Teledyne 2103 Installation And Operation Manual
Read Online
Read the Teledyne 2103 Installation And Operation Manual online using the built-in PDF viewer.
Text of the manual
Installation and Operation Guide
Part #69-2003-637 of Assembly #60-2004-637
Copyright © 2011. All rights reserved, Teledyne Isco
Revision A, April 6, 2012
This instruction manual is designed to help you gain a thorough understanding of the
operation of the equipment. Teledyne Isco recommends that you read this manual
completely before placing the equipment in service.
Although Teledyne Isco designs reliability into all equipment, there is always the possi-
bility of a malfunction. This manual may help in diagnosing and repairing the malfunc-
tion.
If the problem persists, call or e-mail the Teledyne Isco Technical Service Department
for assistance. Simple difficulties can often be diagnosed over the phone.
If it is necessary to return the equipment to the factory for service, please follow the
shipping instructions provided by the Customer Service Department, including the
use of the Return Authorization Number specified. Be sure to include a note
describing the malfunction. This will aid in the prompt repair and return of the
equipment.
Teledyne Isco welcomes suggestions that would improve the information presented in
this manual or enhance the operation of the equipment itself.
Teledyne Isco is continually improving its products and reserves the right to
change product specifications, replacement parts, schematics, and instruc-
tions without notice.
Contact Information
Customer Service
Phone:
(800) 228-4373
(USA, Canada, Mexico)
(402) 464-0231
(Outside North America)
Fax:
(402) 465-3022
Email:
[email protected]
Technical Support
Phone:
(800) 775-2965
(Analytical)
(866) 298-6174
(Samplers and Flow Meters)
Email:
[email protected]
Return equipment to:
4700 Superior Street, Lincoln, NE 68504-1398
Other Correspondence
Mail to:
P.O. Box 82531, Lincoln, NE 68501-2531
Email:
[email protected]
Web site:
www.isco.com
Revised March 17, 2009
2103 Modem Module
Table of Contents
Section 1 Introduction
1.1 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-1
1.2 Product Description. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-1
1.3 Identifying Module Components . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-2
1.4 Safety Symbols and Hazard Alerts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-6
1.5 Technical Service. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-6
Section 2 Preparation and Installation
2.1 Unpacking Instructions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-1
2.1.1 2103Ci Label Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-1
2.2 Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-2
2.2.1 Latches - Locking and Unlocking . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-2
2.2.2 Communication Connectors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-3
2.2.3 Stacking Modules . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-3
2.3 Telephone Line Connection (2103 only). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-4
2.3.1 Modem Cable Connection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-5
2.3.2 Connection Without the Modem Cable . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-5
2.4 Antenna (2103Ci Only) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-6
2.5 Connecting to Flowlink . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-7
2.6 Modem Site Connection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-12
2.6.1 2103 Modem Connection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-12
2.6.2 2103Ci Modem Connection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-13
2.7 Setting Up Text Messaging (2103Ci) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-16
2.8 Power Conservation for Cellular Modems (2103Ci) . . . . . . . . . . . . . . . . . . . . . . . 2-17
2.8.1 Equation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-19
2.9 Pushed Data Capability . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-20
Section 3 Modbus Protocol
3.1 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-1
3.2 Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-1
3.2.1 Establishing Communication . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-2
3.2.2 Module Addressing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-2
3.3 Configurations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-3
3.4 Glossary of Terms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-4
3.5 Common Acronyms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-5
3.6 Register Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-5
Section 4 Maintenance
4.1 Maintenance Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-1
4.1.1 Cleaning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-1
4.2 Maintenance Kit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-1
4.3 Desiccant . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-2
4.3.1 Replacing the Desiccant . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-2
4.3.2 Reactivating the Desiccant . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-2
4.4 Hydrophobic Filter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-3
2103 Modem Module
Table of Contents
iv
4.5 O-Rings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-3
4.6 How to Obtain Service . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-3
Appendix A Replacement Parts
A.1 Replacement Parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-1
A.1.1 2103 Modem Module . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-2
A.1.2 2103Ci Cellular Module . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-5
A.1.3 Antenna . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-8
Appendix B Accessories
B.1 How to Order. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . B-1
B.2 General Accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . B-1
Appendix C Material Safety Data Sheets
C.1 Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . C-1
List of Figures
1-1 2103 Modem Components - Top View . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-2
1-2 2103 Modem Components - Bottom View . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-3
1-3 2103 Communication Connector Pins . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-5
1-4 2103 Modem Cable Connector . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-5
2-1 2103Ci label . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-1
2-2 Connecting the Modem Cable . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-5
2-3 2103 Modem Cable Connector . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-6
2-4 Antenna connector (left) and antenna connected to 2103Ci (right) . . . . . . . . . . . . 2-6
2-5 Flowlink connect screen (version 5.10.616)
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-8
2-6 Site resolution screen . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-9
2-7 Site Information screen . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-10
2-8 Devices screen - connected to site . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-11
2-9 Devices screen - 2103 module . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-12
2-10 Modem screen - configuring the cellular modem . . . . . . . . . . . . . . . . . . . . . . . . 2-13
2-11 Devices screen - configuring the modem for Serial-Over-IP communication . . 2-14
2-12 Devices screen - configuring the modem for CSD communication . . . . . . . . . . 2-15
2-13 Setting up text messaging and alarm conditions . . . . . . . . . . . . . . . . . . . . . . . . 2-16
2-14 Wireless power control window . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-18
2-15 Equation Builder for power control . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-19
2-16 Condition Builder . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-19
2-17 Data tab . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-20
3-1 Configuration example . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-3
List of Tables
1-1 2103 Modem Module Components - Top View . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-2
1-2 2103 Modem Module Components - Bottom View . . . . . . . . . . . . . . . . . . . . . . . . . 1-3
1-3 2103 Modem Module Technical Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-4
1-4 Specifications – 2191 Battery Module . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-4
1-5 2103 Communication Connector Pins (2103 only) . . . . . . . . . . . . . . . . . . . . . . . . . 1-5
3-1 Modbus ASCII Address 1 Register Definitions . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-6
3-2 Modbus ASCII Address 2+(N-1) Register Definitions . . . . . . . . . . . . . . . . . . . . . . 3-6
3-3 2100 Measurement Parameters by Model Number* . . . . . . . . . . . . . . . . . . . . . . . 3-8
2103 Modem Module
Section 1 Introduction
1.1 Introduction
This instruction manual is designed to help you gain a thorough
understanding of the operation of the 2103 and 2103Ci Modem
Modules. Teledyne Isco recommends that you read this manual
completely before placing the equipment into service.
Information in this manual pertains to both the 2103 phone line
modem and the 2103Ci cellular modem. Where there are specific
differences between the modules, they are identified.
1.2 Product Description
The 2103 Modem is a data communications unit designed to
transmit data from Isco’s 2100 Series flow modules, which
measure parameters of open channel flow streams. It works in
conjunction with Isco’s Flowlink software.
The 2103 Modem can be located anywhere within a stack of up to
three other 2100 Series networked modules, using the same
locking mechanism that connects the 2100 Series modules to
each other. The 2103 is compatible with Isco’s 2150 Area Velocity
flow module, 2110 Ultrasonic flow module, 2101 Field Wizard,
and 2102 wireless module. It is powered by Isco’s 2191 battery
module.
All enclosures are rated NEMA 4X, 6P(IP68). The permanently
sealed enclosures are designed to meet the environmental
demands of many sewer flow monitoring applications. All connec-
tions between modules, sensors, and communication cables lock
in place. The locking mechanisms strongly secure the compo-
nents and ensure a watertight seal.
2103 Modem Module
Section 1
Introduction
1-2
1.3 Identifying Module
Components
Figures 1-1 and 1-2 identify the key components of the 2103
Modem Module.
Figure 1-1 2103 Modem Components - Top View
1
3
4
5
6
2
Table 1-1 2103 Modem Module Components - TopView
Item No.
Name
Description
1
Communication Connector
This port is used to connect the 2103 Modem to other modules
in a stack.
2
Connector Cap
Insert into the communication connector when it is not in use to
protect the connector from moisture damage. When the connec-
tor is in use, store the connector cap in the cap holder.
3
Cap Holder
Stores the connector cap when the communication connector is
in use.
4
Modem Cable Connector
(2103 only)
Used to connect the modem cable to the 2103 Modem Module.
(On the 2103Ci module, this side connector has a slightly differ-
ent appearance, and is used to connect the magnetic mount
antenna.)
5
Modem Cable Connector Cap
Insert into the modem cable connector when the connector is
not in use to protect the connector from damage.
6
Latch Release
Push in to unlock the module from a stack.
2103 Modem Module
Section 1
Introduction
1-3
Figure 1-2 2103 Modem Components - Bottom View
2
1
3
4
5
Table 1-2 2103 Modem Module Components - BottomView
Item No.
Name
Description
1
Communication Connector This connects the 2103 Modem to other 2100 Series modules in the
stack and is used to transfer data.
2
Connector Plug
Insert into the communication connector when not in use to protect the
connector from moisture damage. When the connector is in use, store
the connector plug in the holder.
3
Plug Holder
Stores the connector cap when the communication connector is in use.
4
Desiccant Cartridge and
Hydrophobic Filter
Prevents moisture from entering the unit.
5
Latch
Push in to lock the module in a stack.
