD3 (6) Solvent curve: Highest permissible amount of solvent Gtotal [g] as a function of the drying tempera- ture 2.3 Solvent curve FDL 115 0,00 5,00 10,00 15,00 20,00 25,00 30,00 35,00 50 60 70 80 90 100 110 120 130 140 150 160 170 180 190 200 210 220 230 240 250 260 270 280 290 300 Temperature [°C] Gtotal [g] Figure 6: Solvent curve FDL 115 The diagram shows the highest permissible solvent quantity Gtotal [g] in the steam room in correspon- dence to the drying temperature. This is based on the calculation acc. to EN 1539:2009 considering the unit specific data, an assumed molecular weight of the solvent of 100g/Mol, and a lower explosion limit of 40g/m3 at 20 °C / 68 °F and at 760 Torr (1013 hPa) (assumptions for unknown solvents acc. to EN 1539:2009). RESET AIR

FDL (E2) 03/2011

page 20/68 In case of too high drying temperature and /or too high quantity of solvent in the steam room it could be possible that the concentration of the solvent steam will lead to an explosion. The permitted maximum of solvent brought into the drying oven and the maximal temperature must not be exceeded. DANGER Excessive drying temperature and/or excessive solvent quantity. Danger of explosion. Danger of death. ∅ Do NOT exceed the maximum solvent quantity. ∅ Do NOT exceed the maximum drying temperature for the solvent quantity. Adjust the temperature safety device according to the selected set-point (chap. 12.1). 3. Completeness of delivery, transportation, storage, and installa- tion 3.1 Unpacking, and checking equipment and completeness of delivery After unpacking, please check the unit and its optional accessories, if any, based on the delivery receipt for completeness and for transportation damage. Inform the carrier immediately if transportation damage has occurred. The final tests of the manufacturer may have caused traces of the shelves on the inner surfaces. This has no impact on the function and performance of the unit. Please remove any transportation protection devices and adhesives in/on the unit and on the doors and remove the operating manuals and accessory equipment. CAUTION Sliding or tilting of the unit. Damage to the unit. ∅ Do NOT lift or transport the unit using the door or the handle. � Lift the unit from the pallet at the four lower corners with the aid of four people. If you need to return the unit, please use the original packing and observe the guidelines for safe lifting and transportation (chap. 3.2). For disposal of the transport packing, see chap. 15.1. Note on second-hand units (Ex-Demo-Units): Second-hand units are units that were used for a short time for tests or exhibitions. They are thoroughly tested before resale. BINDER ensures that the chamber is technically sound and will work flawlessly. Second-hand units are marked with a sticker on the unit door. Please remove the sticker before commis- sioning the unit.

FDL (E2) 03/2011

page 21/68 3.2 Guidelines for safe lifting and transportation After operation, please observe the guidelines for temporarily decommissioning the unit (chap. 15.2). CAUTION Sliding or tilting of the unit. Damage to the unit. ∅ Do NOT lift or transport the unit using the door or the handle. � Transport the unit in its original packaging only. � For moving or shipping, secure the oven with transport straps. � Lift the unit at the four lower corners with the aid of 4 people and place it on a transport pallet with wheels. Push the pallet to the desired site and then lift the unit at its four lo- wer corners from the pallet with the aid of four people. • Permissible ambient temperature range during transport: -10 °C / 14 °F to +60 °C / 140 °F. You can order transport packing and pallets for moving or shipping purposes from BINDER service. 3.3 Storage Intermediate storage of the unit is possible in a closed and dry room. Observe the guidelines for tempo- rary decommissioning (chap. 15.2). • Permissible ambient temperature range during storage: -10 °C / 14 °F to +60 °C / 140 °F. • Permissible ambient humidity: max. 70 % r.H., non-condensing When after storage in a cold location you transfer the unit to its warmer installation site, condensation may form. Before start-up, wait at least one hour until the chamber has attained ambient temperature and is completely dry. 3.4 Location of installation and ambient conditions Set up the safety / paint drying oven FDL on a flat, even and non-flammable surface, free from vibration and in a well-ventilated, dry location and align it using a spirit level. The site of installation must be capa- ble of supporting the unit’s weight (see technical data, chap. 17.3). The chambers are designed for set- ting up inside a building (indoor use). CAUTION Danger of overheating. Damage to the unit. ∅ Do NOT set up units in non-ventilated recesses. � Ensure sufficient ventilation for dispersal of the heat. • Permissible ambient temperature range during operation: +18 °C / 64.4 °F up to +40 °C / 104 °F. At elevated ambient temperature values, fluctuations in temperature can occur. The ambient temperature should not be substantially higher than the indicated ambient tem- perature of +25 °C / 77°F to which the specified technical data relate. For other ambient condi- tions, deviations from the indicated data are possible. • Permissible ambient humidity: 70 % r.H. max., non-condensing. • Installation height: max. 2000 m / 6.6 ft. above sea level. When placing several units of the same size side by side, maintain a minimum distance of 250 mm / 9.84 in between each unit. Wall distances: rear 100 mm / 3.94 in, sides 160 mm / 6.29 in.

FDL (E2) 03/2011

page 22/68 CAUTION Danger by stacking. Damage to the units. ∅ Do NOT place safety / paint drying ovens on top of each other. To completely separate the unit from the power supply, you must disconnect the power plug. Install the unit in a way that the power plug is easily accessible and can be easily pulled in case of danger. Do not install or operate the oven in potentially explosive areas. DANGER Explosion hazard. Danger of death. ∅ Do NOT operate the unit in potentially explosive areas. ∅ KEEP explosive dust or air-solvent mixtures AWAY from the vicinity of the unit. The standard equipped safety / paint drying oven FDL with housing protection IP 33 acc. to DIN 40050 (see type plate) must NOT be installed and operated in explosive and fire-endangered areas. If the safety / paint drying oven is optionally built up with housing protection IP 54, it can be operated in fire-endangered areas. Decisive is always the indication on the type plate. When operating the safety / paint drying oven FDL with housing protection IP 54 in fire- endangered areas, observe DIN VDE 0100-482:2003 (for Germany). Operation of the cooling slots must on no account be impaired. The vapors produced when the heating the content must be extracted from the safety / paint drying oven through non-combustible exhaust gas or exhaust air ducts. The exhaust duct (nominal diameter 100 mm / 3.9 in) on the rear of the chamber serves for this purpose, to which a suitable exhaust air duct can be connected, e.g., a corrugated alumi- num hose. The exhaust connection must be made via a draught limiter; the exhaust air must not be guided into ducts for combustible gas. CAUTION The rear exhaust duct will become hot during operation. Danger of burning. ∅ Do NOT touch the exhaust duct during operation. 4. Installation of the equipment 4.1 Operating instructions Depending on the application and location of the unit, the operator of the safety / paint drying oven must provide the relevant information for safe operation of the unit in a set of operating instructions. Keep these operating instructions with the unit at all times in a place where they are clearly visible. They must be comprehensible and written in the language of the employees.

FDL (E2) 03/2011

page 23/68 4.2 Electrical connection The safety / paint drying oven FDL 115 is supplied ready for connection. • Power connection: fixed power connection cable 1800 mm / 5.9 ft in length with a shockproof plug • Power supply voltage 230 V (1N~) +/- 10 %, 50/60 Hz • Housing protection type according to EN 60529:2000 : IP 33 (optional IP 54) • Electrical protection: protection class I (with grounding conductor connection) • Prior to connection and start-up, check the power supply voltage. Compare the values to the specified data located on the unit’s type plate (unit front behind the door, bottom left-hand, see chap. 1.4). • When connecting, please observe the regulations specified by the local electricity supply company as well as the VDE directives (for Germany) • Pollution degree (acc. to IEC 61010-1): 2 • Over-voltage category (acc. to IEC 61010-1): II CAUTION Danger of incorrect power supply voltage. Damage to the equipment. � Check the power supply voltage before connection and start-up. � Compare the power supply voltage with the data indicated on the type plate. See also electrical data (chap. 17.3). To completely separate the unit from the power supply, you must disconnect the power plug. Install the unit in a way that the power plug is easily accessible and can be easily pulled in case of danger. 4.3 Connection to a suction plant (optional) When directly connecting a suction plant the spatial temperature exactitude, the heating-up and the re- covering times and the maximum temperature will be negatively influenced. So no suction plant should be directly connected to the exhaust duct (nominal diameter 100 mm / 3.9 in) on the rear of the chamber. Connect a suitable exhaust air duct, e.g., a corrugated aluminum hose, to the chamber exhaust duct. The exhaust connection must be made via a draught limiter; the exhaust air must not be guided into ducts for combustible gas. Active suction from the oven must only be performed together with extraneous air. Perforate the connecting piece to the suction device or place an exhaust funnel at some distance to the outgoing air pipe. CAUTION The exhaust duct will become hot during operation. Danger of burning. ∅ Do NOT touch the exhaust duct during operation.

FDL (E2) 03/2011

page 24/68 5. Start up After connecting the electrical supply (chap. 4.2), turn on the unit via the main power switch (1). Set the main power switch (1) to position “I”. Indicator light (3) and an audible signal indicate that the minimal volumetric flow rate of the technical ventilation has not yet been reached. Once the fan has reached its operating speed, indicator light (3) goes off. Now press RESET key (4). The audible alarm stops. After turning on the main power switch (1), the heating is automatically enabled with a delay time of 4 minutes to ensure the required pre-ventilation acc. to EN1539:2009. Set the temperature set point (chap. 6). Detach the plastic cover over the temperature safety device class 2 (2) with a suitable Phillips screw- driver, and then set the temperature safety device (2) to the maximum permissible drying temperature (chap. 12.1) and restore the plastic cover to prevent misadjusting. After reaching the set drying temperature, the heating keeps it constant by turning on and off regularly. You can verify this on the controller display. When loading the oven with solvent-containing material, do not exceed the maximally permitted solvent quantity for the selected drying temperature. Refer to the solvent curve at the oven front, chap. 2.3. Adjust the temperature safety device according to the selected set-point (chap. 12.1). Warming chambers may release odors in the first few days after commissioning. This is not a quality defect. To reduce odors quickly we recommend heating up the chamber to its nominal temperature for one day and in a well ventilated location. 5.1 Settings at the RD3 program controller After turning the unit on with the main power switch (1) the controller is in Normal Display / fixed value operation mode. Depending on the temperature value entered before LED (3a) is lit if the heating is active, or no LED if the actual temperature is equal to or above the set-point. In Display 1 of the controller the actual temperature Value is shown. • With inactive week program timer: In Display 2 of the controller the actual date and time are displayed. Example: 15.01.07 13:52 • With active week program timer: In Display 2 of the controller the actual date and time and the states of the week program timer chan- nels are displayed. Examples: 15.01.07 13:52 - - 15.01.07 13:52 - � 15.01.07 13:52 � - 15.01.07 13:52 � � Channel 1 Off, Channel 2: Off Channel 1: Off, Channel 2: On Channel 1: On, Channel 2: Off Channel 1 On, Channel 2: On

