Sequoia LISST-200X Operation & User’s Manual
Text of the manual
Particle Size Analyzer
Including
LISST-HAB & LISST-Black
User’s Manual
Version 2.2
December, 2021
2700 Richards Road, Suite 107
Bellevue, WA 98005
USA
Telephone: (+1) 855-753-3313
Store Software
USB Card Here
LISST-200X User’s Manual
Page v
Welcome to the LISST-200X Particle Size Analyzer
Using this manual
I. LISST-200X Introduction & Quick Start provides an
overview of the LISST-200X’s operating principles and basic
functions.
II. LISST-200X Operation Details provides details and step-by-
step instructions for using and caring for the instrument.
The Appendices contain specialized details, specifications, and
information about options.
Warranty
See the product warranty at:
http://sequoiasci.com/support/warranty
Technical assistance
For technical assistance please contact your local Distributor or
Sequoia. Please include the instrument serial number with any
correspondence.
Sequoia Scientific, Inc. contact information:
Telephone: (+1) 855-753-3313
Email: [email protected]
LISST-200X User’s Manual
Page vii
Table of Contents
I.
LISST-200X INTRODUCTION & QUICK START ................................................................... 1
A.
LASER DIFFRACTION AND THE LISST-200X ............................................................................ 1
B.
LISST-200X GENERAL DESCRIPTION .................................................................................... 4
C.
LISST-200X QUICK START TUTORIAL .................................................................................... 7
II.
LISST-200X OPERATION DETAILS .................................................................................... 26
A.
STEP BY STEP PROCEDURES ............................................................................................... 27
1.
Installing LISST-200X Software .................................................................................... 27
2.
Establishing Communication with the LISST-200X ...................................................... 28
3.
Saving and Evaluating Clean Water Backgrounds ....................................................... 29
4.
Configuring Data Collection .......................................................................................... 32
5.
Using the Small External Battery Housing .................................................................... 36
6.
Using the Large External Battery Housing ................................................................... 39
7.
Offloading and Deleting Data Files from Internal Memory ........................................... 42
8.
Processing a Single Raw Data File .............................................................................. 43
9.
Batch Processing Multiple Raw Data Files ................................................................... 45
10.
View Processed Data File ........................................................................................ 47
11.
Data Quality Control ................................................................................................. 48
12.
Real-Time Data Processing with LISST-200X Software .......................................... 50
13.
Autonomous Real-time Data Processing ................................................................. 52
14.
Collecting Data from an External Analog Sensor ..................................................... 55
15.
Configuring the LISST-200X as a Sensor for a CTD ............................................... 56
B.
START AND STOP CONDITIONS ............................................................................................. 57
C.
INSTRUMENT COMMUNICATION ............................................................................................. 59
D.
LISST-200X COMMAND SUMMARY ...................................................................................... 60
E.
LISST-200X COMMAND DETAILS ......................................................................................... 61
1.
Display Commands ....................................................................................................... 61
2.
Setup Commands ......................................................................................................... 64
3.
Acquisition/Action commands ....................................................................................... 69
F.
PERFORMANCE OPTIMIZATION ............................................................................................. 71
G.
INSTRUMENT MOUNTING, DEPLOYMENT ORIENTATION, STORAGE AND MAINTENANCE ............. 74
APPENDIX A: TECHNICAL SPECIFICATIONS ........................................................................... 76
APPENDIX B: PARTICLE SIZE BINS .......................................................................................... 78
APPENDIX C: DATA FILE FORMATS ......................................................................................... 79
APPENDIX D: MATLAB DATA PROCESSING ............................................................................ 82
APPENDIX E: CONNECTORS ..................................................................................................... 84
APPENDIX F: PARTICLE SHAPE MODELS ............................................................................... 89
APPENDIX G: LISST-BLACK & LISST-HAB ............................................................................... 90
APPENDIX H: BIOBLOCK ANTI-FOULING WIPER .................................................................... 93
APPENDIX I: PATH REDUCTION MODULE ................................................................................ 96
APPENDIX J: TROUBLESHOOTING & SUPPORT ................................................................... 103
REVISION HISTORY ................................................................................................................... 104
LISST-200X User’s Manual
1
LISST-200X Introduction & Quick Start
I. LISST-200X Introduction & Quick Start
A. Laser Diffraction and the LISST-200X
Measurement
Principle and
Optics
Particle sizing by laser diffraction is currently the most widely prevalent
method in research and industry. It is a multi-parameter measurement.
Just as a set of 3 equations can be solved for 3 unknowns, a
measurement of light scattering at multiple angles can be solved for
equally as many concentrations in different sizes. For example, the
LISST-200X measures scattering into 36 angles; consequently, one
obtains concentrations in 36 size classes of particles. This is called the
particle size distribution (PSD).
To measure scattering at multiple angles, the LISST-200X sends a
collimated beam of light into the water. Laser light scattering at an
angle from the beam arrives, through a daylight rejection filter, at the
focal plane of a receive lens at the same angle from lens axis (see
figure). Distance from lens axis in the lens focal plane corresponds to
scattering angle. The photodetectors in the LISST series instruments
are a series of silicon rings spanning 60-degree arcs. Each ring covers
a small range of scattering angles. Rings are used to achieve stability
of inversion – converting of set of multi-angle scattering measurements
to the PSD.
Notice that the laser beam itself is focused by the receive lens and
passes through a small aperture in the ring detector, centered on the
rings. This beam is sensed by a photodiode placed behind the ring
detector. This is the transmission sensor. As light is removed from the
laser beam by scattering, the beam is attenuated, i.e. the light
transmitted through water is reduced in intensity. A similar attenuation
also affects the light that is scattered by particles and sensed by ring
detectors. Thus, the transmission sensor provides a vital measurement
to de-attenuate the measured scattered light.
LISST-200X Introduction & Quick Start
2
LISST-200X User’s Manual
Clean Water
Backgrounds
Scattering originates not only from particles in the water, but also from
the water itself, and from the windows and internal optics of the LISST-
200X. This non-particle scattering constitutes a Background. (Formerly
this was called zscat, meaning scattering by zero concentration of
scatterers). Accurate measurements require subtracting the
background scattering from the total signal seen by ring detectors.
The background scattering typically changes gradually throughout the
life of the instrument, due to small changes in the laser and optics, and
wear on the exposed window surfaces. It is also very sensitive to
cleanliness of the windows. Therefore, high-quality measurements
require careful background measurements by the user. A factory
background acts as a reference for comparison. The LISST-200X
stores its factory background data in its memory, as well as the user’s
acquired background data.
Background
Quality Control
Because the background measurement is so important, the LISST-
200X firmware and software include procedures for measuring and
evaluating the background scattering. The software checks for signs of
problems such as laser degradation, optics misalignment,
contamination of data by thermal microstructure in the water used for
background, bubbles or particles in the supposedly clean water,
scratches on windows etc.
For details on the background measurement procedure, see Saving
and Evaluating Clean Water Backgrounds on page 29.
Ambient Light
Rejection
The laser diffraction method requires that the light arriving on the
detector be due entirely to scattering of light that originates with the
laser beam. Light from ambient sunlight can distort the results. The
LISST physically shields the detector from much ambient light, but not
all, so daytime measurements at shallow depths can include some
sunlight. However, the LISST-200X (starting with firmware version 1.4,
introduced in May, 2017) can measure the ambient light independently
of scattered light, and remove its effects. This Ambient Light Rejection
(ALR) function is enabled by default. Also see section 0, page 71.
Quick Estimates
of Total
Concentration
and Mean Size
Converting the LISST-200X’s measurements of scattering to PSD
requires extensive computations that are time-consuming for a small
microprocessor. However, a much simpler computation, easily done in
real time, provides estimates of the total particle volume concentration
and mean particle size. These quantities are not as accurate as the
fully-processed PSD, but are useful for quick characterization of water
conditions. For example, if used in conjunction with a CTD package in
profiling applications, the CTD software could display the concentration
and mean size as a depth profile, revealing any vertical structure in the
particle distribution. (Note that real-time interface to a CTD requires
proper configuration of the LISST-200X’s auxiliary connector; see
Configuring the LISST-200X as a Sensor for a CTD on page 56.)
LISST-200X User’s Manual
3
LISST-200X Introduction & Quick Start
The quick estimates are based on weighted sum of the net scattered
light. One set of weight factors yields the total volume concentration;
the other yields the area concentration of particles. The volume/area
ratio provides mean diameter, also known as the Sauter Mean
Diameter (SMD). Notably, SMD can be quite different from D50 in broad
or multi-modal size distribution situations. The SMD output is set to
zero in very clear waters, i.e. when transmission is above 98%,
because the SMD can become erratic when the relative scattering
signal is small. [For an explanation of this method of SMD derivation,
see: Shaped Focal Plane Detectors for Particle Concentration and
Mean Size Observations; Agrawal, Y.C. and O.A. Mikkelsen, (2009),
Optics Express, v 17, n 25, pp 23066-23077].
Calculating the
full PSD in real
time
Newer LISST-200X’s, starting with serial number 2131 (and older units
with upgraded circuitry), contain more powerful processors that can do
full PSD processing in real time. The processed data can be stored in
the instrument’s memory, or transmitted through its serial port. This is
especially useful for unattended deployments, such as buoys or remote
moorings, where it is not possible to connect a Windows computer to
the instrument for data offloads. For full details, see Autonomous Real-
time Data Processing on page 52.
Particle Shape
Models –
Spheres or
Irregular Shape
The multi-angle scattering can be interpreted via inversion as arising
from spherical particles, or from irregularly shaped particles. Provided
software gives you the choice and the resulting PSD files are named
differently to distinguish them. As to which particle model to use, we
suggest that when working with natural waters, use the irregularly
shaped model. The spherical model is appropriate only in exceptional
circumstances. For more details see Appendix F: Particle Shape
Models on page 89.
LISST-200X Introduction & Quick Start
4
LISST-200X User’s Manual
B. LISST-200X General Description
Instrument
Description
The primary measurements delivered by the LISST-200X are the small-
angle scattering properties of particles in water, the laser optical
transmission, depth, and temperature. The LISST-200X is capable of
autonomous operation when used with external batteries. Windows
software is provided to program the instrument for a specific sampling
schedule. An auxiliary connector can be configured for recording data
from an external devices such as fluorometer, or for producing
summary data (mean size and total concentration) as analog voltages.
For details of the possible configurations of the Auxiliary Port, see
Appendix E: Connectors on page 84.
After recovery of the instrument, small-angle scattering data are off-
loaded from the instrument and subsequently inverted mathematically
on a computer to produce the particle size distribution. The inversion
function is included in the Windows software. For MATLAB users,
inversion scripts are available for download from Sequoia’s website:
www.SequoiaSci.com.
The LISST-200X instrument is a laser diffraction device. It consists of
optics for producing a collimated laser beam, a specially constructed
detector array, electronics for signal pre-amplification and processing,
data storage and scheduling computer, and an external battery system.
The principal measurement—angular scattering distribution— is
obtained over 36 ring-detectors whose radii increase logarithmically
from 102 to 20,000 microns. The detector is placed in the focal plane of
the receiving lens. The rings cover an angular range from 0.00085 to
0.34 radians. This angular range corresponds to size ranges from 1.00
to 500 µm. See Appendix B: Particle Size Bins on page 78 for more
information.
The LISST-200X consists of the following parts: a solid-state diode
laser operating at 670 nm wavelength and fiber-optically connected to a
laser beam collimating system, a beam manipulation and orienting
system, a two pressure windows, a scattered-light receiving lens, the
custom designed 36-ring detector, preamplifier electronics, a ring-
selecting multiplexer circuitry, and a digital controller with data memory.
All these components are inside the black-anodized pressure housing.
The LISST-200X does not have internal batteries. Its watertight
housing need not be opened for any normal use.
LISST-200X User’s Manual
5
LISST-200X Introduction & Quick Start
Data Storage
and Interface
The LISST-200X includes high-capacity data logging and storage. The
logging functions are programmed via the provided software. It can be
programmed with different start and stop conditions as well as different
sampling rates and average durations. The data logger stores the data
in non-volatile Compact Flash memory which can be later downloaded
and processed into size distributions and concentration using the
provided software.
The data logger will also accept commands via the RS232 interfaces.
These commands can be used to program the instrument or to
exchange data with another instrument.
Depth and
Temperature
In addition to measuring the particle size and concentration, the LISST-
200X also has depth and temperature sensors. The depth sensor has
a 1000 psi full-scale range and is calibrated to 600 meters depth. A
stainless-steel fitting and tube filled with silicone oil protect the pressure
transducer from contamination and corrosion. The temperature is
measured using a high precision thermistor imbedded into a stainless
steel probe on the Connector end cap. Both values are stored
automatically in the LISST-200X data file.
External I/O Port
The LISST-200X also has an auxiliary 6-pin connector for interfacing to
other devices This port can be configured to produce analog outputs, or
to receive analog inputs.
The Analog Output Configuration allows the LISST-200X to receive
power from a CTD, and to send two analog voltages indicating the
mean size and total concentration to the CTD, while the detailed size
distribution data is stored internally for later downloading. For more
details about this application, see Configuring the LISST-200X as a
Sensor for a CTD on 56.
The Analog Input Configuration enables operation with additional
sensors such as fluorometers and turbidity sensors. It provides 12V
regulated power, and accepts two or three analog inputs (0 to 5 V) from
sensors. The digitized analog values are stored in the same files as the
LISST-200X’s particle size data. This configuration also supports the
instrument packages described in Appendix G: LISST-Black & LISST-
HAB on page 90.
Battery-powered
operation
The LISST-200X does not contain internal batteries (except a small
battery to maintain the real-time clock), but can be powered from Small
or Large Battery Housings (provided with the instrument) or an external
instrument such as a CTD.
The standard Small Battery Housing supplied with the LISST-200X
uses two NiMH rechargeable D-cell batteries to provide about 12 hours
of continuous sampling.
The Large Battery Housing holds 16 standard alkaline D-cells. For
profiling applications, where the instrument is sampling continuously,
the Large Battery Housing has about 200 hours of sampling time or 200
days of stand-by. Powering down the instrument between samples can
LISST-200X Introduction & Quick Start
6
LISST-200X User’s Manual
greatly extend the deployment times. The LISST-200X software
contains a battery life calculator on the Configuration page. This feature
provides rough estimates of the battery run time for a given sampling
program.
USB Power and
Data connection
For laboratory or tethered usage, power can be supplied through the
communications connector on the endcap. A 2-meter USB cable is
provided that will power the instrument from the computers USB port.
No additional power is required to operate the instrument.
Cables up to 50 meters can be provided to supply external power and
communication with the instrument. This can allow real-time
observation of the size distributions.
Included
Accessories
The instrument is shipped pre-aligned and tested. A small chamber is
provided for obtaining measurements of background scattered light
from optical surfaces. This background is subtracted from actual
particle scattering measurements to obtain the true particulate
scattering. Additionally, small tools used to open endcaps, spare
batteries, and communication cable is supplied so that a user need only
provide a PC running Windows.
For extended laboratory applications various chambers are available
from Sequoia.
Software is provided to communicate with the instrument, schedule an
experiment, offload the data, and invert the measurements to obtain
particle size distribution and volume concentration. For laboratory use
or for monitoring the progress of an experiment, the software can be
used for real-time processing.
