VWR® P4 Spectrophotometer
Instruction Manual
European Catalogue Number
EU/UK/CH Plug: 634-1089
Version:
1.0.0
Issued:
24, June 2019
Legal Address of Manufacturer
Europe
VWR International bvba
Researchpark Haasrode 2020
Geldenaaksebaan 464
B-3001 Leuven
+ 32 16 385011
http://be.vwr.com
Made in China
Table of Contents
Warning
1
Avoiding Electrical Shock
1
Package Contents
2
Unpacking
2
Installation
2
Intended use
2
Symbols and conventions
2
Product Specifications
2
Overview
3
Description
3
Getting Started
4
Turn On and Self-check
4
Important Guidelines
4
General Operating Instructions
4
Touch Screen Using Tips
4
Select Application
4
Basic Operation
5
Measurement Results Operation
6
Files Operation
7
Calibration and System Settings
8
Performance Verification
11
Operation
13
Photometry
14
Quantitation
15
Spectrum
19
Troubleshooting
22
Repair and maintenance
22
Daily Maintain
22
Spare Parts Replacement
22
User replaceable accessories and spare parts
25
Technical service
26
Warranty
26
Compliance with local laws and regulations
26
Equipment disposal
27
Functions
28
Installation
28
Install and Uninstall VWR Software Multi Wave Pro
28
Run VWR Software Multi Wave Pro
31
Introduction
32
Main Interface
32
Menus Bar and Tools Bar
33
General Operating Instructions
33
Operation
36
Quantitative Analysis
36
Kinetics Analysis
38
Spectrum Scan
39
Multi-wavelength Analysis
40
DNA/Protein Analysis
41
Energy Scan
42
Other Functions
43
APPENDIX 1
43
APPENDIX 2
44
Avoiding Electrical Shock
Please follow the guidelines below, and read this manual in its entirety to ensure safe operation of the unit.
VWR recommends against the use of P4 Spectrophotometer.
To avoid electrical shock
Do not open the device.
Disconnect the device from the mains supply before carrying out maintenance work or
changing the fuses.
The inside of the device is a high-voltage area Danger!
Do not use the device if it is damaged, especially if the main power cable way is in any
damaged or defective.
Repairs may only be carried out by the service technicians from your local VWR office and
authorized contractual partners.
The device must be connected to a power outlet that has a protective ground connection.
If the equipment is used in a manner not specified by the manufacturer, the protection
provided by the equipment may be impaired.
Avoiding Damage to the Instrument
Do not allow any liquid to enter into the device.
Do not operate the device in a hazardous location or potentially explosive environment.
Power Cord ( Euro Plug )
1PC
on request
Power Cord ( UK Plug )
1PC
on request
Power Cord ( CH Plug )
1PC
on request
Software CD-ROM
1PC
N/A
USB Cable
1PC
N/A
Instruction Manual
1PC
N/A
Dust Cover
1PC
on request
Unpacking
Open the package, according to carefully check the packaging packing list items, if found inside the packaging are
missing or damaged items please contact your local VWR office and authorized contractual partners.
Installation
Placement
Place the instrument on the stable table carefully.
Install printer (Optional)
Check to confirm instrument power switch is turned off, connect the printer’s data cable to the Instrument’s
serial/USB port.
Connect the power cord
Check to confirm instrument power switch is turned off, the power cord plug into two separate power interface and
power supply socket apparatus.
Intended use
For research use only. Not intended for any animal or human therapeutic or diagnostic use.
Symbols and conventions
The following chart is an illustrated glossary of the symbols that are used in this manual.
CAUTION This symbol indicates a potential risk and alerts you to proceed with caution^
CAUTION This symbol indicates the presence of high voltage and warns the user to proceed
with caution
CAUTION This symbol indicates risks associated with hot surfaces
Product Specifications
Optical System
Single Beam
Wavelength Range
190-1100nm
Band Width
2nm
Stray Light
≤0.05%T @ 220nm & 360nm
Photometric Range
0 to 200%T, -0.3 to 3.0A, 0 to 9999C
Wavelength Accuracy
±0.5nm
Photometric Accuracy
±0.4%T or ±0.004A @ 1A
0.002A/h @ 500nm
Memory
236KB
Language
English, French, German, Spanish, Portuguese, Chinese
Display
TFT Colour LCD (5”, Touch Screen)
Interface
USB-A, USB-B,RS-232
Measuring Procedure
Photometry, Quantitation, Spectrum Scan
Power Supply
AC 100~240 V, 50~60 Hz
Dimension
456×360×185
Weight
10.7kg
Work Environment
15 to 35℃, 15 to 70% relative humidity
Store Environment
-10 to 50℃, 15 to 70% relative humidity
This instrument is compliant to the European Directives on Low Voltage Directive 2014/35/EU,
Electromagnetic compatibility 2014/30/EU.
Restriction on use of Hazardous Substances RoHS 2011/65/EU and their amendments.
Overview
P4 Spectrophotometer used in Chemistry, Pharmaceuticals, Biochemical, Metallurgy, Light Industry, Textile,
Material, Environments, Medical, Education and some other fields for Quality Control laboratories.
Description
Front View
Touch
screen
Lid of the
Compartment
Pull Rod
Lid of Lamp
Room
Right View
USB Port
(for USB Disk/
USB Printer)
Power
Switcher
Rear View
Turn On and Self-check
Switch on the power. Self-check includes the following steps: Turn On Lamp → Locating Filter Disc → Locating
Automatic Sample Holder (If Installed) → Get Dark Current → Locating Wavelength → Check Energy → Check
System baseline.