2103 Modem Module
Section 1
Introduction
1-4
Table 1-3 2103 Modem ModuleTechnical Specifications
Dimensions
Length = 10.5 inches (26.7 cm)
Width = 7.5 inches (19 cm)
Height = 2.9 inches (7.4 cm)
Weight
2 lbs. (.9 Kg)
Material
High-impact molded polystyrene
Enclosure
NEMA 4X, 6P, IP68
Power
6.6 to 16.6 VDC, 141 mA typical at 12 VDC, 0.41 mA standby
Operating Temperature
-4° to 140°F (-20° to 60°C)
Storage Temperature
-40° to 140°F (-40° to 60°C)
Typical Battery Life
291 days*
Modulation Standards Supported
(2103 only)
Bell 103, Bell 212, V.21, V.22, V.22 bis, V.23, V.32, V.32 bis, V.34
Communication Speeds Supported
300, 1200, 2400, 4800, 7200, 9600, 12000, 14400, 16800, 19200,
21600, 24000, 26400, 28800, 31200, 33600 bps
Optional Cellular Communication
Serial Over IP, or CSD (2103CI)
Error Correction Standards Supported
V.42 LAPM, MNP-2, MNP-4, MNP-10
Data Compression Standards Supported
V.42 bis, MNP-5
* Actual battery life will vary depending upon configuration. The figure given assumes interrogation with Flowlink once a week, with a site configuration
of a 2103, 2150, and 2191 (using Energizer 529 batteries) and a connection speed of 33600 baud. The 2150 was configured to record level, velocity,
flow rate every 15 minutes, total flow, and battery voltage every 24 hours.
Table 1-4 Specifications – 2191 Battery Module
Size (HWD)
6.0 9.6 7.6 in.
15.2 24.4 19.3 cm
Weight (without batteries)
3.2 lbs.
1.4 kg
Materials
ABS plastic, stainless steel
Enclosure (self-certified)
NEMA 4X, 6P
IP68
Batteries
6V alkaline lantern or lead-acid lantern, quantity 2
Capacity
Alkaline Lantern Batteries 25 Ahrs
Lead-acid Lantern
Batteries
5 Ahrs
2103 Modem Module
Section 1
Introduction
1-5
Figure 1-3 2103 Communication Connector Pins
Figure 1-4 2103 Modem Cable Connector
A
B
C
D
E
F
G
Table 1-5 2103 Communication Connector Pins (2103 only)
Pin
Name
Description
A
LONA
Neuron differential transceiver Data A
B
LONB
Neuron differential transceiver Data B
C
VIN+
Positive power supply voltage input (+12 VDC nominal)
D
VIN–
Negative power supply voltage input (0 VDC nominal)
E
RCVUP
PC data receiver inverted input
F
XMTUP
PC data transmit inverted output
G
Key
Aligns connector pins
2103 Modem Module
Section 1
Introduction
1-6
1.4 Safety Symbols and
Hazard Alerts
This icon identifies a general hazard and is accompanied
with details about the hazard. The instruction manual
identifies the hazardous condition and any steps nec-
essary to correct the condition. The manual presents this infor-
mation in one of two ways:
CAUTION
Cautions identify a potential hazard, which if not avoided, may
result in minor or moderate injury. This category can also warn
you of unsafe practices, or conditions that may cause property
damage.
WARNING
Warnings indicate potentially hazardous conditions. If you
do not avoid these risks, they could cause you death or
serious injury.
1.5 Technical Service
Although Teledyne Isco designs reliability into all of its
equipment, there is always the possibility of a malfunction
occurring. You can use this manual to help in diagnosing and
repairing any malfunctions. If the malfunction persists, call or
write the Teledyne Isco Technical Service Department for assis-
tance:
Teledyne Isco
Technical Service Department
P.O. Box 82531
Lincoln, NE 68501
866-298-6174 or 402-464-0231
FAX: 402-465-3001
e-mail: [email protected]
Simple difficulties can often be diagnosed over the phone. If it is
necessary to return the equipment to the factory for service,
please follow the shipping instructions provided by the Technical
Service Department, including the use of the Return Authori-
zation Number specified. Be sure to include a note describing the
malfunction. This will aid in the prompt repair and return of the
equipment.
2103 Modem Module
Section 2 Preparation and Installation
2.1 Unpacking
Instructions
When the system arrives, inspect the contents for any damage. If
there is damage, contact the delivery company and Teledyne Isco
(or its agent) immediately.
WARNING
If there is any evidence that any items may have been
damaged in shipping, do not attempt to install the unit.
Please contact Teledyne Isco (or its agent) for advice.
When you unpack the system, check the items against the
packing list. If any parts are missing, contact the delivery
company and Teledyne Isco’s Customer Service Department.
When you report missing part(s), please indicate them by part
number. In addition to the main packing list, there may be other
packing lists for various sub-components.
It is recommended that you retain the shipping cartons as they
can be used to ship the unit in the event that it is necessary to
transport the system.
Please complete the registration card and return it to Teledyne
Isco.
2.1.1 2103Ci Label
Information
If you have a 2103Ci, you should record the IP address and phone
number printed on the label, found on the back of the unit
(Figure 2-1). You will need this information when you configure
the unit.
Figure 2-1 2103Ci label
Teledyne Isco
Customer Service Dept.
P.O. Box 82531
Lincoln, NE 68501 USA
Phone: (800) 228-4373
Outside USA & Canada call:
(402) 464-0231
FAX:
(402) 465-3022
E-mail: [email protected]
2103 Modem Module
Section 2
Preparation and Installation
2-2
2.2 Installation
The 2100 Series components are often installed in confined
spaces. Some examples of confined spaces include manholes,
pipelines, digesters, and storage tanks. These spaces may become
hazardous environments that can prove fatal for those unpre-
pared. These spaces are governed by OSHA 1910.146 and require
a permit before entering.
WARNING
Avoid hazardous practices! If you use these instruments in
any way not specified in this manual, the protection
provided by the instruments may be impaired; this will
increase your risk of injury.
WARNING
The installation and use of this product may subject you
to hazardous working conditions that can cause you
serious or fatal injuries. Take any necessary precautions
before entering a worksite. Install and operate this product
in accordance with all applicable safety and health
regulations, and local ordinances.
Follow the instructions below to install your 2103 Modem. Most
of these instructions are similar for the 2103Ci. Where there are
differences, they are specifically identified.
2.2.1 Latches - Locking and
Unlocking
Latches must be operated to stack and unstack the modules in a
Series 2100 stack. The mechanisms are the same for the 2103
Modem and other 2100 Series modules. Detailed instructions
with photos can be found in your 2150 instruction manual.
Take a moment to familiarize yourself with operating the latches.
You must unlock the latch to place the module on top of another
module in a stack. The latch is unlocked by pushing in the latch
release on the right side of the module. To lock the latch, push in
the latch on the left side of the module.
CAUTION
The latch can be damaged by applying too much force. Never
press on both sides at the same time. Do not force the latch if it
is obstructed. While some degree of pressure must be applied
to slide the latch, the ends of the latches should never bend
more than 1/8”.
Note
Latches will “click” when they are fully locked and unlocked.
2103 Modem Module
Section 2
Preparation and Installation
2-3
2.2.2 Communication
Connectors
Connecting the 2103 Modem module involves uncapping and
capping communication connectors. When a communication con-
nector is not in use, the connector should always be capped. The
cap will seal the connector to prevent corrosion, and will improve
communications.
When a communication connector is in use, store the cap on the
holder next to the connector. The communication connector will
be sealed by its mating connector.
Detailed instructions and photos can be found in your 2150 or
2110 instruction manual.
CAUTION
Caps PUSH ON and PULL OFF. Do not rotate the caps to
remove them from the connectors.
Note
For modules to correctly stack and lock together, protective
caps between the modules must be stored on the holders.
2.2.3 Stacking Modules
The 2103 Modem Module can be located anywhere within a stack
of up to three 2100 Series networked modules. It will draw its
power from the battery module located in the stack.
To connect the 2103 with a 2100 Series module, refer to the fol-
lowing instructions.
1. On the top of the 2100 Series module, remove the cap and
stow it on the holder. This exposes the communication con-
nector on the module.
2. Inspect the module’s communication connector. It should
be clean and dry. Damaged O-rings must be replaced.
3. Unlock the 2103’s latch by pressing in on the latch release
(right side).
4. Underneath the 2103, remove the cap from the lower com-
munication connector and stow it in the holder.
5. Lock the latch. Locking the latch correctly seats and aligns
the lower cap in its holder.
6. Position the 2103 over the 2100 Series module. Align the
connectors and lower the 2103 onto the other module.
7. Unlock the 2103’s latch by pressing in on the latch release
(right side).
8. Firmly press the modules together and lock the 2103’s
latch (left side).
The communications indicator will blink during the start-up
routine to indicate the 2103 is operating.
2103 Modem Module
Section 2
Preparation and Installation
2-4
2.3 Telephone Line
Connection (2103
only)
The FCC (Federal Communications Commission) governs com-
munications over telephone lines. Your local telephone company
will provide you with the line between the 2103 Modem and your
computer. Contact them for connection information.
The 2103 Modem Module is in compliance with FCC part 68
rules. However:
Note
In tests conducted by the Communication Certification Labora-
tory, the 2103 Modem did not pass conducted RF testing on
telephone line at 5 MHz, 3V RMS. For details, see international
standard EN 61000-4-6. In the event of this failure, the modem
may be subject to disconnection.
Accordingly, the FCC requires the following information
be published:
Note
The 2103 Modem is designed to be used on standard device
telephone lines. It connects to the telephone by means of a
standard jack called the USOC RJ-11C. Connection to tele-
phone-company-provided coin service (central office imple-
mented systems) is prohibited, and connection to party lines
service is subject to state tariffs.
Changes in Attestation Procedure for Plugs and Jacks
Teledyne Isco attests that the network interface plugs or jacks
used on this equipment comply with and will continue to com-
ply with the mechanical requirements specified in Part 58,
sub-part F, specifically the dimensions, tolerances and metallic
plating requirements. The compliance of these connectors will
be assured by purchase specifications and incoming inspec-
tion. Documentation of such specifications and/or inspections
will be provided to the FCC within 30 days of their request for
the same.
Telephone Company Procedures
The goal of the telephone company is to provide you with the
best service it can. In order to do this, it may occasionally be
necessary for them to make changes in their equipment, oper-
ations or procedures. If these changes might affect your ser-
vice or the operation of your equipment, the telephone
company will give you notice, in writing, to allow you to make
any changes necessary to maintain uninterrupted service.
