Excell ESW Operation & User’s Manual
Read Online
Read the Excell ESW Operation & User’s Manual online using the built-in PDF viewer.
Text of the manual
Waterproof Weighing scale
ESW
© Excell Precision Limited 2014. All rights reserved Worldwide.
The information contained herein is the property of Excell Precision Limited and is
supplied without liability for errors or omissions. No part may be reproduced or used
except as authorized by contract or other written permission. The copyright and the
foregoing restriction on reproduction and use extend to all media in which the
information may be embodied.
EXCELL PRECISION CO., LTD.
02005311
ZSME3000000272
Table of Content
INSTRUCTION FOR USE ...................................................................................................2
PREPARING TO USE THE SCALE.....................................................................................2
CHAPTER 1 INTRODUCTION ............................................................................................4
1-1 FEATURES AND SPECIFICATIONS.....................................................................................4
1-2 POWER SUPPLY............................................................................................................5
1-3 PANEL AND KEYBOARD INTRODUCTION ...........................................................................5
1-4 OPERATING THE SCALE .................................................................................................6
1-5 SELF-TEST MODE.........................................................................................................7
1-6 ERROR MESSAGES .......................................................................................................8
1-7 WEIGHT UNIT ...............................................................................................................8
CHAPTER 2 ADVANCED FUNCTIONS ..............................................................................9
2-1 ADVANCED FUNCTION SETTING TABLE............................................................................9
2-2 ADVANCED FUNCTION SETTING WORKFLOW .................................................................10
2-3 GENERAL FUNCTION SETTING
..............................................................11
2-3-1 Automatic Backlight Function Setting
.......................................12
2-3-2 Automatic Power-off Timer Setting
...........................................13
2-3-3 Hi/Lo/OK Function Setting
........................................................14
2-3-4 Restore to the Default Setting
...................................................15
2-3-5 Noise Filter Setting
...................................................................16
2-3-6 Hold Function Setting
..............................................................17
2-3-7 Auto Unit Weight Averaging Setting
........................................18
2-4 WEIGHT CALIBRATION
..............................................................................19
APPENDIX 1 7-SEGMENT DISPLAY CHARACTERS ......................................................20
EXCELL PRECISION CO., LTD.
02005311
ZSME3000000272
Instruction for Use
1. Please keep the scale in a cool dry place. Do not store it at high temperature.
2. Avoid objects impacting with the scale. Do not drop loads onto the scale or subject the
weigh pan to any strong shock loads.
3. The load placed on the weigh pan must not exceed the maximum weighing capacity of
the scale.
4. If the scale is not going to be used for some time, please clean it and store it in a plastic
bag in dry conditions. A desiccant sachet may be included to prevent any moisture
build up.
Preparing To Use the Scale
1. Adjust the four levelling feet (if fitted) to set the scale pan level using the spirit level
bubble located at the front of the scale.
2. Avoid operating the scale in direct sunlight or drafts of any kind.
3. If possible avoid connecting the scale to ac power outlet sockets which are adjacent to
other appliances to minimise the possibility of interference affecting the performance of
the scale.
4. Remove any weight that might be on the weigh pan before the scale is switched on and
avoid leaving weight on the pan for long periods of time
5. All goods weighed should be placed in the centre of the weigh pan for accurate
weighing. The overall dimensions of the goods being weighed should not exceed the
dimension of the weigh pan.
6. We suggest to warm up the scale for 15~20 minutes before operation to ensure best
accuracy.
7. Please note when the symbol keeps flashing on the screen, the batteries need to
be replaced.
8. Introduction of Storage Battery
Thank you for purchasing EXCELL waterproof
weighing scale. In order to use the scale properly,
please read this instruction carefully before use. If
you have a problem concerning the scale, please
contact your supplier.
EXCELL PRECISION CO., LTD.
02005311
ZSME3000000272
Due to the storage battery adopt the advanced free-maintaining technique, customers
need not to replenish electrolyte.
The scale should be recharged every 3 months to prevent failure of the internal
rechargeable battery.
1. The battery should be charged for 8~10 hours.
2. The temperature of battery should below 45℃.
Maintaining
1. Please do not discharge with over-current when using the battery. Please charge the
battery after discharging current.
2. Please take down the battery when the scale is not used for a long time or break the
connection of cathode.
