wtw Oxi 197i Operating Manual
Text of the manual
RUN / ENTER
AUTO READ
CAL
STO
M
RCL
ON / OFF
Oxi 197i
mg/l
O
7
1 7 6
92
°C
ARng
TP
REL1
Portable oxygen measuring instrument
going to press
The use of advanced technology and the high quality standard of our
instruments are the result of a continuous development. This may re-
sult in differences between this operating manual and your instrument.
Also, we cannot guarantee that there are absolutely no errors in this
manual. Therefore, we are sure you will understand that we cannot ac-
cept any legal claims resulting from the data, figures or descriptions.
Copyright
© Weilheim 2004, WTW GmbH
Reproduction in whole - or even in part - is prohibited without the ex-
press written permission of WTW GmbH, Weilheim.
Printed in Germany.
List of contents
3
1
Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
1.1
Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
1.2
Keypad . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
1.3
Socket field . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
2
Safety . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
2.1
Authorized use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
2.2
General safety instructions . . . . . . . . . . . . . . . . . . . . . . . 10
3
Commissioning. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
3.1
Scope of delivery . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
3.2
Power supply . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
3.3
Initial commissioning . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
3.4
Sensor quiver . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
4
Operation. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
4.1
Switching on the measuring instrument . . . . . . . . . . . . . 15
4.2
Measuring . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
4.2.1
General information . . . . . . . . . . . . . . . . . . . . . . 16
4.2.2
Measuring the D. O. concentration . . . . . . . . . . 17
4.2.3
Measuring the D. O. saturation . . . . . . . . . . . . . 18
4.2.4
Automatic selection of the measuring range,
AutoRange . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
4.2.5
AutoRead AR (drift control) and hold function . . 19
4.2.6
Sensor with stirrer (StirrOx G) . . . . . . . . . . . . . . 20
4.3
Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
4.3.1
Calibrating with the air calibration vessel . . . . . 23
4.3.2
Calibrating via a comparison measurement . . . 25
4.3.3
Entering the salt content (salinity) . . . . . . . . . . . 27
4.4
Saving . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
4.4.1
Saving manually . . . . . . . . . . . . . . . . . . . . . . . . 28
4.4.2
Saving automatically . . . . . . . . . . . . . . . . . . . . . 30
4.4.3
Outputting the data storage . . . . . . . . . . . . . . . . 32
4.4.4
Clearing the memory . . . . . . . . . . . . . . . . . . . . . 36
4.5
Transmitting data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
4.5.1
Data transmission interval (Int 2) . . . . . . . . . . . 37
4.5.2
Recorder (analog output) . . . . . . . . . . . . . . . . . 39
4.5.3
PC/external printer (RS232 interface) . . . . . . . . 40
4.5.4
Remote control . . . . . . . . . . . . . . . . . . . . . . . . . 40
4.6
Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 41
4.7
Reset . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44
5
Maintenance, cleaning, disposal. . . . . . . . . . . . . . . . . 47
5.1
Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 47
List of contents
Oxi 197i
4
5.2
Cleaning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 47
5.3
Disposal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 47
6
What to do if... . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 49
7
Technical data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 53
7.1
General data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 53
7.2
Measuring ranges, resolutions, accuracies . . . . . . . . . . 55
8
Lists . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 57
The portable Oxi 197i measuring instrument enables you to carry out
oxygen measurements rapidly and reliably.
The Oxi 197i provides the maximum degree of operating comfort, reli-
ability and measuring certainty for all applications.
The proven OxiCal® calibration procedure supports you in your work
with the measuring instrument. The special AutoRead function enables
precise measurements.
1
Keypad
2
Display
3
Integrated, exchangeable sensor quiver
4
Socket field
5
Carrying and positioning handle
RUN / ENTER
AUTO READ
CAL
STO
M
RCL
ON / OFF
Oxi 197i
mg/l
O
7
1 7 6
92
°C
ARng
TP
REL1
1
2
3
4
5
If you need further information or application notes, you can obtain the
following material from WTW:
� Application reports
� Primers
� Safety datasheets.
You will find information on available literature in the WTW catalog or
via the Internet.
1.1
Display
Sal
pH
mV/
m
cm
S/
cm
1/
pH
nLF
TEC
Tref25
Tref20
REL2
REL1
Lin
/K
%
Sal
%mg/l
S
O
8
8 8 8 8
8 8
1
°C
Time
Year
LoBat
Oxi
ARng
AR
RCL
Ident
Store
TP
No.
Day.Month
Baud
Auto
Cal
Status display indicator
Measured value
Function and
Temperature display
Sensor symbol
display
Switch measuring instrument on/off
You can use the following sensors with the Oxi 197i:
� Oxygen depth armature
� D. O. sensor CellOx 325, or
� D. O. sensor DurOx 325, or
� Self-stirring D. O. sensor StirrOx G
Connectors
Warning
Only connect sensors to the measuring instrument that cannot return
any voltages or currents that are not allowed (> SELV and > current cir-
cuit with current limiting). Nearly all sensors - especially WTW sensors
- fulfill these conditions.
Sensor / Instrument
Socket
Oxygen sensor
1
Integrated stirrer StirrOx G
2
Printer or PC (serial interface, RS232)
3
Plug-in power supply unit
4
Watertight valve for internal pressure equalization
5
2
1
3
4
5
This operating manual contains basic instructions that you must follow
during the commissioning, operation and maintenance of the measur-
ing instrument . Consequently, all responsible personnel must read this
operating manual before working with the measuring system. The op-
erating manual must always be available within the vicinity of the mea-
suring system.
Target group
The measuring instrument was developed for work in the field and in
the laboratory.
We assume that, as a result of their professional training and experi-
ence, the operators will know the necessary safety precautions to take
when handling chemicals.
Safety
instructions
The individual chapters of this operating manual use safety instructions
such as the label shown below to indicate various hazards or dangers:
Warning
indicates instructions that must be followed precisely in order to avoid
the possibility of slight injuries or damage to the instrument or the envi-
ronment.
Further notes
Note
indicates notes that draw your attention to special features.
Note
indicates cross-references to other documents, e.g. operating manu-
als.
2.1
Authorized use
The authorized use of the measuring instrument consists exclusively of
the measurement of the dissolved oxygen (D. O.) content of liquid me-
dia in the field and laboratory.
The technical specifications as given in chapter 7 TECHNICAL DATA must
be observed. Only the operation and running of the measuring instru-
ment according to the instructions given in this operating manual is au-
thorized.
Any other use is considered to be unauthorized.
General safety instructions
This instrument is built and inspected according to the relevant guide-
lines and norms for electronic measuring instruments (see chapter
7 TECHNICAL DATA).
It left the factory in a safe and secure technical condition.
Function and operating
safety
The smooth functioning and operational safety of the measuring instru-
ment can only be guaranteed if the generally applicable safety mea-
sures and the specific safety instructions in this operating manual are
followed during operation.
The smooth functioning and operational safety of the measuring instru-
ment can only be guaranteed under the environmental conditions that
are specified in chapter 7 TECHNICAL DATA.
If the instrument was transported from a cold environment to a warm
environment, the formation of condensate can lead to the faulty func-
tioning of the instrument. In this event, wait until the temperature of the
instrument reaches room temperature before putting the instrument
back into operation.
