Text of the manual

Shaking device SV1524 | SV3550

Pumping device CPM650 Cooling unit CDP115 Translation of the original operating manual

About this manual

Purpose and target audience This manual describes the design, function, transport, operation and maintenance of accessories for WTB water baths. It is intended for use by trained personnel of the owner, who have the task of operating and/or maintaining the respective accessories. If you are asked to work on the accessories, read this manual carefully before starting. Familiarise yourself with the safety regulations. Only carry out work described in this manual for the respective accessories. If there is something you do not understand, or certain information is missing, ask your manager or contact the manufacturer. Do not do anything without authorisation. Versions Several accessories of different sizes and functions are available. The functions de- scribed in this manual refer to the latest firmware version. Due to individual sizes, illus- trations in this manual may be slightly different to the actual appearance. Other documents that you must observe: ■ the general operating manual for all water baths WTB ■ for service and repair work, observe the separate service manual Storage and resale This operating manual is part of the accessories and must always be stored so that persons to work with the respective accessories have access to them. The owner's responsibility is to ensure that persons who work or are to work with the respective accessories are informed of where these operating manu- al are located. We recommend that you always keep them in a protected place near the accessories. Make sure that the operating manual is not damaged by heat or humidity. If an accessory is resold or transported and then set up again at a different location, the operating manual must remain with it. The current ver- sion of this operating manual in PDF format is also available for download from http://www.memmert.com/ Manufacturer address and customer service MEMMERT GmbH + Co. KG Willi-Memmert-Str. 90-96 D-91186 Büchenbach Germany Phone: +49 (0)9122 925-0 Fax: +49 (0)9122 14585 E-mail: [email protected] Internet: www.memmert.com Customer service: Hotline: +49 (0)9171 9792 911 Service fax: +49 (0)9171 9792 979 Email: [email protected] For service enquiries, please always specify the appliance number given on the name plate.

Terms and signs used ............................................................................................................................................................................ 4

1.2 Product safety and dangers ................................................................................................................................................................... 5 1.3 Requirements of the operating personnel .............................................................................................................................................. 5 1.4 Responsibility of the owner .................................................................................................................................................................... 6 1.1 Applied directives and standards ........................................................................................................................................................... 6 1.5 Changes and alterations ......................................................................................................................................................................... 6 1.6 Behaviour in case of malfunctions and irregularities ............................................................................................................................. 6 1.7 Switch off in an emergency .................................................................................................................................................................... 7 2.� Sloping�and�flat�lids� 8 2.1 Sloping lid ............................................................................................................................................................................................... 8 2.2 Flat lid ................................................................................................................................................................................................... 10 3. Test tube racks 11 4. Shelf reversible 13 5. Shaking Device 15 5.1 Assembly and description .................................................................................................................................................................... 15 5.2 Delivery, Transport and Setting Up ....................................................................................................................................................... 19 5.3 Putting into operation ........................................................................................................................................................................... 23 5.4 Operation and control ........................................................................................................................................................................... 25 5.5 Faults, Warning & Error Messages ....................................................................................................................................................... 28 5.6 Cleaning and decontamination ............................................................................................................................................................. 29 5.7 Repairs and service ..............................................................................................................................................................................30 5.8� Storage�and�disposal ............................................................................................................................................................................31 6. Pumping device CPM 32 6.1 Assembly and description .................................................................................................................................................................... 32 6.2 Delivery, Transport and Setting Up ....................................................................................................................................................... 36 6.3 Putting into operation ........................................................................................................................................................................... 40 6.4 Operation and control ........................................................................................................................................................................... 41 6.5 Dismantling the hoses .......................................................................................................................................................................... 41 6.6 Faults, Warning & Error Messages ....................................................................................................................................................... 42 6.7 Cleaning and decontamination ............................................................................................................................................................. 43 6.8� Repairs�and�service ..............................................................................................................................................................................44 6.9 Storage and disposal ............................................................................................................................................................................44 7. Cooling unit CDP 45 7.1 Assembly and description .................................................................................................................................................................... 45 7.2 Delivery, Transport and Setting Up ....................................................................................................................................................... 49 7.3 Putting into operation ........................................................................................................................................................................... 53 7.4 Operation and control ........................................................................................................................................................................... 54 7.5 Dismantling the hoses .......................................................................................................................................................................... 54 7.6 Faults, Warning & Error Messages ....................................................................................................................................................... 55 7.7 Cleaning and decontamination ............................................................................................................................................................. 56 7.8� Repairs�and�service ..............................................................................................................................................................................57 7.9 Storage and disposal ............................................................................................................................................................................57 8.� Attachment� 58

Terms and signs used

In this manual and on the appliance itself, certain common terms and signs are used to warn you of possible dangers or to give you hints that are important in avoiding injury or damage. Observe and follow these notes and regulations to avoid accidents and damage. These terms and signs are explained below. 1.1.1 Terms used DANGER warns of a dangerous situation that leads directly to death or serious (irreversible) injuries. WARNING warns of a dangerous situation that might le ad to death or serious injuries. CAUTION warns of a dangerous situation that might lead to mo derate or minor injuries. NOTICE warns of material damage 1.1.2 Signs used Symbol Meaning Symbol Meaning General warning sign - Observe operating manual Gases / vapours Danger of electric shock Danger of explosion Flammable substance warning Caution: hot steam Caution: hot Disconnect the mains plug Observe information in separate manual Use gloves

Product safety and dangers

The accessories are technically sophisticated, manufactured using high-quality materials and tested for many hours in the factory. They reflect the state of the art and comply with recognised technical safety regulations. However, there are still risks involved, even when the appliances are used as intended. These are described below. DANGER Danger due to electric shock. Penetration of liquid into the appliance can cause electric shocks and short circuits. – Protect the appliance from splashing water. – Switch off the appliance and disconnect the mains plug before cleaning and maintenance work. – The appliance must not be cleaned and disinfected wet. Allow the appliance to dry completely before putting it back into operation. WARNING When loading the appliance with an unsuitable chamber load, poisonous or explosive vapours or gases may be produced. This could cause the appliance to explode, and persons could be severely injured or poisoned. The appliance may only be loaded with materials / test objects that do not emit any poisonous or explosive vapours when heated up. WARNING Depending on operation, the surfaces of the appliance and the chamber load may still be very hot after the appliance is switched off. Touching these sur- faces can cause burns. Wear heat-resistant protective gloves or wait until the appliance cools down before touching. 1.3 Requirements of the operating personnel The accessories may only be operated and maintained by persons who are of legal age and have been instructed accordingly. Personnel who are to be trained, instructed or who are undergoing general training may only work with the appliance under the contin- uous supervision of an experienced person. Repairs may only be performed by qualified electricians. The regulations in the separate service manual must be observed.