In certain circumstances, it may be necessary for the tele-
phone company to request information from you concerning
the equipment which you have connected to your telephone
line. Upon request of the telephone company, provide the FCC
registration number and the ringer equivalence number (REN);
both of these items are listed on the equipment label. The sum
of all the RENs on your telephone lines should be less than
five in order to assure proper service from the telephone com-
pany. In some cases, a sum of five may not be usable on a
given line. Consult your telephone provider.
2103 Modem Module
Section 2
Preparation and Installation
2-5
If Problems Arise: If any of your telephone equipment is not
operating properly, you should immediately remove it from your
telephone line, as it may cause harm to the telephone network.
If the telephone company notes a problem, they may temporar-
ily discontinue service. When practical, they will notify you in
advance of this disconnection. If advance notice is not feasible,
you will be notified as soon as possible. When you are notified,
you will be given the opportunity to correct the problem and will
be informed of your right to file a complaint with the FCC. Con-
tact your local telephone service provider if you have any ques-
tions about your phone line.
In the event repairs are needed on the 2103 Modem, they
should be performed by Teledyne Isco or its authorized repre-
sentative. For information, contact the Teledyne Isco Customer
Service Department at (866) 298-6174 or (402) 464-0231.
2.3.1 Modem Cable
Connection
After you have installed the 2103 on the stack, you need to attach
the modem cable so the module can be connected to a phone line.
Remove the connector cap from the 5-pin circular modem cable
connector on the right hand side of the 2103. Attach the modem
cable to the connector (Figure 2-1), and then connect the other
end of the modem cable to a standard telephone jack (USOC
RJ-11C).
Figure 2-2 Connecting the Modem Cable
2.3.2 ConnectionWithout
the Modem Cable
If desired, the 2103 can be connected to a standard telephone line
by attaching the telephone line cable to a connector that mates
with the 2103’s 5-pin circular modem cable connector (Figure
2-2). Be sure that you have a water tight seal on your wire con-
nections.
2103 Modem Module
Section 2
Preparation and Installation
2-6
For your reference, the modem cable uses a 5-pin amphenol
socket, MS3106A-5S.
Figure 2-3 2103 Modem Cable Connector
2.4 Antenna(2103CiOnly)
The 2103Ci uses a magnetic mount antenna and Serial Over
Internet Protocol (SOIP) communication to connect with other
units. (The module is also capable of CSD communication, but
this service is being phased out. See Section 2.6.2 for complete
information.) Phone service with a static IP address, rather than
a land line, provides efficient communication.
Note
The 2103Ci only works in a CDMA cellular service area with
1xRTT. See your local CDMA service provider for availability.
As you face the unit, the 2103Ci antenna connector socket is
located on the right hand side of the module (Figure 2-4). Remove
the connector cap from the socket and attach it to the cap on the
end of the antenna cable. To connect the antenna, press down on
the silver metal latch on the side of the module and push the
antenna’s connector into the socket.
Figure 2-4 Antenna connector (left) and antenna connected to 2103Ci (right)
2103 Modem Module
Section 2
Preparation and Installation
2-7
CAUTION
Only use this antenna with 2100 Series Ci cellular modem
modules. Do not use this antenna with non-Ci equipment.
Note
For the transmitter to comply with FCC Maximum Permissible
Exposure (MPE) regulations, the antenna must be located a
minimum of 30 centimeters (12 inches) from the human body.
Note
When any communication connector is not in use, it should
always be capped. The cap will seal the connector to prevent
corrosion, prevent moisture from entering the unit, and improve
communications.
2.5 Connecting to
Flowlink
Once the system is installed, you will configure it in a new or
existing site using Isco’s Flowlink software.
Note
The 2103 and 2103Ci Modems require Flowlink 5.10.412a or
later. Earlier versions do not support the modems.
The first time you connect to the site, you must connect your
computer directly to the stack using Isco’s RS232 connect cable
(part #60-2004-046)
or
USB
port
connect
cable
(part #60-2004-507). Open Flowlink and go to the Connect screen
(Figure 2-5) by either selecting it from the pull down menu or
clicking on the Quick Connect icon.
Make sure the connection Type is Direct, and click on the 2100
Instrument icon to connect. Upon initial connection, Flowlink
creates a site file and adds it to the database. If the system
detects the addition of a new module to an existing site, it will
display the Site Resolution screen (Figure 2-6). Otherwise, it will
display the Site Info screen (Figure 2-7).
2103 Modem Module
Section 2
Preparation and Installation
2-8
Figure 2-5 Flowlink connect screen (version 5.10.616)
2103 Modem Module
Section 2
Preparation and Installation
2-9
Figure 2-6 Site resolution screen
To add the new module to an existing site, select the appropriate
site and click OK. To create a new site, select Create a New Site.
Click in the name field, enter the name for the site, and click OK.
Upon connection, the Site Info tab will appear.
2103 Modem Module
Section 2
Preparation and Installation
2-11
When the module has been added to the system, you will see the
Devices screen (Figure 2-8).
Figure 2-8 Devices screen - connected to site
This window displays all of the modules connected to the site.
2103 Modem Module
Section 2
Preparation and Installation
2-12
2.6 Modem Site
Connection
Connection Information for the site is set up on the Devices tab.
For the 2103Ci, you will first configure the modem on the Modem
tab (refer to Section 2.6.2). For the 2103, go straight to the
Devices tab.
2.6.1 2103 Modem
Connection
Select the “Modem” radio button. The default modem will appear
in the Modem field. To select a different modem for this site only,
click the arrow to view a pull-down list containing all modems
recognized by your Windows operating system. Enter the 2103’s
phone number and click Apply to save your settings.
Figure 2-9 Devices screen - 2103 module
2103 Modem Module
Section 2
Preparation and Installation
2-13
2.6.2 2103Ci Modem
Connection
To begin using the 2103Ci module’s cellular modem, you must set
up the modem’s operation on the Modem screen. The Modem
screen is only used to set up the cellular modem.
The 2103Ci’s digital cellular modem provides a choice of two dif-
ferent connection methods: Serial Over IP (IP) or
Circuit-Switched Data (CSD). The default Contact Connection
Mode is IP; this mode of communication is much faster, and does
not require analog infrastructure.
Note
CSD communication is being phased out by the service pro-
vider, Verizon. As of July 2012, no new CSD lines can be
added, and all CSD service is scheduled to end in 2014.
The default TCP port is 1700, but you can change the port
number if necessary.
The APN settings are for GSM modems only, and do not apply to
this equipment.
Figure 2-10 Modem screen - configuring the cellular modem
Select IP (recommended)
or CSD
TCP port number
(default = 1700)
2103 Modem Module
Section 2
Preparation and Installation
2-14
Connection Information for the site is set up on the Devices
screen. Click the Disconnect button to activate the fields.
Serial over IP connection
To set up a serial-over-IP connection, select the “TCP” radio
button on the Devices tab. In the TCP address field, to the left of
the colon, enter the 2103Ci modem’s IP address, found on the
serial tag on the back of the module. Click Apply to save your set-
tings.
Figure 2-11 Devices screen - configuring the modem for Serial-Over-IP communication
Select TCP
radio button.
Enter IP
Address
2103 Modem Module
Section 2
Preparation and Installation
2-15
CSD connection
To set up a CSD connection, select the “Modem” radio button.
The default modem will appear in the Modem field. To select a
different modem for this site only, click the arrow to view a
pull-down list containing all modems recognized by your
Windows operating system. Lastly, enter the phone number of
the 2103Ci’s modem, found on the serial tag on the back of the
module. Click Apply to save your settings.
Figure 2-12 Devices screen - configuring the modem for CSD communication
Select Modem
radio button.
Enter cell phone
number.
Select Windows
modem from list.
2103 Modem Module
Section 2
Preparation and Installation
2-16
If your cellular service is set up and ready for use through your
service provider, you can now connect to your interface module
via cellular connection.
Note
For systems using the cellular modem, a Power Control
method is highly recommended to conserve battery power.
Refer to Section 2.8.
2.7 Setting Up Text
Messaging (2103Ci)
To program the 2103Ci for text messaging, you must have the
Short Message Service (SMS) server number or the Telocator
Alphanumeric Protocol (TAP) and the parameter settings of your
cell phones and/or pagers.
1. Make sure you are connected to the site, and select the
Alarms tab (Figure 2-13).
Figure 2-13 Setting up text messaging and alarm conditions
2. Under Alarm Notification, select the Alarm type (SMS or
TAP). SMS is recommended. (The TAP option has been
retained for legacy systems; you may be unable to obtain a
number or service.)
3. In the Phone Number list box, type the contact telephone
numbers. These must be valid SMS or TAP access num-
bers. You must enter at least one number; you can enter as
many as five. When an alarm condition is triggered, the
system will try dialing each number in the list.
2103 Modem Module
Section 2
Preparation and Installation
2-17
a. To find a TAP number, and the communications param-
eter settings, consult with your pager service, or go to
http://avtech.com/Support/TAP/index.htm or to
http://www.notepage.net/tap-phone-numbers.htm
4. If you selected TAP, the phone number list area will be
slightly different. For each phone number entered, enter a
Pager ID, which is the phone number of the cell phone or
pager that will receive the text message. (The ID will come
from your pager service.) You can enter an optional pass-
word of up to six characters in length. In the Comm field,
select 7E1 or 8N1 for the communications control specifica-
tion (your pager service has this information).
5. You must enter a message to be sent to the pager. In the field
labeled Message, type the outgoing text message (maximum
of 32 characters).
6. Under “Alarm condition,” select the retry number and
interval, and define the alarm condition using the Equa-
tion Builder. Defined alarm conditions will appear on the
pull-down list.
7. Click the Apply button or press F9 when you are done, to
update the module’s settings.