FDL (E2) 03/2011

page 25/68 Display 1 Display 2 Figure 7: RD3 program controller LED function indications and their signification: (3a) (yellow) Heating active (3b) (yellow) No function (3c) (yellow) No function (3d) (green) Illuminates: program operation Flashes: exceeding of the tolerance limits in “Fixed value entry mode” or in “Program operation”. In “Program operation”: program interruption. The program controller RD3 permits programming of temperature cycles. Two programs with up to 10 sections each or one program with up to 20 sections can be entered (setting in the user level, chap. 10). When changing from 2 programs to 1 program or vice-versa, existing programs are deleted The maximum length of an individual program section can be set to either 99 hs 59 min or to 999 hs 59 min (setting in the user level, chap. 10). This setting is then valid for all program sections. Programming can be done directly via the controller keyboard or graphically at the computer using the software APT-COM™ 3 DataControlSystem (option, chap. 13.1) specially developed by BINDER. 5.2 General indications The program controller RD3 offers several functional levels: Normal Display / fixed value operation: • Display of the actual value of temperature (display 1) and of the actual date and time (display 2). • The chamber is in fixed value operating mode, adjusting to the entered set-points. Fixed value entry mode (chap. 6) • Entry of the temperature set-point for fixed value operating mode • Entry of temperature set-points 1 and 2 for week program operation

FDL (E2) 03/2011

page 26/68 Program editor (chap. 8) • Two programs with up to 10 sections each or one program with up to 20 sections can be entered (se- lection in the user level, chap. 10). Entry of temperature set-points in all program sections (chap. 8.2). • Deleting a program section (chap. 8.4) Program start level (chap. 9) • Selection of an entered program • Entry of settings affecting the program course, as “start delay time” or “number of program cycles” • Program start Week program editor (chap.7) • Setting the shift points User level (chap. 10) • User specific controller settings • Setting the real time clock Normal Display / Fixed value operation 5 seconds Fixed value entry mode Program editor Program start level 5 seconds Week program editor Menu visible only if week program timer is activated in the user level 5 seconds User level If no button is touched within more than 120 sec. the controller returns from the current level to Normal Display.

FDL (E2) 03/2011

page 27/68 6. Fixed value entry mode If you do not want to use the week program timer, deactivate it (factory setting, setting in the user level, chap. 10) before entering any set-points. Basic entry principle: Access the individual parameters with button X/W one after the other. Enter the values with the arrow keys. A value flashing once after 2 seconds indicates that it has been applied by the controller. Normal Display Display 1 shows e.g. 39.8 (actual temperature value) Display 2 shows e.g. 15.01.07 13:52 - - (actual date and time) (actual switching state of week program timer channel 1: Off, channel 2: Off, visible only if week program timer is activated in the user level, chap. 10) Press key Display 1 shows e.g. 40.0 (actual temperature set-point 1) Display 2 shows SP1 TEMPERATURE (variable: temperature in °C) Enter the temperature set-point in °C using arrow keys Value is shown in display 1. Press key Display 1 shows e.g. 90.0 (actual temperature set-point 2) (visible only if week program timer is activated in the user level, chap. 10) Display 2 shows SP2 TEMPERATURE (variable: temperature in °C) Enter temperature set-point in °C using the arrow keys Value is shown in display 1. Press key If no button is pressed within more than 120 sec, or if the EXIT button is pressed, the controller changes to Normal Display. When changing the set-point, check the setting of the temperature safety device (chap. 12.1). The values entered in fixed-value entry mode remain valid after program run-off and are then equilibrated. If the week program timer is active, depending on the running week program another set-point (SP2) may be targeted. Temperatures too high for the introduced solvent quantity can occur. Deactivate the week program timer if you do not use it (default setting, setting in the User level, chap. 10). DANGER Too high temperature. Danger of explosion. Danger of death. � Deactivate the week program timer if you do not use it.

FDL (E2) 03/2011

page 28/68 7. Week program editor The Week program editor permits defining up to 4 shift point for each week day. A shift point defines a moment and the switching state ON or OFF of the channels that become active in this instance. Channel function: • Channel 1 On = Set-point 2 is equilibrated. • Channel 1 Off = Set-point 1 is equilibrated • Channel 2 = reserve The week program timer is initially set to inactive (factory setting). Therefore, you need to activate the week program timer in the user level (chap. 10). Normal Display Display 1 shows e.g. 39.8 (actual temperature value) Display 2 shows e.g. 15.01.07 13:52 - - (actual date and time, actual state of week program timer channel 1: Off, channel 2: Off) Press down key for 5 sec Display 1 shows e.g. 0000 Display 2 shows PROGRAM EDITOR (you are in the program editor) Press down key for 5 sec Display 1 shows 0000 Menu visible only if week program timer is activated in the user level (chap. 10) Display 2 shows WEEK PROG. EDITOR (you are in the week program editor) Press program key Display 1 shows 0000 Display 2 shows UserCod? 0000 (enter user code, display flashes) Enter the user code using arrow keys e.g. 0001 (basic setting, adjustable in the user level, chap. 10). Value is shown in both displays. Automatically forward after 2 sec Display 1 shows 0000 Display 2 shows Monday (selection of day of the week) (actual selection: Monday) Select the day of the week (Monday up to Sunday) with key Day of the week is shown in display 2. Press program key

FDL (E2) 03/2011

page 29/68 Display 1 shows 0000 Display 2 shows Shiftpt. (no function) Press program key Display 1 shows 0000 Display 2 shows Shiftpt. 1 (selection of the shift point) (actual shift point: 1) Select the shift point (1 up to 4) with key Value is shown in display 2. Press program key Display 1 shows e.g. --.-- (time of the selected shift point) Display 2 shows S1: --:-- (actual selection of the shift point: S1) (actual setting: shift point not programmed) Press program key Display 1 shows --.-- (time of the selected shift point) Display 2 shows Time --:-- (entry of the time of the selected shift point) (actual setting: shift point not programmed) Enter the time (hh:mm) using arrow keys Value is shown in display 2. Press key Display 1 shows 0000 Display 2 shows Ch1 = SP2: Off (entry of the state of channel 1) (actual setting: Off) Enter the state of channel 1 (On or Off) using arrow keys Setting is shown in display 2. Press key Display 1 shows 0000 Display 2 shows Channel 2: Off (entry of the state of channel 2) (actual setting: Off) (no function) Enter the state of channel 2 (On or Off) using arrow keys Setting is shown in display 2. Press key Press key EXIT Display 1 shows e.g., 08.30 (time of the selected shift point) Display 2 shows S1: 08:30 - - (actual selection of the shift point: S1) (actual setting: time 08.30, channels Off) Press program key Press key EXIT twice Select the next shift point Select the next day of the week To exit the menu, press several times key EXIT or wait for 120 seconds. Controller returns to normal display.

FDL (E2) 03/2011

page 30/68 7.1 Program table template for Week program Editor Program editor Program title Project Date: Time Channel 1 (tem- perature) Channel 2* Day of the week hh:mm AM PM ON = SP2 OFF = SP1 ON OFF S1 S2 S3 Monday S4 S1 S2 S3 Tuesday S4 S1 S2 S3 Wednesday S4 S1 S2 S3 Thursday S4 S1 S2 S3 Friday S4 S1 S2 S3 Saturday S4 S1 S2 S3 Sunday S4 * Channel 2 is non-functional in the standard unit

FDL (E2) 03/2011

page 31/68 8. Program editor 8.1 Selecting between set-point ramp and set-point step You can program various kinds of temperature transitions. In the user level (chap. 10) you can select between the settings “Ramp” (default setting) and “Step”. Setting “Ramp” permits programming all kinds of temperature transitions. With setting “Step” the controller will equilibrate only to constant temperatures; programming ramps is no longer possible. Switching between settings “Ramp” and “Step” will influence all programs. Please note that this can cause the time courses of existing programs to change significantly. 8.1.1 Programming with setting “Ramp” (default setting) Set-points always refer to the start of a program section, i.e., at the beginning of each program section, the entered set-point will be reached. During program section operation, the temperature gradually passes to the set-point entered for the subsequent program section. You can program all kinds of temperature transitions by the appropriate design of the program section timing: • Gradual temperature changes “set-point ramp“ The set-point gradually moves from one set-point to the one of the following program section during a given interval. The actual temperature value (X) follows the continually moving set-point (W) at any time. • Program sections with constant temperature The initial values of two subsequent program sections are identical; therefore the temperature is kept constant during the whole time of the first program section. • Sudden temperature changes “set-point step” Steps are temperature changes (ramps) that occur during a very short interval. Two program sections with an identical set-point are followed by a section with a different set-point. If the duration of this transitional program section is very short (minimum entry 1 min), the temperature change will proceed rapidly in the minimum amount of time. 01 02 03 04 05 06 07 08 09 W t Figure 8: Possible temperature transitions (with default setting “Ramp” in the user level (chap. 10)

FDL (E2) 03/2011

page 32/68 Program entry as set-point ramp (example): W/°C t/min. S01 S02 S03 S04 0 20 40 60 80 100 30 120 180 380 S05 Program table corresponding to the diagram (with default setting “Ramp”): Section SEC Temperature set-point [ °C] TEMP Section length [hh.mm] TIME S01 40 00:30 S02 60 01:30 S03 90 01:00 S04 90 03:20 S05 110 00:01 The values of such a program table can now be entered to the RD3 program controller (chap. 8.2). Program entry as set-point step (example): W/°C t/min. S01 S02 S03 S04 S05 S06 S07 S08 0 20 40 60 80 100 30 120 180 380