LISST-200X User’s Manual
7
LISST-200X Introduction & Quick Start
C. LISST-200X Quick Start Tutorial
This section gives step by step instructions to unpack your LISST-
200X, load software, and acquire data in the lab. For more detailed
instructions on specific steps, background on how the instrument
works, or specific technical information such as cable pinouts please
refer to the full User’s manual following this Quick Start Tutorial.
Contents of
Shipping Case
Let’s assume that you are opening the LISST-200X shipping case for
the first time. Inside you will find the following:
• User’s Manual,
• USB memory card (credit card size) with the software,
• LISST-200X instrument,
• Plastic Instrument stands,
• Small Volume Test Chamber,
• USB Communications cable,
• Insulated stainless steel clamps,
• Small Battery Housing (without batteries),
• Larger Battery Housing with alkaline batteries installed,
• LISST-200X-to-Battery-Housing cable,
• NiMH batteries and charger.
Step 1: Remove
Instrument from
Shipping Case.
Start by removing the white plastic instrument stands and set them on
a flat working surface. Remove the LISST-200X from the case and set
it on the stands. The LISST-200X has two distinct ends that we will
refer to as the Optics endcap and the Connector endcap.
LISST-200X Introduction & Quick Start
8
LISST-200X User’s Manual
Optics Endcap
The optics endcap contains the optical windows that the laser beam
passes through to make a measurement. The internal optics and laser
electronics are mounted to the inside of this endcap.
Connector
Endcap
The connector endcap has three underwater connectors that are used
for communication, external power, and connecting to optional
accessories or instruments:
3-pin connector: Used for BioBlock, an anti-biofouling accessory
5-pin connector: Used for serial communication and external power.
6-pin connector: Used for analog in/out and digital in/out signals.
See Appendix E: Connectors on page 84 for a full description of the
wiring of the underwater connectors and mating cables.
A temperature sensor is located between the 3-pin and 6-pin
connectors. In the center of the endcap is an LED that blinks when the
instrument is sampling. The stainless steel fitting with small tube is the
port for the depth sensor. This fitting can be removed to allow the port
to be cleaned should sediment or salt deposit build up.
LISST-200X User’s Manual
9
LISST-200X Introduction & Quick Start
Also on the Connector endcap is a white plastic lever. This
mechanical lever has a strong magnet embedded in the plastic. This
magnet can be used to trigger a digital switch inside the instrument
which can be programmed to start and stop the LISST-200X sampling.
The final item on the endcap is the zinc anode. The sacrificial zinc
anode protects the instrument from corrosion during long periods of
time in salt water.
Step 2: Check
for Clean
Windows
At this time, check the optical windows to make sure that they are
clean. There are two windows: The receive window (or large window)
and the transmit window (or small window). Both window need to be
very clean in order to get good measurements.
The best way to check the windows is by using a flashlight. By shining
light from one side and viewing from the other the surface of the
windows can be easily checked for cleanliness.
If there is dirt or fingerprints on the windows clean them first by rinsing
them with lukewarm water and a mild soap solution (e.g. mild hand
soap, liquid dish soap) and then rinsing off all soap residue with clean,
particle free water such as deionized water, distilled water or bottled
drinking water. The windows can also be wiped clean with a soft cloth
(e.g. a lens cloth) or glass cleaner. It is not recommended to use
stronger solvents, such as acetone or toluene. Also, do not use any
abrasive cleaners or wipes. Treat the windows as you would an
expensive camera lens.
Step 3: Attach
Communication
and Power Cable
Remove the Communications cable from the plastic accessory case
within the shipping case. It is the 3 meter cable with the USB
connector on one end and the 5-pin underwater connector on the
other. Remove the underwater cap from the Communications
connector. The connectors will all look similar when the protective cap
is installed. The Communication connector is the only 5-pin connector
and is located next to the stainless steel pressure sensor fitting. After
removing the cap install the cable making sure that proper alignment
of the cable is maintained, so that the connector pins are not bent.
Plug the USB cable into the computer. Please note that USB drivers
may automatically install the first time the USB cable is plugged into
LISST-200X Introduction & Quick Start
10
LISST-200X User’s Manual
the computer. If the driver installed correctly, you should see the
green light on the USB cable blinking. For more information on
establishing and troubleshooting communication with the LISST-200X
see page 27.
Step 4: Install
the Horizontal
test chamber
Remove the Small Volume Horizontal test chamber from the
Accessory Case if it is not already installed on to the instrument. This
one-piece chamber is designed to slip between the optical windows of
the instrument such that the space between the windows can be filled
with water for testing or calibration.
We can now fill the chamber with clean, particle-free water. Sequoia
uses steam-distilled bottled water filtered through a 0.2 micron filter.
Tap water may contain too many particles and may also contain
dissolved gas that can release and form small bubbles on the optical
surfaces. Pressurized filtered water can also contain dissolved gas.
Often the best source of clean water is bottled drinking water.
Step 5: Install
LISST-200X
Software
At this point the instrument is ready to go. We now need to install the
software that is required for operation of the instrument. A USB
memory card the size of a credit card is included with each instrument.
In addition to the communication and processing program the disk also
contains digital copies of this manual and other support files. Insert the
memory card into a USB port on your PC to install the software. You
must install the software on a computer running Windows XP or later
(it is not compatible with Mac or Linux operating systems).
On the memory card you will find the ‘LISST-200X_Installer.exe.’
Double click the installer executable to begin installing the software.
Follow the onscreen instructions and the installer will transfer the
necessary files to your computer and place a shortcut on your desktop
and start menu. Do not remove the memory card from your computer
until the installation is fully completed.
LISST-200X User’s Manual
11
LISST-200X Introduction & Quick Start
Please note the following with respect to the LISST-200X software:
• On a 64-bit Windows PC, the software will be installed in the
following folder: C:\Program Files (x86)\Sequoia\LISST200X,
while on 32-bit Windows machines, the install folder will be
C:\Program Files\Sequoia\LISST200X
• The LISST-200X software WILL NOT work with the older
LISST-100 or LISST-100X instruments. It is NOT possible to
use the program to offload data from a LISST-100 or LISST-
100X instrument, only a LISST-200X will work with this
software.
Step 6: Start
LISST-200X
Application
Start the LISST-200X software by selecting the shortcut the installer
placed on your desktop.
Step 7:
Establish
Communication
with the LISST-
200X
The LISST-200X does not have an internal battery. Power is supplied
to the instrument via a battery housing or USB cable. Please note that
the LISST-200X USB cable has a USB-to-serial convertor and special
power supply that converts the +5V to +11V to run the instrument built
into cable. When using this cable no other power source is required.
Power can also be provided by the Small and Large External Battery
Housings or another external power source such as a CTD. When
using an external battery the serial communication passes through the
battery housing. However, the power from the USB is not used power
the instrument. Power will be drawn from the battery only.
The LISST-200X software will automatically detect and connect to a
LISST-200X instrument if it is connected to the computer.
If an instrument is found, it will display ‘Connected’ in the lower left
corner.
LISST-200X Introduction & Quick Start
12
LISST-200X User’s Manual
If your instrument is connected to the computer, but not found by the
software, click on the ‘No Instrument Connected’ message to see list
of options. Click on the ‘Communication Help’ button for
troubleshooting help.
Step 8: Open
Terminal
window
The LISST-200X software has a terminal window that allows the user
to communicate directly with the instrument using a set of commands.
To open the Terminal window click on the connected instrument
message in the lower left, then click on the wrench icon and select
‘Open Terminal’
A command can be typed into the box in the lower section of the
window. When the Enter key is pressed the command will be sent to
the instrument. A list of the commands that can be sent to the LISST-
200X is displayed when the terminal opens or they can also be found
in the section LISST-200X Command Summary on page 62.
LISST-200X User’s Manual
13
LISST-200X Introduction & Quick Start
Step 9:
Acquiring Clean
Water
Background
Measurement
The background scattering measurement is critical to good instrument
performance. It is especially critical for clean water applications where
the optical transmission is greater than 90% over the 2.5cm path. The
background scattering will also check the overall health of the
instrument. It will verify that all of the systems are functioning and that
the optics is still in alignment. The current background will be acquired
and displayed relative to the factory background scattering and the
most recently saved background that is stored on the instrument. The
image below shows an example of this display. It is opened by
selecting the Collect/View Background button.
The factory background file will be automatically acquired from the
instrument and displayed as a red line on the screen. The ‘Current
Background’ is the last background that was collected and saved on
the instrument (the date of collection is displayed in the legend). Tips
for collecting a background are also displayed on the screen. Pressing
the ‘Background Help’ button will open a PDF with additional
information about background measurements. When the Collect New
Background button is pressed, 20 samples will be collecting. The
average of these measurements will be displayed as a black line on
the screen.
The graph shows the value of the 36 light scattering detectors. The
red line is the factory values, the black line is the average of all 20
measurements. If the background is close to factory levels the
message displayed will indicate a pass or acceptable background.
If the water or windows are not clean or if there is a problem with the
instrument, error messages and suggested actions will be displayed.
LISST-200X Introduction & Quick Start
14
LISST-200X User’s Manual
Dirty water or windows will generally cause higher values across the
middle rings. Large bubbles or particles in the water can cause higher
values on the inner rings or left hand side of the display. High values
on the inner rings combined with a lower transmitted laser power value
can also be an indication of optical misalignment.
If needed you can update the background, such as after cleaning the
windows or replacing the water, by pressing the Collect New
Background button again.
In general, the lower the background values the better the
background. The goal is to get values that are at the same values as
factory line. However, as the instrument is used the background may
increase due to small scratches and slight alignment changes. It may
not be possible to get the background down to the original factory
values.
If the values are acceptable the values can be saved both onboard the
instrument and to a file on your computer. The LISST-200X handles
the background files differently than the LISST-100X. The background
file is stored on the instrument and is saved as part of every data file
that is recorded. When processing a data file the background will be
automatically extracted and used during processing. Saving, tracking
and selecting a background file to use for processing is no longer
required. The background stored on the instrument will continue to be
saved in new data files until a new background is recorded.
If you wish to store a background file on your computer as well as on
the instrument, click the ‘Save as .BGT file’ after you have accepted
and saved the background. It is not necessary to store the background
on your computer, however, these saved backgrounds can be used
during processing instead of the background saved in the data file.
Step 10:
Configuring
Instrument for
Deployment
The Configure Instrument window is used to configure the deployment
parameters. To open the window, choose Configure Instrument from
the home page.
A window similar to the one shown below will appear. The window has
four accordion style tabs: Sampling Schedule, Start Condition, Stop
Condition, and Accessories.
LISST-200X User’s Manual
15
LISST-200X Introduction & Quick Start
With the accordion tabs collapsed, the page shows a summary of the
instrument’s current configuration.
Step 11: Setting
Operating Mode
By selecting the Sampling Schedule Tab at the top of the main window
the screen below appears. This screen is used to set the type of
sampling; Fixed Sample Rate or Burst. You can also select the
samples per average and sample rates on this screen.
Burst and Fixed
Rate Modes
The Burst and Fixed Sample Rate modes are used to save data to a
raw data file on board the instrument. The icons next to the various
values give a better understanding of their meaning. The software
automatically checks the values entered to make sure that there is
no conflict. For example, when a ‘Sample to be Average of’ value is
entered, the minimum sample interval is computed. If this value is
LISST-200X Introduction & Quick Start
16
LISST-200X User’s Manual
less than the minimum permitted the value will be changed to the
minimum and the text will turn red.
The LISST-200X measures internally at ~20 Hz, but the data cannot
be stored to the data logger at this rate. The individual measurements
are averaged into a sample, and it is this sample average that is
stored at a maximum sample rate of 1 Hz. For the example shown
above, the instrument is set to sample in the Fixed Sample Rate mode
at a 1 Hz rate with 10 measurements per average. This average is
obtained in 0.53 seconds.
Step 12: Setting
Start Conditions
After selecting the sampling schedule, the start and stop conditions
can be selected. Choose the Start Condition tab.
There are five options: Depth, Time, External Mechanical Switch,
External Digital Input, and Time Delay. Select the mode by clicking on
the button next to its label. Select the correct parameters as required.
For this example let’s select the External Mechanical Switch Start
Condition.
AutoStart
If the AutoStart check box is selected, the LISST-200X will start the
sampling program when power is applied to the instrument. NOTE: for
AutoStart to detect the application of power, all power sources
(including the USB cable) must first be disconnected for at least 6
seconds. If AutoStart is not selected the user must select the ‘Run
Sample Program’ button from the instrument status bar or send the
‘GO’ command to start the sampling programing after applying power.
Step 13: Setting
Stop Conditions
Similarly, the Stop conditions can also be selected. Click on the Stop
Condition Tab to open the Stop Condition window. The available stop
conditions are: Depth, Time. External Mechanical Switch, External
Digital Input, Fixed number of Samples, and Memory Full/Battery Low
LISST-200X User’s Manual
17
LISST-200X Introduction & Quick Start
For this example, choose the External Mechanical Switch as the Stop
condition.
Step 14:
Configure
Accessories
The accessories tab allows you to select which LISST-200X
accessories are connected to the instrument. The accessories must be
selected here for them to function properly. If there are no accessories,
select ‘None’.
Step 15: Upload
Configuration to
Instrument
After configuring the instrument settings, they must be uploaded to the
instrument. A message will be displayed on the top of the screen when
you have unsaved changes to the instrument configuration
LISST-200X Introduction & Quick Start
18
LISST-200X User’s Manual
Click on the ‘Upload to Instrument’ button to load the configuration
onto the LISST-200X. A confirmation message will be displayed when
the upload is finished.
Step 16: Start
Instrument
The instrument is now configured for deployment, however, it is not
yet running.
You MUST explicitly start the samping program you just
configured. If you do not do that, the LISST will never start
sampling or react to the start and stop conditions.
To start the sampling program. Click on the ‘Run Sample Program…’
button in the instrument status bar.
You will be asked to confirm your sampling program, start/stop
conditions, accessories, and background date.
LISST-200X User’s Manual
19
LISST-200X Introduction & Quick Start
After the sampling program is started, you will see the status bar
change to ‘Sample Program Running’ as shown below. At this point
the instrument is running and ready for deployment.
Note that even when the sampling program is running, sampling will
not necessarily start immediately. For example, if you have selected
the mechanical switch as the start condition, the instrument will only
react to the swtich after the sampling program has been started. Only
then will actual sampling begin. If you want sampling to begin
immediately, select a start condition of “Delay” with zero as the delay
time.
While waiting for the start condition the LED on the connector endcap
will double blink every few seconds to alert the user that the instrument
is running but not yet sampling. During sampling the LED will
illuminate during the actual measurement of an averaged sample. For
example, if the sampling is set for 1 Hz sampling with measurements
per average set to 20 then the LED will blink once per second. If
settings are set for a 30 second average every 15 minutes then the
LED will illuminate for 30 seconds every 15 minutes.
Step 17:
Collecting Data
For our example we can move the white lever on the endcap to the “1“
position. Sampling will start. Data collection will continue until the Stop
Conditions are met. In our case, until the switch lever is returned to the
“0” position.
If the Start and Stop Conditions are set to Depth, External Mechanical
Switch, or External Digital Input the program will return to checking for
the Start Condition. This will only be true if the Start and Stop
conditions match. For example, Depth Start and Depth Stop. For non-
matching Start and Stop Conditions, such as fixed number of samples
or Time Stop, the program will terminate and the instrument will go into
a low power sleep mode.
To stop a running program, or wake it from low power sleep mode, use
the ‘Stop / Wake Instrument’ button from the status bar.
