Vaisala Drycap DMT152 Operation & User’s Manual
Read Online
Read the Vaisala Drycap DMT152 Operation & User’s Manual online using the built-in PDF viewer.
Text of the manual
Phone (int.): +358 9 8949 1
P.O. Box 26
Fax:
+358 9 8949 2227
FIN-00421 Helsinki
Finland
Visit our Internet pages at http://www.vaisala.com/
© Vaisala 2008
No part of this manual may be reproduced in any form or by any means, electronic or
mechanical (including photocopying), nor may its contents be communicated to a third
party without prior written permission of the copyright holder.
The contents are subject to change without prior notice.
Please observe that this manual does not create any legally binding obligations for
Vaisala towards the customer or end user. All legally binding commitments and
agreements are included exclusively in the applicable supply contract or Conditions of
Sale.
VAISALA ________________________________________________________________________ 1
Table of Contents
CHAPTER 1
GENERAL INFORMATION............................................................................ 5
About This Manual ................................................................... 5
Contents of This Manual ....................................................... 5
Version Information............................................................... 6
Related Manuals ................................................................... 6
General Safety Considerations ............................................. 6
Feedback............................................................................... 6
Product Related Safety Precautions ...................................... 7
Recycling .................................................................................. 7
Regulatory Compliances ......................................................... 7
Trademarks............................................................................... 7
Warranty.................................................................................... 8
CHAPTER 2
PRODUCT OVERVIEW.................................................................................. 9
Introduction to Dewpoint Transmitter DMT152..................... 9
Basic Features and Options.................................................. 10
Transmitter Structure............................................................. 10
CHAPTER 3
FUNCTIONAL DESCRIPTION..................................................................... 13
Advanced DRYCAP® Technology......................................... 13
Auto-Calibration .................................................................. 13
Sensor Purge ...................................................................... 14
Sensor Warming ................................................................. 14
CHAPTER 4
INSTALLATION............................................................................................ 15
Selecting the Location........................................................... 15
Installing the Transmitter ..................................................... 16
Wiring ...................................................................................... 18
Connection Cables.............................................................. 19
Power Supply Requirements............................................... 20
Sampling from a Process ...................................................... 21
Pressure......................................................................... 21
Leak tightness and materials ......................................... 21
Flow................................................................................ 22
Sampling Cell Types .............................................................. 22
DMT242SC Sampling Cell .................................................. 22
DMT242SC2 Sampling Cell with Swagelok Connectors .... 22
DSC74 Sampling Cell with Quick Connector and
Leak Screw.......................................................................... 23
User's Guide ______________________________________________________________________
2 ___________________________________________________________________ M210865EN-D
DSC74B Two Pressure Sampling Cell................................24
DSC74C Two-Pressure Sampling Cell with Coil............26
CHAPTER 5
OPERATION.................................................................................................29
Serial Communication............................................................29
Connecting to the Serial Interface.......................................29
Installing the Driver for the USB Cable ..........................29
Terminal Application Settings..............................................30
List of Serial Commands .......................................................32
Device Information and Status..............................................33
Show Device Information ....................................................33
Show Transmitter Status .....................................................34
Show Firmware Version ......................................................34
Configuring Analog Outputs .................................................35
Limit Alarm...........................................................................36
Self-Diagnostic Alarm..........................................................37
Malfunction Alarm................................................................38
Select Analog Output Quantity and Scaling........................38
Configuring Serial Line Operation........................................39
Set Serial Line Settings .......................................................39
Set Serial Line Response Time...........................................39
Set Transmitter Address......................................................40
Format the Measurement Output ........................................40
Select Unit ...........................................................................42
Set Serial Interface Mode....................................................42
Configuring Measurement Parameters ................................43
Enable or Disable Sensor Purge.........................................43
Set Pressure Value for ppm Calculation .............................43
Set Molar Mass for ppmw Calculation.................................45
Serial Line Output Commands..............................................45
Start Measurement Output..................................................45
Stop Measurement Output ..................................................46
Set Output Interval...............................................................46
Output a Reading Once.......................................................47
Troubleshooting and Maintenance Commands ..................47
Display the Currently Active Errors .....................................47
Test Analog Outputs............................................................47
Calibrate Analog Output......................................................48
Extend Analog Output Range..............................................49
Other commands ....................................................................49
Opening the Transmitter in POLL Mode .............................49
Closing the Connection to a Transmitter in POLL Mode.....50
Display Command List ........................................................50
Set Time ..............................................................................50
Reset Transmitter................................................................51
Restore Factory Settings.....................................................51
CHAPTER 6
MAINTENANCE............................................................................................53
Cleaning the Transmitter .......................................................53
VAISALA ________________________________________________________________________ 3
Changing the Filter................................................................. 53
Calibration and Adjustment .................................................. 55
Field Check using DM70..................................................... 55
Troubleshooting..................................................................... 57
Error State........................................................................... 58
Error Codes......................................................................... 59
Unknown Serial Settings..................................................... 60
Technical Support.................................................................. 60
Vaisala Service Centers......................................................... 61
CHAPTER 7
TECHNICAL DATA ...................................................................................... 63
Measured Variables................................................................ 63
Operating Environment ......................................................... 63
Outputs.................................................................................... 64
General .................................................................................... 64
Output Cable Specifications ................................................. 65
Spare Parts and Accessories................................................ 65
Dimensions in mm (inches) .................................................. 66
User's Guide ______________________________________________________________________
4 ___________________________________________________________________ M210865EN-D
List of Figures
Figure 1
Dewpoint Transmitter DMT152.................................................11
Figure 2
DMT152 Installed Directly to Pipeline ......................................16
Figure 3
Removing the Transport Protection Cap..................................16
Figure 4
Installing the Transmitter ..........................................................17
Figure 5
Connecting the Cable...............................................................17
Figure 6
Connectors I and II ...................................................................18
Figure 7
Connector Pinout......................................................................19
Figure 8
Cable with Snap-On Connector................................................20
Figure 9
Cable with Threaded Connector...............................................20
Figure 10
Sampling Cells DMT242SC2 (left) and DMT242SC (right) ......23
Figure 11
DSC74 Sampling Cell with Accessories...................................24
Figure 12
DSC74B....................................................................................25
Figure 13
Removing the Leak Screw........................................................25
Figure 14
Default Assembly of DSC74C ..................................................26
Figure 15
Alternative Assembly of DSC74C (for Tight Spaces)...............27
Figure 16
PuTTY Terminal Application.....................................................31
Figure 17
Filter Structure ..........................................................................54
Figure 18
Supply Voltage Too Low for mA Output Error..........................56
Figure 19
DMT152 Dimensions................................................................66
List of Tables
Table 1
Manual Revisions .......................................................................6
Table 2
Related Manuals.........................................................................6
Table 3
Quantities Measured by the DMT152.........................................9
Table 4
DMT152 Default Serial Interface Setting..................................30
Table 5
DMT152 Serial Commands ......................................................32
Table 6
FORM Command Quantities ....................................................41
Table 7
FORM Command Modifiers......................................................41
Table 8
Pressure Conversion Coefficients ............................................44
Table 9
Error codes ...............................................................................59
Chapter 1 ________________________________________________________ General Information
VAISALA ________________________________________________________________________ 5
CHAPTER 1
GENERAL INFORMATION
This chapter provides general notes for the manual and the Dewpoint
Transmitter DMT152.
About This Manual
This manual provides information for installing, operating, and
maintaining Dewpoint Transmitter DMT152.
Contents of This Manual
This manual consists of the following chapters:
- Chapter 1, General Information, provides general notes for the
manual and the Dewpoint Transmitter DMT152.
- Chapter 2, Product Overview, introduces the features and
advantages of Dewpoint Transmitter DMT152.
- Chapter 3, Functional Description, describes the advanced
functionality of Dewpoint Transmitter DMT152, including the
auto-calibration, sensor purge and sensor warming functions.
- Chapter 4, Installation, provides you with information that helps
you install the Dewpoint Transmitter DMT152.
- Chapter 5, Operation, contains the operating instructions and serial
line commands for the Dewpoint Transmitter DMT152.
- Chapter 6, Maintenance, provides information on the maintenance
of Dewpoint Transmitter DMT152.
- Chapter 7, Technical Data, provides the technical data of Dewpoint
Transmitter DMT152.
User's Guide ______________________________________________________________________
6 ___________________________________________________________________ M210865EN-D
Version Information
Table 1
Manual Revisions
Manual Code
Description
M210865EN-A
March 2008 – First version.
M210865EN-B
July 2008 – Changed default serial line mode.
Added instructions for field checking with
DM70. Added troubleshooting instructions.
M210865EN-C
August 2008 – Updated analog output ranges.
M210865EN-D
October 2008 – Revised instructions for alarm
configuration in chapter 5.
Related Manuals
Table 2
Related Manuals
Manual Code
Manual Name
M210866EN-A
Vaisala DRYCAP® Dewpoint Transmitter
DMT152 Quick Reference Guide
General Safety Considerations
Throughout the manual, important safety considerations are
highlighted as follows:
WARNING
Warning alerts you to a serious hazard. If you do not read and follow
instructions very carefully at this point, there is a risk of injury or
even death.
CAUTION
Caution warns you of a potential hazard. If you do not read and
follow instructions carefully at this point, the product could be
damaged or important data could be lost.
NOTE
Note highlights important information on using the product.