FDL (E2) 03/2011

page 33/68 Program table corresponding to the diagram (with default setting “Ramp”): Section SEC Temperature set-point [ °C] TEMP Section length [hh.mm] TIME S01 40 00:30 S02 40 00:01 S03 60 01:30 S04 60 00:01 S05 80 01:00 S06 80 00:01 S07 110 03:20 S08 110 00:01 The values of such a program table can now be entered to the RD3 program controller (chap. 8.2). The end point of the desired cycle must be programmed with an additional section (in our examples S05 for set-point ramp and S08 for set-point step) with a section time of at least one minute. Otherwise, the program will stop one section too early because the program line is incomplete. 8.1.2 Programming with setting “step” With setting “Step” selected, you don’t need to program the transition section in the Program Editor. With setting “step” the controller will equilibrate only to constant temperatures; programming ramps is no longer possible. The set-points are maintained constant for the duration of a program section. At the start of each program section, the unit heats up with the maximum speed in order to attain the entered set-point. Program entry as set-point step (example): W/°C t/min. S01 S02 S03 S04 S05 S06 S07 S08 0 20 40 60 80 100 30 120 180 380

FDL (E2) 03/2011

page 34/68 Program table corresponding to the diagram (with setting “Step”): Section SEC Temperature set-point [ °C] TEMP Section length [hh.mm] TIME S01 40 00:30 S02 60 01:30 S03 80 01:00 S04 110 03:20 The values of such a program table can now be entered to the RD3 program controller (chap. 8.2). 8.1.3 General notes on programming temperature transitions If the tolerance limits set in the user level (chap. 10) are exceeded, the program is halted until the actual temperature value returns to within the tolerance range. During this program interruption, the LED (3d) flashes. Therefore, the duration of the program might be extended due to the programming of tolerances The programming is saved even in case of a power failure or after turning off the unit. After program rundown the controller returns to fixed value operation showing Normal Display and equili- brates to the temperature value previously entered in fixed value entry mode. Before starting the program, check the set-point value entered in fixed value entry mode. After program rundown temperature will equilibrate to this value. Deactivate the week program timer (factory setting, setting in the user level, chap. 10) before starting a program. 8.2 Set-point entry for program operation From Normal Display, press down button X/W for 5 sec to access the program editor. Then enter the set- points one after the other in all program sections of a selected program. You can enter two programs with up to 10 sections each or one program with up to 20 sections (setting in the user level, chap. 10). In order to avoid incorrect programming, we recommend entering the values of the program course into a table (template in chap. 8.3). Example of program table (with default setting “Ramp”): Section SEC Temperature set-point [ °C] TEMP Section length [hh.mm] TIME S01 40 00:30 S02 60 01:30 S03 90 01:00 S04 90 03:20 S05 110 00:01 The values of the program table can now be entered to the RD3 program controller.

FDL (E2) 03/2011

page 35/68 Step 1 – Selecting the program and the program section: Normal Display Display 1 shows e.g. 39.8 (actual temperature value) Display 2 shows e.g. 15.01.07 13:52 - - (actual date and time, actual state of week program timer channel 1: Off, channel 2: Off) Press down key for 5 sec. Display 1 shows e.g. 0000 Display 2 shows PROGRAM EDITOR (you are in the program editor) Press program key Display 1 shows 0000 Display 2 shows UserCod? 0000 (enter user code) Enter user code using arrow keys e.g. 0001 (basic setting, adjustable in the user level, chap. 10). Value is shown in both displays. Automatically forward after 2 sec. Display 1 shows e.g. 01 (program P01 selected) --- : --- PRG. (program can be selected) Display 2 shows alternating CONTINUE X/W (information: to 1st program section with X/W) Select program P01 or P02 using arrow keys Value is shown in display 1. Press key In the selected program P01 or P02, program sections can be selected: Display 1 shows e.g. 01 (section S01 selected) P01: --- SEC. Display 2 shows alternating CONTINUE X/W section S01 has already been created. enter new set-points for the individual variables with button X/W or: Display 1 shows e.g. 01 (section S01 selected) P01: --- SEC. Display 2 shows alternating NEW SEC. X/W section S01 has not yet been created. enter set-points for the individual variables with button X/W Select sections S01 to S10 or to S20 using arrow keys As long as no program section has been entered, the display switches back to 01 in case of any entry > 01, because all sections need to be entered one after the other, and each new section is created as NEWSEC. Example: If three programs sections have been already entered, the next section to be entered is S04. Before this, no section > S04 can be selected.

FDL (E2) 03/2011

page 36/68 Next step – set-point entry in the desired program sections: Basic entry principle: Access the parameters of individual program sections with button X/W one after the other. Enter the values of the individual parameters with the arrow keys. A value flashing once after 2 seconds indicates that it has been adopted by the controller. If several parameters are to be skipped (e.g. in order to change a parameter in a posterior program section), the parameters can be rapidly jumped over by holding down the X/W key. If no button is pressed for more than 120 sec the controller switches back to Normal Display. The program entered to this point remains stored. Press key Display 1 shows e.g. 40.0C (actual temperature set-point) S01: TEMP 40.0 (variable: temperature in °C) Display 2 shows alternating CONTINUE X/W (information: go on with X/W) Enter temperature set-point of S01 in °C using arrow keys Press key Display 1 shows e.g. 00.30 (actual section length set-point) S01: TIME 00:30 (variable: section length in hh:mm) Display 2 shows alternating CONTINUE X/W (information: go on with X/W) Enter section length set-point of S01 in hh.mm using arrow keys Value is shown in both displays. Press key Selecting the next program sections to be entered Display 1 shows e.g. 02 (section S02 selected) P01: --- SEC. Display 2 shows alternating CONTINUE X/W Section S02 has already been created. enter new set-points for the individual parameters with X/W. or: Display 1 shows e.g. 02 (section S02 selected) P01: --- SEC. Display 2 shows alternating NEW SEC. X/W Section S02 has not yet been created. enter set-points for the individual parameters with X/W Select the next section to be entered using arrow keys Display 1 shows e.g. 60.0C (actual temperature set-point) S02:TEMP 60.0 (variable: temperature in °C) Display 2 shows alternating CONTINUE X/W (information: go on with X/W) Enter the temperature set-point of S02 in °C using arrow keys Etc. If all sections up to S10 or up to S20 have been programmed, section S01 follows again. In order to quit the entry mode, press the “EXIT” button several times or wait 120 sec → the controller will then return to Normal Display. When changing the set-point, check the setting of the safety device (chap. 12.1).

FDL (E2) 03/2011

page 37/68 8.3 Program table template Program editor Program title Project Program No. Date: Section SEC Temperature set-point [ °C] TEMP Section length [hh.mm] TIME S01 S02 S03 S04 S05 S06 S07 S08 S09 S10 S11 S12 S13 S14 S15 S16 S17 S18 S19 S20

FDL (E2) 03/2011

page 38/68 8.4 Deleting a program section A program section is deleted from the program by setting the section duration to Zero. Normal display Press down key for 5 sec. Display 1 shows e.g. 0000 Display 2 shows PROGRAM EDITOR (you are in the program editor) Press program key Display 1 shows 0000 Display 2 shows UserCod? 0000 (enter user code) Enter the user code using arrow keys e.g. 0001 (basic setting, adjustable in the user level, chap. 10). Value is shown in both displays. Automatically forward after 2 sec Display 1 shows e.g. 01 (program P01 selected) --- : --- PRG. (program can be selected) Display 2 shows alternating CONTINUE X/W (information: to 1st program section with X/W) Select the desired program, e.g. P01, using arrow keys Value is shown in display1. Press key In the selected program P01 or P02, program sections can be selected: Display 1 shows e.g. 01 (actual selection of the section: S01) P01: --- SEC. (program section can be selected) Display 2 shows alternating CONTINUE X/W (information: set-point entry with X/W) Select desired section, e.g. S03, using arrow keys (omitted if section S01 shall be deleted). Press key Display 1 shows e.g. 90.0C (actual temperature set-point) S03:TEMP 90.0 (variable: temperature) Display 2 shows alternating CONTINUE X/W (information: continue with X/W) No entry Press key Display 1 shows e.g. 01.00 (actual section length) S03:TIME 01:00 (variable: section length) Display 2 shows alternating CONTINUE X/W (information: continue with X/W) Enter set-point Zero for section length in hh:mm of S03 using arrow keys Value is shown in display 2 or in both displays (display depends on maximum time setting in the user level chap. 10) Display 1 shows e.g. 00.00 (actual section length) S03:TIME 00:00 (variable: section length) Display 2 shows alternating DELETE SEC. X/W (information: delete section with X/W) Press key

FDL (E2) 03/2011

page 39/68 The following section (in our example now S03) is displayed: Display 1 shows e.g. 03 (actual selection of the section: S03) P01:S03 (program section can be selected) Display 2 shows alternating CONTINUE X/W (information: set-point entry with X/W) Press key EXIT or wait 120 sec Controller returns to Normal Display If you delete a program section which is followed by further sections, those following move up in place of the deleted section. In our example, section S03 has been deleted. If sections S04, S05, etc. have been programmed earlier, they will now replace the preceding sections, i.e., S04 is now called S03 etc. Deletion leads to overwriting the section by the following one. It is therefore not possible to temporarily inactivate a program section. To enter a section later to a program, all the sections following the new one must be entered again. 9. Program start level Before starting the program, check the set-point entered in Fixed value operation mode. After end of the program, the temperature will equilibrate to this value. This value must not exceed the permitted drying temperature for the used solvent quantity. DANGER Too high temperature after the program ends. Danger of explosion. Danger of death. ∅ Set-point of Fixed value operation must NOT exceed the maximum drying temperature suitable for the solvent quantity. � Check the set-point of Fixed value operation and if necessary adapt it. After the program ends, the temperature will equilibrate to the set-point entered in Fixed value operation mode. If the week program timer is active, another set-point (SP2) might be targeted according to pro- gramming. Temperatures too high for the introduced solvent quantity can occur. Deactivate the week program timer before starting the program (default setting, setting in the User level, chap. 10). DANGER Too high temperature after the program ends. Danger of explosion. Danger of death. � Deactivate the week program timer before starting the program. In the first step, select a program. This is on condition that a program has been entered previously (chap. 8.2) and that program type “2 programs with 10 sections each” has been selected in the user level (chap. 10).

