RKI Instruments GX-6000 Operator's Manual
Text of the manual
Single Sensor PID Model
Operator’s Manual
Part Number: 71-0363
Revision: 0
Released: 8/10/17
GX-6000 Operator’s Manual
WARNING
Read and understand this instruction
manual before operating instrument.
Improper use of the gas monitor could
result in bodily harm or death.
Periodic calibration and maintenance of
the gas monitor is essential for proper
operation and correct readings. Please
calibrate and maintain this instrument
regularly! Frequency of calibration
depends upon the type of use you have
and the sensor types. Typical calibration
frequencies for most applications are
between 1 and 3 months, but can be
required more often or less often based
on your usage.
GX-6000 Operator’s Manual
Table of Contents
Table of Contents
Chapter 1: Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
About the GX-6000 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
About this Manual . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Chapter 2: Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Instrument Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Case . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
LCD . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Control Buttons . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Flashlight LED . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Alarm LEDs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Infrared Communications Port . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Buzzer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Vibrator . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Printed Circuit Boards (PCBs) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Pump . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Flow Chamber . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Sensors. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
Inlet Filter Holder . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
Batteries. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
Included Accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
Tapered Rubber Nozzle . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
Belt Clip. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
Rubber Boot. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
Wrist Strap . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
Sample Hose and Probe. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
Screen Protector . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
Other Accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
VOC Zero Filter. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
DIN Rail Mounting Assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
Table of Contents
GX-6000 Operator’s Manual
Chapter 3: Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
Start Up . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
Turning On the GX-6000 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
Performing a Demand Zero . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
Turning Off the GX-6000 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
Measuring Mode, Normal Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
Monitoring an Area . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
Using Optional Sample Hoses . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
Snap Log Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
Measuring Mode, Alarms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
Alarm Indications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
Responding to Alarms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
Display Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38
Tips for Using Display Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39
PID Gas Name Screen . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39
Peak Screen . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42
STEL Screen. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43
TWA Screen . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43
View Alarm Settings Screen . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44
Time in Operation Screen . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 46
Date/Time, Battery Voltage Screen . . . . . . . . . . . . . . . . . . . . . . . . . . 46
Log Time Remaining Screen . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 47
Clear Data Logging Screen . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 47
Pump Off Screen . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 49
Select User ID Screen . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 50
Select Station ID Screen. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 52
Viewing Snap Logger Data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 54
Peak Bar Screen. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 55
Gas Display Screen . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 56
LCD Flip Screen . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 56
LCD Background Color Flip Screen . . . . . . . . . . . . . . . . . . . . . . . . . 57
Language Screen . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 57
Data Logging . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 58
GX-6000 Operator’s Manual
Table of Contents
Chapter 4: Calibration Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 60
Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 60
Calibration Supplies and Equipment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 60
Entering Calibration Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 61
Calibrating Using the Auto Calibration Method . . . . . . . . . . . . . . . . . . . . . 61
Setting the Fresh Air Reading . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 61
Performing a Span Adjustment in Auto Calibration . . . . . . . . . . . . . 63
Returning to Measuring Mode. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 68
Calibrating Using the Single Calibration Method . . . . . . . . . . . . . . . . . . . . 69
Setting the Fresh Air Reading . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 69
Performing a Span Adjustment in Single Calibration . . . . . . . . . . . 70
Performing a Bump Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 74
Chapter 5: User Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 80
Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 80
Tips for Using User Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 80
Entering User Mode. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 81
Setting the Date and Time . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 82
Setting the Date Format . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 83
Configuring the Channels . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 83
Updating the Language Setting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 83
Viewing the ROM/SUM of the Instrument . . . . . . . . . . . . . . . . . . . . . . . . . 84
Turning the Password Function On or Off. . . . . . . . . . . . . . . . . . . . . . . . . . 85
Exiting User Mode. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 86
Chapter 6: Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 87
Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 87
Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 87
Replacing or Recharging the Batteries . . . . . . . . . . . . . . . . . . . . . . . . . . . . 89
Replacing the Alkaline Batteries . . . . . . . . . . . . . . . . . . . . . . . . . . . . 89
Replacing the Lithium Ion Battery Pack . . . . . . . . . . . . . . . . . . . . . . 91
Recharging the Lithium Ion Battery Pack . . . . . . . . . . . . . . . . . . . . . 93
Recharging the Lithium Ion Battery Pack Out of the Instrument . . . 95
Replacing the Probe’s Particle Filter and Hydrophobic Filter Disk . . . . . . 98
Replacing the PID Sensor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 100
Replacing the Hydrophobic Filter and Wire Mesh Disk . . . . . . . . . . . . . . 102
Table of Contents
GX-6000 Operator’s Manual
PID Sensor Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 103
Cleaning the PID Sensor’s Lamp . . . . . . . . . . . . . . . . . . . . . . . . . . 103
Replacing the PID Sensor’s Lamp. . . . . . . . . . . . . . . . . . . . . . . . . . 107
Replacing the PID Sensor’s Electrode Stack. . . . . . . . . . . . . . . . . . 110
General Parts List . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 113
Appendix A: Maintenance Mode . . . . . . . . . . . . . . . . . . . . . . . . . . 116
Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 116
Tips for Using Maintenance Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 119
Using Maintenance Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 120
Setting the Date and Time . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 123
Setting the Date Format . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 124
Performing a Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 124
Performing a Bump Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 124
Updating Calibration Settings. . . . . . . . . . . . . . . . . . . . . . . . . . . . . 125
Updating Bump Test Settings. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 127
Updating Alarm Parameters . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 132
Turning the Pump On/Off Display On or Off . . . . . . . . . . . . . . . . . 134
Turning the User/Station ID Function On or Off . . . . . . . . . . . . . . 135
Updating the Backlight Time Setting . . . . . . . . . . . . . . . . . . . . . . . . 135
Updating the Language Setting . . . . . . . . . . . . . . . . . . . . . . . . . . . . 135
Updating the Lunch Break Setting. . . . . . . . . . . . . . . . . . . . . . . . . . 136
Updating the Auto Zero Setting . . . . . . . . . . . . . . . . . . . . . . . . . . . . 136
Updating the Demand Zero Setting . . . . . . . . . . . . . . . . . . . . . . . . . 137
Zero Follower Setting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 137
Zero Suppression Settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 137
Updating the Autoranging Setting. . . . . . . . . . . . . . . . . . . . . . . . . . 137
Updating the Confirmation Alert Setting. . . . . . . . . . . . . . . . . . . . . 137
Turning the Key Tone Noise On or Off . . . . . . . . . . . . . . . . . . . . . . 138
Turning Inert Mode On or Off . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 138
Updating the Leak Check/Bar Hole Mode Setting. . . . . . . . . . . . . . 138
Setting the Bar Hole Measurement Time . . . . . . . . . . . . . . . . . . . . . 138
Updating the CO Display Setting . . . . . . . . . . . . . . . . . . . . . . . . . . 139
Updating the Man Down Settings . . . . . . . . . . . . . . . . . . . . . . . . . . 139
Updating the Datalogging Parameters . . . . . . . . . . . . . . . . . . . . . . 141
Adjusting the Low Flow Setpoint. . . . . . . . . . . . . . . . . . . . . . . . . . . 143
Viewing the ROM/SUM of the Instrument . . . . . . . . . . . . . . . . . . . . 144
Turning the Password Function On or Off . . . . . . . . . . . . . . . . . . . 144
GX-6000 Operator’s Manual
Table of Contents
Restoring the Default Settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 145
Exiting Maintenance Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 147
Appendix B: 10.0 eV/Benzene PID Sensor . . . . . . . . . . . . . . . . . . . 148
Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 148
Sensor Description. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 148
Tube and Tube Holder . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 149
Tube . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 149
Tube Holder . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 149
Start Up . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 150
Display Mode. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 150
Measuring Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 151
Benzene Select Mode. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 151
Viewing Benzene Select Mode Data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 157
Calibrating the 10.0 eV/Benzene PID Sensor . . . . . . . . . . . . . . . . . . . . . . 160
Important Calibration Information . . . . . . . . . . . . . . . . . . . . . . . . . 160
Performing an AUTO CAL . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 161
Performing a SINGLE CAL. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 161
Performing a TUBE CAL. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 161
Entering a CAL CODE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 168
Maintenance. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 170
PID Sensor Maintenance. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 170
Replacing a Tube . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 170
Parts List . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 172
WARNING: Understand manual before operating. Substitution of
components may impair intrinsic safety. To prevent ignition
of a hazardous atmosphere, batteries must only be changed
or charged in an area known to be nonhazardous. Not tested
in oxygen enriched atmospheres (above 21%).
NOTE: RKI Instruments, Inc. recommends that you refer to ISA-RP12.13,
Part II-1987 or an equivalent international recommended practice
for guidance in the use of combustible gas detection instruments.
1 • Overview
GX-6000 Operator’s Manual
Chapter 1: Introduction
Overview
This chapter briefly describes the GX-6000 Single Sensor PID Model gas
monitor. This chapter also describes the GX-6000 Single Sensor PID Model
Operator’s Manual (this document). Table 1 at the end of this chapter lists the
specifications for the GX-6000 Single Sensor PID Model (hereafter referred
to as the GX-6000).
About the GX-6000
The GX-6000’s rugged, reliable, and easy-to-use design makes it ideally
suited for a wide range of applications, including sewage treatment plants,
utility manholes, tunnels, hazardous waste sites, power stations,
petrochemical refineries, mines, paper mills, drilling rigs, and fire fighting
stations. The GX-6000 offers a full range of features including:
•
VOC monitoring
•
Sample-drawing pump with up to 50 foot range
•
Liquid crystal display (LCD) for complete and understandable
information at a glance
•
Ultrabright alarm LEDs
•
Distinctive audible/vibrating alarms for dangerous gas conditions and
audible alarms for unit malfunction
•
Microprocessor control for reliability, ease of use, and advanced
capabilities
•
Data logging functions
•
Alarm trend data
•
STEL and TWA (for high range PID only) and over range alarms
•
Peak reading
•
Built-in time function
•
Lunch break feature
•
RF shielded high impact plastic case
•
CSA classification for Class I, Division I, Groups A, B, C, and D
hazardous atmospheres
GX-6000 Operator’s Manual
About the GX-6000 • 2
WARNING: The Model GX-6000 detects elevated levels of VOC gases
which can be dangerous or life threatening. When using the
GX-6000, you must follow the instructions and warnings in
this manual to assure proper and safe operation of the unit
and to minimize the risk of personal injury. Be sure to
maintain and periodically calibrate the GX-6000 as described
in this manual.
3 • Specifications
GX-6000 Operator’s Manual
Specifications
Table 1: Sensor Specifications/Alarm Points
Table 2: GX-6000 Specifications
VOCs, Isobutylene Calibration Standard
Low Range PID
High Range PID
Detection Range
0 - 50000 ppb
0 - 6000 ppm
Reading Increment
• 0 - 5000: 1 ppb
• 5000 - 50000: 10 ppb
• 0 - 600 ppm: 0.1 ppm
• 600 - 6000 ppm: 1 ppm
Alarm 1
Factory Setting
5000 ppb
400 ppm
Alarm 2
Factory Setting
10000 ppb
1000 ppm
STEL Alarm
n/a
60.0 ppm
TWA Alarm
n/a
42.0 ppm
Sampling
Method
Sample Draw
Response
Time
T90 Within 30 Seconds
Display
Graphics LCD Display
Operating
Temperature
& Humidity
-20°C to 50°C/Below 95% RH (Without Condensation)
Safety/
Regulatory
• ATEX: II 1 G Ex ia IIC T4 Ga
Certificate Number Presafe 15ATEX6171
• IECEx: Ex ia IIC T4 Ga
Certificate Number PRE 15.0011
• CSA classified as Intrinsically Safe. Exia. Class I, Groups A, B, C, & D.
Temperature Code T3A.
186718
Power
Supply
• Three AA size alkaline batteries
OR
• One lithium ion battery pack
GX-6000 Operator’s Manual
Specifications • 4
Continuous
Operating
Hours
@ 25 °C
• Alkaline Batteries: 20 Hours (Non Alarm Operation, Fully Charged)
• Li-ion Battery Pack: 25 Hours (Non Alarm Operation, Fully Charged)
Case
High-impact Plastic, RF Shielded, Dust and Weather Proof
Included
Accessories
• Belt clip
• Red rubber boot
• Wrist strap
• Tapered nozzle
• Screen protector
• 3 foot hose and probe
• VOC zero filter (included for instruments with a low range PID sensor)
Other
Accessories
•
Alkaline Battery Pack
•
Rechargeable Li-ion Battery Pack
•
115 VAC Charger
•
12 VDC Charger
•
Hose/Probes of Various Lengths, See “General Parts List” on page 113.
•
VOC Zero Filter (optional for instruments with a high range PID sensor)
•
DIN Rail Mounting Assembly (for mounting chargers to the wall)
•
Product CD, includes Data Logger Management Program
•
IrDA/USB Cable for connecting to a computer when using the Data Logger
Management Program (not needed if computer has an infrared port)
Dimensions
and Weight
Approximately 200(H) x 68(W) x 52(D) mm (7.9”H x 2.7”W x 2.0”D)
Approximately 400 g (14 oz.)
5 • About this Manual
GX-6000 Operator’s Manual
About this Manual
Although the GX-6000 can support up to 6 sensors, this manual specifically
describes an instrument with only a PID sensor installed.
The GX-6000 Single Sensor PID Model Operator’s Manual uses the
following conventions for notes, cautions, and warnings.
NOTE: Describes additional or critical information.
CAUTION: Describes potential damage to equipment.
WARNING: Describes potential danger that can result in injury or death.
The GX-6000 Single Sensor PID Model Operator’s Manual is organized as
follows:
•
Chapter 1 is an introduction to the GX-6000.
•
Chapter 2 describes the components of the GX-6000.
•
Chapter 3 describes the operation of the GX-6000.
•
Chapter 4 describes Calibration Mode which allows you to calibrate the
GX-6000.
•
Chapter 5 describes the GX-6000’s User Mode.
•
Chapter 6 describes the GX-6000’s maintenance requirements and
procedures.
•
Appendix A describes Maintenance Mode which allows you to configure
different parameters of the GX-6000.
•
Appendix B describes the 10.0 eV/benzene PID sensor.
GX-6000 Operator’s Manual
Overview • 6
Chapter 2: Description
Overview
This chapter describes the GX-6000 instrument and accessories.
