CFM/CMM Thermo Anemometer +

InfraRed Thermometer Model AN200 UNITS AREA RECALL cf m/cmm Anemometer Thermometer AN200 NEXT PROBE TEMPERATURE

Extech Instruments AN200 AN200 V2.1 05/07

2 Introduction Congratulations on your purchase of the Extech AN200 CFM/CMM Thermo Anemometer with InfraRed Thermometer. This instrument measures Air Velocity, Air Flow (volume), Air Temperature (with probe) and Surface Temperature (with the InfraRed function). The large, easy-to-read backlit LCD includes primary and secondary displays plus numerous status indicators. The InfraRed feature includes a laser pointer for convenient targeting
AN200 V2.1 05/07 3 Specifications Air Velocity Range Resolution Accuracy m/s (meters per sec) 0.40 - 30.00 m/s 0.01 m/s ± (3%rdg + 0.20 m/s) km/h (kilometers/hour) 1.4 - 108.0 km/h 0.1 km/h ± (3%rdg + 0.8 km/hr) ft/min (feet per minute) 80 – 5900 ft/min 1 ft/min ± (3%rdg + 40 ft/m) mph (miles per hour) 0.9 – 67.0 mph 0.1 mph ± (3%rdg + 0.4 MPH) knots (nautical MPH) 0.8 to 58.0 knots 0.1 knots ± (3%rdg + 0.4 knots) Air Flow Range Resolution Area
AN200 V2.1 05/07 4 Meter Description 1. Power ON/OFF button 2. Probe input jack 3. Laser pointer 4. IR Sensor 5. Rubber holster 6. LCD Display 7. IR thermometer measurement button 8. Airflow buttons (4) 9. Air Temperature function buttons (2) 10. Vane 11. Airflow Average button 12. Backlight button Note: The battery compartment, tilt stand and tripod mount are located on the rear of the instrument Keypad • Press to turn the meter ON or OFF • I
AN200 V2.1 05/07 5 Display Layout • MAX (top of LCD): Max Hold function engaged for the Air Temperature function • HOLD (top of LCD): Data Hold function engaged for the Air Temperature function • PROBE TEMP: Reminder that the top LCD digits represent Air (Vane) Temperature • : Indicates that the laser pointer is on. • IR TEMP: Indicates that the larger LCD digits represent IR temperature measurement • VEL: indicates that meter is in
AN200 V2.1 05/07 6 Operation Connecting the Vane 1. The vane plug is inserted in the meter’s sensor jack at the top of the meter. The plug and jack are keyed so that the plug can only fit in the jack one way. 2. Turn the plug carefully until it lines up with the jack and then firmly push the plug in place. Do not apply undue force or try to twist the plug side-to-side. 3. If the vane is not connected to the meter or if the sensor is defective, the LCD
AN200 V2.1 05/07 7 Air Flow Measurements (CMM / CFM) 1. Turn on the meter using the ON/OFF button 2. Press the UNITS button to select the desired air flow units: CMM (cubic meters per minute) or CFM (cubic feet per minute). NOTE: At power up the meter will display the last unit of measure previously entered. 3. To begin entering the area in m 2 or ft 2, press and hold the AREA button until it beeps twice. The leftmost digit of the bottom display wi
AN200 V2.1 05/07 8 Data Hold (Air Velocity/Air Flow) 1. While taking measurements, press the HOLD button to freeze the air velocity/air flow reading for later viewing. 2. The HOLD indicator will appear in the bottom of the LCD display. 3. Press HOLD again to return to normal operation. MAX/MIN/AVG Record (Air Velocity/Air Flow) This allows the user to record and view the highest (MAX), lowest (MIN) and average (AVG) readings. 1. Press the button MAX/MIN
AN200 V2.1 05/07 9 Temperature Units 1. Remove the meter’s rubber protective jacket and select the desired temperature units using the °F/°C slide switch located in the battery compartment. 2. Replace the protective jacket and connect the sensor to the sensor input jack on top of the meter. Data Hold (Air Temperature) 1. While taking measurements, press the PROBE TEMPERATURE HOLD button to freeze the air temperature reading. 2. The HOLD indicator wil
AN200 V2.1 05/07 10 InfraRed (Non-Contact) Temperature Measurements 1. The IR sensor is located at the top of the meter. 2. Point the sensor toward the surface to be measured. 3. Press and hold the red IR button to begin measuring the surface temperature of a desired target. IR TEMP and will appear on the display. The laser pointer will switch on to help aim the meter. 4. The measured IR surface temperature will appear at the center of the LCD (lar
AN200 V2.1 05/07 11 InfraRed Measurement Considerations • When taking IR measurements the meter automatically compensates for ambient temperature changes. Note that it may take up to 30 minutes to adjust to extremely wide ambient changes. • Low temperature measurements quickly followed by high temperature measurements may require several minutes to stabilize as a result of the IR sensor cooling process. • If the surface of the object under test is
AN200 V2.1 05/07 12 Thermal Emissivity Table for Common Materials Material Emissivity Material Emissivity Asphalt 0.90 to 0.98 Cloth (black) 0.98 Concrete 0.94 Human skin 0.98 Cement 0.96 Leather 0.75 to 0.80 Sand 0.90 Charcoal (powder) 0.96 Earth 0.92 to 0.96 Lacquer 0.80 to 0.95 Water 0.67 Lacquer (matt) 0.97 Ice 0.96 to 0.98 Rubber (black) 0.94 Snow 0.83 Plastic 0.85 to 0.95 Glass 0.85 to 1.00 Timber 0.90 Ceramic 0.90 to 0.94 Paper 0.70 to 0.94 Marble

Extech Instruments Thermometer Owners Instructions:

Device Model Type of Document Pages Size Views Downloads Updated
433201

433201 Instruction manual - 81L67D

7 1451 320 01-10-2026
42535

42535 Operation & user’s manual - HPH4AP

4 108 26 29-06-2026
HD200

Thermometer #JNMDP8

9 368 56 15-09-2026
42520

Manual: Extech Instruments 42520 (F22W7E, Upd.20-06-2026)

4 449 99 20-06-2026

Thermometer: Similar Manuals