Sequoia LISST-200X Operation & User’s Manual

Text of the manual

Particle Size Analyzer

Including LISST-HAB & LISST-Black User’s Manual Version 2.2 December, 2021 2700 Richards Road, Suite 107 Bellevue, WA 98005 USA Telephone: (+1) 855-753-3313 Store Software USB Card Here

LISST-200X User’s Manual

Page v Welcome to the LISST-200X Particle Size Analyzer Using this manual I. LISST-200X Introduction & Quick Start provides an overview of the LISST-200X’s operating principles and basic functions. II. LISST-200X Operation Details provides details and step-by- step instructions for using and caring for the instrument. The Appendices contain specialized details, specifications, and information about options. Warranty See the product warranty at: http://sequoiasci.com/support/warranty Technical assistance For technical assistance please contact your local Distributor or Sequoia. Please include the instrument serial number with any correspondence. Sequoia Scientific, Inc. contact information: Telephone: (+1) 855-753-3313 Email: [email protected]

LISST-200X User’s Manual

Page vii Table of Contents I. LISST-200X INTRODUCTION & QUICK START ................................................................... 1 A. LASER DIFFRACTION AND THE LISST-200X ............................................................................ 1 B. LISST-200X GENERAL DESCRIPTION .................................................................................... 4 C. LISST-200X QUICK START TUTORIAL .................................................................................... 7 II. LISST-200X OPERATION DETAILS .................................................................................... 26 A. STEP BY STEP PROCEDURES ............................................................................................... 27 1. Installing LISST-200X Software .................................................................................... 27 2. Establishing Communication with the LISST-200X ...................................................... 28 3. Saving and Evaluating Clean Water Backgrounds ....................................................... 29 4. Configuring Data Collection .......................................................................................... 32 5. Using the Small External Battery Housing .................................................................... 36 6. Using the Large External Battery Housing ................................................................... 39 7. Offloading and Deleting Data Files from Internal Memory ........................................... 42 8. Processing a Single Raw Data File .............................................................................. 43 9. Batch Processing Multiple Raw Data Files ................................................................... 45 10. View Processed Data File ........................................................................................ 47 11. Data Quality Control ................................................................................................. 48 12. Real-Time Data Processing with LISST-200X Software .......................................... 50 13. Autonomous Real-time Data Processing ................................................................. 52 14. Collecting Data from an External Analog Sensor ..................................................... 55 15. Configuring the LISST-200X as a Sensor for a CTD ............................................... 56 B. START AND STOP CONDITIONS ............................................................................................. 57 C. INSTRUMENT COMMUNICATION ............................................................................................. 59 D. LISST-200X COMMAND SUMMARY ...................................................................................... 60 E. LISST-200X COMMAND DETAILS ......................................................................................... 61 1. Display Commands ....................................................................................................... 61 2. Setup Commands ......................................................................................................... 64 3. Acquisition/Action commands ....................................................................................... 69 F. PERFORMANCE OPTIMIZATION ............................................................................................. 71 G. INSTRUMENT MOUNTING, DEPLOYMENT ORIENTATION, STORAGE AND MAINTENANCE ............. 74 APPENDIX A: TECHNICAL SPECIFICATIONS ........................................................................... 76 APPENDIX B: PARTICLE SIZE BINS .......................................................................................... 78 APPENDIX C: DATA FILE FORMATS ......................................................................................... 79 APPENDIX D: MATLAB DATA PROCESSING ............................................................................ 82 APPENDIX E: CONNECTORS ..................................................................................................... 84 APPENDIX F: PARTICLE SHAPE MODELS ............................................................................... 89 APPENDIX G: LISST-BLACK & LISST-HAB ............................................................................... 90 APPENDIX H: BIOBLOCK ANTI-FOULING WIPER .................................................................... 93 APPENDIX I: PATH REDUCTION MODULE ................................................................................ 96 APPENDIX J: TROUBLESHOOTING & SUPPORT ................................................................... 103 REVISION HISTORY ................................................................................................................... 104

LISST-200X User’s Manual

1 LISST-200X Introduction & Quick Start I. LISST-200X Introduction & Quick Start A. Laser Diffraction and the LISST-200X Measurement Principle and Optics Particle sizing by laser diffraction is currently the most widely prevalent method in research and industry. It is a multi-parameter measurement. Just as a set of 3 equations can be solved for 3 unknowns, a measurement of light scattering at multiple angles can be solved for equally as many concentrations in different sizes. For example, the LISST-200X measures scattering into 36 angles; consequently, one obtains concentrations in 36 size classes of particles. This is called the particle size distribution (PSD). To measure scattering at multiple angles, the LISST-200X sends a collimated beam of light into the water. Laser light scattering at an angle from the beam arrives, through a daylight rejection filter, at the focal plane of a receive lens at the same angle from lens axis (see figure). Distance from lens axis in the lens focal plane corresponds to scattering angle. The photodetectors in the LISST series instruments are a series of silicon rings spanning 60-degree arcs. Each ring covers a small range of scattering angles. Rings are used to achieve stability of inversion – converting of set of multi-angle scattering measurements to the PSD. Notice that the laser beam itself is focused by the receive lens and passes through a small aperture in the ring detector, centered on the rings. This beam is sensed by a photodiode placed behind the ring detector. This is the transmission sensor. As light is removed from the laser beam by scattering, the beam is attenuated, i.e. the light transmitted through water is reduced in intensity. A similar attenuation also affects the light that is scattered by particles and sensed by ring detectors. Thus, the transmission sensor provides a vital measurement to de-attenuate the measured scattered light.

LISST-200X Introduction & Quick Start

2 LISST-200X User’s Manual Clean Water Backgrounds Scattering originates not only from particles in the water, but also from the water itself, and from the windows and internal optics of the LISST- 200X. This non-particle scattering constitutes a Background. (Formerly this was called zscat, meaning scattering by zero concentration of scatterers). Accurate measurements require subtracting the background scattering from the total signal seen by ring detectors. The background scattering typically changes gradually throughout the life of the instrument, due to small changes in the laser and optics, and wear on the exposed window surfaces. It is also very sensitive to cleanliness of the windows. Therefore, high-quality measurements require careful background measurements by the user. A factory background acts as a reference for comparison. The LISST-200X stores its factory background data in its memory, as well as the user’s acquired background data. Background Quality Control Because the background measurement is so important, the LISST- 200X firmware and software include procedures for measuring and evaluating the background scattering. The software checks for signs of problems such as laser degradation, optics misalignment, contamination of data by thermal microstructure in the water used for background, bubbles or particles in the supposedly clean water, scratches on windows etc. For details on the background measurement procedure, see Saving and Evaluating Clean Water Backgrounds on page 29. Ambient Light Rejection The laser diffraction method requires that the light arriving on the detector be due entirely to scattering of light that originates with the laser beam. Light from ambient sunlight can distort the results. The LISST physically shields the detector from much ambient light, but not all, so daytime measurements at shallow depths can include some sunlight. However, the LISST-200X (starting with firmware version 1.4, introduced in May, 2017) can measure the ambient light independently of scattered light, and remove its effects. This Ambient Light Rejection (ALR) function is enabled by default. Also see section 0, page 71. Quick Estimates of Total Concentration and Mean Size Converting the LISST-200X’s measurements of scattering to PSD requires extensive computations that are time-consuming for a small microprocessor. However, a much simpler computation, easily done in real time, provides estimates of the total particle volume concentration and mean particle size. These quantities are not as accurate as the fully-processed PSD, but are useful for quick characterization of water conditions. For example, if used in conjunction with a CTD package in profiling applications, the CTD software could display the concentration and mean size as a depth profile, revealing any vertical structure in the particle distribution. (Note that real-time interface to a CTD requires proper configuration of the LISST-200X’s auxiliary connector; see Configuring the LISST-200X as a Sensor for a CTD on page 56.)

LISST-200X User’s Manual

3 LISST-200X Introduction & Quick Start The quick estimates are based on weighted sum of the net scattered light. One set of weight factors yields the total volume concentration; the other yields the area concentration of particles. The volume/area ratio provides mean diameter, also known as the Sauter Mean Diameter (SMD). Notably, SMD can be quite different from D50 in broad or multi-modal size distribution situations. The SMD output is set to zero in very clear waters, i.e. when transmission is above 98%, because the SMD can become erratic when the relative scattering signal is small. [For an explanation of this method of SMD derivation, see: Shaped Focal Plane Detectors for Particle Concentration and Mean Size Observations; Agrawal, Y.C. and O.A. Mikkelsen, (2009), Optics Express, v 17, n 25, pp 23066-23077]. Calculating the full PSD in real time Newer LISST-200X’s, starting with serial number 2131 (and older units with upgraded circuitry), contain more powerful processors that can do full PSD processing in real time. The processed data can be stored in the instrument’s memory, or transmitted through its serial port. This is especially useful for unattended deployments, such as buoys or remote moorings, where it is not possible to connect a Windows computer to the instrument for data offloads. For full details, see Autonomous Real- time Data Processing on page 52. Particle Shape Models – Spheres or Irregular Shape The multi-angle scattering can be interpreted via inversion as arising from spherical particles, or from irregularly shaped particles. Provided software gives you the choice and the resulting PSD files are named differently to distinguish them. As to which particle model to use, we suggest that when working with natural waters, use the irregularly shaped model. The spherical model is appropriate only in exceptional circumstances. For more details see Appendix F: Particle Shape Models on page 89.

LISST-200X Introduction & Quick Start

4 LISST-200X User’s Manual B. LISST-200X General Description Instrument Description The primary measurements delivered by the LISST-200X are the small- angle scattering properties of particles in water, the laser optical transmission, depth, and temperature. The LISST-200X is capable of autonomous operation when used with external batteries. Windows software is provided to program the instrument for a specific sampling schedule. An auxiliary connector can be configured for recording data from an external devices such as fluorometer, or for producing summary data (mean size and total concentration) as analog voltages. For details of the possible configurations of the Auxiliary Port, see Appendix E: Connectors on page 84. After recovery of the instrument, small-angle scattering data are off- loaded from the instrument and subsequently inverted mathematically on a computer to produce the particle size distribution. The inversion function is included in the Windows software. For MATLAB users, inversion scripts are available for download from Sequoia’s website: www.SequoiaSci.com. The LISST-200X instrument is a laser diffraction device. It consists of optics for producing a collimated laser beam, a specially constructed detector array, electronics for signal pre-amplification and processing, data storage and scheduling computer, and an external battery system. The principal measurement—angular scattering distribution— is obtained over 36 ring-detectors whose radii increase logarithmically from 102 to 20,000 microns. The detector is placed in the focal plane of the receiving lens. The rings cover an angular range from 0.00085 to 0.34 radians. This angular range corresponds to size ranges from 1.00 to 500 µm. See Appendix B: Particle Size Bins on page 78 for more information. The LISST-200X consists of the following parts: a solid-state diode laser operating at 670 nm wavelength and fiber-optically connected to a laser beam collimating system, a beam manipulation and orienting system, a two pressure windows, a scattered-light receiving lens, the custom designed 36-ring detector, preamplifier electronics, a ring- selecting multiplexer circuitry, and a digital controller with data memory. All these components are inside the black-anodized pressure housing. The LISST-200X does not have internal batteries. Its watertight housing need not be opened for any normal use.

LISST-200X User’s Manual

5 LISST-200X Introduction & Quick Start Data Storage and Interface The LISST-200X includes high-capacity data logging and storage. The logging functions are programmed via the provided software. It can be programmed with different start and stop conditions as well as different sampling rates and average durations. The data logger stores the data in non-volatile Compact Flash memory which can be later downloaded and processed into size distributions and concentration using the provided software. The data logger will also accept commands via the RS232 interfaces. These commands can be used to program the instrument or to exchange data with another instrument. Depth and Temperature In addition to measuring the particle size and concentration, the LISST- 200X also has depth and temperature sensors. The depth sensor has a 1000 psi full-scale range and is calibrated to 600 meters depth. A stainless-steel fitting and tube filled with silicone oil protect the pressure transducer from contamination and corrosion. The temperature is measured using a high precision thermistor imbedded into a stainless steel probe on the Connector end cap. Both values are stored automatically in the LISST-200X data file. External I/O Port The LISST-200X also has an auxiliary 6-pin connector for interfacing to other devices This port can be configured to produce analog outputs, or to receive analog inputs. The Analog Output Configuration allows the LISST-200X to receive power from a CTD, and to send two analog voltages indicating the mean size and total concentration to the CTD, while the detailed size distribution data is stored internally for later downloading. For more details about this application, see Configuring the LISST-200X as a Sensor for a CTD on 56. The Analog Input Configuration enables operation with additional sensors such as fluorometers and turbidity sensors. It provides 12V regulated power, and accepts two or three analog inputs (0 to 5 V) from sensors. The digitized analog values are stored in the same files as the LISST-200X’s particle size data. This configuration also supports the instrument packages described in Appendix G: LISST-Black & LISST- HAB on page 90. Battery-powered operation The LISST-200X does not contain internal batteries (except a small battery to maintain the real-time clock), but can be powered from Small or Large Battery Housings (provided with the instrument) or an external instrument such as a CTD. The standard Small Battery Housing supplied with the LISST-200X uses two NiMH rechargeable D-cell batteries to provide about 12 hours of continuous sampling. The Large Battery Housing holds 16 standard alkaline D-cells. For profiling applications, where the instrument is sampling continuously, the Large Battery Housing has about 200 hours of sampling time or 200 days of stand-by. Powering down the instrument between samples can

LISST-200X Introduction & Quick Start

6 LISST-200X User’s Manual greatly extend the deployment times. The LISST-200X software contains a battery life calculator on the Configuration page. This feature provides rough estimates of the battery run time for a given sampling program. USB Power and Data connection For laboratory or tethered usage, power can be supplied through the communications connector on the endcap. A 2-meter USB cable is provided that will power the instrument from the computers USB port. No additional power is required to operate the instrument. Cables up to 50 meters can be provided to supply external power and communication with the instrument. This can allow real-time observation of the size distributions. Included Accessories The instrument is shipped pre-aligned and tested. A small chamber is provided for obtaining measurements of background scattered light from optical surfaces. This background is subtracted from actual particle scattering measurements to obtain the true particulate scattering. Additionally, small tools used to open endcaps, spare batteries, and communication cable is supplied so that a user need only provide a PC running Windows. For extended laboratory applications various chambers are available from Sequoia. Software is provided to communicate with the instrument, schedule an experiment, offload the data, and invert the measurements to obtain particle size distribution and volume concentration. For laboratory use or for monitoring the progress of an experiment, the software can be used for real-time processing.