FDL (E2) 03/2011

page 40/68 Then define the settings for the program course. Two parameters can be set: • Program delay time, i.e. a defined time before a program starts. It can be entered with a precision of 1 minute, and its maximum value is 99.59 (99 hs 59 min). If the value is 00.00, the program will start immediately. • Number of program cycles, i.e. the desired number of program repeats. Values from 1 to 99 can be entered. If the program is not going to be repeated, enter the value “0”. For infinite repeats enter the value “–1”. The program is repeated as a whole; it is not possible to repeat individual sections. In the last step start the selected program. These steps must be carried out subsequently. Deactivate the week program timer (factory setting, setting in the user level, chap. 10) before starting a program. Step 1 – Program selection (only with program type “2 programs” set): Normal Display Press program key Display 1 shows e.g. 1 (actual selection of the program) Display 2 shows SEL.PRG. (select program 1 or 2) Enter program number 1 or 2 using arrow keys Value is shown in display 1. Next step – entry of program course settings Press program key Display 1 shows e.g. 00.00 (entered delay time hh.mm) Display 2 shows RUN TIME (enter delay time of program start) Set delay time in hh.mm using arrow keys Value is shown in display 1. Press program key Display 1 shows e.g. -1 (actual selection of the number of program cycles) Display 2 shows REPEAT (enter number of program cycles) Select number of cycles -1, 0, 1 etc. using arrow keys Value is shown in display 1. Last step – program start: Press program key Display 1 shows e.g. 1 (selected program) Display 2 shows RUN PRG. (Question: start selected program?) Press program key Display 1 shows e.g. 25.5C actual temperature value Display 2 shows P01:S01 00:29:39 (time running backwards) (actual program P01, actual section S01, and remaining time of program section S01) Program is running. The green LED (3d) lights up. In addition to the green LED (3d) indicating a running program, the LED (3a) is lit if the heating is active, or no LED if the actual temperature equals the set-point.

FDL (E2) 03/2011

page 41/68 During program course the arrow keys and the EXIT button are not functional. By pressing the program key for 3 seconds, you can terminate the program course. If you press button during program course, the entered set-point of the actually running program sec- tion is shown for 5 seconds: Display 1 shows e.g. 65.5C (actual temperature value) Display 2 shows P01:S03 00:47:12 (actual program P01, actual section S03, and remaining time of program section S03) Press key Display 1 shows e.g. 90 (actual temperature set-point 1) Display 2 shows SP1 TEMPERATURE 5 seconds Display 1 shows e.g. 30 (actual temperature set-point 2) Display 2 shows SP2 TEMPERATURE (no function during program operation) 5 seconds After program runoff (and, if appropriate, of the program repeats) the controller returns to fixed value op- eration showing Normal Display and adjusting to the temperature value that has been previously entered in the fixed value entry mode. 10. User level In this menu the following parameters can be se (in brackets the corresponding abbreviated information given in display 2): • Unit address (Adress) Set the controller address (1 to 255) for operation with the communication software APT-COM™. • User code (User-cod) Modification of the user code setting (factory setting 0001) for access to the user level and the pro- gram editor. Keep in mind any modification of the user code. There is no access to these levels without a valid user code. • Decimal point position (Decimal) Selection if integer values or one position after the decimal point can be entered. The integer repre- sentation is shown in Display 2 (set-point entry) while the actual value in Display 1 is always shown with one decimal point. • Audio Alert (Buzzer) Inactive: no audible signal (buzzer) in case of an alarm event. Active: in case of an alarm event (see chap. 11.2) an audible signal (buzzer) will sound. It can be re- set by pressing the “EXIT” button. • Selection of controller menu language (Language) German, English, or French can be selected.

FDL (E2) 03/2011

page 42/68 • Counter of operating hours (Oper.hs) Information about the number of operating hours currently reached or since the last reset. (no setting, display only). • Max. number of operating hours (Op.limit) Enter a limit number of operating hours, i.e., the maximum number of operating hours that can be run. Maximum setting: 9999. Reaching the limit has no effect. • Reset operating hours (Op.back) Reset operating hours to zero. • Interface protocol (Protocol) “Modbus”: The chamber interface can be used as a communication interface to connect it to a com- puter. This serves to control the chamber by the communication software APT-COM™. It is possible to read and write the values of all parameters. “Printer”: A protocol printer (option) for data printouts can be connected to the chamber interface. The printer regularly protocols the actual temperature value with fixed formatting and with adjustable print intervals (chap. 13.2). In both cases an interface converter RS 422 / RS 232 is used. • Print interval (Prt.-Inv.) Set the print interval in minutes. Function is available only if setting “Printer” has been selected in the previous menu point. • Display illumination (Disp.LED) Select between continuous display illumination and limited illumination, which will automatically go off 300 sec after the last entry. • Program type selection (PrgSelec) Select between entry of two programs with up to 10 sections each or of one program with up to 20 sections. When changing from 2 programs to 1 program or vice-versa, existing programs are deleted in the program editor. • Maximum section duration (Prg.Time) The maximum length of an individual program section can be set to either 99 hs 59 min or to 999 hs 59 min. This setting is then valid for all program sections. When changing the maximum duration setting, pre-existing programs will be deleted in the program editor. • Set-point programming type (Setp.sim) Select between “Ramp” and “Step”. With setting “Step” selected, you don’t need to program the transi- tion section in the Program Editor. If you select setting “Step”, the controller will equilibrate only to constant temperatures; programming ramps becomes impossible. A change between settings “ramp” and “step” will influence all programs. Note that signifi- cant change in time courses may arise in existing programs. • Tolerance limit range (Tol.band) Entry of a tolerance limit value in °C. If the actual value of temperature exceeds the set-point of a pro- gram section by more than the entered tolerance limit value, the program is halted (LED (3d) flashing) until the actual temperature value is again within the tolerance range. Entry of “0” means tolerance limits are off.

FDL (E2) 03/2011

page 43/68 • Activating or inactivating the week program timer (Prog.Clk) „Inactive“: The week program timer is turned off (factory setting). The corresponding setting menu (chap. 7) is not visible, nor is set-point 2 in the “Fixed value entry mode” (chap. 6). „Active“: The week program timer is activated. When deactivating the week program timer, any programming made in advance will re- main in memory and take effect when the week program timer is activated again.. Deactivate the week program timer before staring a program (chap. 9). • Display mode (12h/24h) Select between 12 hours (display “AM” or “PM”) or 24 hours. • Date of the real time clock (Date) Main menu. Use the program key to access the settings of year, month, and day in the corresponding submenus. • Year of the real time clock (Year) Enter the year (2006 up to 2050) • Month of the real time clock (Month) Enter the month (1 up to 12). • Day of the real time clock (Day) Enter the day (1 up to 31). • Time of the real time clock (Time) Main menu. Use the program key to access the settings of hour and minute in the corresponding sub- menus. There is no automatic switch between daylight saving time and regular time. • Hour of the real time clock (Hour) Enter the hour (0 up to 23). • Minute of the real time clock (Minute) Enter the minute (0 up to 59).

FDL (E2) 03/2011

page 44/68 Normal Display Display 1 shows e.g. 19.8 (actual temperature value) Display 2 shows e.g. 15.05.06 13:52 - - (actual date and time, actual switching state of week pro- gram timer channel 1: Off, channel 2: Off) Press down key for 5 sec Display 1 shows e.g. 0000 Display 2 shows PROGRAM EDITOR (you are in the program editor) Press down key for 5 sec Display 1 shows 0000 Menu visible only if week program timer is activated. Display 2 shows WEEK PROG. EDITOR (you are in the week program editor) Press down key for 5 sec Display 1 shows 0000 Display 2 shows USER – LEVEL (you are in the user level) Press program key Display 1 shows 0000 Display 2 shows UserCod? 0000 (enter user code, display flashes) Enter the user code using arrow keys e.g. 0001 (basic setting, or the valid code in case it has been previously changed in this menu). Value is shown in both displays. Automatically forward after 2 sec Display 1 shows 1 (actual address: 1) Display 2 shows Adress 1 (entry of unit address) (actual address: 1) Enter the unit address (1 up to 254) using arrow keys Address is shown in both displays. Press key Display 1 shows 1 (actually valid user code: 1) Display 2 shows User-cod 1 (change user code) (actually set: 1) Enter a new value using arrow keys Value is shown in both displays. Press key Display 1 shows 0000 (no function) Display 2 shows Saf.mode: Limit (no function) Press key Display 1 shows 0 (no function) Display 2 shows Saf.setp 0 (no function) Press key Display 1 shows 0000 (no function) Display 2 shows Decimal: XXX.X (setting of decimal point position) (actual setting: XXX.X) Select decimal point position using arrow keys Decimal point position XXX.X or XXXX. is shown in display 2.

FDL (E2) 03/2011

page 45/68 Press key Display 1 shows 0000 (no function) Display 2 shows Buzzer : Active (setting of the alarm buzzer) (actual setting: „Active“) Select between “Active” and “Inactiv” using arrow keys Setting is shown in display 2. Press key Display 1 shows 0000 (no function) Display 2 shows Language : English (selection of controller language) (actual setting: English) Select between languages German, English, and French using arrow keys Setting is shown in display 2. Press key Display 1 shows e.g. 0004 (unit operating hours) Display 2 shows Oper.hs 0004:28 (operating hours up to now hhhh:mm) (actually displayed: 4 hs 28 min) Press key Display 1 shows 1000 (actual setting: 1000 hs) Display 2 shows Op.limit 1000:00 (maximum number of operating hours hhhh:mm (actual setting: 1000 hs) Set value using arrow keys Setting is shown in both displays. Press key Display 1 shows 0000 (no function) Display 2 shows Op.back : No (reset counter of operating hours ?) (actual setting: No) Select between “Yes” and “No” using arrow keys Setting is shown in display 2. Press key Display 1 shows 0000 (no function) Display 2 shows Protocol: MODBUS (Selection of interface mode) (actual setting: Modbus) Select between protocols “MODBUS” and “Printer” using arrow keys Setting is shown in display 2. Press key Display 1 shows e.g. 3 (actual setting: 3 min) Display 2 shows Prt-Inv. 3 (print interval) (actual setting: 3 min) Select value between 0 and 255 minutes using arrow keys Setting is shown in displays 1 and 2. Press key Display 1 shows 0000 Display 2 shows Disp.LED: No (continuous display illumination?) (actual setting: No) Select between “Yes” and “No” using arrow keys Setting is shown in display 2.