Responsibility of the owner

The owner of the appliance ■ is responsible for the flawless condition of the appliance and for it being operated in accordance with its intended use; ■ is responsible for ensuring that persons who are to operate or service the appliance are qualified to do this, have been instructed accordingly and are familiar with the operating manual at hand; ■ must know about the applicable guidelines, requirements and operational safety regu- lations, and train staff accordingly; ■ is responsible for ensuring that unauthorised persons have no access to the appliance; ■ is responsible for ensuring that the maintenance plan is adhered to and that maintenance work is properly carried out; ■ has to ensure that the appliance and its surroundings are kept clean and tidy, for exam- ple through corresponding instructions and inspections; ■ is responsible for ensuring that personal protective clothing is worn by operating per- sonnel, e.g., work clothes, safety shoes and protective gloves. 1.1 Applied directives and standards EU Declaration of Conformity for the appliance can be found at the end of the document and online at: www.memmert.com/de/downloads/downloads/eu-konformitaetserklaerung Under the REACH regulation, Memmert provides the information on chemical sub- stances in Memmert appliances online at https://www.memmert.com/de/reach-rohs/ 1.5 Changes and alterations The accessories may be used only in connection with a Memmert water bath WTB. The accessories must not be converted or modified without authorisation. No parts may be added or inserted which have not been approved by the manufacturer. Unauthorised modifications or changes to the appliance will void any manufacturer's responsibility regarding the CE conformity of the product. The manufacturer is not liable for any damage, danger or injuries that result from un- authorised changes or alterations, or from non-compliance with the provisions in this manual. 1.6 Behaviour in case of malfunctions and irregularities The accessories may only be used in a flawless condition. If you, as the operator, notice irregularities, malfunctions or damage, immediately turn off the accessories and inform your superior.

Switch off in an emergency

Press the main switch on the respective accessories. This disconnects the appliance from the power supply at all poles. Illustration Accessories Shaking device - SV Cooling unit - CDP115 Pumping device - CPM650

Sloping and flat lids

2. Sloping and flat lids You may suffer bruising injuries to your hands. Always wear protective gloves. Do not let the mounted slanted or flat lid of the water bath fall shut. Note that the slanted or flat lid heats up to the temperature control liquid temperature during operation. 2.1 Sloping lid To avoid unwanted evaporation of the temperature control liquid and to achieve the most even temperature distribution possible, the appliance should always be operated with the slanted lid closed. The shape of the lid ensures that condensate does not drip into the chamber load. Water bath size Stainless steel sloping lid WTB6 WTB11 B39844 WTB15 WTB24 B39845 WTB35 WTB50 B39846 CAUTION CAUTION

Sloping and flat lids

2.1.1 Assembly 1. Loosen the blind plugs at the fastening points (left + right) for the lid on the water bath. 2. Remove the dummy plug and store it. 3. Put�on�the�fastening�bracket�of�the�cover�and�fix�it�with�screws. 4. Tighten�screws�and�check�that�the�lid�fits�tightly.

Sloping and flat lids

2.2 Flat lid The flat lid, available as an accessory, can be used to place test flasks on the surface of the temperature control liquid. By removing or inserting rings, the recesses are adapted to the piston size. The rings may only be inserted or removed when they have cooled down. Water bath size Stainless steel flat lid Stainless steel special flat lid WTB6 WTB11 B48804 B48807 WTB15 WTB24 B48805 B48808 WTB35 WTB50 B48806 B48809

Test tube racks

3. Test tube racks The test tube racks differ in the number and diameter of the individual bores. They are hooked into the water bath from above. The test tube racks may only be inserted or removed when they have cooled down. Ø Bores WTB6 / WTB11 WTB15 / WTB24 WTB35 / WTB50 Ø10mm 34 bores B48822 58 bores B48823 74 Bores B50656 Ø12mm 30 bores B48824 58 bores B48825 66 Bores B50657 Ø14.5mm 26 Bores B48826 45 Bores B48827 58 bores B50658 Ø18mm 22 bores B48828 38 Bores B48829 50 Bores B50659 33 Bores B48834 57 Bores B48835 75 Bores B50662

Test tube racks

Ø Bores WTB6 / WTB11 WTB15 / WTB24 WTB35 / WTB50 Ø18mm 44 bores B48836 76 Bores B48837 100 Bores B50663 Ø21mm 18 Bores B48830 34 bores B48831 44 bores B50660 Ø32mm 14 Bores B48832 22 bores B48833 30 bores B50661

Maximum loading of the waterbaths with reversible shelf

Size WTB Maximum Loading WTB 6/11 25 kg WTB 15/24 50 kg WTB 35/50 100 kg Conversion of reversible shelf 60 mm 1. Reversible shelves can be set to two heights: 60mm and 30 mm 2. Unscrew screws to remove feet 3. Remove feet and rotate depending on desired height (in this case 30mm)

4. Position rotated feet on holders of floor grate and fix with screws

30 mm 5. Reversible shelf now has a height of 30 mm

Assembly and description

5.1.1 Design Description Overview 1. Water bath 1 2 3 4 2. Recording frame 3. Shaking Device 4. Display shaker 5.1.2 Description Brief technical description The shaking device enables controlled shaking of test flasks or test tubes in a water bath. The shaking device performs a back and forth movement with a frequency of ap- prox. 30 to 150 strokes per minute. The shaking frequency is electronically controlled and adjustable via touch screen. The same ambient conditions apply for use as for the basic units. NOTICE If the shaking device remains in a switched-on water bath, the shaking device must always be switched on, as in the switched-off state, the overpressure steam barrier is inoperative, and condensate can enter the shaking device housing. 5.1.3 Intended use The Memmert shaking device is intended exclusively for a Memmert water bath WTB15, WTB24, WTB35 or WTB50. It is used for the controlled back and forth movement of samples at 30 to 150 rpm (strokes per minute), with one of approx. 15mm. The sample holder is in direct contact with the temperature control medium. 5.1.4 Use not in accordance with the intended purpose Depending on the loading of the shaking device and the selected shaking frequency, the temperature control medium and/or the shaking material could spill over. The user's responsibility is to adjust the shaking frequency or the fill level but without going below the minimum fill level.

The shaker complies with the current requirements of the RoHS Directive.

You can find more information on this and material compliance on our homepage: https://www.memmert.com/de/downloads/ Components Material Shaking Device Stainless steel 1.4301 - ASTM 304 Support basket Stainless steel 1.4301 - ASTM 304 Screws and hinges Stainless steel 1.4301 - ASTM 304 i The chamber load for the appliance must be carefully checked for chemical compati- bility with the materials mentioned. 5.1.6 Electrical equipment ■ Operating voltage and current consumption: See name plate or technical data ■ Protection class I, i.e. operating insulation with PE conductor in accordance with EN 60664-1 ■ Protection type IP 20 acc. to EN 60 529 ■ Interference suppression acc. to EN 55011 class B ■ Appliance fuse: Fuse F15H250V 5.1.7 Electrical connection The appliance is intended for operation on a power supply network with a system im- pedance Zmax at the transfer point (house connection) of a maximum of 0.292 Ohm. The operator must ensure that the appliance is only operated on power supply network that meets these requirements. If necessary, you can ask your local energy supply com- pany what the system impedance is. When connecting, observe the country-specific regulations (e.g. in Germany DIN VDE 0100 withFI protective circuit).

