Narda Safety Test Solutions GmbH

Sandwiesenstraße 7 72793 Pfullingen, Germany © 2017 ® Names and Logo are registered trademarks of Narda Safety Test Solutions GmbH and L3 Communications Holdings, Inc. Trade names are trademarks of the owners. Order no.: 3006/98.22 Version: 31/03.2017, S/N A ... (Firmware V1.5.0 4) Previous version: 03/10.2014, S/N A ... (Firmware V1.4.0 4) Subject to change. Our normal guarantee and delivery terms apply. Printed in Germany

Command Reference Guide

Narda SRM-3006 I Contents 1 General information ...........................................................................................................................1 Backus-Naur format .....................................................................................................................1 Get commands (query) ................................................................................................................1 Set commands (send) .................................................................................................................1 Responses ...................................................................................................................................1 Remote ON operation ..................................................................................................................1 Data types ....................................................................................................................................2 Error codes ..................................................................................................................................2 2 Command groups ..............................................................................................................................3 General ........................................................................................................................................3 Mode LEVEL................................................................................................................................3 Mode SAFETY (Multi Channel Power) ........................................................................................3 Mode SPECTRUM ......................................................................................................................4 Mode UMTS - Option required ....................................................................................................4 Mode LTE - Option required ........................................................................................................4 Mode LTE TDD - Option required ...............................................................................................4 Clock ............................................................................................................................................5 Data Logger .................................................................................................................................5 Setups ..........................................................................................................................................5 Service tables (channel tables) ...................................................................................................5 Remote Desktop ..........................................................................................................................5 3 Command Descriptions......................................................................................................................6 DATE ...........................................................................................................................................6 DATE? .........................................................................................................................................7 DEV_ID? ......................................................................................................................................8 DEV_INFO? .................................................................................................................................9 DEV_OPTION........................................................................................................................... 10 DEV_OPTION? ........................................................................................................................ 11 DL_CLR .................................................................................................................................... 12 DL_CLR_ALL............................................................................................................................ 13 DL_DATA? (General) ............................................................................................................... 14 DL_DATA? (LevelRecording) ................................................................................................... 15 DL_DATA? (SafetyEvaluation, condensed) ............................................................................. 19 DL_DATA? (SafetyEvaluation, detailed) .................................................................................. 24

Command Reference Guide

II SRM-3006 Narda DL_DATA? (Spectrum) ............................................................................................................. 29 DL_DATA? (Scope) .................................................................................................................. 33 DL_DATA? (UMTS) .................................................................................................................. 37 DL_DATA? (LTE) ...................................................................................................................... 41 DL_DATA? (LTE TDD) ............................................................................................................. 46 DL_INFO? ................................................................................................................................. 51 DL_INFO_SUB? ....................................................................................................................... 52 DL_MEMORY? ......................................................................................................................... 53 DL_NUMBER? .......................................................................................................................... 54 DL_STORE ............................................................................................................................... 55 DL_VOICE? .............................................................................................................................. 56 ERROR? ................................................................................................................................... 57 LEVEL? ..................................................................................................................................... 58 LEVEL_AVG_CONFIG ............................................................................................................. 59 LEVEL_AVG_CONFIG? ........................................................................................................... 60 LEVEL_AVG_LIST?.................................................................................................................. 61 LEVEL_CONFIG ....................................................................................................................... 62 LEVEL_CONFIG? ..................................................................................................................... 63 LIVESCREEN? ......................................................................................................................... 64 LTE?.......................................................................................................................................... 65 LTE_AVG_CONFIG .................................................................................................................. 68 LTE_AVG_CONFIG? ................................................................................................................ 69 LTE_AVG_LIST? ...................................................................................................................... 70 LTE_CONFIG ........................................................................................................................... 71 LTE_CONFIG? ......................................................................................................................... 72 LTE_RESET_TABLE ................................................................................................................ 73 LTE_TDD? ................................................................................................................................ 74 LTE_TDD_AVG_CONFIG ........................................................................................................ 77 LTE_TDD_AVG_CONFIG? ...................................................................................................... 78 LTE_TDD_AVG_LIST? ............................................................................................................. 79 LTE_TDD_CONFIG .................................................................................................................. 80 LTE_TDD_CONFIG? ................................................................................................................ 81 LTE_TDD_RESET_TABLE ...................................................................................................... 82 MEAS_START .......................................................................................................................... 83 MEAS_STOP ............................................................................................................................ 84 MODE ....................................................................................................................................... 85

Command Reference Guide

Narda SRM-3006 III MODE? ..................................................................................................................................... 86 MR_LIST? ................................................................................................................................. 87 PROG_EXIT ............................................................................................................................. 88 RBW_LIST? .............................................................................................................................. 89 REMOTE .................................................................................................................................. 90 REMOTE? ................................................................................................................................ 91 RESULT_TYPE_RESET .......................................................................................................... 92 RESULT_TYPE_RESET_MIN_MAX ....................................................................................... 93 SAFETY? .................................................................................................................................. 94 SAFETY_AVG_CONFIG .......................................................................................................... 97 SAFETY_AVG_CONFIG? ........................................................................................................ 98 SAFETY_AVG_LIST? .............................................................................................................. 99 SAFETY_CONFIG .................................................................................................................. 100 SAFETY_CONFIG? ................................................................................................................ 101 SCR_CLR ............................................................................................................................... 102 SCR_CLR_ALL ...................................................................................................................... 103 SCR_DATA?........................................................................................................................... 104 SCR_INFO?............................................................................................................................ 105 SCR_NUMBER? .................................................................................................................... 106 SEND_KEY............................................................................................................................. 107 SEND_ROT_KNOB ................................................................................................................ 108 SPECTRUM?.......................................................................................................................... 109 SPECTRUM_AVG_CONFIG .................................................................................................. 111 SPECTRUM _AVG_CONFIG? ............................................................................................... 112 SPECTRUM_AVG_LIST? ...................................................................................................... 113 SPECTRUM_BI_VALUE? ...................................................................................................... 114 SPECTRUM_CONFIG ........................................................................................................... 115 SPECTRUM_CONFIG? ......................................................................................................... 116 SPECTRUM_MRK_HIGHEST? ............................................................................................. 117 SPECTRUM_MRK_IDX_VALUE?.......................................................................................... 118 SPECTRUM_MRK_VALUE? .................................................................................................. 119 SPECTRUM_PKT_TABLE? ................................................................................................... 120 SRV_CLR ............................................................................................................................... 121 SRV_CLR_ALL ....................................................................................................................... 122 SRV_DS ................................................................................................................................. 123 SRV_DS? ............................................................................................................................... 124

Command Reference Guide

IV SRM-3006 Narda SRV_LIST? ............................................................................................................................. 125 SRV_SEL ................................................................................................................................ 126 SRV_SEL? .............................................................................................................................. 127 SU_CLR .................................................................................................................................. 128 SU_CLR_ALL ......................................................................................................................... 129 SU_DEFAULT ......................................................................................................................... 130 SU_DS .................................................................................................................................... 131 SU_DS? .................................................................................................................................. 132 SU_LIST? ............................................................................................................................... 133 SU_RECALL ........................................................................................................................... 134 SU_STORE ............................................................................................................................. 135 SWEEP_STATE? ................................................................................................................... 136 TIME........................................................................................................................................ 137 TIME?...................................................................................................................................... 138 UMTS? .................................................................................................................................... 139 UMTS_AVG_CONFIG ............................................................................................................ 141 UMTS _AVG_CONFIG? ......................................................................................................... 142 UMTS_AVG_LIST? ................................................................................................................. 143 UMTS_CONFIG ...................................................................................................................... 144 UMTS_CONFIG? .................................................................................................................... 145 UMTS_RESET_TABLE .......................................................................................................... 146 UNIT ........................................................................................................................................ 147 UNIT? ...................................................................................................................................... 148 UNIT_LIST? ............................................................................................................................ 149 VBW_LIST? ............................................................................................................................ 150 VERSION? .............................................................................................................................. 151 4 Remote application examples ....................................................................................................... 152 Time synchronization with a PC ............................................................................................. 152 Synchronizing queries of measured values with the sweep state .......................................... 154 Carrier signal monitoring with alarms ..................................................................................... 156

Command Reference Guide

Narda SRM-3006 1 1 General information Backus-Naur format Symbol Meaning < > Defined element | Exclusive OR {..} Element group, at least one element required [..] Optional field ( ) Comment Get commands (query) Get commands are used above all to query measurement values and settings. They always include ‘?’ in the body of the command. Commas separate the parameters. Syntax: [PARA1] , [PARA2] , [PARAn] ; Set commands (send) Set commands are used to configure and control the device and are characterized by the fact that there is no question mark in the body of the command. Commas separate the parameters. Syntax: [PARA1] , [PARA2] , [PARAn] ; Responses A response from the device always follows a Set or a Get-command. The response always contains an error code and, optionally, the requested data. Syntax [DATA1] , [DATA2] , [DATAn] , ; Remote ON operation Almost all the commands can only be processed by the device when it is in Remote ON mode. The REMOTE ON; command activates remote operation. Once activated, remote operation can be deactivated by the REMOTE OFF; command.

Command Reference Guide

2 SRM-3006 Narda Data types Abbreviation Description Example String ASCII character string “Funny Fox” Enum Enumerator, fixed texts ON, OFF, REMOTE Float/Double Floating point number 12.34 or 1.234e1 Byte 8 bit whole number without sign 123 ShortInt 16 bit whole number with sign -1234 Integer 32 bit whole number with sign -12345 DWord 32 bit whole number without sign 12345 LongInt 64 bit whole number with sign -123456 Date ASCII character string, format: dd:mm:yy 21.04.81 Time ASCII character string, format: hh:mm:ss 10:08:59 VersionString V[d]d.[d]d.[d]d[blankxxx] “V0.0.0" 4 "V99.99.99” Error codes Code Description 0 no error 401 remote command is not implemented in the remote module 402 invalid parameter 403 invalid count of parameters 404 invalid parameter range 405 last command is not completed 406 answer time between remote module and application module is too high 407 invalid or corrupt data 408 error while accessing the hardware 409 command is not supported in this version of the application module 410 remote is not activated (please send "REMOTE ON;" first) 411 command is not supported in the selected mode 412 memory of data logger is full 413 option code is invalid 414 incompatible version 415 subindex full 416 file counter full 417 data lost 418 checksum error 419 programming not successfull 420 path not found 421 break detected 422 low battery 423 file open error 424 data verify error

Command Reference Guide

Narda SRM-3006 3 2 Command groups General DEV_ID? Queries the device identification number DEV_INFO? Queries device information that describes the individual device DEV_OPTION Activates a device option DEV_OPTION? Queries the options prepared and activated in the device ERROR? Queries the current error code MEAS_START Starts the cyclical measurement in the current operating mode MEAS_STOP Stops the cyclical measurement in the current operating mode MODE Sets the current operating mode MODE? Queries the current operating mode MR_LIST? Queries the current possible reference level settings PROG_EXIT Ends the measurements and switches the device to standby RBW_LIST? Queries the current possible RBW settings REMOTE Activates / deactivates remote communication REMOTE? Queries readiness for remote operation RESULT_TYPE_RESET Resets all measurement runs (traces) to default values RESULT_TYPE_RESET_MIN_MAX Resets the Min and Max traces to default values UNIT Sets the current unit setting UNIT? Queries the current unit setting UNIT_LIST? Queries the current possible unit settings VBW_LIST? Queries the current possible VBW settings VERSION? Queries the versions of the current firmware modules Mode LEVEL LEVEL? Queries the different measurement runs (traces) LEVEL_AVG_CONFIG Sets the current averaging method LEVEL_AVG_CONFIG? Queries the current averaging method LEVEL_AVG_LIST? Queries the current possible averaging settings LEVEL_CONFIG Sets the current measurement parameters for LEVEL mode LEVEL_CONFIG? Queries the current measurement parameters for LEVEL mode SWEEP_STATE? Queries the status and progress of the current measurement Mode SAFETY (Multi Channel Power) SAFETY? Queries one or more measurement runs (traces) in Safety mode SAFETY_AVG_CONFIG Sets the current averaging method SAFETY_AVG_CONFIG? Queries the current averaging method SAFETY_AVG_LIST? Queries the current possible averaging settings SAFETY_CONFIG Sets the current measurement parameters for Safety mode SAFETY_CONFIG? Queries the current measurement parameters for Safety mode SWEEP_STATE? Queries the status and progress of the current measurement