Instrument Description
The GX-6000 includes the case, LCD, control buttons, flashlight LED, alarm
LEDs, infrared communication port, buzzer, vibrator, printed circuit board,
pump, flow chamber, sensor, filters, inlet filter holder, and batteries.
Figure 1: Component Location
Case
The GX-6000’s sturdy, high-impact plastic case is radio frequency (RF)
resistant and is suitable for use in many environmental conditions, indoors
and out. The case is dust proof and water resistant. A clear plastic window is
located on the front of the case. The battery pack and flow chamber are
located on the back of the GX-6000. The inlet filter holder is located on the
top of the GX-6000 case.
Inlet Filter
Holder
Flow
Chamber
Battery
Pack
Flashlight LED
Alarm LEDs
RKI GX-6000
DISP
LOCK
(PANIC)
AIR
S HIFT
RESE T
Infrared
Communications
Port
Control Buttons
LCD
P OW ER/E NTER
7 • Instrument Description
GX-6000 Operator’s Manual
LCD
A digital LCD (liquid crystal display) is visible through a clear plastic
window in the top case. The LCD shows the gas reading. The LCD also
shows information for each of the GX-6000’s operating modes.
Control Buttons
Five control buttons are located below the LCD. They are, from left to right
and top to bottom, DISP/LOCK, ▲ AIR, SHIFT ▼ (PANIC), POWER/
ENTER, and RESET.
Flashlight LED
A white LED is visible through a round, raised, frosted lens in the top of the
case. This LED can be used to provide extra light, if necessary. The RESET
button turns this LED on and off.
Table 3: GX-6000 Control Button Functions
Button
Function(s)
DISP/LOCK
• activates Display Mode
• enters instructions into the GX-6000’s microprocessor
• allows you to exit a menu without saving changes
• locks screen orientation in any mode if INVERSION SELECT
is set to ON
▲ AIR
• activates the demand zero function (adjusts the GX-6000’s
fresh air reading)
• enters instructions into the GX-6000’s microprocessor
• moves the cursor on the LCD up the screen
• increases the value of a parameter available for adjustment
• scrolls through parameter options
SHIFT ▼ (PANIC)
• enters instructions into the GX-6000’s microprocessor
• moves the cursor on the LCD down the screen
• decreases the value of a parameter available for adjustment
• scrolls through parameter options
• initiates a Panic alarm in any mode if PANIC is set to ON
POWER/ENTER
• turns the GX-6000 on and off
• enters instructions, values, and settings into the GX-6000’s
microprocessor
RESET
• silences and resets audible alarm if ALARM LATCHING is set
to LATCHING and ALARM SILENCE is set to ON
• turns the flashlight LED on and off
GX-6000 Operator’s Manual
Instrument Description • 8
Alarm LEDs
Five sets of red alarm LEDs (light emitting diodes) border the LCD. The
alarm LEDs alert you to gas, low battery, and failure alarms.
Infrared Communications Port
An infrared (IR) communications port is located on the front of the case,
below the POWER/ENTER button. The data transmitted through the port is
in standard IrDA protocol. A computer’s infrared port or an IrDA/USB cable
connected to a USB port can be used to download data saved by the GX-6000
to a computer using the GX-6000 Data Logger Management Program. See
the GX-6000 Data Logger Management Program operator’s manual for data
logging and downloading instructions.
Buzzer
One solid-state electronic buzzer is located inside the case. Holes on the top
front of the case allow the sound to exit the case. The buzzer sounds for gas
alarms, malfunctions, low battery voltage, and as an indicator during use of
the GX-6000’s many display and adjustment options.
Vibrator
A vibrating motor inside the GX-6000 case vibrates for gas alarms, unit
malfunctions, and as an indicator during normal use of the various modes of
the GX-6000.
Printed Circuit Boards (PCBs)
The GX-6000 printed circuit boards analyze, record, control, store, and
display the information collected. The circuit boards are located inside the
case. They are not user serviceable.
Pump
A diaphragm pump inside the GX-6000 draws the sample to the sensor. It can
draw sample from as far as 50 feet from the GX-6000. The pump is not user
serviceable.
CAUTION: Sample hose lengths of more than 50 feet are not recommended
for the GX-6000 because of flow rate reduction.
Flow Chamber
The flow chamber is on the back of the GX-6000 and is held in place by four
Phillips screws. The flow chamber seals to the rubber sensor gasket which
seals to the sensor face inside the GX-6000 and routes flow from the pump to
the sensor and to the exhaust port (on the top of the GX-6000 case).
9 • Instrument Description
GX-6000 Operator’s Manual
Sensors
PID Sensor
Two types of PID sensors can be used with the GX-6000, a low range (higher
sensitivity) sensor and a high range (lower sensitivity) sensor (see Table 1 for
specifications).
The PID sensor is a cylindrical sensor with a diffusion opening on the front
and 3 pins on the back. It is installed in a white housing that has three sockets
on the bottom that mate with the GX-6000 instrument. The PID sensor must
always be installed in the first smart sensor position which is located in the
top left corner of the sensor block.
The standard calibration for a PID channel is to isobutylene. A PID channel
can be factory setup for and calibrated to other gases. Consult RKI
Instruments, Inc. for other available PID configurations and to specify the
desired PID configuration when a unit is ordered.
Dummy Sensors
This version of the GX-6000 has 5 dummy sensors installed in the unused
sensor positions. Dummy sensors are factory installed. The flat top of the
dummy sensor should face up and the bottom hollow side should face down.
Inlet Filter Holder
The filter holder is a clear plastic dome shaped piece on the top of the case. A
male quick connect fitting is located on the inlet filter holder. This is the
GX-6000’s inlet fitting. The filter holder may be removed by turning it
counterclockwise and pulling it away from the case. One flat membrane disk
hydrophobic filter, a wire mesh disk, and a rubber filter retaining gasket are
held in place by the filter holder and are located in the bottom of the case
chamber where the filter holder is installed.
Batteries
Three AA-size alkaline batteries or a rechargeable lithium ion battery pack
(4.1 VDC) power the GX-6000. Instrument run time is dependent upon
battery type. At 25°C the alkaline batteries last at least 20 hours and the
lithium ion battery pack lasts at least 25 hours. The battery icon in the upper
right of the LCD shows remaining battery life.
When the GX-6000 detects a low battery voltage, a low battery warning is
activated. When battery voltage is too low for Measuring Mode, the
GX-6000 sounds a dead battery alarm.
The alkaline batteries can be replaced by removing the battery cover on the
back of the case. The lithium ion pack can be replaced by removing the entire
battery pack. The battery pack release latch is located on the bottom of the
GX-6000 Operator’s Manual
Included Accessories • 10
instrument. When viewing the instrument from the bottom with the LCD
facing down, push the battery pack release latch toward the right to release
the pack.
The lithium ion battery pack can be recharged by placing the GX-6000 in its
battery charging station or by placing the battery pack in the charging station.
NOTE: Use of batteries or battery chargers not specified by RKI
Instruments, Inc. will compromise the CSA classification and may
void the warranty. See “Replacing or Recharging the Batteries” on
page 89.
WARNING: To prevent ignition of a hazardous atmosphere, batteries
must only be changed or charged in an area known to be
nonhazardous.
AVERTISSEMENT:Pour éviter l’inflammation d’une atmosphère
dangereuse, les batteries doivent uniquement être modifiés
ou facturés dans une zone connue comme non dangereuse.
Included Accessories
Included accessories consist of the tapered rubber nozzle, belt clip, rubber
boot, wrist strap, and the sample hose/probe.
Tapered Rubber Nozzle
A cone shaped 4 inch long rubber nozzle is included with the GX-6000 as
standard. It can be installed on the inlet fitting by pushing the larger end over
it. The smaller end can be inserted through a hole in a wall or some other
access to an enclosed area to sample the environment.
Belt Clip
A belt clip can be mounted to the back of the case using 3 Phillips head
screws. The belt clip allows the GX-6000 to be securely attached to a belt.
Rubber Boot
A red protective rubber boot can be installed over the GX-6000.
Wrist Strap
A wrist strap is included with the GX-6000 and can be attached to the right or
left wrist strap installation feature on the GX-6000 case.
11 • Included Accessories
GX-6000 Operator’s Manual
Sample Hose and Probe
A 3 foot sample hose with an attached probe is standard with the GX-6000.
When desired, the rubber nozzle may be removed and the sample hose and
probe may be connected to the inlet fitting. Sample hose lengths are available
from 3 feet to 50 feet (see “General Parts List” on page 113). The quick
connect end of the sample hose connects to the inlet fitting of the GX-6000.
The probe is integral with the hose and connects to it with a tube fitting.
CAUTION: Sample hose lengths of more than 50 feet are not recommended
for the GX-6000 because of flow rate reduction.
The probe includes a replaceable particle filter and hydrophobic filter disk
that prevent particulates and water from entering the GX-6000’s flow system.
See “Replacing the Probe’s Particle Filter and Hydrophobic Filter Disk” on
page 98 for instructions to replace the particle filter and hydrophobic filter
disk.
Figure 2: Sample Hose and Probe
GX-6000 Operator’s Manual
Included Accessories • 12
Screen Protector
The clear screen protector can be installed over the GX-6000’s LCD to
prevent it from getting scratched.
1 .
Remove the GX-6000’s rubber boot.
2 .
Orient the GX-6000 so that the LCD is as horizontal as possible.
3 .
Clean the LCD with rubbing alcohol.
4 .
Place 1 drop of water in the center of the LCD.
5 .
Grasp the tab at the top of the screen protector’s backing and pull it away
from the screen protector.
6 .
Use needle-nose pliers or tweezers to handle the screen protector.
7 .
Align the top of the screen protector with the top of the “RKI GX-6000”
logo.
8 .
Press down on the center of the screen protector. Water should spread out
along the entire surface of the screen protector between the screen
protector and the LCD. Some of the water may come out the edges.
9 .
If you are unhappy with the initial placement of the screen protector, lift
the screen protector up from a corner. Quickly rearrange the screen
protector and set it back down on the LCD.
10 . Use a small, stiff piece of plastic and, working your way out from center,
remove any air bubbles. The screen protector may shift during this
process so be sure to hold it securely with your other hand.
11 . Allow enough time for the water between the screen protector and the
LCD to dry before handling the GX-6000.
12 . Reinstall the rubber boot.
13 • Other Accessories
GX-6000 Operator’s Manual
Other Accessories
Several other accessories are available for the GX-6000. This section
describes the VOC zero filter and the DIN rail. Data logging accessories are
briefly described in “Data Logging” on page 58.
VOC Zero Filter
The VOC zero filter scrubs out low levels of VOC gases using charcoal. A
VOC zero filter is included as standard with GX-6000s that include a low
range PID sensor since this sensor is more susceptible to low levels of VOCs.
A VOC zero filter is not included with GX-6000s that include a high range
PID sensor. Use the VOC zero filter when performing an air adjust on a PID
sensor in an area that may have a low-level VOC background. The tubing
stub on the end of the filter should be used to connect the filter to the inlet
fitting or probe. Be sure the arrow on the filter is pointing toward the GX-
6000.
Figure 3: VOC Zero Filter
DIN Rail Mounting Assembly
Two different DIN rail mounting assemblies are available for the GX-6000’s
charger. Each assembly has two end clamps but one assembly is long enough
to accommodate 1 charger and the other assembly is long enough to
accommodate 4 chargers.
Figure 4: DIN Rail Dimensions
1 .
Use #10 screws to mount the DIN rail to the wall.
VOC Zero Filter
Tubing Stub
4" for 1-Charger Model
14" for 4-Charger Model
Use #10 Screws to Mount
1.38"
GX-6000 Operator’s Manual
Other Accessories • 14
2 .
Install the charger(s) on the DIN rail. Seat the top of the DIN rail in the
top slot on the back of the charger. Press the bottom of the charger
toward the DIN rail until it locks into place.
Figure 5: Installing a Charger on the DIN Rail
3 .
Slide the charger(s) along the DIN rail into the desired position.
4 .
Install the end clamps on the left and right side of the charger or bank of
chargers to prevent sliding.
Figure 6: Clamp Installation
DIN Rail
Charger
Clamp
DIN Rail
15 • Other Accessories
GX-6000 Operator’s Manual
Figure 7: 1-Charger Installation
Figure 8: 4-Charger Installation
5 .
To remove any charger(s) from the DIN rail:
a. Remove the clamps by pushing on the bottom of the clamp and
pulling on the top of the clamp.
b. Slide the charger(s) off the DIN rail.
Figure 9: Charger Removal
GX-6000 Operator’s Manual
Overview • 16
Chapter 3: Operation
Overview
This chapter explains how to use the GX-6000 to perform confined space
entry monitoring or general area monitoring in Normal Mode. There are three
operational modes in Normal Mode: Measuring Mode, Display Mode, and
Calibration Mode. While in Normal Mode, the unit is normally operating in
Measuring Mode. Display Mode and Calibration Mode are accessible from
Measuring Mode. Display Mode is described in this chapter. Calibration
Mode is described in “Chapter 4: Calibration Mode” on page 60.
Start Up
This section explains how to start up the GX-6000, get it ready for operation,
and turn it off.
Turning On the GX-6000
To illustrate certain functions, the following description of the GX-6000 start
up sequence assumes that the following menu items in Maintenance Mode
are turned on: LUNCH BREAK, CAL REMINDER, BUMP
REMINDER, and ID DISPLAY. If any of these items are turned off, then
the corresponding screens will not appear.
The GX-6000 may be used with a sample hose and probe or with the tapered
rubber nozzle. Determine which configuration works best for your
application.
NOTE: When the sample hose is not being used, its outgassing
characteristics may result in a small buildup of gas to which the PID
sensor will respond. If a sample hose has been sitting unused for a
period of time, when that sample hose is connected to a GX-6000,
the PID channel may temporarily show a reading. The reading will
return to a fresh air reading after all of the built up gas has been
drawn out of the sample hose.
1 .
Connect the tapered rubber nozzle or the sample hose and probe to the
GX-6000’s quick connect inlet fitting.
2 .
Press and briefly hold down the POWER/ENTER button. Release the
button when you hear a beep.
17 • Start Up
GX-6000 Operator’s Manual
3 .
If LUNCH BREAK is turned on (see “Updating the Lunch Break
Setting” on page 136), the Lunch Break Screen appears. The unit counts
down from 5 seconds at the top of the screen.
