Memmert SV1524 Operating Manual
Read Online
Read the Memmert SV1524 Operating Manual online using the built-in PDF viewer.
Text of the manual
Shaking device SV1524 | SV3550
Pumping device CPM650
Cooling unit CDP115
Translation of the original operating manual
About this manual
Purpose and target audience
This manual describes the design, function, transport, operation and maintenance of
accessories for WTB water baths. It is intended for use by trained personnel of the
owner, who have the task of operating and/or maintaining the respective accessories.
If you are asked to work on the accessories, read this manual carefully before starting.
Familiarise yourself with the safety regulations. Only carry out work described in this
manual for the respective accessories. If there is something you do not understand, or
certain information is missing, ask your manager or contact the manufacturer. Do not
do anything without authorisation.
Versions
Several accessories of different sizes and functions are available. The functions de-
scribed in this manual refer to the latest firmware version. Due to individual sizes, illus-
trations in this manual may be slightly different to the actual appearance.
Other documents that you must observe:
■
the general operating manual for all water baths WTB
■
for service and repair work, observe the separate service manual
Storage and resale
This operating manual is part of the accessories and must always be stored so
that persons to work with the respective accessories have access to them.
The owner's responsibility is to ensure that persons who work or are to work
with the respective accessories are informed of where these operating manu-
al are located. We recommend that you always keep them in a protected place
near the accessories. Make sure that the operating manual is not damaged by
heat or humidity. If an accessory is resold or transported and then set up again at
a different location, the operating manual must remain with it. The current ver-
sion of this operating manual in PDF format is also available for download from
http://www.memmert.com/
Manufacturer address and customer service
MEMMERT GmbH + Co. KG
Willi-Memmert-Str. 90-96
D-91186 Büchenbach
Germany
Phone: +49 (0)9122 925-0
Fax: +49 (0)9122 14585
E-mail: [email protected]
Internet: www.memmert.com
Customer service:
Hotline: +49 (0)9171 9792 911
Service fax: +49 (0)9171 9792 979
Email: [email protected]
For service enquiries, please always specify the appliance number
given on the name plate.
Terms and signs used ............................................................................................................................................................................ 4
1.2
Product safety and dangers ................................................................................................................................................................... 5
1.3
Requirements of the operating personnel .............................................................................................................................................. 5
1.4
Responsibility of the owner .................................................................................................................................................................... 6
1.1
Applied directives and standards ........................................................................................................................................................... 6
1.5
Changes and alterations ......................................................................................................................................................................... 6
1.6
Behaviour in case of malfunctions and irregularities ............................................................................................................................. 6
1.7
Switch off in an emergency .................................................................................................................................................................... 7
2.� Sloping�and�flat�lids�
8
2.1
Sloping lid ............................................................................................................................................................................................... 8
2.2
Flat lid ................................................................................................................................................................................................... 10
3. Test tube racks
11
4. Shelf reversible
13
5. Shaking Device
15
5.1
Assembly and description .................................................................................................................................................................... 15
5.2
Delivery, Transport and Setting Up ....................................................................................................................................................... 19
5.3
Putting into operation ........................................................................................................................................................................... 23
5.4
Operation and control ........................................................................................................................................................................... 25
5.5
Faults, Warning & Error Messages ....................................................................................................................................................... 28
5.6
Cleaning and decontamination ............................................................................................................................................................. 29
5.7
Repairs and service ..............................................................................................................................................................................30
5.8�
Storage�and�disposal ............................................................................................................................................................................31
6. Pumping device CPM
32
6.1
Assembly and description .................................................................................................................................................................... 32
6.2
Delivery, Transport and Setting Up ....................................................................................................................................................... 36
6.3
Putting into operation ........................................................................................................................................................................... 40
6.4
Operation and control ........................................................................................................................................................................... 41
6.5
Dismantling the hoses .......................................................................................................................................................................... 41
6.6
Faults, Warning & Error Messages ....................................................................................................................................................... 42
6.7
Cleaning and decontamination ............................................................................................................................................................. 43
6.8�
Repairs�and�service ..............................................................................................................................................................................44
6.9
Storage and disposal ............................................................................................................................................................................44
7. Cooling unit CDP
45
7.1
Assembly and description .................................................................................................................................................................... 45
7.2
Delivery, Transport and Setting Up ....................................................................................................................................................... 49
7.3
Putting into operation ........................................................................................................................................................................... 53
7.4
Operation and control ........................................................................................................................................................................... 54
7.5
Dismantling the hoses .......................................................................................................................................................................... 54
7.6
Faults, Warning & Error Messages ....................................................................................................................................................... 55
7.7
Cleaning and decontamination ............................................................................................................................................................. 56
7.8�
Repairs�and�service ..............................................................................................................................................................................57
7.9
Storage and disposal ............................................................................................................................................................................57
8.� Attachment�
58
Terms and signs used
In this manual and on the appliance itself, certain common terms and signs are used to
warn you of possible dangers or to give you hints that are important in avoiding injury
or damage. Observe and follow these notes and regulations to avoid accidents and
damage. These terms and signs are explained below.
1.1.1
Terms used
DANGER
warns of a dangerous situation that leads directly to death or serious (irreversible)
injuries.
WARNING
warns of a dangerous situation that might le
ad to death or serious injuries.
CAUTION
warns of a dangerous situation that might lead to mo
derate or minor injuries.
NOTICE
warns of material damage
1.1.2
Signs used
Symbol
Meaning
Symbol
Meaning
General warning sign - Observe
operating manual
Gases / vapours
Danger of electric shock
Danger of explosion
Flammable substance warning
Caution: hot steam
Caution: hot
Disconnect the mains plug
Observe information in separate
manual
Use gloves
Product safety and dangers
The accessories are technically sophisticated, manufactured using high-quality
materials and tested for many hours in the factory. They reflect the state of the
art and comply with recognised technical safety regulations. However, there are
still risks involved, even when the appliances are used as intended. These are
described below.
DANGER
Danger due to electric shock.
Penetration of liquid into the appliance can cause electric shocks and short circuits.
–
Protect the appliance from splashing water.
–
Switch off the appliance and disconnect the mains plug before cleaning and
maintenance work.
–
The appliance must not be cleaned and disinfected wet. Allow the appliance to
dry completely before putting it back into operation.
WARNING
When loading the appliance with an unsuitable chamber load, poisonous or explosive
vapours or gases may be produced. This could cause the appliance to explode, and
persons could be severely injured or poisoned. The appliance may only be loaded with
materials / test objects that do not emit any poisonous or explosive vapours when
heated up.
WARNING
Depending on operation, the surfaces of the appliance and the chamber load
may still be very hot after the appliance is switched off. Touching these sur-
faces can cause burns. Wear heat-resistant protective gloves or wait until the
appliance cools down before touching.
1.3
Requirements of the operating personnel
The accessories may only be operated and maintained by persons who are of legal age
and have been instructed accordingly. Personnel who are to be trained, instructed or
who are undergoing general training may only work with the appliance under the contin-
uous supervision of an experienced person. Repairs may only be performed by qualified
electricians. The regulations in the separate service manual must be observed.
