Organization of this manual
This manual describes the features and operation of the Vibration Meter VM-83.
To ensure safe use, the manual first lists important safety precautions. Be sure to read
this section thoroughly.
The manual contains the following sections.
Outline
Gives basic information on the configuration and features of the unit, and con-
tains a block diagram.
Controls and Features
Briefly identifies and explains all parts of the unit.
Preparations
Describes installation, connections, inserting batteries, pre-use checks and other
steps.
Measurement
Describes the steps for measurement.
Comparator
Describes how to use the comparator function.
Printer
Gives information on a printer that can be connected to the unit.
Serial Interface
Describes how to use the serial interface for connection to a computer.
Technical Information
Provides additional information that is helpful for using the unit.
Specifications
Lists the technical specifications of the unit.
Index
Index to the contents of the manual.
All company names and product names mentioned in this manual are trademarks or registered
trademarks of their respective owners.
www.64817.com
This item is classified as "a regulated product or regulated technol-
ogy in the Foreign Exchange and Foreign Trade Control Act".
Export of this item from Japan requires an export license from the
Government of Japan.
The product described in this manual is in conformity with the
following European standards;
EN 61000-6-3:2001
EN 61000-6-1:2001
*EN 61000-6-2:2001
Note: CE requirements are met provided that a core filter is fitted
to every cable.
*The measurement result may be influenced when the instrument is
used in a radio-frequency electromagnetic (RFE) field.
To conform to the EU requirement of the Directive
2002/96/EC on Waste Electrical and Electronic
Equipment, the symbol mark on the right is shown
on the instrument.
www.64817.com
In this manual, important safety instructions are specially marked as shown below. To
prevent the risk of death or injury to persons and severe damage to the unit or periph-
eral equipment, make sure that all instructions are fully understood and observed.
Caution
Important
Disregarding instructions
printed here incurs the risk of
damage to the product.
Disregarding instructions
printed here incurs the risk of
injury to persons and/or dam-
age to peripheral equipment.
Mentioned about the tips to use
this unit properly. (This mes-
sages do not have to do with
safety.)
Note
1234567890123456789012345678901212345678901234567890123456789012123456789012
1234567890123456789012345678901212345678901234567890123456789012123456789012
1234567890123456789012345678901212345678901234567890123456789012123456789012
1234567890123456789012345678901212345678901234567890123456789012123456789012
www.64817.com
� Operate the unit only as described in this manual.
� Take care not to drop the unit, and protect it from shocks and vibrations.
� Do not store or use the unit in locations where the unit may be subject to
- splashes of water or high levels of dust,
- air with high salt or sulphur content, or other gases or chemicals,
- high temperature or humidity (above 50°C, 90% RH), or direct sun-
light,
- directly transmitted vibrations or shock.
� Observe the following precautions after using the unit:
- Always switch off the power.
- When the unit is not to be used for an extended period, remove the
batteries to prevent possible damage caused by battery leakage.
� Do not disassemble the unit or attempt internal alterations.
� Have the unit checked and serviced regularly. (Sensitivity calibration
can be performed at the factory for a fee.)
� When powering the unit externally, use only the specified AC adapter
(option). Using a different adapter may cause malfunction or damage.
� Do not tap the LCD panel for example with your finger or a pen, to
prevent possible malfunction or damage.
� In case of malfunction, do not attempt any repairs. Note the condition of
the unit clearly and contact the supplier.
� When disconnecting cables, always hold the plug and do not pull the
cable. Do not apply excessive force.
� When disposing of the unit or the accessories, follow national and local
regulations regarding waste disposal.
www.64817.com
Organization of this manual ....................................................... i
FOR SAFETY .......................................................................... iii
Precautions .................................................................................v
Outline........................................................................................1
Controls and Features.................................................................3
Front Panel ............................................................................3
Rear Panel..............................................................................5
Side Panel ..............................................................................7
Display...................................................................................8
Preparations..............................................................................11
Power Supply.......................................................................11
Inserting Batteries................................................................11
AC Adapter..........................................................................12
Using the Prop-up Feet ........................................................13
Selecting an Accelerometer .................................................14
Connecting the Accelerometer ............................................15
Setting the INPUT SELECT Switch ...................................17
Accelerometer Sensitivity Calibration ................................18
Measurement ............................................................................20
Power-On.............................................................................20
Measurement Mode Setting.................................................21
Display Characteristics Setting............................................22
High-Pass Filter Setting.......................................................23
Low-Pass Filter Setting .......................................................24
Level Range Setting ............................................................25
Menu Settings ......................................................................27
Calibration ...........................................................................31
Maximum Value Hold .........................................................33
Peak Hold ............................................................................36
www.64817.com
Serial Interface .........................................................................47
Transmission Principle ........................................................47
Local Mode/Remote Mode ..................................................48
Transfer Protocol .................................................................51
Error Processing ..................................................................53
Control Operation in Remote Mode ....................................53
Command Format ................................................................54
Communication with Multiple Units using SC-31M/SC-31S ......66
Technical Information ..............................................................69
Noise Level and Measurement Range .................................69
Group Delay ........................................................................73
Rack Mounting ....................................................................73
Accelerometer Installation...................................................74
Display Range .....................................................................76
Aliasing Effect .....................................................................77
Filter Characteristics............................................................78
VM-83PB1 Software ...........................................................80
Specifications ...........................................................................81
Index ........................................................................................93
www.64817.com
The VM-83 is a vibration meter designed for measurement and evaluation of
vibrations, using a piezoelectric accelerometer or a servo accelerometer. It pro-
vides four types of input connectors and allows selection of acceleration, veloc-
ity, and displacement measurement. With the optional servo accelerometer, even
very low frequency vibrations in the range of 0.1 to 1 Hz can be measured,
something that is very hard to achieve with conventional piezoelectric acceler-
ometers.
Display characteristics can be switched to rms, equivalent peak, and equivalent
peak-to-peak. Peak hold and max. hold functions are also available, and a com-
parator with level evaluation output is also available.
AC output, DC output, and a serial interface are provided as standard equip-
ment.
www.64817.com
� � � � � � � � � � � �
� � � �
� � �
� � �
� � � � � � � � � � � �
� � �
� � � � �
� � �
� �
� � � � � � �
� � � �
� � � � � � � �
� � � � �
� � �
� � �
�
� � �
� � �
� �
� � � � � �
� � � � �
� � � � �
�
� � � � � � � �
� � � � � �
� �
� �
� � � � � � � �
� � � � � �
� �
� � � � � �
� �
� � �
� � � � � � � � � � � � � � � �
� � � � � � � � � � � � � � � � � � � � �
� � � � � � � � � � � � � � �
�
� � � � � � �
� � � � � � � � � � � �
� � �
� � � � � � � � � � � �
� � � � � � � � � � � � � �
� � � � � � � � � � � � � � � � � � �
�
� � � � � � � � � � � � � � � � � �
� � � � � � � � � � � � � � � � � � � �
� � � � � � � � � � � �
�
� � � � � � � � � � � � � � � �
� � � � � � � � � � � �
� � � � � � � � � � � � � � � � � � � � � � �
� � � � � � � � � � � � �
� � � � � � �
� � � � � � � � � � � � � �
� � � � � � � � � � � � � � �
� � � � � � � � � � � � � �
� � � � � � � � � � � � � � �
Controls and Features
Front Panel
LEVEL RANGE keys
Serve for switching the level range up (
) or down (
).
Display backlight key
Serves for turning the display backlight on and off.
When the unit is operating on batteries, the display automatically turns itself
off after 60 seconds. When the batteries are run down, the display will be
darker.
Calibration key (CAL key)
This key turns the calibration signal supplied at the AC and DC outputs on
and off. The signal is used to perform calibration of external equipment.
Accelerometer sensitivity key (PICKUP SENS key)
Serves for setting the accelerometer sensitivity. When the accelerometer sen-
sitivity setting mode is on, the indicator above the key lights.
www.64817.com
Controls and Features
MENU key
Allows changing the menu by selecting a menu number. If this key is pressed
in remote control mode, the remote mode is canceled and the unit returns to
the local mode.
Power switch
Serves to turn the unit on and off.
COMP ON/OFF key
Serves to turn the comparator function on and off.
HOLD RESET key
Serves for resetting maximum hold data and peak hold data, as well as the
comparator output.
Low-pass filter key
Serves for activating the low-pass filter (indicator above the key lights).
MAX HOLD ON/OFF key
Serves to turn the maximum value hold function on and off.
High-pass filter key
Serves for activating the high-pass filter (indicator above the key lights).
Accelerometer input connector
The piezoelectric accelerometer is to be connected here.
Setting keys
Serve to set the measurement mode, display characteristics, low-pass filter,
high-pass filter, and accelerometer sensitivity. The keys will affect the func-
tion for which the key indicator is lit. The keys are also used for changing
menu settings.
Measurement mode selector key
Serves to change the measurement mode (indicator above the key lights).
Display characteristics selector key
Serves to change the display (detection) characteristics (indicator above the
key lights).
Display
Shows measurement values and various key settings.
Controls and Features
Rear Panel
Battery holder
Four IEC R14 (size C) batteries are inserted here.
Battery holder button
Push this button to lock or unlock the battery holder.
AC output connector
An AC signal corresponding to the measurement value is available at this
connector (full-scale point: 2 V, output impedance approx. 600 ohms).
DC output connector
A DC signal corresponding to the measurement value is available at this
connector (full-scale point: 2 V, output impedance approx. 600 ohms).
External power supply connector
The AC adapter (option) can be connected here.
Controls and Features
Comparator output terminals
The comparator signal is available at these terminals. The maximum applied
voltage is 24 V, and the maximum drive current varies as follows depending
on the impressed voltage.
50 mA
(when the impressed voltage is 24 V)
25 mA
(when the impressed voltage is 12 V)
10 mA
(when the impressed voltage is 5 V)
Input select switch
Serves to select the input source. Available settings are ACCELEROMETER /
PREAMP 1 / PREAMP 2 / SERVO / SERVO
.
Serial interface connector
Serves for connection of a computer or printer.
Servo input connector
The servo accelerometer (LS-10C or LS-20C) can be connected here.
Preamplifier 2 input connector
A piezoelectric accelerometer with integrated preamplifier can be connected
here.
Preamplifier 1 input connector
A piezoelectric accelerometer can be connected here via a preamplifier.
