BIRD 4314B Operating Instructions Manual
Read Online
Read the BIRD 4314B Operating Instructions Manual online using the built-in PDF viewer.
Text of the manual
� Copyright 2000 by Bird Electronic Corporation
Instruction Book Part Number 920-4314B Rev. C
Thruline ® is a Registered Trademark of Bird Electronic Corporation
Electronic Corporation
Cleveland (Solon) Ohio USA
The following are general safety precautions that are not necessarily related
to any specific part or procedure and do not necessarily appear elsewhere in
this publication.
Keep Away From Live Circuits
Operating personnel must at all times observe normal safety precautions. Do
not replace components inside equipment with the high voltage supply
turned on. To avoid casualties, always remove power.
Do Not Service Or Adjust Alone
Under no circumstances should any person reach into an enclosure for the
purpose of service or adjustment of equipment except in the presence of some-
one who is capable of rendering aid.
Safety Earth Ground
An uninterruptible earth safety ground must be supplied from the main
power source to test instruments. Grounding one conductor of a two conduc-
tor power cable is not sufficient protection. Serious injury or death can occur
if this grounding is not properly supplied.
Shock Hazard
Do not attempt to remove a RF transmission line while power is present. Ra-
diated RF power is a potential health hazard.
Resuscitation
Personnel working with or near high voltages should be familiar with modern
methods of resuscitation.
i
Warning notes call attention to a procedure, which if not correctly
performed could result in personal injury.
CAUTION
Caution notes call attention to a procedure, which if not correctly
performed could result in damage to the instrument.
This symbol appears on the equipment indicating there is
important information in the instruction manual regard-
ing that particular area.
� Note: Calls attention to supplemental information.
4314B Warning Statements
The following warnings appear in the text where there is danger to operating
and maintenance personnel, and are repeated here for emphasis.
WARNING
Exposure to RF power radiation and the possibility of RF shock or
burns exists with some operating conditions. Always be sure to turn off
the transmitter when connecting or disconnecting a wattmeter. Be sure
the transmission line is terminated into a load or antenna. When a
Plug-In Element is removed from the RF line socket, the line section
center conductor is exposed. Do not put fingers or other objects into the
Plug-In Element socket while RF power is applied.
WARNING
Always be sure transmitter power is off before disconnecting the unit
from the transmission line.
ii
4314B Caution Statements
The following cautions appear in the text whenever a procedure, if not prop-
erly followed, could put the equipment in danger of damage and are repeated
here for emphasis.
CAUTION
Do not drop the Thruline Wattmeter or its elements or subject them to
hard blows. The microammeter is shock mounted in the wattmeter
housing, but its delicate mechanism may be damaged by severe impact.
CAUTION
Severe damage to the Plug-In Element or wattmeter can result from
exposure to excessive RF power. Make sure the Plug-In Element
installed in the wattmeter has a sufficiently high wattage rating to
handle the line load when it is first applied to the RF line.
CAUTION
Use reasonable care in handling. Do not drop or subject the wattmeter
or elements to hard blows as accuracy may be impaired or other
damage may result.
CAUTION
Do not apply RF power to the wattmeter in excess of the full scale
range of the element.
CAUTION
When making low reflection readings using a more sensitive element,
take care to insert the element so that it senses reflected power only.
Do not rotate the element in the socket so that it is subjected to
forward power. This can result in damage to the Plug-In Element, the
microammeter, or both.
CAUTION
If other than Female N type connectors are used, limit power and
frequency to the capabilities of the RF coaxial cable or connectors used.
Damage to connectors or errors in reading could result.
Model 4314B Safety Statements
USAGE
ANY USE OF THIS INSTRUMENT IN A MANNER NOT
SPECIFIED BY THE MANUFACTURER MAY IMPAIR
THE INSTRUMENT’S SAFETY PROTECTION.
iii
EL USO DE ESTE INSTRUMENTO DE MANERA NO
ESPECIFICADA POR EL FABRICANTE, PUEDE
ANULAR LA PROTECCIÓN DE SEGURIDAD DEL
INSTRUMENTO.
BENUTZUNG
WIRD DAS GERÄT AUF ANDERE WEISE
VERWENDET ALS VOM HERSTELLER
BESCHRIEBEN, KANN DIE GERÄTESICHERHEIT
BEEINTRÄCHTIGT WERDEN.
UTILISATION
TOUTE UTILISATION DE CET INSTRUMENT QUI
N’EST PAS EXPLICITEMENT PRÉVUE PAR LE FABRI-
CANT PEUT ENDOMMAGER LE DISPOSITIF DE PRO-
TECTION DE L’INSTRUMENT.
IMPRIEGO
QUALORA QUESTO STRUMENTO VENISSE
UTILIZZATO IN MODO DIVERSO DA COME
SPECIFICATO DAL PRODUTTORE LA PROZIONE DI
SICUREZZA POTREBBE VENIRNE COMPROMESSA.
SERVICE
SERVICING INSTRUCTIONS ARE FOR USE BY SER-
VICE -TRAINED PERSONNEL ONLY. TO AVOID DAN-
GEROUS ELECTRIC SHOCK, DO NOT PERFORM ANY
SERVICING UNLESS QUALIFIED TO DO SO.
SERVICIO
LAS INSTRUCCIONES DE SERVICIO SON PARA USO
EXCLUSIVO DEL PERSONAL DE SERVICIO
CAPACITADO. PARA EVITAR EL PELIGRO DE
DESCARGAS ELÉCTRICAS, NO REALICE NINGÉN
SERVICIO A MENOS QUE ESTÉ CAPACITADO PARA
HACERIO.
WARTUNG
ANWEISUNGEN FÜR DIE WARTUNG DES GERÄTES
GELTEN NUR FÜR GESCHULTES FACHPERSONAL.
iv
ZUR VERMEIDUNG GEFÄHRLICHE, ELEKTRISCHE
SCHOCKS, SID WARTUNGSARBEITEN
AUSSCHLIEßLICH VON QUALIFIZIERTEM
SERVICEPERSONAL DURCHZUFÜHREN.
ENTRENTIEN
L’EMPLOI DES INSTRUCTIONS D’ENTRETIEN DOIT
ÊTRE RÉSERVÉ AU PERSONNEL FORMÉ AUX
OPÉRATIONS D’ENTRETIEN. POUR PRÉVENIR UN
CHOC ÉLECTRIQUE DANGEREUX, NE PAS
EFFECTUER D’ENTRETIEN SI L’ON N’A PAS ÉTÉ
QUALIFIÉ POUR CE FAIRE.