FDL (E2) 03/2011

page 46/68 Press key Display 1 shows 0000 Display 2 shows PrgSelec: 2Prg10S (1 program with max. 20 sections or 2 programs with max. 10 sections each?) (actual setting: 2Prg10S) Select between “2Prg10S” and “1Prg20S” using arrow keys Setting is shown in display 2. Press key Display 1 shows 0000 Display 2 shows Prg.Time: 99:59 (max. section length 99:59 or 999:59?) (actual setting: 99:59) Select between 99:59 in hh:mm or 999:59 in hhh:mm using arrow keys Setting is shown in display 2. Press key Display 1 shows 0000 Display 2 shows Setp.sim Ramp (ramp or step?) (actual setting: ramp) Select between Ramp and Step using arrow keys Setting is shown in display 2. Press key Display 1 shows 0000 Display 2 shows Tol.band 0 (Tolerance limits in °C) (actual setting: 0) Set value in °C using arrow keys Setting is shown in display 2. Press key Display 1 shows 0000 Display 2 shows Prog.Clk Inactive (Week program timer active or inactive? (actual setting: Inactive) Select between “Active” and “Inactive” using arrow keys Setting is shown in display 2. Press key Display 1 shows 0000 Display 2 shows 12h/24h 24h (Display mode 12 hours or 24 hours? (actual setting: 24h) Select between 12 hours and 24 hours us- ing arrow keys Setting is shown in display 2.

FDL (E2) 03/2011

page 47/68 Press key Display 1 shows 0000 Display 2 shows Date (Main menu: Setting the date of the real time clock) Press program key Display 1 shows e.g. 2006 (Actual setting: 2006) Display 2 shows Year 2006 (Setting the year of the real time clock) Set year (2006 up to 2050) using arrow keys Setting is shown in display 2. Press key Display 1 shows e.g. 5 (Actual setting: may) Display 2 shows Month 5 (Setting the month of the real time clock) Set month (1 up to 12) using arrow keys Setting is shown in display 2. Press key Display 1 shows e.g. 15 (Actual setting: 15) Display 2 shows Day 15 (Setting the day of the real time clock) Set day (1 up to 31) using arrow keys Setting is shown in display 2. Press key EXIT Press key Display 1 shows 0000 Display 2 shows Time (Main menu: Setting the time of the real time clock) Press program key Display 1 shows e.g. 13 (Actual setting: 13, i.e. 1 p.m.) Display 2 shows Hour 13 (Setting the hour of the real time clock) Set hour (0 up to 23) using arrow keys Setting is shown in display 2. Press key Display 1 shows e.g. 30 (Actual setting: 30 minutes) Display 2 shows Minute 30 (Setting the minute of the real time clock) Set minute(0 up to 59) using arrow keys Setting is shown in display 2. Press several times key EXIT or wait for 120 seconds Controller returns to normal display.

FDL (E2) 03/2011

page 48/68 11. Behavior in case of failures 11.1 Behavior after power failure Power failure during fixed-value operation (Normal Display): the entered parameters remain saved. After power returns, operation continues with the set parameters. The heating starts operating after a delay time of 4 minutes. Power failure during program operation: After power return program course continues with the set- points that have been reached previously during program operation. The heating starts operating after a delay time of 4 minutes. 11.2 Alarm messages Alarm messages, e g. “RANGE ERROR CH1” in case of sensor rupture, are shown in Display 2 only in Normal Display. A buzzer can be activated / deactivated in the user level (chap. 10). It can be reset by pressing the EXIT button. The alarm text shown in Normal Display goes off only if the cause of the alarm does not exist any longer. 12. Safety devices 12.1 Temperature safety device class 2 (DIN 12880:2007) The temperature safety device class 2 protects the unit, its environment and the charging material from exceeding the maximum permissible temperature. Please observe the guideline BGI/GUV-I 850-0 on safe working in laboratories (formerly BGR/GUV-R 120 or ZH 1/119 laboratory guidelines issued by the employers’ liability insurance association) (for Ger- many). In the event of a fault in the temperature controller, the safety device (2) permanently turns off the unit. Only the fan continues running for safety reasons. This status is reported visually by the indicator lamp (2a) and acoustically by the buzzer. The operation of the safety device (2) is checked by moving it slowly counter-clockwise until it is turned off. The safety device cut-off is reported visually by the indicator lamp (2a) and acoustically by the buzzer. Then release the safety device by pressing the reset button (2b) and turn on the unit is as described pre- viously. Figure 9: Safety thermostat class 2 Function: The safety thermostat class 2 is functionally and electrically independent of the temperature control de- vice and turns off the unit at all poles. When the control knob (2) is set to the end stop (position 10), the safety thermostat class 2 acts as a unit protection device. If it is set somewhat higher than the set-point temperature selected on the controller, it acts as a material protection device. (2) (2a) (2b)

FDL (E2) 03/2011

page 49/68 When the safety device has turned off the unit, identifiable by the red alarm lamp (2a) lighting up, proceed as follows: • Disconnect the unit from the power supply. • Have the cause of the fault examined and rectified by a technician. • Release safety thermostat class 2 by pressing reset button (2b). • Restart the unit as described in chap. 5. Setting: The diagram given in chap. 2.3 indicates the drying temperature to be set on the controller in relation to the quantity of the introduced solvents. This must not be exceeded. For this reason, adjust the tempera- ture safety device according to the selected set-point. DANGER Excessive drying temperature. Danger of explosion. Danger of death. ∅ Do NOT exceed the maximum drying temperature for the solvent quantity. To check at which temperature the safety device activates, turn on the oven and set the required nominal value on the temperature controller. The scale division from 1 to 10 corresponds to the temperature range from 30 °C / 86°F up to 320 °C / 608°F and serves as a setting aid. • Detach the plastic cover over the temperature safety device class 2 (2) with a suitable Phillips screwdriver. • Turn the control knob (2) of the safety device using a coin to its end-stop (position 10) (unit protection). • When the set point is reached, turn back the control knob (2) until its trip point (turn it counter-clockwise) is reached. • The red alarm lamp (2a) lighting up identifies the trip point; and the reset button (2b) pops out. • The optimum setting of the safety device is obtained by turning the knob clockwise by approx. one graduation mark on the scale. • Push the reset button (2b) in again. • Restore the plastic cover to prevent misadjusting Figure 10: Setting the safety device class 2 The unit is only active when the reset button (2b) is pushed in. When the safety thermostat class 2 responds, the red alarm lamp (2a) illuminates, the reset button (2b) pops out and the unit turns off permanently at all poles. Only the fan keeps turning for safety reasons. Check the safety thermostat regularly and adjust it following changes of the set-point. Fixed value operation: Adapt the temperature safety device every time the set-point for temperature is changed. Set the set-point of temperature safety device by about 5 °C to 10 °C above the controller temperature set-point. Program operation: Adapt the temperature safety device to the highest temperature set- point value of the program actually used. Functional check: Test the functional capability of the safety thermostat at appropriate intervals. It is recommended having this check performed by the authorized service personnel, e.g. before the beginning of a long working process.

FDL (E2) 03/2011

page 50/68 12.2 Exhaust air monitoring If the volumetric flow rate of the exhaust air is too low, the heating turns off immediately for safety rea- sons. The indicator light (3) lights up and an audible signal is given. The indicator light and the buzzer can only be turned off by pushing key (4). If the cause of the low volume of exhaust air is not resolved when the key is released, the error messages will be given again. RESET (4) (3) (3) Red indicator light, loss of technical ventilation (4) RESET key for the heating cut-out due to failure of tech- nical ventilation Figure 11: Instrument panel (detail) It is the operator’s responsibility to ensure that the doors of drying ovens are opened immediately on fail- ure of the technical ventilation system (GUV-R 500 chap. 2.28). On failure of the technical ventilation, open the door of the safety / paint drying oven. 13. Options 13.1 Communication software APT-COM™ 3 DataControlSystem (option) The unit is regularly equipped with a serial interface RS 422 that can connect the BINDER communica- tion software APT-COM™ 3 DataControlSystem. The connection to a computer is established using the FDL interface via an interface converter RS 422 / RS 232. Confirm that the interface mode is correctly set to “Modbus” in the user level (chap. 10). The actual temperature values are given at adjustable intervals. Programming can be performed graphi- cally via PC. Up to 30 chambers with RS 422 interface can be cross linked. For further information, please refer to the operating manual of the BINDER communication software APT-COM™ 3. Pin allocation of the RS 422 interface at the rear of the incubator: Pin 2: RxD (+) Pin 3: TxD (+) Pin 4: RxD (-) Pin 5: TxD (-) Pin 7: Ground 13.1 Ethernet interface (option) With this option, the chamber is equipped with an Ethernet interface that can connect the BINDER com- munication software APT-COM™ 3 DataControlSystem. The actual temperature values are given at ad- justable intervals. Programming can be performed graphically via PC. Up to 30 chambers with RS 422 interface can be cross-linked. The MAC Address is indicated below the Ethernet interface. For further information, please refer to the operating manual of the BINDER communication software APT-COM™ 3. The additional RS422 interface is only used for service purposes. Do NOT connect it to any network. The interface is labeled accordingly.

FDL (E2) 03/2011

page 51/68 13.2 Protocol printer (option) The protocol printer is connected using the FDL interface via an interface converter RS 422 / RS 232. Confirm that the interface mode is correctly set to “Printer” in the user level (chap. 10). Figure 12: Protocol printer Interface configuration: Baud rate: 9600 Stop bit: 1 Parity: none The actual temperature values are put out regularly with fixed formatting. Printout: one printed line for each print interval with relative time stamp, temperature value with one deci- mal point, curve representation (see “Manual for Setting Matrix Printer Epson LX-300+“, Art. No. 7001- 0041). In every 5th line the set print interval “Ptime” in minutes is noted. The printer interval is set in the user level (chap. 10). Example: Temp.: 34.7 * Temp.: 35.6 * Temp.: 32.8 * Temp.: 30.1 * Ptime : 001. * Temp.: 27.4 * Temp.: 26.9 * Temp.: 26.6 * Temp.: 26.4 * Ptime : 001. * Temp.: 26.2 * In this example the print interval is set to 1 min, i.e., every minute a new temperature value is printed. In conjunction with the specified print interval, real-time reference can be achieved by noting down the starting time of recording. For operating instructions of the printer, please refer to the operating manual of the printer manufacturer.