Note that the above options will only be available if you started the
sample program from the software. If you plug in an instrument that is
already executing a sampling program, you will need to manually wake
the instrument by selecting the ‘Wake Instrument…’ button and
selecting the COM port the instrument is connected to.
LISST-200X Introduction & Quick Start
20
LISST-200X User’s Manual
Step 18:
Downloading
Data
The instrument has now stored data in the on-board memory card.
Select ‘Offload Or Erase Files’. A list of files will appear:
Choose the files to offload by clicking on them while holding down the
CTRL key. The Shift key can also be used to select a range of files.
Choose a location to offload the files to on the right side.
The files will be saved with names in the following format:
Ldddhhmm.RBN, where ddd is the day of the year, hh is the hour, and
mm is the minute that the file was first written to. As the data is
downloading a the status bars at the bottom will update. The data is
offloaded at 115K baud.
Step 19:
Processing Raw
Data
We now have the data transferred from the instrument to the PC. To
process the data file choose Process Data Files from the home page.
LISST-200X User’s Manual
21
LISST-200X Introduction & Quick Start
You will be prompted to select the raw data file to open (selecting
multiple files will open up the batch processing window). Raw data files
have the extension .RBN, and are offloaded directly from the LISST-
200X data logger. Below is list of the file types you can expect to see
when using the LISST-200X.
Extension
Discription
Format
.RBN
Raw Data
Binary
.RTX
Raw Data
ASCII
.CSV
Processed Data
ASCII
.BGT
Background File
ASCII
Every LISST-200X data file contains all the necessary information to
process the file. Therefore, the software will automatically determine
the instrument serial number, factory background, current background
and other instrument specific parameters.
A raw data file display will be generated for your selected data file.
The range of samples to process can be selected by moving the
sliders at the bottom of the display. This can be used to exclude data
from the beginning or the end of the file. The default selection is to
process the complete file.
LISST-200X Introduction & Quick Start
22
LISST-200X User’s Manual
Select the output directory and choose a base file name. Press the
Process File button to convert the raw file into processed size
distributions. For more details on the available options when this
window is open, see the detailed Instructions starting on page 43.
When the processing is complete a new tab will open, showing the
processed results. You can navigate back to the raw data tab if you’d
like to change the processing settings and process the data file again.
Step 20: Viewing
Processed
Results
Processed data files are stored as ASCII files (.CSV). When a raw file
is processed the resulting .CSV file is displayed automatically. You can
also open any LISST-200X .CSV file by selecting the View Process
File button from the home page.
A display similar to the one below will open. A list of available
parameters is displayed on the left. Clicking one of the parameters will
open a new window where the parameter will be displayed. The
parameter windows can be reorganized by dragging and dropping.
The slider at the bottom is used to scroll through the data file. A red
point on some graphs will indicate which sample is being displayed.
Step 21:
Opening a Real
Time Session
The LISST-200X software also supports the ability to acquire, process,
and display size data it in real time. The Start and Stop conditions will
not be used and therefore their settings are not relevant.
Before opening a real time session, you should collect a new
background and store it onboard the instrument. The real time session
LISST-200X User’s Manual
23
LISST-200X Introduction & Quick Start
will use the last background stored on the instrument to process the
data in real time. Therefore, you must ensure the background on the
instrument is up to date before continuing.
To open the Real-Time session, choose Collect Live Data from the
home page.
Choose a location to save the data by filling out the output directory
and the base file name. The sample interval can also be adjusted by
entering the number of seconds between sample in the sample
interval text box.
Even though data is displayed to the screen, it is not saved until you
select the Start Logging Button.
The Spherical / Irregular Shape radio buttons can be selected in order
to display the results as being processed under the assumption that
the particles are spheres or randomly shaped (natural grains) particles.
Data using both models is saved and the buttons only represent what
is displayed on the screen.
A list of available parameters is displayed on the left. Clicking one of
the parameters will open a new window where the parameter will be
displayed. The parameter windows can be reorganized by dragging
and dropping.
LISST-200X Introduction & Quick Start
24
LISST-200X User’s Manual
Deploying the
LISST-200X in
the Field
The above steps should have given you a good understanding on the
how to operate the LISST-200X. More details on all of the steps can
be found in the User’s Manual following this tutorial. The following
steps go into deploying the instrument in the field.
Step 22:
Preparing for
Transport
The LISST-200X has been designed to be a robust field instrument.
However, it is still a highly sensitive optical instrument and needs to be
treated with care. This is especially true during shipping. The LISST-
200X is shipped in a custom case with specially designed foam
cushioning. Anytime the instrument is transported it should be in this
case. We highly recommend using air shipping when possible to
eliminate the extended vibrations that ground shipping can cause.
Because the LISST-200X does not have an internal battery there is no
need to worry about putting the instrument into low power sleep mode.
You can simply disconnect it from the battery housing.
The Large External Battery Housing, which has alkaline batteries, can
be stored for extended period when fresh batteries are installed.
The Small External Battery Housing, which has NiMH rechargeable
batteries, should be opened and the batteries removed during
transport. There are not restrictions on shipping NiMH batteries.
However, the batteries will have a small drain when installed in the
Small Battery housing. Therefore it is recommended that they be
removed until the battery is ready to be used. The batteries should
also be stored in a fully charged state.
The LISST-200X can be pre-programmed with the desired sampling
program such as start and stop conditions and sampling rates. These
settings will remain in non-volatile memory until power is applied.
Therefore it is possible to prepare the instrument for sampling before
shipping it out. It is even possible to program the instrument to
automatically start upon power up. This is described in the next step.
Step 23:
Configuring
AutoStart
The LISST-200X can be configured to start running the sampling
program upon power up. In some situations this ability can be quite
helpful as you do not need to connect the instrument to the PC to start
the sampling. The instrument can be preconfigured with the desired
settings and the wait for the battery to be plugged in or external power
from the CTD to be present before looking for the Start Condition.
To configure the LISST-200X to start its sampling program upon power
up use the LISST-200X software or the configure auto start from the
terminal window. If using the terminal window, use the AS command
(short for Auto Start). The AS command can be issued when the
200X> prompt is displayed in the Terminal Window. When AutoStart
is enabled a message will be displayed as part of the Status command
or the Query instrument results.
LISST-200X User’s Manual
25
LISST-200X Introduction & Quick Start
When the LISST-200X is powered up a message will be displayed to
the Terminal window (or via the RS232 interface) that will prompt the
user that the instrument will auto start in 5 seconds. Pressing the stop
key or sending two CTRL-C characters will stop the instrument from
starting and return to the 200X> prompt.
Step 24:
Connecting to
External
Sensors or
Instruments
The LISST-200X had the ability to connect to external sensors and
instruments such as other dataloggers or CTDs through the 6-pin
Auxiliary connector on the endcap. The details of using the Auxiliary
connector can be found in the Step-by-Step instructions on page 52.
In short, the Auxiliary connector can be configured in one of four ways:
1) Analog and Digital inputs, 2) Dual Analog Inputs, 3) Dual Analog
Outputs, or 4) Triple analog input. When using the Analog or Digital
inputs it is possible to provide power to an external sensor. By default
this power output is switched off. It will need to be enabled before
power will be available. You can also configure the warmup time
required for the sensor after power is applied before sampling will
begin.
The Analog Output configuration sends out two analog voltages, Mean
Size and Total Concentration. This summary information is updated
each time data for the full size distributions is recorded. When using
the Analog Output configuration power can be received through the
Auxiliary connector. The wiring of the connector is designed to match
the SeaBird CTD Auxiliary Input connector simplifying the connection.
LISST-200X Operation Details
26
LISST-200X User’s Manual
II. LISST-200X Operation Details
Section
Organization
This section contains detailed instructions for performing various
procedures. These are either in the form of step-by-step instructions or
detailed descriptions of the various aspects of instrument operation
(e.g. command list, instrument mounting and deployment).
General
Precautions
• LISST-200X is a sensitive optical instrument – please handle it
gently as you would handle a very expensive camera.
• Critical alignments may be disturbed if the instrument is subjected
to shock or rough handling.
• Evidence of shock/rough handling will void the warranty.
• Whenever in transit, store the instrument in the provided padded
shipping case.
• If placing the instrument vertically on the standoffs, be sure to do
so gently as the Compact Flash Memory Card inside may
otherwise come loose.
WARNING-Class 3R laser – AVOID DIRECT EYE EXPOSURE
The LISST-200X uses a laser diode emitting a maximum of 1 mW of
visible (red) light at a wavelength of 670nm. Under normal
circumstances the laser travels only within the 2.5 cm sample volume.
However, if reflective objects are placed in the path of the laser beam,
the light could be redirected. Avoid conditions that could direct the
beam toward an eye.
LISST-200X User’s Manual
27
LISST-200X Operation Details
A. Step by Step Procedures
1.
Installing LISST-200X Software
Software for the PC is used to configure the LISST-200X and for downloading and
processing the size distributions.
STEP
ACTION
RESULT
1
• The LISST-200X comes with a USB memory card.
Plug the provided memory card into a USB port on
your computer. Locate the ‘LISST-200X_Installer.exe’
executable on the memory card. Note that this
software it is not compatible with Mac or Linux
operating systems. Your operating system must be
Window XP or newer to run the software.
Installer was found on
LISST-200X memory
card
2
• Double click the ‘LISST-200X_Installer.exe.’ Follow
the onscreen instructions and the installer will transfer
the necessary files to your computer and place a
shortcut on your desktop and start menu. Do not
remove the memory card from your computer until the
installation is fully completed:
Software installation
complete
LISST-200X Operation Details
28
LISST-200X User’s Manual
2.
Establishing Communication with the LISST-200X
Establish communication with LISST-200X using the supplied software and USB cable.
Communicating with the LISST-200X via the software should be automatic. Should there be
a problem you can use these step-by-step instructions to troubleshoot the problem.
STEP
ACTION
RESULT
1
• If the LISST-200X software is open, close the
program, then plug the USB communication cable
from your computer.
LISST-200X will be
powered on
2
• After plugging the USB cable back into your computer,
the drivers for the USB to serial converter should
install automatically. You should see a notification on
the task bar that drivers are being installed. If you are
unsure if the drivers installed, you can install the
drivers manually by running ‘CDM USB Drivers.exe’,
located on the memory card that came with your
instrument.
Computer is now set up
to communicate with
the LISST-200X
3
• After the driver installation is compete, open the
LISST-200X software.
Software appears
onscreen
4
• The software should automatically search and locate
your LISST-200X instrument among the serial ports on
your computer
• If no instrument is found you will see “No Instrument
Connected” displayed in the lower left corner.
• Try selecting the ‘Retry Connection’ button or try to
wake the instrument using the ‘Wake Instrument’
button.
Software will attempt to
automatically connect
to the instrument
5
• If further troubleshooting is needed click on the
‘Communication Help’ button to explore the
communication troubleshooting document.
Open Troubleshooting
Document
LISST-200X User’s Manual
29
LISST-200X Operation Details
3.
Saving and Evaluating Clean Water Backgrounds
In order to properly compute the size distribution, it is necessary to remove the light
scattering from the internal optics and window surfaces so that only the light scattering from
the particles of interest are used to compute the size distribution.
STEP
ACTION
RESULT
1
• Connect the instrument to the computer and establish
communication in the LISST-200X program (page 28)
Software open,
communicating with
LISST-200X.
2
• Clean instrument and install Clean Water Background
Test chamber.
• Fill with clean filtered water and make sure no bubbles
are in the water or on the windows.
• The water and instrument should be at the same
temperature. Using water a significantly different
temperature from the water can impact the quality of
the background.
• Stir the water well before obtaining a background to
make sure the water is well mixed.
• For the best Background, make sure the instrument is
clean; the water you use is filtered, and free of
bubbles. Temperature fluctuations in water will
seriously degrade the background. Use water at the
same temperature as the instrument. Stir well.
Optics submerged in
water
3
• Select Collect/View Background from the software
home page. The factory background file will be
automatically acquired from the instrument and
displayed on the screen.
• The current signal from the 36 ring detectors will also
be displayed in real time.
• When the Collect New Background button is pressed
20 samples will be collected and the average
displayed to the screen.
Background collected
and displayed on the
screen
LISST-200X Operation Details
30
LISST-200X User’s Manual
• The graph shows the value of the 36 light
scattering detectors. The red line is the factory
values, the blue line current background saved on
the instrument. A newly collected background will
be displayed as a solid black line. If the
background is close to factory levels the message
displayed will indicate a pass or acceptable
background. Even if your new background passes,
but is significantly higher than factory, clean the
windows and repeat to verify you have acquired
the best possible result.
Acceptable Background
• If the water or windows are not clean or if there is a
problem with the instrument, error messages and
suggested actions will be displayed. Dirty water or
windows will generally cause higher values across the
middle rings. Large bubbles or particles in the water
can cause higher values on the inner rings or left hand
side of the display. High values on the inner rings
combined with a lower transmitted laser power value
can also be an indication of optical misalignment. If
needed you can update the background, such as after
cleaning the windows or replacing the water, by
pressing the Collect New Background button again.
Unacceptable
Background
4
• In general, the lower the background values the better
the background. The goal is to get values that are at
the same values as factory line. However, as the
instrument is used the background may increase due
to small scratches and slight alignment changes. It
may not be possible to get the background down to
the original factory values.
LISST-200X User’s Manual
31
LISST-200X Operation Details
• Multiple cleanings or using better filtered water may be
required to get the best possible background.
• When you have an acceptable background press the
Save button to store the background on the
instrument. You also have the option to save the
background to a file on your computer. After the
background is saved, select the ‘Save as .BGT File’
button. The LISST-200X handles the background files
differently than the LISST-100X. The background file
is stored on the instrument and is saved as part of
every data file that is recorded. When processing a
data file the background will be automatically
extracted and used during processing. Saving,
tracking and selecting a background file to use for
processing is no longer required. The background
stored on the instrument will continued to be saved in
new data files until a new background is recorded.
Background Accepted
and saved.
LISST-200X Operation Details
32
LISST-200X User’s Manual
4.
Configuring Data Collection
The LISST-200X software can be used to configure the Start and Stop Conditions, Fixed
Rate or Burst sampling and sample rates.
STEP
ACTION
RESULT
1
• Connect the instrument to the computer and establish
communication in the LISST-200X program (page 28).
Software open,
communicating
with LISST-200X.
2
• Open the Sampling Programs window by choosing
Instrument Configuration from the home page.
• Certain instruments can compute the PSD onboard in real
time. If your instrument has this feature, an additional
configuration ‘Data Output’ will appear on this page. (See
Autonomous Real-time Data Processing on page 52 for
more information).
• To set the LISST-200X data logger clock to the computer
clock, open the status bar, click the wrench icon, and select
‘Set Instrument Clock’.
• You can also zero the depth sensor from this same menu.
Instrument Status
Displayed.
Clock set
3
• By selecting the Sampling Schedule Tab at the top of the
main window the screen below appears. This screen is
used to set the type of sampling: Fixed Rate or Burst. You
can also select the samples per average and sample rates
on this screen.
Operating Mode
set
LISST-200X User’s Manual
33
LISST-200X Operation Details
• The Burst and Fixed sample rate modes are used to set up
a sampling program that will save data onboard the
instrument. The icons next to the various values give a
better understanding of their meaning. The software
automatically checks the values entered to make sure that
there is no conflict. For example, when a Samples per
Average value is entered, the minimum sample interval is
computed. If this value is less than the minimum permitted
the value will be changed to the minimum and the text will
turn red. A similar test is done on the Burst Interval setting.