•
To continue accumulating peak and time-weighted average (TWA)
readings (high range PID only) from the last time the GX-6000 was
used, press and release the POWER/ENTER button before the
countdown reaches 0 or allow the countdown to reach 0. If you do not
press the POWER/ENTER button within the 5 second countdown,
the GX-6000 automatically resumes accumulating the peak and TWA
readings. The GX-6000 will also continue to keep track of operating
time including the operating time from the last time the GX-6000 was
used. See “Time in Operation Screen” on page 46 for more
information about how the GX-6000 tracks the operating time. The
short-term exposure limit (STEL) reading is reset each time the
GX-6000 is turned on.
•
To reset the accumulation of these measurements, press and release
the DISP/LOCK button before the countdown reaches 0.
4:07
LUNCH BREAK
5 SEC
YES:ENTER
NO:DISP
GX-6000 Operator’s Manual
Start Up • 18
4 .
If CAL REMINDER is set to ON, the screen that appears next depends
on how CAL EXPRD is set in Maintenance Mode (see “Updating the
Calibration Expired Setting” on page 126).
•
If the unit is due for calibration and CAL EXPRD is set to
CONFIRM TO USE, then the following screen displays and the
buzzer sounds in a double pulsing pattern.
To perform a calibration, press and release the POWER/ENTER
button. The GX-6000 will enter Calibration Mode and the LCD will
show the Calibration Mode main menu. See “Chapter 4: Calibration
Mode” on page 60 for instructions to calibrate the GX-6000. When
you are done with the calibration and exit Calibration Mode, the unit
will begin the startup sequence. If the calibration was successful, the
screen above will not appear again until the unit is due for
calibration. If the calibration was not successful, the screen above
will again appear in the startup sequence.
To continue without performing a calibration, press and release the
RESET button.
•
If the unit is due for calibration and CAL EXPRD is set to
CANNOT USE, then the following screen displays and the buzzer
sounds in a double pulsing pattern.
4:07
CAL DATE PAST
CONFIRM TO USE
CAL MODE : ENTER
NO : RESET
4:07
CAL DATE PAST
CANNOT USE
CAL MODE : ENTER
19 • Start Up
GX-6000 Operator’s Manual
The GX-6000 cannot be used until a successful calibration has been
performed. Press and release the POWER/ENTER button to enter
Calibration Mode. See “Chapter 4: Calibration Mode” on page 60 for
instructions to calibrate the GX-6000.
NOTE: In this situation, even if the User password function has been turned
on, no password is required to perform a calibration.
When you are done with the calibration and exit Calibration Mode,
the unit will begin the startup sequence. If the calibration was
successful, the screen above will not appear again until the unit is due
for calibration. If the calibration was not successful, the screen above
will again appear in the startup sequence.
•
If the unit is due for calibration and CAL EXPRD is set to NO
EFFECT, then the following alert screen displays and the buzzer
sounds in a double pulsing pattern.
If you want to enter Calibration Mode, press and release the
POWER/ENTER button. If you do not want to enter Calibration
Mode, wait for the instrument to continue with its startup sequence.
4:07
CAL DATE PAST
NO EFFECT
CAL MODE : ENTER
GX-6000 Operator’s Manual
Start Up • 20
5 .
If BUMP REMINDER is set to ON (factory setting is OFF), the screen
that appears next depends on how BUMP EXPRD is set in Maintenance
Mode (see “Updating the Bump Expired Setting” on page 130).
•
If the unit is due for bump testing and BUMP EXPRD is set to
CONFIRM TO USE, then the following screen displays and the
buzzer sounds in a double pulsing pattern.
To perform a bump test, press and release the POWER/ENTER
button. The GX-6000 will enter Calibration Mode and the LCD will
show the Calibration Mode main menu. See “Chapter 4: Calibration
Mode” on page 60 for instructions to bump test the GX-6000. When
you are done with the bump test and exit Calibration Mode, the unit
will begin the startup sequence. If the bump test was successful, the
screen above will not appear again until the unit is due for bump
testing. If the bump test was not successful, the screen above will
again appear in the startup sequence.
To continue without performing a bump test, press and release the
RESET button.
NOTE: If BUMP DISP is set to OFF, the bump test menu item will not
appear in Calibration Mode even though the instrument is
prompting you to perform a bump test. A bump test can always be
performed in Maintenance Mode, if necessary.
4:07
BUMP DATE PAST
CONFIRM TO USE
CAL MODE : ENTER
NO : RESET
21 • Start Up
GX-6000 Operator’s Manual
•
If the unit is due for bump testing and BUMP EXPRD is set to
CANNOT USE, then the following screen displays and the buzzer
sounds in a double pulsing pattern.
The GX-6000 cannot be used until a successful bump test has been
performed. Press and release the POWER/ENTER button to enter
Calibration Mode. See “Chapter 4: Calibration Mode” on page 60 for
instructions to bump test the GX-6000.
NOTE: In this situation, even if the User password function has been turned
on, no password is required to perform a bump test.
NOTE: If BUMP DISP is set to OFF, the bump test menu item will not
appear in Calibration Mode even though the instrument is
prompting you to perform a bump test. A bump test can always be
performed in Maintenance Mode, if necessary.
When you are done with the bump test and exit Calibration Mode,
the unit will begin the startup sequence. If the bump test was
successful, the screen above will not appear again until the unit is due
for bump testing. If the bump test was not successful, the screen
above will again appear in the startup sequence.
4:07
BUMP DATE PAST
CANNOT USE
CAL MODE : ENTER
GX-6000 Operator’s Manual
Start Up • 22
•
If the unit is due for bump testing and BUMP EXPRD is set to NO
EFFECT, then the following alert screen displays and the buzzer
sounds in a double pulsing pattern.
If you want to enter Calibration Mode, press and release the
POWER/ENTER button. If you do not want to enter Calibration
Mode, wait for the instrument to continue with its startup sequence.
NOTE: If CAL REMINDER or BUMP REMINDER or both are set to
OFF, a WARM-UP screen will display before the warm up
sequence continues.
6 .
The Date/Time Screen appears for a few seconds.
4:07
BUMP DATE PAST
NO EFFECT
CAL MODE : ENTER
4:07
DATE
4/ 1/15
DD/MM/YYYY
4:07
23 • Start Up
GX-6000 Operator’s Manual
7 .
The Battery Voltage Screen appears for a few seconds.
8 .
The Active Gases Screen appears for a few seconds indicating the active
channel and its target gas.
4:07
BATTERY
VOLTAGE
4.5V
BATTERY TYPE
ALKALINE
LATCHING
GAS
P
ppm
VOC
GAS
SMART 1
Isobutylene
10.6 eV
GX-6000 Operator’s Manual
Start Up • 24
9 .
The full scale value and the gas alarm setpoints are displayed by five
screens in sequence: the Full Scale Screen, the Low Alarm Screen, High
Alarm Screen, STEL Alarm Screen (high range PID only), and TWA
Alarm Screen (high range PID only). Each screen remains on the LCD
for three seconds.
10 . If ID DISPLAY is set to ON (see “Turning the User/Station ID Function
On or Off” on page 135), the User ID Screen appears for a few seconds,
followed by the Station ID Screen.
If ID DISPLAY is set to OFF, these screens will not appear.
F.S.
VOC P
ppm
6000
AL1
VOC P
ppm
400.0
AL2
VOC P
ppm
1000
STEL
VOC P
ppm
60.0
TWA
VOC P
ppm
42.0
4:07
USER ID
KIMBERLY
4:07
STATION ID
TANK 5
25 • Start Up
GX-6000 Operator’s Manual
11 . If the PID sensor fails during start up, the screen indicates the failure and
the buzzer sounds a double pulsing tone once per second.
The GX-6000 cannot be used if the PID sensor fails. Turn the instrument
off and replace the sensor.
12 . The GX-6000 is now monitoring for gas in Measuring Mode. The
Measuring Mode Screen appears displaying the current gas reading.
Performing a Demand Zero
Before using the GX-6000, it is recommended to set the fresh air reading by
performing a demand zero. This will set the PID channel to 0 ppm.
1 .
Find a fresh-air environment. This is an environment free of toxic or
combustible gases and of normal oxygen content (20.9%).
2 .
Turn on the unit as described above in “Turning On the GX-6000”.
3 .
If you suspect any low-level VOC background in the area, attach the
tubing stub of the VOC zero filter to the inlet fitting or probe and let the
instrument draw through it for 1 minute before continuing.
4 .
Press and hold the ▲ AIR button. The LCD prompts you to continue
holding the ▲ AIR button and the buzzer will pulse while you hold the
button.
SENSOR
VOC P
ppm
FAIL
4:07
VOC
ppm
0.0
GX-6000 Operator’s Manual
Measuring Mode, Normal Operation • 26
5 .
Continue to hold the ▲ AIR button until the LCD prompts you to release
it. The GX-6000 will set the fresh air reading. Start up is complete and
the unit is now ready for monitoring.
Turning Off the GX-6000
1 .
Press and hold the POWER/ENTER button.
2 .
TURN OFF will appear on the display and the buzzer will pulse for
about five seconds.
3 .
Release the button when TURN OFF disappears from the display.
Measuring Mode, Normal
Operation
When the GX-6000 completes its startup sequence, it is in Measuring Mode.
In Measuring Mode the GX-6000 continuously monitors the sampled
atmosphere and displays the gas concentration present. In a low-light
environment, press and release any button to turn on the display backlight.
See “Updating the Backlight Time Setting” on page 135 to program backlight
duration. If CONFIRMATION is set to ON in the Maintenance Mode menu
(see “Updating the Confirmation Alert Setting” on page 137), the GX-6000
beeps periodically to confirm that it’s operating.
Monitoring an Area
1 .
Start up the GX-6000 as described above in “Start Up” on page 16. It is
now in Measuring Mode.
2 .
Take the GX-6000 to the monitoring area.
4:07
VOC
ppm
0.0
27 • Measuring Mode, Normal Operation
GX-6000 Operator’s Manual
3 .
Put the probe tip in the area to be monitored.
NOTE: If the particle filter or hydrophobic filter become dirty or clogged,
replace them. If water enters the probe, dry out or replace the
particle filter (if installed) and shake any water out of the probe and
off of the hydrophobic filter. If you notice that water has entered the
flow system through the probe, replace the probe’s hydrophobic
filter. See “Replacing the Probe’s Particle Filter and Hydrophobic
Filter Disk” on page 98 for instructions to replace the particle filter
and the hydrophobic filter.
4 .
Wait 10 - 15 seconds and observe the display for gas reading. If a reading
is observed, allow the reading to stabilize to determine the gas
concentration present.
NOTE: Response time increases with the length of the sample hose. Long
sample hoses will require more time to show a response at the
GX-6000. The maximum sample hose length recommended for the
GX-6000 is 50 feet. Consult RKI Instruments, Inc. for longer
sample hose lengths.
5 .
Keep the following in mind:
•
The PID channel will indicate an upscale reading if one of a variety
of combustible gases is present. If %LEL concentrations of one of
these combustible gases is present, the PID channel may indicate an
overscale reading.
•
If concentrations of methane greater than 10% LEL are present in the
monitoring environment, the PID channel’s reading will be
suppressed.
•
The PID sensor will also respond to H2S and NH3. If H2S or NH3 is
present, the PID channel may indicate an upscale reading depending
on the concentration present.
6 .
If a gas alarm occurs, take appropriate action. See “Responding to
Alarms” on page 33.
GX-6000 Operator’s Manual
Measuring Mode, Normal Operation • 28
Using Optional Sample Hoses
The standard sample hose for the GX-6000 is 3 feet long. Optional samples
hoses and probes with longer hoses are available from 5 - 50 feet in 5 foot
increments (see “General Parts List” on page 113). If you are considering
using a hose and probe with a longer hose, keep in mind that a longer hose
will increase the GX-6000’s response time and the flowrate may decrease
close to the low flow alarm point.
CAUTION: Sample hose lengths of more than 50 feet are not recommended
for the GX-6000 because of flow rate reduction and increased
response time. Consult RKI Instruments, Inc. for hose lengths
longer than 50 feet.
The chart below illustrates how response time is affected by the sample hose
length.
Table 4: GX-6000 Response Time vs. Sample Hose Length
Hose Used
Typical Time to 90%
of Response (T90)
Probe & 3 Foot Hose (standard)
10 seconds
Probe & 10 Foot Hose
15 seconds
Probe & 30 Foot Hose
30 seconds
Probe & 50 Foot Hose
40 seconds
29 • Measuring Mode, Normal Operation
GX-6000 Operator’s Manual
Snap Log Mode
The snap logging function in Snap Log Mode allows the user to record data at
a specific time and have it saved to the data logger. The data is assigned a
snap log ID and is saved with the station ID that was in use when the data
was taken.
To enter Snap Log Mode and record snap log data:
1 .
While the unit is in Measuring Mode, press and hold the SHIFT ▼
(PANIC) button, then press and hold the ▲ AIR button and hold both
until you hear a beep. The unit will cycle through the following screens.
The first screen displays what snap log ID will be given to this particular
set of data and that you should press the POWER ENTER button to save
a set of snap log data. The snap log ID number increases sequentially
with each set of snap log data taken. The second screen displays what
Station ID will be associated with this snap log and that you should press
the DISP/LOCK button to exit Snap Log Mode without saving a set of
snap log data. The third screen displays the year, month, day, and time of
REC DATA
REC. DATA
NUMBER 001
YES : ENTER
REC DATA
REC. DATA
STATION ID
TANK 5
NO : DISP
REC DATA
REC. DATA
05/20/2015
MM/DD/YYYY
3:38
REC DATA
VOC
ppm
0.0
GX-6000 Operator’s Manual
Measuring Mode, Normal Operation • 30
the snap log. The fourth screen displays the current gas reading.
NOTE: If the GX-6000 detects an alarm condition while in Snap Log Mode,
it will automatically exit Snap Log Mode and return to Measuring
Mode. You may then reenter Snap Log Mode and take snap logs
while the instrument is still in alarm.
2 .
You can change the Station ID to be used with the snap log by pressing
the SHIFT ▼ (PANIC) button and then pressing the DISP/LOCK button.
The Station ID Select Screen will appear and the current Station ID will
be flashing.
3 .
Use the ▲ AIR and SHIFT ▼ (PANIC) buttons to scroll to the desired
Station ID, then press and release the POWER/ENTER button to return
to the Snap Logging Screen sequence.
To return to the Snap Logging Screen sequence without changing the
Station ID, press and release the DISP/LOCK button.
4 .
To take a snap log of the current gas reading, press and release the
POWER/ENTER button. The unit will display SAVED along the bottom
of the screen before returning to the Snap Logging Screen sequence.