Command Reference Guide

4 SRM-3006 Narda Mode SPECTRUM SPECTRUM? Queries one or more measurement runs in Spectrum mode SPECTRUM _AVG_CONFIG Sets the current averaging method SPECTRUM _AVG_CONFIG? Queries the current averaging method SPECTRUM _AVG_LIST? Queries the current possible averaging settings SPECTRUM _BI_VALUE? Band Integration over Frequency SPECTRUM _MRK_HIGHEST? Queries the marker values for the highest measurement value SPECTRUM _MRK_IDX_VALUE? Queries the marker values for a selected marker SPECTRUM _MRK_VALUE? Queries the marker values for a specified frequency point SPECTRUM _PKT_TABLE? Queries a configurable peak table SPECTRUM_CONFIG Sets the current measurement parameters for Spectrum mode SPECTRUM_CONFIG? Queries the current measurement parameters for Spectrum SWEEP_STATE? Queries the status and progress of the current measurement Mode UMTS - Option required UMTS? Queries the different measurement runs (traces) UMTS _AVG_CONFIG Sets the current averaging method UMTS _AVG_CONFIG? Queries the current averaging method UMTS_AVG_LIST? Queries the current possible averaging settings UMTS_CONFIG Sets the current measurement parameters for UMTS mode UMTS_CONFIG? Queries the current measurement parameters for UMTS mode UMTS_RESET_TABLE Resets the UMTS channel table SWEEP_STATE? Queries the status and progress of the current measurement Mode LTE - Option required The LTE commands applies for the LTE FDD optional mode LTE? Queries the different measurement runs (traces) LTE_AVG_CONFIG Sets the current averaging method LTE_AVG_CONFIG? Queries the current averaging method LTE_AVG_LIST? Queries the current possible averaging settings LTE_CONFIG Sets the current measurement parameters for LTE mode LTE_CONFIG? Queries the current measurement parameters for LTE mode LTE_RESET_TABLE Resets the table for LTE results (Cell IDs) SWEEP_STATE? Queries the status and progress of the current measurement Mode LTE TDD - Option required LTE_TDD? Queries the different measurement runs (traces) LTE_TDD_AVG_CONFIG Sets the current averaging method LTE_TDD_AVG_CONFIG? Queries the current averaging method LTE_TDD_AVG_LIST? Queries the current possible averaging settings LTE_TDD_CONFIG Sets the current measurement parameters for LTE TDD mode LTE_TDD_CONFIG? Queries the current measurement parameters for LTE TDD mode LTE_TDD_RESET_TABLE Resets the table for LTE TDD results (Cell IDs) SWEEP_STATE? Queries the status and progress of the current measurement

Command Reference Guide

Narda SRM-3006 5 Clock DATE Sets the current date setting for the device DATE? Queries the current date setting for the device TIME Sets the current system time of the device TIME? Queries the current system time of the device Data Logger DL_CLR Deletes a selected data set DL_CLR_ALL Deletes all data sets DL_DATA? Queries a sub data set DL_INFO? Queries overview information for a data set DL_INFO_SUB? Queries overview information for a sub data set DL_MEMORY? Queries the free memory space available for storing data sets DL_NUMBER? Queries the number of saved data sets DL_STORE Saves a data set in the current measurement mode DL_VOICE? Queries the voice comment of a data set in PCM wave format SCR_CLR Deletes a selected screenshot SCR_CLR_ALL Deletes all screenshots SCR_DATA? Queries a selected screenshot SCR_INFO? Queries the additional information for a screenshot SCR_NUMBER? Queries the number of existing screenshots Setups SU_CLR Deletes a selected device setup SU_CLR_ALL Deletes all device setups SU_DEFAULT Resets the device to the default (works) settings SU_DS Transfers a setup file to the device SU_DS? Reads a setup file out of the device SU_LIST? Lists the names of the setups stored in the device SU_RECALL Activates a selected device setup SU_STORE Saves the current device settings as a new setup in the device Service tables (channel tables) SRV_CLR Deletes a service table SRV_CLR_ALL Deletes all service tables SRV_DS Creates a service table SRV_DS? Queries a service table SRV_LIST? Queries all service tables without details SRV_SEL Sets the current active service table SRV_SEL? Queries the current active service table Remote Desktop LIVESCREEN? Queries the current device display as a screenshot SEND_KEY Activates an operating panel key by remote communication SEND_ROT_KNOB Activates rotary control by remote communication

Command Reference Guide

6 SRM-3006 Narda 3 Command Descriptions DATE Description Sets the current date setting for the device Input DATE ; Output ; Parameter List Parameter Type Description Date dd.mm.yy DWord Error code Example (Input) DATE 06.07.12; (Output) 0;

Command Reference Guide

Narda SRM-3006 7 DATE? Description Queries the current date setting for the device Input DATE?; Output ,; Parameter List Parameter Type Description Date dd.mm.yy DWord Error code Example (Input) DATE?; (Output) 03.05.10,0;

Command Reference Guide

8 SRM-3006 Narda DEV_ID? Description Queries the device identification number. Input DEV_ID?; Output < DeviceID >,; Parameter List Parameter Type Description < DeviceID> String Device identification number DWord Error code Example (Input) DEV_ID?; (Output) "F89AEF31CD344840",0;

Command Reference Guide

Narda SRM-3006 9 DEV_INFO? Description Queries all device information that describes the individual device. Input DEV_INFO?; Output ,,,, ,, ,,; Parameter List Parameter Type Description String Product name (e.g. “SRM-3006”) String Production identification number String Serial number String Device identification number VersionString Firmware version Date Firmware date Date Date of the calibration Date Recommended date for recalibration DWord Error code Example (Input) DEV_INFO?; (Output) "SRM-3006","SW0003","A-1234","F89AEF31CD344840", "V1.1.2",29.04.10,12.03.10,12.03.11,0;

Command Reference Guide

10 SRM-3006 Narda DEV_OPTION Description Activates a device option. Options are additional measurement modes that the device is prepared for. Each device needs a separate activation code for each option, which can be purchased. The current device firmware determines whether an option is prepared and can be activated. Input DEV_OPTION ,; Output ; Parameter List Parameter Type Description Short Option index, valid values: 0 to 63 String Key code provided by Narda for activation of an option. The code consists of 16 hexadecimal numbers. Use capital letters for A – F. Exact spelling required. DWord Error code Example (Input) DEV_OPTION 2,“0123456789ABCDEF“; (Output) 0; Example (Input) DEV_OPTION 3,“0123456789ABCDEF“; (Output) 409;

Command Reference Guide

Narda SRM-3006 11 DEV_OPTION? Description Queries the options prepared and activated in the device. Options are additional measurement modes that the device is prepared for. Each device needs a separate activation code for each option, which can be purchased. The current device firmware determines whether an option is prepared and can be activated. Input DEV_OPTION? ; Output ,,; Parameter List Parameter Type Description Short Option index, valid values: 0 to 63 Enum FREE; CLOSED; UNKNOWN String Option name DWord Error code Example (Input) DEV_OPTION? 0; (Output) FREE,"UMTS C-CPICH Demodulation",0; Example (Input) DEV_OPTION? 1; (Output) FREE,"Scope",0; Example (Input) DEV_OPTION? 2; (Output) UNKNOWN,"",0;

Command Reference Guide

12 SRM-3006 Narda DL_CLR Description Deletes a selected data set (DataLogger). All associated sub data sets are also deleted. Input DL_CLR ; Output ; Parameter List Parameter Type Description Integer Index of the selected dataset DWord Error code Example (Input) DL_CLR 3; (Output) 0;

Command Reference Guide

Narda SRM-3006 13 DL_CLR_ALL Description Deletes all data sets (DataLogger) and all associated sub data sets. Input DL_CLR_ALL; Output ; Parameter List Parameter Type Description DWord Error code Example (Input) DL_CLR_ALL; (Output) 0;

Command Reference Guide

14 SRM-3006 Narda DL_DATA? (General) Description Queries a sub data set (DataLogger) A data set always comprises one or more sub data sets. The store mode determines whether each data set is saved separately or as a group of sub data sets. Data sets can be substantial in size and contain various types of information depending on the operating mode used to compile them. Structure of DL information Section Description GENERAL Describes device specific information and is the same in all operating modes. SETUP COMMON Describes non-operating mode specific device settings and is the same in all operating modes. SETUP MODE Describes operating mode specific device settings. TRACE COMMON Describes information that is the same for all measurement runs in an operating mode. TRACE SPECIFIC Describes information that differs from one measurement run to another. This also includes the individual measured values in each measurement run, among other things. ADDITIONAL DATA Additional overload indication in SCOPE mode

Command Reference Guide

Narda SRM-3006 15 DL_DATA? (LevelRecording) Description Queries a sub data set (DataLogger) in LevelRecording mode Input DL_DATA? ,; Output ,,,,,, ,,,, ,,, ,,,,, ,,,, ,,,,,,,, , ,,,,,,, ,, ,,,, {,,,,} ; Input Parameter List Parameter Type Description Short Index of the selected dataset Short Index of the selected sub-dataset

Command Reference Guide

16 SRM-3006 Narda Output Parameter List, Section: GENERAL Parameter Type Description DWord Unique ID of the sub-dataset Enum LEVEL Enum MAN Manual storing, executed by the user COND_FIRST Auto-Store at first TRUE condition COND_ALL Auto-Store for all TRUE conditions TIME Auto-Store, Timer controlled MR_USER Storing triggered by measurement routine, with user interaction required MR_NUM Storing triggered by measurement routine, event controlled (numerical counter) MR_TIME Storing triggered by measurement routine, event controlled by timer Date dd.mm.yy Time hh:mm:ss Enum YES, NO; Yes, if there is at least one result in the dataset exceeding the range limit. Overdriven results may have a very high uncertainty. Enum NO No GPS signal available ACTUAL GPS signal available FROZEN GPS signal disappeared, using last valid coordinates Enum GPS, DGPS (Differential GPS) Enum 3D, 2D Short 0 to 36 Double -9999.9 to +9999.9 (meter) Double -90.000 00 to +90.000 00 (decimal degrees) Double -180.000 00 to +180.000 00 (decimal degrees) Enum Voice comment available: YES, NO; String Text comment up to 40 characters String Device serial number Date Device calibration date, dd.mm.yy String Device firmware version String Cable serial number Date Cable calibration date, dd.mm.yy String Antenna serial number Date Antenna calibration date, dd.mm.yy

Command Reference Guide

Narda SRM-3006 17 Output Parameter List, Section: SETUP COMMON Parameter Type Description Float Reference level (Measurement Range) [Unit] Enum Unit Enum Unit category: A, B, C, D A: Referred to RF Input 1 W (@ 50 Ohms) B: For connected antennas 1 W/m2 (@ 377 Ohms) C: For connected antennas 1 (equal to 100%) and referred to a standard D: For connected current clamps 1 W (@ 1 Ohm) Float Value required to determine Yref when changing between unit categories Enum X, Y, Z: Measurement of one selected spatial axis using a 3-axis antenna RSS: Isotropic measurement using a 3-axis antenna (Root Sum Square) SINGLE: Measurement without antenna or with a single axis antenna String Name of the safety standard being used String Name of the service table being used String Name of the cable being used String Name of the antenna being used Output Parameter List, Section: SETUP MODE Parameter Type Description Double Center frequency [Hz] Double Resolution bandwidth [Hz] Enum Video filter: ON, OFF; Double Video bandwidth [Hz] Double RMS averaging time [s] Integer 0, 3, 6, 10, 15, 20 [dB] Increase of the reference noise level related to the typical noise power level. Enum OFF Measurement results are always provided. ON Measurement results are provided only when the reference noise level is exceeded. Otherwise the result is indicated as “< value of reference noise level”. Float Upper limit of the display range [Unit] Float Y-scale of the display range: 20, 40, 60, 80, 100, 120 [dB]

Command Reference Guide

18 SRM-3006 Narda Output Parameter List, Section: TRACE COMMON Parameter Type Description Long Number of sweeps performed Short 0 – 100 [%] Progress of averaging Long 0 – 999 999 Number of sweeps used for spatial averaging Short Number of traces Output Parameter List, Section: TRACE SPECIFIC Parameter Type Description Enum RMS, MAX_RMS, PEAK, MAX_PEAK Enum YES, NO; Yes, if at least one trace exceeded the range limit. Overdriven results may have a very high uncertainty. Enum UNCHECKED When Noise Suppression is OFF LOW Result is below the reference noise level OK Result is equal or above the reference noise level Float Measurement result [Unit] DWord Error code Example (Input) DL_DATA? 29,1; (Output) 20044097,LEVEL,MAN,10.05.10,15:04:58,NO,NO,GPS,2D,3,0,48.4579766667, 9.23011166667,NO,"", "SW-0003",12.03.10,"V0.0.2 beta1","",01.01.01,"",01.01.01, -30,dBm,A,0,SINGLE,"ICNIRP 1998 general public","Österreich UMTS","","", 1500000000,5000000,OFF,50000,2.4,0,OFF,-30,120, 60,100,0,4, RMS,NO,UNCHECKED,-81.10704, MAX_RMS,NO,UNCHECKED,-81.10155, PEAK,NO,UNCHECKED,-69.45588, MAX_PEAK,NO,UNCHECKED,-67.52631, 0;