Controls and Features
Side Panel
� � � �
� � � � � � � � � � � �
� � � �
Controls and Features
� � �
�
� � �
� � �
� � � �
� � � � �
� � �
� � � �
�
� �
� � � �
� � � �
� �
�
� � � � � �
� � � � �
� �
� � �
� � �
�
�
� �
�
�
�
�
�
�
�
�
�
� � �
� � � � � � � � � � �
� � �
� � � �
� � � � � � � � �
� � � � � � � � � � � � � �
� � � � � � � � � � � � � � �
� � � � � � � � �
� � � � � � � � � � � � � � � �
� � � � � � � � � � � � � � � � �
� � � � � � � � �
� �
� � � � � � � � � �
� � � � � � � � � � � � � � � � �
�
� � � � � � � � � � � � � � � �
� � � � �
�
� � � � � � � �
� � � � � � � � � � � � � � �
� � � � � � � � �
� � � � � � � � � � � � � � �
� � � � � � � � � � � � � � � � �
� � � � � � � � �
� � � � � � � � � � � � �
� � � � � � � � � � � �
� � � � � � � � �
Display
For explanation purposes, the illustration below shows all display elements.
In actual use, not all elements will be seen together.
COMP indicator
This indicator appears when the comparator function is used.
REMOTE
This indicator appears when the unit is being controlled remotely over the
serial link.
OVER indicator
This indicator appears when the input signal has caused overload.
Controls and Features
Units indicator
The measurement unit applicable to the current measurement mode is shown
here.
m/s2
When a piezoelectric accelerometer is used, this indicator ap-
pears during acceleration measurement.
mm/s2
When a servo accelerometer is used, this indicator appears
during acceleration measurement.
mm/s
This indicator appears during velocity measurement.
mm
This indicator appears during displacement measurement.
pC/(m/s2)
This indicator appears during sensitivity setting for the pi-
ezoelectric accelerometer.
Low-pass filter indicator
Shows the low-pass filter setting.
High-pass filter indicator
Shows the high-pass filter setting.
RMS/EQ PEAK/EQ P-P indicators
Shows the display (detection) characteristics setting.
Measurement value
Shows the current measurement value, maximum hold value, peak hold value,
menu settings, etc.
PEAK/MAX HOLD/OVER indicators
When the maximum hold function has been activated, the MAX HOLD in-
dicator is displayed.
When the peak hold function has been activated, the PEAK HOLD indicator
is displayed.
When the hold value has exceeded the measurement maximum, the OVER
indicator is displayed.
Bar graph
When the level range setting is 10000, 1000, 100, 10, 1, 0.1, or 0.01, the
indication follows the upper scale:
0, 2, 4, 6, 8, 10.
When the level range setting is 3000, 300, 30, 3, 0.3, 0.03, or 0.003, the
indication follows the lower scale:
0, 1, 2, 3.
The bar graph indication corresponds to the input level.
Controls and Features
� � � � � � � � � � � � � � � � � � � � �
� � � � � � � �
� � � � � � � � � �
� � �
� � � � � � � �
� � � � � � � �
� � � � � � � � �
UNDER indicator
This indicator appears when the input signal has fallen below the measure-
ment threshold.
Battery capacity indicator
Shows the remaining capacity of the batteries.
CAL indicator
This indicator appears during calibration.
Level range indicator
Shows which level range has been selected.
� � �
� � � � � � � � � � � � � �
Preparations
Power Supply
This unit can be operated either on four IEC R14 (size C) batteries or an AC
adapter (option).
Inserting Batteries
Unlock the battery holder by pressing
the battery holder button and pull out
the battery holder. After replacing the
batteries, push the battery holder back
into the unit. Be sure to lock the holder
by pressing the battery holder button
again.
Important
- While inserting or removing batteries, support the battery holder by hand.
- Take care not to insert batteries with wrong polarity.
- While not using the unit, the batteries should be removed.
- When opening the battery holder, take care that no foreign objects
or dirt get deposited in the holder.
- Do not mix old and new batteries, or different types of batteries.
� � � � � � � � �
� � � � � � � � � � � � � �
� � � � � � � � � � � � � � � � � � � � �
� � � � � � � � � � � � � � � � � � � � �
� � � � � � � �
� � � � � � � � �
� � � �
� � � � � � � �
� � � � � � � � � � � � � � � � �
Battery life
Alkaline batteries
LR14
approx. 20 hours
Manganese batteries
R14P
approx. 9 hours
*
The above values were determined under the following conditions.
20°C 50% RH, accelerometer PV-85, ACC, HPF OFF, LPF OFF, back-
light OFF, switch LEDs OFF, communications OFF (MENU 0-0), con-
tinuous operation
*
The actual battery life will differ, depending on the accelerometer, set-
tings, ambient conditions, and battery type.
When the backlight is used, battery consumption will be about 20% higher.
When the servo accelerometer is connected, current consumption increases by
about 15%.
Remaining battery capacity indication
The following indication appears on the display, depending on the battery sta-
tus.
Note
When the battery indicator start to flash, replace the
batteries immediately. While the indicator is flash-
ing, the measurement will not be accurate.
AC Adapter
The unit can be powered from the following adapter: NC-98 (option)
Connect the cable from the AC adapter to the external power supply connector
on the rear panel.
� � � � � � � � � � �
� � � � � � � � � � � �
Using the Prop-up Feet
The unit can be tilted by flipping out the prop-up feet. Tilt the feet until they
snap in place.
Important
When using the prop-up feet, do not place ob-
jects on top of the VM-83 and do not push down
on the unit.
Selecting an Accelerometer
The VM-83 can use piezoelectric accelerometer or servo accelerometer. Select
the type of accelerometer according to the intended measurement range vibra-
tion frequency, environmental conditions, and other related factors.
-
Some major piezoelectric accelerometers made by Rion (representative
values)
General
High-output
Light weight Heat-resistant
3-axis
Model
PV-85
PV-87
PV-90B
PV-65
PV-93
Charge sensitivity
6
40
0.15
7
0.7
(pC/(m/s2) )
Mounting resonance
24
9
70
25
30*2
frequency (kHz) *1
Frequency range (Hz)
1 to 7000
1 to 3000
1 to 25000
1 to 9000
1 to 8000*2
Temperature range (°C) -50 to +160
-50 to +160
-50 to +160
-50 to +260
-50 to +160
Mass (g)
23
115
1.2
26
28
Construction principle
Shear type
Shear type
Shear type
Shear type
Shear type
*1 When mounted with specified screws at specified torque
*2 Vertical direction value
-
Servo accelerometer
The servo accelerometer model that can be used with the VM-83 is the
LS-10C or LS-20C.
Model
LS-10C
LS-20C
Voltage sensitivity (V/(m/s2))
0.300
0.3000
Vibration frequency range
DC to 100 Hz
DC to 100 Hz
Temperature range (°C)
-20 to +60
-20 to +60
Mass (g)
220
125
� � � � �
� � � �
� � � � � � � � � � � � �
� � � � � �
� � � � �
� � � �
� � � � �
� � � �
� � � � � �
� � � �
� � � � � �
� � � � � � � � � � � � � � � � � � � �
� � � � � � � � � � � �
� � � � �
� � � � � � � � � � � � � � �
� �
� � � � � � � � � � � � � �
� � � � � � � � � � � � � � �
� � � � � � � �
� � � � �
� � �
� � �
� � � � �
� � � � �
�
� � � � � � � �
� � � � � �
� � � �
� � � � � � � �
� � � � � �
� � � � � �
� � � � � � � � � � � � � � � � � � � �
� �
� � � � � � � � � � � � � � �
� � � � � � � � � � � � � � �
Connecting the Accelerometer
Note
Before connecting the accelerometer, turn power to
the unit OFF.
Piezoelectric accelerometer
Direct connection
Use the cable supplied with the accelerometer (VP-51 series) to connect the
accelerometer to the ACCELEROMETER input connector on the VM-83.
Because the input circuit is configured as a charge amplifier, accelerometer
cable length has almost no effect on sensitivity.
However, a long cable may lead to an increase in
internal noise. The cable run should therefore be
kept as short as possible.
If the cable length exceeds 5 meters, the use of a
preamplifier is recommended.
Connection via a preamplifier
When the accelerometer is to be installed at a remote location, a preampli-
fier should be used to minimize noise and susceptibility to interference. The
preamplifier should be located close to the ac-
celerometer, and the longer cable run should be
between the preamplifier and the VM-83.
Use the cable supplied with the accelerometer
(VP-51 series) to connect the accelerometer to
the input of the preamplifier (VP-26A etc.).
Connect the output of the preamplifier to the
PREAMP 1 input connector on the VM-83, us-
ing a cable of the EC-02 series.
� � � � �
� � � �
� � � � � � � � � � � � �
� � � � � �
� � � � �
� � � �
� � � � �
� � � �
� � � � � �
� � � �
� � � � � �
� � � � � �
� � � � � � � � � � � � � � � � � � � �
� �
� � � � � � � � � � � � � �
� � � � � � � � � � � � � � �
� � � � �
� � � �
� � � � � � � � � � � � �
� � � � � �
� � � � �
� � � �
� � � � � �
� � � � � �
� � � � � �
� �
� � � � � �
� � � � � � � � � � � � � � � � � � � �
� �
� � � � � � � � � � � � � �
EC-02 series cables are available in six different lengths, as listed below. Choose
the cable that is most suitable for your application.
EC-02:
3 m
EC-02A:
5 m
EC-02B:
10 m
EC-02C:
30 m (with reel)
EC-02D:
50 m (with reel)
EC-02E:
100 m (with reel)
Connection of an accelerometer with built-in preamplifier
When using an accelerometer with built-in pream-
plifier (PV-41, PV-42), use the cable supplied with
the accelerometer (VP-51 series) to connect the ac-
celerometer to the PREAMP 2 input connector on
the VM-83.
Servo accelerometer LS-10C or LS-20C
Connecting the servo accelerometer LS-10C or LS-20C
Use a cable of the EC-40 series to connect the accel-
erometer to the SERVO (LS-10C or LS-20C) input
connector on the VM-83.
EC-40A: 5 m
EC-40B: 10 m
EC-40C: 30 m (with reel)
EC-40D: 50 m (with reel)
Setting the INPUT SELECT Switch
Set the INPUT SELECT switch as required for the type of input that is being
used.
ACCELEROMETER
:
Signal from ACCELEROMETER connector on front panel
PREAMP1
:
Signal from PREAMP 1 connector on rear panel
PREAMP2
:
Signal from PREAMP 2 connector on rear panel
SERVO
:
Signal from SERVO connector on rear panel. Servo accel-
erometer set to vertical axis.
SERVO
:
Signal from SERVO connector on rear panel. Servo accel-
erometer set to horizontal axis.
Note
During calibration, comparator operation, maximum
hold and peak hold, changing the setting of the IN-
PUT SELECT switch has no effect.
Accelerometer Sensitivity Calibration
Important
Different accelerometers have different sensi-
tivity. Be sure to perform sensitivity calibration.
Sensitivity calibration
1. Set the INPUT SELECT switch as required.
2. Set the power switch to ON.
3. Press the ACCELEROMETER SENS key on the front panel. (The indi-
cator above the key lights up and the sensitivity value flashes.)