Feedback
Vaisala Customer Documentation Team welcomes your comments
and suggestions on the quality and usefulness of this publication. If
you find errors or have other suggestions for improvement, please
indicate the chapter, section, and page number. You can send
comments to us by e-mail: [email protected]
Chapter 1 ________________________________________________________ General Information
VAISALA ________________________________________________________________________ 7
Product Related Safety Precautions
The Dewpoint Transmitter DMT152 delivered to you has been tested
for safety and approved as shipped from the factory. Note the
following precautions:
CAUTION
Do not modify the unit. Improper modification can damage the
product or lead to malfunction.
Recycling
Recycle all applicable material.
Dispose of batteries and the unit according to statutory regulations.
Do not dispose of with regular household refuse.
Regulatory Compliances
Vaisala DRYCAP® Dewpoint Transmitter DMT152 is in conformity
with the following EU directives, including the latest amendments,
and with national legislation implementing these directives:
- EMC-Directive (2004/108/EC)
- RoHS-Directive (2002/96/EEC)
The conformity is shown by compliance with the following standards:
- EN 61326-1:1997 + A1:1998 + A2:2001 + A3:2003, Electrical
equipment for measurement, control and laboratory use - EMC
requirements - industrial environment.
- CISPR 22: 2006 Class B, EN 61000-4-2: 2001, EN 61000-4-3:
2002, EN 61000-4-4: 2004, EN 61000-4-5: 2001, EN 61000-4-6:
2003
- EN 50419 (2005-01)
Trademarks
DRYCAP® is a registered trademark of Vaisala.
User's Guide ______________________________________________________________________
8 ___________________________________________________________________ M210865EN-D
Warranty
Vaisala hereby represents and warrants all Products
manufactured by Vaisala and sold hereunder to be
free from defects in workmanship or material
during a period of twelve (12) months from the date
of delivery save for products for which a special
warranty is given. If any Product proves however to
be defective in workmanship or material within the
period herein provided Vaisala undertakes to the
exclusion of any other remedy to repair or at its
own option replace the defective Product or part
thereof free of charge and otherwise on the same
conditions as for the original Product or part
without extension to original warranty time.
Defective parts replaced in accordance with this
clause shall be placed at the disposal of Vaisala.
Vaisala also warrants the quality of all repair and
service works performed by its employees to
products sold by it. In case the repair or service
works should appear inadequate or faulty and
should this cause malfunction or nonfunction of the
product to which the service was performed Vaisala
shall at its free option either repair or have repaired
or replace the product in question. The working
hours used by employees of Vaisala for such repair
or replacement shall be free of charge to the client.
This service warranty shall be valid for a period of
six (6) months from the date the service measures
were completed.
This warranty is however subject to following
conditions:
a) A substantiated written claim as to any alleged
defects shall have been received by Vaisala
within thirty (30) days after the defect or fault
became known or occurred, and
b) The allegedly defective Product or part shall,
should Vaisala so require, be sent to the works
of Vaisala or to such other place as Vaisala may
indicate in writing, freight and insurance
prepaid and properly packed and labelled,
unless Vaisala agrees to inspect and repair the
Product or replace it on site.
This warranty does not however apply when the
defect has been caused through
a) normal wear and tear or accident;
b) misuse or other unsuitable or unauthorized use
of the Product or negligence or error in storing,
maintaining or in handling the Product or any
equipment thereof;
c) wrong installation or assembly or failure to
service the Product or otherwise follow
Vaisala's service instructions including any
repairs or installation or assembly or service
made by unauthorized personnel not approved
by Vaisala or replacements with parts not
manufactured or supplied by Vaisala;
d) modifications or changes of the Product as well
as any adding to it without Vaisala's prior
authorization;
e) other factors depending on the Customer or a
third party.
Notwithstanding the aforesaid Vaisala's liability
under this clause shall not apply to any defects
arising out of materials, designs or instructions
provided by the Customer.
This warranty is expressly in lieu of and excludes
all other conditions, warranties and liabilities,
express or implied, whether under law, statute or
otherwise, including without limitation any implied
warranties of merchantability or fitness for a
particular purpose and all other obligations and
liabilities of Vaisala or its representatives with
respect to any defect or deficiency applicable to or
resulting directly or indirectly from the Products
supplied
hereunder,
which
obligations
and
liabilities are hereby expressly cancelled and
waived.
Vaisala's
liability
shall
under
no
circumstances exceed the invoice price of any
Product for which a warranty claim is made, nor
shall Vaisala in any circumstances be liable for lost
profits or other consequential loss whether direct or
indirect or for special damages.
Chapter 2 __________________________________________________________ Product Overview
VAISALA ________________________________________________________________________ 9
CHAPTER 2
PRODUCT OVERVIEW
This chapter introduces the features and advantages of Dewpoint
Transmitter DMT152.
Introduction to Dewpoint Transmitter DMT152
The Vaisala DRYCAP® dewpoint transmitter DMT152 measures
dewpoint temperature accurately in a measurement range from
-80 ºC to -30 ºC (-112 … -22 °F). The maximum output range of the
transmitter is -100 … +20 °C (-148 … +68 °F). If the primary
dewpoint measurement range of the application is above -30 °C
(-22 °F), other Vaisala transmitters may be more suitable.
The DMT152 is designed for a wide range of OEM applications. The
transmitter is easy to install, and the mechanics have been designed
for harsh environments requiring protection against dust, dirt and
splashed water.
The excellent stability and reliability of the transmitter’s performance
is based on advanced DRYCAP® polymer sensor technology. The
DRYCAP® technology has low maintenance needs due to its excellent
long term stability and durability against condensation. For more
information on DRYCAP® technology, refer to chapter Functional
Description on page 13.
Table 3
Quantities Measured by the DMT152
Quantity
Abbreviation
Metric Unit
Non Metric Unit
Dewpoint/frost point
temperature
TDF
ºC
ºF
Parts per million
H2O
ppmv/ ppmw
ppmv/ ppmw
User's Guide ______________________________________________________________________
10 __________________________________________________________________ M210865EN-D
Basic Features and Options
- Calibrated dewpoint measurement range -80 ... -30 °C Td,
output range -100 ... +20 °C Td
- Output in Td* and ppm
- DRYCAP® polymer sensor with auto-calibration and sensor purge
ensures long term stability
- Sensor warming keeps the sensor dry in abnormally high humidity
situations
- Small size to fit in tight installations
- Two analog output channels, selectable from 0 … 20 mA /
4 ... 20 mA and 0 … 5 V / 0 … 10 V
- Configurable alarm modes for analog outputs
- Digital output: RS-485
- Serial line connection also via USB serial interface cable (Vaisala
order code 219690)
- Simultaneous use of analog and digital output possible
- Low maintenance requirements due to excellent long term stability
- Field check suitability with DM70 hand-held meter
- Optional sampling cells with different installation options available
as DMT152 installation accessories
*When the dewpoint is below 0 °C, the transmitter outputs frostpoint
for Td
Transmitter Structure
The structure of the DMT152 is shown in Figure 1 on page 11. The
transmitter body does not have user serviceable parts inside, and is not
designed to be opened. Opening the transmitter will void the warranty.
When the transmitter is delivered, the filter is protected by a yellow
transport protection cap that keeps the sensor dry. The transport
protection cap should be left on the transmitter during storage.
Remove the transport protection cap before installing the transmitter.
Chapter 2 __________________________________________________________ Product Overview
VAISALA _______________________________________________________________________ 11
0710-151
Figure 1
Dewpoint Transmitter DMT152
where
1
=
4-pin M8 connector I: analog output channels and operating
power
2
=
4-pin M8 connector II (shown with protective cap):
digital output (RS-485) and operating power
3
=
30mm nut
4
=
DRYCAP® sensor protected with stainless steel mesh filter
(see Figure 17 on page 54)
5
=
Connection thread: ISO G1/2", UNF 3/4"-16 2A, or
NPT 1/2"
When installing a DMT152 transmitter with ISO or UNF thread, use a
sealing ring to achieve a tight installation. Three copper sealing rings
are provided with the transmitter, and more can be ordered from
Vaisala. Note that the sealing rings for ISO and UNF threads are
different. See section Spare Parts and Accessories on page 65.
NOTE
Do not use a sealing ring with the NPT thread. Instead, seal the thread
using PTFE tape or PTFE paste sealant.
User's Guide ______________________________________________________________________
12 __________________________________________________________________ M210865EN-D
This page intentionally left blank.
Chapter 3 ______________________________________________________ Functional Description
VAISALA _______________________________________________________________________ 13
CHAPTER 3
FUNCTIONAL DESCRIPTION
This chapter describes the advanced functionality of Dewpoint
Transmitter DMT152, including the auto-calibration, sensor purge and
sensor warming functions.
Advanced DRYCAP® Technology
Dewpoint Transmitter DMT152 utilizes an advanced, patented
measurement technology to ensure accurate measurement with
excellent long term stability. This results in very low maintenance
requirements for the transmitter. The lasting performance is achieved
with microprocessor technology and software that automatically
performs self-diagnostic functions in addition to the normal dewpoint
measurement. The self-diagnostic procedures that are conducted are
called auto-calibration, sensor purge and sensor warming.
Auto-Calibration
The auto-calibration feature of the DMT152 transmitter is an
automatic procedure which greatly reduces the possible drift in the dry
end of the dewpoint measurement. The auto-calibration is performed:
- at 45 minute intervals
- when the transmitter is powered up
- when the measured environment changes rapidly
During auto-calibration the sensor is warmed for a short period
(< 1 min) and the sensor capacitance values are evaluated at the
elevated temperature. The possible dry end drift is then corrected to
correspond to the calibrated values. During the auto-calibration the
User's Guide ______________________________________________________________________
14 __________________________________________________________________ M210865EN-D
transmitter outputs the Td value that was measured prior to the
procedure. When the auto-calibration is done, normal measurement
and output is resumed.