Command Reference Guide

Narda SRM-3006 19 DL_DATA? (SafetyEvaluation, condensed) Description Queries a sub data set (DataLogger) in SafetyEvaluation mode and in ‘condensed’ format. Services are grouped together as a result of service names with the same prefix. Input DL_DATA? ,; Output ,,,,,, ,,,, ,,, ,,,,, ,,,, ,,,,,,,, , ,,,,,, ,,,,,, ,,,, , ,,,, , ,,,,,, ,{,,,} ; Input Parameter List Parameter Type Description Short Index of the selected dataset Short Index of the selected sub-dataset

Command Reference Guide

20 SRM-3006 Narda Output Parameter List, Section: GENERAL Parameter Type Description DWord Unique ID of the sub-dataset Enum SAFETY_CONDENSED Enum MAN Manual storing, executed by the user COND_FIRST Auto-Store at first TRUE condition COND_ALL Auto-Store for all TRUE conditions TIME Auto-Store, Timer controlled MR_USER Storing triggered by measurement routine, with user interaction required MR_NUM Storing triggered by measurement routine, event controlled (numerical counter) MR_TIME Storing triggered by measurement routine, event controlled by timer Date dd.mm.yy Time hh:mm:ss Enum YES, NO; Yes, if there is at least one result in the dataset exceeding the range limit. Overdriven results may have a very high uncertainty. Enum NO No GPS signal available ACTUAL GPS signal available FROZEN GPS signal disappeared, using last valid coordinates Enum GPS, DGPS (Differential GPS) Enum 3D, 2D Short 0 to 36 Double -9999.9 to +9999.9 (meter) Double -90.000 00 to +90.000 00 (decimal degrees) Double -180.000 00 to +180.000 00 (decimal degrees) Enum Voice comment available: YES, NO; String Text comment up to 40 characters String Device serial number Date Device calibration date, dd.mm.yy String Device firmware version String Cable serial number Date Cable calibration date, dd.mm.yy String Antenna serial number Date Antenna calibration date, dd.mm.yy

Command Reference Guide

Narda SRM-3006 21 Output Parameter List, Section: SETUP COMMON Parameter Type Description Float Reference level (Measurement Range) [Unit] Enum Unit Enum Unit category: A, B, C, D See DL_Data? (Level) Float Value required to determine Yref when changing between unit categories Enum X, Y, Z: Measurement of one selected spatial axis using a 3-axis antenna RSS: Isotropic measurement using a 3-axis antenna (Root Sum Square) SINGLE: Measurement without antenna or with a single axis antenna String Name of the safety standard being used String Name of the service table being used String Name of the cable being used String Name of the antenna being used Parameter List, Output Section: SETUP MODE Parameter Type Description Enum YES Frequency gaps between channels will be ignored NO Frequency gaps between channels will be measured and summarized as “Others” Enum MANUAL, AUTO, INDIVIDUAL for more details see SAFETY_CONFIG Double Resolution bandwidth setting [Hz] Enum NUMBER, TIME Result averaging over a certain number of samples or over a certain period of time Integer Averaging time 60 – 1800 [s] Integer Number of averaging samples: 4, 8, 16, 32, 64, 128, 256 Integer 0, 3, 6, 10, 15, 20 [dB] Increase of the reference noise level related to the typical noise power level. Enum OFF Measurement results are always provided. ON Measurement results are provided only when the reference noise level is exceeded. Otherwise the result is indicated as “< value of reference noise level”. Float Upper limit of the display range [Unit] Float Y-scale of the display range: 20, 40, 60, 80, 100, 120 [dB] Enum DETAIL Separate results for each channel (service) CONDENSED Combined results for channels (services) having similar names Enum TABLE Numerical results shown as a table BARGRAPH Results shown as a bar graph

Command Reference Guide

22 SRM-3006 Narda Output Parameter List, Section: TRACE COMMON Parameter Type Description Long Number of sweeps performed Integer Sweep Time [ms] Short 0 – 100 [%] Progress of averaging Long 0 – 999 999 Number of sweeps used for spatial averaging Integer Number of channels (services) being measured and listed String Channel name (service name) Double Resolution bandwidth [Hz] being used for this channel (service) Double Lower frequency of the channel (service) [Hz] Double Upper frequency of the channel (service) [Hz] Integer Index of the group the channel (service) is associated with Short Number of traces, 1 - 7 Output Parameter List, Section: TRACE SPECIFIC Parameter Type Description Enum ACT, AVG, MAX, MAX_AVG, MIN, MIN_AVG, STD Enum YES, NO; Yes, if at least one trace exceeded the range limit. Overdriven results may have a very high uncertainty. Float Total power level [Unit] for the frequency bands of all channels (services) Enum Noise flag related to the total power of all channels: UNCHECKED When Noise Suppression is OFF LOW Result is below the reference noise level OK Result is equal or above the reference noise level Float Power level [Unit] of the frequency gaps between the channels Enum Nose flag related to the power level of the gaps (Others): UNCHECKED, LOW, OK see above Long Number of measurement values per trace Float Power level [Unit] of the channel group (service group) Enum Noise flag related to a specific channel group: UNCHECKED When Noise Suppression is OFF LOW Result is below the reference noise level OK Result is equal or above the reference noise level String Name of the channel group (service group) DWord Error code

Command Reference Guide

Narda SRM-3006 23 Example (Input) DL_DATA? 25,1; (Output) 8068156,SAFETY_CONDENSED,MAN,10.05.10,11:43:30,NO,NO,GPS,2D,0,0,0,0,NO,"", "SW-0003",12.03.10,"V0.0.2 beta1","",01.01.01,"Serial No",21.09.09, 0.0071,V/m,B,-13.0103,SINGLE,"ICNIRP 1998 general public","Österreich UMTS","","Ant Zero", NO,AUTO,1000000,TIME,360,8,0,ON,0.07099999,120,CONDENSED,TABLE, 19974,118,100,0, 9, "Mobilkom",300000,1900100000,1910100000,1, "T-Mobile AT",300000,1910100000,1915100000,2, "Hutchison",300000,1915100000,1920100000,3, "T-Mobile AT",300000,2019900000,2024700000,2, "Mobilkom",300000,2110300000,2120100000,1, "Hutchison",300000,2120100000,2134900000,3, "Orange",300000,2134900000,2149700000,4, "Mobilkom",300000,2149700000,2154700000,1, "T-Mobile AT",300000,2154700000,2169700000,2, 3, ACT,NO,0.0001548696,OK,0.0001287497,OK,4, 5.234514E-005,OK,"Mobilkom", 3.428021E-005,LOW,"T-Mobile", 5.097328E-005,OK,"Hutchison", 3.365057E-005,OK,"Orange", MAX_AVG,NO,0.0001605316,OK,0.0001346133,OK,4, 5.080223E-005,OK,"Mobilkom", 4.21289E-005,OK,"T-Mobile", 4.570597E-005,OK,"Hutchison", 3.535929E-005,OK,"Orange", STD,NO,59.93644,OK,60.25105,OK,4, 59.93644,OK,"Mobilkom", 60.09395,OK,"T-Mobile", 60.17255,OK,"Hutchison", 61,OK,"Orange", 0;

Command Reference Guide

24 SRM-3006 Narda DL_DATA? (SafetyEvaluation, detailed) Description Queries a sub data set (DataLogger) in SafetyEvaluation mode and in ‘detailed’ format. Input DL_DATA? ,; Output ,,,,,, ,,,, ,,, ,,,,, ,,,, ,,,,,,,, , ,,,,,, ,,,,,, ,,,,, ,,,,,, ,{,,,,,,} ; Input Parameter List Parameter Type Description Short Index of the selected dataset Short Index of the selected sub-dataset

Command Reference Guide

Narda SRM-3006 25 Output Parameter List, Section: GENERAL Parameter Type Description DWord Unique ID of the sub-dataset Enum SAFETY Enum MAN Manual storing, executed by the user COND_FIRST Auto-Store at first TRUE condition COND_ALL Auto-Store for all TRUE conditions TIME Auto-Store, Timer controlled MR_USER Storing triggered by measurement routine, with user interaction required MR_NUM Storing triggered by measurement routine, event controlled (numerical counter) MR_TIME Storing triggered by measurement routine, event controlled by timer Date dd.mm.yy Time hh:mm:ss Enum YES, NO; Yes, if there is at least one result in the dataset exceeding the range limit. Overdriven results may have a very high uncertainty. Enum NO No GPS signal available ACTUAL GPS signal available FROZEN GPS signal disappeared, using last valid coordinates Enum GPS, DGPS (Differential GPS) Enum 3D, 2D Short 0 to 36 Double -9999.9 to +9999.9 (meter) Double -90.000 00 to +90.000 00 (decimal degrees) Double -180.000 00 to +180.000 00 (decimal degrees) Enum Voice comment available: YES, NO; String Text comment (up to 40 chr) String Device serial number (up to 15 chr) Date Device calibration date, dd.mm.yy String Device firmware version (up to 15 chr) String Cable serial number (up to 15 chr) Date Cable calibration date, dd.mm.yy String Antenna serial number (up to 15 chr) Date Antenna calibration date, dd.mm.yy

Command Reference Guide

26 SRM-3006 Narda Output Parameter List, Section: SETUP COMMON Parameter Type Description Float Reference level (Measurement Range) [Unit] Enum Unit Enum Unit category: A, B, C, D See DL_Data? (Level) Float Value required to determine Yref when changing between unit categories Enum X, Y, Z: Measurement of one selected spatial axis using a 3-axis antenna RSS: Isotropic measurement using a 3-axis antenna (Root Sum Square) SINGLE: Measurement without antenna or with a single axis antenna String Name of the safety standard being used (up to 35 chr) String Name of the service table being used (up to 35 chr) String Name of the cable being used (up to 35 chr) String Name of the antenna being used (up to 35 chr) Parameter List, Output Section: SETUP MODE Parameter Type Description Enum YES Frequency gaps between channels will be ignored NO Frequency gaps between channels will be measured and summarized as “Others” Enum MANUAL, AUTO, INDIVIDUAL for more details see SAFETY_CONFIG Double Resolution bandwidth setting [Hz] Enum NUMBER, TIME Result averaging over a certain number of samples or over a certain period of time Integer Averaging time 60 – 1800 [s] Integer Number of averaging samples: 4, 8, 16, 32, 64, 128, 256 Integer 0, 3, 6, 10, 15, 20 [dB] Increase of the reference noise level related to the typical noise power level. Enum OFF Measurement results are always provided. ON Measurement results are provided only when the reference noise level is exceeded. Otherwise the result is indicated as “< value of reference noise level”. Float Upper limit of the display range [Unit] Float Y-scale of the display range: 20, 40, 60, 80, 100, 120 [dB] Enum DETAIL Separate results for each channel (service) CONDENSED Combined results for channels (services) having similar names Enum TABLE Numerical results shown as a table BARGRAPH Results shown as a bar graph

Command Reference Guide

Narda SRM-3006 27 Output Parameter List, Section: TRACE COMMON Parameter Type Description Long Number of sweeps performed Integer Sweep Time [ms] Short 0 – 100 [%] Progress of averaging Long 0 – 999 999 Number of sweeps used for spatial averaging Short Number of traces, 1 - 7 Output Parameter List, Section: TRACE SPECIFIC Parameter Type Description Enum ACT, AVG, MAX, MAX_AVG, MIN, MIN_AVG, STD Enum YES, NO; Yes, if at least one trace exceeded the range limit. Overdriven results may have a very high uncertainty. Float Total power level [Unit] for the frequency bands of all channels (services) Enum Noise flag related to the total power of all channels: UNCHECKED When Noise Suppression is OFF LOW Result is below the reference noise level OK Result is equal or above the reference noise level Float Power level [Unit] of the frequency gaps between the channels Enum Nose flag related to the power level of the gaps (Others): UNCHECKED, LOW, OK see above Long Number of measurement values per trace Float Power level [Unit] of the channel (service) Enum Noise flag related to a specific channel (service): UNCHECKED When Noise Suppression is OFF LOW Result is below the reference noise level OK Result is equal or above the reference noise level String Name of the channel (service), up to 15 chr Double Resolution bandwidth [Hz] being used for this channel (service) Double Lower frequency of the channel (service) [Hz] Double Upper frequency of the channel (service) [Hz] DWord Error code