Responsibility of the owner
The owner of the appliance
■
is responsible for the flawless condition of the appliance and for it being operated in
accordance with its intended use;
■
is responsible for ensuring that persons who are to operate or service the appliance
are qualified to do this, have been instructed accordingly and are familiar with the
operating manual at hand;
■
must know about the applicable guidelines, requirements and operational safety regu-
lations, and train staff accordingly;
■
is responsible for ensuring that unauthorised persons have no access to the appliance;
■
is responsible for ensuring that the maintenance plan is adhered to and that maintenance
work is properly carried out;
■
has to ensure that the appliance and its surroundings are kept clean and tidy, for exam-
ple through corresponding instructions and inspections;
■
is responsible for ensuring that personal protective clothing is worn by operating per-
sonnel, e.g., work clothes, safety shoes and protective gloves.
1.1
Applied directives and standards
EU Declaration of Conformity for the appliance can be found at the end of the document
and online at:
www.memmert.com/de/downloads/downloads/eu-konformitaetserklaerung
Under the REACH regulation, Memmert provides the information on chemical sub-
stances in Memmert appliances online at
https://www.memmert.com/de/reach-rohs/
1.5
Changes and alterations
The accessories may be used only in connection with a Memmert water bath WTB.
The accessories must not be converted or modified without authorisation. No parts
may be added or inserted which have not been approved by the manufacturer.
Unauthorised modifications or changes to the appliance will void any manufacturer's
responsibility regarding the CE conformity of the product.
The manufacturer is not liable for any damage, danger or injuries that result from un-
authorised changes or alterations, or from non-compliance with the provisions in this
manual.
1.6
Behaviour in case of malfunctions and irregularities
The accessories may only be used in a flawless condition. If you, as the operator, notice
irregularities, malfunctions or damage, immediately turn off the accessories and inform
your superior.
Switch off in an emergency
Press the main switch on the respective accessories. This disconnects the appliance
from the power supply at all poles.
Illustration
Accessories
Shaking device - SV
Cooling unit - CDP115
Pumping device - CPM650
Sloping and flat lids
2.
Sloping and flat lids
You may suffer bruising injuries to your hands. Always wear protective gloves. Do not
let the mounted slanted or flat lid of the water bath fall shut.
Note that the slanted or flat lid heats up to the temperature control liquid temperature
during operation.
2.1
Sloping lid
To avoid unwanted evaporation of the temperature control liquid and to achieve the
most even temperature distribution possible, the appliance should always be operated
with the slanted lid closed. The shape of the lid ensures that condensate does not drip
into the chamber load.
Water bath size
Stainless steel sloping lid
WTB6
WTB11
B39844
WTB15
WTB24
B39845
WTB35
WTB50
B39846
CAUTION
CAUTION
Sloping and flat lids
2.1.1
Assembly
1. Loosen the blind plugs at the fastening points (left + right)
for the lid on the water bath.
2. Remove the dummy plug and store it.
3. Put�on�the�fastening�bracket�of�the�cover�and�fix�it�with�screws.
4. Tighten�screws�and�check�that�the�lid�fits�tightly.
Sloping and flat lids
2.2
Flat lid
The flat lid, available as an accessory, can be used to place test flasks on the surface of
the temperature control liquid. By removing or inserting rings, the recesses are adapted
to the piston size. The rings may only be inserted or removed when they have cooled
down.
Water bath size
Stainless steel flat lid
Stainless steel special flat lid
WTB6
WTB11
B48804
B48807
WTB15
WTB24
B48805
B48808
WTB35
WTB50
B48806
B48809
Test tube racks
3.
Test tube racks
The test tube racks differ in the number and diameter of the individual bores. They are
hooked into the water bath from above. The test tube racks may only be inserted or
removed when they have cooled down.
Ø Bores
WTB6 / WTB11
WTB15 / WTB24
WTB35 / WTB50
Ø10mm
34 bores
B48822
58 bores
B48823
74 Bores
B50656
Ø12mm
30 bores
B48824
58 bores
B48825
66 Bores
B50657
Ø14.5mm
26 Bores
B48826
45 Bores
B48827
58 bores
B50658
Ø18mm
22 bores
B48828
38 Bores
B48829
50 Bores
B50659
33 Bores
B48834
57 Bores
B48835
75 Bores
B50662
Test tube racks
Ø Bores
WTB6 / WTB11
WTB15 / WTB24
WTB35 / WTB50
Ø18mm
44 bores
B48836
76 Bores
B48837
100 Bores
B50663
Ø21mm
18 Bores
B48830
34 bores
B48831
44 bores
B50660
Ø32mm
14 Bores
B48832
22 bores
B48833
30 bores
B50661
Maximum loading of the waterbaths with reversible shelf
Size WTB
Maximum Loading
WTB 6/11
25 kg
WTB 15/24
50 kg
WTB 35/50
100 kg
Conversion of reversible shelf
60 mm
1. Reversible shelves can be set to two heights: 60mm and 30 mm
2. Unscrew screws to remove feet
3. Remove feet and rotate depending on desired height (in this case 30mm)
4. Position rotated feet on holders of floor grate and fix with screws
30 mm
5. Reversible shelf now has a height of 30 mm
Assembly and description
5.1.1
Design
Description
Overview
1.
Water bath
1
2
3
4
2.
Recording frame
3.
Shaking Device
4.
Display shaker
5.1.2
Description
Brief technical description
The shaking device enables controlled shaking of test flasks or test tubes in a water
bath. The shaking device performs a back and forth movement with a frequency of ap-
prox. 30 to 150 strokes per minute. The shaking frequency is electronically controlled
and adjustable via touch screen. The same ambient conditions apply for use as for the
basic units.
NOTICE
If the shaking device remains in a switched-on water bath, the shaking device must always be switched on, as in the switched-off
state, the overpressure steam barrier is inoperative, and condensate can enter the shaking device housing.
5.1.3
Intended use
The Memmert shaking device is intended exclusively for a Memmert water bath WTB15,
WTB24, WTB35 or WTB50. It is used for the controlled back and forth movement of
samples at 30 to 150 rpm (strokes per minute), with one of approx. 15mm. The sample
holder is in direct contact with the temperature control medium.
5.1.4
Use not in accordance with the intended purpose
Depending on the loading of the shaking device and the selected shaking frequency,
the temperature control medium and/or the shaking material could spill over. The user's
responsibility is to adjust the shaking frequency or the fill level but without going below
the minimum fill level.
The shaker complies with the current requirements of the RoHS Directive.
You can find more information on this and material compliance on our homepage:
https://www.memmert.com/de/downloads/
Components
Material
Shaking Device
Stainless steel 1.4301 - ASTM 304
Support basket
Stainless steel 1.4301 - ASTM 304
Screws and hinges
Stainless steel 1.4301 - ASTM 304
i The chamber load for the appliance must be carefully checked for chemical compati-
bility with the materials mentioned.
5.1.6
Electrical equipment
■
Operating voltage and current consumption: See name plate or technical data
■
Protection class I, i.e. operating insulation with PE conductor in accordance with
EN 60664-1
■
Protection type IP 20 acc. to EN 60 529
■
Interference suppression acc. to EN 55011 class B
■
Appliance fuse: Fuse F15H250V
5.1.7
Electrical connection
The appliance is intended for operation on a power supply network with a system im-
pedance Zmax at the transfer point (house connection) of a maximum of 0.292 Ohm.
The operator must ensure that the appliance is only operated on power supply network
that meets these requirements. If necessary, you can ask your local energy supply com-
pany what the system impedance is. When connecting, observe the country-specific
regulations (e.g. in Germany DIN VDE 0100 withFI protective circuit).
