truFreeze® Passive Venting Set-up Instructions
1.
Position the console as
desired for procedure and
lock the wheels by pushing
the red lever to either side.
2. Remove the power cord from
the holder and plug into
standard wall power. Ensure
the power cord is connected
to the console.
Make sure the power cord
is dry.
3. Power on the console using
the green switch on the back
of the device.
4. Ensure that the red
emergency stop button is off
(NOT lit).
3
1
6
truFreeze® Passive Venting Set-up Instructions 11. If using the remote, ensure the remote control
has two (2) working AA batteries installed by
selecting the run tab and attempting to adjust
the site counter. Reset site to “1”.
10. Locate the sterile truFreeze catheters.
Make sure the catheters have not passed their expiration dates.
8. Select the “run” screen.
9
9. Confirm that the device has sufficient
liquid nitrogen via the indicator on screen