truFreeze® Passive Venting Set-up Instructions

1. Position the console as desired for procedure and lock the wheels by pushing the red lever to either side. 2. Remove the power cord from the holder and plug into standard wall power. Ensure the power cord is connected to the console. Make sure the power cord is dry. 3. Power on the console using the green switch on the back of the device. 4. Ensure that the red emergency stop button is off (NOT lit). 3 1 6
truFreeze® Passive Venting Set-up Instructions 11. If using the remote, ensure the remote control has two (2) working AA batteries installed by selecting the run tab and attempting to adjust the site counter. Reset site to “1”. 10. Locate the sterile truFreeze catheters. Make sure the catheters have not passed their expiration dates. 8. Select the “run” screen. 9 9. Confirm that the device has sufficient liquid nitrogen via the indicator on screen