Important Guidelines
Reagents and dilution buffers can cause cauterization and other damage to health.
Samples (nucleic acids, proteins, bacteria cultures) can be infectious and cause serious damage to
health.
During sample preparation, measuring procedures and maintenance and cleaning work, observe all local
laboratory safety precautions (e.g. wear protective clothing and gloves, use of disinfectant) regarding the
handling of sample material.
Dispose of measuring solutions and cleaning and disinfectant materials in accordance with the relevant
local laboratory regulations.
General Operating Instructions
Touch Screen Using Tips
The entire screen can be started with a touch. To make a choice, use your nails, fingertips, pencil, or stylus to
press the screen. Don't press the screen with sharp objects (such as ball point).
Select Application
Main Interface, press the icon to select application.
Photometry Measure the absorbance or transmittance of the sample.
Quantitation Establish the standard curve and measure the concentration of the sample.
Spectrum Scan the sample in a wavelength range.
File Manage files stored in the instrument or USB disk.
System System calibration and setup.
Performance verification Verify the performance of the instrument.
Basic Operation
Home Back to main interface.
Return Back to the previous interface.
/
Page Up/Down Go to previous/next page.
Measurement Results Operation
Open Open result(s) from internal/USB memory.
Save Save result(s) to internal/USB memory.
Print Print result(s).
Delete Delete selected result(s).
Rename, Print and Delete Results
Rename a Sample:
List interface, press the area Name, key in the sample name (Up to
8 characters).
Print
the
Measurement
Report:
List interface, press the icon
.
Delete sample(s):
List interface, press the Check Box, and press the icon
.
Open Results
Open:
1.
List interface, press the icon
.
2.
Press the icon internal memory/USB memory to select the
memory which the file saved.
Press file lists to select, press the button Open.
Save Results
Save:
1.
List interface, press the icon Save.
2.
Press the icon
/
to select the Internal/USB
memory which the file to save.
3.
Type in the file name, press the button Save.
Files Operation
Internal Memory Internal memory of the spectrophotometer.
USB Memory USB extended mass memory.
Copy Copy the selected file(s) from internal /USB memory to USB/internal memory.
Export csv Export file(s) To *.csv format
Export txt Export file(s) To *.txt format
Delete Delete the selected file(s).
Rename, Import, Export and Delete Files
Rename a File:
File management interface, press the area Name, key in the file
name (Up to 8 characters).
Copy
File(s)
From/To
Internal
Memory/USB
Memory:
File management interface, press the Check Box, press the button
(Need USB disk).
Export File(s) To *.csv
Format
File management interface, press the Check Box, press the button
(Need USB disk).
Export File(s) To
*.txt
Format
File management interface, press the Check Box, press the button
(Need USB disk).
Delete File(s):
File management interface, press the Check Box, and press the
icon
.
Calibration and System Settings
Select the icon
in the main interface. Display options to calibrate the system and configure the basic
instrument settings.
Calibration
Calibrate Start to do calibration.
Select Tab Calibration in the System interface. Remove something in the measurement channel, close the
sample chamber cover, select the item Dark current, Wavelength or System baseline, press the icon
to
do calibration.
Settings of Light Source
Tungsten lamp reset Reset the Tungsten lamp usage.
Deuterium lamp reset Reset the Deuterium lamp usage.
Select Tab Light source in the System interface. The light source information is displayed on the screen.
Press the icon
to turn on/off the Tungsten lamp/Deuterium lamp.
Change the lamp switching
point:
Press the value of lamp switching point. Input the new value.
Reset the lamp usage:
Press the icon
/
to reset the Tungsten lamp/Deuterium
lamp usage time.
Important information
1. If only one of the light sources is used in for a long period of time, please turn off another light
source to save energy.
2. If the lamp switching point is changed, the system baseline must be recalibrated.
Edit Clock
Accept Accept the new value.
Select Tab Clock in the System interface. Press the value of year, month, date, hour, minute or second to change.
Press the icon
to accept new value.
Memory Management
Format Internal Memory Format the internal memory of the spectrophotometer.
Format USB Memory Format the USB mass storage.
Select Tab Memory in the System interface. The use of the internal and USB memory (If inserted) are show. Press
the icon
/
to format internal memory/USB memory.
Language Selection
Accept Accept the new language.
Select Tab Language in the System interface. Select a language, press the icon
to change.
General Options
Select Tab General Options in the System interface.
Beep:
Press the icon
to turn on/off the beep.
Brightness:
Press the icon
/
to decrease/increase the brightness of
the LCD display.
Close display:
Press the icon
to turn on/off. If turned on, the display will
close automatically if no operation for 30 minutes.
Sample holder:
If you have a new automatic sample holder installed, you need to set
the sample holder type by pressing the icon
/
(AC-5: 5 cells automatic sample holder, AC-8:8 cells
automatic sample holder).
Restore Defaults
Restore Restore the parameters to factory settings.
Select Tab Restore defaults in the System interface. Select an item, press the icon
to restore.
Performance Verification
Select the icon
in the main interface. Display options to verify the performance of the instrument.
Before verifying the performance, the instrument needs to be
preheated for 30 minutes, and then re-measure dark current.
Verifying Wavelength Accuracy and Wavelength Repeatability
Select Tab Wavelength accuracy in the Performance verification interface.