Safe operation
If safe operation is no longer possible, the instrument must be taken out
of service and secured against inadvertent operation!
Safe operation is no longer possible if the measuring instrument:
� has been damaged in transport
� has been stored under adverse conditions for a lengthy period of
time
� is visibly damaged
� no longer operates as described in this manual.
If you are in any doubt, please contact the supplier of the instrument.
Obligations
of the purchaser
The purchaser of the measuring instrument must ensure that the fol-
lowing laws and guidelines are observed when using dangerous sub-
stances:
� EEC directives for protective labor legislation
� National protective labor legislation
� Safety regulations
� Safety datasheets of the chemical manufacturers.
Scope of delivery
� Portable measuring instrument, Oxi 197i with integrated recharge-
able battery
� Carrying and positioning handle
� Carrying strap
� Sensor quiver
� Plug-in power supply unit
� Operating manual
3.2
Power supply
Mains operation and
charging the battery
You can either operate the measuring instrument with the integrated re-
chargeable battery or with the plug-in power supply. The plug-in power
supply supplies the measuring instrument with low voltage (12 V DC).
At the same time, the rechargeable battery is charged.
Note
The automatic stirring function of the self-stirring D. O. sensor
StirrOx G can only be used when the plug-in power supply is connect-
ed.
Charging time of the
battery
approx. 16 hours. The battery is charged even when the instrument is
switched off. The LoBat display indicator appears when the battery is
nearly empty and has to be charged as soon as possible.
Warning
The line voltage at the operating site must lie within the input voltage
range of the original plug-in power supply (see chapter 7 TECHNICAL DA-
TA).
Warning
Use original plug-in power supplies only (see
chapter 7 TECHNICAL DATA).
Connecting the plug-in
power supply unit
3.3
Initial commissioning
Perform the following activities:
� For mains operation and charging the battery: Connect the plug-in
power supply (see section 3.2 POWER SUPPLY).
� Set the date and time.
Setting the date and
time
1
Connect the plug (1) to the socket (2) of the measuring instru-
ment.
2
Connect the original WTW plug-in power supply (3) to an easily
accessible mains socket.
2
1
3
1
Press the key and hold it down.
2
Press the key.
The display test appears briefly on the display.
3
Press the key repeatedly until the date flashes
on the display (Day.Month display indicator).
Set the date of the current day with .
5
Confirm with .
The date (month) flashes in the display.
6
Set the current month with .
7
Confirm with .
The year appears on the display.
8
Set the current year with .
9
Confirm with .
The hours flash on the display.
10
Set the current time with .
11
Confirm with .
The minutes flash on the display.
12
Set the current time with .
13
Confirm with .
The instrument switches to the measuring mode.
REL1
Tref25
5
t 2
Time
Cal
I
O
AR
cm
1/
TP
ARng
nLF
°C
cm
S/
Sal
RCL
%
S
%
m
mg/l
Oxi
Year
Ident
No.
Tref20
LoBat
Lin
Sal
pH
TEC
Auto
Store
Baud
K
/
REL2
/pH
mV
Day.Month
0 9 0 4
To store the sensors during field operation and to keep the sensor ele-
ment moist, the quiver tip contains a sponge rubber insert that can be
moistened with deionized water.
Note
For further details on proper storage, refer to the operating manual of
the sensor.
Moistening the
quiver insert
1
Press the quiver out of the holder from the back side of the in-
strument and pull it out completely.
2
Pull off the quiver tip and moisten the sponge rubber with
deionized water.
Switching on the measuring instrument
Note
The measuring instrument has an energy saving feature to avoid un-
necessary battery depletion. The energy saving feature switches the
measuring instrument off if no key has been pressed for an hour. The
energy saving feature is not active when the AutoStore function is ac-
tive.
The energy saving feature is also not active
� if the power is supplied by the plug-in power supply,
� if the communication cable and a PC with a running communication
program are connected,
� if the recorder cable is connected,
� if the printer cable is connected (for external printers).
1
Connect the D. O. sensor to the measuring instrument.
2
Press the key.
The display test appears briefly on the display.
The relative slope for the sensor type that was just connected
subsequently appears for approx. one second.
The measuring instrument then automatically switches to the
measuring mode that was last selected. The display shows the
relevant measured value.
You can measure the following variables:
� D. O. concentration
� D. O. saturation
The measuring instrument is supplied with the following functions:
� AutoRange (automatic switchover of the measurement range). If a
measuring range is exceeded, AutoRange causes the measuring in-
strument to change automatically to the next higher measuring
range and back again. Therefore, the instrument always measures
in the measuring range with the highest possible resolution. The
function can be switched off.
� The AutoRead function (drift control) for checking the stability of the
measurement signal. This ensures the reproducibility of the measur-
ing signal. For details of how to switch the AutoRead function on/off,
see page 19.
Preparatory activities
Perform the following preparatory activities when you want to measure:
Note
Incorrect calibration of D. O. sensors will result in incorrect measured
values. Calibrate at regular intervals.
Note
The measuring instrument automatically recognizes the type of the D.
O. sensor (CellOx 325 or DurOx 325).
Temperature sensor
The D. O. sensor has an integrated temperature sensor that always
measures the current temperature of the test sample.
Warning
When connecting an earthed PC/printer, measurements cannot be per-
formed in earthed media as incorrect values would result. The RS232
interface is not galvanically isolated.
1
Connect the D. O. sensor to the measuring instrument (see
section 1.3 SOCKET FIELD).
2
Calibrate or check the measuring instrument with the D. O.
sensor. How to calibrate is described in section 4.3.
3
Select the measuring mode with .
Measuring the D. O. concentration
Note
When measuring the concentration of test samples with a salt content
of more than 1 g/l, a salinity correction is required. For this, you have to
measure and input the salinity of the measured medium first. This is de-
scribed in section 4.3.3 on page 27.
Switching the salt content correction on or off, see below.
To measure the D. O. concentration with and without salt content cor-
rection, proceed as follows:
Switching on/off the salt
content correction
Proceed as follows to switch on the salt content correction:
1
Perform the preparatory activities according to section 4.2.1.
2
Immerse the D. O. sensor in the test sample.
3
Press the key repeatedly until the D. O. concentration in
mg/l appears on the display.
REL1
Sal
S
pH
AR
1
LoBat
nLF
Lin
Oxi
Tref25
Tref20
Cal
TEC
Auto
Store
8
Time
Year
Ident
No.
Day.Month
Baud
1/cm Sal
/K
%
RCL
REL2
/pH
mV
m
%S/cm
7
1 7 6
9 2
°C
TP
O
ARng
mg/l
1
In the D. O. concentration measuring mode, switch on the salt
content correction with . The SAL display indicator ap-
pears on the display. The value that was entered is taken into
consideration during the measurement (see section 4.3.3).
2
Switch the salt content correction off with . The SAL dis-
play indicator is no longer displayed.
REL1
Sal
S
pH
AR
1
LoBat
nLF
Lin
Oxi
Tref25
Tref20
Cal
TEC
Auto
Store
8
Time
Year
Ident
No.
