Key Pages of Document

The lowest negative

voltage bottom Maximum The highest positive voltage peak Peak to Peak Max – Min Low The most frequent value below the median High The most frequent value above the median Amplitude High – Low On positive oscillation 100  Amplitude High Max On negative oscillation 100  Amplitude Min Low Frequency Period 1 Average frequency Period N of N complete periods Duration Average duration of a complete cycle calculated on as many periods as possible Rising edge T1 = 10% Amplitude T2 = 90% Amplitude Trise = ...

The DAS 240 and DAS220 is a programmable recorder that makes it possible to measure and

record up to 200 channels(DAS240) or 10 channels (DAS220), voltages (from 1mV to 100V), currents, temperatures (thermocouple, Pt100, Pt1000), resistance etc… as well as logical channels. The dialog between the user and the recorder is made easier thanks to an intuitive interface on a wide capacitive and color 10’’1 touchscreen. The measurement parameters are easy to program. The recording is made directly on an internal 32 GB memory card. The DAS240 is composed wi ...

This chapter describes all possible triggers of the appliance.

They are used by: - the « setup » menu, with the A and B alarms - the « triggers » menu, with the start and stop parameters of the recording.  Edge/Level trigger  Edge trigger: you need a change of status o Example: Channel A1, rising edge, threshold = 0V: triggers only if the signal status changes from negative to positive  Level trigger: no need to pass the threshold o Example: Channel A1, high level, threshold = 0V: triggers only if the signal is positive  Alarms only have level triggers.  ...

4.4. Channels menu

By pressing the menu « channels », you access the table of all available analogic and temperature channels of your recorder. This table gives you a quick insight of the various adjustments of the channels. You can change these adjustments directly in this table by pressing the parameter you want to update. You can display 10 channels maximum. or keys : to display other channels You can access the setting of parameters channel by channel by pressing the name of the channel at the top of the table (A1, A2…) or by pressing on the bargraph of the channel (right si ...

TABLE OF CONTENT

1. IMPORTANT INFORMATION 1.1. SPECIFIC CAUTIONS ............................................................................................................................................... 1.1 1.2. SAFETY INSTRUCTIONS .......................................................................................................................................... 1.1 1.2.1. Conformity and limits of the appliance ........................................................................................................... 1.2 2. PRESENTATION 2.1. GENERAL ................................ ...

The accuracy of the thermocouples is specified in append.

Compensation of the cold welding: ± 0.5°C 11.4. Pt100/Pt1000 Record  Current generator:  Pt100: 1mA  Pt1000: 100µA  Domain of use: -200°C to 850°C  2 wires or 3 wires  Correction resistance in 2-wire mode: max. 30Ω  max. resistance in 3-wire mode: 50Ω  Accuracy at 20°C room temperature: ±0.3°C ± 0.1% of the measurement value 11.5. Resistance Probe Record 2 input ranges:  1kΩ:  Current generator: 1mA  Accuracy 1Ω  10kΩ  Current generator: 100µA  Accuracy 10Ω 11.6. Samp ...

SEFRAM DAS240 user guide recommended for: SPX50, MW 3517BF, 87, DAS220BAT, 9893.

⚠️ Download Safety

The Measuring Instruments Operation & user’s manual (80 pages) is completely safe to download (last scan date: 16/02/2026).

Related PDF Instructions for SEFRAM DAS240:

Measuring Instruments Equipment: Frequently Requested

Have some Questions about SEFRAM DAS240 Device?

Get valuable and actionable advices from other SEFRAM DAS240 device users. Describe your problem or question, regarding the using of DAS240, in as much detail as possible, and other users will be able to answer you.