About this manual
Purpose and target audience
This manual describes the construction, function, transport operation and maintenance
of constant climate chambers HPPeco and cooled incubators IPPecoplus. It is intended
for use by trained personnel of the owner, who have the task of operating and/or main-
taining the respective appliance. If you are asked to work on the appliance, read this
manual carefully before starting. Familiarise yourself with the safety regulations. Only
perform work that is described in this manual. If there is something you do not under-
stand, or certain information is missing, ask your manager or contact the manufacturer.
Do not do anything without authorisation.
Versions
The appliances are available in different confi gurations and sizes. If specifi c equipment
features or functions are available only for certain confi gurations, this is indicated at
the relevant points in this manual. The functions described in this manual refer to the
latest fi rmware version. Due to individual confi gurations and sizes, illustrations in this
manual may be slightly different to the actual appearance. Function and operation are
identical.
Other documents to be observed:
■ For operation of the appliance with MEMMERT AtmoCONTROL, observe the separate
software manual. To open the AtmoCONTROL software manual, click on “Help“ in the
AtmoCONTROL menu bar.
■ For service and repair work, observe the separate service manual
Storage and resale
This operating manual belongs with the appliance and should always be stored where
persons working on the appliance have access to it. It is the owner's responsibility to
ensure that persons who are working on or are going to work on the appliance know
where to fi nd the operating manual. We recommend that it is always stored in a protect-
ed location close to the appliance. Make sure that the operating manual is not damaged
by heat or humidity. If the appliance is resold or transported and then set up again at a
different location, the operating manual must remain with it. The current version of this
operating manual in PDF format is also available for download from www.memmert.
com/de/downloads/
Manufacturer address and customer service
MEMMERT GmbH + Co. KG
Willi-Memmert-Str. 90-96
91186 Büchenbach
Germany
Phone: +49 (0)9122 925-0
Fax: +49 (0)9122 14585
E-mail: [email protected]
Internet: www.memmert.com
Customer service:
Hotline: +49 (0)9171 9792 911
Service fax: +49�(0)9171�9792�979
Email: [email protected]
For service enquiries, please always specify the appliance num-
ber given on the type plate.
Terms and signs used .................................................................................................................................................................. 5
1.2
Product safety and dangers ......................................................................................................................................................... 6
1.3
Requirements of the operating personnel ................................................................................................................................... 7
1.4
Responsibility of the owner ......................................................................................................................................................... 7
1.5
Intended use ................................................................................................................................................................................ 7
1.6
Changes and alterations .............................................................................................................................................................. 8
1.7
Behaviour in case of malfunctions and irregularities .................................................................................................................. 8
1.8
Switching off the appliance in an emergency .............................................................................................................................. 8
2. Construction and description
9
2.1
Design .......................................................................................................................................................................................... 9
2.2
Description ................................................................................................................................................................................. 10
2.3
Working range of constant climate chambers HPPeco ............................................................................................................. 11
2.4
Material ...................................................................................................................................................................................... 12
2.5
Electrical equipment .................................................................................................................................................................. 12
2.6
Connections and interfaces ....................................................................................................................................................... 12
2.7
Designation (nameplate) ............................................................................................................................................................ 13
2.8
Technical data ............................................................................................................................................................................ 14
2.9
Applied directives and standards .............................................................................................................................................. 15
2.10 Ambient conditions .................................................................................................................................................................... 15
2.11 Scope of delivery ........................................................................................................................................................................ 15
3. Delivery, Transport and Setting Up
16
3.1
Safety ......................................................................................................................................................................................... 16
3.2
Delivery ....................................................................................................................................................................................... 16
3.3
Transport .................................................................................................................................................................................... 17
3.4
Unpacking .................................................................................................................................................................................. 17
3.5
Disposing of packaging material ............................................................................................................................................... 17
3.6
Storage after delivery ................................................................................................................................................................. 17
3.7
Setting up ................................................................................................................................................................................... 18
3.8
Installation options .................................................................................................................................................................... 19
3.9
Level and secure the device against rolling away (Sizes 1400 and 2200) ................................................................................ 20
3.10 Tilt protection ............................................................................................................................................................................. 21
3.11 Adjusting doors ......................................................................................................................................................................... 21
4. Putting into operation
22
4.1
Connecting the appliance .......................................................................................................................................................... 22
4.2
Switching on............................................................................................................................................................................... 24
5. Operation and control
25
5.1
Operating personnel ................................................................................................................................................................... 25
5.2
Operating the appliance ............................................................................................................................................................. 27
5.3
Operating modes ........................................................................................................................................................................ 29
5.4
Monitoring function .................................................................................................................................................................... 34
5.5
Ending operation ........................................................................................................................................................................ 39
6. Malfunctions, warning and error messages
40
6.1
Warning message of the monitoring function ........................................................................................................................... 40
6.2
Power failure .............................................................................................................................................................................. 42
7. Menu mode
43
7.1
Overview ..................................................................................................................................................................................... 43
7.2
Basic operation in menu mode using the example of language selection ................................................................................ 44
7.3
Setup .......................................................................................................................................................................................... 45
7.4
Date and Time ............................................................................................................................................................................ 49
7.5
Calibration .................................................................................................................................................................................. 50
7.6
Program ..................................................................................................................................................................................... 55
7.7
Sound ......................................................................................................................................................................................... 56
7.8
Protocol ...................................................................................................................................................................................... 57
7.9
USER ID ...................................................................................................................................................................................... 58
8. Maintenance and Servicing
59
8.1
Cleaning ..................................................................................................................................................................................... 59
8.2
Regular maintenance ................................................................................................................................................................. 60
8.3
Repairs and service .................................................................................................................................................................... 60
9. Storage and disposal
61
9.1
Storage ....................................................................................................................................................................................... 61
9.2
Disposal ..................................................................................................................................................................................... 61
Terms and signs used
In this manual and on the appliance itself, certain common terms and signs are used to
warn you of possible dangers or to give you hints that are important in avoiding injury
or damage. Observe and follow these notes and regulations to avoid accidents and
damage. These terms and signs are explained below.
1.1.1
Terms used
WARNING
Warns of a dangerous situation that might lead to death or serious injuries.
CAUTION
Warns of a dangerous situation that might lead to moderate or minor injuries.
NOTICE
Warns of material damage
1.1.2
Signs used
Symbol
Meaning
Symbol
Meaning
Caution - hazard area
Gases�/�vapours
Caution - current
Prohibited - do not wear
Warning - fl ammable substances
Prohibited - do not enter
Caution - hot surfaces
Prohibited - do not tip
Danger of explosion
Note - disconnect the mains
plug
Note - observe manual
Note - wear gloves
Note - wear safety shoes
Note - observe number of
people
Product safety and dangers
The appliances described in this manual are technically sophisticated, manufactured
using high-quality materials and subject to many hours of testing in the factory. They
refl ect the state of the art and comply with recognised technical safety regulations.
However, there are still risks involved, even when the appliances are used as intended.
These are described below.
Live components may be exposed once the covers have been removed. Touching these
can lead to an electric shock. Disconnect the mains plug before removing any covers.
Work on the electrical system must only be performed by qualifi ed electricians.
Leaving the door open during operation can cause the appliance to overheat or pose
a fi re hazard. Do not leave the door open during operation.
When loading the appliance with an unsuitable load, poisonous or explosive vapours
or gases may be produced. This could cause the appliance to explode, and persons
could be severely injured or poisoned. The appliance may only be loaded with mate-
rials / test objects that do not emit any poisonous or explosive vapours when heated
up.
With appliances above a specifi c size, you could become accidentally locked inside,
which could put you at risk of death.
Do not climb into the appliance!
Depending on operation, the surfaces in the interior of the appliance and the chamber
load may still be very hot after the appliance is switched off. Touching these surfaces
can cause burns. Wear heat-resistant protective gloves or wait until the appliance
cools down after switching off before touching.
WARNING
WARNING
WARNING
WARNING
CAUTION
Requirements of the operating personnel
The appliance may only be operated and maintained by persons who are of legal age
and have been instructed accordingly. Personnel who are to be trained, instructed or
who are undergoing general training may only work with the appliance under the contin-
uous supervision of an experienced person. Repairs may only be performed by qualifi ed
electricians. The regulations in the separate service manual must be observed.
1.4
Responsibility of the owner
The owner of the appliance
■ is responsible for the fl awless condition of the appliance and for it being operated in
accordance with its intended use
■ is responsible for ensuring that persons who are to operate or service the appliance
are qualifi ed to do this, have been instructed accordingly and are familiar with the
operating instructions at hand
■ must know about the applicable guidelines, requirements and operational safety regu-
lations, and train staff accordingly
■ is responsible for ensuring that unauthorised persons have no access to the appli-
ance
■ is responsible for ensuring that the maintenance plan is adhered to and that mainte-
nance work is properly carried out
■ has to ensure that the appliance and its surroundings are kept clean and tidy, for
example through corresponding instructions and inspections
■ is responsible for ensuring that personal protective clothing is worn by operating
personnel, e.g. work clothes, safety shoes and protective gloves.
1.5
Intended use
Constant climate chambers HPPeco and cooled incubators IPPecoplus may be used
exclusively for temperature and climate testing of materials and substances in the con-
text of the procedures and specifi cations described in this manual. Any other use is
improper and may result in hazards and damage.
The appliance is not explosion-proof (does not comply with the German occupational
health and safety regulation VBG 24). The appliance may only be loaded with materials
and substances which cannot form any toxic or explosive vapours at the set tempera-
ture and which cannot explode, burst or ignite.
The appliance may not be used to dry, vaporise or brand materials for which the pur-
chasing or its components constitutes a risk of fi re and/or explosion, especially if the
solvents of these materials could form an explosive mixture when combined with air. If
there is any doubt as to the composition of materials, they must not be loaded into the
appliance. Potentially explosive gas-air mixtures must not form, neither in the chamber
nor in the direct vicinity of the appliance.
Changes and alterations
No unauthorised changes or alterations may be made to the appliance. No parts may be
added or inserted which have not been approved by the manufacturer.
Unauthorised changes or alterations result in the CE declaration of conformity losing its
validity, and the appliance may no longer be operated.
The manufacturer is not liable for any damage, danger or injuries that result from un-
authorised changes or alterations, or from non-compliance with the provisions in this
manual.
1.7
Behaviour in case of malfunctions and irregularities
The appliance may only be used in a fl awless condition. If you as the operator notice
irregularities, malfunctions or damage, immediately take the appliance out of service
and inform your superior.