To exit Snap Log Mode without taking a snap log or when you are
finished recording snap logs, press and release the DISP/LOCK button.
The unit will immediately return to the Measuring Mode Screen.
The data recorded in Snap Log Mode can be viewed in Display Mode.
See “Viewing Snap Logger Data” on page 54 for more information.
4:07
STATION ID
VALVE 3
31 • Measuring Mode, Alarms
GX-6000 Operator’s Manual
Measuring Mode, Alarms
This section covers alarm indications in Measuring Mode. It also describes
how to reset the GX-6000 after an alarm has occurred and how to respond to
an alarm condition.
NOTE: False alarms may be caused by radio frequency (RF) or
electromagnetic (EMI) interference. Keep the GX-6000 away from
RF and EMI sources such as radio transmitters or large motors.
Alarm Indications
The GX-6000 buzzer will sound an alarm, the LEDs will flash, and the
vibrator will pulse when any sort of alarm condition or failure is encountered.
NOTE: If an alarm condition occurs while you are in Display Mode, the
GX-6000 will automatically bring up the alarm screen instead.
The table below summarizes the types of alarms produced by the GX-6000
and their indications.
None of the Man Down alarm indications will happen if MAN DOWN is set
to OFF in Maintenance Mode (factory setting).
Table 5: Alarm Types and Indications
Alarm Type
Visual Indications
Other Indications
Low Alarm
Concentration of gas rises
above the Low Alarm setting.
• Display alternates between
gas reading and AL1
• Alarm LED arrays flash in
circle sequence once per
second
• Backlight turns on
• High-low tone
sounding twice per
second
• Vibrator pulses once
per second
High Alarm
Concentration of gas rises
above the High Alarm
setting.
• Display alternates between
gas reading and AL2
• Alarm LED arrays flash in
circle sequence twice per
second
• Backlight turns on
• High-low tone
sounding four times
per second
• Vibrator pulses twice
per second
GX-6000 Operator’s Manual
Measuring Mode, Alarms • 32
TWA or STEL
Concentration of high range
VOC rises above the TWA or
STEL alarm setting.
• Display alternates between
gas reading and TWA or STEL
• Alarm LED arrays flash in
circle sequence once per
second
• Backlight turns on
• High-low tone
sounding twice per
second
• Vibrator pulses once
per second
Over Range
• Display alternates between
OVER displayed where the
gas reading normally is and
OVER displayed where the
reading units normally are
• Alarm LED arrays flash in
circle sequence twice per
second
• Backlight turns on
• High-low tone
sounding four times
per second
• Vibrator pulses twice
per second
Low Flow
• The display indicates FAIL
LOW FLOW
• Alarm LED arrays flash once
per second
• Backlight turns on
Double pulsing tone
once per second
Low Battery Warning
• The last bar in the battery icon
disappears and the battery
icon starts flashing
None
Dead Battery Alarm
• Gas reading disappears and
FAIL BATTERY appears at the
top of the screen
• Alarm LED arrays flash once
per second
Double pulsing tone
once per second
Sensor Failure
• SENSOR appears at the top of
the screen and FAIL appears
under the gas name.
• Alarm LED arrays flash once
per second
Double pulsing tone
once per second
Clock Failure
• FAIL CLOCK appears at the
top of the screen
• Alarm LED arrays flash once
per second
Double pulsing tone
once per second
System Failure
• FAIL SYSTEM appears at the
top of the screen and an error
code displays below it
• Alarm LED arrays flash once
per second
Double pulsing tone
once per second
Table 5: Alarm Types and Indications
Alarm Type
Visual Indications
Other Indications
33 • Measuring Mode, Alarms
GX-6000 Operator’s Manual
Responding to Alarms
This section describes response to gas, over range, battery, sensor failure,
clock failure, system failure, man down, and panic alarms.
Responding to Gas Alarms
1 .
Determine which gas alarm has been activated.
2 .
Follow your established procedure for an increasing gas condition.
Man Down Warning 1
The WARNING 1 TIME
defined in Maintenance
Mode has passed since the
instrument detected
movement.
• Alarm LED arrays flash once
per second
Single pulsing tone
once per second
Man Down Warning 2
The WARNING 2 TIME
defined in Maintenance
Mode has passed since the
instrument detected
movement.
• Alarm LED arrays flash twice
per second
Single pulsing tone
twice per second
Man Down Alarm
The ALARM TIME defined in
Maintenance Mode has
passed since the instrument
detected movement.
• MAN DOWN flashes at top of
screen in place of time
• Alarm LED arrays flash in
circle sequence twice per
second
• High-low tone
sounding twice per
second
• Vibrator pulses twice
per second
Panic
User presses and holds the
SHIFT ▼ (PANIC) button.
• Screen unaffected for 5
seconds before alarm starts
• MAN DOWN flashes at top of
screen in place of time
• Alarm LED arrays flash in
circle sequence twice per
second
• Backlight turns on
• Single pulsing tone
twice per second for 5
seconds before alarm
starts
• High-low tone
sounding twice per
second
• Vibrator pulses twice
per second
Table 5: Alarm Types and Indications
Alarm Type
Visual Indications
Other Indications
GX-6000 Operator’s Manual
Measuring Mode, Alarms • 34
3 .
Reset or silence the alarm as necessary or allowed. Table 6 below
summarizes resetting and silencing alarms for all ALARM LATCHING
and ALARM SILENCE combinations that are possible. See “Updating
the Alarm Latching Setting” on page 133 and “Updating the Alarm
Silence Setting” on page 134 for further descriptions of these
parameters.
Responding to an Over Range Alarm
WARNING: An over range condition may indicate an extreme
combustible gas concentration. Confirm a normal condition
with a different GX-6000 or with another gas detecting
device.
CAUTION: A high off-scale reading may indicate an explosive
concentration.
PRUDENCE:Des lectures élevées hors échelle peuvent indiquer une
concentration explosive.
1 .
Follow your established procedure for an extreme gas condition.
2 .
Reset the alarm using the RESET button once the alarm condition has
cleared.
Table 6: Resetting and Silencing Alarms
ALARM LATCHING: LATCHING
ALARM LATCHING: SELF RESET
ALARM
SILENCE:
ON
(factory setting)
• Press and release the RESET
button to silence the buzzer.
• If the gas concentration was still
above the alarm level when the
button was pressed, the LED arrays
continue to flash, the vibrator
continues to pulse, and the
GX-6000 continues to display the
current alarm level.
• The gas reading must fall below an
alarm setting before you can reset
the alarm, the LEDs, and the
vibrator using the RESET button.
• Press and release the RESET
button to silence the buzzer.
• The RESET button will not reset the
alarm.
• Alarm, LEDs, and vibrator will
automatically reset when gas
reading falls below an alarm
setpoint.
ALARM
SILENCE:
OFF
• RESET button will not silence
buzzer.
• The gas reading must fall below an
alarm setting before you can reset
the alarm condition using the
RESET button.
• RESET button will not silence or
reset alarm.
• Alarm condition will automatically
reset when gas reading falls below
an alarm setpoint.
35 • Measuring Mode, Alarms
GX-6000 Operator’s Manual
3 .
Calibrate the GX-6000 as described in “Chapter 4: Calibration Mode” on
page 60.
4 .
If the over range condition continues or if you are not able to
successfully calibrate the unit, you may need to replace the sensor.
5 .
If the over range condition continues after you have replaced the sensor,
contact RKI Instruments, Inc. for further instructions.
Responding to Battery Alarms
WARNING: The GX-6000 is not operational as a gas monitoring device
during a dead battery alarm. Take the GX-6000 to a non-
hazardous area and replace or recharge the batteries as
described in “Replacing or Recharging the Batteries” on
page 89.
The GX-6000 is fully functional during a low battery warning. However,
only a limited amount of operating time remains, approximately 1 - 2 hours.
The amount of time depends on how often the LCD backlight is used and
how often the unit is responding to alarm conditions. Recharge the Li-ion
battery pack or replace the alkaline batteries as soon as possible as described
in “Replacing or Recharging the Batteries” on page 89.
NOTE: Alarms and the LCD back light consume battery power and reduce
the amount of operating time remaining.
Responding to a Sensor Failure Alarm
1 .
Try calibrating the sensor as described in “Chapter 4: Calibration Mode”
on page 60 before replacing it.
2 .
If the sensor failure continues, replace the sensor as described in
“Replacing the PID Sensor” on page 100.
3 .
If the sensor failure condition continues after you have replaced the
sensor, contact RKI Instruments, Inc. for further instructions.
GX-6000 Operator’s Manual
Measuring Mode, Alarms • 36
Responding to Clock Failure Alarms
A clock failure alarm occurs if the unit’s internal clock malfunctions.
1 .
Press and release the RESET button to continue into Measuring Mode.
CAUTION: There will be no datalogging function if you operate the
instrument after a clock failure.
2 .
Attempt to change the date using the DATE menu item in Maintenance
Mode. See “Setting the Date and Time” on page 123.
3 .
If the date cannot be set correctly, contact RKI Instruments, Inc. as soon
as possible.
Responding to System Failure Alarms
1 .
If a system failure occurs, the system failure screen will display an error
code as shown below:
2 .
The error code meanings are shown in the table below:
Table 7: Error Code Explanation
Error Code
Explanation
000
ROM failure
4:07
FAIL CLOCK
4:07
FAIL SYSTEM
0 2 2
37 • Measuring Mode, Alarms
GX-6000 Operator’s Manual
3 .
If the error code is anything but 022 as shown above, the instrument
cannot be used. Contact RKI Instruments, Inc. as soon as possible.
If the error code is 022, you may press and release the RESET button to
continue into Measuring Mode if the instrument must be used
temporarily.
CAUTION: There will be no datalogging function if you operate the
instrument after a 022 system failure. Contact RKI Instruments,
Inc. as soon as possible.
Responding to a Man Down Warning 1 and Warning 2
The Man Down Warning 1 and Warning 2 alarms occur after the WARNING
1 TIME and WARNING 2 TIME, respectively, has passed since the last
movement of the instrument. See “Updating the WARNING 1 TIME Setting”
on page 140 and “Updating the WARNING 2 TIME Setting” on page 140 for
instructions to change these values.
1 .
Follow your established procedure for a man down warning.
2 .
To silence the alarm and reset the Man Down clock, move the instrument
or press and release the RESET button.
Responding to a Man Down Alarm
The Man Down Alarm alarms occur after the ALARM TIME has passed
since the last movement of the instrument. See “Updating the ALARM TIME
Setting” on page 141 for instructions to change this value.
1 .
Follow your established procedure for a man down alarm.
2 .
To silence the alarm and reset the Man Down clock, press and release the
RESET button. Moving the instrument will not reset the alarm or Man
Down clock.
Responding to a Panic Alarm
If the user is in a dangerous situation or feels that others must be alerted to
any sort of problem, holding the SHIFT ▼ (PANIC) button will initiate a
panic alarm.
1 .
Press and release the RESET button to silence and reset the alarm.
010
RAM failure
021
FRAM failure
022
FLASH memory failure
Table 7: Error Code Explanation
Error Code
Explanation
GX-6000 Operator’s Manual
Display Mode • 38
Display Mode
Two other operating modes are accessible when the GX-6000 is in Measuring
Mode. They are Display Mode and Calibration Mode. This section describes
using the GX-6000 in Display Mode. See Table 8 below for a list of Display
Mode’s menu items, a short description of each item, and the page number
for further description.
Table 8: Display Mode Items
Menu Item (page number)
Description
SELECT PID LIST (pg.39)
Define the PID channel’s gas name
PEAK (pg.42)
View and clear the peak reading
STEL (pg.43)
View the STEL reading (high range PID only)
TWA (pg.43)
View the TWA reading (high range PID only)
ALARM POINTS (pg.44)
View the alarm points
OPERATION TIME (pg.46)
View the time in operation
DATE AND BATTERY (pg.46)
View the date/time, battery voltage, and battery type
LOG REMAIN (pg.47)
View the log time remaining
LOG CLEAR (pg.47)
Clear the log data
PUMP OFF (pg.49)
Turn the pump on or off (if the PUMP OFF DISP item in
Maintenance Mode is set to ON)
USER ID (pg.50)
Select the user ID (if the ID DISPLAY item in Maintenance
Mode is set to ON)
STATION ID (pg.52)
Select the station ID (if the ID DISPLAY item in
Maintenance Mode is set to ON)
REC. DATA DISP (pg.54)
View snap logger data
PEAK BAR (pg.55)
Turn the peak bar function on or off
GAS DISP (pg.56)
N/A
INVERT SELECT (pg.56)
Turn the LCD flip function on or off
LCD BACKGROUND (pg.57)
Reverse the black/white of the LCD
LANGUAGE (pg.57)
Change the instrument’s language back to English (only
appears if the language has been changed to something
other than English)
39 • Display Mode
GX-6000 Operator’s Manual
Tips for Using Display Mode
•
To enter Display Mode and scroll from one menu item to the next or skip
an item when a question is asked, press and release the DISP/LOCK
button.
•
To enter an item, press and release the POWER/ENTER button.
•
To change a flashing parameter, use either the ▲ AIR button or SHIFT ▼
(PANIC) button.
•
To return to Measuring Mode from the top level of any menu item, press
and release the RESET button.
NOTE: With the exception of the Pump Off Screen, each screen displays for
20 seconds. If you do not press a button within 20 seconds, the
GX-6000 automatically returns to Measuring Mode. If the Pump Off
Screen is used to turn the instrument’s pump off, the Pump Off
Screen will display until you turn the pump back on.
PID Gas Name Screen
The standard PID channel is configured for and calibrated to isobutylene. If
calibration to a different gas is required for an application, the PID channel
can also be configured for other gases. If you configure the instrument for
another gas, the instrument will retain that configuration until you change it
again. Turning the instrument off and on does not change this setting.
1 .
With the PID Gas Name Screen displayed, press and release the
POWER/ENTER button. The top level of the PID Gas Name List will be
displayed and the cursor will be next to the letter corresponding to the
first letter of the current setting. In the example below, the PID channel
is configured for isobutylene so the cursor is next to “I”. Each letter has a
list of gas names behind it. At the top of the letter list, there is a User List
item and a Recent List item. The User List can hold 30 gas names and
DISP
SELECT PID LIST
PID SET TO
Isobutylene
10.6 eV
CHANGE :ENTER
KEEP :DISP
EXIT :RESET
GX-6000 Operator’s Manual
Display Mode • 40
the GX-6000 Data Logger Management Program is required to
configure this list. The Recent List can hold up to 8 gas names and is
made up of the 8 most recently used gas names. The Recent List is not
user editable.