GmbH+Co.KG D-91126 Schwabach FRG
Äußere Rittersbacher Str. 38
Made in Germany
memmert
Schutzart DIN EN 60529 - IP 20
1
2
3
4
5
6
8
7
Typ: SV1524
~ 230 V 50/60 Hz
F-Nr.: P321.0001
0,1 A 8 W
2.
Operating voltage
3.
Protection type
4.
CE conformity
5.
Address of manufacturer
6.
Disposal note
7.
Connection / power ratings
8.
Appliance number
5.1.9
Technical data – Shaking device
Gerätegröße
15/24
35/50
Dimensions
Length
mm
706
826
Width
mm
297
500
Height
mm
190
190
electrical data
Shaking frequence
Number of
strokes per
minute
35 bis 150
Operating voltage
V
115/ 230
Frequency
Hz
50/ 60
Power
W
8
12
Current
consumption
50/60 Hz
115V
A
0,1
0,1
230V
0,1
0,1
Packaging material
Net weight
kg
6,1
8,7
Gross weight (in cardboard
box)
kg
8,1
11,7
Width
cm
84
98
Height
cm
47
70
Depth
cm
33
33
+5 °C to 40 °C
Air humidity
max. 80 % rh (non-condensing)
Overvoltage category
II
Pollution degree
2
Altitude of installation
max. 2,000 m above sea level
Maximum mains voltage fluctuations
AC 115 (+/– 10 %)
AC 230 (+/– 10 %)
■
The appliance may only be used in enclosed areas and under the ambient conditions
listed below
■
The appliance may not be used in areas where there is a risk of explosion. The sur-
rounding air must not contain any explosive dusts, gases, vapours or gas-air mixtures.
The appliance is not explosion-proof.
■
Heavy dust production or aggressive vapours in the vicinity of the appliance could lead
to sedimentation in the interior and, as a consequence, could result in short circuits or
damage to electrical parts. For this reason, sufficient measures to prevent large clouds
of dust or aggressive vapours from developing should be taken.
5.1.11
Scope of delivery
Standard
■
Shaking Device
■
Recording frame
Optional
■
Support basket
■
Sloping lid
Delivery, Transport and Setting Up
5.2.1
Safety
WARNING
Condensation in the electrical components may cause short circuits. After
transporting or storing the appliance under humid conditions, remove it from its pack-
aging and let it ventilate for at least 24 hours in normal ambient conditions. Do not
connect the appliance to the mains power during this time.
CAUTION
You may suffer crush injuries to your hands when transporting and setting up the
accessories. Always wear protective gloves. Grip the appliance downwards only at the
sides
CAUTION
Danger of cuts due to sharp edges.
Always wear gloves when working on the appliance.
5.2.2
Delivery/unpacking
■
The accessories are packed in a cardboard box
■
Do not unpack the accessories until they are in their installation location to avoid dam-
age
■
Carefully cut open the cardboard packaging along one edge.
5.2.3
Product control
■
Check the delivery note to ensure that the scope of delivery is complete
■
Check the accessories for damage
If you notice deviations from the scope of delivery, damage or irregularities, do not put
the accessories into operation, but inform the carrier and the manufacturer's factory.
5.2.4
Disposing of packaging material
■
Dispose of the packaging material (cardboard, wood, foil) in accordance with the appli-
cable disposal regulations for the respective material in your country
The appliance may fall from the installation location due to unevenness and injure you or
someone else. Always place the appliance on an even, load-bearing and non-slip sur-
face.
CAUTION
The surfaces of the appliance can overheat during operation and overheat heat-sen-
sitive surfaces in close proximity to the appliance. Always place the appliance on a
heat-resistant surface and maintain the specified clearance around the appliance.
The distance between the wall and the rear of the appliance must be at least 15 cm. The
clearance from the ceiling must not be less than 75 cm and the side clearance from
walls or nearby appliances must not be less than 10 cm. Sufficient air circulation in the
vicinity of the appliance must be guaranteed at all times.
10 cm
10 cm
75 cm
10 cm
15 cm
Assembly shaking device
1. Loosen the blind plugs at the fixing points (on the
side) for the shaking device on the water bath.
2. Place the holding frame on the side of the water bath
and fix it with screws.
3. Place the shaking device on the water bath frame from above.
4. Fix the shaking device to the mounting frame with screws.
5. The support basket is inserted into the shaking device
and fixed with knurled screws.
6. Insert the support basket from above into the holders
of the shaking device.
7. Fix the support basket to the shaker with knurled screws.
Putting into operation
5.3.1
Putting into operation for the first time
NOTICE
When putting into operation the accessories for the first time, please do not leave them unattended until they have reached a steady
state.
5.3.2
Connect the appliance to the power supply
Place the power cable so that
■
it does not represent a trip hazard
■
it cannot come into contact with any hot parts
■
it is easily accessible at all times and can be pulled off quickly, for example in case of
interference or an emergency
5.3.3
Filling the water bath
WARNING
If the water bath level is too high, the water may spill over during the operation of
the shaker. This can cause burns. Adjust the level of the water bath to the feed of the
shaker and the shaking frequency.
Please note that the maximum filling level of the water bath is reduced by the attach-
ment of the shaking device and the chamber load
The water bath must be filled so that the liquid level does not exceed the maximum fill-
ing level, including the shaking device. This is marked by the edge on the water bathtub.
MIN
MAX
min. 4 cm
11
15
min. 2 cm
24
35
50
NOTICE
Depending on the loading of the shaking device and the selected shaking frequency, the water bath filling may spill over. Here, the
shaking frequency or the filling level must be adjusted without falling below the minimum filling level.
5.3.4
Switching on appliance
Press main switch on the back of the appliance.
Operation and control
WARNING
During the operation of the shaker, individual components may heat up. Touching its
surface can cause burns. Use temperature-resistant protective gloves.
CAUTION
As the shaker is a moving part, you may suffer crushing injuries while operating the shaker.
Load the basket before starting the shaker, and do not reach into the appliance during
operation.
NOTICE
The mounting frame is only intended for placing the shaker.
Please place no other appliances or parts of appliances on the frame and not support them.
5.4.1
Quick shutdown in case of overload
The appliance has an overload-controlled quick shut-off for safety. You can tell this
safety device has responded when the error code - 139 is shown on the touch display
(see chapter Störungen, Warn- & Fehlermeldungen). Here, it must be checked whether
an overload or an object has been jammed in the device.
After the overload has been removed, the shaker can be restarted by tapping on the
display.
5.4.2
Home screen
CustomView will appear first after switching the appliance on (4.4.4)
To access the home screen, touch any spot on the TouchDisplay
► To set the set shaking frequency, select the set display Set .
► Select the symbol
to set the timer
► To start the shaking frequency, select
.
► Touch the
symbol for other appliance settings.
00:00 h
Set 30 rpm0 rpm
Setting the shaking frequency and time
149
rpm
150
Set the individual values by scrolling vertically on the
touch display:
► Entering the shaking frequency in rpm
► Set time for the timer in minutes/hours
Press
to confirm entry. The setting is not saved with
5.4.4
Settings
Info
CustomView
Other settings are possible under the Settings menu item
:
► Appliance information overview
► CustomView setting
Press
to return to the home screen
5.4.5
CustomView
Under the Settings menu item, select CustomView to set up the individual
display
Selecting the set shaking frequency and timer time activates or deactivates the
respective element.
The selected element will be greyed out when deactivated
00:50 h
Set 30 rpm0 rpm
Tapping the set shaking frequency and timer time activates or deactivates the
respective element.