2.8 Power Conservation
for Cellular Modems
(2103Ci)
If you are using a 2103Ci, you may want to conserve battery
power by setting up a Power Control method using the equation
builder. Rather than have the internal modem continuously
enabled, you can specify conditions for when it is powered up. To
begin configuring the power settings, click the Wireless Power
Control tab (Figure 2-14). Then click on Set Equation. (The Set
Wireless Schedule button is there only for legacy systems.)
Note
Power Control applies only to the internal modem, not the
module itself. However, during the specified periods when the
modem is disabled, you will only be able to communicate with
the module via direct connection.
2103 Modem Module
Section 2
Preparation and Installation
2-18
Figure 2-14 Wireless power control window
2103 Modem Module
Section 2
Preparation and Installation
2-19
2.8.1 Equation
The Equation Builder screen lets you control the modem’s On/Off
time through the use of equations (Figure 2-15). When equation
conditions are true, the site will turn the modem on. When condi-
tions are false, the modem is turned off, conserving power.
To build an equation (Figure 2-15), select the Module Name the
equation will affect. Choose an equation from the Equation List,
or click Add and type in a descriptive name. Select a condition
from the list on the bottom half of the screen and click on Edit
Condition or, if no conditions are defined, click Edit Condition.
Figure 2-15 Equation Builder for power control
Selecting Edit Condition opens up the Condition Builder screen
where you can define a new condition or edit an existing one
(Figure 2-14). Define your parameters and press OK.
Figure 2-16 Condition Builder
2103 Modem Module
Section 2
Preparation and Installation
2-20
2.9 Pushed Data
Capability
The 2103 can automatically send data to a designated server
running Isco Flowlink Pro software, using 1xRTT
packet-switched data transmission (2103Ci), or a modem listener
feature for bridging landline connections (2103). The user-spec-
ified primary data transmission interval (5 minutes to 24 hours)
can automatically change to a secondary interval when specific
site conditions occur at the monitoring site. An Oracle® or
Microsoft® SQL database is required to use this feature. Contact
the factory for additional information.
To use the data push capability, you need to make sure you are
connected to the modem module, and then select the Data tab
(Figure 2-17).
Figure 2-17 Data tab
Click the Pushed Data button to set up a schedule for the data to
be pushed to a Flowlink Pro server.
Detailed Flowlink instructions are beyond the scope of this
manual. Flowlink’s operating instructions are available in a
Windows Help format. You can access the Help topics for an
active window by clicking its Help button or by pressing F1 on
your computer’s keyboard. You can also access Help topics by
selecting Help from the Flowlink menu.
2103 Modem Module
Section 3 Modbus Protocol
Sections 3.1 through 3.5 give an overview of the basic capabilities
and operation of Modbus protocol as it applies to Isco 2100 Series
flow modules.
For a Glossary of Terms and Common Acronyms, see sections 3.4
and 3.5.
For Modbus technical specifications, turn to section 3.6.
3.1 Introduction
Modbus is a simple command/response mechanism to read from
and write to specific memory locations called registers. A register
is a holding place for a piece of digital information within the
equipment. There are three standard protocols for Modbus:
Modbus RTU, Modbus TCP/IP, and Modbus ASCII. The Isco 2100
Series devices use Modbus ASCII protocol, the method discussed
in this manual. Modbus ASCII has more liberal communication
timing requirements. Modbus communication for the Isco 2100
Series provides a standard protocol that can be used to retrieve
real-time data from a single module or stack of modules at a site,
or multiple sites, over a wide area. The data can be sent to a
central computer for display, data collection, or process control.
Modbus implementation is independent of Flowlink and cannot
alter the Flowlink-programmed configuration of the module.
Modbus cannot be used to retrieve historical data from a
module’s memory.
Due to the wide variety of configurations that can be made with
Modbus, it is impossible to cover every usable application. This
section will discuss the overall capabilities and operation of
Modbus.
3.2 Operation
There are many standard, third party Modbus drivers and OPC
servers that may be used to link a remote Modbus device, such as
a 2100 Series module, to SCADA or process control software,
such as Wonderware™or Intellution™. The OPC server commu-
nicates with the remote instrumentation and accesses registers.
The definition of what information is contained and where (the
register number, or address) is decided by the manufacturer
(Teledyne Isco).
In a 2100 module, the registers hold, but are not limited to, the
current real-time value of the meter’s level, velocity, flow, input
voltage, temperature, and total flow readings, stored in specified
register locations. A list of the 2100 register addresses, and what
parameters are held where, is available in section 3.6.
2103 Modem Module
Section 3
Modbus Protocol
3-2
By accessing these registers you can obtain the current value of
whatever parameter you desire. The reading(s) can then be dis-
played or stored wherever you designate as a destination; for
example, a process control computer.
Note
Level, flow, velocity, and temperature data is stored in metric
units only.
Not all registers are limited to read-only data storage. You can
also use some registers for control purposes. For example, by
writing a “1” value to register 24 (“Identify Module” register), you
will tell a 2100 module to light the LED on the front of the
module.
3.2.1 Establishing
Communication
There are several different communications protocols supported
in the 2100 series that require auto-baud rate detection. Because
of this, each time a modbus connection is made, the module uses
a polling mechanism to repeatedly send a command until a
response is received. It may take up to 20 command retries
before the module has identified the baud rate and a response is
received.
3.2.2 Module Addressing
When connecting to a site via a Modbus OPC server, you use a
dedicated line of communication to that module or stack from the
OPC server, which can be a dedicated communications cable
(direct connection) or a dedicated phone number (modem).
When you are using a direct connection, you are dedicating a
specified COM port on the computer, and that COM port deter-
mines the site to which you are connecting.
When you are using a modem, the dedicated line is defined by
the site's phone number.
If you connect more than one 2100 Series module at a site, the
Modbus OPC server, while using the shared communication line
for all of the modules within the network, must have some way to
differentiate between the modules. When sending a command to
a specific module, the command has an address field. This allows
the server software to talk to, as well as control, the specified
module, while ignoring other modules in the same stack or site.
Each module capable of Modbus Protocol communication will
automatically create its own specific ASCII address within the
site, using:
• The model numbers of the modules
• The user-defined module names
2103 Modem Module
Section 3
Modbus Protocol
3-3
3.3 Configurations
A variety of configurations can be made with Modbus, either
through direct connection or through a modem.
In the example shown in Figure 3-1, you are direct-connecting a
server PC to two individual 2100 sites through Modbus, using
the COM ports on the OPC Server, which are directly connected
to the remote sites.
Connection to the module is made through the RS-232 communi-
cation port on the top of the module.
Note
For low power operation, we recommend connecting the mod-
ule(s) to the computer using the straight-through cable
(Isco part number 60-5314-529), which consumes less power,
instead of our standard interrogation cable.
In Figure 3-1, the OPC Server PC must have two COM ports.
Modbus requires one COM port each, for direct connection of
each 2150.
Figure 3-1 Configuration example
The operation sequence for the example above can be summa-
rized in the following steps:
2150:
1. 2150s take readings from probes.
2. 2150s store readings (level, velocity, flow rate, etc.) in their
specified registers.
Process Control:
3. The user requests data through Process Control.
4. Process Control asks the OPC server to gather informa-
tion.
2150
2150
OPC
Server
Process
Control
COM
port 1
COM
port 2
2103 Modem Module
Section 3
Modbus Protocol
3-4
5. OPC connects to the 2150 stack through the cable (direct
connection), takes register data from the specified 2150,
and populates the OPC server's holding index.
6. Process Control takes data from the OPC server's holding
index and gives data to the user.
Note that Process Control can be either manual or automated in
this example, and that the OPC server and Process Control may
be located physically on the same computer.
3.4 Glossary of Terms
ASCII – Short for American Standard Code for Information
Interchange, ASCII is a code that represents English characters
with numbers. Most computers represent text with ASCII code,
making it possible for one computer or device to share data with
another.
2100 modules support Modbus ASCII protocol.
Dedicated Line – A telecommunications path reserved for com-
munication between two specified points and not shared among
multiple points.
Modbus Protocol – Modbus Protocol is a messaging structure
used to establish master-slave/client server communications
between
intelligent
devices.
Modbus
is
a
simple
command/response mechanism to read from and write to reg-
isters.
OPC – OPC (OLE for Process Control) means open connectivity
via open (free for use) standards. It is a series of software stan-
dards specifications that fill a need in automation (like printer
drivers did for Windows), acting as a translator for data trans-
mission and process control.
The specification defines a standard set of objects, interfaces, and
methods for use in process control and manufacturing auto-
mation applications to facilitate interoperability. There are hun-
dreds of OPC Data Access servers and clients.
Registers – Registers are locations in memory that have specific
data stored for retrieval or are used for control functions. A reg-
ister is a holding place for a piece of digital information within
the equipment. The definition of what is contained and where
(the registry number, or address) is decided by the manufacturer
(in this case Teledyne Isco).
SCADA – SCADA (Supervisory Control And Data Acquisition)
is a computer system for gathering and analyzing real-time data.
SCADA systems are used to monitor and control plant operation,
or equipment in industries such as telecommunications, water
and waste control, energy, oil and gas refining, and transpor-
tation.
The SCADA system transfers the information (for example,
where a leak has occurred in a pipeline), back to a central site,
alerting the home station of the leak, performing necessary
analysis and control (such as determining if the leak is critical),
and displaying the information in a logical and organized
manner.
2103 Modem Module
Section 3
Modbus Protocol
3-5
SCADA systems can be relatively simple, such as one that mon-
itors the environmental conditions of a small office building, or
very complex, such as a system that monitors all the activity in a
nuclear power plant or a municipal water system.
3.5 Common Acronyms
ASCII – American Standard Code for Information Interchange
DCS – Distributed Control Systems
MTU – Master Terminal Unit
OPC – Object Linking and Embedding (OLE) for Process Control
PLC – Programmable Logic Controller
RTU – Remote Terminal Unit
SCADA – Supervisory Control And Data Acquisition
TCP/IP – Transmission Control Protocol/Internet Protocol
3.6 Register Specifications
All numbers in the Modbus registers are stored most significant
byte first. If the polling device has a byte ordering of least signif-
icant byte first (an Intel-based PC, for example), the bytes will
need to be reversed after they are received.