Day.Month
Baud
1/cm
/K
%
RCL
REL2
/pH
mV
m
%S/cm
5
1 7 6
4 2
°C
TP
O
ARng
mg/l
Sal
Measuring the D. O. saturation
You can measure the D. O. saturation as follows:
4.2.4
Automatic selection of the measuring range, AutoRange
Sensors
You can use the sensors listed in section 1.3 on page 8. The sensor
type is automatically recognized and the resolution is adjusted if nec-
essary (see chapter 7 TECHNICAL DATA).
AutoRange
Two measuring ranges are available for each measuring mode. If a
measuring range is exceeded, AutoRange causes the measuring in-
strument to change automatically to the next higher measuring range
and back again. Therefore, the instrument always measures in the
measuring range with the highest possible resolution.
To switch the automatic measurement range selection on/off, see
section 4.6 CONFIGURATION. When the AutoRange function is switched
on, ARng appears on the display.
1
Perform the preparatory activities according to section 4.2.1
2
Immerse the D. O. sensor in the test sample.
3
Press the key repeatedly until the saturation in % appears
on the display.
Sal
REL1
mg/l
/
mV
m
Sal
S
pH
AR
1
LoBat
nLF
Lin
Oxi
Tref25
Tref20
Cal
TEC
Auto
Store
8
Time
Year
Ident
No.
Day.Month
Baud
1/cm
/K
%
RCL
REL2
pH
S/cm
8
1 7 6
9 2
°C
TP
O
ARng
%
Sal
S
pH
1
LoBat
nLF
Lin
Oxi
Tref25
Tref20
Cal
TEC
Auto
Store
8
Time
Year
Ident
No.
Day.Month
Baud
1/cm Sal
/K
%
RCL
REL2
/pH
mV
m
%S/cm
7
1 7 6
0 8
°C
TP
REL1
O
ARng
mg/l
AR
AutoRead AR (drift control) and hold function
The AutoRead function (drift control) checks the stability of the mea-
surement signal. The stability has a considerable impact on the repro-
ducibility of the measured values. With the aid of the hold function the
measured value display is frozen.
Criteria
With identical measurement conditions, the following criteria are valid
for the AutoRead function:
Use the AutoRead function and hold function like this with the sensors
� Oxygen depth armature
� CellOx 325
� DurOx 325
Note
The special features of the self-stirring D. O. sensor StirrOx G can be
found in section 4.2.6 on page 20.
Measuring mode
Reproducibility
Response time
D. O. concentra-
tion
better than 0.05 mg/l
> 10 seconds
D. O. saturation
index
better than 0.6 %
> 10 seconds
1
Select the measuring mode with .
2
Activate the AutoRead function with . The current mea-
sured value is frozen (hold function).
3
Start AutoRead with .
AR flashes until a stable measured value is reached.
4
If necessary, start the next AutoRead measurement with
.
REL1
Sal
S
pH
1
LoBat
nLF
Lin
Oxi
Tref25
Tref20
Cal
TEC
Auto
Store
8
Time
Year
Ident
No.
Day.Month
Baud
1/cm Sal
/K
%
RCL
REL2
/pH
mV
m
%S/cm
7
1 7 6
0 8
°C
TP
O
ARng
mg/l
AR
Sensor with stirrer (StirrOx G)
Connection
Connect the 2-pole stirrer plug to the stirrer output of the measuring in-
strument (designation REL1).
Note
The automatic stirring function of the self-stirring D. O. sensor
StirrOx G can only be used when the plug-in power supply is connect-
ed.
Stirrer function
The D. O. sensor, StirrOx G, has a temperature sensor and stirrer inte-
grated. Using the key on the sensor you switch the stirrer on.
Measuring with
AutoRead
You can switch on the stirrer and start an AutoRead measurement at
the same time. To do so, proceed as follows:
5
To terminate the AutoRead function: Press the key.
1
Select the measuring mode with .
2
Activate the AutoRead function with . The current mea-
sured value is frozen (hold function).
3
Submerse the sensor in the test sample and press the stirrer
button. The stirrer is startet. At the same time, the AutoRead
measurement starts. AR flashes until a stable measured value
is achieved. Subsequently, the stirrer switches itself off.
REL1
Sal
S
pH
1
LoBat
nLF
Lin
Oxi
Tref25
Tref20
Cal
TEC
Auto
Store
8
Time
Year
Ident
No.
Day.Month
Baud
1/cm Sal
/K
%
RCL
REL2
/pH
mV
m
%S/cm
7
1 7 6
0 8
°C
TP
O
ARng
mg/l
AR
D. O. sensors age. This changes the slope of the D. O. sensor. Calibra-
tion determines the current slope of the sensor and stores this value in
the instrument.
Thus, you should calibrate at regular intervals.
Calibration data records
The Oxi 197i administrates two sets of calibration data:
Sensors of different classes can be calibrated separately. When a sen-
sor of one class is calibrated the calibration data of the other class re-
main stored. The Oxi 197i recognizes the class of the connected
sensor and automatically uses the correct calibration data.
Calibration procedures
The Oxi 197i provides 2 calibration procedures:
� Calibration in water vapor-saturated air.
To calibrate the CellOx 325, use theOxiCal®-SL air calibration ves-
sel (accessory), to calibrate the DurOx 325, use the OxiCal®-D air
calibration vessel (contained in the scope of delivery of the sensor).
� Calibration via a comparison measurement (e. g. Winkler titration
according to DIN EN 25813 or ISO 5813). At the same time, the rel-
ative slope is adapted to the comparison measurement by a correc-
tion factor.
When to calibrate?
� After connecting another D. O. sensor
� If the sensor symbol flashes, i. e. after the calibration interval has ex-
pired
Calibration interval
The calibration interval (Int 3) is set to 14 days in the factory. You can
select the interval in the range of 1 ... 999 days.
AutoRead
The calibration procedure automatically activates the AutoRead func-
tion.
Displaying calibration
data
Each time the instrument is switched on, the calibration data are shown
on the display for a short time (see section 4.1 SWITCHING ON THE MEA-
SURING INSTRUMENT). In order to view the calibration data, switch the
measuring instrument off and switch it on again.
� Set 1, for sensor class 1:
– Oxygen depth armature
– CellOx 325, or
– StirrOx G
� Set 2, for sensor class 2:
– DurOx 325
The calibration protocol contains the calibration data of the current cal-
ibration. You can transmit the calibration protocol to a printer via the se-
rial interface (see page 35).
Note
You can automatically print a calibration record after the calibration. To
do so, connect a printer to the interface according to section 4.5.3 be-
fore calibrating. After a valid calibration, the record is printed.
Sample printout:
Sensor evaluation
After the calibration, the measuring instrument evaluates the current
status of the sensor against the relative slope. The evaluation appears
on the display. The relative slope has no effect on the measuring accu-
racy. Low values indicate that the electrolyte will soon be depleted and
the sensor will have to be regenerated.
CALIBRATION PROTOCOL
02.03.02 14:19
Device No.: 12345678
CALIBRATION 02
Cal time: 02.03.02 / 14:19
Cal interval: 14d
OxiCal Tauto AR
Relative Slope: 0,88
Probe: +++
Display
Relative slope
S = 0.8 ... 1.25
S = 0.7 ... 0.8
S = 0.6 ... 0.7
E3
Eliminate the error according to
chapter 6 WHAT TO DO IF...