4
• Select the Start Conditions tab to configure when the
instrument will begin sampling.
• There are five options: Depth, Time, External Mechanical
Switch, External Digital Input, and Time Delay. Select the
mode by clicking on the button next to the its label. Select
the correct parameters as required.
Start Condition
Set
•
If the AutoStart check box is selected, the LISST-200X will
start the sampling program when power is applied to the
instrument. If this is not selected the user must send the
LISST-200X Operation Details
34
LISST-200X User’s Manual
‘GO’ command to start the sampling programing after
applying power.
5
• Choose the Stop Conditions Tab to select the conditions
when sampling should stop.
• The available stop conditions are: Depth, Time, External
Mechanical Switch, External Digital Input, Fixed number of
samples, and Memory Full.
• If the Stop Condition is Depth, External Mechanical Switch,
or External Digital Input and the Start Condition is also one
of these three options the Base Program on the instrument
will return to waiting for the Start condition. For example, if
the start and stop conditions were set to External
Mechanical Switch then multiple sampling sessions can be
obtained by flipping the switch on and off. This is useful
when doing profiles. It eliminates the need to communicate
with the instrument between profiles.
Stop Condition
Set
6
• Choose the Accessories tab to configure LISST-200X
accessories. The accessories must be selected here in
order for them to function properly.
•
LISST-200X User’s Manual
35
LISST-200X Operation Details
7
After configuring the instrument settings, they must be uploaded
onto the instrument. A message will be displayed on the top of
the screen when you have unsaved changes to the instrument
configuration
Click on the ‘Upload to Instrument’ button to load the
configuration onto the LISST-200X. A confirmation message will
be displayed when the upload is finished.
Upload
Configuration
8
• The instrument is now configured for deployment, however,
the sampling program is not yet running.
• You MUST run the sampling program in order to collect
data. But even when running, the sampling program will wait
for the programmed start condition before collecting data.
See Start and Stop Conditions on page 57, and the steps
above, for details about start conditions.
• When the sampling program is running, you will receive a
confirmation message that the program has started, and the
instrument will become disconnected from the computer. At
this point the instrument is ready to be deployed.
• The green LED on the connector endcap will indicate the
instrument status:
-
If waiting for the start condition, every few seconds it will
blink twice quickly.
-
If logging, it will light while collecting each sample.
Run Sampling
Program
LISST-200X Operation Details
36
LISST-200X User’s Manual
5.
Using the Small External Battery Housing
The LISST-200X comes with a Small External Battery Housing that can be used for short
term deployments. The instructions in this section describe how to use the Small Battery
including installing and charging the NiMH rechargeable batteries. NOTE: the Small
Battery Housing cannot be used to power the LISST-HAB or LISST-Black (see page
90) or BioBlock (page 93).
STEP
ACTION
RESULT
1
• The LISST-200X Small External Battery Housing uses
two rechargeable NiMH D-cell batteries. Four of the
1.5V 10000mAh batteries and a charger are included
with the housing. The batteries are not installed when
the instrument a new instrument is shipped. It is
recommend that the batteries be removed from the
housing when not in use. Leaving the batteries in the
housing will cause the batteries to discharge and
could cause damage to the cells.
Batteries fully charged.
• Fully charge the batteries using the provided charger.
Fully discharged batteries will take about 8 hours to
charge.
NiMH battery charger
2
• The Small Battery Housing is made of plastic and has
an endcap with two underwater connectors and two
captive thumbscrews on one end.
• Loosen the thumbscrews. Note that the screws will
not need to be fully removed. They are designed to
be removed from the main housing but still remain
attached to the endcap. The screws can be removed
if you continue to rotate the screws through the
threaded holes in the endcap.
Battery open and ready
for batteries
LISST-200X User’s Manual
37
LISST-200X Operation Details
• Slide the endcap with battery holder attached out of
the main housing.
3
• Slide the clear cover off of the battery holder and
install the two D cell batteries.
• Install with the positive “button” end toward the end
cap of the housing.
• Slide the clear cover over the batteries such that the
rubber pad is pushing against the batteries as shown
in the picture above
Batteries installed.
4
• Check that the o-ring in the endcap and the mating
surface on the main housing are clean. Lightly grease
with silicone grease as needed.
• Carefully insert the assembly back into the main
housing until the o-ring makes contact with the
housing. Align the thumbscrews with the holes in the
main housing and push the endcap on until it is flush
with the housing.
• Tighten the thumbscrews by hand (DO NOT USE
ANY TOOL). DO NOT OVERTIGHTEN.
Battery assembled and
ready for use.
LISST-200X Operation Details
38
LISST-200X User’s Manual
5
• The battery housing can be connected to the LISST-
200X using the provided 5-pin Male to 5-pin Female
cable.
• Connect the Male end of the cable to the Female
Bulkhead connector on the housing.
• Connect the other end of the cable to the 5-pin Male
bulkhead on the LISST-200X.
• Power is now going to the LISST-200X. If AutoStart is
enabled the instrument may be active and waiting for
the start condition.
Instrument is powered
and ready to collect
data
6
• If desired connect the Communications cable to the
male 5-pin bulkhead connector on the Small Battery
Housing. Communications will pass through the
housing to the LISST-200X.
• The battery and communications cables are shown
below.
LISST-200X User’s Manual
39
LISST-200X Operation Details
6.
Using the Large External Battery Housing
The Large External Battery Housing is used to provide power to the LISST-200X for long
term deployments or extended profiling operations.
STEP
ACTION
RESULT
1
• The LISST-200X Large External Battery is shipped
with 16 Alkaline D batteries pre-installed and is ready
for use. The following steps describe how to open the
battery and replace the batteries. Skip to Step X to
connect the battery to the LISST-200X.
• The new alkaline batteries can be left in the battery
housing for storage. However, it is not recommended
that fully discharged batteries be left in the battery
housing during long term storage.
Battery ready to use
when shipped.
2
• To access and replace the batteries in the Large
Battery housing you will need remove the endcap with
the connectors and handle.
• Remove the four 6-32 X ¾ long socket head cap
screws from the connector endcap using the provided
7/64” ball driver. Remove the screws uniformly or hold
the cap down against the light spring force.
Endcap ready for
removal
3
• Pull out the connector endcap. There are no wires or
other connections. Set the endcap to the side.
Endcap removed and
batteries replaced.
LISST-200X Operation Details
40
LISST-200X User’s Manual
• Remove the batteries making note of their orientation.
• Insert the batteries into the pressure case. Labels on
the inside surface of the center divider will guide you
on the orientation of the batteries for each stack.
4
• Inspect the o-ring on the endcap and the seating
surface on the pressure case to make sure there is no
debris that could affect the seal. Lightly grease the
surfaces as needed.
• Align the tab in the center of the endcap to the slot in
the post in the center of the pressure case
Endcap re-installed
5
• Screw the connector endcap onto the pressure case
with the four 6-32 x ¾ socket head cap screws. If
necessary, apply anti-seize compound onto the
threads.
Battery assembled and
ready for use.
6
• The battery housing can be connected to the LISST-
200X using the provided 5-pin Male to 5-pin Female
cable.
• Connect the Male end of the cable to the Female
Bulkhead connector on the Battery Housing. The
connector will be labeled with “TO LISST”.
Instrument is powered
and ready to collect
data
LISST-200X User’s Manual
41
LISST-200X Operation Details
• Connect the other end of the cable to the 5-pin Male
bulkhead on the LISST-200X.
• Power is now going to the LISST-200X. If AutoStart is
enabled, the instrument may be active and waiting for
the start condition.
7
• If desired, connect the Communications cable to the
male 5-pin bulkhead connector on the Large Battery
Housing which is labeled “TO PC”. Communications
will pass through the battery to the LISST-200X.
• The battery and communications cables are shown
below.
LISST-200X Operation Details
42
LISST-200X User’s Manual
7.
Offloading and Deleting Data Files from Internal Memory
Data collected using the LISST-200X is stored on the internal data logger. This section
describes how to offload and delete files from the internal data logger.
STEP
ACTION
RESULT
1
• Connect the instrument to the computer and establish
communication in the LISST-200X program (page 28).
Software open,
communicating with
LISST-200X.
2
• Select Offload Or Erase Files from the home page.
Shows a list of files
currently stored
onboard the instrument
3
• Select a directory to save the offloaded files on the
right side. Select individual or multiple files on the
LISST to delete or offload (by holding down either shift
or ctrl). Press the green arrow.
Select files to delete or
offload
4
• The progress bars at the bottom of the screen will
update as files are offloaded or deleted
Files are offloaded or
deleted
LISST-200X User’s Manual
43
LISST-200X Operation Details
8.
Processing a Single Raw Data File
Data that has been downloaded from the datalogger is in a raw binary file (.RBN extension).
It must be processed into particle size by the LISST-200X program.
STEP
ACTION
RESULT
1
• Start the LISST-200X program
Program started.
2
• Select Process Data Files from the home page. Select
the file you want to process. A display similar to the
one shown below will appear.
• A summary of auxiliary data is shown on the screen.
More detailed plots can be viewed by clicking on the
parameter.
Select file types to be
created.
3
• You must also select the inversion model to be used
when processing the data: Spherical, randomly
shaped or both.
• For a description of the particle models, see Appendix
F: Particle Shape Models on page 89.
• If you have selected Randomly shaped Particle
Inversion Model, the processed files will have an’ _rs’
suffix
Select inversion
method(s) to be used –
spherical, randomly
shaped or both.
4
• Standard raw data processing will use the background
file contained in the .RBN file. If you would like to use
a different background file than the one in the RBN
file, or use a data record as a background, select the
‘View/Select Background’ button
Select to process a file
using the background in
the data file or an
external .BGT file
LISST-200X Operation Details
44
LISST-200X User’s Manual
You can view and make background selections from
this window.
5
• Next, select the output data file name. Fill in the output
directory and the base file name. Several files will be
created.
• The CSV file is a space delimited ASCII file containing
the fully processed and calibrated data. This is useful
for loading into spreadsheets such as Excel etc.
• The RTX file is a space delimited ASCII file containing
the raw data from the LISST-200X data logger. This is
useful for looking at the raw data in a spreadsheet
such as Excel etc.
Name out output file.
6
• To select the range of raw data to process, move the
sliders at the bottom of the display.
Range to process
selected.
7
• After selecting the range to process, press the
Process File button. The display will show the
processing progress and the resulting data will be
displayed when processing is complete.
Processing Completed.
LISST-200X User’s Manual
45
LISST-200X Operation Details
9.
Batch Processing Multiple Raw Data Files
Data that has been downloaded from the datalogger is in a raw binary file (.RBN extension).
It must be processed into particle size by the LISST-200X program. If you have multiple raw
data files to process (e.g. from a series of profiling deployments), follow these processing
steps that allows you to process all files in one operation.
STEP
ACTION
RESULT
1
• Start the LISST-200X program
Program started.
4
• Select Process Data Files from the home page.
Select multiple files to process (by holding down either
shift or ctrl).
Select Files
5
• By default, batch processing will use the background
stored in the raw data files to process the data. If you
would like to use an external .BGT file, press the
‘Select .BGT Background File’ button.
Select source of
background files
6
• You must also select the inversion model to be used
with the real time data processing: Spherical,
randomly shaped or both.
• For a description of the two particle models, see
Appendix F: Particle Shape Models on page 89.
• If you have selected Randomly shaped Particle
Inversion Model, the processed files will have an’ _rs’
suffix
Select inversion
method(s) to be used –
spherical, randomly
shaped or both..
LISST-200X Operation Details
46
LISST-200X User’s Manual
7
• Select the Output Directory for the processed files.
• Files will be named with the same name as the raw
data files but with different extensions. Several files
will be created.
• The CSV file is a space delimited ASCII file
containing all the processed data.
• The RTX file is a space delimited ASCII file
containing the raw data from the LISST-200X data
logger.
Select Output Directory
• Press the “Process Files” button.
• The files will be processed in order and the
progress bars will indicate the status of each file
Process files
LISST-200X User’s Manual
47
LISST-200X Operation Details
10. View Processed Data File
After processing of the data into a .CSV, the data can be viewed to the screen. This step-by-
step procedure covers the viewing of data and optional displays.
STEP
ACTION
RESULT
1
• Start LISST-200X program
Program started.
2
• Select View Processed File from the home page
3
• Select the processed data file (.CSV) from the file
selection window. Double click the file or type the file
name and press Open.
Processed data file
selected.
4
• After selecting the processed file, a window similar to
the one shown below will appear.
Particle size
distribution displayed
on screen
• A list of available parameters is displayed on the left.
Clicking one of the parameters will open a new window
where the parameter will be displayed. The parameter
windows can be reorganized by dragging and dropping.
The slider at the bottom is used to scroll through the
data file. A red point on some graphs will indicate which
sample is being displayed.
5
• Additional processed files can be opened
simultaneously by pressing the plus button next the
open tab at the top left of the screen.
• When multiple tabs are open, the tabs can be dragged
out into their own windows, allowing for easy
comparison between files.
Open additional Files
LISST-200X Operation Details
48
LISST-200X User’s Manual
11. Data Quality Control
Making sure your data make sense.
STEP
ACTION
RESULT
1
• Follow steps 1-5 in the previous section: View
Processed Data File
Data file loaded
2
3
• The first value to inspect is the Optical
Transmission. By definition, transmission must be a
number between 0 and 1 (except that electronic
noise may cause it to be slightly higher than 1 in
particle-free water)
• Tip: The transmission value is in column 60 of the
.CSV file.
• If transmission shows up as being larger than 1
(one), then your measurement is most likely taken
in very clear water and/or you have a bad
background measurement obtained with dirty water
and/or dirty windows.
• If your background measurement was obtained
using dirty water, it may be possible to redo the
background using clean water and re-process the
data, then check to see if the transmission values
drop below 1 (one).
Transmission values
much greater than 1 or
less than 0 indicate
bad data.
4
• If your transmission values generally are in the
0.98-0.995 range, your measurements are taken in
very clear water.
• This means that the signal-to-noise ratio will be low,
and the data may have a lot of noise in them, but
can most likely still be used.
Transmission values
very close to 1 yield
noisy data.
LISST-200X User’s Manual
49
LISST-200X Operation Details
6
• If your transmission values are < 0.10 (or 10%), the
water is too turbid. Disregard these data.
Data with transmission
values < 10% should
be disregarded.
7
• If your transmission values are > 0.995 (or 99.5%),
the water is too clear. Disregard these data.
Data with transmission
values > 99.5% should
be disregarded.
8
Summarizing
• Disregard data if transmission is > 0.995 (> 99.5%).
• Disregard data if transmission is < 0.10 (< 10%).
• Be wary of data collected at transmission values
between 0.98 and 0.995 – low signal-to-noise ratio.
• Be wary of data collected at transmission values
between 0.30 and 0.10 – generally decreasing data
quality as the transmission decreases below 0.30
(30%).
LISST-200X Operation Details
50
LISST-200X User’s Manual
12. Real-Time Data Processing with LISST-200X Software
This procedure covers the acquisition, display and storage of processed data in real time,
using the LISST-200X software. For on-board processing by the LISST-200X firmware, see
Autonomous Real-time Data Processing on page 52.