GmbH+Co.KG D-91126 Schwabach FRG

Äußere Rittersbacher Str. 38 Made in Germany memmert Schutzart DIN EN 60529 - IP 20 1 2 3 4 5 6 8 7 Typ: SV1524 ~ 230 V 50/60 Hz F-Nr.: P321.0001 0,1 A 8 W 2. Operating voltage 3. Protection type 4. CE conformity 5. Address of manufacturer 6. Disposal note 7. Connection / power ratings 8. Appliance number 5.1.9 Technical data – Shaking device Gerätegröße 15/24 35/50 Dimensions Length mm 706 826 Width mm 297 500 Height mm 190 190 electrical data Shaking frequence Number of strokes per minute 35 bis 150 Operating voltage V 115/ 230 Frequency Hz 50/ 60 Power W 8 12 Current consumption 50/60 Hz 115V A 0,1 0,1 230V 0,1 0,1 Packaging material Net weight kg 6,1 8,7 Gross weight (in cardboard box) kg 8,1 11,7 Width cm 84 98 Height cm 47 70 Depth cm 33 33

+5 °C to 40 °C

Air humidity max. 80 % rh (non-condensing) Overvoltage category II Pollution degree 2 Altitude of installation max. 2,000 m above sea level Maximum mains voltage fluctuations AC 115 (+/– 10 %) AC 230 (+/– 10 %) ■ The appliance may only be used in enclosed areas and under the ambient conditions listed below ■ The appliance may not be used in areas where there is a risk of explosion. The sur- rounding air must not contain any explosive dusts, gases, vapours or gas-air mixtures. The appliance is not explosion-proof. ■ Heavy dust production or aggressive vapours in the vicinity of the appliance could lead to sedimentation in the interior and, as a consequence, could result in short circuits or damage to electrical parts. For this reason, sufficient measures to prevent large clouds of dust or aggressive vapours from developing should be taken. 5.1.11 Scope of delivery Standard ■ Shaking Device ■ Recording frame Optional ■ Support basket ■ Sloping lid

Delivery, Transport and Setting Up

5.2.1 Safety WARNING Condensation in the electrical components may cause short circuits. After transporting or storing the appliance under humid conditions, remove it from its pack- aging and let it ventilate for at least 24 hours in normal ambient conditions. Do not connect the appliance to the mains power during this time. CAUTION You may suffer crush injuries to your hands when transporting and setting up the accessories. Always wear protective gloves. Grip the appliance downwards only at the sides CAUTION Danger of cuts due to sharp edges. Always wear gloves when working on the appliance. 5.2.2 Delivery/unpacking ■ The accessories are packed in a cardboard box ■ Do not unpack the accessories until they are in their installation location to avoid dam- age ■ Carefully cut open the cardboard packaging along one edge. 5.2.3 Product control ■ Check the delivery note to ensure that the scope of delivery is complete ■ Check the accessories for damage If you notice deviations from the scope of delivery, damage or irregularities, do not put the accessories into operation, but inform the carrier and the manufacturer's factory. 5.2.4 Disposing of packaging material ■ Dispose of the packaging material (cardboard, wood, foil) in accordance with the appli- cable disposal regulations for the respective material in your country

The appliance may fall from the installation location due to unevenness and injure you or

someone else. Always place the appliance on an even, load-bearing and non-slip sur- face. CAUTION The surfaces of the appliance can overheat during operation and overheat heat-sen- sitive surfaces in close proximity to the appliance. Always place the appliance on a heat-resistant surface and maintain the specified clearance around the appliance. The distance between the wall and the rear of the appliance must be at least 15 cm. The clearance from the ceiling must not be less than 75 cm and the side clearance from walls or nearby appliances must not be less than 10 cm. Sufficient air circulation in the vicinity of the appliance must be guaranteed at all times. 10 cm 10 cm 75 cm 10 cm 15 cm

Assembly shaking device

1. Loosen the blind plugs at the fixing points (on the side) for the shaking device on the water bath. 2. Place the holding frame on the side of the water bath and fix it with screws. 3. Place the shaking device on the water bath frame from above.

4. Fix the shaking device to the mounting frame with screws.

5. The support basket is inserted into the shaking device and fixed with knurled screws. 6. Insert the support basket from above into the holders of the shaking device. 7. Fix the support basket to the shaker with knurled screws.

Putting into operation

5.3.1 Putting into operation for the first time NOTICE When putting into operation the accessories for the first time, please do not leave them unattended until they have reached a steady state. 5.3.2 Connect the appliance to the power supply Place the power cable so that ■ it does not represent a trip hazard ■ it cannot come into contact with any hot parts ■ it is easily accessible at all times and can be pulled off quickly, for example in case of interference or an emergency 5.3.3 Filling the water bath WARNING If the water bath level is too high, the water may spill over during the operation of the shaker. This can cause burns. Adjust the level of the water bath to the feed of the shaker and the shaking frequency. Please note that the maximum filling level of the water bath is reduced by the attach- ment of the shaking device and the chamber load The water bath must be filled so that the liquid level does not exceed the maximum fill- ing level, including the shaking device. This is marked by the edge on the water bathtub. MIN MAX

min. 4 cm

11 15 min. 2 cm 24 35 50 NOTICE Depending on the loading of the shaking device and the selected shaking frequency, the water bath filling may spill over. Here, the shaking frequency or the filling level must be adjusted without falling below the minimum filling level. 5.3.4 Switching on appliance Press main switch on the back of the appliance.

Operation and control

WARNING During the operation of the shaker, individual components may heat up. Touching its surface can cause burns. Use temperature-resistant protective gloves. CAUTION As the shaker is a moving part, you may suffer crushing injuries while operating the shaker. Load the basket before starting the shaker, and do not reach into the appliance during operation. NOTICE The mounting frame is only intended for placing the shaker. Please place no other appliances or parts of appliances on the frame and not support them. 5.4.1 Quick shutdown in case of overload The appliance has an overload-controlled quick shut-off for safety. You can tell this safety device has responded when the error code - 139 is shown on the touch display (see chapter Störungen, Warn- & Fehlermeldungen). Here, it must be checked whether an overload or an object has been jammed in the device. After the overload has been removed, the shaker can be restarted by tapping on the display. 5.4.2 Home screen CustomView will appear first after switching the appliance on (4.4.4) To access the home screen, touch any spot on the TouchDisplay ► To set the set shaking frequency, select the set display Set . ► Select the symbol to set the timer ► To start the shaking frequency, select . ► Touch the symbol for other appliance settings. 00:00 h Set 30 rpm0 rpm

Setting the shaking frequency and time

149 rpm 150 Set the individual values by scrolling vertically on the touch display: ► Entering the shaking frequency in rpm ► Set time for the timer in minutes/hours Press to confirm entry. The setting is not saved with 5.4.4 Settings Info CustomView Other settings are possible under the Settings menu item : ► Appliance information overview ► CustomView setting Press to return to the home screen 5.4.5 CustomView Under the Settings menu item, select CustomView to set up the individual display Selecting the set shaking frequency and timer time activates or deactivates the respective element. The selected element will be greyed out when deactivated 00:50 h Set 30 rpm0 rpm