Auto-calibration is carried out only if several criteria for the
measurement environment are fulfilled. This ensures the reliability of
the adjustments, and maintains the excellent long term stability that
the patented technology offers.
Sensor Purge
Sensor purge is also an automatic procedure that minimizes the drift at
the wet end readings of the dewpoint measurement. Sensor purge is
performed once a week or when the power is switched on. The sensor
is heated for several minutes which will then evaporate all excess
molecules out of the sensor polymer. This, together with the auto-
calibration, results in a very small drift. Due to the very linear
behavior of the polymer technology, corrections made at the dry end
and the wet end correct the reading in the whole range.
Sensor Warming
The DRYCAP® sensor of the DMT152 transmitter is warmed, which
prevents the sensor and filter from becoming wet in high humidities.
High humidity may be present when the dewpoint temperature rises
close to the gas temperature.
The advantage of sensor warming is the rapid response of dewpoint
measurement. A wet sensor and filter would otherwise result in a
dewpoint equal to ambient temperature (that is RH = 100 %). If the
sensor gets soaked, it will recover fully back to normal operation after
it dries out.
Chapter 4 _______________________________________________________________ Installation
VAISALA _______________________________________________________________________ 15
CHAPTER 4
INSTALLATION
This chapter provides you with information that helps you install the
Dewpoint Transmitter DMT152.
Selecting the Location
It is important that the conditions at the point of installation represent
well the gas to be measured. Temperature changes do not affect the
dewpoint measurement, but pressure changes will have an effect on
the measurement. All leaks in the system must be eliminated to avoid
ambient humidity affecting the measurement.
Direct installation to the measured gas is the recommended
installation method if the temperature of the gas is suitable for
DMT152 and no additional filtering is needed due to very dusty or
oily gas. Oil as such is not harmful for the DRYCAP® sensor, but
response time may be slowed down if there are oil particles in the
system, or if oil is collected to the filter over a longer period of time.
The maximum measurement pressure is 50 bara / 725 psia (absolute
pressure) for direct measurement.
If the gas temperature is higher than the specified maximum operating
temperature of the transmitter, gas sampling and cooling it to ambient
temperature (e.g. 10 °C / 18 °F) is recommended. Note that the
dewpoint temperature must be clearly lower than the ambient
temperature to avoid condensation in the sampling line. Sampling
from the process is easy by using Vaisala sampling cell options; see
section Sampling from a Process on page 21.
The DMT152 is light in weight which means that it can be installed in
a sample pipeline in the sampling cells without the need of any
additional mechanical support.
User's Guide ______________________________________________________________________
16 __________________________________________________________________ M210865EN-D
Max 50 bar
0710-101
Figure 2
DMT152 Installed Directly to Pipeline
Installing the Transmitter
After selecting a suitable measurement location, follow the procedure
below to install the transmitter:
1.
Remove the yellow transport protection cap from the
transmitter. Do not touch the filter with your hands.
0710-128
Figure 3
Removing the Transport Protection Cap
2.
Prepare the sealing of the connection:
-
If the transmitter has a parallel ISO G1/2" thread or
UNF 3/4"-16 thread, place the sealing ring at the base of the
thread. Always use a new sealing ring; do not re-use a previously
installed one.
- If the transmitter has an NPT 1/2" thread, do not use the sealing
ring. Instead, apply PTFE tape or suitable paste sealant to the
thread. Follow the application instructions of the sealant.
Chapter 4 _______________________________________________________________ Installation
VAISALA _______________________________________________________________________ 17
3.
Make sure that the threads on the mounting point are of the
correct type, and fasten the transmitter to the measurement
point. Use your hands to turn the probe until it feels tight. Do
not use force at this point, and check that the sealing ring
(if used) remains centered.
0710-129
Figure 4
Installing the Transmitter
4.
Use a 30mm wrench to tighten the connection to 25 N m, as
shown in Figure 4. If you do not have a 30mm wrench, use a
1 3/16” wrench or an adjustable wrench.
CAUTION
Only tighten the probe from the 30mm nut. Do NOT apply force to
other points in the probe body.
5.
Connect the wires of the connection cable. When using cables
provided with DMT152 refer to the wiring section on page 18.
See the power supply requirements on page 20.
0710-130
Figure 5
Connecting the Cable
User's Guide ______________________________________________________________________
18 __________________________________________________________________ M210865EN-D
6.
Plug in the cable to the transmitter. Be sure to use the correct
connector; refer to section Wiring on page 18.
7.
Cover the unused connector on the transmitter with the rubber
plug that is attached to the transmitter.
8.
Turn on the power supply. The DMT152 performs self
diagnostics at start-up, and the output is frozen for several
minutes. The transmitter will output an approximated reading
after one minute, but accurate measurement will only be
available after the start-up procedure is complete.
Normal operation is typically achieved in 10 … 15 minutes. If
the measurement environment changes during the startup, the
start-up may take up to 15 minutes.
NOTE
The transmitter always performs the self-diagnostic procedure and
sensor purge when starting up. Keep the transmitter continuously
powered to optimize its measurement performance and availability.
Wiring
Pin
I
II
Wire colour
1
VDC supply+
VDC supply+
Brown
2
Signal Ch 1+
RS-485 - / B
White
3
GND
GND
Blue
4
Signal Ch 2+
RS-485 + / A
Black
0707-035
Figure 6
Connectors I and II
Chapter 4 _______________________________________________________________ Installation
VAISALA _______________________________________________________________________ 19
0707-036
Figure 7
Connector Pinout
CAUTION
The power supply lines are internally connected. You can use either
one of them, but do not connect more than one supply voltage in
permanent installations. Temporary simultaneous use with the USB
serial interface cable or DM70 handheld dewpoint meter (which also
provide operation power) is OK.
Connection Cables
The following connection cable options are available for DMT152:
- Unshielded cable with M8 female straight snap-on connector
- Shielded cables with M8 female straight threaded connector
- USB serial interface cable
For the order codes of the cables, refer to section Spare Parts and
Accessories on page 65.
NOTE
Avoid using long cables (over 2m) with voltage output.
User's Guide ______________________________________________________________________
20 __________________________________________________________________ M210865EN-D
0510-036
Figure 8
Cable with Snap-On Connector
0510-037
Figure 9
Cable with Threaded Connector
Power Supply Requirements
The DMT152 transmitters are designed to operate with a supply
voltage of 15 ... 28 VDC (voltage output version) or 21 ... 28 VDC
(current output version). When only the RS-485 output is used,
11 ... 28 VDC is enough in typical measurement conditions.
When measuring in pressures 20 ... 50 bara (290 ... 725 psia) or
temperatures -40 ... 0 °C (-40 ... 32 °F), a supply voltage of
21 ... 28 VDC is required.
The power supply should maintain the voltage for all load conditions.
Current consumption during normal operation is 20 mA. The
consumption increases during the sensor self-diagnostics (auto-
calibration and sensor purge). The maximum current consumption is
220 mA pulsed current.
Chapter 4 _______________________________________________________________ Installation
VAISALA _______________________________________________________________________ 21
Sampling from a Process
Sampling is needed when direct measurement of the air or gas is not
possible or desirable. This may be related to, for example, a high
process temperature, protecting the sensor from water spikes, the need
to easily detach the instrument from a pressurized process without
running the whole process down, or wanting to make the measurement
at a more convenient location. To get a representative sample of the
process gas and avoid error sources caused by the sampling system,
the following aspects should be considered:
Pressure
It is important to keep in mind that changing the pressure of a gas
changes the dewpoint temperature of the gas. If the sensor is at a
different pressure than the process itself, an error of several tens of
degrees dewpoint may be present.
You can use the Vaisala Humidity Calculator to simulate the effect of
pressure change to dewpoint. The Humidity Calculator can be found
at:
http://www.vaisala.com/humiditycalculator
Leak tightness and materials
Due to the typically very low level of humidity, dewpoint
measurements are sensitive to ambient moisture leaks. It is therefore
important to have a leak tight sampling system construction.
The system is likewise sensitive to diffusion of water molecules
through the materials, as well as the absorption of moisture into the
materials. The piping should preferably be made of metal, e.g.
stainless steel with a good surface finish. Hygroscopic materials such
as plastics are undesirable and should be avoided.
Sample tubing should be as short in length as possible. "Dead ends"
should be avoided. Also minimizing the number of connections helps
in avoiding leaks. If possible, prepare for measurements by flushing
the sampling lines with dry gas.
User's Guide ______________________________________________________________________
22 __________________________________________________________________ M210865EN-D
Flow
Vaisala DRYCAP® sensor is not flow dependent as such. For this
reason, the flow of the system is not critical to the measurement,
although a higher flow makes the response time faster. Typically a
flow of 1-2 l/min (0.035-0.7 ft3/min) is enough.
Sampling Cell Types
DMT242SC Sampling Cell
The DMT242SC is a basic sampling cell with only the main sampling
cell body. The inlet and outlet are female threaded sample connectors
(inlet G3/8", outlet G1/4" ISO). Refer to Figure 10 on page 23.
The DMT242SC is suitable for users requiring only a sampling cell to
fit the probe into and doing further assembly (piping into inlet &
outlet, valves, possible flow meter) by themselves.
DMT242SC2 Sampling Cell with
Swagelok Connectors
The DMT242SC2 is similar to DMT242SC, but connections are made
easy. The sampling cell includes welded swagelok connectors at both
the inlet and outlet that fit directly to 1/4" tubing.