1.8
Switching off the appliance in an emergency
Press the main switch on the ControlCOCKPIT and disconnect the power plug. This
disconnects the appliance from the power supply at all poles.
Construction and description
2.
Construction and description
2.1
Design
Description
Overview - HPP110ecoplus
1. ControlCOCKPIT with capacitive func-
tion keys and LCD displays
1
2
3
4
5
6
7
8
9
10
2. Main switch
3. Turn control with confi rmation key
4. Peltier element
5. Inner glass door
6. Slide-in units
7. Peltier dehumidifi er
8. Nameplate
9. Door-handle
10. USB interface
Description
Overview - HPP1400plus
1. ControlCOCKPIT with capacitive function
keys and LCD displays
1
2
3
4
5
6
7
8
9
2. Main switch
3. Turn control with confi rmation key
4. Slide-in units
5. Heated full-sight glass door
6. Lockable castors with extendable feet
7. Nameplate
8. Door-handle
9. USB interface
Construction and description
2.2
Description
Appliance size 110 to 1060
Description
HPPeco & IPPecoplus
The appliance can heat the interior up to +70 °C and cool it down to +5 °C. Low-noise,
long-life and energy-saving Peltier cooling and heating technology is used for this. In
heating operation, a part of the required energy is extracted from the surroundings
(heat pump principle).
HPPeco
Additionally, the humidity in the interior can be regulated between 10% rh and 90% rh
(rh = relative humidity). The humidity is increased by the evaporation of water from
a tank which is then fed into the interior and reduced by condensation on a Peltier
module.
Optionally, the appliance can be equipped with a light module, making it possible to
adjust the interior lighting in steps of 1 %.
Appliance size 1400 to 2200
Description
HPPeco & IPPecoplus
The appliance can heat the interior up to 60 ºC and cool it down to 15 ºC. Low-noise,
long-life and energy-saving Peltier cooling and heating technology is used for this. In
heating operation, a part of the required energy is extracted from the surroundings
(heat pump principle).
Constant climate chambers HPPeco
Additionally, the humidity in the interior can be regulated between 15% rh and 80% rh
(rh = relative humidity). The humidity is increased by the evaporation of water from
a tank which is then fed into the interior and reduced by condensation on a Peltier
module.
Construction and description
2.3
Working range of constant climate chambers HPPeco
The temperature-humidity diagram specifi es at what range of temperature and humidity
a permanent, condensation-free operation of the constant climate chamber HPPeco is
possible.
If the chamber is in operation at the upper limit or outside the working range for long
periods, puddles of water may form inside the chamber and water may force its way out
of the door seal.
Climate points in the threshold range of the climate diagram can only be reached
with the correct dehumidifi cation time interval. For optimal time interval setting, see
chapter 7.3.7
Range
Diagram
■
Range A:
In this range, temperature and humidity
can be combined as you please, without
resulting in any signifi cant condensation.
In extreme ambient conditions, the work-
ing range may be restricted.
0
10
20
30
40
50
60
70
80
90
100
0
5
10
15
20
25
30
35
40
45
50
55
60
65
70
75
Temperature °C
Humidity % rH
C
A
B
■
Range B:
If the specifi ed range is exceeded up-
wards, e.g. 80 %� rh at 60 °C, the hot
steam fed in will immediately condense,
due to the dew-point, at the coldest point
in the appliance.
■
Range C:
At low temperatures and low relative air
humidity, the effective range is heavily
dependent on the degree of humidity of
the chamber load.
Construction and description
2.4
Material
These appliances fulfi l the current requirements of the RoHS Directive. For more infor-
mation about this and about the Material Compliance of these Memmert appliances in
general, please visit our homepage at www.memmert.com.
Components
Material
Housing (lid + side panel)
Stainless steel 1.4016 - ASTM 430
Housing (rear panel)
Galvanised steel plate
Interior (incl. covers)
Stainless steel 1.4301 - ASTM 304
Accessories (perforated sheet, steel grid)
Stainless steel 1.4301 - ASTM 304
Door seal
Silicone
Inner glass door
Glass
The chamber load for the appliance must be carefully checked for chemical compatibil-
ity with the materials mentioned.
2.5
Electrical equipment
■ Operating voltage and current consumption: See nameplate
■ Protection class I, i.e. operating insulation with PE conductor in accordance with EN
61010
■ Protection type IP 20 acc. to EN 60 529
■ Interference suppression acc. to EN 55011 class B
■ Appliance fuse: Safety fuse 250 V/15 A, quick-blow
■ The temperature controller is protected with a miniature fuse 100 mA
(160 mA at 115 V)
2.6
Connections and interfaces
2.6.1
Electrical connection
This appliance is intended for operation on an electrical power system with a system
impedance Zmax at the point of transfer (service line) of a maximum of 0.292 Ohm.
The operator must ensure that the appliance is operated only on an electrical power
system that meets these requirements. If necessary, you can ask your local energy sup-
ply company what the system impedance is. Observe the country-specifi c regulations
when connecting (e.g. in Germany DIN VDE 0100 with residual current circuit breaker).
2.6.2
Communication interfaces
The interfaces are intended for appliances which meet the requirements of IEC 60950-1.
USB interface
The appliance is fi tted by default with a USB port in accordance with the USB specifi ca-
tion. With this you can:
■ transfer software stored on a USB storage medium to the appliance
■ export protocol logs from the appliance to a USB storage medium
■ transfer user ID data stored on a USB storage medium to the appliance
i
Construction and description
Ethernet interface
Via Ethernet interface, the appliance can be connected to a network, so that pro-
grammes created with the AtmoCONTROL software can be transferred to the appliance
and protocols read out.
For identifi cation purposes, each appliance connected must have its own unique IP
address. Setting the IP address is described in chapter 7.3.2 .
You will fi nd a description of how to transfer programs via Ethernet in the enclosed
AtmoCONTROL manual.
The appliance can be directly connected to a computer/laptop using an optional USB
to Ethernet converter.
2.7
Designation (nameplate)
The nameplate provides information about the appliance model, manufacturer and
technical data. It is attached to the front of the appliance, on the right behind the door.
Description
Overview
1. Type designation
GmbH+Co.KG D-91126 Schwabach FRG
Äußere Rittersbacher Str. 38
Made in Germany
�������
Schutzart DIN EN 60529 - IP 20
1
2
3
4
5
6
10
9
8
7
Typ: HPP110eco
100-240 V~ 50/60 Hz
DIN12880-2007-Kl.:3.1
F-Nr.: H418.3008
2.0-4.5 A 420 W
Nenntemp.: 70 °C
2. Operating voltage
3. Applicable standard
4. Protection type
5. CE conformity
6. Address of manufacturer
7. Disposal note
8. Temperature range
9. Connection / power ratings
10. Appliance number
Construction and description
2.8
Technical data
Appliance size
110
260
410
750
1060
1400
2200
Appliance width (D)
mm
745
824
1224
1435
2157
Appliance height (E)
mm
864
1183
1720
1913
Appliance depth (F)
mm
555
655
755
1005
905
Depth of door lock
mm
56
Chamber width (A)
mm
560
640
1040
1250
1972
Chamber height (B)
mm
480
800
1200
1450
Chamber depth (C)
mm
400
500
600
850
750
Chamber volume
mm
108
256
384
749
1060
1360
2140
Weight (including packaging)
kg
102
173
213
279
424
639
730
Power
HPPeco
W
420
700
1400
2000
IPPeco
W
320
600
1300
1900
Current
con-
sumption
HPPeco
115 V
A
4,5
7,5
14,5
-
230 V
A
2
3
6
10,5
IPPeco
115 V
A
3,5
6,5
13,5
-
230 V
A
1,5
2,5
5,5
10
max number steel grids/ perforated sheet
unit
5
9
14
28
42
max load per steel grid/ perforated sheet
kg
20
30
20
30
max load per appliance
kg
150
200
250
Temperature
Adjustment range
°C
+5 bis +70
+15 bis +60
Adjustment
preci-
sion
K
0,1
Humidity adjustment range
%
10 bis 90
15 bis 80
E
D
120
A
B
93
93
111
F
56
30
88
C
32
�������
MENU
ON
OFF
ON
ON
Construction and description
2.9
Applied directives and standards
Based on the standards and guidelines listed in the following, the products described in
this manual have received a CE label from the company Memmert:
■ Directive 2004/108/EC amended (Directive of the European Parliament and of the
Council on the approximation of the laws of the Member States relating to electromag-
netic compatibility)
■ Directive 2006/95/EC amended (Directive of the European Parliament and of the Coun-
cil on the harmonisation of the laws of Member States relating to electrical equipment
designed for use within certain voltage limits)
2.9.1
Declaration of conformity
You can download the EC declaration of conformity of the appliance online under
www.memmert.com/de/downloads/
2.10
Ambient conditions
Ambient criteria
Ambient temperature
+16 °C to +40 °C
Air humidity
max. 70 % rh non-condensing
Overvoltage category
II
Pollution degree
2
Altitude of installation
max. 2,000 m above sea level
■ The appliance may only be used in enclosed areas and under the ambient conditions
listed below
■ The appliance may not be used in areas where there is a risk of explosion. The ambi-
ent air must not contain any explosive dusts, gases, vapours or gas-air mixtures. The
appliance is not explosion-proof.
■ Heavy dust production or aggressive vapours in the vicinity of the appliance could lead
to sedimentation in the interior and, as a consequence, could result in short circuits
or damage to electrical parts. For this reason, suffi cient measures to prevent large
clouds of dust or aggressive vapours from developing should be taken.
2.11
Scope of delivery
Standard delivery
■ Power cable
■ Tilt protection
■ Steel grid (number depends on appliance size)
■ USB storage medium with software and AtmoCONTROL manual
■ Operating manual
■ Calibration certifi cate
In addition for constant climate chambers HPPeco
■ Water tank with connection hose
■ Tank holder (only for appliances of size 410 and up)
Optional accessories
■ USB- Ethernet converter. This makes it possible to connect the Ethernet connection
interface to the USB port of a computer/laptop.
■ Reinforced steel grid with a load capacity of 60 kg (size 110 and up)
Delivery, Transport and Setting Up
3.
Delivery, Transport and Setting Up
3.1
Safety
The appliance could fall over and seriously injure you. Never tilt the appliance. Only
transport in upright position and without load (except standard accessories such as
steel grids or shelves). Appliances with castors always have to be moved by two peo-
ple.