2 .
Use the ▲ AIR and SHIFT ▼ (PANIC) buttons to scroll through the letter
and list options. To view the gas names behind a letter, User List, or
Recent List, press and release the POWER/ENTER button when the
cursor is next to the desired letter.
3 .
Use the ▲ AIR and SHIFT ▼ (PANIC) buttons to scroll through the gas
names.
DISP
SELECT PID 11/24
F -
G -
H -
I-
J -
K -
L -
M -
DISP
I -
I s o b u t a n o l
I s o b u t y l a c e t a t
I s o b u t y l a c r y l a
I s o b ut y l e n e
I s o b u t y l e n e epo
I s o b u ty r a l d e hy d
I s o b u t y r ic a c i d
I s o d e c a n o l
41 • Display Mode
GX-6000 Operator’s Manual
4 .
To view the information for each gas name, press and release the
POWER/ENTER button when the cursor is next to the desired gas name.
The number in the upper left corner is the gas’s number in relation to all
other gas names that are stored in the GX-6000. The chemical formula is
displayed next to the gas number. The response factor relative to
isobutylene is listed on the second line. The gas name is listed on the
third line.
5 .
To select a gas name, press and release the POWER/ENTER button
while you are viewing the parameters. The selection will be in effect
until you change it again.
To return to the gas name list without saving any changes, press and
release the DISP/LOCK button.
6 .
To return to the top level PID Gas Name List that has all of the letter
choices, press and release the DISP/LOCK button.
7 .
To return to the PID Gas Name Screen, press and release the DISP/
LOCK button one more time.
DISP
I - 17/46
#396 C4H8
VOC RF 1.00
I s o b ut y l e n e
GX-6000 Operator’s Manual
Display Mode • 42
Peak Screen
The peak screen displays the highest concentration detected since the
GX-6000 was turned on. The peak reading is are stored in the GX-6000’s
memory until a higher level is detected, the peak reading is cleared, or the
GX-6000 is turned off.
The lunch break feature enables the GX-6000 to save the peak reading when
it is turned off so it can continue them when it is turned on again. See
“Turning On the GX-6000” on page 16.
To clear the peak reading, do the following:
1 .
With the Peak Screen displayed, press and hold the RESET button. The
following screen sequence will occur.
2 .
Release the RESET button. The peak reading will be reset and the unit
will return to the Peak Screen.
If you do not want to clear the peak reading, release the RESET button
before the above screen sequence occurs. The unit will return to the Peak
Screen.
PEAK
VOC
ppm
4300
PEAK
PEAK CLEAR
HOLD
PEAK
PEAK CLEAR
RELEASE
43 • Display Mode
GX-6000 Operator’s Manual
STEL Screen
The STEL Screen displays the short term exposure limit (STEL) reading
(high range PID only). The STEL reading is the average reading over the last
15 minutes.
TWA Screen
The TWA Screen displays the time weighted average (TWA) reading (high
range PID only).
The TWA reading is the average reading over the last 8 hours. If 8 hours have
not elapsed since the last time the TWA reading was cleared, the average is
still calculated over 8 hours. The missing time is assigned a 0 value for
readings. If LUNCH BREAK is set to OFF, the TWA is cleared when the
GX-6000 is turned off.
If LUNCH BREAK is set to ON, the GX-6000 will remember the TWA
reading when it is turned off so it can continue it when it is turned on again.
See “Turning On the GX-6000” on page 16.
STEL
VOC
ppm
10.0
TWA
VOC
ppm
5.0
GX-6000 Operator’s Manual
Display Mode • 44
View Alarm Settings Screen
The View Alarm Settings Screen gives you the option to view the gas alarm
settings.
To view the gas alarm settings, do the following:
1 .
With the View Alarm Settings Screen displayed, press and release the
POWER/ENTER button. The Full Scale Settings screen will appear. Use
the ▲ AIR and SHIFT ▼ (PANIC) buttons to scroll through the Warning,
Alarm, STEL (high range PID only), and TWA (high range PID only)
settings.
DISP
ALARM POINTS
YES :ENTER
NO :DISP
EXIT :RESET
45 • Display Mode
GX-6000 Operator’s Manual
2 .
Pressing and releasing the POWER/ENTER button while in a settings
screen will cause the instrument to simulate that condition. The buzzer,
LED’s, and vibration will activate. Press and release the POWER/
ENTER button again to stop the simulated alarm.
3 .
Press and release the DISP/LOCK button to return to the View Alarm
Settings Screen.
▲ AIR
▲ AIR
▲ AIR
▲ AIR
▲ AIR
SHIFT ▼ (PANIC)
F.S.
VOC
ppm
6000
AL1
VOC
ppm
400.0
AL2
VOC
ppm
1000
STEL
VOC
ppm
60.0
TWA
VOC
ppm
42.0
GX-6000 Operator’s Manual
Display Mode • 46
Time in Operation Screen
The Time In Operation Screen displays the length of time since the GX-6000
was turned on if the lunch break feature is turned off. With LUNCH BREAK
set to OFF, the time in operation is reset when the GX-6000 is turned off. See
“Updating the Lunch Break Setting” on page 136 for a description of the
lunch break feature.
If LUNCH BREAK is set to ON, the time in operation will only be reset if
you do not choose to resume the peak and TWA measurements when the
GX-6000 is turned on in the Lunch Break Screen described in Step 3 on
page 17 in “Turning On the GX-6000”. If you choose to resume the peak and
TWA measurements during startup, the GX-6000 will include the time in
operation when the unit was last turned off in the current time in operation.
Date/Time, Battery Voltage Screen
The Battery Voltage Screen displays the current battery voltage. Fully
charged alkaline batteries typically indicate around 6.0 volts; a fully charged
Li-ion battery pack typically indicate 4.1 volts. This screen also displays
during the startup sequence.
DISP
OPERATION TIME
0:02
DISP
DATE AND BATTERY
5/12/2015
MM/DD/YYYY
1:57
4.1V
BATTERY TYPE
ALKALINE
47 • Display Mode
GX-6000 Operator’s Manual
Log Time Remaining Screen
The Data Logging screen displays the time remaining until the data logger
memory is full.
LOG CLEAR Screen
CAUTION: Once you clear the data logger, you cannot retrieve any data
previously stored in the data logger.
The LOG CLEAR screen asks if you want to clear the data logger memory.
To continue through Display Mode without clearing the log data, press and
release the DISP/LOCK button.
DISP
LOG REMAIN
129 HOUR
DISP
LOG CLEAR
YES :ENTER
NO :DISP
EXIT :RESET
GX-6000 Operator’s Manual
Display Mode • 48
To clear the data logger memory, do the following:
1 .
With the LOG CLEAR Screen displayed, press and release the
POWER/ENTER button. The following screen appears asking you to
confirm that you want to clear the data logger memory.
2 .
If you do not want to clear the data logger memory, press and release the
DISP/LOCK button. The unit will return to the LOG CLEAR screen in
Display Mode.
3 .
If you want to clear the data logger memory, press and release the
POWER/ENTER button. The unit will display the following screen as it
clears the data.
4 .
When the instrument has cleared the data, it will return to the LOG
CLEAR screen in Display Mode.
DISP
LOG CLEAR
YES : ENTER
NO : DISP
CONFIRM ?
DISP
LOG CLEAR
CLEAR
49 • Display Mode
GX-6000 Operator’s Manual
Pump Off Screen
WARNING: The GX-6000 is not a gas monitoring device while the pump
is off.
NOTE: The item appears in Display Mode only if the PUMP OFF DISP
menu item in Maintenance Mode is turned ON. The factory setting
for PUMP OFF DISP is OFF.
This feature can be used to conserve battery life if there is a significant period
of time until the next gas monitoring task, but you do not want to wait for the
warm-up sequence to complete if you turn the instrument off and on again.
1 .
In this screen, press and release the POWER/ENTER button to turn the
pump off. The fan symbol in the upper right corner will disappear. Once
the pump is off, you cannot exit this screen and return to Measuring
Mode or continue in Display Mode until you turn the pump back on.
2 .
To turn the pump back on and continue in Display Mode or return to
Measuring Mode, press and release the DISP/LOCK button.
DISP
PUMP OFF
YES :ENTER
NO :DISP
EXIT :RESET
1:44
PUMP OFF
OFF
GX-6000 Operator’s Manual
Display Mode • 50
Select User ID Screen
This screen only appears in Display Mode if the ID DISPLAY item in
Maintenance Mode is set to ON. Use this screen to select a user ID from the
user ID list in the GX-6000’s memory. The current user ID is displayed. A
user ID can be up to 16 characters long. The GX-6000 can store up to 128
user IDs.
The user ID provides a way to identify the GX-6000 user during a data
logging session. If the user ID is changed during an operating session, a new
data session is initiated with the new user ID attached to it. This allows you to
change the user ID during operation and have each user ID that was used
during an operating session saved for the corresponding data. See the
GX-6000 Data Logger Management Program Operator’s Manual for a
detailed description of data logging and the user ID.
The user ID list cannot be edited using the GX-6000 user interface. The
GX-6000 Maintenance Data Loader Program is required to define or change
user IDs in the user ID list. For a detailed description of editing the list of
user IDs stored in the GX-6000, see the GX-6000 Maintenance Data Loader
Program Operator’s Manual.
DISP
USER ID
KIMBERLY
51 • Display Mode
GX-6000 Operator’s Manual
To select a different user ID:
1 .
With the Select User ID Screen displayed, press and release the
POWER/ENTER button. The current user ID will be displayed and it
will be flashing.
2 .
Use the ▲ AIR and SHIFT ▼ (PANIC) buttons to scroll through the
available user IDs. The list will wrap around if you get to the top or
bottom. Any of the user IDs in the list that have not been changed from
the factory setting will be shown as USER_ID_XXX where the “XXX”
is a number from 001 to 128.
3 .
When the desired user ID is displayed, press and release the POWER/
ENTER button.
NOTE: To exit the selection screen without saving a change, press and
release the DISP/LOCK button. You will return to the Select User
ID screen without saving the user ID change.
4 .
The unit will save the selected user ID as the current one and will return
you to the Select User ID Screen.
DISP
USER ID
KIMBERLY
GX-6000 Operator’s Manual
Display Mode • 52
Select Station ID Screen
This screen only appears in Display Mode if the ID DISPLAY item in
Maintenance Mode is set to ON. Use this screen to select a station ID from
the station ID list in the GX-6000’s memory. The current station ID is
displayed. A station ID can be up to 16 characters long. The GX-6000 can
store up to 128 station IDs.
The station ID provides a way to identify a location where monitoring was
done during a data logging session. If the station ID is changed during an
operating session, a new data session is initiated with the new station ID
attached to it. This allows you to change the station ID during operation and
have each station ID that was used during an operating session saved for the
corresponding data. See the GX-6000 Data Logger Management Program
Operator’s Manual for a detailed description of data logging and the station
ID.
The station ID list cannot be edited using the GX-6000 user interface. The
GX-6000 Maintenance Data Loader Program is required to define or change
station IDs in the station ID list. For a detailed description of editing the list
of station IDs stored in the GX-6000, see the GX-6000 Maintenance Data
Loader Program Operator’s Manual.
DISP
STATION ID
TANK 5
53 • Display Mode
GX-6000 Operator’s Manual
To select a different station ID:
1 .
With the Select Station ID Screen displayed, press and release the
POWER/ENTER button. The current user ID will be displayed and it
will be flashing.
2 .
Use the ▲ AIR and SHIFT ▼ (PANIC) buttons to scroll through the
available station IDs. The list will wrap around if you get to the top or
bottom. Any of the station IDs in the list that have not been changed
from the factory setting will be shown as STATION_ID_XXX where the
“XXX” is a number from 001 to 128.
3 .
When the desired station ID is displayed, press and release the POWER
ENTER button.
NOTE: To exit the selection screen without saving a change, press and
release the DISP/LOCK button. You will return to the Select Station
ID screen without saving the station ID change.
4 .
The unit will save the selected station ID as the current one and return
you to the Select Station ID Screen.
DISP
STATION ID
TANK 5
GX-6000 Operator’s Manual
Display Mode • 54
Viewing Snap Logger Data
The Snap Logging Screen allows you to view data from previous snap logs.
For information on how to use the snap logging feature, see “Snap Log
Mode” on page 29.
1 .
With the snap logging screen displayed, press and release the POWER/
ENTER button. The screen that appears will depend on whether or not
any snap logs have been taken.
2 .
If snap logs have been taken, the screen indicates the year, month, day,
and time that the most recent snap log was taken.
The number near the bottom of the screen indicates the snap log ID
number. The first snap log that is taken is given an ID of 001. The next
snap log ID is 002. The ID number increases sequentially with each set
of snap log data.
The last line of the screen indicates the Station ID that was used for the
snap log.
3 .
Use the ▲ AIR and SHIFT ▼ (PANIC) buttons to scroll through different
snap log IDs.
DISP
REC. DATA DISP
YES :ENTER
NO :DISP
EXIT :RESET
REC DATA
REC. DATA
NO DATA
REC DATA
DATE
5/20/2015
MM/DD/YYYY
3:40
NUMBER 001
TANK 5
No snap logs taken Snap logs taken
55 • Display Mode
GX-6000 Operator’s Manual
4 .
To view the data in a snap log ID, press and release the POWER/ENTER
button.
The gas reading that was taken during the snap log is displayed. You can
use the ▲ AIR and SHIFT ▼ (PANIC) buttons to scroll through the
different snap logs. The gas reading will change as you view different
snap log data but the snap log ID is not visible from this screen. To
scroll through the data by viewing the snap log ID number, press
and release POWER/ENTER.
5 .
To return to the Snap Logging Screen, press and release the DISP/LOCK
button.
Peak Bar Screen
The Peak Bar Screen allows you to turn the peak bar on or off. If the function
is turned on, the peak bar appears along the right side of the gas reading in
Measuring Mode and on the Peak Screen in Display Mode. It shows the peak
reading in bar graph format.
1 .
With the Peak Bar Screen displayed, press and release the POWER/
ENTER button. The current setting will be displayed.
2 .
Use the ▲ AIR and SHIFT ▼ (PANIC) buttons to display the desired
setting.
REC DATA
VOC
ppm
0.0
DISP
PEAK BAR
YES :ENTER
NO :DISP
EXIT :RESET
GX-6000 Operator’s Manual
Display Mode • 56
3 .