STEP
ACTION
RESULT
1
• Connect the instrument to the computer and establish
communication in the LISST-200X program (page 28).
Communication
established.
2
• The background currently stored onboard the instrument
will be used to process the data as it is collected in real
time.
• Be sure to collect a new background and store it onboard
the instrument before beginning a real time session
Background is
current
3
• Select Collect Live Data from the home page
Open Instrument
selected.
4
• A display will appear on the screen. It is similar to the
main window of the View Processed Data function.
• A list of available parameters is displayed on the left.
Clicking one of the parameters will open a new window
where the parameter will be displayed. The parameter
windows can be reorganized by dragging and dropping.
• The ‘Sample Interval’ text box allows you to adjust the
time between samples in seconds.
LISST-200X User’s Manual
51
LISST-200X Operation Details
• The Sample Number value displayed underneath the
plots will increment each time a sample is saved.
• The buttons labeled ‘Spherical’ and ‘Irregular’ allow you to
display the volume distribution calculated with either the
spherical or random shape particle model. The buttons
only effect the display and have no effect on the data that
is saved in the .CSV file. Data from both particle models
is always saved.
5
• Select an output directory and a base file name. Several
files will be created.
• The CSV file is a space delimited ASCII file containing all
the processed data.
• The RTX file is a space delimited ASCII file containing the
raw data from the LISST-200X data logger.
Select output
directory and file
name
6
• Data will not be saved until the ‘Start Logging Data’ button
is pressed.
Start Data Logging
LISST-200X Operation Details
52
LISST-200X User’s Manual
13. Autonomous Real-time Data Processing
Newer LISST-200X’s (starting with serial number 2131), or older units with upgraded
electronics, have enough computational power to do PSD processing autonomously, without
requiring a connection to a Windows computer.
You may choose to save the processed data files (.CSV) in the LISST-200X memory, or
transmit them in real time through the serial port, or both. Note that processing slows the
maximum sampling rate. Without it, the LISST-200X can sample continuously once per
second. Depending on the settings you select, the time per sample may be up to 2 seconds.
To determine whether your LISST-200X supports real-time processing
STEP
ACTION
RESULT
1
• Connect the instrument to the computer and establish
communication in the LISST-200X program (page 28).
Communication
established.
2
• Open the information panel at the bottom of the window,
and check the firmware version:
• If the firmware is 2.1 or greater, your instrument
supports autonomous processing.
• If the firmware is less than 2.1 but at least 2, a user-
installable firmware upgrade can enable autonomous
processing. Contact Sequoia Scientific support for
instructions.
• If the firmware version is less than 2, your instrument’s
hardware does not support autonomous processing.
Contact Sequoia Scientific sales regarding upgrade
options.
If your instrument
supports real-time
processing,
proceed.
To use real-time processing
STEP
ACTION
RESULT
1
• Connect the instrument to the computer and establish
communication in the LISST-200X program. (see
Establishing Communication with the LISST-200X on
page 28).
Communication
established.
LISST-200X User’s Manual
53
LISST-200X Operation Details
2
• Navigate to the Configure Instrument page. If your
instrument’s hardware and firmware support it (see
above), you will see a Data Output category.
Configure
Instrument page
open.
3
• Click on Data Output to make its options visible.
o ‘Save PSD Files On Instrument’ – if checked:
▪ PSD is computed onboard in real time
▪ PSD is saved in a file onboard the
instrument, in .CSV format (see Appendix
C: Data File Formats on page 79)
o ‘Transmit PSD In Real Time’ – if checked:
▪ PSD is computed onboard in real time
▪ A PSD data record is transmitted over the
serial port immediately after each
measurement is collected
o ‘Inversion Model’
▪ Select the inversion model that will be used
for onboard PSD processing (see Appendix
F: Particle Shape Models on page 89)
Instrument
configuration
selected.
4
• IMPORTANT: after you have selected the desired
configuration, press the ‘Send Config. To Instrument’
button to load the configuration onto the instrument.
Configuration
uploaded to
instrument.
LISST-200X Operation Details
54
LISST-200X User’s Manual
• A message box will confirm the upload has been
completed.
Configuration
complete.
LISST-200X User’s Manual
55
LISST-200X Operation Details
14. Collecting Data from an External Analog Sensor
The LISST-200X can supply regulated 12V power to an external sensor, and measure one
to three analog voltages, depending on its hardware configuration and firmware settings. If
the 12V power output is activated, it will turn on each time the 200X collects a sample or
burst of samples, and turn off each time the 200X goes into a waiting state of more than a
few seconds. If using the 12V output, you can specify how long it should turn on before
sampling, as a warmup time for the sensor.
To provide these functions, the auxiliary connector on the LISST-200X must be in one of
three configurations: Analog & Digital Input (which provides a single analog input), Dual
Analog Input configuration (which provides two), or Triple Analog Input configuration (which
provides three).
Remember that power consumed by an external sensor will reduce battery endurance, in a
battery-powered deployment. It might also exceed the capacity of the USB-powered data
cable. If so, you can augment the USB cable’s power by connecting the large battery
housing.
STEP
ACTION
RESULT
1
• Determine current configuration of Auxiliary connector: in
LISST-200X Software, open the instrument status bar. Look
for the Endcap Configuration. If it is not one of the analog-
input configurations, it must be changed; contact Sequoia
for assistance.
Auxiliary
Connector
configuration
determined
2
• Make the appropriate connections between the 6-pin
auxiliary connector of the LISST-200X and your sensor. To
see the pin assignments, type the CONFIG command in the
LISST-200X terminal window; page 61). Or see Appendix E:
Connectors on page 84.
External sensor
connected
3
• Type SENSORPOWER into the terminal window, and
respond to its prompt to select the appropriate option.
• Type SENSORWARM in the terminal window to set a
warmup time (in seconds).
• WARNING: a USB cable may not provide enough power
for external sensors. Use a battery housing or other power
source to provide enough power to run external sensors.
Configured for
analog collection
4
• Set the standard sampling parameters to control the
sampling schedule and other options (see
• Configuring Data Collection on page 32).
Ready to collect
data
LISST-200X Operation Details
56
LISST-200X User’s Manual
15. Configuring the LISST-200X as a Sensor for a CTD
The LISST-200X can operate as an analog-output auxiliary sensor for a CTD. For this
purpose the LISST-200X’s auxiliary six-pin connector must be in its “Dual Analog Output”
configuration. The two analog outputs represent the total volume concentration of particles,
and the Sauter mean diameter (SMD). These values are approximations for the
convenience of real-time display. For highest-quality data you will still need to offload and
process the detailed data from the LISST-200X after deployment.
The LISST-200X updates its analog outputs each time it collects a sample, at a maximum of
1 Hz.
STEP
ACTION
RESULT
1
• Determine current configuration of Auxiliary connector: in
LISST-200X software, open the instrument status bar. Look
for the Endcap Configuration. If it is not in the Dual Analog
Output configuration, it must be changed; contact Sequoia
for assistance.
Auxiliary
Connector
configuration
determined
2
• Make the appropriate connections between the 6-pin
auxiliary connector of the LISST-200X and your sensor. To
see the pin assignments, type the CONFIG command in the
LISST-200X terminal window (see LISST-200X Command
Details on page 61), or see Appendix E: Connectors.
Connected to
CTD
3
• In LISST-200X software, set the sampling parameters as
follows (see
• Configuring Data Collection on page 32):
o Operating mode: Fixed sample rate
o Sample to be average of 20 measurements
o Sample interval: 1 second
o Automatically start sampling program upon power up
Start condition: delay 0 minutes
o Stop condition: memory full
200X will collect
data at 1 Hz
whenever the
CTD supplies
power to it
LISST-200X User’s Manual
57
LISST-200X Operation Details
B. Start and Stop Conditions
Overview
The LISST-200X can be set up to start and stop data collection
according to various conditions. The most obvious is the white switch
lever on the connector end cap. However, the switch is only one of
several options. The start condition and stop condition can be different
from each other. Also see Configuring Data Collection on page 32.
Start Conditions
The five possible start conditions are:
1. Depth,
2. Time,
3. External Mechanical Switch,
4. External Digital Input, and
5. Time Delay.
The details of each are described below. Note that regardless of which
start condition is selected, the LISST-200X will not start checking for
the start condition until the sample program is run from the Windows
software (see page 32), by the GO command (see Acquisition/Action
commands on page 69), or by automatic start after power-up.
Depth Start
The LISST-200X checks its depth every 3 seconds. Each time it
checks, the green LED on the end cap will double-flash. If the depth
exceeds the threshold, the program will start data collection. When
setting the threshold, remember that atmospheric pressure,
temperature, and even the orientation of the LISST-200X can affect the
offset of the pressure sensor. Therefore, it is best to zero the sensor
(see page 32) immediately before starting, and to set the threshold to at
least 1 meter, to avoid incorrect triggering.
Time Start
The LISST-200X will go into lower-power sleep until the programmed
time. (If the programmed time has already passed, it will start
immediately.) While waiting, it will quickly flash the LED once every 30
seconds.
External
Mechanical
Switch Start
The LISST-200X will wait until the white plastic lever on its endcap is
moved to the “1” position. (If it is already in that position, it will start
immediately.) While waiting, it will quickly flash the LED twice every few
seconds.
External Digital
Input Start
If the auxiliary connector is configured for analog and digital inputs (see
Appendix E: Connectors), pin 2 of the 6-pin connector is the digital
input. The program will check the status of the digital input once a
second. If the digital input is greater than 2 volts (relative to digital
ground, pin1), data collection will start. While waiting, the LED will
quickly flash twice every few seconds.
Time Delay Start
The time delay start condition will cause the program to wait the
specified number of minutes. While waiting it will quickly flash the LED
once every 30 seconds.
LISST-200X Operation Details
58
LISST-200X User’s Manual
Stop Conditions
The possible stop conditions are:
1. Depth,
2. Time,
3. External Mechanical Switch,
4. External Digital Input,
5. Fixed number of samples, and
6. Memory full or power lost
When storing data in the Fixed Sample Rate mode the Stop conditions
are checked after each averaged sample has been saved. When
storing data in the Burst mode the Stop conditions are only checked
after a full Burst has been completed. The Start and Stop conditions
have no effect on the real-time sampling mode (Real-Time Data
Processing with LISST-200X Software on page 50). The details of each
condition are described below.
Depth Stop
If the depth is less than the threshold, sampling will stop. If the Start
Condition is Depth Start the program will return to checking for the start
depth, and will start a new file each time the start threshold his crossed.
For all other Start Conditions the when the current depth is less than
the threshold the LISST-200X will stop and go into deep sleep.
If using depth for both start and stop conditions, consider whether the
start and stop thresholds should be different, depending on your
deployment scenario. For example, if the instrument might be
suspended from a winch at a depth near the start and stop thresholds,
there could be “false” starts and stops due to wave motion. This could
be prevented by setting the start depth deeper than the stop depth.
Time Stop
After each sample or burst, if the given date and time have passed, the
LISST-200X will stop and go into low-power sleep.
External
Mechanical
Switch Stop
After each sample or burst, if the switch lever is in the off or “0”
position, sampling will stop. If the Start Condition is Switch Start, the
program will return to checking the start condition. For all other Start
conditions the instrument will stop and go into deep sleep.
External Digital
Input Stop
The status of the digital 1 input is checked after each sample or burst. If
the voltage at the input is less than 0.7 volts the sampling will stop. If
the Start Condition is a Digital Input Start the program will return to
checking the start condition. For all other Start conditions the
instrument will stop and go into deep sleep.
Fixed Number of
Samples
When the number of samples to be saved has been reached the
program will stop and go into deep sleep (regardless of start condition).
Memory Full or
Voltage Low
Sampling will continue until the memory is full, or the power input falls.
If the memory fills, the instrument will go into deep sleep.
NOTE: the memory capacity is large enough that it should never be
necessary to fill it. We recommend deleting files from the LISST-200X
as soon as they have been offloaded to a backed-up computer or other
storage.
LISST-200X User’s Manual
59
LISST-200X Operation Details
C. Instrument Communication
Overview
The cables Sequoia supplies with the LISST-200X send data through a
standard USB port. However, the LISST-200X itself communicates
through RS-232 serial protocol. The 5-pin male connector on the
LISST-200X carries the RS-232 signals (and power). The cables
incorporate adapters that convert between RS-232 and USB.
If you use a different cable that does not include the RS-232-to-USB
adapter, you can use an external adapter, or connect directly to an RS-
232 port if your computer has one (rare on any recent computer). In
Windows, both RS-232 and USB connections will appear as COM
ports.
If interfacing to a data logger or controller system other than a Windows
computer, you will likely also use a direct RS-232 connection.
The RS232 link communicates at 9600 baud, 8 data bits, No parity, and
1 stop bit. For offloading data files, the LISST-200X software uses
YMODEM protocol at 115K baud. The transfer rate can be changed in
the settings of the LISST-200X software. A slower speed may be
required when downloading data over cables longer than a few meters.
Using the LISST-
200X Terminal
Window
In the LISST-200X software, you can open a terminal window to directly
view communications with the LISST-200X, and enter commands.
When the LISST-200X is connected and the terminal window is the
front window, the LISST-200X should respond to pressing the enter key
with the L200X:> prompt. If the instrument is in the deep sleep mode,
you can wake it by pressing the Stop button.
Direct
Commands
In most cases, LISST-200X software will be used to configure and
operate the LISST-200X. However, some functions are available only
through direct commands in the terminal window. Also, in some
applications the LISST-200X may need to communicate to another
datalogger or custom program. For this purpose, a large set of
commands is available to operate the instrument. See the following
sections for detailed descriptions of each command.
LISST-200X Operation Details
60
LISST-200X User’s Manual
D. LISST-200X Command Summary
Display Commands
CONFIG
Display instrument hardware configuration information
DD, DIR
Display current disk directory
DEPTH
Display current depth
DS, DS 1
Display current instrument status information
DT
Display current time and date
HELP, ?
Display general help messages and command list
SW
Display mechanical switch position
TEMP
Display current temperature
VIN
Display input supply voltage
Setup Commands
AS x
Set autostart
BI x
Set burst interval
MA x
Set samples per average equal to x
MODE, OM x
Set operating mode (burst or fixed-rate)
PD x, STOPDATA x
Set stop condition Data
PSDMODEL x
Set model used for PSD processing (if active)
SAVEPSD x
Set whether PSD data will be saved in on-board memory
SI x
Set sample interval
SP x, STOPCOND x
Set stop condition
ST x, STARTCOND x
Set start condition.
SB x
Set samples per burst
SC yyyy/mm/dd hh:mm:ss
Set clock with time and date
SENDDATA x
Control real-time transmission of data via RS232
SENSORPOWER x t
Set whether power output turns on during logging
SENSORWARM t
Set external sensor power warmup time
TD x, STARTDATA x
Set start condition data
ZD
Reset depth sensor offset
Acquisition/Action Commands
DL filename
Delete file
GO
Start data collection using current parameters
GX
Grab sample and transmit it
XR
Transmit raw data when average saved
ZS
Acquire 20 samples and evaluate them for quality.
ZZ
Go into deep sleep mode (minimum power consumption)
LISST-200X User’s Manual
61
LISST-200X Operation Details
E. LISST-200X Command Details
NOTE: Command are shown in upper case for clarity, but are not case-sensitive.
1.
