Omega RD100B Operation & User’s Manual
Read Online
Read the Omega RD100B Operation & User’s Manual online using the built-in PDF viewer.
Text of the manual
For latest product manuals:
omegamanual.info
Shop online at
User’s Guide
Servicing North America:
U.S.A.:
One Omega Drive, P.O. Box 4047
ISO 9001 Certified
Stamford, CT 06907-0047
TEL: (203) 359-1660
FAX: (203) 359-7700
e-mail: [email protected]
Canada:
976 Bergar
Laval (Quebec) H7L 5A1, Canada
TEL: (514) 856-6928
FAX: (514) 856-6886
e-mail: [email protected]
For immediate technical or application assistance:
U.S.A. and Canada: Sales Service: 1-800-826-6342 / 1-800-TC-OMEGA®
Customer Service: 1-800-622-2378 / 1-800-622-BEST®
Engineering Service: 1-800-872-9436 / 1-800-USA-WHEN®
Mexico:
En Espan˜ol: (001) 203-359-7803
e-mail: [email protected]
FAX: (001) 203-359-7807
[email protected]
Servicing Europe:
Benelux:
Postbus 8034, 1180 LA Amstelveen, The Netherlands
TEL: +31 (0)20 3472121
FAX: +31 (0)20 6434643
Toll Free in Benelux: 0800 0993344
e-mail: [email protected]
Czech Republic:
Frystatska 184, 733 01 Karviná, Czech Republic
TEL: +420 (0)59 6311899
FAX: +420 (0)59 6311114
Toll Free: 0800-1-66342
e-mail: [email protected]
France:
11, rue Jacques Cartier, 78280 Guyancourt, France
TEL: +33 (0)1 61 37 2900
FAX: +33 (0)1 30 57 5427
Toll Free in France: 0800 466 342
e-mail: [email protected]
Germany/Austria:
Daimlerstrasse 26, D-75392 Deckenpfronn, Germany
TEL: +49 (0)7056 9398-0
FAX: +49 (0)7056 9398-29
Toll Free in Germany: 0800 639 7678
e-mail: [email protected]
United Kingdom:
One Omega Drive, River Bend Technology Centre
ISO 9002 Certified
Northbank, Irlam, Manchester
M44 5BD United Kingdom
TEL: +44 (0)161 777 6611
FAX: +44 (0)161 777 6622
Toll Free in United Kingdom: 0800-488-488
e-mail: [email protected]
OMEGAnet® Online Service
Internet e-mail
omega.com
[email protected]
It is the policy of OMEGA Engineering, Inc. to comply with all worldwide safety and EMC/EMI
regulations that apply. OMEGA is constantly pursuing certification of its products to the European New
Approach Directives. OMEGA will add the CE mark to every appropriate device upon certification.
The information contained in this document is believed to be correct, but OMEGA accepts no liability for any
errors it contains, and reserves the right to alter specifications without notice.
WARNING: These products are not designed for use in, and should not be used for, human applications.
Thank you for purchasing the OMEGA RD100B/RD1800B Recorder.
This user’s manual describes the functions of the Ethernet interface and the RS-422A/
485 communication interface. To ensure correct use, please read this manual
thoroughly before beginning operation.
After reading the manual, keep it in a convenient location for quick reference whenever a
question arises during operation.
The following five manuals, including this one, are provided as manuals for the RD100B/
RD1800B Recorder. Please read all of them.
The figures used in this manual are mostly of the RD100B. If you are using the
RD1800B, refer to the figures for reference.
• Paper Manual
Manual Title
Manual No.
Description
RD100B Recorder
M-4232
Explains the basic operations of the RD100B
Operation Guide
recorder.
RD1800B Recorder
M-4243
Explains the basic operations of the RD1800B
Operation Guide
recorder.
• Electronic Manuals Provided on the Accompanying CD-ROM
Manual Title
Manual No.
Description
RD100B Recorder
M-4231
Explains all the functions and procedures of the
User’s Manual
RD100B recorder excluding the communication
functions.
RD1800B Recorder
M-4242
Explains all the functions and procedures of the
User’s Manual
RD1800B recorder excluding the
communication
functions.
RD100B/RD1800B
M-4233
This manual. Explains the functions of the
Communication Interface
Ethernet interface and the RS-422A/485
User’s Manual
communication interface.
Notes
•
The contents of this manual are subject to change without prior notice as a result of
continuing improvements to the instrument’s performance and functions. The figures
given in this manual may differ from those that actually appear on your screen.
•
Every effort has been made in the preparation of this manual to ensure the accuracy
of its contents. However, should you have any questions or find any errors, please
contact OMEGA.
•
Copying or reproducing all or any part of the contents of this manual without
permission is strictly prohibited.
•
The TCP/IP software of this product and the document concerning the TCP/IP
software have been developed/created based on the BSD Networking Software,
Release 1 that has been licensed from the University of California.
Revisions
•
1st Edition
December 2004
•
2nd Edition
March 2005
•
3rd Edition
August 2005
Microsoft, MS-DOS, Windows, Windows NT, and Windows XP are either registered
trademarks or trademarks of Microsoft Corporation in the United States and/or other
countries.
•
Adobe, Acrobat, and PostScript are trademarks of Adobe Systems incorporated.
•
For purposes of this manual, the TM and ® symbols do not accompany their
respective trademark names or registered trademark names.
•
Other company and product names are trademarks or registered trademarks of their
respective holders.
Functional Enhancement of the RD100B Recorder
Functions have been added or modified on the RD100B Recorder since system version
1.11. You can check the system version on the system display. For details, see the
RD100B User's Manual (M-4231).
Suffix Code Added or Modified Functions
Reference
Section 4.4: SA command
Section 4.4: TD command
Section 4.5: XN command
Section 4.5: YS command
Section 3.2
Section 2.5
(Changed) Expanded the selectable range of alarm values during linear
scaling (including 1-5V and SQRT) to −5% to 105% of the scale.
(Changed) The procedure to set the start/end date and time of Daylight
Saving Time (DST) has been changed. The TD command
can be used on the µR20000 and the RD100B with system
version 1.11. The SS command can be used on the
RD100B with system version 1.02 or earlier.
(Added)
The print/display format of the date can be changed.
(Changed) Modbus slave protocol can be used.
Two-wire sstem.
(Changed) Users with the same user name cannot be registered.
-
-
-
/C3
/C7
How to Use This Manual
Structure of the Manual
This user’s manual consists of the following sections.
Chapter 1
Overview of the Communication Functions
Gives an overview of the communication functions.
Chapter 2
Using the Ethernet Interface (/C7 Option)
Explains the specifications of the Ethernet interface and how to use the interface.
Chapter 3
Using the RS-422A/485 Communication Interface (/C3 Option)
Explains the specifications of the RS-422A/485 communication interface and how to use the
interface.
Chapter 4
Commands
Explains each command that is available.
Chapter 5
Responses
Explains the responses that the recorder returns and the output format of the setup data and
measured/computed data.
Chapter 6
Status Information
Explains the registers that indicate the recorder statuses.
Appendix
Provides an ASCII character code table, flow charts for outputting data from the recorder,
login procedure, and a list of error messages.
Index
Index of contents.
Conventions Used in This Manual
• Unit
•
k: Denotes 1000. Example: 5 kg, 100 kHz
•
K: Denotes 1024. Example: 640 KB
• Note
The following markings are used in this manual.
Improper handling or use can lead to injury to the user or damage
to the instrument. This symbol appears on the instrument to
indicate that the user must refer to the user’s manual for special
instructions. The same symbol appears in the corresponding place
in the user’s manual to identify those instructions. In the manual,
the symbol is used in conjunction with the word “WARNING” or
“CAUTION.”
WARNING
Calls attention to actions or conditions that could cause serious or
fatal injury to the user, and precautions that can be taken to prevent
such occurrences.
CAUTION
Calls attentions to actions or conditions that could cause light injury
to the user or damage to the instrument or user’s data, and
precautions that can be taken to prevent such occurrences.
Note
Calls attention to information that is important for proper operation
of the instrument.
• Bold Characters
Bold characters are mainly characters and numbers that appear on the display.
• Subheadings
On pages that describe the operating procedures in Chapter 2 and 3, the following
symbols are used to distinguish the procedures from their explanations.
Explanation
This subsection describes the setup parameters and the limitations
on the procedures.
Procedure
Follow the numbered steps. All procedures are written with
inexperienced users in mind; depending on the operation, not all
steps need to be taken.
How to Use This Manual
Names of Parts and Basic Key Operations
Display and Keys
You use the panel keys and the display to configure the communication functions. For a
description of other parts of the recorder, see section 3.1 in the Recorder User’s Manual.
(The figure below is of the RD100B Recorder.)
Display
Front
Rear Panel
Keys
Power
switch
Optional terminal block
RS-422A/485
communication
terminals
Ethernet port
Indicator
Keys
RCD key
Starts/stops recording.
MENU key
Hold this key down for 3 seconds to enter Setting mode. Hold this key down for
3 seconds also to exit from Setting mode.
DISP key
Switches the screen in the main display.
FUNC key
Used when executing manual printout, message printout, etc.
DISP MENU key
Hold this key down for 3 seconds to switch to the data display
setup screen. Hold this key down for 3 seconds also to exit
from the data display setup screen.
FEED key
Feeds the chart paper.
CHARACTER Key: Changes the character type when entering a character. Press
this key while holding down the SHIFT key to switch the character type in reverse
order.
UP/DOWN Key: Switches the setup item or the value.
Press this key while holding down the SHIFT key to switch the setup item
or the value in reverse order.
LEFT/RIGHT Key: Moves the cursor to the right when entering a value
or character. Press this key while holding down the SHIFT key to move
the cursor to the left..
SHIFT Key: Used with the key, key, or the
CHARACTER key.
ENTER Key: Confirms the setup item or value.
CH UP key
Switches the displayed channel.
(when manual switching is specified)
RCD
MENU
FEED
CH UP
FUNC
MENU
DISP
1
CHARACTER
ESC/?
SHIFT
ESC Key: Cancels the operation.
When pressed with the SHIFT key, the display of the comment
on the setting turns ON/OFF.
Names of Parts and Basic Key Operations
Basic Key Operations
This section describes basic operations on the front panel keys to change various settings.
• Execution Modes
The recorder has the following execution modes.
•
Operation mode: A mode used to perform recording and monitoring.
•
Setting mode: A mode used to set the input range, alarms, chart speed, and other
parameters.
•
Basic Setting mode: A mode used to set the basic specifications of functions with
the recording operation stopped.
* In the explanation of commands in (chapter 4), Run mode collectively refers to
Operation mode and Setting mode.
Settings related to communications are configured in Basic Setting mode. You
cannot enter Basic Setting mode while the recorder is recording or while
computation is in progress on the computation function (/M1 option).
• Entering Basic Setting Mode
Hold down the MENU key for 3 seconds.
The Setting mode display appears.
Set=Range
Input range and s
The panel keys are set to the functions marked above the keys as shown below.
RCD
MENU
FEED
CH UP
FUNC
MENU
DISP
1
CHARACTER
ESC/?
SHIFT
Hold down both the
( DISP ) key and the
(
FUNC ) key for 3 seconds.
The Basic Setting mode display appears. The top and bottom lines are the setup item
and comment, respectively. The section that is blinking in the setup item that you
change. In this manual, the section that you change appears shaded.
The comment line shows useful information such as a description of the setup item
and the range of selectable values. Read the comment and change the items as
necessary.
The item to be controlled blinks.
Setup item
Comment
Basic=Alarm
Auxiliary alarm
• Selecting the Setup Item and Value
The selected item change each time you press the
( DISP ) key. The selected
item change in reverse order if you press the
( DISP ) while holding down the
SHIFT ( FEED ) key.
Selections
Basic=Ethernet
This manual denotes the operation of pressing a key while holding down the SHIFT
( FEED ) key as SHIFT + the other key (for example: SHIFT +
key).
After you make a selection, press the
( CH UP ) key. The next screen appears.
When the Setting Complete screen appears, the changed item is applied.
Ethernet host
Setting complete
• Using the ESC Key
If you press the ESC (
1 MENU
) key, the operation is cancelled, and the display returns to
a higher level menu. If you do not show the Setting Complete screen, the changes
you made up to that point are discarded.
You can show and hide the comment on the bottom line by pressing the ESC (
1 MENU
)
key while holding down the SHIFT ( FEED ) key.
• Entering Values
Use the
key or SHIFT +
key to move the cursor. Use the
key or
SHIFT +
key to change a digit value. You repeat these steps to enter the value.
SHIFT key + key
key
Target digit
A=000.000.000.000
SHIFT key + key
key
A=000.000.000.000
When you press the
key, the change is applied and the next setup item is
displayed.
• Entering Characters
Use the
key or SHIFT +
key to move the cursor.
Use the CHARACTER key or SHIFT + CHARACTER key to select the character type.
Use the
key or SHIFT +
key to select a character. You repeat these steps
to set the character string.
Select the character type
CHARACTER key or
SHIFT key + CHARACTER key
Cursor
Host=ABCDefgh2004
CHR:A-Z
The character type changes in the following order: uppercase alphabet, lowercase
alphabet, numbers, and symbols.
key
SHIFT key + key
Host=ABCDefgh2004
CHR:A-Z
When you press the
key, the change is applied and the next screen is displayed.
Inserting Characters
Press the CHARACTER key or SHIFT + CHARACTER key to show Ins DISP and
then press the
key. A space for one character is inserted. Enter the character.
Deleting a Character
Press the CHARACTER key or SHIFT + CHARACTER key to show Del DISP and
then press the
key. The character is deleted.
Deleting an Entire Character String
Press the CHARACTER key or SHIFT + CHARACTER key to show Clear DISP and
then press the
key. The entire character string is deleted.
Copying & Pasting a Character String
Show the copy source character string.
Press the CHARACTER key or SHIFT + CHARACTER key to show Copy DISP and
then press the
key. The character string is saved to the memory.
Show the copy destination.
Press the CHARACTER key or SHIFT + CHARACTER key to show Paste DISP and
then press the
key. The character string is pasted.
Names of Parts and Basic Key Operations
• Exiting from Basic Setting Mode
Press the ESC key several times to return to the Basic= screen.
Press the
key or SHIFT +
key to select End and then press the
key.
The setup save screen appears.
Basic=End
Save Setting
Press the
key to select Store and then press the
key.
The setting is applied and the Operation mode screen appears.
If you select Abort and press the
key, the setting is discarded and the Operation
mode screen appears.
End=Store
Save settings and
Names of Parts and Basic Key Operations
Foreword ......................................................................................................................................... i
Functional Enhancement of the RD100B Recorder ........................................................................ii
How to Use This Manual ................................................................................................................ iii
Names of Parts and Basic Key Operations .................................................................................... v
Chapter 1 Overview of the Communication Functions
1.1
Communication Functions Using the Ethernet Interface (/C7 Option) ............................. 1-1
Functional Construction ................................................................................................... 1-1
Setting/Measurement Server ........................................................................................... 1-1
Maintenance/Test Server ................................................................................................. 1-2
Instrument Information Server ......................................................................................... 1-2
Other Functions ............................................................................................................... 1-3
1.2
Communication Functions Using the RS-422A/485 Communication Interface
(/C3 Option) ..................................................................................................................... 1-4
Functional Construction ................................................................................................... 1-4
Setting/Measurement Server ........................................................................................... 1-4
Modbus Slave .................................................................................................................. 1-4
Chapter 2 Using the Ethernet Interface (/C7 Option)
2.1
Ethernet Interface Specifications ..................................................................................... 2-1
Basic Specifications ......................................................................................................... 2-1
The Maximum Number of Simultaneous Connections and the Number
of Simultaneous Use ........................................................................................................ 2-1
2.2
Connecting the Ethernet Interface ................................................................................... 2-2
When Connecting Only the Recorder and a PC .............................................................. 2-2
When Connecting to a Preexisting Network .................................................................... 2-2
2.3
Configuring the Ethernet Interface ................................................................................... 2-3
2.4
Checking the Connection Status...................................................................................... 2-7
2.5
Registering Users ............................................................................................................ 2-8
2.6
Setting the Communication Timeout and Keepalive ...................................................... 2-11
Chapter 3 Using the RS-422A/485 Communication Interface (/C3 Option)
3.1
RS-422A/485 Communication Interface Specifications ................................................... 3-1
3.2
Terminal Arrangement and Signal Names and the Connection Procedure
of the RS-422A/485 Communication Interface ................................................................ 3-2
Terminal Arrangement and Signal Names ....................................................................... 3-2
Connection Procedure ..................................................................................................... 3-2
Connection Example with a Host Computer .................................................................... 3-3
3.3
The Bit Structure of One Character and the Operation of the Receive Buffer ................. 3-7
The Bit Structure of One Character ................................................................................. 3-7
Receive Buffer and Received Data .................................................................................. 3-7
3.4
Modbus Slave Protocol Specifications ............................................................................. 3-8
Registers .......................................................................................................................... 3-9
Modbus Error Response ................................................................................................ 3-10
3.5
Setting the Serial Interface............................................................................................. 3-11
Chapter 4 Commands
4.1
Command Syntax ............................................................................................................ 4-1
Response ......................................................................................................................... 4-2
4.2
A List of Commands ......................................................................................................... 4-3
Execution Modes and User Levels .................................................................................. 4-3
Setting Commands .......................................................................................................... 4-3
Basic Setting Commands................................................................................................. 4-4
Control Commands .......................................................................................................... 4-5
Output Commands ........................................................................................................... 4-5
RS-422A/485 Dedicated Commands ............................................................................... 4-5
Maintenance/Test Commands ......................................................................................... 4-6
Instrument Information Output Commands ...................................................................... 4-6
4.3
Parameter Values ............................................................................................................ 4-7
Input Range ..................................................................................................................... 4-7
Miscellaneous .................................................................................................................. 4-8
4.4 Setting Commands .................................................................................................................... 4-10
4.5 Basic Setting Commands .......................................................................................................... 4-18
4.6 Control Commands ................................................................................................................... 4-25
4.7 Output Commands .................................................................................................................... 4-26
4.8 RS-422A/485 Dedicated Commands ........................................................................................ 4-28
4.9 Maintenance/Test Commands (Available when using the maintenance/test
server function via Ethernet communications) ............................................................... 4-29
4.10
Instrument Information Output Commands (Available when using the instrument
information server function via Ethernet communications) ............................................ 4-30
Chapter 5 Responses
5.1
Response Syntax ............................................................................................................. 5-1
Affirmative Response ....................................................................................................... 5-1
Single Negative Response............................................................................................... 5-1
Multiple Negative Responses .......................................................................................... 5-1
ASCII Output .................................................................................................................... 5-2
BINARY Output ................................................................................................................ 5-2
RS-422A/485 Dedicated Commands and Responses ..................................................... 5-5
5.2
Output Format of ASCII Data ........................................................................................... 5-6
Setting/Basic Setting data ................................................................................................ 5-6
Decimal Point Position/Unit Information........................................................................... 5-7
Measured/computed Data................................................................................................ 5-8
Report Data Generated by the Periodic Printout ........................................................... 5-10
Status Information .......................................................................................................... 5-12
User Information ............................................................................................................ 5-12
5.3
Output Format of BINARY Data ..................................................................................... 5-13
Measured/Computed Data and FIFO Data .................................................................... 5-13
Chapter 6 Status Information
6.1
Status Information and Filter ............................................................................................ 6-1
6.2
The Bit Structure of the Status Information ...................................................................... 6-2
Status Information 1 ......................................................................................................... 6-2
Status Information 2 ......................................................................................................... 6-2
Status Information 3 ......................................................................................................... 6-2
Status Information 4 ......................................................................................................... 6-2
Contents
Appendix 1 ASCII Character Codes ........................................................................................App-1
Appendix 2 Output Flow of FIFO Data ....................................................................................App-2
Appendix 3 Login Procedure ...................................................................................................App-4
Appendix 4 A List of Error Messages ......................................................................................App-6
Setting Errors ...............................................................................................................App-6
Operation Errors ..........................................................................................................App-7
Communication Errors .................................................................................................App-7
Warning Messages ......................................................................................................App-8
System Errors ..............................................................................................................App-8
Index
Overview of the Communication Functions
1.1
Communication Functions Using the Ethernet
Interface (/C7 Option)
The recorder can be equipped with an optional Ethernet interface. For details on how to
use the Ethernet interface, see chapter 2.
Functional Construction
The following figure shows the relationship between the communication function of the
recorder and the Ethernet interface. Perform communication according to the respective
protocol.
* Protocol is a set of rules that two computers use to communicate via a communication line (or
network).
PC
Connect the recorder and the
PC using an Ethernet cable.
Maintenance/
Test Server
Setting/
Measurement Server
Instrument
Information Server
Login (user authentication/access
privileges granting)
Upper layer
protocol
Dedicated protocol
Lower layer
protocol
TCP
IP
Interface
Ethernet (10BASE-T)
UDP
Communication functions of the recorder
Application
TCP (Transmission Control Protocol)
UDP (User Datagram Protocol)
IP (Internet Protocol)
Setting/Measurement Server
•
You can specify settings that are approximately equivalent to those specified by front
panel key operations. However, you cannot turn the power ON/OFF, set the user
name and password for communications, nor set the key lock.
•
The data below can be output.
Data Type
Output Format
Measured/computed data
BINARY/ASCII
Setup data
ASCII
Periodic printout and the most recent TLOG computation data
ASCII
Status information
ASCII
Information on connected users
ASCII
•
The commands that can be used are Setting commands, Basic Setting commands,
Control commands, and Output commands.
•
Ethernet interface settings: Section 2.3
•
Commands: Section 4.2
•
Data output format: Chapter 5
Chapter 1 Overview of the Communication Functions
Outputs Ethernet communication information such as connection information and
network statistics from the recorder.
•
The commands that can be used Maintenance/Test commands.
•
Ethernet interface settings: Section 2.3
•
Commands: Section 4.2
Instrument Information Server
•
Outputs the serial number, model name, and other information about the recorder
connected via the Ethernet network.
•
The commands that can be used Instrument Information Output commands.
•
Ethernet interface settings: Section 2.3
•
Commands: Section 4.2
1.1 Communication Functions Using the Ethernet Interface (/C7 Option)
Overview of the Communication Functions
Other Functions
• Login Function
Only users that are registered in advance can access the Setting/Measurement and
Maintenance/Test servers.
•
Users are identified by their user name and password.
•
You can register one administrator and six users.
Administrator privileges
The administrator can use all the functions on the Setting/Measurement and
Maintenance/Test servers
User Privileges
•
Setting/Measurement server
Users can output measured data, setup data, scheduled printing, and the most
recent TLOG computation data. Users cannot control the recorder.
•
Maintenance/Test server
Users cannot disconnect communications between the recorder and other PCs.
All other operations are allowed.
•
There is a maximum number of simultaneous connections that can be established
with the recorder.
•
Login function settings: Section 2.5
•
Maximum number of simultaneous connections: Section 2.1
•
Commands available to the administrator and users: Section 4.2
• Communication Timeout
This function drops the connection with the PC if there is no data transmission for a
given time at the application level (see “Functional Construction”). For example, this
function prevents a PC from being connected to the recorder indefinitely which would
prohibit other users from making new connections for data transfer.
•
Communication timeout setting: Section 2.6
• Keepalive
This function drops the connection if there is no response to the inspection packet that
is periodically transmitted at the TCP level.
•
Keepalive setting: Section 2.6
1.1 Communication Functions Using the Ethernet Interface (/C7 Option)
Communication Functions Using the RS-422A/
485 Communication Interface (/C3 Option)
The recorder can be equipped with an optional RS-422A/485 communication interface.
For details on how to use the RS-422A/485 communication interface, see chapter 3.
Functional Construction
The following figure shows the relationship between the communication function of the
recorder and the RS-422A/485 communication interface. Perform communication
according to the respective protocol.
* Protocol is a set of rules that two computers use to communicate via a communication line (or
network).
PC
Connect the recorder and the
PC using a serial cable.