S < 0.6 or S > 1.25
Calibrating with the air calibration vessel
Proceed as follows to calibrate the instrument:
Note
The sponge in the air calibration vessel must be moist (not wet). Follow
the instructions in the OxiCal®-SL or DurOx 325 operating manual.
1
Connect the D. O. sensor to the measuring instrument.
2
Keep an air calibration vessel ready for use.
3
Put the D. O. sensor into the air calibration vessel.
4
Press the key repeatedly until O2 CAL appears.
5
Press the key. AutoRead is active, AR flashes.
6
As soon as a stable value is achieved, AR stops flashing. Cal-
ibration is finished; the sensor symbol indicates the relative
slope determined and the sensor evaluation (see page 22).
REL1
S
AR
Year
Ident
No.
Day.Month
Tref25
Tref20
LoBat
Lin
nLF
mg/l
ARng
Sal
pH
1
TEC
Auto
Store
8
Time
Baud
1/cm Sal
/K
%
RCL
REL2
/pH
mV
m
%S/cm
C
1 7 6
A L
°C
TP
O
Oxi
Cal
REL1
Year
Ident
No.
Day.Month
Tref25
Tref20
LoBat
Lin
nLF
mg/l
ARng
Sal
pH
1
TEC
Auto
Store
8
Time
Baud
1/cm Sal
/K
%
RCL
REL2
/pH
mV
m
%S/cm
0
1 7 6
8 8
°C
TP
O S
AR
Oxi
Cal
In chapter 6 WHAT TO DO IF... page 49, you will find the measures to
take for error elimination.
7
Switch to the measuring mode with .
REL1
Year
Ident
No.
Day.Month
Tref25
Tref20
LoBat
Lin
nLF
mg/l
ARng
Sal
pH
1
TEC
Auto
Store
8
Time
Baud
1/cm Sal
/K
%
RCL
REL2
/pH
mV
m
%S/cm
0
1 7 6
8 8
°C
TP
O S
AR
Oxi
Cal
Calibrating via a comparison measurement
Proceed as follows to calibrate the instrument:
1
Connect the D. O. sensor to the measuring instrument.
2
Press the key repeatedly until O2 FAC appears.
3
Immerse the D. O. sensor in the calibrating solution.
4
Press the key.
The AutoRead measurement begins. If the measured value is
stable, the instrument displays the determined measured value
and the correction factor of the relative slope (initial value =
1.000).
5
Remove the D. O. sensor from the calibrating solution.
6
Perform a comparison measurement, e. g. Winkler titration,
with the calibrating solution (determine the nominal value).
Leave the measuring instrument switched on until the nominal
value is determined.
7
Set the displayed value on the measuring instrument to the
nominal value with . This changes the related correc-
tion factor of the relative slope.
REL1
S
AR
Year
Ident
No.
Day.Month
Tref25
Tref20
LoBat
Lin
nLF
mg/l
ARng
Sal
pH
1
TEC
Auto
Store
8
Time
Baud
1/cm Sal
/K
%
RCL
REL2
/pH
mV
m
%S/cm
F
1 7 6
A C
°C
TP
O
Oxi
Cal
TP
REL1
S
Year
Ident
No.
Day.Month
Tref25
Tref20
LoBat
Lin
nLF
Sal
pH
1
TEC
Auto
Store
Time
Baud
1/cm Sal
/K
%
RCL
REL2
/pH
mV
m
%S/cm
4
0 0 0
4 2
°C
O
1
AR
ARng
Oxi
Cal
mg/l
If the correction factor deviates from 1.000, the display of the measured
value units flashes.
8
Switch to the measuring mode with or .
TP
REL1
S
Year
Ident
No.
Day.Month
Tref25
Tref20
LoBat
Lin
nLF
Sal
pH
1
TEC
Auto
Store
Time
Baud
1/cm Sal
/K
%
RCL
REL2
/pH
mV
m
%S/cm
4
0 0 7
4 5
°C
O
1
AR
ARng
Oxi
Cal
mg/l
Entering the salt content (salinity)
A salt content correction is required in the oxygen concentration mea-
surement of samples with a salt content of more than 1 g/l. To do this,
you have to enter the salinity equivalent (the measured salinity) of the
test sample (range 0.0 - 70.0) and to switch on the salinity correction.
Entering the
salt content
Note
How to switch on the salt content correction is described on page 17.
Parameter
Value range
Salinity
0.0 ... 70.0 in steps of 0.1
1
Determining the salinity of the test sample (any method).
2
Press the key repeatedly until Sal appears on the dis-
play.
3
Enter the salt content with .
4
Switch to the measuring mode with .
REL1
TP
Oxi
Cal
AR
°C
S
Year
Ident
No.
Day.Month
Tref25
Tref20
LoBat
Lin
nLF
Sal
pH
1
TEC
Auto
Store
8
Time
Baud
1/cm
/K
%
RCL
REL2
/pH
mV
m
%S/cm
5
6 9 6
5 0
O
mg/l
Sal
ARng
The measuring instrument has an internal data memory. It can store up
to 500 datasets.
A complete data record consists of:
� Storage location
� Date/time
� Measured value
� Temperature
� Temperature measuring procedure
� ID number
You can transmit measured values (data records) to the data storage
in two ways:
� Save manually
� Switch on AutoStore (Int 1).
4.4.1
Saving manually
You can transmit a measured value to the data storage as follows:
1
Press the key.
The current number (location number No.) of the next free stor-
age location appears under the current measured value on the
display.
2
Confirm with .
The display switches to entering the ID number.
RCL
REL1
2 5
°C
TP
Sal
S
pH
AR
1
LoBat
nLF
Lin
Oxi
Tref25
Tref20
Cal
TEC
Auto
8
Time
Year
Ident
Day.Month
Baud
1/cm Sal
/K
%
REL2
/pH
mV
m
%S/cm
8
4
3 7
O
mg/l
ARng
No.
Store
This message appears when all of the 500 storage locations are occu-
pied.
You have the following options:
3
Using , enter the required ID number
(1 ... 999).
4
Confirm with .
The instrument changes to the measuring mode.
No.
RCL
REL1
2 5
°C
TP
Sal
S
pH
AR
1
LoBat
nLF
Lin
Oxi
Tref25
Tref20
Cal
TEC
Auto
8
Time
Year
Day.Month
Baud
1/cm Sal
/K
%
REL2
/pH
mV
m
%S/cm
8
2
3 7
O
mg/l
ARng
Store
Ident
Saving the current measured value.
The oldest measured value (storage location 1)
will be overwritten by this
Press
Returning to the measuring mode without
saving
press any key
Outputting the data storage
see
section 4.4.3
Clearing the memory
see
section 4.4.4
The save interval (Int 1) determines the chronological interval between
automatic save processes.
After the fixed interval has expired, the current data record is transmit-
ted to the storage and to the interface.
Setting the
save interval:
The default setting for the save interval (Int 1) is OFF.
By this, the AutoStore function is switched off.
To switch the function on, set an interval (5 s, 10 s, 30 s, 1 min, 5 min,
10 min, 15 min, 30 min, 60 min):
1
Press the key and hold it down.