Tapping the set shaking frequency and timer time activates or deactivates the

respective element. Set shaking frequency and timer time are deactivated here confirms the selection; changes will not be accepted If the selection was confirmed with , tap to return to the Home screen, after 30 seconds, the personalised CoustomView appears 5.4.6 Information SV3550 - V1.0.0 - 20210510 0.0.0 Memmert GmbH + Co. KG Äußere Rittersbacher Straße 38 91126 Schwabach / Germany [email protected] Phone: +49 9122 9250 Under menu item Settings, click on Info to view the item information of the shaking device 00:50 h Set 30 rpm0 rpm 100 rpm

Faults, Warning & Error Messages

CAUTION If the shaking device is used strangely, for example, uncontrolled shaking, you may suffer bruising injuries to your hands. Disconnect the appliance from the power supply by pulling out the mains plug. Description Cause Action Motor lock-up error ■ Restart movement, ■ Reduce load. E-111 130 rpm Software error ■ Restart appliance ■ Contact Service in case of frequent occurrence E-132 130 rpm Memory access error ■ Restart appliance ■ Contact Service in case of frequent occurrence E-135 130 rpm CRC error (internal memory error) ■ Restart appliance ■ Contact Service in case of frequent occurrence E-139 130 rpm

Cleaning and decontamination

DANGER Danger due to electric shock. Penetration of liquid into the appliance can cause electric shocks and short circuits. – Protect the appliance from splashing water. – Before cleaning and maintenance work, switch off the appliance and disconnect the mains plug. – The appliance must not be cleaned and disinfected wet. Allow the appliance to dry completely before putting into operation again. CAUTION Danger of cuts due to sharp edges. Always wear gloves when working on the appliance. 5.6.1 Cleaning Interior and metal surfaces The metal surfaces of the shaker can be cleaned with commercially available stainless steel cleaning agents. Make sure that no rusting objects come into contact with the stain- less steel housing. Rust deposits can lead to an infection of the stainless steel. If rust spots appear due to contamination, the affected areas must be cleaned and polished immediately. Plastic parts Do not clean plastic parts of the water bath with caustic or solvent-based cleaning agents. Glass surfaces Glass surfaces can be cleaned with a commercially available glass cleaner.

Skin and eye contact with decontaminants may cause irritation of the hands and eyes

or allergic reactions and chemical burns. – Wear chemical-resistant gloves during decontamination. – For correct application, follow the instructions for use on the respective container of the decontamination agent Memmert appliances can become contaminated by loading infectious substances or by environmental influences at the place of installation. All parts of the affected appliance must be decontaminated both before sending to our service and before disposing of the appliance. ■ For decontamination, we recommend commercially available decontaminants based on isopropanol or ethanol. If you have any questions about using a decontamination agent for decontaminating Memmert appliances, please contact our service department. After cleaning and decontamination, ventilate the installation site and allow the appli- ance to dry completely. 5.7 Repairs and service DANGER Danger due to electric shock. Penetration of liquid into the appliance can cause electric shocks and short circuits. – Protect the appliance from splashing water. – Before cleaning and maintenance work, switch off the appliance and disconnect the mains plug. – The appliance must not be cleaned and disinfected wet. Allow the appliance to dry completely before putting into operation again. WARNING Live components may be exposed once the covers have been removed. Touch- ing these can lead to an electric shock. Disconnect the mains plug before re- moving any covers. Work on the electrical system must only be performed by qualified electricians. Repairs and service work are described in a separate service manual.

Storage and disposal

5.8.1 Storage The appliance may only be stored under the following conditions: ■ in a dry and enclosed, dust-free room ■ frost-free ■ disconnected from the power supply 5.8.2 Disposal This product is subject to Directive 2012/19/EC on Waste Electrical and Electronic Equipment (WEEE) of the European Parliament and of the Council of Ministers. This ap- pliance was placed on the market after 13 August 2005 in countries which have already integrated this Directive into their national laws. It may not be disposed of in normal household waste. For disposal, please contact your dealer or the manufacturer. Any appliances that are infected, infectious or contaminated with materials hazardous to health are excluded from return. Please also observe all other regulations applicable in this context.

Pumping device CPM

6. Pumping device CPM 6.1 Assembly and description 6.1.1 Design Description Overview front 1. Pumping device 1 2 3 2. Housing fan 3. Water bath 6.1.2 Description brief technical description By using a pumping device, the water can circulate in the tub. The delivery volume is 1.2 litres per minute. The temperature control of the WTB water bath is optimised for standing water. The operation of a CPM650 circulation pump may require readjustment of the temperature control, as flowing water has a higher heat transfer coefficient than standing water. See the general operating manual for the WTB water bath. A warranty of the installed pump is only given when using the pumping device up to 80°C water temperature. If the pumping equip- ment is used at temperatures above 80°C, no warranty can be claimed if damage occurs or a defect to the pump. Description Pumping device 1. Water bath 1 2 3 4 5 2. Water bath pumping device CPM650 3. Water return from pumping device to water bath 4. Water inlet from the water bath to the pumping device NOTICE NOTICE

Pumping device CPM

6.1.3 Intended use The Memmert circulation pump is intended exclusively for use with a Memmert water bath WTB. It is used for the internal circulation of the temperature control medium in the water bath. 6.1.4 Use not in accordance with the intended purpose The circulation pump is used exclusively for internal circulation in the water bath and is not intended for temperature control of external circuits. The circulation pump must be used only for pumping water and must not be operated with other liquids. 6.1.5 Material This pumping device complies with the current requirements of the RoHS Directive. You can find more information on this and material compliance on our homepage. https://www.memmert.com/de/downloads/ Components Material Circulating pump housing Stainless steel 1.4301 - ASTM 304 Screws and hinges Stainless steel 1.4301 - ASTM 304 Silicone tube Silicone - Shore A translucent i The chamber load for the appliance must be carefully checked for chemical compati- bility with the materials mentioned. 6.1.6 Electrical equipment ■ Operating voltage and current consumption: See name plate or technical data ■ Protection class I, i.e. operating insulation with PE conductor in accordance with EN 60664-1 ■ Protection type IP 20 acc. to EN 60 529 ■ Interference suppression acc. to EN 55011 class B ■ Appliance fuse: Fuse F15H250V 6.1.7 Connections and interfaces This appliance is intended for operation on an electrical power system with a system impedance Zmax at the point of transfer (service line) of a maximum of 0.292 Ohm. The operator must ensure that the appliance is operated only on an electrical power system that meets these requirements. If necessary, you can ask your local energy supply com- pany what the system impedance is. When connecting, observe the country-specific regulations (e.g. in Germany DIN VDE 0100 withFI protective circuit).