LISST-200X User’s Manual

7 LISST-200X Introduction & Quick Start C. LISST-200X Quick Start Tutorial This section gives step by step instructions to unpack your LISST- 200X, load software, and acquire data in the lab. For more detailed instructions on specific steps, background on how the instrument works, or specific technical information such as cable pinouts please refer to the full User’s manual following this Quick Start Tutorial. Contents of Shipping Case Let’s assume that you are opening the LISST-200X shipping case for the first time. Inside you will find the following: • User’s Manual, • USB memory card (credit card size) with the software, • LISST-200X instrument, • Plastic Instrument stands, • Small Volume Test Chamber, • USB Communications cable, • Insulated stainless steel clamps, • Small Battery Housing (without batteries), • Larger Battery Housing with alkaline batteries installed, • LISST-200X-to-Battery-Housing cable, • NiMH batteries and charger. Step 1: Remove Instrument from Shipping Case. Start by removing the white plastic instrument stands and set them on a flat working surface. Remove the LISST-200X from the case and set it on the stands. The LISST-200X has two distinct ends that we will refer to as the Optics endcap and the Connector endcap.

LISST-200X Introduction & Quick Start

8 LISST-200X User’s Manual Optics Endcap The optics endcap contains the optical windows that the laser beam passes through to make a measurement. The internal optics and laser electronics are mounted to the inside of this endcap. Connector Endcap The connector endcap has three underwater connectors that are used for communication, external power, and connecting to optional accessories or instruments: 3-pin connector: Used for BioBlock, an anti-biofouling accessory 5-pin connector: Used for serial communication and external power. 6-pin connector: Used for analog in/out and digital in/out signals. See Appendix E: Connectors on page 84 for a full description of the wiring of the underwater connectors and mating cables. A temperature sensor is located between the 3-pin and 6-pin connectors. In the center of the endcap is an LED that blinks when the instrument is sampling. The stainless steel fitting with small tube is the port for the depth sensor. This fitting can be removed to allow the port to be cleaned should sediment or salt deposit build up.

LISST-200X User’s Manual

9 LISST-200X Introduction & Quick Start Also on the Connector endcap is a white plastic lever. This mechanical lever has a strong magnet embedded in the plastic. This magnet can be used to trigger a digital switch inside the instrument which can be programmed to start and stop the LISST-200X sampling. The final item on the endcap is the zinc anode. The sacrificial zinc anode protects the instrument from corrosion during long periods of time in salt water. Step 2: Check for Clean Windows At this time, check the optical windows to make sure that they are clean. There are two windows: The receive window (or large window) and the transmit window (or small window). Both window need to be very clean in order to get good measurements. The best way to check the windows is by using a flashlight. By shining light from one side and viewing from the other the surface of the windows can be easily checked for cleanliness. If there is dirt or fingerprints on the windows clean them first by rinsing them with lukewarm water and a mild soap solution (e.g. mild hand soap, liquid dish soap) and then rinsing off all soap residue with clean, particle free water such as deionized water, distilled water or bottled drinking water. The windows can also be wiped clean with a soft cloth (e.g. a lens cloth) or glass cleaner. It is not recommended to use stronger solvents, such as acetone or toluene. Also, do not use any abrasive cleaners or wipes. Treat the windows as you would an expensive camera lens. Step 3: Attach Communication and Power Cable Remove the Communications cable from the plastic accessory case within the shipping case. It is the 3 meter cable with the USB connector on one end and the 5-pin underwater connector on the other. Remove the underwater cap from the Communications connector. The connectors will all look similar when the protective cap is installed. The Communication connector is the only 5-pin connector and is located next to the stainless steel pressure sensor fitting. After removing the cap install the cable making sure that proper alignment of the cable is maintained, so that the connector pins are not bent. Plug the USB cable into the computer. Please note that USB drivers may automatically install the first time the USB cable is plugged into

LISST-200X Introduction & Quick Start

10 LISST-200X User’s Manual the computer. If the driver installed correctly, you should see the green light on the USB cable blinking. For more information on establishing and troubleshooting communication with the LISST-200X see page 27. Step 4: Install the Horizontal test chamber Remove the Small Volume Horizontal test chamber from the Accessory Case if it is not already installed on to the instrument. This one-piece chamber is designed to slip between the optical windows of the instrument such that the space between the windows can be filled with water for testing or calibration. We can now fill the chamber with clean, particle-free water. Sequoia uses steam-distilled bottled water filtered through a 0.2 micron filter. Tap water may contain too many particles and may also contain dissolved gas that can release and form small bubbles on the optical surfaces. Pressurized filtered water can also contain dissolved gas. Often the best source of clean water is bottled drinking water. Step 5: Install LISST-200X Software At this point the instrument is ready to go. We now need to install the software that is required for operation of the instrument. A USB memory card the size of a credit card is included with each instrument. In addition to the communication and processing program the disk also contains digital copies of this manual and other support files. Insert the memory card into a USB port on your PC to install the software. You must install the software on a computer running Windows XP or later (it is not compatible with Mac or Linux operating systems). On the memory card you will find the ‘LISST-200X_Installer.exe.’ Double click the installer executable to begin installing the software. Follow the onscreen instructions and the installer will transfer the necessary files to your computer and place a shortcut on your desktop and start menu. Do not remove the memory card from your computer until the installation is fully completed.

LISST-200X User’s Manual

11 LISST-200X Introduction & Quick Start Please note the following with respect to the LISST-200X software: • On a 64-bit Windows PC, the software will be installed in the following folder: C:\Program Files (x86)\Sequoia\LISST200X, while on 32-bit Windows machines, the install folder will be C:\Program Files\Sequoia\LISST200X • The LISST-200X software WILL NOT work with the older LISST-100 or LISST-100X instruments. It is NOT possible to use the program to offload data from a LISST-100 or LISST- 100X instrument, only a LISST-200X will work with this software. Step 6: Start LISST-200X Application Start the LISST-200X software by selecting the shortcut the installer placed on your desktop. Step 7: Establish Communication with the LISST- 200X The LISST-200X does not have an internal battery. Power is supplied to the instrument via a battery housing or USB cable. Please note that the LISST-200X USB cable has a USB-to-serial convertor and special power supply that converts the +5V to +11V to run the instrument built into cable. When using this cable no other power source is required. Power can also be provided by the Small and Large External Battery Housings or another external power source such as a CTD. When using an external battery the serial communication passes through the battery housing. However, the power from the USB is not used power the instrument. Power will be drawn from the battery only. The LISST-200X software will automatically detect and connect to a LISST-200X instrument if it is connected to the computer. If an instrument is found, it will display ‘Connected’ in the lower left corner.

LISST-200X Introduction & Quick Start

12 LISST-200X User’s Manual If your instrument is connected to the computer, but not found by the software, click on the ‘No Instrument Connected’ message to see list of options. Click on the ‘Communication Help’ button for troubleshooting help. Step 8: Open Terminal window The LISST-200X software has a terminal window that allows the user to communicate directly with the instrument using a set of commands. To open the Terminal window click on the connected instrument message in the lower left, then click on the wrench icon and select ‘Open Terminal’ A command can be typed into the box in the lower section of the window. When the Enter key is pressed the command will be sent to the instrument. A list of the commands that can be sent to the LISST- 200X is displayed when the terminal opens or they can also be found in the section LISST-200X Command Summary on page 62.

LISST-200X User’s Manual

13 LISST-200X Introduction & Quick Start Step 9: Acquiring Clean Water Background Measurement The background scattering measurement is critical to good instrument performance. It is especially critical for clean water applications where the optical transmission is greater than 90% over the 2.5cm path. The background scattering will also check the overall health of the instrument. It will verify that all of the systems are functioning and that the optics is still in alignment. The current background will be acquired and displayed relative to the factory background scattering and the most recently saved background that is stored on the instrument. The image below shows an example of this display. It is opened by selecting the Collect/View Background button. The factory background file will be automatically acquired from the instrument and displayed as a red line on the screen. The ‘Current Background’ is the last background that was collected and saved on the instrument (the date of collection is displayed in the legend). Tips for collecting a background are also displayed on the screen. Pressing the ‘Background Help’ button will open a PDF with additional information about background measurements. When the Collect New Background button is pressed, 20 samples will be collecting. The average of these measurements will be displayed as a black line on the screen. The graph shows the value of the 36 light scattering detectors. The red line is the factory values, the black line is the average of all 20 measurements. If the background is close to factory levels the message displayed will indicate a pass or acceptable background. If the water or windows are not clean or if there is a problem with the instrument, error messages and suggested actions will be displayed.

LISST-200X Introduction & Quick Start

14 LISST-200X User’s Manual Dirty water or windows will generally cause higher values across the middle rings. Large bubbles or particles in the water can cause higher values on the inner rings or left hand side of the display. High values on the inner rings combined with a lower transmitted laser power value can also be an indication of optical misalignment. If needed you can update the background, such as after cleaning the windows or replacing the water, by pressing the Collect New Background button again. In general, the lower the background values the better the background. The goal is to get values that are at the same values as factory line. However, as the instrument is used the background may increase due to small scratches and slight alignment changes. It may not be possible to get the background down to the original factory values. If the values are acceptable the values can be saved both onboard the instrument and to a file on your computer. The LISST-200X handles the background files differently than the LISST-100X. The background file is stored on the instrument and is saved as part of every data file that is recorded. When processing a data file the background will be automatically extracted and used during processing. Saving, tracking and selecting a background file to use for processing is no longer required. The background stored on the instrument will continue to be saved in new data files until a new background is recorded. If you wish to store a background file on your computer as well as on the instrument, click the ‘Save as .BGT file’ after you have accepted and saved the background. It is not necessary to store the background on your computer, however, these saved backgrounds can be used during processing instead of the background saved in the data file. Step 10: Configuring Instrument for Deployment The Configure Instrument window is used to configure the deployment parameters. To open the window, choose Configure Instrument from the home page. A window similar to the one shown below will appear. The window has four accordion style tabs: Sampling Schedule, Start Condition, Stop Condition, and Accessories.

LISST-200X User’s Manual

15 LISST-200X Introduction & Quick Start With the accordion tabs collapsed, the page shows a summary of the instrument’s current configuration. Step 11: Setting Operating Mode By selecting the Sampling Schedule Tab at the top of the main window the screen below appears. This screen is used to set the type of sampling; Fixed Sample Rate or Burst. You can also select the samples per average and sample rates on this screen. Burst and Fixed Rate Modes The Burst and Fixed Sample Rate modes are used to save data to a raw data file on board the instrument. The icons next to the various values give a better understanding of their meaning. The software automatically checks the values entered to make sure that there is no conflict. For example, when a ‘Sample to be Average of’ value is entered, the minimum sample interval is computed. If this value is

LISST-200X Introduction & Quick Start

16 LISST-200X User’s Manual less than the minimum permitted the value will be changed to the minimum and the text will turn red. The LISST-200X measures internally at ~20 Hz, but the data cannot be stored to the data logger at this rate. The individual measurements are averaged into a sample, and it is this sample average that is stored at a maximum sample rate of 1 Hz. For the example shown above, the instrument is set to sample in the Fixed Sample Rate mode at a 1 Hz rate with 10 measurements per average. This average is obtained in 0.53 seconds. Step 12: Setting Start Conditions After selecting the sampling schedule, the start and stop conditions can be selected. Choose the Start Condition tab. There are five options: Depth, Time, External Mechanical Switch, External Digital Input, and Time Delay. Select the mode by clicking on the button next to its label. Select the correct parameters as required. For this example let’s select the External Mechanical Switch Start Condition. AutoStart If the AutoStart check box is selected, the LISST-200X will start the sampling program when power is applied to the instrument. NOTE: for AutoStart to detect the application of power, all power sources (including the USB cable) must first be disconnected for at least 6 seconds. If AutoStart is not selected the user must select the ‘Run Sample Program’ button from the instrument status bar or send the ‘GO’ command to start the sampling programing after applying power. Step 13: Setting Stop Conditions Similarly, the Stop conditions can also be selected. Click on the Stop Condition Tab to open the Stop Condition window. The available stop conditions are: Depth, Time. External Mechanical Switch, External Digital Input, Fixed number of Samples, and Memory Full/Battery Low

LISST-200X User’s Manual

17 LISST-200X Introduction & Quick Start For this example, choose the External Mechanical Switch as the Stop condition. Step 14: Configure Accessories The accessories tab allows you to select which LISST-200X accessories are connected to the instrument. The accessories must be selected here for them to function properly. If there are no accessories, select ‘None’. Step 15: Upload Configuration to Instrument After configuring the instrument settings, they must be uploaded to the instrument. A message will be displayed on the top of the screen when you have unsaved changes to the instrument configuration

LISST-200X Introduction & Quick Start

18 LISST-200X User’s Manual Click on the ‘Upload to Instrument’ button to load the configuration onto the LISST-200X. A confirmation message will be displayed when the upload is finished. Step 16: Start Instrument The instrument is now configured for deployment, however, it is not yet running. You MUST explicitly start the samping program you just configured. If you do not do that, the LISST will never start sampling or react to the start and stop conditions. To start the sampling program. Click on the ‘Run Sample Program…’ button in the instrument status bar. You will be asked to confirm your sampling program, start/stop conditions, accessories, and background date.

LISST-200X User’s Manual

19 LISST-200X Introduction & Quick Start After the sampling program is started, you will see the status bar change to ‘Sample Program Running’ as shown below. At this point the instrument is running and ready for deployment. Note that even when the sampling program is running, sampling will not necessarily start immediately. For example, if you have selected the mechanical switch as the start condition, the instrument will only react to the swtich after the sampling program has been started. Only then will actual sampling begin. If you want sampling to begin immediately, select a start condition of “Delay” with zero as the delay time. While waiting for the start condition the LED on the connector endcap will double blink every few seconds to alert the user that the instrument is running but not yet sampling. During sampling the LED will illuminate during the actual measurement of an averaged sample. For example, if the sampling is set for 1 Hz sampling with measurements per average set to 20 then the LED will blink once per second. If settings are set for a 30 second average every 15 minutes then the LED will illuminate for 30 seconds every 15 minutes. Step 17: Collecting Data For our example we can move the white lever on the endcap to the “1“ position. Sampling will start. Data collection will continue until the Stop Conditions are met. In our case, until the switch lever is returned to the “0” position. If the Start and Stop Conditions are set to Depth, External Mechanical Switch, or External Digital Input the program will return to checking for the Start Condition. This will only be true if the Start and Stop conditions match. For example, Depth Start and Depth Stop. For non- matching Start and Stop Conditions, such as fixed number of samples or Time Stop, the program will terminate and the instrument will go into a low power sleep mode. To stop a running program, or wake it from low power sleep mode, use the ‘Stop / Wake Instrument’ button from the status bar. Note that the above options will only be available if you started the sample program from the software. If you plug in an instrument that is already executing a sampling program, you will need to manually wake the instrument by selecting the ‘Wake Instrument…’ button and selecting the COM port the instrument is connected to.

