ZIROX SGM7.2.4 Manual
Text of the manual
ZIROX Sensoren & Elektronik GmbH
Am Koppelberg 21, 17489 Greifswald, Germany �+49 3834 8309 -00, Fax: -29 E-mail: [email protected]
ZIROX - Oxygen Measuring Systems
SGM7.2.4
SGM7.2.6
Oxygen Measuring Device
For
Reflow Soldering Systems
Oxygen Measuring Device
For
Reflow Soldering Systems
Range: 2 · 10
5 ... 10 vol-ppm Power supply: 100…240 V, 47…63 Hz
Operating hour counter
Indication of maintenance interval for filter change
(depends on pump mode)
Oxygen Monitor SGM7.2.4/7.2.6
Content
4
Table of contents
1
General information .................................................................................... 6
1.1
Introduction ........................................................................................ 6
1.2
Copyright ........................................................................................... 6
1.3
Commonly used symbols ................................................................... 7
2
Application field .......................................................................................... 8
3
Safety regulations ....................................................................................... 9
4
Functional description .............................................................................. 10
4.1
Measuring principle ......................................................................... 10
4.2
General recommendation ................................................................ 11
4.2.1 Gas flow rate ........................................................................ 11
4.2.2 Accuracy of the measurement .............................................. 12
4.2.3 Special features (operating hours counter and maintenance
interval) ........................................................................................... 12
5
Technical data ........................................................................................... 13
5.1
Characteristics ................................................................................. 13
5.2
Mechanical data .............................................................................. 13
5.3
Electrical engineering/electronics .................................................... 13
5.3.1 General data ........................................................................ 13
5.4
Interface data .................................................................................. 14
5.4.1 Pump control ........................................................................ 15
6
Composition of the oxygen monitor SGM7.2.4 ....................................... 16
6.1
General composition ........................................................................ 16
6.1.1 General overview ................................................................. 16
6.1.2 Construction principle of the solid electrolyte sensor ............ 17
6.1.3 Electronic data processing ................................................... 18
6.2
Usage and connection elements ...................................................... 19
6.2.1 Current supply ...................................................................... 19
6.2.2 Front .................................................................................... 19
6.2.3 Rear ..................................................................................... 20
7
Installation and initiation .......................................................................... 21
7.1
Installation conditions ...................................................................... 21
7.2
Set-up of operating state ................................................................. 21
8
Operation and parametrization ................................................................. 23
8.1
Operation......................................................................................... 23
8.1.1 Switch-on and measurement indication ................................ 23
8.1.2 Gas flow adjustment ............................................................. 23
8.1.3 Data monitoring .................................................................... 23
8.1.4 Indication of the end of maintenance interval for carbon filter 23
8.1.5 Status and error messages .................................................. 24
8.2
Parametrization ............................................................................... 25
Oxygen Monitor SGM7.2.4/7.2.6
Content
5
8.2.1 Adjustable parameters ......................................................... 25
8.2.2 Programming menus ............................................................ 26
8.3
Calibration ....................................................................................... 29
8.3.1 Zero calibration .................................................................... 29
8.3.2 Span gas calibration ............................................................. 29
8.3.3 Calibration status ................................................................. 30
8.4
Fault clearance ................................................................................ 30
9
Maintenance, overhaul and storage ......................................................... 31
9.1
General instructions ......................................................................... 31
9.2
Replacement of the equipment fuse ................................................ 31
10 Appendix .................................................................................................... 32
10.1
Activated carbon filter: description and application notes .............. 32
10.1.1 Filter construction ............................................................... 32
10.1.2 Application and operation of the filter ................................. 32
10.1.3 Replacement of activated carbon ....................................... 33
10.1.4 Technical data .................................................................... 34
10.2
EU Declaration of conformity ........................................................ 35
10.3
Warranty conditions ...................................................................... 36
11 Your own notes and remarks ................................................................... 37
Oxygen Monitor SGM7.2.4/7.2.6
1 General information
6
1 General information
1.1
Introduction
This operation manual describes the composition, mode of operation and
use of the oxygen monitor SGM7.2.4 of ZIROX Sensoren und Elektronik
GmbH.
If not particularly marked, the specifications are valid for the SGM7.2.6 as
well.
Address of manufacturer:
ZIROX Sensoren & Elektronik GmbH
Am Koppelberg 21
17489 Greifswald
Germany
Phone:
+49 38 34 830900
Fax:
+49 38 34 830929
E-mail:
[email protected]
www.zirox.de
The manufacturer guarantees that this manual was written in accordance
with the functional and technical features of the delivered SGM7.2.4.
This manual is not subject to the amendment service. If the manufacturer
modifies the SGM7.2.4 with the aim of making technical improvements, the
user is responsible for inserting the additional or updated pages supplied.
Proper operation of the SGM7.2.4 can only be ensured if the contents of this
manual are known. Therefore, all chapters of this manual must be read
carefully prior to operating the SGM7.2.4.
The values on the device display in this manual are examples or preset
parameters of the manufacturer. The user must set process-specific
parameters.