� � � � � � � � � � � �
� � � �
� � �
� � �
� � � � � � � � � � � �
� � �
� � � � �
� � �
� �
� � � � � � �
� � � �
� � � � � � � �
� � � � �
� � �
� � �
�
� � �
� � �
� �
� � � � � �
� � � � �
� � � � �
�
� � � � � � � �
� � � � � �
� � � �
� � � � � � � �
� � � � � �
� �
� � � � � �
� �
� � �
� � � � � � � � � � � �
� � � � � � � � � � � � � � � �
� � � � � � � � �
� � � � �
� � � � � � � �
4. Use the setting keys (
,
) to set the display indication to the value indi-
cated as charge sensitivity on the calibration chart of the accelerometer.
When an accelerometer with integrated preamplifier is used, enter the voltage
sensitivity. For example, when the voltage sensitivity is 5.90 mV /(m / s2), use
the value 5.90 pC / (m / s2).
For the servo accelerometer LS-10C, enter 0.300.
For the servo accelerometer LS-20C, enter the value according to the
calibration data (approx. 0.300).
When using the preamplifier VP-26A, set the charge sensitivity.
Holding one of the setting keys down for more than 2 seconds causes the
display value to change quickly.
� � � � � � � � � � � � � � � �
� �
�
� � � �
� � � � � � � � � �
� � � �
� � � � � � � � � � � � � � � � � �
� � � �
� � � � � � � � � � � � � � � � � � � � � �
� � � � � � � � � � �
� � �
� � � � � � � �
� � � �
� � � � � � � � � � � �
� � � � �
� �
� �
�
� �
� �
� � � �
� � � � � � � � � � � � � � � � � �
When the unit is turned on, the same settings as used before the unit was turned
off are reestablished (resume function), and measurement starts. The level range,
high-pass filter setting, low-pass filter setting, and display characteristics are
memorized for each measurement mode.
Settings memorized by
Settings reset at regular
resume function
vibration measurement start*
Measurement mode
CAL status
Level range
Maximum value hold
Display characteristics
Peak hold
High-pass filter
Backlight status
Low-pass filter
Sensitivity setting status
Sensitivity
MENU setting status
MENU0 to MENU2 item settings
Comparator status
MENU4 to MENU8 item settings
Remote status
*
These measurement conditions are not memorized by the resume
function. They will be reset to the default vibration measurement con-
dition when the unit is turned on.
Note
Settings made via the serial interface are not memo-
rized for the resume function.
Measurement Mode Setting
Select the measurement mode from ACC (acceleration), VEL (velocity), or DISP
(displacement).
Setting procedure
1. Press the ACC/VEL/DISP key. (The indicator above the key lights
up.)
2. Use the setting keys (
,
) to select from ACC → VEL → DISP.
ACC (acceleration)
: m/s2, mm/s2
VEL (velocity)
: mm/s
DISP (displacement) : mm
Note
The unit for acceleration (ACC) measurement de-
pends on the setting of the INPUT SELECT switch.
ACCELEROMETER, PREAMP 1, PREAMP 2
: m/s2
SERVO , SERVO
: mm/s2
� � �
�
� � �
� � �
� � � �
� � � � �
� � �
� � � �
�
� �
� � � �
� � � �
� �
�
� � � � � �
� � � � �
� �
� � �
� � �
�
�
� �
�
�
�
�
�
�
�
�
�
�
� � � � � � � � � � � � � � � � � � � �
Display Characteristics Setting
Select the display (detection) characteristics from RMS, EQPEAK, or EQP-P.
Setting procedure
1. Press the RMS/PEAK/P-P key. (The indicator above the key lights up.)
2. Use the setting keys (
,
) to select from RMS → EQPEAK → EQP-P.
RMS
: effective value
EQPEAK
: Equivalent peak value
EQP-P
: Equivalent peak-to-peak value
The equivalent peak value and equivalent peak-to-peak value are cal-
culated according to the following formula.
Equivalent peak value = rms ×
�
Equivalent peak-to-peak value = equivalent peak value ×2
� � �
�
� � �
� � �
� � � �
� � � � �
� � �
� � � �
�
� �
� � � �
� � � �
� �
�
� � � � � �
� � � � �
� �
� � �
� � �
�
�
� �
�
�
�
�
�
�
�
�
�
� � � � � � � � � � � � � � � � � � � � � � �
High-Pass Filter Setting
Setting procedure
1. Press the HPF key. (The indicator above the key lights up.)
2. Use the setting keys (
,
) to select the filter setting. The available set-
tings are listed below.
Piezoelectric accelerometer
1 Hz, 3 Hz, 10 Hz, 20 Hz, 50 Hz, -- (OFF; ACC only)
Servo accelerometer
0.1 Hz, 0.3 Hz, 1 Hz
For more information on the characteristics of the high-pass filter, please
refer to the Technical Information.
� � �
�
� � �
� � �
� � � �
� � � � �
� � �
� � � �
�
� �
� � � �
� � � �
� �
�
� � � � � �
� � � � �
� �
� � �
� � �
�
�
� �
�
�
�
�
�
�
�
�
�
� � � � � � � � � � � � � � � �
Low-Pass Filter Setting
Setting procedure
1. Press the LPF key. (The indicator above the key lights up.)
2. Use the setting keys (
,
) to select the filter setting. The available set-
tings are listed below.
Piezoelectric accelerometer
100 Hz, 300 Hz, 1 kHz, 3 kHz, 10 kHz, -- (OFF; ACC only)
Servo accelerometer LS-10C
50 Hz, 100 Hz
For more information on the characteristics of the low-pass filter, please
refer to the Technical Information.
� � �
�
� � �
� � �
� � � �
� � � � �
� � �
� � � �
�
� �
� � � �
� � � �
� �
�
� � � � � �
� � � � �
� �
� � �
� � �
�
�
� �
�
�
�
�
�
�
�
�
�
� � � � � � � � � � � � � � �
Level Range Setting
Set the level range as follows.
Use the setting keys (
,
) to select an appropriate level range.
The relationship between INPUT SELECT switch, accelerometer sensitiv-
ity, measurement mode, and HPF is shown in the table on the next page.
�
�
�
�
�
�
�
�
When INPUT SELECT switch is set to ACCELEROMETER, PREAMP
1, or PREAMP 2
(*0 HPF: 1 Hz; *1 HPF: 3 Hz; *2 HPF: 10 Hz)
Sensitivity
Measurement
Level range
mode
0.030
ACC
3
10
30
100
300
1000
3000
10000
VEL
30
100
300
1000
3000
10000
-
-
0.999
DISP*0
-
10
30
100
300
1000
3000
10000
pC/(m/s2)
DISP*1
3
10
30
100
300
1000
3000
10000
DISP*2
0.3
1
3
10
30
100
300
1000
1.00
ACC
0.3
1
3
10
30
100
300
1000
VEL
3
10
30
100
300
1000
-
-
9.99
DISP*0
-
1
3
10
30
100
300
1000
pC/(m/s2)
DISP*1
0.3
1
3
10
30
100
300
1000
DISP*2
0.03
0.1
0.3
1
3
10
30
100
10.0
ACC
0.03
0.1
0.3
1
3
10
30
100
VEL
0.3
1
3
10
30
100
-
-
99.9
DISP*0
-
0.1
0.3
1
3
10
30
100
pC/(m/s2)
DISP*1
0.03
0.1
0.3
1
3
10
30
100
DISP*2
0.003
0.01
0.03
0.1
0.3
1
3
10
-
When INPUT SELECT switch is set to SERVO or SERVO
Sensitivity Measurement
Level range
mode
0.100
ACC
10
30
100
300
1000
VEL
1
3
10
30
100
0.999
DISP
0.1
0.3
1
3
10
The settings for the serial interface, printer, peak hold function, and comparator
function are made via menus.
1. Call up the menu screen.
Note
If one of the keys CAL, COMP ON/OFF, or MAX
HOLD ON/OFF is set to ON, the menu screen will
not appear.
� � � � � � � � � � � �
� � � �
� � �
� � �
� � � � � � � � � � � �
� � �
� � � � �
� � �
� �
� � � � � � �
� � � �
� � � � � � � �
� � � � �
� � �
� � �
�
� � �
� � �
� �
� � � � � �
� � � � �
� � � � �
�
� � � � � � � �
� � � � � �
� � � �
� � � � � � � �
� � � � � �
� �
� � � � � �
� �
� � �
� � � � � � � � � � � �
� � � � � � � � � � � � �
�
� � � � � � �
� � �
� � � � � � � � � � � � � � � � � �
�
� � � � � � � � � � � � � � � � � � � � � � � �
2. When the MENU key is pressed, a menu screen as shown below appears.
With each push of the MENU key, the menu number cycles through 0 →
1 → 2 → 3 → 4 → 5 → 6 → 7 → 8 → measurement screen.
3. Select the number of the desired menu, and use the setting keys (
,
) to
change the displayed value.
Note
The new setting will become effective when switch-
ing back from the menu screen to the measurement
screen.
Menu contents
Menu 0
Serial interface and printer function ON/OFF
0-0
Serial interface and printer function OFF. Choose this set-
ting when using neither the serial interface nor the printer
function.
0-1
Serial interface ON
0-2
Printer function ON
Menu 1
Serial interface and printer transfer rate
1-96
Transfer rate 9600 bps
When the printer function is set to ON, select the 1-96
setting.
1-19
Transfer rate 19200 bps
� � � � � �
� �
� � �
�
�
�
�
�
� � � � � � � � � �
�
� � � � � � � � � �
� � � � � � �
Serial interface ID number
2-
Choose a setting from 0 to 15.
Menu 3
Peak hold function ON/OFF
3-0
Peak hold function OFF
3-1
Peak hold function ON
Menu 4
Comparator function and comparator level
Sets the comparator function and chooses a level between 0 and 98%
of the full-scale value, in 2% steps.
4-
Choose a setting from 0 to 98 in steps of 2.
Menu 5
Delay time for comparator function
Sets the delay time in the range from 0 to 9 seconds.
5-
Choose a setting from 0 to 9 in steps of 1.
Menu 6
Comparator function auto reset ON/OFF
6-0
Auto reset function OFF
6-1
Auto reset function ON
Menu 7
Auto reset time for comparator function
When auto reset has been set to ON with menu 6, this setting deter-
mines the auto reset time in the range from 0 to 90 seconds, in 1 sec-
ond steps.
7-
Choose a setting from 0 to 90 in steps of 1.
Comparator output buzzer
This controls buzzer use for the comparator function. When enabled,
the buzzer sounds while the comparator output is active. When dis-
abled, there is no buzzer sound also when the comparator output is
active.