LISST-200X Introduction & Quick Start

20 LISST-200X User’s Manual Step 18: Downloading Data The instrument has now stored data in the on-board memory card. Select ‘Offload Or Erase Files’. A list of files will appear: Choose the files to offload by clicking on them while holding down the CTRL key. The Shift key can also be used to select a range of files. Choose a location to offload the files to on the right side. The files will be saved with names in the following format: Ldddhhmm.RBN, where ddd is the day of the year, hh is the hour, and mm is the minute that the file was first written to. As the data is downloading a the status bars at the bottom will update. The data is offloaded at 115K baud. Step 19: Processing Raw Data We now have the data transferred from the instrument to the PC. To process the data file choose Process Data Files from the home page.

LISST-200X User’s Manual

21 LISST-200X Introduction & Quick Start You will be prompted to select the raw data file to open (selecting multiple files will open up the batch processing window). Raw data files have the extension .RBN, and are offloaded directly from the LISST- 200X data logger. Below is list of the file types you can expect to see when using the LISST-200X. Extension Discription Format .RBN Raw Data Binary .RTX Raw Data ASCII .CSV Processed Data ASCII .BGT Background File ASCII Every LISST-200X data file contains all the necessary information to process the file. Therefore, the software will automatically determine the instrument serial number, factory background, current background and other instrument specific parameters. A raw data file display will be generated for your selected data file. The range of samples to process can be selected by moving the sliders at the bottom of the display. This can be used to exclude data from the beginning or the end of the file. The default selection is to process the complete file.

LISST-200X Introduction & Quick Start

22 LISST-200X User’s Manual Select the output directory and choose a base file name. Press the Process File button to convert the raw file into processed size distributions. For more details on the available options when this window is open, see the detailed Instructions starting on page 43. When the processing is complete a new tab will open, showing the processed results. You can navigate back to the raw data tab if you’d like to change the processing settings and process the data file again. Step 20: Viewing Processed Results Processed data files are stored as ASCII files (.CSV). When a raw file is processed the resulting .CSV file is displayed automatically. You can also open any LISST-200X .CSV file by selecting the View Process File button from the home page. A display similar to the one below will open. A list of available parameters is displayed on the left. Clicking one of the parameters will open a new window where the parameter will be displayed. The parameter windows can be reorganized by dragging and dropping. The slider at the bottom is used to scroll through the data file. A red point on some graphs will indicate which sample is being displayed. Step 21: Opening a Real Time Session The LISST-200X software also supports the ability to acquire, process, and display size data it in real time. The Start and Stop conditions will not be used and therefore their settings are not relevant. Before opening a real time session, you should collect a new background and store it onboard the instrument. The real time session

LISST-200X User’s Manual

23 LISST-200X Introduction & Quick Start will use the last background stored on the instrument to process the data in real time. Therefore, you must ensure the background on the instrument is up to date before continuing. To open the Real-Time session, choose Collect Live Data from the home page. Choose a location to save the data by filling out the output directory and the base file name. The sample interval can also be adjusted by entering the number of seconds between sample in the sample interval text box. Even though data is displayed to the screen, it is not saved until you select the Start Logging Button. The Spherical / Irregular Shape radio buttons can be selected in order to display the results as being processed under the assumption that the particles are spheres or randomly shaped (natural grains) particles. Data using both models is saved and the buttons only represent what is displayed on the screen. A list of available parameters is displayed on the left. Clicking one of the parameters will open a new window where the parameter will be displayed. The parameter windows can be reorganized by dragging and dropping.

LISST-200X Introduction & Quick Start

24 LISST-200X User’s Manual Deploying the LISST-200X in the Field The above steps should have given you a good understanding on the how to operate the LISST-200X. More details on all of the steps can be found in the User’s Manual following this tutorial. The following steps go into deploying the instrument in the field. Step 22: Preparing for Transport The LISST-200X has been designed to be a robust field instrument. However, it is still a highly sensitive optical instrument and needs to be treated with care. This is especially true during shipping. The LISST- 200X is shipped in a custom case with specially designed foam cushioning. Anytime the instrument is transported it should be in this case. We highly recommend using air shipping when possible to eliminate the extended vibrations that ground shipping can cause. Because the LISST-200X does not have an internal battery there is no need to worry about putting the instrument into low power sleep mode. You can simply disconnect it from the battery housing. The Large External Battery Housing, which has alkaline batteries, can be stored for extended period when fresh batteries are installed. The Small External Battery Housing, which has NiMH rechargeable batteries, should be opened and the batteries removed during transport. There are not restrictions on shipping NiMH batteries. However, the batteries will have a small drain when installed in the Small Battery housing. Therefore it is recommended that they be removed until the battery is ready to be used. The batteries should also be stored in a fully charged state. The LISST-200X can be pre-programmed with the desired sampling program such as start and stop conditions and sampling rates. These settings will remain in non-volatile memory until power is applied. Therefore it is possible to prepare the instrument for sampling before shipping it out. It is even possible to program the instrument to automatically start upon power up. This is described in the next step. Step 23: Configuring AutoStart The LISST-200X can be configured to start running the sampling program upon power up. In some situations this ability can be quite helpful as you do not need to connect the instrument to the PC to start the sampling. The instrument can be preconfigured with the desired settings and the wait for the battery to be plugged in or external power from the CTD to be present before looking for the Start Condition. To configure the LISST-200X to start its sampling program upon power up use the LISST-200X software or the configure auto start from the terminal window. If using the terminal window, use the AS command (short for Auto Start). The AS command can be issued when the 200X> prompt is displayed in the Terminal Window. When AutoStart is enabled a message will be displayed as part of the Status command or the Query instrument results.

LISST-200X User’s Manual

25 LISST-200X Introduction & Quick Start When the LISST-200X is powered up a message will be displayed to the Terminal window (or via the RS232 interface) that will prompt the user that the instrument will auto start in 5 seconds. Pressing the stop key or sending two CTRL-C characters will stop the instrument from starting and return to the 200X> prompt. Step 24: Connecting to External Sensors or Instruments The LISST-200X had the ability to connect to external sensors and instruments such as other dataloggers or CTDs through the 6-pin Auxiliary connector on the endcap. The details of using the Auxiliary connector can be found in the Step-by-Step instructions on page 52. In short, the Auxiliary connector can be configured in one of four ways: 1) Analog and Digital inputs, 2) Dual Analog Inputs, 3) Dual Analog Outputs, or 4) Triple analog input. When using the Analog or Digital inputs it is possible to provide power to an external sensor. By default this power output is switched off. It will need to be enabled before power will be available. You can also configure the warmup time required for the sensor after power is applied before sampling will begin. The Analog Output configuration sends out two analog voltages, Mean Size and Total Concentration. This summary information is updated each time data for the full size distributions is recorded. When using the Analog Output configuration power can be received through the Auxiliary connector. The wiring of the connector is designed to match the SeaBird CTD Auxiliary Input connector simplifying the connection.

LISST-200X Operation Details

26 LISST-200X User’s Manual II. LISST-200X Operation Details Section Organization This section contains detailed instructions for performing various procedures. These are either in the form of step-by-step instructions or detailed descriptions of the various aspects of instrument operation (e.g. command list, instrument mounting and deployment). General Precautions • LISST-200X is a sensitive optical instrument – please handle it gently as you would handle a very expensive camera. • Critical alignments may be disturbed if the instrument is subjected to shock or rough handling. • Evidence of shock/rough handling will void the warranty. • Whenever in transit, store the instrument in the provided padded shipping case. • If placing the instrument vertically on the standoffs, be sure to do so gently as the Compact Flash Memory Card inside may otherwise come loose. WARNING-Class 3R laser – AVOID DIRECT EYE EXPOSURE The LISST-200X uses a laser diode emitting a maximum of 1 mW of visible (red) light at a wavelength of 670nm. Under normal circumstances the laser travels only within the 2.5 cm sample volume. However, if reflective objects are placed in the path of the laser beam, the light could be redirected. Avoid conditions that could direct the beam toward an eye.

LISST-200X User’s Manual

27 LISST-200X Operation Details A. Step by Step Procedures 1. Installing LISST-200X Software Software for the PC is used to configure the LISST-200X and for downloading and processing the size distributions. STEP ACTION RESULT 1 • The LISST-200X comes with a USB memory card. Plug the provided memory card into a USB port on your computer. Locate the ‘LISST-200X_Installer.exe’ executable on the memory card. Note that this software it is not compatible with Mac or Linux operating systems. Your operating system must be Window XP or newer to run the software. Installer was found on LISST-200X memory card 2 • Double click the ‘LISST-200X_Installer.exe.’ Follow the onscreen instructions and the installer will transfer the necessary files to your computer and place a shortcut on your desktop and start menu. Do not remove the memory card from your computer until the installation is fully completed: Software installation complete

LISST-200X Operation Details

28 LISST-200X User’s Manual 2. Establishing Communication with the LISST-200X Establish communication with LISST-200X using the supplied software and USB cable. Communicating with the LISST-200X via the software should be automatic. Should there be a problem you can use these step-by-step instructions to troubleshoot the problem. STEP ACTION RESULT 1 • If the LISST-200X software is open, close the program, then plug the USB communication cable from your computer. LISST-200X will be powered on 2 • After plugging the USB cable back into your computer, the drivers for the USB to serial converter should install automatically. You should see a notification on the task bar that drivers are being installed. If you are unsure if the drivers installed, you can install the drivers manually by running ‘CDM USB Drivers.exe’, located on the memory card that came with your instrument. Computer is now set up to communicate with the LISST-200X 3 • After the driver installation is compete, open the LISST-200X software. Software appears onscreen 4 • The software should automatically search and locate your LISST-200X instrument among the serial ports on your computer • If no instrument is found you will see “No Instrument Connected” displayed in the lower left corner. • Try selecting the ‘Retry Connection’ button or try to wake the instrument using the ‘Wake Instrument’ button. Software will attempt to automatically connect to the instrument 5 • If further troubleshooting is needed click on the ‘Communication Help’ button to explore the communication troubleshooting document. Open Troubleshooting Document

LISST-200X User’s Manual

29 LISST-200X Operation Details 3. Saving and Evaluating Clean Water Backgrounds In order to properly compute the size distribution, it is necessary to remove the light scattering from the internal optics and window surfaces so that only the light scattering from the particles of interest are used to compute the size distribution. STEP ACTION RESULT 1 • Connect the instrument to the computer and establish communication in the LISST-200X program (page 28) Software open, communicating with LISST-200X. 2 • Clean instrument and install Clean Water Background Test chamber. • Fill with clean filtered water and make sure no bubbles are in the water or on the windows. • The water and instrument should be at the same temperature. Using water a significantly different temperature from the water can impact the quality of the background. • Stir the water well before obtaining a background to make sure the water is well mixed. • For the best Background, make sure the instrument is clean; the water you use is filtered, and free of bubbles. Temperature fluctuations in water will seriously degrade the background. Use water at the same temperature as the instrument. Stir well. Optics submerged in water 3 • Select Collect/View Background from the software home page. The factory background file will be automatically acquired from the instrument and displayed on the screen. • The current signal from the 36 ring detectors will also be displayed in real time. • When the Collect New Background button is pressed 20 samples will be collected and the average displayed to the screen. Background collected and displayed on the screen

LISST-200X Operation Details

30 LISST-200X User’s Manual • The graph shows the value of the 36 light scattering detectors. The red line is the factory values, the blue line current background saved on the instrument. A newly collected background will be displayed as a solid black line. If the background is close to factory levels the message displayed will indicate a pass or acceptable background. Even if your new background passes, but is significantly higher than factory, clean the windows and repeat to verify you have acquired the best possible result. Acceptable Background • If the water or windows are not clean or if there is a problem with the instrument, error messages and suggested actions will be displayed. Dirty water or windows will generally cause higher values across the middle rings. Large bubbles or particles in the water can cause higher values on the inner rings or left hand side of the display. High values on the inner rings combined with a lower transmitted laser power value can also be an indication of optical misalignment. If needed you can update the background, such as after cleaning the windows or replacing the water, by pressing the Collect New Background button again. Unacceptable Background 4 • In general, the lower the background values the better the background. The goal is to get values that are at the same values as factory line. However, as the instrument is used the background may increase due to small scratches and slight alignment changes. It may not be possible to get the background down to the original factory values.

LISST-200X User’s Manual

31 LISST-200X Operation Details • Multiple cleanings or using better filtered water may be required to get the best possible background. • When you have an acceptable background press the Save button to store the background on the instrument. You also have the option to save the background to a file on your computer. After the background is saved, select the ‘Save as .BGT File’ button. The LISST-200X handles the background files differently than the LISST-100X. The background file is stored on the instrument and is saved as part of every data file that is recorded. When processing a data file the background will be automatically extracted and used during processing. Saving, tracking and selecting a background file to use for processing is no longer required. The background stored on the instrument will continued to be saved in new data files until a new background is recorded. Background Accepted and saved.