Pages, charts and figures are numbered consecutively.
1.2
Copyright
This operation manual is copyright protected.
It must not be partially or completely reproduced, copied, or distributed,
without prior written permission of the manufacturer. The use for competitive
advantages or the distribution to third parties are not authorized either.
All rights reserved.
Oxygen Monitor SGM7.2.4/7.2.6
1 General information
7
1.3
Commonly used symbols
Symbol for imminent danger:
This symbol refers to imminent danger to persons’ life and health.
In case of disregard, fatal injuries may result.
Symbol for indirect danger:
This symbol indicates indirect danger.
The degree of the damage depends on the circumstances and the
actions of the persons involved.
In case of disregard, destruction or damage of the SGM, its single
components or other material assets as well as minor injuries may
result.
Symbol for proper handling:
This symbol appears where the manual refers to the adherence to
rules, instructions and proper operation.
In case of disregard, damage or destruction of the SGM or its single
components may result.
NOTE
danger
attention
Oxygen Monitor SGM7.2.4/7.2.6 2 Application field
8
2
Application field
The protective gas monitor SGM7.2.4 serves the continuous measuring of
the oxygen concentration in industrial, laboratory and protective gases as
well as in the process of mixing and producing of special forming gases in
industry. The measuring of the free oxygen concentration in inert gases and
also the measuring of bound oxygen in gas compounds is possible.
Designate
use
The main application fields of the oxygen monitor are:
• processes for mixing and producing of forming gases
• production processes of welding and soldering, especially gas-
shielded arc welding in steel and container production as well as plant
construction
• production processes of electronic components under buffer gas.
The introduction of explosive gas compounds, high concentrations of
halogens or sulphuric gases (e.g. SO2) into the SGM7.2.4 is not permitted.
The contact of the SGM7.2.4 with siliconic or phosphoric compounds is not
permitted either.
The SGM7.2.4
Functions
• measures and indicates the oxygen concentration in a measuring gas
continuously;
• indicates any deviations of the oxygen concentration in the measuring
gas from an adjustable set value;
• monitors the course of a particular production process under buffer
gas;
• checks the purity of buffer gases and determines whether the
requested protective effect of the buffer gas is reached.
The requirements and limit values provided in "Technical data" must be
strictly observed.
Any other use is treated as non-authorized use.
Oxygen Monitor SGM7.2.4/7.2.6 3 Safety regulations
9
3 Safety regulations
The following regulations for industrial safety provide basic information about
potential danger during the operation of the buffer gas monitor SGM7.2.4.
Therefore, they must be observed and strictly followed by the responsible
staff.
• A failure-free and functional operating of the SGM can only be
guaranteed with knowledge of this manual. Therefore, all chapters of
this manual must be read carefully before the installation and initiation
of the SGM.
• The SGM is to be used for the functional operation only (see chapter
2).
• The SGM is to be installed, operated, and serviced by trained staff only.
• The SGM is to be connected to an isolated ground socket (Schuko-
socket) with the supplied cable.
Explosive gas compounds, halogens in high concentration, and sulphuric
gases (e.g. SO2) are not permitted to be measured by this SGM7.2.4.
Due to the high operation temperature of the sensor, the SGM produces a
lot of heat.
Make sure the device does not overheat because of covering.
A vertical installation may cause heat accumulation and
damage to the device.
Switch off and disconnect the device from the power supply before opening
the housing cover of the SGM.
Special safety instructions for potential danger during certain working
processes are given in relevant text passages.
Oxygen Monitor SGM7.2.4/7.2.6 4 Functional description
10
4 Functional description
4.1
Measuring principle
In industry and laboratories, the measurement of the oxygen concentration
in gases is often required. Mostly, it is measured in gases, which have a
considerable, temperature-independent oxygen concentration.
The NERNST equation is used as a basis for determining the oxygen
concentration in gases with the oxygen monitor SGM7.2.4.
NERNST
equation
U
RT
F
p
p
O
air
O
meas gas
= 4
2
2
ln
,
,
.
(I)
U – cell voltage in V
R – molar gas constant, R = 8.314 J/(mol · K)
T – measuring temperature in K
F – Faraday-constant, F = 9.648 · 104 C/mol
pO2, air – partial pressure of the oxygen at the reference electrode
in dry air in Pa
pO2, meas.gas – partial pressure of the oxygen at the measuring electrode
in the measuring gas in Pa.
The sensor of the SGM7.2.4 is based on the conductivity of oxide ions in a special
ceramic substance (zirconium dioxide) with stabilizing additions. The conductivity
of these oxide ions increases exponentially with the temperature and reaches a
sufficiently high temperature above 600°C.
The gas to be measured passes through the ceramic oxide ion conductor, which
is a gas-tight tube. The ceramic tube is situated axially symmetrically in a thermally
well-insulated heater. The electrodes of the galvanic sensor are made from
platinum. The electrode on the outside of the tube, surrounded by dry air, is used
as a reference electrode with a constant, known electrode potential. (composition
see chapter 6.1.2)
Oxygen Monitor SGM7.2.4/7.2.6
4 Functional description
11
Based on the assumption that the total pressures of the gases are almost
the same at both electrodes (in this case the volume concentrations may
be used in the calculation instead of the partial pressures) and replacing
the parameters by numbers in equation (I) the following equation applies:
Equation
for oxygen
concentr.