3. Do not short the battery terminals to check whether there is current. Please check
whether the connection point is firm to guarantee good connection.
4. The battery should be replaced by specialized person. No reverse-battery or the
product will be damaged.
a) Anode of battery should be connected with Anode of product battery(usually red
cable)
b) Cathode of battery should be connected with Cathode of product battery (usually
brown cable or black cable)
c) See the picture following
Safety warnings
1. The electrolyte of battery is caustic which causes metal, cotton, etc to corrode.
2. The hydrogen will be resolved when using or charging the battery and it will cause
explosion when approaches fire.
Brown cable(or black cable)
connected with Anode of
battery
Red cable connected with
Cathode of battery
EXCELL PRECISION CO., LTD.
02005311
ZSME3000000272
Chapter 1 Introduction
1-1 Features and Specifications
Features:
Sealed waterproof silica gel strip blocks water from infiltrating into the scale.
Surrounded by waterproof grade sheeting to ensure the water free.
1/3,000~1/6,000 display resolution available.
Aluminum frames coated with anti-corrosion treatment for longer lifespan.
Stainless steel made feet and platter for anti-rust.
High speed of 24bits AD fast reacts and shortens the weighing operation duration.
Selectable units: Kilogram (kg), gram (g), ounce (oz), and pound (lb) weighing units
available.
Built-in rechargeable battery can be easily replaced.
Low power indication and auto power off.
Well-designed protection point for transportation.
Concave handle for easy carrying.
Specifications:
Model
Capacity
Division
Resolution
ESW-3
3kg
0.5g
1g
1/6,000
1/3,000
ESW-6
6kg
1g
2g
1/6,000
1/3,000
ESW-15
15kg
2g
5g
1/7,500
1/3,000
ESW-30
30Kg
5g
10g
1/6,000
1/3,000
Operating Temperature: Non-approval model: 0°C ~ 40°C (32°F ~ 104°F)
Approval model: -10°C ~ 40°C (14°F ~ 104°F)
Dimensions: 240 x 120 x 280 mm ( W x H x D)
Weight of the scale: Around 3.5kg
240mm
280mm
120mm
EXCELL PRECISION CO., LTD.
02005311
ZSME3000000272
.
. .
. . .
. . . .
Pcs
1-2 Power Supply
Power Supply Selection
1. DC 6V / 4Ah rechargeable battery
2. AC 110V~240V (±10%) adapter
Power Consumption
14mA (system no backlight;Single display) about 275 hours
19mA (system backlight;single display) about 200 hours
Low Battery Warning
Please note when the ( ) symbol keeps flashing on the display, the batteries should
be changed.
The scale will turn off automatically after a few hours when the low battery warning
symbol shows up. The scale must be fully charged, before operating again.
1-3 Panel and Keyboard Introduction
Icon Introduction
STABLE
: The weight is stable.
GROSS
: The scale is in the gross mode. The display shows the goods and any
container weight. This Gross status indication is on when the TARE
function is used.
: The unit weight is not sufficient. When the icon is on, the counting
function is operational but the count may contain errors.
: The sampling size is not sufficient When the icon is on, the counting
function is operational but the count may contain errors.
Keyboard Function
ON/ZERO KEY
This key possesses two functions: Power On and Zero function.
OFF KEY
When the scale is switched on, press the OFF key, the scale will switch off.
TARE KEY
The tare function will not operate during the following conditions:
EXCELL PRECISION CO., LTD.
02005311
ZSME3000000272
1. When the scale powers on if the weight is negative and after a container is placed on
the weigh pan if the weight is still below zero.
2. The tare value is over the full scale capacity.
UNITS KEY
Press the UNITS key to switch weight units; the icons will indicate the active units.
NET/GROSS KEY
In the Tare mode, the screen displays the “TARE” icon; press the NET/GROSS key to
switch between the “Net value” and the “Gross value”.
1-4 Operating the Scale
POWER ON
When the scale is off, press the ON/ZERO key, the scale will switch on.
POWER OFF
When the scale is on, press the OFF key, the scale will switch off.
ZERO
When the weigh pan is empty (free of load) and the display is not showing zero, press
the ON/ZERO key to zero the scale. At zero, the “Æ0Å” indication is on.