To fit 6 mm tubing to the connectors, an adapter such as Swagelok®
Reducer SS-6M0-R-4 (not supplied by Vaisala) can be used.
DMT242SC2 is the suitable choice in e.g. plastics drying systems,
where the measurement is made by tapping off the dryer system and
bringing a small air stream to the sensor. The swagelok connectors of
DMT242SC2 easily connect to a cooling coil or tubing providing the
essential function of cooling the dry air to ambient temperature before
it reaches the sensor.
Chapter 4 _______________________________________________________________ Installation
VAISALA _______________________________________________________________________ 23
0801-069
Figure 10
Sampling Cells DMT242SC2 (left) and DMT242SC
(right)
where
1
=
Male pipe welded connector Swagelok 1/4"
2
=
G1/2"
3
=
G1/4"
4
=
G3/8"
DSC74 Sampling Cell with Quick
Connector and Leak Screw
The DSC74 has been designed especially for compressed air lines.
The sampling cell contains an adjustable leak screw that allows
keeping up the pipeline pressure at the sensor. The leak screw is
opened and closed with a screwdriver. The leak screw should be 1/2
turn open. This can be verified by first closing the leak screw, then
turning 1/2 turn to open it. Then the air coming out will make a barely
audible little hiss and it can be lightly felt when putting a hand in front
of the air stream.
NOTE
Do not open the leak screw more than 1/2 turn to avoid reducing the
pressure in the sampling cell.
The DSC74 comes with a quick connector that fits to industry
standard compressed air line connectors (suitable for type D, Quick08,
NIP08). This allows for easy installation and detachment of the
dewpoint transmitter without having to shut down the process.
Alternative ways to connect are through the two different thread
User's Guide ______________________________________________________________________
24 __________________________________________________________________ M210865EN-D
adapters (G3/8" to G1/2" and G3/8" to G1/4" ISO) that are supplied
with each DSC74 unit.
When using the DSC74, seal the threads of the quick connector or
thread adapter carefully with PTFE thread seal tape. Tighten with a
fork spanner.
0507-105
Figure 11
DSC74 Sampling Cell with Accessories
where
1
=
Thread adapter type G3/8" - G1/2"
2
=
Thread adapter type G3/8" - G1/4"
3
=
Quick connector
4
=
Leak screw
5
=
Sampling cell body (DMT242SC)
DSC74B Two Pressure Sampling
Cell
The DSC74B and DSC74C sampling cells are enhanced versions of
the DSC74. The body is designed in a way that it is possible to
measure at both system pressure and atmospheric pressure.
The DSC74B limits the flow rate with a fixed leak screw. The flow is
optimized for pressures 3 ... 10 barg. The fixed leak screw eliminates
the risk of opening the leak screw fully by accident and this way
emptying a gas vessel. The maximum flow can be increased, if
needed, by removing the leak screw and adjusting the flow manually
with the valve. Harmful gases can be recovered by connecting a
collection system at the outlet (not available from Vaisala).
Chapter 4 _______________________________________________________________ Installation
VAISALA _______________________________________________________________________ 25
In the basic operation of the DSC74B, the gas flows to the sensor from
the front and the outlet is on the side. To have the measurement done
at atmospheric pressure, the inlet and outlet are reversed. Then the
reducing parts supplied (G3/8" -G1/2" or G3/8" - G1/4") on the outlet
side help to protect the sensor from ambient humidity coming in.
DSC74B parts are:
- Sampling cell, thread 3/8"G
- Connection part with a needle valve and an integrated leak screw
- Reducing Nipple (thread adapter), G3/8" - G1/2"
- Reducing Adapter (thread adapter), G3/8" - G1/4"
0510-032
Figure 12
DSC74B
where
1
=
Gas goes in
2
=
Gas comes out
0510-033
Figure 13
Removing the Leak Screw
User's Guide ______________________________________________________________________
26 __________________________________________________________________ M210865EN-D
DSC74C Two-Pressure Sampling Cell with Coil
The DSC74C with an outlet coil is designed for the most critical
measurements at atmospheric pressure. In the DSC74C the coil is
connected to the sampling cell outlet to protect the sensor from
ambient humidity disturbing the measurement. The coil can also be
used as a cooling coil for sampling from high temperature processes.
DSC74C parts are:
- Sampling cell, thread G3/8"
- Connection part with a needle valve and an integrated leak screw
- Reducing Nipple (thread adapter), G3/8" - G1/2"
- Reducing Adapter (thread adapter), G3/8" - G1/4"
- Diffusion coil (for measurements in atmospheric pressure)
0510-034
Figure 14
Default Assembly of DSC74C
where
1
=
Gas goes in. The coil can also be used here.
2
=
Gas comes out
3
=
Coil
4
=
Valve
Chapter 4 _______________________________________________________________ Installation
VAISALA _______________________________________________________________________ 27
0403-113
Figure 15
Alternative Assembly of DSC74C (for Tight Spaces)
where
1
=
Gas comes out
2
=
Coil
3
=
Thread, max. size 7 mm
4
=
Gas goes in
5
=
Valve
The thread size cannot exceed 7 mm. Use the provided adapter to
avoid damage to the probe.
User's Guide ______________________________________________________________________
28 __________________________________________________________________ M210865EN-D
This page intentionally left blank.
Chapter 5 ________________________________________________________________ Operation
VAISALA _______________________________________________________________________ 29
CHAPTER 5
OPERATION
This chapter contains the operating instructions and serial line
commands for the Dewpoint Transmitter DMT152.
Serial Communication
Connecting to the Serial Interface
The DMT152 can be connected to a PC using the RS-485 line on
Port II. You can use, for example, the USB serial interface cable
(Vaisala order code 219690). The cable also provides operation power
to the transmitter from the USB port. It is OK to connect the USB
serial interface cable even if the transmitter is already supplied with
operation power from another cable.
If you have not used the DMT152 USB serial interface cable before,
install the driver that came with the cable. Refer to section Installing
the Driver for the USB Cable below for detailed instructions.
NOTE
It is recommended to only connect one USB serial interface cable and
DMT152 to a PC. Depending on the USB implementation, your PC
may not be able to power more than one DMT152 at a time.
Installing the Driver for the USB Cable
Before taking the USB cable into use, you must install the provided
USB driver on your PC. When installing the driver, you must
acknowledge any security prompts that may appear. The driver is
compatible with Windows 2000, Windows XP, Windows Server
2003, and Windows Vista.
User's Guide ______________________________________________________________________
30 __________________________________________________________________ M210865EN-D
1.
Check that the USB cable is not connected. Disconnect the cable
if you have already connected it.
2.
Insert the media that came with the cable, or download the
driver from www.vaisala.com.
3.
Execute the USB driver installation program (setup.exe), and
accept the installation defaults. The installation of the driver
may take several minutes.
4.
After the driver has been installed, connect the USB cable to a
USB port on your PC. Windows will detect the new device, and
use the driver automatically.
5.
The installation has reserved a COM port for the cable. Verify
the port number, and the status of the cable, using the Vaisala
USB Instrument Finder program that has been installed in the
Windows Start menu.
Windows will recognize each individual cable as a different device,
and reserve a new COM port. Remember to use the correct port in the
settings of your terminal program.
There is no reason to uninstall the driver for normal use. However, if
you wish to remove the driver files and all Vaisala USB cable devices,
you can do so by uninstalling the entry for Vaisala USB Instrument
Driver from the Add or Remove Programs (Programs and
Features in Windows Vista) in the Windows Control Panel.
Terminal Application Settings
The default settings of the DMT152 serial interface are presented in
Table 4. If the settings of your DMT152 have been changed and you
do not know what they are, refer to section Unknown Serial Settings
on page 60.
Table 4
DMT152 Default Serial Interface Setting
Property
Description / Value
Baud rate
19200
Parity
None
Data bits
8
Stop bits
1
Flow control
Xon/Xoff (none)
The steps below describe how to connect to the DMT152 using the
PuTTY terminal application for Windows (available for download at
www.vaisala.com) and a USB serial interface cable:
Chapter 5 ________________________________________________________________ Operation
VAISALA _______________________________________________________________________ 31
1.
Connect the USB serial interface cable between your PC and the
RS-485 port of the DMT152 (port II).
2.
Start the PuTTY application.
3.
Select the Serial settings category, and check that the correct
COM port is selected in the Serial line to connect to field.
Note: You can check which port the USB cable is using with the
Vaisala USB Instrument Finder program that has been
installed in the Windows Start menu.
4.
Check that the other serial settings are correct for your
connection, and change if necessary. Flow control should be set
to None unless you have a reason to change it.
5.
Click the Open button to open the connection window and start
using the serial line.
Note: If PuTTY is unable to open the serial port you selected, it
will show you an error message instead. If this happens, restart
PuTTY and check the settings.
6.
You may need to adjust the Local echo setting in the Terminal
category to see what you are typing on the serial line. To access
the configuration screen while a session is running, click the
right mouse button over the session window, and select Change
Settings... from the pop-up menu.
0807-004
Figure 16
PuTTY Terminal Application
User's Guide ______________________________________________________________________
32 __________________________________________________________________ M210865EN-D
List of Serial Commands
All commands can be issued either in uppercase or lowercase.
The notation refers to pressing the carriage return (Enter) key on
your computer keyboard. Enter a to clear the command buffer
before starting to enter commands.
NOTE
The DMT152 does not output the “>” prompt that is used in the
output examples of this manual. The prompt has been added for
clarity of the examples.