ASSISTENZA TECNICA
LE ISTRUZIONI RELATIVE ALL’ASSISTENZA SONO
PREVISTE ESCLUSIVAMENTE PER IL PERSONALE
OPPORTUNAMENTE ADDESTRATO. PER EVITARE
PERICOLOSE SCOSSE ELETTRICHE NON
EFFETTUARRE ALCUNA RIPARAZIONE A MENO CHE
QUALIFICATI A FARLA.
RF VOLTAGE MAY BE PRESENT IN RF ELEMENT
SOCKET - KEEP ELEMENT IN SOCKET DURING OP-
ERATION.
DE LA TENSION H.F. PEAT ÊTRE PRÉSENTE DANS
LA PRISE DE L’ÉLÉMENT H.F. -CONSERVER
L’ÉLÉMENT DANS LA PRISE LORS DE L’EMPLOI.
HF-SPANNUNG KANN IN DER
HF-ELEMENT-BUCHSE ANSTEHEN -ELEMENT
WÄHREND DES BETRIEBS EINGESTÖPSELT
LASSEN.
PUEDE HABER VOLTAJE RF EN EL ENCHUFE DEL
ELEMENTO RF -MANTENGA EL ELEMENTO EN EL
ENCHUFE DURANTE LA OPERACION.
IL PORTAELEMENTO RF PUÒ PRESENTARE
VOLTAGGIO RF - TENERE L’ELEMENTO NELLA
PRESA DURANTE IL FUNZIONAMENTO.
v
About This Manual
This instruction book guides the user through the operation and maintenance
of the Bird Model 4314B RF Directional Thruline Wattmeter.
This instruction book is arranged so that the essential information about
safety is in the front of the book. Reading the Safety Precautions section be-
fore operating the equipment is strongly advised.
The remainder of this manual is divided into chapters and sections. At the
beginning of each chapter, a general overview describes the contents of that
chapter.
Operation
First time users should read Chapter 1 - Introduction, Chapter 2 -
Installation, and Chapter 3 - Theory of Operation, to get an overview of
equipment capabilities. An experienced operator can refer to Chapter 4 - Op-
erating Instructions. All instructions necessary to operate the equipment are
in this chapter.
Maintenance
All personnel should be familiar with preventive maintenance found in Chap-
ter 5 - Maintenance. If a failure should occur, the troubleshooting section will
aid in isolating and repairing the failure.
Parts
For the location of major assemblies or parts, refer to the parts list in Chap-
ter 5 - Maintenance.
Changes To This Manual
We have made every effort to ensure this manual is accurate. If you should
discover any errors, or if you have suggestions for improving this manual,
please send your comments to our factory. This manual may be periodically
updated. When inquiring about updates to this manual refer to the part num-
ber and revision level on the title page.
vi
Table of Contents
Safety Precautions .
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
i
Warning Statements .
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
ii
Caution Statements .
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
iii
About This Manual
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
vi
Introduction .
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
1
Purpose and Function
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
1
Description .
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
1
Performance Characteristics and Capabilities
.
.
.
.
.
.
.
.
.
1
Power Requirements .
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
2
Additional Equipment
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
2
Standard Plug-In Elements .
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
2
Specifications .
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
3
Installation
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
5
Portability .
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
5
Connections.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
5
Theory of Operation .
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
9
Traveling Wave Viewpoint
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
9
Coupling Circuit.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
9
Peak Power Operation.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
10
Component Description
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
14
Operating Instructions
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
17
Zero Adjust Meter
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
17
Plug-In Element Selection and Insertion .
.
.
.
.
.
.
.
.
.
.
17
CW Power Measurements .
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
18
Peak Power Measurements.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
18
Load power
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
18
Determining VSWR .
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
19
Making Low Reflection Readings .
.
.
.
.
.
.
.
.
.
.
.
.
.
22
Testing Lines, Connectors, Filters and Related Components.
.
.
22
Attenuation By Open Or Short Circuit.
.
.
.
.
.
.
.
.
.
.
.
23
Measuring Percentage Of Positive Modulation
.
.
.
.
.
.
.
.
23
Performance Notes .
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
25
Shutdown .
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
26
Maintenance .
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
29
Cleaning
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
29
Preventative Maintenance .
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
29
Line Section Care
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
29
Inspection .
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
30
Bird Model 4314B RF Directional Thruline Wattmeter
Trouble Shooting .
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
30
Troubleshooting Guide.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
31
Battery Test .
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
33
Battery Replacement
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
33
Meter Replacement .
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
34
Instrumentation Module Replacement .
.
.
.
.
.
.
.
.
.
.
.
36
RF Connector Replacement.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
38
Calibration
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
38
Customer Service .
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
40
Replacement Parts .
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
41
Available QC Type Connectors .
.
.
.
.
.
.
.
.
.
.
.
.
.
.
41
The Model 4314B Thruline Wattmeter is an inser-
tion-type RF wattmeter designed to measure power flow
and load match in 50 ohm coaxial transmission lines. It
is intended for use on CW, AM, FM, SSB, TV, and pulse
signals only.
Description
The Model 4314B Thruline Wattmeter is a portable unit
contained in a die cast aluminum housing. The unit is
equipped with a plastic strap for carrying, four rubber
shock feet on the base, and four rubber bumpers on the
back which allow the unit to stand vertical or lie flat in a
horizontal position when used. For additional protection,
the microammeter is specially shock mounted. A slotted
screw is provided on the lower front face of the meter for
zeroing the pointer. The RF line section face below the
meter protrudes slightly from the wattmeter housing with
the Plug-In Element socket in the center. The circuit
module is mounted on the line section inside the housing.
It has an attached phosphor bronze spring contact finger,
which protrudes through a lateral hole and into the
Plug-In Element socket on the RF line section. The finger
has a button on its end which makes connection with the
contacts of the Plug-in Element. The silver-plated brass
RF line section is precision made to provide the best pos-
sible impedance match to the coaxial RF transmission
line in which the wattmeter is inserted. At each end of
the line section are Bird Quick-Change type RF connec-
tors, which may be quickly interchanged with any other
Bird “QC” connectors.