Set shaking frequency and timer time are deactivated here
confirms the selection;
changes will not be accepted
If the selection was confirmed with
,
tap to return to the
Home screen,
after 30 seconds, the personalised CoustomView appears
5.4.6
Information
SV3550 - V1.0.0 - 20210510
0.0.0
Memmert GmbH + Co. KG
Äußere Rittersbacher Straße 38
91126 Schwabach / Germany
[email protected]
Phone: +49 9122 9250
Under menu item Settings, click on Info to view the item information of the
shaking device
00:50 h
Set 30 rpm0 rpm
100 rpm
Faults, Warning & Error Messages
CAUTION
If the shaking device is used strangely, for example, uncontrolled shaking, you may
suffer bruising injuries to your hands. Disconnect the appliance from the power supply
by pulling out the mains plug.
Description
Cause
Action
Motor lock-up error
■
Restart movement,
■
Reduce load.
E-111
130 rpm
Software error
■
Restart appliance
■
Contact Service
in case of frequent occurrence
E-132
130 rpm
Memory access error
■
Restart appliance
■
Contact Service
in case of frequent occurrence
E-135
130 rpm
CRC error (internal memory error)
■
Restart appliance
■
Contact Service
in case of frequent occurrence
E-139
130 rpm
Cleaning and decontamination
DANGER
Danger due to electric shock.
Penetration of liquid into the appliance can cause electric shocks and short circuits.
–
Protect the appliance from splashing water.
–
Before cleaning and maintenance work, switch off the appliance and disconnect
the mains plug.
–
The appliance must not be cleaned and disinfected wet. Allow the appliance to
dry completely before putting into operation again.
CAUTION
Danger of cuts due to sharp edges.
Always wear gloves when working on the appliance.
5.6.1
Cleaning
Interior and metal surfaces
The metal surfaces of the shaker can be cleaned with commercially available stainless steel
cleaning agents. Make sure that no rusting objects come into contact with the stain-
less steel housing. Rust deposits can lead to an infection of the stainless steel. If rust
spots appear due to contamination, the affected areas must be cleaned and polished
immediately.
Plastic parts
Do not clean plastic parts of the water bath with caustic or solvent-based cleaning
agents.
Glass surfaces
Glass surfaces can be cleaned with a commercially available glass cleaner.
Skin and eye contact with decontaminants may cause irritation of the hands and eyes
or allergic reactions and chemical burns.
–
Wear chemical-resistant gloves during decontamination.
–
For correct application, follow the instructions for use on the respective
container of the decontamination agent
Memmert appliances can become contaminated by loading infectious substances or by
environmental influences at the place of installation. All parts of the affected appliance
must be decontaminated both before sending to our service and before disposing of
the appliance.
■
For decontamination, we recommend commercially available decontaminants based on
isopropanol or ethanol.
If you have any questions about using a decontamination agent for decontaminating
Memmert appliances, please contact our service department.
After cleaning and decontamination, ventilate the installation site and allow the appli-
ance to dry completely.
5.7
Repairs and service
DANGER
Danger due to electric shock.
Penetration of liquid into the appliance can cause electric shocks and short circuits.
–
Protect the appliance from splashing water.
–
Before cleaning and maintenance work, switch off the appliance and disconnect
the mains plug.
–
The appliance must not be cleaned and disinfected wet. Allow the appliance to
dry completely before putting into operation again.
WARNING
Live components may be exposed once the covers have been removed. Touch-
ing these can lead to an electric shock. Disconnect the mains plug before re-
moving any covers. Work on the electrical system must only be performed by
qualified electricians.
Repairs and service work are described in a separate service manual.
Storage and disposal
5.8.1
Storage
The appliance may only be stored under the following conditions:
■
in a dry and enclosed, dust-free room
■
frost-free
■
disconnected from the power supply
5.8.2
Disposal
This product is subject to Directive 2012/19/EC on Waste Electrical and Electronic
Equipment (WEEE) of the European Parliament and of the Council of Ministers. This ap-
pliance was placed on the market after 13 August 2005 in countries which have already
integrated this Directive into their national laws. It may not be disposed of in normal
household waste. For disposal, please contact your dealer or the manufacturer. Any
appliances that are infected, infectious or contaminated with materials hazardous to
health are excluded from return. Please also observe all other regulations applicable
in this context.
Pumping device CPM
6.
Pumping device CPM
6.1
Assembly and description
6.1.1
Design
Description
Overview front
1.
Pumping device
1
2
3
2.
Housing fan
3.
Water bath
6.1.2
Description
brief technical description
By using a pumping device, the water can circulate in the tub. The delivery volume is 1.2
litres per minute.
The temperature control of the WTB water bath is optimised for standing water. The operation of a CPM650 circulation pump may
require readjustment of the temperature control, as flowing water has a higher heat transfer
coefficient than standing water. See the general operating manual for the WTB water bath.
A warranty of the installed pump is only given when using the pumping device up to 80°C water temperature. If the pumping equip-
ment is used at temperatures above 80°C, no warranty can be claimed if damage occurs or a defect to the pump.
Description
Pumping device
1.
Water bath
1
2
3
4
5
2.
Water bath pumping device CPM650
3.
Water return from pumping device to
water bath
4.
Water inlet from the water bath to the
pumping device
NOTICE
NOTICE
Pumping device CPM
6.1.3
Intended use
The Memmert circulation pump is intended exclusively for use with a Memmert water
bath WTB. It is used for the internal circulation of the temperature control medium in
the water bath.
6.1.4
Use not in accordance with the intended purpose
The circulation pump is used exclusively for internal circulation in the water bath and is
not intended for temperature control of external circuits. The circulation pump must be
used only for pumping water and must not be operated with other liquids.
6.1.5
Material
This pumping device complies with the current requirements of the RoHS Directive.
You can find more information on this and material compliance on our homepage.
https://www.memmert.com/de/downloads/
Components
Material
Circulating pump housing
Stainless steel 1.4301 - ASTM 304
Screws and hinges
Stainless steel 1.4301 - ASTM 304
Silicone tube
Silicone - Shore A translucent
i The chamber load for the appliance must be carefully checked for chemical compati-
bility with the materials mentioned.
6.1.6
Electrical equipment
■
Operating voltage and current consumption: See name plate or technical data
■
Protection class I, i.e. operating insulation with PE conductor in accordance with
EN 60664-1
■
Protection type IP 20 acc. to EN 60 529
■
Interference suppression acc. to EN 55011 class B
■
Appliance fuse: Fuse F15H250V
6.1.7
Connections and interfaces
This appliance is intended for operation on an electrical power system with a system
impedance Zmax at the point of transfer (service line) of a maximum of 0.292 Ohm. The
operator must ensure that the appliance is operated only on an electrical power system
that meets these requirements. If necessary, you can ask your local energy supply com-
pany what the system impedance is. When connecting, observe the country-specific
regulations (e.g. in Germany DIN VDE 0100 withFI protective circuit).
Pumping device CPM
6.1.8
Designation (nameplate)
Description
Overview
1.
Type designation
GmbH+Co.KG D-91126 Schwabach FRG
Äußere Rittersbacher Str. 38
Made in Germany
memmert
Schutzart DIN EN 60529 - IP 20
1
2
3
4
5
6
8
7
Typ: CPM650
~ 230 V 50 Hz
F-Nr.: J221.0004
0,43 A 43 W
2.
