Fluke PPC2 PLUS Manual 

Specifications

Brand: Fluke
Category: Other
Model: PPC2 PLUS (Manual )
File Type: PDF
File Size: 1.46 Mb
Pages: 166

Read Online

Read the Fluke PPC2 PLUS Manual  online using the built-in PDF viewer.

Text of the manual

PPC2+™ Operation and Maintenance Manual

©2001 DH Instruments, Inc.                 2SQRZ\O]]_\OVS[_SN]KXNQK]O]K\OZY^OX^SKVVcRKdK\NY_]/XO\Qc]^Y\ONSX^RO]OVS[_SN]KXNQK]O]MKX LO \OVOK]ON _XObZOM^ONVc KXN aS^R Ob^\OWO PY\MO  2SQR Z\O]]_\O ]c]^OW] ]RY_VN LO K]]OWLVON KXN YZO\K^ONYXVcLcZO\]YXXOVaRYRK`OLOOXSX]^\_M^ONSXZ\YZO\]KPO^cZ\KM^SMO]  © 1998 - 2001 DH Instruments, Inc. All rights reserved. Information in this document is subject to change without notice. No part of this document may be reproduced or transmitted in any form or by any means, electronic or mechanical, for any purpose, without the express written permission of DH Instruments, Inc. 4765 East Beautiful Lane Phoenix AZ 85044-5318 USA. DH Instruments makes sincere efforts to ensure accuracy and quality of its’ published materials; however, no warranty, expressed or implied, is provided. DH Instruments disclaims any responsibility or liability for any direct or indirect damages resulting from the use of the information in this manual or products described in it. Mention of any product does not constitute an endorsement by DH Instruments of that product. This manual was originally composed in English and was subsequently translated into other languages. The fidelity of the translation cannot be guaranteed. In case of conflict between the English version and other language versions, the English version predominates. DH Instruments, DH, DHI, PPC, PPC2, PPC2+, and AutoZ are trademarks, registered and otherwise. Document No. 550091f 010403 Printed in the USA

PPC2+™ Operation and Maintenance Manual

Page i ©2001 DH Instruments, Inc. >>++,,66//9 900--9 98 8>>//8 8>>== TABLE OF CONTENTS ................................................................................i TABLES ................................................................................................... iv FIGURES .................................................................................................. v ABOUT THIS MANUAL .............................................................................. vi 1. INTRODUCTION .................................................................................... 1 1.1 PRODUCT OVERVIEW............................................................................................................................. 1 1.2 SPECIFICATIONS..................................................................................................................................... 2 1.2.1 GENERAL SPECIFICATIONS .......................................................................................................................2 1.2.2 PRESSURE MEASUREMENT SPECIFICATIONS ........................................................................................2 1.2.2.1 REFERENCE PRESSURE TRANSDUCER (RPT) SPECIFICATIONS ..........................................3 1.2.2.2 ON-BOARD BAROMETER.............................................................................................................4 1.2.3 PRESSURE CONTROL SPECIFICATIONS...................................................................................................4 1.2.4 ANALOG INPUT SPECIFICATION................................................................................................................5 2. INSTALLATION ..................................................................................... 7 2.1 UNPACKING AND INSPECTION.............................................................................................................. 7 2.1.1 REMOVING FROM PACKAGING..................................................................................................................7 2.1.2 INSPECTING CONTENTS .............................................................................................................................7 2.2 SITE REQUIREMENTS............................................................................................................................. 8 2.3 INITIAL SETUP ......................................................................................................................................... 8 2.3.1 PREPARING FOR OPERATION....................................................................................................................8 2.3.2 FRONT AND REAR PANELS ........................................................................................................................9 2.3.2.1 FRONT PANEL ..............................................................................................................................9 2.3.2.2 REAR PANEL ASSEMBLY .......................................................................................................... 10 2.3.3 POWER CONNECTION............................................................................................................................... 10 2.3.4 CONNECTING TO A PRESSURE SUPPLY (SUPPLY PORT) .................................................................... 11 2.3.5 CONNECTING A VACUUM PUMP (EXHAUST PORT) ............................................................................... 11 2.3.6 CONNECTING TO THE DEVICE UNDER TEST (TEST AND TEST(-) PORTS) ...................................................... 11 2.3.6.1 INSTALLING THE SPLT .............................................................................................................. 12 2.3.7 THE VENT CONNECTION (VENT PORT) ................................................................................................... 13 2.3.8 THE ATM REF PASS THROUGH (UNLABELED)....................................................................................... 13 2.4 POWER-UP AND VERIFICATION .......................................................................................................... 13 2.4.1 SWITCH POWER ON................................................................................................................................... 13 2.4.2 CHECK PROPER PRESSURE MEASUREMENT OPERATION.................................................................. 14 2.4.2.1 CHECKING ABSOLUTE MODE PRESSURE MEASUREMENT.................................................. 14 2.4.2.2 CHECKING GAUGE MODE PRESSURE MEASUREMENT ........................................................ 14 2.4.3 LEAK CHECK.............................................................................................................................................. 14 2.4.4 PURGE ........................................................................................................................................................ 15 2.4.5 CHECK PROPER PRESSURE CONTROL OPERATION............................................................................ 15 2.5 SHORT TERM STORAGE ...................................................................................................................... 15 3. OPERATION........................................................................................ 17 3.1 GENERAL/MANUAL OPERATION......................................................................................................... 17 3.1.1 MAIN RUN SCREEN.................................................................................................................................... 17 3.1.2 GENERAL OPERATING PRINCIPLES........................................................................................................ 18 3.1.2.1 MAIN 4 X 4 KEYPAD LAYOUT AND PROTOCOL ....................................................................... 18 3.1.2.2 DIRECT PRESSURE CONTROL KEYS....................................................................................... 19 3.1.2.3 AUTOMATED PRESSURE CONTROL ........................................................................................ 20 3.1.2.4 PRESSURE READY (<*>)/NOT READY (<↑> OR <↓>)............................................................... 22

PPC2+™ Operation and Maintenance Manual

©2001 DH Instruments, Inc. Page ii  3.1.2.5 MULTIPLE PRESSURE RANGES ............................................................................................... 24 3.2 DIRECT FUNCTION KEYS ..................................................................................................................... 25 3.2.1 RANGE ........................................................................................................................................................ 25 3.2.1.1 RANGE SPECIFIC FUNCTIONS AND SETTINGS ...................................................................... 27 3.2.2 UNIT............................................................................................................................................................. 27 3.2.3 CONTROL.................................................................................................................................................... 29 3.2.3.1 SYSTEM DEFAULT CONTROL PARAMETERS.......................................................................... 30 3.2.3.2 EXECUTING AUTOMATED PRESSURE CONTROL COMMANDS............................................. 30 3.2.4 UPPER LIMIT............................................................................................................................................... 33 3.2.4.1 OVERPRESSURE FUNCTION .................................................................................................... 34 3.2.5 SEQ.............................................................................................................................................................. 34 3.2.5.1 SETTING UP/EDITING A QUICK SEQUENCE ............................................................................ 34 3.2.5.2 SETTING UP/EDITING A FILE SEQUENCE................................................................................ 36 3.2.5.3 RUNNING A SEQUENCE ............................................................................................................ 36 3.2.6 CUSTOM CONTROL ................................................................................................................................... 38 3.2.6.1 TURNING-OFF CUSTOM CONTROL PARAMETERS................................................................. 40 3.2.7 HEAD........................................................................................................................................................... 40 3.2.8 LEAK CHECK/PURGE ................................................................................................................................ 41 3.2.8.1 LEAK CHECK/PURGE MENU AND FUNCTIONS FLOW CHART ............................................... 42 3.2.8.2 LEAK CHECK SCREEN............................................................................................................... 43 3.2.8.3 PURGE FUNCTION ..................................................................................................................... 44 3.2.8.4 SETTING PRESSURES FOR LEAK CHECK/PURGE FUNCTION .............................................. 46 3.2.8.5 TYPICAL USE OF THE LEAK CHECK/PURGE FUNCTION........................................................ 47 3.2.9 ANALOG...................................................................................................................................................... 48 3.2.9.1 VIEWING ANALOG MEASUREMENTS ....................................................................................... 49 3.3 SETUP MENU KEY................................................................................................................................. 49 3.3.1 CONTROLREF............................................................................................................................................. 49 3.3.2 RESLTN....................................................................................................................................................... 51 3.3.3 SEQUENCE ................................................................................................................................................. 51 3.3.4 DRIVERS ..................................................................................................................................................... 55 3.4 SPECIAL MENU KEY ............................................................................................................................. 56 3.4.1 AUTOZ......................................................................................................................................................... 56 3.4.1.1 VIEW AUTOZ............................................................................................................................... 59 3.4.1.2 EDIT AUTOZ................................................................................................................................ 60 3.4.1.3 RUN AUTOZ ................................................................................................................................ 60 3.4.2 HEAD........................................................................................................................................................... 64 3.4.3 PRESU......................................................................................................................................................... 65 3.4.4 INTERNAL ................................................................................................................................................... 66 3.4.4.1 CONFIG ....................................................................................................................................... 66 3.4.4.2 CAL.............................................................................................................................................. 67 3.4.4.3 REMOTE...................................................................................................................................... 67 3.4.4.4 TIME ............................................................................................................................................ 68 3.4.4.5 RESET ......................................................................................................................................... 69 3.4.4.6 LEVEL.......................................................................................................................................... 71 3.4.4.7 ATM ............................................................................................................................................. 74 3.4.4.8 PURGE ........................................................................................................................................ 75 3.4.4.9 SCRSAV ...................................................................................................................................... 75 4. REMOTE OPERATION.......................................................................... 77 4.1 OVERVIEW ............................................................................................................................................. 77 4.2 INTERFACING ........................................................................................................................................ 77 4.2.1 RS-232 INTERFACE.................................................................................................................................... 77 4.2.1.1 COM1........................................................................................................................................... 77 4.2.1.2 IEEE-488...................................................................................................................................... 78 4.2.1.3 COM2........................................................................................................................................... 78 4.3 PROGRAMMING FORMATS .................................................................................................................. 79 4.3.1 CLASSIC PROGRAM MESSAGE FORMAT ............................................................................................... 79 4.3.2 ENHANCED PROGRAM MESSAGE FORMAT........................................................................................... 79 4.3.2.1 ENHANCED PROGRAM MESSAGE USING THE COMMAND TYPE ......................................... 80 4.3.2.2 ENHANCED PROGRAM MESSAGE USING THE QUERY TYPE ............................................... 80 4.4 COMMANDS ........................................................................................................................................... 81 4.4.1 PROGRAM MESSAGES.............................................................................................................................. 81 4.4.2 ERROR MESSAGES ................................................................................................................................... 83

PPC2+™ Operation and Maintenance Manual

Page iii ©2001 DH Instruments, Inc.  4.4.3 PROGRAM MESSAGE DESCRIPTION OVERVIEW................................................................................... 84 4.4.4 PROGRAM MESSAGE DESCRIPTIONS .................................................................................................... 85 4.5 STATUS REPORTING SYSTEM .......................................................................................................... 116 4.5.1 ERROR QUEUE......................................................................................................................................... 116 4.5.2 STATUS BYTE REGISTER ....................................................................................................................... 117 4.5.3 STANDARD EVENT REGISTER ............................................................................................................... 119 4.5.4 READY STATUS REGISTER..................................................................................................................... 120 4.6 IEEE STD. 488.2 COMMON AND STATUS PROGRAM MESSAGES................................................. 120 4.6.1 PROGRAM MESSAGE DESCRIPTIONS .................................................................................................. 121 5. MAINTENANCE, ADJUSTMENTS AND CALIBRATION........................... 125 5.1 OVERVIEW ........................................................................................................................................... 125 5.2 CALIBRATION OF REFERENCE PRESSURE TRANSDUCERS ........................................................ 126 5.2.1 PRINCIPLE ................................................................................................................................................ 126 5.2.1.1 PA/PM COEFFICIENTS.............................................................................................................127 5.2.1.2 SETTING ZNATERR.................................................................................................................. 127 5.2.1.3 ORDER OF OPERATIONS ........................................................................................................ 128 5.2.1.4 AS RECEIVED/AS LEFT DATA ................................................................................................. 128 5.2.2 EQUIPMENT REQUIRED .......................................................................................................................... 128 5.2.3 SET-UP AND PREPARATION................................................................................................................... 129 5.2.4 RPT CALIBRATION USING CALTOOL FOR RPTS SOFTWARE............................................................. 130 5.2.5 EDITING AND VIEWING RPT CALIBRATION INFORMATION................................................................. 130 5.2.6 SETTING ZNATERR.................................................................................................................................. 131 5.2.7 RPT CALIBRATION/ADJUSTMENT WITHOUT CALTOOL FOR RPTS SOFTWARE .............................. 132 5.3 ADJUSTMENT OF ON-BOARD BAROMETER.................................................................................... 133 5.4 CALIBRATION OF ANALOG INPUT.................................................................................................... 134 5.4.1 CALIBRATION PRINCIPLE....................................................................................................................... 134 5.4.1.1 AUTOMATED INTERNAL CALIBRATION.................................................................................. 134 5.4.1.2 USER ADJUSTMENT OF ANALOG MEASUREMENT RANGES AND CHANNELS.................. 135 5.4.2 ADJUSTING ANALOG MEASUREMENT CHANNELS ............................................................................. 135 5.5 PNEUMATIC CONTROL MODULE CONFIGURATION ....................................................................... 136 5.6 OVERHAUL........................................................................................................................................... 138 5.7 RELOADING EMBEDDED SOFTWARE INTO FLASH MEMORY....................................................... 138 5.8 SUBASSEMBLY DESCRIPTION AND LOCATION.............................................................................. 139 5.8.1 INTERNAL VIEW ....................................................................................................................................... 139 5.8.1.1 ON-BOARD BAROMETER......................................................................................................... 140 5.8.1.2 POWER SUPPLY....................................................................................................................... 140 5.8.1.3 TRANSDUCER SELECTION MODULE ..................................................................................... 140 5.8.1.4 DRIVER BOARD........................................................................................................................ 140 5.8.1.5 MICRO CARD ............................................................................................................................ 140 5.8.1.6 REFERENCE PRESSURE TRANSDUCERS............................................................................. 140 5.8.1.7 PNEUMATIC MODULE .............................................................................................................. 141 6. TROUBLESHOOTING ......................................................................... 143 6.1 OVERVIEW ........................................................................................................................................... 143 7. APPENDIX ........................................................................................ 149 7.1 DRIVERS............................................................................................................................................... 149 7.2 ANALOG ............................................................................................................................................... 151 7.2.1 DETAILED SIGNAL DESCRIPTIONS........................................................................................................ 151 7.3 UNIT CONVERSION ............................................................................................................................. 152 7.3.1 PRESSURE................................................................................................................................................ 152 8. WARRANTY STATEMENT................................................................... 153 8.1 OVERVIEW ........................................................................................................................................... 153 9. GLOSSARY....................................................................................... 155

PPC2+™ Operation and Maintenance Manual

©2001 DH Instruments, Inc. Page iv >>++,,66//== Table 1. Reference Pressure Transducer (RPT) Designations and Ranges ...........................................3 Table 2. Default Pressure Control Parameters ........................................................................................4 Table 3. Analog Input Specifications........................................................................................................5 Table 4. PPC2+ Packing List ...................................................................................................................7 Table 5. PPC2+ Range Identification Summary.....................................................................................25 Table 6. Lo or Hi Reference Transducer Range 1, 2, 3 .........................................................................30 Table 7. System Default Control Parameters.........................................................................................39 Table 8. UNIT Function - Available Units...............................................................................................65 Table 9. Security Levels.........................................................................................................................73 Table 10. Pin Designations and Connections ..........................................................................................78 Table 11. List of PPC2+ COM2 DB-9F Pin Designations ........................................................................78 Table 12. Program Message List .............................................................................................................81 Table 13. Error #’s and Descriptions........................................................................................................83 Table 14. 8 Bit Status Byte Register ......................................................................................................117 Table 15. 8 Bit Standard Event Register................................................................................................119 Table 16. 8 Bit Ready Status Register ...................................................................................................120 Table 17. Program Message List ...........................................................................................................120 Table 18. Troubleshooting Guide...........................................................................................................143 Table 19. Driver Output..........................................................................................................................149 Table 20. External Drivers......................................................................................................................150 Table 21. Analog Port/Signal Descriptions.............................................................................................151 Table 22. Conversion Coefficients .........................................................................................................152 Table 23. Warranty Service Centers......................................................................................................154

PPC2+™ Operation and Maintenance Manual

Page v ©2001 DH Instruments, Inc. 00331 1? ?<)/Not Ready (<↑> or <↓>) Static Control Mode......................................................23 Figure 8. Ready (<*>)/Not Ready (<↑> or <↓>) in Dynamic Pressure Control Mode..............................23 Figure 9. Hi and Low RPT Pneumatic Circuit..........................................................................................24 Figure 10. Leak Check/Purge Menu Flow Chart .......................................................................................43 Figure 11. Status Register Schematic.....................................................................................................117 Figure 12. Internal View ..........................................................................................................................139 Figure 13. Pneumatic Module (Schematic) .............................................................................................141 Figure 14. Cable Schematic....................................................................................................................150

PPC2+™ Operation and Maintenance Manual

©2001 DH Instruments, Inc. Page vi ++,,9 9? ?>>>>2 233==7 7++8 8? ?++66 Manual Conventions This manual is intended to provide the user with the basic information necessary to operate a PPC2+ pressure controller/calibrator. It also includes a great deal of additional information provided to allow you to optimize its use and take full advantage of its many features and functions. Before using the manual, take a moment to familiarize yourself with the table of contents structure: Sections 1, 2 and 3 should be read by all first time PPC2+ users. Section 3 is most important for those using the local front panel interface but should be read over by all users to familiarize themselves with general PPC2+ operating principles. Section 4 is for remote operation from an external computer. Section 5 provides maintenance and calibration information. Section 6 is a quick troubleshooting guide. Use it to troubleshoot unexpected PPC2+ behavior based on the symptom of that behavior.  0Y\^RY]OYPcY_aRYNYX^\OKNWKX_KV]€QYNS\OM^Vc^Y=OM^SYX^Y]O^_ZcY_\::-KXN^ROX QY^Y=OM^SYXPY\ZYaO\_ZKXN`O\SPSMK^SYX>RS]aSVVQO^cY_\_XXSXQ[_SMUVcaS^RKWSXSWKV\S]UYP MK_]SXQNKWKQO^YcY_\]OVPY\cY_\XOa::->2/8|aROXcY_RK`O[_O]^SYX]Y\]^K\^^YaYXNO\ KLY_^KVV^ROQ\OK^POK^_\O]cY_WSQR^LOWS]]SXQQO^SX^Y^ROWKX_KV    ::- ,1 RK] K ]OZK\K^O WKX_KV aS^R WYNOV ]ZOMSPSM SXPY\WK^SYX  3P cY_ K\O_]SXQK::- ,1_]O^RS]WKX_KVSXMYXT_XM^SYXaS^R^RO::-,19ZO\K^SYXKXN7KSX^OXKXMO7KX_KV Certain words and expressions have specific meaning as they pertain to PPC2+. The Glossary Section is useful as a quick reference for exact definition of specific words and expressions as they are used in the manual. -+?>398S]_]ONSXWKX_KV^YSNOX^SPc_]O\aK\XSXQ]KXNMK_^SYX]   89>/S]_]ONSX^ROWKX_KV^YSNOX^SPcYZO\K^SXQKXNKZZVSMK^SYX]KN`SMOKXNKNNS^SYXKVObZVKXK^SYX] [ ] indicates direct function keys (e.g., [RANGE]). < > indicates PPC2+ screen displays (e.g., <1yes>).

PPC2+™ Operation and Maintenance Manual

Page 1 ©2001 DH Instruments, Inc.  38><9.?->398 1.1 PRODUCT OVERVIEW PPC2+ is a stand-alone, microprocessor driven, pressure controller/calibrator intended to apply and measure accurate values of gas pressure for the calibration and testing of pressure measuring instruments. It has been designed to provide very high performance combined with maximum versatility and ease of use. PPC2+ measures pressures using one or two high accuracy reference pressure transducers (RPTs) and an on-board barometer. Pressures control is achieved by a pneumatic module based on solenoid valves and differential pressure regulators. PPC2+ can be controlled locally by the operator using a front panel display, keypad and special function keys or remotely by a computer using ASCII character command strings over an RS-232 or IEEE-488 interface. PPC2+ models are available to measure and control pressure in ranges up to 0 to 1 500 psi (0 to 10 MPa) in both gauge and absolute measurement modes.

PPC2+™ Operation and Maintenance Manual

©2001 DH Instruments, Inc. Page 2  1.2 SPECIFICATIONS 1.2.1 GENERAL SPECIFICATIONS Power Requirements: 85 to 264 VAC, 50/60 Hz, 30 VA max consumption Operating Temperature Range: 15 to 35 °C Storage Temperature Range: -20 to 70 °C Vibration: Meets MIL-T-28800D Weight: 12.8 kg (28.2 lb) Dimensions: 7.1 in. H x 12.6 in. W x 15.8 in. D (18 cm x 32 cm x 40 cm) Microprocessor: Motorola 68302, 16 MHz Communication Ports: RS-232 (COM1), RS-232 (COM2), IEEE-488 Pressure Ranges: Up to six pressure ranges from 0 to 5 psi (33 kPa) to 0 to 1 500 psi (10 000 kPa) using one or two reference pressure transducers and a barometer Operating Medium: Any clean, dry, non-corrosive gas Pressure Connections: Supply: 1/8 in. NPT F Test(+): 1/4 in. NPT F Test(-): 1/4 in. NPT F (with gauge RPT only) Vent: 1/4 in. NPT F Exhaust: 1/4 in. NPT F Atm Ref: Pass through Pressure Limits: Maximum Working Test Pressure: Range H3+5 % Maximum Test Pressure without Damage: Range H3+10 % Recommended Supply Pressure: Range H3+10 % Maximum Supply Pressure without Damage: If Range H3>1 000 psi (7 000 kPa): 2 000 psi (14 000 kPa) If 300 psi (2 000 kPa) ) or two (<6 ranges>) Reference Pressure Transducers (RPTs) and a barometer. All RPTs with ranges over 30 psi (200 kPa) are of the absolute pressure type using an evacuated, permanently sealed reference. Absolute RPTs can measure both absolute and gauge pressure. Gauge pressures with an absolute RPTs are defined by offsetting atmospheric pressure with dynamic compensation for atmospheric changes using the on- board barometer. Gauge RPTs cannot measure absolute pressure.

PPC2+™ Operation and Maintenance Manual

Page 3 ©2001 DH Instruments, Inc.  Table 1. Reference Pressure Transducer (RPT) Designations and Ranges 1.2.2.1 REFERENCE PRESSURE TRANSDUCER (RPT) SPECIFICATIONS   +66@+6?/=+2/+->3@/<+81/?86/==9>2//. Warm Up Time: None required Resolution: To 1 ppm, user settable by individual range Temperature Effect: Fully compensated with active independent temperature measurement from -20 to 100 °C ± 0.005 % maximum temperature effect in normal ambient 15 to 35°C operating range Acceleration Affect: ± 0.008 % /g maximum, worst axis Allows operation at ± 20° from reference plane without significant effect Precision 1: ± 0.005 % Stability: 90 day 1 year Gauge Mode (w/Autozero): 0.003 % 0.009 % Absolute Mode (w/Autozero): 0.003 % 0.009 % Absolute Mode (w/out Autozero): 0.006 % 0.015 % Measurement Accuracy 2: 90 day 1 year Gauge Mode (w/Autozero): 0.008 % 0.012 % Absolute Mode (w/Autozero): 0.008 % 0.012 % Absolute Mode (w/out Autozero): 0.010 % 0.017 %  1 Measurement Precision: Combined linearity, hysteresis, repeatability of measurements made by the reference pressure transducer. 2 Measurement Accuracy: Maximum deviation of the reference pressure transducer indication from the true value of measured pressure including precision, stability, temperature effect and calibration standard accuracy of ± 0.0035 % of reading.  NOTE: When using an absolute reference pressure transducer for gauge mode measurements in a gauge range of < 30 psi (200 kPa), add ± 0.001 psi (8 Pa) to the measurement specification to take into account the resolution of the on-board barometer.