The Modbus ASCII address is used to index the data by modules.
Modbus ASCII address 1 contains information related to the site.
The first register contains a 16-bit integer count of the number of
modules that have data to report. The maximum number of
modules that can be supported is 4.
Modbus ASCII addresses 2 through the number of the module in
the stack (N) minus 1 contain data from the individual modules.
The Modbus ASCII addresses will be sorted by the model
number, and then by module name, which is entered by the user
through Flowlink. This allows the user to control the ordering of
the addresses and easily predict what data will be in specific reg-
isters.
Every measured parameter has a corresponding status and mea-
surement time that are updated with each measurement.
The maximum number of supported measurements from all
modules in the system is 28.
The Modbus registers are assigned within 30 seconds after the
2100 module is powered up. To conserve power for the users who
do not use Modbus communications, no Modbus registers will be
updated with sensor readings until a Modbus master communi-
cates with the 2100 module.
2103 Modem Module
Section 3
Modbus Protocol
3-6
The register definitions for the Site Information device (Modbus
ASCII address 1) are in Table 3-1 below:
The register definitions for the individual modules (Modbus
ASCII addresses 2+(N-1)) are in Table 3-2 below:
Table 3-1 Modbus ASCII Address 1 Register Definitions
Register
Number(s)
Name
Data type
Units
Read/Write
1
Number of module (N) (1-4)
16 bit integer
None
Read
2-20
Site name
38-byte string
None
Read
Table 3-2 Modbus ASCII Address 2+(N-1) Register Definitions
Register Number(s)
Name
Data Type
Units
Read/Write
1-4
Model number
8-byte string
None
Read
5-23
Module name
38-byte string
None
Read
24a
Identify module
16 bit integer
None
Read/Write
25b
Take reading flag
16 bit integer
None
Read/Write
26c
Update interval
16 bit integer
Seconds
Read/Write
27d
Active flag 1
16 bit field
None
Read
28
Active flag 2
16 bit field
None
Read
29
Active flag 3
16 bit field
None
Read
30
Active flag 4
16 bit field
None
Read
40,41
Level
4-byte float
Meters
Read
42
Level status codee
16-bit integer
Read
43-48
Level time record
Timef
Read
55,56
Velocity
4-byte float
Meters/second
Read
57
Velocity status code
16-bit integer
Read
58-63
Velocity time record
Time
Read
70,71
Flow
4-byte float
Cubic Meters/sec
Read
72
Flow status code
16-bit integer
Read
73-78
Flow time record
Time
Read
85,86
Flow 1
4-byte float
Cubic Meters/sec
Read
87
Flow 1 status code
16-bit integer
Read
88-93
Flow 1 time record
Time
Read
100,101
Volume
4-byte float
Cubic Meters
Read
102
Volume status code
16-bit integer
Read
103-108
Volume time record
Time
Read
115,116
Volume 1
4-byte float
Cubic Meters
Read
2103 Modem Module
Section 3
Modbus Protocol
3-7
117
Volume 1 status code
16-bit integer
Read
118-123
Volume 1 time record
Time
Read
130,131
Voltage
4-byte float
Volts
Read
132
Voltage status code
16-bit integer
Read
133-138
Voltage time record
Time
Read
145,146
Temperature
4-byte float
Degrees Celsius
Read
147
Temperature status code
16-bit integer
Read
148-153
Temperature time record
Time
Read
160,161
Internal Temp
4-byte float
Degrees Celsius
Read
162
Internal Temp status code
16-bit integer
Read
163-168
Internal Temp time record
Time
Read
175,176
Analog channel 1
4-byte float
0-100 percent
Read
177
Analog channel 1 status code
16-bit integer
Read
178-183
Analog channel 1 time record
Time
Read
190,191
Analog channel 2
4-byte float
0-100 percent
Read
192
Analog channel 2 status code
16-bit integer
Read
193-198
Analog channel 2 time Record
Time
Read
205,206
Analog channel 3
4-byte float
0-100 percent
Read
207
Analog channel 3 status code
16-bit integer
Read
208-213
Analog channel 3 time record
Time
Read
220,221
Analog channel 4
4-byte float
0-100 percent
Read
222
Analog channel 4 status code
16-bit integer
Read
223-228
Analog channel 4 time record
Time
Read
235,236
Analog channel 5
4-byte float
0-100 percent
Read
237
Analog channel 5 status code
16-bit integer
Read
238-243
Analog channel 5 time record
Time
Read
250,251
Analog channel 6
4-byte float
0-100 percent
Read
252
Analog channel 6 status code
16-bit integer
Read
253-258
Analog channel 6 time record
Time
Read
265,266
Analog channel 7
4-byte float
0-100 percent
Read
267
Analog channel 7 status code
16-bit integer
Read
268-273
Analog channel 7 time record
Time
Read
280,281
Analog channel 8
4-byte float
0-100 percent
Read
282
Analog channel 8 status code
16-bit integer
Read
283-288
Analog channel 8 time record
Time
Read
1765, 1766
Signal strengthg
4-byte float
0-100 percent
Read
Table 3-2 Modbus ASCII Address 2+(N-1) Register Definitions (Continued)
Register Number(s)
Name
Data Type
Units
Read/Write
2103 Modem Module
Section 3
Modbus Protocol
3-8
1767
Signal strength status code
16-bit integer
Read
1768-1779
Signal strength time record
Time
Read
1780, 1781
Spectrum strengthg
4-byte float
0-100 percent
Read
1782
Spectrum strength status code
16-bit integer
Read
1783-1794
Spectrum strength time record
Time
Read
1795, 1796
Spectrum ratiog
4-byte float
0-100 percent
Read
1797
Spectrum ratio status code
16-bit integer
Read
1798-1809
Spectrum ratio time record
Time
Read
a.
A write to the Identify module register will cause the module to perform the identify operation which may be a steady
LED for a few seconds or a beep in the Field Wizard.
b.
Setting the Take Reading flag to 1 will cause the module to update the registers with current data readings. It will
be set to zero when the readings have all been updated. This may be used to initiate readings and poll for when
they are ready to be read. It may take up to 50 seconds to update all the readings, depending upon the flow condi-
tions. Setting the Take Reading flag to 2 causes an automatic, 15 second update of readings when a Modbus mas-
ter is polling the 2100.
c.
The Update Interval specifies an interval in seconds that the registers are automatically updated. It defaults to zero,
which indicates that no automatic updating will occur.
d.
The Active Flag (1-4) bit fields specify what fields/registers are active in the list. This provides support for a maxi-
mum of 64 fields. For example, if bit 0 of register 27 is set, the Level (registers 40,41) is active. If bit 1 of register
27 is set, then the Velocity (registers 55,56) is active. If bit 0 of register 28 is set, the Analog channel 7 (registers
265,266) is active.
e.
A non-zero status code indicates a measurement problem.
f.
Time is represented in a series of registers: Order is from lowest address to highest - Seconds (0-59), Minutes
(0-59), Hours (0-23), Days (1-31), Month (1-12) and Year (1977-2099).
g.
Optional software feature. Contact Teledyne Isco to order enable code for the module used for communication.
Table 3-2 Modbus ASCII Address 2+(N-1) Register Definitions (Continued)
Register Number(s)
Name
Data Type
Units
Read/Write
Table 3-3 2100 Measurement Parameters by Model Number*
2103, 2103c/g, 2105, 2105c/g 2108
2110
2150, 2151, 2151P
Voltage
Analog channel 1 Level
Level
Analog channel 2 Flow
Velocity
Analog channel 3 Volume
Flow
Voltage
Flow 1
Temperature Volume
Volume 1
Voltage
Temperature
Signal strength
Spectrum strength
Spectrum ratio
*Single module only; does not include any additional connected devices. Subject to change.
2103 Modem Module
Section 4 Maintenance
4.1 Maintenance
Overview
This section explains the maintenance requirements of the 2103
Modem Module.
The 2103 is designed to perform reliably in adverse conditions
with a minimal amount of routine service requirements. To keep
your system working properly, you should check the desiccant
and channel conditions at regular intervals.
Maintenance intervals are affected by many variables. Humidity
levels obviously affect the service life of the desiccant, and the
amount of debris in the stream can drastically alter the channel
conditions.
Experience is often the best tool to use when establishing
minimum maintenance intervals for your system. Until you have
gained an understanding of the AV Module’s operation under dif-
fering environmental conditions, a weekly maintenance interval
is recommended.
4.1.1 Cleaning
The 2103 Module case may be cleaned using a soft cloth, warm
water, and a mild detergent. Do not use an abrasive cleanser, or
you might scratch the surface of the case.
Before cleaning, make sure that all the protective connector caps
are in place to avoid damage to any of the connectors. You should
also ensure that no water or cleanser enters the desiccant unit.
4.2 Maintenance Kit
Some of the parts mentioned in the Maintenance section of this
manual are available in a maintenance kit.
Kit number 60-2009-004 contains three O-rings for the communi-
cation connectors, silicone lubricant to apply to the O-rings, a des-
iccant assembly, and a container of silica gel desiccant. A
maintenance instruction sheet is included with the kit.
You can order the kit by calling Teledyne Isco’s Customer Service
Department.
2103 Modem Module
Section 4
Maintenance
4-2
4.3 Desiccant
The 2103 Modem uses desiccant to protect the internal compo-
nents from moisture damage. The cartridge is filled with indi-
cating silica gel, which is blue or yellow when dry. As the
desiccant becomes saturated, the color changes from blue to pink,
or from yellow to green. Replace the desiccant before the entire
length of the cartridge turns pink or green.