Operating voltage
3.
Protection type
4.
CE conformity
5.
Address of manufacturer
6.
Disposal note
7.
Connection / power ratings
8.
Appliance number
6.1.9
Technical data - Pumping device
Zubehör
CPM650
Dimensions
Length
mm
296
Width
mm
140
Height
mm
180
electrical data
Conveyance volume
l/min
1,2
Operating voltage
V
230
Power
W
43
Current consumption
230V/50 Hz
A
0,43
Packaging material
Net weight
kg
2,8
Gross weight (in cardboard
box)
kg
3,7
Width
cm
47
Height
cm
42
Depth
cm
22
Pumping device CPM
6.1.10
Ambient conditions
Ambient temperature
+5 °C to 40 °C
Air humidity
max. 80 % rh (non-condensing)
Overvoltage category
II
Pollution degree
2
Altitude of installation
max. 2,000 m above sea level
Maximum mains voltage fluctuations
AC 230 (+/– 10 %)
■
The appliance may only be used in enclosed areas and under the ambient
conditions listed below
■
The appliance may not be used in areas where there is a risk of explosion. The sur-
rounding air must not contain any explosive dusts, gases, vapours or gas-air mixtures.
The appliance is not explosion-proof.
■
Heavy dust production or aggressive vapours in the vicinity of the appliance could lead
to sedimentation in the interior and, as a consequence, could result in short circuits or
damage to electrical parts. For this reason, sufficient measures to prevent large clouds
of dust or aggressive vapours from developing should be taken.
6.1.11
Scope of delivery and required accessories
■
Pumping device
■
Installation and connection set (consisting of inlet and outlet pipe, two hinges)
Pumping device CPM
6.2
Delivery, Transport and Setting Up
6.2.1
Safety
WARNING
Condensation in the electrical components may cause short circuits. After transport-
ing or storing the appliance under humid conditions, remove it from its packaging and
let it ventilate for at least 24 hours in normal ambient conditions. Do not connect the
appliance to the mains power during this time.
CAUTION
You may suffer crushing injuries to your hands when transporting and setting up the
appliance. Always wear protective gloves.
6.2.2
Delivery/unpacking
■
The accessories are packed in a cardboard box
■
Do not unpack the accessories until they are in their installation location to avoid
damage
■
Carefully cut open the cardboard packaging along one edge.
6.2.3
Product control
■
Check the delivery note to ensure that the scope of delivery is complete
■
Check the accessories for damage
If you notice deviations from the scope of delivery, damage or irregularities,
do not put the accessories into operation, but inform the carrier and the manufacturer's
factory.
6.2.4
Disposing of packaging material
■
Dispose of the packaging material (cardboard, wood, foil) in accordance with the appli-
cable disposal regulations for the respective material in your country
Pumping device CPM
6.2.5
Setting up
WARNING
The tube is supplied in a predefined length. This is possibly too long. Please make
sure that the hoses do not hang down. You could get caught during operation and be
scalded or burned by the heated temperature control medium.
CAUTION
The appliance may fall from the installation location due to unevenness and injure you
or someone else. Always place the appliance on an even, load-bearing and non-slip
surface.
CAUTION
The surfaces of the appliance can overheat during operation and overheat heat-sen-
sitive surfaces in close proximity to the appliance. Always place the appliance on a
heat-resistant surface and maintain the specified clearance around the appliance.
The distance between the wall and the rear of the appliance must be at least 15 cm. The
clearance from the ceiling must not be less than 75 cm and the side clearance from
walls or nearby appliances must not be less than 10 cm. Sufficient air circulation in the
vicinity of the appliance must be guaranteed at all times.
memmert
10 cm
10 cm
10 cm
15 cm
75 cm
10 cm
Pumping device CPM
6.2.6
Installation options pumping device
The pumping device can be placed either next to or behind the water bath . When
placing the pumping device, take care not to kink the hoses.
When setting up the pumping unit, make sure that
■
the specified wall distances are observed.
■
the hoses of the pumping device to the water bath are cut as short as possible.
NOTICE
Determine the installation location before you install the pumping device in the water bath. The water bath and pumping device may
only be transported individually.
6.2.7
Installation pumping device
1. Attach brackets for the pipe connection to the back of the water bath.
2. Attach the pipe connections to the brackets and hang them in the water bath.
Pumping device CPM
Pump connection IN (suction line): Attach the tube to the long pipe connec-
tion on the water bath, as this is where the water is conveyed from the water
bath to the pumping device.
Pump connection OUT (pressure line): Attach the tube to the short pipe
connection on the water bath, as this is where the water from the pump is
conveyed back to the water bath.
3. Fit the hoses onto the pipe connections
NOTICE
Pumping device CPM
6.3
Putting into operation
i There must not be any coarse-grained particles in the water bath. This could damage
the pumping device.
6.3.1
Putting into operation for the first time
NOTICE
When putting the appliance into operation for the first time, do not leave it unattended until it has reached a steady state.
6.3.2
Connect the appliance to the power supply
Place the power cable so that
■
it does not represent a trip hazard
■
it cannot come into contact with any hot parts
■
it is easily accessible at all times and can be pulled off quickly, for example in case of
interference or an emergency
6.3.3
Filling the water bath
NOTICE
To operate the pumping device, a minimum filling level of 70 mm is necessary. When filling and charging the water bath, please pay
attention to the maximum filling height, which must not be exceeded.
memmert
MIN
MAX
Pumping device CPM
6.4
Operation and control
WARNING
During operation of the pumping device, individual components may heat up.
Touching its surface can cause burns. Use temperature-resistant protective gloves.
6.4.1
Switching on appliance
Press main switch on the back of the appliance.
The pumping capacity of the pumping device is constant and cannot be regulated.
While switched on, the pumping device operates at a steady rate. The temperature is
controlled via the setting on the water bath.
6.4.2
Function test
■
To check that the unit is pumping, pull off the suction tube from the tube suspender and
place a finger on it to test for suction. If there is no suction effect: See chapter Faults,
warning and error messages.
6.5
Dismantling the hoses
WARNING
During operation of the pumping device, individual components may heat up.
Touching its surface can cause burns. Let the heated components cool down first.
1. Allow the water�in�the�water�bath�and�the�hoses�to�cool�down�first.
2. Drain the water from the water bath using the drain tap
until�the�raw�joints�are�exposed.
3. Pull the hoses off the pipe connections
NOTICE
Pumping device CPM
6.6
Faults, Warning & Error Messages
WARNING
During operation of the pumping device, individual components may heat up. Touch-
ing its surface can cause burns. Use temperature-resistant protective gloves.
Description
Cause
Action
CPM cannot be switched off
Main switch defective
■
Disconnect the mains plug
■
Contact customer service
Tempering medium circulates little
or not at all
■
Water level too low
■
Pump defective
■
Monitor water level
■
Contact customer service
Pumping device CPM
6.7
Cleaning and decontamination
DANGER
Danger due to electric shock.
Penetration of liquid into the appliance can cause electric shocks and short circuits.
–
Protect the appliance from splashing water.
–
Before cleaning and maintenance work, switch off the appliance and disconnect
the mains plug.
–
The appliance must not be cleaned and disinfected wet. Allow the appliance to
dry completely before putting into operation again.
CAUTION
Danger of cuts due to sharp edges.
Always wear gloves when working on the appliance.