LISST-200X Operation Details

32 LISST-200X User’s Manual 4. Configuring Data Collection The LISST-200X software can be used to configure the Start and Stop Conditions, Fixed Rate or Burst sampling and sample rates. STEP ACTION RESULT 1 • Connect the instrument to the computer and establish communication in the LISST-200X program (page 28). Software open, communicating with LISST-200X. 2 • Open the Sampling Programs window by choosing Instrument Configuration from the home page. • Certain instruments can compute the PSD onboard in real time. If your instrument has this feature, an additional configuration ‘Data Output’ will appear on this page. (See Autonomous Real-time Data Processing on page 52 for more information). • To set the LISST-200X data logger clock to the computer clock, open the status bar, click the wrench icon, and select ‘Set Instrument Clock’. • You can also zero the depth sensor from this same menu. Instrument Status Displayed. Clock set 3 • By selecting the Sampling Schedule Tab at the top of the main window the screen below appears. This screen is used to set the type of sampling: Fixed Rate or Burst. You can also select the samples per average and sample rates on this screen. Operating Mode set

LISST-200X User’s Manual

33 LISST-200X Operation Details • The Burst and Fixed sample rate modes are used to set up a sampling program that will save data onboard the instrument. The icons next to the various values give a better understanding of their meaning. The software automatically checks the values entered to make sure that there is no conflict. For example, when a Samples per Average value is entered, the minimum sample interval is computed. If this value is less than the minimum permitted the value will be changed to the minimum and the text will turn red. A similar test is done on the Burst Interval setting. 4 • Select the Start Conditions tab to configure when the instrument will begin sampling. • There are five options: Depth, Time, External Mechanical Switch, External Digital Input, and Time Delay. Select the mode by clicking on the button next to the its label. Select the correct parameters as required. Start Condition Set • If the AutoStart check box is selected, the LISST-200X will start the sampling program when power is applied to the instrument. If this is not selected the user must send the

LISST-200X Operation Details

34 LISST-200X User’s Manual ‘GO’ command to start the sampling programing after applying power. 5 • Choose the Stop Conditions Tab to select the conditions when sampling should stop. • The available stop conditions are: Depth, Time, External Mechanical Switch, External Digital Input, Fixed number of samples, and Memory Full. • If the Stop Condition is Depth, External Mechanical Switch, or External Digital Input and the Start Condition is also one of these three options the Base Program on the instrument will return to waiting for the Start condition. For example, if the start and stop conditions were set to External Mechanical Switch then multiple sampling sessions can be obtained by flipping the switch on and off. This is useful when doing profiles. It eliminates the need to communicate with the instrument between profiles. Stop Condition Set 6 • Choose the Accessories tab to configure LISST-200X accessories. The accessories must be selected here in order for them to function properly. •

LISST-200X User’s Manual

35 LISST-200X Operation Details 7 After configuring the instrument settings, they must be uploaded onto the instrument. A message will be displayed on the top of the screen when you have unsaved changes to the instrument configuration Click on the ‘Upload to Instrument’ button to load the configuration onto the LISST-200X. A confirmation message will be displayed when the upload is finished. Upload Configuration 8 • The instrument is now configured for deployment, however, the sampling program is not yet running. • You MUST run the sampling program in order to collect data. But even when running, the sampling program will wait for the programmed start condition before collecting data. See Start and Stop Conditions on page 57, and the steps above, for details about start conditions. • When the sampling program is running, you will receive a confirmation message that the program has started, and the instrument will become disconnected from the computer. At this point the instrument is ready to be deployed. • The green LED on the connector endcap will indicate the instrument status: - If waiting for the start condition, every few seconds it will blink twice quickly. - If logging, it will light while collecting each sample. Run Sampling Program

LISST-200X Operation Details

36 LISST-200X User’s Manual 5. Using the Small External Battery Housing The LISST-200X comes with a Small External Battery Housing that can be used for short term deployments. The instructions in this section describe how to use the Small Battery including installing and charging the NiMH rechargeable batteries. NOTE: the Small Battery Housing cannot be used to power the LISST-HAB or LISST-Black (see page 90) or BioBlock (page 93). STEP ACTION RESULT 1 • The LISST-200X Small External Battery Housing uses two rechargeable NiMH D-cell batteries. Four of the 1.5V 10000mAh batteries and a charger are included with the housing. The batteries are not installed when the instrument a new instrument is shipped. It is recommend that the batteries be removed from the housing when not in use. Leaving the batteries in the housing will cause the batteries to discharge and could cause damage to the cells. Batteries fully charged. • Fully charge the batteries using the provided charger. Fully discharged batteries will take about 8 hours to charge. NiMH battery charger 2 • The Small Battery Housing is made of plastic and has an endcap with two underwater connectors and two captive thumbscrews on one end. • Loosen the thumbscrews. Note that the screws will not need to be fully removed. They are designed to be removed from the main housing but still remain attached to the endcap. The screws can be removed if you continue to rotate the screws through the threaded holes in the endcap. Battery open and ready for batteries

LISST-200X User’s Manual

37 LISST-200X Operation Details • Slide the endcap with battery holder attached out of the main housing. 3 • Slide the clear cover off of the battery holder and install the two D cell batteries. • Install with the positive “button” end toward the end cap of the housing. • Slide the clear cover over the batteries such that the rubber pad is pushing against the batteries as shown in the picture above Batteries installed. 4 • Check that the o-ring in the endcap and the mating surface on the main housing are clean. Lightly grease with silicone grease as needed. • Carefully insert the assembly back into the main housing until the o-ring makes contact with the housing. Align the thumbscrews with the holes in the main housing and push the endcap on until it is flush with the housing. • Tighten the thumbscrews by hand (DO NOT USE ANY TOOL). DO NOT OVERTIGHTEN. Battery assembled and ready for use.

LISST-200X Operation Details

38 LISST-200X User’s Manual 5 • The battery housing can be connected to the LISST- 200X using the provided 5-pin Male to 5-pin Female cable. • Connect the Male end of the cable to the Female Bulkhead connector on the housing. • Connect the other end of the cable to the 5-pin Male bulkhead on the LISST-200X. • Power is now going to the LISST-200X. If AutoStart is enabled the instrument may be active and waiting for the start condition. Instrument is powered and ready to collect data 6 • If desired connect the Communications cable to the male 5-pin bulkhead connector on the Small Battery Housing. Communications will pass through the housing to the LISST-200X. • The battery and communications cables are shown below.

LISST-200X User’s Manual

39 LISST-200X Operation Details 6. Using the Large External Battery Housing The Large External Battery Housing is used to provide power to the LISST-200X for long term deployments or extended profiling operations. STEP ACTION RESULT 1 • The LISST-200X Large External Battery is shipped with 16 Alkaline D batteries pre-installed and is ready for use. The following steps describe how to open the battery and replace the batteries. Skip to Step X to connect the battery to the LISST-200X. • The new alkaline batteries can be left in the battery housing for storage. However, it is not recommended that fully discharged batteries be left in the battery housing during long term storage. Battery ready to use when shipped. 2 • To access and replace the batteries in the Large Battery housing you will need remove the endcap with the connectors and handle. • Remove the four 6-32 X ¾ long socket head cap screws from the connector endcap using the provided 7/64” ball driver. Remove the screws uniformly or hold the cap down against the light spring force. Endcap ready for removal 3 • Pull out the connector endcap. There are no wires or other connections. Set the endcap to the side. Endcap removed and batteries replaced.

LISST-200X Operation Details

40 LISST-200X User’s Manual • Remove the batteries making note of their orientation. • Insert the batteries into the pressure case. Labels on the inside surface of the center divider will guide you on the orientation of the batteries for each stack. 4 • Inspect the o-ring on the endcap and the seating surface on the pressure case to make sure there is no debris that could affect the seal. Lightly grease the surfaces as needed. • Align the tab in the center of the endcap to the slot in the post in the center of the pressure case Endcap re-installed 5 • Screw the connector endcap onto the pressure case with the four 6-32 x ¾ socket head cap screws. If necessary, apply anti-seize compound onto the threads. Battery assembled and ready for use. 6 • The battery housing can be connected to the LISST- 200X using the provided 5-pin Male to 5-pin Female cable. • Connect the Male end of the cable to the Female Bulkhead connector on the Battery Housing. The connector will be labeled with “TO LISST”. Instrument is powered and ready to collect data

LISST-200X User’s Manual

41 LISST-200X Operation Details • Connect the other end of the cable to the 5-pin Male bulkhead on the LISST-200X. • Power is now going to the LISST-200X. If AutoStart is enabled, the instrument may be active and waiting for the start condition. 7 • If desired, connect the Communications cable to the male 5-pin bulkhead connector on the Large Battery Housing which is labeled “TO PC”. Communications will pass through the battery to the LISST-200X. • The battery and communications cables are shown below.

LISST-200X Operation Details

42 LISST-200X User’s Manual 7. Offloading and Deleting Data Files from Internal Memory Data collected using the LISST-200X is stored on the internal data logger. This section describes how to offload and delete files from the internal data logger. STEP ACTION RESULT 1 • Connect the instrument to the computer and establish communication in the LISST-200X program (page 28). Software open, communicating with LISST-200X. 2 • Select Offload Or Erase Files from the home page. Shows a list of files currently stored onboard the instrument 3 • Select a directory to save the offloaded files on the right side. Select individual or multiple files on the LISST to delete or offload (by holding down either shift or ctrl). Press the green arrow. Select files to delete or offload 4 • The progress bars at the bottom of the screen will update as files are offloaded or deleted Files are offloaded or deleted

LISST-200X User’s Manual

43 LISST-200X Operation Details 8. Processing a Single Raw Data File Data that has been downloaded from the datalogger is in a raw binary file (.RBN extension). It must be processed into particle size by the LISST-200X program. STEP ACTION RESULT 1 • Start the LISST-200X program Program started. 2 • Select Process Data Files from the home page. Select the file you want to process. A display similar to the one shown below will appear. • A summary of auxiliary data is shown on the screen. More detailed plots can be viewed by clicking on the parameter. Select file types to be created. 3 • You must also select the inversion model to be used when processing the data: Spherical, randomly shaped or both. • For a description of the particle models, see Appendix F: Particle Shape Models on page 89. • If you have selected Randomly shaped Particle Inversion Model, the processed files will have an’ _rs’ suffix Select inversion method(s) to be used – spherical, randomly shaped or both. 4 • Standard raw data processing will use the background file contained in the .RBN file. If you would like to use a different background file than the one in the RBN file, or use a data record as a background, select the ‘View/Select Background’ button Select to process a file using the background in the data file or an external .BGT file

LISST-200X Operation Details

44 LISST-200X User’s Manual You can view and make background selections from this window. 5 • Next, select the output data file name. Fill in the output directory and the base file name. Several files will be created. • The CSV file is a space delimited ASCII file containing the fully processed and calibrated data. This is useful for loading into spreadsheets such as Excel etc. • The RTX file is a space delimited ASCII file containing the raw data from the LISST-200X data logger. This is useful for looking at the raw data in a spreadsheet such as Excel etc. Name out output file. 6 • To select the range of raw data to process, move the sliders at the bottom of the display. Range to process selected. 7 • After selecting the range to process, press the Process File button. The display will show the processing progress and the resulting data will be displayed when processing is complete. Processing Completed.

LISST-200X User’s Manual

45 LISST-200X Operation Details 9. Batch Processing Multiple Raw Data Files Data that has been downloaded from the datalogger is in a raw binary file (.RBN extension). It must be processed into particle size by the LISST-200X program. If you have multiple raw data files to process (e.g. from a series of profiling deployments), follow these processing steps that allows you to process all files in one operation. STEP ACTION RESULT 1 • Start the LISST-200X program Program started. 4 • Select Process Data Files from the home page. Select multiple files to process (by holding down either shift or ctrl). Select Files 5 • By default, batch processing will use the background stored in the raw data files to process the data. If you would like to use an external .BGT file, press the ‘Select .BGT Background File’ button. Select source of background files 6 • You must also select the inversion model to be used with the real time data processing: Spherical, randomly shaped or both. • For a description of the two particle models, see Appendix F: Particle Shape Models on page 89. • If you have selected Randomly shaped Particle Inversion Model, the processed files will have an’ _rs’ suffix Select inversion method(s) to be used – spherical, randomly shaped or both..

LISST-200X Operation Details

46 LISST-200X User’s Manual 7 • Select the Output Directory for the processed files. • Files will be named with the same name as the raw data files but with different extensions. Several files will be created. • The CSV file is a space delimited ASCII file containing all the processed data. • The RTX file is a space delimited ASCII file containing the raw data from the LISST-200X data logger. Select Output Directory • Press the “Process Files” button. • The files will be processed in order and the progress bars will indicate the status of each file Process files

LISST-200X User’s Manual

47 LISST-200X Operation Details 10. View Processed Data File After processing of the data into a .CSV, the data can be viewed to the screen. This step-by- step procedure covers the viewing of data and optional displays. STEP ACTION RESULT 1 • Start LISST-200X program Program started. 2 • Select View Processed File from the home page 3 • Select the processed data file (.CSV) from the file selection window. Double click the file or type the file name and press Open. Processed data file selected. 4 • After selecting the processed file, a window similar to the one shown below will appear. Particle size distribution displayed on screen • A list of available parameters is displayed on the left. Clicking one of the parameters will open a new window where the parameter will be displayed. The parameter windows can be reorganized by dragging and dropping. The slider at the bottom is used to scroll through the data file. A red point on some graphs will indicate which sample is being displayed. 5 • Additional processed files can be opened simultaneously by pressing the plus button next the open tab at the top left of the screen. • When multiple tabs are open, the tabs can be dragged out into their own windows, allowing for easy comparison between files. Open additional Files

LISST-200X Operation Details

48 LISST-200X User’s Manual 11. Data Quality Control Making sure your data make sense. STEP ACTION RESULT 1 • Follow steps 1-5 in the previous section: View Processed Data File Data file loaded 2 3 • The first value to inspect is the Optical Transmission. By definition, transmission must be a number between 0 and 1 (except that electronic noise may cause it to be slightly higher than 1 in particle-free water) • Tip: The transmission value is in column 60 of the .CSV file. • If transmission shows up as being larger than 1 (one), then your measurement is most likely taken in very clear water and/or you have a bad background measurement obtained with dirty water and/or dirty windows. • If your background measurement was obtained using dirty water, it may be possible to redo the background using clean water and re-process the data, then check to see if the transmission values drop below 1 (one). Transmission values much greater than 1 or less than 0 indicate bad data. 4 • If your transmission values generally are in the 0.98-0.995 range, your measurements are taken in very clear water. • This means that the signal-to-noise ratio will be low, and the data may have a lot of noise in them, but can most likely still be used. Transmission values very close to 1 yield noisy data.

LISST-200X User’s Manual

49 LISST-200X Operation Details 6 • If your transmission values are < 0.10 (or 10%), the water is too turbid. Disregard these data. Data with transmission values < 10% should be disregarded. 7 • If your transmission values are > 0.995 (or 99.5%), the water is too clear. Disregard these data. Data with transmission values > 99.5% should be disregarded. 8 Summarizing • Disregard data if transmission is > 0.995 (> 99.5%). • Disregard data if transmission is < 0.10 (< 10%). • Be wary of data collected at transmission values between 0.98 and 0.995 – low signal-to-noise ratio. • Be wary of data collected at transmission values between 0.30 and 0.10 – generally decreasing data quality as the transmission decreases below 0.30 (30%).