When the weight value is within the zero range, the zero function operates to zero the
scale or cancel the tare function.
Zero range: The OIML & NTEP models have a zero range of ± 2% of Full Scale. The
Sri Lanka model has a zero range of ± 4% of Full Scale.
SWITCHING UNITS
Press the UNITS key to switch weight units, the icons or arrows will indicate the
active units as appropriate. The units available are dependent on the exact scale
model.
After power off, the scale will memorize the active units. When the scale is powered on
again, it displays the previously active units.
TARE FUNCTION
(1) Put a container on the weigh pan and after the weight is stable, press the TARE
key to zero the weight of the container. The screen displays the “Net” icon.
(2) Put the goods in the container, the screen displays the net weight value of the
goods.
(3) Remove the full container; the screen displays the negative weight value of the
container. At this time pressing the TARE key again will cancel the tare and the
scale reverts back to zero. The “Net” icon is switched off.
The tare function can be operated continually to the full weighing capacity of the scale.
Continual tare operation is adding or removing tare objects on weigh pan and pressing
the TARE key each time.
NET/GROSS FUNCTION
In the Tare mode, the screen displays the “Net” icon, press the NET/GROSS key to
switch between the “Net value” and the “Gross value”.
When the GROSS icon is on, the weight value on the display is the total amount of
the tare value and net value.
At the Gross status, only OFF and NET/GROSS keys are functional.
NET/GROSS key is only used in Tare mode.
EXCELL PRECISION CO., LTD.
02005311
ZSME3000000272
Pcs
.
. .
. . .
. . . .
Pcs
.
. .
. . .
. . . .
SIMPLE COUNTING FUNCTION
(1) Use the UNITS key to enter into the “PCS” mode
(2) Press the NET/GROSS key to select the counting sample size
(S = 10, S = 20, S = 50, S = 100, S=200).
The LCD shows
,
,
,
,
in order.
(3) Put the samples on the weigh pan and press the UNITS key, the screen displays
“------”. After the sampling process is complete, put the goods on the weigh pan
and the screen shows the quantity of the items.
The sample weight should be heavier then the minimum capacity of the scale (20d), If
not the arrow pointing to the icon will be activated.
The weight of a sample should be heavier than the 0.2d (d=division), or the arrow
pointing to the icon will be on.
When the or are indicated, the scale is still operational but the count may
contain errors.
To power off in this mode, the scale will memorize the “Pcs” unit. When the scale is
powered on again, it directly enters the simple counting mode.
While the “Auto unit weight average” function is available in the Advanced Function,
the goods on the weigh pan are 5pcs greater than the sample size and less than
double the sample size, the scale will automatically re-sample the unit weight.
1-5 Self-Test Mode
Set the switch SWA1 on the bottom of machine to the LOCK position.When power is off,
hold NET/GROSS , and press ON/ZERO key, Wait till display shows
to
enter “Self-Test Mode”.
Press UNITS key
Press ON/ZERO key
Press ON/ZERO key
Press ON/ZERO key
Press ON/ZERO key
Press UNITS key
Press UNITS key
Press UNITS key
EXCELL PRECISION CO., LTD.
02005311
ZSME3000000272
1-6 Error Messages
⇒ Unable to read the EEPROM
The EEPROM is not mounted yet, or the circuit near EEPROM is broken.
⇒ Initial zero is higher than the zero range when switching the indicator on.
(Approval models)
⇒ Initial zero is lower than the zero range when switching the indicator on.
(Approval models)
⇒ Internal value is below Zero.
⇒ A/D IC malfunction (cannot read A/D value). The load cell may not be connected to
the scale correctly.
⇒ The weight of the object is over 9 divisions of the maximum capacity.
⇒ A/D IC value is over the maximum range.
1-7 Weight Unit
(kg)
1 g = 0.001 kg
(g)
1 g = 1 g
(lb)
1 g = 0.002204623 lb
(lb/oz) 1 g = 0.03527396 oz
INTERNAL VALUE MODE (must hook up Load Cell to test)
c Press TARE to enter, and the display shows internal value
d Please check whether the internal value has changed obviously with weight
changing.
e Please check the backlight.
f Press ON/ZERO key to back to the last screen , the display shows
KEYPAD TEST MODE
c Press TARE to enter, display shows
Keypad’s internal code: TARE key= 07 UNITS key= 06 NET/GROSS key=05
d Press ON/ZERO key to back to the last screen , the display shows
FIRMWARE VERSION DISPLAY MODE
c Press TARE to enter , display shows the firmware version
d Press TARE key again, the display shows maintenance number
for 2 secs
e Press ON/ZERO key to back to the last screen, display shows
BACK TO THE LAST SCREEN
Press TARE key to exit self-test mode, the scale will restart automatically.