Performance
Character-
istics and
Capabilities
To make measurements, insert a Plug-In Element into
the line section socket and rotate it to either stop. Point
the element arow in the source to load direction of power
flow and the meter will display a reading of forward
power. Rotate the element so that the arrow is pointed in
the opposite direction to obtain a reflected power reading.
With known values of forward and reflected power the
load match can now be determined by calculating the
VSWR or by using the graphs supplied in this manual
(pages 20-21).
1
Setting the front panel toggle switch to peak mode gives
the wattmeter the capability of indicating peak pulse or
envelope power.
Power
Require-
ments
The Model 4314B can operate from two different power
sources. The wattmeter comes supplied with two 9 V al-
kaline batteries which typically provide 20 hours of oper-
ation. AC adapters are also available for use with either
a 115 or 230 Vac line.
Additional
Equipment
The only additional equipment required by the Model
4314B is a Plug-In Element (see table below). Most of
them are designed to read both average and peak power.
Some are designed exclusively for peak reading applica-
tion.
Standard Plug-In Elements
Bird Model 4314B RF Directional Thruline Wattmeter
2
Use
Frequency (MHz)
Power Ranges (Watts)
Average of
Peak
0.45-2.5
2-30
25-60, 50-125, 100-250,
200-500, 400-1000
60-80, 80-95 95-125,
110-160, 150-250, 200-300,
275-450, 425-850, 800-950
60-80, 80-95, 95-150,
150-250, 200-300, 250-450,
400-850, 800-950
950-1260, 1100-1800,
1700-200, 2200-2300
1000, 2.5 kW, 10 kW
50, 100, 250, 500, 1000,
2500, 5000
5, 10, 25, 50, 100, 250, 500,
1000
1 Watt
2.5 Watts
1, 2.5, 5, 10, 25
Peak Only
25-60, 50-125, 100-250,
200-500, 400-1000
950-1260
2.5 kW, 5 kW, 10 kW
50, 100, 250, 500, 2.5 kW,
5 kW
50 ohms nominal
Insertion VSWR
With “N” Connectors
1.05 Max. to 1000 MHz, 1.1 Max.
to 2300 MHz
Connectors
Bird Quick Change “QC” Female N
normally supplied
RF Power Ranges
Average (CW) Mode
100 mW to 10 kilowatts (� 5% of
full scale)
Peak-pulse or envelope-power
mode
100 mW to 10 kilowatts (� 8% of
full scale)
Frequency Range
0.45 to 2300 MHz (either mode)
Pulse Parameters
Minimum duty factor
1.0 x 10-4
Minimum repetition rate
30 pps
Minimum pulse width
0.4 microseconds at 100-2300 MHz
1.5 microseconds at 26-99 MHz
15 microseconds at 2-25 MHz
Weight
3 lb. (1.4 kg)
Dimensions
3-5/8”L x 4”W x 7”H
(92.1 x 102 x 178 mm)
AC Power requirements (using
adapter)1
115 Vac. � 10%, 60 Hz
220 Vac, � 10%, 50Hz
Batteries
Two 9 V alkaline
Battery Life
20 hours of operation, typically
1 Use only adapter supplied by Bird Electronic Corporation
Introduction
3
The Model 4314B is a portable instrument; the housing is
not designed for fixed mounting. A strap is provided for
carrying purposes.
CAUTION
Do not drop the Thruline Wattmeter or its elements or
subject them to hard blows. The microammeter is shock
mounted in the wattmeter housing, but its delicate
mechanism may be damaged by severe impact.
When transporting the Thruline Wattmeter, it is best to
insert the original dust plug, or a Plug-In Element with
the arrow pointed down, in the measuring socket, and se-
cure it with the catch. This will shunt the meter circuit
and serve to protect it by dampening the pointer action
during handling or shipping. Also, set the wattmeter con-
trols to ON and CW modes. In this position, any unneces-
sary battery drain will be avoided. Handle the Plug-In
Elements with care at all times as their calibration could
be disturbed if they are dropped.
WARNING
Exposure to RF power radiation and the possibility of
RF shock or burns exists with some operating
conditions. Always be sure to turn off the transmitter
when connecting or disconnecting a wattmeter. Be sure
the transmission line is terminated into a load or
antenna. When a Plug-In Element is removed from the
RF line socket, the line section center conductor is
exposed. Do not put fingers or other objects into the
Plug-In Element socket while RF power is applied.
Connec-
tions
The Model 4314B Wattmeter is normally supplied with
two Female N type connectors which are of the Bird
Quick-Change “QC” design. Other “QC” connectors are
available as listed in the Replacement Parts list at the
back of this manual.
Quick-Change connectors can be replaced by removing the
four 8-32 screws from each corner of the square flange of
5
Bird Model 4314B RF Directional Thruline Wattmeter
6
1-5/16"
(33.3mm)
3-3/8"
3-5/8”
(92.1mm)
(85.7mm)
6-7/8"
(174.6mm)
4"
(101.6mm)
1-3/4"
(44.5mm)
Female "N" Q.C.
Connectors (2)
(Normally
Supplied)
ON
BATTERY
CW
PEAK
15
80
F.S. = 30 UA.
PN 2080 - 002
W
ELECTRONIC
20
0
BIRD
10
5
30
40
TT
A
THRULINE
60
S
CORP.
R
20
10
40
50
100
20
Figure 1
Model 4314B
Outline
Drawing
the connector and then pulling it straight outward. In-
stall the new connector by reversing this procedure.
Insert the Model 4314B Thruline Wattmeter in coaxial
transmission lines of 50 ohms nominal impedance. The
signal source and load connections can be made to either
side of the wattmeter with no difference in effect. Use a
coaxial transmission line fitted with suitable matching RF
connectors. If cables of other than 50 ohms impedance are
used, a mismatch will occur which could cause serious in-
accuracies in the readings.
CAUTION
Severe damage to the Plug-In Element or wattmeter
can result from exposure to excessive RF power. Make
sure the Plug-In Element installed in the wattmeter
has a sufficiently high wattage rating to handle the
line load when it is first applied to the RF line.
Installation
7
Theory of Operation
Traveling
Wave
Viewpoint
The operation of this wattmeter is based on the travelling
wave concept of RF transmission. As RF power is applied
to a transmission line, there is a forward wave travelling
from the transmitter to the load, and a reflected wave
travelling from the load to the transmitter. The closer the
load is matched to the transmission line, the smaller the
reflected wave will be. To determine the amount of power
dissipated in a load resistor, it is necessary to determine
the wattage of the forward wave and the wattage of the
reflected wave. The difference between the two indicates
load power.