8-0
Buzzer disabled
8-1
Buzzer enabled
Note
The settings for menus 0 to 2, and 4 to 8 will be re-
tained also while the unit is turned off.
Menu number
Setting
0-0
Serial interface and printer OFF
0
0-1
Serial interface ON
0-2
Printer ON
1
1-96
9600 bps
1-19
19200 bps
2
2-
ID number setting
0 to 15
3
3-0
Peak hold OFF
3-1
Peak hold ON
4
4-
Comparator level setting 0 to 98 (in 2 steps)
5
5-
Delay time setting
0 to 9
6
6-0
Auto reset OFF
6-1
Auto reset ON
7
7-
Auto reset time setting 0 to 90
8
8-0
Buzzer disabled
8-1
Buzzer enabled
When using external equipment to record the AC output signal or DC output
signal of the VM-83, perform calibration as follows.
In the CAL (calibration) state, the range full-scale signal is output from the AC
output and DC output.
� � � � � � � � � � � �
� � � �
� � �
� � �
� � � � � � � � � � � �
� � �
� � � � �
� � �
� �
� � � � � � �
� � � �
� � � � � � � �
� � � � �
� � �
� � �
�
� � �
� � �
� �
� � � � � �
� � � � �
� � � � �
�
� � � � � � � �
� � � � � �
� � � �
� � � � � � � �
� � � � � �
� �
� � � � � �
� �
� � �
� � � � � � � � � � � � �
� � �
� � � � � � � � � � � � � � � � � �
�
� � � � � � � � � � � � � � � � � � � � � � � �
� � �
� � � � � �
� �
� � �
� � � �
� �
�
�
�
�
�
(1) When INPUT SELECT switch on rear panel is set to ACCELEROM-
ETER, PREAMP 1, or PREAMP 2
AC output (AC OUT) :
80 Hz, 2 V*
DC output (DC OUT) :
2 V
(2) When INPUT SELECT switch on rear panel is set to SERVO or
SERVO
AC output (AC OUT) :
1 Hz, 2 V *
DC output (DC OUT) :
2 V
Pressing the CAL key again turns the calibration mode off.
*
AC output signal in calibration mode
Display characteristics
RMS
:
2 Vrms output
EQPEAK :
2 Vpeak output
EQP-P
:
2 Vp-p output
Note
In calibration mode, all controls except the LIGHT
key, CAL key, and power switch are inactive.
To enable correct output of the calibration signal
when the CAL key is pressed, the following settings
are required.
- COMP ON/OFF
: OFF
- MAX HOLD ON/OFF
: OFF
- Menu 3, Peak hold ON/OFF : OFF
When the range setting is 0.03, 0.3, 3, 30, 300, or
3000, the range full-scale value is 0.0316, 0.3162,
3.162, 31.62, 316.2, or 3162, respectively.
Maximum Value Hold
This function serves for memorizing the maximum measured value.
Setting procedure
1. Press the MAX HOLD ON/OFF key to set the function to ON.
2. When the maximum hold function is ON, the indication MAX HOLD
appears on the display, and the indicated numeric value is the maximum
value measured up to that point. When a higher value occurs, the display
is updated.
� � � � � � � � � � � �
� � � �
� � �
� � �
� � � � � � � � � � � �
� � �
� � � � �
� � �
� �
� � � � � � �
� � � �
� � � � � � � �
� � � � �
� � �
� � �
�
� � �
� � �
� �
� � � � � �
� � � � �
� � � � �
�
� � � � � � � �
� � � � � �
� � � �
� � � � � � � �
� � � � � �
� �
� � � � � �
� �
� � �
� � � � �
�
� �
� � � �
� � � � � �
� �
� � �
�
�
�
�
�
3. When the maximum hold value is higher than the overload threshold, the
indication MAX HOLD OVER is shown.
4. The HOLD RESET key can be used to reset the maximum hold value at
any time.
�
� �
� � � �
� � � � � �
� �
� � �
�
�
�
�
�
� � � �
� � � � � � � � � � � �
� � � �
� � �
� � �
� � � � � � � � � � � �
� � �
� � � � �
� � �
� �
� � � � � � �
� � � �
� � � � � � � �
� � � � �
� � �
� � �
�
� � �
� � �
� �
� � � � � �
� � � � �
� � � � �
�
� � � � � � � �
� � � � � �
� � � �
� � � � � � � �
� � � � � �
� �
� � � � � �
� �
� � �
� � � � � � � � � � � � � �
5. Pressing the MAX HOLD key again turns the maximum hold mode off.
The unit returns to normal measurement.
Note
While the maximum hold function is ON, all con-
trols except the LIGHT key, MAX HOLD ON/OFF
key, HOLD RESET key, and power switch are inac-
tive.
The maximum hold function applies only to the nu-
meric display. The bar graph indication continues to
function as during normal measurement.
The peak hold function uses 51.2 kHz sampling on the vibration input signal to
determine the peak value.
Setting procedure
1. Use the MENU key to call up menu 3.
2. Use the setting keys to select the setting 3-1 (peak hold ON).
3. Return to the measurement screen.
4. When the peak hold function has been activated, the indication PEAK
HOLD appears on the display, and the indicated numeric value is the
peak value measured up to that point. When a higher peak occurs, the
display is updated.
When the peak hold value is higher than the overload threshold, the indi-
cation PEAK HOLD OVER is shown.
5. The HOLD RESET key can be used to reset the peak hold value at any
time.
� �
� � �
�
�
�
�
� � � �
� � � �
�
� � � � � � � � � � � �
� � � �
� �
� � �
�
�
�
�
� � � �
� � � �
�
� �
�
The peak hold function applies only to the numeric
display. The bar graph indication continues to func-
tion as during normal measurement.
While the peak hold function is ON, all controls ex-
cept the LIGHT key, MENU key, HOLD RESET
key, and power switch are inactive.
Turning the peak hold function off
Set the menu 3 item to 3-0.
Note
The peak hold function is available only for ACC
(acceleration) measurement.
The bar graph continues to operate with the display
characteristics as selected before activating peak
hold.
The comparator works by constantly monitoring the measured vibration level
and comparing it to a preset reference level (comparator level). If the compara-
tor level is exceeded, the comparator output becomes active (open collector
circuit, LCD flashing, buzzer sounds).
To turn the comparator function on and off, use the COMP ON/OFF key.
Pressing the COMP ON/OFF key once turns the function on and pressing it
once more turns it off again.
The various settings for the comparator function are made via menu screens
(see "Menu Settings" on page 27).
Setting items
� Comparator level
0 to 98% in 2% steps
� Delay time
0 to 9 seconds in 1 s steps
� Auto reset function
� Auto reset time
0 to 90 seconds in 1 s steps
� Buzzer
� � � � � � � � � � � �
� � � �
� � �
� � �
� � � � � � � � � � � �
� � �
� � � � �
� � �
� �
� � � � � � �
� � � �
� � � � � � � �
� � � � �
� � �
� � �
�
� � �
� � �
� �
� � � � � �
� � � � �
� � � � �
�
� � � � � � � �
� � � � � �
� � � �
� � � � � � � �
� � � � � �
� �
� � � � � �
� �
� � �
� � �
� � � � � � � � � � � �
While comparator function is operating, the COMP indicator is shown.
In the bar graph indication, the bar corresponding to the comparator reference
level remains constantly on.
Comparator level
The comparator level is the threshold where the comparator output becomes
active. This setting is made with menu 4. (See "Menu settings" on page 27.)
For example, if the comparator level is set to 48%, the level corresponding to
48% of the level scale value is the threshold, and the 48% point of the bar graph
scale is on.
� � �
�
� � � � � �
� �
� � �
�
� �
�
�
�
�
�
� � �
Note
While the comparator function is ON, all controls
except the LIGHT key, HOLD RESET key, COMP
ON/OFF key, and power switch are inactive.
Auto reset function
The auto reset function is controlled with menu screen 6 (see "Menu Settings"
on page 27).
When the function is set to ON and the vibration level has fallen again below
the comparator level, the comparator output will automatically be reset (turned
off) after the auto reset time has elapsed. When the function is set to OFF, the
comparator output remains on until the HOLD RESET key is pressed or the
comparator function is turned off with the COMP ON/OFF key.
Auto reset time
The auto reset time is set with menu screen 7 (see "Menu Settings" on page 27).
When the function is set to ON and the vibration level falls below the compara-
tor level, the comparator output will be reset after this time.
The auto reset time can be set from 0 to 90 seconds in 1 second intervals.
This setting is used when the auto reset function has been set to ON.
Buzzer
This setting controls operation of the buzzer when the comparator is activated.
The setting is made with menu screen 8 (see "Menu Settings" on page 27).
� � � � � � � � � � � � � � � �
� � � � � � � � � � � � � � � � � � � � � � � � � � �
� � � � � � � � � �
� � � � � � � � � �
� � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � �
� � � � � � � � � � � � � � � � � � � � � � � � �
� � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � �
� � � � � � � � � � � � � � � � � � � � � � � � � � � � � �
� � � � � � � � � � � � � � � � � � � � � � � �
� � � � � � � � � � � � � � �
�
� � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � �
� � � � � � � � � � � � � � � � � � � � � � � � � � � � � �
When the comparator output was activated, it can be reset in three ways.
(1) Auto reset
As described above, when auto reset is ON, the comparator output
will be turned off automatically after activation when the auto reset
time has elapsed.
(2) HOLD RESET key
Pressing the HOLD RESET key on the front panel immediately resets
the comparator output. This function is independent of the auto reset
ON/OFF setting.
(3) COMP ON/OFF key
Turning the comparator function off by pressing the COMP ON/OFF
key on the front panel immediately resets the comparator output.
With methods (1) and (2), the comparator function remains ON also after reset.
With method (3), the unit returns to normal measurement.
The comparator output has three elements.
(1) Buzzer is heard (long intermittent beeps). This applies only if the buzzer
function has been set to ON with menu 8.
(2) LCD display flashes.
Measurement value indicator section flashes in 0.5 second intervals.
(3) Open collector circuit operates, causing the comparator output terminals
on the rear panel to close.
Circuit block diagram
Maximum drive current:
50 mA
(when impressive volume is 24 V)
25 mA
(when impressive volume is 12 V)
10 mA
(when impressive volume is 5 V)
Maximum applied voltage: 24 V
�
� � �
� � � � �
You can connect an optional printer to produce hard copy of measurement re-
sults. The following printers are compatible with the VM-83:
CP-10, CP-11, DPU-414
Activating printer operation
1. Turn power to the VM-83 and the printer off.
2. Use a separately purchased cable to connect the serial interface connector
on the rear panel of the VM-83 to the printer (option).
Cable:
Straight serial cable (generic)
Connector type on VM-83:
9-pin D-sub, male
Connector type at printer:
25-pin D-sub, female (For DPU- 414, use
the connector adapter supplied with the
printer.)