Setting/Measurement Server
Dedicated protocol
Protocol
Interface
RS-422A/485
Modbus Slave
Modbus protocol
Communication functions of the recorder
Application
Setting/Measurement Server
The functions are the same as those of the Setting/Measurement server of the Ethernet
interface. See page 1-1.
•
RS-422A/485 communication interface settings: Section 3.5
•
Commands: Section 4.2
•
RS-422A/485 dedicated commands: Section 4.8
•
Data output format: Chapter 5
Modbus Slave
•
The Modbus protocol can be used to read the measured/computed data on your PC
by reading the input registers of the recorder. The communication input data can be
written or read by writing/reading the hold register of the recorder.
•
For details on the Modbus function codes that the recorder supports, see section 3.4.
•
This function can be used only when communicating via the serial interface (option).
•
For a description on the settings required in using this function, see section 3.5.
Data from slave device
Data to slave device
Modbus master device
Modbus slave device
Serial communication
Recorder
Using the Ethernet Interface (/C7 Option)
2.1
Ethernet Interface Specifications
Basic Specifications
Item
Specifications
Electrical and mechanical specifications
Conforms to IEEE 802.3
(Ethernet frames are of DIX specification)
Transmission medium type
10BASE-T
Protocol
TCP, IP, UDP, ICMP, and ARP
The Maximum Number of Simultaneous Connections and the Number of Simultaneous
Use
The following table shows the maximum number of simultaneous connections, the
number of simultaneous users, and the port numbers of the recorder.
Function
Maximum Number
Number of Simultaneous Users
Port Number†
of Connections
Setting/Measurement 3
1
2††
34260/tcp
server
Maintenance/Test
1
1
1††
34261/tcp
server
Instrument
-
-
-
34264/udp
Information server
†
The port numbers are fixed.
†† For details on administrator and user privileges, see “Login Function” in section 1.1.
Chapter 2 Using the Ethernet Interface (/C7 Option)
Connecting the Ethernet Interface
When Connecting Only the Recorder and a PC
Connect the recorder and the PC via a HUB as in the following figure.
Hub
RD100B/RD1800B Recorder
RD100B/RD1800B Recorder
PC
10BASE-T
straight cable
10BASE-T
straight cable
When Connecting to a Preexisting Network
The following figure illustrates an example in which a recorder and a PC are connected
to the network. When connecting the recorder or the PC to a preexisting network, the
transfer rate, connector type, etc. must be matched. For details, consult your system or
network administrator.
Hub, router, etc.
Network
RD100B/RD1800B Recorder
PC
10BASE-T
straight cable
Note
• Depending on the reliability of the network or the volume of network traffic, all the transferred
data may not be retrieved by the PC.
• Communication performance deteriorates if multiple PCs access the recorder simultaneously.
Using the Ethernet Interface (/C7 Option)
2.3
Configuring the Ethernet Interface
Set the host name and IP address of the recorder. You do not have to set the DNS
(domain name system).
Setup Items
DNS
Ethernet
End
Local IP
A
M
G
IP address
Subnet mask
Host name
Domain name
Default gateway
DNS
P
S
DNS On/Off
Server
(primary)
Server
(secondary)
Host
Domain
Suffix_P
Suffix_S
Domain suffix
(primary)
Domain suffix
(secondary)
Host
Procedure
For a description of the basic setup operations, see “Basic Key Operations” on page v.
Entering Basic Setting Mode
Hold down the MENU key for 3 seconds to display the Setting mode screen.
Next, hold down both the
( DISP ) key and the
(
FUNC ) key for 3 seconds to
display the Basic Setting mode screen.
Note
To cancel an operation, press the ESC key.
Host Name and Domain Name
1.
Press the
key to select Ethernet and then press the
key.
Basic=Ethernet
2.
Press the
key to select Host and then press the
key.
Ethernet=Host
3.
Set the host name of the recorder and then press the
key.
Key operations
•
Use the
key to select the digit for entering a character.
•
Use the CHARACTER key to select the character type.
•
Use the
key to select the character you wish to enter.
Host=
4.
Set the domain name and press the
key in the same fashion as in step 3.
The setting complete screen appears.
Domain=
Ethernet host
Setting complete
5.
Press the ESC key to return to the higher level menu. To save the settings and
exit from Basic Setting mode, proceed to “Saving the Settings.”
* When the key, key, or key is pressed while holding down the SHIFT key,
the operation is reversed as when the respective key is pressed by itself.
CHARACTER
IP Address, Subnet Mask, and Default Gateway
1.
Press the
key to select Ethernet and then press the
key.
Basic=Ethernet
2.
Press the
key to select Local IP and then press the
key.
Ethernet=Local IP
3.
Set the IP address of the recorder and then press the
key.
Key operations
•
Use the
key to select the digit for entering a value.
•
Use the
key to select the value you wish to enter.
A=
4.
Set the IP address of the subnet mask and then press the
key in the same
fashion as in step 3.
M=
5.
Set the IP address of the default gateway and then press the
key in the
same fashion as in step 3.
The local IP setting complete screen appears.
G=
Ethernet address
Setting complete
6.
Press the ESC key to return to the higher level menu. To save the settings and
exit from Basic Setting mode, proceed to “Saving the Settings.”
DNS (Domain Name System)
1.
Press the
key to select Ethernet and then press the
key.
Basic=Ethernet
2.
Press the
key to select DNS and then press the
key.
Ethernet=DNS
3.
Press the
key to select On and then press the
key.
DNS=On
4.
Set the IP address of the primary DNS server and then press the
key.
Key operations
•
Use the
key to select the digit for entering a value.
•
Use the
key to select the value you wish to enter.
P=
5.
Set the IP address of the secondary DNS server and then press the
key in
the same fashion as in step 4.
S=
2.3 Configuring the Ethernet Interface
* When the key, key, or key is pressed while holding down the SHIFT key,
the operation is reversed as when the respective key is pressed by itself.
CHARACTER
Using the Ethernet Interface (/C7 Option)
6.
Set the primary domain suffix and then press the
key.
Key operations
•
Use the
key to select the digit for entering a character.
•
Use the CHARACTER key to select the character type.
•
Use the
key to select the character you wish to enter.
Suffix_P=
7.
Set the secondary domain suffix and then press the
key in the same fashion
as in step 6.
The DNS setting complete screen appears.
Suffix_S=
Ethernet DNS
Setting complete
8.
Press the ESC key to return to the higher level menu. To save the settings and
exit from Basic Setting mode, proceed to “Saving the Settings.”
Saving the Settings
1.
Press the ESC key to return to the Basic= screen.
2.
Press the
key to select End and then press the
key.
3.
Press the
key to select Store and then press the
key.
The settings are activated, and the Operation mode screen appears.
Explanation
For details on the settings, consult your system or network administrator.
• Host Name
Set the recorder’s host name and the domain name of the network to which the
recorder belongs. Be sure to set these items when using the DNS.
• Host
Set the recorder’s host name using up to 64 alphanumeric characters.
• Domain
Set the network domain name to which the recorder belongs using up to 64
alphanumeric characters.
• IP Address, Subnet Mask, and Default Gateway
• IP address
•
Set the IP address to assign to the recorder. The default value is 0.0.0.0.
•
The IP address is used to distinguish between the various devices connected to
the Internet when communicating using the TCP/IP protocol. The address is a
32-bit value normally expressed with four values (0 to 255), each separated by a
period as in 192.168.111.24.
• M (Subnet Mask)
•
Specify the mask that is used to determine the network address from the IP
address. The default value is 0.0.0.0.
•
Set this value according to the system or the network to which the recorder
belongs. In some cases, this setting may not be necessary.
2.3 Configuring the Ethernet Interface
* When the key, key, or key is pressed while holding down the SHIFT key,
the operation is reversed as when the respective key is pressed by itself.
CHARACTER
• G (Default Gateway)
•
Set the IP address of the gateway (router, etc.) used to communicate with other
networks. The default value is 0.0.0.0.
•
Set this value according to the system or the network to which the recorder
belongs. In some cases, this setting may not be necessary.
• Setting the DNS (Domain Name System)
The DNS is a system that correlates the host name/domain name to the IP address.
The host name/domain name can be used instead of the IP address when accessing
the network. The DNS server manages the database that contains the host name/
domain name and IP address correlation.
• On/Off
Select On when using the DNS.
• P (Primary DNS Server)
Set the IP address of the primary DNS server. The default value is 0.0.0.0.
• S (Secondary DNS Server)
Set the IP address of the secondary DNS server. The default value is 0.0.0.0. If
the primary DNS server is down, the secondary server is used to search the host
name and IP address correlation.
• Suffix_P (Primary Domain Suffix), Suffix_S (Secondary Domain Suffix)
When the recorder searches another server using the DNS server, the domain
name of the recorder is appended to the host name as a possible domain name if it
is omitted. If the IP address corresponding to the server name is not found on the
DNS server, then it may be that the system is configured to use another domain
name for searching. This alternate domain name is specified as the domain suffix.
•
Set the domain suffix using up to 64 alphanumeric characters.
•
Up to two domain suffixes can be specified (primary and secondary).
• Saving the Settings
To activate the settings made in the Basic Setting mode, the settings must be saved.
2.3 Configuring the Ethernet Interface
Using the Ethernet Interface (/C7 Option)
2.4
Checking the Connection Status
The connection status of the Ethernet interface can be confirmed with the indicator that
is located to the left of the Ethernet port on the recorder.
Indicator
Connection Status of the Ethernet Interface
ON (Green)
The Ethernet interface is electrically connected.
Blinking (Green)
Transmitting data.
OFF
The Ethernet interface is not electrically connected.
Optional terminal block
Ethernet port
Indicator
Users that can access the recorder via the Ethernet network must be registered. This
function is called login function.
Setup Items
LoginSet
Ethernet
End
Level
Register
User
Admin/User
On/Off
Use/Not
User name
Login
Password
Login
Procedure
For a description of the basic operations, see “Basic Key Operations” on page v.
Entering Basic Setting Mode
Hold down the MENU key for 3 seconds to display the Setting mode screen.
Next, hold down both the
( DISP ) key and the
(
FUNC ) key for 3 seconds to
display the Basic Setting mode screen.
Note
To cancel an operation, press the ESC key.
Enabling/Disabling the Login Function
1.
Press the
key to select Ethernet and then press the
key.
Basic=Ethernet
2.
Press the
key to select Login and then press the
key.
Ethernet=Login
3.
Press the
key to select Use and then press the
key.
The setting complete screen appears.
Login=On
4
Press the ESC key to return to the Ethernet menu.
Registering Users
5.
Press the
key to select LoginSet and then press the
key.
Ethernet=LoginSet
6.
Press the
key to select Admin (administrator) or User1 to User6, and then
press the
key.
Level=Admin
7.
Press the
key to select On and then press the
key.
Register=On
* When the key, key, or key is pressed while holding down the SHIFT key,
the operation is reversed as when the respective key is pressed by itself.
CHARACTER
Using the Ethernet Interface (/C7 Option)
8.
Set the user name and then press the
key.
Key operations
•
Use the
key to select the digit for entering a character.
•
Use the CHARACTER key to select the character type.
•
Use the
key to select the character you wish to enter.
User=
9.
Set the password and then press the
key in the same fashion as in step 8.
The setting complete screen appears.
Password=
Ethernet login
Setting complete
To register other users, press the
key to return to step 6 and repeat steps 8,
8, and 9.
Saving the Settings
1.
Press the ESC key to return to the Basic= screen.
2.
Press the
key to select End and then press the
key.
3.
Press the
key to select Store and then press the
key.
The settings are activated, and the Operation mode screen appears.
Explanation
You can limit the users that can access the Setting/Measurement and Maintenance/Test
servers on the recorder via the Ethernet interface.
• Enabling/Disabling the Login Function
Set whether to use the login function.
• Registering Users
• User level
Select either of the user levels, administrator or user.
• Administrator
One administrator can be registered. An administrator has the authority to use all
Setting/Measurement server and Maintenance/Test server commands.
• User
Six users can be registered. A user has limited authority to use the commands.
See section 4.2.
• Selecting Whether to Register (On/Off) the User
If On is selected, set the user name and password.
• Setting the User Name
•
Set the user name using up to 16 alphanumeric characters.
•
The same user name can not be registered.
•
Since the word “quit” is reserved as a command on the recorder, the user name
“quit” is not allowed.
• Setting the Password
Set the password using up to 4 alphanumeric characters and spaces.
2.5 Registering Users
* When the key, key, or key is pressed while holding down the SHIFT key,
the operation is reversed as when the respective key is pressed by itself.
CHARACTER
• The relationship between the login function and the user name that is used when accessing
the recorder is as follows:
• When the login function is set to “Use”
• The registered user name and password can be used to login to the recorder.
• The user level is the level that was specified when the user name was registered.
• When the login function is set to “Not”
• The user name “admin” can be used to login to the recorder as an administrator.
Password is not necessary.
• The user name “user” can be used to access the recorder as a user. Password is not
necessary.
• There are limitations on the number of simultaneous connections or simultaneous uses of the
recorder from the PC (see section 2.1).
• For a description of the login process of the Setting/Measurement server and Maintenance/
Test server, see appendix 3.
• Saving the Settings
To activate the settings made in the Basic Setting mode, the settings must be saved.
2.5 Registering Users
Using the Ethernet Interface (/C7 Option)
2.6
Setting the Communication Timeout and
Keepalive
The communication timeout function and the keepalive function can be configured.
Setup Items
K.Alive
Ethernet
End
Keep alive
On/Off
On/Off
Timeout
Timeout time
Duration
Timeout
Procedure
For a description of the basic operations, see “Basic Key Operations” on page v.
Entering Basic Setting Mode
Hold down the MENU key for 3 seconds to display the Setting mode screen.
Next, hold down both the
( DISP ) key and the
(
FUNC ) key for 3 seconds to
display the Basic Setting mode screen.
Note
To cancel an operation, press the ESC key.
Communication Timeout
1.
Press the
key to select Ethernet and then press the
key.
Basic=Ethernet
2.
Press the
key to select Timeout and then press the
key.
Ethernet=Timeout
3.
Press the
key to select On and then press the
key.
Timeout=On
4.
Set the timeout time and then press the
key.
Key operations
•
Use the
key to select the digit for entering a value.
•
Use the
key to select the value you wish to enter.
Duration=
Ethernet timeout
Setting complete
5
Press the ESC key to return to the higher level menu. To save the settings and
exit from Basic Setting mode, proceed to “Saving the Settings.”
* When the key, key, or key is pressed while holding down the SHIFT key,
the operation is reversed as when the respective key is pressed by itself.
CHARACTER
key to select Ethernet and then press the
key.
Basic=Ethernet
2.
Press the
key to select K.Alive and then press the
key.
Ethernet=K. Alive
3.
Press the
key to select On and then press the
key.
The setting complete screen appears.
Keep alive=On
Keep alive
Setting complete
5
Press the ESC key to return to the higher level menu. To save the settings and
exit from Basic Setting mode, proceed to “Saving the Settings.”
Saving the Settings
1.
Press the ESC key to return to the Basic= screen.
2.
Press the
key to select End and then press the
key.
3.
Press the
key to select Store and then press the
key.
The settings are activated, and the Operation mode screen appears.
Explanation
The communication timeout function and the keepalive function can be configured.
• Communication Timeout
• Selecting On or Off
If On is selected, set the timeout time.
• Timeout Time
If communication timeout is enabled, the connection is dropped if no data transfer
is detected over a time period specified here.
Selectable range: 1 to 120 minutes
• Enabling (On)/Disabling (Off) Keepalive
Select On to enable the keepalive function.
• Saving the Settings
To activate the settings made in the Basic Setting mode, the settings must be saved.
Keepalive: Section 1.1
2.6 Setting the Communication Timeout and Keepalive
* When the key, key, or key is pressed while holding down the SHIFT key,
the operation is reversed as when the respective key is pressed by itself.
CHARACTER
Using the RS-422A/485 Communication Interface (/C3 Option)
3.1
RS-422A/485 Communication Interface
Specifications
This section describes the RS-422A/485 communication interface specifications.
Item
Specifications
Terminal block type
Number of terminals: 6, terminal attachment screws: ISO
M4/nominal length of 6 mm
Electrical and mechanical
Complies with the EIA-422A(RS-422A) and
specifications
EIA-485(RS-485) standards
Connection
Multidrop Four-wire system
1:32
Two-wire system
1:31 (Modbus slave protocol)
Transmission mode
Half-duplex
Synchronization
Start-stop synchronization
Baud rate
Select from 1200, 2400, 4800, 9600, 19200, and 38400 [bps].
Start bit
Fixed to 1 bit
Data length
Select 7 or 8 bits
Parity
Select Odd, Even, or None (no parity).
Stop bit
Fixed to 1 bit
Received buffer length
2047 bytes
Escape sequence
Open and close
Electrical characteristics
6 points consisting of FG, SG, SDB, SDA, RDB, and RDA
The SG, SDB, SDA, RDB, and RDA terminals and the
internal circuitry of the recorder are functionally isolated.
The FG terminal is the frame ground.
Communication distance
Up to 1.2 km
Terminal resistance
120 Ω, 1/2 W
Chapter 3 Using the RS-422A/485 Communication Interface (/C3 Option)
Terminal Arrangement and Signal Names and
the Connection Procedure of the RS-422A/485
Communication Interface
Terminal Arrangement and Signal Names
L
N
ALARM
CHART
REMOTE
1
2
3
RS-422A
SD SD SG
A
B
G
RD RD FG
A
B
G
4
5
C
NO
C
NC
NO
C
NC
NO
C
NC
NO
C
NC
NO
C
NC
NO
C
NC
RS-422A
SD SD SG
A
B
RD RD FG
A
B
SDA
SDB
SG
RDA RDB
FG
Rear panel
Terminal Name
Description
FG (Frame Ground)
Case ground of the recorder.
SG (Signal Ground)
Signal ground.
SDB (Send Data B)
Send data B (+).
SDA (Send Data A)
Send data A (–).
RDB (Received Data B)
Receive data B (+).
RDA (Received Data A)
Receive data A (–).
Connection Procedure
• Cable
Use the cable that meets the conditions below.
Item
Conditions
Cable
Shielded twisted pair cable
3 pairs AWG24-14 (Four-wire system),
2 pairs 24 AWG or more (Two-wire system)
Characteristic impedance
100 Ω
Capacitance
50 pF/m
Cable length
Up to 1.2 km*
*
The transmission distance of the RS-422A/485 interface is not the straight-line distance, but
rather the total length of the (shielded twisted-pair) cable.
• Connecting the Cable
As shown in the following figure, attach a crimp-on lug with isolation sleeves for 4 mm
screws to the end of the cable. Keep the exposed section from the end of the shield
within 5 cm.
SDA SDB
SG
RDA RDB
FG
Electric
potential of
the shield
Shield
Four-wire system
Two-wire system
SDA SDB
SG
RDA RDB
FG
Electric
potential of
the shield
Shield
Using the RS-422A/485 Communication Interface (/C3 Option)
WARNING
To prevent the possibility of electric shock, connect the cables with the power
turned OFF.
Note
•
Connect the RD pin to the SD (TD) pin on the PC (converter) end and the SD pin to the RD
pin on the PC end.
•
The two-wire system can be used only when using the Modbus protocol.
Connection Example with a Host Computer
A connection can be made with a host computer having a RS-232, RS-422A, or RS-485
port.
•
In the case of RS-232, a converter is used.
•
For recommended converters, see “Serial Interface Converter” on the next page.
•
The two-wire system can be used only when using the Modbus protocol. For the
configuration procedure, see section 3.5
• Four-Wire System
Generally, a four-wire system is used to connect to a host computer. In the case of a
four-wire system, the transmission and reception lines need to be crossed over.
Terminator (external) 120 Ω 1/2W or greater
#1
Do not connect terminators to #1 through #n-1.
RS-422A/485
terminal on the
recorder
#2
#n
(#n ≤ 32)
Terminator (external)
Host
computer
SG
RDB( + )
RDA( - )
SDB( + )
SDA( - )
FG
SG
RD B
RD A
SD B
SD A
(SG)
(RD B)
(RD A)
(SDB)
(SDA)
FG
SG
RD B
RD A
SD B
SD A
(SG)
(RD B)
(RD A)
(SDB)
(SDA)
FG
SG
RD B
RD A
SD B
SD A
(SG)
(RD B)
(RD A)
(SDB)
(SDA)
3.2 Terminal Arrangement and Signal Names and the Connection Procedure of the RS-422A/485 Communication Interface
(The following figure illustrates the case when the host computer’s interface is RS-232.)
Terminator (external) 120 Ω 1/2W or greater
#1
Do not connect terminators to #1 through #n-1.
RS-422A/485
terminal on the
recorder
#2
#n
(#n ≤ 32)
Terminator (external)
SHIELD
RD( + )
RD( - )
TD( + )
TD( - )
FG
SG
RD B
RD A
SD B
SD A
Host
computer
Converter
(SG)
(RD B)
(RD A)
(SDB)
(SDA)
FG
SG
RD B
RD A
SD B
SD A
(SG)
(RD B)
(RD A)
(SDB)
(SDA)
FG
SG
RD B
RD A
SD B
SD A
(SG)
(RD B)
(RD A)
(SDB)
(SDA)
RS-232
• Two-Wire System
Connect the transmission and reception signals with the same polarity on the RS-
422A/485 terminal block. The two-wire system can be used only when using the
Modbus protocol.
Terminator (externally attached) 120 Ω, 1/2 W or more
#1
Do not connect terminator to #1 to #n–1
#2
#n
(#n ≤ 31)
Terminator (externally attached)
Host
computer
SG
RDB( + )
RDA( – )
SDB( + )
SDA( – )
FG
SG
RD B
RD A
SD B
SD A
(SG)
(B)
(A)
(B)
(A)
FG
SG
RD B
RD A
SD B
SD A
(SG)
FG
SG
RD B
RD A
SD B
SD A
(SG)
(B)
(A)
RS-422A/485
terminal on the
recorder
3.2 Terminal Arrangement and Signal Names and the Connection Procedure of the RS-422A/485 Communication Interface
Using the RS-422A/485 Communication Interface (/C3 Option)
(The following figure illustrates the case when the host computer’s interface is RS-232.)
Terminator (externally attached) 120 Ω, 1/2 W or more
#1
Do not connect terminator to #1 to #n–1
#2
#n
(#n ≤ 31)
Terminator (externally attached)
FG
SG
RD B
RD A
SD B
SD A
(SG)
(B)
(A)
(B)
(A)
FG
SG
RD B
RD A
SD B
SD A
(SG)
FG
SG
RD B
RD A
SD B
SD A
(SG)
(B)
(A)
SHIELD
TD( + )
TD( – )
RD( + )
RD( – )
Host
computer
Converter
RS-232
RS-422A/485
terminal on the
Recorder
Note
•
The method used to eliminate noise varies depending on the situation. In the connection
example, the shield of the cable is connected only to the recorder’s ground (one-sided
grounding). This is effective when there is a difference in the electric potential between the
computer’s ground and the recorder’s ground. This may be the case for long distance
communications. If there is no difference in the electric potential between the computer’s
ground and the recorder’s ground, the method of connecting the shield also to the computer’s
ground may be effective (two-sided grounding). In addition, in some cases, using two-sided
grounding with a capacitor connected in series on one side is effective. Consider these
possibilities to eliminate noise.
•
When using the two-wire type interface (Modbus protocol), the 485 driver must be set to high
impedance within 3.5 characters after the last data byte is sent by the host computer.
• Serial Interface Converter
CAUTION
Some converters have FG and SG pins that are not isolated. In this case, do
not connect anything to the converter’s FG and SG pins (unlike the figure on the
previous page). Especially in the case of long distance communications, the
potential difference that appears may damage the recorder or cause
communication errors. For converters that do not have the SG pin, they can be
used without using the signal ground. For details, see the manual that came
with the converter.
On some non-recommended converters, the signal polarity may be reversed (A/B or
+/- indication). In this case, reverse the connection.