Standard Sample:
Holmium oxide solution or equivalent filter
Measurement:
1.
Remove something in the measurement channel, close the sample
chamber cover, press the wavelength value, type in the wavelength
of measurement, press the button Zero;
2.
Put the Standard Sample in the measurement channel, press the
button Measure;
3.
Repeat step 2 to do measurement three times. The difference
between the average of the three measurements and the standard
value is the single-point wavelength tolerance. The difference
between the maximum and minimum values of the three
measurements is single point wavelength repeatability;
4.
Repeat step 1-3 to do measurement single-point wavelength
tolerance one by one. The maximum value in the single-point
wavelength tolerance is wavelength accuracy. The maximum value
in the single-point wavelength reproducibility is wavelength
repeatability.
Verifying Photometric Accuracy and Photometric Repeatability
Select Tab Photometric accuracy in the Performance verification interface.
Standard Sample:
NIST 930D or equivalent filter
Measurement:
1.
Remove something in the measurement channel, close the sample
chamber cover, press the wavelength value, type in the wavelength
of measurement, press the button Zero;
2.
Put the Standard Sample in the measurement channel, press the
button Measure;
3.
Repeat step 2 to do measurement three times. The difference
between the average of the three measurements and the standard
value is the single-point photometric tolerance. The difference
between the maximum and minimum values of the three
measurements is single point photometric repeatability;
4.
Repeat step 1-3 to do measurement single-point photometric
tolerance one by one. The maximum value in the single-point
photometric tolerance is photometric accuracy. The maximum value
in the single-point photometric reproducibility is photometric
repeatability.
Verifying Stray Light
Select Tab Stray light in the Performance verification interface.
Standard Sample:
10g/L NaI solution or equivalent filter (220nm), 50g/L NaNo2 solution
or equivalent filter (340 or 360nm)
Measurement:
1.
Remove something in the measurement channel, close the sample
chamber cover, press the wavelength value, type in the wavelength of
measurement;
2.
Put the Reference in the measurement channel, press the button Zero;
3.
Put the Standard Sample in the measurement channel, press the
button Measure, the result is the stray light of this wavelength.
Verifying Noise (0A)
Select Tab Noise (0A) in the Performance verification interface.
Standard Sample:
None
Measurement:
1.
Remove something in the measurement channel, close the sample
chamber cover, press the wavelength value, type in the wavelength 500,
press the button Zero;
2.
Press the button Measure, the result is the noise of this wavelength.
Verifying Noise (0%T)
Select Tab Noise (0%T) in the Performance verification interface.
Standard Sample:
Block
Measurement:
1.
Remove something in the measurement channel, close the sample
chamber cover, press the wavelength value, type in the wavelength 500,
press the button Zero;
2.
Put the Block in the measurement channel, press the button Measure,
the result is the dark noise of this wavelength.
Verifying Stability
Select Tab Noise in the Performance verification interface.
Standard Sample:
None
Measurement:
1.
Remove something in the measurement channel, close the sample
chamber cover, press the wavelength value, type in the wavelength 500,
press the button Zero;
2.
Press the button Measure, the result is the noise of 500nm.
Verifying Bandwidth
Select Tab Dark Bandwidth in the Performance verification interface.
Standard Sample:
None
Measurement:
1.
Remove something in the measurement channel, close the sample
chamber cover, press the wavelength value, type in the wavelength
656.1;
2.
Press the button Measure, the result is the bandwidth.
Operation
Important information
The cuvettes must be clear and there’s no remains of the
samples on the surface of it. Only Silicon (Quartz) cuvettes are permitted to be used in the range of
UV area.
Photometry mode is used to measure the absorbance or transmissivity of the sample.
1.
Main interface, press the icon
to start a Photometry application.
Mode Switch measurement mode to %T, Abs or Energy.
Wavelength Set measurement wavelength.
Zero Do 0Abs/100%T.
Read Measure sample and record the result.
List View the result(s) list.
Increase/Decrease Increase/Decrease the gain of signal. Only for Energy mode.
2.
Press the icon
to switch to the measurement mode.
Abs
Measure absorbance value of the sample(s).
%T
Measure transmittance value of the sample(s).
E
Measure energy value of the sample(s).
3.
Press the icon
to set wavelength, key in the measurement wavelength.
4.
Put the reference in the measurement channel, press the icon
to do zero.
5.
Put the sample in the measurement channel, press the icon
to measure a sample and record the
result.
6.
Press the icon
to browse the result(s).
Quantitation mode is used to measure the concentration of the sample(s).
1
Main interface, press the icon
to start a Quantitation application.
2
Establish Method
2.1
Quantitation interface, press the button Establish method.
A=A1: Absorbance is equal to the measured absorbance value of the
measured wavelength 1
A=A1-m*A2: Absorbance is equal to the difference between the
absorbance value of the measured absorbance at the wavelength 1 and the
wavelength 2, m is the Coefficient
A=A1/A2: Absorbance is equal to the ratio of the measured absorbance
value of the measured wavelength 1 and 2
Wavelength 1
Measurement wavelength 1
Wavelength 2
Measurement wavelength 2
Fitting
LIN-0: Linear to zero
LIN: Linear.
QUA: Quadratic.
Unit
- (No Unit), %, ppm, ppb, g/L, mg/L, μg/L, ng/L, g/dL, mg/dL, μg/dL, mg/mL,
μg/mL, ng/mL, μg/μL, ng/μL, mol/L, mmol/L, IU, Custom(User input, Up to 8
characters).