Press and release the POWER/ENTER button to save the setting and
return to the Peak Bar Screen.
To return to the Peak Bar Screen without saving any changes, press and
release the DISP/LOCK button.
Gas Display Screen
This menu item does not apply to this version of the GX-6000.
LCD Flip Screen
If INVERT SELECT is set to ON, the instrument’s LCD will automatically
flip if the instrument is turned upside down. This allows for the LCD to be
read with the instrument right side up or upside down. If desired, holding
down the DISP/LOCK button will lock the screen in its current position and
keep it from rotating. A LOCK symbol will flash at the top of the screen to
indicate that the LCD’s position is fixed. To unlock the LCD, hold the DISP/
LOCK button until the LOCK symbol disappears.
NOTE: Even if INVERT SELECT is set to ON, the screens in Calibration
Mode, User Mode, and Maintenance Mode will not flip if the
instrument is turned upside down.
If INVERT SELECT is set to OFF, the instrument’s LCD will not flip and
can only be read when the instrument is right side up. A LOCK symbol will
appear at the top of the screen to indicate that the LCD’s position is fixed.
1 .
With the Gas Display Screen showing, press and release the POWER/
ENTER button. The current setting will be displayed.
2 .
Use the ▲ AIR and SHIFT ▼ (PANIC) buttons to display the desired
setting.
3 .
Press and release the POWER/ENTER button to return to the LCD Flip
Screen.
To return to the LCD Flip Screen without saving any changes, press and
release the DISP/LOCK button.
DISP
INVERT SELECT
YES :ENTER
NO :DISP
EXIT :RESET
57 • Display Mode
GX-6000 Operator’s Manual
LCD Background Color Flip
With LCD BACKGROUND set to ON, the LCD colors will flip. The
background will be black instead of white and the text will be white instead
of black. With LCD BACKGROUND set to OFF (factory setting), the LCD
background will be white and the LCD text will be black.
1 .
With the LCD Background Flip Screen showing, press and release the
POWER/ENTER button. The current setting will be displayed.
2 .
Use the ▲ AIR and SHIFT ▼ (PANIC) buttons to display the desired
setting.
3 .
Press and release the POWER/ENTER button to return to the LCD
Background Flip Screen.
To return to the LCD Background Flip Screen without saving any
changes, press and release the DISP/LOCK button.
Language Screen
If the instrument’s language is changed to something other than English, the
Language Screen will appear in Display Mode. The Language Screen will
not appear if the instrument’s language is set to English. The Language
Screen allows the user to change the instrument’s language back to English.
DISP
LCD BACKGROUND
YES :ENTER
NO :DISP
EXIT :RESET
DISP
LANGUAGE
CHANGE
TO ENGLISH
YES :ENTER
NO :DISP
EXIT :RESET
GX-6000 Operator’s Manual
Data Logging • 58
To change the language back to English, do the following:
1 .
With the Language Screen showing, press and release the POWER/
ENTER button. The following screen will appear asking you to confirm
that you want to change the language back to English.
2 .
Press and release the POWER/ENTER button again to confirm the
change.
To return to the Language Screen without changing the instrument’s
language to English, press and release the DISP/LOCK button.
Data Logging
The GX-6000 features the ability to log data to its internal memory and
download it to a computer via the infrared communications port on the front
of the unit. It logs gas readings in Measuring Mode, alarm data, and
calibration data.
NOTE: The data logging function is separate from the snap logging
function. For a description of the snap logging function, see “Snap
Log Mode” on page 29.
To utilize the GX-6000’s downloading capability, you will need the GX-6000
Data Logger Management Program and a computer with an infrared port or a
USB port that runs one of the following operating systems: Windows® 7,
Windows® 8, or Windows® 10. If your computer has an infrared port, then
no additional accessories are needed to download data from the GX-6000. If
your computer does not have an infrared port but does have a USB port, a
USB/IrDA adapter cable can be used to download data from the GX-6000
using the USB port. The GX-6000 Data Logger Management Program is
available at www.rkiinstruments.com/gx6000. The USB/IrDA adapter cable
is also available from RKI Instruments, Inc.
DISP
LANGUAGE
CHANGE
TO ENGLISH
CONFIRM ?
59 • Data Logging
GX-6000 Operator’s Manual
The data logging capacity depends on how often the GX-6000 stores data and
how often the GX-6000 is turned on and off. The table below illustrates how
much data logging time is available for the various interval times. It assumes
that the unit is only turned on once and that there are no alarm occurrences.
See “Updating the Data Log Interval Setting” on page 142 for instructions on
setting the data logging interval time.
For a complete description of the Data Logger Management Program and
procedures for downloading data to a computer, see the GX-6000 Data
Logger Management Program Operator’s Manual.
Table 9: Data Logging Capacity
Interval
Time
Data Logging Time
10 seconds
10 hours
20 seconds
20 hours
30 seconds
30 hours
1 minute
60 hours
3 minutes
180 hours
5 minutes
300 hours
10 minutes
600 hours
GX-6000 Operator’s Manual
Overview • 60
Chapter 4: Calibration Mode
Overview
This section describes the GX-6000 in Calibration Mode. In Calibration
Mode, you can move through a menu of screens to do the following:
•
Perform a fresh air (zero) adjustment
•
Perform a span adjustment using auto calibration
•
Perform a span adjustment using single calibration
•
Perform a bump test
NOTE: You can set up the GX-6000 to alert you during the startup sequence
when calibration is due. See “Updating the Calibration Reminder
Setting” on page 125.
The unit should be calibrated periodically. The optimum frequency of
calibration depends heavily on how the GX-6000 is used. For example,
instruments used daily may need to be calibrated weekly or monthly, while
instruments that are used only a few times a year may need to be calibrated
before each use. Typical calibration frequencies range from monthly to
quarterly.
Calibration Supplies and
Equipment
To calibrate the GX-6000, you will need:
•
Known calibrating samples of the gas being detected. RKI Instruments,
Inc. normally recommends a calibration concentration between 10% and
50% of the full scale value but in this case, RKI Instruments, Inc.
recommends a calibration concentration of 10 ppm IBL for the low range
sensor and 100 ppm IBL for the high range sensor.
•
A demand-flow regulator to provide adequate sample gas flow
•
Non-absorbent tubing
WARNING: If you are using a calibration kit that includes a gas bag and
a fixed flow regulator or dispensing valve, do not apply gas
directly to the GX-6000 with the regulator or dispensing valve
or damage to the pump will result.
61 • Entering Calibration Mode
GX-6000 Operator’s Manual
Entering Calibration Mode
To enter Calibration Mode, do the following:
1 .
Find a fresh-air environment. This is an environment free of toxic or
combustible gases and of normal oxygen content (20.9%).
2 .
While in Measuring Mode, press and hold the SHIFT ▼ (PANIC) button,
then press the DISP/LOCK button and release both buttons.
3 .
If the unit prompts you for the password, enter it by using the ▲ AIR and
SHIFT ▼ (PANIC) buttons to select each password number and then
pressing and releasing POWER/ENTER to enter the number and move
on to the next one.
NOTE: You will only be prompted to enter a password if the PASSWORD
menu item in User Mode is set to ON. See “Turning the Password
Function On or Off” on page 85 for instructions to update the
PASSWORD setting and to choose a password.
4 .
The Calibration Mode Screen displays with the cursor next to AIR CAL.
Calibrating Using the Auto
Calibration Method
This method is the quickest and easiest method to calibrate the GX-6000. See
Table 13 on page 113 for available cylinders. Make sure your calibration
cylinder is appropriate for the PID detection range.
Setting the Fresh Air Reading
1 .
Find a fresh-air environment. This is an environment free of toxic or
combustible gases and of normal oxygen content (20.9%).
2 .
While in Measuring Mode, press and hold the SHIFT ▼ (PANIC) button,
CAL
CAL MODE
AIR CAL
AUTO CAL
SINGLE CAL
NORMAL MODE
GX-6000 Operator’s Manual
Calibrating Using the Auto Calibration Method • 62
then press the DISP/LOCK button and release both buttons.
3 .
If the unit prompts you for the password, enter it by using the ▲ AIR and
SHIFT ▼ (PANIC) buttons to select each password number and then
pressing and releasing POWER/ENTER to enter the number and move
on to the next one.
NOTE: You will only be prompted to enter a password if the PASSWORD
menu item in User Mode is set to ON. See “Turning the Password
Function On or Off” on page 85 for instructions to update the
PASSWORD setting and to choose a password.
4 .
The Calibration Mode Screen displays with the cursor next to AIR CAL.
5 .
Press and release the POWER/ENTER button. The Fresh Air Reading
screen will display.
6 .
If you suspect any low-level VOC background in the area, attach the
tubing stub of the VOC zero filter to the inlet fitting or probe and let the
instrument draw through it for 1 minute before continuing.
7 .
To continue with the fresh air adjustment, press and hold the ▲ AIR
button.
If you do not want to continue, press and release the DISP/LOCK button
and the unit will return to the Calibration Mode Screen.
CAL
CAL MODE
AIR CAL
AUTO CAL
SINGLE CAL
NORMAL MODE
AIR
VOC
ppm
0.0
63 • Calibrating Using the Auto Calibration Method
GX-6000 Operator’s Manual
8 .
The GX-6000 will indicate that it is adjusting the zero reading for a few
seconds and then it will prompt you to release the ▲ AIR button.
9 .
Release the ▲ AIR button. The fresh air adjustment will finish, the fresh
air reading will be displayed momentarily, and then the instrument will
return to the Calibration Mode Screen.
Performing a Span Adjustment in Auto Calibration
1 .
Install the demand flow regulator onto the calibration cylinder.
2 .
Connect the sample tubing to the demand flow regulator.
AIR
AIR CAL
HOLD AIR KEY
AIR
AIR CAL
HOLD AIR KEY
ADJUSTING ZERO
AIR
AIR CAL
RELEASE AIR KEY
ZERO ADJUSTED
AIR
AIR CAL
ZERO ADJUSTED
CAL
CAL MODE
AIR CAL
AUTO CAL
SINGLE CAL
NORMAL MODE
AIR
AIR CAL
END
AIR
VOC
ppm
0.0
GX-6000 Operator’s Manual
Calibrating Using the Auto Calibration Method • 64
3 .
Install the sample hose and probe on the GX-6000 inlet fitting. Make
sure the probe’s two halves are tightened firmly together to avoid leaks
that can affect the calibration. See Figure 21, “Replacing the Particle
Filter and Hydrophobic Filter Disk” on page 99 for an illustration of the
internal parts of the probe.
4 .
Move the cursor next to the AUTO CAL menu item by using the ▲ AIR
or SHIFT ▼ (PANIC) button.
5 .
Press and release the POWER/ENTER button to display the Auto Cal
Menu.
CAL
CAL MODE
AIR CAL
AUTO CAL
SINGLE CAL
NORMAL MODE
AUTO
AUTO CAL
CONCENTRATION
GAS SELECT
ESCAPE
65 • Calibrating Using the Auto Calibration Method
GX-6000 Operator’s Manual
6 .
To check or update the gas concentration, use the ▲ AIR and SHIFT ▼
(PANIC) buttons to put the cursor next to CONCENTRATION and
press and release the POWER/ENTER button to display the Calibration
Gas Values Screen.
The gas concentration displayed in the Calibration Gas Values Screen
must match the gas concentration listed on the PID cylinder.
If the concentration matches, go to Step 11.
If the concentration does not match, continue with Step 7.
If you do not want to continue with the concentration adjustment, press
and release the DISP/LOCK button to return to the Auto Cal Menu.
7 .
The cursor will already be next to the PID channel.
8 .
Press and release the POWER/ENTER button. The auto cal value will
begin to flash.
9 .
Use the ▲ AIR and SHIFT ▼ (PANIC) buttons to adjust the calibration
gas setting to the desired value.
10 . Press and release the POWER/ENTER button to save the change. The
calibration gas value stops flashing.
11 . When you are done adjusting the calibration gas value, move the cursor
to ESCAPE.
AUTO
CONCENTRATION
VOC 100.0 ppm
ESCAPE
AUTO
CONCENTRATION
VOC 100.0 ppm
ESCAPE
GX-6000 Operator’s Manual
Calibrating Using the Auto Calibration Method • 66
12 . Press and release the POWER/ENTER button. The instrument returns to
the Auto Cal Menu.
13 . Use the ▲ AIR and SHIFT ▼ (PANIC) buttons to move the cursor next to
GAS SELECT.
14 . Press and release the POWER/ENTER button. The PID sensor screen
will be displayed. In addition to the PID sensor screen, there is an
ESCAPE screen that will take you back to the Auto Cal menu. Use the
▲ AIR and SHIFT ▼ (PANIC) buttons to scroll between the screens.
AUTO
AUTO CAL
CONCENTRATION
GAS SELECT
ESCAPE
AUTO
GAS SELECT
VOC 100.0 ppm
AUTO
GAS SELECT
ESCAPE
▲ AIR or
SHIFT ▼ (PANIC)
67 • Calibrating Using the Auto Calibration Method
GX-6000 Operator’s Manual
15 . With the PID sensor screen displayed, press and release the POWER/
ENTER button. The target gas selection screen will display. This screen
allows you to select the gas you want to use for calibration. Isobutylene
will always be displayed on the top line. The second line will show the
gas that is currently selected in the PID Gas Name Screen in Display
Mode. In the example below, acetone is the currently selected gas. If
isobutylene is the selected gas, then isobutylene will appear twice in this
list.
16 . Use the ▲ AIR and SHIFT ▼ (PANIC) buttons to move the cursor next to
the gas you want to use for calibration.
17 . Press and release the POWER/ENTER button to proceed to the
Calibration In Process Screen with APPLYGAS and the gas reading
flashing.
If you do not want to proceed with the calibration, press and release the
DISP/LOCK button to return to the Gas Select Screen.
If you do want to continue with the calibration, proceed to the next step.
18 . Connect the tubing from the demand flow regulator to the rigid tube on
the probe. Allow the GX-6000 to draw gas for one minute.
AUTO
GAS SELECT
Isobutylene
Acetone
ESCAPE
APPLYGAS
VOC
ppm
100.0
GX-6000 Operator’s Manual
Calibrating Using the Auto Calibration Method • 68
19 . Press and release the POWER/ENTER button to set the span adjustment
to the programmed value.
20 . If the PID channel passed calibration, the following screen sequence
occurs.
21 . If the PID sensor cannot be adjusted to the proper value, a screen
displays that indicates a calibration failure.