3.2 Terminal Arrangement and Signal Names and the Connection Procedure of the RS-422A/485 Communication Interface
For a two-wire system, the host computer must control the transmission driver of the
converter in order to prevent collisions of transmit and received data. When using the
recommended converter, the driver is controlled using the RS (RTS) signal on the RS-
232.
• When Instruments That Support Only the RS-422A Interface Exist in the System
When using the four-wire system, up to 32 recorders can be connected to a single
host computer. However, this may not be true if instruments that support only the RS-
422A interface exist in the system.
When the instrument that support only the RS-422A interface exist in the
system
The maximum number of connection is 16. Some of OMEGA’s conventional recorder
only support the RS-422A driver. In this case, only up to 16 units can be connected.
Note
In the RS-422A standard, 10 is the maximum number of connections that are allowed on one
port (for a four-wire system).
• Terminator
When using a multidrop connection (including a point-to-point connection), connect a
terminator only to the recorder on the end of the chain. In addition, turn the terminator
on the host computer ON (see the computer’s manual). If a converter is being used,
turn ON its terminator. The terminator must be attached externally to the
recommended converters.
3.2 Terminal Arrangement and Signal Names and the Connection Procedure of the RS-422A/485 Communication Interface
Using the RS-422A/485 Communication Interface (/C3 Option)
3.3
The Bit Structure of One Character and the
Operation of the Receive Buffer
The Bit Structure of One Character
The serial interface on the recorder communicates using start-stop synchronization. In
start-stop synchronization, a start bit is added every time a character is transmitted. The
start bit is followed by the data bits, parity bit, and stop bit. (See the figure below.)
Data bit
(7 or 8 bits)
1 character
Stop bit
Parity bit
Even, Odd,
or None
Start bit
Circuit idle
state
Return to the idle
state (dotted line)
or the start bit of
the next data
character
(solid line)
Receive Buffer and Received Data
The data received from the PC is first placed in the receive buffer of the recorder. When
the received buffer becomes full, all of the data that overflow are discarded.
Modbus Slave Protocol Specifications
The Modbus slave protocol specifications of the recorder are as follows:
Item
Specifications
Transmission medium
RS-422A/485
Flow control
None only
Baud rate
Select from 1200, 2400, 4800, 9600, 19200, or 38400 [bps]
Start bit
Fixed to 1 bit
Stop bit
Fixed to 1 bit
Parity check
Select odd, even, or none (no parity).
Transmission mode
RTU (Remote Terminal Unit) mode only
• Data length: 8 bits
• Data interval: 24 bits or less*
• Error detection: Uses CRC-16
* Determines message termination with a time interval to 3.5
characters or more.
Maximum number of
Four-wire system: 32 slave devices
connected units
Two-wire system: 31 slave devices
The function codes of the Modbus slave protocol that the recorder supports are shown
below. The recorder does not support broadcast commands.
Function Code
Specifications
Operation
3
Read the hold register (4xxxx).
The master device can read
the communication input data
written using function codes 6
and 16.
4
Read the input register (3xxxx).
The master device loads the
computed, measured, alarm,
and time data of the recordr.
6
Single write to hold register (4xxxx)
The master device writes to
the communication input data
of the recorder.
8
Loopback test
The master device performs a
loopback test of the recorder.
The recorder only supports
message return (test code
0x00*)
16
Write to the hold register (4xxxx)
The master device writes to
the communication input data
of the recorder.
*
Hexadecimal ”00”
Using the RS-422A/485 Communication Interface (/C3 Option)
3.4 Modbus Slave Specifications
Registers
The registers for using the Modbus slave protocol are listed below.
The register data does not contain unit and decimal point position information. Set them
on the Modbus master. Binary values are stored to the register in order from the highest
byte.
Input register
Data
30001
Measured data of CH01
:
:
30024
Measured data of CH24
• The data is a 16-bit signed integer. The value is the same as the measured data in binary
output format (see page 5-13).
• The decimal point and unit information varies depending on the range setting of each
channel.
• Valid channels vary depending on the device. An error response (code 2) occurs when an
invalid channel is read.
31001
Alarm status of the measured data of CH01
:
:
31024
Alarm status of the measured data of CH24
• The data is a 16-bit signed integer. The value is the same as the alarm status in binary
output format (see page 5-13). The data is entered in the "A2A1A4A3" order in the register.
• Valid channels vary depending on the device. An error response (code 2) occurs when an
invalid channel is read.
32001
Computed data of CH0A (lower word)
32002
Computed data of CH0A (upper word)
32003
Computed data of CH0B (lower word)
32004
Computed data of CH0B (upper word)
:
:
32048
Computed data of CH1P (upper word)
• Registers corresponding to models with the /M1 computation function option.
• The data is a 32-bit signed integer. Two registers are assigned for each data point. The
value is the same as the computed data in binary output format (see page 5-13).
• The decimal point and unit information varies depending on the span setting of each
channel.
• Valid channels vary depending on the device. An error response (code 2) occurs when an
invalid channel is read.
33001
Alarm status of the computed data of CH0A
:
:
33024
Alarm status of the computed data of CH1P
• Registers corresponding to models with the /M1 computation function option.
• Data type and value are the same as those of the alarm status of the measured data.
• Valid channels vary depending on the device. An error response (code 2) occurs when an
invalid channel is read.
36001
List of alarms of the measured data of CH01 to CH04
36002
List of alarms of the measured data of CH05 to CH08
36003
List of alarms of the measured data of CH09 to CH12
36004
List of alarms of the measured data of CH13 to CH16
36005
List of alarms of the measured data of CH17 to CH20
36006
List of alarms of the measured data of CH21 to CH24
1 register (1 word)
Level 1
Level 2
Level 3
Level 4
The bit is set to 1 when an alarm occurs.
Example of Register 36001
CH4
CH3
CH2
CH1
36007 to 36020
Always 0
List of alarms of the measured data of CH0A to CH0D
36022
List of alarms of the measured data of CH0E to CH0J
36023
List of alarms of the measured data of CH0K to CH0P
36024
List of alarms of the measured data of CH1A to CH1D
36025
List of alarms of the measured data of CH1E to CH1J
36026
List of alarms of the measured data of CH1K to CH1P
• The register details are same as the list of alarms of measured data.
• Registers 36001 to 35026 can be read with a single command.
39001
Year (4 digits)
39002
Month
39003
Day
39004
Hour
39005
Minute
39006
Second
39007
Millisecond
39008
Summer/Winter time
Hold register
Data
40001
Communication input data of C01
:
:
40024
Communication input data of C24
• Pen model: C01 to C08, dot model: C01 to C12 (RD100B), C01 to C24 (RD1800B)
• The data is a 16-bit signed integer.
Modbus Error Response
The recorder returns the following error codes to the master device. For the error
messages related to communications that the recorder displays, see appendix 4.
Code
Meaning
Cause
1
Bad function code
Unsupported function request.
2
Bad register number
Attempted to read/write to a register that
has no corresponding channel.
3
Bad number of registers
The specified number of registers is less
than or equal to 0 or greater than or
equal to 126 (when reading)/124 (when
writing).
However, no response is returned for the following cases.
•
CRC error
•
Errors other than those in the table above.
3.4 Modbus Slave Protocol Specifications
Using the RS-422A/485 Communication Interface (/C3 Option)
3.5
Setting the Serial Interface
The serial interface must be configured.
Setup Items
RS422/485
Address
End
Baud rate
Data length
Parity
Protocol
Procedure
For a description of the basic operations, see “Basic Key Operations” on page v.
Entering Basic Setting Mode
Hold down the MENU key for 3 seconds to display the Setting mode screen.
Next, hold down both the
( DISP ) key and the
(
FUNC ) key for 3 seconds to
display the Basic Setting mode screen.
Note
To cancel an operation, press the ESC key.
1.
Press the
key to select RS422/485 and then press the
key.
Basic=RS422/485
2.
Press the
key to select the recorder’s address and then press the
key.
Address=1
3.
Press the
key to select the Baud rate value and then press the
key.
Baud rate=38400
4
Press the
key to select the Data length value and then press the
key.
Data length=8
5
Press the
key to select the Parity value and then press the
key.
Parity=Even
6.
Press the
key to select the NORMAL value and then press the
key.
The setting complete screen appears.
Protocol=NORMAL
RS422/485
Setting complete
Saving the Settings
1.
Press the ESC key to return to the Basic= screen.
2.
Press the
key to select End and then press the
key.
3.
Press the
key to select Store and then press the
key.
The settings are activated, and the Operation mode screen appears.
Select the address from the following range.
01 to 32
• Baud rate
Select the baud rate from the following:
1200, 2400, 4800, 9600, 19200, or 38400
• Data length
Select the data length from below. To output data in BINARY format, be sure to set
the data length to 8 bits.
7 or 8
• Parity (Parity check method)
Select the parity check from the following:
Odd, Even, or None
• Protocol
Select the protocol when using the Modbus slave protocol.
NORMAL: Standard protocol
MODBUS: Modbus slave protocol
• Saving the Settings
To activate the settings made in the Basic Setting mode, the settings must be saved.
3.5 Setting the Serial Interface
The syntax of the setting/basic setting/output commands (see sections 4.4 to 4.7) of the
instrument is given below. ASCII codes (see appendix 1) are used for the character codes.
For the Maintenance/Test command syntax, see section 4.9.
For the Instrument Information server command syntax, see section 4.10.
Command name
Parameter
?
Terminator
Delimiter
Sub delimiter
SR 02,SKIP;SR 03,VOLT,2V,-1500,1800
Parameter (02 SKIP·····1800)
Delimiter (,)
Command name (SR)
Sub delimiter (;)
Command example
• Command Name
Defined using two alphabet characters.
• Parameter
•
Command parameters.
•
Set using alphabet characters or numerical values.
•
Parameters are separated by delimiters (commas).
•
When the parameter is a numerical value, the valid range varies depending on the
command.
•
Spaces before and after of the parameter are ignored (except for parameters that
are specified using an ASCII character string (unit, tag, and message string), when
spaces are valid.)
•
You can omit the parameters that do not need to be changed from their current
settings. However, delimiters cannot be omitted.
Example
SR 01,,2V
If multiple parameters are omitted and delimiters occur at the end of the command,
those delimiters can be omitted.
Example
SR 01,VOLT,,, → SR 01,VOLT
•
The number of digits of the parameters below is fixed. If the number of digits is not
correct when entering the command, a syntax error results.
•
Date
YY/MM/DD (8 characters)
YY: Year (Enter the lower two digits of the year.)
MM: Month
DD: Day
•
Time
HH:MM:SS (8 characters)
HH: Hour
MM: Minute
SS: Second
•
Channel number: 2 characters (Example: 01, 0A)
•
Relay number: 3 characters (Example: I01)
•
Communication input data: 3 characters (Example: C02)
•
Constants used in the computation function (/M1 option): 3 characters
(Example: K03)
•
Remote control (/R1 option) input terminal status: 3 characters (Example: D04)
Chapter 4 Commands
A question mark is used to specify a query.
•
By placing a query after a command or parameter, the setting information of the
corresponding command can be queried. Some commands cannot execute
queries. For the query syntax of each command, see sections 4.4 to 4.7.
Example 1
SR[ p1]?
SR? or SR p1? can be executed.
Example 2
SA[ p1[,p2]]?
SA?, SA p1?, or SA p1,p2? can be executed.
• Delimiter
•
A comma is used as a delimiter.
•
Parameters are separated by delimiters.
• Sub Delimiter
•
A semicolon is used as a sub delimiter.
•
By separating each command with a sub delimiter, up to 10 commands can be
specified one after another. However, the following commands and queries cannot
be specified one after another. Use them independently.
•
Output commands other than BO, CS, and IF commands.
•
YE command
•
Queries
* If there are consecutive sub delimiters, they are considered to be single. In addition, sub
delimiters at the front and at the end are ignored.
Example
;SR01,VOLT;;;SR02,VOLT; is taken to be
SR01,VOLT;SR02,VOLT.
• Terminator (Terminating Character)
Use either of the following two characters for the terminator.
•
CR + LF (0DH 0AH in ASCII code.)
•
LF (0AH in ASCII code.)
Note
• The total data length from the first character to the terminator must be less than 2047 bytes. In
addition, the character string length of 1 command must be less than 512 bytes.
• Commands are not case sensitive (with the exception of user-specified character strings).
• All the commands that are listed using sub delimiters are executed even if one of the
commands is erroneous.
• Spaces that are inserted before and after a parameter are ignored. However, if spaces are
inserted before a command, after a sub delimiter, or after a query, an error occurs.
Response
The recorder returns a response (affirmative/negative response) to a command that is
separated by a single terminator.* The controller should follow the one command to one
response format. When the command-response rule is not followed, the operation is not
guaranteed.
For the response syntax, see section 5.1.
* The exceptions are the RS-422A/485 dedicated commands (see section 4.8).
Note
When using the RS-422A/485 interface, allow at least 1 ms before sending the next command
after receiving a response. Otherwise, the command may not be processed correctly.
4.1 Command Syntax
A List of Commands
Execution Modes and User Levels
Execution Modes
The recorder has two execution modes. Each command is specified to be used in a
particular execution mode. If you attempt to execute a command in a mode that is
different from the specification, a syntax error occurs. Use the DS command to switch to
the appropriate mode, and then execute the command. Query commands can be
executed in either mode.
•
Basic Setting mode
Measurement/computation is stopped and settings are changed in this mode.
•
Run mode
Run mode collectively refers to Operation mode and Setting mode of the recorder.
User Levels
The administrator and user specifications in the table indicate the user level that is
specified using the login function for Ethernet communications. For details, see section
1.1.
Setting Commands
Command
Function
Execution Mode
Administrator
User
Page
Name
SR
Sets the input range.
Run mode
Yes
No
4-10
SO
Sets the computing equation (/M1 option).
Run mode
Yes
No
4-11
VB
Sets the bias.
Run mode
Yes
No
4-11
SA
Sets the alarm.
Run mode
Yes
No
4-11
SN
Sets the unit.
Run mode
Yes
No
4-12
SC
Sets the chart speed.
Run mode
Yes
No
4-12
SD
Sets the date and time.
Run mode
Yes
No
4-13
VT
Sets the dot printing interval (dot model).
Run mode
Yes
No
4-13
SZ
Sets zone recording.
Run mode
Yes
No
4-13
SP
Sets the partial expanded recording.
Run mode
Yes
No
4-13
VR
Turns ON/OFF the recording on each channel.
Run mode
Yes
No
4-13
ST
Sets the tag.
Run mode
Yes
No
4-14
SG
Sets the message.
Run mode
Yes
No
4-14
SE
Sets the secondary chart speed
Run mode
Yes
No
4-14
(used by the remote control function (/R1 option)).
SV
Sets the moving average (dot model).
Run mode
Yes
No
4-14
SF
Sets the input filter (pen model).
Run mode
Yes
No
4-14
BD
Sets the alarm delay time.
Run mode
Yes
No
4-14
VF
Sets the brightness of the display (VFD) and internal
Run mode
Yes
No
4-14
illumination.
TD
Sets the DST.
Run mode
Yes
No
4-14
SS
Sets the DST. (Can be used on the RD100B with system
Run mode
Yes
No
4-15
version 1.02 or earlier)
SK
Sets the computation constant (/M1 option).
Run mode
Yes
No
4-15
SJ
Sets the timer used in TLOG computation (/M1 option).
Run mode
Yes
No
4-15
CM
Sets the communication input data (/M1 option).
Run mode
Yes
No
4-15
FR
Sets the acquiring interval to the FIFO buffer.
Run mode
Yes
No
4-15
VD
Sets the data display screen.
Run mode
Yes
No
4-16
Yes: Command usable
No: Command not usable
Basic Setting Commands
Note
• In order to activate the settings that are changed using the basic setting commands, the
settings must be saved using the YE or XE command.
• The settings that are returned in response to a query in the basic setting mode will contain the
new settings even if they are not saved.
• When the settings are saved with the XE command, communication is not dropped. The
settings that are changed using the YS/YB/YA/YN/YD/YQ/YK command are activated after
power-cycling the recorder.
• When the YE command is executed, communication is dropped. The response to the YE
command is not returned.
Command Function
Execution Mode
Administrator
User
Page
Name
XA
Sets alarm related settings.
Basic Setting mode
Yes
No
4-18
XI
Sets the A/D integral time.
Basic Setting mode
Yes
No
4-18
XB
Sets the burnout detection.
Basic Setting mode
Yes
No
4-19
XJ
Sets the RJC.
Basic Setting mode
Yes
No
4-19
UC
Changes the dot color (dot model).
Basic Setting mode
Yes
No
4-19
UO
Sets the pen offset compensation (pen model).
Basic Setting mode
Yes
No
4-19
UP
Sets the items to be printed.
Basic Setting mode
Yes
No
4-19
UR
Sets the periodic printout interval.
Basic Setting mode
Yes
No
4-19
UM
Sets the types of report data that are output to the
Basic Setting mode
Yes
No
4-20
periodic printout.
UB
Sets the display mode of the bar graph.
Basic Setting mode
Yes
No
4-20
UI
Sets whether to use moving average (dot model).
Basic Setting mode
Yes
No
4-20
UJ
Sets whether to use the input filter (pen model).
Basic Setting mode
Yes
No
4-20
UK
Sets whether to use of partial expanded recording.
Basic Setting mode
Yes
No
4-20
UL
Selects the display/record language.
Basic Setting mode
Yes
No
4-20
XN
Selects the date format.
Basic Setting mode
Yes
No
4-20
XT
Selects the temperature unit.
Basic Setting mode
Yes
No
4-21
UF
Sets whether to use the extended functions.
Basic Setting mode
Yes
No
4-21
UT
Selects the time printout format.
Basic Setting mode
Yes
No
4-21
XR
Sets the remote control input (/R1 option).
Basic Setting mode
Yes
No
4-21
YS
Sets the RS-422A/485 interface (/C3 option).
Basic Setting mode
Yes
No
4-21
XQ
Sets the TLOG timer (/M1 option).
Basic Setting mode
Yes
No
4-22
UN
Changes the assignment of channels to the recording pen Basic Setting mode
Yes
No
4-22
(pen model, /M1 option).
US
Sets the computation error procedure (/M1 option).
Basic Setting mode
Yes
No
4-22
YB
Sets the host name and domain name (/C7 option).
Basic Setting mode
Yes
No
4-22
YA
Sets the IP address (/C7 option).
Basic Setting mode
Yes
No
4-22
YN
Sets the DNS (/C7 option).
Basic Setting mode
Yes
No
4-23
YD
Sets whether to use the login function via communication
Basic Setting mode
Yes
No
4-23
(/C7 option).
YQ
Sets the communication timeout (/C7 option)
Basic Setting mode
Yes
No
4-23
YK
Sets keepalive (/C7 option).
Basic Setting mode
Yes
No
4-23
UA
Sets the record position.
Basic Setting mode
Yes
No
4-23
YE
Exits from Basic Setting mode.
Basic Setting mode
Yes
No
4-24
XE
Exits from Basic Setting mode.
Basic Setting mode
Yes
No
4-24
Yes: Command usable
No: Command not usable
4.2 A List of Commands
Switches the execution mode.
All modes
Yes
No
4-25
RCD
PS
Starts/Stops recording.
Run mode
Yes
No
4-25
DISP
UD
Switches the screen/switches the channel.
Run mode
Yes
No
4-25
FUNC AK
Executes alarm acknowledge (AlarmACK)
Run mode
Yes
No
4-25
FUNC TL
Starts/stops/resets computation (/M1 option).
Run mode
Yes
No
4-25
FUNC MP
Starts/Stops manual print.
Run mode
Yes
No
4-25
FUNC LS
Starts/Stops the list (setting information) printout.
Run mode
Yes
No
4-25
FUNC SU
Starts/Stops the setup list (basic setting information)
Run mode
Yes
No
4-25
printout.
FUNC MS
Executes the message printout.
Run mode
Yes
No
4-25
FUNC AC
Clears the alarm printout buffer.
Run mode
Yes
No
4-25
FUNC MC
Clears the message printout buffer.
Run mode
Yes
No
4-26
FUNC VG
Resets the report data of the periodic printout.
Run mode
Yes
No
4-26
-
YC
Initializes the settings.
Basic Setting mode
Yes
No
4-26
-
UY
Stops the record position adjustment.
Basic Setting mode
Yes
No
4-26
Yes: Command usable
No: Command not usable
Output Commands
Command
Command Function
Execution Mode
Administrator
User
Page
Type
Name
Control
BO
Sets the byte output order.
All modes
Yes
Yes
4-26
CS
Sets the check sum
All modes
Yes
Yes
4-26
(usable only during serial communications).
IF
Sets the status filter
All modes
Yes
Yes
4-26
CC
Disconnects an Ethernet connection
All modes
Yes
Yes
4-26
(usable only during Ethernet communications)
Setup, measurement, and computation data output
FE
Outputs decimal point position, unit information,
All modes
Yes
Yes
4-27
and setup data.
FD
Outputs the most recent measured/computed data.
Run mode
Yes
Yes
4-27
FY
Outputs the statistical computation results.
Run mode
Yes
Yes
4-27
FF
Outputs FIFO data.
Run mode
Yes
Yes
4-27
IS
Outputs status information.
All modes
Yes
Yes
4-28
FU
Outputs user information.
All modes
Yes
Yes
4-28
Yes: Command usable
No: Command not usable
RS-422A/485 Dedicated Commands
Command Function
Execution Mode
Administrator
User
Page
Name
Esc O
Opens the instrument.
All modes
Yes
Yes
4-28
Esc C
Closes the instrument.
All modes
Yes
Yes
4-28
Yes: Command usable
No: Command not usable
4.2 A List of Commands
These commands can be used only when using Ethernet communications.
Command Name
Function
Execution Mode
Administrator
User
Page
close
Disconnects the connection between other instruments.
All modes
Yes
No
4-29
con
Outputs connection information.
All modes
Yes
Yes
4-29
eth
Output Ethernets statistical information.
All modes
Yes
Yes
4-29
help
Outputs help.
All modes
Yes
Yes
4-29
net
Outputs network statistical information.
All modes
Yes
Yes
4-29
quit
Disconnects the connection of the instrument being
All modes
Yes
Yes
4-30
operated.
Yes: Command usable
No: Command not usable
Instrument Information Output Commands
These commands can be used only when using Ethernet communications.
Parameter
Function
Page
serial
Outputs the serial number.
4-30
host
Outputs the host name.
4-30
ip
Outputs the IP address.
4-30
4.2 A List of Commands
This section explains frequently used parameters.
Input Range
The following tables show the input types (VOLT, TC, RTD, DI, and 1-5V), range types,
and the ranges for the leftmost and rightmost values of the span.