Display Commands
CONFIG
Display instrument hardware configuration
Syntax:
CONFIG
Description:
The current hardware configuration including Auxiliary I/O connector
configuration
Example:
L200x:> CONFIG
Serial Number: 2001
Firmware Version: 1100 (1.100)
VCC: 59371
Full Path Length: 2500 (25 mm)
Eff. Path Length: 2500 (25 mm)
BioBlock Installed: 0
Start Condition: Digital input (4)
Start Condition Data: 0
Stop Condition: Switch (3)
Stop Condition Data: 0
Measurements Per Average: 10
Sample Interval: 1 seconds
Burst Mode: 0
Samples Per Burst: 5
Burst Interval: 25 minutes
Transmit Raw Data: 0
Lifetime Sample Count: 529554
Lifetime Laser On Time: 1703980 seconds
Endcap Configuration: 0 (Analog & digital in)
Wiring of MCBH6MP connector:
1: Ground
2: Digital input 1
3: Digital input 2
4: +12V output
5: Analog input 1
6: Ground
DD, DIR
Display File Directory
Syntax:
DD or DIR
Description:
Display current disk directory in DOS type format. Includes total bytes
used and bytes available.
Example:
L200x:> DD
. 2099-03-14 03:35:48
.. 2099-03-14 03:35:48
L2390658.RBN 1,612,440 2019-08-29 14:55:28
L2411501.RBN 3,253,800 2019-09-03 07:58:20
L2480807.RBN 736,800 2019-09-06 09:40:14
3 file(s) 7,608,360 bytes
2 dir(s) 1,016,381,440 bytes free
LISST-200X Operation Details
62
LISST-200X User’s Manual
DEPTH
Display current depth
Syntax:
DEPTH
Description:
The current depth is displayed to the screen.
Example:
L200x:> depth
0.000 m (680 counts)
DS
Display current status information
Syntax:
DS
Description:
The instrument settings and status are displayed to the screen. The
format may change with different firmware and hardware versions.
Example:
L200x:> DS
LISST-200X Status and Settings
Serial number: 2021
Endcap configuration: 2 (Dual analog out)
Firmware Version: 1.470 Feb 8 2021 21:06:43
Current Date/Time: 2019/01/09 22:12:50
Current Day of the Year: 009
Supply voltage: 14.74 V
Operating mode: Fixed rate mode
AutoStart: OFF.
Delay Start with 0 minute delay
Stop on full memory or low battery
Measurements per Average: 32
Ambient light subtraction: on.
Sample Interval: 1
Path length: 25 mm
Standard status messages will be sent during logging.
Data will be saved in files of unlimited size.
AUV mode: off
Free memory: 1016 Mbytes
DS 1
Display current status information, compact form
Syntax:
DS 1
Description:
Instrument status displayed in simplified format for reading by software
Example:
L200x:> ds 1
LISST-200X Current Status and Settings
SN = 0
OM = 0
ST = 5
TD = 0
SP = 6
PD = 0
MA = 32
SI = 1
BI = 1
SB = 1
AS = 0
BIOBLOCK = 0
STUBE = 0
LISST-200X User’s Manual
63
LISST-200X Operation Details
SENSORPOWER = 0
SENSORWARM = 0
Current Time = 1987/01/09 22:14:00
Battery= 1474
Switch= 0
Memory= 1016381440 Bytes
Endcap= 2
PATHLENGTH = 25
Firmware = 1.470
DT
Display current time and date
Syntax:
DT
Description:
Displays current time and date to the screen.
Example:
L200x:> DT
Date/time = 2021/01/01 18:16:06
HELP, ?
Display general help messages and command list
Syntax:
HELP or ?
Description:
Displays the list of command to the screen.
Example:
L200x:> HE
LISST-200X Commands -------------------------------------
[list of commands]
SW
Display current Mechanical Switch position
Syntax:
SW
Description:
The current position of the Mechanical Switch is displayed to the screen.
Example:
L200x:> SW
Switch: 0
TEMP
Display current temperature
Syntax:
TEMP
Description:
The current temperature is displayed to the screen.
Example:
L200x:> temp
25.09 C (26629 counts)
VIN
Display Battery Voltage
Syntax:
VIN
Description:
The current supply voltage is displayed to the screen.
Example:
L200x:> vin
14.74 V (2058 counts)
LISST-200X Operation Details
64
LISST-200X User’s Manual
2.
Setup Commands
AS
Enable Autostart
Syntax:
AS x, where x is 1 (yes) or 0 (no)
Description:
With Autostart enabled, the firmware will immediately start the sampling
program when power is applied to the instrument.
Example:
L200x:> AS 1
Autostart upon power-up is on
Cautions:
If Autostart is enabled, the user cannot communicate with the instrument
when powering it up – it will immediately start sampling according to the
SD defaults. In order to stop sampling and establish normal
communication, the user must issue a stop command, either from the
LISST-200X software’s graphical interface, or by pressing CTRL-C in the
terminal window.
BI
Set Burst Interval
Syntax:
BI x, where x is the number of seconds between the start of 2
consecutive bursts, from 1 to 10,000 seconds.
Description:
In Burst Mode (MODE = 1), the burst interval is the number of seconds
between two consecutive bursts, each burst composed of a number of
samples per burst (specified by the SB command).
Example:
L200x:> BI 900
New Seconds between Bursts: 900
MA
Set measurements per average
Syntax:
MA x
The exact result depends on whether ambient light rejection (ALR) is
active. Starting with firmware version 1.4 (May, 2017), ALR is on by
default, and the actual number of samples averaged will be approximately
x/10. That is, MA 32 will result in 3 ALR cycles per average.
With or without ALR, 32 is the maximum MA value that allows 1 sample
per second.
Description:
Each recorded or displayed measurement is based on an average of
measurements. MA sets the number of measurements per.
If no value follows command, prompts will be displayed for the value.
Example:
L200x:> ma 32
Measurements per average: 32 (3 ambient subtractions)
OM
Set Operating Mode
Syntax:
MODE x or OM x
Description:
Sets the Operating Mode to one of the following types:
0 = Fixed rate
1 = Burst Mode
OM command only without a parameter will initiate a prompt for the
Operating mode.
Example:
L200x:> OM 1
Operating mode: Burst mode (1)
LISST-200X User’s Manual
65
LISST-200X Operation Details
PD,
STOPDATA
Set Stop Condition Data
Syntax:
PD x or STOPDATA x where x is the stop condition data as described
below
Description:
The PD command sets the stop condition data to be used when the
collection data. The stop condition data is used with the Stop Condition
settings as follows.
• If the Stop Condition is Depth Stop (option 1) the input will be stop
depth in meters.
• If the stop condition is set to Time/Date Stop (option 2) the input for PD
will be the stop date and time.
• If the Stop Condition is Fixed Number Stop (option 5) the input will be
the number of samples to collect before stopping.
• The PD setting is ignored for Mechanical Switch Stop (option 3),
Digital Input Stop (option 4) or Maximum memory or Low Battery Stop
(option 6).
Example:
If Stop Condition =1 (Depth Stop):
L200x:> PD 3
Depth stop at 3 meters
If Stop Condition = 2 (Time/Date Stop):
L200x:> PD 2016/12/31 23:59:59
Time stop at 2016/12/31 23:59:59
If Stop Condition = 5 (Fixed Number of Samples):
L200x:> PD 100
Stop after 100 samples.
PSDMODEL
Set particle model used for on-board size distribution calculations
(firmware 2.1 or higher)
Syntax:
PSDMODEL x
Description:
If x = 0, PSD processing with use the spherical particle model.
If x = 1, the irregular (also called “random”) particle model will apply.
This setting applies only when real-time PSD processing is active. It is
available only in firmware 2.1 and higher. Also see SAVEPSD,
SENDDATA, and Autonomous Real-time Data Processing on page 52.
Example:
L200x:> PSDMODEL 1
Optical Model: 1 (irregular)
SAVEDATA
Save Data Setting—USE WITH CAUTION
Syntax:
SAVEDATA x, where x is 1 or 0
Description:
SAVEDATA 0 turns off storing of raw data in the instrument’s memory.
This is only for unusual circumstances.
Example:
L200x:> SAVEDATA 1
1: Raw data will be saved during logging.
L200x:> SAVEDATA 0
WARNING: Data will NOT be saved during logging!
Is this really what you want [N] ? n
LISST-200X Operation Details
66
LISST-200X User’s Manual
Cautions:
Be very careful with this setting!
Issue a DS command to verify the status of the store mode setting. A
warning will be displayed as part of the DS status if the store mode is
disabled:
SAVEPSD
Control saving of processed PSD files (firmware 2.1 or higher)
Syntax:
SAVEPSD x
Description:
If x = 1, during sampling the instrument will generate a .CSV file
containing processed particle size distributions, in addition to the normal
raw .RBN file.
If x = 0, only the normal .RBN files will be saved.
For more details, see Autonomous Real-time Data Processing on page
52. This command requires firmware version 2.1 or higher.
Example:
L200x:> SAVEPSD
PSD data will be saved during logging.
SB
Set Samples per Burst
Syntax:
SB x, where x is the number of samples per burst.
Each sample is taken at the sample interval (in seconds) set by the SI
command.
If no value follows command, prompts will be displayed for the value.
Example:
L200x:> SB 10
New Samples per Burst: 10
SC
Set Clock with time and date
Syntax:
SC yyyy/mm/dd hh:mm:ss
Where yyyy=year, mm=month, dd=day, hh=hour (24 hour format),
mm=minute, ss=seconds
If no values follow the “SC” or “sc” command, prompts for entering the
time and date will be displayed.
Example:
L200x:> SC 2020/12/20 13:20:18
Time set to 2020/12/20 13:20:18
SENDDATA
Transmit data during sampling (firmware 2.1 or higher)
Syntax:
SENDDATA x
Description:
The value of x sets whether data will be transmitted from the instrument’s
RS232 interface during sampling. Only options 0 and 4 are suitable for
normal operation.
0 = Do not send data (default)
1 = Multi-line troubleshooting data
2 = NMEA-format brief data (for glider applications)
3 = Single-line raw (CANNOT be processed by LISST-200X software)
4 = Processed PSD data; activates real-time processing and sends
processed data after each sample. For details, see Autonomous Real-
time Data Processing on page 52. This option requires firmware version
2.1 or higher.
Example:
L200x:> SENDDATA 0
Setting: 0 (no data)
LISST-200X User’s Manual
67
LISST-200X Operation Details
SENSORPOWER
Control 12V power to auxiliary connector during logging
Syntax:
SENSORPOWER x, where x is 1 or 0
Description:
If x = 1, the 12V power output on the auxiliary connector (if it is
suitably configured) will turn on during data collection with the GO
command. If running in burst mode, the power will turn off between
bursts. Logging will also incorporate a warmup time, set with the
SENSORWARM command.
If SENSORPOWER is entered without an x argument, it will prompt
for a value and for a sensor warmup time.
Example:
L200x:> SENSORPOWER 1
Power output will turn on during logging, with 5 second
warmup
SENSORWARM
Warmup time when powering external sensors
Syntax:
SENSORWARM x, where x is in seconds
Description:
If SENSORPOWER is on, SENSORWARM sets the number of
seconds the LISST-200X will wait for external sensors to warm up
before it proceeds with sampling.
Example:
L200x:> SENSORWARM 10
Power output, when used, will turn on with 10 second
warmup.
Power output during logging is on.
SI
Set Sample Interval
Syntax:
SI x, where x is the number of seconds between samples, from 1 to
10,000.
Description:
In either fixed-rate or burst mode, the sample interval is the number of
seconds between two consecutive samples, each composed as an
average of a number of measurements (specified by the MA command).
Example:
L200x:> SI 5
Seconds between samples: 5
SP,
STOPCOND
Set Stop Condition
Syntax:
SP x or STOPDATA x, where x is the stop condition code described below
Description:
The SP command sets the stop condition to be used when collecting data.
The stop condition options are:
1 = Depth Stop
2 = Time/Date Stop
3 = Mechanical Switch Stop
4 = Digital Input Stop
5 = Number of Samples Stop
6 = Maximum memory or Low Battery Stop
If no value follows command, prompts will be displayed for the value. After
setting a stop condition of Depth, Time/Date or Number of Samples, use
STOPDATA to enter an appropriate parameter value.
Example:
L200x:> SP 5
Stop condition: 5
LISST-200X Operation Details
68
LISST-200X User’s Manual
ST,
STARTCOND
Set Start Condition
Syntax:
ST x or STARTCOND x where x is the start condition code described
below
Description:
The ST command sets the start condition to be used when the GO
command is issued. The start condition options are:
1 = Depth Start
2 = Time/Date Start
3 = Mechanical Switch Start
4 = Digital Input Start
5 = Delay Start
If no value follows command, prompts will be displayed for the value.
After setting the start condition to Depth, Time or Delay, use STARTDATA
to set the applicable condition value.
Example:
L200x:> ST 5
Start condition: 5
TD,
STARTDATA
Set Start Condition Data
Syntax:
TD x or STARTDATA x where x is the start condition data described below
Description:
The TD command sets the start condition data to be used when the GO
command is issued. The start condition data is used with the Start
Condition setting as follows:
• If the Start Condition is Depth Start (option 1) the input will be start
depth in meters.
• If the start condition is set to Time/Date Start (option 2) the input for TD
will be the start date and time.
• If the Start Condition is Delay Start (option 5) the input will be time
delay in minutes.
• The TD setting is ignored for Mechanical Switch Start (option 3) or
Digital Input Start(option 4).
If no value follows command, prompts will be displayed for the value.
Example:
If Start Condition = 1 (Depth Start)
L200x:> STARTDATA 3
Start condition: Depth Start at 3 meters
If Start Condition = 2 (Time/Date Start)
L200x:> STARTDATA 2020/12/31 23:59:59
Start condition: Time Start at 2020/12/31 23:59:59
If Start Condition = 5 (Delay Start)
L200x:> STARTDATA 2
Start condition: Delay Start with 2 minute delay
ZD
Reset Depth Sensor Offset
Syntax:
ZD
LISST-200X User’s Manual
69
LISST-200X Operation Details
Description:
The ZD (or zd) command resets the depth sensor offset so that the sensor
reads a depth of 0m at zero depth (in air). You must issue the ZD
command, then select 1 (yes) or 0 (no) to reset depth sensor
Example:
L200x:> ZD
Depth sensor offset adjust.
Instrument must be at zero depth and similar temperature to
field conditions.
Reset depth sensor offset? (1=yes,0=no): [0] ? 1
Previous offset was -140.65. New offset is -4.37.
Previous depth was -136.28 meters.
New depth using corrected offset is 0.00 meters.
3.
Acquisition/Action commands
DL
Delete file from LISST memory
Syntax:
DL filename, where filename is the name of the file to be deleted.
Description:
DL command is used to delete file from memory.
Example:
L200x:> DL L3231513.RBN
Delete the file 'L3231513.RBN'?
Y for Yes or N for No: [N] ? Y
Deleting L3231513.RBN
Deleted.
Cautions:
WARNING: Make sure the file being deleted has already been offloaded.
Once the file is deleted it cannot be recovered.
GO
Start Data Collection using current Settings
Syntax:
GO
Description:
Starts Fixed Rate or Burst Mode Data collection using current settings.
Example:
L200x:> GO
Starting sampling.
Start condition: Delay Start with 2 minute delay
Stop condition: Stop after 100 samples
Low-power sleep until 2021/12/10 17:39:56
zzz...Waiting until 2021/12/10 17:39:56
Cautions:
To stop data acquisition before it is complete, send 2 Ctrl-C characters.