PPC2+™ Operation and Maintenance Manual

©2001 DH Instruments, Inc. Page 4  1.2.2.2 ON-BOARD BAROMETER The barometric sensor is NOT used as a source of absolute accuracy. It is used ONLY to measure changes in atmospheric pressure to provide dynamic compensation of the RPTs atmospheric pressure offset when using an absolute RPT to make gauge pressure measurements. 1.2.3 PRESSURE CONTROL SPECIFICATIONS  +VV`KV_O]K\OÝ0=YP^ROKM^S`O\KXQO_XVO]]Y^RO\aS]OSXNSMK^ON Modes and Ready (<*>) Indication: Static: Sets pressure within hold limit and stops active control until hold limit is exceeded Pressure is Ready (<*>) when within hold limit and stability test is met Dynamic: Sets pressure within hold limit and continuously adjusts as close as possible to target Pressure is Ready (<*>) when within hold limit Control Parameters: Hold limit, stability limit (optimum values set by default, can be customized independently for each measurement range) Table 2. Default Pressure Control Parameters Lo REFERENCE TRANSDUCER RANGE 1, 2, 3 Hi REFERENCE TRANSDUCER RANGE 1, 2, 3 HOLD LIMIT STABILITY LIMIT HOLD LIMIT STABILITY LIMIT STATIC MODE ± 1 % FS of the active pressure range ± 0.005 % FS/sec of the active pressure range ± 1 % FS of the active pressure range ± 0.005 % FS/sec of the active pressure range. DYNAMIC MODE ± 0.005 % FS of the active pressure range or ± 2 ppm of range H3, whichever is larger ± 0.005 % of the active pressure range ± 0.005 % FS of the active pressure range ± 0.005 % FS of the active pressure range

PPC2+™ Operation and Maintenance Manual

Page 5 ©2001 DH Instruments, Inc.   Control Precision 1: ± 0.001 % (with 30:1 maximum ratio between highest and lowest range) Normal Test Volume: Lo (max range < 300 psi): 100 to 1 000 cc (500 cc optimal) Hi (max range ˜ 300 psi): 0 to 500 cc (250 cc optimal) Control Speed: Slew Rate (0 to Controller FS) in Optimal Volume with No Control: 10 to 30 seconds Pressure Setting (Typical Time To Ready (<*>) Indication in Dynamic Mode): 20 to 30 seconds (lower for low, higher for high ranges) Reduced by increasing hold limit Minimum Control Pressure: Gauge Mode: None when EXHAUST at vacuum Absolute Mode: 0.2 % FS of range H3, will also set zero absolute (see Section 3.2.3.1). Delivered Pressure Accuracy 2: 90 day 1 year Gauge Mode: 0.010 % 0.013 % Absolute Mode (w/Autozero): 0.010 % 0.013 % Absolute Mode (w/out Autozero): 0.012 % 0.018 %  1 Control Precision: Minimum useable hold limit in dynamic control mode. 2 Delivered Pressure Accuracy: Maximum deviation from the true value of pressure applied to the device under test in dynamic control mode with default control limits and assuming measurement accuracy as in Section 1.2.2.1. In static control mode, control errors can be eliminated making accuracy of delivered pressure the same as measurement accuracy.  1.2.4 ANALOG INPUT SPECIFICATION Table 3. Analog Input Specifications MEASUREMENT RANGE DISPLAY RESOLUTION A/D RESOLUTION ACCURACY (+/- % FS) 0 to 100 mV 0.001 0.014 0.05 0 to 5 V 0.001 0.7 0.03 0 to 10 V 0.001 0.14 0.03 4 to 20 mA 0.001 0.003 0.05 24 V Power Source: Four channels, each output independently can sink 250 mA at 24 V, but maximum output of all four channels must not exceed 250 mA.

PPC2+™ Operation and Maintenance Manual

Page 7 ©2001 DH Instruments, Inc.  38=>+66+>398 2.1 UNPACKING AND INSPECTION 2.1.1 REMOVING FROM PACKAGING PPC2+ is delivered in a custom corrugated container with high density polyethylene inserts to hold it in place; or in the optional molded medium density polyethylene shipping case with a custom foam insert for holding the PPC2+ and SPLT (if included). Remove the PPC2+ and its accessories from the shipping container and remove each element from its protective plastic bag. 2.1.2 INSPECTING CONTENTS Check that all items are present and have NO visible damage. A standard PPC2+ includes all items indicated in Table 4. Table 4. PPC2+ Packing List DESCRIPTION PART NO. PPC2+ Pressure Controller/Calibrator FAM0002 Calibration Certificate 550100 Test Report 550103 Accessories: 401357 or 401357-CE Operation and Maintenance Manual 550091 Power Cord (7.5 ft.) 100770 or 100770-CE (6) Rubber Feet Caps 400203 General Accessories Disk (Important: Includes system support software and documentation.) 102987

PPC2+™ Operation and Maintenance Manual

©2001 DH Instruments, Inc. Page 8  2.2 SITE REQUIREMENTS The PPC2+ can be installed on any flat, stable surface at a convenient height. The front feet can be extended so that the unit can be inclined slightly for easier viewing. The PPC2+ can also be mounted in a standard 19-in. rack mount using the optional rack mount kit. Minimizing the length of tubing connecting the PPC2+ to the device or system to be tested will enhance control performance and reduce pressure setting times. Ready access to the PPC2+ rear panel should be considered to facilitate making and breaking pressure connections. The Self Purging Liquid Trap (SPLT), if used, should be mounted vertically at the low point of the pneumatic system. Support facilities required include: • An electrical power source of 85 to 264 VAC, 47 to 440 Hz. • A continuous regulated pressure supply of clean, dry, non-corrosive gas at maximum PPC2+ range (H3) +10 % to be connected to the SUPPLY port. Lower gas pressure supply can be used but should exceed the maximum desired test output pressure by 10 to 20 %. • A vacuum source of less than 1 psia (7 kPa) and with displacement of at least 3 cfm (90 lm) if control of pressures under 3 psi (20 kPa) gauge is desired. 2.3 INITIAL SETUP 2.3.1 PREPARING FOR OPERATION To prepare PPC2+ for check out and operation: • Remove the plastic caps from the PPC2+ rear panel pressure connections. • Remove the protective plastic sheet from the front panel display. • Install the rubber feet caps onto the bottom case feet, if desired. • Familiarize yourself briefly with the front and rear panel (see Section 2.3.2).

PPC2+™ Operation and Maintenance Manual

Page 9 ©2001 DH Instruments, Inc.  2.3.2 FRONT AND REAR PANELS 2.3.2.1 FRONT PANEL The front panel assembly provides a 2 x 20 vacuum fluorescent display of PPC2+ operating status and a membrane keypad for local user control. 1. Multi-Function Keypad 3. Remote Indicator 2. Direct Pressure Control Keys 4. Display Figure 1. Front Panel 2 3 4

PPC2+™ Operation and Maintenance Manual

©2001 DH Instruments, Inc. Page 10  2.3.2.2 REAR PANEL ASSEMBLY The rear panel assembly provides pressure connections, communications interfaces, the power ON/OFF module and product labeling. Pressure fittings are internally secured to prevent loosening when making and breaking connections. 1. Label, Product 9. Fuse 2. Pressure Connection, SUPPLY 10. Pressure Connection, TEST 3. Analog Option Connector 11. Pressure Connection, TEST(-) 4. Drivers Option Connector (models with gauge RPT only) 5. IEEE-488 Connector 12. Pressure Connection, EXHAUST 6. COM1 Connector 13. Pass Through, ATM REF 7. COM2 Connector 14. Pressure Connection, VENT 8. Power Switch Figure 2. Rear Panel Assembly 2.3.3 POWER CONNECTION • Check that the PPC2+ power switch is OFF. • Connect the supplied power cable to the rear panel power module. • Connect the other end of the power cable to an electrical supply of 85 to 264 VAC, 47 to 440 Hz. 5 10 11 12 6 7 8 9 13 14

PPC2+™ Operation and Maintenance Manual

Page 11 ©2001 DH Instruments, Inc.  2.3.4 CONNECTING TO A PRESSURE SUPPLY (SUPPLY PORT) Using a pressure connecting hose of appropriate pressure rating, connect the pressure supply to the SUPPLY connection on the rear panel of PPC2+. The PPC2+ SUPPLY connection is 1/8 in. NPT female. The supply pressure should be equal to the maximum PPC2+ range +10 %. Lower gas pressure sources can be used but should exceed the maximum desired test output pressure by 10 to 20 %.  8O`O\ MYXXOM^ K Z\O]]_\O ]_ZZVc Q\OK^O\ ^RKX   Y`O\ ^RO WKbSW_W \KXQO 2 YP ^RO ::-WYNOVcY_K\O_]SXQ 2.3.5 CONNECTING A VACUUM PUMP (EXHAUST PORT) For PPC2+ to set pressures under atmosphere and/or to reliably set pressure under 3 psi (20 kPa) gauge other than zero gauge, a vacuum pump must be connected to the EXHAUST port. 8O`O\MYXXOM^KZ\O]]_\O]_ZZVc^YY\ZV_Q^RO::-/B2+?=>ZY\^   >Y K`YSN L_SVNSXQ _Z Z\O]]_\O YX ^RO /B2+?=> ZY\^ KXNY\ ^RO `KM__W Z_WZ ^RO `KM__W ]Y_\MO]RY_VNOS^RO\LOMYX^SX_Y_]Vc98Y\]RY_VNLcZK]]^YK^WY]ZRO\OaROX^RO`KM__W ]Y_\MOS]900>RS]S]LOMK_]OaROXK]_ZZVcZ\O]]_\OS]KZZVSON^Y^RO::-=?::6CZY\^ KXN^RO::-S]89>SX^RO`OX^98MYXNS^SYX^RO\OS]KMYX]^KX^QK]ObRK_]^^R\Y_QR^RO ::-/B2+?=>ZY\^^YaRSMR^RO`KM__W]Y_\MOaSVVLOMYXXOM^ON   >YK]]_\OYZ^SW_WZ\O]]_\OMYX^\YVaROXMRKXQSXQ^ROZ\O]]_\OKZZVSON^Y^RO/B2+?=>ZY\^ P\YW`KM__W^YK^WY]ZRO\OY\`SMO`O\]KLO]_\O^YMRKXQO^ROMYX^\YV\OPO\OXMO]O^^SXQSP ^RO]O^^SXQS]89>SX+?>9WYNO]OO=OM^SYX  2.3.6 CONNECTING TO THE DEVICE UNDER TEST (TEST AND TEST(-) PORTS) If you are using a SPLT, see Section 2.3.6.1 before proceeding to connect the device under test. Using a pressure connecting hose of appropriate pressure rating, connect the device or system to be tested to the TEST port of the PPC2+. The PPC2+ TEST connection is 1/4 in. NPT female. If PPC2+ is equipped with a gauge RPT (GXXXX), it has a TEST port and a TEST(-) port. The TEST(-) port is connected to the low side of the gauge RPT. This connection is normally left open to atmosphere. It can also be connected to the low side of a differential device that is being calibrated or tested. With very low pressure devices, this will enhance the calibration results by helping assure that PPC2+ and device under test reference ports are at the same pressure.

PPC2+™ Operation and Maintenance Manual

©2001 DH Instruments, Inc. Page 12   8/@/<KZZVcZ\O]]_\O^Y^RO>/=>ZY\^aS^RY_^RK`SXQKZ\O]]_\O]_ZZVcO[_KV^YY\Q\OK^O\ ^RKX ^RO KZZVSON Z\O]]_\O MYXXOM^ON ^Y ^RO =?::6C ZY\^  AROX MYX^\YVVSXQ Z\O]]_\O Ob^O\XKVVc Y\ WOK]_\SXQ Ob^O\XKV Z\O]]_\O ^R\Y_QR ^RO =?::6C ZY\^ 8/@/< MK_]O ^RO Z\O]]_\O^YMRKXQOK^K`O\c\KZSN\K^O0Y\ObKWZVONYXY^`OX^]_NNOXVcLcYZOXSXQKX Ob^O\XKV`KV`O3X^O\XKVNKWKQO^Y^RO::-WKc\O]_V^   8/@/< MYXXOM^ K Z\O]]_\O ]_ZZVc ^Y ^RO >/=> ZY\^  >RO Z\O]]_\O KZZVSON ^Y ^RS] ZY\^ ]RY_VNLOWKSX^KSXONK^K^WY]ZRO\SMZ\O]]_\OÝZ]SU:K/bMOONSXQ^RO]OVSWS^]WKc NKWKQO^RO<:>   9ZO\K^SXQ^RO::-MYXXOM^ON^YK]c]^OWaS^RVS[_SNMYX^KWSXKX^]aS^RY_^^KUSXQZ\YZO\ Z\OMK_^SYX]^YZ_\QO^RO]c]^OWKXN^O]^VSXOWKcMK_]OMYX^KWSXK^SYXYP^RO::-^RK^ aSVV\O[_S\OXYXaK\\KX^c]O\`SMO   7SXSWSdSXQ ^RO VOXQ^R YP ^RO ^O]^ MYXXOM^SYX ^_LSXQ aSVV OXRKXMO MYX^\YV ZO\PY\WKXMO KXN \ON_MOZ\O]]_\O]O^^SXQ^SWO0Y\XY\WKVYZO\K^SYX^RO^Y^KV`YV_WOYP^RONO`SMOY\]c]^OW _XNO\^O]^SXMV_NSXQMYXXOM^SXQ^_LSXQ]RY_VNLOVO]]^RKXMM SX_Z^YZ]S U:KKXNVO]]^RKXMMSXY`O\Z]SU:K   7SXSWSdSXQ ^RO VOXQ^R YP ^RO ^O]^ MYXXOM^SYX ^_LSXQ aSVV OXRKXMO MYX^\YV ZO\PY\WKXMO KXN \ON_MOZ\O]]_\O]O^^SXQ^SWO0Y\XY\WKVYZO\K^SYX^RO^Y^KV`YV_WOYP^RONO`SMOY\]c]^OW _XNO\^O]^SXMV_NSXQMYXXOM^SXQ^_LSXQ]RY_VNLOVO]]^RKXMM SX_Z^YZ]S U:KKXNVO]]^RKXMMSXY`O\Z]SU:K  ::- Z\O]]_\O MYX^\YV aSVV XY^ YZO\K^O Z\YZO\Vc SP ^RO\O K\O ObMO]]S`O VOKU] SX ^RO ^O]^]c]^OW  3X QOXO\KV ^RO WKbSW_W KMMOZ^KLVO VOKU \K^O PY\ YZ^SWKV ::- K_^YWK^ON Z\O]]_\OMYX^\YVYZO\K^SYXKXN^YK]]_\OSX^YVO\KXMOWOK]_\OWOX^]aS^RNOPK_V^Z\O]]_\O MYX^\YVZK\KWO^O\]S]ÝYPKM^S`O\KXQO0=WSX_^O3X.C8+73--98><96WYNO^YRKXNVO RSQRO\ ^O]^ ]c]^OW VOKU \K^O] SXM\OK]O ^RO RYVN VSWS^ _]SXQ -?=>97 -98><96 ]OO =OM^SYX    ::- Z\O]]_\O MYX^\YV WKc LO KN`O\]OVc KPPOM^ON SP ^RO ^O]^ MYXXOM^SYX ^_LSXQ S] ^YY\O]^\SM^S`O  0Y\ YZ^SW_W \O]_V^] ^RO SXXO\ NSKWO^O\ YP ^RO MYXXOM^SXQ RY]O]RY_VNLO (!SX!WW  2.3.6.1 INSTALLING THE SPLT The SPLT (optional) is intended to collect and exhaust liquid that may be present in the device or system under test so that they do not return to contaminate the PPC2+. The SPLT is installed in the test connection line at a low point between PPC2+ and the device or system under test. See the SPLT manual for more complete instructions on SPLT installation.

PPC2+™ Operation and Maintenance Manual

Page 13 ©2001 DH Instruments, Inc.  2.3.7 THE VENT CONNECTION (VENT PORT) The PPC2+ VENT connection is the system vent to atmosphere point used to set zero gauge pressure as well as to obtain reference pressure transducer measurements of atmospheric pressure. Though a pressure hose can be connected to the VENT connection to direct the vented gas flow, a completely unobstructed connection to atmosphere must be maintained for PPC2+ reference pressure measurements to operate normally. The PPC2+ VENT connection is 1/4 in. NPT female.  8/@/<ZV_QYL]^\_M^Y\MYXXOM^K]_ZZVcZ\O]]_\O^Y^RO::-`OX^MYXXOM^SYX 2.3.8 THE ATM REF PASS THROUGH (UNLABELED) The ATM REF pass through is connected to the on-board barometer and to the valve that opens the Lo RPT (if present) to atmosphere when it is not in use. The ATM REF pass through assures that the Lo RPT and on-board barometer are truly subject to atmospheric pressure rather than to the pressure inside the PPC2+ case which may be slightly different from ambient pressure.  .Y XY^ ZV_Q Y\ YL]^\_M^ ^RO +>7  2.4 POWER-UP AND VERIFICATION 2.4.1 SWITCH POWER ON Actuate the power switch on the PPC2+ rear panel. Observe the front panel display as PPC2+ initializes, error checks and goes to the main run screen (see Section 3.1.1). PPC2+ operating condition at power-up is VENT valve closed unless the pressure measured by the RPT was within 3 psi (20 kPa) of atmospheric pressure. If the PPC2+ fails to reach the main run screen, service is required. Record the sequences of operation and displays observed.  3P^ROKM^S`O\KXQOYXZYaO\NYaXaK]K2S\OPO\OXMO^\KX]N_MO\\KXQO^RO]KWO\KXQOaSVVLO KM^S`OYX^ROXOb^ZYaO\_Z3P^ROKM^S`OZYaO\NYaX\KXQOaK]K6Y\OPO\OXMO^\KX]N_MO\ \KXQO\KXQO2aSVVLOKM^S`OYXZYaO\_Z>RS]S]^YZ\Y^OM^^RO6Y\OPO\OXMO^\KX]N_MO\P\YW KMMSNOX^KVY`O\Z\O]]_\O]OO=OM^SYX

PPC2+™ Operation and Maintenance Manual

©2001 DH Instruments, Inc. Page 14  2.4.2 CHECK PROPER PRESSURE MEASUREMENT OPERATION 2.4.2.1 CHECKING ABSOLUTE MODE PRESSURE MEASUREMENT If the PPC2+ has an absolute RPT (designated AXXXX), check that it operates properly in absolute mode. If the PPC2+ is not vented (vent LED OFF), press the [VENT] direct pressure control key to vent the PPC2+ (VENT LED ON) (see Section 3.1.2.2). Using the [RANGE] function key to change ranges if necessary (see Section 3.2.1) select a range of the absolute RPT. Press the [UNIT] function key and select mode. Change the pressure unit if desired (see Section 3.2.2). Observe the current value of atmospheric pressure. Check that the value agrees with the local value of atmospheric pressure. Repeat this process for all the ranges on both RPTs if the PPC2+ has two absolute RPTs. Check that the values of atmospheric pressure measured by the different ranges agree with each other within PPC2+ measurement tolerances (see Section 1.2.2). If they do not agree within tolerances, PPC2+ may need calibration or repair. 2.4.2.2 CHECKING GAUGE MODE PRESSURE MEASUREMENT If the PPC2+ is not vented, press the [VENT] direct pressure control key to vent it (see Section 3.1.2.2). Press the [UNIT] function key and select mode. Change the pressure unit if desired (see Section 3.2.2). Observe that, within ten seconds, zero is indicated. It is normal for PPC2+ to indicate a value other than zero for up to ten seconds when first entering gauge mode. Using the [RANGE] function key to change ranges, observe that zero is indicated for each range within 10 seconds. It is normal for PPC2+ to indicate a value other than zero when vented when gauge mode is first entered or ranges are changed. After about ten seconds, the VENT LED should flash and zero should be indicated. If this does not occur, check that the AUTOZERO function is ON (see Section 3.4.1). If AUTOZERO is ON and display will not zero when vented, PPC2+ may need repair. 2.4.3 LEAK CHECK If desired, perform a leak check of the test system (see Section 3.2.8).

PPC2+™ Operation and Maintenance Manual

Page 15 ©2001 DH Instruments, Inc.  2.4.4 PURGE If the SPLT is included and installed in the test line, perform a purge of the Device Under Test (DUT). This will rid the DUT of contaminated liquids and enable descending pressures to be controlled (see Section 3.2.8).  9ZO\K^SXQ^RO::-MYXXOM^ON^YK]c]^OWaS^RVS[_SNMYX^KWSXKX^]aS^RY_^^KUSXQZ\YZO\ Z\OMK_^SYX]^YZ_\QO^RO]c]^OWKXN^O]^VSXOWKcMK_]OMYX^KWSXK^SYXYP^RO::-^RK^ aSVV\O[_S\OXYXaK\\KX^c]O\`SMO 2.4.5 CHECK PROPER PRESSURE CONTROL OPERATION Select a pressure range using the RANGE function (see Section 3.2.1). Press the [CONTROL] function key and select (see Section 3.2.3). Press [ENTER]. Specify a target pressure within the active range and press [ENTER] again (see Section 3.2.3). PPC2+ should set the target pressure and indicate Ready (<*>) (see Section 3.1.2.4) continuously in 15 to 60 seconds. If it does not, see Section 6. 2.5 SHORT TERM STORAGE The following is recommended for short term storage of PPC2+: • Vent the PPC2+ test pressure. • Turn the power OFF using the rear panel power switch. • Shut OFF or disconnect the pressure supply. Shut OFF or disconnect the vacuum supply.

PPC2+™ Operation and Maintenance Manual

Page 17 ©2001 DH Instruments, Inc.  9:/<+>398 3.1 GENERAL/MANUAL OPERATION PPC2+ is designed to offer a balance between simple, straight forward operation and the availability of a wide variety of functions with a high level of operator discretion, if desired. The local operator interface is through a 2 x 20 display, a 4 x 4 keypad and four direct pressure control keys. 3.1.1 MAIN RUN SCREEN The PPC2+ main run screen is its home display that is reached on power-up and from which other functions and menus are accessed. It is the very top level of all menu structures. The main run screen is where the operator works with PPC2+ to set and read pressures. It provides complete information on the system’s current configuration and operating status. 1. Ready/Not Ready Indication (see Section 3.1.2.4): <*> when ready, <↑> or <↓> indicating direction of measured pressure evolution if not ready. 2. PRESSURE1: Numerical value and sign of pressure measured by active RPT and range. 3. UNIT Current unit of measure of pressure indication (see Section 3.2.2). 4. Pressure measurement mode: gauge, absolute (see Section 3.2.2). 5. Indicates whether a head correction is applied. if applied, < > (blank) if not (see Section 3.2.7). 6. Indicates whether the autozero function is ON or OFF. if ON; < > (blank) if OFF (see Section 3.4.1). 7. SR: Indicates active RPT ( for high, for low) and range (<1> = low, <2> = mid, <3> = hi) (see Section 3.2.1). 8. Used only when running a sequence. Indicates number of current increment over total number of increments in the sequence or [ENTER] when ready. Changes to countdown time during count down when next increment of sequence is on timed delay rather than enter. 9. Pressure control mode ( for static, for dynamic) with a appended if custom control settings are in use. These control indicator characters flash when the system is controlling. In static mode, if control is active they flash even if NO valve is operating (e.g., while inside the hold limit and waiting for stability). Characters flash while controlling pressure down to vent but NOT when vent valve is open (see Sections 3.2.3 and 3.2.6). 10. PRESSURE2: Numerical sign and value of either target pressure , rate of change of pressure or deviation from target . Target value and deviation are in the same pressure unit as the current unit in Step 3 of this section. Rate of change of pressure is in current unit per second (see Section 3.2.3). 2 3 7 TPRESSURE2 CC NN/NN 7

PPC2+™ Operation and Maintenance Manual

©2001 DH Instruments, Inc. Page 18  11. Indicates whether the number following is target pressure , rate of change of pressure or current deviation from target value ) (see Section 3.2.3). Target pressure is indicated when: • Dynamic control is active and pressure is Not Ready (<↑> or <↓>) • Static control is active and control valves are operating. Deviation is indicated when: • Dynamic control is active and pressure is Ready (<*>). Rate of change is indicated when: • No control is active (e.g., control is suspended or system is vented.) • Static control is active, and NO valves are active.  ::-RK]K]M\OOX]K`O\P_XM^SYXaRSMRMK_]O]^RONS]ZVKc^YNSWSPXYUOcS]Z\O]]ONPY\ WSX_^O]:\O]]SXQKUOc\O]^Y\O]P_VVZYaO\^Y^RONS]ZVKc>RO]M\OOX]K`O\^SWOMKXLO MRKXQONY\]M\OOX]K`SXQMKXLOMYWZVO^OVc]_ZZ\O]]ON]OO=OM^SYX 3.1.2 GENERAL OPERATING PRINCIPLES 3.1.2.1 MAIN 4 X 4 KEYPAD LAYOUT AND PROTOCOL The PPC2+ has a 4 x 4 keypad for local operator access to direct functions, function menus and for data entry. 1. The Function/Data keys allow very commonly used functions to be accessed directly by a single keystroke when pressed from the main run screen. The name of the function is on the bottom half of the key (see Section 3.2). These keys enter numerical values when editing. 2. The Editing and Execution keys are for causing execution, suspending execution, backing up in menus and editing entries. 3. The Menu/Data keys provide access to function menus when pressed from the main run screen. The menu name is on the bottom half of the key. The SETUP menu is for more frequently used functions (see Section 3.3). The SPECIAL menu is for functions that are not generally used as a part of day to day operation (see Section 3.4). These keys enter numerical values when editing. Figure 3. Keypad Layout 2 . SETUP 0 SPECIAL 1 2 LEAK CHECK 3 ANALOG 4 UPPER LIMIT 5 SEQ 6 CUSTOM CONTROL 7 RANGE 8 9 + / - ESCAPE ENTER CONTROL 3