2
Press the key. Int 1 appears on the display.
3
Set the required interval between the saving procedures with
.
4
Confirm with .
The number of free memory locations appears on the display.
5
As soon as all of the 500 storage locations are occupied,
AutoStore is terminated (Int 1 = OFF).
If there are not enough storage locations available for your
measurements:
– Output and backup the data storage (see page 32) and
– clear the memory (see page 36).
REL1
pH
Ident
No.
Time
Cal
d
O
AR
cm
1/
Tref25
TP
ARng
nLF
°C
cm
S/
Sal
RCL
%
S
%
m
mg/l
Oxi
Year
Day.Month
Tref20
LoBat
Lin
Sal
TEC
Baud
K
/
REL2
/pH
mV
F
1
n t
Store
I
F
O
Auto
REL1
I
pH
Ident
No.
Time
Cal
O
AR
cm
1/
Tref25
TP
ARng
nLF
°C
cm
S/
Sal
RCL
%
S
%
m
mg/l
Oxi
Year
Day.Month
Tref20
LoBat
Lin
Sal
TEC
Baud
K
/
REL2
/pH
mV
E
9
4 9
Store
E
R
Auto
F
The AutoStore function is interrupted if you start other functions, e.g.
output the data storage.
After the function is finished, the AutoStore function is continued. By
this, however, temporal gaps in the recording of the measured values
will occur.
Switching off AutoStore
Switch AutoStore off by:
� setting the save interval (Int 1) to OFF, or
� switching the measuring instrument off and then on again.
6
Confirm with .
The prompt for the ID number appears on the display.
7
Set the required ID number with .
8
Confirm with .
The instrument switches to the Oxi measuring mode and starts
the measuring and saving process.
AutoStore flashes on the display.
REL1
9
4 9
R E
F
I
pH
No.
Time
Cal
O
AR
cm
1/
Tref25
TP
ARng
nLF
°C
cm
S/
Sal
RCL
%
S
%
m
mg/l
Oxi
Year
Day.Month
Tref20
LoBat
Lin
Sal
TEC
Baud
K
/
REL2
/pH
mV
1
Store
Auto
Ident
Outputting the data storage
You can output the contents of the data storage:
� Stored data on the display
� Calibration data on the display
� Stored data on the serial interface
� Calibration protocol on the interface
Outputting stored data
on the display
You can perform the following activities:
1
Press the key repeatedly until StO dISP appears on the
display.
2
Press the key.
A measured value appears on the display.
The storage location of the data record is displayed for approx.
2 s, then the respective temperature appears.
Display further elements of the data record
(ID number, date, time, storage location)
Press
Advance one data record (storage location)
Press
Go back one data record (storage location)
Press
REL1
Store
Auto
Ident
I
pH
No.
Time
Cal
O
AR
cm
1/
Tref25
TP
ARng
nLF
°C
cm
S/
Sal
%
S
%
m
mg/l
Oxi
Year
Day.Month
Tref20
LoBat
Lin
Sal
TEC
Baud
K
/
REL2
/pH
mV
o
S T
I S
D
RCL
P
Sal
REL1
Sal
S
pH
AR
1
LoBat
nLF
Lin
Oxi
Tref25
Tref20
Cal
TEC
Auto
Store
8
Time
Year
Ident
No.
Day.Month
Baud
1/cm
/K
%
REL2
/pH
mV
m
%S/cm
6
2 4 7
0 3
°C
TP
O
ARng
mg/l
RCL
If you want to search for a certain element (e.g. date), proceed as fol-
lows:
Outputting stored data
to the interface
Note
You can cancel the transmission with or .
After the instrument number, the printout contains the complete storage
contents in ascending order of the storage location numbers.
1
Using , select the element (e.g. date).
2
Press or repeatedly until the required date appears
on the display.
After approx. 2 s the temperature of the displayed measured
value appears.
1
Press the key repeatedly until Sto SEr appears on the
display.
2
Press the key.
The complete storage content is transmitted to the interface;
during the data transmission the numbers of the currently
transmitted storage locations run through. After the data trans-
mission, the measuring instrument automatically switches to
the measuring mode.
REL1
D
Store
Auto
Ident
I
pH
No.
Time
Cal
O
AR
cm
1/
Tref25
TP
ARng
nLF
°C
cm
S/
Sal
%
S
%
m
mg/l
Oxi
Year
Day.Month
Tref20
LoBat
Lin
Sal
TEC
Baud
K
/
REL2
/pH
mV
o
S T
E
S
RCL
r
on the display
Device No.: 99990000
No. 1:
09.03.02 17:10
101.7 % 17.6° C
Tauto AR
Ident : 10
No. 2:
09.03.02 17:11
7.11 mbar 17.6° C
Tauto AR
Ident : 10
No. 3:
09.03.02 17:12
7.88 mg/l 17.6° C
Tauto
Ident : 10
No. 4:
09.03.02 17:15
7.11 mg/l 17.8° C
Tauto
SAL = 17.9
Ident : 7
...
1
Press the key repeatedly until CAL dISP appears on
the display.
2
Press the key.
The relative slope appears on the display:
REL2
RCL
REL1
Sal
mg/l
ARng
O
TP
Oxi
Cal
AR
°C
S
Year
Ident
No.
Day.Month
Tref25
Tref20
LoBat
Lin
nLF
Sal
pH
C
TEC
Auto
Store
Time
Baud
1/cm
/K
%
/pH
mV
m
%S/cm
C
I S P
A L
d
mV
Oxi
TEC
Auto
pH
pH
/
TP
1/cm Sal
% /K
°C
Year
Ident
1
Store
AR
LoBat
nLF
Lin
Tref25
Tref20
ARng
8
Time
No.
Day.Month
Baud
REL2
m
S/cm
%mg/l
Sal
8 8
REL1
Cal
S L
RCL
0
S
O
0
on the interface
Note
You will find a sample calibration protocol in PRINTING THE CALIBRATION
RECORD, page 22.
3
Using or , you can switch back to the
measuring mode.
1
Press the key repeatedly until CAL SEr appears on the
display.
2
Press the key.
The calibration protocol is transmitted to the interface.
After the data transmission, the measuring instrument auto-
matically switches to the measuring mode.
REL2
TP
AR
/K
%
REL1
d
Sal
mg/l
ARng
O
Oxi
Cal
°C
S
Year
Ident
No.
Day.Month
Tref25
Tref20
LoBat
Lin
nLF
Sal
pH
C
TEC
Auto
Store
Time
Baud
cm
1/
/pH
mV
m
%S/cm
C
S E R
A L
RCL
Clearing the memory
With this function, you can delete the stored data records. 500 storage
locations will then be available again.
Note
The Clear memory function only appears when there are data records
stored in the memory. Otherwise, the measuring instrument automati-
cally switches to the measuring mode.
Proceed as follows to clear all data records:
Note
The calibration data remain stored and can be called up.
1
Switch off the measuring instrument.
2
Press the key and hold it down.
3
Press the key.
The display test appears briefly on the display.
Subsequently, Sto CLr appears.
4
Confirm the clearing process with .
Pressing any other key prevents the clearing, the data records
will remain stored.