3. Set the DIP switches on the printer as required, then turn on the printer
and set it to ON LINE.
4. Set the power switch of the VM-83 to ON.
5. Using the menu screens, select the menu settings 0-2 and 1-96. When you
switch back from the menu screen to the measurement screen, printing
starts.
� � � � � � � � � � � � �
� � � � � �
� � �
� � � � � �
� � � � �
� � � �
� � � � � �
� � � � �
� � � � �
� � � � �
� � � �
�
� � �
� � � � � � �
� � � � � � � � � � � � � � � � �
� � � � � � � � �
Printer
Stopping printer operation
1. Using the menu screens, select the menu setting 0-0.
2. Disconnect the cable.
Note
printer operation is paused during calibration, com-
parator operation, sensitivity setting, and while us-
ing the menu screens.
DIP switch settings for printers CP-11, CP-10
Set the DIP switches for the respective printers as shown below.
Note
Switches 7 and 8 of DIP switch bank 2 of printer CP-11
are set individually at the factory before shipping.
�
� �
�
�
�
�
�
�
�
�
� �
�
�
�
�
�
�
�
� � � � � � � � � � � � � � � � � � � � � � � � � � � � � �
� � � � � � � � � � � � � � � � � � � � � � � � � � � � � �
� � � � �
�
� �
�
�
�
�
�
�
�
�
� �
�
�
�
�
�
� � � � � � � � � � � � � � � � � � � � � � � � � � � � � �
� � � � � � � � � � � � � � � � � � � � � � � � � � � � � �
� � � � �
DIP switch settings for printer DPU-414
When connecting the printer DPU-414, set the DIP switches 1 to 3 for the printer
software as follows. For more information on making these settings, please
refer to the instruction manual and the guide sheet of the DPU-414.
Vibration Meter VM-83 (9600 bps)
1
OFF
2
ON
3
ON
SW-1
4
OFF
5
ON
6
OFF
7
ON
8
ON
1
ON
2
ON
3
ON
SW-2
4
ON
5
ON
6
ON
7
ON
8
ON
1
ON
2
ON
3
OFF
SW-3
4
ON
5
OFF
6
ON
7
ON
8
ON
A set of five measurement values for every 2 seconds is printed out every 10
seconds as one line.
" " is used to indicate an OVER occurrence and "#" an UNDER occurrence.
� � � � � � � � � �
� � � � � � � � � � �
� � � � � � � � � � � � � � � � � � �
� � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � �
� � � � � � � � � �
�
� � � �
� � �
� � � �
� � �
� ��
�
� �� �
� � � �
� � � � �
� � � �
� � �
�
� � � � �
� � � �
� � � � �
� � � �
� � � �
�
� � � � �
� � � �
� � � � �
� � � �
� � �
�
� � � �
� � � �
� � � �
� � � �
� � �
�
� � � �
� � � �
� � � � �
� � � �
� � � �
�
� � � � �
� � � �
� � � �
� � �
� � �
�
� � � � �
� � � �
� � � � �
� � � �
� � �
The VM-83 incorporates a serial interface that can be used to set measurement
parameters and control measurement using commands sent from a computer.
Measurement results can also be sent to the computer.
Using SC-31M or SC-31S adapters (option), a single computer can be used to
control up to 16 VM-83 units.
Transmission Principle
Transfer principle
:
asynchronous, half-duplex
Data word length
:
8 bit
Stop bits
:
2
Parity
:
none
Baud rate
:
9600 bps, 19200 bps
Cable type
:
Generic cross-wired serial cable (null modem)
Connector on VM-83 9-pin D-sub, male
(When using an SC-31 adapter, cross-wired cable is
not required.)
Local Mode/Remote Mode
� Local mode
In this mode, the controls on the panel of the VM-83 are used to operate
the unit.
Immediately after being turned on, the unit is always in local mode.
� Remote mode
In this mode, the VM-83 operates in response to commands sent from a
computer. Only the MENU key and LIGHT key on the VM-83 are active,
all other controls are disabled. The indication REMOTE appears on the
display of the VM-83 when the unit is in remote mode.
� Switching between local mode and remote mode
The RMT command is used to switch between local mode and remote
mode.
� � �
� � � � � �
� �
� � �
� � � � � �
� �
� � �
� � � � � �
� �
� � �
� � � � � �
� � � � � �
� � � � � � � �
� �
� � �
� � � � � �
� � � � � � � �
� � � � � � � � � � � � � � � � �
� � � � � � � � �
� � � � � � � � � � � � � � � � � �
� � � � � �
� � � � � � � � � � � � � � � � � � � � � � � �
� � � � � � � � � � � � �
Preparation
1. Turn power to the VM-83 and the computer off.
2. Use a separately purchased cable to connect the serial interface connector
on the rear panel of the VM-83 to the computer.
Cable
:
Generic cross-wired serial cable (null
modem)
Connector type on VM-83:
9-pin D-sub, male
1:1 connection
1:N connection
When wishing to connect multiple VM-83 units (up to 16) to one com-
puter, the optional SC-31M/SC-31S adapters and LAN cables (generic
10BaseT cable) are required.
Connection example
Maximum allowable wiring distance: 400 m
3. Set the power switch of the VM-83 to ON.
4. Make the required menu settings at the VM-83.
For details, please refer to "Menu Settings" on page 27.
Menu 0-1
Menu 1-96 or 1-19
Set transfer rate.
Menu 2-
Set ID number.
Note
ID number
The ID number serves to identify the VM-83 unit
when multiple units are connected to one computer.
In such a setup, the ID number must be set to a
unique setting for each VM-83.
Even if only one VM-83 unit is connected to the
computer, the ID number must be set.
Sending of commands
In order to control the VM-83 from a computer or to retrieve measurement data,
certain commands must be sent to the VM-83. The data exchange must be per-
formed according to certain rules, to ensure that both the VM-83 and the com-
puter recognize the commands and data properly.
To send commands to the VM-83, the following procedure must be observed.
The following explanation assumes that the preparations described on page 49
have been completed.
** stands for the ID number.
Example : If the ID number 0 is set, ** stands for 00.
1. The computer sends RMT1** to the VM-83.
2. When RMT1** has been received, the VM-83 returns
READY to the computer. The link is now estab-
lished, and the indication REMOTE appears on the display of the
VM-83.
3. To change a setting at the VM-83, the computer sends (setting command)
to the VM-83.
4. When a valid setting command is received by the VM-83, it carries out
the respective processing steps. When these are completed successfully,
the VM-83 returns an READY to the computer.
5. To receive data, the computer sends (data request command)
to the VM-83.
6. When the VM-83 receives a command which requests data, it carries out
the respective processing steps and sends (data) to
the computer.
7. To terminate the link, the computer sends RMT0** to the
VM-83.
** is the ID number.
In order to ensure correct data exchange between the VM-83 and the computer,
the rules described above must be observed. If an error occurs, the following
steps should be taken.
-
The computer has sent RMT1** but the REMOTE indicator
on the VM-83 does not light.
Send RMT1** again after about 4 seconds. Repeat this 2
to 3 times. If the REMOTE indicator still does not light, one of the
following conditions may exist:
-
Transfer parameters do not match. (Check settings made with menu
screens.)
-
Interface cable is defective or not properly connected.
-
VM-83 is not powered.
-
ID number setting does not match.
When the computer has sent a wrong command or the parameter is out of range,
the VM-83 disregards that command. If the VM-83 is in receiving mode but
there is no command from the computer for 4 seconds or more, the receive
mode times out.
Control Operation in Remote Mode
While the VM-83 is in remote mode, only the LIGHT key and MENU key are
active. The LIGHT key allows turning the display backlight on and off, and the
MENU key in remote mode serves for switching back to local mode. This al-
lows manual override of remote mode at any time.
Commands that can be used by the VM-83 consist of 3 characters (3 bytes),
usually followed by a parameter which specifies the action range of the com-
mand.
There are two types of parameters:
-
Parameters for changing function settings
-
Parameter for requesting data
The parameters for changing function settings are numerals, and the parameter
for requesting data is "?". In the following command description, the function
setting parameters are denoted by "n". The data output by the VM-83 in re-
sponse to the data requesting parameter are denoted by "p".
Note
The VM-83 cannot process multiple commands sent
together.
� � � �
� � � �
� � � � � � �
� � � � � � � � �
� � � � � � � � � �
RMT n1 n2 n3
Select local mode/remote mode
56
RMT?
Get ID number and remote mode status
56
RNG n
Set level range
56
UNT n
Set measurement mode
57
DET n
Set display characteristics
58
SNS n1 n2 n3 n4
Set sensitivity
58
LPF n
Set low-pass filter setting
59
HPF n
Set high-pass filter setting
59
CAL n
Set calibration mode
60
CAL?
Get calibration mode status
60
BAT?
Get battery capacity status
60
DOD?
Get measurement data
60
DOF n
Output instantaneous value
61
every 0.1 seconds in non-protocol mode
MAX n
Set maximum hold function to On/Off
62
MAX?
Get maximum hold function status
62
PEK n
Set peak hold function to On/Off
62
PEK?
Get peak hold function status
62
RST
Reset hold value
62
CMP n
Set comparator function to On/Off
62
CMP?
Get comparator function status
62
CMS n1 n2 n3 n4 n5 n6 n7
Make comparator function settings
63
CMS?
Get comparator function settings
63
STS?
Get setting information
64
Note
Settings made via the serial interface are not memo-
rized for the resume function (see page 20).
RMT n1 n2 n3
Select local mode/remote mode
n1 = 0
Local mode
n1 = 1
Remote mode
n2 n3 = 00 to 15, FF
Sets ID number. VM-83 specified by ID number receives
the command. FF specifies all connected VM-83.
RMT?
Get ID number and remote mode status
Output data format
p1 p2 p3
p1, p2 is ID number
p3 = 1
Remote mode
RNG n
Set level range
Value of n is determined by INPUT SELECT switch set-
ting, measurement mode, sensitivity information, and HPF
setting, as shown in table below.