ϕO2= 20.64 · e (-46,42 ·
)
(II)
ϕO2 – oxygen concentration in the measuring gas in vol%
U – potential difference in mV
T – measuring temperature in K
20.64 – oxygen concentration in air with a relative
humidity of 50% in vol%
4.2
General recommendation
The oxygen may be in free or bound form inside the measuring gas1. (see
chapter 10.1)
Thereby, the following dependencies are valid:
– for free oxygen
– for bound oxygen
The equation (II) for calculating the oxygen concentration is valid for
measuring gases with free oxygen as well as for reducing gas compounds in
which oxygen only exists in bound form (e.g. in H2/H2O- or CO/CO2-
compounds).
4.2.1 Gas flow rate
The constant flow rate of the measuring gas is ensured by an internal pump.
It can be chosen between HIGH (35 l/h, maintenance interval runs 3 times
faster than in the LOW mode) and LOW (8 l/h).
1 Different conditions of oxygen in the measuring gas must be distinguished:
Free oxygen: Oxygen molecules in the gas are independent without a bond to other gas components
(inert gases such as N2 or Ar).
Bound oxygen: Free oxygen molecules do not exist in the gas, only in bound form e.g. as water
vapor. Higher temperatures cause a dissociation and oxygen molecules are available. Since the
dissociation degree increases with the temperature, the measurement result depends on the
temperature.
Possibly, free oxygen can react with potential burnable gases at the hot platinum electrode. The result
can be a reducing gas.
U
T
U
T
~
U
T
~ 1
Oxygen Monitor SGM7.2.4/7.2.6
4 Functional description
12
4.2.2 Accuracy of the measurement
The manufacturer guarantees a measuring error of less than 5 % (relative
error) only with measurements of oxygen concentrations within a range of 2
· 105 ... 10 ppm. For measurements of oxygen concentrations of 10 ... 10-3
ppm, the relative error is less than 5 % if the gas inlet tube has no leaks or
permeability.
For measurements of oxygen concentrations less than 10 ppm, the following
aspects must be taken into account during evaluation:
• composition of the measuring gas (e.g. presence of burning gases)
• specific characteristics of the production process (e.g.
material used)
• temperature of the measuring gas.
To reduce the measuring error in low oxygen concentrations, the following
conditions must be provided:
• The measuring gas must be extracted where a formation of layers can
be avoided.
• The tube from the measuring point to the SGM7.2.4 must be as short
as possible in order to avoid a change in the chemical balance in the
tube.
• All gas inlet and outlet tubes must be free of leaks.
• If the measuring gas contains reducing components (e.g. alcohol), the
concentration of free oxygen cannot be determined correctly as
chemical reactions occur at the electrode. In such cases, the
measuring gas should be filtered by an activated carbon filter before
entering the SGM7.2.4 (see chapter 10.1).
4.2.3 Special features (operating hours counter and maintenance interval)
Open the menu 'OPERATING HOURS' by pressing any key.
TOTAL TIME: 1234 h
CHANGE CARBON
FILTER IN: 0 h
MAINT.INTERVAL: 600 h
RESET
USER CODE
RETURN
First, enter the password, then set the time of the preset maintenance interval
to the next filter change (choose line 'RESET' and press 'Enter').
In pump mode HIGH the maintenance interval runs 3 times faster than
in mode LOW!
NOTE
Oxygen Monitor SGM7.2.4/7.2.6 5 Technical data
13
5
Technical data
5.1
Characteristics
Description............................................. Protective gas monitor SGM7.2.4
Application ............................................. Measuring of the oxygen concentration
in inert gases
Measuring range .................................... 10 vol-ppm…20.6 vol% O2
Accuracy at normal pressure ................. Relative measuring error < 5 %
Measuring gas flow rate ......................... 8 resp. 35 l/h (low, high)
Max. permitted pressure of the gas ........ 20 mbar over pressure
Max. permitted temp. of the gas ............. 80 °C
∆p over measuring cell........................... Approx. 1 kPa (100 mm WS) at 10 l/h
Working conditions ................................ 10…45 °C, rH < 80 % at 20 °C
Storage conditions ................................. -20…60 °C, rH < 95 % at 20 °C
Degree of protection .............................. IP40
5.2
Mechanical data
Dimensions ............................................ 135 x 100 x 240 mm
Weight ................................................... 3 kg
Gas input ............................................... Plug connection 6 mm
Gas output ............................................. Tube nipple 4 mm
5.3
Electrical engineering/electronics
5.3.1 General data
Current supply
Voltage ....................................... 100…240 V AC, 47…63 Hz
Current consumption .................. 35 VA
Heater measuring cell................. 24 V DC, approx. 10 W
(controlled internally)
Keypad and display
Keypad .................................................. 3 keys
Text display ........................................... LCD dot-matrix
Oxygen Monitor SGM7.2.4/7.2.6 5 Technical Data
14
5.4
Interface data
Serial interface RS-232
Transfer rate .................................. max. 19200 Baud, adjustable
Stop bits ........................................ 1
Data bits ........................................ 8
Parity ............................................. none
Handshake .................................... without
Pin-configuration connector SUB-D 9-pol. F:
Chart 1:
Serial interface protocol (CR = Carriage Return)
Command
Response
Description
M2CR
M2x.xxExxCR
Oxygen concentration in ppm
A1CR
A1xxxCR
Cell voltage in mV
A2CR
A2xxxCR
Measuring cell temperature in °C
Chart 2:
Error indication
Error code
Description
0
ERROR0
Transmission error
1
ERROR1
Warm-up
2
ERROR2
Cell temperature too low
3
ERROR3
Thermocouple defective
4
ERROR4
Operation hour meter run out!