FDL (E2) 03/2011

page 52/68 13.3 Additional measuring channel for digital object temperature indicator with magnetic sensor or pincer-type sensor (option) The object temperature display enables the determination of the actual temperature of the specimen dur- ing the whole process. The supplied pincer-type or magnetic sensor is used to display the object tem- perature at Display 2 of the RD3 controller. The object temperature data are put out together with the data of the temperature controller to the RS 422 interface as a second measuring channel and can be documented by the communication software APT-COM™ developed by BINDER (option, chap. 13.1). Socket (inner chamber Pincer-type sensor Material sample Magnetic sensor Figure 13: Sensor probes for optional temperature display With this option the chamber is equipped with an analog output 4-20 mA for temperature. This output permits transmitting data to external data registration systems or devices. The connection is realized as a DIN socket at the rear of the chamber as following: ANALOG OUTPUT 4-20 mA DC PIN 1: temperature – PIN 2: temperature + Temperature range: 0 °C / 32 °F to +300 °C / 572 °F A suitable DIN plug is enclosed. Figure 14: Pin allocation of DIN socket

FDL (E2) 03/2011

page 53/68 14. Maintenance, cleaning, and service 14.1 Maintenance intervals, service DANGER Electrical hazard. Danger of death. ∅ The unit must NOT become wet during operation or maintenance work. ∅ Do NOT remove the rear panel of the unit. � Disconnect the unit before conducting maintenance work. Disconnect the power plug. � Ensure all maintenance work is conducted by licensed electricians or experts author- ized by BINDER. Ensure regular maintenance work is performed at least once a year. The warranty becomes void if maintenance work is conducted by non-authorized personnel. It is particularly important to test the function of the flow monitor once a year. Record the results of the maintenance tests in a service log Replace the door gasket only when cold. Otherwise, the door gasket may become damaged. We recommend taking out a maintenance agreement. Please consult BINDER Service. BINDER telephone hotline: +49 (0) 7462 2005 555 BINDER fax hotline: +49 (0) 7462 2005 93555 BINDER e-mail hotline: [email protected] BINDER service hotline USA: +1 866 885 9794 or +1 631 224 4340 (toll-free in the USA) BINDER service hotline Spain +34 9492 677 23 BINDER service hotline Asia Pacific: +852 39070500 or +852 39070503 BINDER service hotline Russia and CIS +7 495 98815 17 BINDER Internet website http://www.binder-world.com BINDER address BINDER GmbH, post office box 102, D-78502 Tuttlingen International customers, please contact your local BINDER distributor.

FDL (E2) 03/2011

page 54/68 14.2 Replacing the intake filter The fresh air intake filter (fine particle filter for particles 1µm up to 10 µm, class F6/EU6 acc. to EN 779:2002) at the top right-hand side must be replaced from time to time, depending on the degree of soiling. Pull out the slide and replace the filter cartridge. Intake filter Slide Figure 15: Fresh air intake filter 14.3 Cleaning and decontaminating the safety / paint drying oven DANGER Electrical hazard. Danger of death. ∅ Do NOT spill water or cleaning agents over the inner and outer surfaces. � Disconnect the chamber before cleaning. Disconnect the power plug. � Completely dry the appliance before turning it on again. Cleaning Always keep clean the inner parts of the safety / paint drying oven including the drip pans, drip trays and exhaust-air ducts. Remove the remnants of coating material at suitable intervals. Disconnect the chamber from the power supply before cleaning. Disconnect the power plug. Wipe the surfaces with a moistened towel. In addition, you can use the following cleaning agents: Exterior surfaces inner chamber racks door gaskets Standard commercial cleaning detergents free from acid or halogenides. Alcohol based solutions. We recommend using the neutral cleaning agent Art. No. 1002-0016. Instrument panel Standard commercial cleaning detergents free from acid or halogenides. We recommend using the neutral cleaning agent Art. No. 1002-0016. For surface protection, perform cleaning as quickly as possible. After cleaning completely remove any cleaning agents from the surfaces with a moistened towel. Soapsuds may contain chlorides and must therefore NOT be used for cleaning.

FDL (E2) 03/2011

page 55/68 Decontamination Disconnect the chamber from the power supply prior to decontamination. Pull the power plug. You can use the following disinfectants: Inner chamber Standard commercial surface disinfectants free from acid or halogenides. Alcohol based solutions. We recommend using disinfectant Art. No. 1002-0022. In case of contamination of the interior by biologically or chemically hazardous goods, there are three possible procedures depending on the type of contamination and charging material. (1) The safety / paint drying ovens FDL can be hot air sterilized at 190 °C / 374°F for at least 30 minutes. All inflammable goods must be removed from the interior before. (2) Spray the inner chamber with an appropriate disinfectant. Before start-up, the unit must be absolutely dry and ventilated, as explosive gases may form during the decontamination process. (3) If necessary, have strongly contaminated inner chamber parts removed by an engineer for cleaning, or have them exchanged. Sterilize the inner chamber parts in a sterilizer or autoclave. With every decontamination method, always ensure adequate personal safety. CAUTION Danger of corrosion. Damage to the unit. ∅ Do NOT use acidic or chlorine cleaning detergents. We recommend using the neutral cleaning agent Art. No. Art. Nr. 1002-0016 for a thorough and mild cleaning. Any corrosive damage that may arise following use of other cleaning agents is excluded from liability by BINDER GmbH. 14.4 Sending the unit back to BINDER GmbH If you return a BINDER product to us for repair or any other reason, we will only accept the product upon presentation of an authorization number that has previously been issued to you. An authorization number will be issued after receiving your complaint either in writing or by telephone prior to your sending the BINDER product back to us. The authorization number will be issued following receipt of the information below: • BINDER product type and serial number • Date of purchase • Name and address of the dealer from which you bought the BINDER product • Exact description of the defect or fault • Complete address, contact person and availability of that person • Exact location of the BINDER product in your facility • A contamination clearance certificate (chap. 18) must be faxed in advance The authorization number must be applied to the packaging in such a way that it can be easily recog- nized or be recorded clearly in the delivery documents. For security reasons we cannot accept a unit delivery if it does not carry an authorization num- ber.

FDL (E2) 03/2011

page 56/68 15. Disposal 15.1 Disposal of the transport packing Packing element Material Disposal Straps to fix packing on pallet Plastic Plastic recycling Non-wood (compressed match- wood, IPPC standard) Wood recycling Wooden transport box (option) with metal screws Metal Metal recycling Solid wood (IPPC standard) Wood recycling Pallet with foamed plastic stuffing PE foam Plastic recycling Cardboard Paper recycling Shipping box with metal clamps Metal Metal recycling Edge protection Styropor® or PE foam Plastic recycling Protection of doors and racks PE foam Plastic recycling Bag for operating manual PE foil Plastic recycling Insulating air cushion foil (packing of optional accessories) PE foil Plastic recycling If recycling is not possible, all packing parts can also be disposed of with normal waste. 15.2 Decommissioning Turn off the main power switch (1). Disconnect the unit from the power supply. When switching off the main power switch ON / OFF (1), the stored parameters remain saved. • Temporal decommissioning: See indications for appropriate storage, chap. 3.3. • Final decommissioning: Dispose of the unit as described in chap. 15.3 to 15.5. 15.3 Disposal of the unit in the Federal Republic of Germany According to directive 2002/96/EC of the European Parliament and of the Council on waste electrical and electronic equipment (WEEE), BINDER devices are classified as “monitoring and control instruments” (category 9) only intended for professional use“. They must not be disposed of at public collecting points. The safety / paint drying oven FDL bears the symbol for the marking of electrical and elec- tronic equipment manufactured / placed on the market in the EC after 13 August 2005 and be disposed of in separate collection according to the directive 2002/96/EC on waste elec- trical and electronic equipment (WEEE) and German national law for electrical and elec- tronic equipment (Elektro- und Elektronikgerätegesetz, ElektroG). WEEE marking: crossed- out wheeled bin with solid bar under. A significant part of the materials must be recycled in order to protect the environment. At the end of the device’s service life, have the device disposed of according to the German national law for electrical and electronic equipment (Elektro- und Elektronikgerätegesetz, ElektroG) from 23 March 2005, BGBl. I p. 762 or contact BINDER service who will organize taking back and disposal of the unit according to the German national law for electrical and electronic equipment (Elektro- und Elektronik- gerätegesetz, ElektroG) from 23 March 2005, BGBl. I p. 762.

FDL (E2) 03/2011

page 57/68 CAUTION Violation against existing law. ∅ Do NOT dispose of BINDER devices at public collecting points. � Have the device disposed of professionally at a recycling company that is certified ac- cording to the German national law for electrical and electronic equipment (Elektro- und Elektronikgerätegesetz, ElektroG) from 23 March 2005, BGBl. I p. 762. or � Instruct BINDER service to dispose of the device. The general terms of payment and delivery of the BINDER GmbH apply, which were valid at the time of purchasing the unit. Certified companies disassemble waste (used) BINDER equipment in primary substances for recycling according to directive 2002/96/EC. The devices must be free from toxic, infectious or radioactive sub- stances in order to eliminate any health hazards to the employees of the recycling companies. Prior to handing the unit over to a recycling company, it is the user’s responsibility that it is free from toxic, infectious or radioactive substances. − Prior to disposal clean all introduced or residual toxic substances from the unit. − Prior to disposal disinfect the unit from all sources of infection. Be aware that sources of infection may also be located outside the inner chamber. − If you cannot safely remove all toxic substances and sources of infection from the unit, dispose of it as special waste according to national law. − Fill out the contamination clearance certificate (chap. 18) and enclose it with the unit. WARNING Contamination of the device with toxic, infectious or radioactive substances. Danger of intoxication. Danger of infection. ∅ NEVER take a unit contaminated with toxic substances or sources of infection for recy- cling according to directive 2002/96/EC. � Prior to disposal, remove all toxic substances and sources of infection from the unit. � A unit from which all toxic substances or sources of infection cannot be safely removed must be considered as “special” waste according to national law. Dispose of it accord- ingly. 15.4 Disposal of the unit in the member states of the EC except for the Federal Republic of Germany According to directive 2002/96/EC of the European Parliament and of the Council on waste electrical and electronic equipment (WEEE), BINDER devices are classified as “monitoring and control instruments” (category 9) only intended for professional use“. They must not be disposed of at public collecting points. The safety / paint drying oven FDL bears the symbol for the marking of electrical and elec- tronic equipment manufactured / placed on the market in the EC after 13 August 2005 and be disposed of in separate collection according to the directive 2002/96/EC on waste elec- trical and electronic equipment (WEEE). WEEE marking: crossed-out wheeled bin with solid bar under.