GX
Grab sample and transmit it
Syntax:
GX
Description:
Acquires single averaged sample and displays the result to the screen.
Example:
Input: GX
Output: {
…. 36 ring values + 24 Aux parameters
}
Notes:
The GX command does not store the sample to a datafile.
LISST-200X Operation Details
70
LISST-200X User’s Manual
ZS
Collect and transmit background scattering data
Syntax:
ZS
Description:
Acquires 20 averaged sample and displays the result to the screen.
Evaluates the data for quality as a background measurement.
Example:
Input: ZS
Output: {
…. 36 ring values + 24 Aux parameters
}
repeat 20 times, then messages about the quality of the data.
Notes:
The ZS command does not store the sample to a datafile.
ZZ
Go into deep sleep mode (minimum power consumption)
Syntax:
ZZ
Description:
Sends LISST-200X in to low power sleep mode. Instrument will send a
message every 30 seconds to indicate it is sleeping. It can be wakened at
any time by sending 2 control-C characters about ½ second apart.
Example:
L200x:> ZZ
Low-power sleep.
zzz...Press Stop or C to wake up.
zzz...Press Stop or C to wake up.
LISST-200X User’s Manual
71
LISST-200X Operation Details
F.
Performance Optimization
Background
Measurement
and its
Importance
As discussed in earlier sections, the LISST-200X measures light
scattered at small angles, to characterize particles in water. However, a
certain amount of scattering is also generated by pure water, and by
the LISST-200X’s own optics and windows. We call this the
background scattering. In order to determine the scattering contributed
by particles, the background scattering must be accurately measured
and subtracted from the total scattering measured. The measurement
of background scattering is sometimes called a “zscat”, meaning the
signal measured in water with zero “scatterers” or particles.
The minimum concentration of particles measurable by the LISST-200X
is very sensitive to the quality of the background scatter file. The lower
the concentration in the water being measured, the more important the
quality of water and the care used when measuring the background.
When measuring in water with high particle concentrations, the relative
signal-to-background noise ratio increases, and final results will be less
sensitive to flaws the background. However, a background should
always be collected before an experiment.
Techniques for
measuring
background
Cleanliness
It is essential to use clean, bubble-free water for the background
measurement. The water can be fresh or salt water. For most
applications, steam-distilled water works well. Steam-distilled water is
typically available in one-gallon containers. We have found that this
distilled water tends to be a bit cleaner than typical bottled or packaged
water. We buy steam-distilled drinking water and filter it through a
0.2µm filter.
A small volume horizontal chamber has been provided for submerging
the optics while acquiring a background. It is inserted between the
windows of the optics. The instrument should be placed horizontally on
the supplied white plastic supports. However, plastic bags, Tupperware
or clean containers of any kind can be used to hold water for a
background measurement. The instrument’s optics end can be
submerged in them vertically, as long as the optics are completely
submerged and there is no blockage of the windows by bubbles or
other objects.
Because the area surrounding the windows is submerged, it is
important to thoroughly clean and rinse this part of the instrument
before acquiring a background.
Toothbrush, liquid soap and water works well for cleaning the optical
end of the instrument. Do not use abrasive powders, which will scratch
the windows and degrade instrument performance.
LISST-200X Operation Details
72
LISST-200X User’s Manual
Watch for
Outgassing and
Bubbles
Water may contain dissolved air that develops into bubbles as the
water warms. Bubbles, especially if they cling to the windows of the
LISST-200X, will greatly modify the scattering pattern, rendering the
background useless. Be sure to remove any bubbles from the windows
before acquiring a background. Use a squirt bottle or pipette to blow
the bubbles off the window.
Ambient Light
Rejection
Starting with firmware version 1.4 in May, 2017, the LISST-200X
automatically measures and subtracts light that is not produced by
scattering from its laser. This Ambient Light Rejection (ALR) prevents
sunlight or other bright sources from contaminating the scattering
measurements. Without ALR, as in earlier LISST instruments, high
ambient light levels, if not physically blocked from the LISST’s optical
aperture, could falsely elevate the apparent concentration of small
particles.
You can verify that ALR is active by connecting to the instrument with
LISST-200X, then opening the terminal window. Type DS and press
enter, and look for “Ambient light subtraction: on” in the output. If it is
not shown, you should upgrade the firmware. Contact Sequoia
Scientific for assistance.
ALR works by measuring the signal on the LISST-200X’s 36 detectors
first with the laser on, then with the laser off. This on-off process is
repeated 3 times per second. Because of the time required for the laser
and electronics to change between the on and off states, rapidly
changing ambient light may not be completely rejected. Therefore, if
working in bright sunlight, near the water surface, and in conditions
where the sunlight is rapidly modulated by waves, ambient light could
still have an effect. ALR will reduce the average effect, but the
changing ambient conditions could increase sample-to-sample noise. In
those worst-case conditions it is still advisable to shade the LISST-
200X optics from ambient light. The same is true if operating under
intense artificial light modulated by AC power.
Optical
Alignment
The background measurement (described step-by-step on page 29) is
the best source of information on the current health of the LISST-200X.
It provides information on the current functionality of all of the major
systems including laser, ring detector, data collection electronics, and
optical alignment. The LISST-200X is a sensitive optical device. The
laser must be aligned such that the focused spot is centered on the
hole in the center of the ring detector. If this alignment in not correct
the instrument will not function correctly. The LISST-200X software
automatically detects and warns of serious misalignment, during the
background measurement process.
The first indicator of an alignment problem is a severe drop in the
transmitted laser power when compared to the factory laser power.
The laser power and laser reference values will change over time but
they should track together.
LISST-200X User’s Manual
73
LISST-200X Operation Details
The second indictor of misalignment is high value on the inner rings.
The inner rings are shown on the left side of the background display.
The most important rings for misalignment indication are rings 1-4 in
the background scatter file. If the focused laser starts to move away
from the center of the ring detector it will scatter more light onto the
inner rings. This will cause the inner ring values to be much higher
than the factory values, and cause the transmitted laser power to
decrease.
It must be noted that low laser power or high inner rings may not
always indicate misalignment. Low laser power can also occur
because the windows or water is dirty or if there are bubbles on the
window. Large particles or bubbles can cause the inner rings to be
higher than factory values. All of the other possibilities must be
eliminated before it can be concluded that the instrument is misaligned.
If necessary, the LISST-200X can be aligned by Sequoia Scientific or
an authorized service representative.
LISST-200X Operation Details
74
LISST-200X User’s Manual
G. Instrument Mounting, Deployment Orientation,
Storage and Maintenance
Horizontal
Mounting
The LISST-200X measures scattered light to obtain particle size
distributions. Therefore, it is important to keep particles from sticking
to the window surface. The first line of defense it to keep the window
surface vertical by mounting the instrument horizontally. This is most
critical when the water motion is slow. When working in rivers or in a
towed or profiling application the orientation is not as critical.
The center section of the main pressure case has a slightly smaller
diameter than the ends. The smaller diameter section is the
suggested mounting area. The U-shaped mounting block provided
with the instrument fit this diameter and can be used for mounting the
instrument. The clamps supplied with the instrument can also be
used.
Spare clamps for mounting the instrument are available from
Sequoia.
Electrical
Isolation
To reduce the corrosion of the aluminum parts, a zinc anode is
attached to the connector endcap. This anode must be exposed to
the water for it to be effective.
When mounting the instrument be sure to electrically isolate the
instrument from all other metal. Any contact with other metal can
greatly increase the rate of corrosion. Isolate the instrument with
rubber or plastic to keep the LISST-200X from being the sacrificial
anode for the mounting hardware. Failure to properly isolate the
instrument from all other metal will void the warranty.
A set of stainless steel clamps with rubberized interiors ships as part
of the instrument package. Use these or the plastic clamps for
mounting, and make sure that no metal is in direct contact with the
pressure housing or other components of the instrument. If you lose
the spare clamps, replacement clamps are available for purchase
from Sequoia.
Cleaning the
Optical Windows
The condition of the windows is critical to the performance of the
LISST-200X. Care must be taken when cleaning the windows. The
windows and the instrument should be rinsed thoroughly with fresh
water after each deployment. The windows should be cleaned with a
soft cloth or lens tissue. Liquid detergent/soap and water may be
used. For removing grease spots, finger prints, etc., alcohol may be
used. Do not use stronger solvents such as Acetone or Toluene.
LISST-200X User’s Manual
75
LISST-200X Operation Details
Abrasive powders must never be used near the optics windows; they
will scratch the windows and degrade instrument performance.
Orientation
Relative to
Current Flow
In situations with significant current flow, it is best to orient the
instrument perpendicular to the current, so water can flow as directly
as possible through the optics end of the instrument. The diagram
below shows the optimum orientation with respect to current flow
Storage
For full protection, always store the LISST-200X in its foam-lined
shipping case.
Calibrations and
Adjustments
The LISST-200X should not require any adjustment or calibrations.
The pressure sensors can be re-calibrated by Sequoia, if desired.
The performance of the instrument can be checked with the use of a
sample of particles of a known size distribution. Standard particles
can be obtained from manufacturers of standard particles, such as
Whitehouse Scientific (http://www.whitehousescientific.com/) or PTI –
Powder Technology Inc. (http://www.powdertechnologyinc.com/).
Comparison of the LISST-200X’s performance with that of other
particle-size measuring devices that are not based on laser diffraction
may result in inconsistencies because the instruments may measure
different properties of the particles to interpret particle size.
Appendix A: Technical Specifications
76
LISST-200X User’s Manual
Appendix A: Technical Specifications
• Parameters measured/derived:
o Particle Size distribution
o Total Volume Concentration
o Optical Transmission
o Depth
o Temperature
o Volume Scattering Function
• Particle size range: 1.00 – 500 µm in 36 log-spaced classes
• Optical path length: 2.5 cm
• Optical transmission: 12-bit resolution
• Analog inputs (availability depends on configuration)
o Measurement range: 0 to 5.2 V
o Resolution: 0.1 mV (inputs 1 and 2); 1.5 mV (input 3)
o Uncertainty: 1.5% of reading + 2 mV
o Input impedance: 500 kΩ
o Time constant: 250 ms
• Digital input (availability depends on configuration)
o Nominal logic levels: 0 and 3V
o Voltage tolerance without damage: -12 to 12V
• Analog outputs (availability depends on configuration)
o Output range: 0.1 to 3.0 V
o Resolution: < 1 mV
o Output impedance: 1 kOhm
o Uncertainty: 1.5% of reading + 2 mV
• Data storage memory: 1GB (~12 million measurements)
• Maximum sample speed: 1 size distribution per second
• Temperature sensor
o Range: -5 to 45 °C
o Resolution: 0.001 °C
o Uncertainty: approximately 1 °C
• Depth Sensor
o Range: 0 to 600 m of sea water
o Resolution: 1 cm
o Uncertainty: approximately 1% of reading (if atmospheric offset zeroed)
• Input power:
o Operating range: 9 to 24 V
o Current during active sampling:
▪ LISST-200X: 100 mA @ 12V
▪ LISST-Black: 220 mA @ 12V
▪ LISST-HAB: 180 mA @ 12V
o Current while waiting for user command: < 50 mA @ 12V
o Current during sleep: < 3 mA
• Dimensions: 10.03 cm (3.95”) diameter x 63.9 cm (26.18”) L
• Weight: 5.4 kg (11.8 lbs.) in air, 1.7 kg (3.8 lbs.) in water
• Depth rating: 600 m
Appendix B: Particle Size Bins
78
LISST-200X User’s Manual
Appendix B: Particle Size Bins
There are 36 size ranges logarithmically placed from 1.00 – 500 microns in diameter. The
upper size in each bin is approximately 1.18 times the lower, with the exception of bin 1. The
table below shows the lower and upper limit of each size bin in microns, together with the
median size (also in microns) for each size bin. The sizes are the same for both Spherical
and Randomly Shaped inversions.
Size bin #
Lower
Upper
Median
1
1.00
1.48
1.21
2
1.48
1.74
1.60
3
1.74
2.05
1.89
4
2.05
2.42
2.23
5
2.42
2.86
2.63
6
2.86
3.38
3.11
7
3.38
3.98
3.67
8
3.98
4.70
4.33
9
4.70
5.55
5.11
10
5.55
6.55
6.03
11
6.55
7.72
7.11
12
7.72
9.12
8.39
13
9.12
10.8
9.90
14
10.8
12.7
11.7
15
12.7
15.0
13.8
16
15.0
17.7
16.3
17
17.7
20.9
19.2
18
20.9
24.6
22.7
19
24.6
29.1
26.7
20
29.1
34.3
31.6
21
34.3
40.5
37.2
22
40.5
47.7
43.9
23
47.7
56.3
51.9
24
56.3
66.5
61.2
25
66.5
78.4
72.2
26
78.4
92.6
85.2
27
92.6
109
101
28
109
129
119
29
129
152
140
30
152
180
165
31
180
212
195
32
212
250
230
33
250
297
273
34
297
354
324
35
354
420
386
36
420
500
459
LISST-200X User’s Manual
79
Appendix C: Data File Formats
Appendix C: Data File Formats
File Types and Extensions Used by the LISST-200X.
Extension
Description
Format
Comments
.RBN
Raw Data
Binary
Raw data file produced by the LISST-200X. In
addition to raw scattering data, RBN files
contain a complete description of the
instrument configuration, its most recent
background scattering, and the factory
background.
.CSV
Processed Data
ASCII
Comma-delimited file containing size
distributions and meta data. Details on page
80. For information about generating these
files in the LISST-200X, see Autonomous
Real-time Data Processing on page 52.
Secondary files not used for standard processing
.BGT
Background File
ASCII
Comma-delimited file containing the raw ring
values from a single background measurement
(1 column, 59 rows)
.RTX
Raw Data
ASCII
Comma delimited ASCII representation of raw
data (59 columns, 1 measurement per row).
NOTE: Sequoia software does not process
RTX files. They are only for troubleshooting or
special applications.
Appendix C: Data File Formats
80
LISST-200X User’s Manual
Processed Data
Format
The values in the processed data file (.CSV extension) are stored in the
order shown below. Each sample is stored in one row.
Column #
Parameter
1:36
Volume concentration for size class 1 through 36 [µL/L]
37
Laser transmission Sensor [mW]
38
Supply voltage in [V]
39
External analog input 1 [V] (fluorometer 1 in LISST-HAB & LISST-Black)
40
Laser Reference sensor [mW]
41
Depth in [m of sea water]
42
Temperature [C]
43
Year
44
Month
45
Day
46
Hour
47
Minute
48
Second
49
External analog input 2 [V] (fluorometer 2 in LISST-HAB & LISST-Black)
50
Mean Diameter [μm] (calculated from fully processed size distribution)
51
Total Volume Concentration [PPM] (calculated from fully processed size
distribution)
52
Relative Humidity [%]
53
Accelerometer X [not presently calibrated or used]
54
Accelerometer Y [not presently calibrated or used]
55
Accelerometer Z [not presently calibrated or used]
56
Raw pressure [most significant bit]
57
Raw pressure [least significant 16 bits]
58
Ambient Light [counts – not calibrated]
59
External analog input 3 [V] (fluorometer 3 in LISST-HAB & LISST-Black)
60
Computed optical transmission over path [dimensionless]
61
Beam-attenuation (c) [m-1].