Table 5
DMT152 Serial Commands
Command
Description
?
Show device information
??
Show device information (overrides poll mode)
ACAL [1/2]
Calibrate analog channel
ADDR [0 … 99]
Set the transmitter address for POLL mode
AERR
Set analog output device malfunction error mode
ALARM
Set analog output limit alarm modes
AMODE
Set analog output mode
AOVER [ON/OFF]
Allow analog outputs to exceed their range 10%
ASEL
Configure analog output quantities and scaling
ATEST
Test analog outputs
CLOSE
Close the temporary connection and return the
device to POLL mode
ERRS
Display the currently active errors
FORM
Change the format of the measurement output
FRESTORE
Restore factory settings
HELP
Display the command list
INTV [0 … 255
S/MIN/H]
Set the continuous output interval (for RUN mode
and R command)
MOL
Set the molar mass of the process gas (used in
ppmw calculation)
OPEN [0 … 99]
Open temporary connection to a device that is in
POLL mode
PRES
Set value for pressure compensation
R
Start the continuous outputting
RESET
Reset the transmitter
S
Stop the continuous outputting
SDELAY [0 … 255]
Set response sending delay in milliseconds
SEND [0 … 99]
Output the reading once
SERI
Set the serial interface settings
SMODE
[RUN/POLL/STOP]
Set the serial interface mode
STAT
Display transmitter status
STATUS
Set analog output self-diagnostic alarm
TIME
Set time for transmitter clock
Chapter 5 ________________________________________________________________ Operation
VAISALA _______________________________________________________________________ 33
Command
Description
UNIT
Select metric or non-metric units
VERS
Display transmitter software version
XPRES
Set value for pressure compensation (temporarily,
is cleared at reset)
XPUR [ON/OFF]
Enable or disable sensor purge (temporarily, is
enabled at reset)
Device Information and Status
Show Device Information
The ? command outputs a listing of device information.
?
Example:
>?
DMT152/1.10
Serial number : D1930026
Batch number : D1330070
Module number : ????????
Sensor number : C1420000
Cal. date : 20080512
Cal. info : Vaisala/HEL
Serial mode : STOP
Baud P D S : 19200 N 8 1
Output interval: 1 S
Address : 0
Pressure : 1.0132 bar
Filter : 1.000
Ch1 output : 0 ... 5 V
Ch2 output : U ALARM
Ch1 Tdf lo : -80.00 'C
Ch1 Tdf hi : 20.00 'C
Ch2 Tdf lo : 0.00 'C
Ch2 Tdf hi : 0.00 'C
If the transmitter is in poll mode, but a connection has not been
opened using the OPEN command, issue the ?? command. For a
description of the serial interface modes, see section Set Serial
Interface Mode on page 42.
??
User's Guide ______________________________________________________________________
34 __________________________________________________________________ M210865EN-D
Show Transmitter Status
The STAT command displays the status of the transmitter in four
characters.
STAT
The status codes are as follows:
0000 Normal measurement
1000 Sensor purge
1010 Settling period after sensor purge
1100 Auto-calibration after sensor purge
1110 Settling period after purge + auto-calibration
0100 Auto-calibration (e.g. after interval has elapsed)
0110 Settling period after auto-calibration
0001 Autocalibration failed (will retry automatically)
Example:
>stat
0000
>
NOTE
It is normal for auto-calibration to fail if the measurement
environment is changing rapidly; the transmitter will retry the auto-
calibration automatically. Auto-calibration should succeed when the
environment has stabilized.
Show Firmware Version
VERS
Example:
>vers
DMT152 / 1.10
>
Chapter 5 ________________________________________________________________ Operation
VAISALA _______________________________________________________________________ 35
Configuring Analog Outputs
The DMT152 has two analog output channels. These channels can
function either as normal measurement channels, or produce an
ON/OFF output if the measurement exceeds a given alarm level (limit
alarm) or the transmitter is performing an auto-calibration or purge
(self-diagnostic alarm).
-
In the normal operation mode, a channel outputs voltage or current
that corresponds to the reading of the selected quantity.
- In the ON/OFF output mode, a channel outputs a specified voltage
or current level. This mode is used by the limit alarm and the self-
diagnostic alarm.
Additionally, DMT152 has a malfunction alarm that is always enabled
in normal operation mode. The malfunction alarm can also be enabled
for a channel in ON/OFF output mode. See section Malfunction
Alarm on page 38.
When the transmitter is ordered from Vaisala, the modes are
configured as specified in the order form. You can change the
operation settings after delivery using the AMODE command:
AMODE [c1 c2]
where
c1 =
Channel 1 analog output mode. Available options are:
1 = 0 ... 20 mA
2 = 4 ... 20 mA
4 = 0 ... 5 V
5 = 0 ... 10 V
7 = I alarm (ON/OFF output using current)
8 = U alarm (ON/OFF output using voltage)
Modes 1 ...5 are normal measurement modes, where
malfunction alarm is always enabled, and limit alarm and
self-diagnostic alarm are disabled.
c2 =
Channel 2 analog output mode. Available options are the
same as for channel 1.
Example:
>amode 2 1
Ch1 output : 4 ... 20 mA
Ch2 output : 0 ... 20 mA
>
User's Guide ______________________________________________________________________
36 __________________________________________________________________ M210865EN-D
Limit Alarm
The limit alarm is activated when the measured quantity goes below
the low limit, or above the high limit. The alarm can be specified
during ordering for a specified parameter, or configured on the serial
line using the ALARM command.
When ordering a limit alarm, the alarm levels are set as follows:
- For current output, an active alarm is indicated by the current
being at the maximum of the scale (20 mA).
- For voltage output, an active alarm is indicated by the voltage
being at the maximum of the selected scale. For example, if the
voltage output scale is 0 … 5 V, an output of 5 V indicates the
alarm is active.
Using the ALARM command you can also set a hysteresis value to
prevent the repeated triggering of the alarm when the measurement
fluctuates around a limit value.
ALARM
The following example shows that a high limit alarm has been enabled
for channel 1. If the measured value rises above the -40.00 °C Td limit,
analog channel 1 is set to 5 V (Limit Hi = -40.00, Level Hi = 5.00).
The malfunction alarm (aerr) is disabled for both channels.
>alarm ?
Channel 1:
Low Limit : OFF
High Limit : ON
Aerr : OFF
Quantity : Tdf
Limit Lo : 0.00
Limit Hi : -40.00
Hysteresis : 0.00
Level Lo : 0.00
Level Hi : 5.00
Channel 2:
Low Limit : OFF
High Limit : OFF
Aerr : OFF
Quantity : ppm
Limit Lo : 0.00
Limit Hi : 0.00
Hysteresis : 0.00
Level Lo : 0.00
Level Hi : 0.00
>
Chapter 5 ________________________________________________________________ Operation
VAISALA _______________________________________________________________________ 37
NOTE
Avoid enabling more than one alarm type for a channel.
Self-Diagnostic Alarm
When the transmitter is performing a sensor purge or auto-calibration,
the measurement output is frozen (unchanging) until the procedure is
complete. The self-diagnostic alarm indicates this state by setting the
analog output at a defined level, in the same way as the limit alarm
(see page 36).
Use the STATUS command to enable/disable the alarm, and to define
the voltage/current levels (see example below). The channel is set to
the value of Level Hi when purge or auto-calibration is active, and to
Level Lo during normal operation.
STATUS
The following example shows the self-diagnostic alarm enabled for
channel 1:
>status
Channel 1:
Status : ON ?
Level Lo : 0.00 ?
Level Hi : 5.00 ?
Channel 2:
Status : OFF ?
Level Lo : 0.00 ?
Level Hi : 0.00 ?
>
Level Lo and Level Hi are shared between the STATUS and
ALARM commands. For example, if you change the levels using the
alarm command, it will also change when viewing the status
command.
User's Guide ______________________________________________________________________
38 __________________________________________________________________ M210865EN-D
Malfunction Alarm
If the transmitter is malfunctioning, the channel output is set to a pre-
defined level. This overrides the normal measurement output of the
channel. You can specify the level using the AERR command:
AERR
Example:
>aerr
Ch1 error out : 20.000 mA ?
Ch2 error out : 0.000 V ?
>
The current value is shown after the colon “:”. To set a new value,
enter if after the question mark. The error output value must be within
the valid range for the selected output mode.
NOTE
The error output value is displayed only when there are minor
electrical faults such as a humidity sensor damage. When there is a
severe device malfunction, the error output value is not necessarily
shown.
The malfunction alarm is always enabled for channels that are in the
normal measurement modes (see section Configuring Analog Outputs
on page 35), but you can enable or disable it using the ALARM
command for channels that are in the ON/OFF output mode.
Select Analog Output Quantity and
Scaling
Use the ASEL command to select the quantities output by the analog
channels. Refer to Table 6 on page 41 for the selectable quantities.
After entering the command, you will be prompted to enter the scaling
of the quantities for both channels. Press enter to accept the suggested
scaling values.
ASEL [xxx yyy]
where
xxx =
Analog channel 1 output quantity
zzz =
Analog channel 2 output quantity
Chapter 5 ________________________________________________________________ Operation
VAISALA _______________________________________________________________________ 39
Example:
>asel tdf ppm
Ch1 Tdf lo : -80.00 'C ?
Ch1 Tdf hi : 20.00 'C ?
Ch2 ppm lo : 0.00 ppm ?
Ch2 ppm hi : 1000000.00 ppm ?