Calibration
Coe K: Input equation coefficient.
Std M: Measure standard sample(s)
Std I: Input standard sample(s)
Standard quantity
Standard sample number (Up to 10)
2.2
Press the item to set measurement parameters.
2.3
After all the parameters are set up, press the button Next to start establishing the standard curve. If the
item Calibration is set to the parameter Coe K, Std M or Std I, please refer to 2.3.1, 2.3.2 or 2.3.3.
2.3.1 Input equation coefficient to establish standard curve.
(1) Input equation coefficient K0 ~ K3. Press button Next.
2.3.2 Measure standard sample to establish standard curve
(1) Put the reference in the measurement channel, press the button Zero to do zero.
(2) Put the 1# standard sample in the measurement channel, press the button Read to measure.
(3) Repeat step 3.3 to measure other of the standard samples.
(4) Press the item to input concentration of standard samples, press the button Next.
2.3.3 Input standard sample to establish standard curve
(1) Press the item Abs and Conc to input absorbance and concentration of standard samples,
press the button Next.
2.4
Finished establish method. Press the button Save to save the method, press the button Measure to
accept the new method and go to the measurement interface, Press the button Finish to exit.
Quantitation interface, press the button Measure sample.
Method Select measurement method.
Zero Do 0Abs/100%T.
Read Measure the sample and record the result.
List View the result(s) list.
3.2
Press the icon
to select method.
Press the button Open to load measurement method stored in the internal memory/USB disk.
3.4
Press the button Measure to accept the new measurement method and back to measurement
interface.
3.5
Put the reference in the measurement channel, press the icon
to do zero.
3.6
Put the sample in the measurement channel, press the icon
to measure a sample and record the
result.
3.7
Press the icon
to browse the result(s).
Spectrum
Spectrum mode is used to scan the absorbance or transmissivity of the sample in a wavelength range.
1.
Main interface, press the icon
to start a Spectrum application.
Method Set the measurement parameters.
Zero Scan baseline.
Read Scan the sample and draw curve.
Stop Stop scanning.
List View the result(s) list.
2.
Press the icon
to setup the measurement parameters.
Start wavelength
Scan start wavelength
End wavelength
Scan end wavelength
Step
Scan interval: 0.1,0.2, 0.5, 1.0,2.0, 5.0,10.0 nm
Speed
HS: High speed
MS: Medium speed
LS: Low speed
Photometry mode
Abs: absorbance
%T: transmissivity
Y minimum
Minimum ordinate
Press the item to select or key in the parameters, press the button Measure to accept the new parameters
and back to measurement interface.
4.
Put the reference in the measurement channel, press the icon
to scan baseline.
5.
Put the sample in the measurement channel, press the icon
to scan a sample and record the result.
6.
Press the icon
to browse the curve and result(s).
Scale Set the Coordinate value.
Left Moves the cursor to the left point(peak) to point(peak).
Right Moves the cursor to the left point(peak) to point(peak).
Mode %T Change the mode to %T.
Mode Abs Change the mode to Abs.
Peak Change the search mode point/peak.
Review the information in the table below to troubleshoot operating problems.
Problem
Cause
Solution
Power on, no response
Power cord connection is not reliable
Improve connectivity
Fuse burning
Replace fuse
Measurement uncertainty
Glass cuvettes used in UV Range
Use quartz cuvettes
Sample is not Stable
Improve the sample
The concentration of sample is too high Diluted sample
Power Supply Voltage Low or not
Stable
Improve the Power Supply
Lamp damage or lamp life maturity
Replace lamp
Dark Current Error when self-
check
The lid of the compartment is open
during self-check
Close the lid, restart
System Calibrate Failed
Something block the Light path
Remove it, calibrate again
Measurements inaccurate
Cuvettes were contaminated
Clean cuvettes
Samples were contaminated
Improve samples
Worse matching of the cuvettes
Improve the matching of the cuvettes
Dark current error
Resample dark current
Repair and maintenance
Daily Maintain
Check the compartment
After measurement, the cuvettes with sample solutions should be taken out of the compartment in time. Or the
volatilization of the solution would make the mirror go moldy. Users must pay more attention to the corrosive
sample and liquid easy to volatilize. Any solution remains in the compartment should be wipe off immediately.
Surface clean
The cover of the instrument is with paint. Please use wet towel to wipe off the drips on the surface immediately.
Organic solution is forbidden to be used to clean the cover. Please wipe off the dirt on the cover timely.
Clean the cuvettes
After every test or after a solution change, the cuvettes should be cleaned carefully, or the remains on the surface
would cause measuring error.
Spare Parts Replacement
Replace the fuse
Danger! Be sure to switch off the power and unplug the socket before replacement!
1. Tools preparation
Prepare a 3×75 Flat Blade screwdriver.
2. Switch Off the power supply
Switch off the power supply, and unplug the socket.
3. Take out the Fuse Seat
Push the fuse case by using the screwdriver, and turn it counterclockwise, the fuse seat will pop out when
released.
4. Replace a new fuse
Pick out the spare fuse (3.15A/250V) and replace it.
5. Reset the fuse seat
Replace the fuse seat in the power socket. Push the fuse case by using the screwdriver, and turn it clockwise,
the fuse seat will be locked when released.
6. Switch on the power
Plug the socket and switch on the power.
Replace lamps
Hot! Wait 20 minutes before open the lamp chamber after power off to avoid scald!