Pumping device CPM

6.1.8 Designation (nameplate) Description Overview 1. Type designation GmbH+Co.KG D-91126 Schwabach FRG Äußere Rittersbacher Str. 38 Made in Germany memmert Schutzart DIN EN 60529 - IP 20 1 2 3 4 5 6 8 7 Typ: CPM650 ~ 230 V 50 Hz F-Nr.: J221.0004 0,43 A 43 W 2. Operating voltage 3. Protection type 4. CE conformity 5. Address of manufacturer 6. Disposal note 7. Connection / power ratings 8. Appliance number 6.1.9 Technical data - Pumping device Zubehör CPM650 Dimensions Length mm 296 Width mm 140 Height mm 180 electrical data Conveyance volume l/min 1,2 Operating voltage V 230 Power W 43 Current consumption 230V/50 Hz A 0,43 Packaging material Net weight kg 2,8 Gross weight (in cardboard box) kg 3,7 Width cm 47 Height cm 42 Depth cm 22

Pumping device CPM

6.1.10 Ambient conditions Ambient temperature +5 °C to 40 °C Air humidity max. 80 % rh (non-condensing) Overvoltage category II Pollution degree 2 Altitude of installation max. 2,000 m above sea level Maximum mains voltage fluctuations AC 230 (+/– 10 %) ■ The appliance may only be used in enclosed areas and under the ambient conditions listed below ■ The appliance may not be used in areas where there is a risk of explosion. The sur- rounding air must not contain any explosive dusts, gases, vapours or gas-air mixtures. The appliance is not explosion-proof. ■ Heavy dust production or aggressive vapours in the vicinity of the appliance could lead to sedimentation in the interior and, as a consequence, could result in short circuits or damage to electrical parts. For this reason, sufficient measures to prevent large clouds of dust or aggressive vapours from developing should be taken. 6.1.11 Scope of delivery and required accessories ■ Pumping device ■ Installation and connection set (consisting of inlet and outlet pipe, two hinges)

Pumping device CPM

6.2 Delivery, Transport and Setting Up 6.2.1 Safety WARNING Condensation in the electrical components may cause short circuits. After transport- ing or storing the appliance under humid conditions, remove it from its packaging and let it ventilate for at least 24 hours in normal ambient conditions. Do not connect the appliance to the mains power during this time. CAUTION You may suffer crushing injuries to your hands when transporting and setting up the appliance. Always wear protective gloves. 6.2.2 Delivery/unpacking ■ The accessories are packed in a cardboard box ■ Do not unpack the accessories until they are in their installation location to avoid damage ■ Carefully cut open the cardboard packaging along one edge. 6.2.3 Product control ■ Check the delivery note to ensure that the scope of delivery is complete ■ Check the accessories for damage If you notice deviations from the scope of delivery, damage or irregularities, do not put the accessories into operation, but inform the carrier and the manufacturer's factory. 6.2.4 Disposing of packaging material ■ Dispose of the packaging material (cardboard, wood, foil) in accordance with the appli- cable disposal regulations for the respective material in your country

Pumping device CPM

6.2.5 Setting up WARNING The tube is supplied in a predefined length. This is possibly too long. Please make sure that the hoses do not hang down. You could get caught during operation and be scalded or burned by the heated temperature control medium. CAUTION The appliance may fall from the installation location due to unevenness and injure you or someone else. Always place the appliance on an even, load-bearing and non-slip surface. CAUTION The surfaces of the appliance can overheat during operation and overheat heat-sen- sitive surfaces in close proximity to the appliance. Always place the appliance on a heat-resistant surface and maintain the specified clearance around the appliance. The distance between the wall and the rear of the appliance must be at least 15 cm. The clearance from the ceiling must not be less than 75 cm and the side clearance from walls or nearby appliances must not be less than 10 cm. Sufficient air circulation in the vicinity of the appliance must be guaranteed at all times. memmert 10 cm 10 cm 10 cm 15 cm 75 cm 10 cm

Pumping device CPM

6.2.6 Installation options pumping device The pumping device can be placed either next to or behind the water bath . When placing the pumping device, take care not to kink the hoses. When setting up the pumping unit, make sure that ■ the specified wall distances are observed. ■ the hoses of the pumping device to the water bath are cut as short as possible. NOTICE Determine the installation location before you install the pumping device in the water bath. The water bath and pumping device may only be transported individually. 6.2.7 Installation pumping device 1. Attach brackets for the pipe connection to the back of the water bath. 2. Attach the pipe connections to the brackets and hang them in the water bath.

Pumping device CPM

Pump connection IN (suction line): Attach the tube to the long pipe connec- tion on the water bath, as this is where the water is conveyed from the water bath to the pumping device. Pump connection OUT (pressure line): Attach the tube to the short pipe connection on the water bath, as this is where the water from the pump is conveyed back to the water bath. 3. Fit the hoses onto the pipe connections NOTICE

Pumping device CPM

6.3 Putting into operation i There must not be any coarse-grained particles in the water bath. This could damage the pumping device. 6.3.1 Putting into operation for the first time NOTICE When putting the appliance into operation for the first time, do not leave it unattended until it has reached a steady state. 6.3.2 Connect the appliance to the power supply Place the power cable so that ■ it does not represent a trip hazard ■ it cannot come into contact with any hot parts ■ it is easily accessible at all times and can be pulled off quickly, for example in case of interference or an emergency 6.3.3 Filling the water bath NOTICE To operate the pumping device, a minimum filling level of 70 mm is necessary. When filling and charging the water bath, please pay attention to the maximum filling height, which must not be exceeded. memmert MIN MAX

Pumping device CPM

6.4 Operation and control WARNING During operation of the pumping device, individual components may heat up. Touching its surface can cause burns. Use temperature-resistant protective gloves. 6.4.1 Switching on appliance Press main switch on the back of the appliance. The pumping capacity of the pumping device is constant and cannot be regulated. While switched on, the pumping device operates at a steady rate. The temperature is controlled via the setting on the water bath. 6.4.2 Function test ■ To check that the unit is pumping, pull off the suction tube from the tube suspender and place a finger on it to test for suction. If there is no suction effect: See chapter Faults, warning and error messages. 6.5 Dismantling the hoses WARNING During operation of the pumping device, individual components may heat up. Touching its surface can cause burns. Let the heated components cool down first. 1. Allow the water�in�the�water�bath�and�the�hoses�to�cool�down�first. 2. Drain the water from the water bath using the drain tap until�the�raw�joints�are�exposed. 3. Pull the hoses off the pipe connections NOTICE

Pumping device CPM

6.6 Faults, Warning & Error Messages WARNING During operation of the pumping device, individual components may heat up. Touch- ing its surface can cause burns. Use temperature-resistant protective gloves. Description Cause Action CPM cannot be switched off Main switch defective ■ Disconnect the mains plug ■ Contact customer service Tempering medium circulates little or not at all ■ Water level too low ■ Pump defective ■ Monitor water level ■ Contact customer service

Pumping device CPM

6.7 Cleaning and decontamination DANGER Danger due to electric shock. Penetration of liquid into the appliance can cause electric shocks and short circuits. – Protect the appliance from splashing water. – Before cleaning and maintenance work, switch off the appliance and disconnect the mains plug. – The appliance must not be cleaned and disinfected wet. Allow the appliance to dry completely before putting into operation again. CAUTION Danger of cuts due to sharp edges. Always wear gloves when working on the appliance. 6.7.1 Cleaning Interior and metal surfaces Conventional stainless steel cleaning agents can be used to clean the metal surfaces of the pump unit. Make sure that no rusting objects come into contact with the stain- less steel housing. Rust deposits can lead to an infection of the stainless steel. If rust spots appear due to contamination, the affected areas must be cleaned and polished immediately. Plastic parts Do not clean plastic parts with abrasive or solvent-based cleaning agents. 6.7.2 Decontamination CAUTION Skin and eye contact with decontaminants may cause irritation of the hands and eyes or allergic reactions and chemical burns. – Wear chemical-resistant gloves during decontamination. – For correct application, follow the instructions for use on the respective contain- er of the decontamination agent Memmert appliances can become contaminated by loading infectious substances or by environmental influences at the place of installation. All parts of the affected appliance must be decontaminated both before sending to our service and before disposing of the appliance. ■ For decontamination, we recommend commercially available decontaminants based on isopropanol or ethanol. If you have any questions about using a decontamination agent for decontaminating Memmert appliances, please contact our service department. After cleaning and decontamination, ventilate the installation site and allow the appli- ance to dry completely.