REL1
RCL
d
Sal
mg/l
ARng
O
TP
Oxi
Cal
AR
°C
S
Year
Ident
No.
Day.Month
Tref25
Tref20
LoBat
Lin
nLF
Sal
pH
C
TEC
Auto
Time
Baud
1/cm
/K
%
REL2
/pH
mV
m
%S/cm
S
C L R
T o
Store
You have the following possibilities of transmitting data:
� One of the following options:
– With the AutoStore function (page 30), measured values are
periodically saved internally (save interval Int 1) and output on the
interface.
– With the Data transmission interval function (Int 2), measured
values are periodically output on the interface (see below).
� With the Output data storage function (page 32), calibration data or
saved measured values are output on the interface.
� Via the analog recorder output (page 39), measured values are out-
put as voltage values.
� With the KOM pilot communication kit (accessory), data can be
transmitted bidirectionally (page 40).
Note
If you connect a recorder (analog output), the output on the digital in-
terface is switched off.
4.5.1
Data transmission interval (Int 2)
The interval for the data transmission (Int 2) determines the chronolog-
ical interval between automatic data transmissions. After the selected
interval expires, the current data record is transmitted to the interface.
Note
When the AutoStore function is active, the data transmission is per-
formed according to the setting of the save interval (Int 1). Set the save
interval (Int 1) to OFF to activate the Data transmission interval (Int 2).
The default setting for the interval is OFF.
To start the data transmission, set an interval (5 s, 10 s, 30 s, 1 min, 5
min, 10 min, 15 min, 30 min, 60 min):
1
Press the key and hold it down.
2
Press the key. Int 2 appears on the display.
3
Set the required interval between the saving procedures with
.
4
Confirm with .
The measuring instrument automatically switches to the mea-
suring mode.
REL1
Store
RCL
d
Sal
mg/l
ARng
O
TP
Oxi
Cal
AR
°C
S
Year
Ident
No.
Day.Month
Tref25
Tref20
LoBat
Lin
nLF
Sal
pH
TEC
Auto
Time
Baud
1/cm
/K
%
REL2
/pH
mV
m
%S/cm
n
O F F
T 2
I
Recorder (analog output)
You can transmit data to a recorder via the analog output. Connect the
analog output to the recorder via the AK323 interface cable.
The data output automatically switches to Recorder output.
Socket assignment
Note
The analog output is activated automatically in the cable by connecting
2 and 3.
Signal range
The signal range of the analog output depends on the measured vari-
able and the measuring range:
Concentration
Saturation
1 free
2 Plug coding
3 Ground
4 Analog output
1
2
3
4
RS 232
REC
(internal resistance < 5 Ohm)
Measuring range
Voltage
Resolution
0 ... 19.99 mg/l
0 ... 1999 mV
0.01 mg/l per 1 mV
0 ... 90.0 mg/l
0 ... 900 mV
0.1 mg/l per 1 mV
Measuring range
Voltage
Resolution
0 ... 199.9
0 ... 1999 mV
0.1 % per 1 mV
0 ... 600 %
0 ... 600 mV
1 % per 1 mV
PC/external printer (RS232 interface)
Via the RS 232 interface, you can transmit the data to a PC or an ex-
ternal printer.
Use the AK340/B (PC) or AK325/S (ext. printer) cable to connect the
interface to the devices.
The data output automatically switches to the RS232 interface.
Warning
The RS232 interface is not galvanically isolated.
When connecting an earthed PC/printer, measurements cannot be per-
formed in earthed media as incorrect values would result.
Set up the following transmission data at the PC/printer:
Socket assignment
4.5.4
Remote control
The measuring instrument can be remotely controlled from a PC. This
requires the KOM pilot communication kit. It is available as an acces-
sory.
The instrument is then controlled via commands that simulate key-
strokes and request the current display contents.
Note
A more detailed description is provided within the scope of delivery of
the communication kit.
Baud rate
selectable between: 1200, 2400, 4800, 9600
Handshake
RTS/CTS + Xon/Xoff
Parity
none
Data bits
8
Stop bits
1
1 CTS
2 RxD
3 Ground
4 TxD
1
2
3
4
RS 232
REC
You can adapt the measuring instrument to your individual require-
ments. To do this, the following parameters can be changed (the status
on delivery is marked in bold):
Note
You can leave the configuration menu at any time with . The pa-
rameters that have already been changed are stored.
Baud rate
Baud rate
1200, 2400, 4800, 9600
Air pressure display
Current value in mbar
(no input possible)
Calibration interval (Int 3)
1 ... 14 ... 999 d
AutoRange ARng
On or off
Date/time
Any
1
Switch off the measuring instrument.
2
Press the key and hold it down.
3
Press the key.
The display test appears briefly on the display. The measuring
instrument then switches automatically to the setting of the
baud rate.
4
Select the required baud rate with .
5
Confirm with . On the display, the current air
pressure in mbar appears.
REL1
Store
RCL
Sal
mg/l
ARng
O
TP
Oxi
Cal
AR
°C
S
Year
Ident
No.
Day.Month
Tref25
Tref20
LoBat
Lin
nLF
Sal
pH
TEC
Auto
Time
1/cm
/K
%
REL2
/pH
mV
m
S/cm
%
8 0 0
4
Baud
selection of the
measurement range
6
Confirm with . Int 3 appears on the display.
REL1
RCL
Cal
TEC
d Auto
C A
pH
pH
/
Sal
mg/l
ARng
O
TP
Oxi
AR
°C
S
Year
Ident
No.
Day.Month
Tref25
Tref20
LoBat
Lin
nLF
SalC
Store
Time
Baud
1/cm
/K
%
REL2
mV
m
S/cm
%
9 4 9
P
7
Set the required interval in days (d) with .
8
Confirm with . ARng appears on the display.
AR
cm
1/
4
Tref25
TP
ARng
nLF
°C
cm
S/
Sal
RCL
%
S
%
m
mg/l
Oxi
Year
Ident
No.
Day.Month
Tref20
LoBat
Lin
Sal
pH
TEC
Auto
Store
Baud
K
/
REL2
/pH
mV
1 4
3
n t
REL1
d
I
O
Time
Cal
9
Using , switch between no and YES.
YES: Switch on AutoRange.
no: Switch off AutoRange.
10
Confirm with .
The date (day) flashes in the display.
I
Time
d
Cal
TEC
pH
REL1
AR
cm
1/
Tref25
TP
nLF
°C
cm
S/
Sal
RCL
%
S
%
m
mg/l
Oxi
Year
Ident
No.
Day.Month
Tref20
LoBat
Lin
Sal
Auto
Store
Baud
K
/
REL2
/pH
mV
S
n
A R
E
y
ARng
O
Date and time
11
Set the date of the current day with .
12
Confirm with .
The date (month) flashes in the display.
13
Set the current month with .
14
Confirm with .
The year appears on the display.
15
Set the current year with .
16
Confirm with .
The hours flash on the display.
17
Set the current time with .
18
Confirm with .
The minutes flash on the display.
19
Set the current time with .
20
Confirm with .
The measuring instrument automatically switches to the mea-
suring mode.
REL1
Tref25
5
t 2
Time
Cal
I
O
AR
cm
1/
TP
ARng
nLF
°C
cm
S/
Sal
RCL
%
S
%
m
mg/l
Oxi
Year
Ident
No.