The buzzer and alarm LEDs activate in a double pulsing pattern. Press
and release the RESET button to reset the alarm and continue to the Auto
Cal Menu. Attempt to calibrate the PID channel again. If the failure
continues, investigate the cause. See “Troubleshooting” on page 87.
Returning to Measuring Mode
1 .
Disconnect the tubing from the probe.
2 .
Unscrew the demand flow regulator from the calibration cylinder.
3 .
Use the SHIFT ▼ (PANIC) button to move the cursor next to ESCAPE,
then press and release the POWER/ENTER button to return to the
Calibration Mode Menu.
4 .
Use the SHIFT ▼ (PANIC) button to place the cursor next to the
NORMAL MODE menu option, then press and release the POWER/
ENTER button to return to Measuring Mode.
AUTO
PASS
AUTO
AUTO CAL
CONCENTRATION
GAS SELECT
ESCAPE
AUTO
VOC
ppm
100.0
SENSOR
VOC
ppm
FAIL
69 • Calibrating Using the Single Calibration Method
GX-6000 Operator’s Manual
Calibrating Using the Single
Calibration Method
Setting the Fresh Air Reading
1 .
While in the Calibration Mode Screen, move the cursor to the AIR CAL
menu item by using the SHIFT ▼ (PANIC) button.
2 .
Press and release the POWER/ENTER button. The Fresh Air Reading
Screen will be displayed.
3 .
If you suspect any low-level VOC background in the area, attach the
tubing stub of the VOC zero filter to the inlet fitting or probe and let the
instrument draw through it for 1 minute before continuing.
4 .
To continue with the fresh air adjustment, press and hold the ▲ AIR
button.
If you do not want to continue, press and release the DISP/LOCK button
and the unit will return to the Calibration Mode Screen.
CAL
CAL MODE
AIR CAL
AUTO CAL
SINGLE CAL
NORMAL MODE
AIR
VOC
ppm
0.0
GX-6000 Operator’s Manual
Calibrating Using the Single Calibration Method • 70
5 .
The GX-6000 will indicate that it is adjusting the zero reading for a few
seconds and then it will prompt you to release the ▲ AIR button.
6 .
Release the ▲ AIR button. The fresh air adjustment will finish, the fresh
air reading will be displayed momentarily, and then the instrument will
return to the Calibration Mode Screen.
Performing a Span Adjustment in Single Calibration
1 .
Install the demand flow regulator onto the calibration cylinder.
2 .
Connect the sample tubing to the demand flow regulator.
AIR
AIR CAL
HOLD AIR KEY
AIR
AIR CAL
HOLD AIR KEY
ADJUSTING ZERO
AIR
AIR CAL
RELEASE AIR KEY
ZERO ADJUSTED
AIR
AIR CAL
ZERO ADJUSTED
CAL
CAL MODE
AIR CAL
AUTO CAL
SINGLE CAL
NORMAL MODE
AIR
AIR CAL
END
AIR
VOC
ppm
0.0
71 • Calibrating Using the Single Calibration Method
GX-6000 Operator’s Manual
3 .
Install the sample hose and probe on the GX-6000 inlet fitting. Make
sure the probe’s two halves are tightened firmly together to avoid leaks
that can affect the calibration. See Figure 21, “Replacing the Particle
Filter and Hydrophobic Filter Disk” on page 99 for an illustration of the
internal parts of the probe.
4 .
Move the cursor next to the SINGLE CAL menu item by using the
▲ AIR or SHIFT ▼ (PANIC) button.
5 .
Press and release the POWER/ENTER button. The Select Sensor Screen
appears.
6 .
The cursor will already be next to the PID channel.
If you do not want to proceed with the span adjustment, press and release
the DISP/LOCK button or place the cursor next to ESCAPE and press
and release POWER/ENTER to return to the Calibration Mode Screen.
To proceed with the calibration, continue with the next step.
CAL
CAL MODE
AIR CAL
AUTO CAL
SINGLE CAL
NORMAL MODE
SINGLE
SINGLE CAL
VOC
ESCAPE
GX-6000 Operator’s Manual
Calibrating Using the Single Calibration Method • 72
7 .
Press and release the POWER/ENTER button. The target gas selection
screen will display. This screen allows you to select the gas you want to
use for calibration. Isobutylene will always be displayed on the top line.
The second line will show the gas that is currently selected in the PID
Gas Name Screen in Display Mode. In the example below, acetone is the
currently selected gas. If isobutylene is the selected gas, then isobutylene
will appear twice in this list.
8 .
Use the ▲ AIR and SHIFT ▼ (PANIC) buttons to move the cursor next to
the gas you want to use for calibration.
9 .
Press and release the POWER/ENTER button to proceed to the Single
Calibration Apply Gas Screen. APPLYGAS and the gas reading will be
flashing.
10 . Connect the tubing from the demand flow regulator to the rigid tube on
the probe. Allow the GX-6000 to draw gas for one minute.
11 . If necessary, adjust the reading to match the cylinder concentration with
the ▲ AIR and SHIFT ▼ (PANIC) buttons. If the reading cannot be
adjusted to match the cylinder concentration, see “Troubleshooting” on
page 87.
SINGLE
GAS SELECT
Isobutylene
Acetone
ESCAPE
APPLYGAS
VOC
ppm
100.0
73 • Calibrating Using the Single Calibration Method
GX-6000 Operator’s Manual
12 . Press and release the POWER/ENTER button to save the span value.
The LCD will indicate that the calibration has ended, then the Select
Sensor Screen appears.
13 . Disconnect the tubing from the GX-6000’s probe.
14 . Unscrew the demand flow regulator from the calibration cylinder.
15 . With the Select Sensor Screen displayed, place the cursor next to
ESCAPE using the ▲ AIR or SHIFT ▼ (PANIC) button.
16 . Press and release the POWER/ENTER button to return to the Calibration
Mode Screen.
17 . Use the SHIFT ▼ (PANIC) button to place the cursor next to the
NORMAL MODE menu item, then press and release the POWER/
ENTER button to return to Measuring Mode.
SINGLE
PASS
SINGLE
SINGLE CAL
VOC
ESCAPE
SINGLE
SINGLE CAL
VOC
ESCAPE
GX-6000 Operator’s Manual
Performing a Bump Test • 74
Performing a Bump Test
NOTE: The BUMP DISP menu item in Maintenance Mode is factory set to
OFF. The BUMP TEST menu item will not appear unless BUMP
DISP is set to ON. See “Updating the Bump Display Setting” on
page 127 for instructions.
NOTE: The GX-6000 uses parameters defined in the BUMP
PARAMETERS menu item of Maintenance Mode while
performing a bump test. To view or update these parameters, see
“Updating the Bump Parameters” on page 128.
1 .
Install the demand flow regulator onto the calibration cylinder.
2 .
Connect the sample tubing to the demand flow regulator.
3 .
Install the sample hose and probe on the GX-6000 inlet fitting. Make
sure the probe’s two halves are tightened firmly together to avoid leaks
that can affect the calibration. See Figure 21, “Replacing the Particle
Filter and Hydrophobic Filter Disk” on page 99 for an illustration of the
internal parts of the probe.
4 .
While in Measuring Mode, press and hold the SHIFT ▼ (PANIC) button,
then press the DISP/LOCK button and release both buttons.
5 .
If the unit prompts you for the password, enter it by using the ▲ AIR and
SHIFT ▼ (PANIC) buttons to select each password number and then
pressing and releasing POWER/ENTER to enter the number and move
on to the next one.
NOTE: You will only be prompted to enter a password if the PASSWORD
menu item in User Mode is set to ON. See “Turning the Password
Function On or Off” on page 85 for instructions to update the
PASSWORD setting and to choose a password.
75 • Performing a Bump Test
GX-6000 Operator’s Manual
6 .
The Calibration Mode Screen displays with the cursor next to AIR CAL.
7 .
Move the cursor next to the BUMP TEST menu item by using the
SHIFT ▼ (PANIC) button.
8 .
Press and release the POWER/ENTER button. The PID sensor screen
will be displayed. In addition to the PID sensor screen there is an
ESCAPE screen that will take you back to the Calibration Mode menu.
Use the ▲ AIR and SHIFT ▼ (PANIC) buttons to display the PID sensor
screen.
CAL
CAL MODE
AIR CAL
AUTO CAL
SINGLE CAL
BUMP TEST
NORMAL MODE
CAL
CAL MODE
AIR CAL
AUTO CAL
SINGLE CAL
BUMP TEST
NORMAL MODE
AUTO
GAS SELECT
VOC 100.0 ppm
AUTO
GAS SELECT
ESCAPE
▲ AIR or
SHIFT ▼ (PANIC)
GX-6000 Operator’s Manual
Performing a Bump Test • 76
9 .
Before proceeding, confirm that the bump test gas value is the same as
the concentration in the calibration cylinder. If it is not, adjust the bump
test gas value by entering the AUTO CAL menu item in Calibration
Mode, changing the value there, and reentering the BUMP TEST menu
item.
10 . With the PID sensor screen displayed, connect the tubing from the
demand flow regulator to the rigid tube on the probe then quickly press
and release the POWER/ENTER button to proceed to the Bump In
Progress Screen. The instrument will count down.
If you do not want to proceed with the bump test, press and release the
DISP/LOCK button to return to the Gas Select Screen.
If you do want to continue with the bump test, proceed to the next step.
30BUMP
VOC
ppm
100.0
77 • Performing a Bump Test
GX-6000 Operator’s Manual
11 . When AUTO CAL in the BUMP PARAMETERS menu item in
Maintenance Mode is set to OFF:
a. The instrument will indicate whether or not the PID channel passed
or failed the bump test with an OK (pass) or an NG (fail) to the right
of the gas. You can scroll between the bump test result and the bump
test gas reading with the ▲ AIR and SHIFT ▼ (PANIC) button.
b. To return to the Gas Select Screen, press and release the POWER/
ENTER button. If the PID channel failed the bump test, the buzzer
will sound a double pulsing tone until you press and release the
POWER/ENTER button.
If the PID sensor failed the bump test, see “Troubleshooting” on
page 87 to investigate the cause of the failure and replace the failed
sensor, if necessary.
c. Disconnect the tubing from the probe.
▲ AIR or
SHIFT ▼ (PANIC)
BUMP
VOC
ppm
100.0
BUMP
VOC OK
GX-6000 Operator’s Manual
Performing a Bump Test • 78
12 . When AUTO CAL in the BUMP PARAMETERS menu item in
Maintenance Mode is set to ON:
If the PID channel passes the bump test:
a. The results screen appears. You can scroll between the bump test
result and the bump test gas reading with the ▲ AIR and SHIFT ▼
(PANIC) buttons.
b. To return to the Gas Select Screen, press and release the POWER/
ENTER button.
c. Disconnect the tubing from the probe.
If the P ID channel fails the bump test:
a. The instrument will beep and a calibration is immediately and
automatically started. Continue to apply the calibration gas. CAL will
appear at the top of the screen along with a countdown.
The calibration time counted down during a calibration initiated
because of a failed bump test is the difference between the gas time
and the calibration time defined in the BUMP SETTING menu item
in Maintenance Mode.
▲ AIR or
SHIFT ▼ (PANIC)
BUMP
VOC
ppm
100.0
BUMP
VOC OK
60CAL
VOC
ppm
100.0
79 • Performing a Bump Test
GX-6000 Operator’s Manual
b. At the end of the calibration, the instrument displays the results from
both the bump test and the calibration. Use the ▲ AIR and SHIFT ▼
(PANIC) button to scroll between the calibration/bump test results,
the bump test gas reading, and the calibration gas reading.
c. To return to the Gas Select Screen, press and release the POWER/
ENTER button at any time. If the PID channel failed the calibration,
the buzzer will sound a double pulsing tone until you press and
release the POWER/ENTER button.
d. Disconnect the tubing from the probe.
13 . Unscrew the demand flow regulator from the calibration cylinder.
14 . Use the SHIFT ▼ (PANIC) button to display the ESCAPE screen, then
press and release the POWER/ENTER button to return to the Calibration
Mode Menu.
15 . Use the SHIFT ▼ (PANIC) button to place the cursor next to the
NORMAL MODE menu option, then press and release the POWER/
ENTER button to return to Measuring Mode.
▲ AIR or
▲ AIR or
▲ AIR or
SHIFT ▼ (PANIC)
SHIFT ▼ (PANIC)
SHIFT ▼ (PANIC)
BUMP
VOC NG OK
CAL
VOC
ppm
100.0
BUMP
VOC
ppm
27.0
GX-6000 Operator’s Manual
Overview • 80
Chapter 5: User Mode
Overview
This section describes the GX-6000 in User Mode. See Table 10 below for a
list of the items found in User Mode, the page that the menu item’s
instructions can be found on, and a short description of the menu item.
Tips for Using User Mode
•
When in the main menu, the cursor (>) indicates which menu item will be
selected if the POWER/ENTER button is pressed and released.
•
Use the SHIFT ▼ (PANIC) button to move the cursor down through the
main menu and submenu items, and to lower values or change the setting
in a specific option.
•
Use the ▲ AIR button to move the cursor up through the main menu and
submenu items, and to raise values or change the setting in a specific
option.
•
Use the POWER/ENTER button to enter a selected menu item with the
cursor next to it and to enter and save settings during programming.
•
An adjustable parameter that is flashing can be adjusted with the ▲ AIR
and SHIFT ▼ (PANIC) buttons.
•
Press the DISP/LOCK button while in a screen where you are entering or
updating parameters to exit the screen without saving any changes.
Table 10: User Mode Menu Items
Menu Item (page number)
Description
DATE (pg.82)
Set the instrument’s date and time
DATE FORMAT (pg.83)
Set the date format
CONFIGURE CH (pg.83)
N/A
LANGUAGE (pg.83)
Change the instrument’s language
ROM/SUM (pg.84)
View the instrument’s ROM number and checksum
PASSWORD (pg.85)
Turn the User Mode password function on or off
and update the password
START MEASURE (pg.86)
Return to Measuring Mode
81 • Entering User Mode
GX-6000 Operator’s Manual
Entering User Mode
WARNING: The GX-6000 is not in operation as a gas detector while in
User Mode.
1 .
Take the GX-6000 to a non-hazardous location and turn it off if it is on.
2 .
Press and hold the ▲ AIR and SHIFT ▼ (PANIC) buttons, then press and
hold the POWER/ENTER button. When you hear a beep, release the
buttons.
3 .
The screen that appears will depend on the setting of User Mode’s
PASSWORD item.