Because of the heavy weight of the appliance, you could injure yourself if you try to
lift it. At least four people are needed to carry appliance sizes 110 and 260. Applianc-
es larger than that may not be carried, but must be transported using a manual pallet
jack or forklift truck.
110 and 260
410 and up
You may get your hands or feet squashed when transporting and installing the appli-
ance. Wear protective gloves and safety boots.
Only grasp the appliance at the sides of the base:
�
3.2
Delivery
Appliance size 110 to 1060
■ The appliance is packed in cardboard and is delivered on a wooden palette.
Appliance size 1400 to 2200
■ The appliance is packaged in a wooden crate and delivered on a wooden pallet.
WARNING
CAUTION
CAUTION
Delivery, Transport and Setting Up
3.3
Transport
The appliance can be transported in three ways:
■ With a forklift truck or a manual pallet jack; move the forks of the truck entirely under
the pallet
■ On its own castors, in case of the corresponding confi guration, for which the catch on
the (front) castors must be released
3.4
Unpacking
■ To avoid damage, do not unpack the appliance until you reach the installation site.
■ Remove the cardboard packaging by pulling it upwards or carefully cutting along an
edge or unscrew and remove wooden crate
3.4.1
Checking for completeness and transport damage
■ Check the delivery note to ensure that the scope of delivery is complete
■ Check the appliance for damage
If you notice deviations from the delivery note, damage or irregularities, do not put the
appliance into operation but inform the haulage company and the manufacturer.
3.4.2
Remove the transportation lock
■ Remove the transportation lock. It is located between the door hinge, door and frame
and has to be removed after opening the door.
3.5
Disposing of packaging material
■ Dispose of the packaging material (cardboard, wood, foil) in accordance with the
applicable disposal regulations for the respective material in your country
3.6
Storage after delivery
If the appliance is fi rst to be stored after delivery:
■ Observe storage conditions, see chapter 9.1
Delivery, Transport and Setting Up
3.7
Setting up
Due to its centre of gravity, the appliance can fall over to the front and injure you or
other people. Always attach the appliance to a wall with the tilt protection. In case
there is not enough space, do not put the appliance into operation and do not open the
door. Contact the Memmert service.
3.7.1
Preconditions
The installation site must be fl at and horizontal and able to reliably bear the weight of
the appliance. Do not place the appliance on a fl ammable surface.
Depending on the model (see nameplate), a 230�V or 115�V power connection must be
available at the installation site.
The distance between the wall and the rear of the appliance must be at least 15 cm.
The clearance from the ceiling must not be less than 20 cm and the side clearance from
walls or nearby appliances must not be less than 5 cm. Suffi cient air circulation in the
vicinity of the appliance must be guaranteed at all times.
For appliances with castors, these need to be positioned in forward direction at all
times.
10 cm
5 cm
15 cm
> 20 cm
> 20 cm
5 cm
WARNING
Delivery, Transport and Setting Up
3.8
Installation options
Setting up
Comments
110
260
410
750
1060
1400
2200
Bottom
Table
Check the load capacity
fi rst
Stacked
two
appliances
maxi-
mum; mounting material
(feet) provided
Sub frame
with/without castors
Castor
frame
Height adjustable
feet
Delivery, Transport and Setting Up
3.9
Level and secure the device against rolling away (Sizes 1400 and 2200)
The height of the appliance can be adjusted using the heavy-duty castors attached to
the bottom of the appliance.
It can also be secured against rolling away or being shifted. To do this, the feet must
be extended.
1. To unlock the adjustment mechanism, pull out the ratchet lever on the ring
2. To retract and extend the stand, actuate the ratchet lever
The direction of movement (up/down) can be adjusted using the rocker above the
ratchet lever:
■
Press in the rocker on the right side to extend the stand with the ratchet lever
■
Press in the rocker on the left side to retract the stand with the ratchet lever
In the end position, push in the ratchet lever again to fi x the stand.
Use a spirit level to ensure that all four feet are adjusted to the same height.
To move the cabinet, all four feet must be retracted.
i
Delivery, Transport and Setting Up
3.10
Tilt protection
Attach the appliance to a wall with the tilt protection. The tilt protection is included in
the scope of delivery.
1. Tighten the tilt protection onto the back of the appliance as illustrated.
i
Depending on the ambient conditions, the tilt protection can be fastened to either
of the two holes in the appliance.
15 cm
2. Bend the tilt protection upwards by 90° in the desired distance to the wall (ob-
serve the minimum distance to the wall).
3. Drill a hole, insert a dowel and screw the tilt protection to a suitable wall.
3.11
Adjusting doors
You can adjust the doors if necessary, for example if they are warped due to uneven
fl ooring. There are two adjusting screws each at the top and the bottom of each door for
this purpose. First, adjust the setting at the top of the door and, if this is not suffi cient,
adjust the screws at the bottom of the door.
Door adjustment is also available as a service video:
https://www.memmert.com/de/downloads/media/service-videos/
1.
Open the door.
2.
Loosen the screws.
3.
Adjust the position of the door.
4.
Tighten the screws again.
5.
Check the position of the door.
6.
Readjust if required.
i
Putting into operation
4.
Putting into operation
Condensation in the electrical components may cause short circuits. After transport-
ing or storing the device under humid conditions, remove it from its packaging and let
it ventilate for at least 24 hours in normal environmental conditions. Do not connect
the device to the mains power during this time.
When putting the appliance into operation for the fi rst time, do not leave it unattended
until it has reached a steady state.
4.1
Connecting the appliance
Observe the country-specifi c regulations when making connections (e.g. DIN VDE
0100 with earth leakage circuit breaker, in Germany). Observe the connection and pow-
er ratings (see name plate and the "Technical Data"chapters 2.7 and 2.8). Make sure to
establish a safe PE conductor connection.
Plug the provided power cable into the rear of the appliance and connect it to the
power supply.
Place the power cable so that
■ it is easily accessible at all times and can be pulled off quickly, for example in case of
interference or an emergency
■ it does not represent a trip hazard
■ it cannot come into contact with any hot parts
WARNING
CAUTION
Putting into operation
4.1.1
Filling and connecting the water tank (for constant climate chambers HPPeco only)
Water specifi cations
Only water with the following specifi cations may be used in Memmert appliances:
■ demineralised/deionised
■ Conductivity of 5 – 10 μS/cm
■ pH value between 5 and 7
■ chlorine-free
The use of ultrapure water or DI water with an electrical conductance level below 5 μS/
cm can damage silicone tubing and cause pitting on the stainless steel components
installed. Unsuitable water also creates favourable conditions for limescale in the
steam generators and steam pipes.
Connection
Fill the supplied water tank with water and use the enclosed tube to connect it to the
“H2O” connection on the rear of the chamber.
For appliances of size 410 or greater, the tank can be attached to the appliance with
the included tank holder. To do this, fasten the tank holder to the rear panel using four
screws. Then connect the water tank to the steam generator.
i
Putting into operation
4.1.2
Hook drip pan
Holders for the drip pan are next to each Peltier dehumidifi er. The drip pan is hooked
into the holders from above
�
4.2
Switching on
Press the main switch on the front of the appliance.
The start-up process is shown by three animated white dots
.
If the dots have another colour, an error has occurred (error messages see chapter 6).
The appliance displays are in English by default when the appliance is switched on for
the fi rst time. How you can change the languages is described in chapter 7.2.
However, to get a basic overview of operating the appliance, you should read the
following chapter fi rst.
Operation and control
5.
Operation and control
5.1
Operating personnel
The appliance may only be operated by persons who are of legal age and have been in-
structed accordingly. Personnel who are to be trained, instructed or who are undergoing
general training may only work with the appliance under the continuous supervision of
an experienced person.
Opening the door
With appliances above a specifi c size, you could become accidentally locked inside,
which could put you at risk of death. Do not climb into the appliance!
To open the door, pull the door handle sideways (to the left or to the right, depending
on the door version, and open the door wide.
To close the appliance, push the door closed and push the door handle sideways.
WARNING
Operation and control
Loading the appliance
When loading the appliance with an unsuitable load, poisonous or explosive vapours
or gases may be produced. This could cause the appliance to explode, and persons
could be severely injured or poisoned. The appliance may only be loaded with ma-
terials which do not form any toxic or explosive vapours when heated up, and which
cannot ignite. If there is any doubt as to the composition of materials, they must not
be loaded into the appliance.
Check the chamber load for chemical compatibility with the materials of the appliance.
Insert the sliding steel grids or sliding shelves. The maximum number or grids / shelves
and the load capacity are specifi ed in the technical data overview in chapter 7.3.5 . To
achieve optimal temperature distribution, the type of slide-in unit used – steel grid or
shelf – must be set in the menu under SETUP.
The chamber must not be loaded too tightly, so that proper air circulation in the interior
is guaranteed.
Do not place any load on the fl oor, on the side walls or under the ceiling of the interior
In case of improper loading (not enough space between the items), the set temperature
may be exceeded or it may take longer until it is reached.
(see also the “correct loading” sticker on the appliance).
WARNING
Operation and control
5.2
Operating the appliance
5.2.1
ControlCOCKPIT
In manual mode, the desired parameters are entered in the ControlCOCKPIT on the front
of the appliance. You can also make basic settings here (menu mode). Additionally,
warning messages are displayed, e.g. if the temperature is exceeded. In program mode,
the parameters defi ned, the program description, the program segment currently active
and program duration remaining are displayed.
HPPeco and IPPecoplus
ON
ONN
ON
O
Holz trocknen
aufheizen
09:12h
Fr 20.10.2010
20:31
min
000°C
ALARM of °C
max
000°C
auto off
+ - 99K
0
12
%rh
°C
GRAPH
off
ON
LIGHT
ON
ON
LIGHT
ON
LIGHT
ON
ONN
ON
O
HUMIDITY
30.0%rh
Set 30.0%rh
TEMP
°C
22.4
TEMP
°C
Set 37.0
LIGHT
%
100
ALARM
max
38.5°C
auto
+/- 0.0K
min
35.5°C
Manual Mode
12.Sept.2012
13:44
3
11
12
13
14
1
2
17
18
9
10
15
16
4
5
6
7
8
TIMER
30m
04h
End
23.Nov 13:30
ON
ONN
ON
O
Holz trocknen
aufheizen
09:12h
Fr 20.10.2010
20:31
min
000°C
ALARM of °C
max
000°C
auto off
+ - 99K
0
12
%rh
°C
GRAPH
off
180
TEMP
ON
LIGHT
ON
ON
LIGHT
ON
LIGHT
ON
ONN
ON
O
LÜFTER
0%
°C
22.4
TEMP
°C
Set 37.0
TIMER
30m
04h
Ende 23.Nov 13:30
LIGHT
%
100
ALARM
max
38.5°C
+/- 2.0K
min
36.5°C
auto
Manual Mode
12.Sept.2012
13:44
3
11
12
13
14
1
2
17
18
9
10
15
16
4
5
6
1. Activation key for temperature set-
point adjustment
2. Setpoint and actual temperature
display
3. Menu key
4. Activation key digital backwards coun-
ter with target time setting
5. Main switch
6. Digital backwards counter with target
time setting
7.