Change filter or carbon, then reset! (see manual)
5
6
ERROR6
System error
Analog output
SGM7.2.4: Voltage output, 0/2 - 10V, potential-free, range adjustable
SGM7.2.6: Current output, 0/4 – 20 mA, potential-free, range adjustable
Chart 3:
Limit value relay
Charge
Resistive charge (cosϕϕϕϕ = 1)
Max. switching voltage
125 V AC, 60 V DC
Max. switching current
1A
Max. switching power
62.5 VA, 30 W
Min. permissible charge
1 mA at 5 V DC
MD
TXD
RXD
Oxygen Monitor SGM7.2.4/7.2.6 5 Technical Data
15
5.4.1 Pump control
A powerful pump is integrated in the SGM7.2.4, which can be controlled via
interface RS232:
Chart 4:
pump control via RS232
Command from
PC
Reply of SGM7.2.4
Remark
P0CR
P0CR
pump off
P1CR
P1CR
pump low
P2CR
P2CR
pump high
pCR
pXCR
scan pump state
X=0 pump off
X=1 pump low
X=2 pump high
Control input for pump relay: 24 V DC (internal inverse-polarity protection
and free-wheeling diode)
Chart 5:
Pin config. for 7-pol. socket and connection cable with plug
(1.5 m long)
Pin
Description
Cable color
1
limit relay NC contact
blue
2
limit relay NO contact
brown
3
limit relay central contact
green
4
pump control
white
5
pump control
grey
6
output +
pink
7
output -
yellow
Oxygen Monitor SGM7.2.4/7.2.6 6 Composition
16
6 Composition of the oxygen monitor SGM7.2.4
6.1
General composition
6.1.1 General overview
The SGM7.2.4 is a portable device. The general composition is shown in Fig. 1.
Fig. 1:
General composition of the SGM7.2.4
The measuring gas gets into the sensor by means of little overpressure at
the inlet or the optional pump ingests the measuring gas. In the first case,
the optional pressure reducer or needle valve at the gas inlet can be used to
control the gas flow. If the pump is installed, the gas flow is controlled by a
flow control of the pump.
Oxygen Monitor SGM7.2.4/7.2.6 6 Composition
17
6.1.2 Construction principle of the solid electrolyte sensor
1
2
5
6
7
3
4
Measurig Gas
Fig. 2:
Composition of the solid electrolyte sensor
The measuring cell (sensor) consists of a tube made of zirconium
dioxide (2/1) with two platinum electrodes. The measuring electrode
is inside the tube (2/3). The reference electrode (2/4) is located on
the outside of the tube and has a constant electrode potential. The
electrodes and the ceramic tube form a galvanic solid electrolyte
measuring cell.
Measuring cell
(sensor)
In order to gain a higher oxide ion conductivity of the zirconium
dioxide tube, the sensor is heated to 750°C. This also avoids
interfering reactions with combustible components of the measuring
gas at the electrode caused by chemical unbalances. A thermocouple
(2/5) inside the measuring cell determines the actual electrode
temperature. A regulator ensures a constant temperature.
Sensor heater
The heated measuring cell produces thermal energy. Therefore, the
SGM7.2.4 must not be covered.
1
Ceramic tube
2
Ceramic cover of reference
electrode
3
Measuring electrode
4
Reference electrode
5
Thermocouple
6
Connecting wire of
reference electrode
7
Connecting wire of
measuring electrode
Oxygen Monitor SGM7.2.4/7.2.6 6 Composition
18
6.1.3 Electronic data processing
The following block diagram illustrates the data processing.
Fig. 3:
Block diagram of the SGM7.2.4
Sensor
Amplifier for
Cell Voltage
Heater
Control
Microcontroller
Sensor Surveillance
Pow er
Supply
1.0 A
0.8 A
Amplifier for
Thermocouple
Display
Keyboard
DC-Output
Relay Outputs
RS-232
110-230 V/50-60 Hz
Analog out
Oxygen Monitor SGM7.2.4/7.2.6 6 Composition
19
6.2
Usage and connection elements
6.2.1 Current supply
The SGM7.2.4 is firmly connected to the current supply by a connecting
cable.