LISST-200X Operation Details

50 LISST-200X User’s Manual 12. Real-Time Data Processing with LISST-200X Software This procedure covers the acquisition, display and storage of processed data in real time, using the LISST-200X software. For on-board processing by the LISST-200X firmware, see Autonomous Real-time Data Processing on page 52. STEP ACTION RESULT 1 • Connect the instrument to the computer and establish communication in the LISST-200X program (page 28). Communication established. 2 • The background currently stored onboard the instrument will be used to process the data as it is collected in real time. • Be sure to collect a new background and store it onboard the instrument before beginning a real time session Background is current 3 • Select Collect Live Data from the home page Open Instrument selected. 4 • A display will appear on the screen. It is similar to the main window of the View Processed Data function. • A list of available parameters is displayed on the left. Clicking one of the parameters will open a new window where the parameter will be displayed. The parameter windows can be reorganized by dragging and dropping. • The ‘Sample Interval’ text box allows you to adjust the time between samples in seconds.

LISST-200X User’s Manual

51 LISST-200X Operation Details • The Sample Number value displayed underneath the plots will increment each time a sample is saved. • The buttons labeled ‘Spherical’ and ‘Irregular’ allow you to display the volume distribution calculated with either the spherical or random shape particle model. The buttons only effect the display and have no effect on the data that is saved in the .CSV file. Data from both particle models is always saved. 5 • Select an output directory and a base file name. Several files will be created. • The CSV file is a space delimited ASCII file containing all the processed data. • The RTX file is a space delimited ASCII file containing the raw data from the LISST-200X data logger. Select output directory and file name 6 • Data will not be saved until the ‘Start Logging Data’ button is pressed. Start Data Logging

LISST-200X Operation Details

52 LISST-200X User’s Manual 13. Autonomous Real-time Data Processing Newer LISST-200X’s (starting with serial number 2131), or older units with upgraded electronics, have enough computational power to do PSD processing autonomously, without requiring a connection to a Windows computer. You may choose to save the processed data files (.CSV) in the LISST-200X memory, or transmit them in real time through the serial port, or both. Note that processing slows the maximum sampling rate. Without it, the LISST-200X can sample continuously once per second. Depending on the settings you select, the time per sample may be up to 2 seconds. To determine whether your LISST-200X supports real-time processing STEP ACTION RESULT 1 • Connect the instrument to the computer and establish communication in the LISST-200X program (page 28). Communication established. 2 • Open the information panel at the bottom of the window, and check the firmware version: • If the firmware is 2.1 or greater, your instrument supports autonomous processing. • If the firmware is less than 2.1 but at least 2, a user- installable firmware upgrade can enable autonomous processing. Contact Sequoia Scientific support for instructions. • If the firmware version is less than 2, your instrument’s hardware does not support autonomous processing. Contact Sequoia Scientific sales regarding upgrade options. If your instrument supports real-time processing, proceed. To use real-time processing STEP ACTION RESULT 1 • Connect the instrument to the computer and establish communication in the LISST-200X program. (see Establishing Communication with the LISST-200X on page 28). Communication established.

LISST-200X User’s Manual

53 LISST-200X Operation Details 2 • Navigate to the Configure Instrument page. If your instrument’s hardware and firmware support it (see above), you will see a Data Output category. Configure Instrument page open. 3 • Click on Data Output to make its options visible. o ‘Save PSD Files On Instrument’ – if checked: ▪ PSD is computed onboard in real time ▪ PSD is saved in a file onboard the instrument, in .CSV format (see Appendix C: Data File Formats on page 79) o ‘Transmit PSD In Real Time’ – if checked: ▪ PSD is computed onboard in real time ▪ A PSD data record is transmitted over the serial port immediately after each measurement is collected o ‘Inversion Model’ ▪ Select the inversion model that will be used for onboard PSD processing (see Appendix F: Particle Shape Models on page 89) Instrument configuration selected. 4 • IMPORTANT: after you have selected the desired configuration, press the ‘Send Config. To Instrument’ button to load the configuration onto the instrument. Configuration uploaded to instrument.

LISST-200X Operation Details

54 LISST-200X User’s Manual • A message box will confirm the upload has been completed. Configuration complete.

LISST-200X User’s Manual

55 LISST-200X Operation Details 14. Collecting Data from an External Analog Sensor The LISST-200X can supply regulated 12V power to an external sensor, and measure one to three analog voltages, depending on its hardware configuration and firmware settings. If the 12V power output is activated, it will turn on each time the 200X collects a sample or burst of samples, and turn off each time the 200X goes into a waiting state of more than a few seconds. If using the 12V output, you can specify how long it should turn on before sampling, as a warmup time for the sensor. To provide these functions, the auxiliary connector on the LISST-200X must be in one of three configurations: Analog & Digital Input (which provides a single analog input), Dual Analog Input configuration (which provides two), or Triple Analog Input configuration (which provides three). Remember that power consumed by an external sensor will reduce battery endurance, in a battery-powered deployment. It might also exceed the capacity of the USB-powered data cable. If so, you can augment the USB cable’s power by connecting the large battery housing. STEP ACTION RESULT 1 • Determine current configuration of Auxiliary connector: in LISST-200X Software, open the instrument status bar. Look for the Endcap Configuration. If it is not one of the analog- input configurations, it must be changed; contact Sequoia for assistance. Auxiliary Connector configuration determined 2 • Make the appropriate connections between the 6-pin auxiliary connector of the LISST-200X and your sensor. To see the pin assignments, type the CONFIG command in the LISST-200X terminal window; page 61). Or see Appendix E: Connectors on page 84. External sensor connected 3 • Type SENSORPOWER into the terminal window, and respond to its prompt to select the appropriate option. • Type SENSORWARM in the terminal window to set a warmup time (in seconds). • WARNING: a USB cable may not provide enough power for external sensors. Use a battery housing or other power source to provide enough power to run external sensors. Configured for analog collection 4 • Set the standard sampling parameters to control the sampling schedule and other options (see • Configuring Data Collection on page 32). Ready to collect data

LISST-200X Operation Details

56 LISST-200X User’s Manual 15. Configuring the LISST-200X as a Sensor for a CTD The LISST-200X can operate as an analog-output auxiliary sensor for a CTD. For this purpose the LISST-200X’s auxiliary six-pin connector must be in its “Dual Analog Output” configuration. The two analog outputs represent the total volume concentration of particles, and the Sauter mean diameter (SMD). These values are approximations for the convenience of real-time display. For highest-quality data you will still need to offload and process the detailed data from the LISST-200X after deployment. The LISST-200X updates its analog outputs each time it collects a sample, at a maximum of 1 Hz. STEP ACTION RESULT 1 • Determine current configuration of Auxiliary connector: in LISST-200X software, open the instrument status bar. Look for the Endcap Configuration. If it is not in the Dual Analog Output configuration, it must be changed; contact Sequoia for assistance. Auxiliary Connector configuration determined 2 • Make the appropriate connections between the 6-pin auxiliary connector of the LISST-200X and your sensor. To see the pin assignments, type the CONFIG command in the LISST-200X terminal window (see LISST-200X Command Details on page 61), or see Appendix E: Connectors. Connected to CTD 3 • In LISST-200X software, set the sampling parameters as follows (see • Configuring Data Collection on page 32): o Operating mode: Fixed sample rate o Sample to be average of 20 measurements o Sample interval: 1 second o Automatically start sampling program upon power up Start condition: delay 0 minutes o Stop condition: memory full 200X will collect data at 1 Hz whenever the CTD supplies power to it

LISST-200X User’s Manual

57 LISST-200X Operation Details B. Start and Stop Conditions Overview The LISST-200X can be set up to start and stop data collection according to various conditions. The most obvious is the white switch lever on the connector end cap. However, the switch is only one of several options. The start condition and stop condition can be different from each other. Also see Configuring Data Collection on page 32. Start Conditions The five possible start conditions are: 1. Depth, 2. Time, 3. External Mechanical Switch, 4. External Digital Input, and 5. Time Delay. The details of each are described below. Note that regardless of which start condition is selected, the LISST-200X will not start checking for the start condition until the sample program is run from the Windows software (see page 32), by the GO command (see Acquisition/Action commands on page 69), or by automatic start after power-up. Depth Start The LISST-200X checks its depth every 3 seconds. Each time it checks, the green LED on the end cap will double-flash. If the depth exceeds the threshold, the program will start data collection. When setting the threshold, remember that atmospheric pressure, temperature, and even the orientation of the LISST-200X can affect the offset of the pressure sensor. Therefore, it is best to zero the sensor (see page 32) immediately before starting, and to set the threshold to at least 1 meter, to avoid incorrect triggering. Time Start The LISST-200X will go into lower-power sleep until the programmed time. (If the programmed time has already passed, it will start immediately.) While waiting, it will quickly flash the LED once every 30 seconds. External Mechanical Switch Start The LISST-200X will wait until the white plastic lever on its endcap is moved to the “1” position. (If it is already in that position, it will start immediately.) While waiting, it will quickly flash the LED twice every few seconds. External Digital Input Start If the auxiliary connector is configured for analog and digital inputs (see Appendix E: Connectors), pin 2 of the 6-pin connector is the digital input. The program will check the status of the digital input once a second. If the digital input is greater than 2 volts (relative to digital ground, pin1), data collection will start. While waiting, the LED will quickly flash twice every few seconds. Time Delay Start The time delay start condition will cause the program to wait the specified number of minutes. While waiting it will quickly flash the LED once every 30 seconds.

LISST-200X Operation Details

58 LISST-200X User’s Manual Stop Conditions The possible stop conditions are: 1. Depth, 2. Time, 3. External Mechanical Switch, 4. External Digital Input, 5. Fixed number of samples, and 6. Memory full or power lost When storing data in the Fixed Sample Rate mode the Stop conditions are checked after each averaged sample has been saved. When storing data in the Burst mode the Stop conditions are only checked after a full Burst has been completed. The Start and Stop conditions have no effect on the real-time sampling mode (Real-Time Data Processing with LISST-200X Software on page 50). The details of each condition are described below. Depth Stop If the depth is less than the threshold, sampling will stop. If the Start Condition is Depth Start the program will return to checking for the start depth, and will start a new file each time the start threshold his crossed. For all other Start Conditions the when the current depth is less than the threshold the LISST-200X will stop and go into deep sleep. If using depth for both start and stop conditions, consider whether the start and stop thresholds should be different, depending on your deployment scenario. For example, if the instrument might be suspended from a winch at a depth near the start and stop thresholds, there could be “false” starts and stops due to wave motion. This could be prevented by setting the start depth deeper than the stop depth. Time Stop After each sample or burst, if the given date and time have passed, the LISST-200X will stop and go into low-power sleep. External Mechanical Switch Stop After each sample or burst, if the switch lever is in the off or “0” position, sampling will stop. If the Start Condition is Switch Start, the program will return to checking the start condition. For all other Start conditions the instrument will stop and go into deep sleep. External Digital Input Stop The status of the digital 1 input is checked after each sample or burst. If the voltage at the input is less than 0.7 volts the sampling will stop. If the Start Condition is a Digital Input Start the program will return to checking the start condition. For all other Start conditions the instrument will stop and go into deep sleep. Fixed Number of Samples When the number of samples to be saved has been reached the program will stop and go into deep sleep (regardless of start condition). Memory Full or Voltage Low Sampling will continue until the memory is full, or the power input falls. If the memory fills, the instrument will go into deep sleep. NOTE: the memory capacity is large enough that it should never be necessary to fill it. We recommend deleting files from the LISST-200X as soon as they have been offloaded to a backed-up computer or other storage.

LISST-200X User’s Manual

59 LISST-200X Operation Details C. Instrument Communication Overview The cables Sequoia supplies with the LISST-200X send data through a standard USB port. However, the LISST-200X itself communicates through RS-232 serial protocol. The 5-pin male connector on the LISST-200X carries the RS-232 signals (and power). The cables incorporate adapters that convert between RS-232 and USB. If you use a different cable that does not include the RS-232-to-USB adapter, you can use an external adapter, or connect directly to an RS- 232 port if your computer has one (rare on any recent computer). In Windows, both RS-232 and USB connections will appear as COM ports. If interfacing to a data logger or controller system other than a Windows computer, you will likely also use a direct RS-232 connection. The RS232 link communicates at 9600 baud, 8 data bits, No parity, and 1 stop bit. For offloading data files, the LISST-200X software uses YMODEM protocol at 115K baud. The transfer rate can be changed in the settings of the LISST-200X software. A slower speed may be required when downloading data over cables longer than a few meters. Using the LISST- 200X Terminal Window In the LISST-200X software, you can open a terminal window to directly view communications with the LISST-200X, and enter commands. When the LISST-200X is connected and the terminal window is the front window, the LISST-200X should respond to pressing the enter key with the L200X:> prompt. If the instrument is in the deep sleep mode, you can wake it by pressing the Stop button. Direct Commands In most cases, LISST-200X software will be used to configure and operate the LISST-200X. However, some functions are available only through direct commands in the terminal window. Also, in some applications the LISST-200X may need to communicate to another datalogger or custom program. For this purpose, a large set of commands is available to operate the instrument. See the following sections for detailed descriptions of each command.