Humidity control display
8.
Humidity control activation key
9.
Turn control for setpoint adjust-
ment
10. Confi rmation key
11. Interior lighting activation key (only
for models with light module)
12. Interior lighting display (only for
models with light module)
13. Appliance state and programme
display
14. Activation key for the appliance
state
15. Activation key for temperature
monitoring
16. Monitoring display
17. Graphical representation
18. Activation key for graphical rep-
resentation
Operation and control
5.2.2
Basic operation
In general, all settings are made according to the following pattern:
.5°C
100
TEMP
TIMER
TIMER
--
--
m
h
2
TEMP
Set
Activate the desired parameter (e.g. temperature). To do so, press the corresponding
activation key on the left or right or the respective display. The activated display is
lined in colour, the other displays are dimmed. The set value is highlighted in colour.
°
22.4
TEMP
Set 37.0
By turning the turn control to the left or right, adjust the set value (e.g. to 37.0 ºC).
°C
22.4
TEMP
°C
Set 37.0
Save the set value by pressing the confi rmation key.
The display returns to normal and the appliance begins adjusting to the defi ned set
value.
■ Additional parameters can be set accordingly
■ If no new values are entered or confi rmed for approx. 30 seconds, the appliance auto-
matically restores the former values
T
T
-
T
If you want to cancel the setting procedure, press the activation key on the left or right
of the display that you want to exit. The appliance restores the former values. Only the
settings that you have confi rmed by pressing the confi rmation key before cancelling
the setting procedure are accepted.
Operation and control
5.3
Operating modes
The appliance can be operated in different modes:
Manual mode
The appliance runs in permanent operation at the values set on the ControlCOCKPIT.
Operation in this mode is described in chapter 5.3.1 .
Timer operation
Operation with digital backwards counter with target time setting, adjustable from 1
minute to 99 days (timer): The appliance will run at the values set until the set time
has elapsed. Operation in this mode is described in chapter 5.3.2 .
Program mode
The appliance automatically runs program sequences which have been defi ned us-
ing AtmoCONTROL software at a computer / laptop and then transferred to the
appliance from a USB stick or via Ethernet. Operation in this mode is described in
chapter 5.3.3 .
Remote control mode
via remote control (chapter 7.3.8)
The status display shows which operating mode or operating state the appliance is
currently in. The current operating state is highlighted in colour and indicated by the
text display:
Manual Mode
12.Sept.2012
13:44
Appliance is in program mode
program is stopped
Appliance is in manual mode
The example shows the appliance in manual mode, identifi ed by the coloured hand
symbol.
Timer active
12.Sept.2012
13:44
When the appliance is in timer mode, Timer active is displayed:
23.2°C
TEMP
Set 38.0°C
If the appliance is in remote control mode,
the
symbol appears in the temperature display:
Operation and control
5.3.1
Manual mode
In this operating mode, the appliance runs in permanent operation at the values set on
the ControlCOCKPIT.
Adjustment options
As described in chapter 5.2.2 , you can set the following parameters after pressing the
corresponding activation key (in any sequence):
Temperature
°C
22.4
TEMP
°C
Set 37.0
Adjustment range depends on appliance (see name plate chapter 2.7 and technical
data chapter 2.8)
Heating operation is indicated by the
symbol.
Cooling is indicated by the
symbol.
You can select °C or °F as the temperature unit displayed.
i
The minimum temperature that can be reached depends on the surrounding condi-
tions. The devices can cool down to 20 °C below room temperature. For this purpose,
the Peltier module needs suffi cient ventilation
Humidity (for constant climate chambers HPPeco only)
HUMIDITY
70 %rh
Set34% rh
Adjustment range: 10 to 90 % rh
Humidifi cation is indicated by the
symbol.
Dehumidifi cation is indicated by the
symbol.
i
A high level of air humidity in the interior can only be achieved without condensa-
tion if the interior is thoroughly heated. For this reason, how fast the humidity is
dynamically adjusted to approach the setpoint depends on the interior temperature.
Interior lighting (only for models with light module)
LIGHT
0 %
Temp too high
LIGHT
40 %
Adjustment range: 0 to 100 % in steps of 1%
i
The interior light is only active at temperatures of up to 40°. If this temperature is
exceeded, the interior light switches off automatically. The light display will then
show “Temp too high”.
Operation and control
5.3.2
Operation with digital backwards counter with target time setting, adjustable from 1 minute to 99 days
In timer operation, you can adjust the time the appliance runs at the set values. The
appliance has to be in manual operating mode for this.
Up to a duration of 23 hours 59 minutes, the time is displayed in hh:mm (hours:minutes)
format. For 24 hours and more, the format dd:hh (days:hours) is used. The maximum
duration adjustable is 99 days 00 hours.
TIMER
-
Ende 9:00 23.11.
--
h-
m
1. Press the activation key to the left of the timer display. The timer display is acti-
vated.
TIMER
End
0
13:30 23.11.
04
m
h 3
2. Turn the turn control until the desired duration is displayed – in this example 4
hours 30 minutes. The approximate end time is shown beneath, in a smaller font.
3. Press the confi rmation key to confi rm.
TIMER
30m
04h
End
13:30 23.11.
Timer active
12.Sept.2012
13:44
The display now shows the remaining time in a large font and the approximate end
time in a smaller font beneath. The status display shows Timer active.
4. Now, as described under 5.2.2 , set the individual values which you want the ap-
pliance to operate at. The set values can be changed at any time while the timer
elapses. The changes are effective immediately.
In Setup, you can choose if the timer should run setpoint-dependent or not. This deter-
mines whether the timer should not start until a tolerance band around the set temper-
ature is reached or if it should start right after activation. The
symbol on the timer
display indicates that the timer is set to setpoint-dependent.
TIMER
00m
00h
End
13:30 23.11.
Once the timer has fi nished, the display shows 00h:00m. All functions (heating etc.)
are switched off. In addition, an acoustic alarm sounds, which can be turned off by
pressing the confi rmation key.
TIMER
--m
--h
End
9:00 23.11.
To- the timer, open the timer display by pressing the activation key again and then
turning the turn control to reduce the timer setting until --:-- is displayed. Press the
confi rmation key to confi rm.
i
i
Operation and control
5.3.3
Program mode
In this operating mode, programmes saved in the appliance can be started with
different combinations of individual parameters (temperature, humidity, interior
lighting) at staggered intervals, which the appliance then automatically process-
es in sequence. These programs are not created directly at the appliance but ex-
ternally at a computer / laptop and using AtmoCONTROL software. Transfer to
the appliance is possible using the provided USB storage medium or via Ethernet.
■ A description of how to create and save programs can be found in the separate
AtmoCONTROL software manual
Starting a programme
0 %
manueller Betrieb
%
Fr 20.10.2010 20:31
Manual mode
17:44
13.Sept.2012
Activate
1. Press the activation key on the right of the status display. The current operating
mode is highlighted automatically, in this example Manual Mode (
).
Test 012
ready
10:44
12.Sept.2012
2. Turn the turn control until the start symbol is highlighted. The current program
is displayed, in this example Test 012.
i
Only the program currently- in menu mode and shown in the display can be used.
If you want to process another program, you need to activate it in menu mode fi rst
(description in chapter 7).
Test 012
Ramp 1
10:44
12.Sept.2012
3. To start the program, press the confi rmation key. The program is executed. The
display shows:
■
the program description (in this example Test 012)
■
the programme segment description, in this example Ramp 1
■
the current run (in case of loops)
i
You cannot change any parameters (e.g. the temperature) at the appliance while
a program is running. However, the displays ALARM and GRAPH can still be used.
Operation and control
Cancel program
You can cancel an active program at any time.
T
00 %
manueller Betrieb
T
%
Fr 20.10.2010 20:31
Test 012
Ramp 3
10:44
12.Sept.2012
1. Press the activation key to the right of the status display. The status display is
automatically highlighted.
Cancel program
Test 012
10:48
12.Sept.2012
2. Turn the turn control until the ■ stop symbol is highlighted.
End
Test 012
10:49
12.Sept.2012
3. Press the confi rmation key to confi rm. The program is cancelled.
i
A cancelled program cannot be resumed at the point it was cancelled. It must be
restarted from the beginning.
End of program
End
Test 012
10:49
12.Sept.2012
End is shown on the display to indicate that the program has fi nished normally.
You can now...
Manual Mode
12.Sept.2012
13:44
■
restart the program as described
■
select another program to run in menu mode and run it as described
■
return to manual mode. To do so, reactivate it by pressing the activation key
next to the status display, then turn the turn control until the hand symbol
is
highlighted in colour and press the confi rmation key
Operation and control
5.4
Monitoring function
5.4.1
Temperature monitoring
The appliance is equipped with a multiple overtemperature protection in accordance
with DIN 12 880. This serves to avoid damage to the chamber load and/or appliance in
case of a malfunction:
■ electronic temperature monitoring (TWW)
■ automatic temperature monitor (ASF)
The
ALARM
max
38.5°C
1.0K
min
35.5°C
auto
+-
monitoring temperature of the electronic temperature monitoring is measured via a
separate PT100 temperature sensor in the chamber. Temperature monitoring settings
are made via the ALARM display. The settings made apply to all operating modes.
If temperature monitoring has been triggered, this is indicated on the temperature dis-
play: the actual temperature is highlighted in red and a
warning symbol is shown
The type of temperature monitoring triggered (TWW in this example) is shown beneath
the temperature.
If the alarm sound has been activated in the menu mode (sounds, chapter 7.7, indicated
by the speaker symbol
in the alarm display), the alarm is additionally signalled by an
intermittent acoustic signal, which can be turned off by pressing the confi rmation key.