FDL (E2) 03/2011

page 58/68 At the end of the device’s service life, notify the distributor who sold you the device, who will take back and dispose of the unit according to the directive 2002/96/EC of 27 January 2003 on waste electrical and electronic equipment (WEEE). CAUTION Violation against existing law. ∅ Do NOT dispose of BINDER devices at public collecting points. � Have the device disposed of professionally at a recycling company that is certified ac- cording to conversion of the directive 2002/96/EC into national law. or � Instruct the distributor who sold you the device to dispose of it. The agreements apply that were agreed with the distributor when purchasing the unit (e.g. his general terms of payment and delivery). � If your distributor is not able to take back and dispose of the unit, please contact BINDER service. Certified companies disassemble waste (used) BINDER equipment in primary substances for recycling according to directive 2002/96/EC. The devices must be free from toxic, infectious or radioactive sub- stances in order to eliminate any health hazards to the employees of the recycling companies. Prior to handing the unit over to a recycling company, it is the user’s responsibility that it is free from toxic, infectious or radioactive substances. − Prior to disposal, clean all introduced or residual toxic substances from the unit. − Prior to disposal, disinfect the unit from all sources of infection. Be aware that sources of infection may also be located outside the inner chamber. − If you cannot safely remove all sources of infection and toxic substances from the unit, dispose of it as special waste according to national law. − Fill out the contamination clearance certificate (chap. 18) and enclose it with the unit. WARNING Contamination of the device with toxic, infectious or radioactive substances. Danger of intoxication. Danger of infection. ∅ NEVER take a unit contaminated with toxic substances or sources of infection for recy- cling according to directive 2002/96/EC. � Prior to disposal, remove all toxic substances and sources of infection from the unit. � A unit from which all toxic substances or sources of infection cannot be safely removed must be considered as “special” waste according to national law. Dispose of it accord- ingly.

FDL (E2) 03/2011

page 59/68 15.5 Disposal of the unit in non-member states of the EC CAUTION Alteration of the environment. � For final decommissioning and disposal of the unit, please contact BINDER service. � Follow the statutory regulations for appropriate, environmentally friendly disposal. The main board of the safety / paint drying oven includes a lithium cell. Please dispose of it according to national regulations. 16. Troubleshooting Fault description Possible cause Required measures Heating Unit door not properly closed. Completely close unit door. Door gasket defective. Replace door gasket, Set-point temperature is not reached after specified time. LED (3a) (loss of technical venti- lation) not lit. Controller not adjusted. Calibrate and adjust controller. Controller defective. Pt 100 sensor defective. Semiconductor relay defective Contact BINDER service. Chamber heating permanently, set-point not maintained. Controller not adjusted. Calibrate and adjust controller. Safety device has turned off the oven. Limit temperature reached. Safety device class 2 (chap. 12.1) set too low. Allow the oven to cool down the oven and press the “RESET” button. Check temperature set-point and setting of safety device (chap. 12.1). If appropriate, select suitable limit value. Semiconductor relay defective. Controller defective. Chamber doesn’t heat up. LED (3a) (loss of technical venti- lation) not lit. LED (2a) of safety device lit. Controller display off. Fan turning. Safety device defective. Contact BINDER service. Check connection to power supply. No power supply. Check if the main power switch (1) is turned on. Unit without any function. Controller defective. Contact BINDER service. Deviations from the indicated heating-up times. Oven fully loaded. Charge the oven less or consider longer heating-up times. Chamber just turned on, nec- essary fan speed not yet reached. Wait a few minutes. Fan defective. Contact BINDER service. Exhaust air channel blocked. Flap of exhaust air channel defective. Check exhaust air system (customer side). Chamber doesn’t heat up. LED (3a) (loss of technical venti- lation) lit. Intake opening blocked (soiled filter). Change filter.

FDL (E2) 03/2011

page 60/68 Fault description Possible cause Required measures Ventilation No volumetric flow rate for fresh air and forced-air circulation. LED (3a) (loss of technical venti- lation) lit. Fan defective. Contact BINDER service. No volumetric flow rate for fresh air. LED (3a) (loss of technical venti- lation) lit. Intake opening blocked (soiled filter). Change filter. Exhaust air channel blocked. No fresh air or exhaust air flow. LED (3a) (loss of technical venti- lation) lit. Flap of exhaust air channel defective. Check exhaust air system (customer side). No pressure differential signal at the pipe ends The measuring pipe in the exhaust passage blocked. Contact BINDER service. Pressure differential not sufficient to switch the pressure switch Rupture of pressure differential measuring hose. Contact BINDER service. Controller Program duration longer than programmed. Inappropriate tolerances have been programmed. For rapid transition phases, do NOT program tolerance limits in order to permit maximum heating speed. Program stops one section too early. Program line is incomplete. When programming, define the end value of the desired cycle by adding an additional section with a section time of at least one minute (with setting set-point ramp). Programs have been deleted. Change from 2 programs to 1 program or vice-versa When changing, ensure that the programs are no longer needed. The controller returns to Normal Display from any level. No button was pressed for more than 120 sec. Repeat entries, enter the values rapidly. Message RANGE ERROR CH1 in Normal Display in Display 2 Sensor rupture between sen- sor and controller Contact BINDER service. Ramp temperature transitions are only realized as steps. Set-point programming type set to “Step” in the User level (chap. 10). Set the set-point programming type ti setting “Ramp” in the User level (chap. 10). Only qualified service personnel authorized by BINDER must perform repair. Repaired units must comply with the BINDER quality standards. Every fire and explosion in relation to a lacquer dryer must be reported to the employer’s liabil- ity insurance association (for Germany).

FDL (E2) 03/2011

page 61/68 17. Technical description 17.1 Factory calibration and adjustment This unit was calibrated and adjusted in the factory. Calibration and adjustment were performed using standardized test instructions, according to the QM DIN EN ISO 9001 system applied by BINDER (certi- fied since December 1996 by TÜV CERT). All test equipment used is subject to the administration of measurement and test equipment that is also constituent part of the BINDER QM DIN EN ISO 9001 sys- tems. They are controlled and calibrated to a DKD-Standard at regular intervals. 17.2 Definition of usable volume The usable volume illustrated below is calculated as follows: b c c a b a a C B A A, B, C = internal dimensions (W, H, D) a, b, c = wall separation a = 0.1*A b = 0.1*B c = 0.1*C VUSE = (A - 2 * a) * (B - 2 * b) * (C - 2 * c) Figure 16: Determination of the usable volume The technical data refers to the so defined usable volume. Do NOT place samples outside this usable volume. Do NOT load this volume by more than half to enable sufficient airflow inside the chamber. Do NOT divide the usable volume into separate parts with large area samples. Do NOT place samples too close to each other in order to permit circulation between them and thus obtain a homogenous distribution of temperature and humidity. 17.3 Technical data Exterior Dimensions Width mm / inch 835 / 32.87 Height (including feet) mm / inch 800 / 31.50 Depth mm / inch 685 / 26.97 Additional depth of door handle mm / inch 50 / 1.97 Wall clearance rear mm / inch 100 / 3.94 Wall clearance side mm / inch 160 / 6.30 Exhaust duct, outer diameter mm / inch 100 / 3.94 Total steam space volume 1) l / cu.ft. 156 / 5.51 Interior dimensions Width mm / inch 600 / 23.62 Height mm / inch 435 / 17.13 Depth mm / inch 435 / 17.13 Interior volume l / cu.ft 115 / 4.1 Number of racks, chrome-plated (regular / max.) 2/5 Load per rack Kg /lbs 20 / 44 Permitted total load Kg /lbs 50 / 110 Weight of the unit (empty) Kg /lbs 90 / 199

FDL (E2) 03/2011

page 62/68 Temperature data Temperature range, 5 °C above ambient up to °C / °F 300 / 572 Maximum temperature of the heating surface °C / °F 750 / 1382 Maximum permitted oven temperature °C / °F 300 / 572 Temperature fluctuation ± K 0.3 at 70 °C / 158°F ± K 1.5 at 150 °C / 302°F ± K 2.5 Temperature uniformity (variation) at 300 °C / 572°F ± K 4 to 70 °C / 158°F min 7 to 150 °C / 302°F min 17 Heating-up time 2) to 300 °C / 572°F min 44 to 70 °C / 158°F min 1 to 150 °C / 302°F min 3 Recovery time after door was opened for 30 sec. 2) to 300 °C / 572°F min 6 Ventilation data Air change approx. x/min. 3 Air circulation approx. x/min. 40 approx. l/min 400 Minimum volumetric flow rate of exhaust air 3) approx. m3/h 24.0 Air flow rate m/sec 0.8 up to 1.2 Solvent data Highest permitted solvent quantity 4) (at T-180 °C, M-100g/mol, U-40g/m3, K=0,5) g 6.65 Electrical data Housing protection acc. to EN 60529 IP 33 Nominal voltage (+10 %) 50/60 Hz V 230 1 N ~ Nominal current Amp 13.0 Nominal power kW 2.90 Sound-pressure level dB(A) < 70 at 70 °C / 158°F Wh/h 388 at 150 °C / 302°F Wh/h 1098 Energy consumption 5) at 300 °C / 572°F Wh/h 2023 Power plug shock proof plug Over-voltage category acc. to IEC 61010-1 II Pollution degree acc. to IEC 61010-1 2 Legend: 1) If the volume of the dry goods inserted in the dryer exceeds 10% of the total steam area, it must be deducted in the calculation of the total steam volume. 2) Up to 98 % of the set value 3) Procedure: "calculation of the volumetric flow rate through the exhaust duct based on the velocity of flow measured there with a propeller anemometer " 4) At a drying temperature of 180 degrees Celsius and an average molecular weight of the solvent of M=100g/mole. For a different drying temperature, the highest permissible solvent quantity must be re- calculated. The calculation should be made on the basis of the "principles for the calculation of ventila- tion of chamber dryers and continuous dryers" in accordance with industrial standard EN 1539:2009, Appendix B. 5) Use this value for sizing air condition systems. All technical data is specified for units with standard equipment at an ambient temperature of +25 °C / 77°F and a power supply voltage fluctuation of ±10. The temperature data is determined in accordance to BINDER factory standard following DIN 12880, observing the recommended wall clearances of 10 % of the height, width and depth of the inner chamber. All indications are average values, typical for units produced in series. We reserve the right to change technical specifications at any time. If the cabinet is fully loaded, the specified heating up times may vary according to the load.