PPC2+™ Operation and Maintenance Manual

Page 19 ©2001 DH Instruments, Inc.  Pressing the [ENTER] key generally causes execution or forward movement in the menu tree. Pressing [ENTER] from the main run screen allows an automated pressure control command to be given. Pressing the [ESCAPE] key moves back in the menu tree and/or causes execution to cease or suspend. Pressing [ESCAPE] repeatedly always eventually brings you back to the main run screen and from there will allow a momentary viewing of the PPC2+ introduction screen. Pressing the [+/-] key changes a numerical sign when editing. It also toggles through multiple screens when available. Pressing the [←] and [→] keys allow reverse and forward cursor movement when editing data entry. They also allow the target pressure value to be jogged up and down while controlling or when editing target value.  =YWO]M\OOX]QYLOcYXN^RO^aYVSXO]Z\Y`SNONLc^RONS]ZVKc>RS]S]SXNSMK^ONLcK PVK]RSXQK\\YaSX^RO]OMYXNVSXOYP^RONS]ZVKc:\O]]^ROE←GKXNE→GUOc]^YWY`O ^ROM_\]Y\^YKMMO]]^ROVSXO]^RK^K\OXY^`S]SLVOY\NS\OM^VcOX^O\^ROX_WLO\YP^RO RSNNOXWOX_MRYSMOSPcY_UXYaS^ 3.1.2.2 DIRECT PRESSURE CONTROL KEYS The direct pressure control keys provide direct manual control of pressure increase, decrease and vent. They can be useful in adjusting pressure when automated pressure control to a target value is not needed. The direct pressure control keys interrupt and override automated pressure control. Pressing the [VENT] key causes PPC2+ to control pressure to near atmospheric pressure and then open the system vent valve (see Section 5.8.1.7). Vent valve open is indicated by lighting a RED LED just above the [VENT] key. The vent valve will remain open until the [VENT] key is pressed again, another direct pressure control key is pressed, or an automated pressure control command is given. Pressing the [VENT] key also causes the purge function to execute when the PPC2+ LEAK CHECK/PURGE function is active (see Section 3.2.8).  >RO::-,1@/8>]O[_OXMONSPPO\]P\YWY^RO\::-WYNOV]N_O^Y^RO_]O YP K >/=>>/=> LcZK]] `KV`O  =OO ^RO ::- ,1 9ZO\K^SYX KXN 7KSX^OXKXMO7KX_KV=OM^SYX:<38-3:6/ Pressing the [↑] and [↓] keys cause pressure to increase or decrease at the slow slew rate. Pressing the [FAST] key while pressing an [↑] or [↓] key will cause the pressure increase or decrease speed to change from slow to fast. Figure 4. Direct Pressure Control Keys

PPC2+™ Operation and Maintenance Manual

©2001 DH Instruments, Inc. Page 20  3.1.2.3 AUTOMATED PRESSURE CONTROL PPC2+ automated pressure control provides automated adjustment and control of pressure to a user designated target value. This feature allows pressure values to be set simply and precisely by simple numerical entry. Pressing [ENTER] from the main run screen allows a pressure control target value to be entered and executed. Pressing [ESCAPE] or a direct pressure control key will cause pressure control to be interrupted. PPC2+ supports two pressure control modes to meet different pressure setting and controlling requirements: static and dynamic. Pressure control parameters for each control mode are automatically set to optimal default values for each PPC2+ range when the control mode is selected in the active range (see Section 3.2.3). If desired, control parameters can be customized by the user using the CUSTOM CONTROL function (see Section 3.2.6). Control parameters: • Target Value: The pressure setpoint specified by the operator. • Hold Limit: A symmetrical positive and negative limit around the target value within which the controlled pressure is to be maintained. • Stability Limit: A rate of change of pressure limit in units of pressure/second used as a criterion for the Ready (<*>)/Not Ready (<↑> or <↓>) condition in static control or when automated control is not active. See Section 3.1.2.3, Static Control and Dynamic Control for a detailed explanation of each control mode and its advantages, the default control parameters and the Custom Control options.  +]O\SO]YPZ\O]]_\OMYX^\YV^K\QO^`KV_O]MKXLOZ\YQ\KWWONK^YXO^SWOSXMV_NO Y^RO\ Z\O]O^ YZO\K^SXQ ZK\KWO^O\] _]SXQ ^RO =/;?/8-/ P_XM^SYX ]OO =OM^SYX Static Control Static control mode is designed to set the pressure near the target value and then cease active control to allow pressure to stabilize naturally. The advantage of this control mode is that pressure can be set and/or measured without interference from a pressure control system. In a system without excessive leaks, the pressure stability achieved (especially at low pressures) may be greater than the stability with which the pressure control system can actively control pressure. Using static control to control pressures near the desired set point and then measuring back the stabilized pressure without interference of the control function can allow control errors to be eliminated. However, the final pressure achieved will be different from the target value. During static pressure control, the hold limit is active. If the pressure goes outside of the hold limit, a Not Ready (<↑> or <↓>) condition will occur (see Section 3.1.2.4) and pressure will be readjusted to the target value (see Section 3.2.3 for default hold and stability limit values and Section 3.2.6 for setting the hold and stability limits to user defined values).

PPC2+™ Operation and Maintenance Manual

Page 21 ©2001 DH Instruments, Inc.   >K\QO^:\O]]_\O : < / = = ? < / =O^^SXQ :\O]]_\O^Y >K\QO^@KV_O :\O]]_\OSX 2YVNVSWS^ 8Y:\O]]_\O -YX^\YV :\O]]_\O AOX^9_^ YP2YVN VSWS^ K\QO^ @KV_O :\O]]_\OSX 2YVN6SWS^ 8Y:\O]]_\O -YX^\YV >SWO 2YVN 6SWS^ Figure 5. Static Pressure Control Operation Dynamic Control Dynamic control mode is designed to set the pressure to the target value and control continuously to keep pressure within the hold limit and as close to the target value as possible. The advantage of this control mode is that the final pressure achieved will be the same as the target value. However, unlike static control, in dynamic control a control error will be included in the pressure achieved. The maximum value of the control error is equal to the hold limit. The average value of the control error is generally well inside the hold limit. During dynamic pressure control, the hold limit is active. If the pressure goes outside of the hold limit, a Not Ready (<↑> or <↓>) condition will occur. See Sections 3.1.2.4 and 3.2.3 for default hold limit values. See Section 3.2.6 for setting the hold limit.

PPC2+™ Operation and Maintenance Manual

©2001 DH Instruments, Inc. Page 22   :38?9?=6C+.4?=>/.>9>2/>+<1/>@+6?/ >SWO 2YVN 6SWS^ >K\QO^:\O]]_\O : < / = = ? < / Figure 6. Dynamic Pressure Control Operation 3.1.2.4 PRESSURE READY (<*>)/NOT READY (<↑> OR <↓>) The far left character of the main run screen provides a pressure Ready (<*>)/Not Ready (<↑> or <↓>) indication. This indication is intended to provide the user with a clear and objective criterion for determining when a valid pressure measurement can be made. Ready (<*>)/Not Ready (<↑> or <↓>) indications are: • <*>: Pressure Ready. • <↓>: Pressure Not Ready and decreasing. • <↑>: Pressure Not Ready and increasing. When pressure control is NOT active: A Ready (<*>) condition will occur any time NO control valve is operating and the pressure rate of change is inside the stability limit. The stability limit is defined in terms of rate of change of pressure in current pressure units per second. When pressure control is active: The criteria for determining the Ready (<*>)/Not Ready (<↑> or <↓>) condition depend on whether the current control mode is static or dynamic. Pressure Ready (<*>)/Not Ready (<↑> or <↓>) parameters are set by default when a control mode is selected and or can be customized if desired (see Sections 3.2.3 and 3.2.6).

PPC2+™ Operation and Maintenance Manual

Page 23 ©2001 DH Instruments, Inc.  Static Control Ready (<*>)/Not Ready (<↑> or <↓>) When static pressure control mode is active a Ready (<*>) condition will occur when: • NO control valve is operating. • The current measured pressure is inside the hold limit. • The rate of change of pressure is less than the current stability limit. >K\QO^:\O]]_\O : < / = = ? < / 8Y^K\QO^@KV_O 8Y^ SWO 2YVN 6SWS^ )/Not Ready (<↑> or <↓>) Static Control Mode Dynamic Control Ready (<*>)/Not Ready (<↑> or <↓>) When dynamic pressure control mode is active, a Ready (<*>) condition will occur when: • The current measured pressure is inside the hold limit. Figure 8. Ready (<*>)/Not Ready (<↑> or <↓>) in Dynamic Pressure Control Mode Ready (<*>)/Not Ready (<↑> or <↓>) When Automated Pressure Control is Not Active

PPC2+™ Operation and Maintenance Manual

©2001 DH Instruments, Inc. Page 24  When NO automated pressure control is active, Ready (<*>) is indicated whenever the rate of change of pressure is less than the current stability limit. 3.1.2.5 MULTIPLE PRESSURE RANGES PPC2+ has one or two reference pressure transducers (RPT) each of which has three ranges for a total of three or six pressure ranges. This multi-ranging feature allows accuracy to be optimized for the range of pressure in which you are working. Generally, the best range to select (see Section 3.2.1) is that whose full scale is closest to, but greater than, the maximum pressure of the device or system under test. PPC2+ handles all of the data and internal valving operations needed to make range changes occur transparently to the operator when the RANGE function is used to select a range. Ranges available can be viewed at any time, but changing ranges can only occur when PPC2+ is vented. When ranges are changed, the upper limit is automatically changed to the default for that range or to the last upper limit that was set for that range (see Section 3.2.4). In addition, most other functions and settings are range specific (see Section 3.2.1).   /KMR ::- RK] ^R\OO Y\ ]Sb \KXQO]  3X QOXO\KV _]O\ ]O^^SXQ] KXN YZO\K^SYXKV KNT_]^WOX^]K\O]ZOMSPSM^Y^RO\KXQOM_\\OX^VcSX_]OK]SPcY_RKN]SbSX]^\_WOX^] \K^RO\^RKXYXO The PPC2+ internal pneumatic layout for handling two RPTs if two RPTs are present is: Figure 9. Hi and Low RPT Pneumatic Circuit ATM REF Port N/C Valve N/O Valve to Pneumatic Module Hi Ref Lo Ref

PPC2+™ Operation and Maintenance Manual

Page 25 ©2001 DH Instruments, Inc.  Ranges and Identification The currently active reference transducer and range is continuously displayed in the upper right hand corner of the main run screen and most other screens. See Table 1 for a complete listing of the RPTs available and their ranges. Hi RPT: The RPT in a single RPT PPC2+ or the RPT with the highest maximum range in a dual RPT PPC2+ is referred to as the Hi RPT. Lo RPT: The RPT with the lower maximum range in a dual RPT PPC2+ is referred to as the Lo or secondary RPT. Range 1, 2 or 3: The three ranges of an RPT are referred to as 1 = lo range, 2 = mid range, 3 = hi range. Table 5. PPC2+ Range Identification Summary REFERENCE PRESSURE TRANSDUCER AND RANGE DESIGNATION DISPLAY SYMBOL * Lo RPT, Lo range Lo, 1 L1 Lo RPT, Mid range Lo, 2 L2 Lo RPT, Hi range Lo, 3 L3 Hi RPT, Lo range Hi, 1 H1 Hi RPT, Mid range Hi, 2 H2 Hi RPT, Hi range Hi, 3 H3 * The display symbol is included in the upper, right hand corner of most PPC2+ menu displays as a convenient indicator of active range. 3.2 DIRECT FUNCTION KEYS 3.2.1 RANGE m PURPOSE To view and/or change the pressure measurement range. m PRINCIPLE Each PPC2+ has three or six ranges (see Section 3.1.2.5). Pressing the [RANGE] function key allows the ranges to be viewed and selected, including automated switching of RPTs when necessary, with built-in logic to prevent accidental overpressure of the Lo RPT if present. PPC2+ functions and settings such as pressure unit of measure (UNIT), control mode (CONTROL) and upper limit (UPPER LIMIT) are range specific. Changes made while in one range apply to that range only and will not affect the other ranges. Only the HEAD function is not range specific (see Section 3.2.7).

PPC2+™ Operation and Maintenance Manual

©2001 DH Instruments, Inc. Page 26  m OPERATION Pressing the [RANGE] function key activates the range viewing and selecting function. Pressing the [+/-] key or the [RANGE] key again while in the RANGE functions steps through displays of available ranges: Lo to Hi. Pressing [ENTER] while in the RANGE function when PPC2+ is vented will cause the currently displayed range to become the active range. Pressing [ESCAPE] while in the range function returns to the main run screen with no range change having been made. When the [RANGE] function key is first pressed, the active reference pressure transducer (RPT) and range are displayed. For example: 1. Identifies RPT currently in use ( or ). 2. Identifies current range (<1>, <2> or <3>) of the RPT currently in use. 3. Range designator. 4. Full scale pressure value in the current units for the RPT and range number when used in gauge or absolute mode. If the RPT is a gauge-only RPT, there is no absolute range indicated. 5. Units of measure (the units of measure currently active for this range). Pressing the [+/-] key or pressing the [RANGE] key again causes the screen to step through the other available ranges in sequence: Lo to Hi. For example: 1. Identifies RPT ( or ). 2. Identifies range of the RPT (<1>, <2> or <3>). 3. Range designator. 4. Full scale pressure value in the active units of measure for the RPT and range number when used in gauge or absolute mode. If the RPT is a gauge-only RPT, there is no absolute range indicated. 5. Units of measure (the units of measure currently active for the range).  Y Z\Y^OM^ KQKSX]^ Y`O\Z\O]]_\O YP ^RO 6Y Z\O]]_\O \OPO\OXMO ^\KX]N_MO\ KXN K`YSN KMMSNOX^KVVcObMOONSXQ\KXQO_ZZO\VSWS^]^ROKM^S`O\KXQOMKXYXVcLOMRKXQONaROX^RO ]c]^OWS]`OX^ON3PE/8>/ `OX^ON^RONS]ZVKcaSVVSXNSMK^O&@OX^]c]^OWP_VVc^YMRKXQO\KXQO(:\O]]SXQ^ROE@/8>G UOc aSVV MK_]O ::- ^Y `OX^ KXN ^ROX \O^_\X ^Y ^RO <+81/ P_XM^SYX aROX ^RO @/8> S]MYWZVO^O 2 3 psi 35g/50a 2 3 psi 300g/300a

PPC2+™ Operation and Maintenance Manual

Page 27 ©2001 DH Instruments, Inc.  3.2.1.1 RANGE SPECIFIC FUNCTIONS AND SETTINGS In general, PPC2+ functions and settings are range specific. They are set and stored for each range so that changing them when in one range does not change them in the other ranges. When you return to a range, you will find the settings you left. Functions and settings that are NOT range specific are: • Functions: HEAD, ANALOG • Setup Menu: <1ControlRef>, <4Drivers> 3.2.2 UNIT m PURPOSE To specify the pressure unit and measurement mode in which PPC2+ displays pressure values. See also Section 3.4.3. m PRINCIPLE PPC2+ allows the pressure measurement unit and mode to be changed. Internally, PPC2+ always operates in Pascal (Pa), the SI unit of pressure. Values of pressure are represented in other units by the application of conversion factors to convert from Pa (see Section 7.3). RPTs designated AXXXX intrinsically measure absolute pressure relative to an evacuated and sealed reference. These can also be used to measure in gauge mode by the subtraction of atmospheric pressure. PPC2+ supports extensive on-board logic and measurements to assure the precise atmospheric compensation of absolute RPTs to deliver gauge measurements (see Section 3.4.1). This allows simple, one step switching between gauge and absolute measurement modes by the operator without special procedures or hardware. RPTs designated GXXXX intrinsically measure gauge pressure and can be used in gauge pressure measurement mode only. m OPERATION To change the pressure unit and/or the measurement mode, press the [UNIT] function key. The display is: Select <1> to change pressure units or <2> to change measurement mode. If a measurement mode change is not currently possible (if the active RPT is gauge only or if the current pressure unit is an altitude unit which can only be absolute), the <2mode> selection is not offered. If <2mode> was selected the next screen offers the choice between <1gauge> and <2absolute>. Making a selection returns to the main run screen with the selected measurement mode active. 1unit 2mode

PPC2+™ Operation and Maintenance Manual

©2001 DH Instruments, Inc. Page 28  If <1unit> was selected from the measurement screen, the units screen is presented. The cursor will be on the number corresponding to the active unit for the active range. To change the pressure unit for the active range, select the desired unit. Pressing [ENTER] activates the unit and returns to the main run screen. If the pressure unit selected is inWa, the reference temperature for water density must be specified. When inWa is selected as the unit, the next display is: Select the desired reference temperature for water density using the [←] or [→] key to move the cursor. Pressing [ENTER] returns to the main run screen with inWa based on water density at the selected reference temperature as the pressure unit. The current inWa reference temperature can be viewed by observing the position of the cursor in the reference temperature screen.  >RO\O K\O ^aY ZY]]SLVO NOPK_V^ Z\O]]_\O _XS^ ]M\OOX] YXO ?= KXN YXO =3 NOZOXNSXQ YX aRO^RO\^RO::-RK]LOOXPKM^Y\c]O^K]?=Y\=3ARO^RO\^RO::-S]]O^_ZK]?=Y\=3 S]SXNSMK^ONLc^RO\SQR^^aYMRK\KM^O\]YX^RO^YZVSXOYP^RO::-SX^\YN_M^SYX]M\OOXaRSMR MKXLO`SOaONLcZ\O]]SXQE/=-+:/GP\YW^ROWKSX\_X]M\OOXY\aROXZYaO\SXQ_Z::- 0Y\KX =3 ::- ^RO NOPK_V^ _XS^] K\O$  &U:K( &:K( &7:K( &LK\( &WLK\(& WWAK(   8Y \OPO\OXMO ^OWZO\K^_\O ]OVOM^SYX S] XOMO]]K\c PY\ ^RO _XS^ WWAK K] ^RO YXVc \OPO\OXMO ^OWZO\K^_\OMYWWYXVc_]ONPY\WWAKS]Ù-   >ROZ\O]]_\OWOK]_\OWOX^_XS^]OVOM^ONS]\KXQO]ZOMSPSMAROXSXKQS`OX\KXQOKVVP_XM^SYX] KXN ]O^^SXQ] aSVV LO \OZ\O]OX^ON SX ^RO M_\\OX^ WOK]_\OWOX^ _XS^ PY\ ^RK^ \KXQO 2YaO`O\MO\^KSXSX^O\XKVKXNY\WO^\YVYQSMKVP_XM^SYX]OQ<:>MKVSL\K^SYXMYOPPSMSOX^]K\O KVaKc]\OZ\O]OX^ONSX:K\OQK\NVO]]YP^ROM_\\OX^\KXQO_XS^3XKNNS^SYXaROX^ROM_\\OX^ _XS^ S] KX KV^S^_NO _XS^ ^RO \KXQO KXN _ZZO\ VSWS^ SXNSMK^SYX] K\O SX U:K SP ^RO _XS^ S] WO^O\]WKXNZ]SSP^RO_XS^S]POO^P^   =OO=OM^SYX!PY\^KLVO]YP^ROMYX`O\]SYXPKM^Y\]_]ONLc::-   >RONOPK_V^Z\O]]_\O_XS^]K`KSVKLVO_XNO\^RO?83>P_XM^SYXNOZOXNYXaRO^RO\^RO::-RK] LOOX]O^_ZK]KX=3Y\?=`O\]SYXSXNSMK^ONLc=3Y\?=K^^YZ\SQR^YPSX^\YN_M^SYX]M\OOX >ROMRYSMOYP]Sb_XS^]K`KSVKLVO_XNO\^RO?83>P_XM^SYXMKXLOWYNSPSONP\YWKaSNO\]OVOM^SYX Lc^RO_]O\]OO=OM^SYX>RO_XS^]K`KSVKLVO_XNO\^RO?83>P_XM^SYXMKXLO\O]O^^Y NOPK_V^ Lc Z\O]]SXQ E=:/-3+6G KXN ]OVOM^SXQ &3X^O\XKV( &

PPC2+™ Operation and Maintenance Manual

Page 29 ©2001 DH Instruments, Inc.  3.2.3 CONTROL m PURPOSE To set the automated pressure control mode for the currently active range and activate default control parameters. m PRINCIPLE The CONTROL function allows the active control mode for the current range to be set to either static control or dynamic control (see Section 3.1.2.3). The control mode set is specific to the active range and will be maintained with that range when ranges are changed. When the control mode is set using the CONTROL function, CONTROL parameters are set to system default parameters. The default control parameters have been determined to be those most suitable for the typical user to operate within PPC2+ pressure control and measurement specifications (see Section 3.2.3.1). If desired, control mode parameters can be customized using the CUSTOM CONTROL function (see Section 3.2.6). For a complete description of the operation and purpose of static and dynamic control modes (see Section 3.1.2.3). m OPERATION To select the control mode for the currently active range and set control parameters to the default control parameters, press the [CONTROL] function key from the main run screen. The display is: Selecting <1static> or <2dynamic> causes the selected pressure control mode to be activated with default control parameters and returns to the main run screen.  -YX^\YVWYNO]O^^SXQS]\KXQO]ZOMSPSM+MRKXQOSXMYX^\YVWYNOWKNOaRSVOSXYXO::- \KXQONYO]89>KPPOM^^ROMYX^\YVWYNO]O^^SXQSXY^RO\\KXQO]]OO=OM^SYX   >ROM_\\OX^MYX^\YVWYNOS]SXNSMK^ONLc^aYMRK\KM^O\]SX^ROWSNNVOYP^ROLY^^YWVSXO YP^ROWKSX\_X]M\OOX&=(PY\]^K^SMKXN&.(PY\NcXKWSMaS^R&-(KZZOXNONSPM_]^YW MYX^\YV]O^^SXQ]K\OSX_]O=OO=OM^SYX   3XNcXKWSMMYX^\YVWYNOaROXO`O\^ROZ\O]]_\OS]RS]S]LOMK_]OSXNcXKWSM MYX^\YV WYNO aROX ^RO Z\O]]_\O S] RS] S] XY^ ^\_O aROX ]O^^SXQ dO\Y SX KL]YV_^O WOK]_\OWYNO]OO=OM^SYX+_^YWK^ON:\O]]_\O-YWWKXN]PY\DO\Y:\O]]_\O Active Range 1dynamic 2static

PPC2+™ Operation and Maintenance Manual

©2001 DH Instruments, Inc. Page 30  3.2.3.1 SYSTEM DEFAULT CONTROL PARAMETERS Table 6. Lo or Hi Reference Transducer Range 1, 2, 3 HOLD LIMIT STABILITY LIMIT STATIC MODE ± 1 % FS of the active pressure measurement range ± 0.005 % FS/sec of the active pressure measurement range. DYNAMIC MODE ± 0.005 % FS of the active range or ± 2 ppm FS of range H3, whichever is larger ± 0.005 % FS/sec of the active pressure measurement range   ::- ,1 NOPK_V^ MYX^\YV VSWS^] K\O NSPPO\OX^ P\YW ]^KXNK\N ::- WYNOV NOPK_V^MYX^\YVVSWS^]=OO^RO::-,19ZO\K^SYXKXN7KSX^OXKXMO7KX_KV 1 3.2.3.2 EXECUTING AUTOMATED PRESSURE CONTROL COMMANDS m PURPOSE To use the automated pressure control functions of PPC2+ to set target pressure values. m OPERATION To set a pressure, press [ENTER] from the main run screen. The display is: 1. Ready (<*>)/Not Ready (<↑>) or <↓>) indication, current pressure, units, measurement mode, range (regular first line of main run screen). 2. Entry field for target pressure value to be set automatically. Comes up with last entry or zero if pressure measurement units have been changed, range has been changed, or system has been vented. Use the numerical keys and editing keys to enter the target pressure value desired (see Section 3.2.3.2.2) for information on zero pressure commands. Also, the [←] and [→] keys can be pressed to jog the current target pressure value displayed down or up by the current pressure control hold limit. Press [ENTER] to start pressure control to the target pressure and return to the main run screen. The main run screen is: 1. Ready (<*>)/Not Ready (<↑> or <↓>) indication, current pressure, units, measurement mode, range (regular first line of main run screen). 2. Current control mode indicator: Flashes when automated pressure control is ON. for dynamic control, for static control with added if custom control parameters are in use. 3. Target pressure value when controlling and in Not Ready (<↑> or <↓>) condition; pressure rate of change value when in static control mode and waiting for stability; current deviation from the target value when in dynamic control mode and Ready (<*>) condition. 2 Target:135.000 2 3 T 250.000 DC

PPC2+™ Operation and Maintenance Manual

Page 31 ©2001 DH Instruments, Inc.  Observe the Ready (<*>)/Not Ready (<↑> or <↓>) indication for indication of when ready conditions have been satisfied (see Section 3.1.2.4). PPC2+ will continue controlling following static or dynamic control operation protocol (see Section 3.1.2.3) until automated pressure control is interrupted.  +^K\QO^Z\O]]_\OOX^\c^RK^ObMOON]^ROM_\\OX^_ZZO\VSWS^]OO=OM^SYX Y\^RK^S]Y_^YP\KXQOaSVVXY^LOKMMOZ^ONK]^RO^K\QO^`KV_OaROXE/8>/ROWSXSW_WKXNWKbSW_WSX\KXQO^K\QO^ aSVVLO]O^   =OO =OM^SYX  SP ::- S] _XKLVO ^Y MYX^\YV Z\O]]_\O Y\ KZZOK\] ^Y MYX^\YV Z\O]]_\OZYY\Vc   0Y\::-^Y]O^Z\O]]_\O]_XNO\K^WY]ZRO\OKXNY\^Y\OVSKLVc]O^Z\O]]_\O] _XNO\ Z]SU:K QK_QO Y^RO\ ^RKX dO\Y QK_QO K `KM__W Z_WZ W_]^ LO MYXXOM^ON^Y^RO/B2+?=>ZY\^]OO=OM^SYXKXN::-W_]^LO]O^_Z PY\MYX^\YV_]SXQK`KM__W\OPO\OXMO]OO=OM^SYXDO\YQK_QOZ\O]]_\O MKXKVaKc]LO]O^   :\O]]SXQ^ROE←GKXNE→GONS^SXQUOc]YX^RObUOcZKNMKXLO_]ONaRSVO MYX^\YVVSXQ ^Y TYQ ^RO Z\O]]_\O MYX^\YV ^K\QO^ `KV_O NYaX Y\ _Z Lc ^ROM_\\OX^ RYVN VSWS^  >RO]O UOc] KV]Y TYQ ^RO ^K\QO^ `KV_O aROX SX ^RO /8>/< ^K\QO^ `KV_O]M\OOX Interrupting Automated Pressure Control Automated pressure control is interrupted by: • Pressing [ESCAPE]: Suspends control and remains in main run screen. • Pressing any direct pressure control key (see Section 3.1.2.2): Suspends control and executes direct pressure control. • Pressing any function key: Suspends control and goes to the selected function. • Pressing [ENTER]: Suspends control and goes to ENTER target pressure value screen. To resume automated pressure control, press [ENTER] and insert a target pressure value.