Information on what to do in this case is provided in chapter 6.
Before reading how to adjust temperature monitoring (from chapter 5.4.4), please read
the description of the individual monitoring functions here.
5.4.2
Electronic temperature monitoring (TWW)
The manually set monitoring temperature min and max of the overtemperature control
is monitored by an adjustable over/undertemperature controller (TWW) of protection
class 3.3 according to DIN 12 880. If the manually set monitoring temperature max is
exceeded, the TWW takes over temperature control and begins to regulate the monitor-
ing temperature
t
°C
Setting MAX
Set temperature
Controller error
Emergency operation
min
000°C
AL
ON
LIGHT
ON
LIGHT
FEUCHTE
30.0%rh
Set 30.0%rh
22 4
TEMP
°C
°C
22 4
TEMP
°C
LI
AL
auto
min
35.5
TIMER
30m
04h
Ende 23.Nov 13:30
°C
TEMP
TWW Set 38.5 °C38.9
TIMER
FE
22.4
37.0
TEMP
Set
°C
°C
°C
22.4
TEMP
°C
Set 37.0
TIMER
°C
TEMP
TWW Set 38.5 °C38.9
Operation and control
5.4.3
Automatic temperature monitor (ASF)
ASF is a monitoring device that automatically follows the set temperature setpoint with-
in an adjustable tolerance band.
The ASF – if switched on – is automatically activated as soon as the actual temperature
value reaches 50 % of the set tolerance band of the setpoint (in the example: 50 °C ±
1 K) for the fi rst time (section A).
If the set tolerance band around the setpoint (50 °C ± 2 K) is left – e.g. if the door is
opened during operation (section B of illustration) – the alarm will be set off. The ASF
alarm is automatically terminated as soon as 50 % of the set tolerance band of the set-
point (in the example: 50 °C ± 1 K) are reached again (section C).
If the temperature setpoint is altered, the ASF is automatically disabled temporarily (in
this example: The setpoint is changed from 50 °C to 25 °C, section D), until it reaches
the tolerance range of the new temperature setpoint (section E).
52°C
48°C
ASF active
52°C
48°C
ASF alarm
ASF active
ASF active
27°C
23°C
°C
50°C
25°C
t
A
B
C
D
E
AUTO
AUTO
AUTO
5.4.4
Adjusting temperature monitoring
min
000°C
ALARM
max
000°C
auto off
- 99K
+
ALARM
+- 0.0K
min
°C
auto
max
120.0°C
60 0
.
For IPPecoplus cooled incubators:
■
Press the activation key to the left of the ALARM display. The min set-
ting
(undertemperature
protection)
is
automatically
activated.
Continue with item 1
min
000°C
ALARM
max
000°C
auto off
- 99K
+
ALARM
max
40.0°C
auto
0.0K
min
15.0°C
+-
For constant climate chambers HPPeco:
■
Press the activation key to the left of the ALARM display. The temperature moni-
toring setting is automatically activated ( ).
ALARM
max
auto
40
0.0K
min
15.0°C
.0°C
+-
Save the selection by pressing the confi rmation key. The min setting (undertemper-
ature protection) is automatically activated.
Operation and control
ALARM
0.0K
min
°C
auto
max
°C
35 5.
40 0.
+-
1. By turning the turn control, adjust the desired lower alarm limit value, in the exam-
ple on the right 35.5 °C.
i
The lower alarm limit value cannot be set higher than the top one. If no undertemper-
ature protection limit is required, set the lowest temperature.
ALARM
0.0K
min
°C
auto
max
°C
35 5.
40 0.
+-
2. Press the confi rmation key to confi rm. The max display (overtemperature protec-
tion) is activated.
ALARM
0.0K
min
°C
auto
max
°C
35.5
38 5.
+-
3. By turning the turn control, adjust the desired upper alarm limit value, in the
example on the left 38.5 °C.
i
The monitoring temperature must be set suffi ciently high above the maximum set
temperature. We recommend 1 to 3�K.
ALARM
max
38.5°C
0.0K
min
35.5°C
auto
+-
4. Accept the upper alarm limit value by pressing the confi rmation key. The setting of
the automatic temperature monitor (ASF) is automatically activated (auto).
ALARM
max
38.5°C
0.0K
min
35.5°C
auto
+-
5. With the turn control, select ON () or OFF ().
ALARM
K
min
°C
auto
max
38.5°C
35 5.
0.0
+-
6. Press the confi rmation key to confi rm. The ASF tolerance band setting is activated.
ALARM
K
min
°C
auto
max
38.5°C
35 5.
1.0
+-
7. With the turn control, adjust the desired tolerance band, e.g. 2.0 K.
i
We recommend a tolerance band of 1 to 3�K.
ALARM
max
38.5°C
1.0K
min
35.5°C
auto
+-
8. Press the confi rmation key to confi rm. Temperature monitoring is now active.
Operation and control
5.4.5
Humidity monitoring (for constant climate chambers HPPeco only)
If humidity monitoring was triggered, this is indicated by the humidity display: the actual
humidity is highlighted in red and a warning symbol
is shown ( ). If the alarm sound
has been activated in the Menu mode (sounds, chapter 7.7, indicated by the speaker
symbol
), the alarm is additionally signalled by an intermittent acoustic signal. Infor-
mation on what to do in this case is provided in chapter 6.
Adjusting humidity monitoring
min
000°C
ALARM
max
000°C
auto off
- 99K
+
ALARM
max
40.0°C
auto
0.0K
min
15.0°C
+-
1. Press the activation key to the left of the ALARM display. The temperature monitor-
ing setting is automatically activated.
ALARM
min
40.0%rh
max
60.0%rh
2. Turn the turn control until the humidity monitoring entry
is highlighted.
ALARM
min
40.0%rh
max
60.0%rh
3. Accept the selection by pressing the confi rmation key. The lower humidity alarm
limit is automatically highlighted.
ALARM
min
50.0%rh
max
60.0%rh
4. By turning the turn control, adjust the desired lower alarm limit, in the example on
the left 50 % rh.
ALARM
min
50.0%rh
max
60.0%rh
5. Accept the selection by pressing the confi rmation key. The upper humidity alarm
limit is automatically highlighted.
ALARM
min
50.0%rh
max
70.0%rh
6. By turning the turn control, adjust the desired lower alarm limit, in the example on
the left 70 % rh.
ALARM
min
50.0%rh
max
70.0%rh
7. Accept the selection by pressing the confi rmation key and exit the Alarm display by
pressing the activation key on the side. Humidity monitoring is now active.
O2
10.0 %
10.0 %
Set
O2
Set 15.0%
%
12.0
FEUCHTE
2.4°C
Set 37.0°C
O2
10.0 %
10.0 %
Set
O2
Set 15.0%
12.0%
Operation and control
Graph
The GRAPH display provides an overview of the chronological sequence of the set val-
ues and the actual values as a curve.
To close the graphical representation, press the activation key you used to activate it
again.
5.4.6
Temperature profi le
0
4
8
12
16
20
24
°C
40
20
60
80
100
Fr 20.10.2010 20:34
12.09.2012
14.00
16.00
18.00
38
39
40
1. Press the activation key to the right of the GRAPH display. The display is enlarged
and the temperature profi le shown.
0
4
8
12
16
°C
40
20
60
80
100
Fr 2
12.0
14.00
16.00
38
39
40
2. To change the time frame to be displayed: Press the activation key near the arrow
symbols . The time frame to be displayed can now be changed by turning the
turn control.
20:34
3. To zoom the graph in or out: Press the activation key next to the magnifying glass
symbol. Select whether you want to zoom in or out (+/–) with the turn control and
confi rm your selection by pressing the confi rmation key.
i
Operation and control
5.4.7
Humidity Humidity profi le (for constant climate chambers HPPeco only)
0
4
8
12
16
20
24
°C
40
20
60
80
100
Fr 20.10.2010 20
12.09.2012
14.00
16.00
18.00
38
39
40
1. Activate graphic representation as described above and then press the activation
key next to the parameter selection.
°C
40
20
60
80
100
38
39
40
2. Set the humidity with the turn control.
°C
40
20
60
80
100
40
60
80
%rh
3. Press the confi rmation key to confi rm. The humidity profi le is displayed. You can
change the display range as described above, as well as extend or reduce it.
5.5
Ending operation
■ Switch off active appliance functions (turn back the set values).
■ Remove the chamber load
■ For constant climate chambers HPPeco: Check the water tank and fi ll up if necessary
(chapter 4.1.1)
■ Switch off the appliance with the main switch
Malfunctions, warning and error messages
6.
Malfunctions, warning and error messages
Live components may be exposed once the covers have been removed. Touching
these can lead to an electric shock. Malfunctions requiring work inside the appliance
may only be rectifi ed by electricians. Observe the separate service manual for this.
Do not try to rectify appliance errors yourself but contact the MEMMERT customer ser-
vice department or an authorised service point.
In case of enquiries, please always specify the model and appliance number on the
nameplate (see chapter 2.7).
6.1
Warning message of the monitoring function
If the alarm sound has been activated in the menu mode ( sound, chapter 7.7, indicated
by the speaker symbol
), the alarm is additionally signalled by an intermittent acous-
tic signal. If the confi rmation key is pressed, the acoustic alarm can be temporarily
switched off until the next alarm event occurs.
6.1.1
Temperature monitoring
Description
Cause
Action
Temperature alarm and "ASF" are
displayed
TEMP
ASF Set 38.5 °C
40.4°C
Automatic temperature monitor (ASF)
was triggered.
■ Check if the door is closed. Closing the
door
■ Extend the ASF tolerance band
■ If the alarm continues: Contact custom-
er service
Temperature alarm and "TWW" are
displayed
TEMP
TWW Set 38.5 °C
42.4°C
The adjustable temperature controller
(TWW) has assumed heating control.
■ Increase the difference between the
monitoring and setpoint temperature –
by either increasing the max value of the
temperature monitoring or decreasing
the setpoint temperature.
■ If the alarm continues:
Contact customer service
WARNING
Malfunctions, warning and error messages
6.1.2
Humidity monitoring (for constant climate chambers HPPeco only)
Error description
Cause of error
Troubleshooting
Error display symbol
HUMIDITY
Set 55.0%rh
55.4%rh
Water tank empty
■ Fill the water tank with water and press
the confi rmation key
Alarm display (MaxAl)
HUMIDITY
MaxAl Set 70.0%rh
75.4%rh
Upper humidity limit exceeded
■ Open the door for 30 sec. and wait to
see if the appliance reliably adjusts to
the setpoint
■ If the error occurs again, contact cus-
tomer service.