1. Tools preparation
Prepare a 6×150mm Flat Blade screwdriver and a pair of glove.
2. Power Off
Switch off the power supply and unplug the socket.
3. Open the cover
Loosen the indicated two screws and remove the lamp cover.
4. Replace the D2 lamp
Unplug the connector(No. 1) .Unscrew the 2 screws on the D2 Flange (No.2) and remove the D2 lamp. Draw
on the cotton glove and replace a new lamp. Fix the 2 screws and plug the connector again.
5. Replace W lamp
The Tungsten lamp is equipped with a blue-grey silicon coating by manufacturer. This coating is only a
transport safety device. It can be removed with the first exchange of lamp.
Pull out the defected W lamp and draw on the cotton glove. Insert the new W lamp as deep as possible on the
lamp seat.
Tungsten
lamp
6. Adjust the position of the W lamp
Switch on the power(the Switch Mirror should be placed to the position as indicates). Observe the entrance facular,
and it should in the center of the entrance hole. If the facular deviate to Left or Right, then loosen the two screws
and move the lamp seat to Left or Right until it focus on the center of the slot. Then fix the screws.
Reset the cover of the light chamber and fix the screws. Reset the cover of the lamp room and fix the screws
User replaceable accessories and spare parts
Description
Quantity
Cat. No.
CELL HOLDER, 4-CELL, 10MM
1PC
634-6003
CELL HOLDER, 4-CELL,10 TO 50MM
1PC
634-6004
CELL HOLDER, 4-CELL, 10 TO 100MM
1PC
634-6005
CELL HOLDER, FOR CYLINDRICAL CELL
1PC
634-6006
CELL HOLDER, WATER-JACKETED, 1 CELL, 10MM
1PC
634-1135
CELL HOLDER FOR MICRO CELLS,BEAM HEIGHT:15MM
1PC
634-0687
CELL HOLDER FOR TEST TUBES, D=16mm
1PC
634-1133
CELL HOLDER,5-POSITION AUTO CELL CHANGER, 10 TO 100MM
1PC
634-1136
CELL HOLDER,8-POSITION AUTO CELL CHANGER, 10MM
1PC
634-6010
CELL HOLDER, SOLID SAMPLE
1PC
634-6011
CELL HOLDER, WATER-JACKED, 4 CELL, 10MM
1PC
634-1134
CUVETTE, SQUARE.GLASS,10MM
4PCS
634-6013
CUVETTE, SQUARE.GLASS,20MM
2PCS
634-6014
CUVETTE, SQUARE.GLASS,30MM
2PCS
634-6015
CUVETTE, SQUARE.GLASS,50MM
2PCS
634-6016
CUVETTE, SQUARE.GLASS,100MM
1PC
634-6017
CUVETTE, SQUARE.QUARTZ,10MM
2PCS
634-6018
CUVETTE, SQUARE.QUARTZ,20MM
2PCS
634-6019
CUVETTE, SQUARE.QUARTZ,30MM
2PCS
634-6020
CUVETTE, SQUARE.QUARTZ,50MM
2PCS
634-6021
CUVETTE, SQUARE.QUARTZ,100MM
1PC
634-6022
CELL,QUARTZ, 100UL, 10MM, BEAM HEIGHT:15MM
1PC
634-0688
CELL,QUARTZ, 200UL, 10MM, BEAM HEIGHT:15MM
1PC
634-0689
CELL,QUARTZ, 500UL, 10MM
1PC
634-6025
FLOW CELL, 5 MM, GLASS, BEAM HEIGHT:15MM
1PC
634-0690
FLOW CELL, 10 MM, GLASS, BEAM HEIGHT:15MM
1PC
634-0691
FLOW CELL, 20 MM, GLASS, BEAM HEIGHT:15MM
1PC
634-0692
FLOW CELL, 30 MM, GLASS, BEAM HEIGHT:15MM
1PC
634-0693
FLOW CELL, 5 MM, QUARTZ, BEAM HEIGHT:15MM
1PC
634-0694
FLOW CELL, 10 MM, QUARTZ, BEAM HEIGHT:15MM
1PC
634-0695
FLOW CELL, 20 MM, QUARTZ, BEAM HEIGHT:15MM
1PC
634-0696
FLOW CELL, 30 MM, QUARTZ, BEAM HEIGHT:15MM
1PC
634-0697
PELTIER UNIT, A-1101
1PC
634-1140
SIPPER UNIT WITHOUT TEMP. CONTROL, A-1001
1PC
634-1138
SIPPER UNIT WITH PELTIER TEMP. CONTROL, A-1201
1PC
634-1139
PRINTER, THERMAL PRINTER, RD-TH32-SC
1PC
634-1137
THERMAL PAPER
1PC
634-6043
LAMP, HALOGEN, 12V/20W
1PC
634-1142
SAMPLE ROOM PANEL WITH PORT
1PC
634-1143
FUSE, 3.15A/250V
1PC
634-0651
SOFTWARE CD-ROOM
1PC
634-1141
Technical service
Web Resources
Visit the VWR website at www.vwr.com for:
• Complete technical service contact information
• Access to the VWR Online Catalogue, and information about accessories and related products
• Additional product information and special offers
Contact us For information or technical assistance contact your local VWR representative or visit.
www.vwr.com.