PPC2+™ Operation and Maintenance Manual

©2001 DH Instruments, Inc. Page 32  Automated Pressure Commands for Zero Pressure Zero in gauge mode: A command for automated pressure control to a target value of zero when in gauge pressure measurement mode is interpreted in the same manner as if the [VENT] direct pressure control key had been pressed (see Section 3.1.2.2). A Ready (<*>) condition will occur when the vent valve has been opened and the pressure meets the current stability test. Zero in absolute mode: When attempting to set a target pressure of zero in absolute measurement mode, PPC2+ opens its down control valves fully and closes a bypass interrupt valve to interrupt the natural flow of gas out the EXHAUST port through its pressure regulators once pressure is under 3.5 psia (25 kPa). However, even with a vacuum pump connected to its EXHAUST port, there is a limit to how low the pressure in PPC2+ can be pulled down. This is a function of the quality of the vacuum pump used, the characteristics of the volume connected to the TEST port and PPC2+’s internal restrictions. When a command for automated pressure control to a target value of zero is given when in absolute pressure measurement mode, PPC2+ responds by reducing the pressure as far as possible using down fast pressure control. A Ready (<*>) condition will be indicated when the stability limit is reached and the control down fast valve will remain open keeping the vacuum pump connected to the PPC2+ pneumatic module and TEST port. When pressure control is suspended or a non-zero pressure target is executed, the bypass interrupt valve is opened.  ._O^Y^ROWKXXO\SXaRSMR::-RKXNVO]KXK_^YWK^ONZ\O]]_\OMYWWKXNPY\dO\Y SX KL]YV_^O Z\O]]_\O WOK]_\OWOX^ WYNO K ROM_\\OX^NS]ZVKcON Z\O]]_\O ]RY_VN LO _]ON K] ^RO `KV_O YP Z\O]]_\O KZZVSON ^Y ^RONO`SMOY\]c]^OW _XNO\^O]^  AROX::-S]QS`OXKMYWWKXN^Y]O^dO\YSXKL]YV_^OWOK]_\OWOX^WYNO^RO\K^O YP MRKXQO YP Z\O]]_\O aSVV NOM\OK]O K] ^RO Z\O]]_\O NOM\OK]O] KXN ^RO \K^O aSVV O`OX^_KVVcLOXOK\dO\YaROX^RO`KM__WZ_WZRK]Z_VVONNYaX^ROZ\O]]_\OSX^RO ::- KXN ^RO ^O]^ `YV_WO K] PK\ K] ZY]]SLVO  >Y WKUO LO]^ _]O YP ^RO ZY\^RK]\ON_MONZ\O]]_\OK]PK\K]ZY]]SLVO

PPC2+™ Operation and Maintenance Manual

Page 33 ©2001 DH Instruments, Inc.  3.2.4 UPPER LIMIT m PURPOSE To set the upper limit pressure value for a pressure range and measurement mode. m PRINCIPLE The UPPER LIMIT function allows the setting of a maximum pressure not to be exceeded when using a specific range and measurement mode. Automated pressure control target values above the upper limit will not be accepted and direct pressure control keys will shut off when the upper limit is reached. Upper limit settings are specific to each range and measurement mode (gauge or absolute). The UPPER LIMIT function is most often used to protect the device or system connected to the PPC2+ TEST port from accidental overpressure.  3P XY _ZZO\ VSWS^ `KV_O S] OX^O\ON Lc ^RO _]O\ ^RO _ZZO\ VSWS^ PY\ OKMR \KXQO KXN WOK]_\OWOX^WYNOS]]O^LcNOPK_V^^YYPOKMR\KXQOKXNWOK]_\OWOX^WYNOPY\ \KXQO]6KXN2KXN^YYPOKMR\KXQOKXNWOK]_\OWOX^WYNOPY\KVVY^RO\\KXQO] m OPERATION When the [UPPER LIMIT] function key is pressed from the main run screen the display is: 1. Entry field to enter desired upper limit value. Enter the desired upper limit value and PPC2+ returns to the main run screen with a new upper limit value active. When the UPPER LIMIT has been exceeded, the display of current pressure flashes and a buzzer sounds for 3 seconds every 2 seconds. Reduce pressure using direct pressure control keys or an automated pressure command to return to normal operation.  ?ZZO\VSWS^`KV_O]K\O]ZOMSPSM^YOKMR\KXQOKXNWOK]_\OWOX^WYNO,OMK\OP_VXY^^Y K]]_WO ^RK^ ^RO _ZZO\ VSWS^ ]O^ SX YXO WOK]_\OWOX^ WYNO aSVV KZZVc ^Y ^RO Y^RO\ 0Y\ObKWZVOSPcY_]O^!Z]SK]^RO_ZZO\VSWS^SXKL]YV_^OWYNO^RO_ZZO\VSWS^aSVVXY^ LO!Z]SKK^WY]ZRO\SMZ\O]]_\OSXQK_QOWYNO>ROQK_QOWYNO_ZZO\VSWS^PY\^RK^ \KXQOaSVV]^SVVLO^RONOPK_V^VSWS^Y\KXY^RO\VSWS^^RK^RK]LOOX]O^SXQK_QOWYNO   ?ZZO\VSWS^]K\OKVaKc]]ZOMSPSONKXNNS]ZVKcONSX^ROM_\\OX^Z\O]]_\O_XS^ObMOZ^PY\ KV^S^_NO_XS^]AROXSXKV^S^_NO_XS^]_ZZO\VSWS^]K\OObZ\O]]ONSXU:KSP^ROKV^S^_NO_XS^ S]WO^O\]WKXNZ]SSP^ROKV^S^_NO_XS^S]POO^P^   ::-RK]?::/<KXN69A/<VSWS^]XY^T_]^KX?::/<6373>=OO^RO::-,1 9ZO\K^SYXKXN7KSX^OXKXMO7KX_KV 105 psi a

PPC2+™ Operation and Maintenance Manual

©2001 DH Instruments, Inc. Page 34  3.2.4.1 OVERPRESSURE FUNCTION In addition to the UL function, PPC2+ has an overpressure function. This function executes when the maximum RPT range is exceeded by 25 %. The function causes all pressure control to be interrupted and disabled, changes the range to H3 and causes the display to flash. The overpressure function also logs the time and date of the overpressure condition to assist in incident diagnosis. To recover from an overpressure condition, cycle PPC2+ power. 3.2.5 SEQ m PURPOSE To execute a programmed sequence of automated pressure control target values: Either a to run in the current range with all current settings or a specifying the range and settings to use. m PRINCIPLE The sequence function is intended to facilitate the execution of a series of pressure control target values such as is commonly needed when PPC2+ is used to run a typical calibration sequence on a device or system being tested. There are two types of sequences: • Quick Sequence (see Section 3.2.5.1) allows quick definition and execution of a sequence in the current range and using all of the current settings (e.g., pressure units, control mode, control settings). • File Sequence (see Section 3.2.5.2) allows execution of a preprogrammed sequence script including the PPC2+ range to use, measurement mode, control mode, target values and other operational settings. m OPERATION When the [SEQ] function key is pressed from the main run screen, the display is: Select the sequence type desired. See Section 3.2.5.1 for quick sequence or Section 3.2.5.2 for file sequence set ups. See Section 3.2.5.3 for running a sequence. 3.2.5.1 SETTING UP/EDITING A QUICK SEQUENCE m OPERATION When <1Quick> is selected from the SEQ function menu, the display is: This selection provides the choice between running the quick sequence that is currently set up or changing the quick sequence set up. Selecting <1Yes> will cause the currently defined quick sequence to run (see Section 3.2.5.3). Selecting <2No> will go to the quick sequence set up/edit routine. 1quick 2file

PPC2+™ Operation and Maintenance Manual

Page 35 ©2001 DH Instruments, Inc.  When <2No> is selected from the Run Current Quick Sequence menu, the set-up/edit routine is accessed. The first display is: 1. Entry field for sequence full scale value. Recalls value last entered. This entry determines the maximum pressure (furthest from zero) of the quick pressure sequence. Enter the desired sequence maximum pressure and press [ENTER].  =O[_OXMO 0= W_]^ LO VO]] ^RKX ^RO Z\O]]_\O _ZZO\ VSWS^ YP ^RO M_\\OX^Vc KM^S`O Z\O]]_\O \KXQO Y\ KX O\\Y\ aSVV YMM_\ aROX K^^OWZ^SXQ ^Y \_X ^RO ]O[_OXMO ]OO =OM^SYX>Y\_XK]O[_OXMOSXMV_NSXQKZ\O]]_\OQ\OK^O\^RKX^ROM_\\OX^_ZZO\ VSWS^SXM\OK]O^RO_ZZO\VSWS^LOPY\O\_XXSXQ^RO]O[_OXMOY\MRKXQO^YKRSQRO\\KXQO SPK`KSVKLVO The next display is: 1. Entry field for sequence pressure increment in % of full scale. Recalls last value entered. This selection determines the size of the pressure increment in terms of % of the full scale (FS) value entered in the previous screen. The sequence will run between zero and full scale in pressure steps of (full scale x increment %) starting from zero and ending with the uneven increment if applicable. Enter the desired pressure increment and press [ENTER]. The next display is: This selection determines the order in which the sequence increments will execute. Selecting <1Up> will cause the sequence to execute starting at zero and going to the full scale value. Selecting <2Down> will cause the sequence to execute starting at the full scale value and going to zero. Selecting <3Both> will cause the sequence to execute from zero to full scale and back to zero. After making the increment order selection the next display is: This selection determines under what conditions, when the sequence is run, the pressure target value will change to the next value in the sequence. If <1ENTER> is selected, the sequence will proceed to the next target value whenever, and only whenever, [ENTER] is pressed. 100 psi a increment: 20 % FS 1Up 2Down 3Both 1ENTER 2Timer

PPC2+™ Operation and Maintenance Manual

©2001 DH Instruments, Inc. Page 36  If <2Timer> is selected, a timer will be started each time a pressure Ready (<*>) condition is achieved (see Section 3.1.2.4) and the sequence will continue to the next pressure control target value without operator intervention once the timer has expired. After selecting <2Timer>, an additional screen is presented for entry of the timer value in seconds (minimum 1, maximum 999). After making the next increment on selection the next display is: 1. Indication of maximum (furthest from zero) pressure included in the sequence which is about to run. Press [ENTER] to run the quick sequence that has just been set up/edited (see Section 3.2.5.3). Press [ESCAPE] to return to the main run screen without running a sequence. 3.2.5.2 SETTING UP/EDITING A FILE SEQUENCE Setting up and editing file sequences is done by pressing [SETUP] and selecting <3Sequence> (see Section 3.3.3). 3.2.5.3 RUNNING A SEQUENCE m PURPOSE Selecting <1Yes> in the Run current quick sequence menu causes the current quick sequence to execute. Entering a valid sequence file number from the Run Sequence File # menu causes a file sequence to execute. A Quick Sequence executes in the current range with the current settings. A File Sequence sets up the range and other settings as specified in the sequence file (see Sections 3.2.5 and 3.3.3).  AROXSXS^SKVSdSXQObOM_^SYXYPK[_SMU]O[_OXMO::-MROMU]$KSP^RO]O[_OXMO P_VV ]MKVO Z\O]]_\O S] aS^RSX ^RO M_\\OX^ _ZZO\ VSWS^% L SP ::- S] SX KL]YV_^O WOK]_\O WYNO ^RK^ ^RO ]O[_OXMO NYO] XY^ MYX^KSX XOQK^S`O `KV_O]  3P OS^RO\ MYXNS^SYXS]XY^WO^KXO\\Y\WO]]KQOaSVVLONS]ZVKcONKXN^RO]O[_OXMOaSVVXY^ LO\_X   >Y\_XKPSVO]O[_OXMO^RK^]ZOMSPSO]K<:>\KXQOY^RO\^RKX^ROM_\\OX^\KXQO ::- W_]^ LO `OX^ON LOMK_]O ^RO \KXQO MKXXY^ LO MRKXQON aROX ::- S] XY^`OX^ON   >Y\_XKPSVO]O[_OXMO^RO]O[_OXMOPSVOX_WLO\]OVOM^ONW_]^KV\OKNcRK`OLOOX ]O^_ZLcZ\O]]SXQE=/>?:GKXN]OVOM^SXQ&=O[_OXMO(]OO=OM^SYX 100 psi a FS

PPC2+™ Operation and Maintenance Manual

Page 37 ©2001 DH Instruments, Inc.  m OPERATION The first display is: 1. Indication of maximum (furthest from zero) pressure included in the sequence which is about to run. This screen is offered as a means for the user to check, prior to execution, that the correct sequence is selected and that it does not exceed the range of the device or system being tested. Pressing [ESCAPE] returns to the main run screen without sequence execution. Pressing [ENTER] causes sequence execution to proceed. Pressure control will begin to set the first target value of the sequence (or vent will execute if the first increment is zero gauge pressure or a vent command). The display is the Main Run Screen (see Section 3.1.1) with indication of the sequence progress: 1. Indication of sequence increment that is currently being controlled/total increments in sequence. Changes to ENTER or countdown time remaining once Ready (<*>) condition has been reached at target value. Except for the bottom right hand corner, the sequence run display is identical to the normal main run screen (see Section 3.1.1). The bottom right corner provides information on sequence execution. This information updates as each sequence increment is executed. If the sequence was set up with , execution will proceed automatically from increment to increment after a Ready <*> condition has occurred at the increment and the countdown timer has expired. If the sequence was set up with . Press [ENTER] to continue to the next increment.  /<Z\O]]SXQE/8>/RS] POK^_\O MKX LO _]ON ^Y ]USZ YXO Y\ ]O`O\KV SXM\OWOX^] SPNO]S\ON To interrupt the sequence, press [ESCAPE]. Pressure control will immediately cease and the display will go to: 100 psi a FS TPRESSURE2 CC NN/NN 1yes 2no

PPC2+™ Operation and Maintenance Manual

©2001 DH Instruments, Inc. Page 38  Selecting <1Yes> causes the sequence to abort and returns to the main run screen. Selecting <2No> causes the sequence to resume where it left off.  +VV;_SMU=O[_OXMO]OXNaS^RK@/8>MYWWKXNKP^O\^ROVK]^ZYSX^>YK`YSN^RO @/8>Z\O]]E/=-+:/GKP^O\^ROVK]^ZYSX^ 3.2.6 CUSTOM CONTROL m PURPOSE To customize the control parameters of the current automated pressure control mode for the current range. m PRINCIPLE Automated pressure control and the associated pressure Ready (<*>)/Not Ready (<↑> or <↓>) condition are characterized by hold limit and stability limit parameters which determine how PPC2+ will control and under what conditions a Ready (<*>) indication will occur (see Sections 3.1.2.3 and 3.1.2.4). PPC2+ provides system default values for the hold limit and stability limit parameters for each range which have been determined to be those most suitable for the typical user to operate within PPC2+ pressure control and measurement specifications. The default parameters are automatically set each time the control mode is selected using the [CONTROL] function key (see Table 7). The CUSTOM CONTROL function allows the hold and stability limit parameters to be adjusted by the user. This can be used to increase control speed (time to Ready (<*>)), usually by decreasing control precision, or to increase precision which usually decreases control speed (see Section 3.1.2.4). For example, changing the dynamic pressure control hold limit from its default value of ± 0.005 % FS of the current range to ± 0.01 % FS of the current range will decrease the time required to set a pressure (achieve a Ready (<*>) condition) since the limit in which the pressure must be set has increased by a factor of two but it will also increase by a factor of two the maximum possible control error included when Ready (<*>) is indicated.  +NT_]^SXQNOPK_V^MYX^\YVZK\KWO^O\]_]SXQ^RO-?=>97-98><96P_XM^SYXWKc\O]_V^SX MYX^\YVZK\KWO^O\]^RK^^RO::-S]_XKLVO^YWOO^]Y^RK^KRS]NYO]XY^SXNSMK^O::-WKVP_XM^SYXT_]^^RK^^ROMYX^\YVZK\KWO^O\] XOON^YLO\OVKbONY\]O^LKMU^YNOPK_V^ZK\KWO^O\]

PPC2+™ Operation and Maintenance Manual

Page 39 ©2001 DH Instruments, Inc.  Table 7. System Default Control Parameters Lo REFERENCE TRANSDUCER RANGE 1, 2, 3 Hi REFERENCE TRANSDUCER RANGE 1, 2, 3 HOLD LIMIT STABILITY LIMIT HOLD LIMIT STABILITY LIMIT STATIC MODE ± 1 % FS of the active pressure range ± 0.005 % FS/sec of the active pressure range ± 1 % FS of the active pressure range ± 0.005 % FS/sec of the active pressure range. DYNAMIC MODE ± 0.005 % FS of the active pressure range or ± 2 ppm of range H3, whichever is larger ± 0.00 % of the active pressure range ± 0.005 % FS of the active pressure range ± 0.005 % FS of the active pressure range m OPERATION To set custom control parameters, press the [CUSTOM CONTROL] key from the main run screen. The display is: 1. Current reference transducer ( for high, for Lo) and range (<1>, <2> or <3>). 2. Current control mode ( for static, for dynamic). 3. Entry field for setting the desired hold limit. Recalls the default hold limit or the last custom hold limit set for the currently active range. Enter the desired hold limit. The next display is: 1. Current reference transducer ( for high, for Lo) and range (<1>, <2> or <3>). 2. Current control mode ( for static, for dynamic). 3. Entry field for setting the desired stability limit. Recalls the default stability limit or the last custom stability limit for the currently active range. Enter the desired stability limit. The display returns to the main run screen with the custom control parameters active.  >Y \O]O^ NOPK_V^ MYX^\YV ZK\KWO^O\] PY\ K \KXQO KXN MYX^\YV WYNO ]OVOM^ ^RO MYX^\YV WYNO_]SXQ^RO-98><96P_XM^SYXaRSVOSX^RK^\KXQO   -_]^YWMYX^\YVZK\KWO^O\]K\O\KXQOKXNMYX^\YVWYNO]ZOMSPSM-RKXQO]WKNOSXYXO \KXQOKXNMYX^\YVWYNOK\O]^Y\ONPY\^RK^\KXQOKXNMYX^\YVWYNOKXNaSVVLO\OMKVVON aROX^RK^\KXQOKXNMYX^\YVWYNOS]\O^_\XON^Y-RKXQO]WKNOSXYXO\KXQOKXNMYX^\YV WYNONYXY^KPPOM^KXcY^RO\\KXQOKXNMYX^\YVWYNO   AROX M_]^YW MYX^\YV ZK\KWO^O\] K\O KM^S`O K - S] KZZOXNON ^Y ^RO MYX^\YV WYNO SXNSMK^SXQMRK\KM^O\&=(PY\]^K^SM&.(PY\NcXKWSMSX^ROWSNNVOYP^ROLY^^YWVSXOYP ^ROWKSX\_X]M\OOX 3 0.005 psi D 3 0.005 psi D

PPC2+™ Operation and Maintenance Manual

©2001 DH Instruments, Inc. Page 40  3.2.6.1 TURNING-OFF CUSTOM CONTROL PARAMETERS To return to default pressure control parameters for the active range and control mode, select the control mode using the CONTROL function (see Section 3.2.3). 3.2.7 HEAD m PURPOSE To cause a pressure value representing a difference in height to be added to the pressure measured by the PPC2+ reference pressure transducer. m PRINCIPLE PPC2+’s reference pressure transducers measure gauge or absolute pressure at the height of the rear panel TEST port. Frequently, when performing a calibration or test, the device or system under test is at a different height than the PPC2+’s TEST port. This difference in height, frequently called head, can cause a significant difference between the pressure measured by the PPC2+ at its TEST port height and the pressure actually applied to the device under test which is at a different height. In this case, it is useful to make a head correction to the pressure measured by the PPC2+ RPT at its test port in order to accurately predict the pressure actually applied at a different height. PPC2+ can accurately calculate head pressures for nitrogen, helium and air as the test gas over its working pressure range. The HEAD function allows the height difference to be specified and causes the resulting head pressure to be added to the pressure measured at the TEST port. The [HEAD] function key is used to turn the head correction function ON and OFF and to specify the height difference between the PPC2+ TEST port and another height. The height units and the test gas are specified under by pressing [SPECIAL] and selecting <2Head> (see Section 3.2.7).  ?]OYP^RO2/+.P_XM^SYX^YK]]_\OSX^YVO\KXMOWOK]_\OWOX^]S]WY]^SWZY\^KX^SXVYa KL]YV_^OZ\O]]_\O\KXQO]=ZOMSPcSXQ^ROROKNROSQR^aS^RSXÝSXMWS]KNO[_K^O ^Y K]]_\O ^RK^ O`OX SX ^RO aY\]^ MK]O ^RO _XMO\^KSX^c YX ^RO ROKN MY\\OM^SYX aSVV LO SX]SQXSPSMKX^\OVK^S`O^Y^RO^YVO\KXMOYX^ROWOK]_\OWOX^

PPC2+™ Operation and Maintenance Manual

Page 41 ©2001 DH Instruments, Inc.  m OPERATION To access the HEAD function, press the [HEAD] function key. The display is: 1. Test gas currently specified for the head correction. 2. Entry field for head height. Entering a value of zero turns the HEAD function OFF. Entering a value other than zero turns the HEAD function ON using the height entered. Pressing [ESCAPE] returns to the main run screen with no change to the current head setting.  >RO2/+.P_XM^SYXS]89>\KXQO]ZOMSPSM>RO2/+.98Y\900]^K^_]\OWKSX]^RO]KWO K]\KXQO]K\OMRKXQON/NS^]WKNO^Y^ROROKN]ZOMSPSMK^SYX]K\OSXNOZOXNOX^YP\KXQO   AROX^RO2/+.P_XM^SYXS]98ROKN`KV_Ož3^S]SXNSMK^ONLc&R(SX^RO^YZVSXOYP ^RO WKSX \_X ]M\OOX ]OO =OM^SYX   AROX ^RO 2/+. P_XM^SYX S] 900 ^RO &R( S] XY^]RYaX   >YMRKXQO_XS^]YPROKNROSQR^LO^aOOXSXMRO]KXNMOX^SWO^O\]KXN^YMRKXQO^RO^O]^ QK]]ZOMSO]Z\O]]E=:/-3+6GKXN]OVOM^&2OKN(]OO=OM^SYX!   >RO \OPO\OXMO ROSQR^ YP ^RO ::- Z\O]]_\O WOK]_\OWOX^ S] ^RO WSNNVO YP ^RO ::- >/=>ZY\^>ROROKNROSQR^]RY_VNLOOX^O\ONK]KZY]S^S`O`KV_OSP^RONO`SMOY\]c]^OW _XNO\^O]^S]RSQRO\^RKX^RO::-KXNXOQK^S`OSPS^S]VYaO\ 3.2.8 LEAK CHECK/PURGE m PURPOSE The [LEAK CHECK] function key accesses two functions: • The LEAK CHECK function to assist in checking the pressurized connections and device or system under test connected to the PPC2+ TEST port for leaks (see Section 3.2.8.2). • The PURGE function to use the optional Self Purging Liquid Trap (SPLT) to collect and exhaust liquid contaminants from the device or systems under test before they reach the PPC2+ (see Section 3.2.8.3). 95 cm N2

PPC2+™ Operation and Maintenance Manual

©2001 DH Instruments, Inc. Page 42  m PRINCIPLE Leaks in the test system can have a detrimental effect on measurement accuracy and, when large enough, can cause the PPC2+ to be unable to set and control pressures reliably. The LEAK CHECK function is provided as a means of checking and quantifying the leaks that may be present in the system. PPC2+ is designed to precisely set, control and measure gas pressures. Liquid contamination of the PPC2+ internal pneumatic control module and reference pressure transducers can cause poor pressure control and interfere with accurate pressure measurement. Liquid contaminants that may be present in test devices or systems that are connected to the PPC2+ TEST port are likely to make their way back to PPC2+’s internal pneumatic system as the test device or system is pressurized and depressurized by PPC2+. To every extent possible, only clean hoses and connections should be used in connecting PPC2+ to test devices or system and those devices and systems should be free of liquid contaminants. However, in the event that PPC2+ will need to be used to test some devices or systems that contain small amounts of liquid contaminants, a Self Purging Liquid Trap (SPLT) accessory supported by the PURGE function is available. As the LEAK CHECK and PURGE functions are generally performed together at the beginning of a test and there are commonalties to the procedures, PPC2+’s LEAK and PURGE functions are designed to operate together under the single [LEAK CHECK] function key. m OPERATION To access the LEAK CHECK/PURGE function, from the main run screen press the [LEAK CHECK] function key. The function can be accessed directly at any pressure in any range. The first screen is the leak check screen (see Section 3.2.8.2). 3.2.8.1 LEAK CHECK/PURGE MENU AND FUNCTIONS FLOW CHART The flowchart of the LEAK CHECK/PURGE menu and functions is shown in Figure 10.