4.3.1 Replacing the
Desiccant
The desiccant is contained in a cartridge located on the left side
of the 2103 Modem. To remove the cartridge, unscrew the collar
and slide the cartridge out of the 2103 Modem. The clear tube
reveals the silica gel desiccant inside.
To replace the silica gel desiccant:
1. Hold the cartridge upright with the collar at the top.
2. As shown in the margin, push the collar off the cartridge.
3. Empty the saturated silica gel beads or granules.
4. Fill the tube with new (Isco P/N 099-0011-03) or reacti-
vated (see section 4.3.2) silica gel desiccant.
5. Press the collar onto the tube.
6. Slide the cartridge into the 2103 Modem. Tighten the collar
to seal the cartridge in place.
4.3.2 Reactivating the
Desiccant
Silica gel beads and granules of desiccant can be reactivated.
CAUTION
Desiccant may produce irritating fumes when heated. Observe
the following precautions:
•
Use a vented oven in a well ventilated room.
•
Do not remain in the room while the regeneration is taking
place.
•
Use the recommended temperature. Avoid heating the
desiccant at higher than recommended temperatures.
There is the potential of irritating fumes coming from the des-
iccant during reactivation. Because of this, we urge you to use
caution, and to heat the desiccant in a well ventilated room.
Material Safety Data Sheets are in the back of this manual.
The desiccant’s ability to remove moisture may lessen with each
saturation/reactivation cycle, resulting in a need for more fre-
quent service. After several cycles, the desiccant may no longer
be effective as it saturates too quickly. At this point, replace the
desiccant.
Collar
Cartridge
2103 Modem Module
Section 4
Maintenance
4-3
Silica gel
To reactivate the silica gel desiccant, pour the spent desiccant
into a heat resistant container. Never heat the cartridge
assembly; it will melt. Heat the silica gel in a vented convection
oven at 212° to 350°F (100° to 175°C) for two to three hours, or
until the blue or yellow color returns. Allow the desiccant to cool
and store it in an airtight container until ready for use.
4.4 Hydrophobic Filter
If the 2103 is in a humid location or submerged, a hydrophobic
filter prevents water from entering the desiccant cartridge and
reference line. Any amount of water will plug the filter and it
must be rinsed with clean water and allowed to dry, or replaced
so that the case can be reliably ventilated.
Remove the hydrophobic filter with a 5/8" or 16mm socket. Gently
screw in the replacement filter (Isco part #209-0093-93).
If the hydrophobic filter frequently requires replacement, con-
sider relocating the modules so that they are better protected.
4.5 O-Rings
The communication connectors on the top and bottom of the 2103
Modem contain O-rings that need periodic lubrication and
replacement. The maintenance kit described in section 3.2 con-
tains O-rings and silicone lubricant.
Whenever you replace the O-rings, or have removed them from
the connectors for some reason, you should lubricate the O-rings
by applying lubricant around the circumference of the ring.
Note
Do not use petroleum-based lubricants. Petroleum-based
lubricants will cause the O-ring to swell and eventually deterio-
rate. Aerosol silicone lubricant sprays often use petroleum
based propellants. If you are using an aerosol spray, allow a
few minutes for the propellant to evaporate before proceeding.
4.6 How to Obtain Service
The internal components of the 2103 Modem are not user-ser-
viceable. The case is completely sealed to protect the internal
components. To repair the unit, the case must be broken open
and replaced. If you think your module requires repair, contact
Isco’s Technical Service Department.
Corresponding with a Teledyne Isco Technical Service Represen-
tative can often resolve the problem without the need to return
the item. If the difficulty cannot be resolved you will be issued a
Return Authorization Number (RAN) and information on
returning it to the factory.
Teledyne Isco
Technical Service Department
P.O. Box 82531
Lincoln, NE 68501
866-298-6174 or 402-464-0231
FAX: 402-465-3001
e-mail:
[email protected]
2103 Modem Module
Appendix A Replacement Parts
A.1 Replacement Parts
Replacement parts are called out in the following pages. Refer to
the call-out in the adjacent table to determine the part number
for the item.
Replacement parts can be purchased by contacting Teledyne
Isco’s Customer Service Department.
Teledyne Isco
Customer Service Department
P.O. Box 82531
Lincoln, NE 68501 USA
Phone: (800) 228-4373
(402) 464-0231
FAX:
(402) 465-3022
E-mail: [email protected]
2103 Modem Module
Appendix B Accessories
B.1 How to Order
Accessories can be purchased by contacting Teledyne Isco’s Cus-
tomer Service Department.
Teledyne Isco
Customer Service Dept.
P.O. Box 82531
Lincoln, NE 68501 USA
Phone: (800) 228-4373
(402) 464-0231
FAX:
(402) 465-3022
E-mail: [email protected]
B.2 General Accessories
2103 Instruction manual . . . . . . . . . . . . . . . . . . . . . . .69-2003-637
Flowlink Software . . . . . . . . . . . . . . . . . . . . . . . . . . . .(call factory)
Maintenance Kit . . . . . . . . . . . . . . . . . . . . . . . . . . . . .60-2009-004
Silica Gel Desiccant . . . . . . . . . . . . . . . . . . . . . . . . . . .099-0011-03
Hydrophobic Filter. . . . . . . . . . . . . . . . . . . . . . . . . . . .209-0093-93
Modem Cable for analog 2103 module . . . . . . . . . . . .60-3214-020
Magnetic Mount Dual Band Antenna. . . . . . . . . . . . .60-2004-564
CAUTION
Only use this antenna with 2100 Series Ci cellular modem
modules. Do not use this antenna with non-Ci equipment.
2103 Modem Module
Appendix C Material Safety Data Sheets
C.1 Overview
This appendix to the manual provides Material Safety Data
Sheets for the desiccant used by the 2103 Modem Module.
Teledyne Isco cannot guarantee the accuracy of the data. Specific
questions regarding the use and handling of the products should
be directed to the manufacturer listed on the MSDS.
2103 Modem Module
Appendix C Material Safety Data Sheets
C-2
*Optional
Indicating Silica Gel
Material Safety Data Sheet
Indicating Silica Gel
Identity (Trade Name as Used on Label)
Manufacturer
:
MULTISORB TECHNOLOGIES, INC.
(formerly Multiform Desiccants, Inc.)
MSDS Number* :
M75
Address:
325 Harlem Road
Buffalo, NY 14224
CAS Number* :
Phone Number (For Information):
716/824-8900
Date Prepared:
July 6, 2000
Emergency Phone
Number:
716/824-8900
Prepared By* :
G.E. McKedy
Section 1 - Material Identification and Information
Components - Chemical Name & Common Names
(Hazardous Components 1% or greater; Carcinogens 0.1% or
greater)
%*
OSHA
PEL
ACGIH
TLV
OTHER LIMITS
RECOMMENDE
D
Silica Gel SiO2
98.0
6mg/m3
(total dust)
10mg/m3
(total dust)
Cobalt Chloride
>2.0
0.05mg/m3
(TWA cobalt
metal dust &
fume)
.05mg/m3
(Cobalt, TWA)
Non-Hazardous Ingredients
TOTAL
100
Section 2 - Physical/Chemical Characteristics
Boiling
Point
N/A
Specific Gravity
(H20 = 1)
2.1
Vapor Pressure
(mm Hg and Temperature
N/A
Melting
Point
N/A
Vapor
Density
(Air =1)
N/A
Evaporation Rate
(__________=1)
N/A
Solubility
in Water
Insoluble, but will adsorb moisture.
Water
Reactive
Not reactive, but will adsorb moisture.
Appearance
and Odor
Purple crystals, no odor.
Section 3 - Fire and Explosion Hazard Data
Flash Point and
Methods Used
N/A
Auto-Ignition
Temperature
N/A
Flammability Limits in
Air % by Volume
N/A
LEL
UEL
Extinguisher
Media
Dry chemical, carbon dioxide and foam can be used.
Special Fire
Fighting Procedures
Water will generate heat due to the silica gel which will adsorb water and liberate heat.
Unusual Fire and
Explosion Hazards
When exposed to water, the silica gel can get hot enough to reach the boiling point of water. Flooding with
water will reduce the temperature to safe limits.
Section 4 - Reactivity Hazard Data
STABILITY
Stable
Unstable
Conditions
To Avoid
Moisture and high humidity environments.
Incompatibility
(Materials to Avoid)
Water.
Hazardous
Decomposition
Products
Carbon dioxide, carbon monoxide, water
HAZARDOUS POLYMERIZATION
May Occur
Conditions
To Avoid
None.
2103 Modem Module
Appendix C Material Safety Data Sheets
C-3
*Optional
Indicating Silica Gel
Page 2
Section 5 - Health Hazard Data
PRIMARY ROUTES
Inhalation
Ingestion
OF ENTRY
Skin Absorption
Not Hazardous
CARCINOGEN
NTP
OSHA
LISTED IN
IARC Monograph
Not Listed
HEALTH HAZARDS
Acute
May cause eye, skin and mucous membrane irritation.
Chronic
Prolonged inhalation may cause lung damage.
Signs and Symptoms
of Exposure
Drying and irritation.
Medical Conditions
Generally Aggravated by Exposure
Asthma.
EMERGENCY FIRST AID PROCEDURES - Seek medical assistance for further treatment, observation and support if necessary.
Eye Contact
Flush with water for at least 15 minutes.
Skin
Contact
Wash affected area with soap and water.
Inhalation
Remove affected person to fresh air.
Ingestion
Drink at least 2 glasses of water.
Section 6 - Control and Protective Measures
Respiratory Protection
(Specify Type)
Use NIOSH approved dust mask or respirator.
Protective
Gloves
Light cotton gloves.
Eye Protection
Safety glasses.
VENTILATION
Local Exhaust
Mechanical (General)
Special
TO BE USED
Other (Specify)
Other Protective
Clothing and Equipment
None.
Hygienic Work
Practices
Avoid raising dust. Avoid contact with skin, eyes and clothing.