LISST-200X Operation Details

60 LISST-200X User’s Manual D. LISST-200X Command Summary Display Commands CONFIG Display instrument hardware configuration information DD, DIR Display current disk directory DEPTH Display current depth DS, DS 1 Display current instrument status information DT Display current time and date HELP, ? Display general help messages and command list SW Display mechanical switch position TEMP Display current temperature VIN Display input supply voltage Setup Commands AS x Set autostart BI x Set burst interval MA x Set samples per average equal to x MODE, OM x Set operating mode (burst or fixed-rate) PD x, STOPDATA x Set stop condition Data PSDMODEL x Set model used for PSD processing (if active) SAVEPSD x Set whether PSD data will be saved in on-board memory SI x Set sample interval SP x, STOPCOND x Set stop condition ST x, STARTCOND x Set start condition. SB x Set samples per burst SC yyyy/mm/dd hh:mm:ss Set clock with time and date SENDDATA x Control real-time transmission of data via RS232 SENSORPOWER x t Set whether power output turns on during logging SENSORWARM t Set external sensor power warmup time TD x, STARTDATA x Set start condition data ZD Reset depth sensor offset Acquisition/Action Commands DL filename Delete file GO Start data collection using current parameters GX Grab sample and transmit it XR Transmit raw data when average saved ZS Acquire 20 samples and evaluate them for quality. ZZ Go into deep sleep mode (minimum power consumption)

LISST-200X User’s Manual

61 LISST-200X Operation Details E. LISST-200X Command Details NOTE: Command are shown in upper case for clarity, but are not case-sensitive. 1. Display Commands CONFIG Display instrument hardware configuration Syntax: CONFIG Description: The current hardware configuration including Auxiliary I/O connector configuration Example: L200x:> CONFIG Serial Number: 2001 Firmware Version: 1100 (1.100) VCC: 59371 Full Path Length: 2500 (25 mm) Eff. Path Length: 2500 (25 mm) BioBlock Installed: 0 Start Condition: Digital input (4) Start Condition Data: 0 Stop Condition: Switch (3) Stop Condition Data: 0 Measurements Per Average: 10 Sample Interval: 1 seconds Burst Mode: 0 Samples Per Burst: 5 Burst Interval: 25 minutes Transmit Raw Data: 0 Lifetime Sample Count: 529554 Lifetime Laser On Time: 1703980 seconds Endcap Configuration: 0 (Analog & digital in) Wiring of MCBH6MP connector: 1: Ground 2: Digital input 1 3: Digital input 2 4: +12V output 5: Analog input 1 6: Ground DD, DIR Display File Directory Syntax: DD or DIR Description: Display current disk directory in DOS type format. Includes total bytes used and bytes available. Example: L200x:> DD . 2099-03-14 03:35:48 .. 2099-03-14 03:35:48 L2390658.RBN 1,612,440 2019-08-29 14:55:28 L2411501.RBN 3,253,800 2019-09-03 07:58:20 L2480807.RBN 736,800 2019-09-06 09:40:14 3 file(s) 7,608,360 bytes 2 dir(s) 1,016,381,440 bytes free

LISST-200X Operation Details

62 LISST-200X User’s Manual DEPTH Display current depth Syntax: DEPTH Description: The current depth is displayed to the screen. Example: L200x:> depth 0.000 m (680 counts) DS Display current status information Syntax: DS Description: The instrument settings and status are displayed to the screen. The format may change with different firmware and hardware versions. Example: L200x:> DS LISST-200X Status and Settings Serial number: 2021 Endcap configuration: 2 (Dual analog out) Firmware Version: 1.470 Feb 8 2021 21:06:43 Current Date/Time: 2019/01/09 22:12:50 Current Day of the Year: 009 Supply voltage: 14.74 V Operating mode: Fixed rate mode AutoStart: OFF. Delay Start with 0 minute delay Stop on full memory or low battery Measurements per Average: 32 Ambient light subtraction: on. Sample Interval: 1 Path length: 25 mm Standard status messages will be sent during logging. Data will be saved in files of unlimited size. AUV mode: off Free memory: 1016 Mbytes DS 1 Display current status information, compact form Syntax: DS 1 Description: Instrument status displayed in simplified format for reading by software Example: L200x:> ds 1 LISST-200X Current Status and Settings SN = 0 OM = 0 ST = 5 TD = 0 SP = 6 PD = 0 MA = 32 SI = 1 BI = 1 SB = 1 AS = 0 BIOBLOCK = 0 STUBE = 0

LISST-200X User’s Manual

63 LISST-200X Operation Details SENSORPOWER = 0 SENSORWARM = 0 Current Time = 1987/01/09 22:14:00 Battery= 1474 Switch= 0 Memory= 1016381440 Bytes Endcap= 2 PATHLENGTH = 25 Firmware = 1.470 DT Display current time and date Syntax: DT Description: Displays current time and date to the screen. Example: L200x:> DT Date/time = 2021/01/01 18:16:06 HELP, ? Display general help messages and command list Syntax: HELP or ? Description: Displays the list of command to the screen. Example: L200x:> HE LISST-200X Commands ------------------------------------- [list of commands] SW Display current Mechanical Switch position Syntax: SW Description: The current position of the Mechanical Switch is displayed to the screen. Example: L200x:> SW Switch: 0 TEMP Display current temperature Syntax: TEMP Description: The current temperature is displayed to the screen. Example: L200x:> temp 25.09 C (26629 counts) VIN Display Battery Voltage Syntax: VIN Description: The current supply voltage is displayed to the screen. Example: L200x:> vin 14.74 V (2058 counts)

LISST-200X Operation Details

64 LISST-200X User’s Manual 2. Setup Commands AS Enable Autostart Syntax: AS x, where x is 1 (yes) or 0 (no) Description: With Autostart enabled, the firmware will immediately start the sampling program when power is applied to the instrument. Example: L200x:> AS 1 Autostart upon power-up is on Cautions: If Autostart is enabled, the user cannot communicate with the instrument when powering it up – it will immediately start sampling according to the SD defaults. In order to stop sampling and establish normal communication, the user must issue a stop command, either from the LISST-200X software’s graphical interface, or by pressing CTRL-C in the terminal window. BI Set Burst Interval Syntax: BI x, where x is the number of seconds between the start of 2 consecutive bursts, from 1 to 10,000 seconds. Description: In Burst Mode (MODE = 1), the burst interval is the number of seconds between two consecutive bursts, each burst composed of a number of samples per burst (specified by the SB command). Example: L200x:> BI 900 New Seconds between Bursts: 900 MA Set measurements per average Syntax: MA x The exact result depends on whether ambient light rejection (ALR) is active. Starting with firmware version 1.4 (May, 2017), ALR is on by default, and the actual number of samples averaged will be approximately x/10. That is, MA 32 will result in 3 ALR cycles per average. With or without ALR, 32 is the maximum MA value that allows 1 sample per second. Description: Each recorded or displayed measurement is based on an average of measurements. MA sets the number of measurements per. If no value follows command, prompts will be displayed for the value. Example: L200x:> ma 32 Measurements per average: 32 (3 ambient subtractions) OM Set Operating Mode Syntax: MODE x or OM x Description: Sets the Operating Mode to one of the following types: 0 = Fixed rate 1 = Burst Mode OM command only without a parameter will initiate a prompt for the Operating mode. Example: L200x:> OM 1 Operating mode: Burst mode (1)

LISST-200X User’s Manual

65 LISST-200X Operation Details PD, STOPDATA Set Stop Condition Data Syntax: PD x or STOPDATA x where x is the stop condition data as described below Description: The PD command sets the stop condition data to be used when the collection data. The stop condition data is used with the Stop Condition settings as follows. • If the Stop Condition is Depth Stop (option 1) the input will be stop depth in meters. • If the stop condition is set to Time/Date Stop (option 2) the input for PD will be the stop date and time. • If the Stop Condition is Fixed Number Stop (option 5) the input will be the number of samples to collect before stopping. • The PD setting is ignored for Mechanical Switch Stop (option 3), Digital Input Stop (option 4) or Maximum memory or Low Battery Stop (option 6). Example: If Stop Condition =1 (Depth Stop): L200x:> PD 3 Depth stop at 3 meters If Stop Condition = 2 (Time/Date Stop): L200x:> PD 2016/12/31 23:59:59 Time stop at 2016/12/31 23:59:59 If Stop Condition = 5 (Fixed Number of Samples): L200x:> PD 100 Stop after 100 samples. PSDMODEL Set particle model used for on-board size distribution calculations (firmware 2.1 or higher) Syntax: PSDMODEL x Description: If x = 0, PSD processing with use the spherical particle model. If x = 1, the irregular (also called “random”) particle model will apply. This setting applies only when real-time PSD processing is active. It is available only in firmware 2.1 and higher. Also see SAVEPSD, SENDDATA, and Autonomous Real-time Data Processing on page 52. Example: L200x:> PSDMODEL 1 Optical Model: 1 (irregular) SAVEDATA Save Data Setting—USE WITH CAUTION Syntax: SAVEDATA x, where x is 1 or 0 Description: SAVEDATA 0 turns off storing of raw data in the instrument’s memory. This is only for unusual circumstances. Example: L200x:> SAVEDATA 1 1: Raw data will be saved during logging. L200x:> SAVEDATA 0 WARNING: Data will NOT be saved during logging! Is this really what you want [N] ? n

LISST-200X Operation Details

66 LISST-200X User’s Manual Cautions: Be very careful with this setting! Issue a DS command to verify the status of the store mode setting. A warning will be displayed as part of the DS status if the store mode is disabled: SAVEPSD Control saving of processed PSD files (firmware 2.1 or higher) Syntax: SAVEPSD x Description: If x = 1, during sampling the instrument will generate a .CSV file containing processed particle size distributions, in addition to the normal raw .RBN file. If x = 0, only the normal .RBN files will be saved. For more details, see Autonomous Real-time Data Processing on page 52. This command requires firmware version 2.1 or higher. Example: L200x:> SAVEPSD PSD data will be saved during logging. SB Set Samples per Burst Syntax: SB x, where x is the number of samples per burst. Each sample is taken at the sample interval (in seconds) set by the SI command. If no value follows command, prompts will be displayed for the value. Example: L200x:> SB 10 New Samples per Burst: 10 SC Set Clock with time and date Syntax: SC yyyy/mm/dd hh:mm:ss Where yyyy=year, mm=month, dd=day, hh=hour (24 hour format), mm=minute, ss=seconds If no values follow the “SC” or “sc” command, prompts for entering the time and date will be displayed. Example: L200x:> SC 2020/12/20 13:20:18 Time set to 2020/12/20 13:20:18 SENDDATA Transmit data during sampling (firmware 2.1 or higher) Syntax: SENDDATA x Description: The value of x sets whether data will be transmitted from the instrument’s RS232 interface during sampling. Only options 0 and 4 are suitable for normal operation. 0 = Do not send data (default) 1 = Multi-line troubleshooting data 2 = NMEA-format brief data (for glider applications) 3 = Single-line raw (CANNOT be processed by LISST-200X software) 4 = Processed PSD data; activates real-time processing and sends processed data after each sample. For details, see Autonomous Real- time Data Processing on page 52. This option requires firmware version 2.1 or higher. Example: L200x:> SENDDATA 0 Setting: 0 (no data)

LISST-200X User’s Manual

67 LISST-200X Operation Details SENSORPOWER Control 12V power to auxiliary connector during logging Syntax: SENSORPOWER x, where x is 1 or 0 Description: If x = 1, the 12V power output on the auxiliary connector (if it is suitably configured) will turn on during data collection with the GO command. If running in burst mode, the power will turn off between bursts. Logging will also incorporate a warmup time, set with the SENSORWARM command. If SENSORPOWER is entered without an x argument, it will prompt for a value and for a sensor warmup time. Example: L200x:> SENSORPOWER 1 Power output will turn on during logging, with 5 second warmup SENSORWARM Warmup time when powering external sensors Syntax: SENSORWARM x, where x is in seconds Description: If SENSORPOWER is on, SENSORWARM sets the number of seconds the LISST-200X will wait for external sensors to warm up before it proceeds with sampling. Example: L200x:> SENSORWARM 10 Power output, when used, will turn on with 10 second warmup. Power output during logging is on. SI Set Sample Interval Syntax: SI x, where x is the number of seconds between samples, from 1 to 10,000. Description: In either fixed-rate or burst mode, the sample interval is the number of seconds between two consecutive samples, each composed as an average of a number of measurements (specified by the MA command). Example: L200x:> SI 5 Seconds between samples: 5 SP, STOPCOND Set Stop Condition Syntax: SP x or STOPDATA x, where x is the stop condition code described below Description: The SP command sets the stop condition to be used when collecting data. The stop condition options are: 1 = Depth Stop 2 = Time/Date Stop 3 = Mechanical Switch Stop 4 = Digital Input Stop 5 = Number of Samples Stop 6 = Maximum memory or Low Battery Stop If no value follows command, prompts will be displayed for the value. After setting a stop condition of Depth, Time/Date or Number of Samples, use STOPDATA to enter an appropriate parameter value. Example: L200x:> SP 5 Stop condition: 5

LISST-200X Operation Details

68 LISST-200X User’s Manual ST, STARTCOND Set Start Condition Syntax: ST x or STARTCOND x where x is the start condition code described below Description: The ST command sets the start condition to be used when the GO command is issued. The start condition options are: 1 = Depth Start 2 = Time/Date Start 3 = Mechanical Switch Start 4 = Digital Input Start 5 = Delay Start If no value follows command, prompts will be displayed for the value. After setting the start condition to Depth, Time or Delay, use STARTDATA to set the applicable condition value. Example: L200x:> ST 5 Start condition: 5 TD, STARTDATA Set Start Condition Data Syntax: TD x or STARTDATA x where x is the start condition data described below Description: The TD command sets the start condition data to be used when the GO command is issued. The start condition data is used with the Start Condition setting as follows: • If the Start Condition is Depth Start (option 1) the input will be start depth in meters. • If the start condition is set to Time/Date Start (option 2) the input for TD will be the start date and time. • If the Start Condition is Delay Start (option 5) the input will be time delay in minutes. • The TD setting is ignored for Mechanical Switch Start (option 3) or Digital Input Start(option 4). If no value follows command, prompts will be displayed for the value. Example: If Start Condition = 1 (Depth Start) L200x:> STARTDATA 3 Start condition: Depth Start at 3 meters If Start Condition = 2 (Time/Date Start) L200x:> STARTDATA 2020/12/31 23:59:59 Start condition: Time Start at 2020/12/31 23:59:59 If Start Condition = 5 (Delay Start) L200x:> STARTDATA 2 Start condition: Delay Start with 2 minute delay ZD Reset Depth Sensor Offset Syntax: ZD

LISST-200X User’s Manual

69 LISST-200X Operation Details Description: The ZD (or zd) command resets the depth sensor offset so that the sensor reads a depth of 0m at zero depth (in air). You must issue the ZD command, then select 1 (yes) or 0 (no) to reset depth sensor Example: L200x:> ZD Depth sensor offset adjust. Instrument must be at zero depth and similar temperature to field conditions. Reset depth sensor offset? (1=yes,0=no): [0] ? 1 Previous offset was -140.65. New offset is -4.37. Previous depth was -136.28 meters. New depth using corrected offset is 0.00 meters. 3. Acquisition/Action commands DL Delete file from LISST memory Syntax: DL filename, where filename is the name of the file to be deleted. Description: DL command is used to delete file from memory. Example: L200x:> DL L3231513.RBN Delete the file 'L3231513.RBN'? Y for Yes or N for No: [N] ? Y Deleting L3231513.RBN Deleted. Cautions: WARNING: Make sure the file being deleted has already been offloaded. Once the file is deleted it cannot be recovered. GO Start Data Collection using current Settings Syntax: GO Description: Starts Fixed Rate or Burst Mode Data collection using current settings. Example: L200x:> GO Starting sampling. Start condition: Delay Start with 2 minute delay Stop condition: Stop after 100 samples Low-power sleep until 2021/12/10 17:39:56 zzz...Waiting until 2021/12/10 17:39:56 Cautions: To stop data acquisition before it is complete, send 2 Ctrl-C characters. GX Grab sample and transmit it Syntax: GX Description: Acquires single averaged sample and displays the result to the screen. Example: Input: GX Output: { …. 36 ring values + 24 Aux parameters } Notes: The GX command does not store the sample to a datafile.