PPC2+™ Operation and Maintenance Manual

Page 43 ©2001 DH Instruments, Inc.  Figure 10. Leak Check/Purge Menu Flow Chart  >RO :?<1/ P_XM^SYX W_]^ RK`O LOOX OXKLVON Lc Z\O]]SXQ E=:/-3+6G KXN ]OVOM^SXQ &3X^O\XKV( &":_\QO( ]OO =OM^SYX" PY\ ^RO :?<1/ P_XM^SYX ^Y ObOM_^O 3P^RO:?<1/P_XM^SYXRK]XY^LOOXOXKLVON^ROE@/8>GNS\OM^Z\O]]_\OMYX^\YVUOc aSVVLORK`OXY\WKVVcP\YW^ROVOKUMROMU]M\OOX 3.2.8.2 LEAK CHECK SCREEN The leak check screen appears when the [PURGE/LEAK CHECK] function key has been pressed from the main run screen. 1. Current measured pressure and pressure units. 2. Active reference pressure transducer and range. 3. Current rate of change of pressure (leak rate) in current pressure units/minute (updated every 10 seconds). 4. Current rate of change of pressure (leak rate) in current pressure units/second. This screen is for observing the current leak rate. Pressing [ESCAPE] returns to the main run screen. MAIN RUN SCREEN LEAK CHECK SCREEN /=-+:/ @/8> /8>/< :?<1/6/+5 MAIN RUN SCREEN PURGE FUNCTION IF PURGE ACTIVE SET PRESSURE FUNCTION 2 3 0.001/s 0.06/m

PPC2+™ Operation and Maintenance Manual

©2001 DH Instruments, Inc. Page 44  If the PURGE function is enabled (see Section 3.4.4.8), pressing [VENT] when in the leak check screen causes the PURGE function to execute (see Section 3.2.8.3). Pressing [ENTER] when in the leak check screen goes to the set pressure screen (see Sections 3.2.3.1 and 3.2.8.4).  3XQOXO\KV^ROWKbSW_WKMMOZ^KLVOVOKU\K^OPY\YZ^SWKV::-K_^YWK^ONZ\O]]_\O MYX^\YV YZO\K^SYX KXN ^Y K]]_\O SX ^YVO\KXMO WOK]_\OWOX^] aS^R NOPK_V^ Z\O]]_\O MYX^\YVZK\KWO^O\]S]YPM_\\OX^\KXQO0=WSX_^O3XNcXKWSMMYX^\YVWYNO^Y RKXNVORSQRO\^O]^]c]^OWVOKU\K^O]SXM\OK]O^RORYVNVSWS^_]SXQ-?=>97-98><96 ]OO=OM^SYX    -RKXQSXQ^ROZ\O]]_\OSX^RO^O]^]c]^OWMK_]O]KNSKLK^SM^OWZO\K^_\OMRKXQO]SX ^ROQK]^RK^XOON^YRK`ONS]]SZK^ONLOPY\OK`KVSNVOKUWOK]_\OWOX^MKXLOWKNO 3XQOXO\KVK^YWSX_^OaKS^NOZOXNSXQYX^ROWKQXS^_NOYP^ROZ\O]]_\OMRKXQO KXN ^RO ]SdO YP ^RO ^Y^KV Z\O]]_\SdON `YV_WO S] KNO[_K^O ^Y KVVYa ^RO KNSKLK^SM ^OWZO\K^_\OMRKXQO^YNS]]SZK^OKXN`KVSNVOKUWOK]_\OWOX^]^YLOWKNO  3.2.8.3 PURGE FUNCTION   0Y\^ROZ_\QOP_XM^SYX^YYZO\K^O^RO:?<1/P_XM^SYXW_]^LOOXKLVONLcZ\O]]SXQ E=:/-3+6GKXN]OVOM^SXQ&3X^O\XKV(&":_\QO(]OO=OM^SYX"KXNKXYZ^SYXKV =OVP:_\QSXQ6S[_SN>\KZ=:6>W_]^LOMY\\OM^VcSX]^KVVONSX^ROVSXOMYXXOM^SXQ^RO ::->/=>ZY\^^Y^RONO`SMOY\]c]^OW_XNO\^O]^]OO=OM^SYX    >RO:?<1/P_XM^SYXWKUO]_]OYP^RO::-`KV`ON\S`O\YZ^SYX@KV`ON\S`O\X_WLO\ "S]_]ON^YKM^_K^O^RO=:6>ObRK_]^`KV`O>RO=:6>ObRK_]^`KV`O3]KM^_K^ON aROX ZO\PY\WSXQ ^RO :?<1/ P_XM^SYX KXN SX ^RO LKMUQ\Y_XN O`O\c ^SWO ^RO ::- `OX^`KV`OS]YZOXONAROXWKUSXQ_]OYP`KV`ON\S`O\]XY^O^RK^aROX^RO:?<1/ P_XM^SYX S] OXKLVON `KV`O N\S`O\ X_WLO\ " aSVV LO KM^_K^ON SXNOZOXNOX^Vc YP `KV`O N\S`O\MYWWKXN]  If the PURGE function is enabled (see Section 3.4.4.8), pressing the [VENT] direct pressure control key (see Section 3.1.2.2) from the LEAK CHECK screen will cause the PURGE function to execute. Execution of the PURGE function is automated and proceeds as follows: • If current pressure is greater than 100 psi (700 kPa) gauge, pressure is controlled down to less than 100 psi (700 kPa) gauge.   >ROZ_\QOP_XM^SYX]RY_VNLOObOM_^ONP\YW\Y_QRVcP_VV]MKVOYP^RONO`SMOY\]c]^OW _XNO\ ^O]^ Y\ <Z]S 7:K aRSMRO`O\ S] VYaO\  +`YSN ObOM_^SXQ ^RO :?<1/ P_XM^SYXP\YWZ\O]]_\O]KLY`OZ]S7:K

PPC2+™ Operation and Maintenance Manual

Page 45 ©2001 DH Instruments, Inc.  • Pressure control stops and a 5 second wait occurs. • SPLT exhaust valve is opened. • Wait until pressure is less than 3 psi (20 kPa) gauge or 16 psi (110 kPa) absolute. • Vent valve is opened with SPLT exhaust valve left open. When the PURGE function is initiated, if the current pressure is over 100 psi (700 kPa) gauge, the display is: 1. Current measured pressure and pressure units. 2. Current reference pressure transducer and range. 3. Indication that PPC2+ is reducing pressure to the purge pressure of less than 100 psi (700 kPa) gauge.  :\O]]SXQ E/=-+:/G K^ KXc ^SWO N_\SXQ Z_\QO \Y_^SXO ObOM_^SYX MK_]O] ^RO Z_\QO P_XM^SYX^YKLY\^KXN\O^_\X]^Y^ROVOKUMROMU]M\OOX]OO=OM^SYX" Once the current pressure is, or if the purge function was initiated, under 100 psi (700 kPa) gauge the display is: 1. Current measured pressure and pressure units. 2. Current reference pressure transducer and range. 3. Indication that PPC2+ is waiting 5 seconds and preparing to open the SPLT exhaust valve. Indication flashes. After a five (5) second wait the display is: 1. Current measured pressure and pressure units. 2. Current reference pressure transducer and range. 3. Indication that PPC2+ has opened the SPLT exhaust valve and is waiting for pressure to go to zero gauge. This indication flashes. 2 3 Descending to purge 2 3 PREPARING TO PURGE 2 3 PURGING

PPC2+™ Operation and Maintenance Manual

©2001 DH Instruments, Inc. Page 46  Once the pressure reaches less than 3 psi (20 kPa) gauge or 16 psi (110 kPa) absolute, PPC2+ opens its internal vent valve and returns to the main run screen in its normal vented condition. AROX^RO=:6>ObRK_]^`KV`OYZOX]K^Z\O]]_\ON_\SXQ^ROZ_\QO\Y_^SXOKXcVS[_SN MYX^KWSXKX^MYVVOM^ONaSVVLO\OVOK]ON^R\Y_QR^RO=:6>ObRK_]^^_LO3PWOK]_\KLVO KWY_X^] YP VS[_SN MYX^KWSXK^SYX K\O ObRK_]^ON P\YW ^RO =:6> ObRK_]^ `KV`O S^ S] \OMYWWOXNON ^RK^ ^RO Z_\QO P_XM^SYX LO \OZOK^ON _X^SV XY XY^SMOKLVO KWY_X^ YP VS[_SNMYX^KWSXK^SYXS]YL]O\`ONObRK_]^SXQP\YW^RO=:6>ObRK_]^`KV`O   >ROSX^O\XKV`YV_WOYP^RO=:6>S]KLY_^MM>RO=:6>aSVVXY^YZO\K^OOPPOM^S`OVcSP ^RO`YV_WOYPVS[_SNMYX^KWSXKX^]MYVVOM^ONS]Q\OK^O\^RKXKLY_^MM>RO=:6>S] NO]SQXON^YZ\Y^OM^^RO::-KQKSX]^YXVc\O]SN_KVVS[_SNMYX^KWSXK^SYX/`OXSPKX =:6> S] LOSXQ _]ON Z\OMK_^SYX] ]RY_VN LO ^KUOX ^Y \ON_MO ZY^OX^SKV VS[_SN MYX^KWSXK^SYX P\YW ^RO NO`SMO Y\ ]c]^OW _XNO\ ^O]^ K] W_MR K] ZY]]SLVO LOPY\O WKUSXQ^ROMYXXOM^SYX^Y^RO::->/=>ZY\^   .YXY^ZV_Q^RO=:6>ObRK_]^`KV`O>ROZ_\QO\Y_^SXOMKXXY^LOMYWZVO^ONaS^R^RO =:6>ObRK_]^`KV`OZV_QQONKXN^RO=:6>]Z\Y^OM^S`OP_XM^SYXaSVVXY^LOOPPOM^S`O   3P ^RO =:6> S] OVOM^\SMKVVc MYXXOM^ON ^Y ^RO ::- S^] ObRK_]^ `KV`O aSVV KVaKc] LO YZOXaROX^RO::-`OX^`KV`OS]YZOX3^WKcLOMVY]ONWKX_KVVc_]SXQ^RO.<3@/< P_XM^SYX]OO=OM^SYX  3.2.8.4 SETTING PRESSURES FOR LEAK CHECK/PURGE FUNCTION The direct pressure control keys are active when in the LEAK CHECK screen and can be used to adjust pressure as desired (see Section 3.1.2.2) keeping in mind that the [VENT] key will cause the PURGE function to execute if PURGE is enabled (see Section 3.4.4.8). Pressing the [ENTER] key from the leak check screen will cause the regular set pressure screen for automated pressure control to appear (see Section 3.2.3.2). An automated pressure control command can then be executed.

PPC2+™ Operation and Maintenance Manual

Page 47 ©2001 DH Instruments, Inc.  3.2.8.5 TYPICAL USE OF THE LEAK CHECK/PURGE FUNCTION The most common use of the LEAK CHECK/PURGE function is at the beginning of a test just after the device or system under test has been connected to the PPC2+ TEST port. Generally, the purge process should be completed before leak checking is performed. This may not be the case when the PPC2+ will be used to apply only ascending pressures. To purge: • Press the [LEAK CHECK] function key. • Use the direct manual pressure control keys to set the desired purge pressure (generally full scale of the device or system under test or 150 psi (1 MPa), whichever is lower). • Wait for pressure to stabilize. • Press the [VENT] key to cause the purge function to execute. Wait for purge function to execute completely. Observe whether any liquid contaminants exhaust from the SPLT exhaust valve. • Repeat from all steps until no noticeable amount of liquid contamination is observed exhausting from the SPLT exhaust valve. To leak check: • Press the [LEAK CHECK] function key. • Use the direct manual pressure control keys to set the desired leak check pressure (generally full scale of the device or system under test). • Wait for pressure to stabilize. • Observe the leak rates displayed in the leak check screen to determine if the leak rate is within acceptable limits. If leak rate is not acceptable, identify the source of the leak, vent system and eliminate the leak if possible. • Repeat from all steps until leak rate observed in is acceptable.  >RO VOKU MROMU P_XM^SYX MKX KV]Y LO _]ON K^ KXc ^SWO N_\SXQ YZO\K^SYX ^Y MROMU M_\\OX^VOKU\K^OKXNY\MK_]O^ROZ_\QO\Y_^SXO^YObOM_^O   >RO\OS]`O\cVS^^VO\S]UYPMYX^KWSXK^SYX^Y::-SPS^aSVVLO_]ON^YKZZVcK]MOXNSXQ Z\O]]_\O] YXVc ^Y ^RO NO`SMO Y\ ]c]^OW _XNO\ ^O]^ ]SXMO PVYaSXQ QK] LKMU ^Y ^RO ::- MKX LO K`YSNON  3X^RS] MK]O ZK\^SM_VK\Vc SP aY\USXQ aS^R K ]SQXSPSMKX^Vc MYX^KWSXK^ON NO`SMO Y\ ]c]^OW _XNO\ ^O]^ S^ WKc LO Z\OPO\KLVO ^Y XY^ _]O ^RO :?<1/P_XM^SYX_X^SV^ROOXNYP^RO^O]^KP^O\^ROVK]^K]MOXNSXQZ\O]]_\OSXM\OWOX^ RK]LOOX]O^+V]YLO]_\OXY^^YObOM_^O^RO:?<1/P_XM^SYXK^Z\O]]_\OKLY`O Z]S7:K

PPC2+™ Operation and Maintenance Manual

©2001 DH Instruments, Inc. Page 48  3.2.9 ANALOG m PURPOSE To select the current measurement range of the optional analog input function. This key has no function if the analog input function is not present in the PPC2+. m PRINCIPLE The PPC2+ is available with an analog input option that allows the user to measure up to four external voltage or current sources. Four ranges of measurement are provided: • 0 to 100 mV. • 0 to 5 V. • 0 to 10 V. • 4 - 20 mA. The selected range is effective on all of the inputs together. The measurement is available on the PPC2+ front panel or remotely. m OPERATION To access the ANALOG function, press the [ANALOG] function key from the main run screen. If the PPC2+ is equipped with the analog input option, the display is: Input range: 1:100 mV 2:5V 3:10V 4:4-20 mA The cursor is on the number of the currently active measurement range. Make the desired selection. The display returns to previous screen with the new analog measurement range active.  >RO OVOM^\SMKV MYXXOM^SYX] ^Y ^RO KXKVYQ SXZ_^ YZ^SYX K\O WKNO K^ K MYXXOM^Y\ YX ^RO ::-\OK\ZKXOV]OO=OM^SYX!

PPC2+™ Operation and Maintenance Manual

Page 49 ©2001 DH Instruments, Inc.  3.2.9.1 VIEWING ANALOG MEASUREMENTS To view the current reading of the four channels of the optional analog input press the [+/-] key from the main run screen. The display is: The displayed values update about every 0.75 seconds. To return to the main run screen from the analog input display screen press [+/-] or [ESCAPE].   >RO_XS^SXNSMK^SYXS]@PY\^RO^YKXN^Y@\KXQO]%W@PY\^RO^YW@ \KXQOKXNW+PY\^RO^YW+\KXQO   AROXO`O\^RO::-S]PS\]^^_\XON98Y\^ROKXKVYQSXZ_^YZ^SYX\KXQOS]MRKXQON KXNY\O`O\cWSX_^O]YPYZO\K^SYX]OO=OM^SYX^ROKXKVYQP_XM^SYXZO\PY\W] KX K_^YMKVSL\K^SYX McMVO  >RS] McMVO MYWZVO^O] SX KLY_^  ]OMYXN] N_\SXQ aRSMR KXKVYQWOK]_\OWOX^]K\OSX^O\\_Z^ONKXN^ROKXKVYQ`KV_OPSOVN]NS]ZVKcNK]RO] 3.3 SETUP MENU KEY m PURPOSE The [SETUP] menu key accesses a menu of functions and features commonly used in setting up PPC2+ for use. m OPERATION Press the [SETUP] key from the main run screen to access the SETUP menu. The display is: 3.3.1 CONTROLREF m PURPOSE To specify whether PPC2+ should automatically determine whether a vacuum source or atmospheric pressure is connected to its EXHAUST port or to manually set the EXHAUST port condition. -0.001 V 0.001 V Channel 2 Channel 3 - Channel 4 3Sequence 4Drivers

PPC2+™ Operation and Maintenance Manual

©2001 DH Instruments, Inc. Page 50  m PRINCIPLE The PPC2+ exhaust (down) control valves and the output of the exhaust pressure regulator are connected to the PPC2+ EXHAUST port (see Section 5.8.1.7). The response of the pneumatic module used to control pressure, particularly when controlling pressure near and under atmospheric pressure, changes depending on whether the EXHAUST port is connected to a vacuum source or left open to atmosphere. To use the proper pressure control algorithms, PPC2+ must know the condition of the exhaust port. The ControlRef function allows the PPC2+ to know whether its exhaust port is at vacuum or atmospheric pressure. An independent exhaust pressure sensor measures the pressure present at the EXHAUST port. Normally, the exhaust sensor output is used automatically by PPC2+ to determine the exhaust port condition. If desired, this automatic determination can be overridden and the exhaust port condition set manually. The ControlRef function allows setting of whether the EXHAUST port condition will be determined automatically using the exhaust sensor or set manually. It also allows viewing of the current status of the EXHAUST port.  >RO XY\WKV MYXNS^SYX YP ^RO -YX^\YV ZY\^ Z\O]]_\O MYXNS^SYX] K] ::- aSVV NO^O\WSXO ^RO ObS]^SXQ MYXNS^SYX K_^YWK^SMKVVc  >RO Y^RO\ ]O^^SXQ] K\O XY\WKVVc _]ON YXVc PY\ ]ZOMSKV KZZVSMK^SYX] Y\ ^\Y_LVO ]RYY^SXQ :\O]]_\OMYX^\YV XOK\ KXN _XNO\ K^WY]ZRO\SM Z\O]]_\O aSVV XY^ YZO\K^O Z\YZO\Vc SP -YX^\YV. The display is: Make the appropriate selection. Selecting <1auto> sets the PPC2+ to determine the EXHAUST port condition automatically using its internal exhaust sensor. This is the normal control ref setting. Selecting <2vac> sets the PPC2+ to assume that a vacuum supply is connected to the EXHAUST port. Selecting <3atm> sets the PPC2+ to assume that the EXHAUST port is open to atmospheric pressure. 2vac 3atm 4view

PPC2+™ Operation and Maintenance Manual

Page 51 ©2001 DH Instruments, Inc.  Selecting <4view> allows viewing of the current EXHAUST port condition assumed by PPC2+ and how it was determined. ,OPY\O MYXXOM^SXQ K `KM__W Z_WZ Y\ `KM__W ]Y_\MO ^Y ^RO ::- /B2+?=> ZY\^ =OO=OM^SYXPY\SXPY\WK^SYXYX`KM__WMYXXOM^SYX\O[_S\OWOX^]KXNZ\OMK_^SYX]   >RO-YX^\YVRO]O^^SXQWKNO SX -YX^\YV. The display is: Enter the desired pressure display resolution in % FS of the currently active range. The display returns to the main run screen with the entered resolution setting active.  >RO]c]^OWNOPK_V^Z\O]]_\ONS]ZVKc\O]YV_^SYXS]0=  >RO\O]YV_^SYX]O^^SXQS]\KXQO]ZOMSPSM+\O]YV_^SYX]O^^SXQWKNOSXYXO\KXQONYO]XY^ KPPOM^Y^RO\\KXQO]  3.3.3 Sequence m PURPOSE To create, store and edit the sequence files used by the SEQ function (see Section 3.2.5). m PRINCIPLE The PPC2+ FILE SEQUENCE function allows a sequence of automated pressure control targets in a specific range with predetermined control and operating parameters to be executed by recalling and running a sequence file. The files containing the information necessary to run a file sequence are created, stored and edited by pressing [SETUP] and selecting <3Sequence>. 0.0010% FS

PPC2+™ Operation and Maintenance Manual

©2001 DH Instruments, Inc. Page 52  Setting up a sequence file requires specifying: • The sequence file number to be created or edited. • The RPT to be used (Hi, Lo). • The range on the RPT (1, 2 or 3). • The pressure units of measure and measurement mode (gauge or absolute). • The number of pressure targets in the sequence. • The numerical value of each target pressure. • Whether a VENT function should be included at the end of the sequence. • Whether sequence execution should continue from one pressure target to the next on operator pressing [ENTER] or automatically on a timer and, if on a timer, the timer duration. • Pressure control mode to be used (static or dynamic). • Whether to use custom control parameter settings and the value of those settings if custom control is selected. • The file number under which the sequence definition should be stored. m OPERATION   =O[_OXMOPSVO]O^_Z\O[_S\O]^RO_]O\^Y]ZOMSPcKVK\QOX_WLO\YPZK\KWO^O\]SXK]O\SO] YP NK^K OX^\c ]M\OOX] NO]M\SLON LOVYa  0Y\ K ]_WWK\c YP ^RO ]O[_OXMO NOPSXS^SYX \O[_S\OWOX^]SX^ROY\NO\^ROcK\OONS^ON=OO:<38-3:6/YP^RS]]OM^SYX  To access the Sequence file building function, press [SETUP] and select <3Sequence>. The display is: Selecting <1view> allows sequence file characteristics to be viewed only. Selecting <2edit> allows a sequence file to be edited or created. The instructions below assume that <2edit> has been selected.  AROXNOPSXSXQK]O[_OXMO^ROE/=-+:/GUOcMKXLO_]ON^YLKMU^R\Y_QR^RONOPSXS^SYX KXN \OOX^O\ SXPY\WK^SYX _X^SV ^RO X_WO\SMKV `KV_O YP ^RO Z\O]]_\O ^K\QO^] RK] LOOXNOPSXON>ROX/=-+:/aSVVQY^YK=K`O+]€]M\OOXSXaRSMR^RO]O[_OXMOK]ONS^ON ]YPK\MKXLO]K`ON^YK]O[_OXMOPSVOX_WLO\>RS]KVVYa]SXPY\WK^SYXSX^ROLOQSXXSXQYP K]O[_OXMOPSVO^YLOONS^ONaS^RY_^RK`SXQ^YQY^R\Y_QR^ROOX^S\O]O[_OXMONOPSXS^SYX 1Edit 2View