VSWR (Voltage Standing Wave Ratio) has become a
widely used tool in the travelling wave concept. The
standing waves which are produced by transmission line
mismatch cannot be read directly. However, the VSWR
can be easily determined by the use of the VSWR Con-
version Nemographs (figures 5-6, pages 20-21), without
requiring the use of unwieldy and expensive slotted line
equipment. Many users find that the ratio of
watts reflected
watts forward
is more useful than VSWR.
Coupling
Circuit
When the wattmeter is connected into the system, the RF
power is directed through the unit’s line section. The air
line is a short, uniform section which does not impair the
impedance of the RF coaxial line into which it is inserted.
When the Plug-In Element is installed in the socket of
the line section, the RF waves travelling through the line
produce energy in the coupling circuit of the Plug-In Ele-
ment by both inductance and capacitance (figure 2). The
inductive currents will flow according to the direction of
the travelling waves producing them. Therefore, assuming
that the Plug-In Element remains stationary, the coupling
currents produced from the waves of one direction add in
phase, while those produced from the waves of the oppo-
site direction subtract in phase. The additive or “arrow”
direction is assigned to the forward wave.
The element is
designed so that the wave components travelling in the
opposite direction of the arrow will nearly cancel each
other out, making the element highly insensitive to the
reverse wave direction. Because of the highly directional
characteristics of the element, the resultant direct current
9
sensed by the microammeter indicates the power level of
only the RF waves travelling in the arrow direction.
The Plug-In Element is designed so that it can be rotated
180 degrees in its socket in the line section. When it is
rotated, the meter indicates the power in a direction op-
posite that of the initial reading. If the forward direction
power is read first, the reflected direction power will be
read after the Plug-In Element is rotated. The energy re-
sulting from the inductively coupled component of the for-
ward wave will bring about cancellation as described
above.
Peak Power
Operation
The Model 4314B is designed to measure peak power in
addition to average power. When the toggle switch is
placed in peak mode, battery power energizes the ampli-
fier system. If the ac adapter is being used, it energizes
relay (K1) and disengages the batteries (figure 3).
The battery test circuit is energized when the pushbutton
of switch S2 is depressed. If the voltages of the two 9 V
batteries are within the limits necessary to properly oper-
ate the amplifier circuit, the pointer of the meter deflects
beyond the battery test mark on the meter scale. If the
pointer fails to reach the mark, the batteries need re-
placed. Replacement of both batteries simultaneously will
assure that sufficient battery power will be available for
extended periods of time (see Chapter 5, Maintenance).
The amplifier circuit is designed to provide current to the
meter which indicates at a steady state the peak of the
power applied to resistor R5. Resistor R5 exactly matches
the resistance of meter M so that the existing circuit in
the Plug-In Element is loaded exactly the same as during
the average reading mode (figure 2).
When switch S1 is pressed, the dc input which is nor-
mally applied to the meter is interrupted and is applied
instead to resistor R5 and differential amplifier Q1. This
results in a current gain ratio of approximately 100:1.
This output is applied to a resistance bridge consisting of
resistors R6, R7, R8, R9, and R10. Variable resistor R8 of
this bridge permits zero calibration of the amplifier cir-
cuit. Bridge output is applied to differential amplifier U1
which provides a voltage gain of about 1000:1. This ratio
represents the surge of voltage from the amplifier which
charges capacitor C3 and indicates the minimum pulse
duration. Capacitor C3 applies a potential to differential
amplifiers Q2 and Q3 as long as the capacitor remains
Bird Model 4314B RF Directional Thruline Wattmeter
10
charged. These amplifiers provide a massive current gain
ratio, but unity or slightly less voltage gain. This is ap-
plied to meter M to indicate the peak power in the line.
A portion of the dc output of amplifiers Q2 and Q3 is fed
back to resistor R4 and to amplifier Q1. Resistor R4 is
matched to resistor R5 to provide unity gain to the am-
plifier. This feedback circuit maintains the output from
Q1 and U1 to keep capacitor C3 charged to that value
which yields unity closed loop gain.
The capacitor continues to energize Q2 and Q3, to main-
tain the reading of the meter even though the peak of
the pulse is no longer applied to the amplifier assembly
input. Only the peak power is indicated even though
there is a wide fluctuation of input power.
Because of capacitor leakage, diode back resistance, and
transistor input current, there will be a decay in the cir-
cuit that limits the time the amplifier system will retain
its output level. As the circuit decays, the meter will re-
turn to zero if no additional pulses are received at resis-
tor R5.
11
Theory of Operation
Bird Model 4314B RF Directional Thruline Wattmeter
12
RF Coaxial Line
XMTR
or
LOAD
LOAD
or
XMTR
Directional Coupling
Diode
Detector Element
Meter
DC Contact
Circuit Board
Figure 2
Schematic
Diagram -
Element
Indicates in watts the transmission line power
flow sensed by a Plug-In Element. Contains
three scales which correspond to the full scale
ratings of various Plug-In Elements.
2
ON/BATTERY
Switch
When the switch is in the ON mode, CW or
peak power will be indicated on the meter.
With switch in BATTERY mode, the current
state of the battery will be indicated on the me-
ter.
3
CW/PEAK Switch
When switch is in the CW mode, average power
will be indicated on the meter. With switch in
the PEAK mode, peak pulse or peak envelope
power is indicated on meter.
4
Peak LED
Indicates wattmeter is in PEAK mode and
power is supplied to the peak sensing circuitry.
5
Plug-In Element
Determines whether forward or reflected power
is being sensed. When arrow points to load, the
instrument senses forward power. When arrow
points to transmitter, the instrument senses re-
flected power.
6
Meter Zero Adjust
Provides a zero adjustment for the meter.
7
DC Jack
Input connector for supplying power to peak
sensing circuit when the wattmeter is used in
conjunction with ac adapter.
Bird Model 4314B RF Directional Thruline Wattmeter
14
Theory of Operation
Thruline
WATTMETER
MODEL 4314B
SERIAL XXXX
BIRD ELECTRONIC CORPORATION
CLEVELAND (SOLON), OHIO
DANGER
RF VOLTAGE
KEEP ELEMENT IN LINE
SOCKET DURING OPERATION
15
80
F.S. = 30 UA.