Warranty
VWR warrants that this product will be free from defects in material and workmanship for a period of two (2) years
from date of delivery. If a defect is present, VWR will, at its option and cost, repair, replace, or refund the purchase
price of this product to the customer, provided it is returned during the warranty period. This warranty does not
apply if the product has been damaged by accident, abuse, misuse, or misapplication, or from ordinary wear and
tear. If the required maintenance and inspection services are not performed according to the manuals and any
local regulations, such warranty turns invalid, except to the extent, the defect of the product is not due to such non-
performance.
Items being returned must be insured by the customer against possible damage or loss. This warranty shall be
limited to the aforementioned remedies. IT IS EXPRESSLY AGREED THAT THIS WARRANTY WILL BE IN LIEU
OF ALL WARRANTIES OF FITNESS AND IN LIEU OF THE WARRANTY OF MERCHANTABILITY.
Compliance with local laws and regulations
The customer is responsible for applying for and obtaining the necessary regulatory approvals or other
authorisations necessary to run or use the Product in its local environment. VWR will not be held liable for any
related omission or for not obtaining the required approval or authorisation, unless any refusal is due to a defect of
the product.
This equipment is marked with the crossed out wheeled bin symbol to indicate that this equipment must not be
disposed of with unsorted waste.
Instead it's your responsibility to correctly dispose of your equipment at lifecycle -end by handling it over to an
authorized facility for separate collection and recycling. It's also your responsibility to decontaminate the equipment
in case of biological, chemical and/or radiological contamination, so as to protect from health hazards the persons
involved in the disposal and recycling of the equipment.
For more information about where you can drop off your waste of equipment, please contact your local dealer from
whom you originally purchased this equipment.
By doing so, you will help to conserve natural and environmental resources and you will ensure that your
equipment is recycled in a manner that protects human health.
Thank you
We’ll introduce the main functions of VWR Software Multi Wave Pro in this chapter.
Quantitative Analysis
Provide 2 methods to establish a Standard Curve;
Up to 20 standard samples to establish a new curve or input the coefficients to make a standard curve;
There are three methods, which are Linear zero , Linear and Square to fit the standard curve.
Kinetics Analysis
We can choose the Time Intervals(0.5, 1.0, 2.0, 5.0, 10.0, 30.0 or 60.0 Seconds);
You can choose different Photometric Mode to display the curve (Transmittance-Time & Absorbance-Time).
Spectrum Scan
You can choose the Scan Intervals(0.1, 0.2, 0.5, 1.0, 2.0 and 5.0nm);
You can choose the Photometric Mode to display the spectrum (Wavelength-Transmittance, Wavelength-
Absorbance);
Automatically list spectrum peaks;
Mathematics and smooth.
Multi-wavelength Analysis
Up to 20 wavelengths to measure a sample.
DNA/Protein Analysis
We provide 2 methods;
You can input the coefficients by yourself.
Energy Scan
You can switch the light source point(W Lamp, D2 Lamp or automatically switch);
You can choose the scanning intervals(0.1,0.2,0.5,1.0,2.0 and 5.0);
You can set the gain(1 to 8 times);
Automatically list spectrum peaks;
Installation
We will show you how to install VWR Software Multi Wave Pro on your PC in this chapter.
Install and Uninstall VWR Software Multi Wave Pro
PC System Requirements
Pentium or above PC;
CD-ROM;
USB Ports.
512MB Memory(1GB or Above is strongly recommended);
50MB or above hard disc Space;
Microsoft Windows XP/Vista/7/8/10.
Install VWR Software Multi Wave Pro
1.
Put VWR Software Multi Wave Pro disc in the CD-ROM;
2.
Double Click to open the CD-ROM, and then, double click Setup.exe which is under the root directory of CD
to start installation , “Welcome” form appear, click Next;
Input user’s information, click Next;
4.
Choose install path, then click Next to copy files to Hard disc;
5.
Unplug the USB cable if connected before, click Next;
6.
After finished copy all files of VWR Software Multi Wave Pro, it will start install the USB Drive. Click Next;
Select “I accept this agreement”. Click Next to copy files to PC;
8.
Click Finish to finish install USB Drive;
9.
Plug the USB cable, click Next;
10. Click Finish to finish all the setup.
Uninstall VWR Software Multi Wave Pro
1.
Start→All Files→VWR Software→Uninstall VWR Software Multi Wave Pro 1.0, click Next to uninstall.
2.
After all the files were removed, click Finish to end.
Run VWR Software Multi Wave Pro
Double click the icon
on desktop.
Start→All Files→VWR Software→VWR Software Multi Wave Pro 1.0.
Set up Communication Port
Important information
Be sure that the USB line which is connecting PC with instrument
is working and the spectrophotometer is at the standby state interface.
After VWR Software Multi Wave Pro being started, then click Main Menu→Options, a dialog box of Options will
display in the screen, click Search to automatically search the instrument’s communication port, after get the
communication port, click OK to save setting. If you choose the option of Connect to Spectrophotometer When
Start, the instrument will connect to the PC automatically when you run VWR Software Multi Wave Pro next time.
Set User’s Information
Click Main Menu→Options, pop-up Options window, click User’s Information tag, choose it and input relative
user information, click OK to save settings, all this information will appear in your testing report.
Set Data Format
Click Main Menu→Options, a dialog box of Options will display in the screen, click Data Format tag, choose the
displaying format, and click OK to save setting.
Set Interface Style
Click Main Menu→Options, a dialog box of Options will display in the screen, and click Interface tag. You can
custom the style and color schemes of the interface as you like. Click OK to save setting.