PPC2+™ Operation and Maintenance Manual

Page 53 ©2001 DH Instruments, Inc.  The next display is: 1. Entry field for the file number of the sequence to be edited. Enter the desired sequence file number.  ?Z^Y]O[_OXMOPSVO]MKXLO]O^_ZKNN\O]]X_WLO\]^R\Y_QR The next selection determines the RPT that will be used: 1. Indication of the sequence file number that is currently being edited. Select the RPT that is to be used by this sequence. The next selection specifies the range of the selected RPT that is to be used for this sequence: Select the range (see Section 3.1.2.5) that is to be used when running this sequence. The next selections specify the pressure unit and measurement mode in which the sequence pressure target pressures will be defined:  3P ^RO <:> ]OVOM^ON PY\ ^RO ]O[_OXMO S] K QK_QO <:> 1BBBB ^RO WOK]_\OWOX^ WYNO ]OVOM^SYXaSVVXY^LOYPPO\ONK]QK_QO<:>]MKXYXVcWOK]_\OSXQK_QOWYNO Press [ENTER] to proceed to the pressure unit and measurement mode selection. These will operate in the same manner as when the UNIT function is used (see Section 3.2.2). To change the pressure units available, press [SPECIAL] and select <3PresU> (see Section 3.4.3). The next selection specifies the number of pressure targets in the sequence, including zeros and vents: 1. Entry field to edit the total number of targets in the sequence including zeros and vents. Enter the total number of targets in the sequence. nn 1Hi 2Lo Seq01 3Rng3 Seq01 Seq01 11 Seq01

PPC2+™ Operation and Maintenance Manual

©2001 DH Instruments, Inc. Page 54  Next, the numerical values of the pressure targets in the sequence are specified: 1. Number of the target in the sequence whose value is currently being edited. 2. Entry field for the numerical value of the target pressure. Enter the numerical value of the target pressure or press the [VENT] direct pressure control key to cause a VENT command to execute. The display repeats, incrementing through the target numbers until the total number of targets specified has been defined. The next selection defines how the sequence will end, by venting after the last target <1yes> or at the last target <2no>: Selecting <1Yes> will cause a VENT command to execute after the last target of the sequence has been completed. Selecting <2No> will cause the sequence to end at the last target. The next selection defines how the sequence will proceed from one pressure target to the next: If <1ENTER> is selected, when the sequence is executed, PPC2+ will wait at each pressure target until [ENTER] is pressed before proceeding to the next target. If <2Timer> is selected, when the sequence is executed, PPC2+ will start a timer when a Ready (<*>) condition is reached at a pressure target. Each time the timer expires, the sequence will proceed automatically to the next pressure target. If is selected, the dwell time in seconds (minimum 1 and maximum 99) is entered in the next screen. After defining how the sequence will proceed from one pressure target to the next, the control mode is specified: Enter the desired pressure control mode for the sequence (see Section 3.1.2.3 for a description of the control modes). The option of customizing the control mode parameters for the control mode selected is then offered: If <1Yes> is selected, screens will be presented allowing the control hold limit and stability limit to be adjusted (see Sections 3.2.3 and 3.2.6). If <2No> is selected, the control parameters will be the default control parameters for the range and control mode specified for this sequence (see Sections 3.2.3 and 3.2.6). 2 0 psi Seq01 Seq01 1yes 2no 1ENTER 2timer Seq01 1dynamic2static Seq01 1Yes 2No Seq01

PPC2+™ Operation and Maintenance Manual

Page 55 ©2001 DH Instruments, Inc.  Definition of the file sequence is now complete and the save as choice is offered: 1. Entry field for sequence file address number under which to save this sequence. Enter the desired sequence file number or press [ENTER] if the desired file number is already displayed. Pressing [ESCAPE], at , allows all changes to be abandoned. 3.3.4 DRIVERS m PURPOSE To control the output signals of PPC2+’s 8 channel, 12 V external drivers. m PRINCIPLE PPC2+ external drivers are available to drive peripheral equipment in a PPC2+ system, for example, solenoid valves or the optional Self Purging Liquid Trap (SPLT). The driver electrical connections are available from a rear panel connector  =OO=OM^SYX#PY\N\S`O\]ZOMSPSMK^SYX]KXNZSXY_^] m OPERATION To access the driver control function press [SETUP] and select <4drivers>. The display is: Pressing the keypad numerical key driver number turns that driver ON and OFF with either a momentary or a toggled response (see next paragraph). An active driver is indicated by <*> immediately following the driver number. Pressing [ENTER] while in the External drivers menu causes a menu to appear that allows selection of whether the driver actuation by selecting the driver number will be <1momentary> or <2toggle>. >RO::-:?<1/P_XM^SYXWKUO]_]OYP^RO`KV`ON\S`O\YZ^SYX@KV`ON\S`O\X_WLO\"S] _]ON ^Y KM^_K^O ^RO =:6> ObRK_]^ `KV`O  >RO =:6> ObRK_]^ `KV`O 3] KM^_K^ON aROX ZO\PY\WSXQ^RO:?<1/P_XM^SYXKXNSX^ROLKMUQ\Y_XNO`O\c^SWO^RO::-`OX^`KV`O S]YZOXON  AROX WKUSXQ _]O YP `KV`O N\S`O\] XY^O ^RK^ aROX ^RO :?<1/ P_XM^SYX S] OXKLVON`KV`ON\S`O\X_WLO\"aSVVLOKM^_K^ONSXNOZOXNOX^VcYP_]O\SXS^SK^ON`KV`ON\S`O\ MYWWKXN]]OO=OM^SYX" 01 2 3 4 5 6 7 8

PPC2+™ Operation and Maintenance Manual

©2001 DH Instruments, Inc. Page 56  3.4 SPECIAL MENU KEY The [SPECIAL] menu key accesses a menu of functions that are less commonly used or not normally used in regular operation. Press the [SPECIAL] menu key to access the SPECIAL menu. The display is:  3X::-,1^ROE=:/-3+6GWOX_UOcSXMV_NO]K]OVOM^SYX&@KV`O](=OO^RO::-,1 9ZO\K^SYXKXN7KSX^OXKXMO7KX_KV 3.4.1 AUTOZ m PURPOSE To change the offset (zero) of the Lo and Hi reference pressure transducers relative to a reference value in order to compensate for RPT zero drift between full recalibrations.  >Y K]]_\O YZO\K^SYX aS^RSX aS^R K_^YdO\Y€ WOK]_\OWOX^ KMM_\KMc ]ZOMSPSMK^SYX] ]OO =OM^SYXS^S]\OMYWWOXNON^RK^+_^YDLO\_X^RO`KV_OYPD900=/>_ZNK^ONK^ VOK]^O`O\cNKc]Y\aROX::-RK]LOOXObZY]ON^Y^OWZO\K^_\OMRKXQO]ObMOONSXQ ÝÙ- Ù0 m PRINCIPLE How AutoZ Works The main component of the change over time of the PPC2+ RPTs is zero drift or offset, independent of span. Rezeroing PPC2+ RPTs relative to a more stable reference between recalibrations allows higher measurement accuracy specifications to be maintained with less frequent full calibrations. The PPC2+ autozero function (AutoZ) provides full on-board support for the rezeroing process to simplify its application by the user. The autozeroing function uses four parameters: 1. ZSTD: The value of the autozero pressure as indicated by the reference autozero device. For absolute RPTs in absolute measurement mode, the autozero pressure is always atmospheric pressure and the ZSTD value can be supplied either a) by manual entry; b) automatically from a DHI RPM connected to the PPC2+ COM2 port or; c) automatically from a Lo absolute RPT if one is available. For gauge RPTs or absolute RPTs in gauge measurement mode, the autozero pressure is always zero (atmospheric pressure) which is supplied by definition when the RPT is vented to atmosphere. 4Internal

PPC2+™ Operation and Maintenance Manual

Page 57 ©2001 DH Instruments, Inc.  2. ZCURERR: The difference between ZSTD and the RPT indication at the autozero pressure at some time after the calibration of the RPT (ZCURERR = RPT w/out AutoZ - ZSTD). 3. ZNATERR: The difference between ZSTD and the RPT indication at the autozero pressure just after the RPT has been calibrated. This is referred to as the natural error at the autozero pressure. Because no RPT is perfectly linear and ZSTD is not perfectly accurate, the disagreement between the RPT reading and ZSTD at the autozero pressure is unlikely to ever be zero (except for gauge RPTs for which the zero point is forced at calibration). 4. ZOFFSET: ZCURERR corrected for ZNATERR, represents the drift of the RPT relative to the reference (ZOFFSET = ZCURERR - ZNATERR). The current RPT reading, adjusted by ZOFFSET, is the autozeroed RPT reading (e.g., the RPT reading corrected for zero drift since it was calibrated). For an absolute RPT used in gauge mode, ZOFFSET also includes the value of atmospheric pressure that is being subtracted to arrive at gauge pressure. The AutoZ function manages the determination, storage and application of ZNATERR and ZOFFSET individually for each PPC2+ range and measurement mode. AutoZ ON/OFF The AutoZ function can be turned ON and OFF. When AutoZ is ON, the ZOFFSET is always applied to the pressure measured by PPC2+. When AutoZ is OFF, ZOFFSET is not applied (except for an absolute RPT in gauge mode for which ZOFFSET is the current value of atmospheric pressure which is always subtracted to achieve gauge pressure). Gauge Mode with an Absolute RPT, Dynamic Compensation for Atmospheric Pressure PPC2+ allows gauge pressure measurements with an absolute RPT by calculation, subtracting the value of atmosphere (tare) from the RPT’s absolute measurement to arrive at a gauge value. The appropriate tare value changes with the natural evolution of atmospheric pressure at a given location. For this reason, the value of the tare is redetermined each time the RPT is known to be exposed to atmospheric pressure (e.g., the PPC2+ is vented and the pressure is stable). However, if atmospheric pressure changes significantly between opportunities to redetermine the tare, for example if running an extensive test that does not include returns to a vented condition, the changes in atmospheric pressure during that time will not be taken into account and a significant zero offset in the gauge pressure indications could result. To minimize zero errors due to the evolution of atmospheric pressure between taring opportunities when using an absolute RPT in gauge measurement mode, PPC2+ dynamically compensates the atmospheric tare value. PPC2+’s on-board barometer measures atmospheric pressure independently from the RPT. Between opportunities to update the tare value (vented conditions), the difference between the on-board barometer reading at the time of the last tare value and the current on-board barometer reading is used to compensate the tare value. This value is called ATMOFFSET. This technique, which relies only on the resolution and short term stability of the on-board barometer, rather than its absolute accuracy, allows gauge measurements with an absolute RPT with an additional uncertainty due to possible changes in atmospheric pressure limited to ± 0.001 psi (8 Pa).

PPC2+™ Operation and Maintenance Manual

©2001 DH Instruments, Inc. Page 58   AROX _]SXQ KX KL]YV_^O <:> ^Y WKUO VYa QK_QO WYNO WOK]_\OWOX^] ^RO VSWS^ SX \O]YV_^SYX YP ^RO YXLYK\N LK\YWO^O\ _]ON PY\ NcXKWSMMYWZOX]K^SYXPY\K^WY]ZRO\SM Z\O]]_\OWKcMK_]O]SQXSPSMKX^NS]^Y\^SYXYP^RO\O]_V^]0Y\LO]^\O]_V^]aROX_]SXQKX KL]YV_^O<:>SXQK_QOWYNO_XNO\Z]SU:K^_\X^RO+_^YDP_XM^SYX900]OO =OM^SYX  ^Y OVSWSXK^O ^RO OPPOM^ YP ^RO NcXKWSM MYWZOX]K^SYX ]c]^OW KXN `OX^ ::-YP^OX^YUOOZ^ROK^WY]ZRO\SM^K\O`KV_OM_\\OX^ Recommendations for the Use of the AutoZ Function The AutoZ function provides a powerful and easy to use tool for improving the measurement accuracy specifications of a PPC2+ and reducing the calibration recall interval by compensating for zero drift between full recalibrations. The following simple recommendations will help assure that you use this feature to best advantage. • Always leave AutoZ ON when operating in the gauge measurement mode. The only exception is when using an absolute RPT in gauge mode at pressures under 20 psi (150 kPa) (see note above). • Always leave AutoZ ON when operating in the absolute measurement mode if ZNATERR was set properly at calibration and the AutoZ routine using a valid atmospheric reference has been run regularly. • Execute the run AutoZ routine only when a reference whose accuracy is known to be significantly better than that of the PPC2+ RPT is available. Keep range ratios in mind when comparing accuracy. A ± 0.01 % FS barometer is roughly 10 times more accurate than an ± 0.01 % FS 150 psi (1 MPa) RPT range because the RPT/barometer range ratio is 10:1. • Though this may not be practical and generally is not necessary, the best reference with which to run AutoZ is a gas operated piston gauge applying atmospheric pressure to the PPC2+ test port. The best day to day reference is a properly calibrated, barometric range, DHI RPM1, RPM2 or RPM3 interfaced directly to the PPC2+ COM2 port. • Allow the PPC2+ to stabilize at atmospheric pressure and ambient temperature for 5 to 10 minutes before running AutoZ. m OPERATION  >RO +_^YD P_XM^SYX KXN `KV_O] K\O \KXQO +8. WYNO QK_QO Y\ KL]YV_^O ]ZOMSPSM 1K_QO<:>] 1BBBB RK`O XY D8+>/<< KXN XY SXNOZOXNOX^ +_^YD \Y_^SXO K] ^ROc K_^YdO\Y^\KX]ZK\OX^Vc^Y^RO_]O\aROXO`O\^RO::-S]`OX^ON To access the PPC2+ AutoZ function press [SPECIAL] and select <1AutoZ>. The display is: 3edit 4run

PPC2+™ Operation and Maintenance Manual

Page 59 ©2001 DH Instruments, Inc.  Selecting <1off> (or <1on>) is for changing the AutoZ status for the current range and measurement mode from ON to OFF or vice versa. Selecting <1off> turns OFF the AutoZ function for the current range and mode (see AutoZ ON/OFF of this section).  +_^YD 98 S] SXNSMK^ON Lc K &d( SX ^RO WKSX \_X ]M\OOX ^YZ VSXO ^RS\N MRK\KM^O\ P\YW ^RO\SQR^AROX+_^YDS]900^ROMRK\KM^O\S]LVKXU Selecting <2view> allows the current value of ZOFFSET for the current range and mode to be viewed. Selecting <3edit> allows the value of ZOFFSET for the current range and measurement mode to be edited. Selecting <4run> allows the routine for determination and activation of ZOFFSET by measurement of ZSTD to be executed. 3.4.1.1 VIEW AUTOZ To view the current ZOFFSET for the current range and measurement mode press [SPECIAL] and select <1AutoZ>, <2view>. The display is: 1. Current measurement mode. 2. Current range. 3. Display field of current value of ZOFFSET  D900=/> ]RY_VN LO dO\Y aROX ^RO ::- S] XOa Y\ RK] T_]^ LOOX MKVSL\K^ON D900=/>]RY_VN LO \Y_QRVc O[_KV ^Y K^WY]ZRO\SM Z\O]]_\O PY\ KX KL]YV_^O <:> YZO\K^SXQSXQK_QOWYNO   >RO `KV_O YP D900=/> S] KVaKc] NS]ZVKcON KXN OX^O\ON SX :K]MKV :K K] ::-] SX^O\XKVYZO\K^SYX]K\OSX:K 2 3 0.0 Pa

PPC2+™ Operation and Maintenance Manual

©2001 DH Instruments, Inc. Page 60  3.4.1.2 EDIT AUTOZ  >ROONS^+_^YDP_XM^SYX]RY_VNLO_]ONaS^RQ\OK^MK_^SYXK]OX^O\SXQSXKZZ\YZ\SK^O `KV_O] KXN ^_\XSXQ 98 +_^YD WKc \O]_V^ SX SXMY\\OM^ K_^YdO\YSXQ KXN Y_^ YP ^YVO\KXMOWOK]_\OWOX^]  3X XY\WKV YZO\K^SYX ^RO `KV_O YP D900=/> ]RY_VN LO MRKXQON_]SXQ^RO\_X+_^YDP_XM^SYX]OOX=OM^SYX>ROYXVcObZOM^ON_]O YP ONS^ +_^YD S] ^Y ]O^ D900=/> ^Y dO\Y KP^O\ K MKVSL\K^SYX YP ::- <:>] ,OPY\OONS^SXQD900=/>=OO=OM^SYX]:<38-3:6/KXN  To edit the current ZOFFSET for the current range and measurement mode, press [SPECIAL] and select <1AutoZ>, <3edit>. The display is: 1. Current measurement mode. 2. Current range. 3. Edit field for value of ZOFFSET  >RO`KV_OYPD900=/>S]KVaKc]NS]ZVKcONKXNOX^O\ONSX:K]MKV:K 3.4.1.3 RUN AUTOZ Run AutoZ is the function by which the current RPT reading is compared to ZSTD and a new value of ZOFFSET representing RPT zero drift is determined (see Section 3.4.1, PRINCIPLE, How AutoZ Works). To access run AutoZ, press [SPECIAL] and select <1AutoZ>, <4run>. If the measurement mode of the current range is gauge, run AutoZ occurs transparently to the user whenever AutoZ is ON and the PPC2+ is in the vented condition. The display is: 1. Indication of current range. 2 3 0.0 Pa no manual AutoZ

PPC2+™ Operation and Maintenance Manual

Page 61 ©2001 DH Instruments, Inc.  If the measurement mode of the current range is absolute the display is: 1. Current range. 2. Current measurement mode. 3. Selection of source of ZSTD for ZOFFSET determination. Selecting <1Entry> allows the value of ZSTD to be entered from the front panel keypad. Selecting <2COM2> allows the value of ZSTD to be read automatically from an RPM interfaced with PPC2+’s COM2 communications port. Selecting <3RngL3> is available only if the current RPT is a Hi absolute RPT and there is a Lo absolute RPT. Allows the value of ZSTD to be read automatically from the Lo RPT.  +VVYa^RO::-^Y]^KLSVSdOK^K^WY]ZRO\SMZ\O]]_\OKXNKWLSOX^^OWZO\K^_\OPY\ ^YWSX_^O]LOPY\O\_XXSXQ+_^YD   3P\_XXSXQ+_^YD\O]_V^]SX`KV_O]YPD900=/>^RK^K\OQ\OK^O\^RKXÝ0=YP ^ROM_\\OX^::-WOK]_\OWOX^\KXQO^RO::-KXNY\^RO]Y_\MOYPD=>.WKcLO Y_^YP^YVO\KXMOY\^RO+_^YDZ\YMO]]WKcRK`OLOOXPK_V^c,OPY\OKM^S`K^SXQKXOa D900=/>Q\OK^O\^RKXÝ0=YP^ROM_\\OX^::-\KXQOMROMU^YLO]_\O^RK^ LY^R^RO::-KXN^RO]Y_\MOYPD=>.aO\OSXQYYNaY\USXQY\NO\Z\YZO\Vc`OX^ON^Y ]^KLVOK^WY]ZRO\SMZ\O]]_\OK^^RO]KWOROSQR^KXN\OKNSXQSX^RO]KWOZ\O]]_\O _XS^]aROX+_^YDaK]\_X   AROX^RO\_X+_^YD]OVOM^SYXS]WKNOSPK2/+.MY\\OM^SYXS]M_\\OX^VcKM^S`O]OO =OM^SYX! ^RO ROKN MY\\OM^SYX aSVV LO NS]KLVON ^Y K`YSN K_^YdO\YSXQ€ ^RO ROKN`KV_O    >RO`KV_OYPD900=/>aSVVXY^LOObKM^VcO[_KV^Y^RONSPPO\OXMOLO^aOOX^ROM_\\OX^ \OKNSXQ KXN D=>. SP D8+>/<< S] XY^ dO\Y ]OO =OM^SYX :<38-3:6/ 2Ya +_^YDAY\U]   >RO`KV_OYPD900=/>S]KVaKc]NS]ZVKcONKXNOX^O\ONSX:K]MKV:K AutoZ H3 abs by: 1Entry 2COM2 3RngL3

PPC2+™ Operation and Maintenance Manual

©2001 DH Instruments, Inc. Page 62  Run AutoZ by Entry AutoZ by entry allows the value of ZSTD (See Section 3.4.1, PRINCIPLE, How AutoZ Works) to be entered manually. This provides a simple way of autozeroing relative to an independent reference device such as a house barometer that does not interface directly with PPC2+. To access run AutoZ by entry press [SPECIAL] and select <1AutoZ>, <4run>, <1Entry>. The display is: 1. Pressure reading, units and mode of the active RPT range. 2. Active range indicator. 3. Entry field for the value of ZSTD. Enter the value of ZSTD in the current units. The next display is: The new ZOFFSET is the ZOFFSET calculated from this execution of run AutoZ. Press [ENTER] to activate the new ZOFFSET and return to the main run screen. Press [ESCAPE] to maintain the old ZOFFSET and return to the main run screen.  >RO`KV_OYPD900=/>S]KVaKc]NS]ZVKcONKXNOX^O\ONSX:K]MKV:K  >RO`KV_OYPD=>.S]OX^O\ONSX^ROM_\\OX^Z\O]]_\O_XS^] 2 3 ZSTD: 0.0000 New ZOFFSET: 6.5 Pa

PPC2+™ Operation and Maintenance Manual

Page 63 ©2001 DH Instruments, Inc.  Run AutoZ by COM2 AutoZ by COM2 allows the value of ZSTD (see Section 3.4.1, PRINCIPLE, How AutoZ Works) to be read automatically from a DHI RPM1, RPM2 or RPM3 connected by RS-232 interface to the PPC2+ COM2 port. To access run AutoZ by COM2 press [SPECIAL] and select <1AutoZ>, <4run>, <2COM2>. The display is: 1. Pressure reading, units and mode of the active RPT range. 2. Active range indicator. 3. Pressure reading of the RPM connected to PPC2+’s COM2 port. When Ready (<*>), press [ENTER]. The next display is: The new ZOFFSET is the ZOFFSET calculated from this execution of run AutoZ. Press [ENTER] to activate the new ZOFFSET and return to the main run screen. Press [ESCAPE] to maintain the old ZOFFSET and return to the main run screen.  0Y\::-^YMYWW_XSMK^OaS^RKX<:7MYXXOM^ON^YS^]-97ZY\^^RO::-KXN ^RO <:7 <= SX^O\PKMO] W_]^ LO ]O^ _Z Z\YZO\Vc ]OO =OM^SYX   3P^RO ::- S] _XKLVO ^Y VYMK^O KX <:7 YX -97 aROX \_XXSXQ +_^YD Lc -97 KP^O\ ]OMYXN]S^aSVV^SWOY_^KXNNS]ZVKc&<:7XY^NO^OM^ON(   >RO`KV_OYPD900=/>S]KVaKc]NS]ZVKcONKXNOX^O\ONSX:K]MKV:K Run AutoZ by RngL3 The AutoZ by RngL3 choice is only available if the current RPT is an absolute, Hi RPT and there is an absolute Lo RPT in the PPC2+. Run AutoZ by RngL3 allows AutoZ to be run on a Hi RPT range using Lo RPT range 3 (L3) as the source of ZSTD (see Section 3.4.1, PRINCIPLE, How AutoZ Works). To access run AutoZ by RngL3 press [SPECIAL] and select <1AutoZ>, <4run>, <3RngL3>. The display is: 1. Pressure reading, units and mode of the active Hi RPT range. 2. Active range indicator. 3. Pressure reading of Lo RPT range 3 (L3). 2 3 ZSTD: 14.102 2 3 ZSTD: 14.102 New ZOFFSET: 13.8 Pa

PPC2+™ Operation and Maintenance Manual

©2001 DH Instruments, Inc. Page 64  When, press [ENTER]. The next display is: The new ZOFFSET is the ZOFFSET calculated from this execution of run AutoZ. Press [ENTER] to activate the new ZOFFSET and return to the main run screen. Press [ESCAPE] to maintain the old ZOFFSET and return to the main run screen.  3PcY_K\O\_XXSXQ+_^YDYX^RO2SKXN6Y<:>]YPK::-KXN_]SXQ\_X+_^YDLc LO]_\O^Y\_X+_^YDYX^RO6Y<:>PS\]^   ,OPY\O \_XXSXQ +_^YD Lc 6Y <:> LO ]_\O ^RK^ \KXQO 6 YP ^RO 6Y <:> S] MY\\OM^Vc MKVSL\K^ONKXNK_^YdO\YON   >RO`KV_OYPD900=/>S]KVaKc]NS]ZVKcONKXNOX^O\ONSX:K 3.4.2 HEAD m PURPOSE To specify the configuration of the HEAD function (see Section 3.2.7) including the length unit of measure for head height entry and the test gas species for head pressure calculations (see Section 3.2.7). m OPERATION From the main run screen, press [SPECIAL] and select <2Head>. The display is: Select the desired head height unit. The next display is: Select the type of gas that is being supplied to PPC2+. The characteristics of the gas selected will be used by PPC2+ in calculating head pressures. New ZOFFSET: 13.8 Pa 1in 2cm 1N2 2He 3Air