• DC Voltage (VOLT), Square Root (SQRT), Difference between Channels (DELTA)
Range Type
Parameter for
Range of Leftmost and
Range of Leftmost and
the SR Command
Rightmost Values of Span
Rightmost Values of Span
of the SR Command
20 mV
20 mV
–20.00 to 20.00 mV
–2000 to 2000
60 mV
60 mV
–60.00 to 60.00 mV
–6000 to 6000
200 mV
200 mV
–200.0 to 200.0 mV
–2000 to 2000
2 V
2 V
–2.000 to 2.000 V
–2000 to 2000
6 V
6 V
–6.000 to 6.000 V
–6000 to 6000
20 V
20 V
–20.00 to 20.00 V
–2000 to 2000
50 V
50 V
–50.00 to 50.00 V
–5000 to 5000
• 1-5V
Range Type
Parameter for
Range of Leftmost and
Range of Leftmost and
the SR Command
Rightmost Values of Span
Rightmost Values of Span
of the SR Command
1-5V
1-5V
Leftmost value:
800 to 1200
0.8000 to 1.200 V
Rightmost value:
4800 to 5200
4.8000 to 5.200 V
• Thermocouple (TC)
Range Type
Parameter for
Range of Leftmost and
Range of Leftmost and
the SR Command
Rightmost Values of Span
Rightmost Values of Span
of the SR Command
R
R
0.0 to 1760.0°C
0 to 17600
S
S
0.0 to 1760.0°C
0 to 17600
B
B
0.0 to 1820.0°C
0 to 18200
K
K
–200.0 to 1370.0°C
–2000 to 13700
E
E
–200.0 to 800.0°C
–2000 to 8000
J
J
–200.0 to 1100.0°C
–2000 to 11000
T
T
–200.0 to 400.0°C
–2000 to 4000
N
N
0.0 to 1300.0°C
0 to 13000
W
W
0.0 to 2315.0°C
0 to 23150
L
L
–200.0 to 900.0°C
–2000 to 9000
U
U
–200.0 to 400.0°C
–2000 to 4000
WRe
WRe
0.0 to 2400.0°C
0 to 24000
Difference between channels (DELTA)
R
-
–1760.0 to 1760.0°C
–17600 to 17600
S
-
–1760.0 to 1760.0°C
–17600 to 17600
B
-
–1820.0 to 1820.0°C
–18200 to 18200
K
-
–1570.0 to 1570.0°C
–15700 to 15700
E
-
–1000.0 to 1000.0°C
–10000 to 10000
J
-
–1300.0 to 1300.0°C
–13000 to 13000
T
-
–600.0 to 600.0°C
–6000 to 6000
N
-
–1300.0 to 1300.0°C
–13000 to 13000
W
-
–1999.9 to 2315.0°C
–19999 to 23150
L
-
–1100.0 to 1100.0°C
–11000 to 11000
U
-
–600.0 to 600.0°C
–6000 to 6000
WRe
-
–1999.9 to 2400.0°C
–19999 to 24000
• Resistance Temperature Detector (RTD)
Range Type
Parameter for
Range of Leftmost and
Range of Leftmost and
the SR Command
Rightmost Values of Span
Rightmost Values of Span
of the SR Command
Pt100
PT
–200.0 to 600.0°C
–2000 to 6000
JPt100
JPT
–200.0 to 550.0°C
–2000 to 5500
Difference between channels (DELTA)
Pt100
-
–800.0 to 800.0°C
–8000 to 8000
JPt100
-
–750.0 to 750.0°C
–7500 to 7500
• ON/OFF input (DI)
Range Type
Parameter for
Range of Leftmost and
Range of Leftmost and
the SR Command
Rightmost Values of Span
Rightmost Values of Span
of the SR Command
Level
LEVEL
0 to 1†
0 to 1
Contact
CONT
0 to 1††
0 to 1
†
“0” when less than 2.4 V, “1” when greater than or equal to 2.4 V.
†† “0” when contact is OFF, “1” when contact is ON.
Miscellaneous
Channel Number
• Pen model
Measurement channel: 01 to 04
Computation channel: 0A, 0B, 0C, 0D, 0E, 0F, 0G, 0J
• Dot model
Measurement channel:
RD100B: 01 to 06
RD1800B: 01 to 24
Computation channel:
RD100B: 0A, 0B, 0C, 0D, 0E, 0F, 0G, 0J, 0K, 0M, 0N, 0P
RD1800B: 0A, 0B, 0C, 0D, 0E, 0F, 0G, 0J, 0K, 0M, 0N, 0P,
1A, 1B, 1C, 1D, 1E, 1F, 1G, 1J, 1K, 1M, 1N, 1P
Relay Number (/A1, /A2, /A3, /A4, and /A5 Options)
•
Models with the /A1 option: I01, I02
•
Models with the /A2 option: I01 to I04
•
Models with the /A3 option: I01 to I06
•
Models with the /A4 option: I01 to I06, I11 to I16 (RD1800B)
•
Models with the /A5 option: I01 to I06, I11 to I16, I21 to I26, I31 to I36 (RD1800B)
Communication Input Data
RD100B
Pen model: C01 to C08, dot model: C01 to C12
µR20000
Pen model: C01 to C08, dot model: C01 to C24
Computation Constant (/M1 option).
K01 to K30
Remote Control Input Terminal (/R1 Option)
D01 to D05
4.3 Parameter Values
Chart Speed on the Pen Model
5
6
8
9
10
12
15
16
18
20
24
25
30
32
36
40
45
48
50
54
60
64
72
75
80
90
96
100
120
125
135
150
160
180
200
225
240
250
270
300
320
360
375
400
450
480
500
540
600
675
720
750
800
900
960
1000
1080
1200
1350
1440
1500
1600
1800
2000
2160
2250
2400
2700
2880
3000
3600
4000
4320
4500
4800
5400
6000
7200
8000
9000
10800 12000
mm/h
4.3 Parameter Values
4.4 Setting Commands
4.4
Setting Commands
SR
Sets the input range.
When setting channels to skip
Syntax
SR p1,p2
p1
Channel number
p2
Measurement mode (SKIP)
Query
SR[ p1]?
Example
Set channel 01 to skip.
SR 01,SKIP
Description •
This command cannot be specified while
computation is in progress.
•
Measurements are not made on channels that
are set to SKIP.
When setting channels to voltage, TC, RTD, or ON/
OFF input
Syntax
SR p1,p2,p3,p4,p5
p1
Channel number
p2
Measurement mode (Input type)
VOLT
DC voltage
TC
Thermocouple
RTD
Resistance temperature
detector
DI
ON/OFF input
p3
Range type
p4
Leftmost value of span
p5
Rightmost value of span
Query
SR[ p1]?
Example
Measure 0°C to 1760.0°C on channel 01 using
thermocouple type R.
SR 01,TC,R,0,17600
Description •
This command cannot be specified while
computation is in progress.
•
Set p3 according to the table in section 4.3.
•
For p4 and p5, enter an integer value of 5
digits or less according to the table in section
4.3. The decimal position is fixed to the
position indicated in the table in section 4.3.
When setting channels to 1-5V range
Syntax
SR p1,p2,p3,p4,p5,p6,p7,p8
p1
Channel number
p2
Measurement mode (Input type) (1-5V)
p3
Leftmost value of span
p4
Rightmost value of span
p5
Leftmost value of scaling (−20000 to 30000)
p6
Rightmost value of scaling (−20000 to
30000)
p7
Scale decimal point position (0 to 4)
p8
Whether to use 1-5V low-cut function (ON,
OFF)
Query
SR[ p1]?
Example
Set channel 01 to 1-5V range and scale the input
value in the range 0.0 to 1200.0. Do not use the
low-cut function.
SR 01,1-5V,1000,5000,0,12000,1,OFF
Description •
This command cannot be specified while
computation is in progress.
•
For p3 and p4, enter an integer value of 5
digits or less according to the table in section
4.3. The decimal position is fixed to the
position indicated in the table in section 4.3.
•
Be sure that p6 is greater than p5.
•
Parameter p8 is valid only when the low-cut
function is enabled (see the UF command).
When computing the difference between channels
Syntax
SR p1,p2,p3,p4,p5
p1
Channel number
p2
Measurement mode (DELTA)
p3
Reference channel
p4
Leftmost value of span
p5
Rightmost value of span
Query
SR[ p1]?
Example
Set channel 03 to channel difference
computation with respect to channel 01
(reference channel). Set the leftmost and
rightmost values of span to −200.0 and 200.0,
respectively.
SR 03,DELTA,01,−2000,2000
Description •
This command cannot be specified while
computation is in progress.
•
The reference channel must be a channel that
is smaller in channel number than itself.
•
The range type is the same as that of the
reference channel.
•
For p4 and p5, enter an integer value of 5
digits or less according to the table in section
4.3. The decimal position is fixed to the
position indicated in the table in section 4.3.
When setting the linear scaling
Syntax
SR p1,p2,p3,p4,p5,p6,p7,p8,p9
p1
Channel number
p2
Measurement mode (SCALE)
p3
Input type
VOLT
DC voltage
TC
Thermocouple
RTD
Resistance temperature
detector
DI
ON/OFF input
p4
Range type
p5
Leftmost value of span
p6
Rightmost value of span
p7
Leftmost value of scaling (−20000 to 30000)
p8
Rightmost value of scaling (−20000 to
30000)
p9
Scaling decimal point position (0 to 4)
Query
SR[ p1]?
Example
Scale channel 02 whose input range is 0 to 10 V
to −100.0 to 500.0.
SR 02,SCALE,VOLT,20V,0,1000,−1000,
5000,1
This command cannot be specified while
computation is in progress.
•
Set p4 according to the table in section 4.3.
•
For p5 and p6, enter an integer value of 5
digits or less according to the table in section
4.3. The decimal position is fixed to the
position indicated in the table in section 4.3.
•
Be sure that p8 is greater than p7.
When setting the square root
Syntax
SR p1,p2,p3,p4,p5,p6,p7,p8,p9,p10
p1
Channel number
p2
Measurement mode (SQRT)
p3
Range type
p4
Leftmost value of span
p5
Rightmost value of span
p6
Leftmost value of scaling (−20000 to 30000)
p7
Rightmost value of scaling (−20000 to
30000)
p8
Scaling decimal point position (0 to 4)
p9
Low-cut function (ON, OFF)
p10 Low-cut value
Query
SR[ p1]?
Example
Given channel 01 whose input range is 0 to 10 V,
take the square root of the input value and scale
the result in the range 0.00 (m3/s) to 100.00 (m3/
s). When the input value is less than or equal to
5.0% of the recording span, use the low-cut
function.
SR 01,SQRT,20V,0,1000,0,10000,2,ON,
50
Description •
This command cannot be specified while
computation is in progress.
•
Set p3 according to the table in section 4.3.
•
For p4 and p5, enter an integer value of 5
digits or less according to the table in section
4.3. The decimal position is fixed to the
position indicated in the table in section 4.3.
•
Be sure that p7 is greater than p6.
•
Parameters p9 and p10 is valid only when the
low-cut function is enabled (see the UF
command).
SO
Sets the computing equation
(/M1 option).
Syntax
SO p1,p2,p3,p4,p5,p6
p1
Computation channel number
p2
Turn ON/OFF the computing equation (ON,
OFF)
p3
Computing equation (up to 120 characters)
p4
Leftmost value of span (−9999999 to
99999999)
p5
Rightmost value of span (−9999999 to
99999999)
p6
Span decimal point position (0 to 4)
Query
SO[ p1]?
Example
Calculate the sum of channel 01 and 02 on
computation channel 0A. Set the leftmost and
rightmost values of span to –10.00 and 15.00,
respectively.
SO 0A,ON,01+02,−1000,1500,2
Description •
This command cannot be specified while
computation is in progress.
•
For a description on the computing equations,
see the RD100B Recorder User’s Manual.
•
For p4 and p5, enter a value using 7 digits or
less ,excluding the decimal, for negative
numbers and 8 digits or less for positive
numbers.
VB
Sets the bias.
Syntax
VB p1,p2,p3
p1
Channel number
p2
Bias ON/OFF (ON, OFF)
p3
Bias value
Query
VB[ p1]?
Example
Given channel 03 whose range type is 2V
(measurable range: −2.000 to 2.000 V), add a
bias of 0.1 V.
VB 03,ON,100
Description •
Set p3 in the range of ±10% of the span of the
measurable range at the range type or ±10%
of the scaling span. Specify the value with an
integer (see section 4.3).
•
The bias setting is valid only when the bias
function is enabled (see the UF command).
SA
Sets the alarm.
When not using the alarm
Syntax
SA p1,p2,p3
p1
Channel number
p2
Alarm number (1 to 4)
p3
Alarm ON/OFF state (OFF)
Query
SA[ p1[,p2]]?
Example
Do not use alarm number 4 of channel 01.
SA 01,4,OFF
When using the alarm
Syntax
SA p1,p2,p3,p4,p5,p6,p7
p1
Channel number
p2
Alarm number (1 to 4)
p3
Alarm ON/OFF state (ON)
p4
Alarm type
H High limit alarm
L Low limit alarm
h Difference high limit alarm
l Difference low limit alarm
R High limit on rate-of-change alarm
r Low limit on rate-of-change alarm
T Delay high limit alarm
t Delay low limit alarm
(Characters are case-sensitive.)
p5
Alarm value
4.4 Setting Commands
4.4 Setting Commands
p6
Relay output
ON
Relay ON
OFF
Relay OFF
p7
Relay number (RD100B: I01 to I06,
RD1800B: I01 to I06, I11 to I16, I21 to I26,
I31 to I36)
Query
SA[ p1[,p2]]?
Example
Set a high limit alarm (alarm value = 1000) on
alarm number 1 of channel 02 and output to
relay I01.
SA 02,1,ON,H,1000,ON,I01
Description •
When the input range is set to SKIP (SR
command), p3 cannot be turned ON.
•
The alarm settings are all turned OFF for the
following cases.
•
When the input type is changed (VOLT,
TC, etc).
•
When the range type is changed.
•
When the span and scaling values are
changed during linear scaling (includes
changing the decimal position).
•
The h and l settings of p4 are valid only when
the measurement range is set to computation
between channels.
•
The hysteresis of alarm ON/OFF (valid when
p4 is H or L) is set using the XA command.
•
If p4 is set to R or r, set the interval for the
high/low limit on the rate-of-change using the
XA command.
•
The T and t settings of p4 can be specified
when the alarm delay function is enabled (UF
command).
•
If p4 is set to T or t, set the alarm delay time
for the delay high/low limit alarm using the BD
command.
•
Parameter p5 for the high limit alarm/low limit
alarm and delay high limit alarm/delay lowlimit
alarm
•
For DC voltage, TC, and RTD input: Values
within the measurable range in the
specified range (example: –2.000 to 2.000
V for the 2 V range).
•
For ON/OFF input (DI): 0 or 1.
•
For scaling (1-5V, scaling, and square
root): −5 to 105% of the scale span (except
within −20000 to 30000).
•
Parameter p5 for the difference high limit
alarm/difference low limit alarm: Values in the
measurable range can be specified (example:
–1760.0 to 1760.0°C for the TC type R).
•
Parameter p5 for the high limit on rate-of-
change alarm/low limit on rate-of-change
alarm: A value greater than or equal to 1 digit
can be specified. For example, 1 digit
corresponds 0.001 for the 2 V range
(measurable range: –2.000 to 2.000 V). The
maximum value that can be specified is the
width of the measurable range (4.000 V for 2
V range). For ON/OFF input, only “1” can be
specified.
•
On models with the computation function (/M1
option), alarms can be set on computation
channels.
•
When the computation equation setting
(SO command) is turned OFF, p3 cannot
be turned ON.
•
For computation channels, the alarm types
that can be specified are H (high limit
alarm), L (low limit alarm), T (delay high
limit alarm), and t (delay low limit alarm). T
and t can be specified when the alarm
delay function is enabled (UF command).
•
If p4 is set to T or t, set the alarm delay
time for the delay high/low limit alarm using
the BD command.
•
Set p5 within the range –9999999 to
99999999 excluding the decimal using an
integer.
•
The alarm ON/OFF hysteresis is set using
the XA command.
•
If the computation channel ON/OFF state,
the computing equation, or the span value
is changed, all alarm settings of that
channel are turned OFF.
SN
Sets the unit.
Syntax
SN p1,p2
p1
Channel number
p2
Unit string (up to 6 characters)
Query
SN[ p1]?
Example
Set the unit of channel 02 to M/H.
SN 02,M/H
Description •
The unit setting is valid on channels set to 1-
5V, scaling, or square root.
•
For the characters that can be used for the
units, see appendix 1, “ASCII Character
Codes.”
•
On models with the computation function (/M1
option), unit can be set on computation
channels.
SC
Sets the chart speed.
Syntax
SC p1
p1
Chart speed
Query
SC?
Example
Set the chart speed to 25 mm/h.
SC 25
Description Select the chart speed from the list of choices
below.
Pen model
5 to 12000 mm/h (82 levels, see section 4.3)
Dot model
1 to 1500 mm/h (1 mm steps)
Sets the date and time.
Syntax
SD p1
p1
Date and time (fixed to the YY/MM/DD
HH:MM:SS format)
YY
Year (00 to 99, the lower 2
digits)
MM
Month (01 to 12)
DD
Day (01 to 31)
HH
Hour (00 to 23)
MM
Minute (00 to 59)
SS
Second (00 to 59)
Query
SD?
Example
Set the recorder clock to 13:00:00, December 1,
2004.
SD 04/12/01 13:00:00
Description •
The format of p1 is fixed to 17 characters. An
error results if a space is inserted in positions
other than the specified positions.
•
On the RD100B with system version 1.02 or
earlier, the parameter p2 can be used.
p2
Standard time or DST (winder, summer)
Winter
Standard time
Summer
DST
VT
Sets the dot printing interval
(dot model).
Syntax
VT p1
p1
Dot printing interval setting.
AUTO
Automatically adjust the dot
printing interval according to
the chart speed.
FIX
Record at the fastest printing
interval.
Query
VT?
Example
Record at the fastest printing interval.
VT FIX
Description When trend recording is performed on 6
channels, the fastest printing interval is 10
seconds.
SZ
Sets zone recording.
Syntax
SZ p1,p2,p3
p1
Channel number
p2
Leftmost position of the zone (RD100B: 0 to
95, RD1800B: 0 to 175) [mm]
p3
Rightmost position of the zone (RD100B: 5
to 100, RD1800B: 0 to 180) [mm]
Query
SZ[ p1]?
Example
Display channel 02 in a zone between 30 and 50
mm.
SZ 02,30,50
Description •
Be sure that p3 is greater than p2 and that the
zone width (p3 – p2) is greater than or equal
to 5 mm.
•
On models with the computation function (/M1
option), computation channels can be
specified.
SP
Sets the partial expanded
recording.
Syntax
SP p1,p2,p3,p4
p1
Channel number
p2
Partial expanded recording ON/OFF (ON,
OFF)
p3
Boundary position (1 to 99) [%]
p4
Boundary value
Query
SP[ p1]?
Example
Set the 25% position of channel 01 to 1.000 V.
SP 01,ON,25,1000
Description •
When the input range is set to SKIP (SR
command), p2 cannot be turned ON.
•
Set p3 as a percentage where 100 mm is
assumed to be 100%.
•
Parameter p4 can be set in the range
(leftmost value of span + 1) to (rightmost
value of span – 1). If scaling is enabled, the
range is (leftmost value of scaling – 1) to
(rightmost value of scaling + 1). Set the
parameter using an integer.
•
This setting is possible when partial expanded
recording is enabled (UK command).
•
This command cannot be used if the partial
expanded recording range does not exist
(when the span width is set to 1, for example).
•
On models with the computation function (/M1
option), computation channels can be
specified.
•
When the computation equation setting (SO
command) is turned OFF, p2 cannot be
turned ON.
VR
Turns ON/OFF the recording on
each channel.
Dot model
Syntax
VR p1,p2,p3
p1
Channel number
p2
Trend recording ON/OFF (ON, OFF)
p3
Periodic printout ON/OFF (ON, OFF)
Query
VR[ p1]?
Example
Turn trend recording ON and turn periodic
printout OFF on channel 06.
VR 06,ON,OFF
Description On models with the computation function (/M1
option), computation channels can be specified.
Pen model
Syntax
VR p1,p2
p1
Channel number
p2
Periodic printout ON/OFF (ON, OFF)
Query
VR[ p1]?
Example
Turn periodic printing ON on channel 01.
VR 01,ON
Description On models with the computation function (/M1
option), computation channels can be specified.
4.4 Setting Commands
4.4 Setting Commands
ST
Sets the tag.
Syntax
ST p1,p2
p1
Channel number
p2
Tag string (up to 7 characters)
Query
ST[ p1]?
Example
Set the tag of channel 02 to TI-2.
ST 02,TI-2
Description •
For the characters that can be used for the
tags, see appendix 1, “ASCII Character
Codes.”
•
On models with the computation function (/M1
option), computation channels can be
specified.
SG
Sets the message.
Syntax
SG p1,p2
p1
Message number (1 to 5)
p2
Message string (up to 16 characters)
Query
SG[ p1]?
Example
Set character string “START” in message
number 1.
SG 1,START
Description For the characters that can be used for the
messages, see appendix 1, “ASCII Character
Codes.”
SE
Sets the secondary chart speed
(used by the remote control
function (/R1 option)).
Syntax
SE p1
p1
Secondary chart speed
See the explanation.
Query
SE?
Example
Set the chart speed to 50 mm/h.
SE 50
Description Select the chart speed from the list of choices
below.
Pen model
5 to 12000 mm/h (82 levels, see section 4.3)
Dot model
1 to 1500 mm/h (1 mm steps)
SV
Sets the moving average
(dot model).
Syntax
SV p1,p2
p1
Channel number
p2
Number of samples for computing the
moving average (OFF, 2 to 16) [times]
Query
SV[ p1]?
Example
Set the number of samples for computing the
moving average of channel 02 to 12.
SV 02,12
Description •
This setting is available on the dot model.
•
This setting is possible when moving average
is enabled (UI command).
SF
Sets the input filter (pen model).
Syntax
SF p1,p2
p1
Channel number
p2
Filter time constant (OFF, 2s, 5s, 10s)
Query
SF[ p1]?
Example
Set the filter of channel 02 to 2 s.
SF 02,2s
Description This setting is possible when the input filter is
enabled (UJ command).
BD
Sets the alarm delay time.
Syntax
BD p1,p2
p1
Channel number
p2
Alarm delay (1 to 3600) [s]
Query
BD[ p1]?
Example
Set the alarm delay of channel 01 to 120 s.
BD 01,120
Description •
This setting is possible when the alarm delay
function is enabled (UF command).
•
On models with the computation function (/M1
option), computation channels can be
specified.
VF
Sets the brightness of the display
(VFD) and internal illumination.
Syntax
VF p1,p2
p1
VFD brightness (1 to 8)
p2
Internal illumination (OFF, 1 to 4)
OFF
Turns OFF the internal
illumination.
Query
VF?
Example
Set the display (VFD) brightness to 2 and the
internal illumination to 1.
VF 2,1
Description The brightness increases as the value increases.
TD
Sets the DST.
Syntax
TD p1,p2,p3,p4,p5,p6,p7,p8,p9
p1
Use/Not use DST (USE, NOT)
p2
DST start time: month (Jan, Feb, Mar, Apr,
May, Jun, Jul, Aug, Sep, Oct, Nov, or Dec)
p3
DST start time: number of the week in the
month (1st, 2nd, 3rd, 4th, or Last)
p4
DST start time: day of the week (Sun, Mon,
Tue, Wed, Thu, Fri, or Sat)
p5
DST start time: hour (0 to 23)
p6
DST end time: month (Jan, Feb, Mar, Apr,
May, Jun, Jul, Aug, Sep, Oct, Nov, or Dec)
p7
DST end time: number of the week in the
month (1st, 2nd, 3rd, 4th, or Last)
p8
DST end time: day of the week (Sun, Mon,
Tue, Wed, Thu, Fri, or Sat)
p9
DST end time: hour (0 to 23)
Query
TD[ p1]?
Enable DST and set the DST start time to hour 0
on the 2nd Sunday of June and the DST end
time to hour 0 on the 2nd Sunday of December.
TD Use,Jun,2nd,Sun,0,Dec,2nd,Sun,0
Description •
The DST start time and end time cannot be
set to the same time.
•
The TD command can be used on the
RD100B and RD1800B with system version
1.11 or later. On the RD100B with system
version 1.02 or earlier, use the SS command.
SS
Sets the DST.
Syntax
SS p1,p2,p3
p1
Standard time/DST selection
Winter
Standard time
Summer
DST
p2
Switch time setting ON/OFF (ON, OFF)
p3
Date and time (fixed to the YY/MM/DD HH
format)
YY
Year (00 to 99, the lower 2 digits)
MM
Month (01 to 12)
DD
Day (01 to 31)
HH
Hour (00 to 23)
Query
SS[ p1]?
Example
Switch to DST on 0 hour of June 1st, 2005.
SS Summer,ON,05/06/01 00
Description •
Parameter p3 is not available when p2 is OFF.
•
The SS command can be used on the
RD100B with system version 1.02 or earlier.
SK
Sets the computation constant
(/M1 option).
Syntax
SK p1,p2
p1
Constant number (K01 to K30)
p2
Constant (up to 11 characters)
The range is –9.9999E+29 to –1.0000E–30,
0, 1.0000E–30 to 9.9999E+29.