EXCELL PRECISION CO., LTD.
02005311
ZSME3000000272
Chapter 2 Advanced Functions
2-1 Advanced Function Setting Table
Below is an overview of the advanced functions. For detailed settings refer to the following
sections:
DISPLAY
LEVEL 1 FUNCTION
DISPLAY
LEVEL 2 FUNCTIONS
Exit the ADVANCED
FUNCTION setting mode
---
---
Return to theADVANCED
FUNCTION setting menu
Automatic backlight
function setting
Automatic power-off
timer setting
Hi/Lo/OK function setting
Restore the default
settings
Noise filter setting
Hold function setting
General Function setting
mode
Auto unit weight
averaging setting
External Weight
Calibration
---
---
Reserved
Reserved
Reserved
EXCELL PRECISION CO., LTD.
02005311
ZSME3000000272
2-2 Advanced Function Setting Workflow
In the weighing mode, press the NET/GROSS and ON/ZERO keys at the same time to
enter the Advanced Function setting mode. The LCD shows .
Overall workflow of the Advanced Function setting mode:
⇒ General Function setting mode
⇒ External Weight Calibration
⇒ Reserved
⇒ Exit the Advanced Function setting mode
Refer to the following sections for the detailed operation
procedures of each function setting.
Press UNITS key
Press ON/ZERO key
Press ON/ZERO key
Press ON/ZERO key
Press ON/ZERO key
Press UNITS key
Press UNITS key
Press UNITS key
EXCELL PRECISION CO., LTD.
02005311
ZSME3000000272
2-3 General Function Setting
There are 7 functions in the general function setting mode from
to
.
Workflow of the General Function setting:
ACTIONS
DISPLAY
NOTE
⇒ Return to the Advanced Function Setting Mode Menu
⇒ Automatic Backlight Function Settings
⇒ Automatic Power-off Timer Settings
⇒ Hi/Lo/OK Settings
⇒ Restore the Default Settings
⇒ Noise Filter Settings
⇒ Hold Function Settings
⇒ Auto Unit Weight Averaging Setting
Refer to the following sections for detailed operation procedures of each
setting.
Press TARE key
Use the keys to select
the function
Key in “00” and press TARE
key to return to the Advanced
function setting menu
ON/ZERO key ⇒ Upward key (from 0 to 9)
UNITS key ⇒ Downward key (from 9 to 0)
TARE key ⇒ Move the cursor to right
NET/GROSS key ⇒ Move the cursor to left
EXCELL PRECISION CO., LTD.
02005311
ZSME3000000272
2-3-1 Automatic Backlight Function Setting
Select in the General Function setting mode to
change the backlight function setting.
Automatic backlight function
When the weight is over 10d, the display backlight will be on. After the weight
is stable for 10 seconds or when the scale returns to zero, the display
backlight switches off.
DISPLAY
ACTIONS
NOTE
Key in the parameter “00”
Press TARE key
Use ON/ZERO orUNITS
key to select function
“on” or “off
Press TARE key
Press TARE key to
complete the general
setting procedures
LCD displays the last status
EXCELL PRECISION CO., LTD.
02005311
ZSME3000000272
Automatic power-off timer setting
Use ON/ZERO or UNITS key to key in
parameter
0 ⇒ No auto power-off
1 ⇒ When the scale is idle for 1
minute, the scale
automatically switches off
2 ⇒ When the scale is idle for 2
minutes, the scale
automatically switches off
9 ⇒ When the scale is idle for 9
minutes, the scale
automatically switches off
2-3-2 Automatic Power-off Timer Setting
Select
in the General Function setting mode to
change the automatic power-off timer setting.
LCD displays the last status
Automatic power-off function
When the weight on weigh pan is less than 10d or keeps idle for the set time, the
scale will automatically switch off.