LISST-200X Operation Details

70 LISST-200X User’s Manual ZS Collect and transmit background scattering data Syntax: ZS Description: Acquires 20 averaged sample and displays the result to the screen. Evaluates the data for quality as a background measurement. Example: Input: ZS Output: { …. 36 ring values + 24 Aux parameters } repeat 20 times, then messages about the quality of the data. Notes: The ZS command does not store the sample to a datafile. ZZ Go into deep sleep mode (minimum power consumption) Syntax: ZZ Description: Sends LISST-200X in to low power sleep mode. Instrument will send a message every 30 seconds to indicate it is sleeping. It can be wakened at any time by sending 2 control-C characters about ½ second apart. Example: L200x:> ZZ Low-power sleep. zzz...Press Stop or C to wake up. zzz...Press Stop or C to wake up.

LISST-200X User’s Manual

71 LISST-200X Operation Details F. Performance Optimization Background Measurement and its Importance As discussed in earlier sections, the LISST-200X measures light scattered at small angles, to characterize particles in water. However, a certain amount of scattering is also generated by pure water, and by the LISST-200X’s own optics and windows. We call this the background scattering. In order to determine the scattering contributed by particles, the background scattering must be accurately measured and subtracted from the total scattering measured. The measurement of background scattering is sometimes called a “zscat”, meaning the signal measured in water with zero “scatterers” or particles. The minimum concentration of particles measurable by the LISST-200X is very sensitive to the quality of the background scatter file. The lower the concentration in the water being measured, the more important the quality of water and the care used when measuring the background. When measuring in water with high particle concentrations, the relative signal-to-background noise ratio increases, and final results will be less sensitive to flaws the background. However, a background should always be collected before an experiment. Techniques for measuring background Cleanliness It is essential to use clean, bubble-free water for the background measurement. The water can be fresh or salt water. For most applications, steam-distilled water works well. Steam-distilled water is typically available in one-gallon containers. We have found that this distilled water tends to be a bit cleaner than typical bottled or packaged water. We buy steam-distilled drinking water and filter it through a 0.2µm filter. A small volume horizontal chamber has been provided for submerging the optics while acquiring a background. It is inserted between the windows of the optics. The instrument should be placed horizontally on the supplied white plastic supports. However, plastic bags, Tupperware or clean containers of any kind can be used to hold water for a background measurement. The instrument’s optics end can be submerged in them vertically, as long as the optics are completely submerged and there is no blockage of the windows by bubbles or other objects. Because the area surrounding the windows is submerged, it is important to thoroughly clean and rinse this part of the instrument before acquiring a background. Toothbrush, liquid soap and water works well for cleaning the optical end of the instrument. Do not use abrasive powders, which will scratch the windows and degrade instrument performance.

LISST-200X Operation Details

72 LISST-200X User’s Manual Watch for Outgassing and Bubbles Water may contain dissolved air that develops into bubbles as the water warms. Bubbles, especially if they cling to the windows of the LISST-200X, will greatly modify the scattering pattern, rendering the background useless. Be sure to remove any bubbles from the windows before acquiring a background. Use a squirt bottle or pipette to blow the bubbles off the window. Ambient Light Rejection Starting with firmware version 1.4 in May, 2017, the LISST-200X automatically measures and subtracts light that is not produced by scattering from its laser. This Ambient Light Rejection (ALR) prevents sunlight or other bright sources from contaminating the scattering measurements. Without ALR, as in earlier LISST instruments, high ambient light levels, if not physically blocked from the LISST’s optical aperture, could falsely elevate the apparent concentration of small particles. You can verify that ALR is active by connecting to the instrument with LISST-200X, then opening the terminal window. Type DS and press enter, and look for “Ambient light subtraction: on” in the output. If it is not shown, you should upgrade the firmware. Contact Sequoia Scientific for assistance. ALR works by measuring the signal on the LISST-200X’s 36 detectors first with the laser on, then with the laser off. This on-off process is repeated 3 times per second. Because of the time required for the laser and electronics to change between the on and off states, rapidly changing ambient light may not be completely rejected. Therefore, if working in bright sunlight, near the water surface, and in conditions where the sunlight is rapidly modulated by waves, ambient light could still have an effect. ALR will reduce the average effect, but the changing ambient conditions could increase sample-to-sample noise. In those worst-case conditions it is still advisable to shade the LISST- 200X optics from ambient light. The same is true if operating under intense artificial light modulated by AC power. Optical Alignment The background measurement (described step-by-step on page 29) is the best source of information on the current health of the LISST-200X. It provides information on the current functionality of all of the major systems including laser, ring detector, data collection electronics, and optical alignment. The LISST-200X is a sensitive optical device. The laser must be aligned such that the focused spot is centered on the hole in the center of the ring detector. If this alignment in not correct the instrument will not function correctly. The LISST-200X software automatically detects and warns of serious misalignment, during the background measurement process. The first indicator of an alignment problem is a severe drop in the transmitted laser power when compared to the factory laser power. The laser power and laser reference values will change over time but they should track together.

LISST-200X User’s Manual

73 LISST-200X Operation Details The second indictor of misalignment is high value on the inner rings. The inner rings are shown on the left side of the background display. The most important rings for misalignment indication are rings 1-4 in the background scatter file. If the focused laser starts to move away from the center of the ring detector it will scatter more light onto the inner rings. This will cause the inner ring values to be much higher than the factory values, and cause the transmitted laser power to decrease. It must be noted that low laser power or high inner rings may not always indicate misalignment. Low laser power can also occur because the windows or water is dirty or if there are bubbles on the window. Large particles or bubbles can cause the inner rings to be higher than factory values. All of the other possibilities must be eliminated before it can be concluded that the instrument is misaligned. If necessary, the LISST-200X can be aligned by Sequoia Scientific or an authorized service representative.

LISST-200X Operation Details

74 LISST-200X User’s Manual G. Instrument Mounting, Deployment Orientation, Storage and Maintenance Horizontal Mounting The LISST-200X measures scattered light to obtain particle size distributions. Therefore, it is important to keep particles from sticking to the window surface. The first line of defense it to keep the window surface vertical by mounting the instrument horizontally. This is most critical when the water motion is slow. When working in rivers or in a towed or profiling application the orientation is not as critical. The center section of the main pressure case has a slightly smaller diameter than the ends. The smaller diameter section is the suggested mounting area. The U-shaped mounting block provided with the instrument fit this diameter and can be used for mounting the instrument. The clamps supplied with the instrument can also be used. Spare clamps for mounting the instrument are available from Sequoia. Electrical Isolation To reduce the corrosion of the aluminum parts, a zinc anode is attached to the connector endcap. This anode must be exposed to the water for it to be effective. When mounting the instrument be sure to electrically isolate the instrument from all other metal. Any contact with other metal can greatly increase the rate of corrosion. Isolate the instrument with rubber or plastic to keep the LISST-200X from being the sacrificial anode for the mounting hardware. Failure to properly isolate the instrument from all other metal will void the warranty. A set of stainless steel clamps with rubberized interiors ships as part of the instrument package. Use these or the plastic clamps for mounting, and make sure that no metal is in direct contact with the pressure housing or other components of the instrument. If you lose the spare clamps, replacement clamps are available for purchase from Sequoia. Cleaning the Optical Windows The condition of the windows is critical to the performance of the LISST-200X. Care must be taken when cleaning the windows. The windows and the instrument should be rinsed thoroughly with fresh water after each deployment. The windows should be cleaned with a soft cloth or lens tissue. Liquid detergent/soap and water may be used. For removing grease spots, finger prints, etc., alcohol may be used. Do not use stronger solvents such as Acetone or Toluene.

LISST-200X User’s Manual

75 LISST-200X Operation Details Abrasive powders must never be used near the optics windows; they will scratch the windows and degrade instrument performance. Orientation Relative to Current Flow In situations with significant current flow, it is best to orient the instrument perpendicular to the current, so water can flow as directly as possible through the optics end of the instrument. The diagram below shows the optimum orientation with respect to current flow Storage For full protection, always store the LISST-200X in its foam-lined shipping case. Calibrations and Adjustments The LISST-200X should not require any adjustment or calibrations. The pressure sensors can be re-calibrated by Sequoia, if desired. The performance of the instrument can be checked with the use of a sample of particles of a known size distribution. Standard particles can be obtained from manufacturers of standard particles, such as Whitehouse Scientific (http://www.whitehousescientific.com/) or PTI – Powder Technology Inc. (http://www.powdertechnologyinc.com/). Comparison of the LISST-200X’s performance with that of other particle-size measuring devices that are not based on laser diffraction may result in inconsistencies because the instruments may measure different properties of the particles to interpret particle size.

Appendix A: Technical Specifications

76 LISST-200X User’s Manual Appendix A: Technical Specifications • Parameters measured/derived: o Particle Size distribution o Total Volume Concentration o Optical Transmission o Depth o Temperature o Volume Scattering Function • Particle size range: 1.00 – 500 µm in 36 log-spaced classes • Optical path length: 2.5 cm • Optical transmission: 12-bit resolution • Analog inputs (availability depends on configuration) o Measurement range: 0 to 5.2 V o Resolution: 0.1 mV (inputs 1 and 2); 1.5 mV (input 3) o Uncertainty: 1.5% of reading + 2 mV o Input impedance: 500 kΩ o Time constant: 250 ms • Digital input (availability depends on configuration) o Nominal logic levels: 0 and 3V o Voltage tolerance without damage: -12 to 12V • Analog outputs (availability depends on configuration) o Output range: 0.1 to 3.0 V o Resolution: < 1 mV o Output impedance: 1 kOhm o Uncertainty: 1.5% of reading + 2 mV • Data storage memory: 1GB (~12 million measurements) • Maximum sample speed: 1 size distribution per second • Temperature sensor o Range: -5 to 45 °C o Resolution: 0.001 °C o Uncertainty: approximately 1 °C • Depth Sensor o Range: 0 to 600 m of sea water o Resolution: 1 cm o Uncertainty: approximately 1% of reading (if atmospheric offset zeroed) • Input power: o Operating range: 9 to 24 V o Current during active sampling: ▪ LISST-200X: 100 mA @ 12V ▪ LISST-Black: 220 mA @ 12V ▪ LISST-HAB: 180 mA @ 12V o Current while waiting for user command: < 50 mA @ 12V o Current during sleep: < 3 mA • Dimensions: 10.03 cm (3.95”) diameter x 63.9 cm (26.18”) L • Weight: 5.4 kg (11.8 lbs.) in air, 1.7 kg (3.8 lbs.) in water • Depth rating: 600 m

Appendix B: Particle Size Bins

78 LISST-200X User’s Manual Appendix B: Particle Size Bins There are 36 size ranges logarithmically placed from 1.00 – 500 microns in diameter. The upper size in each bin is approximately 1.18 times the lower, with the exception of bin 1. The table below shows the lower and upper limit of each size bin in microns, together with the median size (also in microns) for each size bin. The sizes are the same for both Spherical and Randomly Shaped inversions. Size bin # Lower Upper Median 1 1.00 1.48 1.21 2 1.48 1.74 1.60 3 1.74 2.05 1.89 4 2.05 2.42 2.23 5 2.42 2.86 2.63 6 2.86 3.38 3.11 7 3.38 3.98 3.67 8 3.98 4.70 4.33 9 4.70 5.55 5.11 10 5.55 6.55 6.03 11 6.55 7.72 7.11 12 7.72 9.12 8.39 13 9.12 10.8 9.90 14 10.8 12.7 11.7 15 12.7 15.0 13.8 16 15.0 17.7 16.3 17 17.7 20.9 19.2 18 20.9 24.6 22.7 19 24.6 29.1 26.7 20 29.1 34.3 31.6 21 34.3 40.5 37.2 22 40.5 47.7 43.9 23 47.7 56.3 51.9 24 56.3 66.5 61.2 25 66.5 78.4 72.2 26 78.4 92.6 85.2 27 92.6 109 101 28 109 129 119 29 129 152 140 30 152 180 165 31 180 212 195 32 212 250 230 33 250 297 273 34 297 354 324 35 354 420 386 36 420 500 459

LISST-200X User’s Manual

79 Appendix C: Data File Formats Appendix C: Data File Formats File Types and Extensions Used by the LISST-200X. Extension Description Format Comments .RBN Raw Data Binary Raw data file produced by the LISST-200X. In addition to raw scattering data, RBN files contain a complete description of the instrument configuration, its most recent background scattering, and the factory background. .CSV Processed Data ASCII Comma-delimited file containing size distributions and meta data. Details on page 80. For information about generating these files in the LISST-200X, see Autonomous Real-time Data Processing on page 52. Secondary files not used for standard processing .BGT Background File ASCII Comma-delimited file containing the raw ring values from a single background measurement (1 column, 59 rows) .RTX Raw Data ASCII Comma delimited ASCII representation of raw data (59 columns, 1 measurement per row). NOTE: Sequoia software does not process RTX files. They are only for troubleshooting or special applications.

Appendix C: Data File Formats

80 LISST-200X User’s Manual Processed Data Format The values in the processed data file (.CSV extension) are stored in the order shown below. Each sample is stored in one row. Column # Parameter 1:36 Volume concentration for size class 1 through 36 [µL/L] 37 Laser transmission Sensor [mW] 38 Supply voltage in [V] 39 External analog input 1 [V] (fluorometer 1 in LISST-HAB & LISST-Black) 40 Laser Reference sensor [mW] 41 Depth in [m of sea water] 42 Temperature [C] 43 Year 44 Month 45 Day 46 Hour 47 Minute 48 Second 49 External analog input 2 [V] (fluorometer 2 in LISST-HAB & LISST-Black) 50 Mean Diameter [μm] (calculated from fully processed size distribution) 51 Total Volume Concentration [PPM] (calculated from fully processed size distribution) 52 Relative Humidity [%] 53 Accelerometer X [not presently calibrated or used] 54 Accelerometer Y [not presently calibrated or used] 55 Accelerometer Z [not presently calibrated or used] 56 Raw pressure [most significant bit] 57 Raw pressure [least significant 16 bits] 58 Ambient Light [counts – not calibrated] 59 External analog input 3 [V] (fluorometer 3 in LISST-HAB & LISST-Black) 60 Computed optical transmission over path [dimensionless] 61 Beam-attenuation (c) [m-1].

