RP-3R Pump for
GX-3R and GX-3R Pro
Operator’s Manual
Part Number: 71-0495
Revision: P3
Released: 12/2/21
www.GlobalTestSupply.com
Find Quality Products Online at:
[email protected]
RKI Instruments, Inc. warranties gas alarm equipment sold by us to be free from defects in materials,
workmanship, and performance for a period of one year from date of shipment from RKI Instruments,
Inc. Any parts found defective within that period will be repaired or replaced, at our option, free of
charge. This warranty does not apply to those items which by their nature are subject to deterioration
or consumption in normal service, and which must be cleaned, repaired, or replaced on a routine basis.
Examples of such items are:
Warranty is voided by abuse including mechanical damage, alteration, rough handling, or repair
procedures not in accordance with the operator’s manual. This warranty indicates the full extent of our
liability, and we are not responsible for removal or replacement costs, local repair costs, transportation
costs, or contingent expenses incurred without our prior approval.
THIS WARRANTY IS EXPRESSLY IN LIEU OF ANY AND ALL OTHER
WARRANTIES AND REPRESENTATIONS, EXPRESSED OR IMPLIED, AND
ALL OTHER OBLIGATIONS OR LIABILITIES ON THE PART OF RKI
INSTRUMENTS, INC., INCLUDING BUT NOT LIMITED TO, THE WARRANTY
OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. IN NO
EVENT SHALL RKI INSTRUMENTS, INC. BE LIABLE FOR INDIRECT,
INCIDENTAL, OR CONSEQUENTIAL LOSS OR DAMAGE OF ANY KIND
CONNECTED WITH THE USE OF ITS PRODUCTS OR FAILURE OF ITS
PRODUCTS TO FUNCTION OR OPERATE PROPERLY.
This warranty covers instruments and parts sold to users by authorized distributors, dealers, and
representatives as appointed by RKI Instruments, Inc.
We do not assume indemnification for any accident or damage caused by the operation of this gas
monitor, and our warranty is limited to the replacement of parts or our complete goods.
a) Absorbent cartridges
d) Batteries
b) Pump diaphragms and valves
e) Filter elements
c) Fuses
www.GlobalTestSupply.com
Find Quality Products Online at:
[email protected]
Table of Contents
Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Specifications. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Description. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Pump Unit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Rubber Sample Nipple. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Optional Hose & Probe . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Normal Operation. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Switching Flow Rates . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
Alarms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
Replacing the Battery. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
Replacing the Hydrophobic Disk Filter and Wire Mesh Disk . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
Parts List . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
WARNING:
Understand manual before operating and for proper battery type. Substitution of
components may impair intrinsic safety. To prevent ignition of a hazardous
atmosphere, batteries must only be changed in an area known to be nonhazardous.
Not tested in oxygen enriched atmospheres (above 21%).
CAUTION:
Read and understand the GX-3R and GX-3R Pro Operator’s Manual before using the
RP-3R.
www.GlobalTestSupply.com
Find Quality Products Online at:
[email protected]
This manual describes: the RP-3R pump, RP-3R maintenance, and operation of the RP-3R with the
GX-3R and GX-3R Pro. A parts list at the end of this manual lists replacement parts and accessories
for the RP-3R.
Specifications
Table 1 lists specifications for the RP-3R pump.
Table 1: Specifications
Power Supply
1 AA size alkaline
Operating Time
Higher-flow mode: 10 hours minimum at 25°C
Lower-flow mode: 16 hours minimum at 25°C
Sample Flow Rate
Higher-flow mode: 0.50 LPM (liters per minute)
Lower-flow mode: 0.35 LPM (liters per minute)
Maximum Tube Length
for Optional Sample Hose
100 feet
Operating Temperature &
Humidity
-20°C to +50°C, 0 - 95% Relative Humidity (non-condensing)
Dimensions
29 (D) x 82 (W) x 96 (H) mm, rubber sample nipple not included
Weight
Approximately 120 g (including alkaline battery)
Alarm Indications
Low Flow & Low Battery
Safety/Regulatory (when
used with the GX-3R or
GX-3R Pro)
ATEX (Certificate Number Presafe 17 ATEX 11584)
•
II1G Ex ia IIC T4 Ga (-20°C to +50°C Toshiba LR6)
•
II1G Ex ia IIC T4 Ga (-20°C to +40°C Duracell MN1500)
•
II1G Ex ia IIC T3 Ga (-20°C to +50°C Duracell MN1500)
IECEx (Certificate Number IECEx PRE 17.0070)
•
Ex ia IIC T4 Ga (-20°C to +50°C Toshiba LR6)
•
Ex ia IIC T4 Ga (-20°C to +40°C Duracell MN1500)
•
Ex ia IIC T3 Ga (-20°C to +50°C Duracell MN1500)
www.GlobalTestSupply.com
Find Quality Products Online at:
[email protected]
The RP-3R pump is a sample drawing accessory for the GX-3R and GX-3R Pro. It consists of the
pump unit and the rubber sample nipple.