(The + sign of “E+” can be omitted.)
Query
SK[ p1]?
Example
Set 1.0000E−10 to constant K01.
SK K01,1.0000E−10
Description This command cannot be specified while
computation is in progress.
SJ
Sets the timer used in TLOG
computation (/M1 option).
Syntax
SJ p1,p2,p3
p1
Computation channel number
p2
Timer number (Periodic, 1, 2)
p3
Sum scale (for TLOG.SUM)
OFF
Integrates the data per scan interval.
/s
Integrates the physical amount that
are in unit of /s.
/min Integrates the physical amount that
are in unit of /min.
/h
Integrates the physical amount that
are in unit of /hour.
/day Integrates the physical amount that
are in unit of /day.
Query
SJ[ p1]?
Example
Enable timer 1 on computation channel 0B. No
sum scale designation.
SJ 0B,1,OFF
Description •
This command cannot be specified while
computation is in progress.
•
About p3
The data for sum computation are sampled
every scan interval. For data with units such
as /s, /min, /h, and /day as in a flow rate, the
data can be summed over the unit time as
shown below. Parameter p3 is valid only for
sum values.
OFF
Σ (measured value)
/s
Σ (measured value) × scan interval
/min
Σ (measured value) × scan interval/60
/h
Σ (measured value) × scan interval/
3600
/day
Σ (measured value) × scan interval/
(3600×24)
The unit of the scan interval is seconds.
CM
Sets the communication input
data (/M1 option).
Syntax
CM p1,p2
p1
Communication input data number (see
section 4.3)
p2
Communication input data
The selectable range is –9.9999E+29 to –
1.0000E–30, 0, 1.0000E–30 to 9.9999E+29.
(The + sign of “E+” can be omitted.)
Query
CM?
Example
Set 1.0000E–10 to communication input data
number C01.
CM C01,1.0000E−10
FR
Sets the interval for acquiring
data to the FIFO buffer
Syntax
FR p1
p1
Pen model: 125ms, 250ms, 500ms, 1s, 2s,
2.5s, 5s, 10s
Dot model: 1s, 2s, 2.5s, 5s, 10s
Query
FR?
Example
Set the acquiring interval to the FIFO buffer to 1 s.
FR 1s
Description •
Set the acquiring interval to an integer multiple
of the scan interval that is greater than the
scan interval.
•
The scan interval on the pen model is 125 ms.
•
The scan interval on the dot model varies
depends on the integration time of the A/D
converter as follows:
When the integration time is 16.7 ms or 20 ms;
6 channels: 1 s
12, 18, or 24 channels: 2.5 s
4.4 Setting Commands
When the integration time is 100 ms;
6 channels: 2.5 s
12 channels: 5 s
18 and 24 channels: 10 s
VD
Sets the data display screen.
On screens without detailed parameters
Syntax
VD p1,p2
p1
Screen number (01 to 15)
p2
Display type
4CH bargraph
4-channel bar graph display (pen
model)
6CH bargraph
6-channel bar graph display (RD100B
dot model)
Channel alarm status
Channel alarm status display
Time/Chart speed
Date/Time/Chart speed display
DI/DO
DI/DO status display
STATUS
Status display
System
System display
Light out
Screen OFF
SKIP
Skip
Query
VD[ p1]?
Example
Assign the status display to screen 02.
VD 02,STATUS
On screens with the displayed channel switching
interval
Syntax
VD p1,p2,p3
p1
Screen number (01 to 15)
p2
Display type
1CH digital
1-channel digital display
2CH digital
2-channel digital display
4CH digital
4-channel digital display
6CH digital
6-channel digital display (dot model)
12CH digital
12-channel digital display (RD1800B
12, 18 and 24 dot model)
1CH digital+1CH bargraph
1-channel digital display + 1 channel
bar graph display
1CH digital+4CH bargraph
1-channel digital display + 4 channel
bar graph display (pen model)
2CH digital+2CH bargraph
2-channel digital display + 2 channel
bar graph display
Tag_1CH digital
1-channel digital display (tag)
Tag_2CH digital
2-channel digital display (tag)
Tag_1CH digital+1CH bargraph
1-channel digital display + 1 channel
bar graph display (tag)
Tag_1CH digital+4CH bargraph
1-channel digital display + 4 channel
bar graph display (tag, pen model)
p3
Displayed channel switching interval
MANUAL
Fixed display channels, data
update rate (pen model): 2 s,
data update rate (dot model):
scan interval
AUTO1S
Switch the displayed
channels and data every
second.
AUTO2S
Switch the displayed
channels and data every 2
seconds.
AUTO3S
Switch the displayed
channels and data every 3
seconds.
AUTO4S
Switch the displayed
channels and data every 4
seconds.
AUTO5S
Switch the displayed
channels and data every 5
seconds.
Query
VD[ p1]?
Example
Assign the 2-channel digital display to screen 04
and automatically switch the displayed channels
every 5 seconds.
VD 04,2CH digital,AUTO5S
For flag display
Syntax
VD p1,p2,p3
p1
Screen number (01 to 15)
p2
Display type
FLAG
Flag display (fixed)
p3
Display ON/OFF for each channel
Specify the ON/OFF state of each channel
with 1 or 0 (see the explanation).
Query
VD[ p1]?
Example 1 On the 6-dot model, assign the flag display to
screen 03 and display the flags of channel 01
and 02.
VD 03,FLAG,110000
Example 2 On the 2-pen model with the computation
function (/M1 option), assign the flag display to
screen 03 and display the flags of measurement
channel 02 and computation channels 0A, 0B,
0E, and 0F.
VD 03,FLAG,01-11001100
4.4 Setting Commands
Description Use the format below to specify the display ON/
OFF state of each channel.
•
List the values one after another for all
channels and separate the measurement
channels and computation channels with a
hyphen.
•
Channels cannot be omitted.
•
The characters after the hyphen are valid only
on models with the computation function (/M1
option).
For multiple display (display in which different
screens can be assigned to the top and bottom
sections
Syntax
VD
p1,p2,p3,p4,p5,p6,p7,p8,p9,p10,p11,p12
p1
Screen number (01 to 15)
p2
Display type
Multiple display
Display in which different screens can
be assigned to the top and bottom
sections
p3
Top section display type
1CH digital
1-channel digital display
2CH digital
2-channel digital display
3CH digital
3-channel digital display (RD1800B)
TIME
Date/Time (RD100B)
Chart speed
Chart speed (RD100B)
TIME/Chart speed
Date/Time/Chart speed (RD1800B)
Channel alarm status
Channel alarm status display
STATUS
Status display
Light out
Screen OFF
Tag_1CH digital
1-channel digital display (tag)
Note
Parameters that become invalid due to p3, p4, p8, or p9
setting are skipped, and the next parameter is brought
forward.
p4
Displayed channel switching interval
Condition: Set this parameter when p3 is
1CH digital or 2CH digital.
MANUAL
Fixed display channels, data
update rate (pen model): 2 s,
data update rate (dot model):
scan interval
AUTO1S
Switch the displayed
channels and data every 1
second.
AUTO2S
Switch the displayed
channels and data every 2
seconds.
AUTO3S
Switch the displayed
channels and data every 3
seconds.
AUTO4S
Switch the displayed
channels and data every 4
seconds.
AUTO5S
Switch the displayed
channels and data every 5
seconds.
p5
Channel number
Condition: This parameter can be specified
when p3 is 1CH digital, 2CH digital,
or Tag_1CH digital and p4 is MANUAL.
p6
2nd channel number
Condition: This parameter is specified when
p3 is 2CH digital and p4 is MANUAL.
p7
3rd channel number
Condition: This parameter is specified when
p3 is 3CH digital and p4 is MANUAL.
p8
Bottom section display type
Same as p3
p9
Displayed channel switching interval
Same as p4
p10 Channel number
Same as p5
p11 2nd channel number
Same as p6
p12 3rd channel number
Same as p7
Query
VD[ p1]?
Example
Assign the display in which different screens can
be assigned to the top and bottom sections to
screen 09. Set the top section to status display
and the bottom section to 1-channel digital
display. Switch the displayed channel every 3
seconds (parameters p5 to p7, p10, and p12
explained above are omitted).
VD 09,Multiple display,STATUS,1CH
digital,AUTO3S
4.4 Setting Commands
4.5 Basic Setting Commands
4.5
Basic Setting Commands
•
In order to activate the settings that are
changed using the basic setting commands,
the settings must be saved using the YE or
XE command.
•
The settings that are returned in response to a
query in the basic setting mode will contain
the new settings even if they are not saved.
Note
•
The settings that are changed using the YS/YB/YA/YN/
YD/YQ/YK command and saved using the XE command
are activated after power-cycling the recorder.
•
When the YE command is executed, communication is
dropped.
XA
Sets alarm related settings.
Syntax
XA p1,p2,p3,p4,p5,p6,p7,p8,p9,p10
p1
Fault diagnosis output ON/OFF (ON, OFF)
p2
Reflash operation ON/OFF (ON, OFF)
p3
Relays that are to operate using AND logic
NONE
None (all relays operate using
OR logic)
I01
I01
I01-I02
I01, I02
I01-I03
I01 to I03
I01-I04
I01 to I04
I01-I05
I01 to I05
I01-I06
I01 to I06
I01-I11
I01 to I11 (RD1800B)
I01-I12
I01 to I12 (RD1800B)
I01-I13
I01 to I13 (RD1800B)
I01-I14
I01 to I14 (RD1800B)
I01-I15
I01 to I15 (RD1800B)
I01-I16
I01 to I16 (RD1800B)
I01-I21
I01 to I21 (RD1800B)
I01-I22
I01 to I22 (RD1800B)
I01-I23
I01 to I23 (RD1800B)
I01-I24
I01 to I24 (RD1800B)
I01-I25
I01 to I25 (RD1800B)
I01-I26
I01 to I26 (RD1800B)
I01-I31
I01 to I31 (RD1800B)
I01-I32
I01 to I32 (RD1800B)
I01-I33
I01 to I33 (RD1800B)
I01-I34
I01 to I34 (RD1800B)
I01-I35
I01 to I35 (RD1800B)
I01-I36
I01 to I36 (RD1800B)
p4
Energize/De-energize the relay
ENERGIZE
Energize the relay when
an alarm is detected
DE_ENERGIZE
De-energize the relay
when an alarm is
detected
p5
Hold/Not hold the relay
HOLD
Hold the relay output until an
alarm acknowledge operation
is executed
NONHOLD
Reset the relay output when
the alarm is cleared.
p6
Hold/Not hold the alarm status display
HOLD
Hold the display until an
alarm acknowledge operation
is executed
NONHOLD
Clear the display output when
the alarm is cleared.
p7
Interval for the high limit on rate-of-change
alarm (01 to 15)
p8
Interval for the low limit on rate-of-change
alarm (01 to 15)
p9
Alarm hysteresis on measurement channels
(OFF, 0.1%, 0.2%, 0.3%, 0.4%, 0.5%, 0.6%,
0.7%, 0.8%, 0.9%, or 1.0%)
p10 Alarm hysteresis on computation channels
(OFF, 0.1%, 0.2%, 0.3%, 0.4%, 0.5%, 0.6%,
0.7%, 0.8%, 0.9%, or 1.0%)
Query
XA?
Example
Enable fault diagnosis output. Disable reflash
and AND operations. Set the relay operation to
energize and hold. Set the alarm status display
to hold. Set the interval for the high limit of rate-
of-change alarm to 10, the interval for the low
limit of rate-of-change alarm to 12, the
measurement alarm hysteresis to 0.5%, and the
computation alarm hysteresis to OFF.
XA ON,OFF,NONE,ENERGIZE,HOLD,HOLD,
10,12,0.5%,OFF
Description •
The interval for the high limit of rate-of-change
alarm is equal to “scan interval × p7” and the
interval for the low limit of rate-of-change
alarm is equal to “scan interval × p8.”
•
Parameter p10 can be specified on models
with the computation function (/M1 option).
•
For the details on the settings, see the
Recorder User’s Manual.
XI
Sets the A/D integral time.
Syntax
XI p1
p1
Integration time of the A/D converter
AUTO
Automatically set in synch with
the power supply frequency.
50HZ
12.5 ms
60HZ
16.7 ms
100MS
100 ms (dot model)
Query
XI?
Example
Set the A/D integral time to 50 Hz.
XI 50HZ
Description 100 ms is available only on the dot model.
Sets the burnout detection.
Syntax
XB p1,p2
p1
Channel number
p2
The procedure taken when a sensor
burnout is detected
OFF
Disable the burnout detection.
UP
Set the recording off the scale to the
right when a burnout is detected.
DOWN Set the recording off the scale to the
left when a burnout is detected.
Query
XB[ p1]?
Example
Set the recording off the scale to the right when a
sensor burnout is detected on channel 01.
XB 01,UP
XJ
Sets the RJC.
When using the internal compensation circuit
Syntax
XJ p1,p2
p1
Channel number
p2
Reference junction compensation selection
(INTERNAL)
Query
XJ[ p1]?
Example
Use the internal compensation circuit on channel
01.
XJ 01,INTERNAL
When using an external RJC
Syntax
XJ p1,p2,p3
p1
Channel number
p2
Reference junction compensation selection
(EXTERNAL)
p3
Compensation voltage (–20000 to 20000)
[µV]
Query
XJ[ p1]?
Example
Set the reference junction compensation of
channel 02 to external and set the compensation
voltage to 0 µV.
XJ 02,EXTERNAL,0
Description The unit of p3 is µV (microvolts).
UC
Changes the dot color (dot model).
Syntax
UC p1,p2
p1
Channel number
p2
Dot color
PURPLE
RED
GREEN
BLUE
BROWN
BLACK
Query
UC[ p1]?
Example
Set the dot color of channel 06 to purple.
UC 06,PURPLE
Description On models with the computation function (/M1
option), the dot color of computation channels
can be changed.
UO
Sets the pen offset
compensation (pen model).
Syntax
UO p1
p1
Pen offset compensation ON/OFF (ON, OFF)
Query
UO?
Example
Use the pen offset compensation.
UO ON
UP
Sets the items to be printed.
Pen model
Syntax
UP p1,p2,p3,p4,p5,p6
p1
Channel number/tag selection
CHANNEL
Print the channel number
TAG
Print the tag
p2
Alarm printout
ON1
Print at alarm occurrence and release
ON2
Print at alarm occurrence
OFF
Do not print
p3
Record start time printout ON/OFF (ON, OFF)
p4
New chart speed printout ON/OFF (ON, OFF)
p5
Scale printout ON/OFF (ON, OFF)
p6
Recording color printout ON/OFF (ON, OFF)
Query
UP?
Example
Print tags. Print all items.
UP TAG,ON1,ON,ON,ON,ON
Dot model
Syntax
UP p1,p2,p3,p4,p5,p6
p1
Channel number/tag selection
CHANNEL
Print the channel number
TAG
Print the tag
p2
Channel number printout ON/OFF (ON,
OFF) by the trend recording
p3
Alarm printout
ON1
Print at alarm occurrence and release
ON2
Print at alarm occurrence
OFF
Do not print
p4
Record start time printout ON/OFF (ON, OFF)
p5
New chart speed printout ON/OFF (ON, OFF)
p6
Scale printout ON/OFF (ON, OFF)
Query
UP?
Example
Print the channel numbers. Print all items.
UP CHANNEL,ON,ON1,ON,ON,ON
UR
Sets the periodic printout interval.
Syntax
UR p1,p2,p3,p4
p1
Print interval mode
Auto
Automatically set the interval
according to the chart speed
Manual
Specify the interval
p2
Reference time (00 to 23 [hour])
p3
Interval (10min, 12min, 15min, 20min,
30min, 1h, 2h, 3h, 4h, 6h, 8h, 12h, 24h)
(when p1 is Manual)
4.5 Basic Setting Commands
4.5 Basic Setting Commands
p4
Periodic printout mode
OFF
Disable periodic printout
INST
Print instantaneous values.
REPORT
Print the report data over the
interval.
Query
UR?
Example
Carry out periodic printout every 2 hours with
hour 0 as the reference. Print the measured
values (instantaneous values) at each interval.
UR Manual,00,2h,INST
Description •
When p1 is set to Auto, the periodic printout
interval synchronizes to the chart speed.
•
If p4 is set to REPORT, set the type of report
data on each channel using the UM command.
UM
Sets the types of report data that
are output to the periodic
printout.
Syntax
UM p1,p2,p3
p1
Channel number
p2
Type of report data
INST
Instantaneous value
AVE
Average value
MIN
Minimum value
MAX
Maximum value
SUM
Sum value
MIX
Average + minimum +
maximum values
p3
Sum scale
OFF
Integrates the data per scan
interval.
/s
Integrates the physical amount
that are in unit of /s.
/min
Integrates the physical amount
that are in unit of /min.
/h
Integrates the physical amount
that are in unit of /hour.
/day
Integrates the physical amount
that are in unit of /day.
Query
UM[ p1]?
Example
Print the average values of channel 03 using
periodic printout.
UM 03,AVE
Description •
This setting is possible when printing of report
data (p4 is REPORT) is specified using the
UR command.
•
About p3
The data for sum computation are sampled
every scan interval. For data with units such
as /s, /min, /h, and /day as in a flow rate, the
data can be summed over the unit time.
Parameter p3 is valid only for sum values.
•
On models with the computation function (/M1
option), computation channels can be
specified.
UB
Sets the display mode of the bar
graph.
Syntax
UB p1,p2
p1
Channel number
p2
Bar graph display mode
NORMAL
The reference position is set
to the smaller of the two
values, leftmost value and
rightmost value.
CENTER
The reference position is set
to the 50% position.
Query
UB[ p1]?
Example
Display channel 01 using a bar graph with the
reference position set to the 50% position.
UB 01,CENTER
Description On models with the computation function (/M1
option), computation channels can be specified.
UI
Sets whether to use moving
average (dot model).
Syntax
UI p1
p1
Use/Not use moving average (USE, NOT)
Query
UI?
Example
Use moving average.
UI USE
UJ
Sets whether to use the input
filter (pen model).
Syntax
UJ p1
p1
Use/Not use the input filter (USE, NOT)
Query
UJ?
Example
Use the input filter.
UJ USE
UK
Sets whether to use of partial
expanded recording.
Syntax
UK p1
p1
Use/Not use partial expanded recording
(USE, NOT)
Query
UK?
Example
Use partial expanded recording.
UK USE
UL
Selects the display/record
language.
Syntax
UL p1
p1
Language (ENGLISH, JAPANESE)
Query
UL?
Example
Use English.
UL ENGLISH
XN
Selects the date format.
Syntax
UL p1
p1
Date format for displaying and printing
Y/M/D: (example) 2005/08/31
M/D/Y: (example) 08/31/2005
D/M/Y: (example) 31/08/2005
D.M.Y: (example) 31.08.2005
M.D.Y: (example) Aug.31.2005
Query
XN?
Example
Set to the "2005/08/31" format.
XN Y/M/D
Description This setting applies to the format of the date
shown on the display and the date printed on the
periodic printout, manual printout, alarm printout,
message printout, recording start printout, and
new chart speed printout.
Note
How this setting applies to the time printout format of the
alarm printout, message printout, recording start printout,
and new chart speed printout: The setting specifies the
format of the "M/D" and "Y/M/D" section when the time
printout format is set to "M/D H:M," "M/D H:M:S," or "Y/M/
D H:M:S."
XT
Selects the temperature unit.
Syntax
XT p1
p1
Temperature unit
C °C
F °F
Query
XT?
Example
Use °F.
XT F
UF
Sets whether to use the
extended functions.
Syntax
UF p1,p2,p3,p4
p1
Bias function (USE, NOT)
p2
Square root low-cut function (USE, NOT)
p3
1-5V input low-cut function (USE, NOT)
p4
Alarm delay function (USE, NOT)
Query
UF?
Example
Use all the extended functions.
UF USE,USE,USE,USE
Description •
The input offset is set using the VB command.
•
The low-cut function is set using the SR
command.
•
The delay alarm is set using the SA and BD
commands.
UT
Selects the time printout format.
Syntax
UT p1,p2,p3,p4
p1
Time printout format of the alarm printout
The available selections are the same as
those of p2. None is not available.
p2
Time printout format of the message printout
HH:MM
Hour:Minute
HH:MM:SS
Hour:Minute:Second
M/D H:M
Month Day Hour:Minute
M/D H:M:S
Month Day Hour:Minute:
Second
YMD H:M:S
Month Day Year Hour:Minute:
Second
None
No time printout
p3
Time printout format of the record start time
printout
The available selections are the same as
those of p2. None is not available.
p4
Time printout format of the new chart speed
printout
The available selections are the same as
those of p2. None is not available.
Query
UT?
Example
Set the all time printouts to Hour:Minute format.
UT HH:MM,HH:MM,HH:MM,HH:MM
XR
Sets the remote control input
(/R1 option).
Syntax
XR p1,p2
p1
Remote control input terminal number (1 to
5)
p2
Action
NONE
No action
Record On/Off
Start/Stop recording.
Chart speed
Change the chart
speed.
Time adjust
Reset the internal
clock to the nearest
hour.
Math start/stop
Start/Stop
computation.
Math reset
Reset computation.
Manual print
Execute manual print.
Alarm Ack
Execute alarm
acknowledge.
Message1
Print message 1
Message2
Print message 2
Message3
Print message 3
Message4
Print message 4
Message5
Print message 5
Query
XR[ p1]?
Example
Assign the printout of message 2 to the remote
control input of terminal number 1.
XR 1,Message2
Description Math start/stop and Math reset can be specified
on models with the computation function (/M1
option).
YS
Sets the RS-422A/485 interface
(/C3 option).
Syntax
YS p1,p2,p3,p4,p5
p1
Address (1 to 32)
p2
Baud rate (1200, 2400, 4800, 9600, 19200,
38400)
p3
Data length (7, 8)
p4
Parity check (ODD, EVEN, NONE)
p5
Protocol (NORMAL, MODBUS)
4.5 Basic Setting Commands
Set the address to 2, the baud rate to 9600, the
data length to 8, the parity check to ODD, and
the NORMAL protocol.
YS 2,9600,8,ODD,NORMAL
Description •
The settings specified by this command and
saved using the XE command take effect after
the recorder is power cycled.
•
If p5 is set to MODBUS, Modbus slave is
selected.
XQ
Sets the TLOG timer
(/M1 option).
When not using the timer
Syntax
XQ p1,p2
p1
Timer number (1 or 2)
p2
Timer type (OFF)
Query
XQ[ p1]?
Example
Disable the number 1 timer.
XQ 1,OFF
When using an absolute timer
Syntax
XQ p1,p2,p3,p4,p5,p6
p1
Timer number (1 or 2)
p2
Timer type (ABSOLUTE)
p3
Interval (10min, 12min, 15min, 20min,
30min, 1h, 2h, 3h, 4h, 6h, 8h, 12h, 24h)
p4
Reference time (fixed to HH format)
HH
Hour (00 to 23)
p5
Reset/not reset the data when the timer
expires (ON/OFF)
p6
Printout ON/OFF (ON, OFF)
Query
XQ[ p1]?
Example
Set an absolute timer to timer number 1. Set the
interval to 30 minutes, the reference time to hour
7, reset the data when the timer expires, and
disable printout.
XQ 1,ABSOLUTE,30min,07,ON,OFF
Description Each time the interval specified by p3 elapses
from the time specified by p4, the operations set
with p5 and p6 are performed.
When using the relative timer
Syntax
XQ p1,p2,p3,p4,p5
p1
Timer number (1 or 2)
p2
Timer type (RELATIVE)
p3
Interval (fixed to HH:MM format)
HH
Hour (00 to 24)
MM
Minute (00 to 59)
Set the interval in the range 00:01 to 24:00
p4
Reset/not reset the data when the timer
expires (ON/OFF)
p5
Printout ON/OFF (ON, OFF)
Query
XQ[ p1]?
Example
Set a relative timer to timer number 1. Set the
interval to 1 hour 15 minutes, reset the data
when the timer expires, and disable printout.