PN 2080 - 002
W
ELECTRONIC
20
0
BIRD
10
5
30
40
TT
A
THRULINE
60
S
CORP.
R
20
10
40
50
100
20
ON
BATTERY
CW
PEAK
1
2
3
4
5
6
7
18V DC
20 mA
Figure 4
Component
Diagram
Before taking any readings with the wattmeter, it is nec-
essary to zero the meter under no-power conditions.
Rotate Plug-In Element so that the arrow points down.
Set wattmeter’s controls to ON and CW modes.
Using a small screwdriver, turn the meter’s zero adjust
screw clockwise or counterclockwise so that the meter
pointer aligns with the zero mark on the meter scale.
WARNING
Exposure to RF power radiation and the possibility of
RF shock or burns exists with some operating
conditions. Always be sure to turn off the transmitter
when connecting or disconnecting a wattmeter. Be sure
the transmission line is terminated into a load or
antenna. When a Plug-In Element is removed from the
RF line socket, the line section center conductor is
exposed. Do not put fingers or other objects into the
Plug-In Element socket while RF power is applied.
CAUTION
Use reasonable care in handling. Do not drop or subject
the wattmeter or elements to hard blows as accuracy
may be impaired or other damage may result.
CAUTION
Do not apply RF power to the wattmeter in excess of
the full scale range of the element.
Plug-In
Element
Selection
and
Insertion
The elements determine the power range to be read on
the meter scale. The major markings (50 W, 100 W, etc.)
are the full scale power value for that element. Elements
are also marked for frequency range. The transmitter fre-
quency must be within the band of the element used.
RF power measurements are made with the insertion of a
Plug-In Element. Insert the element into the line section’s
receptacle. Forward power is indicated when the arrow on
the element points in the direction of power flow; from
17
transmitter to load. Reflected power measurements are
made with the element rotated 180� and the arrow point-
ing towards the transmitter. The element must be rotated
fully when power measurements are being made. The in-
dex pin protruding from the element’s body should rest
against the stop on the line section in both the forward
and reflected position. The small catch in the lower left
hand corner of the casting face should press down on the
shoulder of the Plug-In Element. This will keep the
Plug-In Element in proper alignment and insure good
contact with the dc pickup, the lower edge of the element
and the line section body.
CW Power
Measure-
ments
Set the wattmeter controls to the ON and CW modes,
and read the appropriate meter scale corresponding to the
element full-scale value.
Peak Power
Measure-
ments
Set the ON/BATTERY switch to the BATTERY mode. If
the pointer of the meter indicates below the battery test
zone, the batteries must be replaced. (see chapter 5,
Maintenance) Return ON/BATTERY switch to the ON
position. Set CW/PEAK switch to PEAK mode, and read
the appropriate meter scale corresponding to the element
full-scale value.
Load power
Where an appreciable amount of power is reflected, as
with an antenna, it is necessary to subtract the reflected
from the forward power to obtain the effective load
power. Power delivered to and radiated by an antenna is
given by:
Wl = Wf - Wr
Where: Wl = load power
Wf = forward power
Wr = reflected power
A load closely matched to 50 ohms resulting in a VSWR
of 1.2 or less, will require less than a one percent correc-
tion. A good RF load resistor, such as a Bird Termaline
Load Resistor, will produce this sort of negligible or un-
readable reflected power.
The Model 4314B Thruline Wattmeter used with a Bird
Termaline Load Resistor of proper power rating forms a
highly useful absorption wattmeter. Since the reflected
power will be negligible, it will be unnecessary to rotate
the element from the forward direction.
Bird Model 4314B RF Directional Thruline Wattmeter
18
The Model 4314B Thruline Wattmeter is not designed to
provide direct VSWR readings. In most cases, operators
find that the ratio of forward to reflected power is of
equal use. However, VSWR readings can be determined
very easily by the use of the provided graphs as follows:
Determine the forward and reflected power as described
above.
Using the appropriate VSWR Conversion Nomograph
(pages 20-21), determine the intersection of the forward
and reflected power values. The slanted line passing clos-
est to this point indicates the VSWR.
19
Operating Instructions
Bird Model 4314B RF Directional Thruline Wattmeter
20
10
15
20
0.5
1.0
1.5
2.0
2.5 3.0
4.0
5.0
25 30
40
50
100
150
200
300
400 500
.2
.4
.6
.8
1.0
1.2
1.4
1.6
1.8
2.0
4
6
8
10
12
14
16
18
20
VSWR=
1+
Reflected Power
Forward Power
Reflected Power
Forward Power
1
FORWARD POWER - WATTS
REFLECTED POWER - WATTS
3.0
2.5
2.0
1.6
1.5
1.4
1.3
1.25
1.20
1.16
1.14
1.12
1.10
1.09
1.08
1.07
1.06
1.05
1.8
10
4.0
VSWR
Figure 5
VSWR Conversion Nomograph
FORWARD POWER - WATTS
REFLECTED POWER - WATTS
10
4.0
VSWR
Figure 6
VSWR Conversion Nemograph
It is sometimes desirable to make more accurate low re-
flection readings. This can be done using a more sensitive
element, provided that care is taken to prevent applica-
tion of high forward power to the low-power element.
This procedure is limited to use with higher power trans-
mitters only.
Measure watts forward, using the proper Plug-In Ele-
ment. Reverse the arrow direction of the element to de-
termine the general level of reflected power.
Remove the Plug-In Element and insert a lower power el-
ement which has a maximum rating of more than the re-
flected power indicated in the first measurement.
Carefully insert the element so that it reads reflected
power only.
CAUTION
When making low reflection readings using a more
sensitive element, take care to insert the element so
that it senses reflected power only. Do not rotate the
element in the socket so that it is subjected to forward
power. This can result in damage to the Plug-In
Element, the microammeter, or both.
Read the reflected power on the meter in the usual man-
ner.
When using the two-element method of reading low re-
flected power, we do not recommend using a pair of ele-
ments which has a full scale power differential of greater
than 10 to 1.
Testing
Lines,
Connectors,
Filters and
Related
Components
Lines, connectors, filters, and related components can be
tested using the Thruline Wattmeter. The method of test-
ing depends upon the circumstances involved for any par-
ticular test. Some of these tests are listed below:
The standing wave ratio or the reflected-power to
forward-power ratio of a line can be determined by termi-
nating the line with a good load resistor, such as a Bird
Termaline Load Resistor. Proceed as described under de-
termining VSWR. Low reflected power may be measured
as described above.
