Installation & Operating
Guide
Models 4680 & 4685
EXFG Oxygen Transmitters
4600
4600
A1
A2
% Oxygen
20.9
% Oxygen
25.0
A1
A2
ABB 4680 We are an established world force in the design and manufacture of instrumentation for industrial
process control, flow measurement, gas and liquid analysis and environmental applications.
As a part of ABB, a world leader in process automation technology, we offer customers
application expertise, service and support worldwide.
We are committed to teamwork, high quality manufacturing, advanced technology and unrivalled
service and support.
T
ABB 4680 Controls & Displays
Operation
Programming
I
I
I
I
I
I
Part No.
IM/EXFG–4600
INTERFACE UNIT
Certification
Product Identification
Mechanical Installation
Electrical Installation
Calibration
Fault Finding
I
I
I
I
I
I
Part No.
IM/EXFG–INT
PROBE
I
I
I
I
I
I
I
Part No.
IM/EXFG–PB
SPECIFICATION SHEET
Full Specification
I
Part No.
SS/EXFG
Certification
Product Identification
NON HAZARDOUS AREA
HAZARDOUS AREA
(Zone 1, Class IIB)
Process
20°C to
600°C
1.1 Bar
Absolute
Maximum
(40in WG)
EXFG Probe
EEx d IIB T3 (Tamb –20°C to 55°C)
to EN30014 1993 and EN30018 1995
Mains Supply
Control Thermocouple
Signal
4 to 20mA Output Signal
(25% to 0.25% O2)
or
4680 Transmitter
Mains Supply
Flameproof Interface Electronics Unit
EEx d IIB T6 Conforming to BS5501 Part 5: 1977
Flue
4685 Transmitter
Retransmission
Output – see
Sectio
out of case
(Panel-mounted Model)
4685/500
Undo captive
screw
Remove Plug
(if fitted)
1
2
3
(Wall- /Pipe-mounted Model)
4680/500
IP66
C – Within Environmental Limits
55°C
Max.
–20°C
Min.
B – Within Temperature Limits
1000m
(3281 ft) max.
A – Maximum Distance Between
Instrument and Interface Unit
2
PREPARATION
3
MECHANICAL INSTALLATION
2.1
Checking the Instrument Type
2.1.1
Wall-mounted Instruments – Fig. 2.1
2.1.2
Panel-mount
Mark fixing centres
(see Fig. 3.2)
Drill suitable
holes
Secure instrument to
wall using
suitable fixings
1
2
3
Position ‘U’ bolts on pipe
Position plates over ‘U’ bolts
Secure transmitter to mounting plate
Secure plates
1
2
3
4
A – Wall-mounting
B – Pipe-mounting
68 (2.68)
214
(8.43)
232
(9.13)
Fixing Centres
160 (6.3)
250
(9.84)
69 (2.72)
Fixing Centres
Allowance for
Cable Bends
200 (7.9)
42
(1.65)
61 (23/8) O.D. Vertical or Horizontal Post