Command Reference Guide

28 SRM-3006 Narda Example (Input) DL_DATA? 22,1; (Output) 2996901,SAFETY,MAN,10.05.10,10:01:34,NO,NO,GPS,2D,0,0,0,0,NO,"", "SW-0003",12.03.10,"V0.0.2 beta1","",01.01.01,"Serial No",21.09.09, 0.0071,V/m,B,-13.0103,SINGLE,"ICNIRP 1998 general public","Swisscom","","Ant Zero", NO,AUTO,50000,TIME,360,8,0,ON,0.07099999,120,DETAIL,TABLE, 31,97,0,0,7, ACT,NO,1.392834E-005,OK,1.29509E-005,OK,2, 3.078454E-006,LOW,"BCCH1 - CH 12",50000,937400000,937600000, 4.102984E-006,OK,"BCCH2 - CH 17",50000,938400000,938600000, MAX,NO,1.744996E-005,OK,1.538582E-005,OK,2, 9.887214E-006,OK,"BCCH1 - CH 12",50000,937400000,937600000, 7.87721E-006,OK,"BCCH2 - CH 17",50000,938400000,938600000, MAX_AVG,NO,1.426396E-005,OK,1.174674E-005,OK,2, 6.181964E-006,OK,"BCCH1 - CH 12",50000,937400000,937600000, 5.22092E-006,OK,"BCCH2 - CH 17",50000,938400000,938600000, AVG,NO,1.426396E-005,OK,1.174674E-005,OK,2, 6.181964E-006,OK,"BCCH1 - CH 12",50000,937400000,937600000, 5.22092E-006,OK,"BCCH2 - CH 17",50000,938400000,938600000, MIN_AVG,NO,1.426396E-005,OK,1.174674E-005,OK,2, 6.181964E-006,OK,"BCCH1 - CH 12",50000,937400000,937600000, 5.22092E-006,OK,"BCCH2 - CH 17",50000,938400000,938600000, MIN,NO,1.049014E-005,OK,8.40833E-006,OK,2, 3.078454E-006,LOW,"BCCH1 - CH 12",50000,937400000,937600000, 3.078454E-006,LOW,"BCCH2 - CH 17",50000,938400000,938600000, STD,NO,42.09836,OK,42.10304,OK,2, 42.09836,OK,"BCCH1 - CH 12",50000,937400000,937600000, 42.12057,OK,"BCCH2 - CH 17",50000,938400000,938600000, 0; Example (Input) DL_DATA? 23,1; (Output) 3402638,SAFETY,MAN,10.05.10,10:09:27,NO,NO,GPS,2D,0,0,0,0,NO,"", "SW-0003",12.03.10,"V0.0.2 beta1","",01.01.01,"Serial No",21.09.09, 0.0071,V/m,B,-13.0103,SINGLE,"ICNIRP 1998 general public","Swisscom","","Ant Zero", NO,AUTO,50000,TIME,360,8,10,ON,0.07099999,120,DETAIL,TABLE, 839,115,22,0,2, ACT,NO,2.38456E-005,LOW,1.946985E-005,LOW,2, 9.734926E-006,LOW,"BCCH1 - CH 12",50000,937400000,937600000, 9.734926E-006,LOW,"BCCH2 - CH 17",50000,938400000,938600000, MAX_AVG,NO,2.38456E-005,LOW,1.946985E-005,LOW,2, 9.734926E-006,LOW,"BCCH1 - CH 12",50000,937400000,937600000, 9.734926E-006,LOW,"BCCH2 - CH 17",50000,938400000,938600000, 0;

Command Reference Guide

Narda SRM-3006 29 DL_DATA? (Spectrum) Description Queries a sub data set (DataLogger) in Spectrum mode. Input DL_DATA? ,; Output ,,,,,, ,,,, ,,, ,,,,, ,,,, ,,,,,,,, , ,,,,,,,, ,, ,,,,,, ,,,{,} ; Input Parameter List Parameter Type Description Short Index of the selected dataset Short Index of the selected sub-dataset

Command Reference Guide

30 SRM-3006 Narda Output Parameter List, Section: GENERAL Parameter Type Description DWord Unique ID of the sub-dataset Enum SPECTRUM Enum MAN Manual storing, executed by the user COND_FIRST Auto-Store at first TRUE condition COND_ALL Auto-Store for all TRUE conditions TIME Auto-Store, Timer controlled MR_USER Storing triggered by measurement routine, with user interaction required MR_NUM Storing triggered by measurement routine, event controlled (numerical counter) MR_TIME Storing triggered by measurement routine, event controlled by timer Date dd.mm.yy Time hh:mm:ss Enum YES, NO; Yes, if there is at least one result in the dataset exceeding the range limit. Overdriven results may have a very high uncertainty. Enum NO No GPS signal available ACTUAL GPS signal available FROZEN GPS signal disappeared, using last valid coordinates Enum GPS, DGPS (Differential GPS) Enum 3D, 2D Short 0 to 36 Double -9999.9 to +9999.9 (meter) Double -90.000 00 to +90.000 00 (decimal degrees) Double -180.000 00 to +180.000 00 (decimal degrees) Enum Voice comment available: YES, NO; String Text comment up to 40 characters String Device serial number Date Device calibration date, dd.mm.yy String Device firmware version String Cable serial number Date Cable calibration date, dd.mm.yy String Antenna serial number Date Antenna calibration date, dd.mm.yy

Command Reference Guide

Narda SRM-3006 31 Output Parameter List, Section: SETUP COMMON Parameter Type Description Float Reference level (Measurement Range) [Unit] Enum Unit Enum Unit category: A, B, C, D See DL_Data? (Level) Float Value required to determine Yref when changing between unit categories Enum X, Y, Z: Measurement of one selected spatial axis using a 3-axis antenna RSS: Isotropic measurement using a 3-axis antenna (Root Sum Square) SINGLE: Measurement without antenna or with a single axis antenna String Name of the safety standard being used String Name of the service table being used String Name of the cable being used String Name of the antenna being used Output Parameter List, Section: SETUP MODE Parameter Type Description Double Lower limit of the frequency range [Hz] Double Upper limit of the frequency range [Hz] Double Resolution bandwidth [Hz] Enum Video filter: ON, OFF; Double Video bandwidth [Hz] Enum NUMBER, TIME Integer Time [s]: 60, 120, 180, 240, 300, 360, 420, 480, 540, 600, 660, 720, 780, 840, 900, 960, 1020, 1080, 1140, 1200, 1260, 1320, 1380, 1440, 1500, 1560, 1620, 1680, 1740, 1800 Integer Number of averaging samples: 4, 8, 16, 32, 64, 128, 256 Float Upper limit of the display range [Unit] Float Y-scale of the display range: 20, 40, 60, 80, 100, 120 [dB] Output Parameter List, Section: TRACE COMMON Parameter Type Description Long Number of sweeps performed Integer Sweep Time [ms] Short 0 – 100 [%] Progress of averaging Long 0 – 999 999 Number of sweeps used for spatial averaging Double Frequency resolution of the FFT [Hz] Short Number of traces, 1 - 7

Command Reference Guide

32 SRM-3006 Narda Output Parameter List, Section: TRACE SPECIFIC Parameter Type Description Enum Available traces: ACT Actual value AVG Averaged value MAX Maximum value MAX_AVG Maximum of the averaged values MIN Minimum value MIN_AVG Minimum of the averaged values STD Values of the selected standard Enum YES, NO; Yes, if at least one trace exceeded the range limit. Overdriven results may have a very high uncertainty. Long Number of measurement values per trace Float Power level [Unit] of the channel group (service group) DWord Error code Example (Input) DL_DATA? 1,3; (Output) 347188,SPECTRUM,TIME,28.04.10,14:30:22,NO,NO,GPS,2D,0,0,0,0,NO,"", "SW-0003",12.03.10,"V1.1.2 beta25","",01.01.01,"A-0015",13.09.04, -47,dBA/m,B,34.9055,RSS,"ICNIRP 1998 general public", "Singapore UMTS Downlink","","Three-axis Antenna 25 MHz - 3GHz", 993282300,1006717700,1000000,OFF,10000,NUMBER,360,128,-27,120, 1,285,100,0,500000,3, ACT,NO,28, -99.03781,-94.57047,-95.16548,-98.94504,-100.8946,-100.7684,-104.6847,-100.5939, -98.37939,-100.0578,-103.2881,-100.1873,-97.25618,-97.85872,-97.76405,-96.51723, -97.82542,-99.08546,-100.5912,-104.1949,-102.8588,-98.62962,-97.80673,-99.97719, -103.6932,-103.6553,-102.5543,-103.196, MAX,NO,28, -99.03781,-94.57047,-95.16548,-98.94504,-100.8946,-100.7684,-104.6847,-100.5939, -98.37939,-100.0578,-103.2881,-100.1873,-97.25618,-97.85872,-97.76405,-96.51723, -97.82542,-99.08546,-100.5912,-104.1949,-102.8588,-98.62962,-97.80673,-99.97719, -103.6932,-103.6553,-102.5543,-103.196, STD,NO,28, -18.78984,-18.78765,-18.78547,-18.78329,-18.7811,-18.77892,-18.77674,-18.77456, -18.77238,-18.77021,-18.76803,-18.76586,-18.76369,-18.76151,-18.75934,-18.75717, -18.755,-18.75283,-18.75067,-18.7485,-18.74633,-18.74417,-18.74201,-18.73985, -18.73769,-18.73553,-18.73337,-18.73121, 0;

Command Reference Guide

Narda SRM-3006 33 DL_DATA? (Scope) Description Queries a sub data set (DataLogger) in Scope mode. Input DL_DATA? ,; Output ,,,,,, ,,,, ,,, ,,,,, ,,,, ,,,,,,,, , ,,,,,, ,,,,,, ,, ,,, {,,,{,} } ,{,} ; Parameter List Parameter Type Description Short Index of the selected dataset Short Index of the selected sub-dataset

Command Reference Guide

34 SRM-3006 Narda Output Parameter List, Section: GENERAL Parameter Type Description DWord Unique ID of the sub-dataset Enum SCOPE Enum MAN Manual storing, executed by the user COND_FIRST Auto-Store at first TRUE condition COND_ALL Auto-Store for all TRUE conditions TIME Auto-Store, Timer controlled MR_USER Storing triggered by measurement routine, with user interaction required MR_NUM Storing triggered by measurement routine, event controlled (numerical counter) MR_TIME Storing triggered by measurement routine, event controlled by timer Date dd.mm.yy Time hh:mm:ss Enum YES, NO; Yes, if there is at least one result in the dataset exceeding the range limit. Overdriven results may have a very high uncertainty. Enum NO No GPS signal available ACTUAL GPS signal available FROZEN GPS signal disappeared, using last valid coordinates Enum GPS, DGPS (Differential GPS) Enum 3D, 2D Short 0 to 36 Double -9999.9 to +9999.9 (meter) Double -90.000 00 to +90.000 00 (decimal degrees) Double -180.000 00 to +180.000 00 (decimal degrees) Enum Voice comment available: YES, NO; String Text comment up to 40 characters String Device serial number Date Device calibration date, dd.mm.yy String Device firmware version String Cable serial number Date Cable calibration date, dd.mm.yy String Antenna serial number Date Antenna calibration date, dd.mm.yy

Command Reference Guide

Narda SRM-3006 35 Output Parameter List, Section: SETUP COMMON Parameter Type Description Float Reference level (Measurement Range) [Unit] Enum Unit Enum Unit category: A, B, C, D See DL_Data? (Level) Float Value required to determine Yref when changing between unit categories Enum X, Y, Z: Measurement of one selected spatial axis using a 3-axis antenna RSS: Isotropic measurement using a 3-axis antenna (Root Sum Square) SINGLE: Measurement without antenna or with a single axis antenna String Name of the safety standard being used String Name of the service table being used String Name of the cable being used String Name of the antenna being used Output Parameter List, Section: SETUP MODE Parameter Type Description Double Center frequency [Hz] Double Resolution bandwidth [Hz] Enum Video filter: ON, OFF; Double Video bandwidth [Hz] Float Upper limit of the display range [Unit] Float Y-scale of the display range: 20, 40, 60, 80, 100, 120 [dB] Enum Trigger mode: FREE_RUN, USER, TIME, SINGLE, MULTI Float Trigger level [Unit] Float Trigger delay, -90 to 200 [%] Enum Trigger edge: POSITIV, NEGATIV Date dd.mm.yy Time hh:mm:ss Double Time span, 1e-9 bis 86400 [s] Double Time resolution, 1e-9 bis 5400 [s] Output Parameter List, Section: TRACE COMMON Parameter Type Description Integer Number of measurement values per trace Reserved for storing of uncompleted datasets Enum NO, TRIGGERED, STOPPED, WAIT_ARMING, ARMED Number of measurement values per trace Short Number of traces

