Transcat PROCISION TCL Series Startup Manual

Specifications

Brand: Transcat
Category: Scales
Model: PROCISION TCL Series
File Type: PDF
File Size: —
Pages: 22

Read Online

Read the Transcat PROCISION TCL Series Startup Manual online using the built-in PDF viewer.

Text of the manual

Models: TCL-164 | TCL-224 | TCL-314

Weighing pan x 1 Balance x 1 Power adapter x 1 Draft shield x 1 Grounding foot x 1 Foot x 3 Models: TCL-363 | TCL-603 | TCL-1003 Weighing pan x 1 Balance x 1 Foot x 3 Grounding foot x 1 Models: TCL-2102 Models: TCL-4502 | TCL-6002 Power adapter x 1 Balance ×1 Weighing pan ×1 Draft shield 1 ×1 Power adapter x 1

2. WORKROOM AND BASIC OPERATION GUIDELINES

3. UNPACKING Place the balance away from heat sources. Avoid exposing the balance to the sunlight. Avoid air drafts and air movements at the workstation. Avoid exposing the balance to a magnetic field. Do not weigh magnetic substances. Make sure that the balance is placed on an even surface. Do not place the balance in a hazardous area. Do not weigh explosive materials. Do not place the balance on an unstable ground exposed to shocks and vibrations. +104oF +40 oC +50 oF +10 oC Operate the device in a room where the temperature ranges between 10–40 °C (50–104 °F) and where the relative humidity is below 80%. Place the balance on a solid surface to ensure stability. To obtain stable and repeatable weighing results, an anti-vibration table is recommended. Keep the packaging should a warranty claim or service be required. TOP 1 2 3

Models: TCL-363 | TCL-603 | TCL-1003 | TCL-2102

Models: TCL-363 | TCL-603 | TCL-1003 | TCL-2102 4. ACTIVITIES TO BE DONE PRIOR TO OPERATION 4.1. Remove transport lock Models: TCL-64 | TCL-164 | TCL-224 | TCL-314 Grounding foot Keep the transport lock should a warranty claim or service be required. Keep the transport lock should a warranty claim or service be required. Check the grounding spring. Make sure that the grounding spring juts slightly out of the hole. Check the grounding spring. Make sure that the grounding spring juts slightly out of the hole. 4.2. Grounding spring check Models: TCL-64 | TCL-164 | TCL-224 | TCL-314 1 2

5. COMPONENTS ASSEMBLY

2 3 4 1 2 3 1 2 1 Install the grounding foot in a freely selected opening. Install the grounding foot in a freely selected opening. Models: TCL-2102 Models: TCL-164 | TCL-224 | TCL-314 Models: TCL-64 1 2 3 1 Models: TCL-4502 | TCL-6002 2 Models: TCL-363 | TCL-603 | TCL-1003

6. PREPARING FOR WORK

6.1. Balance temperature stabilization time Prior to switching the device on it is necessary to ensure that it has reached a temperature equal to room temperature. Connect the power adapter to DC connector. Connect the power adapter to the mains. To level the precision balances, turn its feet and keep turning the feet until the air bubble takes the central position. The leveling vial is located at the rear of the unit. Balance temperature stabilization time ranges between 1 - 8 hours. B US 1 USB 2 Ethern To level the semi-micro and analytical balances, turn its feet and keep turning the feet until the air bubble takes the central position. The leveling vial is located in the front of the weighing chamber.