Connect to PC
After enter the Main Interface, click
in the Tool Bar to connect with PC. After connection, the icon is on the
pitched on state, you can release PC if you click the icon again.
Introduction
We will introduce the VWR Software Multi Wave Pro in this chapter.
Main Interface
This is the main interface after start.
Menus Bar and Tools Bar
Menus Bar and Tools Bar provide three different ways to operate this software.
You can finish all the operation through Keyboard or Mouse.
Most of the functions can be finished through the Shortcut Bar.
Most of the functions can be finished through the menu by Right-hand button.
General Operating Instructions
Set Wavelength
Click the icon
on the Tools Bar to set wavelength. Input the value of wavelength in the frame of Set
Wavelength, click Goto to confirm. Then the instrument will move the wavelength to the point you want, then it
blanks automatically.
Put reference into the light path, click the icon
in the Tool Bar.
New a Testing
Main Menu→File→New, select a testing to click.
Lists of the file’s type
▪
Quantity Analysis File
*.qua
▪
Kinetics Analysis File
*.kin
▪
Spectrum Scan File
*.wls
▪
Multi-wavelength File
*.mti
▪
DNA/Protein Analysis File
*.dna
▪
Energy Scan File
*.ens
▪
Standard Curve
*.std
Test a Sample
Put sample into the light path, click the icon
in the Tool Bar.
Set Test Parameters
Click the button of
to set the parameters.
Modify a Sample
1. Choose the data frame you want to modify or set the curve as current curve;
2. Pull the unknown sample in light path;
3. Click the icon
in Tools Bar to re-measure it, and save the current result to substitute the original one.
Delete a Test Result
1. Choose the data frame you want to modify or set the curve as current curve;
2. Click the icon
in the Tools Bar to delete the data.
Name a Sample
1. Select the line you want to rename.
2. Double-click Sample Name, input the sample’s name, and press Enter on the keyboard to confirm.
Save a Testing
Click the icon
in Tools Bar, input the file name and click Save to confirm in the dialog box.
Open a Testing
Click the icon
in Tools Bar, choose the file name you need in the dialog box, then click Open to confirm.
Print Test Report
Click the icon
in Tools Bar, then set Print Parameters in the dialog box, press Print to confirm.
Export Data to Excel (the Microsoft Excel should be installed)
Click Main Menu→File→Export to Microsoft Excel, Excel will be started up, and data will be stored in it.
Export Curve as bmp Picture
Click Main Menu→File→Export Bitmap, input the file name and click save to confirm the dialogue box.
Switch Display Mode
Click the icon
or
in the Tools Bar to switch the display mode (Abs. or Transmit).
Zoom Selected Area
Click the icon
in Tools Bar, choose the area you want to magnify, then you’ll see the magnified one, the
Mouse’s cursor turns into the form of “+”. Click the icon
again to cancel the current status.
Modify Scale
Click the icon
on Tools Bar to display coordinate as you like.
Auto Fitting
Click the icon on Tools Bar to change the scale to appropriate value.
Select as Current Curve
Click Main Menu→Curve Operate→Choose the curve as you like.
Change the Color of Current Spectrum
Click Main Menu→Curve Operate→Choose the color as you like.
Search Peaks
Click the icon
in Tools Bar to search the peaks, they labeled automatically on the curve, and the peaks will be
listed in the Right Frame. You can set the Peak height by clicking
icon.
Mathematics Function
Click the icons
or
in the Tools Bar, a new dialog box will display on the screen respectively.
You can add, subtract, multiply or divide two spectrums.
We use this function to wipe off the interferer of noise during the course of scan, so as to make the spectrum look
much smoother. Click Main Menu→Spectrum→Spectrum Smooth.
Operation
We will introduce you how to use VWR Software Multi Wave Pro.
Quantitative Analysis
VWR Software Multi Wave Pro use Standard Curve to test the sample’s concentration at a fixed wavelength point.
1. Click the icon
in the Tools Bar shortcut to create a new Quantitative Analysis.
2. Click the icon
in the Tools Bar Shortcut to set the parameters of Quantitative Analysis.
Choose test method;
Input test wavelength in WL Frame;
Choose the Concentration Unit ;
Choose curve fitting method;
Create Standard Curve
Two methods are under your choice to create a new curve.
Method 1: Coefficient
(1) Click Coefficient option;
(2) Click Fit Method to choose Fit Curve method;
(3) Input the curve equation’s coefficients in corresponding frames;
(4) Click OK to finish setting.
Method 2: Use Standard Samples to create a new curve
(1) Click Standard Sampling option;
(2) Click Standard Sample Quantity to choose the quantity(≤20);
(3) Input the sample concentration in corresponding Frame;
(4) Click OK to finish setting;
(5) Pull the Reference into the light path and click the icon
, then the system will move the wavelength
to the one you just set, it blanks then;
(6) Put 1# standard sample into the light path and move the cursor on the fist frame of Abs. The sample
Absorbance value will appear in this frame after click
in the shortcut toolbar;
(7) Measure other standard samples in the same way as step (6) shows. The curve displays automatically
when finish;
You can modify the curve’s coordinates by Clicking the icon
. You can also save the curve by
clicking
, and you can reload it by clicking
next time.
3. Pull the unknown into the light path (if it is Automatic 8-cell Holder in the instruments, click
to set the
location of the relative cell first), Then click the icon, click
in the shortcut toolbar to test, the test result will
display in the data sheet.