6.7.1
Cleaning
Interior and metal surfaces
Conventional stainless steel cleaning agents can be used to clean the metal surfaces
of the pump unit. Make sure that no rusting objects come into contact with the stain-
less steel housing. Rust deposits can lead to an infection of the stainless steel. If rust
spots appear due to contamination, the affected areas must be cleaned and polished
immediately.
Plastic parts
Do not clean plastic parts with abrasive or solvent-based cleaning agents.
6.7.2
Decontamination
CAUTION
Skin and eye contact with decontaminants may cause irritation of the hands and eyes
or allergic reactions and chemical burns.
–
Wear chemical-resistant gloves during decontamination.
–
For correct application, follow the instructions for use on the respective contain-
er of the decontamination agent
Memmert appliances can become contaminated by loading infectious substances or by
environmental influences at the place of installation. All parts of the affected appliance
must be decontaminated both before sending to our service and before disposing of
the appliance.
■
For decontamination, we recommend commercially available decontaminants based on
isopropanol or ethanol.
If you have any questions about using a decontamination agent for decontaminating
Memmert appliances, please contact our service department.
After cleaning and decontamination, ventilate the installation site and allow the appli-
ance to dry completely.
Pumping device CPM
6.8
Repairs and service
DANGER
Danger due to electric shock.
Penetration of liquid into the appliance can cause electric shocks and short circuits.
–
Protect the appliance from splashing water.
–
Before cleaning and maintenance work, switch off the appliance and disconnect
the mains plug.
–
The appliance must not be cleaned and disinfected wet. Allow the appliance to
dry completely before putting into operation again.
WARNING
Live components may be exposed once the covers have been removed. Touch-
ing these can lead to an electric shock. Disconnect the mains plug before re-
moving any covers. Work on the electrical system must only be performed by
qualified electricians.
Repairs and service work are described in a separate service manual.
6.9
Storage and disposal
6.9.1
Storage
The accessories may only be stored under the following conditions:
■
in a dry and enclosed, dust-free room
■
frost-free
■
disconnected from the power supply
Before storage, empty the hoses of the pumping device.
6.9.2
Disposal
This product is subject to Directive 2012/19/EC on Waste Electrical and
Electronic Equipment (WEEE) of the European Parliament and of the Council of Min-
isters. This appliance was placed on the market after 13 August 2005 in countries
which have already integrated this Directive into their national laws. It may not be
disposed of in normal household waste. For disposal, please contact your dealer
or the manufacturer. Any appliances that are infected, infectious or contaminated
with materials hazardous to health are excluded from return. Please also observe all
other regulations applicable in this context.
Cooling unit CDP
7.
Cooling unit CDP
7.1
Assembly and description
7.1.1
Design
Description
Overview
1.
Peltier cooling unit with fan
1
2
3
4
5
2.
Housing fan
3.
On/Off switch
4.
Mains lead
5.
Water bath WTB15
7.1.2
Description
Brief technical description
The CDP water bath cooling unit enables the water in Memmert water baths to be
cooled to temperatures below room temperature. Water is pumped out of the water
bath, cooled by a Peltier module and then returned to the water bath.
The cooling power is constant and cannot be regulated. The temperature control of the
water is taken over by the water bath; its controller automatically takes into account the
cooling power of the cooling unit.
The cooling unit's low-temperature fuse automaticaly switches off the appliance at a
water temperature of approx. 3 ºC.
Functional principle of the water bath cooling unit CDP115
Description
Overview front
1.
Water bath
1
2
3
4
5
2.
Water bath cooling unit CDP115
3.
Peltier cooling module
4.
Water return from cooling unit to the
water bath
5.
Water supply from the water bath to the
cooling unit
Cooling unit CDP
7.1.3
Intended use
The Memmert cooling unit is intended exclusively for use with a Memmert water bath
WTB. It is used to cool the temperature control medium in the water bath to tempera-
tures below room temperature from +10 °C.
7.1.4
Use not in accordance with the intended purpose
The cooling unit is used exclusively for internal cooling in the water bath and is not in-
tended for temperature control of external circuits. The water bath cooling unit must be
used only for pumping and cooling water and must not be operated with other liquids.
7.1.5
Material
This cooling unit complies with the current requirements of the RoHS Directive. You
can find more information on this and on material compliance on our homepage www.
memmert.com.
Components
Material
Cooling unit housing
Stainless steel 1.4301 - ASTM 304
Peltier cooling module
Anodised aluminium
Screws and hinges
Stainless steel 1.4301 - ASTM 304
Inlet and outlet tube
Chloroprene tube LEZ-NEO 4x2 inside ø4, outside ø8
i The chamber load for the appliance must be carefully checked for chemical compati-
bility with the materials mentioned.
7.1.6
Electrical equipment
■
Operating voltage and current consumption: See name plate or technical data
■
Protection class I, i.e. operating insulation with PE conductor in accordance with
EN 60664-1
■
Protection type IP 20 acc. to EN 60 529
■
Interference suppression acc. to EN 55011 class B
■
Appliance fuse: Fuse F15H250V
■
Undertemperature monitor as frost protection: Switching off the cooling unit at < +5 °C
7.1.7
Connections and interfaces
This appliance is intended for operation on an electrical power system with a system
impedance Zmax at the point of transfer (service line) of a maximum of 0.292 Ohm. The
operator must ensure that the appliance is operated only on an electrical power system
that meets these requirements. If necessary, you can ask your local energy supply com-
pany what the system impedance is. When connecting, observe the country-specific
regulations (e.g. in Germany DIN VDE 0100 withFI protective circuit).
Cooling unit CDP
7.1.8
Designation (nameplate)
Description
Overview
1.
Type designation
GmbH+Co.KG D-91126 Schwabach FRG
Äußere Rittersbacher Str. 38
Made in Germany
memmert
Schutzart DIN EN 60529 - IP 20
1
2
3
4
5
6
8
7
Typ: CDP115
~ 230 V 50/60 Hz
F-Nr.: J121.0001
0,7 A 160 W
2.
Operating voltage
3.
Protection type
4.
CE conformity
5.
Address of manufacturer
6.
Disposal note
7.
Connection / power ratings
8.
Appliance number
7.1.9
Technical data - Cooling unit
Zubehör
CPD115
Dimensions
Length
mm
456
Width
mm
205
Height
mm
435
Electrical data
Conveyance volume
l/min
0,6
Effective cooling capacity
W
113
Electrical connection
V
115/230
Power
W
160
Current
consumption
50/60 Hz
115 V
A
0,7
230 V
1,4
Packaging material
Net weight
kg
8,35
Gross weight (in cardboard
box)
kg
9,95
Width
cm
59
Height
cm
41
Depth
cm
33
Cooling unit CDP
7.1.10
Ambient conditions
Ambient temperature
+5 °C to 40 °C
Air humidity
max. 80 % rh (non-condensing)
Overvoltage category
II
Pollution degree
2
Altitude of installation
max. 2,000 m above sea level
Maximum mains voltage fluctuations
AC 115 (+/– 10 %)
AC 230 (+/– 10 %)
■
The appliance may only be used in enclosed areas and under the ambient
conditions listed below
■
The appliance may not be used in areas where there is a risk of explosion. The sur-
rounding air must not contain any explosive dusts, gases, vapours or gas-air mixtures.
The appliance is not explosion-proof.
■
Heavy dust production or aggressive vapours in the vicinity of the appliance could lead
to sedimentation in the interior and, as a consequence, could result in short circuits or
damage to electrical parts. For this reason, sufficient measures to prevent large clouds
of dust or aggressive vapours from developing should be taken.