When INPUT SELECT is set to ACCELEROMETER, PREAMP1,
PREAMP2
(*0 = HPF 1 Hz, *1 = HPF 3 Hz, *2 = HPF 10 Hz and higher)
Sensitivity
Measurement
n =0
n = 1
n = 2
n = 3
n = 4
n = 5
n = 6
n = 7
mode
0.030
ACC
3
10
30
100
300
1000
3000
10000
VEL
30
100
300
1000
3000
10000
-
-
0.999
DISP*0
-
10
30
100
300
1000
3000
10000
pC/(m/s2)
DISP*1
3
10
30
100
300
1000
3000
10000
DISP*2
0.3
1
3
10
30
100
300
1000
1.00
ACC
0.3
1
3
10
30
100
300
1000
VEL
3
10
30
100
300
1000
-
-
9.99
DISP*0
-
1
3
10
30
100
300
1000
pC/(m/s2)
DISP*1
0.3
1
3
10
30
100
300
1000
DISP*2
0.03
0.1
0.3
1
3
10
30
100
10.0
ACC
0.03
0.1
0.3
1
3
10
30
100
VEL
0.3
1
3
10
30
100
-
-
99.9
DISP*0
-
0.1
0.3
1
3
10
30
100
pC/(m/s2)
DISP*1
0.03
0.1
0.3
1
3
10
30
100
DISP*2
0.003
0.01
0.03
0.1
0.3
1
3
10
-
When INPUT SELECT is set to SERVO , SERVO
Sensitivity Measurement
n = 1
n = 2
n = 3
n = 4
n = 5
mode
0.100
ACC
10
30
100
300
1000
VEL
1
3
10
30
100
0.999
DISP
0.1
0.3
1
3
10
UNT n
Set measurement mode
n = 0
ACC (acceleration)
n = 1
VEL (velocity)
n = 2
DISP (displacement)
Set display characteristics
n = 0
RMS
n = 1
EQPEAK
n = 2
EQP-P
SNS n1 n2 n3 n4
Set sensitivity
n1 to n3 :
Set sensitivity to 030 to 999
n4 = 0
:
Set sensitivity to 1/10 of value specified
with n1 to n3
n4 = 1
:
Set sensitivity to 1/100 of value specified
with n1 to n3
n4 = 2
:
Set sensitivity to 1/1000 of value specified
with n1 to n3
Explanation for command SNS n1 n2 n3 n4
� � � �
� � � �
� � � �
� � � �
� � � �
� � � � �
� � � � � � � � � � �
�
� � � � � � � � � �
�
� � � � � � � � � �
�
� � � � � � � � � �
Set low-pass filter setting
Command action depends in INPUT SELECT switch setting.
Piezoelectric accelerometer
:
ACCELEROMETER, PREAMP1,
PREAMP2
Servo accelerometer
:
SERVO , SERVO
Piezoelectric
Servo
n = 0
100
50
n = 1
300
100
n = 2
1 k
100
n = 3
3 k
100
n = 4
10 k
100
n = 5
OFF
100
n = 6*
user filter
user filter
* This is valid only when the user filter is set up by
using VM-83PB1 software (option).
HPF n
Set high-pass filter setting
Command action depends in INPUT SELECT switch setting.
Piezoelectric accelerometer
:
ACCELEROMETER, PREAMP1,
PREAMP2
Servo accelerometer
:
SERVO , SERVO
Piezoelectric
Servo
n = 0
OFF
0.1
n = 1
1
0.3
n = 2
3
1
n = 3
10
1
n = 4
20
1
n = 5
50
1
n = 6*
user filter
user filter
* This is valid only when the user filter is set up by
using VM-83PB1 software (option).
With piezoelectric accelerometer and measurement mode
VEL or DISP, n = 0 setting is not allowed.
Set calibration mode.
n = 0
Calibration OFF
n = 1
Calibration ON
CAL?
Get calibration mode status
Output data format
p
p = 0
Calibration OFF
p = 1
Calibration ON
BAT?
Get battery capacity status
Output corresponds to status of battery capacity indicator
on display.
Output data format
p
p = 0
Flashing
p = 1
1 segment lit
p = 2
2 segments lit
p = 3
3 segments lit
DOD?
Get measurement data
Returns a value corresponding to the display indication.
Output data format
p1 p2 p3 p4 E ± p5, p6
p1 to p4
Effective 4 digits of measurement value
p5 = -6 to +1
p6 = O
(overload occurred)
p6 = _
(no overload occurred) (_ is a space)
* Explanation of p5
p5 stands for 10p5.
For example, when measurement value is 10000, p5
will be 1, because 1000 × 101
* When operating in maximum hold mode, returned
value corresponds to MAX HOLD value.
Same applies to peak hold mode.
When this command is received while there is no
display indication, output will be p1 to p4 = 0000.
DOF n
Output instantaneous value every 0.1 seconds in non-pro-
tocol mode
n = 0
Instantaneous value output OFF
n = 1
Instantaneous value output ON
Data will be output every 0.1 seconds. When DOF 0 is
received, non-protocol mode output is terminated.
For information on output data format, see section on
"DOD?".
DOF communication format
Note
Because this unit supports only half-duplex commu-
nication, it cannot receive and send data at the same
time. When sending the DOF 0 command to the unit,
the computer must use timing which ensures that the
command does not collide with the instantaneous
value data that are being output by the unit.
� � � � � � � � � � � � �
� � � � � � � � � � � � �
� �
� � �
� � � � � � � �
� � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � �
� � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � �
� � � � � � � � � � � � � � � � �
� � � � � � � � � � � � � � � � � � �
� � � � � � � � � � � � � � � � � � �
Set maximum hold function to On/Off
n = 0 Maximum hold OFF
n = 1 Maximum hold ON
MAX?
Get maximum hold function status
Output data format
p
p = 0
Maximum hold OFF
p = 1
Maximum hold ON
PEK n
Set peak hold function to On/Off
n = 0
Peak hold OFF
n = 1
Peak hold ON
PEK?
Get peak hold function status
Output data format
p
p = 0
Peak hold OFF
p = 1
Peak hold ON
RST
Reset hold value
This command is active during maximum hold, peak hold,
and comparator operation.
CMP n
Set comparator function to On/Off
n = 0
Comparator OFF
n = 1
Comparator ON
CMP?
Get comparator function status
Output data format
p
p = 0
Comparator OFF
p = 1
Comparator ON
CMS n1 n2 n3 n4 n5 n6 n7
Make comparator function settings
n1 to n2 :
Comparator level 00 to 98 (steps of 2)
n3
:
Delay time 0 to 9
n4 = 0
:
Auto reset OFF
n4 = 1
:
Auto reset ON
n5 to n6 :
Auto reset time 00 to 90 (steps of 10)
n7 = 0
:
Buzzer OFF
n7 = 1
:
Buzzer ON
For items that are not to be changed, send _ (space).
CMS?
Get comparator function settings
Output data format
p1 p2 p3 p4 p5 p6 p7
p1 to p2
:
Comparator level 00 to 98
p3
:
0 to 9 (delay time)
p4
:
0 to 1 (auto reset ON/OFF)
p5 to p6
:
00 to 90 (auto reset time)
p7
:
0 to 1 (buzzer ON/OFF)
Get setting information
Output data format
p1 p2 p3 p4 p5 p6 p7 p8 p9 p10
p1
:
INPUT SELECT 0 to 4
p1 = 0
:
ACCELEROMETER in-
put
p1 = 1
:
PREAMP 1 input
p1 = 2
:
PREAMP 2 input
p1 = 3
:
SERVO input
p1 = 4
:
SERVO
input
p2
:
Level range 0 to 7 (RNG n)
p3
:
Measurement mode 0 to 2 (UNT n)
p4
:
Display characteristics 0 to 2 (DET n)
p5 to p8
:
Sensitivity (SNS n1 n2 n3 n4)
p9
:
High-pass filter setting (HPF n)
p10
:
Low-pass filter setting (LPF n)
For information on value in brackets, refer to sections on
respective commands.
Allowable remote mode commands in various operation conditions
� indicates that command is allowed (valid).
× indicates that command is not allowed (invalid).
Command
Normal
Calibration
Comparator
Maximum
Peak
DOF
Explanation
measurement
hold
hold
operation
RMT n1n2n3
�
�
�
�
�
�
Cancel remote mode
RMT?
�
×
×
×
×
×
RNG n
�
×
×
×
×
×
Level range
UNT n
�
×
×
×
×
×
Measurement mode
DET n
�
×
×
×
×
×
Display characteristics
SNS n1 to n4
�
×
×
×
×
×
Set sensitivity
LPF n
�
×
×
×
×
×
Low-pass filter
HPF n
�
×
×
×
×
×
High-pass filter
CAL n
�
�
×
×
×
×
Calibration setting
CAL?
�
�
×
×
×
×
BAT?
�
×
×
×
×
×
Battery status
DOD?
�
×
�
�
�
×
Measurement value
DOF n
�
×
×
×
×
�(DOF0)
Instantaneous value
(100 ms)
MAX n
�
×
×
�(MAX0)
×
×
Maximum hold
MAX?
�
×
×
�
×
×
PEK n
�
×
×
×
�(PEK0)
×
Peak hold
PEK?
�
×
×
×
�
×
RST
×
×
�
�
�
×
Reset
CMP n
�
×
�(CMP0)
×
×
×
Comparator
CMP?
�
×
�
×
×
×
CMS n1 to n7
�
×
×
×
×
×
Comparator setting
CMS?
�
×
×
×
×
×
STS?
�
×
×
×
×
×
Setting information
� � �
� � � � � �
� �
� � �
� � � � � �
� �
� � �
� � � � � �
� �
� � �
� � � � � �
� � � � � �
� � � � � � � �
Communication with Multiple Units using SC-31M/SC-31S
Using the optional SC-31M/SC-31S adapters, it is possible to connect up to
16 VM-83 units to a single computer.
Connecting multiple VM-83 units (max. 16) to a computer
Use SC-31M/SC-31S adapters (option) and LAN cables (generic 10BaseT
cable)
Maximum allowable wiring distance: 400 m
Note
(1) When multiple units are connected, commands
can be sent either to a single unit or all units. It is
not possible to send for example a command to
3 out of 5 connected units.
(2) When sending commands to all units, only set-
ting commands are valid. Commands for get-
ting information (parameter "?") and the DOF
command are invalid.
Sending a command to a single unit
To send a command to a single unit when multiple units are connected, the
same procedure as for one-on-one communication is used.
1. Verify that all connected VM-83 units are set to local mode.
2. Send the RMT1** command, where ** is the ID number of
the unit to be controlled.
3. Control the specified VM-83 unit by sending commands for changing
settings and getting information.
4. To switch control to another VM-83 unit, send the RMT0**
command with the ID number of the currently controlled unit, to return
that unit to the local mode. Then send the RMT1** command
with the ID number of the new unit, and repeat the above steps.
Note
Before switching control to a new VM-83 unit, always return
the currently controlled unit to the local mode first. If two or
more units are set to remote mode, output data will be mixed
up.
� � � � � � � �
� � � �
� � � � � � �
� � � �
� � � � � � �
� �
� � �
� � � � � � � � � � �
� �
� � �
� � � � � � � � � �
� �
� � �
� � � � � � � � � � �
Sending a command to all connected units
When sending commands to all connected units, only commands for changing
settings are allowed. To address a command to all units, set all units to remote
mode by sending the RMT1FF command. Note that a VM-83 that has received
the RMT1FF command will not output a response to any command.