XQ 1,RELATIVE,01:15,ON,OFF
Description Each time the interval specified by p3 elapses
from the time the computation is started, the
operations set with p4 and p5 are performed.
UN
Changes the assignment of
channels to the recording pen
(pen model, /M1 option).
Syntax
UN p1,p2
p1
Pen number (1 to 4)
p2
Channel number
Query
UN[ p1]?
Example
Assign computation channel 0A to pen number 4.
UN 4,0A
US
Sets the computation error
procedure (/M1 option).
Syntax
US p1,p2
p1
Procedure taken when an error occurs.
+OVER
Handle error data as +OVER
−OVER
Handle error data as –OVER
p2
Procedure taken when the “input over” is
detected (procedure taken when an “input
over” value is applied to TLOG.SUM or
TLOG.AVE computation)
SKIP
Do not use the data for
computation.
LIMIT
Use the limit value given by
the recorder for computation.
Query
US?
Example
When the computed result is in error, handle the
error data as +OVER. Use the limit value given
by the recorder for computation instead of the
“input over” value.
US +OVER,LIMIT
YB
Sets the host name and domain
name (/C7 option).
Syntax
YB p1,p2
p1
Host name (up to 64 characters)
p2
Domain name (up to 64 characters)
Query
YB?
Example
Set the host name to “ABC” and the domain
name to “recorder.co.jp.”
YB ABC,recorder.co.jp
Description The settings specified by this command and
saved using the XE command take effect after
the recorder is power cycled.
YA
Sets the IP address (/C7 option).
Syntax
YA p1,p2,p3
p1
IP address (0.0.0.0 to 255.255.255.255)
p2
Subnet mask
(0.0.0.0 to 255.255.255.255)
p3
Default gateway
(0.0.0.0 to 255.255.255.255)
Query
YA?
4.5 Basic Setting Commands
Set the IP address to 192.168.111.24, the subnet
mask to 255.255.255.0, and the default gateway
to 0.0.0.0.
YA 192.168.111.24,255.255.255.0,
0.0.0.0
Description The settings specified by this command and
saved using the XE command take effect after
the recorder is power cycled.
YN
Sets the DNS (/C7 option).
When not using the DNS
Syntax
YN p1
p1
Use/Not use the DNS (OFF)
Query
YN?
Example
Do not use the DNS.
YN OFF
When using the DNS
Syntax
YN p1,p2,p3,p4,p5
p1
Use/Not use the DNS (ON)
p2
Address of the primary DNS server
(0.0.0.0 to 255.255.255.255)
p3
Address of the secondary DNS server
(0.0.0.0 to 255.255.255.255)
p4
Domain suffix 1 (up to 64 characters)
p5
Domain suffix 2 (up to 64 characters)
Query
YN?
Example
Use the DNS server at 192.168.0.1.
YN ON,192.168.0.1
Description The settings specified by this command and
saved using the XE command take effect after
the recorder is power cycled.
YD
Sets whether to use the login
function via communication
(/C7 option).
Syntax
YD p1
p1
Use/Not use the login function via
communication (USE, NOT)
Query
YD?
Example
Use the login function via communication.
YD USE
Description The settings specified by this command and
saved using the XE command take effect after
the recorder is power cycled.
YQ
Sets the communication timeout
(/C7 option).
When not using the timeout
Syntax
YQ p1
p1
Enable/Disable communication timeout (OFF)
Query
YQ?
Example
Disable timeout.
YQ OFF
When using the timeout
Syntax
YQ p1,p2
p1
Enable/Disable communication timeout (ON)
p2
Timeout time (1 to 120) [minutes]
Query
YQ?
Example
Enable communication timeout and set the
timeout period to 3 min.
YQ ON,3
Description The settings specified by this command and
saved using the XE command take effect after
the recorder is power cycled.
YK
Sets keepalive (/C7 option).
Syntax
YK p1
p1
Enable/Disable keepalive (ON, OFF)
Query
YK?
Example
Disable keepalive.
YK OFF
Description The settings specified by this command and
saved using the XE command take effect after
the recorder is power cycled.
UA
Sets the record position.
Pen model
Syntax
UA p1,p2,p3
p1
Record position selection
ZERO
0% position
FULL
100% position
p2
Pen number (1 to 4)
p3
Record position adjustment value
When p1 = ZERO
RD100B: An integer
between 0 and 70
RD1800B: An integer
between 0 and 180
When p1 = FULL
RD100B: An integer
between –45 to 15
(reference value:
3026)
RD1800B: An integer
between –165 to 165
(reference value:
5447)
Example
Adjust the 0% position of pen 1. Set the record
position adjustment value to 20.
UA ZERO,1,20
Description •
Check the recorded result and correct the
record position adjustment value.
•
To end the adjustment, execute the UY0
command. If you do not end the adjustment,
you cannot change the execution mode.
•
The unit for p3 is 1/30 mm.
4.5 Basic Setting Commands
Record position selection
ZERO 0% position
FULL 100% position
Hysteresis
The difference in the record position
according to the operating direction of
the printer carriage
p2
Record position adjustment value
When p1 = Hysteresis
An integer between –7 to 7
When p1 = ZERO
RD100B: An integer between 0 and 15
RD1800B: An integer between 0 and 50
When p1 = FULL
RD100B: An integer between –30 to 30
(reference value: 1000)
RD1800B: An integer between –50 to 50
(reference value: 1800)
Example
Adjust the 100% position. Set the record
position adjustment value to 25.
UA FULL,25
Description •
Carry out the adjustment in the following
order: P1 = Hysteresis, ZERO, and then
FULL.
•
Check the recorded result and correct the
record position adjustment value.
•
To end the adjustment, execute the UY0
command. If you do not end the adjustment,
you cannot change the execution mode.
•
The unit for p2 is dot (0.1 mm).
YE
Exits from Basic Setting mode.
When the YE command is executed, communication is
dropped.
Syntax
YE p1
p1
Apply/Discard the settings
STORE
Apply the settings
ABORT
Discard the settings
Example
Apply the basic settings.
YE STORE
Description Applies the settings changed with the basic
setting commands.
XE
Exits from Basic Setting mode.
Communication is not dropped when the XE command is
executed.
Syntax
XE p1
p1
Apply/Discard the settings
STORE
Apply the settings
ABORT
Discard the settings
Example
Apply the basic settings.
XE STORE
Description The settings specified by the following
commands and saved using the XE command
take effect after the recorder is power cycled.
YS, YB, YA, YN, YD, YQ, and YK
4.5 Basic Setting Commands
Switches the execution mode.
Syntax
DS p1
p1
Mode
0 Run mode
1 Basic Setting mode
Example
Switch to Basic Setting mode.
DS 1
Description The execution mode cannot be changed to Basic
Setting Mode while recording or computation is
in progress.
PS
Starts/Stops recording.
Syntax
PS p1
p1
Start/Stop recording.
0 Start
1 Stop
Example
Start recording.
PS 0
UD
Switches the screen/switches
the channel.
Returning to the data display screen
Returns from the Setting mode screen, the screen that
appears when the FUNC key is pressed, or the screen that
appears when the DISP MENU key is pressed back to the
measurement data display screen.
Syntax
UD p1
p1
Fixed to 0.
Example
Return to the data display screen.
UD 0
Displaying the specified data display screen
Syntax
UD p1,p2
p1
Fixed to 1.
p2
Screen number (1 to 15)
Example
Display data display screen 2.
UD 1,2
Switching the display channel (manual switching)
Syntax
UD p1
p1
Fixed to 2.
Example
Switch the display channel.
UD 2
Description •
An error results if the specified screen is set to
SKIP for UD1.
•
UD2 is valid on screens whose display
channel is set to manual switching.
AK
Executes alarm acknowledge
(Alarm ACK).
Syntax
AK p1
p1
Executes alarm acknowledge (0)
Example
Execute alarm acknowledge.
AK 0
TL
Starts/stops/resets computation
(/M1 option).
Syntax
TL p1
p1
Operation type
0 Computation start
1 Computation stop
2 Computation reset
Example
Start the computation.
TL 0
MP
Starts/Stops manual print.
Syntax
MP p1
p1
Operation type
0 Manual print start
1 Manual print stop
Example
Start manual print.
MP 0
LS
Starts/Stops the list (setting
information) printout.
Syntax
LS p1
p1
Operation type
0 List print start
1 List print stop
Example
Start list print.
LS 0
Description List print refers to printing of settings of Setting
mode.
SU
Starts/Stops the setup list (basic
setting information) printout.
Syntax
SU p1
p1
Operation type
0 Setup list print start
1 Setup list print stop
Example
Start setup list print.
SU 0
Description Setup list print refers to printing of settings of
Basic Setting mode.
MS
Prints the message.
Syntax
MS p1
p1
Message number (1 to 5)
Example
Print the message of message number 3.
MS 3
Description The message string is set with the SG command.
AC
Clears the alarm printout buffer.
Syntax
AC p1
p1
Clear the alarm printout buffer (0)
Example
Clear the alarm printout buffer.
AC 0
4.6 Control Commands
Clears the message printout
buffer.
Syntax
MC p1
p1
Clear the message printout buffer (0)
Example
Clear the message printout buffer.
MC 0
VG
Resets the report data of the
periodic printout.
Syntax
VG p1
p1
Operation type
2 Reset the report data of the periodic
printout.
Example
Reset the report data of the periodic printout.
VG 2
Description This setting is valid when the recorder is set to
print the report data using periodic printout.
YC
Initializes settings.
Syntax
YC p1
p1
Initialization type
0
Initialize the Setting mode and Basic
Setting mode settings.
1
Initialize the Setting mode settings.
Example
Initialize the Setting mode and Basic Setting
mode settings.
YC 0
UY
Stops the record position
adjustment.
Syntax
UY p1
p1
Fixed to 0.
Query
UY?
Example
Stop the record position adjustment.
UY 0
Description Returns the execution status of the record
position adjustment in response to a query.
0: Stopped, 1: In progress
4.7
Output Commands
BO
Sets the byte output order.
Syntax
BO p1
p1
Byte order
0 Outputs the data MSB first.
1 Outputs the data LSB first.
Query
BO?
Example
Output the data MSB first.
BO 0
Description This command applies to the byte order of
numerical data during BINARY output.
CS
Sets the checksum (/C3 option).
Syntax
CS p1
p1
Use/Not use the checksum
0 Not use
1 Use
Query
CS?
Example
Use the checksum.
CS 1
Description This command can be used only on the RS-
422A/485 communication interface.
IF
Sets the status filter.
Syntax
IF p1
p1
Status filter value
(0.0.0.0 to 255.255.255.255)
Query
IF?
Example
Set the status filter value to 1.0.4.0.
IF 1.0.4.0
Description For details, see chapter 6.
Note
Initialization of BO/CS/IF Command Settings
•
For serial communications
Settings entered using the BO/CS/IF commands
revert to their initial values when the recorder is reset
(when the recorder is power cycled, or the user exits
the basic setting mode with the YE command).
•
Byte output order, checksum: 0
•
Status filter: 255.255.255.255
If the recorder is reset, you must restore these
settings.
•
For Ethernet communications
Settings entered using the BO/IF commands revert to
their initial values when the connection to the recorder
is cut. After reconnecting the recorder, you must
reenter the settings.
CC
Disconnects an Ethernet
connection (/C7 option).
Syntax
CC p1
p1
Disconnect the connection (0)
4.6 Control Commands/4.7 Output Commands
Disconnect the connection.
CC 0
FE
Outputs decimal point position,
unit information, and setup data.
Syntax
FE p1,p2,p3
p1
Output data type
0 Setup data of Setting mode
1 Decimal point position and unit
information
2 Setup data of Basic Setting mode
4 Setting data file
p2
First output channel number
p3
Last output channel number
Example
Output the setup data of the Setting commands
of channel 01 through 04.
FE 0,01,04
Description •
Be sure to set p2 and p3 so that p3 is greater
than or equal to p2.
•
The settings of p2 and p3 are valid when p1 =
0, 1, or 2.
•
The setting data file can be viewed using the
configuration software.
FD
Outputs the most recent
measured/computed data.
Syntax
FD p1,p2,p3
p1
Output data type
0
Output the most recent measured/
computed data in ASCII format
1
Output the most recent measured/
computed data in BINARY format
p2
First output channel number
p3
Last output channel number
Example
Output the most recent measured/computed data
from channel 01 to 04 in ASCII format.
FD 0,01,04
Description •
The most recent measured/computed data
indicates the most recent measured/computed
data when the recorder receives the FD
command.
•
Be sure to set p2 and p3 so that p3 is greater
than or equal to p2.
FY
Outputs the statistical
computation results.
Syntax
FY p1,p2,p3
p1
Output data type
Inst
Output the most recent periodic
printout data (instantaneous
value)
Report
Output the statistical calculation
data of periodic printout (report
data)
Tlog1
Output the data at the most
recent timeout of TLOG timer 1
Tlog2
Output the data at the most
recent timeout of TLOG timer 2
p2
First output channel number
p3
Last output channel number
Example
Outputs the most recent periodic printout data of
channel 01 to 04.
FY Inst,01,04
Description Tlog1 and Tlog2 are valid on models with the
computation function (/M1 option).
FF
Outputs the FIFO data.
Syntax
FF p1,p2,p3,p4
p1
Operation type
GET
Output the data starting from the
next to the previous read position
RESEND
Retransmit the previous output
RESET
Set the read position (block) to
the most recent acquire position
(block)
GETNEW
Output the most recent data
p2
First output channel number
p3
Last output channel number
p4
The upper limit of number of blocks that are
to be loaded
Pen model
1 to 240
Dot model
1 to 60
If the measured/computed data is less than
the specified number of blocks, only the
available data is transmitted.
Example
Output two blocks of FIFO data from channels 01
to 06.
FF GET ,01,06,2
Description •
The FIFO buffer is of a circular type which
overwrites from the oldest data when it is full.
•
The FR command is used to set the acquiring
interval.
•
There are two FIFO data output formats.
Logging output (GET)
Output the specified number of blocks (p4) of
FIFO data starting from the next to the
previous read position (block).
Make sure to read the data within the
following buffer period to prevent data
dropouts.
Pen model
FIFO buffer length
240 blocks
Maximum buffer period 240 intervals
Dot model
FIFO buffer length
60 blocks
Maximum buffer period 60 intervals
4.7 Output Commands
Newest value output (GETNEW)
Output the specified number of blocks (p4) of
FIFO data back starting from the recent
acquire position (block).
•
Parameters p2, p3, and p4 are valid when p1
is set to GET or GETNEW.
•
If p4 is omitted, all the data of all blocks
acquired in the FIFO buffer are output.
•
Be sure to set p2 and p3 so that p3 is greater
than or equal to p2.
•
For the output flow of FIFO data, see
appendix 4.
IS
Outputs status information.
Syntax
IS p1
p1
Output status information (0)
Example
Output status information.
IS 0
Description The output status can be masked using the
status filter (IF command).
FU
Outputs user information.
Syntax
FU p1
p1
Output user information (0)
Example
Output user information.
FU 0
Description Outputs the information of the user currently
connected to the recorder.
4.8
RS-422A/485 Dedicated
Commands
ESC O
Opens the instrument.
The ASCII code of ESC is 1BH.
Syntax
ESC O p1
p1
Instrument address (01 to 32)
Example
Open the instrument at address 01.
ESC O 01
Description •
Specifies the address of the instrument with
which to communicate.
•
Only one instrument can be opened at any
given time.
•
When an instrument is opened with the ESC
O command, any other instrument that is
currently open is automatically closed.
•
When this command is received correctly, the
recorder transmits the data “ESC 0 xx” (xx:
Instrument’s address).
•
Use CR+LF for the terminator. LF cannot be
used.
ESC C
Closes the instrument.
The ASCII code of ESC is 1BH.
Syntax
ESC C p1
p1
Instrument address (01 to 32)
Example
Close the instrument at address 01.
ESC C 01
Description •
Clears the current connection with the
instrument.
•
When this command is received correctly, the
recorder transmits the data “ESC C xx” (xx:
Instrument’s address).
•
Use CR+LF for the terminator. LF cannot be
used.
4.7 Output Commands/4.8 RS-422A/485 Dedicated Commands
Commands (Available when
using the maintenance/test
server function via
Ethernet communications)
close
Disconnects the connection
between other instruments.
Syntax
close,p1,p2:p3
p1
Port on the recorder side (0 to 65535)
p2
IP address on the PC side
(0.0.0.0 to 255.255.255.255)
p3
Port on the PC side (0 to 65535)
Example
close,34159,192.168.111.24:1054
E0
Description This command cannot be used to disconnect a
server port. Also, it cannot disconnect the
recorder being operated. Use the quit command
for this purpose.
con
Outputs connection information.
Syntax
con
Example
con
EA
04/10/01 12:34:56
Active connections
Proto Local Address
Foreign Address
State
TCP
192.168.111. 24:34159 192.168.111. 24:1053
ESTABLISHED
TCP
0. 0. 0. 0:34155
0. 0. 0. 0: 0
LISTEN
TCP
0. 0. 0. 0:34159
0. 0. 0. 0: 0
LISTEN
TCP
0. 0. 0. 0:34150
0. 0. 0. 0: 0
LISTEN
EN
TCP
Protocol used.
Local Address
The recorder’s socket address.
Displays “IP address:port number.”
Foreign Address
The destination socket address.
Displays “IP address:port number.”
State
Connection status.
ESTABLISHED
Connection established.
4.9 Maintenance/Test Commands
eth
Outputs Ethernet statistical
information.
Syntax
eth
Example
eth
EA
04/10/01 12:34:56
Ethernet Statistics
Name In Pkt In Err Out Pkt
Out Err
16 Coll
lo0
0
0
0
0
0
mb0
74
0
64
0
0
EN
help
Outputs help.
Syntax
help [,p1]
p1
Command name
(close, con, eth, help, net, quit)
Example
help
EA
con
- echo connection information
eth
- echo ethernet information
help
- echo help
net
- echo network status
quit
- close this connection
EN
net
Outputs network statistical
information.
Syntax
net
Example
net
EA
04/10/01 12:34:56
Network Status
APP: power on time = 00/00/00 12:34:56
APP: applalive = disable
APP: genedrops = 0
APP: diagdrops = 0
APP: ftpsdrops = 0
TCP: keepalive = 30 s
TCP: connects = 14
TCP: closed = 0
TCP: timeoutdrop = 0
TCP: keepdrops = 0
TCP: sndtotal = 53
TCP: sndbyte = 0
TCP: sndrexmitpack = 0
TCP: sndrexmitbyte = 1
TCP: rcvtotal = 0
TCP: rcvbyte = 0
DLC: 16 collisions = 0
EN
Keepalive check cycle.
TCP: connects
Total number of connections established.
TCP: closed
Total number of dropped connections.
TCP: timeoutdrop
Total number of dropped connections due to
TCP retransmission timeout. When the
transmitted packet (the unit of transmitted
data) is not received, the packet is
automatically retransmitted at a
predetermined time interval. If the packet is
not received after 14 retransmissions,
timeout occurs and the connection is
dropped.
TCP: keepdrops
Total number of dropped connections due to
TCP keepalive timeout.
TCP: sndtotal
Total number of transmitted packets.
TCP: sndbyte
Total number of transmitted bytes.
TCP: sndrexmitpack
Total number of retransmitted packets.
TCP: sndrexmitbyte
Total number of retransmitted bytes.
TCP: rcvtotal
Total number of received packets.
TCP: rcvbyte
Total number of received bytes.
DLC: 16 collisions
Number of collision incidents. A collision
occurs when two or more instruments on the
network attempt to transmit simultaneously.
The tendency for collisions to occur
increases when the network is congested.
16 collisions would mean 16 consecutive
collision incidents.
quit
Disconnects the connection of
the instrument being operated
Syntax
quit
4.10 Instrument Information
Output Commands
(Available when using the
instrument information
server function via
Ethernet communications)
The instrument information server function
interprets one UDP packet to be one command
and returns a single packet (containing the
recorder’s information) in response to the
command.
Port number
34264/udp
(see section 2.1)
Transfer data
ASCII
Received buffer length
128
Transmit buffer length
512
Maximum number of parameters
32
In the command packet, parameters
corresponding to the desired information are
placed one after another.
Parameter Description
serial
Outputs the serial number.
host
Outputs the host name (the host name specified
in section 2.3).
ip
Outputs the IP address (the IP address specified
in section 2.3).
Example
Query the IP address and host name. (Of the
two frames below, the top frame represents the
command packet, the bottom frame represents
the response packet.)
ip host
EA
ip = 192.168.111.24
host = ABC
EN
Description •
Separate each parameter with one or more
blanks (space, tab, carriage return, line feed).
•
Parameters are not case sensitive.
•
Undefined parameters are ignored.
•
Parameters beyond the 32nd parameter are
ignored.
4.10 Instrument Information Output Commands (Available when using the instrument information server
function via Ethernet communications)
The following table shows the types of responses for various commands described in the
previous chapter.
The recorder returns a response (affirmative/negative response) to a command that is
separated by a single terminator. The controller should follow the one command to one
response format.
Command
Function
See section 4.9.
See section 4.10.
Setting/Measurement
server
Maintenance/Test Server
Instrument Information server
Response
Command Type
Affirmation
Negation
Setting command
Basic setting command
Control command
Output
command
RS-422A/485 dedicated Dedicated response No response
Setup, measurement,
and computation data
output
ASCII output
BINARY output
Affirmative
response
Single
negative
response
or
Multiple
negative
responses
Note
The “CRLF” used in this section denotes a terminator.
Affirmative Response
When the command is processed correctly, an affirmative response is returned.
• Syntax
E0CRLF
• Example
E0
Single Negative Response
When the command is not processed correctly, a single negative response is returned.
• Syntax
E1_nnn_mmm•••mCRLF
nnn
Error number (001 to 999)
mmm•••m
Message (variable length, one line)
_
Space
• Example
E1 001 "System error"
Multiple Negative Responses
•
If there is an error in any one of the multiple commands that are separated by sub
delimiters, multiple negative responses are returned.
•
The response is generated for each erroneous command.
•
If there are multiple commands that have errors, the negative responses are
separated by commas.
•
The error position number is assigned to the series of commands in order starting with
“1” assigned to the first command.
• Syntax
E2_ee:nnnCRLF
(When there is only one error)
E2_ee:nnn,ee:nnn,•••,ee:nnnCRLF
(When there are multiple errors)
ee
Error position (01 to 10)
nnn
Error number (001 to 999)
_
Space
Chapter 5 Responses
The following types of ASCII data are available. For the data formats, see section 5.2.
Setting data, basic setting data, decimal position/unit information, measured/
computed data, report data generated by the periodic printout, status information, and
user information
• Syntax
EACRLF
•••••••••••••••CRLF
:
•••••••••••••••CRLF
•••••••••••••••CRLF
ENCRLF
BINARY Output
• Conceptual Diagram
'E'
'B'
CR
LF
BINARY data
ASCII
BINARY
Data length
Data sum
Flag
ID
Header sum
1 byte
BINARY header
(12 bytes)
BINARY data
BINARY footer
(2 bytes)
• EBCRLF
Indicates that the data is BINARY.
• Data Length
The byte value of “flag + identifier + header sum + BINARY data + data sum.”
• Header Sum
The sum value of “data length + flag + identifier.”
• BINARY Data
For the output format of various data types, see section 5.3.
• Data Sum
The sum value of “BINARY data.”
Note
The data length of the BINARY header section is output according to the byte order specified
with the BO command.
5.1 Response Syntax
Meaning of the Flag
0
1
7
BO
MSB
LSB
Output byte order
6
CS
No
Yes
Existence of a checksum
5
–
–
–
4
–
–
–
3
–
–
–
2
–
–
–
1
–
–
–
0
Reserved
–
–
Fixed to 1.
•
When the BO flag is “0,” the MSB is output first. When the BO flag is “1,” the LSB
is output first.
•
If the check sum is enabled (parameter = 1) using the CS command parameter,
each sum value is inserted in the header sum and data sum sections in the
“Conceptual Diagram” on the previous page. If the check sum is disabled
(parameter = 0), a zero is inserted in the header sum and data sum sections. For a
sample program that calculates the sum value, see “Calculating the Sum Value” on
the next page.