>
Configuring Serial Line Operation
Set Serial Line Settings
Use the SERI command to set the serial line settings. The new
settings will be taken into use when the transmitter is reset or powered
up. Note that the transmitter will use its default serial line settings if it
is in error state; see section Error State on page 58.
Also note that changing the serial line settings will slightly alter how
the transmitter works with the MI70 indicator. If you are not using the
default settings of 19200 N 8 1, you will have to power off the MI70
before it can find the transmitter (i.e., you cannot connect it on-the-
fly).
SERI [b p d s]
where
B
=
baud rate (300, 600, 1200, 2400, 4800, 9600, or 19200)
P
=
parity (n = none, e = even, o = odd)
D
=
data bits (7 or 8)
S
=
stop bits (1 or 2)
Example:
>seri 9600 e 7 1
Baud P D S : 9600 E 7 1
>
Set Serial Line Response Time
With the SDELAY command you can set delay (response time) of the
serial line, or view the currently set delay value. Increasing the value
by one increases the delay by 4 milliseconds. For example, value of
5 equals ~20 ms answer delay.
User's Guide ______________________________________________________________________
40 __________________________________________________________________ M210865EN-D
SDELAY [x]
Where
x
= Serial line delay (0 … 255)
Example:
>sdelay 1
Serial delay : 1
>
Set Transmitter Address
To operate in the POLL mode, the transmitter must have an address. If
multiple transmitters share the same serial line, each transmitter must
have a different address.
For a description of the serial interface modes, see section Set Serial
Interface Mode on page 42.
ADDR [nn]
where
nn
=
address (0 ... 99)
Example:
>addr 1
Address : 1
>
Format the Measurement Output
Use the serial line command FORM to change the format or select the
quantities for the output commands SEND and R.
You can return to the default format using the ‘FORM /’ command.
FORM [x]
where
x
=
Formatter string
Formatter string consists of quantities and modifiers. The maximum
length of the form string is 74 characters.
When entering the command, use the abbreviations of the quantities.
The quantities are presented in Table 6, and modifiers in Table 7.
Chapter 5 ________________________________________________________________ Operation
VAISALA _______________________________________________________________________ 41
Table 6
FORM Command Quantities
Quantity
Description
TDF
Dewpoint/frost point temperature
PPM
Water content in parts per million
PPB
Water content in parts per billion
PPMW
Water content in parts per million by weight
Note: For accurate ppm, ppb or ppmw reading, you
must compensate for the pressure and molar mass of
the process gas.
Table 7
FORM Command Modifiers
Modifier
Description
x.y
Length modifier (number of digits and decimal places)
#t
Tabulator
#r
Carriage-return
#n
Line feed
“"
String constant
#xxx
ASCII code value (decimal) of a special character;
for example, #027 for ESC
ADDR
Transmitter address with two characters [00...99]
ERR
Currently active transmitter errors in a five bit field:
bit0bit1bit2bit3bit4
bit0 T MEAS error
bit1 F meas error
bit2 Voltage too low error
bit3 Voltage too low for mA output error
bit4 Ambient temperature error
STAT
Transmitter status in 4 characters. The status codes are
the same as for the STAT command; see section Show
Transmitter Status on page 34.
SN
Transmitter serial number
TIME
Current time; see the description of the TIME command
on page 50.
Ux
Shows the name of the measurement unit using “x”
number of characters. For example, U3 shows the
name of the measurement unit with three characters.
Examples:
>form 4.2 TDF #r #n
OK
>
>form 3.1 “H2O= “ ppm “ “ U3 #r #n
OK
>
User's Guide ______________________________________________________________________
42 __________________________________________________________________ M210865EN-D
Select Unit
The DMT152 Transmitter outputs dewpoint temperature Td in either
Celsius or Fahrenheit degrees depending on the unit selection. The
unit selection also affects the unit (bara or psia) for the pressure
compensation.
UNIT [x]
where
x
=
Selects the unit type to output:
m = metric units, e.g. celcius
n = nonmetric units, e.g. fahrenheit
Set Serial Interface Mode
SMODE [xxx]
where
xxx
=
STOP, RUN, or POLL
In STOP mode: outputting only when command is issued, any
command can be used.
In RUN mode: outputting automatically, only command S can be
used. In POLL mode: outputting only when command is issued, any
command can be used after the line has been opened using the OPEN
command.
The polling mode can be used when more than one transmitter is
connected to one serial bus. When set to POLL state, the transmitters
communicate one at a time when the specific transmitter address is
called on the serial line. See descriptions of the commands ADDR on
page 40 and OPEN on page 49.
Examples:
>smode stop
Output mode : STOP
>
>smode run
Output mode : RUN
>
>smode poll
Output mode : POLL
>
Chapter 5 ________________________________________________________________ Operation
VAISALA _______________________________________________________________________ 43
Configuring Measurement Parameters
Enable or Disable Sensor Purge
You can temporarily disable the sensor purge function using the
XPUR command. Note that disabling the sensor purge will affect the
measurement performance of the transmitter.
XPUR [ON/OFF]
Example:
>xpur off
Purge : OFF
>
After giving the XPUR OFF command, the transmitter will not
perform sensor purge until one of the following events takes place:
-
Transmitter is reset or powered up
-
Sensor purge is enabled using the XPUR ON command
Set Pressure Value for ppm
Calculation
The DRYCAP® 180U sensor in the DMT152 is not pressure
dependent. When measuring dewpoint, pressure compensation is not
needed. However, in the calculation of ppm volume concentration, the
pressure of the system must be known. If you did not specify pressure
compensation when ordering the product, use the commands below to
enable the pressure compensation and set the pressure value.
PRES [pp.pppp]
XPRES [pp.pppp]
where
pp.pppp
=
Absolute pressure in the measuring point. The unit is
bara if metric units have been selected, and psia if
non-metric units are used.
User's Guide ______________________________________________________________________
44 __________________________________________________________________ M210865EN-D
Command XPRES should be used if the value is changed frequently.
The value set using the XPRES command overrides the PRES value,
but it is not retained at reset. When the XPRES is set to 0, the last
value set with PRES is used instead.
Example:
>pres 1.01325
Pressure : 1.01325 bar
>
The pressure conversion coefficients from other pressure units to bars
is given in the table below.
Table 8
Pressure Conversion Coefficients
From
To bara
PaN/m²
0.00001
mmHg torr
0.001333224
inHg
0.03386388
mmH2O
0.00009806650
inH2O
0.002490889
atm
1.01325
at
0.980665
psia
0.06894757
Example:
29.9213 inHg = 29.9213 × 0.03386388 = 1.01325 bara
NOTE
Conversions from mmHg and inHg are defined at 0 °C and from
mmH2O and inH2O at 4 °C.
NOTE
You can also set the pressure compensation value when you connect
the DMT152 to the MI70 indicator. When using the MI70, the
pressure unit is bara.
Chapter 5 ________________________________________________________________ Operation
VAISALA _______________________________________________________________________ 45
Set Molar Mass for ppmw
Calculation
Use the MOL command to view and set the molar mass of the gas in
the measurement environment. The molar mass parameter is used in
calculating H2O ppmw. Some typical molar masses are:
- Air
28.96 g/mol
- N2
28.0134 g/mol
- O2
32.00 g/mol
- SF6
146.06 g/mol
MOL [n]
where
n
= Molar mass (in g/mol) of the gas in the measurement
environment
Examples:
>mol 146.06
Mol : 146.0600
>
>mol
Mol : 146.0600 ?
>
NOTE
You can also set the molar mass when you connect the DMT152 to
the MI70 indicator.
Serial Line Output Commands
Start Measurement Output
Use the R command to start the continuous outputting of measurement
values as an ASCII text string to the serial line. The format of the
measurement message is set with the FORM command.
R
User's Guide ______________________________________________________________________
46 __________________________________________________________________ M210865EN-D
Example (measurement message in default format):
>r
Tdf= -2.5 'C H2O= 4919 ppm
Tdf= -2.5 'C H2O= 4918 ppm
Tdf= -2.5 'C H2O= 4917 ppm
…
Outputting the results continues in intervals issued with the command
INTV. You can stop the output by pressing the esc key, or with the
command S. Since the interface is half-duplex, you must enter the
commands when the transmitter is not outputting.
Stop Measurement Output
You can stop the measurement output by pressing ESC, or with the S
command:
S
Set Output Interval
INTV [n xxx]
where
n
=
time interval = 0 ... 255
xxx
=
time unit = "S", "MIN", or "H"
Example:
>intv 1 min
Output interval: 1 MIN
>
The shortest output interval (with n = 0) is approximately one second
due to the internal measurement cycle.
Chapter 5 ________________________________________________________________ Operation
VAISALA _______________________________________________________________________ 47
Output a Reading Once
SEND [aa]
where
aa
= Transmitter address (0 ... 99). Must be specified if the
transmitter is in POLL mode, and a line has not been opened
using the OPEN command.
Example:
>send
Tdf= -2.0 'C H2O= 5152 ppm
>
Troubleshooting and Maintenance Commands
Display the Currently Active Errors
Use the ERRS command to display the currently active transmitter
error codes. Possible errors and their causes are listed in Table 9 on
page 59.
ERRS
Example:
>errs
Voltage too low error
>
Test Analog Outputs
The operation of the analog outputs can be tested with the ATEST
command by forcing the output to a given value which can then be
measured with calibrated multimeter. The command uses the currently
selected analog output types for the channels (voltage or current) so
you do not have to specify the unit.
Before testing, you can verify the currently set analog output modes
using the AMODE command. After testing the output, give the
ATEST command again to exit the test mode.