If PASSWORD is set to OFF (factory setting), continue with Step 4.
If PASSWORD is set to ON, continue with Step 6.
4 .
If PASSWORD has been set to OFF, the Mode Select screen will
appear, prompting you to choose the mode you’d like to enter.
5 .
Use the ▲ AIR and SHIFT ▼ (PANIC) buttons to put the cursor in front
of USER MODE and press and release the POWER/ENTER button.
Continue to Step 8.
6 .
If PASSWORD has been set to ON in User Mode, a password screen
will appear. The first digit will be flashing.
7 .
Enter the password by using the ▲ AIR and SHIFT ▼ (PANIC) buttons
to select each password number and then pressing and releasing the
POWER/ENTER button to enter it and move on to the next number until
all of the numbers are entered. Continue to Step 8.
PASSWORD
MODE SELECT
USER MODE
MAINTENANCE
PASSWORD
INPUT
PASSWORD
0 0 0 0
PASSWORD set to OFF PASSWORD set to ON
GX-6000 Operator’s Manual
Setting the Date and Time • 82
8 .
The main menu displays. It has 7 menu items.
9 .
Use the ▲ AIR or SHIFT ▼ (PANIC) button to move the cursor up and
down the menu items.
Setting the Date and Time
NOTE: The example screen shown below assumes a DATE FORMAT
setting of MM/DD/YYYY (factory setting). If your instrument’s
DATE FORMAT is set to something else, your screen will appear
different.
1 .
From the main menu, place the cursor next to DATE.
2 .
Press and release POWER/ENTER. The date and time will be displayed
with the year flashing.
3 .
Use ▲ AIR or SHIFT ▼ (PANIC) to display the desired year.
4 .
Press and release POWER/ENTER to save the setting. The month setting
flashes.
5 .
Repeat Step 3 and Step 4 to enter the month, day, hours, and minutes
settings. The main menu displays after you enter the seconds setting.
MAINTE
USER
DATE
DATE FORMAT
CONFIGURE CH
LANGUAGE
ROM/SUM
PASSWORD
START MEASURE
DATE
DATE
5/ 1/2015
MM/DD/YYYY
12:55
83 • Setting the Date Format
GX-6000 Operator’s Manual
Setting the Date Format
The date can be displayed in three ways, MM/DD/YYYY (factory setting),
DD/MM/YYYY, or YYYY/MM/DD.
1 .
From the main menu, place the cursor next to DATE FORMAT.
2 .
Press and release POWER/ENTER. The Date Format screen appears
with the current setting flashing.
3 .
Use ▲ AIR or SHIFT ▼ (PANIC) to display the desired setting.
4 .
Press and release POWER/ENTER to save the setting and return to the
main menu.
Configuring the Channels
This menu item does not apply to this version of the GX-6000.
Updating the Language Setting
The LANGUAGE menu item allows you to select the language for the
GX-6000’s user interface. The available choices are English (factory setting),
Japanese, Italian, Spanish, German, French, Portuguese, Russian, and
Korean.
1 .
From the main menu, place the cursor next to LANGUAGE.
2 .
Press and release POWER/ENTER. The Select Language Screen
appears with the cursor in front of the current language.
USER
LANGUAGE
ENGLISH
JAPANESE
ITALIAN
SPANISH
GERMAN
FRENCH
PORTUGUESE
RUSSIAN
GX-6000 Operator’s Manual
Viewing the ROM/SUM of the Instrument • 84
3 .
Use ▲ AIR or SHIFT ▼ (PANIC) to move the cursor in front of the
desired language.
If you do not wish to select a new language, press and release DISP/
LOCK. The unit will return to the main menu.
4 .
Press and release POWER/ENTER to save the new language setting and
return to the main menu. The GX-6000’s user interface will now be in
the newly selected language.
Viewing the ROM/SUM of the Instrument
The ROM/SUM screen shows the firmware version that is loaded in the
instrument and the firmware checksum.
1 .
Use the ▲ AIR or SHIFT ▼ (PANIC) button to place the cursor next to
ROM/SUM.
2 .
Press and release the POWER/ENTER button. The ROM and checksum
values for your unit will be displayed.
3 .
The first line displays the ROM number. The ROM number indicates the
firmware version number. In the above example, the ROM number is
05017. The bottom line displays the firmware file’s checksum, 62C7 in
the above example.
4 .
Press and release the POWER/ENTER button again to return to the main
menu.
ROM/SUM
ROM 05017
SUM 62C7
85 • Turning the Password Function On or Off
GX-6000 Operator’s Manual
Turning the Password Function On or Off
With PASSWORD set to ON, the GX-6000 prompts you for a password
when you enter User Mode and Calibration Mode.
With PASSWORD set to OFF (factory setting), no password is required to
enter User Mode or Calibration Mode.
1 .
From the main menu, place the cursor in front of PASSWORD.
2 .
Press and release POWER/ENTER. The Password Protection Screen
appears.
3 .
Use ▲ AIR or SHIFT ▼ (PANIC) to display the desired setting.
4 .
If you selected OFF, press and release POWER/ENTER to save the
setting and return to the main menu.
If you selected ON, continue with Step 5.
5 .
Press and release POWER/ENTER. The Set Password Screen appears.
0000 is at the bottom of the screen with the first 0 flashing.
6 .
Use ▲ AIR or SHIFT ▼ (PANIC) to display a number from 0 to 9.
USER
USER PASSWORD
ON
USER
USER PASSWORD
INPUT
PASSWORD
0 0 0 0
GX-6000 Operator’s Manual
Exiting User Mode • 86
7 .
Press and release POWER/ENTER to enter the selection and advance to
the next number.
8 .
Repeat Step 6 and Step 7 to select the remaining numbers. When you
press and release POWER/ENTER to enter the last number, the
password is saved and you return to the main menu.
Exiting User Mode
1 .
From the main menu, place the cursor in front of START MEASURE at
the bottom of the menu.
2 .
Press and release POWER/ENTER.
3 .
The unit will begin its start-up sequence.
87 • Overview
GX-6000 Operator’s Manual
Chapter 6: Maintenance
Overview
This chapter describes troubleshooting procedures for the GX-6000. It also
includes procedures for replacing and recharging the batteries and replacing
various consumable parts.
WARNING: RKI Instruments, Inc. recommends that service, calibration,
and repair of RKI Instruments be performed by personnel
properly trained for this work. Replacing the sensor and
other parts with original equipment does not affect the
intrinsic safety of the instrument.
Troubleshooting
The troubleshooting table describes error messages, symptoms, probable
causes, and recommended action for problems you may encounter with the
GX-6000.
Table 11: Troubleshooting the GX-6000
Symptoms
Probable Causes
Recommended Action
• The LCD is blank.
• The unit may have been
turned off.
• The alkaline batteries may
need to be replaced or Li-
ion battery pack may need
to be recharged.
1. To turn on the unit, press and
briefly hold the POWER/ENTER
button.
2. If the unit does not turn on,
replace the alkaline batteries or
recharge the Li-ion battery pack.
3. If the difficulties continue, contact
RKI Instruments, Inc. for further
instruction.
• The LCD shows
an abnormally
high or low reading
but other gas
detection
instruments do
not.
• The GX-6000 may need to
be recalibrated.
• The sensor may need
replacement.
1. Recalibrate the unit.
2. If the difficulties continue, replace
the sensor and calibrate again.
3. If the difficulties continue, contact
RKI Instruments, Inc. for further
instruction.
GX-6000 Operator’s Manual
Troubleshooting • 88
• The unit indicates
flow failure and
does not recover
when RESET is
pressed and
released.
• The probe tube is
clogged.
• The hydrophobic filter disk
in the probe is dirty.
• The sample hose has a
kink or obstruction.
• The pump is
malfunctioning.
1. Inspect the probe tube for any
obstructions.
2. Inspect the hydrophobic filter disk
in the probe and replace if
necessary.
3. Inspect the sample hose for kinks
or obstructions and replace if
necessary.
4. If difficulties continue, contact
RKI Instruments, Inc. for further
instruction.
• Auto calibration or
single calibration
fails.
• The auto calibration value
may not match the
cylinder gas concentration
(auto calibration only).
• The sample gas is not
reaching the sensor
because of a bad
connection.
• The calibration cylinder
may be out of gas or is
outdated.
• The lamp may need to be
cleaned.
• The electrode stack may
need to be replaced.
• The sensor may need
replacement.
1. Check all calibration tubing for
leaks or for any bad connections.
2. Make sure the GX-6000 has
been properly set up for
calibration.
3. Verify that the calibration cylinder
contains an adequate supply of
fresh test sample.
4. Clean the PID sensor’s lamp.
5. Replace the PID sensor’s
electrode stack.
6. If the fail condition continues,
replace the sensor.
7. If the difficulties continue, contact
RKI Instruments, Inc. for further
instruction.
• Heart symbol at
the top of the
screen becomes
steadily on or
disappears
• A microprocessor error
has occurred.
• Contact RKI Instruments, Inc. for
further instruction.
Table 11: Troubleshooting the GX-6000
Symptoms
Probable Causes
Recommended Action
89 • Replacing or Recharging the Batteries
GX-6000 Operator’s Manual
Replacing or Recharging the
Batteries
WARNING: To prevent ignition of a hazardous atmosphere, batteries
must only be changed or charged in an area known to be
nonhazardous.
AVERTISSEMENT:Pour éviter l’inflammation d’une atmosphère
dangereuse, les batteries doivent uniquement être modifiés
ou facturés dans une zone connue comme non dangereuse.
Replace or recharge the batteries when the battery icon indicates that the unit
is in low battery warning. When in low battery warning, the lowest battery
level indication bar disappears and the battery icon will be flashing.
Replacing Alkaline Batteries
NOTE: Use Procell PC 1500 alkaline batteries or RKI Instruments, Inc.
lithium-ion battery pack 49-1619 to maintain the CSA classification
of the GX-6000. Use of other batteries will void the CSA
classification and may void the warranty. Do not mix old/new or
different types of batteries.
NOTE: Utiliser Procell 1500 piles alcalines ou RKI Instruments, Inc. pack
batterie lithium-ion 49-1619 de maintenir la classification CSA de la
GX-6000. L’utilisation d’autres piles annule la classification CSA et
peut annuler la garantie. Ne mélangez pas les anciennes/nouvelles
ou différents types de piles.
1 .
Turn off the GX-6000.
WARNING: Do not remove the batteries while the GX-6000 is on.
2 .
Turn the GX-6000 over so that the flow chamber and battery cover are
facing up and the LCD is facing down.
GX-6000 Operator’s Manual
Replacing or Recharging the Batteries • 90
3 .
Push the battery cover latch toward the bottom of the instrument.
Figure 10: Releasing Battery Cover Latch
4 .
Grasp the end of the battery cover that’s closest to the bottom of the
instrument and lift it up.
Figure 11: Opening Battery Cover
5 .
Carefully remove the old alkaline batteries. Verify that the battery
compartment and electrical contacts are clean.
91 • Replacing or Recharging the Batteries
GX-6000 Operator’s Manual
6 .
Carefully install the new AA alkaline batteries according to the battery
diagram inside the battery compartment.
Figure 12: Installing the Alkaline Batteries
7 .
Close the battery cover and secure the battery cover latch by pushing it
up and toward the top of the instrument.
Replacing the Lithium Ion Battery Pack
NOTE: Use Procell PC 1500 alkaline batteries or RKI Instruments, Inc.
lithium-ion battery pack 49-1619 to maintain the CSA classification
of the GX-6000. Use of other batteries will void the CSA
classification and may void the warranty. Do not mix old/new or
different types of batteries.
NOTE: Utiliser Procell 1500 piles alcalines ou RKI Instruments, Inc. pack
batterie lithium-ion de maintenir la classification CSA de la GX-
6000. L’utilisation d’autres piles annule la classification CSA et
peut annuler la garantie. Ne mélangez pas les anciennes/nouvelles
ou différents types de piles.
1 .
Turn off the GX-6000.
WARNING: Do not remove the battery pack while the GX-6000 is on.
2 .
Remove the rubber boot, if installed.
3 .
Turn the GX-6000 over so that the flow chamber and battery cover are
facing up and the LCD is facing down.
4 .
The battery pack release latch is located on the bottom of the instrument
and is closer to the front of the case than the back.
GX-6000 Operator’s Manual
Replacing or Recharging the Batteries • 92
5 .
Slide the battery pack release latch to the right with one hand and lift up
on the battery pack with the other.
Figure 13: Releasing the Battery Pack
6 .
Slide the battery pack away from the instrument.
7 .
Install a new battery pack. If the battery pack release latch moved back
into its closed position, slide the latch to the right with one hand and
insert a new battery pack with the other.
Figure 14: Installing the Battery Pack
8 .
Release the battery pack release latch. If it does not return to its closed
position automatically, push the latch to the left and toward the
instrument until it is flush with the bottom of the instrument.
9 .
Reinstall the rubber boot, if desired.
2.
1.
Lift Up on Battery Pack
Slide Battery Pack Release
Latch to the Right
2.
1. Be Sure Battery Pack Release
Latch is Pushed to the Right
Push New Battery Pack Down
93 • Replacing or Recharging the Batteries
GX-6000 Operator’s Manual
Recharging the Lithium Ion Battery Pack in the
Instrument
CAUTION: To be used only with lithium ion battery pack p/n 49-1619.
Charge only with RKI charger model BC-6000, p/n 49-2182, or
RKI charger model BC-6000DC, p/n 49-2183. Use of other
rechargeable batteries or chargers or charging of other
rechargeable batteries in the GX-6000 will void the warranty.
PRUDENCE:Pour être utilisé uniquement avec une batterie au lithium-ion
p/n 49-1619. Charge uniquement en fonction du modèle de
chargeur RKI BC-6000, p/n 49-2182, ou modèle de chargeur
BC-6000DC RKI, p/n 49-2183. L’utilisation d’autres piles ou
chargeurs rechargeables ou charger d’autres batteries
recharegables dans le GX-6000 annule la garantie.
The charger can be mounted to the wall using DIN rail, if desired. See “DIN
Rail Mounting Assembly” on page 13 for instructions.
1 .
Make sure the GX-6000 is off.
2 .
Confirm that the power adapter’s DC output cord is plugged into the
charger.
3 .
If using an AC powered charging station, plug the AC adapter into an
electrical outlet.
Figure 15: Connecting the Charging Station’s AC Adapter
To AC
Outlet
Power Supply
Jack (on
bottom of
charger)
DC12V
AC Power
Supply
AC Powered Charging Station