7.1.11
Scope of delivery and required accessories
■
Cooling unit
■
Installation and connection set (consisting of inlet and outlet pipe, two hinges)
Cooling unit CDP
7.2
Delivery, Transport and Setting Up
7.2.1
Safety
WARNING
Condensation in the electrical components may cause short circuits. After transport-
ing or storing the appliance under humid conditions, remove it from its packaging and
let it ventilate for at least 24 hours in normal ambient conditions. Do not connect the
appliance to the mains power during this time.
CAUTION
You may suffer crushing injuries to your hands when transporting and setting up the
appliance. Always wear protective gloves.
7.2.2
Delivery/unpacking
■
The accessories are packed in a cardboard box
■
Do not unpack the accessories until they are in their installation location to avoid
damage
■
Carefully cut open the cardboard packaging along one edge.
7.2.3
Product control
■
Check the delivery note to ensure that the scope of delivery is complete
■
Check the accessories for damage
If you notice deviations from the scope of delivery, damage or irregularities, do not put
the accessories into operation, but inform the carrier and the manufacturer's factory.
7.2.4
Disposing of packaging material
■
Dispose of the packaging material (cardboard, wood, foil) in accordance with the appli-
cable disposal regulations for the respective material in your country
Cooling unit CDP
7.2.5
Setting up
WARNING
The tube is supplied in a predefined length. This is possibly too long. Please make
sure that the hoses do not hang down. You could get caught during operation and
suffer scalds or burns from the heated water.
CAUTION
The appliance may fall from the installation location due to unevenness and injure you
or someone else. Always place the appliance on an even, load-bearing and non-slip
surface.
CAUTION
The surfaces of the appliance can overheat during operation and overheat heat-sen-
sitive surfaces in close proximity to the appliance. Always place the appliance on a
heat-resistant surface and maintain the specified clearance around the appliance.
The distance between the wall and the rear of the appliance must be at least 15 cm. The
clearance from the ceiling must not be less than 75 cm and the side clearance from
walls or nearby appliances must not be less than 10 cm. Sufficient air circulation in the
vicinity of the appliance must be guaranteed at all times.
NOTICE
Determine the installation location before you start installing the cooling unit to the water bath. The water bath and cooling unit may
only be transported individually.
10 cm
10 cm
15 cm
10 cm
75 cm
Cooling unit CDP
7.2.6
Installation cooling unit
1. Attach brackets for the pipe connection to the back of the water bath.
2. Attach the pipe connections to the brackets and hang them in the water bath.
NOTICE
Pump connection IN (suction line): Attach the tube to the long pipe connec-
tion on the water bath, as this is where the water is conveyed from the water
bath to the pumping device.
Pump connection OUT (pressure line): Attach the tube to the short pipe
connection on the water bath, as this is where the water from the pump is
conveyed back to the water bath.
Cooling unit CDP
3. Put the hoses on the pipe connections and connect them to the CDP115.
7.2.7
Installation options Cooling unit
The cooling unit can be placed either next to or behind the water bath. When placing the
cooling unit, take care not to kink the hoses.
When setting up the cooling unit, make sure that
■
the specified wall distances are observed.
■
The hoses of the cooling unit to the water bath are cut as short as possible.
NOTICE
Determine the installation location before you start installing the cooling unit to the water bath. The water bath and cooling unit may
only be transported individually.
Cooling unit CDP
7.3
Putting into operation
i There must not be any coarse-grained particles in the water bath. This could damage
the pumping device.
7.3.1
Putting into operation for the first time
NOTICE
When putting the appliance into operation for the first time, do not leave it unattended until it has reached a steady state.
7.3.2
Connect the appliance to the power supply
Place the power cable so that
■
it does not represent a trip hazard
■
it cannot come into contact with any hot parts
■
it is easily accessible at all times and can be pulled off quickly, for example in case of
interference or an emergency
7.3.3
Filling the water bath
NOTICE
To operate the cooling unit, a minimum filling level of 70 mm is necessary. When filling and charging the water bath, please pay
attention to the maximum filling height, which must not be exceeded.
memmert
MIN
MAX
Cooling unit CDP
7.4
Operation and control
WARNING
During operation of the cooling unit, individual components may heat up. Touching its
surface can cause burns. Use temperature-resistant protective gloves.
7.4.1
Switching on appliance
Press the main switch on the side of the unit.
The cooling power of the cooling unit is constant and cannot be regulated. The cool-
ing unit works at a constant power as long as it is switched on. The temperature is
controlled via the setting on the water bath. Do not operate the cooling unit when
the water bath heater is switched off. This could trigger the automatic shutdown
.
If the set temperature is above room temperature, it is recommended to switch off the
cooling unit.
7.4.2
Automatic switch-off
The cooling unit has a low temperature fuse which switches off the unit automatically
when the water temperature is at about 3ºC, to prevent icing over and possible damage
to the cooling element. When the water temperature reaches about 5 ºC, the cooling
unit switches back on again.
7.4.3
Function test
■
To check that the unit is pumping, pull off the suction tube from the tube suspender and
place a finger on it to test for suction. If there is no suction effect: See next chapter,
Troubleshooting.
7.5
Dismantling the hoses
1. Allow the water�in�the�water�bath�and�the�hoses�to�cool�down�first.
2. Drain the water from the water bath using the drain tap
until�the�raw�joints�are�exposed.
3. Pull the hoses off the pipe connections
NOTICE
Cooling unit CDP
7.6
Faults, Warning & Error Messages
WARNING
During operation of the pumping device, individual components may heat up. Touch-
ing its surface can cause burns. Use temperature-resistant protective gloves.
Description
Cause
Action
CDP115 cannot be switched off
Main switch defective
■
Disconnect the mains plug
■
Contact customer service
Tempering medium circulates little
or not at all
■
Water level too low
■
Pump defective
■
Monitor water level
■
Contact customer service
Cooling unit CDP
7.7
Cleaning and decontamination
DANGER
Danger due to electric shock.
Penetration of liquid into the appliance can cause electric shocks and short circuits.
–
Protect the appliance from splashing water.
–
Before cleaning and maintenance work, switch off the appliance and disconnect
the mains plug.
–
The appliance must not be cleaned and disinfected wet. Allow the appliance to
dry completely before putting into operation again.
CAUTION
Danger of cuts due to sharp edges.
Always wear gloves when working on the appliance.
7.7.1
Cleaning
Metal surfaces
The metal surfaces of the cooling unit can be cleaned with commercially available
stainless steel cleaning agents. Make sure that no rusting objects come into contact
with the stainless steel housing. Rust deposits can lead to an infection of the stainless
steel. If rust spots appear due to contamination, the affected areas must be cleaned
and polished immediately.
Plastic parts
Do not clean plastic parts with abrasive or solvent-based cleaning agents.
7.7.2
Decontamination
CAUTION
Skin and eye contact with decontaminants may cause irritation of the hands and eyes
or allergic reactions and chemical burns.
–
Wear chemical-resistant gloves during decontamination.
–
For correct application, follow the instructions for use on the respective contain-
er of the decontamination agent
Memmert appliances can become contaminated by loading infectious substances or by
environmental influences at the place of installation. All parts of the affected appliance
must be decontaminated both before sending to our service and before disposing of
the appliance.
■
For decontamination, we recommend commercially available decontaminants based on
isopropanol or ethanol.