Line attenuation (power lost as heat in the line) can be
determined by inserting the line of unknown value be-
tween two Thruline Wattmeters. The end of the line must
Bird Model 4314B RF Directional Thruline Wattmeter
22
be terminated by a load resistor. The attenuation of the
line can be determined by comparing readings made at
the two ends. Allowances must be made for normal in-
strument error where very small values of attenuation
are involved. Some correction may be made by direct
rigid connection of the Thruline Wattmeters.
Attenuation
By Open
Or Short
Circuit
Attenuation can also be determined by the open circuit
method. The test set exhibits good equality between for-
ward and reflected readings when the load connector is
open or short circuited. When this is checked on an open
circuit and an open-circuited length of line of unknown
attenuation is connected to the load connector, the ratio
shown is the attenuation in two passes along the line
(down and back). To convert the reading to decibels, di-
vide it in half or double the line length because twice the
line length is being measured (down and back). This mea-
surement must be supplemented with a reflected power to
forward power ratio check or with a dc continuity check
or leakage check, since open circuits or shorts may exist
part of the way along the line.
Attenuation can also be determined as described above by
using a short circuit rather than an open circuit. The
open circuit method is preferred because the initial equal-
ity (forward power to reflected power) is more easily
achieved in an open circuit.
Measuring
Percentage
Of Positive
Modulation
To measure the percentage of positive modulation in an
amplitude modulation system, employ the average and
peak reading characteristics of the unit since:
Percentage pos. mod. = E
E
E
MAX
CARRIER
CARRIER
�
�100
By substitution we get:
Percentage pos. mod. =
P
P
P
P
C
C
�
�
�
���
�
�
�100
Where: PP = peak power as read with CW/PEAK switch
in PEAK mode.
PC = carrier power as read with CW/PEAK
switch in CW mode
Or:
Percentage pos. mod.=
P
P
P
P
P
C
C
C
�
�
�
���
�
�
�100
23
Operating Instructions
Bird Model 4314B RF Directional Thruline Wattmeter
24
Figure 7
Graph Converting Carrier Power and Peak Power to Percent
And by cancellation:
Percentage pos. mod. =
P
P
P
C
�
�
�
���
�
�
1 �
100
After determining average and peak readings, find the in-
tersection of the carrier and peak powers on figure 7 to
determine the percentage of positive modulation.
Performance
Notes
Although the Model 4314B Thruline Wattmeter is
equipped with Quick-Change connectors, it must be re-
membered that the power rating and insertion loss may
be affected if other than “N” type connectors are used.
Power limits must be governed by the type of connector
or transmission line used. For other types of
Quick-Change connectors, see Chapter 5 for the replace-
ment parts list.
When a Model 4314B is used to match a load to a trans-
mitter and a good match is obtained, removing the in-
strument will not cause any change in the conditions. A
well matched 50 ohm load can be placed at the end of a
50 ohm transmission line of any length without altering
conditions at the transmitter.
When the load is not well matched, like an antenna with
a VSWR of 1.5 or 2.0, line length becomes critical since
the length of line between a mismatched load and the
source transforms the impedance of the load as seen at
the source. If the adjustments for maximum power trans-
fer were made with the Model 4314B in place, removing
it shortens the line by four inches plus two connectors.
This still is no cause for concern at low frequencies where
four to five inches is a small fraction of a wavelength. At
higher frequencies; e.g., above 100 MHz, the power out-
put and frequency of the source may be affected.
To insure that the impedance is identical on either side
of ½ wavelength in your transmission line with the Model
4314B either in or out of the line, it is only necessary to
insert or remove a ½ wavelength.
Make up a length of cable which, when added to the
Thruline Wattmeter, equals ½ wavelength at the fre-
quency of measurement. If more than one frequency is in-
volved, then a separate cable is needed for each
frequency. Use figure 8 to determine the cable length re-
quired.
Operating Instructions
25
� Note: When using UHF Plug PL-259, the cable
length is measured from tip to tip of center pin
of plugs.
� When using other connectors, the cable length
is measured from the ends of the outer conduc-
tor of the connectors.
Bird Model 4314B RF Directional Thruline Wattmeter
26
Cable Wavelength Matching Graph
When all measurements are completed, be sure to set the
CW/Peak switch to CW mode and the ON/BATTERY
switch to the ON position. Leaving the unit in other
switch positions will shorten battery life.
WARNING
Always be sure transmitter power is off before
disconnecting the unit from the transmission line.
28
If any of the contacts or line connectors become dirty,
they should be cleaned with a good contact cleaner or dry
cleaning solvent on a cotton swab stick.
It is important to keep the following surfaces clean:
�
Socket bore
�
Element body circumference
�
Bottom rim of the socket in the line section
�
DC contacts on the element
Take care not to disturb the spring finger of the DC con-
tact when cleaning the socket bore.
The spring finger
can be manually adjusted. The button must be pushed
out far enough to make good contact with the element
body but it must not restrict the entry of the element.
Clean the meter and meter housing using a cloth moist-
ened with a mild detergent solution. Do this only when
necessary and take care not to allow water to enter any
of its circuitry as damage may result.
CAUTION
Use reasonable care in handling. Do not drop or subject
the wattmeter or elements to hard blows as accuracy
may be impaired or other damage could result.
Preventative
Maintenance
Care and maintenance required for the 4314B Thruline
Wattmeter is primarily limited to cleaning. Keeping the
Plug-In Element in the socket of the line section as much
as possible serves as an effective seal against the entry of
dust and dirt and to prevent damage to the meter.
An
aluminum dust plug may be used to cover the socket
opening when the element is removed.
Line Section
Care
The RF connectors on the line section should also be pro-
tected against the entry of dust and dirt by keeping them
29
covered when the wattmeter is disconnected from the RF
line.
If there is evidence of contamination inside the line sec-
tion, the reachable portions should be cleaned and the in-
terior carefully blown out.
Do not attempt to remove the
RF center conductor under any circumstances.
Any at-
tempt to do so will damage the assembly
Inspection
Periodic inspections should be performed at six-month in-
tervals.
Inspect the meter for a cracked meter glass.
Inspect wattmeter line section for a damaged or missing
latch and pivot pin assembly.