PPC2+™ Operation and Maintenance Manual

Page 65 ©2001 DH Instruments, Inc.  3.4.3 PresU m PURPOSE To customize the selection of pressure units that will be available for selection from the [UNIT] function key. m PRINCIPLE The [UNIT] function key makes available a choice of six default pressure units (US or SI units depending on whether the PPC2+ has been factory set as US or SI) (see Section 3.2.2). PPC2+ also supports many commonly used units other than those included in the default set up. These units can be made available for active selection by customizing the UNIT function using by pressing [SPECIAL] and selecting <3PresU>. This allows PPC2+ to offer a very wide selection of units while simplifying day to day operation. The typical user will customize the [UNIT] function key to support his/her six most commonly used units. m OPERATION To customize the [UNIT] function key, from the main run screen press [SPECIAL] and select <3PresU>. The display is: 1. Entry field to select which unit position (1 - 6) of the UNIT function key menu is to be changed. Enter the number of the unit position that you would like to change. The display becomes: Select the desired pressure unit category (SI units include units based on SI such as mmHg), then select the desired unit from the unit menu. The units available are listed in Table 8. Table 8. UNIT Function - Available Units <1SI> <2other> <3altitude> <4user> <1Pa> <2kPa> <3MPa> <4mbar> <5bar> <6mmHg> <7mmWa> <1psi> <2psf> <3inHg> <4inWa> <5kcm2> <1feet> <2meters> 3altitude 4user

PPC2+™ Operation and Maintenance Manual

©2001 DH Instruments, Inc. Page 66  If <4user> was selected, the user unit must be defined. The display is: 1. Entry field. Enter the number of user units per Pascal (Pa) in the entry field. Pressing [ENTER] defines the user unit and returns to the Set up unit # screen.  =OO=OM^SYX!PY\^ROZ\O]]_\O_XS^MYX`O\]SYXPKM^Y\]_]ONLc::- 3.4.4 INTERNAL m PURPOSE To view, set, adjust, and maintain various aspects of PPC2+’s internal operation. m OPERATION To access the internal selections press [SPECIAL] and select <4Internal>. The display is: Select the internal function desired. For a detailed description of each internal function. See Sections 3.4.4.1 to 3.4.4.9.  =YWO]M\OOX]]_MRK]^RO=:/-3+63X^O\XKVWOX_QYLOcYXN^RO^aYVSXO]Z\Y`SNONLc ^RO NS]ZVKc  >RS] S] SXNSMK^ON Lc K PVK]RSXQ K\\Ya SX ^RO ]OMYXN VSXO YP ^RO NS]ZVKc :\O]]^ROE←GKXNE→GK\\YaUOc]^YWY`O^ROM_\]Y\^YKMMO]]^ROVSXO]^RK^K\OXY^ `S]SLVOY\NS\OM^VcOX^O\^ROX_WLO\YP^RORSNNOXWOX_MRYSMOSPcY_UXYaS^ 3.4.4.1 CONFIG m PURPOSE To run an automated routine that readjusts internal pressure control characteristics. This function is considered part of PPC2+ maintenance and is therefore covered in the maintenance section of this manual (see Section 5.5). m OPERATION See Section 5.5. 1.0000000 Units/PA 4Time 5Reset 6Level 7Atm 8Purge 9ScrSav

PPC2+™ Operation and Maintenance Manual

Page 67 ©2001 DH Instruments, Inc.  3.4.4.2 CAL m PURPOSE To calibrate the PPC2+ Hi and Lo reference pressure transducers and adjust the on-board barometer. This function is considered part of PPC2+ maintenance and is therefore covered in the maintenance section of this manual (see Section 5.2). m PRINCIPLE See Sections 5.2, 5.3, and 5.4. m OPERATION See Sections 5.2, 5.3, and 5.4. 3.4.4.3 REMOTE m PURPOSE To configure the PPC2+ communication ports including COM1, COM2 and the IEEE-488 interface. m PRINCIPLE The PPC2+ has two RS-232 communications ports referred to as COM1 and COM2 and an IEEE-488 (GPIB) interface. Communication with a host computer is accomplished using either COM1 or the IEEE-488 interface. COM2 is reserved for connecting an external device (e.g., a multimeter, RPM, etc.). The configuration of all the communication ports can be set up through the PPC2+ front panel. m OPERATION To access the port configurations and remote program command mode, select <3Remote> under <4Internal> of the SPECIAL menu. Select , or to view and edit that port’s settings. Select to select the remote programming command format to use (see FORMAT of this section). COM1 and COM2 The COMx ports can be set for the specific settings required by the user. These settings are baud, parity, length and stop bit. The available options are: • Baud: 300, 600, 1 200, 2 400, 4 800, 9 600 • Parity: NONE, ODD or EVEN • Length: 7 or 8 • Stop Bit: 1 or 2 The default is <2400, E, 7,1> for COM1 and COM2.

PPC2+™ Operation and Maintenance Manual

©2001 DH Instruments, Inc. Page 68  The user can also specify one or two termination characters as well as define these characters. These are referred to as “Term1” and “Term2”. These define the characters that mark the end of commands that are sent to the PPC2+ as well as mark the end of replies sent back to the host computer. The PPC2+ typically will use the character to mark the end, and will ignore the character. These characters can be any decimal number from 1 to 255. The default for these are <13> (carriage return) and <10> (line feed) and usually should not be changed. IEEE-488 The IEEE-488 port’s primary address can be set from 1 to 31 in this screen. The factory default value is 10. Secondary addressing is not used or supported. This address must not conflict with the address of any other devices on the same IEEE-488 bus. The receiving terminating character must be a line feed and EOI. Carriage returns are ignored if received. The PPC2+ sends a line feed and asserts the EOI line to terminate a reply. These settings are fixed to agree with IEEE Std. 488. If you change the address, the IEEE interface will reset (PON) and become idle. Format The PPC2+ has two different syntax formats available for the remote program commands. The classic format is compatible with previous PPC products and is the default format. The enhanced format follows the syntax, format, and status reporting features of IEEE Std 488.2. The details of each are covered in Section 4.3. Selecting a format resets the IEEE interface and puts it into an idle state. 3.4.4.4 TIME m PURPOSE To view and edit the PPC2+ internal time and date settings. m OPERATION To access the time function press [SPECIAL] and select <4Internal>, <4Time>. The display is: Select <1time> to edit the time. Edit hours, then minutes, then am/pm by pressing ENTER at each entry. Seconds go to zero when minutes are entered. Select <2date> to edit the date. The date must be specified in YYMMDD format.  >RO::-NK^OKXN^SWOK\O]O^^Y?XS^ON=^K^O]7Y_X^KSX=^KXNK\N>SWOSX ^RO PSXKV ^O]^ KXN SX]ZOM^SYX Z\YMO]]K^^ROPKM^Y\c3PNO]S\ON_]O^RONK^O P_XM^SYX^Y]O^cY_\VYMKV^SWOKXNNK^O 08:32:11 am 970724

PPC2+™ Operation and Maintenance Manual

Page 69 ©2001 DH Instruments, Inc.  3.4.4.5 RESET m PURPOSE To reset various PPC2+ settings to default or factory values. PPC2+ stores its user definable settings in non-volatile memory. The reset menu allows the user to selectively or completely reset these settings to factory defaults. This clears out any settings that the user has set up, and should be used only to restore the PPC2+ to a known state. PPC2+ will go through its reboot sequence after any type of reset is executed.  ::- \O]O^ P_XM^SYX] aSVV MRKXQO M_\\OX^ ]O^^SXQ] ^Y PKM^Y\c NOPK_V^] >RO]OWKcSXMV_NO]O^^SXQ]`S^KV^Y::-YZO\K^SYXKXNKPPOM^SXQ^ROMKVSL\K^SYX YP^RO\OPO\OXMOZ\O]]_\O^\KX]N_MO\], <5Reset>. The display is: Reset - Sets m PURPOSE Sets most general system operating parameters back to default values. Does not affect calibration coefficients or remote interfaces. • Returns each range unit of measure to the first of six available (see Section 3.2.2). • Returns measurement mode to the RPT mode (A or G) (see Section 3.2.2). • Range to H3 (see Section 3.2.1). • Head settings to 0 cm and Nitrogen (see Section 3.2.7). • Automated pressure control parameters to defaults (see Section 3.1.2.3). • All upper limit values to defaults (see Section 3.2.4). • Pressure display resolution to 0.001 % FS (see Section 3.3.2). • Control mode to dynamic (see Section 3.1.2.3). • COM ports and IEEE-488 settings (see Section 3.4.4.3). • Purge disabled (see Section 3.4.4.8). • Autozero enabled (see Section 3.4.1). 4cal 5all

PPC2+™ Operation and Maintenance Manual

©2001 DH Instruments, Inc. Page 70  • Control Ref to auto (see Section 3.3.1). • Screen saver to 10 minutes (see Section 3.4.4.9). Reset - Units m PURPOSE Sets the six pressure units available under the UNIT function to the SI or US default selections depending on whether the PPC2+ has been factory set for SI or US. See Sections 3.2.2 and 3.4.3. Reset - Seq m PURPOSE • Clears all of the user defined file and quick sequences. See Sections 3.2.5 and 3.3.3. • Sets user level control configuration to factory defaults. See Sections 3.4.4.1 and 5.5. Reset - Cal m PURPOSE  >RO \O]O^  MKV P_XM^SYX aSVV NOVO^O <:> MKVSL\K^SYX MYOPPSMSOX^] MRKXQSXQ ^RO ::-MKVSL\K^SYX  Clears all user values affecting the calibration of RPTs and the on-board barometer including: • Sets all user RPT PA/PM values to 0 and 1 (see Section 5.2). • Sets user on-board barometer PA/PM to 0 and 1 (see Section 5.3). • Sets all ZNATERR values back to zero (see Section 5.2.6). • Sets all ZOFFSET values back to zero (see Section 3.4.1).

PPC2+™ Operation and Maintenance Manual

Page 71 ©2001 DH Instruments, Inc.  Reset - All m PURPOSE  >RO \O]O^  KVV P_XM^SYX aSVV MVOK\ KXN NOVO^O VK\QO KWY_X^] YP _]O\ NOPSXON SXPY\WK^SYXSXMV_NSXQM\S^SMKVMKVSL\K^SYXNK^K  Combines all resets in one global reset command that clears the entire user section of non-volatile memory. These include: • Reset - sets (see Section 3.4.4.5.1). • Reset - units (see Section 3.4.4.5.2). • Reset - seq (see Section 3.4.4.5.3). • Reset - cal (see Section 3.4.4.5.4). • Reset - Com1, Com2 and IEEE-488 interface to defaults (see Section 3.4.4.3).  >RO\O S] XY \O]O^ ^cZO ^RK^ KPPOM^] ^RO _]O\ VO`OV Z\Y^OM^SYX ]O^^SXQ] ]OO =OM^SYX  3.4.4.6 LEVEL m PURPOSE To set user protection levels to restrict access to certain functions and to edit the password required for changing user levels. m PRINCIPLE PPC2+’s front panel user interface provides the means to access all PPC2+ user defined data, settings and functions including calibration data. Inadvertent, uninformed or unauthorized altering or deleting of data, settings and functions could require extensive reconfiguration by the user and might cause invalid readings and behavior. For these reasons, depending upon the application in which PPC2+ is being used, it may be desirable to restrict access to certain functions for certain users. The user level function provides a means of restricting access to certain functions. Four different levels of security are available. Access to changing security levels can be left open, or be protected by a password.

PPC2+™ Operation and Maintenance Manual

©2001 DH Instruments, Inc. Page 72  Security Levels The security levels are structured to support typical operating environments as follows: None This level is intended for use only by the system manager and/or calibration facility. It allows access and editing in all areas including critical metrological information. Low This level of security is designed to protect the specific metrological information and system diagnostic and maintenance functions of the system. It is intended for an advanced operator performing many different tasks. Medium This level of security is designed to protect specific metrological information in the system and to assure that the PPC2+ is operated using consistent operational parameters. High This level of security is designed to protect all operating parameters. It is intended to minimize operator choices, for example to perform repeated identical calibrations under consistent conditions.  ::-S]NOVS`O\ONaS^R^RO]OM_\S^cVO`OV]O^K^VYa^YK`YSNSXKN`O\^OX^KV^O\SXQYP M\S^SMKV SX^O\XKV ]O^^SXQ] L_^ aS^R KMMO]] ^Y MRKXQSXQ ]OM_\S^c VO`OV] _X\O]^\SM^ON 3^S]\OMYWWOXNON^RK^^ROVYa]OM_\S^cVO`OVLOWKSX^KSXONK^KVV^SWO]3P^RO\OS]K \S]U YP _XK_^RY\SdON MRKXQSXQ YP ^RO ]OM_\S^c VO`OV MRKXQSXQ K_^RY\S^c ]RY_VN LO ZK]]aY\NZ\Y^OM^ON]OO9:/<+>398YP^RS]]OM^SYX

PPC2+™ Operation and Maintenance Manual

Page 73 ©2001 DH Instruments, Inc.  The security levels are structured to support typical levels of operation as shown in Table 9. Specifically, the security levels prevent execution of the X’d functions: Table 9. Security Levels FUNCTION LOW MEDIUM HIGH [RANGE] key X [UNIT] key X [CONTROL] key X [UPPER LIMIT] key X [SEQ] key, Run current quick sequence, <2No> X [CUSTOM CONTROL] key X X [HEAD] key X [SETUP] menu, <1ControlRef> X [SETUP] menu, <1ControlRef>, change status X X [SETUP] menu, <2Resltn> X X [SETUP] menu, <3Sequence>, <2Edit> X [SETUP] menu, <4Drivers> X [SPECIAL] menu X X [SPECIAL] menu, <1AutoZ> X X [SPECIAL] menu, <1AutoZ>, edit X X X [SPECIAL] menu, <2Head> X X [SPECIAL] menu, <3PresU> X X [SPECIAL] menu, <4Internal> X X [SPECIAL] menu, <1Config> X X X [SPECIAL] menu, <4Internal>, <2Cal> X X X [SPECIAL] menu, <4Internal>, <4Time>, edit X X X [SPECIAL] menu, <4 Internal>, <5Reset>, <4cal> X X X [SPECIAL] menu, <4Internal>, <5Reset>, <5all> X X X m OPERATION PPC2+ is delivered with no active password and access to the User Level menu is open. The user level is set to 1Low. User levels can be changed freely until a password has been created. To access the User Level function, press [SPECIAL] and select <4Internal>, <6Level>. If no password yet exists or if the correct password has been entered. The display is: Selecting <1change> user level brings up the restriction menu: You can then select the current restriction level, or [ESC] back to the RUN screen. 0 disables password 2low 3medium 4high

PPC2+™ Operation and Maintenance Manual

©2001 DH Instruments, Inc. Page 74  The second selection displays the user password and allows it to be edited. Passwords can be up to six numbers in length and cannot start with a zero.  9XMOKZK]]aY\NRK]LOOXOX^O\ON^RO_]O\VO`OVMKXXY^LOMRKXQONaS^RY_^ \OOX^O\SXQ^ROZK]]aY\N If <0> is entered, then the password is made inactive and the user will not be required to enter a password to access the user level menu. This is the factory default with a security level of <2low>. If there is an active password, the PPC2+ password entry screen will appear. The user must enter the user defined password or the factory secondary password to proceed any further: The first field is the serial number of the PPC2+, followed by a second field that represents the number of times that a secondary password has been used. This second field increments each time a secondary password is used. The third field is where the user enters the password.  >ROPKM^Y\c]OMYXNK\cZK]]aY\NS]K`KSVKLVOSXMK]O^RO_]O\]ZK]]aY\NRK] LOOX WS]ZVKMON Y\ PY\QY^^OX  3^ MKX LO YL^KSXON Lc MYX^KM^SXQ .23 -_]^YWO\=O\`SMO  >RO PKM^Y\c ]OMYXNK\c ZK]]aY\N S] NSPPO\OX^ PY\ KVV ::-]KXNMRKXQO]OKMR^SWOS^S]_]ON 3.4.4.7 ATM m PURPOSE To view the value of atmospheric pressure as measured by the on- board barometer. In the PPC2+ BG0002 model, the on-board barometer measures the pressure on the TEST(-) port (see the PPC2+ BG0002 Operation and Maintenance Manual). m PRINCIPLE PPC2+ has an independent on-board barometer. The atmospheric pressure measurements made by the on-board barometer are used only for dynamic compensation of atmospheric pressure when using an absolute reference pressure transducer to make gauge pressure measurements (see Sections 3.4.1, PRINCIPLE and 1.2.6). 2edit password Password: pppppp

PPC2+™ Operation and Maintenance Manual

Page 75 ©2001 DH Instruments, Inc.   >RO YXLYK\N LK\YWO^O\ S] K VYa KMM_\KMc ]OX]Y\ _]ON YXVc PY\ WOK]_\SXQ MRKXQO] SX K^WY]ZRO\SM Z\O]]_\O Y`O\ ]RY\^ ZO\SYN] YP ^SWO ]OO =OM^SYX:<38-3:6/::-WOK]_\OWOX^KMM_\KMcNYO]XY^NOZOXNYX^RO KL]YV_^OKMM_\KMcYP^ROYXLYK\NLK\YWO^O\  m OPERATION To view the current reading of the on-board barometer press [SPECIAL] and select <4Internal>, <7Atm>. The display is in the pressure units of the current PPC2+ range. 3.4.4.8 PURGE m PURPOSE To enable and disable the purge function which makes automated use of the optional SPLT (see Section 3.2.8.3). m OPERATION To access the purge enable/disable function press [SPECIAL] and select <4Internal>, <8Purge>. Select <1yes> to enable purge or <2no> to disable purge.  AROX ^RO Z_\QO P_XM^SYX S] OXKLVON ::- aSVV WKUO LKMUQ\Y_XN _]O YP `KV`O N\S`O\MRKXXOV "SXNOZOXNOX^VcYP_]O\KM^SYX]]OO=OM^SYX 3.4.4.9 SCRSAV m PURPOSE To adjust the PPC2+ screen saver function. m PRINCIPLE PPC2+ has a screen saver function which causes the display to dim after a front panel key is not pressed for a certain amount of time. This function is factory set to activate after 10 minutes of inactivity but can be adjusted by the user. m OPERATION To access the screen saver function press [SPECIAL] and select <4Internal>, <9ScrSav>. Edit the screen saver on time in minutes as desired.  =O^^SXQ]M\OOX]K`O\^SWO^YdO\YOVSWSXK^O]^RO]M\OOX]K`O\P_XM^SYX]Y^RK^ ^RONS]ZVKcaSVVZO\WKXOX^Vc\OWKSXK^P_VVL\SQR^XO]]

PPC2+™ Operation and Maintenance Manual

Page 77 ©2001 DH Instruments, Inc.  /9:/<+>398 4.1 OVERVIEW Most of the PPC2+ front panel functions can also be executed by commands from a remote computer. The host computer can communicate to the PPC2+ using the PPC2+ COM1 RS-232 port or the IEEE-488 port. Before writing test code which makes use of PPC2+ remote commands, familiarize yourself with its operating principles by reading Section 3 of this manual. 4.2 INTERFACING Sending a program message to the PPC2+ will place it into remote mode. The remote indicator in the lower left hand corner of the PPC2+ front panel will light when the PPC2+ is in remote mode. It will also flicker when a program message is received. The menus usually accessed from the front panel are locked out while in remote. The [ESCAPE] key returns the PPC2+ to local operation unless the program message was sent to the unit, which locks out all keypad operation. 4.2.1 RS-232 INTERFACE 4.2.1.1 COM1 The PPC2+ COM1 RS-232 interface is located on the back of the unit. It is a 9-pin male DB-9F connector configured as a DCE device. Data is transmitted out of the unit using pin 2, and is received on pin 3. This allows a normal pin-to-pin DB-9M to DB-9F RS-232 cable to be used to connect to a DTE host. Handshaking is not required or supported. The COM1 receive buffer is 80 bytes deep. If you overflow the buffer by sending too much data, the data will be lost. Because of this, you must send a single program message at a time and you must wait for the PPC2+ to reply from the previous command before issuing another command.

PPC2+™ Operation and Maintenance Manual

©2001 DH Instruments, Inc. Page 78  Table 10. Pin Designations and Connections PPC2+ COM1 DB-9F PIN DESIGNATIONS PIN # FUNCTION DESCRIPTION 2 TxD This pin transmits serial data from the PPC2+ to the host. 3 RxD This pin accepts serial data from the host computer. 5 Grn This pin is the common return for the TxD and RxD signals. IBM PC/XT DB-9F CONNECTIONS IBM PC/XT DB-9M TO PPC2+ DB9F CONNECTION DB-25M DB-9F DB-9M DB-9F 2 3 3 3 3 2 2 2 7 5 5 5 4.2.1.2 IEEE-488 The PPC2+ IEEE-488 interface is located on the back of the unit. The physical and electrical interface conforms to IEEE Std 488.1-1987 Subset E2 and IEEE Std. 488.2-1992. You should not attempt to communicate with the IEEE-488 interface while using the COM1 interface. The IEEE-488 receive buffer is 250 bytes deep. If you attempt to overflow the buffer, the PPC2+ will hold off release of the NRFD handshake line until it can service and empty the receive buffer. This keeps the buffer from overflowing. 4.2.1.3 COM2 The PPC2+ COM2 RS-232 interface is located on the back of the unit. It can be used to allow the Host computer to communicate with another device through the PPC2+. This allows the user to use one Host COM port to communicate with the PPC2+ and an additional RS-232 device. Refer to the remote program message for details. COM2 is a 9-pin female DB-9F connector configured as a DTE device. Data is transmitted out of the unit using pin 3, and is received on pin 2. This allows a normal pin-to-pin DB-9M to DB-9F RS-232 cable to be used to connect to a DCE device. Handshaking is not required or supported. Table 11. List of PPC2+ COM2 DB-9F Pin Designations PIN # FUNCTION DESCRIPTION 2 TxD This pin transmits serial data from the PPC2+ to a device. 3 RxD This pin accepts serial data from the external device. 5 Grn This pin is the common return for the TxD and RxD signals. 4 DTR This pin is Data Terminal Ready (DTR) (held at +5 V).

PPC2+™ Operation and Maintenance Manual

Page 79 ©2001 DH Instruments, Inc.  4.3 PROGRAMMING FORMATS The PPC2+ program message set is downward compatible with the previous PPC2 program message set. PPC2+ also introduces a new enhanced program message format to support IEEE Std 488.2 syntax, format, and status reporting. The user must select whether to use either the original “classic” program message format, or the new “enhanced” program message format. The enhanced format is recommended when possible, as the classic format is obsolete. Selection can be accomplished from the front panel (see Section 3.4.4.3) or remotely by using the “MSGFMT” program message. The default is the classic program message format to allow downward compatibility with existing host software. The basic commands are similar for both the classic and enhanced formats, but the usage, syntax, format and status reporting are different. The PPC2+ enhanced program message format also includes new additional common program messages that are required in every device that complies with the IEEE Std 488.2. These addition common program messages are covered separately in Section 4.6.1. 4.3.1 CLASSIC PROGRAM MESSAGE FORMAT The classic program message format is downward compatible with the previous PPC2 program message set. However, several PPC2 functions are not supported by PPC2 and neither are their corresponding commands (see DEVICE, DPG, ISO, PPC, HOLD, TOUT, TS, CONFIG10). This format is recommended only if you need downward compatibility with older PPC2 controllers. Each program message sent is also a query. You can only send one program message to the PPC2+ at time. After sending any program message, you must wait for the PPC2+ to reply before sending another program message. This reply will contain data, or a numeric error message if the program message was invalid. You must wait for this reply before issuing another program message to the PPC2+. This insures that the PPC2+ has completed the program message. Most remote program messages will return a reply within 500 ms except: PR, PRR, SR, ATM, RATE Up to 2.0 seconds to allow a new measurement. The syntax and format used for each program message in the classic mode is listed next to the keyword ‘Classic:’ in each program message summary in Section 4.4.4. 4.3.2 ENHANCED PROGRAM MESSAGE FORMAT This enhanced program message format uses the IEEE Std. 488.2 format, syntax and status reporting and is recommended for new applications. Errors are reported using the IEEE Std. 488.2 status reporting model. If an error is reported, the error is put into an Error Queue and the “ERR?” query program message can be used to get a text description of the most recent error. If you are using the IEEE-488 port, the service request line can be setup to be asserted if this occurs (see Section 4.5.2). There are two possible program message types for every program message. Each of these two types starts with the same basic text referred to as the program message header.