If you have any questions about using a decontamination agent for decontaminating
Memmert appliances, please contact our service department.
After cleaning and decontamination, ventilate the installation site and allow the
appliance to dry completely.
Cooling unit CDP
7.8
Repairs and service
DANGER
Danger due to electric shock.
Penetration of liquid into the appliance can cause electric shocks and short circuits.
–
Protect the appliance from splashing water.
–
Before cleaning and maintenance work, switch off the appliance and disconnect
the mains plug.
–
The appliance must not be cleaned and disinfected wet. Allow the appliance to
dry completely before putting into operation again.
WARNING
Live components may be exposed once the covers have been removed. Touch-
ing these can lead to an electric shock. Disconnect the mains plug before re-
moving any covers. Work on the electrical system must only be performed by
qualified electricians.
Repairs and service work are described in a separate service manual.
7.9
Storage and disposal
7.9.1
Storage
The accessories may only be stored under the following conditions:
■
in a dry and enclosed, dust-free room
■
frost-free
■
disconnected from the power supply
Before storage, empty the hoses of the cooling unit.
7.9.2
Disposal
This product is subject to Directive 2012/19/EC on Waste Electrical and Electronic
Equipment (WEEE) of the European Parliament and of the Council of Ministers. This
appliance was placed on the market after 13 August 2005 in countries which have al-
ready integrated this Directive into their national laws. It may not be disposed of in
normal household waste. For disposal, please contact your dealer or the manufacturer.
Any appliances that are infected, infectious or contaminated with materials hazardous
to health are excluded from return. Please also observe all other regulations applicable
in this context.
CE declarations of conformity Shaking device SV1524 | SV3550
EC Declaration of Conformity
Manufacturer´s name and address:
Memmert GmbH + Co. KG
Äussere Rittersbacherstraße 38
D-91126 Schwabach
Deutschland
Authorized representative for
compiling the technical documents:
Memmert GmbH + Co.KG
Äußere Rittersbacher Straße 38
D-91126 Schwabach
Deutschland
Product:
Shaking device
Type:
SV…/1524/3550
Including all accessories
Nominal voltage:
AC 230 V 50/60 Hz, alternative AC 115 V 50/60 Hz
This product complies with the provisions of the directive
2006/42/EC
DIRECTIVE 2006/42/EC OF THE EUROPEAN PARLIAMENT AND OF THE COUNCIL
of 17 May 2006 on machinery, and amending Directive 95/16/EC (recast)
Full compliance with the standard below proves the conformity of the designated product with the essential
protection requirements of the above-mentioned EC Directive:
EN ISO 12100:2010
The designated product is in conformity with the European EMC-Directive
2014/30/EU
including amendments
Council Directive of 03 May 1989 on the approximation of the laws of the Member States relating to
electromagnetic compatibility
Full compliance with the standards listed below proves the conformity of the designated product with the essential
protection requirements of the above-mentioned EC Directive:
DIN EN 61326-1:213-07
EN 61326-1:2013
The object described in the declaration above corresponds to the directive 2011/65/EC of the European
Parliament and the Council of 8 June 2011on the restriction of the use of certain hazardous substances in
electrical and electronic equipment.
Schwabach, 04.06.2021
Legally binding signature of the issuer:
Christiane Riefler-Karpa, managing director
This declaration certifies compliance with the above mentioned directives but does not include a property assurance.
The safety note given in the product documentation which are part of the supply, must be observed.
D49191
CE declarations of conformity Pumping device CPM650
EC Declaration of Conformity
Manufacturer´s name and address:
Memmert GmbH + Co. KG
Äussere Rittersbacherstraße 38
D-91126 Schwabach
Deutschland
Authorized representative for
compiling the technical documents:
Memmert GmbH + Co.KG
Äußere Rittersbacher Straße 38
D-91126 Schwabach
Deutschland
Product:
Pump device
Type:
CPM650
Including all accessories
Nominal voltage:
AC 230 V 50/60 Hz, alternative AC 115 V 50/60 Hz
This product complies with the provisions of the directive
2006/42/EC
DIRECTIVE 2006/42/EC OF THE EUROPEAN PARLIAMENT AND OF THE COUNCIL
of 17 May 2006 on machinery, and amending Directive 95/16/EC (recast)
Full compliance with the standard below proves the conformity of the designated product with the essential
protection requirements of the above-mentioned EC Directive:
EN ISO 12100:2010
The designated product is in conformity with the European EMC-Directive
2014/30/EU
including amendments
Council Directive of 03 May 1989 on the approximation of the laws of the Member States relating to
electromagnetic compatibility
Full compliance with the standards listed below proves the conformity of the designated product with the essential
protection requirements of the above-mentioned EC Directive:
DIN EN 61326-1:213-07
EN 61326-1:2013
The object described in the declaration above corresponds to the directive 2011/65/EC of the European
Parliament and the Council of 8 June 2011on the restriction of the use of certain hazardous substances in
electrical and electronic equipment.
Schwabach, 04.06.2021
Legally binding signature of the issuer:
Christiane Riefler-Karpa, managing director
This declaration certifies compliance with the above mentioned directives but does not include a property assurance.
The safety note given in the product documentation which are part of the supply, must be observed.
D49199
CE declarations of conformity Cooling unit CDP115
EC Declaration of Conformity
Manufacturer´s name and address:
Memmert GmbH + Co. KG
Äussere Rittersbacherstraße 38
D-91126 Schwabach
Deutschland
Authorized representative for
compiling the technical documents:
Memmert GmbH + Co.KG
Äußere Rittersbacher Straße 38
D-91126 Schwabach
Deutschland
Product:
Peltier-cooling device
Type:
CDP115
Including all accessories
Nominal voltage:
AC 230 V 50/60 Hz, alternative AC 115 V 50/60 Hz
This product complies with the provisions of the directive
2006/42/EC
DIRECTIVE 2006/42/EC OF THE EUROPEAN PARLIAMENT AND OF THE COUNCIL
of 17 May 2006 on machinery, and amending Directive 95/16/EC (recast)
Full compliance with the standard below proves the conformity of the designated product with the essential
protection requirements of the above-mentioned EC Directive:
EN ISO 12100:2010
The designated product is in conformity with the European EMC-Directive
2014/30/EU
including amendments
Council Directive of 03 May 1989 on the approximation of the laws of the Member States relating to
electromagnetic compatibility
Full compliance with the standards listed below proves the conformity of the designated product with the essential
protection requirements of the above-mentioned EC Directive:
DIN EN 61326-1:213-07
EN 61326-1:2013
The object described in the declaration above corresponds to the directive 2011/65/EC of the European
Parliament and the Council of 8 June 2011on the restriction of the use of certain hazardous substances in
electrical and electronic equipment.
Schwabach, 04.06.2021
Legally binding signature of the issuer:
Christiane Riefler-Karpa, managing director
This declaration certifies compliance with the above mentioned directives but does not include a property assurance.
The safety note given in the product documentation which are part of the supply, must be observed.
D49207
Memmert GmbH + Co. KG | Äußere Rittersbacher Straße 38 | 91126 Schwabach
Tel. +49 9122 925-0 | E-Mail: [email protected] | www.memmert.com
facebook.com/memmert.family | instagram.com/memmert.family
Shaking device SV1524/ SV3550
Pumping device CPM650
Cooling unit CDP115
Accessory guide
D49169 Last updated
06/2021
englisch