Pumping device CPM

6.8 Repairs and service DANGER Danger due to electric shock. Penetration of liquid into the appliance can cause electric shocks and short circuits. – Protect the appliance from splashing water. – Before cleaning and maintenance work, switch off the appliance and disconnect the mains plug. – The appliance must not be cleaned and disinfected wet. Allow the appliance to dry completely before putting into operation again. WARNING Live components may be exposed once the covers have been removed. Touch- ing these can lead to an electric shock. Disconnect the mains plug before re- moving any covers. Work on the electrical system must only be performed by qualified electricians. Repairs and service work are described in a separate service manual. 6.9 Storage and disposal 6.9.1 Storage The accessories may only be stored under the following conditions: ■ in a dry and enclosed, dust-free room ■ frost-free ■ disconnected from the power supply Before storage, empty the hoses of the pumping device. 6.9.2 Disposal This product is subject to Directive 2012/19/EC on Waste Electrical and Electronic Equipment (WEEE) of the European Parliament and of the Council of Min- isters. This appliance was placed on the market after 13 August 2005 in countries which have already integrated this Directive into their national laws. It may not be disposed of in normal household waste. For disposal, please contact your dealer or the manufacturer. Any appliances that are infected, infectious or contaminated with materials hazardous to health are excluded from return. Please also observe all other regulations applicable in this context.

Cooling unit CDP

7. Cooling unit CDP 7.1 Assembly and description 7.1.1 Design Description Overview 1. Peltier cooling unit with fan 1 2 3 4 5 2. Housing fan 3. On/Off switch 4. Mains lead 5. Water bath WTB15 7.1.2 Description Brief technical description The CDP water bath cooling unit enables the water in Memmert water baths to be cooled to temperatures below room temperature. Water is pumped out of the water bath, cooled by a Peltier module and then returned to the water bath. The cooling power is constant and cannot be regulated. The temperature control of the water is taken over by the water bath; its controller automatically takes into account the cooling power of the cooling unit. The cooling unit's low-temperature fuse automaticaly switches off the appliance at a water temperature of approx. 3 ºC. Functional principle of the water bath cooling unit CDP115 Description Overview front 1. Water bath 1 2 3 4 5 2. Water bath cooling unit CDP115 3. Peltier cooling module 4. Water return from cooling unit to the water bath 5. Water supply from the water bath to the cooling unit

Cooling unit CDP

7.1.3 Intended use The Memmert cooling unit is intended exclusively for use with a Memmert water bath WTB. It is used to cool the temperature control medium in the water bath to tempera- tures below room temperature from +10 °C. 7.1.4 Use not in accordance with the intended purpose The cooling unit is used exclusively for internal cooling in the water bath and is not in- tended for temperature control of external circuits. The water bath cooling unit must be used only for pumping and cooling water and must not be operated with other liquids. 7.1.5 Material This cooling unit complies with the current requirements of the RoHS Directive. You can find more information on this and on material compliance on our homepage www. memmert.com. Components Material Cooling unit housing Stainless steel 1.4301 - ASTM 304 Peltier cooling module Anodised aluminium Screws and hinges Stainless steel 1.4301 - ASTM 304 Inlet and outlet tube Chloroprene tube LEZ-NEO 4x2 inside ø4, outside ø8 i The chamber load for the appliance must be carefully checked for chemical compati- bility with the materials mentioned. 7.1.6 Electrical equipment ■ Operating voltage and current consumption: See name plate or technical data ■ Protection class I, i.e. operating insulation with PE conductor in accordance with EN 60664-1 ■ Protection type IP 20 acc. to EN 60 529 ■ Interference suppression acc. to EN 55011 class B ■ Appliance fuse: Fuse F15H250V ■ Undertemperature monitor as frost protection: Switching off the cooling unit at < +5 °C 7.1.7 Connections and interfaces This appliance is intended for operation on an electrical power system with a system impedance Zmax at the point of transfer (service line) of a maximum of 0.292 Ohm. The operator must ensure that the appliance is operated only on an electrical power system that meets these requirements. If necessary, you can ask your local energy supply com- pany what the system impedance is. When connecting, observe the country-specific regulations (e.g. in Germany DIN VDE 0100 withFI protective circuit).

Cooling unit CDP

7.1.8 Designation (nameplate) Description Overview 1. Type designation GmbH+Co.KG D-91126 Schwabach FRG Äußere Rittersbacher Str. 38 Made in Germany memmert Schutzart DIN EN 60529 - IP 20 1 2 3 4 5 6 8 7 Typ: CDP115 ~ 230 V 50/60 Hz F-Nr.: J121.0001 0,7 A 160 W 2. Operating voltage 3. Protection type 4. CE conformity 5. Address of manufacturer 6. Disposal note 7. Connection / power ratings 8. Appliance number 7.1.9 Technical data - Cooling unit Zubehör CPD115 Dimensions Length mm 456 Width mm 205 Height mm 435 Electrical data Conveyance volume l/min 0,6 Effective cooling capacity W 113 Electrical connection V 115/230 Power W 160 Current consumption 50/60 Hz 115 V A 0,7 230 V 1,4 Packaging material Net weight kg 8,35 Gross weight (in cardboard box) kg 9,95 Width cm 59 Height cm 41 Depth cm 33

Cooling unit CDP

7.1.10 Ambient conditions Ambient temperature +5 °C to 40 °C Air humidity max. 80 % rh (non-condensing) Overvoltage category II Pollution degree 2 Altitude of installation max. 2,000 m above sea level Maximum mains voltage fluctuations AC 115 (+/– 10 %) AC 230 (+/– 10 %) ■ The appliance may only be used in enclosed areas and under the ambient conditions listed below ■ The appliance may not be used in areas where there is a risk of explosion. The sur- rounding air must not contain any explosive dusts, gases, vapours or gas-air mixtures. The appliance is not explosion-proof. ■ Heavy dust production or aggressive vapours in the vicinity of the appliance could lead to sedimentation in the interior and, as a consequence, could result in short circuits or damage to electrical parts. For this reason, sufficient measures to prevent large clouds of dust or aggressive vapours from developing should be taken. 7.1.11 Scope of delivery and required accessories ■ Cooling unit ■ Installation and connection set (consisting of inlet and outlet pipe, two hinges)