Section 7 - Precautions for Safe Handling and Use/Leak Procedures
Steps to be Taken if Material
Is
Spilled Or Released
Sweep or vacuum up and place the spilled material in a waste disposal container. Avoid raising dust.
Waste Disposal
Methods
Dispose in an approved landfill according to federal, state and local regulations.
Precautions to be
Taken
In Handling and
Storage
Cover promptly to avoid blowing dust. Wash after handling.
Other Precautions and/or Special
Hazards
Keep in sealed containers away from moisture. The silica gel will readily adsorb moisture.
2103 Modem Module
Appendix C Material Safety Data Sheets
C-4
MATERIAL SAFETY DATA SHEET
Effective Date
March 8, 2005
MSDS Number
M163
Section 1 Product and Company Information
Product Name:
Silica gel, indicating, yellow
Product Use:
Desiccant, absorbent
Grades:
Silica gel, indicating
Synonyms:
Amorphous silica gel, SiO2, silicon dioxide (amorphous)
Company;
Multisorb Technologies, Inc.
Street Address:
325 Harlem Road
City, State, Zip, Country:
Buffalo, NY 14224-1893 USA
Telephone Number:
(716) 824 8900 [USA] Monday - Friday (8:00 - 5:00 EDT)
Fax Number:
(716) 824 4091 [USA]
Website / E-Mail :
multisorb.com
Section 2 Composition / Information on Ingredients
Component Name
CAS Number
% by Weight
Synthetic amorphous silica gel (SiO2)
112926-00-8
100
Phenolphthalein
77-09-08
100 ppm
While this material is not classified, this MSDS contains valuable information critical to the safe handling
and proper use of this product. This MSDS should be retained and available for employees and other users
of this product.
Section 3 Hazard Identification
Emergency Overview:
A yellow bead or granular material that poses little or no immediate hazard.
This material is not combustible.
Potential Health Effects:
Eyes:
Dust and or product may cause eye discomfort and irritation seen as tearing and reddening.
Skin:
The product dust may cause drying of the skin. Silica gel may get hot enough to burn skin
when it adsorbs moisture rapidly. Use an excess of water to cool the silica gel.
Ingestion:
Material is not toxic and will pass through the body normally.
Inhalation: Slight irritation is possible but none is expected.
Medical Effects Generally Aggravated by Exposure:
Respiratory ailments.
Chronic Effects/Carcinogenity:
May cause eye, skin and mucous membrane irritation and drying.
2103 Modem Module
Appendix C Material Safety Data Sheets
C-5
Section 4 First Aid Measures
Eyes:
Rinse the eyes well with water while lifting the eye lids. If irritation persists, consult a
physician.
Skin:
Wash affected area with soap and water.
Ingestion:
Ingestion is unlikely, this material will pass through the body normally.
Inhalation:
Remove the affected person to fresh air and get medical attention if necessary.
Notes to Physician:
Not applicable
Section 5 Fire Fighting Measures
Flammable Properties:
Not flammable
Flash Point:
Not applicable
Method:
Not applicable
Flammable Limits:
Not flammable
Lower Flammability Limit:
Not applicable
Upper Flammability Limit: Not applicable
Autoignition Temperature:
Not applicable
Hazardous Combustion Products:
Not applicable
Extinguishing Media: Use extinguishing media that is appropriate for the surrounding fire. Silica gel is
not combustible.
Fire Fighting Instructions:
Not combustible
Unusual Fire and Explosion Hazards:
None
Section 6 Accidental Release Measures
Spill:
Sweep or vacuum up and place the spilled material in a waste disposal container. Avoid raising dust.
Wash with soap and water after handling.
Section 7 Handling and Storage
Handling:
Avoid raising dust and minimize the contact between worker and the material. Practice
good hygienic work practices.
Storage:
Store in a cool, dry location. Keep in sealed containers away from moisture. The silica gel
will readily adsorb moisture.
2103 Modem Module
Appendix C Material Safety Data Sheets
C-6
Section 8 Exposure Controls/Personal Protection
Engineering Controls:
Use exhaust ventilation to keep the airborne concentrations below the exposure
limits.
Respiratory Protection: Use NIOSH approved respirator when the air quality levels exceed the TLV's.
Skin Protection:
Light gloves will protect against abrasion and drying of the skin.
Eye Protection:
Safety glasses.
Component Name
Exposure Limits
OSHA
PEL
ACGIH
TLV
Other
Recommended
Limits
Silica gel
TWA 20 mppcf
(80 mg / m3 % SiO2)
TWA 10 mg / m3
NIOSH REL
TWA 6 mg / m3
IDLH 3000 mg / m3
Phenolphthalein
Not Applicable
Not Applicable
Not Applicable
Section 9 Physical and Chemical Properties
Appearance:
Yellow beads or granules
Vapor Density:
Not applicable
Odor:
None
Boiling Point:
4046q F (2230q C)
Physical State:
Solid bead
Melting Point:
3110q F (1710q C)
PH:
Not applicable
Solubility:
Insoluble in water
Vapor Pressure:
Not applicable
Specific Gravity:
2.1
Section 10 Stability and Reactivity
Stability:
Stable
Conditions to avoid:
Moisture and high humidity environments.
Incompatibility:
Water, fluorine, oxygen difluoride, chlorine trifluoride
Hazardous Decomposition Products:
None
Hazardous Polymerization:
Will not occur
2103 Modem Module
Appendix C Material Safety Data Sheets
C-7
Section 11 Toxicological Information
This product and its components are not listed on the NTP or OSHA Carcinogen lists.
Animal Toxicology Tests for DOT Hazard classification
( Tests Conducted on finely ground silica gel)
1 - hour LC50 (rat) > 2 mg / l
48 - hour oral LD50 (rat) est. > 31,600 mg / kg
48 - hour dermal LD50 (rabbit) est. > 2,000 mg / kg
Considered an ocular irritant
Human Toxicology Silica gel is a synthetic amorphous silica not to be confused with crystalline silica.
Epidemiological studies indicate low potential for adverse health effects. In the activated form, silica gel
acts as a desiccant and can cause a drying irritation of the mucous membranes and skin in cases of severe
exposure. Multisorb Technologies Inc. knows of no medical conditions that are abnormally aggravated by
exposure to silica gel. The primary route of entry is inhalation of dust.
Section 12 Ecological Information
Not known to have any adverse effect on the aquatic environment. Silica gel is insoluble and non-toxic.
Section 13 Disposal Information
Disposal Information If this product as supplied becomes a waste, it does not meet the criteria of a
hazardous waste as defined under the Resource Conservation and Recovery Act (RCRA) 40 CFR 261.
Materials of a hazardous nature that contact the product during normal use may be retained on the product.
The user of the product must identify the hazards associated with the retained material in order to assess the
waste disposal options. Dispose according to federal, state and local regulations.
Section 14 Transportation Information
U.S. Department of Transportation Shipping Name:
Not classified as a hazardous material. Not regulated.
Section 15 Regulatory Information (Not meant to be all inclusive - selected regulations represented)
TSCA Listed:
Yes
DSL/NDSL (Canadian) Listed:
Yes
OSHA:
TWA 20 mppcf (80 mg / m3 % SiO2) for Silica gel
NIOSH:
REL TWA 6 mg / m3 IDLH 3,000 mg / m3 for silica gel
Animal tests conducted in 1976 - 1978. 18 month exposure at 15 mg / m3 showed silica
deposition in respiratory macrophages and lymph nodes, minimum lung impairment, no silicosis.
ACGIH:
TLV - 10 mg / m3 for Silica gel
DOT:
Not classified as a hazardous material.
2103 Modem Module
Appendix C Material Safety Data Sheets
C-8
Section 16 Other Information
HMIS Hazardous Materials Identification System
HMIS Rating
Health
Flammability
Reactivity
0
0
0
0 - minimal hazard, 1 - slight hazard, 2 - moderate hazard, 3 - serious hazard, 4 - severe hazard
This MSDS was prepared by: George E. Mckedy
Senior Applications Development Specialist
Multisorb Technologies, Inc.
This data and recommendations presented in this data sheet concerning the use of our product and the materials
contained therein are believed to be correct but does not purport to be all inclusive and shall be used only as a guide.
However, the customer should determine the suitability of such materials for his purpose before adopting them on a
commercial scale. Since the use of our products is beyond our control, no guarantee, expressed or implied, is made and
no responsibility assumed for the use of this material or the results to be obtained therefrom. Information on this form
is furnished for the purpose of compliance with Government Health and Safety Regulations and shall not be used for
any other purposes. Moreover, the recommendations contained in this data sheet are not to be construed as a license to
operate under, or a recommendation to infringe, any existing patents, nor should they be confused with state, municipal
or insurance requirements, or with national safety codes.
2103 Modem Module
Index
Numerics
2103Ci label Information, 2-1
A
accessories, B-1
alarms, 2-16
analog modem, 2-4
antenna, 2-6
antenna 2103Ci, 2-6
C
components
bottom view, 1-3
identification, 1-2
top view, 1-2
connectors, 2-3
analog modem cable, 1-5
analog modem communication, 1-5, 2-5
contact information, 2-1, 4-3, A-1, B-1
technical service, 1-6, 4-3
D
desiccant, 4-2
F
FCC compliance, 2-4
Flowlink, 2-7
alarms, 2-16
connection information, 2-7, 2-12
equation builder, 2-19
modem connection, 2-12, 2-13
power conservation, 2-17
pushed data, 2-20
text messaging, 2-16
I
installation, 2-2
L
latches, 2-2
M
maintenance, 4-1
desiccant, 4-2
hydrophobic filter, 4-3
o-rings, 4-3
MODBUS, 3-1
acronyms, 3-5
configurations, 3-3
glossary of terms, 3-4
register specifications, 3-5
modem connection
cellular, 2-13
CSD, 2-15
landline, 2-12
serial over IP, 2-14
modem modules
connecting, 2-5, 2-12
power conservation, 2-17
pushed data, 2-20
text messaging, 2-16
modem screen, 2-13
MSDS, C-1
P
parts, A-1, B-1
power conservation, 2-17
equation, 2-19
modem modules, 2-17
pushed data, 2-20
R
replacement parts, A-1
S
safety information, 1-6, C-1
software configuration, 2-7
specifications, 1-4
stacking, 2-2, 2-3
T
TCP connection, 2-14
technical service, 1-6
text messaging, 2-16
Table Hazmat 2100 Nodes
60-2003-550 Rev.