User's Guide ______________________________________________________________________
48 __________________________________________________________________ M210865EN-D
ATEST [xxx yyy]
Where
xxx
=
Channel 1 output value (V or mA)
yyy
=
Channel 2 output value (V or mA)
The output shows the test values of the analog channels, and
diagnostic information that may be useful to Vaisala Service if there is
a problem with the analog outputs.
Example:
>atest 1 15
1.00 2660
15.00 19425
>
>atest
0.00 79
0.00 20110
>
Calibrate Analog Output
Use the ACAL command to force the analog outputs to the following
values:
- Current output: 2 mA and 18 mA
- Voltage output: 10 % and 90 % of the range
After entering the ACAL command, use a calibrated current/voltage
meter to measure the output, and enter the values. Note that the
calibration is only valid for the currently selected analog output
modes.
ACAL [channel]
where
Channel
=
Analog output channel: 1 or 2
Example:
>acal 1
Ch1:
I1 ( mA ) ? 0.166
I2 ( mA ) ? 18.1
3.61058569E+02 1.40927863E+03
Chapter 5 ________________________________________________________________ Operation
VAISALA _______________________________________________________________________ 49
Extend Analog Output Range
Use the AOVER command to allow the analog output channels to
exceed their specified range by 10%. The scaling of the quantity
remains as before; the extra range is used for additional measurement
range in the wet end.
AOVER [ON/OFF]
The following example illustrates how the analog output is affected.
Channel 1 outputs Td with voltage output 0 … 5 V (-80 °C … -30
°C). After giving the AOVER ON command, the range is 0 … 5.5 V
(-80 °C … -25 °C). Note that the -30 °C Td point is still at 5 V.
Example:
>aover on
: ON
>
Other commands
Opening the Transmitter in POLL
Mode
Use the OPEN command to connect to a transmitter that is in POLL
mode.
OPEN [aa]
where
aa
=
address (0 ... 99)
Resulting (only in POLL-state):
>open 1
DMT152 1 line opened for operator commands
>
User's Guide ______________________________________________________________________
50 __________________________________________________________________ M210865EN-D
Closing the Connection to a
Transmitter in POLL Mode
The CLOSE command closes the connection to a transmitter.
CLOSE
Example:
>close
line closed
>
Display Command List
HELP
Set Time
Use the TIME command to view and set the transmitter clock. The
clock does not have a battery backup – it is reset to 00:00:00 when the
transmitter is powered up or reset.
TIME [hh mm ss]
where
hh
=
Hours (24-hour clock)
mm
=
Minutes
ss
=
Seconds
Example:
>time 12 30 00
12:30:00
>
>time
12:30:05
>
Chapter 5 ________________________________________________________________ Operation
VAISALA _______________________________________________________________________ 51
Reset Transmitter
When the RESET command is given, the transmitter will restart as if
it had just been powered on. All transmitter settings are retained. The
serial line mode of the transmitter will be set to the mode that was set
using the SMODE command.
As the transmitter starts up, it performs the startup sensor purge and
auto-calibration before resuming measurement.
RESET
Restore Factory Settings
Use the FRESTORE command to restore the factory settings to the
transmitter. All user settings will be lost.
FRESTORE
User's Guide ______________________________________________________________________
52 __________________________________________________________________ M210865EN-D
This page intentionally left blank.
Chapter 6 ______________________________________________________________ Maintenance
VAISALA _______________________________________________________________________ 53
CHAPTER 6
MAINTENANCE
This chapter provides information on the maintenance of Dewpoint
Transmitter DMT152.
Cleaning the Transmitter
The body of the transmitter can be cleaned by wiping with a
moistened lint-free cloth. Do not immerse the transmitter in liquid,
and do not use cleaning agents or solvents.
Do not attempt to clean the filter or the sensor element. If the filter is
contaminated, replace it with a new one.
Changing the Filter
If the filter is contaminated, it should be replaced. New filters can be
ordered from Vaisala, with the order code 220957. Be careful when
changing the filter, since it is easy to break the sensor when the filter
is removed.
CAUTION
If the transmitter is installed in a pressurized chamber, always make
sure that the pressure of the chamber is equalized with the ambient
pressure prior to removing the transmitter. When the transmitter is
removed for maintenance, cap the hole with a capped nut. This way,
the process can be running although the transmitter is not in place.
Plugs are available from Vaisala for ISO threaded connections
(Vaisala order code 218773) and NPT threaded connections
(Vaisala order code 222507).
User's Guide ______________________________________________________________________
54 __________________________________________________________________ M210865EN-D
When replacing the filter, note the following:
-
The pores of the steel mesh may become blocked if touched. Do
not touch the filtering surface with your hands or any tools.
Always handle the filter from the top of the filter body.
-
Touching the filter with your fingers will also deposit oil and dirt
on the filter, which may affect the measurement. Always handle
the filter with gloved hands.
- When installing the transmitter after a filter change, always use a
new sealing ring (ISO and UNF threads only).
0711-195
Figure 17
Filter Structure
where
1
=
Vaisala DRYCAP® sensor
2
=
Spring washer
3
=
Stainless steel mesh filter (do not touch)
4
=
Top of the filter
Replace the filter as follows:
1.
Grip the old filter from the top of the filter body using an 11 mm
socket wrench. Turn the filter counterclockwise until it is loose.
2.
Pull the filter straight out carefully; do not damage the sensor,
and do not lose the spring washer.
3.
Take the new filter, and insert it to the filter threads. The spring
washer should be under the filter.
4.
Tighten the new filter by turning it clockwise with the 11 mm
socket wrench. Tighten to 5 N m.
Chapter 6 ______________________________________________________________ Maintenance
VAISALA _______________________________________________________________________ 55
Calibration and Adjustment
The DMT152 is fully calibrated as shipped from factory. The
recommended calibration interval is 2 years. If the transmitter is used
in the wet end of the measurement range, calibration should be done
every year. Calibration should also be done if there is a reason to
believe that device is not within the accuracy specifications.
You can perform a simple field checking operation by comparing the
readings of the DMT152 and the DM70 hand-held dewpoint meter
with a calibrated reference probe. For more information, see section
Field Check using DM70 on page 55.
If there is need for calibration or adjustment, contact a Vaisala Service
Center or your local Vaisala representative. See contact information
on page 61.
Field Check using DM70
To check of the accuracy of the DMT152 in the field, you need the
following equipment:
- A fully charged MI70 indicator (the measurement display that is
included in the DM70 package)
- A calibrated reference probe that is suitable for the intended
measurement environment and dewpoint range. You can use a
DMP74 probe from the DM70 for dewpoint down to -60 °C
(-76 °F), or a second DMT152 for full measurement range.
- Connection cable between DMT152 and the MI70 indicator
(Vaisala order code 219980).
- A reference environment with a stable dewpoint and temperature in
a suitable range for both probes. The normal measurement
environment of the transmitter can also be used, if it meets these
criteria.
NOTE
In the procedure below, the readings of the probes are checked
individually. You can also perform the field check so that you
connect the reference probe simultaneously with the DMT152 you are
checking. In this case, you need two of the connection cables if you
are using a DMT152 also as the reference probe. Also, you must
power one of the transmitters by other means than the MI70.
User's Guide ______________________________________________________________________
56 __________________________________________________________________ M210865EN-D
Perform the check as follows:
1.
Place the DMT152 in the reference environment (unless it is
already). You can perform the check even if the DMT152 is
currently being powered and used, as long as there is no cable
connected to Port II.
2.
Connect the MI70 connection cable between the MI70 indicator
and Port II of the DMT152.
3.
Turn on the MI70 indicator. If the analog channels of your
DMT152 are configured to use the current output, and if the
DMT152 is powered by the MI70, you will see an error message
that indicates the supply voltage is too low for running the actual
analog output. This is expected behavior; press OK to
acknowledge the error and continue the field check.
0805-014
Figure 18
Supply Voltage Too Low for mA Output Error
4.
Check and adjust the environment settings if prompted. You can
change the pressure compensation value (only needed for ppm
calculation), and the molar mass of the measurement gas. The
pressure must be entered in bara; to enter the value in psia, use
the serial line. For more information on these settings, see the
following sections:
- Set Pressure Value for ppm Calculation on page 43
- Set Molar Mass for ppmw Calculation on page 44
5.
After a brief delay, the DMT152 should provide an approximate
measurement reading of the selected quantities. Remember that
it takes up to 15 minutes for the transmitter to start the active
measurement.If the desired quantities are not shown on the
display, select them from the Display – Quantities and units
menu.
6.
The Auto power off function should be disabled on the MI70
indicator to prevent the indicator from powering off in the
middle of the check. While waiting for the active measurement
to start, check the setting in the Settings - User Interface menu,
and change if necessary.
Chapter 6 ______________________________________________________________ Maintenance
VAISALA _______________________________________________________________________ 57
7.
Write down the Td reading after the measurement has stabilized.
8.
Turn off the MI70 indicator and disconnect the DMT152.
9.
Insert the reference probe to the measurement environment, and
connect it to the MI70 indicator.
10.
Turn on the MI70 indicator. Check and adjust the environment
settings if prompted.
11.
Wait for the reference probe to provide a stable reading, and
compare it with the reading from the DMT152.
The difference between the readings will give you an indication
of the approximate drift of the DMT152 and its need for
calibration. As a general rule, the reading from the DMT152
should be within ±4 °C Td of the reading from the reference
probe.
NOTE
Some notes on using the MI70 indicator with the DMT152:
- If you disconnect the DMT152 while the MI70 indicator is on,
the quantities measured will remain on the screen, but no
measurement will be shown.