Alarm display (MinAl)
HUMIDITY
MinAl Set 60 .0%rh
55.4%rh
Humidity below lower limit
■ Check if the door is closed
■ Check the water supply and the fi lling
level of the water tank. If required, refi ll
water
■ If the error occurs again, contact cus-
tomer service.
6.1.3
Malfunctions, operating problems and appliance errors
Error description
Cause of error
Troubleshooting
Displays are dark
External power supply
was interrupted
■ Check the power supply
Miniature fuse, appliance fuse or power
module faulty
■ Contact customer service
Displays cannot be activated
Appliance locked by USER ID
■ Unlock with USER ID
The appliance is in program, timer or
remote control mode (mode "Write" or
"Write + Alarm")
■ Wait until the end of the program or tim-
er mode or switch off the remote control
Displays suddenly look different
Appliance is in "wrong" mode
■ Change to operating or menu mode by
pressing the MENU key
Error message T:E-3 in the temperature
display
TEMP
T:E-3 Set 37.0 °C
37.4°C
Temperature operating sensor is defec-
tive. The monitoring sensor takes over
the measurement function.
■ The appliance can temporarily be kept
in service
■ Contact customer service as soon as
possible
Malfunctions, warning and error messages
Error description
Cause of error
Troubleshooting
Error message AI E-3 in the temperature
display
TEMP
AI E-3 Set 37.0 °C
37.4°C
Temperature monitoring sensor is de-
fective. The operating sensor takes over
the measurement function.
■ The appliance can temporarily be kept
in service
■ Contact customer service as soon as
possible
Error message E-3 in the temperature
display
TEMP
Set 45.0 °C
E-3 °C
Operating and monitoring sensor defective
■ Switch off appliance
■ Remove the chamber load
■ Contact customer service
Error message E-6 in the humidity
display
HUMIDITY
Set 50.0%rh
%rh
E-6
Humidity sensor defective
■ No humidity control possible
■ Contact customer service
When switching on the appliance, the
start animation is displayed in another
colour than white
Cyan
:
Not enough storage space on the SD
card
■ Contact customer service
Red
:
The system fi les could not be loaded
Orange
:
The fonts and images could not be load-
ed
6.2
Power failure
In case of a power failure, the appliance operates as follows:
In manual mode
After power supply has been restored, operation is continued with the parameters set.
The time and the duration of the power failure is documented in the log memory.
In timer or program mode
In case of an interruption of the power supply of less than 60 minutes, the current pro-
gramme is continued from the point at which it was interrupted. For longer interruptions
of the power supply, all appliance functions (heating, fan etc.) are switched off.
In remote control mode
The previous values are restored. If a program has been initiated via remote control, it
is continued.
In menu mode, you can make basic settings, load programs and export protocols, as
well as adjust appliance parameters.
Before changing menu settings, read the description of the respective functions on the
following pages to avoid possible damage to the appliance and/or chamber load.
To enter menu mode, press the MENU key.
To exit the menu mode at any time, press the MENU key again. The appliance then
returns to operating mode. Only changes accepted by pressing the confi rmation key
are saved.
7.1
Overview
SETUP
SIGNALTÖNE
PROG
ZEIT UND DATUM
CALIB
USER ID
PROTOCOL
LANGUAGE
TIME
SOUND
ON
ONN
ON
O
1
2
3
4
5
12
14
15
13
6
7
18
19
8
9
10 11
16
17
1. Language selection activation key
2. Language selection display
3. Date and time display
4. Date and time setting activation key
5. Exit menu mode and return to operat-
ing mode
6. Setup activation key (basic appliance
settings)
7. Setup display (appliance settings)
8. Adjustment display
9. Adjustment activation key
10. Turn control for adjustment
11. Confi rmation key (accepts setting
made with the turn control)
12. program selection activation key
13. program selection display
14. Protocol display
15. Protocol activation key
16. Acoustic signal adjustment activation
key
17. Acoustic signal adjustment display
18. USER ID display
19. USER ID display activation key
i
i
Basic operation in menu mode using the example of language selection
In general, all settings in menu mode are done just like in operating mode: Activate the
respective display, use the turn control for setting and press the confi rmation key to
accept the change. A more detailed description is provided in the following, using the
example of language selection.
All other settings can be made accordingly. The settings possible are described in the
following sections.
If no new values are entered or confi rmed for approx. 30 seconds, the appliance auto-
matically restores the former values
ENGLISH
DEUTSCH
ESPANOL
POLSKI
CESTINA
FRANCAIS
1. Activate the desired parameter (in this example the language). To do so, press
the corresponding activation key on the left or right or the respective display. The
activated display is enlarged.
SETUP
SIGN
LANGUAGE
i
If you want to exit or cancel the settings, again press the activation key which
you have used to activate the display. The appliance returns to the menu over-
view. Only the settings that you have confi rmed by pressing the confi rmation key
before cancelling the setting procedure are accepted.
ENGLISH
DEUTSCH
MAGYAR
ITALIANO
ESPANOL
FRANCAIS
POLSKI
CESTINA
2. With the turn control, select the desired new setting, e.g. Español (Spanish).
ENGLISH
DEUTSCH
MAGYAR
ITALIANO
ESPANOL
FRANCAIS
POLSKI
CESTINA
3. Save the setting by pressing the confi rmation key.
SETUP
SIGNALTÖ
CAL
LANGUAGE
TIM
4. To return to the menu overview, press the activation key again.
SETUP
Balance
Einheit
IP Adresse
+30%
192.168.100.100
Alarm Temp
°C
F
TWW
TWB
Timer Mode
SETUP
Unit
IP address
192.168.100.100
Subnet mask
255.255.0.0
Alarm Temp
°C
F
TWW
TWB
Timer mode
Slide-in unit
Grid Shelf
You can now
■ activate another menu function by pressing the corresponding activation key or
■ return to operating mode by pressing the MENU key.
i
In the SETUP display, you can set the following parameters:
■ the IP address and subnet mask of the appliance's Ethernet interface
(for connection to a network)
■ the unit on the temperature display (°C or °F, chapter 7.3.3)
■ how the digital backwards counter with target time setting works
(Timer Mode, chapter 5.3.2)
■ The type of slide-in unit (grid or shelf, chapter 7.3.5)
■ the heat output distribution (Balance, only for appliance sizes , chapter 7.3.6)
■ Remote control (chapter 7.3.8)
■ Gateway (see page 7.3.9)
If the SETUP menu contains more entries than can be displayed, this is indicated by the
display “1/2”. This means that there is a second „page“ of entries.
To display the hidden entries, use the turn control to scroll beyond the lowest entry. The
page display changes to "2/2".
7.3.2
IP address and subnet mask
If you want to operate one ore more appliances in a network, each appliance must have
its own unique IP address for identifi cation. By default, each appliance is delivered with
the IP address 192.168.100.100.
LAN 1: 192.168.1.233
192.168.1.216
LAN 2: 192.168.1.215
LAN 3: 192.168.1.241
192.168.1
Editor
Programmname
Simulation
Protokoll
- +
- +
INP 250 Test 01
180.0°C
i
HPP 250 Labor
i
37.0°C
44.4%rh
15
15
STAND BY
Programm
Einste ll ungen
Protoko ll
Gerät
H fli e
Drucken
AtmoCONTROL
SETUP
Balance
Einheit
IP Adresse
+30%
192.168.100.100
Alarm Temp
°C
F
TWW
TWB
Timer Mode
SETUP
Unit
IP address
192.168.100.100
Subnet mask
255.255.0.0
Alarm Temp
°C
F
TWW
TWB
Timer mode
Slide-in unit
Grid Shelf
1. Activate the SETUP display. The entry IP address is automatically highlighted.
SETUP
Unit
IP address
192.168.100.100
Subnet mask
255.255.0.0
Alarm temp
°C
F
TWW
TWB
Timer mode
Slide-in unit
Grid Shelf
2. Accept the selection by pressing the confi rmation key. The fi rst three digits of the
IP address are automatically selected.
Setup
Unit
IP adress
255.145.136.225
Subnet mask
255.255.0.0
Alarm Temp
°C
F
Timer mode
TWW
TWB
1/2
3. With the turn control, set the new number, e.g. 255.
SETUP
Unit
IP address
255.168.100.100
Subnet mask
255.255.0.0
Alarm temp
°C
F
TWW
TWB
Timer mode
Slide-in unit
Grid Shelf
4. Accept the selection by pressing the confi rmation key. The next three digits of
the IP address are automatically selected. Setting these is done according to the
description above.
SETUP
Unit
IP address
255.145.136.225
Subnet mask
255.255.0.0
Alarm temp
°C
F
TWW
TWB
Timer mode
Slide-in unit
Grid Shelf
5. After setting the last three digits, accept the new IP address by pressing the con-
fi rmation key. The selection returns to the overview. The subnet mask is set ac-
cordingly.
7.3.3
Unit
Here, you can choose whether the temperature is displayed in °C or °F.
7.3.4
Timer mode
Here, you can choose whether the digital backwards counter with target time
setting (timer, chapter 5.3.2) should run setpoint-dependent or not – this deter-
mines whether the timer should not start until a tolerance band of ±3 K around the
set temperature is reached ( B) or whether it should start right after activation (A).
tTimer
tTimer
A
B
°C/°F
t
6 K
SETUP
Unit
IP address
255.145.136.225
Subnet mask
255.255.0.0
Alarm temp
°C
°F
TWW
TWB
Timer mode
Slide-in unit
Grid Shelf
SETUP
Unit
IP address
255.145.136.225
Subnet mask
255.255.0.0
Alarm temp
°C
F
TWW
TWB
Timer mode
Slide-in unit
Grid Shelf
Slide-in unit type steel grid or shelf
Type of slide-in unit (steel grid or shelf)
Here, you have to select the type of slide-in unit (steel grid or shelf) used. The selection
Shelf enables you to adjust the control function to the different air fl ow characteristics
in the interior when using optional sliding shelves instead of the steel grids that are
provided as standard.
7.3.6
Balance
For appliances of the sizes 260, 410 and 750, application-specifi c correction of the heat
output distribution (balance) between the upper and lower heating groups is possible.
The adjustment range is from –50 % to +50 %.