6.2.2 Front
The indicators and control elements are located on the front of the
SGM7.2.4. Depending on the size, the measuring values are displayed
either in vol% or in vol-ppm.
A green or orange light serves as operating indication depending on the
chosen limit value and actual measuring value. A red light points out
certain operating or warning conditions (see chap. 8.1.4).
A keypad is located on the right side of the front panel. It serves the
programming of the device.
Fig. 4:
Front of the SGM7.2.4
Oxygen Monitor SGM7.2.4/7.2.6 6 Composition
20
6.2.3 Rear
The gas inlet and outlet, the analog output and limit value relay
(connection for pump control relay included), the RS232-Interface, the
pump switch and the power switch are located on the rear panel.
The gas inlet is connected with the carbon filter. The inlet of the filter is
located on the front side.
The line cord is permanently fixed with the device. It cannot be
connected.
Fig. 5:
Rear of the SGM7.2.4
Oxygen Monitor SGM7.2.4/7.2.6 7 Installation
21
7
Installation and initiation
7.1
Installation conditions
The protective gas monitor must be set up in a dry and mainly dust-free
place on a stable, flat surface.
• A socket, preferably as a separate electrical circuit, must be
available in immediate proximity of the installation place, protected
with a 10 A fuse.
• No heat sources or appliances, which produce strong magnetic
fields (e.g. electric motors, transformers), should be put in the
proximity of the installation place.
• The SGM7.2.4 operates in horizontal position.
A vertical placement may cause damage by heat accumulation and is not
permitted.
A liquid entry can lead to severe damage or to the complete
destruction of the device.
7.2
Set-up of operating state
A temperature compensation is required after the transportation of the
SGM7.2.4 from cold surroundings to a site with higher ambient
temperature or humidity. A waiting time of about 2 hours before
switching-on must be considered.
1. Install the SGM7.2.4 in the favored place (see chapter 7.1).
2. Connect point of measurement and places of gas inlet and outlet of
the SGM7.2.4. Pay attention to leak-tightness.
3. If a pressure limit is required, install a needle valve directly in front of
the gas inlet (available from manufacturer of the SGM7.2.4).
4. If the amount of water vapor in the measuring gas is likely to cause a
condensation of water in a cold gas pipe, a water trap must be installed
in front of the gas inlet. (The penetration of water into the hot
measuring cell can destroy it and therefore must be absolutely
prevented.)
5. Connect the SGM7.2.4 to the power supply.
NOTE
NOTE
Oxygen Monitor SGM7.2.4/7.2.6 7 Installation
22
On long transport routes and at unfavorable temperatures the material of
the connecting tubes must rule out any oxygen permeability. The
manufacturer recommends the following materials in dependence on the
measuring conditions:
Material of the
connecting
tubes
Low temperature of measuring gas .................. Thick-wall PVC-hose
Higher temperature of measuring gas
Tygon R 3603 (supplier
e.g. Novodirekt Kehl)
Oxygen concentration of < 100 ppm ................ Stainless steel tube.
Silicone can cause measuring inaccuracies because of its oxygen
permeability. Therefore, the manufacturer advises against a usage.
The measuring gas can also flow through the SGM7.2.4 when it is
off.
NOTE
Oxygen Monitor SGM7.2.4/7.2.6 8 Operation
23
8 Operation and parametrization
8.1
Operation
8.1.1 Switch-on and measurement indication
When the oxygen monitor is in operating state and all lines are connected
according to chapter 7.2, the SGM7.2.4 can be switched on. The
measuring cell reaches its operating temperature of 750 °C after approx.
10 minutes. The current measuring value is displayed but only reaches
the indicated error limits after an hour when the thermal conditions in the
measuring cell are even.
After switch-on, the SGM7.2.4 is in display mode. On the display the
programmed measuring value for output 1 is shown (normally the oxygen
concentration) according to the chosen dimension.
8.1.2 Gas flow adjustment
Two different flows can be adjusted via display (low 8 l/h, high 35 l/h)
For excess pressure of the measuring gas, the manufacturer
recommends the installation of a high-quality needle valve. Appropriate
needle valves can be delivered by the manufacturer of the SGM7.2.4. For
higher pressures of the measuring gas, an additional pressure regulator
is recommended which adjusts a pressure of approx. 100 kPa (1 bar)
excess pressure.
8.1.3 Data monitoring
The SGM7.2.4 can be programmed with a limit value. The relevant
messages are delivered via relay output. The relay is open in active state
(This state is displayed in the status line). The signalization is delayed.
The response time can be adjusted for the data monitoring (limit value
delay time) between 1 and 99 seconds.
8.1.4 Indication of the end of maintenance interval for carbon filter
Before operation, the maintenance interval must be set for an optimum
value in the respective process. 600 hours are preset but on the basis of
experience, a considerably shorter operating time of the filter might be
expected.
Changes and reset of the maintenance interval are password-protected
(only possible via display).
The password in the delivery status is 1234.