PPC2+™ Operation and Maintenance Manual

©2001 DH Instruments, Inc. Page 80  4.3.2.1 ENHANCED PROGRAM MESSAGE USING THE COMMAND TYPE The COMMAND type of program message executes a process and can additionally send data to the PPC2+ in the form of comma delimited arguments. This data is usually a setting of some sort that is stored in the PPC2+. If data is specified, it must be preceded by at least one white space from the program message header and be within the range and format described in the program message description. The keyword “Command:” will appear to the left of the required syntax in each program message description. If you are using the IEEE-488 port, the Command type does not generate a reply so you do not need to wait for a reply. You also may send multiple program messages at once by separating each program message with a semicolon. The commands will be queued and executed in as received order. If you are using the RS-232 port COM1, the Command type will always generate a reply so you must wait for a reply before issuing another program message. Because of this, you can only send one Command program message at a time while using the COM1 port. 4.3.2.2 ENHANCED PROGRAM MESSAGE USING THE QUERY TYPE The Query type of program message just requests data from the PPC2+. You must wait for a reply with a query. If you send any type of program message to the PPC2+ after a Query before receiving a reply, the program message will be discarded and an error will be generated. Errors are reporting using the IEEE Std. 488.2 status reporting model. A Query program message always ends with a question mark. Most queries will return a reply within 200 ms except: PR?, PRR?, SR?, ATM?, RATE? Up to 2.0 seconds to allow a new measurement. The syntax for using a QUERY program message is listed next to the keyword “Query:” in each program message summary in Section 4.4.4.

PPC2+™ Operation and Maintenance Manual

Page 81 ©2001 DH Instruments, Inc.  4.4 COMMANDS 4.4.1 PROGRAM MESSAGES Shown are the basic program messages for the classic and the enhanced formats. There are also additional IEEE Std. 488.2 common commands (they start with an ‘*’) that are common in all IEEE-488.2 compatible devices These commands and the enhanced format status reporting model are covered in Sections 4.5 and 4.6. Table 12. Program Message List # Send a command string out of the PPC2+ COM2 port. ABORT Stop pressure generation. AINCAL To set or retrieve the measurement modifiers for the current range. AINR To request the next available measurement of the analog inputs. AINRNG To set or retrieve the current analog input range. ATM Read the current atmospheric pressure (on-board barometer). AUTOZERO Read or set the status of the automatic zero function. AUTOPURGE Read or set the status of the automatic purge function. AUTOVAC Read or set the status of the control reference (EXHAUST port). COM1 Read or set the configuration of the COM1 port. COM2 Read or set the configuration of the COM2 port. CONFIG Obsolete PPC2 Command. DATE Read or set the current date. DEVICE Obsolete PPC2 Command. DF Decrease the pressure quickly. DP Decrease the pressure a given amount. DPG Obsolete PPC2 Command. DRV Read or set the status of an the external valve drivers. DS Decrease pressure slowly. ERR Read the last error message. HEAD Read or set the head settings. HOLD Obsolete PPC2 Command. HS Read or set the current control hold limit in pressure. HS% Read or set the current control hold limit in percent full scale. IF Increase the pressure quickly. IP Slowly increase the pressure a given amount. IS Increase the pressure slowly. ISO Obsolete PPC2 Command. LL Read or set the lower limit for the current range (PPC2+ BG0002 model only) LOCAL Returns control to the PPC2+ front panel. MEM Read the power-up memory test status. MODE Read or set the current pressure control mode. MSGFMT Read or set the type of program message format to use. NVENT Read or set the status of the lo vent valve (PPC2+ BG0002 model only) PASSTHRU Send a command string out of the PPC2+ COM2 port. PCALd,LO Read or set the user LO RPT calibration information. PCALd,HI Read or set the user HI RPT calibration information. PPC Obsolete PPC2 Command. PR Read the next PPC2+ pressure. PRR Read the next PPC2+ PRT pressure, rate, and ATM. PS Set a new target pressure and start automated pressure control. PSF Set a new target pressure and use only the fast speed to reach the target. PSH Obsolete PPC2 Command.

PPC2+™ Operation and Maintenance Manual

©2001 DH Instruments, Inc. Page 82  Table 12. Program Message List (Continued) PSS Set a new target pressure and use only the slow speed to reach the target. QPRR Read the last PPC2+ RPT pressure, rate and ATM. RANGE Read or set the active range of the PPC2+. RATE Read the next available rate of change of pressure. READY Obsolete PPC2 Command. READYCK Read or set a flag that is cleared by a Not Ready (<↑> or <↓>) condition. REMOTE Enable remote local lockout operation. RES Read or set the pressure display resolution for the current transducer and range. RESET Reset the PPC2+ to the default user parameters. RETURN Start a new automated pressure control set using the current target value. SCRSAV Read or set the front panel screen saver period. SN Read the serial number of the PPC2+. SR Read the next available pressure status. SS Read or set the stability required for a Ready (<*>) condition. SS% Read or set the stability required for a Ready (<*>) condition (% FS/ s). STAT Read the pressure generation status. TIME Read or set the current time of day. TOUT Obsolete PPC2 Command. TP Read the current target pressure. TS Obsolete PPC2 Command. UCOEF Convert a pressure in Pascal to pressure in the current units. UDU Read or set the user defined pressure unit. UL Read or set the upper limit for the current range. UNIT Read or set the pressure unit for the current range. VAC Read or set the exhaust reference status flag. VENT Read, execute or abort a vent process. VER Read the PPC2+ software version. ZNATERRd,HI Read or set the autozero natural error for Hi RPT. ZNATERRd,LO Read or set the autozero natural error for Lo RPT. ZOFFSETd,HI Read or set the autozero offset for Hi RPT. ZOFFSETd,LO Read or set the autozero offset for Lo RPT.

PPC2+™ Operation and Maintenance Manual

Page 83 ©2001 DH Instruments, Inc.  4.4.2 ERROR MESSAGES The PPC2+ will always reply to a program message while using the classic format. If the program message is incorrect or contains invalid data, an error number will be returned in the form where n is an integer number that represents a specific error. This allows for easy error trapping by the host computer. In the enhanced format, each error is placed into the Error Queue as it occurs. The query can then be used to remove from the queue and reply the text description of each error that has occurred. If you are using the COM1 port, the Error Queue is cleared each time a program message (except ) is received. Section 4.5 explains the details of the Error Queue. Table 13 is a list of the possible errors numbers and the error description for each. Table 13. Error #’s and Descriptions REPLY DESCRIPTION ERR #0 “OK” ERR #2 “Text argument is too long” ERR #3 “User defined coefficient cannot be 0” ERR #4 “External device not detected” ERR #5 “External device improperly configured” ERR #6 “Numeric argument missing or out of range” ERR #7 “Missing or improper command argument(s)” ERR #8 “External device time-out error” ERR #9 “Unknown command” ERR #11 “Command missing argument” ERR #12 “System overpressured” or “overpressure may result” ERR #13 “Text detected in numeric field” ERR #14 “User unit not defined” ERR #16 “Not available with gauge units” ERR #18 “Command not yet available” ERR #19 “Not available with absolute units” ERR #20 “Not available with gauge transducer” ERR #21 “User device not defined” ERR #22 “Pressure is not stable” ERR #23 “Option not available or installed” ERR #24 “Unit must be vented” ERR #25 “Transducer out of calibration” ERR #26 “COM port failed to initialize” ERR #27 “Internal device#1 time-out error” ERR #28 “Internal device#2 time-out error” ERR #30 “Must be on range H3” ERR #31 “Upper limit is too low” ERR #32 “Not stable enough” ERR #33 “Failed to generate pressure”

PPC2+™ Operation and Maintenance Manual

©2001 DH Instruments, Inc. Page 84  4.4.3 PROGRAM MESSAGE DESCRIPTION OVERVIEW Each program message description is separated into the following sections: Purpose A brief description of the programs message’s function. Command This is the Enhanced program message syntax to send data to the PPC2+ or to execute a PPC2+ function. The PPC2+ must be set to use the enhanced format (see Section 4.3). It may be sent alone, or followed by at least one white space and additional argument(s) to show that arguments can be passed. If there are multiple arguments, then commas must separate them. If you are using the IEEE-488 port, multiple Command type program messages can be sent in one message if you separate them with a semicolon. There will be no reply from the PPC2+ using the IEEE-488 port. If you are using the COM1 port, the PPC1 will reply and you must wait for this reply. If this field is not listed in the program message description, then the Command type is not supported when using the Enhanced format. Query This is the Enhanced program message syntax to request data from the PPC2+. The PPC2+ must be set to use the enhanced format (see Section 4.3.2). The PPC2+ will always reply to a query. You must wait for this reply before issuing another program message. If this field (Query) is not listed in the program message description, then the Query type for the program message is not supported when using the Enhanced format. Classic This is the Classic program message syntax to send data to the PPC2+, to execute a PPC2+ function, or to query for data. This type of command is not recommended for new applications. The PPC2+ must be set to use the classic format (see Section 4.3). It may be followed by a ‘(=)’ and additional argument characters to show that argument(s) can be passed. If there are multiple arguments, then commas must separate them. The PPC2+ will always reply to a Classic program message. You must wait for this reply before issuing another program message. If this field is not listed in the program message description, then it is not supported when using the classic format. Arguments If the program message can be used to set data inside the PPC2+, then this section will describe the arguments and their limits. Default If the program message can be used to set data inside the PPC2+, then this line will show (using the enhanced format) the default setting from the factory. Remarks This field has the details and remarks about the command. Example Examples are given for the enhanced and classic methods. Enhanced: An example of the use of an enhanced method program message to be sent to the PPC2+ is shown. The message sent to the PPC2+ appears after the “Cmd sent:” label. If only a Query type exists, the “Query sent:” label is shown instead. Directly under this label, “Query reply” shows a typical reply to a query type. “Reply:” shows that a query format does not exist. It may have a short description next to it. Classic: An example of the use of a classic program message to be sent to the PPC2+ is shown. The command sent to the PPC2+ appears after the “Cmd sent:” label. The “Reply” label shows a typical reply to the “Sent” example. It may have a short description next to it. Errors If the program message can report an argument error, the types of errors are listed. If using the classic format or the COM1 port, the error message will be replied after receiving the program message. If using the enhanced format via the IEEE-488 port, the error condition will be handled by the status reporting model which will store the errors in an Error Queue and can be programmed to assert the IEEE-488 SRQ line to signal an error has occurred. In either case, the “ERR” or “ERR?” program message can be used to retrieve a text description of the error. See Also Indicates related program messages (“----“) and refers to manual sections giving detail on PPC2+ operation corresponding to the program message.

PPC2+™ Operation and Maintenance Manual

Page 85 ©2001 DH Instruments, Inc.  4.4.4 PROGRAM MESSAGE DESCRIPTIONS  # Purpose To allow the host PC to communicate with a device connected to the PPC2+ COM2 port. This program message is obsolete. You should instead use the “PASSTHRU” program message. This command is not available when using the enhanced command format. Instead, use the “PASSTHRU” command. Classic “#xx” Arguments xx: The string to send out of the COM2 port. It must be less than 40 characters long. Remarks The PPC2+ COM2 port can be used to communicate to another RS-232 device (such as another PPC2+). This allows the user to use one COM port or IEEE-488 port on the host computer to communicate with the PPC2+ and another device. A carriage return and a line feed () are added to the string. After this program message is issued, the PPC2+ will reply back the first string received by the PPC2+ COM2 port that is terminated with a carriage return. Line feeds are discarded. This will discontinue when the next program message is sent to the PPC2+. There is no other reply from this program message. Prior to using this program message, you must ensure that the PPC2+ COM2 port is correctly set up to communicate with the device on COM2. Refer to the “COM2=“ program message. Example (classic) Sent: Reply: “#VER” “DH INSTRUMENTS, INC PPC2+ us A1000/A0015 Ver2.00 ” This example assumes that a second PPC2+’s COM1 port is connected to the PPC2+ COM2 port. This example gets the version of the second PPC2+. See Also “PASSTHRU”, “COM2” abort Purpose Stops an active pressure generation. All control valves are closed. The exhaust and transducer isolation valves are not affected. Command Classic “ABORT” “ABORT” Remarks This program message has no effect if the PPC2+ is not using automated pressure control. When using automated pressure control, it will abort the control. Example (enhanced) Cmd sent: Reply: “ABORT” “ABORT” (no reply if IEEE-488) Example (classic) Sent: Reply: “ABORT” “ABORT” See Also Sections 3.2.3, 3.1.2.3

PPC2+™ Operation and Maintenance Manual

©2001 DH Instruments, Inc. Page 86   aincal Purpose To set or retrieve the measurement modifiers (adder multiplier) for the current analog input range. Query Classic “AINCALn a,m” “AINCALn?” “AINCALn=a,m” “AINRNG” Default “AINRNGn 0,1” Arguments n: a: m: The input channel to modify (1 through 4). The adder modifier in the current range units (V, mV or mA). The multiplier modifier. Remarks These modifiers allow the user to adjust the measurement displayed. The Adder value is normally set at 0, and the multiplier is set to 1 to display the actual measured value. The displayed measurement is calculated as: Disp = (Meas m) + a Example (enhanced) Cmd sent: Query reply: “AINCAL1.012, 1.002” “0.012, 1.002” Example (classic) Cmd Sent: Reply: “AINCAL=.012, 1.002” “0.012, 1.002” Errors ERR# 6: The argument(s) are invalid. See Also Sections 3.2.9, 5.4 ainR Purpose To request the next available measurement of the analog inputs. Query Classic “AINR” “AINR?” Remarks The PPC2+ measures the analog inputs every 0.75 seconds. This command requests the next measurement cycle results from the PPC2+. The results are returned for all four analog inputs at once, Input #1 measurement is first, followed by the measurement unit. Then a comma is used to separate the first measurement from the next field. This is repeated for inputs #2, #3, and #4. If an auto-calibration cycle is occurring, “BUSY” will be returned instead. Example (enhanced) Cmd sent: Query reply: “AINR?” “0.797 mV, 0.934 mV, -0.027 mV, 1.234 mV” Example (classic) Sent: Reply: “AINR” “0.797 mV, 0.934 mV, -0.027 mV, 1.234mV” See Also Section 3.2.9

PPC2+™ Operation and Maintenance Manual

Page 87 ©2001 DH Instruments, Inc.   AINRNG Purpose To set or retrieve the current analog input range. Query Classic “AINRNG n” “AINRNG?” “AINRNGn” “AINRNG?” Default “AINRNG 3” Arguments n: The input range index 1 for 100 mv range 2 for 5 Volt range 3 for 10 Volt range 4 for 4-20 ma range Remarks The analog input option has 4 ranges. The selected range affects all four of the inputs at once. When a range is changed, an auto-calibration cycle occurs, which takes about 10 seconds. Example (enhanced) Cmd sent: Query reply: “AINRNG 2” “2” Example (classic) Cmd Sent: Reply: “AINRNG=2” “2” Errors ERR# 6: The argument(s) are invalid. See Also Section 3.2.9 ATM Purpose Reads the next measured internal ambient pressure. Query Classic “ATM?” “ATM” Remarks The atmospheric pressure as measured by the PPC2+ on-board barometer will be returned in the current pressure units (always absolute). This measurement is followed by the units text. Example (enhanced) Query sent: Query reply: “ATM? “97.123 kPaa” Example (classic) Sent: Reply: “ATM “97.123 kPaa” See Also Section 3.4.4.7

PPC2+™ Operation and Maintenance Manual

©2001 DH Instruments, Inc. Page 88   AUTOPURGE Purpose Read or set the status of the automatic purge function. Command Query “AUTOPURGE n” “AUTOPURGE?” Classic “AUTOPURGE=n” “AUTOPURGE” Default “AUTOPURGE 0” Arguments n: ’0’ To disable AUTO mode. ’1’ To enable AUTO mode. Remarks The PPC2+ can automatically control the external SPLT functions if desired. Example (enhanced) Cmd sent: Query reply: “AUTOPURGE 1” “0” (if PURGE not enabled) “1” (if PURGE enabled) Example (classic) Sent: Query reply: “AUTOPURGE=1” “AUTOPURGE=0” (if PURGE not enabled) “AUTOPURGE=1” (if PURGE enabled) Errors ERR# 6: The argument was other than a ‘0’ or a ‘1’. See Sections 3.4.4.8, 3.2.8 AUTOVAC Purpose Read or set the status of the automated Control Ref determination mode. Command Query “AUTOVAC n” “AUTOVAC?” Classic “AUTOVAC=n” “AUTOVAC” Default “AUTOVAC 1” Arguments n: ’0’ To disable AUTO mode. ’1’ To enable AUTO mode. This will override the manual selection. Remarks The PPC2+ has an internal sensor that can be used to automatically determine if the EXHAUST port is open to atmosphere or to a vacuum. You can disable it to manually override it. Example (enhanced) Cmd sent: Query reply: “AUTOVAC 1” “0” (if not in AUTO mode) “1” (if in AUTO mode) Example (classic) Sent: Query reply: “AUTOVAC=1” “AUTOVAC=0” (if not in AUTO mode) “AUTOVAC=1” (if in AUTO mode) Errors ERR# 6: The argument was other than a ‘0’ or a ‘1’. See Also Section 3.3.1, “VAC” program message

PPC2+™ Operation and Maintenance Manual

Page 89 ©2001 DH Instruments, Inc.   AUTOZERO Purpose Read or set the status of the automatic zero function. Command Query “AUTOZERO n” “AUTOZERO?” Classic “AUTOZERO=n” “AUTOZERO” Default “AUTOZERO 1” Arguments n: ’0’ Autozero OFF ’1’ Autozero ON Remarks The PPC2+ autozero function can be turned ON and OFF. Example (enhanced) Cmd sent: Query reply: “AUTOZERO 1” “0” If autozero OFF “1” If autozero ON Example (classic) Sent: Query reply: “AUTOZERO=1” “AUTOZERO=0” If autozero OFF “AUTOZERO=1” If autozero ON Errors ERR# 6: The argument was other than a ‘0’ or a ‘1’. See Also Section 3.4.1 COM1 Purpose Read or set the RS-232 settings for the COM1 port. Command Query “COM1 baud, parity, data, stop” “COM1?” Classic “COM1=baud, parity, data, stop” “COM1” Arguments baud: The baud rate. This may be ‘300’, ‘600’,’1200’,’2400’,’4800’, ‘9600’ or ‘19200’ parity: The data parity. This may be ‘O’ for odd, ‘E’ for even, or ‘N’ for none. data: The number of data bits. This may be ‘7’ or ‘8’ stop: The number of stop bits. This may be ‘1’ or ‘0’ Defaults “COM1 2400,E,7,1” Remarks The COM1 port is used to communicate to the PPC2+. When the COM1 port configuration of the PPC2+ is changed, the program message reply (COM1 use only) will be sent at the old COM1 settings, but all subsequent communications will be accomplished at the new COM1 settings. Example (enhanced) Cmd sent: Query reply: “COM1 9600,N,8,1” “9600,N,8,1” Example (classic)) Sent: Reply: “COM1=9600,N,8,1” “9600,N,8,1” Errors ERR# 7: Missing or improper program message argument(s). See Also Section 3.4.4.1, “PASSTHRU”

PPC2+™ Operation and Maintenance Manual

©2001 DH Instruments, Inc. Page 90   COM2 Purpose Read or set the RS-232 settings for the COM2 port. Command Query “COM2 baud,parity,data,stop” “COM2?” Classic “COM2= baud,parity,data,stop” “COM2” Arguments baud: The baud rate. This may be ‘300’, ‘600’,’1200’,’2400’,’4800’, ‘9600’ or ‘19200’ parity: The data parity. This may be ‘O’ for odd, ‘E’ for even, or ‘N’ for none. Data: The number of data bits. This may be ‘7’ or ‘8’ stop: The number of stop bits. This may be ‘1’ or ‘0’ Defaults “COM2 2400,E,7,1” Remarks The COM2 port can be used to allow the host computer to communicate through the PPC2+ to an additional device connected to COM2. This can be useful if the Host computer does not have 2 serial ports available. Example (enhanced) Cmd sent: Query reply: “COM2 9600,N,8,1” “9600,N,8,1” Example (classic) Sent: Reply: “COM2=9600,N,8,1” “9600,N,8,1” Errors ERR# 7: Missing or improper program message argument(s). See Also Section 3.4.4.3, “PASSTHRU” CONFIG This is an obsolete PPC2 Command Purpose PPC2+ does not support a remotely activated configuration routine. The CONFIG command from PPC2 will reply as if CONFIG had occurred. See Section 3.4.4.1 DATE Purpose Read or set the PPC2+ date. Command Query “DATE yymmdd” “DATE?” Classic “DATE=yymmdd” “DATE” Arguments yymmdd: The date in the numerical only format year-month-day with no spaces. Remarks The PPC2+ has an internal real time calendar clock. It is used for date stamping calibrations. Example (enhanced) Cmd sent: Query reply: “DATE 961201” “961201” Example (classic) Sent: Reply: “DATE=961201” “961201” Errors ERR# 7: Missing or improper program message argument(s). See Also Section 3.4.4.4

PPC2+™ Operation and Maintenance Manual

Page 91 ©2001 DH Instruments, Inc.   DEVICE This is an obsolete PPC2 Command. Purpose PPC2+ does not support external devices so the “DEVICE” Command is not supported. Will respond ERR# 9. See None DF Purpose Decrease the pressure quickly. Command “DF n” Classic “DF=n” Arguments ‘0’ ‘1’ Closes the fast down valve. Opens the fast down valve. Remarks Opening the fast down valve will cause the pressure to decrease quickly. Example (enhanced) Cmd sent: Reply: “DF 1” “1” (no reply if IEEE-488) Example (classic) Sent: Reply: “DF=1” “DF=1” Errors ERR# 6: The n argument is a ‘0’ or a ‘1’. See Also Section 5.8.1.7 DP Purpose Decrease the pressure slowly a given amount using the slow speed. Command “DP n” Classic “DP=n” Arguments n The decrease in pressure desired (current pressure units). This can be from 0 to 2 % FS of the current RPT range. Remarks The slow speed will be used for a calculated amount of time (up to 5 seconds) to create the desired change. The PPC2+ will not attempt to control the pressure to a target, so the change in pressure will be approximate. Example (enhanced) Cmd sent: Reply: “DP 2” “2.000 kPa”(no reply if IEEE-488) Example (classic) Sent: Reply: “DP=2” “2.000 kPa” Errors ERR# 6: The n argument is not within given limits. See Also None. DPG This is an obsolete PPC2 Command. Purpose PPC2+ does not support external devices so the “DPG” command is not supported. Will respond ERR# 9. See None

PPC2+™ Operation and Maintenance Manual

©2001 DH Instruments, Inc. Page 92   DRVN Purpose Read or set the status of the external electrical driver. Command Query “DRVn x” “DRVn?” Classic “DRVn=x” “DRVn” Arguments n x The driver to operate. This can be from 1 to 8. The state to change the driver to; ‘0’ to de-activate it, ‘1’ to activate it. Remarks The PPC2+ control has eight optional external drivers. Example (enhanced) Cmd sent: Query reply: “DRV1 1” “1” Example (classic) Sent: Reply: “DRV1=1” “DRV1=1” Errors ERR# 6: The n or x arguments are not within given limits. See Also Section 3.3.4 DS Purpose Decrease the pressure slowly. Command “DS n” Classic “DS=n” Arguments n ‘0’ Closes the slow down valve. ‘1’ Opens the slow down valve. Remarks Opening the slow down valve will cause the pressure to decrease slowly. Example (Enhanced) Cmd sent: Reply: “DS 1” “1” (no reply if IEEE-488) Example (Classic) Sent: Reply: “DS=1” “DS=1” Errors ERR# 6: The n argument is a ‘0’ or a ‘1’. See Also Section 5.8.1.7 ERR Purpose Read the new available error message from the Error Queue. Query “ERR?” Classic “ERR” Remarks This program message obtains additional details about an error that has occurred. If the user receives an “ERR# nn” reply, or the enhanced mode is enabled using the IEEE-488 interface and an error has been detected, the error is put into a FIFO Error Queue. The “ERR” program message pulls and replies the oldest error message available. “OK” will be replied if there are no error messages left. Example (enhanced): Query sent: Query reply: “ERR?” “Numeric argument missing or out of range” Example (classic) Sent: Reply: “ERR” “Numeric argument missing or out of range” See Also Section 4.4.2