Distribution of the heating/cooling power (example):
The –20 % (left) setting causes the lower Peltier elements to work at 20 % less power
than the upper ones. The +30 % (right) setting causes the lower Peltier elements to
work at 30 % more power than the upper ones. The 0 % setting restores the default
distribution settings.
7.3.7
Dehumidifi cation interval
The dehumidifi cation peltier modules behind the rear panel precisely generate cold
spots inside the chamber in order to remove humidity from the appliance in a controlled
way.
If the device is dehumidifying for a long period of time in the lower end of the climate
diagram, the water in the air will freeze at the dehumidifi cation peltier modules. If solid
ice should form at the rear panel around the dehumidifi cation peltier modules, the dehu-
midifi cation interval must be adjusted.
SETUP
Unit
IP adress
255.145.136.225
Subnet mask
255.255.0.0
°C
°F
Timer mode
Alarm temp
TWW
TWB
Slide-in unit
Grid Shelf
upper heat output
upper heat output
lower heat output
-20%
lower heat output
+30%
SETUP
Balance
-20 %
2/2
SETUP
Balance
+30 %
2/2
The dehumidifi cation interval function allows the time spans at which the dehumidifi ca-
tion peltier modules cool at maximum capacity to be adjusted individually. The preset
value of 35 minutes is recommended for basic applications.
Adjustment range:
■ Min. 15 minutes
■ Max. 180 minutes
Example
1. Interval begins – dehumidifi cation peltier modules cool at full power and generate coldest point (-12°C), depending on the set time
interval.
2. Interval duration expired – dehumidifi cation peltier modules are not operated for a short time, resulting in a local rise in tempera-
ture. The ice thaws and the melt water is channelled out.
3. Interval begins again
The ideal setting for the dehumidifi cation interval is when there is hardly any ice forma-
tion on the rear panel and the setpoint humidity value is reached.
■ The interval should be decreased if there is heavy ice formation on the rear panel
■ If the setpoint value (humidity) is not reached, the interval should be increased
■ For climate points in the low temperature range with low humidity, the interval should
be extended
If you change the dehumidifi cation interval, test whether this has a positive effect on
low ice formation in the interior.
7.3.8
Remote control
In the setup entry remote control, you can set whether the appliance should be con-
trolled via remote control and, if so, in which mode. These settings are available:
■ Off
■ Read Only
■ Write + Read
■ Write + Alarm
If the appliance is in remote control mode, the
symbol appears in the temperature
display. In the settings Write + Read and Write + Alarm, the appliance cannot be con-
trolled at the ControlCOCKPIT until the remote control has been switched off (setting
Off) or set to Read.
In order to use the remote control function, programming skills and special libraries are
required.
7.3.9
Gateway
The setup entry gateway is used to connect two networks with different protocols. The
gateway is set the same way as the IP address (chapter 7.3.2).
Dehum. interval
35min
2/2
Setup
Balance
Remote Control Off
Gateway 192.168.5 .1
+30 %
2/2
23.2°C
TEMP
Set 38.0°C
Balance
Remote Control Off
Gateway 192.168.5 .1
+30 %
2/2
Date and Time
In the TIME display, you can set the date and time, time zone and daylight saving time.
Changes can only be made in manual operating mode.
Always set the time zone (and daylight saving time yes/no) before setting the date and
time. Avoid changing the set time after that since this can lead to gaps or overlapping
when recording measured values. If you still need to change the time, you should not
run a program immediately before or after doing so.
Date and time
Time
Time zone
Daylight savings
12.05.2012
12:00
GMT 01:00
Date
1. Activate the time setting. To do so, press the activation key on the right side of
the TIME display. The display is enlarged and the fi rst adjustment option (Date)
automatically highlighted.
TIME
Time
Daylight savings
12.05.2012
12:00
GMT 01:00
Date
Time zone
2. Turn the turn control until Time zone is highlighted.
TIME
Time
Daylight savings
12.05.2012
12:00
GMT
Date
Time zone
01:00
3. Accept the selection by pressing the confi rmation key.
TIME
Time
Daylight savings
12.05.2012
12:00
GMT
Date
Time zone
00:00
4. Use the turn control to adjust the time zone on the appliance location,
■
e.g. 00:00 for Great Britain
■
01:00 for France, Spain or Germany
Accept the selection by pressing the confi rmation key.
TIME
Time
12.05.2012
12:00
GMT 00:00
Date
Time zone
Daylight savings
5. With the turn control, select the Daylight Savings entry.
TIME
Time
12.05.2012
12:00
GMT 00:00
Date
Time zone
Daylight savings
6. Accept the selection by pressing the confi rmation key. The adjustment options
are highlighted.
i
7. Set daylight savings to off () or on () with the turn control – in this case on
(). Save the setting by pressing the confi rmation key.
i
Daylight saving time and standard time are not changed automatically.
For this reason, please keep in mind to adjust them at the beginning of each period.
TIME
Time
Time zone
Daylight savings
.
.2012
12:00
GMT 00:00
Date
27 05
8. Now, set date (day, month year) and time (hours, minutes) in the same way. Ac-
cept each selection by pressing the confi rmation key.
7.5
Calibration
To guarantee perfect control, we recommend calibrating the appliance once a year.
7.5.1
Temperature adjustment
The appliances are temperature calibrated and adjusted at the factory. In case readjust-
ment should be necessary later on – for example due to infl uence of the chamber load
– the appliance can be calibrated customer-specifi cally using three calibration temper-
atures of your choice:
■ Cal1
Temperature calibration at low temperature
■ Cal2
Temperature calibration at medium temperature
■ Cal3
Temperature calibration at high temperature
For temperature adjustment, you will need a calibrated reference measuring device.
CAL 1
+0,5°C
CAL 3
+0,4°C
CAL 2
-0,4°C
factory calibration
20°C
30°C
40°C
50°C
0°C
10°C
i
Example: Temperature deviation at 30°C is to be corrected
Cal1
°C -0,2K
Cal2 100.0 °C +0,1K
Cal3 180.0 °C -0,2K
JUSTIEREN
Temperatur
letzte Justierung 12.10.2012 12:00
40.0C
0,2K
-
C
0,1K
+
C
0,2K
-
Calibration
Last updated 12.10.2012 12:00:00
Cal1
Cal2
Cal3
40.0
60.0
30.0
Temperature
Humidity
1. Press the activation key to the right of the CALIB display. The display is enlarged
and the temperature adjustment option is automatically selected.
C
0,2K
-
C
0,1K
+
C
0,2K
-
Temperature
l t t K lib i
12 10 2012 12 00
Cal1
Cal2
Cal3 40.0
5.0
20.0
2. Press the confi rmation key repeatedly, until the calibration temperature Cal2 is
selected.
C
0,2K
-
C
0,1K
+
C
0,2K
-
JUSTIEREN
Temperature Cal1
Cal2
Cal3 40.0
5.0
30.0
3. With the turn control, set the calibration temperature Cal2 to 30 °C.
C
0,2K
-
C
K
C
0,2K
-
Temperature Cal1
Cal2
Cal3 40.0
5.0
30.0
+0,1
4. Save the setting by pressing the confi rmation key. The corresponding calibration
value is automatically highlighted.
C
0,2K
-
C
K
C
0,2K
-
Temperature Cal1
Cal2
Cal3 40.0
5.0
30.0
0,0
5. Set the calibration value to 0.0 K and accept the setting by pressing the confi r-
mation key.
21.4°C
TEMP
Set 30.0 °C
6. Position the sensor of a calibrated reference instrument centrally in the appli-
ance's working chamber.
7. Close the door and, in manual mode, adjust the set temperature to 30 °C.
Example: Temperature deviation at 30°C is to be corrected
31.6 °C
30.0°C
TEMP
Set 30.0°C
8. Wait until the appliance reaches the set temperature and displays 30 °C. The ref-
erence instrument displays for example 31.6 °C.
C
0,2K
-
C
K
C
0,2K
-
Temperature Cal1
Cal2
Cal3 40.0
5.0
30.0
+1,6
9. In the SETUP, adjust the calibration value Cal2 to +1.6 K (actual value measured
minus setpoint temperature) and save the setting by pressing the confi rmation
key.
30,0 °C
30.0°C
TEMP
Set 30.0°C
10. After the calibration procedure, the temperature measured by the reference in-
strument should now also be 30 °C.
With Cal1, a calibration temperature below Cal2 can be programmed accordingly, and
with Cal3, a temperature above. The minimum difference between the Cal values is 10
K.
If all calibration values are set to 0.0 K, the factory calibration settings are restored.
Humidity Humidity profi le (for constant climate chambers HPPeco only)
Humidity control of the constant climate chamber HPPeco can be adjusted according
to customer requirements by means of three freely selectable balance points. For each
selected balance point a positive or negative compensation correction value can be set
between -10 % and +10 %.
For humidity adjustment, you will need a calibrated reference measuring device.
rh 20
+3,0 %
rh 90
-1,5 %
60%
40%
0%
20%
80%
100%
rh 50
+1,5 %
Factory calibration
Example: Humidity deviation at 60 % should be corrected
Cal1
°C -0,2K
Cal2 100.0 °C +0,1K
Cal3 180.0 °C -0,2K
JUSTIEREN
Temperatur
letzte Justierung 12.10.2012 12:00
40.0C
0,2K
-
C
0,1K
+
C
0,2K
-
Calibration
Last updated 12.10.2012 12:00:00
Cal1
Cal2
Cal3
40.0
60.0
30.0
Temperature
Humidity
1. Press the activation key to the right of the CALIB display. The display is enlarged
and the temperature adjustment option is automatically selected.
%rh
%rh
0,5%
-
1,0%
+
Temperature
l t t K lib i
12 10 2012 12 00
Cal1
Cal2
40.0
50.0
Humidity
%rh
+1,0%
Cal3 80.0
2. Turn the turn control until Humidity is highlighted.
%rh
%rh
0,5%
-
1,0%
+
Temperature Cal1
Cal2
40.0
Humidity
50.0
%rh
+1,0%
Cal3 80.0
3. Press the confi rmation key repeatedly, until the calibration point Cal2 is selected.
%rh
%rh
0,5%
-
1,0%
+
Temperature Cal1
Cal2
40.0
Humidity
60.0
%rh
+1,0%
Cal3 80.0
4. With the turn control, set the calibration point Cal2 to 60 % rh.
i
Example: Humidity deviation at 60 % should be corrected
%rh
%rh
0,5%
-
%
Temperature Cal1
Cal2
40.0
Humidity
60.0
1,0
+
%rh
+1,0%
Cal3 80.0
5. Save the setting by pressing the confi rmation key. The corresponding calibration
value is automatically highlighted.