Command Reference Guide

36 SRM-3006 Narda Output Parameter List, Section: TRACE SPECIFIC Parameter Type Description Enum MAX, AVG, MIN, [STD] or ACT, [STD] Trace types are depending on RBW and time resolution settings. For time resolution >= 4 / RBW only trace types MAX, AVG, MIN, [STD] will always be shown when time resolution >= 4 / RBW. Enum YES, NO; Yes, if at least one trace exceeded the range limit. Overdriven results may have a very high uncertainty. Integer Number of measurement values per trace Float Measurement value [Unit] Output Parameter List, Section: TRACE GROUP Parameter Type Description Integer Number of measurement values per trace Enum YES, NO; Overdriven flag for every single measurement value of the trace group (MIN, MAX, AVG). Yes, if overdriven. Example (Input) DL_DATA? 32,1; (Output) 26205727,SCOPE,MAN,10.05.10,16:47:39,NO,NO,GPS,2D,0,0,48.4531683333, 9.22078833333,NO,"", "SW-0003",12.03.10,"V0.0.2 beta1","",01.01.01,"",01.01.01, -30,dBm,A,0,SINGLE,"ICNIRP 1998 general public","Österreich UMTS","","", 199980000,40000,OFF,400,-10,120,FREE_RUN,-30,0,POSITIV,01.01.01, 0:00:00,2,0.125, 17,WAIT_ARMING,3, MAX,NO,17, -92.2266,-93.04564,-93.04564,-92.15926,-93.04564,-93.75627,-93.29871,-93.12836,- 93.04564,-92.4351,-92.15926,-93.04564,-92.15926,-92.4351,-93.04564,-90.9942,- 93.04564, AVG,NO,17, -102.1236,-102.1236,-102.1236,-102.1236,-102.1236,-102.1236,-102.1236,-102.1236,- 102.1236,-102.1236,-102.1236,-102.1236,-102.1236,-102.1236,-102.1236,-102.1236,- 102.1236, MIN,NO,17, -999,-999,-999,-999,-999,-999,-999,-999,-999,-999,-999,-999,-999,-999,-999,-999,-999, 17,NO,NO,NO,NO,NO,NO,NO,NO,NO,NO,NO,NO,NO,NO,NO,NO,NO, 0;

Command Reference Guide

Narda SRM-3006 37 DL_DATA? (UMTS) Description Queries a sub data set (DataLogger) in UMTS mode. Input DL_DATA? ,; Output ,,,,,, ,,,, ,,, ,,,,, ,,,, ,,,,,,,, , ,,;,,, ,,,, ,,,, {,,,,,, {,,,} } ; Input Parameter List Parameter Type Description Short Index of the selected dataset Short Index of the selected sub-dataset

Command Reference Guide

38 SRM-3006 Narda Output Parameter List, Section: GENERAL Parameter Type Description DWord Unique ID of the sub-dataset Enum UMTS Enum MAN Manual storing, executed by the user COND_FIRST Auto-Store at first TRUE condition COND_ALL Auto-Store for all TRUE conditions TIME Auto-Store, Timer controlled MR_USER Storing triggered by measurement routine, with user interaction required MR_NUM Storing triggered by measurement routine, event controlled (numerical counter) MR_TIME Storing triggered by measurement routine, event controlled by timer Date dd.mm.yy Time hh:mm:ss Enum YES, NO; Yes, if there is at least one result in the dataset exceeding the range limit. Overdriven results may have a very high uncertainty. Enum NO No GPS signal available ACTUAL GPS signal available FROZEN GPS signal disappeared, using last valid coordinates Enum GPS, DGPS (Differential GPS) Enum 3D, 2D Short 0 to 36 Double -9999.9 to +9999.9 (meter) Double -90.000 00 to +90.000 00 (decimal degrees) Double -180.000 00 to +180.000 00 (decimal degrees) Enum Voice comment available: YES, NO; String Text comment up to 40 characters String Device serial number Date Device calibration date, dd.mm.yy String Device firmware version String Cable serial number Date Cable calibration date, dd.mm.yy String Antenna serial number Date Antenna calibration date, dd.mm.yy

Command Reference Guide

Narda SRM-3006 39 Output Parameter List, Section: SETUP COMMON Parameter Type Description Float Reference level (Measurement Range) [Unit] Enum Unit Enum Unit category: A, B, C, D See DL_Data? (Level) Float Value required to determine Yref when changing between unit categories Enum X, Y, Z: Measurement of one selected spatial axis using a 3-axis antenna RSS: Isotropic measurement using a 3-axis antenna (Root Sum Square) SINGLE: Measurement without antenna or with a single axis antenna String Name of the safety standard being used String Name of the service table being used String Name of the cable being used String Name of the antenna being used Output Parameter List, Section: SETUP MODE Parameter Type Description Double Center frequency [Hz] Enum NUMBER, TIME Integer Time [s]: 60, 120, 180, 240, 300, 360, 420, 480, 540, 600, 660, 720, 780, 840, 900, 960, 1020, 1080, 1140, 1200, 1260, 1320, 1380, 1440, 1500, 1560, 1620, 1680, 1740, 1800 Integer Number of averaging samples: 4, 8, 16, 32, 64, 128, 256 Integer 0, 3, 6, 10, 15, 20 [dB] Increase of the reference noise level related to the typical noise power level. Enum OFF Measurement results are always provided. ON Measurement results are provided only when the reference noise level is exceeded. Otherwise the result is indicated as “< value of reference noise level”. Float Upper limit of the display range [Unit] Float Y-scale of the display range: 20, 40, 60, 80, 100, 120 [dB] Enum TABLE BARGRAPH (not yet available) Power correction factor; Default: 1.0 ON, OFF

Command Reference Guide

40 SRM-3006 Narda Output Parameter List, Section: TRACE COMMON Parameter Type Description Long Number of sweeps performed Short Sweep Time [ms] Short 0 – 100 [%] Progress of averaging Short Number of traces Output Parameter List, Section: TRACE SPECIFIC Parameter Type Description Enum Available traces: ACT Actual value AVG Averaged value MAX Maximum value MAX_AVG Maximum of the averaged values MIN Minimum value MIN_AVG Minimum of the averaged values STD Values of the selected standard Enum YES, NO; Yes, if at least one trace exceeded the range limit. Overdriven results may have a very high uncertainty. Float Total power [Unit] Float Analog power [Unit] Enum State of the analog noise flag: UNCHECKED, LOW, OK Integer Number of values (UMTS codes) Short UMTS code no., 0 – 511 Float Measured power for the current UMTS code [Unit] Enum YES, NO DWord Error code Example (Input) DL_DATA? 31,1; (Output) 25357286,UMTS,MAN,10.05.10,16:33:31,NO,NO,GPS,2D,0,0,48.4531683333, 9.22078833333,NO,"", "SW-0003",12.03.10,"V0.0.2 beta1","",01.01.01,"",01.01.01, -30,dBm,A,0,SINGLE,"ICNIRP 1998 general public","Österreich UMTS","","", 2112800000,NUMBER,360,4,0,OFF,-30,120,TABLE,1,OFF, 7746,389,100,3, ACT,NO,-83.22507,-76.62833,UNCHECKED,3,501,-84.0913,YES,339, -90.65247,YES,34,-999,YES, MAX,NO,-79.50149,-72.1468,UNCHECKED,3,501,-79.61308,YES,339, -83.91263,YES,34,-92.3071,YES, MAX_AVG,NO,-79.94458,-73.37815,UNCHECKED,3,501,-80.13036,YES,339, -85.5185,YES,34,-97.9665,YES, 0;

Command Reference Guide

Narda SRM-3006 41 DL_DATA? (LTE) The LTE commands applies for the LTE FDD optional mode Description Queries a sub data set (DataLogger) in LTE mode. Input DL_DATA? ,; Output ,,,,,, ,,,, ,,, ,,,,, ,,,, ,,,,,,,, , ,,,;,,, ,,, ,,,, {,,,} , {,,,,, {,,,, {Value,} } ; Input Parameter List Parameter Type Description Short Index of the selected dataset Short Index of the selected sub-dataset

Command Reference Guide

42 SRM-3006 Narda Output Parameter List, Section: GENERAL Parameter Type Description DWord Unique ID of the sub-dataset Enum LTE Enum MAN Manual storing, executed by the user COND_FIRST Auto-Store at first TRUE condition COND_ALL Auto-Store for all TRUE conditions TIME Auto-Store, Timer controlled MR_USER Storing triggered by measurement routine, with user interaction required MR_NUM Storing triggered by measurement routine, event controlled (numerical counter) MR_TIME Storing triggered by measurement routine, event controlled by timer Date dd.mm.yy Time hh:mm:ss Enum YES, NO; Yes, if there is at least one result in the dataset exceeding the range limit. Overdriven results may have a very high uncertainty. Enum NO No GPS signal available ACTUAL GPS signal available FROZEN GPS signal disappeared, using last valid coordinates Enum GPS, DGPS (Differential GPS) Enum 3D, 2D Short 0 to 36 Double -9999.9 to +9999.9 (meter) Double -90.000 00 to +90.000 00 (decimal degrees) Double -180.000 00 to +180.000 00 (decimal degrees) Enum Voice comment available: YES, NO; String Text comment up to 40 characters String Device serial number Date Device calibration date, dd.mm.yy String Device firmware version String Cable serial number Date Cable calibration date, dd.mm.yy String Antenna serial number Date Antenna calibration date, dd.mm.yy

Command Reference Guide

Narda SRM-3006 43 Output Parameter List, Section: SETUP COMMON Parameter Type Description Float Reference level (Measurement Range) [Unit] Enum Unit Enum Unit category: A, B, C, D See DL_Data? (Level) Float Value required to determine Yref when changing between unit categories Enum X, Y, Z: Measurement of one selected spatial axis using a 3-axis antenna RSS: Isotropic measurement using a 3-axis antenna (Root Sum Square) SINGLE: Measurement without antenna or with a single axis antenna String Name of the safety standard being used String Name of the service table being used String Name of the cable being used String Name of the antenna being used Output Parameter List, Section: SETUP MODE Parameter Type Description Double Center frequency [Hz] Double Channel Bandwidth [Hz] 1.4e6, 3e6, 5e6, 10e6, 15e6, 20e6 Enum NUMBER, TIME Integer Time [s]: 60, 120, 180, 240, 300, 360, 420, 480, 540, 600, 660, 720, 780, 840, 900, 960, 1020, 1080, 1140, 1200, 1260, 1320, 1380, 1440, 1500, 1560, 1620, 1680, 1740, 1800 Integer Number of averaging samples: 4, 8, 16, 32, 64, 128, 256 Integer 0, 3, 6, 10, 15, 20 [dB] Increase of the reference noise level related to the typical noise power level. Enum OFF Measurement results are always provided. ON Measurement results are provided only when the reference noise level is exceeded. Otherwise the result is indicated as “< value of reference noise level”. Float Power correction factor; Default: 1.0 Enum ON, OFF Enum Cyclic Prefix Length: NORMAL, EXTENDED Enum Cell Synchronization: SYNC, NO_SYNC

Command Reference Guide

44 SRM-3006 Narda Output Parameter List, Section: TRACE COMMON Parameter Type Description Long Number of sweeps performed Short Sweep Time [ms] Short 0 – 100 [%] Progress of averaging Short >= 0 Short 0 4 503 Short >= 1 Enum YES, NO Short Number of traces Output Parameter List, Section: TRACE SPECIFIC Parameter Type Description Enum Available traces: ACT Actual value AVG Averaged value MAX Maximum value MAX_AVG Maximum of the averaged values MIN Minimum value MIN_AVG Minimum of the averaged values STD Values of the selected standard Enum YES, NO; Yes, if at least one trace exceeded the range limit. Overdriven results may have a very high uncertainty. Float Analog power [Unit] Enum State of the analog noise flag: UNCHECKED, LOW, OK Short 149 Enum PSS,SSS,RS_AVG,RS_SUM,RS_MAX, RS_0,RS_1,RS_2,RS_3 Enum YES, NO Float Total power [Unit] of selected cell IDs Shorts Number of values = NoOfCellIDs There is one value for each cell ID available. Float Measured power for the current cell ID and Signal [Unit] DWord Error code

Command Reference Guide

Narda SRM-3006 45 Example (Input) DL_DATA? 7,1; (Output) 1913414,LTE,MAN,18.06.14, 9:50:16,NO,ACTUAL,GPS,3D,6,504,48.4585316667,9.23057166667,NO,"", "D-0051",14.09.10,"V1.3.2","",01.01.01,"",01.01.01, -30,dBm,A,0,SINGLE,"ICNIRP 1998 General Public","Germany alle Funkdienste","","", 796000000,1400000,NUMBER,360,4,0,OFF,1,OFF,NORMAL,SYNC, 5,229,100,3, 335,2,YES, 334,2,YES, 333,2,YES, 1, ACT,YES,-62.13,UNCHECKED,9, PSS,YES,-72.78,3,-78.27,-74.23,-999.00, SSS,YES,-73.35,3,-79.88,-74.44,-999.00, RS_AVG,YES,-74.85,3,-80.51,-76.23,-999.00, RS_MAX,YES,-73.12,3,-78.10,-74.77,-999.00, RS_SUM,YES,-71.84,3,-77.50,-73.22,-999.00, RS_0,YES,-75.26,3,-78.10,-78.44,-999.00, RS_1,YES,-74.48,3,-86.37,-74.77,-999.00, RS_2,YES,-999.00,3,-999.00,-999.00,-999.00, RS_3,YES,-999.00,3,-999.00,-999.00,-999.00, 0; Example (Input) DL_DATA? 8,1; (Output) 1936150,LTE,MAN,18.06.14, 9:50:39,NO,ACTUAL,GPS,3D,6,511.5,48.4585966667,9.23044833333,NO,"", "D-0051",14.09.10,"V1.3.2","",01.01.01,"",01.01.01, -30,dBm,A,0,SINGLE,"ICNIRP 1998 General Public","Germany alle Funkdienste","","", 796000000,1400000,NUMBER,360,4,0,OFF,1,OFF,NORMAL,SYNC, 5,235,100,3, 335,2,YES, 334,2,NO, 333,2,NO, 1, ACT,NO,-60.91,UNCHECKED,9, PSS,NO,-74.29,3,-74.29,-72.72,-999.00, SSS,NO,-75.50,3,-75.50,-73.19,-999.00, RS_AVG,YES,-77.68,3,-77.68,-75.45,-999.00, RS_MAX,NO,-76.82,3,-76.82,-74.78,-999.00, RS_SUM,NO,-74.67,3,-74.67,-72.44,-999.00, RS_0,NO,-76.82,3,-76.82,-76.25,-999.00, RS_1,NO,-78.76,3,-78.76,-74.78,-999.00, RS_2,NO,-999.00,3,-999.00,-999.00,-999.00, RS_3,NO,-999.00,3,-999.00,-999.00,-999.00, 0;