FDL (E2) 03/2011

page 63/68 17.4 Equipment and Options To operate the safety / paint drying oven, use only original BINDER accessories or accesso- ries / components from third-party suppliers authorized by BINDER. The user is responsible for any risk arising from using unauthorized accessories. Regular equipment Electronically controlled APT.line™ preheating chamber technology In compliance with all safety requirements according to EN 1539: 2009 and GUV-R 500 Kap. 2.28 „Be- treiben von Trocknern für Beschichtungsstoffe““ ("Dryers for coating materials") (for Germany) Multifunction program controller RD3 with digital display Adjustable ramp functions via program editor Temperature safety controller class 2 acc. to DIN 12880:2007 Communication and printer interface RS 422 Fresh-air monitoring with visual and audible alarms and automatic heating shut-off Replaceable fresh-air filter cartridge (fine-particle filter for particle sizes 1 µm to 10 µm, class F6/EU6 acc. to EN 779:2002) FKM door gasket (for max. temp. 200 °C / 392°F) 2 chrome-plated racks Rear exhaust duct, diameter 100 mm / 3.94 inch Options / accessories Access ports with various diameters, with silicone plug Rack, chrome-plated or stainless steel Perforated rack, stainless steel Rack lockings (4 pieces) Reinforced rack stainless steel, with 1 set rack lockings Reinforced inner chamber with 2 reinforced racks Door with window and touching protection Lockable door Silicone door gasket resistant to high temperatures > 200 °C / 392°F. Attention: Above 250 °C / 482°F the gasket will age faster High temperature resistance for inflammable areas (electrical area according to protection type IP 54) Fresh air replacement filter (class F6/EU6 acc. to EN 779:2002) for particles 1µm to 10µm), with alumi- num frame Measurement of air change rate acc. to ASTM D 5374:1993/2005 Additional measuring channel for digital specimen temperature display (with magnet or clip sensor) with analog output 4-20 mA at DIN socket (DIN plug included) Serial printer with interface, for numerical and graphical temperature registration Factory calibration certificate, measurement in the center Extension of factory calibration certificate (additional value) Measuring protocol acc. to DIN 12880:2007 Qualification folder Neutral cleaning agent (liquid concentrate) Stable table on wheels with castors and locking brakes

FDL (E2) 03/2011

page 64/68 17.5 Spare parts BINDER GmbH is responsible for the safety features of the unit only, provided skilled electri- cians or qualified personnel authorized by BINDER perform all maintenance and repair, and if components relating to chamber safety are replaced in the event of failure with original spare parts. The user is responsible for any risks arising from using unauthorized accesso- ries/components. Accessories and spare parts: Description Art. No. Rack, chrome-plated 6004-0003 Rack, stainless steel 6004-0008 Perforated rack, stainless steel 6004-0030 Rack lockings (4 pieces) 8012-0531 Door gasket made of FKM (temperature resistant up to 200 °C / 392 °F) 6005-0030 Door gasket silicone (high temperature resistant) 8012-0499 Controller RD3 with week program timer 5014-0102 Thermostat class 2 30 °C/ 86 °F to 320 °C / 608 °F 5006-0008 Turning knob for thermostat class 2 8009-0004 Protective cover for thermostat class 2 6002-0077 Pilot lamp red 5008-0003 Temperature sensor Pt 100 bend-off 5002-0006 Temperature sensor magnetic head (option specimen temperature display) 5002-0001 Temperature sensor pincer-type head (option specimen temperature display) 5002-0003 Unit foot, black 6002-0006 Protocol printer 8012-0418 Fresh air replacement filter, class F6/EU6 6014-0001 Measurement of air change rate acc. to ASTM D 5374:1993/2005 8012-0195 Calibration certificate 8012-0030 Extension for calibration certificate (additional value) 8012-0022 Measuring protocol acc. to DIN 12880:2007 8012-0156 Qualification folder DL006031 Neutral cleaning agent, 1 kg 1002-0016 Stable table on wheels with castors and locking brakes 9051-0018

FDL (E2) 03/2011

page 65/68 17.6 Dimensions FDL 115 ∅100 795 22 834 685 50 435 602 435 260 320 R A R A Figure 17: Dimensions FDL 115

FDL (E2) 03/2011

page 66/68 18. Contamination clearance certificate Unbedenklichkeitsbescheinigung Declaration regarding safety and health Erklärung zur Sicherheit and gesundheitlichen Unbedenklichkeit The German Ordinance on Hazardous Substances (GefStofV), and the regulations regarding safety at the workplace, require that this form be filled out for all products that are returned to us, so that the safety and the health of our employees can be guaranteed. Die Sicherheit und Gesundheit unserer Mitarbeiter, die Gefahrstoffverordnung GefStofV und die Vorschriften zur Sicherheit am Arbeitsplatz machen es erforderlich, dass dieses Formblatt für alle Produkte, die an uns zurück- geschickt wird. Note: A repair is not possible without a completely filled out form. Ohne Vorliegen des vollständig ausgefüllten Formblattes ist eine Reparatur nicht möglich. • A completely filled out form must be transmitted via Fax (+49 (0) 7462 2005 93555) or by letter in advance, so that this information is available before the equipment/component part arrives. A second copy of this form must accompany the equipment/component part. In addition, the carrier should be notified. Eine vollständig ausgefüllte Kopie dieses Formblattes soll per Telefax (Nr. +49 (0) 7462 2005 93555) oder Brief vorab an uns gesandt werden, so dass die Information vorliegt, bevor das Gerät/Bauteil eintrifft. Eine weitere Ko- pie soll dem Gerät/Bauteil beigefügt sein. Ggf. ist auch die Spedition zu informieren. • Incomplete information or non-conformity with this procedure will inevitably lead to substantial delays in processing. Please understand the reason for this measure, which lies outside our area of influence and will help us to speed up this procedure. Unvollständige Angaben oder Nichteinhalten dieses Ablaufs führen zwangsläufig zu beträchtlichen Verzögerun- gen in der Abwicklung. Bitte haben Sie Verständnis für Maßnahmen, die außerhalb unserer Einflussmöglichkeiten liegen und helfen Sie mit, den Ablauf beschleunigen. • Please print and fill out this form completely. Bitte unbedingt vollständig ausfüllen! 1. Unit/ component part / type: / Gerät / Bauteil / Typ: 2. Serial No./ Serien-Nr.: 3. Details about utilized substances / biological substances / Einzelheiten über die eingesetzten Sub- stanzen/biologische Materialien: 3.1 Designations / Bezeichnungen: a) ____________________________________________________________________________ b) ____________________________________________________________________________ c) ____________________________________________________________________________ 3.2 Safety measures required for handling these substances / Vorsichtsmaßnahmen beim Umgang mit diesen Stoffen: a) ____________________________________________________________________________ b) ____________________________________________________________________________ c) ____________________________________________________________________________

FDL (E2) 03/2011

page 67/68 3.3 Measures to be taken in case of skin contact or release into the atmosphere / Maßnahmen bei Personenkontakt oder Freisetzung: a) ____________________________________________________________________________ b) ____________________________________________________________________________ c) ____________________________________________________________________________ d) ____________________________________________________________________________ 3.4 Other important information that must be taken into account / Weitere zu beachtende und wichtige Informationen: a) ____________________________________________________________________________ b) ____________________________________________________________________________ c) ____________________________________________________________________________ 4. Declaration on the risk of these substances (please checkmark the applicable items) / Erklärung zur Gefährlichkeit der Stoffe (bitte Zutreffendes ankreuzen) : � 4.1 For non toxic, non radioactive, biologically harmless materials / für nicht giftige, nicht radioak- tive, biologisch ungefährliche Stoffe: We hereby guarantee that the above-mentioned unit / component part… / Wir versichern, dass o.g. Gerät/Bauteil... � Has not been exposed to or contains any toxic or otherwise hazardous substances / weder giftige noch sonstige gefährliche Stoffe enthält oder solche anhaften. � That eventually generated reaction products are non-toxic and also do not represent a hazard / auch evtl. entstandene Reaktionsprodukte weder giftig sind noch sonst eine Gefährdung darstellen. � Eventual residues of hazardous substances have been removed / evtl. Rückstände von Gefahrstoffen entfernt wurden. � 4.2 For toxic, radioactive, biologically harmful or hazardous substances, or any other hazard- ous materials / für giftige, radioaktive, biologisch bedenkliche bzw. gefährliche Stoffe oder anderweitig ge- fährliche Stoffe. We hereby guarantee that … / Wir versichern, dass … � The hazardous substances, which have come into contact with the above-mentioned equip- ment/component part, have been completely listed under item 3.1 and that all information in this re- gard is complete / die gefährlichen Stoffe, die mit dem o.g. Gerät/Bauteil in Kontakt kamen, in 3.1 aufgelistet sind und alle Angaben vollständig sind. � That the unit /component part has not been in contact with radioactivity / das Gerät/Bauteil nicht mit Radioak- tivität in Berührung kam 5. Kind of transport / transporter / Transportweg/Spediteur: Transport by (means and name of transport company, etc.) Versendung durch (Name Spediteur o.ä.) ___________________________________________________________________________________ Date of dispatch to BINDER GmbH / Tag der Absendung an BINDER GmbH: ___________________________________________________________________________________

FDL (E2) 03/2011

page 68/68 We hereby declare that the following measures have been taken / Wir erklären, dass folgende Maßnah- men getroffen wurden: � Hazardous substances were removed from the unit including component parts, so that no hazard exists for any person in the handling or repair of these items / das Gerät/Bauteil wurde von Gefahrstoffen be- freit, so dass bei Handhabung/Reparaturen für die betreffenden Person keinerlei Gefährdung besteht � The unit was securely packaged and properly identified / das Gerät wurde sicher verpackt und vollständig gek- ennzeichnet. � Information about the hazardousness of the shipment (if required) has been provided to the trans- porter / der Spediteur wurde (falls vorgeschrieben) über die Gefährlichkeit der Sendung informiert. We hereby commit ourselves and guarantee that we will indemnify BINDER GmbH for all damages that are a consequence of incomplete or incorrect information provided by us, and that we will exempt BINDER GmbH from eventual damage claims by third parties./ Wir versichern, dass wir gegenüber BINDER für jeden Schaden, der durch unvollständige und unrichtige Angaben entsteht, haften und BINDER gegen eventuell entstehende Schadenansprüche Dritter freistellen. We are aware that, in accordance with Article 823 of the German Civil Code (BGB), we are directly liable with regard to third parties, in this instance especially the employees of BINDER GmbH, who have been entrusted with the handling / repair of the unit / component. / Es ist uns bekannt, dass wir gegenüber Dritten – hier insbesondere mit der Handhabung/Reparatur des Geräts/des Bauteils betraute Mitarbeiter der Firma BINDER - gemäß §823 BGB direkt haften Name: _______________________________________________________________________ Position/Title: _______________________________________________________________________ Date / Datum: _______________________________________________________________________ Signature / Unterschrift: _______________________________________________________________ Company stamp / Firmenstempel: Equipment that is returned to the factory for repair must be accompanied by a completely filled out contamination clearance certificate. For service and maintenance on site, such a contami- nation clearance certificate must be submitted to the service technician before the start of any work. No repair or maintenance of the equipment is possible, without a properly filled out con- tamination clearance certificate.