NOTE
Oxygen Monitor SGM7.2.4/7.2.6 8 Operation
24
8.1.5 Status and error messages
During the measuring, the functions of the measuring cell are
monitored. In case of faults and/or errors, messages are displayed.
The relay output is activated simultaneously.
Self-check of the
SGM7.2.4
Chart 6:
Status / error messages
Status
Display
Remark
0
OK
1
LIMIT
2
RANGE <<<
No alarm (only SGM7.2.4)!
Several reasons – see chapter
8.4
3
RANGE >>>
4
FLOW <<<
< 8 l/h
5
FLOW >>>
> 35 l/h
6
Change filter!
> 600 h (or after period adjusted
by customer)
7
8
9
WARM-UP
Cell temperature too low
(wait, after approx. 15 minutes
750 °C are reached)
10
CELL TEMP.<<<
Set temp. -10° and > 30 min
11
ERROR THERMOCOP. Thermocouple broken
12
13
14
SELF-TEST FAILED
SGM7.2.4
Status 1, 4...8:
warnings, only alarm relay active
Status 2, 3:
warnings, no alarm by alarm relay
Status 9...14:
failure, alarm relay active and analog output
(voltage signal) 10 V
SGM7.2.6
Status 1...8:
warnings, only alarm relay active
Status 9...14:
failure, alarm relay active and analog output
(current signal) 0 mA
Oxygen Monitor SGM7.2.4/7.2.6 8 Operation
25
8.2
Parametrization
8.2.1 Adjustable parameters
Chart 7:
Overview of the adjustable parameters
Parameter
Range
Remarks
Display
O2: 0…21 vol% or in ppm
Range up to 100 vol% on
request
Analog output
0-10 V or 2-10 V (SGM7.2.4)
0...20 mA or 4...20 mA (SGM7.2.6)
Scale of output
Linear
Logarithmic (Basis 10)
Logarithmic scale
recommended if
measuring value exceeds
several decades.
Suppressed zero
point
0.00001…21 vol% or
0.1…210000 ppm
For optimal resolution at a
special range
Max. measuring
value
21.0 vol% or 210000 ppm,
Higher values on request
This value corresponds to
the terminal value of the
analog signal (e.g. 10 V)
Response time t90
1 ... 99 sec
Formed by arithmetical
average of the measured
value
Limit value
0…99.99 vol%
or 0…999999 ppm
Selectable as lower or
upper limit value with the
signs”>” or ”<”
Delay time of the
limit value
1 ... 99 sec
Time to alarm when limit
value is exceeded
Transmission
capacity of RS-232-
interface
4800, 9600, 19200 Baud
Internal measuring
gas pump
Switch on/off by soft key, switch at
the rear side or RS232
High/Low via keypad or external
controlling (RS232) adjustable
Maintenance interval 0…999 h
Count down, at zero
alarm indication for filter
change
Oxygen Monitor SGM7.2.4/7.2.6 8 Operation
26
8.2.2 Programming menus
The keys below the display refer to the following menus (the current meaning of
the respective key is shown on the display). A definite parameter can be chosen,
changed and finally confirmed with the
-key.
A Main display
OXYGEN:
20.64vol%
WARM UP
Status line
PUMP OFF 750°C
Pump status (on/off by key) and cell temperature
The following status can be visualised by LED:
red flashing:
Error
green:
OK
yellow:
Limit value
(constantly – pump off, slow flashing – pump low, fast flashing – pump high)
B Settings
SETTINGS
OPERATING HOURS
GENERAL
LIMIT
ANALOG OUTPUT
CALIBRATION
RETURN
B.1 Operating hours
TOTAL TIME: 90 h Total operating hours of the device
CARBON FILTER
CHANGE IN: 145 h
Time remaining to filter change
MAINT.INTERVAL: 600 h
Filter change after 600 h, adjustable
RESET
Reset maintenance interval
USER CODE
Delivery status: 1234
RETURN
Oxygen Monitor SGM7.2.4/7.2.6 8 Operation
27
B.2 General settings
GENERAL SETTINGS
LANGUAGE: ENGLISH
English, Deutsch
BAUD RATE: 9600
4800,9600,19200
CONTRAST: 0
+- 9
BEEP: ON
OFF
RETURN
B.3 Limit values
LIMIT VALUE
VALUE: > 206400 ppm 0…999999 ppm resp. 0…99,99 vol%
DELAY TIME: 1sec 0...99sec
RETURN
B.4 Output
ANALOG OUTPUT
VALUE: ppm O2 vol% O2, O2[log10]
RANGE: 0-10 V
2-10V(SGM7.2.4), 0-20 or 4-20 mA (SGM7.2.6)
ZERO POINT: 0 ppm
END MARK: 210000 ppm
AVERAGE FACTOR: 1sec 1-99sec
RETURN
B.5 Calibration
CALIBRATION
CALIBR. ZERO POINT
CALIBR. SPAN GAS
RETURN
Oxygen Monitor SGM7.2.4/7.2.6 8 Operation
28
B.5.1 Calibration zero point
CALIBR. ZERO POINT
MEAS. VALUE: 206400 ppm Current measuring value
ZERO GAS: 206400 ppm
ZERO GAS always 20.64 % (surrounding air)
CALIBR: ZERO: WAIT 5 Status
CALIBR. VALUE: - 4.5 Calibration value *1
RETURN
*1 If this line is activated and key Enter is pressed for approx. 3 sec, the
calibration value will be set to 0.0!