Press TARE key
Use ON/ZERO or UNITS
key to key in parameter
Press TARE key
ACTIONS
DISPLAY
NOTE
ON/ZERO key ⇒ Upward key (from 0 to 9)
UNITS key ⇒ Downward key (from 9 to 0)
TARE key ⇒ Move the cursor to right
NET/GROSS key ⇒ Move the cursor to left.
Key in the parameter “00”
Press TARE key
to complete the
general setting
procedures
EXCELL PRECISION CO., LTD.
02005311
ZSME3000000272
lb
lb
lb
lb
2-3-3 Hi/Lo/OK Function Setting
Select
in the General Function setting mode
to
set the Hi/Lo/OK function. This function is available in all unit modes. In one
specific unit mode, enter
to set the Hi/Lo/OK values.
When the high limit and low limit are both set as “0”, the Hi/Lo/OK function is
disabled.
ACTIONS
DISPLAY
NOTE
LCD displays the last status
Press TARE key
Key in the desired high
limit value
Press TARE key
Key in the desired low limit
value
Press TARE key
Key in the parameter
setting for alarm
Press TARE key
LCD displays the last status
LCD displays the last status
.
.
.
.
Alarm setting
(a) (b) (c)
The definition of a, b, c positions:
(a) ⇒ 1 = beep sound on
0 = beep sound off
(b) ⇒ 1 = beep sound on when stable
0 = beep sound on when unstable
(c) ⇒ 1 = beep sound on when the weight is
between high and low limits
0 = beep sound on when the weight is not
between high and low limits and
higher than 10d
ON/ZERO key ⇒ Upward key (from 0 to 9)
UNITS key ⇒ Downward key (from 9 to 0)
TARE key ⇒ Move the cursor to right
NET/GROSS key ⇒ Move the cursor to left.
Key in the parameter
and then press TARE key to go back to the
Advanced function setting menu
EXCELL PRECISION CO., LTD.
02005311
ZSME3000000272
2-3-4 Restore to the Default Setting
Select
in the General Function setting mode
to
restore to the default setting.
The default setting includes the following:
1) External weight calibration
2) HI/LO/OK setting value
3) Noise filter setting (External)
4) Sampling setting for the counting function
For OIML, NTEP or Sri Lanka approved models,
setting is disabled.
ACTIONS
DISPLAY
NOTE
⇒ Return (Cancel the
restoration)
⇒ Restore to the default
setting
Press TARE key
Use ON/ZERO or UNITS
keys to select “return” or
“format”
Press TARE key
ON/ZERO key ⇒ Upward key (from 0 to 9)
UNITS key ⇒ Downward key (from 9 to 0)
TARE key ⇒ Move the cursor to right
NET/GROSS key ⇒ Move the cursor to left.
Key in the parameter “00”
Press TARE key
to complete the
general setting
procedures
EXCELL PRECISION CO., LTD.
02005311
ZSME3000000272
2-3-5 Noise Filter Setting
Select
in the General Function setting mode
to
set the noise filter setting.
Returning to the zero point setting
Use ON/ZERO or UNITS key to key in the
parameters or zero point
Default setting = 0
0 ⇒ No skip 5 ⇒ skip 5d
1 ⇒ skip 1d 6 ⇒ skip 6d
2 ⇒ skip 2d 7 ⇒ skip 7d
3 ⇒ skip 3d 8 ⇒ skip 8d
4 ⇒ skip 4d 9 ⇒ skip 9d
When the weight on the scale is over
1/3 full capacity, the function is on.
Digital switch & Stabilization range
setting
LCD displays the last parameter setting
Filter parameter setting
LCD displays the last parameter setting
Filter parameter setting
Use ON/ZERO or UNITS keys to key
in the parameters. Default setting = 5
Parameter 0 ~ 9, the larger the number,
the faster the filter response. Fast
response can lead to weight instability.
Digital switch & Stabilization range
setting
Use ON/ZERO or UNITS keys to key in
the parameters. Default setting = 0
Parameter 0 ~ 9, the larger the number
the more stable the weight.