•
The bits that have “–” for the flag and flag are not used. The value is undefined.
• Identifier
ID Number
Binary Data Type
Format
0
Undefined file
–
1
Measured/computed data
Section 5.3
1
FIFO data
Section 5.3
10
Setup data file
Undisclosed
•
The table above shows the different types of BINARY Data.
•
Measured/computed data can be output using the FD command.
•
FIFO data can be output using the FF command.
•
The setup data file can be output using the FE command. The setup data file can
be loaded in the cofiguration software.
•
The identifier section in the “Conceptual diagram” on the previous page contains
the ID number shown above.
Note
BINARY data that are not indicated in the above table are considered undefined files.
• Calculating the Sum Value
If you set the parameter of the CS command to “1 (enabled),” the checksum value is
output only during serial communications. The check sum is the same as that used in
the TCP/IP and is derived according to the following algorithm.
Buffer on Which the Sum Value Is Calculated
•
For the header sum, it is calculated from “data length + flag + identifier” (fixed to 6
bytes).
•
For the data sum, it is calculated from “BINARY data.”
1 byte
0
(1)
(2)
(3)
(4)
(6)
(5)
Padding
5.1 Response Syntax
If the data length of the buffer is odd, a “0” is padded so that it is even. (1) through (6)
are summed as unsigned two-byte integers (unsigned short). If the digit overflows a
“1” is added. Finally, the result is bit-wise inverted.
Sample Program
The sum value is determined using the following sample program, and the calculated
result is returned. The sum determined by the sample program can be compared with
the header sum of the output BINARY header section and the data sum of the output
BINARY footer section.
/*
* Sum Calculation Function (for a 32-bit CPU)
*
* Parameters buff: Pointer to the top of the data on which the sum is calculated
*
len: Length of the data on which the sum is calculated
* Return value:
Calculated sum
*/
intcksum(unsigned char *buff, int len)
{
unsigned short *p;
/* Pointer to the next two-byte data word in the buffer
that is to be summed. */
unsigned int csum;
/* Checksum value */
int
i;
int
odd;
csum = 0;
/* Initialize. */
odd = len%2;
/* Check whether or not the number of data points is
even. */
len >>= 1;
/* Determine the number of data points using a
“short” data type. */
p = (unsigned short *)buff;
for(i=0;i>16) & 0xffff)) > 0xffff)
/* Add the overflowed digits */
csum = csum - 0xffff;/* If the digit overflows again, add a 1. */
return((~csum) & 0xffff); /* bit inversion */
}
5.1 Response Syntax
RS-422A/485 Dedicated Commands and Responses
The following table shows dedicated commands for the RS-422A/485 interface and their
responses.
Command Syntax
Meaning
Response
ESC O_xx CRLF
Open the instrument.
• Response from the instrument with the
specified address
ESC O xx CRLF
(_ space)
• Response when the instrument with the
specified address does not exist*
ESC C_xx CRLF
Close the instrument.
• Response from the instrument with the
specified address
ESC C xx CRLF
(_ space)
• Response when the instrument with the
specified address does not exist*
*
The causes that the condition become “The instrument with the specified address does not
exist” is such as a command error, the address not matching that of the instrument, the
instrument is not turned ON, and the instrument not being connected via the serial interface.
•
The “xx” in the table indicates the instrument’s address. Specify the address that
is assigned to the instrument from 01 to 32.
•
Only one instrument can be opened at any one time.
•
When an instrument is opened with the ESC O command, all commands on the
instrument become active.
•
When an instrument is opened with the ESC O command, any other instrument
that is open is automatically closed.
•
Use CR+LF for the terminator. LF cannot be used.
Note
• The ASCII code of ESC is 1BH.
5.1 Response Syntax
Output Format of ASCII Data
The following types of ASCII data are available.
•
Setting data/Basic Setting data
•
Decimal point position/unit information
•
Measured/computed data
•
Report data generated by the periodic printout
•
Status information
•
User information
Note
The “CRLF” used in this section denotes a terminator.
Setting/Basic Setting data
•
The FE command (FE0 or FE2) is used to output the data.
•
The setting/basic setting data are output in the order of the listed commands in the
table in section 4.2, “A List of Commands.” However, the setting information for the
following commands is not output.
•
Setting commands
SD, CM, and FR
•
Basic Setting commands
YE and XE
•
The output format of the setting/basic setting data is the same as the syntax of each
command.
•
Some commands are output in multiple lines. (Example: Commands that are
specified for each channel.)
• Syntax
The two-character command name and the succeeding parameters are output in the
following syntax.
EACRLF
ttsss•••sCRLF
••••••••••••••
ENCRLF
tt
Command name (SR, SA•••, XA, XI•••)
sss•••s Setting, basic setting data (variable length, one line)
• Example
EA
SR01,VOLT,20mV,0,20
SR02,VOLT,20mV,0,20
•••••••••••••••••••
EN
Decimal Point Position/Unit Information
•
The FE command (FE1) is used to output the data.
• Syntax
The data is output for each channel in the following syntax.
EACRLF
s_kccuuuuuu,ppCRLF
••••••••••••••••••
ENCRLF
s
Data status (N, D, S)
N: Normal
D: Differential input
S: Skip (When the input range is set to SKIP for a measurement
channel or when the channel is turned OFF for a computation
channel)
k
Channel type
0: Measurement channel
A: Computation channel
cc
Channel number
RD100B: 01 to 06, 0A to 0P
RD1800B: 01 to 24, 0A to 1P
uuuuuu
Unit information (6 characters, left-justified)
mV____: mV
V_____: V
^C____: °C
xxxxxx: (User-defined character string)
pp
Decimal point position (00 to 04)
No decimal (00000) for 00.
One digit to the right of the decimal (0000.0) for 01.
Two digits to the right of the decimal (000.00) for 02.
Three digits to the right of the decimal (00.000) for 03.
Four digits to the right of the decimal (0.0000) for 04.
_
Space
• Example
EA
N 001mV ,01
N 002mV ,01
EN
5.2 Output Format of ASCII Data
The FD (FD0) or FY (FYInst, FYTlog1, or FYTlog2) command is used to output the
data.
• Syntax
The measured/computed data are output in the following syntax along with the date
and time information for each channel
EACRLF
DATE_yy/mo/ddCRLF
TIME_hh:mi:ss.mmmt_S1S2S3S4S5S6CRLF
s_kcca1a2a3a4uuuuuuf1dddddEf2ppCRLF
••••••••••••••••••••••••••••
ENCRLF
yy
Year (00 to 99)
mo
Month (01 to 12)
dd
Day (01 to 31)
hh
Hour (00 to 23)
mi
Minute (00 to 59)
ss
Second (00 to 59)
mmm
Millisecond (000 to 999. A period is placed between seconds
and milliseconds.)
t
‘S’(=summer) or ‘ ’ space(=winter)
S1S2S3S4S5S6
Data status
Takes on the values below for the FYTlog1 and FYTlog2
commands. For all other cases, they are all spaces.
S1
Time change during TLOG computation: T (occurred) or space
(not occurred)
S2
Power OFF and ON during TLOG computation: P (occurred) or
space (not occurred)
S3
Data reset during the TLOG computation: R (occurred) or
space (not occurred)
S4S5S6
All spaces
s
Channel data status (N, D, S, O, B, E)
N: Normal
D: Differential input
S: Skip
O: Over
B: Burnout
E: Error
k
Channel type
0: Measurement channel
A: Computation channel
cc
Channel number
RD100B: 01 to 06, 0A to 0P
RD1800B: 01 to 24, 0A to 1P
a1a2a3a4
a1
Alarm status (level 1)
a2
Alarm status (level 2)
a3
Alarm status (level 3)
a4
Alarm status (level 4)
5.2 Output Format of ASCII Data
Each status is set to H, L, h, l, R, r, T, t, or space.
H: high limit alarm, L: low limit alarm, h: difference high-limit
alarm, l: difference low-limit alarm, R: high limit on rate-of-
change alarm, r: low limit on rate-of-change alarm, T: delay
high limit alarm, t: delay low limit alarm, space: no alarm
uuuuuu
Unit information (6 characters, left-justified)
mV____: mV
V_____: V
^C____: °C
xxxxxx: (User-defined character string)
f1
Sign (+, –) of mantissa
•
Positive over data, error data, and the burnout data when
“up” is specified are positive.
•
Negative over data and the burnout data when “down” is
specified are negaitive.
ddddd
Mantissa (00000 to 99999, 5 digits)
•
8 digits for computed data.
•
For error data (channel data status is E), over data (channel
data status is O), or burnout data (channel status data is B),
the mantissa is set to 99999 (99999999 for computed data).
f2
Sign (+, –) of exponent
pp
Exponent (00 to 04)
_
Space
• Example
EA
DATE 99/02/23
TIME 19:56:32.500
N 001h mV +12345E-03
N 002 mV -12345E-01
S 003
EN
Note
• Data for non-existing channels are not output (not even the channel number).
• For channels set to skip, output values from alarm status to exponent are spaces.
5.2 Output Format of ASCII Data
Report Data Generated by the Periodic Printout
•
The FY command (FYREPORT) is used to output the data.
•
Report data generated by the periodic printout is output.
Syntax
EACRLF
YY/MO/DD_HH:MI:SS.MMMTCRLF
yy/mo/dd_hh:mi:ss.mmmt_S1S2S3S4S5S6CRLF
s1s2s3s4s5s6kccuuuuuuf1dddddEf2pp_f1dddddEf2pp_f1dddddEf2pp_f1ddddd
Ef2pp_f1ddddddddEf2ppCRLF
••••••••••••••••••
ENCRLF
YY/MO/DD HH:MI:SS.MMMT
Report start time information
yy/mo/dd hh:mi:ss.mmmt
Report end time information
YY, yy
Year (00 to 99)
MO, mo
Month (01 to 12)
DD, dd
Day (01 to 31)
HH, hh
Hour (00 to 23)
MI, mi
Minute (00 to 59)
SS, ss
Second (00 to 59)
MMM, mmm
Millisecond (000 to 999)
T, t
‘S’(=summer) or ‘ ’ space(=winter)
S1S2S3S4S5S6
Data status
S1
Time change during reporting: T (occurred) or space
(not occurred)
S2
Power OFF and ON during reporting: P (occurred) or
space (not occurred)
S3
Data clear during reporting: R (occurred) or space (not
occurred)
S4S5S6
All spaces
s1s2s3s4s5s6
Channel data status
s1
Channel mode at the end of reporting: S (Skipped) or space
(Mode other than skipped)
s2
Range change during reporting: C (occurred) or space (not
occurred)
s3
Error data occurrence during reporting: E (yes) or space (no)
s4
±over data occurrence during reporting: O (yes) or space (no)
s5s6
All spaces
cc
Channel number
RD100B: 01 to 06, 0A to 0P
RD1800B: 01 to 24, 0A to 1P
k
Channel type (0, A)
0: Measurement channel
A: Computation channel
uuuuuu
Unit information (6 characters)
5.2 Output Format of ASCII Data
The data is output in the following order: most recent value,
minimum value, maximum value, average value, and sum
value.
f1dddddEf2pp Most recent value, minimum value, maximum value, and
average value of the measurement channel
f1ddddddddEf2pp
Sum value of the measurement channel or the most recent
value, minimum value, maximum value, average value, and
sum value of the computation channel
f1
Sign (+, –) of mantissa
ddddd
Mantissa (00000 to 99999)
dddddddd Mantissa (00000000 to 99999999)
f2
Sign (+, –) of exponent
pp
Exponent (00 to 04)
• Example
EA
04/08/04 10:22:20.500S
04/08/04 19:56:32.500S TP
001mV +12345E-03 +12345E-03 +12345E-03 +12345E-03 +12345678E-03
C 002mV -12345E-01 -12345E-01 -12345E-01 -12345E-01 -12345678E-01
S 003
S 004
A0A +12345678E-03 +12345678E-03 +12345678E-03 +12345678E-03
+12345678E-03
A0B -12345678E-01 -12345678E-01 -12345678E-01 -12345678E-01 -
12345678E-01
S A0C
S A0D
EN
5.2 Output Format of ASCII Data
The IS command is used to output the data.
•
The operation status of the recorder is output
•
For details on the status information, see section 6.2, “The Bit Structure of the Status
Information.”
• Syntax
EACRLF
ddd.ccc.bbb.aaaCRLF
ENCRLF
aaa
Status information 1 (000 to 255)
bbb
Status information 2 (000 to 255)
ccc
Status information 3 (000 to 255)
ddd
Status information 4 (000 to 255)
• Example
EA
000.000.032.000
EN
User Information
•
The FU command is used to output the data.
•
User name, user level, and other information are output.
• Syntax
EACRLF
p_l_uuu•••uCRLF
ENCRLF
p
Physical layer
E: Ethernet
S: RS-422A/485
l
User level
A: Administrator
U: User
uuu•••u
User name (up to 16 characters)
_
Space
• Example
EA
E A admin
EN
5.2 Output Format of ASCII Data
Output Format of BINARY Data
This section describes the output format of the BINARY data that is disclosed. For the
BINARY output format, see “BINARY Output” on page 5-2. For other BINARY data
types, see “Identifier” on page 5-3.
The measured data is output using signed 16-bit integer; the computed data is output
using signed 32-bit integer. These integers can be understood as physical values by
adding the decimal point and the unit.
Typical Examples to Obtain Physical Values
BINARY Value
Decimal Position Code
Physical Value (Measured Value)
10000
0
10000
10000
1
1000.0
10000
2
100.00
10000
3
10.000
10000
4
1.0000
Measured/Computed Data and FIFO Data
•
The FD (FD1) command is used to output the most recent measured/computed data.
•
The FF (FEGET, FERESEND, and FEGETNEW) command is used to output the
FIFO data. The decimal point position and unit can be determined using the FE
command.
•
The ID number of the output format is 1. See “Identifier” on page 5-3.
2 bytes
Number of bytes
2 bytes
BINARY data
(The BINARY data section
on the “Conceptual
diagram” on page 5-2.)
Number of blocks
Block 1
Block n
...
• Number of Blocks
This is the number of blocks.
• Number of Bytes
This is the size of one block in bytes.
• Block
Year
Measured/
Computed
...
...
Measured/
Computed
...
...
Month
Channel
...
...
Channel
...
...
Day
A2A1
...
...
A2A1
...
...
Hour
A4A3
...
...
A4A3
...
...
Minute
Millisecond
(Reserved)*
Flag
Second
Measured data
...
...
Computed data
...
...
1 byte
1 byte
1 byte
1 byte
1 byte
1 byte
2 bytes
1 byte
1 byte
4 bytes
*
The sections indicated as (Reserved) are not used. The value is undefined.
The meanings of the flags are given on the table below. The flags are valid during
FIFO data output. The flags are undefined for other cases.
Bit
Flag
Flag
Meaning of the Flag
0
1
6
–
–
5
–
–
4
–
–
3
–
–
2
No
Yes
Indicates that the decimal position or unit information was changed
during measurement.
1
No
Yes
Indicates that the FIFO acquiring interval was changed during
measurement.
0
No
Yes
Indicates that the internal process took too much time (computation,
for example) and that FIFO dropout occurred.
The bits that have “–” for the flag column are not used. The value is undefined.
• Block Member
Name
BINARY Value
Year
0 to 99
Month
1 to 12
Day
1 to 31
Hour
0 to 23
Minute
0 to 59
Second
0 to 59
Millisecond
0 to 999
DST
1 (=summer) or 0 (=winter)
Measurement, computation
00H: measurement, 80H: computation
Channel
RD100B: 01 to 06 and 31 to 42
RD1800B: 01 to 24, 31 to 54
Alarm status*
A1 (Bit 0 to 3)
A2 (Bit 4 to 7)
0 to 8
A3 (Bit 0 to 3)
A4 (Bit 4 to 7)
Measured data
0 to FFFFH
Computed data
0 to FFFFFFFFH
*
BINARY value 0 to 8 is entered in the upper and lower 4 bits of a byte (8 bits) for the alarm
status. The binary values 0 to 8 correspond to H (high limit alarm), L (low limit alarm), h
(difference high-limit alarm), l (difference low-limit alarm), R (high limit on rate-of-change
alarm), r (low limit on rate-of-change alarm), T (delay high limit alarm), and t (delay low limit
alarm) as follows:
0: no alarm, 1: H, 2: L, 3: h, 4: l, 5: R, 6: r, 7: T, and 8: t.
• Special Data Value
The measured/computed data take on the following values under special conditions.
Special Data Value
Measured Data
Computed Data
+ Over
7FFFH
7FFF7FFFH
– Over
8001H
80018001H
Skip
8002H
80028002H
Burnout (when “up” is set)
7FFAH
7FFF7FFFH
Burnout (when “down” is set)
8006H
80018001H
Error
8004H
80048004H
Undefined
8005H
80058005H
Note
The number of blocks, number of bytes, and measured/computed data are output according to
the byte order specified with the BO command.
5.3 Output Format of BINARY Data
Status Information and Filter
The following figure illustrates the status information and filter on the recorder.
Status information
Filter
Condition register
4 bytes
4
3
2
1
4
3
2
1
•
When a status indicated on the following page is entered, the corresponding bit in the
condition register is set to “1.” The logical AND of the condition register and the filter
becomes the status information.
•
The IF command can be used to set the filter.
•
The IS command is used to output the status information. Status information 1 and 2
are cleared when they are output. Status information 3 and 4 are not cleared when it
is output, and remains at “1” while the event is occurring.
•
When multiple connections are up, filters can be specified for the individual
connection. Therefore, the status information can be held for each connection.
Chapter 6 Status Information
The Bit Structure of the Status Information
The following four groups of status information are output in response to a status
information output request using the IS command. For the output format, see “Status
Information” in section 5.2, “Output Format of ASCII Data.”
Status Information 1
Bit
Name
Description
0
A/D conversion complete
Set to “1” when the A/D conversion of the
measurement is complete.
1
–
–
2
Periodic printout timeout
Set to “1” when the periodic printout timer expires.
3
TLOG timeout
Set to “1” when the TLOG timer expires.
4
–
–
5
–
–
6
–
–
7
–
–
Status Information 2
Bit
Name
Description
0
Measurement drop
Set to “1” when the measurement process could not
keep up.
1
Decimal point/unit
Set to “1” when the decimal point/unit information is
information change
changed.
2
Command error
Set to “1” when there is a command syntax error.
3
Execution error
Set to “1” when an error occurs during command
execution.
4
–
–
5
–
–
6
–
–
7
–
–
Status Information 3
Bit
Name
Description
0
–
–
1
Chart end
Set to “1” while the recorder is out of chart paper.
2
–
–
3
–
–
4
–
–
5
Chart feeding
Set to “1” while the chart is being fed through the
panel key.
6
–
–
7
–
–
Status Information 4
Bit
Name
Description
0
Basic setting
Set to “1” during Basic Setting mode.
1
Recording
Set to “1” while recording is in progress.
2
Computing
Set to “1” while computation is in progress.
3
Alarm generating
Set to “1” while the alarm is occurring.
4
–
–
5
–
–
6
–
–
7
–
–
ASCII Character Codes
The table below contains the ASCII character codes of characters that can be used. The
table below shows characters each command can use.
Command Used for
Characters
SN
Unit
Alphanumeric characters, signs, and space
ST
Tag
Alphanumeric characters, signs, and space
SG
Message
Alphanumeric characters, signs, and space
YB
Host/Domain name
Alphanumeric characters and signs
YN
Domain suffix
Alphanumeric characters and signs
SO
Computing equation
Alphanumeric characters, signs, and space
Note
“µ”, “Ω”, “2”, “3” , and ”°” are assigned to character codes as follows:
µ: 7BH ({), Ω: 7CH (|), 2: 7DH (}), 3: 7EH (~) , °: 5EH (^)
Characters in the parentheses are characters assigned on a keyboard.
0
1
2
3
4
5
6
7
8
9
A
B
C
D
E
F
0
LF
CR
1
ESC
2
SP
#
%
(
)
*
+
–
.
/
3
0
1
2
3
4
5
6
7
8
9
4
@
A
B
C
D
E
F
G
H
I
J
K
L
M
N
O
5
P
Q
R
S
T
U
V
W
X
Y
Z
°
6
a
b
c
d
e
f
g
h
i
j
k
l
m
n
o
7
p
q
r
s
t
u
v
w
x
y
z
µ
Ω
2
3
8
9
A
B
C
D
E
F
Upper 4 bits
Lower 4 bits
(space)
Appendix
Output Flow of FIFO Data
The recorder has a dedicated internal memory for outputting measured/computed data.
This memory is structured as a FIFO (First-In-First-Out). Measured/computed data are
constantly acquired to the internal memory at the specified acquiring interval (FIFO
acquiring interval, set with the FR command). By using this function, it is possible to
read measured/computed data that have been saved at the specified intervals
regardless of the frequency at which the PC periodically reads the measured/computed
data.
The following example shows the case when the acquiring interval is 1 s and the
buffer capacity is for 8 intervals.
Most recent acquire
Position: WP
Previous read
Position: RP1
Current read
Position: RP2
Block 1
2
3
4
5
6
7
Block 8
• Acquiring of the measured/computed data
•
The measured/computed data are acquired to the internal memory at 1 s intervals.
•
Measured/computed data are acquired to blocks 1 through 8 in order. After
acquiring to block 8, the next acquiring operation returns to block 1.
• Reading the measured/computed data (FF GET command is used, logging
output)
Outputs the data from the next to the previous read position (RP1) to the most recent
acquire position (WP).
In this example, more than 2 s has elapsed from the previous read operation.
Therefore, data in blocks 5 and 6 are output.
• Reading the measured/computed data (FF GETNEW command is used, output
of the most recent value)
Output the specified number of blocks of FIFO data back starting from the recent
acquire position (WP).
In this example, if you specify the number of blocks to “5,” data in blocks 2 to 6 are
output.
The buffer capacity varies depending on the model.
Pen model:
240 intervals (30 s at an acquiring interval of 125 ms)
Dot model:
60 intervals (60 s at an acquiring interval of 1 s)
Example in which the FIFO acquiring interval on the recorder is set to 1 s and the
measured data from channel 01 to 06 are continuously output (logging function)
BINARY header
BINARY data
BINARY footer
FF RESET
YES
NO
END
Send command
Stop measurement?
Receive response
E0
Send command
Receive response
Wait for the FIFO
acquiring interval (1 s)
Move the FIFO read position to the most recent acquire position.
Command description
Command to send
Received response data
Output the measured/computed data using the FIFO.
START
BINARY (see page 5-2 and 5-13)
FF GET,01,06
Note
• The FIFO acquiring interval must be set using the FR command beforehand.
• The FIFO acquiring interval applies to both serial and Ethernet communications.
Appendix 2 Output Flow of FIFO Data
When using the Setting/Measurement server or the Maintenance/Test server via the
Ethernet interface (/C7 option), you must log into the recorder from the PC. If you
complete the procedure successfully up to login complete in the following figure, the
commands in chapter 4 become functional.
When using the login function
E1 403
E1 403
Yes
No
No
No
No
No
No
E1 404
E1 421
Connect
Disconnect
E1 422
Disconnect
E1 420
Disconnect
Login complete
Within 3
retries?††
Within 3
retries?††
Wait 5 s
E1 401
E0
E1 400
Entered
"quit"?
Verification
match?
Within the number of
simultaneous uses?†††
Within 2
minutes?
Within the number of
connections†
Within 2
minutes?
User name
Password
Increment retry count
1
1
1
1
End
Start
Judgment
Transition
Code (for a description of codes and messages, see page App-6)
Response from the µR (message omitted)
For a description of the response format, see section 5.1.
User entry
†
Connections cannot exceed the maximum number of connections (see section 2.1).
†† If you attempt to log in using a wrong password four consecutive times, the communication
is dropped (the number of retries for login is three).
††† If you attempt to log in causing the number of simultaneous uses at the administrator or
user level to be exceeded (see section 2.1) four consecutive times, the communication is
dropped (even if the password is correct).