Command Reference Guide

46 SRM-3006 Narda DL_DATA? (LTE TDD) Description Queries a sub data set (DataLogger) in LTE TDD mode. Input DL_DATA? ,; Output ,,,,,, ,,,, ,,, ,,,,, ,,,, ,,,,,,,, , ,,,;,,, ,,,,, ,,,, {,,,} , {,,,,, {,,,, {Value,} } ; Input Parameter List Parameter Type Description Short Index of the selected dataset Short Index of the selected sub-dataset

Command Reference Guide

Narda SRM-3006 47 Output Parameter List, Section: GENERAL Parameter Type Description DWord Unique ID of the sub-dataset Enum LTE_TDD Enum MAN Manual storing, executed by the user COND_FIRST Auto-Store at first TRUE condition COND_ALL Auto-Store for all TRUE conditions TIME Auto-Store, Timer controlled MR_USER Storing triggered by measurement routine, with user interaction required MR_NUM Storing triggered by measurement routine, event controlled (numerical counter) MR_TIME Storing triggered by measurement routine, event controlled by timer Date dd.mm.yy Time hh:mm:ss Enum YES, NO; Yes, if there is at least one result in the dataset exceeding the range limit. Overdriven results may have a very high uncertainty. Enum NO No GPS signal available ACTUAL GPS signal available FROZEN GPS signal disappeared, using last valid coordinates Enum GPS, DGPS (Differential GPS) Enum 3D, 2D Short 0 to 36 Double -9999.9 to +9999.9 (meter) Double -90.000 00 to +90.000 00 (decimal degrees) Double -180.000 00 to +180.000 00 (decimal degrees) Enum Voice comment available: YES, NO; String Text comment up to 40 characters String Device serial number Date Device calibration date, dd.mm.yy String Device firmware version String Cable serial number Date Cable calibration date, dd.mm.yy String Antenna serial number Date Antenna calibration date, dd.mm.yy

Command Reference Guide

48 SRM-3006 Narda Output Parameter List, Section: SETUP COMMON Parameter Type Description Float Reference level (Measurement Range) [Unit] Enum Unit Enum Unit category: A, B, C, D See DL_Data? (Level) Float Value required to determine Yref when changing between unit categories Enum X, Y, Z: Measurement of one selected spatial axis using a 3-axis antenna RSS: Isotropic measurement using a 3-axis antenna (Root Sum Square) SINGLE: Measurement without antenna or with a single axis antenna String Name of the safety standard being used String Name of the service table being used String Name of the cable being used String Name of the antenna being used Output Parameter List, Section: SETUP MODE Parameter Type Description Double Center frequency [Hz] Double Channel Bandwidth [Hz] 1.4e6, 3e6, 5e6, 10e6, 15e6, 20e6 Enum NUMBER, TIME Integer Time [s]: 60, 120, 180, 240, 300, 360, 420, 480, 540, 600, 660, 720, 780, 840, 900, 960, 1020, 1080, 1140, 1200, 1260, 1320, 1380, 1440, 1500, 1560, 1620, 1680, 1740, 1800 Integer Number of averaging samples: 4, 8, 16, 32, 64, 128, 256 Integer 0, 3, 6, 10, 15, 20 [dB] Increase of the reference noise level related to the typical noise power level. Enum OFF Measurement results are always provided. ON Measurement results are provided only when the reference noise level is exceeded. Otherwise the result is indicated as “< value of reference noise level”. Float Power correction factor; Default: 1.0 Enum ON, OFF Enum 0,1,2,3,4,5,6 Enum Cyclic Prefix Length: NORMAL, EXTENDED Enum Cell Synchronization: SYNC, NO_SYNC

Command Reference Guide

Narda SRM-3006 49 Output Parameter List, Section: TRACE COMMON Parameter Type Description Long Number of sweeps performed Short Sweep Time [ms] Short 0 – 100 [%] Progress of averaging Short >= 0 Short 0 4 503 Short >= 1 Enum YES, NO Short Number of traces Output Parameter List, Section: TRACE SPECIFIC Parameter Type Description Enum Available traces: ACT Actual value AVG Averaged value MAX Maximum value MAX_AVG Maximum of the averaged values MIN Minimum value MIN_AVG Minimum of the averaged values STD Values of the selected standard Enum YES, NO; Yes, if at least one trace exceeded the range limit. Overdriven results may have a very high uncertainty. Float Analog power [Unit] Enum State of the analog noise flag: UNCHECKED, LOW, OK Short 149 Enum PSS,SSS,RS_AVG,RS_SUM,RS_MAX, RS_0,RS_1,RS_2,RS_3 Enum YES, NO Float Total power [Unit] of selected cell IDs Shorts Number of values = NoOfCellIDs There is one value for each cell ID available. Float Measured power for the current cell ID and Signal [Unit] DWord Error code

Command Reference Guide

50 SRM-3006 Narda Example (Input) DL_DATA? 7,1; (Output) 1913414,LTE_TDD,MAN,18.06.14, 9:50:16,NO,ACTUAL,GPS,3D,6,504,48.4585316667,9.23057166667,NO,"", "D-0051",14.09.10,"V1.3.2","",01.01.01,"",01.01.01, -30,dBm,A,0,SINGLE,"ICNIRP 1998 General Public","Germany alle Funkdienste","","", 796000000,1400000,NUMBER,360,4,0,OFF,1,OFF,0,NORMAL,SYNC, 5,229,100,3, 335,2,YES, 334,2,YES, 333,2,YES, 1, ACT,YES,-62.13,UNCHECKED,9, PSS,YES,-72.78,3,-78.27,-74.23,-999.00, SSS,YES,-73.35,3,-79.88,-74.44,-999.00, RS_AVG,YES,-74.85,3,-80.51,-76.23,-999.00, RS_MAX,YES,-73.12,3,-78.10,-74.77,-999.00, RS_SUM,YES,-71.84,3,-77.50,-73.22,-999.00, RS_0,YES,-75.26,3,-78.10,-78.44,-999.00, RS_1,YES,-74.48,3,-86.37,-74.77,-999.00, RS_2,YES,-999.00,3,-999.00,-999.00,-999.00, RS_3,YES,-999.00,3,-999.00,-999.00,-999.00, 0; Example (Input) DL_DATA? 8,1; (Output) 1936150,LTE_TDD,MAN,18.06.14, 9:50:39,NO,ACTUAL,GPS,3D,6,511.5,48.4585966667,9.23044833333,NO,"", "D-0051",14.09.10,"V1.3.2","",01.01.01,"",01.01.01, -30,dBm,A,0,SINGLE,"ICNIRP 1998 General Public","Germany alle Funkdienste","","", 796000000,1400000,NUMBER,360,4,0,OFF,1,OFF,0,NORMAL,SYNC, 5,235,100,3, 335,2,YES, 334,2,NO, 333,2,NO, 1, ACT,NO,-60.91,UNCHECKED,9, PSS,NO,-74.29,3,-74.29,-72.72,-999.00, SSS,NO,-75.50,3,-75.50,-73.19,-999.00, RS_AVG,YES,-77.68,3,-77.68,-75.45,-999.00, RS_MAX,NO,-76.82,3,-76.82,-74.78,-999.00, RS_SUM,NO,-74.67,3,-74.67,-72.44,-999.00, RS_0,NO,-76.82,3,-76.82,-76.25,-999.00, RS_1,NO,-78.76,3,-78.76,-74.78,-999.00, RS_2,NO,-999.00,3,-999.00,-999.00,-999.00, RS_3,NO,-999.00,3,-999.00,-999.00,-999.00, 0;

Command Reference Guide

Narda SRM-3006 51 DL_INFO? Description Queries overview information for a data set. Input DL_INFO? ; Output ,,,,

Command Reference Guide

52 SRM-3006 Narda DL_INFO_SUB? Description Queries overview information for a sub data set. Input DL_INFO? ,<>SubDS_Idx; Output ,,,

Command Reference Guide

Narda SRM-3006 53 DL_MEMORY? Description Queries the free memory space available for storing data sets. The response is a relative number expressed in %. Input DL_MEMORY?; Output ,; Parameter List Parameter Type Description Integer Free memory space, 0 - 100 [%] DWord Error code Example (Input) DL_MEMORY?; (Output) 37,0;

Command Reference Guide

54 SRM-3006 Narda DL_NUMBER? Description Queries the number of saved data sets. Input DL_NUMBER?; Output ,; Parameter List Parameter Type Description Integer Number of stored datasets DWord Error code Example (Input) DL_NUMBER?; (Output) 33,0;

Command Reference Guide

Narda SRM-3006 55 DL_STORE Description Saves a data set in the current measurement mode. Input DL_STORE; Output ; Parameter List Parameter Type Description DWord Error code Example (Input) DL_STORE; (Output) 0;

Command Reference Guide

56 SRM-3006 Narda DL_VOICE? Description Queries the voice comment of a data set in PCM wave format. Input DL_VOICE? ,; Output ,,; Parameter List Parameter Type Description Integer Index of the desired dataset Long Number of characters per line (block-wise transfer) 0 no vertical format 64 64 characters per line up to 65533 characters per line Long Number of binary bytes (before conversion to HEX) ASCII HEX Binary bytes as HEX strings DWord Error code Example (Input) DL_VOICE? 37,32 (Output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

Command Reference Guide

Narda SRM-3006 57 ERROR? Description Queries the current error code. Input ERROR?; Output ,; Parameter List Parameter Type Description Enum Current error code: 0 no error 401 remote command is not implemented in the remote module 402 invalid parameter 403 invalid count of parameters 404 invalid parameter range 405 last command is not completed 406 answer time between remote module and application module is too high 407 invalid or corrupt data 408 error while accessing the hardware 409 command is not supported in this version of the application module 410 remote is not activated (please send "REMOTE ON;" first) 411 command is not supported in the selected mode 412 memory of data logger is full 413 option code is invalid 414 incompatible version 415 sub-index full 416 file counter full 417 data lost 418 checksum error 419 programming not successful 420 path not found 421 break detected 422 low battery 423 file open error 424 data verify error DWord Error code for this command Example (Input) ERROR?; (Output) 401,0;

Command Reference Guide

58 SRM-3006 Narda LEVEL? Description Queries the different measurement runs (traces) that are available in LevelRecorder mode. LevelRecorder mode must be selected as the current operating mode. Input LEVEL? ; Output ,,,, ,,,,; Parameter List Parameter Type Description Enum RMS, MAX_RMS, PEAK, MAX_PEAK, STD, ALL Integer Number of sweeps performed Short 0 – 100 [%] Progress of averaging Integer 0 – 999 999 Number of sweeps used for spatial averaging Short Number of traces Enum YES, NO; Yes, if at least one trace exceeded the range limit. Overdriven results may have a very high uncertainty. Enum UNCHECKED Result has not been compared with noise level LOW Result is below the noise level OK Result is equal or above the noise level Float Measurement result [Unit] DWord Error code Example (Input) LEVEL? RMS; (Output) 74,100,0,1, RMS,NO,UNCHECKED,-31.07009, 0; Example (Input) LEVEL? ALL; (Output) 20672,100,0,4, RMS,NO,UNCHECKED,-30.99041, MAX_RMS,NO,UNCHECKED,-30.96372, PEAK,NO,UNCHECKED,-18.25535, MAX_PEAK,NO,UNCHECKED,-16.39886, 0;

Command Reference Guide

Narda SRM-3006 59 LEVEL_AVG_CONFIG Description Sets the current averaging method. The device has 1 averaging method: • Average over a time period (Time) Input LEVEL_AVG_CONFIG ; Output ; Parameter List Parameter Type Description Float Time [s]: 0.48, 0.96, 1.2, 2.4, 3.6, 6, 12, 18, 30 ,60, 120, 180, 300, 360, 600, 900, 1200, 1800 DWord Error code Example (Input) LEVEL_AVG_CONFIG 2.4; (Output) 0;