ℶ❐₼㦘㹡㦘⹂䓸德㒥⏒侯䤓⚜䱿♙⚺摞
Name and amount of Hazardous Substances or Elements in the product
᳝↦᳝ᆇ⠽䋼ܗ㋴
Hazardous Substances or Elements
捷ↅ⚜䱿
Component Name
䪙
(Pb)
∲
(Hg)
䬝
(Cd)
݁Ӌ䫀
(Cr(VI))
⒈㘨㣃
(PBB)
⒈Ѡ㘨㣃
(PBDE)
兎恾㨎
Circuit Boards
X
O
O
O
O
O
ℶ❐₼㦘㹡㦘⹂䓸德㒥⏒侯䤓⚜䱿♙⚺摞᧶Name and amount of Hazardous Substances or Elements in
the product
O: 嫷䯉年㦘㹡㦘⹂䓸德⦷年捷ↅ㓏㦘⧖德㧟㠨₼䤓⚺摞⧖⦷ST/ 㪖屓⸩䤓棟摞尐㻑ⅴₚᇭ
O: Represent the concentration of the hazardous substance in this component’s any homogeneous pieces is
lower than the ST/ standard limitation.
X᧶嫷䯉年㦘㹡㦘⹂䓸德咂⺠⦷年捷ↅ䤓㩟⧖德㧟㠨₼䤓⚺摞怔⒉ST/ 㪖屓⸩䤓棟摞尐㻑ᇭ
(←₩♾⦷㷳⮓᧨㫈㗽⸭棔㍔⑄⺈ₙ嫷₼㓢“X” 䤓㔏㦾☮⥯扪嫛扪㷴広㢝ᇭ)
X: Represent the concentration of the hazardous substance in this component’s at least one homogeneous
piece is higher than the ST/ standard limitation.
(Manufacturer may give technical reasons to the “X”marks)
䘾≬∎䞷㦮䟀兞洛䫽⸩ᇭ
The Environmentally Friendly Use Period (EFUP) was determined through experience.
䞮ℶ㡴㦮嬺冥䪐⦷侊⒦⚆䪐₼ᇭⓜₘ⇜㟿ⷦ䞮ℶ(207 ⅲ嫷2007 ) ᇭ椞⚝䤓₹ⷦ㹜ⅲ嫷㦗᧶
A 㦗᧨B ℛ㦗᧨䷘䷘ᇭ
The date of Manufacture is in code within the serial number. The first three numbers are the year of
manufacture (207 is year 2007) followed by a letter for the month. "A" is January, "B" is February and so on.
DECLARATION OF CONFORMITY
Application of Council Directive:
89/336/EEC – The EMC Directive
73/23/EEC – The Low Voltage Directive
Manufacturer's Name:
Teledyne Isco, Inc.
Manufacturer's Address:
4700 Superior, Lincoln, Nebraska 68504 USA
Mailing Address: P.O. Box 82531, Lincoln, NE 68501
Equipment Type/Environment:
Laboratory Equipment for Light Industrial/Commercial Environments
Trade Name/Model No:
2103 Modem Module
Year of Issue:
2001
Standards to which Conformity is Declared:
FCC part is Sub part B, IC ES-003, En 55022:1998, EN 55024:1998 EMC
Requirements for Information Technology Equipment
EN 60950 Safety Requirements for Information Technology Equipment
FCC Part 68
Standard
Description
Severity Applied
Performance Criteria
EN61000-4-2
Electrostatic Discharge
Level 2 - 4kV contact discharge
Level 3 - 8kV air discharge
A
A
EN61000-4-3
Radiated RF Immunity
80 MHz to 1000MHz 80% AM at 1kHz
Level 1 – 10V/m
A
EN61000-4-4
Electrical Fast Transient
Level 1 – 1kV on I/O lines
A
EN61000-4-5
Surge on I/O Lines
1kV common mode,
0.5KV differential mode
A
*EN61000-4-6
Conducted RF on I/O lines
150 kHz to 80 MHz,
3V rms, 80% modulated
*C
CISPR11/
EN 55011
RF Emissions
Group 1, Class A Industrial, Scientific, and
Medical Equipment
* Between 5 and 7 mega Hz the modem will inadvertently disconnect from the computer. A manual reconnection procedure is
required to restablish a connection.
We, the undersigned, hereby declare that the design of the equipment specified above conforms to the above Directive(s) and
Standards as of July 1, 2001.
William Foster
USA Representative
William Foster
Director of Engineering
Teledyne Isco, Inc.
4700 Superior Street
Lincoln, Nebraska 68504
Phone: (402) 464-0231
Fax: (402) 464-4543
60-2002-156
Rev. B
DECLARATION OF CONFORMITY
Application of Council Directive:
89/336/EEC The EMC Directive
73/23/EEC The Low Voltage Directive
Manufacturer's Name:
Teledyne Isco, Inc.
Manufacturer's Address:
4700 Superior, Lincoln, Nebraska 68504 USA
Mailing Address: P.O. Box 82531, Lincoln, NE 68501
Equipment Type/Environment:
Laboratory Equipment for Light Industrial/Commercial Environments
Trade Name/Model No:
2191 Battery Module
Year of Issue:
2001
Standards to which Conformity is Declared:
EN 61326-1998 EMC Requirements for Electrical Equipment for
Measurement, Control, and Laboratory Use
EN 61010-1 Safety Requirements for Electrical Equipment for Measurement,
Control and Laboratory use.
Standard
Description
Severity Applied
Performance Criteria
EN61000-4-2
Electrostatic Discharge
Level 2 - 4kV contact discharge
Level 3 - 8kV air discharge
B
B
EN61000-4-3
Radiated RF Immunity
80 MHz to 1000MHz 80% AM at 1kHz
Level 1 10V/m
B
EN61000-4-4
Electrical Fast Transient
Level 1 1kV on I/O lines
B
EN61000-4-5
Surge on I/O Lines
1kV common mode,
0.5KV differential mode
B
EN61000-4-6
Conducted RF on I/O lines
150 kHz to 80 MHz,
3V rms, 80% modulated
B
CISPR11/
EN 55011
RF Emissions
Group 1, Class A Industrial, Scientific, and
Medical Equipment
We, the undersigned, hereby declare that the design of the equipment specified above conforms to the above Directive(s) and
Standards as of July 1, 2001.
William Foster
USA Representative
William Foster
Director of Engineering
Teledyne Isco, Inc.
4700 Superior Street
Lincoln, Nebraska 68504
Phone: (402) 464-0231
Fax: (402) 464-4543
60-2002-158
Rev A
Before returning any instrument for repair, please call, fax, or e-mail the Teledyne Isco Service
Department for instructions. Many problems can often be diagnosed and corrected over the
phone, or by e-mail, without returning the instrument to the factory.
Instruments needing factory repair should be packed carefully, and shipped to the attention of
the service department. Small, non-fragile items can be sent by insured parcel post. PLEASE
BE SURE TO ENCLOSE A NOTE EXPLAINING THE PROBLEM.
Shipping Address:
Teledyne Isco - Attention Repair Service
4700 Superior Street
Lincoln, NE 68504 USA
Mailing Address:
Teledyne Isco
PO Box 82531
Lincoln, NE 68501 USA
Phone:
Repair service:
(800) 775-2965 (lab instruments)
(866) 298-6174 (samplers & flow meters)
Sales & General Information: (800) 228-4373 (USA & Canada)
Fax:
(402) 465-3001
Email:
[email protected]
February 28, 2012 P/N 60-1002-040 Rev G
Teledyne Isco One Year Limited Factory Service Warranty*
This
warranty
exclusively
covers
Teledyne
Isco
instruments, providing a one-year limited warranty
covering parts and labor.
Any instrument that fails during the warranty period due to
faulty parts or workmanship will be repaired at the factory
at no charge to the customer. Teledyne Isco’s exclusive
liability is limited to repair or replacement of defective
instruments. Teledyne Isco is not liable for consequential
damages.
Teledyne Isco will pay surface transportation charges both
ways within the 48 contiguous United States if the
instrument proves to be defective within 30 days of
shipment. Throughout the remainder of the warranty period,
the customer will pay to return the instrument to Teledyne
Isco, and Teledyne Isco will pay surface transportation to
return the repaired instrument to the customer. Teledyne
Isco will not pay air freight or customer’s packing and
crating charges. This warranty does not cover loss, damage,
or defects resulting from transportation between the
customer’s facility and the repair facility.
The warranty for any instrument is the one in effect on date
of shipment. The warranty period begins on the shipping
date, unless Teledyne Isco agrees in writing to a different
date.
Excluded from this warranty are normal wear; expendable
items such as charts, ribbon, lamps, tubing, and glassware;
fittings and wetted parts of valves; and damage due to
corrosion, misuse, accident, or lack of proper maintenance.
This warranty does not cover products not sold under the
Teledyne Isco trademark or for which any other warranty is
specifically stated.
No item may be returned for warranty service without a
return authorization number issued by Teledyne Isco.
This warranty is expressly in lieu of all other warranties
and obligations and Teledyne Isco specifically disclaims
any warranty of merchantability or fitness for a
particular purpose.
The warrantor is Teledyne Isco, 4700 Superior, Lincoln, NE
68504, U.S.A.
* This warranty applies to the USA and countries where Teledyne Isco does not have an authorized dealer.
Customers in countries outside the USA, where Teledyne Isco has an authorized dealer, should contact their
Teledyne Isco dealer for warranty service.