When not using the login function
Login as “admin” or “user.”
•
The user name “admin” can be used to login to the recorder as an administrator.
•
The user name “user” can be used to access the recorder as a user.
E1 403
Yes
No
No
No
No
No
No
E1 404
E1 421
Connect
Disconnect
E1 422
Disconnect
E1 420
Disconnect
Login complete
Within 3
retries?
Within 3
retries?
E0
E1 402
Entered
"quit"?
Verification
match?
Within the number of
simultaneous uses?
Within the number of
connections?
Within 2
minutes?
User name
Increment retry count
1
1
1
1
Appendix 3 Login Procedure
A List of Error Messages
There are cases in which error codes and messages are displayed on the screen during
operation. The error messages and their description are listed below.
Setting Errors
Code
Message
Explanation/Countermeasures
1
System error.
Contact OMEGA.
2
Incorrect date or time setting.
Check the setting.
3
A disabled channel is selected.
The channel does not exist.
4
Incorrect function parameter.
Incorrect communication parameter.
5
The input numerical value exceeds the set range.
–
6
Incorrect input character string.
The entered character cannot be used.
7
Too many characters.
–
8
Incorrect input mode.
Incorrect range mode (Volt, TC, Scale, etc.) setting.
9
Incorrect input range code.
Incorrect range type (2 V, R, PT100, etc.) setting.
10
Format error.
Incorrect character string format.
11
Range settings are not same within the selected
Channels whose range differs cannot be set simultaneously.
channels.
12
An invalid characters.
Contains an invalid character.
13
Ref. CH error.
Specify a channel whose range is set to voltage, TC, or RTD for
the reference channel.
21
Cannot set an alarm for a SKIPPED channel.
–
22
The upper and lower span limits are equal.
This is not allowed.
23
The upper and lower scale limits are equal.
This is not allowed.
24
The lower span limit is greater than the upper span
–
limit.
25
The lower scale limit is greater than the upper scale
–
limit.
26
Bias cannot be set to the SKIPPED channel.
–
27
Bias cannot be set to the DI channel.
–
30
The partial boundary value exceeds the range of the –
span.
31
Partial is invalid on the SKIPPED channel.
–
35
The upper and lower limits of the printing zone
Set the rightmost value of the zone – the leftmost value ≥ 5
are equal.
mm.
36
The lower limit of the printing zone is greater
Set the rightmost value of the zone – the leftmost value ≥ 5
than the upper limit.
mm.
37
The printing zone is narrower than the minimum
Set the rightmost value of the zone – the leftmost value ≥ 5
width (5 mm).
mm.
38
Partial is invalid on the DI channel.
–
47
All items in DISP menu parameters are set to SKIP.
–
48
Start = Finish.
The DST start time and end time cannot be set to the same time.
49
Invalid or missing DST time settings.
Since the time gains one hour when the DST starts, the set-up
time does not exist.
61
There is no channel specified by the MATH
Set a computation channel.
expression.
62
MATH expression grammar is incorrect.
Enter the equation correctly.
63
MATH expression sequence is incorrect.
Enter the equation correctly.
64
MATH upper and lower span values are equal.
This is not allowed.
70
MATH constant description is incorrect.
Incorrect computation constant syntax.
71
The range of the MATH constant is exceeded.
–
72
MATH channel is turned off
–
Appendix 4 A List of Error Messages
Code
Message
Explanation/Countermeasures
81
All space or ‘quit’ string cannot be specified.
–
86
The key-lock release password is incorrect.
Enter the correct password.
87
This key is locked.
–
91
Password is incorrect.
Enter the correct password.
100
IP address doesn’t belong to class A, B, or C.
–
101
The result of the masked IP address is all 0s or 1s.
–
102
SUBNET mask is incorrect.
Set a correct subnet mask.
103
The net part of default gateway is not equal to
Set the correct default gateway.
that of IP address.
151
This action is invalid during calculation.
–
160
This action is invalid during chart end.
–
161
This action is invalid during pen hold.
–
162
Cannot set an number for a skipped data.
–
163
This action is invalid during record.
–
164
This action is invalid during manual printing.
–
165
This action is invalid during list printing.
–
166
This action is invalid during setup list printing.
–
167
This action is invalid during chart feed.
–
169
This action is invalid during ribbon hold.
–
Operation Errors
Code
Message
Explanation/Countermeasures
232
There is no available data.
There is no data for periodic printout or data for calculating TLOG
when the timer expired.
Communication Errors
The messages from 390 to 422 can only be returned via communication, and are not displayed on the recorder.
Code
Message
Explanation/Countermeasures
300
Command is too long.
–
301
Too many number of commands delimited with ’;’.
Keep the number of commands separated by sub delimiters
under 10.
302
This command has not been defined.
–
303
Data request command can not be enumerated with –
sub-delimiter.
350
Command is not permitted to the current user level.
–
351
This command cannot be specified in the current
–
mode.
352
The option is not installed.
–
353
This command cannot be specified in the current
–
setting.
354
This command is not available during calculation.
–
390
Command error.
–
391
Delimiter error.
–
392
Parameter error.
–
393
No permission.
–
394
No such connection.
–
395
Use “quit” to close this connection.
Attempted to disconnect its own connection.
396
Failed to disconnect.
–
397
No TCP control block.
The control block of the specified connection cannot be found.
Appendix 4 A List of Error Messages
Code
Message
Explanation/Countermeasures
400
Input username.
–
401
Input password.
–
402
Select username from ‘admin’ or ‘user’.
If the recorder is configured not to use the user name and
password, use user names ‘admin’ or ‘user’.
403
Login incorrect, try again!
–
404
No more login at the specified level is acceptable.
–
420
Connection has been lost.
–
421
The number of simultaneous connection has been
–
exceeded.
422
Communication has timed-out.
–
Warning Messages
Code
Message
Explanation/Countermeasures
600
Initialized.
Settings and measured data have been initialized.
System Errors
Code
Message
Explanation/Countermeasures
902
RAM failure.
Contact OMEGA.
910
A/D error.
Contact OMEGA.
921
A/D calibration value error.
Contact OMEGA.
922
A/D calibration is in the wrong order.
Contact OMEGA.
930
Memory acquisition failure.
Contact OMEGA.
940
The ethernet module is down.
Contact OMEGA.
950
A/D number error.
Contact OMEGA.
951
EEPROM write error.
Contact OMEGA.
960
Ribbon error
Contact OMEGA.
961
Printer error
Contact OMEGA.
962
Plotter error
Contact OMEGA.
963
Pen 1 error
Contact OMEGA.
964
Pen 2 error
Contact OMEGA.
965
Pen 3 error
Contact OMEGA.
966
Pen 4 error
Contact OMEGA.
host ............................................................................. 4-30
IF ................................................................................ 4-26
ip ................................................................................. 4-30
IS ................................................................................ 4-28
LS ............................................................................... 4-25
MC .............................................................................. 4-26
MP .............................................................................. 4-25
MS .............................................................................. 4-25
net ............................................................................... 4-29
PS ............................................................................... 4-25
quit .............................................................................. 4-30
SA ............................................................................... 4-11
SC ............................................................................... 4-12
SD ............................................................................... 4-13
SE ............................................................................... 4-14
serial ........................................................................... 4-30
SF ............................................................................... 4-14
SG .............................................................................. 4-14
SJ ............................................................................... 4-15
SK ............................................................................... 4-15
SN ............................................................................... 4-12
SO .............................................................................. 4-11
SP ............................................................................... 4-13
SR ............................................................................... 4-10
SS ............................................................................... 4-15
ST ............................................................................... 4-14
SU ............................................................................... 4-25
SV ............................................................................... 4-14
SZ ............................................................................... 4-13
TD ............................................................................... 4-14
TL ............................................................................... 4-25
UA ............................................................................... 4-23
UB ............................................................................... 4-20
UC .............................................................................. 4-19
UD .............................................................................. 4-25
UF ............................................................................... 4-21
UI ................................................................................ 4-20
UJ ............................................................................... 4-20
UK ............................................................................... 4-20
UL ............................................................................... 4-20
UM .............................................................................. 4-20
UN .............................................................................. 4-22
UO .............................................................................. 4-19
UP ............................................................................... 4-19
UR .............................................................................. 4-19
US ............................................................................... 4-22
UT ............................................................................... 4-21
UY ............................................................................... 4-26
VB ............................................................................... 4-11
VD ............................................................................... 4-16
VF ............................................................................... 4-14
VG .............................................................................. 4-26
VR ............................................................................... 4-13
VT ............................................................................... 4-13
XA ............................................................................... 4-18
XB ............................................................................... 4-19
XE ............................................................................... 4-24
XI ................................................................................ 4-18
XJ ............................................................................... 4-19
XN ............................................................................... 4-20
XQ .............................................................................. 4-22
XR ............................................................................... 4-21
YA ............................................................................... 4-22
YB ............................................................................... 4-22
Symbols
Page
1-5V input low-cut ............................................................. 4-21
A
Page
A/D integral time ............................................................... 4-18
address ............................................................................. 3-12
administrator ....................................................................... 2-9
affirmative response ............................................................ 5-1
alarm .......................................................................4-11, 4-18
alarm ACK ......................................................................... 4-25
alarm acknowledge ........................................................... 4-25
alarm delay ....................................................................... 4-21
alarm delay time ................................................................ 4-14
alarm printout buffer .......................................................... 4-25
ASCII character codes ................................................... App-1
ASCII data ........................................................................... 5-6
ASCII output ........................................................................ 5-2
assignment of channels .................................................... 4-22
B
Page
bar graph ........................................................................... 4-20
basic setting commands ..................................................... 4-4
basic setting mode ........................................................ vi, 4-3
baud rate ........................................................................... 3-12
bias .........................................................................4-11, 4-21
BINARY data ..................................................................... 5-13
BINARY output .................................................................... 5-2
bit structure ......................................................................... 3-7
brightness ......................................................................... 4-14
burnout detection .............................................................. 4-19
byte output order ............................................................... 4-26
C
Page
channel number .................................................................. 4-8
chart speed ............................................................... 4-9, 4-12
checksum .......................................................................... 4-26
checksum value .................................................................. 5-3
closes the instrument ........................................................ 4-28
command
AC ............................................................................... 4-25
AK ............................................................................... 4-25
BD ............................................................................... 4-14
BO .............................................................................. 4-26
CC .............................................................................. 4-26
close ........................................................................... 4-29
CM .............................................................................. 4-15
con .............................................................................. 4-29
CS ............................................................................... 4-26
DS ............................................................................... 4-25
ESC C ......................................................................... 4-28
ESC O ........................................................................ 4-28
eth ............................................................................... 4-29
FD ............................................................................... 4-27
FE ............................................................................... 4-27
FF ............................................................................... 4-27
FR ............................................................................... 4-15
FU ............................................................................... 4-28
FY ............................................................................... 4-27
help ............................................................................. 4-29
YC ............................................................................... 4-26
YD ............................................................................... 4-23
YE ............................................................................... 4-24
YK ............................................................................... 4-23
YN ............................................................................... 4-23
YQ .............................................................................. 4-23
YS ............................................................................... 4-21
command syntax ................................................................. 4-1
command-response rule ..................................................... 4-2
communication input data ......................................... 4-8, 4-15
computation error .............................................................. 4-22
computing equation ........................................................... 4-11
connection (ethernet) .......................................................... 2-2
connection (RS-422A/485) .................................................. 3-2
connection example (RS-422A/485) ................................... 3-3
connection information ...................................................... 4-29
connection status of the Ethernet interface ......................... 2-7
constant .................................................................... 4-8, 4-15
control commands ............................................................... 4-5
conventions ........................................................................... iv
converter ............................................................................. 3-5
copying & pasting a character string .....................................vii
D
Page
data display screen ........................................................... 4-16
data length ........................................................................ 3-12
date and time .................................................................... 4-13
decimal point position ....................................................... 4-27
decimal point position output .............................................. 5-7
default gateway ................................................................... 2-6
deleting a character ............................................................. vii
deleting an entire character string .........................................vii
delimiter .............................................................................. 4-2
disconnects an Ethernet connection ................................. 4-26
disconnects the connection ..................................... 4-29, 4-30
display .................................................................................... v
DNS .......................................................................... 2-6, 4-23
DNS server ......................................................................... 2-6
domain name ............................................................ 2-5, 4-22
domain suffix ....................................................................... 2-6
dot color ............................................................................ 4-19
dot printing interval ............................................................ 4-13
DST ......................................................................... 4-14, 4-15
E
Page
entering basic setting mode .................................................. vi
entering characters .............................................................. vii
entering values ......................................................................vii
error messages .............................................................. App-6
ESC key ................................................................................ vi
Ethernet interface ................................................................ 2-1
Ethernet statistical information .......................................... 4-29
execution mode ................................................................. 4-25
execution modes ........................................................... vi, 4-3
exiting from basic setting mode ........................................... viii
exits from basic setting mode ........................................... 4-24
extended functions ............................................................ 4-21
F
Page
FIFO ............................................................................... App-2
FIFO data ................................................................ 4-27, 5-13
filter (status information) ..................................................... 6-1
first-in-first-out ................................................................ App-2
flag ............................................................................ 5-3, 5-14
frequently used parameters ................................................ 4-7
functional construction (ethernet) ........................................ 1-1
functional construction (RS-422A/485) ............................... 1-4
H
Page
help ................................................................................... 4-29
Hold register ...................................................................... 3-10
host name ....................................................... 2-5, 4-22, 4-30
I
Page
identifier .............................................................................. 5-3
indicator .............................................................................. 2-7
initializes settings .............................................................. 4-26
input filte ............................................................................ 4-14
input filter .......................................................................... 4-20
input range ................................................................ 4-7, 4-10
Input register ....................................................................... 3-9
inserting characters .............................................................. vii
instrument information output commands ........................... 4-6
instrument information server ............................................. 1-2
interval for acquiring data to the FIFO buffer .................... 4-15
IP address ....................................................... 2-5, 4-22, 4-30
items to be printed ............................................................ 4-19
K
Page
keepalive ......................................................... 1-3, 2-12, 4-23
key operations ....................................................................... vi
keys ........................................................................................ v
L
Page
language ........................................................................... 4-20
list (setting information) printout ........................................ 4-25
log into the recorder ....................................................... App-4
login function ..................................................... 1-3, 2-9, 4-23
M
Page
maintenance/test commands .............................................. 4-6
maintenance/test server ...................................................... 1-2
manual print ...................................................................... 4-25
maximum number of simultaneous connections ................. 2-1
measured/computed data ................................................. 4-27
measured/computed data output ........................................ 5-8
message ................................................................. 4-14, 4-25
message printout buffer .................................................... 4-26
Modbus ....................................................................... 1-4, 3-8
moving average ...................................................... 4-14, 4-20
multiple negative responses ............................................... 5-1
N
Page
negative response ............................................................... 5-1
network statistical information ........................................... 4-29
O
Page
opens the instrument ........................................................ 4-28
operation mode ..................................................................... vi
output commands ............................................................... 4-5
Index
parameters .......................................................................... 4-1
parity check ....................................................................... 3-12
partial expanded recording ..................................... 4-13, 4-20
password ............................................................................. 2-9
pen offset compensation ................................................... 4-19
periodic printout interval .................................................... 4-19
port number ......................................................................... 2-1
protocol ............................................................................... 2-1
Q
Page
query ................................................................................... 4-2
R
Page
rear panel ............................................................................... v
receive buffer ...................................................................... 3-7
record position .................................................................. 4-23
recording on/off ................................................................. 4-13
relay number ....................................................................... 4-8
remote control input .......................................................... 4-21
remote control input terminal .............................................. 4-8
report data ............................................................... 4-20, 4-26
report data output .............................................................. 5-10
response ............................................................................. 4-2
revisions .................................................................................. i
RJC ................................................................................... 4-19
RS-422A/485 communication interface .............................. 3-1
RS-422A/485 dedicated commands ................................... 4-5
RS-422A/485 dedicated commands and responses ........... 5-5
RS-422A/485 interface ...................................................... 4-21
run mode ....................................................................... vi, 4-3
S
Page
secondary chart speed ...................................................... 4-14
selecting the setup item and value ........................................ vi
serial number .................................................................... 4-30
setting commands ............................................................... 4-3
setting mode ......................................................................... vi
setting/basic setting data output ......................................... 5-6
setting/measurement server (ethernet) ............................... 1-1
setting/measurement server (RS-422A/485) ...................... 1-4
setup data ......................................................................... 4-27
setup list (basic setting information) printout ..................... 4-25
simultaneous users ............................................................. 2-1
special data value ............................................................. 5-14
square root low-cut ........................................................... 4-21
start bit ................................................................................ 3-7
starts/stops recording ........................................................ 4-25
starts/stops/resets computation ........................................ 4-25
statistical computation results ........................................... 4-27
status filter ......................................................................... 4-26
status information ...................................................... 4-28, 6-1
status information (bit structure) ......................................... 6-2
status information output ................................................... 5-12
stop bit ................................................................................ 3-7
stops the record position adjustment ................................ 4-26
sub delimiter ........................................................................ 4-2
subnet mask ........................................................................ 2-5
sum value ............................................................................ 5-4
switches the channel ......................................................... 4-25
switches the screen .......................................................... 4-25
syntax .................................................................................. 4-1
T
Page
tag ..................................................................................... 4-14
temperature unit ................................................................ 4-21
terminator .................................................................... 3-6, 4-2
time printout format ........................................................... 4-21
timeout ...................................................................... 1-3, 4-23
timeout time ...................................................................... 2-12
timer .................................................................................. 4-15
TLOG timer ....................................................................... 4-22
types of responses .............................................................. 5-1
U
Page
unit .................................................................................... 4-12
unit information ................................................................. 4-27
unit information output ........................................................ 5-7
user ..................................................................................... 2-9
user information ................................................................ 4-28
user information output ..................................................... 5-12
user level ..................................................................... 2-9, 4-3
user name ........................................................................... 2-9
Z
Page
zone recording .................................................................. 4-13
OMEGA ENGINEERING, INC. warrants this unit to be free of defects in materials and workmanship for a
period of 13 months from date of purchase. OMEGA’s WARRANTY adds an additional one (1) month
grace period to the normal one (1) year product warranty to cover handling and shipping time. This
ensures that OMEGA’s customers receive maximum coverage on each product.
If the unit malfunctions, it must be returned to the factory for evaluation. OMEGA’s Customer Service
Department will issue an Authorized Return (AR) number immediately upon phone or written request.
Upon examination by OMEGA, if the unit is found to be defective, it will be repaired or replaced at no
charge. OMEGA’s WARRANTY does not apply to defects resulting from any action of the purchaser, includ-
ing but not limited to mishandling, improper interfacing, operation outside of design limits,
improper repair, or unauthorized modification. This WARRANTY is VOID if the unit shows evidence of
having been tampered with or shows evidence of having been damaged as a result of excessive corrosion;
or current, heat, moisture or vibration; improper specification; misapplication; misuse or other operating
conditions outside of OMEGA’s control. Components in which wear is not warranted, include but are not
limited to contact points, fuses, and triacs.
OMEGA is pleased to offer suggestions on the use of its various products. However,
OMEGA neither assumes responsibility for any omissions or errors nor assumes liability for any
damages that result from the use of its products in accordance with information provided by
OMEGA, either verbal or written. OMEGA warrants only that the parts manufactured by the
company will be as specified and free of defects. OMEGA MAKES NO OTHER WARRANTIES OR
REPRESENTATIONS OF ANY KIND WHATSOEVER, EXPRESSED OR IMPLIED, EXCEPT THAT OF
TITLE, AND ALL IMPLIED WARRANTIES INCLUDING ANY WARRANTY OF MERCHANTABILITY
AND FITNESS FOR A PARTICULAR PURPOSE ARE HEREBY DISCLAIMED. LIMITATION OF
LIABILITY: The remedies of purchaser set forth herein are exclusive, and the total liability of
OMEGA with respect to this order, whether based on contract, warranty, negligence,
indemnification, strict liability or otherwise, shall not exceed the purchase price of the
component upon which liability is based. In no event shall OMEGA be liable for
consequential, incidental or special damages.
CONDITIONS: Equipment sold by OMEGA is not intended to be used, nor shall it be used: (1) as a “Basic
Component” under 10 CFR 21 (NRC), used in or with any nuclear installation or activity; or (2) in medical
applications or used on humans. Should any Product(s) be used in or with any nuclear installation or
activity, medical application, used on humans, or misused in any way, OMEGA assumes no responsibility
as set forth in our basic WARRANTY/DISCLAIMER language, and, additionally, purchaser will indemnify
OMEGA and hold OMEGA harmless from any liability or damage whatsoever arising out of the use of the
Product(s) in such a manner.
RETURN REQUESTS/INQUIRIES
Direct all warranty and repair requests/inquiries to the OMEGA Customer Service Department. BEFORE
RETURNING ANY PRODUCT(S) TO OMEGA, PURCHASER MUST OBTAIN AN AUTHORIZED RETURN
(AR) NUMBER FROM OMEGA’S CUSTOMER SERVICE DEPARTMENT (IN ORDER TO AVOID
PROCESSING DELAYS). The assigned AR number should then be marked on the outside of the return
package and on any correspondence.
The purchaser is responsible for shipping charges, freight, insurance and proper packaging to prevent
breakage in transit.
FOR WARRANTY RETURNS, please have the
following information available BEFORE
contacting OMEGA:
1. Purchase Order number under which the prod-
uct was PURCHASED,
2. Model and serial number of the product under
warranty, and
3. Repair instructions and/or specific problems
relative to the product.
FOR NON-WARRANTY REPAIRS, consult OMEGA
for current repair charges. Have the following
information available BEFORE contacting OMEGA:
1. Purchase Order number to cover the COST
of the repair,
2. Model and serial number of the product, and
3. Repair instructions and/or specific problems
relative to the product.
OMEGA’s policy is to make running changes, not model changes, whenever an improvement is possible. This affords our
customers the latest in technology and engineering.
OMEGA is a registered trademark of OMEGA ENGINEERING, INC.
© Copyright 2005 OMEGA ENGINEERING, INC. All rights reserved. This document may not be copied, photocopied, repro-
duced, translated, or reduced to any electronic medium or machine-readable form, in whole or in part, without the prior
written consent of OMEGA ENGINEERING, INC.
Where Do I Find Everything I Need for
Process Measurement and Control?
OMEGA…Of Course!
Shop online at omega.com
TEMPERATURE
�� Thermocouple, RTD & Thermistor Probes, Connectors, Panels & Assemblies
�� Wire: Thermocouple, RTD & Thermistor
�� Calibrators & Ice Point References
�� Recorders, Controllers & Process Monitors
�� Infrared Pyrometers
PRESSURE, STRAIN AND FORCE
�� Transducers & Strain Gages
�� Load Cells & Pressure Gages
�� Displacement Transducers
�� Instrumentation & Accessories
FLOW/LEVEL
�� Rotameters, Gas Mass Flowmeters & Flow Computers
�� Air Velocity Indicators
�� Turbine/Paddlewheel Systems
�� Totalizers & Batch Controllers
pH/CONDUCTIVITY
�� pH Electrodes, Testers & Accessories
�� Benchtop/Laboratory Meters
�� Controllers, Calibrators, Simulators & Pumps
�� Industrial pH & Conductivity Equipment
DATA ACQUISITION
�� Data Acquisition & Engineering Software
�� Communications-Based Acquisition Systems
�� Plug-in Cards for Apple, IBM & Compatibles
�� Datalogging Systems
�� Recorders, Printers & Plotters
HEATERS
�� Heating Cable
�� Cartridge & Strip Heaters
�� Immersion & Band Heaters
�� Flexible Heaters
�� Laboratory Heaters
ENVIRONMENTAL
MONITORING AND CONTROL
�� Metering & Control Instrumentation
�� Refractometers
�� Pumps & Tubing
�� Air, Soil & Water Monitors
�� Industrial Water & Wastewater Treatment
�� pH, Conductivity & Dissolved Oxygen Instruments