Inspect all “QC” connectors for bent, broken, or missing
pins.
Trouble-
shooting
Due to its complexity, field repair of the Model 4314B
Thruline Wattmeter is recommended only for certain mal-
functions.
Only those functions within the scope of nor-
mal maintenance are listed.
The following guide contains troubleshooting information
for problems which can occur during normal operation.
Locate the problem, review the possible cause and per-
form the corrective action listed.
If a malfunction is not listed or not corrected by the
listed corrective actions, notify a qualified service center
for further instructions.
Bird Model 4314B RF Directional Thruline Wattmeter
30
No meter indication
(CW or PEAK)
No radio frequency
Turn on transmitter.
Check transmitter op-
erational conditions,
refer to operating in-
struction manual of
equipment used.
Check connections, in-
terconnecting cables,
and auxiliary in-line
equipment.
Arrow on element
pointing in wrong di-
rection
Rotate element so that
arrow points in direc-
tion of RF power flow.
Meter burned out or
damaged
Replace meter.
(page 34)
No pick-up from dc
contact
Clean line section and
element. (page 25)
Defective Plug-In Ele-
ment
Check element. Re-
place element if defec-
tive.
No meter indication
(PEAK only)
Faulty instrumentation
module
Check module. If
faulty, replace instru-
mentation. (page 36)
Dead or low battery
Depress ON/BATTERY
selector switch to the
BATTERY position. If
the battery test indi-
cates weak batteries,
replace them with new
ones. (page 33)
If pointer does not
move from zero during
the battery test, re-
move the batteries and
measure their voltage.
If less than 6.25 volts,
replace the batteries.
(page 33)
31
Maintenance
Erroneous or inaccu-
rate
Low battery
Depress ON/BATTERY
selector switch to the
BATTERY position. If
the battery test indi-
cates weak batteries,
replace them with new
ones. (page 33)
Shorted or opened
cable
Check connections,
in-line equipment, in-
terconnecting cables,
and auxiliary in-line
equipment.
Wattmeter out of
calibration
Check calibration.
(page 36)
Intermittent or incon-
sistent
Faulty transmission
line, line connections,
antenna, or load condi-
tions
Inspect and correct
meter operation or
load.
Dirty line section, dc
contact, and/or ele-
ment
Clean line section and
element. (page 27)
Sticky meter
Replace meter.
(page 34)
High VSWR or high
percentage reflected
power
Bad load or poor load,
antenna, or connectors
Replace connectors.
(page 36)
Shorted or open trans-
mission line
Replace transmission
line.
Foreign material in
line section or in RF
connector bodies
Check for foreign ma-
terial and clean as re-
quired. (page 27)
32
Bird Model 4314B RF Directional Thruline Wattmeter
The Model 4314B Thruline Wattmeter uses two 9 V alka-
line batteries to supply power for the peak circuitry.
To
test these batteries, set the ON/BATTERY switch to the
BATTERY position.
The meter pointer should travel to
the right of the “BAT.TEST” mark.
If the pointer stops
below the mark, replace the batteries.
Battery
Replacement
Remove the back cover by unscrewing the four 8-32 Phil-
lips flat head screws located two on each side of the me-
ter housing near the back edge at the top and bottom of
the sides. Pull the cover straight off.
Rotate the battery in its clip, then pull upward freeing
the battery.
Remove the snap-on battery connector.
Insert the battery into the clip and replace.
Repeat above procedure for the other battery.
33
Maintenance
Inside Back
+
Figure 9
Battery
Replacement
Remove the back cover by first unscrewing the four 8-32
Phillips flat head screws located two on each side of the
meter housing near the back edge at the top and bottom
of the sides.
Pull the cover straight off.
Loosen the two 8-32 nuts on the back of the meter that
secure the meter leads and remove the leads.
Remove the two 10-32 oval head Phillips screws that se-
cure the meter shock ring located on each side of the me-
ter housing just opposite the meter.
Pull the meter and shock ring assembly out of the hous-
ing from the back.
If the three rubber shock mount but-
tons, P/N 4420-098, that the meter rests on are to be
reused, be careful not to lose them.
However, if the re-
placement shock ring, P/N 4410A261, is to be used in-
stead, the buttons may be discarded.
Remove the meter
retaining ring and shock mount from the meter.
Replace the meter. If the replacement shock ring is used,
peel off the white strip covering the adhesive backing.
Carefully adhere the shock ring to the front bezel flange.
Be sure to observe the polarity when replacing the leads
to the meter, black to negative, red to positive.
Zero adjust the meter.
34
Bird Model 4314B RF Directional Thruline Wattmeter
CLEVELAND (SOLON), OHIO
MULTIPY
SWITCH
SETTING
BY
ELEMENT
FACTOR
!
DANGER
RF VOLTAGE
KEEP ELEMENT IN LINE
SOCKET DURING OPERATION
OFF
O.1
0.3
WATTS
THRULINE
Thruline
WATTMETER
MODEL 4314B
SERIAL NUMBER XXX
BIRD ELECTRONIC CORPORATION
CLEVELAND (SOLON), OHIO
+
Figure 10
Meter Replacement
Remove the back as above in meter replacement.
The in-
strumentation module contains the line section and cir-
cuit board chassis as an integral part.
Using a 7/16 wrench, loosen the dress nut on the
CW/PEAK switch. Care should be taken not to scratch
the label.
Remove the wire assemblies from the batteries, meter,
and dc jack. Remove the battery wires (item 4) first at
the printed circuit board by pulling up on the slip-on con-
nector.
Next remove the dc jack wires (item 5). Remove
the meter wires (item 3) also at the printed circuit board.
Care should be taken not to break the wires.
Note the
wire location for ease of reassembly.
Remove the two oval head Phillips screws located on the
front face of the unit on either side of the line section el-
ement port.
Pull the instrumentation module straight out from the
back of the unit.
Replace the connector.
Zero adjust the meter.
36
Bird Model 4314B RF Directional Thruline Wattmeter
If other than Female N type connectors are used, limit
power and frequency to the capabilities of the RF
coaxial cable or connectors used. Damage to connectors
or errors in reading could result.
RF
Connector
Replace-
ment
Remove the four 8-32 pan head screws from the corners
of the connector flange.
Carefully pull the connector straight off.
Replace the connector.