Cooling unit CDP

7.2 Delivery, Transport and Setting Up 7.2.1 Safety WARNING Condensation in the electrical components may cause short circuits. After transport- ing or storing the appliance under humid conditions, remove it from its packaging and let it ventilate for at least 24 hours in normal ambient conditions. Do not connect the appliance to the mains power during this time. CAUTION You may suffer crushing injuries to your hands when transporting and setting up the appliance. Always wear protective gloves. 7.2.2 Delivery/unpacking ■ The accessories are packed in a cardboard box ■ Do not unpack the accessories until they are in their installation location to avoid damage ■ Carefully cut open the cardboard packaging along one edge. 7.2.3 Product control ■ Check the delivery note to ensure that the scope of delivery is complete ■ Check the accessories for damage If you notice deviations from the scope of delivery, damage or irregularities, do not put the accessories into operation, but inform the carrier and the manufacturer's factory. 7.2.4 Disposing of packaging material ■ Dispose of the packaging material (cardboard, wood, foil) in accordance with the appli- cable disposal regulations for the respective material in your country

Cooling unit CDP

7.2.5 Setting up WARNING The tube is supplied in a predefined length. This is possibly too long. Please make sure that the hoses do not hang down. You could get caught during operation and suffer scalds or burns from the heated water. CAUTION The appliance may fall from the installation location due to unevenness and injure you or someone else. Always place the appliance on an even, load-bearing and non-slip surface. CAUTION The surfaces of the appliance can overheat during operation and overheat heat-sen- sitive surfaces in close proximity to the appliance. Always place the appliance on a heat-resistant surface and maintain the specified clearance around the appliance. The distance between the wall and the rear of the appliance must be at least 15 cm. The clearance from the ceiling must not be less than 75 cm and the side clearance from walls or nearby appliances must not be less than 10 cm. Sufficient air circulation in the vicinity of the appliance must be guaranteed at all times. NOTICE Determine the installation location before you start installing the cooling unit to the water bath. The water bath and cooling unit may only be transported individually. 10 cm 10 cm 15 cm 10 cm 75 cm

Cooling unit CDP

7.2.6 Installation cooling unit 1. Attach brackets for the pipe connection to the back of the water bath. 2. Attach the pipe connections to the brackets and hang them in the water bath. NOTICE Pump connection IN (suction line): Attach the tube to the long pipe connec- tion on the water bath, as this is where the water is conveyed from the water bath to the pumping device. Pump connection OUT (pressure line): Attach the tube to the short pipe connection on the water bath, as this is where the water from the pump is conveyed back to the water bath.

Cooling unit CDP

3. Put the hoses on the pipe connections and connect them to the CDP115. 7.2.7 Installation options Cooling unit The cooling unit can be placed either next to or behind the water bath. When placing the cooling unit, take care not to kink the hoses. When setting up the cooling unit, make sure that ■ the specified wall distances are observed. ■ The hoses of the cooling unit to the water bath are cut as short as possible. NOTICE Determine the installation location before you start installing the cooling unit to the water bath. The water bath and cooling unit may only be transported individually.

Cooling unit CDP

7.3 Putting into operation i There must not be any coarse-grained particles in the water bath. This could damage the pumping device. 7.3.1 Putting into operation for the first time NOTICE When putting the appliance into operation for the first time, do not leave it unattended until it has reached a steady state. 7.3.2 Connect the appliance to the power supply Place the power cable so that ■ it does not represent a trip hazard ■ it cannot come into contact with any hot parts ■ it is easily accessible at all times and can be pulled off quickly, for example in case of interference or an emergency 7.3.3 Filling the water bath NOTICE To operate the cooling unit, a minimum filling level of 70 mm is necessary. When filling and charging the water bath, please pay attention to the maximum filling height, which must not be exceeded. memmert MIN MAX

Cooling unit CDP

7.4 Operation and control WARNING During operation of the cooling unit, individual components may heat up. Touching its surface can cause burns. Use temperature-resistant protective gloves. 7.4.1 Switching on appliance Press the main switch on the side of the unit. The cooling power of the cooling unit is constant and cannot be regulated. The cool- ing unit works at a constant power as long as it is switched on. The temperature is controlled via the setting on the water bath. Do not operate the cooling unit when the water bath heater is switched off. This could trigger the automatic shutdown . If the set temperature is above room temperature, it is recommended to switch off the cooling unit. 7.4.2 Automatic switch-off The cooling unit has a low temperature fuse which switches off the unit automatically when the water temperature is at about 3ºC, to prevent icing over and possible damage to the cooling element. When the water temperature reaches about 5 ºC, the cooling unit switches back on again. 7.4.3 Function test ■ To check that the unit is pumping, pull off the suction tube from the tube suspender and place a finger on it to test for suction. If there is no suction effect: See next chapter, Troubleshooting. 7.5 Dismantling the hoses 1. Allow the water�in�the�water�bath�and�the�hoses�to�cool�down�first. 2. Drain the water from the water bath using the drain tap until�the�raw�joints�are�exposed. 3. Pull the hoses off the pipe connections NOTICE

Cooling unit CDP

7.6 Faults, Warning & Error Messages WARNING During operation of the pumping device, individual components may heat up. Touch- ing its surface can cause burns. Use temperature-resistant protective gloves. Description Cause Action CDP115 cannot be switched off Main switch defective ■ Disconnect the mains plug ■ Contact customer service Tempering medium circulates little or not at all ■ Water level too low ■ Pump defective ■ Monitor water level ■ Contact customer service

Cooling unit CDP

7.7 Cleaning and decontamination DANGER Danger due to electric shock. Penetration of liquid into the appliance can cause electric shocks and short circuits. – Protect the appliance from splashing water. – Before cleaning and maintenance work, switch off the appliance and disconnect the mains plug. – The appliance must not be cleaned and disinfected wet. Allow the appliance to dry completely before putting into operation again. CAUTION Danger of cuts due to sharp edges. Always wear gloves when working on the appliance. 7.7.1 Cleaning Metal surfaces The metal surfaces of the cooling unit can be cleaned with commercially available stainless steel cleaning agents. Make sure that no rusting objects come into contact with the stainless steel housing. Rust deposits can lead to an infection of the stainless steel. If rust spots appear due to contamination, the affected areas must be cleaned and polished immediately. Plastic parts Do not clean plastic parts with abrasive or solvent-based cleaning agents. 7.7.2 Decontamination CAUTION Skin and eye contact with decontaminants may cause irritation of the hands and eyes or allergic reactions and chemical burns. – Wear chemical-resistant gloves during decontamination. – For correct application, follow the instructions for use on the respective contain- er of the decontamination agent Memmert appliances can become contaminated by loading infectious substances or by environmental influences at the place of installation. All parts of the affected appliance must be decontaminated both before sending to our service and before disposing of the appliance. ■ For decontamination, we recommend commercially available decontaminants based on isopropanol or ethanol. If you have any questions about using a decontamination agent for decontaminating Memmert appliances, please contact our service department. After cleaning and decontamination, ventilate the installation site and allow the appliance to dry completely.