8. BASIC OPERATIONS

2 1 Press key to start the adjustment With unloaded weighing pan, press ����� ������������ ������� ����� ����������������������������� ����������������������������� ������������������������ ������������������������� ������������������������������ �������������������������������� ������������������������������ ����������������������� ����������������������������������������� ��������������������������������������� ��������������������������������������� ������������������������������������� ����� ������������ ������� ����� ���-6� ���-16� ����22� ���-31� ���-363 ���-603 ���-1003 ���-2�02 50 100 200 200 200 500 1000 TCL-4502 TCL-6002 2000 2000 5000 key. Wait for completion of the mass determination process Load thh weighing pan with the requested mass standard and press key Wait for completion of the adjustment. Upon completion, unload the weighing pan

Unload the weighing pan and press

"Zeroing" button. The balance has been zeroed. The balance has been tared. 2 1 When the weighing result has been displayed press "Tare" button. 8.2. Zeroing

keys select "Working mode"

and press key 2 8.4. Working modes accessibility 1 Press to enter "Setup" menu. 4 3 Using keys to select working modes for which accessibility is to be set and press key. In the "Accessibility" position, press key 5 Using keys to set accessibility parameter to "Yes" value for a selected working mode and press key to confirm.

keys, select working mode and

2 8.5. Selecting working mode 8.6. Weighing unit accessibility Using keys, select "Units" and press key 2 Press key to enter working modes menu. 1 1 Press to enter "Setup" menu. key to confirm.

Using keys to select unit for which

accessibility is to be set and press key. "Accessi In the "Accessibility" position, press In the 5 Using keys to set accessibility parameter to "Yes" value for a selected working mode and press key to confirm. 1 Press key to select the weighing units 8.7. Weighing unit selection

9.1. Administrator's login

9. SETTINGS Some balance settings are accessible for Administrator exclusively. Prior balance parameters setup, log in as the Administrator. To enter user parameters, press key 2 1 Press key to enter "Database" menu 4 3 Using keys select "1111" and press key to confirm. Using keys select "Admin" and press key to confirm You have logged in as the administrator. 5

9.2. User settings

Prior balance users setup, log in as the Administrator. 2 1 Press key to enter "Database" menu 4 3 keys select "Admin" and press Using To select "Users" parameters, press key Press key to add new user. 5 Using keys to select "View/Edit" and press key to confirm key to confirm

key to add new user.

9 8 Using keys select "Code" and press key 10 Using press Using press keys enter user code and Using keys select "Password" and press 11 Using press keys enter user password and keys to enter user name and key to confirm key to confirm key to confirm key to confirm

Using keys select "Access Level" and

press key to confirm. 14 Using keys select permissions level and press key to confirm. Using keys select "Language" and press key to confirm 15 Using keys to select menu language press key to confirm.

11.1. General Recommendations

11. DEVICE CLEANING Prior to cleaning it is necessary to disconnect the device from the mains. Remove the plug from the outlet, disconnect all cables Use soft cloth made of either natural or synthetic fiber exclusively. Avoid using cloth made of stiff/hard fabric, this could cause scratches on the surface. Apply the cleaner onto the cloth. Mind not to apply the cleaner directly onto the device. Do not use aggressive cleaning agents (e.g. solvents, chlorine preparations, corrosive chemicals, bleach) Do not use substances of sharp structure or intended for scrubbing Make sure that dust nor liquid gets inside the weighing system. 11.2. Cleaning Instructions GLASS PANES AND COMPONENTS can be cleaned using window cleaner. WEIGHING PANS AND STAINLESS STEEL OR ALUMININUM COMPONETS can be cleaned using a cloth slightly soaked in a solution of water and gentle cleanser (e.g. soap or dishwashing liquid) OPERATION PANEL AND HOUSING can be cleaned using a cloth slightly soaked in a solution of water and gentle cleanser (e.g. soap or dishwashing liquid) Dry sample leftovers can be removed using either a brush or hand held vacuum cleaner. Clean components can be dried using a soft cloth or dust-free paper towel, with this the remaining moisture will be absorbed. Install all the components only when they are completely dry.

11.3. Weighing chamber elements disassembly

Prior to disassembling the weighing chamber, make sure the balance is turned off and unplugged form the mains. Press the moldings on the housing and lift the movable element up. Now you can remove the front glass Press the moldings on the housing and lift the movable element up. Now you can remove the upper glass

Remove the back glass. The

housing has recesses to facilitate removing the glass. Slide out the doors. All disassembled elements are prepared to be cleaned. To reassemble draft shield, follow directions in reverse order.