Fluke 1630 Technical Data

Text of the manual

Earth Ground Clamp Meter

•  A wide ground loop resistance range, from 0.025 Ω to1500 Ω, to satisfy all requirements •  Large jaw opening of 35 mm (1.38 in) for tests on ground conductors and/or equipotential bus bars •  Measurement of ground leakage current, from 0.2 mA to 1000 mA, without needing to disconnect - ideal for system troubleshooting •  Wide ac current measurement range from 0.2 A to 30 A allows one instrument for multiple applications •  Continuity loop test - quickly assess whether resistance is less than 40 Ω, with audible alarm •  User-defined HI / LO Alarm limits, for rapid measurement evaluation •  Handy “Display-Hold” button for capturing readings in hard-to-reach places •  Time saving memory function automatically records and stores measured values •  Automatic self calibration ensures correct measurement every time The clamp-on ground testing technique used by the Fluke 1630 simplifies ground loop testing and enables non-intrusive leakage current measurement.   The compact and rugged design makes the Fluke 1630 easy to use in small places and harsh environ- ments, while the “Display-Hold” and the continuity test- ing with an audible alarm function, ensures convenience in use. The novel technique means that earth ground loop testing and continuity testing can be completed without breaking the circuit. Earth Ground loop resistance measurements for commercial, industrial, and utility applications

 Fluke Corporation Fluke 1630 Earth Ground Clamp Meter

Stakeless testing system The Fluke 1630 uses the Stakeless testing method, which eliminates the need to disconnect parallel ground rods and find suitable locations for placing auxiliary ground stakes. This saves time and enables users such as industrial and utility electricians, and field service electricians and contractors, to perform earth ground loop tests in locations where it is not possible to use other techniques, such as inside buildings or on power pylons. With the Stakeless testing method, earth ground stakes are no longer necessary. The Fluke 1630 Earth Ground Clamp is placed around the earth ground rod or the connecting cable. A known voltage is induced by one half of the clamp and the current is measured by the other half. The tester automatically determines the ground loop resistance at this grounding connection. Applications •  Earth Ground loop testing in various installations - for example, high voltage pylons, buildings, cell phone substations, RF transmitters, etc. •  Inspection of lightning protection systems •  Leakage current measurement General specifications Operating error Refers to the operating temperature range and is guaranteed for 2 years Storage temp. range -20 °C to 60 °C (-4 °F to 140 °F ) Reference temp. range 23 °C ± 5 °C (73 °F ± 9 °F ) Storage humidity < 75 % RH Operating humidity < 85 % RH Display 9999 digit LC-display with special symbols Protection IP20 according to IEC 529/EN 60529 Safety 300 V, CAT III / pollution degree 2 IEC 61010-1 Weight 640 g (1.4 lb) Conductor size 35 mm (1.38 in) approximately Dimension (length x width x depth) 257 mm x 100 mm x 47 mm (1 in x 3.9 in x 1.9 in) Emission IEC 1000 4-2, IEC 61326-1 class B Immunity IEC 61000-4-2, 8 kV (air) criteria A IEC 61000-4-3, 3 V/m performance criteria A Range selection Auto Overload indicator “OL” on display Measuring time 0.4 sec Measuring frequency 3.333 kHz Battery type 9 V alkaline (type IEC 6 LR 61) Battery life ≥ 8 hours (continuous operation) Power consumption 40 mA Low battery indication Display symbol

 Fluke Corporation Fluke 1630 Earth Ground Clamp Meter

Ground Loop Resistance (Autorange) Operating temperature: 0 °C to 50 °C (14 °F to 122 °F) Range Resolution Operating error1 0.025 - 0.250 Ω 0.002 Ω ± 1.5 % rdg ± 0.05 Ω 0.250 - 9.999 Ω 0.02 Ω ± 1.5 % rdg ± 0.1 Ω 10.00 - 99.99 Ω 0.04 Ω ± 2.0 % rdg ± 0.3 Ω 100.0 - 199.9 Ω 0.4 Ω ± 3.0 % rdg ± 1.0 Ω 200.0 - 400.0 Ω 2 Ω ± 5.0 % rdg ± 5 Ω 400.0 - 600.0 Ω 5 Ω ± 10 % rdg ± 10 Ω 600.0 - 1500 Ω 20 Ω ± 20 % 1Note: Loop resistance with no inductance, external magnetic field < 50 A/m, external electrical field < 1 V/m, conductor centered. Continuity Beep if resistance < approx. 40 Ω Ground/leakage current mA~ (Autorange, 50/60 Hz, true-rms, crest factor < 3.0) Operating temperature: -10 °C to 50 °C (14 °F to 122 °F) Range Resolution Operating error 0.200 - 1.000 mA 0.001 mA ± 2.5 % rdg ± 0.05 mA 1.00 - 10.00 mA 0.01 mA ± 2.5 % rdg ± 0.05 mA 10.0 - 100.0 mA 0.1 mA ± 2.0 % rdg ± 0.3 mA 100 - 1000 mA 1 mA ± 2.0 % rdg ± 3 mA Current A~ (Autorange, 50/60 Hz, true-rms, crest factor < 3.0 ) Operating temperature: -10 °C to 50 °C (14 °F to 122 °F) Range Resolution Operating error 0.20 - 15.00 A 0.01 A ± 2.0 % rdg ± 0.03 A 15.00 - 30.00 A 0.01 A ± 3.0 % rdg ± 0.03 A HI/LO Limit Evaluation Values from 0 Ω to 1510 Ω may be preset for automatic comparison. If the measured value is outside of the limit an audible and visual signal is given. Recording Measured values can be saved or logged with a defined time interval from 1 to 255 s. Recorded values may then be recalled on the display of the instrument. Fluke. Keeping your world up and running.™ The Fluke 1630 Earth Ground Clamp Meter includes: •  Carrying case with strap •  Resistance test loop •  Users manual in English, French, Spanish, German, Italian, Portuguese and Simplified Chinese

Other Accessory Manuals