ACTIONS
DISPLAY
NOTE
Press TARE key
Use ON/ZERO or UNITS
key to key in the
parameters
Press TARE key
Use ON/ZERO or UNITS
key to key in the
parameters
Press TARE key
Use ON/ZERO or UNITS
key to key in the
parameters
Press TARE key
Returning to zero point setting
LCD displays the last status
Key in the parameter
and then press TARE key to go back to the
Advanced function setting menu
For OIML, NTEP or Sri Lanka approved models,
setting is disabled.
EXCELL PRECISION CO., LTD.
02005311
ZSME3000000272
2-3-6 Hold Function Setting
Select
in the General Function setting mode
to set
the hold function.
Hold function setting
Use ON/ZERO or UNITS keys to key in the
parameters
Default setting = 0
0 ⇒ Hold function disabled
1 ⇒ “Peak hold” mode
2 ⇒ “Stable hold 1” mode
3 ⇒ “Stable hold 2” mode
Hold function setting
LCD displays the last parameter setting
= Hold function disabled
= “Peak hold” mode
Keep displaying the maximum weight when the weight is continually changing
To exit this mode, press any key.
= “Stable hold 1” mode
When the weight is stable, the LCD shows the current weight value. To exit
this mode, press any key.
= “Stable hold 2” mode
When the weight is stable, the LCD shows the current weight value. When the
weight returns to zero (<10d), the hold mode is cancelled automatically.
ACTIONS
DISPLAY
NOTE
ON/ZERO key ⇒ Upward key (from 0 to 9)
UNITS key ⇒ Downward key (from 9 to 0)
TARE key ⇒ Move the cursor to right
NET/GROSS key ⇒ Move the cursor to left.
Press TARE key
Use ON/ZERO or UNITS
key to key in the
parameters
Press TARE key
Key in the parameter “00”
Press TARE key to complete
the general setting procedures
EXCELL PRECISION CO., LTD.
02005311
ZSME3000000272
Press TARE key
Use ON/ZERO or UNITS key
to key in the parameters
Press TARE key
2-3-7 Auto Unit Weight Averaging Setting
Auto unit weight averaging setting
Use ON/ZERO or UNITS keys to key in
the parameters
Default setting = 0
0 ⇒ Auto unit weight averaging
1 ⇒ Non auto unit weight averaging
ON/ZERO key ⇒ Upward key (from 0 to 9)
UNITS key ⇒ Downward key (from 9 to 0)
TARE key ⇒ Move the cursor to right
NET/GROSS key ⇒ Move the cursor to left.
Auto unit weight averaging setting
LCD displays the last parameter setting
Key in the parameter “00”
Press TARE key to
complete the general setting
procedures
EXCELL PRECISION CO., LTD.
02005311
ZSME3000000272
kg
kg
kg
2-4 Weight Calibration
In the weighing mode, press the NET/GROSS and ON/ZERO keys at the same time to
enter the Advanced Function setting mode. The LCD shows
and use
the NET/GROSS or UNITS key to select
to enter the weight calibration
mode.
For approved models,
is disabled.
Weight calibration conditions:
The calibration weight value placed on the weight pan must be over e100, and the
standard deviation of the weight must be within 10%.
The LCD displays the calibration
weight value (the maximum capacity
of 1st unit) and the unit. The digit on
the right side keeps flashing.
The scale is calculating the internal
value of the zero point. After
stabilization, the LCD shows the
calibration weight value.
Press TARE key
When the digit on the right
flashes, ensure that the
weigh pan is empty. Then
Press the TARE
key.
After stabilization
Place the displayed weight
on the weigh pan. Press the
TARE key. The scale
calculates the full capacity
internal value and after
stabilization, the buzzer
beeps 3 times. Remove the
weight from the weigh pan
and the procedure
is completed.
.
.
ACTIONS
DISPLAY
NOTE
ON/ZERO key ⇒ Upward key (from 0 to 9)
UNITS key ⇒ Downward key (from 9 to 0)
TARE key ⇒ Move the cursor to right
NET/GROSS key ⇒ Move the cursor to left.
EXCELL PRECISION CO., LTD.
02005311
ZSME3000000272
Appendix 1 7-Segment Display Characters
Digit
7 segments letter Alphabet
7 segments letter Alphabet 7 segments letter
0
A
N
1
B
O
2
C
P
3
D
Q
4
E
R
5
F
S
6
G
T
7
H
U
8
I
V
9
J
W
K
X
L
Y
M
Z