We will introduce how to measure a sample’s change over a selected period of time.
1. Click the icon
to begin a new test of Kinetics;
2. Click the button of
to set the parameters;
3. Choose Photometric Display Mode, input the Wavelength, Display Time, Total Time, Intervals, Responsible
Mode and Coordinates Values;
4. Click OK to finish and exit setting;
5. Pull the reference into the light path and click icon
in the Tool Bar to do blank.
6. Pull the sample into the light path and click the button of to begin the test.You can click icon in the Tools Bar to
cancel at any time.
We’ll introduce how to create the spectrum while using Spectrum Scan function.
1. Click the icon
on the Tools Bar to start a new Spectrum scan;
2. Click the icon
in the Tool Bar to set the parameters of Wavelength Scan;
3. Choose the Photometric Mode, Wavelength Range for Scanning, Intervals, Repeat Times, Response Mode,
and Coordinate Range;
4. Click OK to finish and exit setting;
5. Put reference into the light path, click the icon
in the Tool Bar, then the dialog box of Baseline displays in
the screen, click Scan to scan baseline, click Cancel to interrupt scanning and exit the course;
6. Put the unknown sample in the light path, click the icon
in the Tools Bar to start scanning, click
to
interrupt and cancel scanning.
We’ll introduce how to measure a sample in different wavelengths (≤20).
1. Click the icon
on the Tools Bar, a new form appears on the screen;
2. Click the icon
on the Tools Bar to set the parameters;
3. Choose the quantity of the wavelength, you need input the value of each wavelength;
4. Click OK to finish and exit setting;
5. Pull the reference into the light path, click the icon
in the Tools Bar to do blank;
6. Pull the sample into the light path, click
to start test. And the test result will be listed in the data sheet.
We will introduce how to operate a DNA/Protein Analysis.
1. Click the icon
on the Tools Bar to begin a DNA/Protein Analysis;
2. Click the icon
to set parameters;
3. Choose the measure method first. We provide you 2 methods to measure, and you can also definite the
coefficients and wavelength by yourself;
4. Click OK to finish and exit the setting;
5. Pull the reference into the light path, click
in the Tools Bar to do blank;
6. Pull the sample in the light path, click the icon
to measure. The result will display in the data sheet.
We will introduce how to scan the variability of energy in the range of wavelength.
1. Click Main Menu→File→New→Energy Scan to build Energy Scanning;
2. Click the icon
on the Tools Bar to set parameter;
3. Select Lamp Switch Mode, Wavelength Range, Intervals, Repeat Times and Coordinate Range;
▪
Automatic Switch Mode---The instrument can choose D2 Lamp or W Lamp according to
setting;
▪
Deuterium Lamp Mode---Using Deuterium Lamp in the whole wavelength;
▪
Tungsten Lamp Mode---Using Tungsten Lamp in whole wavelength.
4. Click OK to finish and exit setting;
5. Pull the unknown sample in the light path, click the icon
in Tools Bar to start scanning, click
to
interrupt scanning and exit the course;
We will introduce the additional function about VWR Software Multi Wave Pro in this chapter.
Turn On/Off W Lamp
Click Main Menu→Device→Turn On/Off W Lamp to open or close W Lamp.
Turn On/Off D2 Lamp
Click Main Menu→Device→Turn On/Off D2 Lamp to open or close D2 Lamp.
Setup Switch Point of Light Source
Click Main Menu→Device→Setup Switch Point of Light Source, input the value of wavelength, and then click
OK to confirm.
Get Dark Current
Click Main Menu→Device→Get Dark Current, the system will reget the dark current and the new data will
replace the old one.
Scan System Baseline
Click Main Menu→Device→System Baseline, the dialogue box will display in the screen, you can click Open to
choose the storage base line to test, be sure that nothing in the light path, you can click Scan to build system base
line. You can click Cancel to interrupt and exit the course. You can choose Save System Base Line option; a
dialogue box will display in the screen and click Save to save system base line for your after use.
You needn’t to build system base line normally, only when it is a very long time that the
instrument wasn’t be used or the environment changed much.
Automatic 8-Position Changer (Accessories)
Click the icon
…
to pull the correspond cell in the light path.
Cell Calibration (Need install auto 8 cells Changer)
Click Main Menu→Device→Cell Calibration,this operation for matching difference of all the cells selected.
APPENDIX 1
Methods of Quantitative Analysis
Single Wavelength Method:
Abs.=A1
Double Wavelengths Method: Abs.=m*A1-n*A2
Three Wavelengths Method:
Abs.=A1-(WL1-WL2)*(A2-A3)/(WL2-WL3)-A3
APPENDIX 2
Methods of DNA/Protein Analysis
Method 1:
CDNA=(A1-Aref)*f1-(A2-Aref)*f2
CProtein= (A2-Aref)*f3-(A1-Aref)*f4
Ratio=(A1-Aref)/ (A2-Aref)
A1=A260nm,A2=A280nm, Aref=A320nm (Optional)
f1=62.9, f2=36.0, f3=1552, f4=757.3
Method 2:
CDNA=(A1-Aref)*f1-(A2-Aref)*f2
CProtein= (A2-Aref)*f3-(A1-Aref)*f4
Ratio=(A1-Aref)/ (A2-Aref)
A1=A260nm, A2=A230nm, Aref=A320nm (Optional)
f1=49.1, f2=3.48, f3=183, f4=75.8