Philips V60 Plus User Manual Addendum
Text of the manual
V60 Plus User Manual Addendum: High-Flow Therapy Safety and Alarm Features ......................................3
cs
Dodatek návodu k obsluze ventilátoru V60 Plus: Bezpečnost vysokoprůtokové terapie a funkce alarmu.... 5
da
V60 Plus -Tillæg til brugervejledning: Behandling med højt flow - Sikkerheds- og alarmfunktioner............... 7
de
Nachtrag
V60 Plus User Manual Addendum: High-Flow Therapy Safety and Alarm Features
3
V60 Plus User Manual Addendum:
High-Flow Therapy Safety and Alarm Features
The V60 Plus ventilator with High-Flow Therapy (HFT) enables clinicians to easily
switch between high-flow therapy and noninvasive ventilation for patients needing
varying levels of respiratory car
V60 Plus User Manual Addendum: High-Flow Therapy Safety and Alarm Features
1150533 B
The following summarizes the sequence of events if the maximum system pressure
limit is reached:
1.
Flow is reduced until pressure decreases to 5 cmH2O.
2.
A visual and audible alarm indicates the ventilator Cannot Reach Target
Flow.
3.
Every 5 seconds the ventilator chec
Dodatek návodu kobsluze ventilátoru V60 Plus: Bezpečnost vysokoprůtokové terapie afunkce alarmu
5
Dodatek návodu k obsluze ventilátoru
V60 Plus: Bezpečnost vysokoprůtokové
terapie a funkce alarmu
Ventilátor V60 Plus s vysokoprůtokovou terapií (HFT) umožňuje klinickým
pracovníkům snadné přepínání mezi vysokoprůtokovou a nein