1. Verify that all connected VM-83 units are set to local mode.
2. Send the RMT1FF command to set all connected VM-83
units to remote mode (linked condition).
3. Control the VM-83 units by sending setting commands.
4. To terminate the linked condition, send the RMT0FF com-
mand.
Note
A VM-83 that has received the RMT1FF
command will switch to remote mode, regardless of
the ID number. When the parameter FF was used,
the VM-83 will not output a response to received
commands.
� �
� � � � � � � � � � � � �
� �
� � � � � � � � � � � � �
� �
� � �
� � � � � � � �
� � � � � � � � � � � � � � � � � � � � � � � � � �
� � � � � � � � � � � � � �
� � � � � � � � � � � � � � � � � � � � � � � � � �
� � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � �
Noise Level and Measurement Range
(1) Noise level using accelerometer input and sensitivity 5.00 pC/(m/s2)
Measurement
Measurement
HPF
LPF
Display
Noise level
mode
range
characteristics
Acceleration
0.3
---
---
RMS
0.004 m/s2
Velocity
3
1 Hz
---
RMS
0.1 mm/s
Displacement
1
1 Hz
---
RMS
0.015 mm
Displacement
0.03
10 Hz
---
RMS
0.0003 mm
(2) Noise level examples using servo accelerometer input
Accelerometer type
Measurement
Measurement
HPF
LPF
Display
Noise level
mode
range
characteristics
Acceleration
0.3
---
---
RMS
0.0034 m/s2
PV-85
Velocity
3
10 Hz
---
RMS
0.004 mm/s
Displacement
0.03
10 Hz
---
RMS
0.0002 mm
Acceleration
3
---
---
RMS
0.133 m/s2
PV-90B
Velocity
30
10 Hz
---
RMS
0.17 mm/s
Displacement
0.3
10 Hz
---
RMS
0.007 mm
Acceleration
0.03
---
---
RMS
0.0005 m/s2
PV-87
Velocity
0.3
10 Hz
---
RMS
0.0006 mm/s
Displacement
0.003
10 Hz
---
RMS
0.00003 mm*
*
The LCD can display measurement values to four digits after the decimal
point.
(3) Measurement range with servo accelerometer LS-10C
Acceleration measurement range
� � �
� � �
� � �
�
�
�
� �
� �
� �
� � �
� � �
� � �
� � � �
� � � �
� � � �
� � � �
� � � �
� � � �
� � �
� � �
� � �
�
�
�
� �
� � � �
� � � �
� � � �
� �
� �
� � �
� � �
� � �
� � � �
� � � � �
� � � � � � � � � � � � � � � � � � � � � � � �
� � � � � � � � � � � � � �
Displacement measurement range
� � �
� � �
� � �
�
�
�
� �
� � � �
� � � �
� � � �
� �
� �
� � �
� � �
� � �
� � � �
� � �
� �
� �
� �
�
�
�
� � �
� � �
� � �
� � � �
� � � �
� � � �
� � � � �
� � � � �
� � � � �
� � � � � �
� � � � � �
� � � � � �
� � � � � � � �
� � � � � � �
� � � � � � � � � �
� � � � � � � � � � � � � � � � � � � � �
� � � � � � � � � � � � � �
� � � �
� � � �
� � �
� �
� � � � �
� � � � � � � �
� �
� � � � � � � � � �
Group Delay
The VM-83 directly converts the vibration input signal from analog into digital
form and performs signal processing in the digital domain, using a DSP. The
signal is then reconverted to analog for the AC and DC output. For this reason,
a certain delay (group delay) is introduced as shown below.
Rack Mounting
The VM-83 can be mounted in a rack, using the four screw holes on the bottom
of the unit.
� � � � � � � � � � � � � � � � � � � � � �
� � � � � � � � � � � � � � � �
� � � � � � � � � � � � � �
The accelerometer can be mounted to the measurement object in one of the four
general ways outlined below. The accelerometer mounting method greatly af-
fects the contact resonance frequency *. The advantages and disadvantages of
various methods are described in this section.
Rigid screw mounting
This mounting principle assures optimum frequency re-
sponse characteristics. The mounting surface should be per-
fectly smooth, and the screw holes (M6) must be drilled in
such a way so as not to disturb perpendicularity of the ac-
celerometer.
Insulation attachment mounting
This mounting principle is used when electrical insulation
between accelerometer and measurement object is required.
The contact resonance frequency will be lower than with
rigid screw mounting. Regarding the screw holes, the same
precautions as for rigid screw mounting apply.
Magnet attachment mounting
Because the contact resonance frequency will be quite low,
this principle is mainly suited for vibration measurements
in the medium to low frequency range. The maximum ac-
celeration that can be measured depends on the accelerom-
eter mass.
Rod attachment mounting
Pressing the accelerometer against the measurement object
with a rod is the simplest method, but it is only suitable for
measurements below 500 Hz, because contact resonance
frequency will be very low. This method should only be
used if the other three mounting methods are not feasible.
* Contact resonance frequency
Contact resonance occurs when the area the mounted accelerometer con-
tacts the measurement object becomes temporarily deformed, causing it to
act like a spring. This spring and the accelerometer mass then form a system
that vibrates at a certain resonance frequency.
The resonance frequency, which varies considerably depending on the mount-
ing method of the accelerometer, limits the upper range of vibration frequen-
cies that can be measured.
The diagram below shows the relationship between mounting method and
high range characteristics. This demonstrates that, besides selecting an ap-
propriate mounting method, it is important to set the measurement range so
that the influence of the contact resonance phenomenon is excluded.
Sample chart showing the influence of accelerometer mounting method on
high-range frequency characteristics
This section provides information on the numeric range of the digital display
used in the VM-83.
For simplification, the range is assumed to be "100" and "30", and the display
characteristics setting EQ PEAK. Because the overload point of this unit is
approximately three times the range full-scale point, the display range at the
range "100" setting is about 300, and at the "30" setting (actually 31.6) is about
90, assuming that there is no overload in the intermediate circuits. The display
resolution is 0.1 and 0.01. Therefore the possible display range is as follows:
Range
100
0.1 to 300.0
30
0.01 to 90.00
The VM-83 directly feeds the raw waveform of the vibration signal to an A/D
converter and then performs filter processing, rms conversion, and other func-
tions using a DSP. When the number of effective digits for display decreases,
display accuracy cannot be maintained. Therefore the indication UNDER ap-
pears on the display when the displayed value falls below 0.5% of the full-scale
point. This warns the user that the current range setting is not appropriate. To
assure accurate readings, the range setting should be made so that the UNDER
indication does not appear. At a level of 0.2% of the full-scale point and lower,
the lowest bit level of the A/D converter is reached, so that readings will be
indeterminate. At a range "30" setting, this corresponds to 0.06% of the full-
scale point. This means that readings below this limit will be highly imprecise,
and that the display reading may not be 0 even if the actual vibration signal is
very low.
The VM-83 directly converts the vibration input signal from analog into digital
form and performs signal processing in the digital domain, using a DSP. This
principle is subject to the so-called aliasing effect. An anti-aliasing filter (ana-
log low-pass filter) serves to counter this effect. In the VM-83, a 5th-order filter
is used for this purpose. Note that the filter has the effect shown below.
In the above illustration, the shaded section represents the area where the aliasing
effect is noticeable. With fs/2 as the center, the two spectrum representations
look like mirror reflections of each other.
The arrow (solid line) indicates a case where a signal 1.56 times the 20 kHz
upper measurement limit is input (31.2 kHz). In this case, aliasing will cause a
spectrum level detection of -50 dB (0.3%) at 20 kHz.
� � � �
� � � � � �
� �
� � � � � �
� � � � � � �
� �
�
�
� � � � � � � � � � � � � � � �
� � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � �
� � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � �
� � � � � � � � � � � �
The above specifications are a representative example for AC output.
�
�
�
� �
� �
� �
� � �
� �
�
� � �
� � �
� � �
� � �
� � �
� � � �
� � � �
� � � �
� � �
� � �
� � �
� � �
� � � �
� � � �
� � � � �
� � � � �
� � � � �
� � � � � � � � � � � � �
� � � � � � � � � � � � � �
� �
� �
� �
� � �
� � �
� � �
� � �
� � �
� � �
� � � �
� � � �
� � � �
� � � � �
� �
�
� � �
� � �
� � �
� � �
� � �
� � � � � �
� � � � � �
� � � � �
� � � � �
� � � � � �
� � �
� � � � � � � � � � � � �
� � � � � � � � � � � � � �
Servo accelerometer LS-10C
High-pass filter
Low-pass filter
The above specifications are a representative example for AC output.
�
�
�
� �
� �
�
� � �
� � �
� � �
� � �
� � �
� � � �
� � � �
� � � �
� � �
� � �
� � �
� � � � � �
� � � �
� � � � � �
� � � � � � � � � � � � �
� � � � � � � � � � � � � �
� �
� �
� �
� � �
� � �
� � �
� � �
� � �
� � �
� � � �
� �
�
� � �
� � �
� � �
� � �
� � �
� � � � �
� � � � � �
� � � � � � � � � � � � �
� � � � � � � � � � � � � �
VM-83PB1 is a software package for Microsoft Windows 95, 98, 98SE, Me,
2000, XP and Microsoft Windows NT4.0 which allows controlling settings
and measurement operation of the VM-83 from the computer, via the serial
port. Measurement data downloaded from the VM-83 can be stored in text
file format, allowing further processing for example using a spreadsheet pro-
gram. The VM-83PB1 software also handles communication with multiple
VM-83 units connected to a single computer, using the SC-31 adapter.
Microsoft and Windows are trademarks or registered trademark of Microsoft
U.S.A. Corporation in the U.S.A. and other countries.