B.5.2 Span gas calibration
CALIBR. SPAN GAS
MEAS. VALUE: 206400 ppm Current measuring value
SPAN GAS: 1000 ppm
Adjustable
CALIBR. SPAN: WAIT 5 Calibration status
CALIBR. VALUE: 1.00 Calibration value *2
RETURN
*2 If this line is activated and key Enter is pressed for approx. 3 sec, the
calibration value will be set to 1.0!
B.5.3 Saving
SAVE VALUES ?
YES
NO
Oxygen Monitor SGM7.2.4/7.2.6 8 Operation
29
8.3
Calibration
For measurements with high accuracy requirements a calibration is highly
recommended (The high stability of the measuring cell requires one
check-up per a year only!).
The device must be in operating state
for a minimum of one hour before the
calibration!
8.3.1 Zero calibration
The zero calibration serves the balance of the offset voltage of the ZrO2-
sensor. The offset is caused by the construction (when the electrode has
a disadvantageous position in the heating field) and can occur due to
aging of the sensor.
Before calibration, the sensor is flushed with ambient air. The gas flow is
set by internal or external pump at the value, which is used for measuring
the measuring gas.
With the zero calibration, the working point in ambient air is calibrated with
20.64 vol.% O2.
First, select “Calibration” and then “Calibration ZERO POINT” via keypad.
Select “CALIBR. ZERO:” and start the calibration by pressing “ENTER”.
After approx. 5 sec the calibration is finished. Leave the calibration by
pressing „RETURN“ and „ENTER“. Save it with „YES“.
8.3.2 Span gas calibration
Certified test gas (oxygen concentration near future measuring
conditions) must flow through the device.
First, select “Calibration” and then “CALIBR. SPAN GAS” via keypad. Set
the test gas oxygen concentration. Select “CALIBR. SPAN:” and start the
calibration by pressing “ENTER”.
The stability of the measuring value is checked during the calibration. The
real calibration process starts as soon as the signal produced by the span
gas is stable. Therefore, the duration of the calibration process can vary
(variation smaller than 1 % in 4 seconds).
If the stability is not given, the calibration is stopped after 60 seconds.
Furthermore, the deviation of measuring value from set value is evaluated.
At zero point calibration +/- 20 mV (cell voltage), at span gas calibration
+/- 20 % of measuring value (cell voltage) are allowed.
For the adjustment the following equation applies: Ucell(korr)=(Ucell+A)*B
Uzell = measured cell voltage
A = cell voltage at zero point
B = factor for final value correction
The calibration is left by „RETURN“ and „ENTER“. The calibration is saved
by „YES“.
Oxygen Monitor SGM7.2.4/7.2.6 8 Operation
30
8.3.3 Calibration status
Chart 8:
Messages calibration status
OK (1.5)
last calibration OK (calibration value)
WAIT ! 5
calibration is running
BREAK
cancelled by pressing key
TIME OUT
stability not reached within 60sec
OUT OF RANGE
range exceeded
FAILED
instrumental error
START
start calibration
8.4
Fault clearance
Chart 9:
Disturbances - causes and clearance
Disturbance
Cause
Clearance
Display off
SGM7.2.4 off
switch on SGM7.2.4
No current supply
Check current supply and cord line
Equipment fuse triggered
Change equipment fuse
Warning:
FLOW <<< (Flow too low)
Gas supply blocked, too long for
chosen cross section or leaky
Check gas supply
Remove blockages
Establish tightness
Pump defective
Replacement by manufacturer
Change filter!
Maintenance interval zero
Change carbon filter, reset interval
Relatively high measuring
value, although a low value
for the oxygen concentration
is expected
Gas flow too low
Increase flow rate
Micro leaks in the gas supply
Check fittings
Measurement is dependent
on the throughput (the
smaller the flow, the higher
the value or vice versa)
Leaks in the measuring gas supply
Check gas supply and fittings
Measuring value is
significantly lower than
expected
Reacting components (at high
temperatures with oxygen) in the
measuring gas (e.g. hydrocarbons)
Check activated carbon filter for
saturation
Warning: “WARM-UP”
Measuring cell has not reached
operating temperature yet
Wait for 5 minutes
Check current temperature on
display
Fuse for sensor heater triggered
Switch off, check after switching on
if fault appears again – in this case
consult service
Heater or regulator defective
Consult service
Warning: ERROR
THERMOCOP.
Thermocouple defective
Consult service
Warning: SELF TEST
FAILED (system error)
Error program- or data memory
Consult service
Oxygen Monitor SGM7.2.4/7.2.6 9 Maintanance and Storage
31
9 Maintenance, overhaul and storage
9.1
General instructions
The electronics and the measuring cell are maintenance-free.