Pump Unit
The pump unit is housed in a red plastic case. The silver inlet fitting on the right side of the pump unit
allows for connection of a rubber sample nipple or a hose and probe. The inlet fitting’s clear holder
retains 2 hydrophobic membranes and a mesh disk.
The black cap at the bottom right of the pump unit allows for battery compartment access. The pump
unit needs one AA alkaline battery.
The exhaust fitting is located on the bottom left of the pump unit, next to the battery compartment.
A POWER button, a buzzer opening, a POWER LED, and a FLOW LED are located along the middle
front of the pump unit.
The sensor adapter at the top of the pump unit mates to the GX-3R or GX-3R Pro. There are different
sensor adapters for each instrument type. When installed, the text on the front of the GX-3R/GX-3R
Pro and RP-3R should be facing the same way.
Figure 1: Component Location
RESET
PO WER
FLOW
Inlet Fitting
Battery Cap
Exhaust Fitting
Buzzer Opening
Rubber Sample
Nipple
Front
Bottom
RP-3R
Retaining Tab Lever
Sensor Adapter
POWER
www.GlobalTestSupply.com
Find Quality Products Online at:
[email protected]
A black retaining tab lever on the left side of the RP-3R makes installing/removing the GX-3R or
GX-3R Pro easier.
Rubber Sample Nipple
A cone-shaped 4-inch long rubber sample nipple is included with the RP-3R. It can be installed by
pushing the larger end over the inlet fitting. The smaller end can be inserted in a hole or some other
access point to sample the environment.
Optional Hose & Probe
A sample hose and probe (optional accessories) may be connected to the inlet fitting. Sample hose
lengths are available from 10 feet to 100 feet (see Parts List at the end of the manual).
CAUTION:
Sample hose lengths of more than 100 feet are not recommended for the
RP-3R because of flow rate reduction.
Normal Operation
1.
Start up the GX-3R or GX-3R Pro and enter normal operation as described in the appropriate
Operator’s Manual.
2.
Press the black retaining tab lever on the left side of the RP-3R.
WARNING:
Installing the GX-3R or GX-3R Pro onto the RP-3R without pressing the black
retaining tab lever may result in damage to the RP-3R.
3.
Be sure the front labels on the GX-3R or GX-3R Pro are facing the same way as the front labels on
the RP-3R.
4.
When viewing the GX-3R or GX-3R Pro and RP-3R from the front, engage the RP-3R’s right-
hand retaining clip with the GX-3R or GX-3R Pro.
5.
Push the left side of the GX-3R or GX-3R Pro down until the instrument is seated in the RP-3R.
Release the black retaining tab lever and ensure that both retaining clips are fully engaged. When
fully engaged, the GX-3R or GX-3R Pro should not move around at all.
CAUTION:
If the RP-3R is not fully engaged, the unit will not respond to the target gases properly.
www.GlobalTestSupply.com
Find Quality Products Online at:
[email protected]
Figure 2: Connecting the RP-3R to the GX-3R
AIR
POWER
RESET
RESET
POWER
POWER
POWER
FLOW
POWER
MO DE
CO/H 2S LEL O 2
GX-3R
3. Push left side down
until fully seated
RP-3R
2. Engage right-hand
retaining clip.
1. Push black retaining tab
lever toward pump unit
www.GlobalTestSupply.com
Find Quality Products Online at:
[email protected]
Attach the rubber sample nipple or optional hose and probe to the inlet fitting.
7.
Turn on the RP-3R by pressing and holding the POWER button until the RP-3R beeps and the
POWER LED turns green.
WARNING:
While the GX-3R or GX-3R Pro is installed in the RP-3R, the pump must be on for
the GX-3R or GX-3R Pro to function as a gas detection monitor.
8.
Go to the area to be sampled.
Rubber sample nipple: Insert the rubber sample nipple into the vessel or enclosed area. Allow the
RP-3R to sample for 90 seconds before observing the reading on the GX-3R or GX-3R Pro.
Hose and probe: Hold the probe in the area to be sampled. If necessary, the hose and probe may be
lowered into the sample area. Allow the pump unit to sample for 90 seconds before observing the
reading on the GX-3R or GX-3R Pro.
NOTE: The hose and probe do not have their own hydrophobic filters. Take care to not draw liquid
through the hose and probe.
9.
Monitor the GX-3R or GX-3R Pro’s gas readings and take appropriate action if gas alarms occur.
10. When sampling is complete, turn off the RP-3R by pressing and holding the POWER button until
the RP-3R sounds a long beep and the POWER LED turns off.
11. Turn off the GX-3R or GX-3R Pro.
12. Press the black retaining tab lever on the left side of the RP-3R.
Figure 3: Connecting the Optional Hose and Probe
GX-3R
POWER
RESET
FLOW
RP-3R
POWER
POWER
POW ER
M ODE
AIR
www.GlobalTestSupply.com
Find Quality Products Online at:
[email protected]
Removing the GX-3R or GX-3R Pro from the RP-3R without pressing the black
retaining tab lever may result in damage to the RP-3R.