�
� �
� � � � � � � � �
� � �
� � � � � � �
� � � � � � �
� � � � �
� � � �
� � � � � � � � �
� � � � �
� � � � � � � � � �
� � � � �
� � � � �
� � � � � �
� � � � � �
� � � � �
� � � � � � � � � � � � �
� � � � � � � � � � � � �
� � �
� � � �
� � � �
�
� � �
� � �
� � � � �
� �
� � � �
� �
�
� � � � � � �
� � � � � � � � �
� � � � � � � � � � � � �
� � �
� � � � � � � � � � � �
� � � �
� � � �
� � � � � � � � � � � � �
� � �
� � �
� � � � �
� � � � � � � � � � � �
� � � � � � � � � �
� � � �
� � � � � � � � � � � � � � � � � �
� � � � �
�
� � � � �
� � � � � � �
� � � �
� � � � � �
� � � � � � �
� � �
� � � � �
�
� � � � � � � � � � � � � � � � � �
� � � � �
�
� � � � �
� � � � � � �
� � � �
� � � � � �
� � � � � � �
� � �
� � � � �
�
� � � � � � � � � � � � � � � � � �
� � � � �
�
� � � � �
� � � � � � �
� � � �
� � � � � �
� � � � � � �
� � �
� � � � �
� � � �
� � � � � � � � � � � � �
�
� � � � � � � � � �
� � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � �
� � � � � � � � � � � �
� � � � � � �
� � �
� � �
�
� � � � � �
� � �
� � �
� � � � � � � � � � �
� � � � � � �
� �
� � � � � �
(1) ACCELEROMETER input
For piezoelectric accelerometer (microdot
connector)
Maximum input charge 30000 pC
(2) PREAMP 1 input
For piezoelectric accelerometer via preamplifier
VP-26A (7-pin female connector PRC-03)
(3) PREAMP 2 input
For piezoelectric accelerometer with integrated
preamplifier
18 V, 2 mA drive (microdot connector)
(4) SERVO input
For servo accelerometer LS-10C or LS-20C
(8-pin female connector)
Measurement modes
(1) Acceleration (ACC) m / s2 (piezoelectric accelerometer), mm / s2 (servo
accelerometer)
(2) Velocity (VEL)
mm / s
(3) Displacement (DISP) mm
Measurement range
Piezoelectric accelerometer
-
Accelerometer sensitivity 0.030 to 0.999 pC / (m / s2)
Acceleration
3, 10, 30, 100, 300, 1000, 3000, 10000 m / s2
Velocity
30, 100, 300, 1000, 3000, 10000 mm / s
Displacement
10, 30, 100, 300, 1000, 3000, 10000 mm
(HPF 1 Hz)
Displacement
3, 10, 30, 100, 300, 1000, 3000, 10000 mm
(HPF 3 Hz)
Displacement
0.3, 1, 3, 10, 30, 100, 300, 1000 mm
(HPF 10 Hz or higher)
Accelerometer sensitivity 1.00 to 9.99 pC/(m/s2)
Acceleration
0.3, 1, 3, 10, 30, 100, 300, 1000 m/s2
Velocity
3, 10, 30, 100, 300, 1000 mm/s
Displacement
1, 3, 10, 30, 100, 300, 1000 mm (HPF 1 Hz)
Displacement
0.3, 1, 3, 10, 30, 100, 300, 1000 mm (HPF 3 Hz)
Displacement
0.03, 0.1, 0.3, 1, 3, 10, 30, 100 mm (HPF 10 Hz
or higher)
-
Accelerometer sensitivity 10.0 to 99.9 pC/(m/s2)
Acceleration
0.03, 0.1, 0.3, 1, 3, 10, 30, 100 m/s2
Velocity
0.3, 1, 3, 10, 30, 100 mm/s
Displacement
0.1, 0.3, 1, 3, 10, 30, 100 mm (HPF 1 Hz)
Displacement
0.03, 0.1, 0.3, 1, 3, 10, 30, 100 mm (HPF 3 Hz)
Displacement
0.003, 0.01, 0.03, 0.1, 0.3, 1, 3, 10 mm
(HPF 10 Hz or higher)
Servo accelerometer
Acceleration
10, 30, 100, 300, 1000 m/s2
Velocity
1, 3, 10, 30, 100 mm/s
Displacement
0.1, 0.3, 1, 3, 10 mm
Vibration frequency range
Piezoelectric accelerometer
Acceleration
1 Hz to 20 kHz ±5%
(AC output: 15 kHz to 20 kHz
)
Velocity
1 Hz to 3 Hz ±10%, 3 Hz to 3 kHz ±5%
Displacement
1 Hz to 3 Hz ±20%, 3 Hz to 500 Hz ±10%
Servo accelerometer
Acceleration
0.1 Hz to 100 Hz ±5%
Velocity
0.1 Hz to 0.3 Hz ±10%, 0.3 Hz to 100 Hz ±5%
Displacement
0.1 Hz to 0.3 Hz ±20%, 0.3 Hz to 100 Hz ±10%
OFF, 1, 3, 10, 20, 50 Hz (-10% point), selectable
LPF
OFF, 100, 300, 1 k, 3 k, 10 kHz (-10% point), selectable
Servo accelerometer
HPF
0.1, 0.3, 1 Hz, selectable
LPF
50, 100 Hz, selectable
Display (detection) characteristics
Effective value (RMS)
True rms
Equivalent peak (EQ PEAK)
rms ×
�
Equivalent peak-to-peak (EQ P-P) rms ×2
Maximum value hold
Maximum display value for selected measurement mode and display
characteristics is held (numeric display only). Reset key allows reset at any
time.
Peak value hold
Peak value of acceleration waveform is held.
Comparator function
Level evaluation based comparator. Output activated when comparator level
is exceeded
-
Comparator level setting: in 2% steps of full-scale value
-
Delay time
0 to 9 s in 1-s steps
-
Auto reset time
0 to 90 s in 1-s steps, ON/OFF
-
Comparator output
open-collector output
max. applied voltage: 24 V,
max. drive current: 50 mA (when the impressed voltage is 24 V)
25 mA (when the impressed voltage is 12 V)
10 mA (when the impressed voltage is 5 V)
buzzer output ON/OFF; LCD flashing
(1) Bar graph, linear scale, 100 ms sampled value
0 to 3.16
0 to 10
(2) 4-digit numeric display; arithmetic average of 20 instantaneous values
taken at 100 ms intervals, updated every 2 seconds
(3) Measurement mode, display characteristics, filter
(4) Battery status; 3-segment indication
Calibration
(1) Accelerometer sensitivity
0.030 to 0.999, 1.00 to 9.99, 10.0 to 99.9 pC/(m/s2)
Shown on display during calibration
(2) Output calibration
Calibration output signal for connected equipment
AC
Piezoelectric accelerometer
80 Hz ±2%, 2 V ±2%
Servo accelerometer
1 Hz ±2%, 2 V ±2%
DC 2 V ±2%
� � �
�
� � �
� � �
� � � �
� � � � �
� � �
� � � �
�
� �
� � � �
� � � �
� �
�
� � � � � �
� � � � �
� �
� � �
� � �
�
�
� �
�
�
�
�
�
�
�
�
�
� � � � � � � � � � � � � � � � � � � � � � � � � � �
(1) AC output; full-scale 2 V, output impedance 600 Ω, BNC connector
Load impedance: 10 kΩ or higher
Output voltage accuracy
Piezoelectric (unit electrical characteristics, 80 Hz)
Velocity:
range full-scale ±2%
Acceleration:
range full-scale ±3%
Displacement:
range full-scale ±5%
Servo accelerometer (overall accuracy with LS-10C 1 Hz)
Velocity:
range full-scale ±3%
Acceleration:
range full-scale ±4%
Displacement:
range full-scale ±6%
Max. output voltage: ±6 V peak
(2) DC output; full-scale 2 V, output impedance 600 Ω, BNC connector
Load impedance: 10 kΩ or higher
Output voltage accuracy
Piezoelectric (unit electrical characteristics, 80 Hz)
Velocity:
range full-scale ±2%
Acceleration:
range full-scale ±3%
Displacement:
range full-scale ±5%
Max. output voltage: 6 V (4.2 V when rms display is selected)
(1) Noise level with accelerometer input, sensitivity 5.00 pC/(m/s2)
Measurement
Measurement
HPF
LPF
Display
Noise level
mode
range
characteristics
Acceleration
0.3
---
---
RMS
0.004 m/s2
Velocity
3
1 Hz
---
RMS
0.1 mm/s
Displacement
1
1 Hz
---
RMS
0.015 mm
Displacement
0.03
10 Hz
---
RMS
0.0003 mm
(2) Noise level example with piezoelectric accelerometer
Accelerometer type
Measurement
Measurement
HPF
LPF
Display
Noise level
mode
range
characteristics
Acceleration
0.3
---
---
RMS
0.0034 m/s2
PV-85
Velocity
3
10 Hz
---
RMS
0.004 mm/s
Displacement
0.03
10 Hz
---
RMS
0.0002 mm
Acceleration
3
---
---
RMS
0.133 m/s2
PV-90B
Velocity
30
10 Hz
---
RMS
0.17 mm/s
Displacement
0.3
10 Hz
---
RMS
0.007 mm
Acceleration
0.03
---
---
RMS
0.0005 m/s2
PV-87
Velocity
0.3
10 Hz
---
RMS
0.0006 mm/s
Displacement
0.003
10 Hz
---
RMS
0.00003 mm*
*
The LCD can display measurement values to four digits after the decimal
point.
(3) Measurement range with servo accelerometer LS-10C
Acceleration measurement range
� � �
� � �
� � �
�
�
�
� �
� �
� �
� � �
� � �
� � �
� � � �
� � � �
� � � �
� � � �
� � � �
� � � �
� � �
� � �
� � �
�
�
�
� �
� � � �
� � � �
� � � �
� �
� �
� � �
� � �
� � �
� � � �
� � � � �
� � � � � � � � � � � � � � � � � � � � � � � �
� � � � � � � � � � � � � �
Displacement measurement range
� � �
� � �
� � �
�
�
�
� �
� � � �
� � � �
� � � �
� �
� �
� � �
� � �
� � �
� � � �
� � �
� �
� �
� �
�
�
�
� � �
� � �
� � �
� � � �
� � � �
� � � �
� � � � �
� � � � �
� � � � �
� � � � � �
� � � � � �
� � � � � �
� � � � � � � �
� � � � � � �
� � � � � � � � � �
� � � � � � � � � � � � � � � � � � � � �
� � � � � � � � � � � � � �
Ambient conditions for operation
-10 to +50°C, 20 to 90% RH (no condensation)
Power requirements
IEC R14 (size C) batteries(×4) or AC adapter (option)
Current consumption
approx. 190 mA (6 V DC) (*1)
approx. 60 mA (using AC adapter) (*1, *2)
Battery life for continuousoperation
Alkaline batteries (approx. 20 hours) (*1)
Manganese batteries (approx. 9 hours) (*1)
*1
Measurement conditions:
20°C, 50% RH, power supply voltage 6 V, Accelerometer input,
ACC, backlight off, communication off, switch LEDs off
*2
Measurement conditions:
Current consumption measured with AC adapter at 100 V AC.
Dimensions
171 (H) × 120 (W) × 234 (D) mm
Weight
Approx. 1.8 kg
Interface
Serial interface
For data output and remote control
5WKR4030 interface cable (option)
9-pin D-sub male connector
Half-duplex communication with protocol
Transfer rate 9600, 19200 bps
SC-31M/SC-31S (option) allow connection of multiple VM-83 units
(max.16) to one computer. Maximum allowable wiring distance: 400 m
Printer output
For data output to printer (CP-10, CP-11, DPU-414)