%rh
%rh
0,5%
-
%
Temperature Cal1
Cal2
40.0
Humidity
60.0
0,0
%rh
+1,0%
Cal3 80.0
6. Set the calibration value to 0.0 % and accept the setting by pressing the confi r-
mation key.
HUMIDITY
29.5%rh
Set 60.0%rh
7. Position the sensor of the calibrated reference instrument centrally in the working
chamber of the appliance.
8. Close the door and, in manual mode, adjust the set humidity to 60 % rh.
58.5 %rh
9. Wait until the appliance reaches the set humidity and displays 60 % rh. The refer-
ence instrument displays for example 58.5 % rh.
%rh
%rh
0,5%
-
%
Temperature
letzte Kalibrierung 12 10 2012 12 00
Cal1
Cal2
40.0
Humidity
60.0
-1.5
%rh
+1,0%
Cal3 80.0
10. In the SETUP, adjust the compensation correction value Cal2 to –1.5% (actual
value measured minus setpoint temperature) and save the setting by pressing the
confi rmation key.
60.0 %rh
HUMIDITY
60.0%rh
Set 60.0%rh
11. After the calibration-, the humidity measured by the reference instrument should
now also be 60 % rh.
In the Program display, programmes created using the AtmoCONTROL software can be
transferred to the appliance and saved on a USB storage medium. Here, you can also
select the programme provided for use (chapter 5.3.3 ) and delete programmes.
To load a programme from a USB storage medium: Connect the USB storage
medium with the saved programme(s) to the interface on the right side of the
ControlCOCKPIT.
Program
Delete
Test 012
Test 013
Test 014
Test 015
Test 017
Test 016
Select
1. Activate the programme display. To do so, press the activation key on the left side
of the Prog display. The display is enlarged and the entry Select automatically
highlighted. The programs available for activation are shown on the right. The
program currently available for use – in this example Test 012 – is highlighted in
orange.
Select
Delete
Test 012
Test 022
Test 013
Test 014
Test 015
Test 022
2. Access the Select function by pressing the confi rmation key. All programs avail-
able are displayed, including the ones saved on the USB storage medium (identi-
fi ed by the USB symbol
). The program currently available for use is highlight-
ed in orange.
Select
Delete
Test 012
Test 022
Test 013
Test 014
Test 015
Test 023
3. With the turn control, select the program you want to make available for use.
Select
Delete
Test 012
Test 022
Test 013
Test 014
Test 015
Test 023
4. Accept the selection by pressing the confi rmation key. The program is now load-
ed, which is indicated by the transfer symbol.
Delete
Test 012
Test 022
Test 013
Test 014
Test 015
Test 023
Select
5. As soon as the program is ready, the selection returns to Select. To start the pro-
gramme: As described in the chapter 5.3.3, return to operating mode by pressing
the MENU key.
You can now remove the USB storage medium.
To delete a program, select Delete with the turn control and select the program to
be deleted the same way you can select a program for activation.
i
In the Sound display, it can be defi ne whether or not the appliance should emit acoustic signals and, if yes, on which events:
■ on the press of a key
■ at the end of a program
■ On alarm
■ if the door is open
Sound
Keysound
At the end
If door open
On alarm
1. Activate the acoustic signal adjustment. To do so, press the activation key on the
left side of the Sound display. The display is enlarged. The fi rst category (in this
case Keysound) is automatically highlighted. On the right, the current settings are
shown on.
Keysound
At the end
If door open
On alarm
Sound
If you want to edit another list entry: Turn the turn control until the respective
entry – e.g. if door open (special confi guration) – is highlighted in colour.
Keysound
At the end
If door open
On alarm
2. Save the selection by pressing the confi rmation key. The adjustment options are
automatically highlighted.
Keysound
At the end
If door open
On alarm
3. With the turn control, select the desired setting – in this example OFF ().
Keysound
At the end
If door open
On alarm
4. Save the setting by pressing the confi rmation key.
i
If an acoustic alarm sounds, it can be turned off by pressing the confi rmation
key.
i
The appliance continually logs all relevant measured values, settings and error messag-
es at 1-minute intervals. The internal log memory is of the continuous memory type. The
logging function cannot be switched off and is always active. The measured data are
stored in the appliance, safe from manipulation. If the power supply is interrupted, the
time of the power failure and voltage recovery are stored in the appliance.
You can read out the log data for different periods to a USB storage medium via the USB
port or via Ethernet and then import them to the AtmoCONTROL program for graphical
representation, printing out or storage.
The log memory of the appliance is not modifi ed or deleted by reading it out.
1. Connect the USB storage medium to the USB port on the right of the
ControlCOCKPIT.
Protocol
Last download 12.10.2012 12:00
1 month
Complete control range
1 week
2. Activate the protocol. To do so, press the activation key on the right side of the
PROTOCOL display. The display is enlarged and the period This month automati-
cally highlighted. To select another logging period, use the turn control.
Protocol
1 Month
Complete control range
1 Week
3. Save your selection by pressing the confi rmation key. The transfer starts and a
status symbol indicates the progress.
Protocol
l
D
l
d 25 10 2012 13 20
1 Month
Complete control range
1 Week
4. As soon as the transfer is complete, a check mark appears in front of the period
selected. You can now remove the USB storage medium.
For a description of how to import and process protocol data in AtmoCONTROL or read
it out via Ethernet, please refer to the separate AtmoCONTROL manual.
i
With the USER ID function, you can lock the settings of individual (e.g. temperature) or
all parameters, so that they cannot be changed at the appliance by accident or unau-
thorised persons. You can also lock setting options in menu mode (e.g. adjustment or
date and time settings) this way.
If adjustment options are locked, this is indicated by the lock symbol in the respective
display.
USER ID data is entered in the AtmoCONTROL software and saved on the USB storage
medium. The USB storage medium is thus acting as a key: Parameters can only be
locked or unlocked if it is connected.
A description of how to create a USER ID in AtmoCONTROL is provided in the separate
AtmoCONTROL manual.
7.9.2
USER ID activation and deactivation
1. Connect the USB storage medium with the USER ID data to the USB port on the
right of the ControlCOCKPIT.
USER-ID
Deactivate
Activate
2. Activate the USER ID. To do so, press the activation key on the right side of the
USER ID display. The display is enlarged and the entry Activate automatically
highlighted.
USER-ID
Deactivate
Activate
3. Confi rm the activation by pressing the confi rmation key. The new USER ID data
are transferred from the USB storage medium and activated. As soon as activa-
tion is complete, a check mark appears in front of the corresponding entry.
4. Remove the USB storage medium. Locked parameters are indicated by the lock
symbol on the respective display.
To unlock the appliance, connect the USB storage medium, activate the USER ID dis-
play and select the entry Deactivate.
22.4
40 4
.
TEMP
Set
°C
°C
i
i
Maintenance and Servicing
8.
Maintenance and Servicing
Risk of electric shock.
Disconnect the mains plug before any cleaning or maintenance work.
With appliances above a specifi c size, you could become accidentally locked inside,
which could put you at risk of death. Do not climb into the appliance!
Danger of cuts due to sharp edges.
Always wear gloves when working in the interior of the chamber.
8.1
Cleaning
8.1.1
Interior and metal surfaces
Regular cleaning of the easy-to-clean interior prevents build up of material remains that
could impair the appearance and functionality of the stainless steel chamber over time.
The metal surfaces of the appliance can be cleaned with normal stainless steel clean-
ing agents. Make sure that no rusty objects come into contact with the interior or with
the stainless steel housing. Rust deposits can lead to an infection of the stainless steel.
If rust spots should appear on the surface of the interior due to impurities, the affected
area must be immediately cleaned and polished.
8.1.2
Plastic parts
Do not clean the ControlCOCKPIT and other plastic parts of the appliance with caustic
or solvent-based cleaning agents.
8.1.3
Glass surfaces
Glass surfaces can be cleaned with a commercially available glass cleaner.
8.1.4
Peltier cooling modules
In order to guarantee perfect function and long lifetime of the Peltier cooling modules,
it is absolutely essential that you remove dust deposits from the heat sink on the rear
of the appliance (with a vacuum cleaner, paintbrush or bottle brush, depending on the
amount).
WARNING
WARNING
CAUTION
Maintenance and Servicing
8.2
Regular maintenance
Once a year, grease the moving parts of the doors (hinges and lock) with thin silicone
grease and check that the hinge screws are not loose.
To guarantee perfect control, we recommend calibrating (chapter 7.5) the appliance
once a year.
8.3
Repairs and service
Live components may be exposed once the covers have been removed. Touching
these can lead to an electric shock. Disconnect the mains plug before removing any
covers. Any work inside the appliance may only be performed by qualifi ed electri-
cians.
Repairs and service work are described in a separate service manual.
WARNING
Storage and disposal
9.
Storage and disposal
9.1
Storage
The appliance may only be stored under the following conditions:
■ in a dry and enclosed, dust-free room
■ frost-free
■ disconnected from the power supply
Before storage, remove water tube and empty the water tank (chapter 4.1.1).
9.2
Disposal
This product is subject to Directive 2012/19/EC on Waste Electrical and Electronic
Equipment (WEEE) of the European Parliament and of the Council of Ministers. This
appliance was placed on the market after 13 August 2005 in countries which have al-
ready integrated this Directive into their national laws. It may not be disposed of in
normal household waste. For disposal, please contact your dealer or the manufacturer.
Any appliances that are infected, infectious or contaminated with materials hazardous
to health are excluded from return. Please also observe all other regulations applicable
in this context.
Before disposing of the appliance, please render the door locking mechanism unusable,
for example to prevent playing children from being locked inside the appliance.
There is a lithium battery in the ControlCOCKPIT of the appliance. Remove it and dis-
pose of it in accordance with the regulations in your country.
■ Note for Germany:
The appliance may not be left at public or communal recycling or collection points
Storage and disposal
Memmert GmbH + Co. KG | Äußere Rittersbacher Straße 38 | 91126 Schwabach
Tel. +49 9122 925-0 | E-Mail: [email protected] | www.memmert.com
facebook.com/memmert.family | instagram.com/memmert.family
Constant climate chamber HPPeco
Peltier cooled incubator IPPeco plus
Operating manual
D49025
Last updated 03/2021
English