Calibration
Zero adjustment of the meter is the only calibration re-
quired for the 4314B. This should be done any time the
meter pointer is not exactly aligned with the zero mark
on the meter scale when no power is being applied. Zero
adjustment must also be performed after the meter or in-
strumentation module has been replaced.
To assure that the meter reads correctly for both average
and peak operation, apply direct current from a battery
source to the input of the amplifier circuit (figures 5 and
6, pages 20-21). The steady direct-current peak level
should be the same as the average current level because
no peaks exist in this current.
Apply a direct current
and adjust as necessary so that the same reading is at-
tained in both average and peak modes.
Install a voltage-dropping variable voltage dividing poten-
tiometer and a battery in series with the input of the am-
plifier.
Make sure the resistor is set for zero output.
Connect the positive (+) side of the circuit to input and
the negative (-) side to ground.
Install a voltmeter across
the circuit.
Slowly operate the potentiometer to adjust
output to apply approximately 21 millivolts.
Use a suit-
able resistive network.
Check the reading of the meter.
The pointer should devi-
ate or travel about half of full scale.
Set the CW/PEAK
switch to PEAK mode while watching the meter pointer.
There should be no deviation of the pointer from average
to peak reading.
38
Bird Model 4314B RF Directional Thruline Wattmeter
If the pointer reads a different value for peak reading
than for average reading, adjust potentiometer R8 until
the two readings coincide.
39
Maintenance
Calibration
Potentiometer
Access Hole
X
Figure 12
Calibration
Potentiometer
Bird Electronic Corporation maintains a complete repair
and calibration department at our corporate headquar-
ters.
This department is set up to provide our customers
with the best possible service of Bird equipment.
All instruments returned for service must be shipped pre-
paid and to the attention of the Customer Service Group.
Bird Electronic Corporation
30303 Aurora Rd
Cleveland (Solon), OH
44139-2794
Phone: (440) 248-1200
Fax:
(440) 248-5426
40
Bird Model 4314B RF Directional Thruline Wattmeter
Battery cable assembly
4314-116
5
1
DC jack wire assembly
4314-117
6
1
Meter wire assembly
4314-119
7
2
9 V battery
5-1375
8
2
“QC” connectors
(Female N normally
supplied)*
9
1
Housing cover assembly
4314-114
10
1
Dress nut
5-1670
11
1
AC to dc
converter
115 Vac
230 Vac
5-1864
5-1940
*Available QC Type Connectors
N-Female
4240-062
BNC-Male
4240-132
N-Male
4240-063
LT-Female
4240-018
HN-Female
4240-268
LT-Male
4240-012
HN-Male
4240-278
C-Female
4240-100
LC-Female
4240-031
C-Male
4240-110
LC-Male
4240-025
UHF-Female
(SO-239)
4240-050
BNC-Female
4240-125
UHF-Male
(PL-259)
4240-179
7/8" EIA Air
Line
4240-002
Items 10 and 11 not shown in figure 13.
41
Maintenance
Bird Model 4314B RF Directional Thruline Wattmeter
Inside Back
+
1
2
3
4
5
6
7
8
9
Figure 13
Replacement Parts
All products manufactured by Seller are warranted to be free from defects in
material and workmanship for a period of one (1) year, unless otherwise spec-
ified, from date of shipment and to conform to applicable specifications, draw-
ings, blueprints and/or samples. Seller’s sole obligation under these
warranties shall be to issue credit, repair or replace any item or part thereof
which is proved to be other than as warranted; no allowance shall be made
for any labor charges of Buyer for replacement of parts, adjustment or re-
pairs, or any other work, unless such charges are authorized in advance by
Seller.
If Seller’s products are claimed to be defective in material or workmanship or
not to conform to specifications, drawings, blueprints and/or samples, Seller
shall, upon prompt notice thereof, either examine the products where they
are located or issue shipping instructions for return to Seller (transporta-
tion-charges prepaid by Buyer). In the event any of our products are proved
to be other than as warranted, transportation costs (cheapest way) to and
from Seller’s plant, will be borne by Seller and reimbursement or credit will
be made for amounts so expended by Buyer. Every such claim for breach of
these warranties shall be deemed to be waived by Buyer unless made in writ-
ing within ten (10) days from the date of discovery of the defect.
The above warranties shall not extend to any products or parts thereof which
have been subjected to any misuse or neglect, damaged by accident, rendered
defective by reason of improper installation or by the performance of repairs
or alterations outside of our plant, and shall not apply to any goods or parts
thereof furnished by Buyer or acquired from others at Buyer’s request and/or
to Buyer’s specifications. In addition, Seller’s warranties do not extend to the
failure of tubes, transistors, fuses and batteries, or to other equipment and
parts manufactured by others except to the extent of the original manufac-
turer’s warranty to Seller.
The obligations under the foregoing warranties are limited to the precise
terms thereof. These warranties provide exclusive remedies, expressly in lieu
of all other remedies including claims for special or consequential damages.
SELLER NEITHER MAKES NOR ASSUMES ANY OTHER WARRANTY
WHATSOEVER, WHETHER EXPRESS, STATUTORY, OR IMPLIED, IN-
CLUDING WARRANTIES OF MERCHANTABILITY AND FITNESS, AND
NO PERSON IS AUTHORIZED TO ASSUME FOR SELLER ANY OBLIGA-
TION OR LIABILITY NOT STRICTLY IN ACCORDANCE WITH THE
FOREGOING.
DECLARATION OF CONFORMITY
Manufacturer:
Bird Electronic Corporation
30303 Aurora Road
Cleveland, Ohio 44139-2794
Product:
Thruline Wattmeter
Models:
4314B
The undersigned hereby declares, on behalf of Bird Electronic Corporation of
Cleveland, Ohio, that the above-referenced product, to which this declaration relates, is
in conformity with the provisions of the following standards;
EN 55011:1991 - Emissions: (Class B)
IEC 801-2:1984, Part 3 - EMF Immunity
IEC 801-4:1988, Part 4 - Fast Transient/Burst Immunity
These standards are in accordance with EMC Directive (89/336/EEC).
European Standard EN 61010-1:1993 - Part 1:General Requirements
This standard is in accordance with Low Voltage Directive (73/23/EEC), 1973
The technical documentation file required by this directive is maintained at the corporate
headquarters of Bird Electronic Corporation, 30303 Aurora Road, Cleveland, Ohio
44139
�
�
�
�
Bob Gardiner
Director of Quality
Bird Electronic Corporation