In case of defects in measuring cell or thermocouple, send the SGM to
the manufacturer for an overhaul.
If not in use, the SGM has to be stored in the original packing (if possible)
in a dry, dust-free room.
storage
9.2
Replacement of the equipment fuse
Turn the SGM off and separate it from the net supply before
replacing the equipment fuse.
The fuse (1 AT) is located inside the device (see arrow). It has to be
replaced by the same type.
After switch-off the housing of the measuring cell keeps a
temperature of approx. 60 °C for a while.
danger
Oxygen Monitor SGM7.2.4/7.2.6 10 Appendix
32
10 Appendix
10.1 Activated carbon filter: description and application notes
10.1.1 Filter construction
The figure above shows the carbon filter. Two caps with tube connections
close the acryl glass container containing the activated carbon. Both caps
are inserted into the tube (sealed by O-seal). The fittings (tube connection
with seal ring) have closing stoppers to keep the carbon inside. The gas way
is protected against carbon particles with a pre-filter and a fine filter (2 filter
paper discs) in the outlet of the carbon filter.
10.1.2 Application and operation of the filter
Organic components of the test gas (e.g. alcohols) are adsorbed by the
carbon. After longer usage the cell voltage or the oxygen concentration can
drop unexpectedly. This is a sign for saturation of the filter. The filter or the
carbon filling must be replaced (see 10.1.3).
Oxygen Monitor SGM7.2.4/7.2.6 10 Appendix
33
10.1.3 Replacement of activated carbon
Please follow the instructions for replacing the activated carbon.
1) Separate gas connections of the filter at gas inlet and outlet
2) Loosen all four M3-screws which connect filter and housing
3) Remove cover on the inlet side from acryl glass tube by twisting it
slightly
4) Dispose of spent carbon
5) Remove cover on the outlet side by twisting it slightly
6) Remove paper filter and fleece from connections in the caps with
tweezers
7) Insert 2 fleece discs with tweezers into the tube connections on the gas
inlet side
8) Insert 3 fleece and 2 filter paper discs on the gas outlet side (fleece-
paper-fleece-paper-fleece)
9) Insert cover (gas outlet) into acryl glass tube by twisting it slightly
10) Refill the container with new activated carbon (use funnel) and compress
it by knocking a piece of wood; screwdriver or the like gently on the
housing
11) Fill in the carbon but leave a gap of 1-2 mm to the second cap
12) Attach assembled acryl filter with the M3-screws to the housing
13) Connect gas outlet of the filter with gas inlet of the measuring device
Make sure to insert the fleece discs!
Do not remove the closing stopper in the Swagelok fittings!
Oxygen Monitor SGM7.2.4/7.2.6 10 Appendix
34
10.1.4 Technical data
Weight
Approx. 120 g
Volume
Approx. 100 ml
Duration of operation
Depending on the components and the concentration of
the adsorbed organic components
e.g. approx. 1…3 months at raw gas from a fermenting
tank (ZKG)
Carbon used
Granulated or pelletized, 1…3mm
Oxygen Monitor SGM7.2.4/7.2.6 10 Appendix
35
10.2 EU Declaration of conformity
Oxygen Monitor SGM7.2.4/7.2.6 10 Appendix
36
10.3 Warranty conditions
ZIROX Sensoren & Elektronik GmbH warrants that the products manufactured and
sold are free from manufacturing and material defects at the time of dispatch. In
case of defects and faults within 12 months (probe) and 24 months (electronics
assembly) respectively after dispatch, ZIROX will clear faults at its own option by
repair or replacement. The purchaser must give prompt written notice to ZIROX.
The purchaser is not entitled to claim other legal remedies based on this warranty.
ZIROX does not warrant supplied products which are subject to normal wear and
tear (e.g. reference gas pump).
Corrosive gases and solid particles may cause damage and require repair or
replacement due to normal wear and tear.
The contact of the products with explosive gas compounds, halogens in high
concentrations and sulphuric gases (e.g. SO2) is not permitted.
The contact of the products with siliconic or phosphoric compounds is not permitted
either.
A connection of ZIROX and non-ZIROX products voids any warranty claims.
Warranty and warranty claims are only accepted if they are in accordance with the
"General Sales and Delivery Conditions" of the manufacturer.
Warranty and liability claims for damage to persons and/or property are void if they
are subject to the following:
• Normal wear and tear
• Improper use of the product
• Disregard of the manual’s instructions
• Improper installation, initiation, operation and
maintenance of the product
• Operation of the product without protective measures
• Unauthorized functional and technical modification of the product
• Dismantling of parts as well as installation of spare parts or additional
units, which are not delivered or permitted by the manufacturer
• Improper repairs or faulty operation
• External impact
• Acts of God
Attention: When installing the equipment, the customer must ensure that all
necessary supply lines are connected and the operating temperature of the probe
is reached. Experience has shown that products installed but not in use may be
damaged by the process or by external influence. ZIROX will not accept any
responsibility for such damage.
Oxygen Monitor SGM7.2.4/7.2.6 11 Your notes
37
11 Your own notes and remarks