Command Reference Guide

60 SRM-3006 Narda LEVEL_AVG_CONFIG? Description Queries the current averaging method. The device has 1 averaging methods: • Average over a time period (Time) Input LEVEL_AVG_CONFIG?; Output ,; Parameter List Parameter Type Description Float Time [s]: 0.48, 0.96, 1.2, 2.4, 3.6, 6, 12, 18, 30 ,60, 120, 180, 300, 360, 600, 900, 1200, 1800 DWord Error code Example (Input) LEVEL_AVG_CONFIG?; (Output) 2.4,0;

Command Reference Guide

Narda SRM-3006 61 LEVEL_AVG_LIST? Description Queries the current possible averaging time settings for LevelRecorder mode. The averaging method for LevelRecorder mode is averaging over a time period. The output is a complete list of valid settings for this mode. can be easily used in list boxes and ensures the same look and feel across Narda’s analyzers. can be used in other commands to change the setting. Input LEVEL_AVG_LIST? ; Output ,{, ,}; Parameter List Parameter Type Description Short Number of subsequent items String Averaging time and its unit as string (in „s“ or „min“) Float Averaging time as value [s] DWord Error code Example (Input) LEVEL _AVG_LIST?; (Output) 18, "0.48 s",0.48, "0.96 s",0.96, "1.2 s",1.2, 4 "15 min",900, "20 min",1200, "30 min",1800, 0;

Command Reference Guide

62 SRM-3006 Narda LEVEL_CONFIG Description Sets the current measurement parameters for LevelRecorder mode. LevelRecorder mode must be selected as the current operating mode. Input LEVEL_CONFIG ,,,,; Output ; Parameter List Parameter Type Description Double Center frequency [Hz] Double Resolution bandwidth [Hz] Enum Video filter: ON, OFF; Double Video bandwidth [Hz] Float Reference level (Measurement Range) [Unit] DWord Error code Example (Input) LEVEL_CONFIG 333000000.7,250000,OFF,50000,0.4; (Output) 0;

Command Reference Guide

Narda SRM-3006 63 LEVEL_CONFIG? Description Queries the current measurement parameters for LevelRecorder mode. LevelRecorder mode must be selected as the current operating mode. Input LEVEL_CONFIG?; Output ,,,,,; Parameter List Parameter Type Description Double Center frequency [Hz] Double Resolution bandwidth [Hz] Enum Video filter: ON, OFF; Double Video bandwidth [Hz] Float Reference level (Measurement Range) [Unit] DWord Error code Example (Input) LEVEL_CONFIG?; (Output) 1500000000,5000000,OFF,50000,20,0;

Command Reference Guide

64 SRM-3006 Narda LIVESCREEN? Description Queries the current device display as a screenshot in PNG format. Individual values are themselves represented by binary numbers in the range 0-255 (0x00 – 0xFF). The values are transmitted in ASCII-HEX. Example: the value 0x89 is transmitted as ‘8’ ‘9’, and the associated HEX representation is 0x38 0x39. The complete and consistent quantity of all values can be converted back into binary data so that the device display can be shown straightaway on a monitor as a PNG file. Note Command also useable in “Remote OFF” mode Input LIVESCREEN? ; Output ,,; Parameter List Parameter Type Description Long Number of characters per line (block-wise transfer) 0 no vertical format 64 64 characters per line up to 65533 characters per line Long Number of binary bytes (before conversion to HEX) ASCII HEX Binary bytes as HEX strings DWord Error code Example (Input) LIVESCREEN? 32; (Output) 26453, 89504E470D0A1A0A0000000D49484452 00000320000001E00802000000D2659E A2000000017352474200AECE1CE90000 4 4 011418A8C0FF076DBD652BB8F321CF00 00000049454E44AE426082 ,0;

Command Reference Guide

Narda SRM-3006 65 LTE? The LTE commands applies for the LTE FDD optional mode Description Queries the different measurement runs (traces) that are available in LTE mode. LTE must be selected as the current operating mode. Input LTE? ; Output ,,,, {,,,} , {,,,,, {,,,, {Value,} } } ; Input Parameter List Parameter Type Description Enum ACT Actual value AVG Averaged value MAX Maximum value MAX_AVG Maximum of the averaged values MIN Minimum value MIN_AVG Minimum of the averaged values STD Values of the selected standard ALL All traces

Command Reference Guide

66 SRM-3006 Narda Output Parameter List Parameter Type Description Long Number of sweeps performed Short Sweep Time [ms] Short 0 – 100 [%] Progress of averaging Short >= 0 Short 0 4 503 Short >= 1 Enum YES, NO Short Number of traces Enum Available traces: ACT Actual value AVG Averaged value MAX Maximum value MAX_AVG Maximum of the averaged values MIN Minimum value MIN_AVG Minimum of the averaged values STD Values of the selected standard Enum YES, NO; Yes, if at least one trace exceeded the range limit. Overdriven results may have a very high uncertainty. Float Analog power [Unit] Enum State of the analog noise flag: UNCHECKED, LOW, OK Short 149 Enum PSS,SSS,RS_AVG,RS_SUM,RS_MAX, RS_0,RS_1,RS_2,RS_3 Enum YES, NO Float Total power [Unit] of selected cell IDs Shorts Number of values = NoOfCellIDs There is one value for each cell ID available. Float Measured power for each cell ID and Signal [Unit] DWord Error code

Command Reference Guide

Narda SRM-3006 67 Example (Input) LTE? ACT; (Output) 5,229,100,3, 335,2,YES, 334,2,YES, 333,2,YES, 1, ACT,YES,-62.13,UNCHECKED,9, PSS,YES,-72.78,3,-78.27,-74.23,-999.00, SSS,YES,-73.35,3,-79.88,-74.44,-999.00, RS_AVG,YES,-74.85,3,-80.51,-76.23,-999.00, RS_MAX,YES,-73.12,3,-78.10,-74.77,-999.00, RS_SUM,YES,-71.84,3,-77.50,-73.22,-999.00, RS_0,YES,-75.26,3,-78.10,-78.44,-999.00, RS_1,YES,-74.48,3,-86.37,-74.77,-999.00, RS_2,YES,-999.00,3,-999.00,-999.00,-999.00, RS_3,YES,-999.00,3,-999.00,-999.00,-999.00, 0; Example (Input) LTE? ACT; (Output) 5,235,100,3, 335,2,YES, 334,2,NO, 333,2,NO, 1, ACT,NO,-60.91,UNCHECKED,9, PSS,NO,-74.29,3,-74.29,-72.72,-999.00, SSS,NO,-75.50,3,-75.50,-73.19,-999.00, RS_AVG,YES,-77.68,3,-77.68,-75.45,-999.00, RS_MAX,NO,-76.82,3,-76.82,-74.78,-999.00, RS_SUM,NO,-74.67,3,-74.67,-72.44,-999.00, RS_0,NO,-76.82,3,-76.82,-76.25,-999.00, RS_1,NO,-78.76,3,-78.76,-74.78,-999.00, RS_2,NO,-999.00,3,-999.00,-999.00,-999.00, RS_3,NO,-999.00,3,-999.00,-999.00,-999.00, 0;

Command Reference Guide

68 SRM-3006 Narda LTE_AVG_CONFIG The LTE commands applies for the LTE FDD optional mode Description Sets the current averaging method. The device has 2 averaging methods: • Average over a number of individual measurements (Number) • Average over a time period (Time) There is one setting parameter for each method, which is selected from a basic set of values. Input LTE_AVG_CONFIG ,,; Output ; Parameter List Parameter Type Description Enum NUMBER TIME Integer Number of averaging samples: 4, 8, 16, 32, 64, 128, 256 Integer Time [s]: 60, 120, 180, 240, 300, 360, 420, 480, 540, 600, 660, 720, 780, 840, 900, 960, 1020, 1080, 1140, 1200, 1260, 1320, 1380, 1440, 1500, 1560, 1620, 1680, 1740, 1800 DWord Error code Example (Input) LTE_AVG_CONFIG NUMBER,64,240; (Output) 0; Example (Input) LTE_AVG_CONFIG TIME,4,180; (Output) 0;

Command Reference Guide

Narda SRM-3006 69 LTE_AVG_CONFIG? The LTE commands applies for the LTE FDD optional mode Description Queries the current averaging method. The device has 2 averaging methods: • Average over a number of individual measurements (Number) • Average over a time period (Time) There is one setting parameter for each method, which is selected from a basic set of values. Input LTE_AVG_CONFIG?; Output ,,,; Parameter List Parameter Type Description Enum NUMBER TIME Integer Number of averaging samples: 4, 8, 16, 32, 64, 128, 256 Integer Time [s]: 60, 120, 180, 240, 300, 360, 420, 480, 540, 600, 660, 720, 780, 840, 900, 960, 1020, 1080, 1140, 1200, 1260, 1320, 1380, 1440, 1500, 1560, 1620, 1680, 1740, 1800 DWord Error code Example (Input) LTE_AVG_CONFIG?; (Output) TIME,4,120,0; Example (Input) LTE_AVG_CONFIG?; (Output) NUMBER,256,180,0;;

Command Reference Guide

70 SRM-3006 Narda LTE_AVG_LIST? The LTE commands applies for the LTE FDD optional mode Description Queries the current possible averaging settings for LTE mode. The device has 2 averaging methods: • Average over a number of individual measurements (Number) • Average over a time period (Time) The output is a complete list of valid settings for this mode. can be easily used in list boxes and ensures the same look and feel across Narda’s analyzers. can be used in other commands to change the setting. Input LTE_AVG_LIST? ; Output ,{, ,}; Parameter List Parameter Type Description Enum NUMBER: Average over a number of individual measurements TIME: Average over a time period (in seconds) Short Number of subsequent items String Averaging parameter as string including the unit Float Averaging parameter as a numerical value DWord Error code Example (Input) LTE_AVG_LIST? TIME; (Output) 30, "1 min",60, "2 min",120, "3 min",180, 4 "28 min",1680, "29 min",1740, "30 min",1800, 0;

Command Reference Guide

Narda SRM-3006 71 LTE_CONFIG The LTE commands applies for the LTE FDD optional mode Description Sets the current measurement parameters for LTE mode. LTE must be selected as the current operating mode. Input LTE_CONFIG ,,,,,,; Output ; Parameter List Parameter Type Description Double Center frequency [Hz] Double Channel Bandwidth [Hz] 1.4e6, 3e6, 5e6, 10e6, 15e6, 20e6 Enum Cyclic Prefix Length: NORMAL, EXTENDED Enum Cell Synchronization: SYNC, NO_SYNC Enum ON, OFF Float 1.0 to 10000.0 Float Reference level (Measurement Range) [Unit] DWord Error code Example (Input) LTE_CONFIG 806000000,1400000,NORMAL,SYNC,OFF,1.0,-10; (Output) 0;

Command Reference Guide

72 SRM-3006 Narda LTE_CONFIG? The LTE commands applies for the LTE FDD optional mode Description Queries the current measurement parameters for LTE mode. LTE must be selected as the current operating mode. A list of all CBWs available can be queried with the RBW_LIST? command. Input LTE_CONFIG?; Output ,,,,,,, ; Parameter List Parameter Type Description Double Center frequency [Hz] Double Channel Bandwidth [Hz] 1.4e6, 3e6, 5e6, 10e6, 15e6, 20e6 Enum Cyclic Prefix Length: NORMAL, EXTENDED Enum Cell Synchronization: SYNC, NO_SYNC Enum ON, OFF Float 1.0 to 10000.0 Float Reference level (Measurement Range) [Unit] DWord Error code Example (Input) LTE_CONFIG?; (Output) 806000000,1400000,NORMAL,SYNC,OFF,1.0,-10,0;

Command Reference Guide

Narda SRM-3006 73 LTE_RESET_TABLE The LTE commands applies for the LTE FDD optional mode Description Resets the table for LTE results (Cell IDs). Input LTE_RESET_TABLE; Output ; Parameter List Parameter Type Description DWord Error code Example (Input) LTE_RESET_TABLE; (Output) 0;

Command Reference Guide

74 SRM-3006 Narda LTE_TDD? Description Queries the different measurement runs (traces) that are available in LTE TDD mode. LTE TDD must be selected as the current operating mode. Input LTE_TDD? ; Output ,,,, {,,,} , {,,,,, {,,,, {Value,} } } ; Input Parameter List Parameter Type Description Enum ACT Actual value AVG Averaged value MAX Maximum value MAX_AVG Maximum of the averaged values MIN Minimum value MIN_AVG Minimum of the averaged values STD Values of the selected standard ALL All traces