LISST-200X User’s Manual
81
Appendix C: Data File Formats
ASCII Raw Data
Format
The values in the raw ASCII data file (.RTX extension) are stored in the
order shown below. Each sample is stored in one row.
RTX files are only useful for troubleshooting or special applications.
Sequoia does not supply software for processing them.
Column #
Parameter
1:36
Raw ring values [counts]
37
Laser transmission Sensor [counts]
38
Supply voltage [0.01 V/count]
39
External Analog input 1 [10-4 V/count] (fluorometer 1 in LISST-HAB & LISST-
Black)
40
Laser Reference [counts]
41
Depth [0.01 m/count + 10 m]
42
Temperature [1 m°C/count + 5 °C]
43
Year
44
Month
45
Day of month
46
Hour
47
Minute
48
Second
49
External Analog input 2 [10-4 V/count] (fluorometer 2 in LISST-HAB & LISST-
Black)
50
Sauter Mean Diameter [0.1 µm/count] (estimated)
51
Total Volume Concentration [0.1 PPM/count] (estimated)
52
Relative Humidity [%]
53
Accelerometer X [counts] [not presently used]
54
Accelerometer Y [counts] [not presently used]
55
Accelerometer Z [counts] [not presently used]
56
Raw pressure [most significant bit]
57
Raw pressure [least significant 16 bits]
58
Ambient light [counts]
59
External Analog input 3 [10-4 V/count] (fluorometer 3 in LISST-HAB & LISST-
Black)
Appendix D: MATLAB Data Processing
82
LISST-200X User’s Manual
Appendix D: MATLAB Data Processing
Two Matlab functions for processing data are provided on the instrument USB card. They
are also available from the ‘Software and Downloads’ tab on the LISST-200X webpage. The
processing is split into two steps: (1) reading in and applying corrections to raw data files
(.RBN) and (2) inverting the corrected scattering to a particle size distribution.
The same results can be achieved by processing data files using the LISST-200X software.
However, the following functions allow Matlab users to write their own processing and
plotting code.
The first function is used to read in raw data from an RBN file. The function will import the
data, apply the necessary corrections, and return the corrected data in a structure. The
syntax is as follows:
‘Datafile’ is the path to a binary .RBN file downloaded from a LISST-200X.
Optionally, you may specify a different clean water background file (.BGT) as a second
argument. The data will then be corrected using the specified background file instead of the
background contained in the RBN file.
‘RBNdata’ is a structure with the following fields:
Field
Description
cscat
Corrected scattering
date
Timestamp in Matlab datenum
transmission
Optical transmission
depth
Depth in meters
temperature
Temperature in degrees Celsius
estMeanDiameter
Estimated Sauter mean diameter (µm)
estTotalConc
Estimated total concentration (µL/L)
Lp
Transmitted laser power (mW)
Lref
Laser power reference (mW)
analog1
Analog input 1 (V)
analog2
Analog input 2 (V)
analog3
Analog input 3 (V)
RBNdata = getscat_L200X(datafile)
RBNdata = getscat_L200X(datafile,backgroundFile)
LISST-200X User’s Manual
83
Appendix D: MATLAB Data Processing
supplyVolts
Supply voltage (V)
humidity
Internal instrument relative humidity (%)
accelXYZ
Accelerometer X, Y, and Z (not presently calibrated or used)
raw
Raw data as it appears in the RBN file
factory_bkgrd
The factory background (corrections applied to aux data)
bkgrd
User collected background (corrections applied to aux data)
ambientLight
Counts of ambient light removed from ring values
config
Structure containing various instrument information
dcal
Ring area coefficients
Ta
Vector to convert cscat to estimated total area concentration
Tv
Vector to convert cscat to estimated total volume concentration
The second function inverts the corrected scattering to a particle size distribution. The
syntax is as follows:
The first argument is the corrected scattering (cscat) from the structure returned from
‘getscat_L200X.’ The following three argument are set to zero or one.
Random – If set to 1, the randomly shaped particle matrix is used to invert the data. If set to
zero, the spherical particle matrix is used. See Appendix G: Spherical and Random Shape
Particle Models for more information.
Sharpen – If set to 1, the function checks the width of the size distribution and increases the
number iterations if the size distribution is wide (recommended).
ShowProgressBar – If set to 1, a progress bar will display the processing status.
The function will return the volume distribution in µL/L (vd) and the midpoint of each size bin
in microns (dias).
[vd, dias] = invert_L200X(Cscat,Random,Sharpen,ShowProgressBar)
Appendix E: Connectors
84
LISST-200X User’s Manual
Appendix E: Connectors
IMPORTANT NOTE
In September 2019, starting with serial number 2116, LISST-200X instruments and cables
use connectors manufactured by SubConn. Previous instruments have connectors from
Teledyne Impulse. Connectors from the two manufacturers are very similar, but are not
guaranteed to be compatible.
For reliable connections, only mate SubConn cables with SubConn bulkhead
connectors. Sequoia’s warranty does not cover any problems arising from
mismatched connectors.
To Identify the connector type, note:
1. SubConn bulkhead connectors have brass, golden-colored bodies. They may
develop a green tint after exposure to sea water.
2. SubConn cables and dummy plugs have red locking sleeves.
3. Impulse bulkhead connectors have stainless steel, grey-colored bodies.
4. Impulse cables and plugs have black locking sleeves.
5. The name of the manufacturer is embossed on the rubber body of each connector.
SubConn
(as of September 2019)
Impulse
(older instruments)
LISST-200X User’s Manual
85
Appendix E: Connectors
The LISST-200X has 3 separate underwater connectors: A 5-pin, a 6-pin, and a 3-pin
connector. The photograph shows the placement of each connector. The following text
describes the detailed wiring for each connector. Note that in new instruments starting in
September 2019, these are SubConn connectors with brass bodies.
LISST-200X Connector Endcap
Small External Battery Housing
Large External Battery Housing
Appendix E: Connectors
86
LISST-200X User’s Manual
Communications and Power Connector (5 pin connector)
Bulkhead connector: SubConn MCBH5M
Mating cable connector: SubConn MCIL5F
Connector Pin #
Use
1
Power/Serial Ground
2
External Power In (+9 to 24 V)
3
Power/Serial Ground
4
Serial Out (to DB-9 Pin 2)
5
Serial In (to DB-9 Pin 3)
BioBlock Connector (3 pin connector) (see page 93 for BioBlock details)
Bulkhead connector: SubConn MCBH3M
Mating Cable Part Number: SubConn MCIL3F
Connector Pin #
Use
1
Ground
2
Signal Out
3
+6V Power Out
LISST-200X User’s Manual
87
Appendix E: Connectors
Auxiliary I/O Connector (6 pin connector)
Bulkhead connector: SubConn MCBH6M
Mating Cable Part Number: SubConn MCIL6F
NOTE: This connector has 3 possible configurations. The configuration in a particular
instrument can be found by issuing the CONFIG command at the command prompt.
Analog and Digital Configuration
Connector Pin #
Use
1
Ground
2
Digital In #1
3
Digital In #2
4
+12V out
5
Analog In (0 to 5V)
6
Ground
Analog Input Configuration
Connector Pin #
Use
1
Ground
2*
Analog In #3 (0 to 5V)*
3
Ground
4
+12V out
5
Analog In #1 (0 to 5V)
6
Analog In #2 (0 to 5V)
* Analog input #3 is available in LISST-200Xs with revision F or later electronics.
This is standard starting with serial number 2151, and available as a factory
upgrade on earlier units. In earlier versions, this pin is not connected.
Analog Output Configuration
Connector Pin #
Use
1
Ground (Power)
2
Analog Output: Mean Size
3
Ground (signal)
4
Analog Output: Total Concentration
5
Ground (signal)
6
Power In (9 to 24VDC)
Appendix E: Connectors
88
LISST-200X User’s Manual
Large and Small External Battery Connectors (5 pin connectors)
Bulkhead connector: SubConn MCBH5F
Mating Cable Part Number: SubConn MCIL5M
To LISST Connector (5-pin Female Bulkhead)
Connector Pin #
Use
1
Ground
2
Battery Power Out (+9V nom. 6-15V)
3
Ground
4
Serial Out (to DB-9 Pin 2)
5
Serial In (to DB-9 Pin 3)
To PC Connector ( 5-pin Male Bulkhead)
Connector Pin #
Use
1
No Connection
2
No Connection
3
Ground
4
Serial Out (to DB-9 Pin 2)
5
Serial In (to DB-9 Pin 3)
LISST-200X User’s Manual
89
Appendix F: Particle Shape Models
Appendix F: Particle Shape Models
Particles of different shapes scatter light differently, so the mathematical model used for
processing must be appropriate to the particles under observation. Two models are offered
by Sequoia: the spherical model, and the random or irregular shape model. The resulting
inversion of data will differ slightly for the two models.
The spherical particle model performs the mathematical inversion of scattering data under
the assumption that the particles that scattered light are all spheres. Light scattering by
spheres of any size and refractive index is modeled by Mie theory. According to Mie theory,
the angular scattering depends on the size of the spherical particle, and its refractive index
relative to water. Sequoia employs the full Mie scattering model, without simplifications, for
inversion of LISST measurement as a distribution of spheres. The Mie solution is a
generalized solution to the scattering of light from spheres and is commonly used as the
standard model by all laboratory laser diffraction instrument manufacturers. However, it is
not completely accurate for non-spherical particles found in nature.
Sequoia provides an alternate model that is derived empirically from natural, irregularly-
shaped mineral particles, rather than assuming spheres. The exact details of how this
scattering model was established are described in a paper by Agrawal et al. [Light scattering
by random shaped particles and consequences on measuring suspended sediments by
laser diffraction. Journal of Geophysical Research, Vol. 113, C04023, doi:10.10-
29/2007JC004403.], which can be downloaded from the library section on Sequoia’s
website (www.SequoiaSci.com/library). The method is also briefly described in this article:
http://www.sequoiasci.com/article/random-shaped-particles-lissts/. It is noteworthy that no
other instrument manufacturer has a scattering model for irregular particles. Instead, they
often use a Mie model with large imaginary refractive index, in effect, assuming the particles
to be highly absorbing (i.e. black). This is an obvious fudge with consequences!
Which model should you use? Sequoia’s software allows you to use both, then compare the
results. But in general, the random model is best for any in-situ application.
Spherical
Random (Irregular)
Best use
Particles known to be
predominantly spherical, e.g.
analytical microspheres
Appropriate for natural mineral
particles
Effect of
misapplication
If used with natural irregular
particles, tends to invent fine
particles.
If used with spherical
particles, tends to
underestimate their size.
Relation to other
instruments and
methods
Usually produces the closest
match to other laser diffraction
particle sizers (although they
may still differ due to index of
refraction)
Usually produces the best
match to analysis by sieving.
Appendix G: LISST-Black & LISST-HAB
90
LISST-200X User’s Manual
Appendix G: LISST-Black & LISST-HAB
LISST-Black and LISST-HAB are integrated packages combining a LISST-200X with
fluorometers optimized for detecting either petroleum (LISST-Black), or harmful algal blooms
(LISST-HAB). The fluorometers are Turner Designs Cyclops-7F’s, mounted with special
brackets and cabling made by Sequoia Scientific. The LISST-200X supplies power to the
fluorometers, and digitizes their analog outputs. The digitized fluorometer outputs are
included in each data sample produced by the LISST, at rates up to one per second.
Fluorometer Output Data
The fluorometer outputs are reported in volts, included in the standard .CSV files produce by
the LISST-200X software. For format details, see Appendix C: Data File Formats. The
outputs are not calibrated to an absolute standard.
Firmware Configuration
To check that the fluorometers will be powered whenever the LISST collects data, type the
DS command at the LISST-200x:> prompt, and look for “External sensor power during
logging: on”. If it is not on, send the command SENSORPOWER,3 . This will enable power
to the fluorometers, with a 3-second warmup before the first sample.
Power Consumption
Because of additional power consumed by the fluorometers, the LISST-Black and -HAB
cannot be reliably powered by a USB cable during sampling. A battery housing or other
power source must be connected.
Standard Fluorometer Sets
Fluorometer target,
Turner Designs part number
Power during
sampling (at 12V)
1
2
3
LISST-Black
Refined fuels
2110-000-G
Crude oil
2110-000-O
Chlorophyll
2110-000-C
2.7 W
LISST-HAB
Phycocyanin
2110-000-P
Phycoerythrin
2100-000-E
Chlorophyll
2110-000-C
2.2 W
LISST-200X User’s Manual
91
Appendix G: LISST-Black & LISST-HAB
LISST-Black/LISST-HAB Mounting instructions
Hardware:
Clamp Screws, 8-32 x ½” Socket Head Cap Screws, 316 Stainless Steel
Sensor Hold-down Set Screws, 8-32 x 3/16” Set Screws, flat tip, 316 Stainless Steel
Cover Screws, 6-32 x ¼” Flat Head Phillips Screws, 316 Stainless Steel
Tools:
5/64-inch hex L-key
9/64-inch hex L-key
#2 Phillips Screwdriver
1. Attach the Front Support to the LISST-200X with the Clamp Strap, cross nut and
clamping screw. Position the tab on the Front Support so it fits flush against the
endcap and aligns to the flat surface. Tighten the Clamp Screw.
2. Position the Mid Support on the LISST-200X so
the back edge is positioned on the pressure case
chamfer, as shown at right, and the holes for the
fluorometers line up with the Front Support.
Slightly tighten the Mid Support clamping screw so
that the Mid Support is held in place but can be
rotated as needed when inserting the sensors.
3. Insert all three Cyclops-7F Sensors. Note the
order of the sensors. The labels shown in the figure correspond to their labels in the
data file (see Appendix C: Data File Formats). When looking at the connector end of
the sensors, sensor 1 will be on the left and sensor 3 will be on the right. Do not
tighten the setscrews holding the sensors yet.
Appendix G: LISST-Black & LISST-HAB
92
LISST-200X User’s Manual
4. Rotate the Mid Support as needed to make the axis of the sensors parallel to the axis
of the LISST-200X. Tighten the Mid Support clamp making sure that the Mid Support
is pushed against the tapered part of the pressure case, as in step 2. The sensors
should be free to rotate and move forward and backward.
5. Attach the Cable Support onto the LISST-200X. Tighten the Cable Support clamping
screw slightly so that the Cable Support is held in place but free to rotate and slide
along the case.
6. If the LISST-HAB/Black will be used with the LISST-200X Frame, place the Frame
Clamp between the sensors and the LISST-200X pressure case before connecting
the LISST-HAB/Black cable. Installing the Frame Clamp after assembly is difficult
and not recommended.
7. Attach the LISST-HAB/Black cable to the sensors.
a. The protrusion in the mold needs to be on the top side.
b. Rotate the sensors to match the cable sockets and insert into the connector.
c. Secure the cable locking sleeves.
d. Slide the Cable Support towards the optics end so that cables are supported.
Do not tighten the clamp yet.
8. Position the sensors so they extend approximately 1/8” (3mm) in front of the Front
Support. Secure the sensors with the Hold-down Set Screws.
9. Place the Cover onto the assembly and adjust the position of the Cable Support so
that the Cover can be easily installed. Insert the Cover Screws into the holes at the
Front Support. Adjust the Cable Support position as need so that the holes in the
Cover line up with the holes in the Cable Support and secure it with the Cover
screws. Tighten the Cover and Cable Support clamp screws.
10. Connect the cable to the 6-pin connector on the LISST-200X.
11. Secure the cable to the handle with cable ties.