Tref20
LoBat
Lin
Sal
pH
TEC
Auto
Store
Baud
K
/
REL2
/pH
mV
Day.Month
0 9 0 4
You can reset (initialize) the measurement parameters and the config-
uration parameters separately from one another.
Measurement
parameters
The following measured parameters (O2 InI) are reset to the default
condition:
Note
The calibration data gets lost when the measuring parameters are re-
set. Recalibrate after performing a reset.
Configuration
parameters
The following configuration parameters (InI) are reset to the delivery
status:
Measuring mode
D. O. concentration
AutoRange automatic measure-
ment range selection
On (YES)
Relative slope
1.00
Correction factor of the relative
slope
1.000
Salinity equivalent
0.0
Salinity correction
Off
Baud rate
4800
Interval 1
(automatic saving)
OFF
Interval 2
(for data transmission)
OFF
Resetting the configura-
tion parameters
1
Press the key and hold it down.
2
Press the key.
3
Using , switch between no and YES.
YES: Resetting the measuring parameters
no: Retaining settings.
4
Confirm with .
The measuring instrument switches to the configuration pa-
rameters.
REL1
pH
Day.Month
0 9
Tref25
Time
Cal
I
AR
cm
1/
TP
ARng
nLF
°C
cm
S/
Sal
RCL
%
S
%
m
mg/l
Oxi
Year
Ident
No.
Tref20
LoBat
Lin
Sal
TEC
Auto
Store
Baud
K
/
REL2
/pH
mV
n o
i
in
O
5
Using , switch between no and YES.
YES: Resetting the configuration parameters
no: Retaining settings.
6
Confirm with .
The measuring instrument automatically switches to the mea-
suring mode.
REL1
O
pH
Day.Month
0 9
Tref25
Time
Cal
I
AR
cm
1/
TP
ARng
nLF
°C
cm
S/
Sal
RCL
%
S
%
m
mg/l
Oxi
Year
Ident
No.
Tref20
LoBat
Lin
Sal
TEC
Auto
Store
Baud
K
/
REL2
/pH
mV
n o
i
in
Maintenance, cleaning, disposal
47
5
Maintenance, cleaning, disposal
5.1
Maintenance
The measuring instrument is maintenance-free.
5.2
Cleaning
Occasionally wipe the outside of the measuring instrument with a
damp, lint-free cloth. Disinfect the housing with isopropanol as re-
quired.
Warning
The housing components are made out of synthetic materials
(polyurethane, ABS and PMMA). Thus, avoid contact with acetone and
similar detergents that contain solvents. Remove any splashes imme-
diately.
5.3
Disposal
Packing
This measuring instrument is sent out in a protective transport packing.
We recommend: Keep the packing material. The original packing pro-
tects the measuring instrument from transport damages.
Rechargeable battery
Remove the rechargeable battery from the instrument and dispose of it
at a suitable facility according to local legal requirements. It is illegal to
dispose of the rechargeable battery with household refuse. Proceed as
follows to disassemble the rechargeable battery:
Ni-Cd
Cd
1
Remove the carrying and positioning handle or the carrying
strap.
2
Unscrew the fixing elements (1) using a hexagon key.
3
Remove the instrument from the enclosure by vigorously
pressing against the socket field.
4
Take out the rechargeable battery (2) and cut off the battery ca-
ble.
1
1
2
Maintenance, cleaning, disposal
Oxi 197i
48
Measuring instrument,
Dispose of the measuring instrument without the rechargeable battery
as electronic waste at an appropriate collection point.
What to do if...
49
6
What to do if...
Display LoBat
Instrument does not
react to keystroke
Error message 0FL
0FL
0FL
0FL
Error message E3
E3
E3
E3
Cause
Remedy
– Battery almost empty
– Charge the battery (see
section 3.2)
Cause
Remedy
– Operating condition undefined
or EMC load unallowed
– Processor reset:
Press the and
keys at the same
time and release them again.
The software version is
displayed.
Cause
Remedy
Display range exceeded
Oxygen sensor:
– Not connected
– Connect the sensor
– Cable broken
– Replace sensor
– Depleted
– Replace sensor
– Short-circuit between gold and
lead electrode
– Clean sensor and replace it if
necessary
Cause
Remedy
Invalid calibration
Oxygen sensor:
– Electrolyte solution depleted
– Regenerate sensor
– Membrane contaminated
– Clean membrane
– Electrode system poisoned
– Regenerate sensor
– Obsolete
– Replace sensor
– Broken
– Replace sensor
What to do if...
Oxi 197i
50
Error message E7
E7
E7
E7
AR flashes
continuously
Measured value too low
Measured value too
high
Display to
to
to
to
Sensor symbol flashes
Cause
Remedy
Membrane damaged
– Membrane damaged
– Regenerate sensor
– Membrane head not screwed
on tight enough
– Screw membrane head tight
Cause
Remedy
No stable measured value
– Membrane contaminated
– Clean membrane
Cause
Remedy
– Insufficient
flow
– Provide flow to the sensor
Cause
Remedy
– High amount of dissolved
substances
– Correct solubility function
using the salinity equivalent
– Air bubbles bump on the
membrane with high velocity
– Avoid direct flow to the
membrane
– The carbon dioxide pressure is
too high (> 1 bar)
– Measuring not possible
Cause
Remedy
– Time-out of the interface
– Check the instrument
connected
Cause
Remedy
– Calibration interval expired
– Recalibrate the measuring
system
What to do if...
51
Message StoFull
StoFull
StoFull
StoFull
You want to know which
software version is in
the instrument
Cause
Remedy
– All memory locations are full
– Output data storage and clear
data storage
Cause
Remedy
– E. g., a question by the WTW
service department
– Simultaneously press the
and keys
and release them again. The
software version is displayed.
approx. 90 x 200 x 190 mm
Weight
approx. 1.5 kg (without plug-in power supply)
Mechanical structure
Type of protection:
IP 67
Electrical safety
Protective class:
III
Ambient
conditions
Power supply
Operation
-10 °C ... + 55 °C
Storage
- 25 °C ... + 65 °C
Climatic class
2
Rechargeable battery
Nickel-cadmium (NiCad)
rechargeable battery
Operational life
approx. 600 hours with one charging
Plug-in power supply
unit
(charging device)
The following applies to all plug-in power
supplies:
Connection max.
Overvoltage category II
Euro plug
FRIWO FW1199, 11.7864
Friwo Part. No. 1762613
Input: 230 V ~ / 50 Hz / 5.6 VA
Output: 12 V = / 130 mA / 1.56 VA
US plug
FRIWO FW1199, 11.7880
Friwo Part. No. 1794043
Input: 120 V ~ / 60 Hz / 6 VA
Output: 12 V = / 150 mA
UK plug
FRIWO FW1199, 11.7872
Friwo Part No. 1816491
Input: 230V ~ / 50 Hz / 5.6 VA
Output: 12 V = / 130 mA / 1.56 VA
Automatic switchover when a PC or a printer is connected via the cable,
AK 340/B or AK 325/S.
Analog output
Automatic switchover when the recorder is connected by the cable,
AK 323.