Command Reference Guide

Narda SRM-3006 75 Output Parameter List Parameter Type Description Long Number of sweeps performed Short Sweep Time [ms] Short 0 – 100 [%] Progress of averaging Short >= 0 Short 0 4 503 Short >= 1 Enum YES, NO Short Number of traces Enum Available traces: ACT Actual value AVG Averaged value MAX Maximum value MAX_AVG Maximum of the averaged values MIN Minimum value MIN_AVG Minimum of the averaged values STD Values of the selected standard Enum YES, NO; Yes, if at least one trace exceeded the range limit. Overdriven results may have a very high uncertainty. Float Analog power [Unit] Enum State of the analog noise flag: UNCHECKED, LOW, OK Short 149 Enum PSS,SSS,RS_AVG,RS_SUM,RS_MAX, RS_0,RS_1,RS_2,RS_3 Enum YES, NO Float Total power [Unit] of selected cell ids Shorts Number of values = NoOfCellIDs There is one value for each cell ID available. Float Measured power for each cell ID and Signal [Unit] DWord Error code

Command Reference Guide

76 SRM-3006 Narda Example (Input) LTE_TDD? ACT; (Output) 5,229,100,3, 335,2,YES, 334,2,YES, 333,2,YES, 1, ACT,YES,-62.13,UNCHECKED,9, PSS,YES,-72.78,3,-78.27,-74.23,-999.00, SSS,YES,-73.35,3,-79.88,-74.44,-999.00, RS_AVG,YES,-74.85,3,-80.51,-76.23,-999.00, RS_MAX,YES,-73.12,3,-78.10,-74.77,-999.00, RS_SUM,YES,-71.84,3,-77.50,-73.22,-999.00, RS_0,YES,-75.26,3,-78.10,-78.44,-999.00, RS_1,YES,-74.48,3,-86.37,-74.77,-999.00, RS_2,YES,-999.00,3,-999.00,-999.00,-999.00, RS_3,YES,-999.00,3,-999.00,-999.00,-999.00, 0; Example (Input) LTE_TDD? ACT; (Output) 5,235,100,3, 335,2,YES, 334,2,NO, 333,2,NO, 1, ACT,NO,-60.91,UNCHECKED,9, PSS,NO,-74.29,3,-74.29,-72.72,-999.00, SSS,NO,-75.50,3,-75.50,-73.19,-999.00, RS_AVG,YES,-77.68,3,-77.68,-75.45,-999.00, RS_MAX,NO,-76.82,3,-76.82,-74.78,-999.00, RS_SUM,NO,-74.67,3,-74.67,-72.44,-999.00, RS_0,NO,-76.82,3,-76.82,-76.25,-999.00, RS_1,NO,-78.76,3,-78.76,-74.78,-999.00, RS_2,NO,-999.00,3,-999.00,-999.00,-999.00, RS_3,NO,-999.00,3,-999.00,-999.00,-999.00, 0;

Command Reference Guide

Narda SRM-3006 77 LTE_TDD_AVG_CONFIG Description Sets the current averaging method. The device has 2 averaging methods: • Average over a number of individual measurements (Number) • Average over a time period (Time) There is one setting parameter for each method, which is selected from a basic set of values. Input LTE_TDD_AVG_CONFIG ,,; Output ; Parameter List Parameter Type Description Enum NUMBER TIME Integer Number of averaging samples: 4, 8, 16, 32, 64, 128, 256 Integer Time [s]: 60, 120, 180, 240, 300, 360, 420, 480, 540, 600, 660, 720, 780, 840, 900, 960, 1020, 1080, 1140, 1200, 1260, 1320, 1380, 1440, 1500, 1560, 1620, 1680, 1740, 1800 DWord Error code Example (Input) LTE_TDD_AVG_CONFIG NUMBER,64,240; (Output) 0; Example (Input) LTE_TDD_AVG_CONFIG TIME,4,180; (Output) 0;

Command Reference Guide

78 SRM-3006 Narda LTE_TDD_AVG_CONFIG? Description Queries the current averaging method. The device has 2 averaging methods: • Average over a number of individual measurements (Number) • Average over a time period (Time) There is one setting parameter for each method, which is selected from a basic set of values. Input LTE_TDD_AVG_CONFIG?; Output ,,,; Parameter List Parameter Type Description Enum NUMBER TIME Integer Number of averaging samples: 4, 8, 16, 32, 64, 128, 256 Integer Time [s]: 60, 120, 180, 240, 300, 360, 420, 480, 540, 600, 660, 720, 780, 840, 900, 960, 1020, 1080, 1140, 1200, 1260, 1320, 1380, 1440, 1500, 1560, 1620, 1680, 1740, 1800 DWord Error code Example (Input) LTE_TDD_AVG_CONFIG?; (Output) TIME,4,120,0; Example (Input) LTE_TDD_AVG_CONFIG?; (Output) NUMBER,256,180,0;;

Command Reference Guide

Narda SRM-3006 79 LTE_TDD_AVG_LIST? Description Queries the current possible averaging settings for LTE TDD mode. The device has 2 averaging methods: • Average over a number of individual measurements (Number) • Average over a time period (Time) The output is a complete list of valid settings for this mode. can be easily used in list boxes and ensures the same look and feel across Narda’s analyzers. can be used in other commands to change the setting. Input LTE_TDD_AVG_LIST? ; Output ,{, ,}; Parameter List Parameter Type Description Enum NUMBER: Average over a number of individual measurements TIME: Average over a time period (in seconds) Short Number of subsequent items String Averaging parameter as string including the unit Float Averaging parameter as a numerical value DWord Error code Example (Input) LTE_TDD_AVG_LIST? TIME; (Output) 30, "1 min",60, "2 min",120, "3 min",180, 4 "28 min",1680, "29 min",1740, "30 min",1800, 0;

Command Reference Guide

80 SRM-3006 Narda LTE_TDD_CONFIG Description Set the current measurement parameters for LTE TDD mode. LTE TDD must be selected as the current operating mode. Input LTE_TDD_CONFIG ,,,,,,,; Output ; Parameter List Parameter Type Description Double Center frequency [Hz] Double Channel Bandwidth [Hz] 1.4e6, 3e6, 5e6, 10e6, 15e6, 20e6 integer 0,1,2,3,4,5,6 Enum Cyclic Prefix Length: NORMAL, EXTENDED Enum Cell Synchronization: SYNC, NO_SYNC Enum ON, OFF Float 1.0 to 10000.0 Float Reference level (Measurement Range) [Unit] DWord Error code Example (Input) LTE_TDD_CONFIG 806000000,1400000,0,NORMAL,SYNC,OFF,1.0,-10; (Output) 0;

Command Reference Guide

Narda SRM-3006 81 LTE_TDD_CONFIG? Description Query the current measurement parameters for LTE TDD mode. LTE TDD must be selected as the current operating mode. A list of all CBWs available can be queried with the RBW_LIST? command. Input LTE_TDD_CONFIG?; Output ,,,,,,,, ; Parameter List Parameter Type Description Double Center frequency [Hz] Double Channel Bandwidth [Hz] 1.4e6, 3e6, 5e6, 10e6, 15e6, 20e6 integer 0,1,2,3,4,5,6 Enum Cyclic Prefix Length: NORMAL, EXTENDED Enum Cell Synchronization: SYNC, NO_SYNC Enum ON, OFF Float 1.0 to 10000.0 Float Reference level (Measurement Range) [Unit] DWord Error code Example (Input) LTE_TDD_CONFIG?; (Output) 806000000,1400000,0,NORMAL,SYNC,OFF,1.0,-10,0;

Command Reference Guide

82 SRM-3006 Narda LTE_TDD_RESET_TABLE Description Resets the table for LTE TDD results (Cell IDs). Input LTE_TDD_RESET_TABLE; Output ; Parameter List Parameter Type Description DWord Error code Example (Input) LTE_TDD_RESET_TABLE; (Output) 0;

Command Reference Guide

Narda SRM-3006 83 MEAS_START Description Starts the cyclical measurement in the current operating mode. A started measurement is recognized by a query, e.g. SPECTRUM? and observation of the SweepCounter, which increments when the measurement is started. Input MEAS_START; Output ; Parameter List Parameter Type Description DWord Error code Example (Input) MEAS_START; (Output) 0;

Command Reference Guide

84 SRM-3006 Narda MEAS_STOP Description Stops the cyclical measurement in the current operating mode. A stopped measurement is recognized by a query, e.g. SPECTRUM? and observation of the SweepCounter, which stops incrementing when the measurement is stopped. Input MEAS_STOP; Output ; Parameter List Parameter Type Description DWord Error code Example (Input) MEAS_STOP; (Output) 0;

Command Reference Guide

Narda SRM-3006 85 MODE Description Sets the current operating mode. Input MODE ; Output ; Parameter List Parameter Type Description Enum SPECTRUM SAFETY UMTS SCOPE LEVEL LTE LTE_TDD DWord Error code Example (Input) MODE SPECTRUM; (Output) 0;

Command Reference Guide

86 SRM-3006 Narda MODE? Description Queries the current operating mode. Input MODE?; Output ,, Parameter List Parameter Type Description Enum SPECTRUM SAFETY UMTS SCOPE LEVEL LTE LTE_TDD DWord Error code Example (Input) MODE?; (Output) SPECTRUM,0;

Command Reference Guide

Narda SRM-3006 87 MR_LIST? Description Queries the current possible reference level settings (measurement range). The requested list reflects all the possible settings and guarantees consistency with the other device settings. The parameter can be used directly in the application to produce the same appearance as in the device (e.g. in List boxes). The parameter is a numerical value, which can be transmitted to the device if required in order to produce the corresponding setting. Input MR_LIST?; Output ,{, ,}; Parameter List Parameter Type Description Short Number of subsequent items String Reference level (measurement range) and unit as string Float Reference level as value [Unit] DWord Error code Example (Input) MR_LIST?; (Output) 51, "67 dBmV",67, "66 dBmV",66, "65 dBmV",65, "64 dBmV",64, 4 4 "19 dBmV",19, "18 dBmV",18, "17 dBmV",17, 0;

Command Reference Guide

88 SRM-3006 Narda PROG_EXIT Description Ends the measurement program and switches the device to standby mode. Input PROG_EXIT; Output ; Parameter List Parameter Type Description DWord Error code Example (Input) PROG_EXIT; (Output) 0;

Command Reference Guide

Narda SRM-3006 89 RBW_LIST? Description Queries the current possible RBW settings. The requested list reflects all the possible settings and guarantees consistency with the other device settings. The parameter can be used directly in the application to produce the same appearance as in the device (e.g. in List boxes). The parameter is a numerical value, which can be transmitted to the device if required in order to produce the corresponding setting. Input RBW_LIST? Output ,{, ,}; Parameter List Parameter Type Description Short Number of subsequent items String Resolution bandwidth RBW and unit as string Double Resolution bandwidth RBW as value [Hz] DWord Error code Example (Input) RBW_LIST?; (Output) 13, "100 kHz",100000, "50 kHz",50000, "30 kHz",30000, "20 kHz",20000, "10 kHz",10000, "5 kHz",5000, "3 kHz",3000, "2 kHz",2000, "1 kHz",1000, "500 Hz",500, "300 Hz",300, "200 Hz",200, "100 Hz",100, 0;

Command Reference Guide

90 SRM-3006 Narda REMOTE Description Activates / deactivates remote communication with the SRM. “REMOTE ON” is the condition required for utilizing remote access through all device interfaces. Note Command also useable in “Remote OFF” mode Input: REMOTE ; Output ; Parameter List Parameter Type Description Enum ON = Enable remote operation mode OFF = Disable remote operation mode DWord Error code Example (Input) REMOTE ON; (Output) 0;

Command Reference Guide

Narda SRM-3006 91 REMOTE? Description Checks communication with the SRM, and queries readiness for remote operation. “REMOTE ON” is the condition required for utilizing remote access through all device interfaces. Note Command also useable in “Remote OFF” mode Input REMOTE?; Output ,; Parameter List Parameter Type Description Enum ON = Remote operation mode is enabled OFF = Remote operation mode is disabled DWord Error code Example (Input) REMOTE?; (Output) OFF, 0;

Command Reference Guide

92 SRM-3006 Narda RESULT_TYPE_RESET Description Resets all measurement runs (traces) to default values as is done at the start of the measurement. Input RESULT_TYPE_RESET; Output ; Parameter List Parameter Type Description DWord Error code Example (Input) RESULT_TYPE_RESET; (Output) 0;

Command Reference Guide

Narda SRM-3006 93 RESULT_TYPE_RESET_MIN_MAX Description Resets the Min and Max measurement runs (traces) to default values as is done at the start of the measurement. Input RESULT_TYPE_RESET_MIN_MAX; Output ; Parameter List Parameter Type Description DWord Error code Example (Input) RESULT_TYPE_RESET_MIN_MAX; (Output) 0;

Command Reference Guide

94 SRM-3006 Narda SAFETY? Description Queries one or more measurement runs (traces) in SafetyEvaluation mode. A measurement run is the number of all the measurement values determined during a measurement cycle (sweep). SafetyEvaluation must be selected as the current operating mode. Traces can only be queried when the device is in “Detailed” table view mode. “Condensed” table view mode is not supported in remote operation. Input SAFETY?; Output ,,,, ,,, ,,,,,, ,{,,,,,};