13. Pull up on the left side of the GX-3R or GX-3R Pro when viewing the instrument from the front.
AIR
RESET
RESET
POW ER
POWER
POWER
FLOW
POW ER
POWER
MODE
CO/H2S LEL O2
GX-3R
2. Pull up on left side
RP-3R
1. Push black retaining tab
lever toward pump unit
Figure 4: GX-3R Removal
www.GlobalTestSupply.com
Find Quality Products Online at:
[email protected]
Switching Flow Rates
The RP-3R can operate in a higher-flow mode (standard) or a lower-flow mode.
1.
With the RP-3R turned on, press and hold the POWER button for 1 second.
2.
The FLOW LED will change state (ie. from flashing to steady or vice versa).
NOTE: At temperatures below -10°C, the RP-3R must operate in higher-flow mode. If operating in
lower-flow mode, the RP-3R will automatically switch to higher-flow mode at -10°C. Any
attempt to switch to lower-flow mode will be blocked.
Alarms
The table below outlines the alarms that may occur on the RP-3R.
Flow Mode
LED Indication
Higher-flow mode
FLOW LED steadily yellow
Lower-flow mode
FLOW LED flashes yellow
Alarm Type
Visual
Indications
Audible
Indications
Other
Indications
Recommended Action
Low Battery
POWER
LED
flashing
none
none
Replace the battery.
Dead Battery
POWER
LED
flashing
constant
buzzer
none
Replace the battery.
Low Flow
NOTE: A low flow alarm
may get combined with a
low or dead battery alarm
since a low battery voltage
may cause the pump to
stop working.
FLOW LED
quickly
flashing
intermittent
buzzer
pump stops
•
Check the rubber sample
nipple and filters to make
sure they are clear.
•
If a hose and probe are
used, check them for
kinks or obstructions.
•
Press and release the
POWER button to reset
the alarm.
•
If the low flow condition
persists, the pump may
need servicing.
www.GlobalTestSupply.com
Find Quality Products Online at:
[email protected]
Replacing The Battery
1.
Push down and rotate the black battery cap counterclockwise until it stops.
2.
Pull the battery cap away from the pump unit.
3.
Remove the old battery.
4.
Insert the new AA alkaline battery + end first into the pump unit. The battery orientation is shown
on the back of the pump unit.
5.
Reinstall the black battery cap by pushing it down and turning it clockwise until it stops.
Replacing the Hydrophobic Disk Filter and Wire Mesh Disk
1.
Grasp the clear filter holder (under the inlet fitting) and turn it 1/4 of a turn counterclockwise.
2.
Pull the filter holder away from the pump unit.
3.
The hydrophobic disk filters and wire mesh disk are located in the pump unit and are retained by a
rubber gasket. Gently pull out the rubber gasket with needle nose pliers taking care not to damage
the gasket. The GX-3R version of the RP-3R has a clear spacer below the gasket that may come
out. The GX-3R Pro version of the RP-3R does not have this spacer.
4.
Remove the old hydrophobic filters and/or wire mesh disk from the gasket. A hydrophobic filter is
located on either side of the wire mesh disk.
5.
Install the new hydrophobic filters and/or wire mesh disk making sure a hydrophobic filter is
located on each side of the wire mesh disk.
Figure 5: Changing the Hydrophobic & Wire Mesh Filters
Filter Holder
Clear Spacer (for GX-3R
Versions Only)
POWER
Hole in
Spacer
FLOW
Hydrophobic Disk Filter
Hydrophobic Disk Filter
Rubber Gasket
RP-3R
RESET
POWER
Wire Mesh Disk
POWER
www.GlobalTestSupply.com
Find Quality Products Online at:
[email protected]
For the GX-3R version: If the clear spacer below the rubber gasket came out when you removed
the gasket, reinstall the spacer. Be sure the hole in the spacer lines up with the hole in the housing.
7.
Reinstall the gasket/filters/mesh into the pump unit.
8.
Align the two tabs on the bottom of the filter holder with the two slots in the pump unit. Push the
filter holder into the pump unit and turn it 1/4 turn clockwise until it snaps into place.
Parts List
Table 2 lists replacement parts and accessories for the RP-3R pump.
Table 2: Parts List
Part Number
Description
07-2006
Gasket, filter/screen retaining
07-6051
O-ring, for battery compartment
17-1001RK
Rubber sample nipple
21-1833RK
Filter holder, clear plastic, with inlet fitting
21-1957
Battery cap
21-1958
Clear spacer underneath the filter/screen retaining gasket, for GX-3R versions only
33-0159RK
Hydrophobic disk filter
33-1112RK
Wire mesh disk filter
49-1120RK
AA alkaline battery
80-0009RK-XX
Sample hose (optional). Replace “XX” with length in feet. Available lengths for the
RP-3R are 10, 15, 20, 25, 30, 35, 40, 45, 50, 60, 75, 100 feet.
80-0150RK
Sample probe with dust filter (optional)
www.GlobalTestSupply.com