Lutron Electronics RT-606 Operation Manual

Specifications

Brand: Lutron Electronics
Category: Test Equipment
Model: RT-606 (Operation Manual)
File Type: PDF
File Size: —
Pages: 8

Read Online

Read the Lutron Electronics RT-606 Operation Manual online using the built-in PDF viewer.

Text of the manual

Model : RT-606

Your purchase of this ROTATION TESTER marks a step forward for you into the field of precision measurement. Although this TESTER is a complex and delicate instrument, its durable structure will allow many years of use if p r o p e r o p e r a t i n g techniques are developed. Please read the following instructions carefully and always keep this manual within easy reach. OPERATION MANUAL

* Do not touch the alligator

clips by hand when make the measurement ! * Risk of electric shock ! Caution : * Do not apply the voltage > 660 ACV to the input terminal ! * Remove test alligator clips before make the fuse replacement! * Cleaning - Only use the dry cloth to clean the plastic case ! Environment Conditions * Installation categories III, 660 Vac. * Pollution Degree 2. * Altitude up to 2000 meters. * Indoor use. * Relative humidity 80% max.

TABLE OF CONTENTS

1. SPECIFICATIONS.......................................................1 2. FRONT PANEL DESCRIPTION..................................... 2 2-1 Alligator clips.......................................................2 2-2 L1-L2-L3 indicator................................................2 2-3 Clockwise indicator.............................................. 2 2-4 Counter clockwise indicator.................................. 2 3. OPERATION PROCEDURES.........................................3 4. MAINTENANCE..........................................................5 4-1 Replacement of Fuse...........................................5 4-2 Cleaning.............................................................5 5. THE ADDRESS OF AFTER SERVICE CENTER ............... 6

1. GENERAL SPECIFICATIONS

Signal Input 100V AC up to 660 AC maximum, Voltage 3 min. limit for over 500V AC. Signal Current Approx. 7 mA per phase. Consumption Signal Freq. Motor rotation tester : Range 45 to 70 Hz. Power Supply None Circuit Structure Electronic ( not mechanical). Operating 0 to 50 (32 to 122 ). ℃ ℉ Temperature Operating Less than 80% RH. Humidity Dimension 100 x 50 x 25 mm. 3.9 x 1.9 x 1.0 inch. Weight 150 g/0.33 lb. Fuse Rating 250 mA/250 V. Standard * Operational manual.................1 PC Accessories * 40 cm length test lead with alligator clip.........................................3 PCs 1

2. FRONT PANEL DESCRIPTION

2-1 2-1 2-1 2-2 2-3 2-4 Fig. 1 2-1 Alligator clips 2-2 L1-L2-L3 indicator 2-3 Clockwise indicator 2-4 Counter clockwise indicator 2

3. OPERATION PROCEDURES

Caution : * Do not touch the alligator clips by hand when make the measurement ! * Risk of electric shock ! Caution : * Do not apply the voltage > 660 ACV to the input terminal ! table 1 1)Plug in the test leads into the " 3 phase rotation tester input terminal " (2-1, Fig. 1) 2)Connect alligator clips of the test leads to the 3 phase AC power source. 3)The tester will show the status & the " Rotation direction " of 3 phase lines, refer table 1. 3

* Remove test leads before

make the fuse & battery replacement. ! * Cleaning - Only use the dry cloth to clean the plastic case ! 4-1 Replacement of Fuse 1)Fuse is used for the input protection of " 3 Phase Rotation Tester " 2)It is necessary to replace the fuse when lamp indicator of L1 or L3 show no connection status, then please check if the fuse is broken ? 3)Open the housing case by loosing the screws, replace the fuse with : Rating : 250 mA/250 V, Size : 5 mm dia. x 20 mm 4)Reinstate the housing box again. 4-2 Cleaning 1)Do not take any cleaning action while the testing. 2)When cleaning, do not wipe with the wetting material. 4