Type
RS232, data output
Baud rate
Can be set to
1200, 2400, 4800, 9600 Baud
Data bits
8
Stop bits
2
Parity
None
Handshake
RTS/CTS + Xon/Xoff
Cable length
Max. 15m
Variable
Voltage
Resolution
Saturation [%] / Range
0 ... 200.0 %
0 ... 600 %
0 ... 2 V
0 ... 600 mV
± 0.1
± 1
Concentration [mg/l] / Range
0...20.00 mg/l
0...100.0 mg/l
0 ... 2 V
0 ... 1 V
0.01 mg/l
0.1 mg/l
Accuracy
± 0.5 % of display value
± 0.1(% saturation)
or
± 0.5 % of the display value ± 0.01 mg/l
Internal resistance
< 5 Ohm (current limited to max. 0.2 mA out-
put current)
and norms used
7.2
Measuring ranges, resolutions, accuracies
Measuring ranges,
resolutions
Note: The values given in brackets apply to the sensor DurOx 325 es-
pecially.
Accuracy
(± 1 digit)
Correction functions
EMC
E.C. guideline 89/336/EEC
EN 61326-1:1997
EN 61000-3-2 A14:2000
EN 61000-3-3:1995
FCC Class A
Instrument safety
E.C. guideline 73/23/EEC
EN 61010-1 A2:1995
Climatic class
VDI/VDE 3540
IP protection
EN 60529:1991
Variable
Measuring range
Resolution
Concentration [mg/l]
0 ... 19.99 (0 ... 19.9)
0 ... 90.0 (0 ... 90)
0.01 (0.1)
0.1 (1)
Saturation [%]
0 ... 199.9 (0 ... 199)
0 ... 600
0.1 (1)
1
T [°C]
0 ... 50.0
0.1
Variable
Accuracy
Concentration [mg/l]
± 0.5 % of measured value
at ambient temperature of + 5 °C ... + 30 °C
Saturation [%]
± 0.5 % of measured value
when measuring in the range of ± 10 K
around the calibration temperature
T [°C]
± 0.1
Temperature com-
pensation
Accuracy better than 2 % at 0 ... + 40 °C
Salinity correction
0 ... 70.0 SAL
Air pressure correc-
tion
Automatic through integrated
pressure sensor in the range of 500 ... 1100
mbar
This chapter provides additional information and orientation aids.
Abbreviations
The list of abbreviations explains the indicators and the abbreviations
that appear on the display and in the manual.
Specialist terms
The glossary briefly explains the meaning of the specialist terms. How-
ever, terms that should already be familiar to the target group are not
described here.
Index
The index helps you to find the topics that you are looking for.
AutoRead (drift control)
ARng
Automatic range switching
Measuring instrument measures with highest reso-
lution
°C
Temperature unit, degrees Celsius
Cal
Calibration
E3
Error message
see chapter 6 WHAT TO DO IF...
InI
Initialization
Resets individual basic functions to the status they
had on delivery
LoBat
Battery almost empty (Low Battery)
OFL
Display range exceeded (Overflow)
OxiCal
Automatic calibration for D. O. measurements
SAL
Salinity
SELV
Safety Extra Low Voltage
TP
Temperature measurement active (Temperature
Probe)
To manipulate a measuring system so that the relevant value (e. g. the
displayed value) differs as little as possible from the correct value or
a value that is regarded as correct, or that the difference remains with-
in the tolerance.
Amperometry
Name of a measuring technique. The signal (depending on the mea-
sured parameter) of the sensor is the electric current. The electrical
voltage remains constant.
AutoRange
Name of the automatic selection of the measuring range.
AutoRead
WTW name for a function to check the stability of the measured value.
Calibration
Comparing the value from a measuring system (e. g. the displayed
value) to the correct value or a value that is regarded as correct. Of-
ten, this expression is also used when the measuring system is adjust-
ed at the same time (see adjusting).
D. O. partial pressure
Pressure caused by the oxygen in a gas mixture or liquid.
Measured parameter
The measured parameter is the physical dimension determined by
measuring, e. g. pH, conductivity or D. O. concentration.
Measured value
The measured value is the special value of a measured parameter to
be determined. It is given as a combination of the numerical value and
unit (e. g. 3 m; 0.5 s; 5.2 A; 373.15 K).
Measuring system
The measuring system comprises all the devices used for measuring,
e. g. measuring instrument and sensor. In addition, there is the cable
and possibly an amplifier, terminal strip and armature.
Molality
Molality is the quantity (in Mol) of a dissolved substance in 1000 g sol-
vent.
OxiCal®
WTW name for a procedure to calibrate D. O. measuring systems in
water vapor saturated air.
Oxygen saturation
Short name for the relative D. O. saturation.
Note: The D. O. saturation value of air-saturated water and the D. O.
saturation value of oxygen-saturated water are different.
Reset
Restoring the original condition of all settings of a measuring system.
Resolution
Smallest difference between two measured values that can be dis-
played by a measuring instrument.
The absolute salinity SA of seawater corresponds to the relationship
of the mass of dissolved salts to the mass of the solution (in g/Kg). In
practice, this dimension cannot be measured directly. Therefore, the
practical salinity is used for oceanographic monitoring. It is deter-
mined by measuring the electrical conductivity.
Salt content
General designation for the quantity of salt dissolved in water.
Sample
Designation of the sample ready to be measured. Normally, a test
sample is made by processing the original sample. The test sample
and original sample are identical if the test sample was not processed.
Slope
The slope of a linear calibration function.
Slope (relative)
Designation used by WTW in the D. O. measuring technique. It ex-
presses the relationship of the slope value to the value of a theoretical
reference sensor of the same type of construction.
Standard solution
The standard solution is a solution where the measured value is
known by definition. It is used to calibrate a measuring system.
Analog output 39
Authorized use 9
AutoRange 18
AutoRead 19
Criteria 19
B
Batteries
Charging 11
Battery
Charging time 11
C
Calibration
start 23
Calibration record 22
Connecting a printer 40
D
Data record 28
Data transmission (interval)
38
Default condition 44
Display 6
Displaying calibration data
21
Drift control 19
E
Electrode evaluation 22
Energy saving feature 15
Error messages 49
I
Initial commissioning 12, 14
Initialization 44
Interval
data transmission 37
save 30
K
Keys 7
L
LoBat 49
M
Mains operation 11
Measuring range selection
18
Measuring the D. O. concen-
tration 17
Measuring the D. O. satura-
tion 18
O
Operating safety 10
P
Plug-in power supply 11
Printing
calibration protocol 22
R
Remote control 40
Reset 44
RS232 interface 40
S
Safety 9
safety precautions 9
Salt content correction
entering the salinity 27
switching on 17
Save interval 30
Scope of delivery 11
Setting the baud rate 41
Setting the date 12, 14, 43
Setting the time 12, 43
Slope 21
T
Temperature sensor 16
Wissenschaftlich-Technische Werkstätten GmbH
Dr.-Karl-Slevogt-Straße 1
D-82362 Weilheim
Germany
Tel:
+49 (0) 881 183-0
+49 (0) 881 183-100
Fax:
+49 (0) 881 183-420
E-Mail:
[email protected]
Internet:
http://www.WTW.com