- The messages from the transmitter (such as the supply voltage
error) will appear in English even if the MI70 has been set to
another language.
Troubleshooting
If you have a problem with the DMT152, first read the following
sections concerning the behavior and error indications of the
transmitter:
- Configuring Analog Outputs on page 35
- Error Codes on page 59
- Error State on page 58
Possible causes of errors include insufficient operation voltage,
physical damage to the transmitter, and incorrect connections.
Connection problems can be caused by misconfiguration of the analog
outputs or the serial line. Also remember that the start up of the
transmitter may take up to 15 minutes.
Some problems can be solved by simply resetting the transmitter. You
can reset the transmitter by disconnecting the power or issuing the
RESET command on the serial line.
User's Guide ______________________________________________________________________
58 __________________________________________________________________ M210865EN-D
If you cannot locate the source of the error and return the transmitter
to the normal state, please contact a Vaisala Service Center or your
local Vaisala representative. See sections Error Codes on page 59 and
Vaisala Service Centers on page 61
Error State
If the voltage supplied to the DMT152 is too low for reliable
operation, or there is some serious problem that prevents
measurement, the DMT152 will enter an error state. The error state is
indicated by the following:
- Analog current output at 0 mA or 4 mA (according to selected
range)
- Analog voltage output at 0V
- Serial line outputs asterisk characters (*** **) instead of
measurement data
The analog output levels for the error state are configurable using the
AERR command; see section Malfunction Alarm on page 38. Note
that the asterisk characters on the serial line may also appear during
the start-up procedure.
When the transmitter is in the error state, it will use the default serial
interface settings, regardless of the currently set serial interface mode.
If you have set a different mode with the SERI command, you will
have to set your terminal program to the default settings of
19200 N 8 1 (also listed in Table 4 on page 30) before you can
connect to the transmitter in error state.
Chapter 6 ______________________________________________________________ Maintenance
VAISALA _______________________________________________________________________ 59
Error Codes
The transmitter has error codes that can be displayed on the serial line
using the ERRS command. See section Display the Currently Active
Errors on page 47.
NOTE
All other errors except “Voltage too low for mA output error” will
cause the DMT152 to enter the error state; see page 58.
Table 9
Error codes
Error Code
Cause
T meas error
Internal error, can be caused by
damage to the sensor.
F meas error
Internal error, can be caused by
damage to the sensor.
Voltage too low error
Supplied operating voltage is too low
for reliable operation.
Voltage too low for mA output error
Supplied operating voltage is too low
for reliable operation of the analog
current output.
Note: It is normal to receive this
error when using the MI70 indicator
or the USB serial interface cable, as
they do not provide a sufficient
operating voltage to operate the
current output.
Ambient temperature error
Ambient temperature out of
operating range (probably too hot).
Flash check sum error
Internal error
Param check sum error
Internal error
INFOA check sum error
Internal error
SCOEFS check sum error
Internal error
CURRENT check sum error
Internal error
User's Guide ______________________________________________________________________
60 __________________________________________________________________ M210865EN-D
Unknown Serial Settings
If you are not sure what the serial settings of your transmitter are,
follow the steps below to view and change them.
1.
Make sure there are no other transmitters on the serial line.
2.
Provide the transmitter with a 9 V supply voltage. This voltage
is enough to start it in the error state.
3.
Connect to the serial line using the error state serial line settings
19200 8 N 1.
4.
Give the ? command to view the transmitter settings, including
the serial settings and address. For an output example, see
section Show Device Information on page 33.
You may have to give the command several times if there are
characters in the buffer.
5.
If necessary, set new serial line settings using the SERI
command. For instructions on using the command, see page 39.
Technical Support
For technical questions, contact the Vaisala technical support:
E-mail
[email protected]
Fax
+358 9 8949 2790
Chapter 6 ______________________________________________________________ Maintenance
VAISALA _______________________________________________________________________ 61
Vaisala Service Centers
Send the device to Vaisala Instruments Service Centre for calibration
and adjustment, see contact information below. Remember to attach
the yellow transport protection cap to the transmitter to protect it
(and keep it dry) during transit.
If the product needs repair, please contact Vaisala Instruments Service
Center in your region before shipping the instrument. Follow the
instructions given to you to speed up the process and to avoid extra
costs.
User's Guide ______________________________________________________________________
62 __________________________________________________________________ M210865EN-D
This page intentionally left blank.
Chapter 7 ____________________________________________________________ Technical Data
VAISALA _______________________________________________________________________ 63
CHAPTER 7
TECHNICAL DATA
This chapter provides the technical data of Dewpoint Transmitter
DMT152.
Measured Variables
Dewpoint temperature
Measurement range
-80 … -30 °C (-112 … -22 °F) Td
Accuracy
± 2 °C (3.6 °F) Td
Non-calibrated range
-100 … +20 °C (-148 … +68 °F) Td
Response time 63% [90%] at +20 °C (+68 °F)
gas temperature and 1 bar pressure
-80 -> -30 °C (-112 … -22 °F) Td
10 s [20 s]
-30 -> -80 °C (-22 … -112 °F) Td
2 min [30 min]
Analog output scalings
option 1
-80 ... +20 °C (-112 ... +68 °F) Td
option 2
-100 ... 0 °C (-148 ... +32 °F) Td
option 3
free scaling
ppm volume concentration
Measurement range
0 … 500 ppm
Accuracy at +20 °C (+68 °F),
1013 mbar
± (0.2 ppm + 20 % of reading)
Operating Environment
Temperature
-40 … +60 °C (-40 … +140 °F)
Relative humidity
0 … 100 %RH (up to +20 °C / +68 °F)
Pressure
0 … 50 bar (725 psia)
Measured gases
non-corrosive gases
Sample flow rate
no effect on measurement accuracy
User's Guide ______________________________________________________________________
64 __________________________________________________________________ M210865EN-D
Outputs
Two analog outputs (scalable)
0 … 20 mA
4 … 20 mA
0 … 5 V
0 … 10 V
Accuracy
voltage output
± 0.01 V
current output
± 0.01 mA
Typical temperature dependence
voltage output
± 0.001 % of span / °C
current output
± 0.005 % of span / °C
Digital output
RS485 (2-wire half-duplex)
ON/OFF output available instead of the 2nd analog output
purge or auto-calibration inactive
0 V or 4 mA
purge or auto-calibration active
5 V, 10 V or 20 mA
OR exceeded Td or ppm limit
user selectable
General
Sensor
Vaisala DRYCAP® 180U
Thin-film capacitive polymer sensor
Startup time
10 .. 15 minutes
Recommended calibration interval
2 years
Operating voltage with
RS485 output
11* … 28 VDC
voltage output
15* … 28 VDC
current output
21 … 28 VDC
* For extended temperature down to -40 °C (-40 °F) or pressure
up to 50 bar (725 psia) the supply voltage must be 21 ... 28 VDC.
Supply current
normal measurement
20 mA + load current
during self-diagnostics
max. 220 mA pulsed
Load for
voltage output
min. 10 kOhm
current output
max. 500 Ohm
Housing material
AISI316L and PPS
Stainless steel mesh filter
AISI303, filter grade 18 μm
Mechanical connections
option 1
ISO G1/2”
option 2
NPT 1/2"
option 3
UNF 3/4"-16
Housing classification
IP65 (NEMA 4)
Storage temperature range
-40 … +80 °C (-40 … +176 °F)
Chapter 7 ____________________________________________________________ Technical Data
VAISALA _______________________________________________________________________ 65
Weight
With ISO thread
190 g (6.70 oz)
With NPT thread
210 g (7.41 oz)
With UNF thread
186 g (6.56 oz)
Complies with EMC standard EN61326-1, Electrical equipment
for measurement, control and laboratory use - EMC requirements;
Industrial environment.
Output Cable Specifications
Connector
Straight, moulded M8
female (IEC 60947-5-2)
Color
Black
Number of contacts
4
Conductors
0.25 mm² (× 4) /
0.00039 inch² (× 4)
Insulation
PVC
Cable diameter
5.0 mm / 0.2 inch
Protection class
IP67 / NEMA 4
Wire color code
EN50044
Spare Parts and Accessories
Description
Order Code
Connection cable for DM70
219980
USB serial interface cable
219690
Sealing ring set ISO (copper, 3 pcs)
221524SP
Sealing ring set UNF (copper, 3 pcs)
221563SP
Stainless steel mesh filter (18 µm)
220957SP
110 … 240 VAC external power supply
POWER-1
Plug Kit (ISO 1/2")
218773
Plug Kit (NPT 1/2")
222507
Analog/RS-485 output cables
2 m (6,5 ft) cable, M8 snap-on connector
211598
0.32 m (1 ft) shielded cable, M8 threaded connector
HMP50Z032
3 m (9.8 ft) shielded cable, M8 threaded connector
HMP50Z300
5 m (16.4 ft) shielded cable, M8 threaded connector
HMP50Z500
10 m (32.8 ft) shielded cable, M8 threaded connector
HMP50Z1000
Sampling cells (available only for ISO G1/2”)
Sampling cell
DMT242SC
Sampling cell with 1/4" male Swagelok connectors
DMT242SC2
Sampling cell with quick connector and leak screw
DSC74
Two-pressure sampling cell
DSC74B
Separate cooling/venting coil for sampling cells
DMCOIL
User's Guide ______________________________________________________________________
66 __________________________________________________________________ M210865EN-D
Dimensions in mm (inches)
0710-133
Figure 19
DMT152 Dimensions