Cooling unit CDP

7.8 Repairs and service DANGER Danger due to electric shock. Penetration of liquid into the appliance can cause electric shocks and short circuits. – Protect the appliance from splashing water. – Before cleaning and maintenance work, switch off the appliance and disconnect the mains plug. – The appliance must not be cleaned and disinfected wet. Allow the appliance to dry completely before putting into operation again. WARNING Live components may be exposed once the covers have been removed. Touch- ing these can lead to an electric shock. Disconnect the mains plug before re- moving any covers. Work on the electrical system must only be performed by qualified electricians. Repairs and service work are described in a separate service manual. 7.9 Storage and disposal 7.9.1 Storage The accessories may only be stored under the following conditions: ■ in a dry and enclosed, dust-free room ■ frost-free ■ disconnected from the power supply Before storage, empty the hoses of the cooling unit. 7.9.2 Disposal This product is subject to Directive 2012/19/EC on Waste Electrical and Electronic Equipment (WEEE) of the European Parliament and of the Council of Ministers. This appliance was placed on the market after 13 August 2005 in countries which have al- ready integrated this Directive into their national laws. It may not be disposed of in normal household waste. For disposal, please contact your dealer or the manufacturer. Any appliances that are infected, infectious or contaminated with materials hazardous to health are excluded from return. Please also observe all other regulations applicable in this context.

CE declarations of conformity Shaking device SV1524 | SV3550

EC Declaration of Conformity Manufacturer´s name and address: Memmert GmbH + Co. KG Äussere Rittersbacherstraße 38 D-91126 Schwabach Deutschland Authorized representative for compiling the technical documents: Memmert GmbH + Co.KG Äußere Rittersbacher Straße 38 D-91126 Schwabach Deutschland Product: Shaking device Type: SV…/1524/3550 Including all accessories Nominal voltage: AC 230 V 50/60 Hz, alternative AC 115 V 50/60 Hz This product complies with the provisions of the directive 2006/42/EC DIRECTIVE 2006/42/EC OF THE EUROPEAN PARLIAMENT AND OF THE COUNCIL of 17 May 2006 on machinery, and amending Directive 95/16/EC (recast) Full compliance with the standard below proves the conformity of the designated product with the essential protection requirements of the above-mentioned EC Directive: EN ISO 12100:2010 The designated product is in conformity with the European EMC-Directive 2014/30/EU including amendments Council Directive of 03 May 1989 on the approximation of the laws of the Member States relating to electromagnetic compatibility Full compliance with the standards listed below proves the conformity of the designated product with the essential protection requirements of the above-mentioned EC Directive: DIN EN 61326-1:213-07 EN 61326-1:2013 The object described in the declaration above corresponds to the directive 2011/65/EC of the European Parliament and the Council of 8 June 2011on the restriction of the use of certain hazardous substances in electrical and electronic equipment. Schwabach, 04.06.2021 Legally binding signature of the issuer: Christiane Riefler-Karpa, managing director This declaration certifies compliance with the above mentioned directives but does not include a property assurance. The safety note given in the product documentation which are part of the supply, must be observed. D49191

CE declarations of conformity Pumping device CPM650

EC Declaration of Conformity Manufacturer´s name and address: Memmert GmbH + Co. KG Äussere Rittersbacherstraße 38 D-91126 Schwabach Deutschland Authorized representative for compiling the technical documents: Memmert GmbH + Co.KG Äußere Rittersbacher Straße 38 D-91126 Schwabach Deutschland Product: Pump device Type: CPM650 Including all accessories Nominal voltage: AC 230 V 50/60 Hz, alternative AC 115 V 50/60 Hz This product complies with the provisions of the directive 2006/42/EC DIRECTIVE 2006/42/EC OF THE EUROPEAN PARLIAMENT AND OF THE COUNCIL of 17 May 2006 on machinery, and amending Directive 95/16/EC (recast) Full compliance with the standard below proves the conformity of the designated product with the essential protection requirements of the above-mentioned EC Directive: EN ISO 12100:2010 The designated product is in conformity with the European EMC-Directive 2014/30/EU including amendments Council Directive of 03 May 1989 on the approximation of the laws of the Member States relating to electromagnetic compatibility Full compliance with the standards listed below proves the conformity of the designated product with the essential protection requirements of the above-mentioned EC Directive: DIN EN 61326-1:213-07 EN 61326-1:2013 The object described in the declaration above corresponds to the directive 2011/65/EC of the European Parliament and the Council of 8 June 2011on the restriction of the use of certain hazardous substances in electrical and electronic equipment. Schwabach, 04.06.2021 Legally binding signature of the issuer: Christiane Riefler-Karpa, managing director This declaration certifies compliance with the above mentioned directives but does not include a property assurance. The safety note given in the product documentation which are part of the supply, must be observed. D49199

CE declarations of conformity Cooling unit CDP115

EC Declaration of Conformity Manufacturer´s name and address: Memmert GmbH + Co. KG Äussere Rittersbacherstraße 38 D-91126 Schwabach Deutschland Authorized representative for compiling the technical documents: Memmert GmbH + Co.KG Äußere Rittersbacher Straße 38 D-91126 Schwabach Deutschland Product: Peltier-cooling device Type: CDP115 Including all accessories Nominal voltage: AC 230 V 50/60 Hz, alternative AC 115 V 50/60 Hz This product complies with the provisions of the directive 2006/42/EC DIRECTIVE 2006/42/EC OF THE EUROPEAN PARLIAMENT AND OF THE COUNCIL of 17 May 2006 on machinery, and amending Directive 95/16/EC (recast) Full compliance with the standard below proves the conformity of the designated product with the essential protection requirements of the above-mentioned EC Directive: EN ISO 12100:2010 The designated product is in conformity with the European EMC-Directive 2014/30/EU including amendments Council Directive of 03 May 1989 on the approximation of the laws of the Member States relating to electromagnetic compatibility Full compliance with the standards listed below proves the conformity of the designated product with the essential protection requirements of the above-mentioned EC Directive: DIN EN 61326-1:213-07 EN 61326-1:2013 The object described in the declaration above corresponds to the directive 2011/65/EC of the European Parliament and the Council of 8 June 2011on the restriction of the use of certain hazardous substances in electrical and electronic equipment. Schwabach, 04.06.2021 Legally binding signature of the issuer: Christiane Riefler-Karpa, managing director This declaration certifies compliance with the above mentioned directives but does not include a property assurance. The safety note given in the product documentation which are part of the supply, must be observed. D49207

Memmert GmbH + Co. KG | Äußere Rittersbacher Straße 38 | 91126 Schwabach

Tel. +49 9122 925-0 | E-Mail: [email protected] | www.memmert.com facebook.com/memmert.family | instagram.com/memmert.family Shaking device SV1524/ SV3550 Pumping device CPM650 Cooling unit CDP115 Accessory guide D49169 Last updated 06/2021 englisch