LISST-200X User’s Manual

81 Appendix C: Data File Formats ASCII Raw Data Format The values in the raw ASCII data file (.RTX extension) are stored in the order shown below. Each sample is stored in one row. RTX files are only useful for troubleshooting or special applications. Sequoia does not supply software for processing them. Column # Parameter 1:36 Raw ring values [counts] 37 Laser transmission Sensor [counts] 38 Supply voltage [0.01 V/count] 39 External Analog input 1 [10-4 V/count] (fluorometer 1 in LISST-HAB & LISST- Black) 40 Laser Reference [counts] 41 Depth [0.01 m/count + 10 m] 42 Temperature [1 m°C/count + 5 °C] 43 Year 44 Month 45 Day of month 46 Hour 47 Minute 48 Second 49 External Analog input 2 [10-4 V/count] (fluorometer 2 in LISST-HAB & LISST- Black) 50 Sauter Mean Diameter [0.1 µm/count] (estimated) 51 Total Volume Concentration [0.1 PPM/count] (estimated) 52 Relative Humidity [%] 53 Accelerometer X [counts] [not presently used] 54 Accelerometer Y [counts] [not presently used] 55 Accelerometer Z [counts] [not presently used] 56 Raw pressure [most significant bit] 57 Raw pressure [least significant 16 bits] 58 Ambient light [counts] 59 External Analog input 3 [10-4 V/count] (fluorometer 3 in LISST-HAB & LISST- Black)

Appendix D: MATLAB Data Processing

82 LISST-200X User’s Manual Appendix D: MATLAB Data Processing Two Matlab functions for processing data are provided on the instrument USB card. They are also available from the ‘Software and Downloads’ tab on the LISST-200X webpage. The processing is split into two steps: (1) reading in and applying corrections to raw data files (.RBN) and (2) inverting the corrected scattering to a particle size distribution. The same results can be achieved by processing data files using the LISST-200X software. However, the following functions allow Matlab users to write their own processing and plotting code. The first function is used to read in raw data from an RBN file. The function will import the data, apply the necessary corrections, and return the corrected data in a structure. The syntax is as follows: ‘Datafile’ is the path to a binary .RBN file downloaded from a LISST-200X. Optionally, you may specify a different clean water background file (.BGT) as a second argument. The data will then be corrected using the specified background file instead of the background contained in the RBN file. ‘RBNdata’ is a structure with the following fields: Field Description cscat Corrected scattering date Timestamp in Matlab datenum transmission Optical transmission depth Depth in meters temperature Temperature in degrees Celsius estMeanDiameter Estimated Sauter mean diameter (µm) estTotalConc Estimated total concentration (µL/L) Lp Transmitted laser power (mW) Lref Laser power reference (mW) analog1 Analog input 1 (V) analog2 Analog input 2 (V) analog3 Analog input 3 (V) RBNdata = getscat_L200X(datafile) RBNdata = getscat_L200X(datafile,backgroundFile)

LISST-200X User’s Manual

83 Appendix D: MATLAB Data Processing supplyVolts Supply voltage (V) humidity Internal instrument relative humidity (%) accelXYZ Accelerometer X, Y, and Z (not presently calibrated or used) raw Raw data as it appears in the RBN file factory_bkgrd The factory background (corrections applied to aux data) bkgrd User collected background (corrections applied to aux data) ambientLight Counts of ambient light removed from ring values config Structure containing various instrument information dcal Ring area coefficients Ta Vector to convert cscat to estimated total area concentration Tv Vector to convert cscat to estimated total volume concentration The second function inverts the corrected scattering to a particle size distribution. The syntax is as follows: The first argument is the corrected scattering (cscat) from the structure returned from ‘getscat_L200X.’ The following three argument are set to zero or one. Random – If set to 1, the randomly shaped particle matrix is used to invert the data. If set to zero, the spherical particle matrix is used. See Appendix G: Spherical and Random Shape Particle Models for more information. Sharpen – If set to 1, the function checks the width of the size distribution and increases the number iterations if the size distribution is wide (recommended). ShowProgressBar – If set to 1, a progress bar will display the processing status. The function will return the volume distribution in µL/L (vd) and the midpoint of each size bin in microns (dias). [vd, dias] = invert_L200X(Cscat,Random,Sharpen,ShowProgressBar)

Appendix E: Connectors

84 LISST-200X User’s Manual Appendix E: Connectors IMPORTANT NOTE In September 2019, starting with serial number 2116, LISST-200X instruments and cables use connectors manufactured by SubConn. Previous instruments have connectors from Teledyne Impulse. Connectors from the two manufacturers are very similar, but are not guaranteed to be compatible. For reliable connections, only mate SubConn cables with SubConn bulkhead connectors. Sequoia’s warranty does not cover any problems arising from mismatched connectors. To Identify the connector type, note: 1. SubConn bulkhead connectors have brass, golden-colored bodies. They may develop a green tint after exposure to sea water. 2. SubConn cables and dummy plugs have red locking sleeves. 3. Impulse bulkhead connectors have stainless steel, grey-colored bodies. 4. Impulse cables and plugs have black locking sleeves. 5. The name of the manufacturer is embossed on the rubber body of each connector. SubConn (as of September 2019) Impulse (older instruments)

LISST-200X User’s Manual

85 Appendix E: Connectors The LISST-200X has 3 separate underwater connectors: A 5-pin, a 6-pin, and a 3-pin connector. The photograph shows the placement of each connector. The following text describes the detailed wiring for each connector. Note that in new instruments starting in September 2019, these are SubConn connectors with brass bodies. LISST-200X Connector Endcap Small External Battery Housing Large External Battery Housing

Appendix E: Connectors

86 LISST-200X User’s Manual Communications and Power Connector (5 pin connector) Bulkhead connector: SubConn MCBH5M Mating cable connector: SubConn MCIL5F Connector Pin # Use 1 Power/Serial Ground 2 External Power In (+9 to 24 V) 3 Power/Serial Ground 4 Serial Out (to DB-9 Pin 2) 5 Serial In (to DB-9 Pin 3) BioBlock Connector (3 pin connector) (see page 93 for BioBlock details) Bulkhead connector: SubConn MCBH3M Mating Cable Part Number: SubConn MCIL3F Connector Pin # Use 1 Ground 2 Signal Out 3 +6V Power Out

LISST-200X User’s Manual

87 Appendix E: Connectors Auxiliary I/O Connector (6 pin connector) Bulkhead connector: SubConn MCBH6M Mating Cable Part Number: SubConn MCIL6F NOTE: This connector has 3 possible configurations. The configuration in a particular instrument can be found by issuing the CONFIG command at the command prompt. Analog and Digital Configuration Connector Pin # Use 1 Ground 2 Digital In #1 3 Digital In #2 4 +12V out 5 Analog In (0 to 5V) 6 Ground Analog Input Configuration Connector Pin # Use 1 Ground 2* Analog In #3 (0 to 5V)* 3 Ground 4 +12V out 5 Analog In #1 (0 to 5V) 6 Analog In #2 (0 to 5V) * Analog input #3 is available in LISST-200Xs with revision F or later electronics. This is standard starting with serial number 2151, and available as a factory upgrade on earlier units. In earlier versions, this pin is not connected. Analog Output Configuration Connector Pin # Use 1 Ground (Power) 2 Analog Output: Mean Size 3 Ground (signal) 4 Analog Output: Total Concentration 5 Ground (signal) 6 Power In (9 to 24VDC)

Appendix E: Connectors

88 LISST-200X User’s Manual Large and Small External Battery Connectors (5 pin connectors) Bulkhead connector: SubConn MCBH5F Mating Cable Part Number: SubConn MCIL5M To LISST Connector (5-pin Female Bulkhead) Connector Pin # Use 1 Ground 2 Battery Power Out (+9V nom. 6-15V) 3 Ground 4 Serial Out (to DB-9 Pin 2) 5 Serial In (to DB-9 Pin 3) To PC Connector ( 5-pin Male Bulkhead) Connector Pin # Use 1 No Connection 2 No Connection 3 Ground 4 Serial Out (to DB-9 Pin 2) 5 Serial In (to DB-9 Pin 3)

LISST-200X User’s Manual

89 Appendix F: Particle Shape Models Appendix F: Particle Shape Models Particles of different shapes scatter light differently, so the mathematical model used for processing must be appropriate to the particles under observation. Two models are offered by Sequoia: the spherical model, and the random or irregular shape model. The resulting inversion of data will differ slightly for the two models. The spherical particle model performs the mathematical inversion of scattering data under the assumption that the particles that scattered light are all spheres. Light scattering by spheres of any size and refractive index is modeled by Mie theory. According to Mie theory, the angular scattering depends on the size of the spherical particle, and its refractive index relative to water. Sequoia employs the full Mie scattering model, without simplifications, for inversion of LISST measurement as a distribution of spheres. The Mie solution is a generalized solution to the scattering of light from spheres and is commonly used as the standard model by all laboratory laser diffraction instrument manufacturers. However, it is not completely accurate for non-spherical particles found in nature. Sequoia provides an alternate model that is derived empirically from natural, irregularly- shaped mineral particles, rather than assuming spheres. The exact details of how this scattering model was established are described in a paper by Agrawal et al. [Light scattering by random shaped particles and consequences on measuring suspended sediments by laser diffraction. Journal of Geophysical Research, Vol. 113, C04023, doi:10.10- 29/2007JC004403.], which can be downloaded from the library section on Sequoia’s website (www.SequoiaSci.com/library). The method is also briefly described in this article: http://www.sequoiasci.com/article/random-shaped-particles-lissts/. It is noteworthy that no other instrument manufacturer has a scattering model for irregular particles. Instead, they often use a Mie model with large imaginary refractive index, in effect, assuming the particles to be highly absorbing (i.e. black). This is an obvious fudge with consequences! Which model should you use? Sequoia’s software allows you to use both, then compare the results. But in general, the random model is best for any in-situ application. Spherical Random (Irregular) Best use Particles known to be predominantly spherical, e.g. analytical microspheres Appropriate for natural mineral particles Effect of misapplication If used with natural irregular particles, tends to invent fine particles. If used with spherical particles, tends to underestimate their size. Relation to other instruments and methods Usually produces the closest match to other laser diffraction particle sizers (although they may still differ due to index of refraction) Usually produces the best match to analysis by sieving.

Appendix G: LISST-Black & LISST-HAB

90 LISST-200X User’s Manual Appendix G: LISST-Black & LISST-HAB LISST-Black and LISST-HAB are integrated packages combining a LISST-200X with fluorometers optimized for detecting either petroleum (LISST-Black), or harmful algal blooms (LISST-HAB). The fluorometers are Turner Designs Cyclops-7F’s, mounted with special brackets and cabling made by Sequoia Scientific. The LISST-200X supplies power to the fluorometers, and digitizes their analog outputs. The digitized fluorometer outputs are included in each data sample produced by the LISST, at rates up to one per second. Fluorometer Output Data The fluorometer outputs are reported in volts, included in the standard .CSV files produce by the LISST-200X software. For format details, see Appendix C: Data File Formats. The outputs are not calibrated to an absolute standard. Firmware Configuration To check that the fluorometers will be powered whenever the LISST collects data, type the DS command at the LISST-200x:> prompt, and look for “External sensor power during logging: on”. If it is not on, send the command SENSORPOWER,3 . This will enable power to the fluorometers, with a 3-second warmup before the first sample. Power Consumption Because of additional power consumed by the fluorometers, the LISST-Black and -HAB cannot be reliably powered by a USB cable during sampling. A battery housing or other power source must be connected. Standard Fluorometer Sets Fluorometer target, Turner Designs part number Power during sampling (at 12V) 1 2 3 LISST-Black Refined fuels 2110-000-G Crude oil 2110-000-O Chlorophyll 2110-000-C 2.7 W LISST-HAB Phycocyanin 2110-000-P Phycoerythrin 2100-000-E Chlorophyll 2110-000-C 2.2 W

LISST-200X User’s Manual

91 Appendix G: LISST-Black & LISST-HAB LISST-Black/LISST-HAB Mounting instructions Hardware: Clamp Screws, 8-32 x ½” Socket Head Cap Screws, 316 Stainless Steel Sensor Hold-down Set Screws, 8-32 x 3/16” Set Screws, flat tip, 316 Stainless Steel Cover Screws, 6-32 x ¼” Flat Head Phillips Screws, 316 Stainless Steel Tools: 5/64-inch hex L-key 9/64-inch hex L-key #2 Phillips Screwdriver 1. Attach the Front Support to the LISST-200X with the Clamp Strap, cross nut and clamping screw. Position the tab on the Front Support so it fits flush against the endcap and aligns to the flat surface. Tighten the Clamp Screw. 2. Position the Mid Support on the LISST-200X so the back edge is positioned on the pressure case chamfer, as shown at right, and the holes for the fluorometers line up with the Front Support. Slightly tighten the Mid Support clamping screw so that the Mid Support is held in place but can be rotated as needed when inserting the sensors. 3. Insert all three Cyclops-7F Sensors. Note the order of the sensors. The labels shown in the figure correspond to their labels in the data file (see Appendix C: Data File Formats). When looking at the connector end of the sensors, sensor 1 will be on the left and sensor 3 will be on the right. Do not tighten the setscrews holding the sensors yet.

Appendix G: LISST-Black & LISST-HAB

92 LISST-200X User’s Manual 4. Rotate the Mid Support as needed to make the axis of the sensors parallel to the axis of the LISST-200X. Tighten the Mid Support clamp making sure that the Mid Support is pushed against the tapered part of the pressure case, as in step 2. The sensors should be free to rotate and move forward and backward. 5. Attach the Cable Support onto the LISST-200X. Tighten the Cable Support clamping screw slightly so that the Cable Support is held in place but free to rotate and slide along the case. 6. If the LISST-HAB/Black will be used with the LISST-200X Frame, place the Frame Clamp between the sensors and the LISST-200X pressure case before connecting the LISST-HAB/Black cable. Installing the Frame Clamp after assembly is difficult and not recommended. 7. Attach the LISST-HAB/Black cable to the sensors. a. The protrusion in the mold needs to be on the top side. b. Rotate the sensors to match the cable sockets and insert into the connector. c. Secure the cable locking sleeves. d. Slide the Cable Support towards the optics end so that cables are supported. Do not tighten the clamp yet. 8. Position the sensors so they extend approximately 1/8” (3mm) in front of the Front Support. Secure the sensors with the Hold-down Set Screws. 9. Place the Cover onto the assembly and adjust the position of the Cable Support so that the Cover can be easily installed. Insert the Cover Screws into the holes at the Front Support. Adjust the Cable Support position as need so that the holes in the Cover line up with the holes in the Cable Support and secure it with the Cover screws. Tighten the Cover and Cable Support clamp screws. 10. Connect the cable to the 6-pin connector on the LISST-200X. 11. Secure the cable to the handle with cable ties.

Other Measuring Instruments Manuals