Fill Water Sample Collecting Bottle (0688) with sample water.

2. Press and hold until colorimeter turns on. 3. Press ENTER to select Testing Menu. 4. Scroll to and select 008 Ammonia-N HR from menu. 5. Rinse tube (0290) with sample water. Fill to the 10 mL line. Cap and wipe dry. 6. Insert tube. Close lid. Select Scan Blank. 7. Remove tube. Add 8 drops of Ammonia Nitrogen Reagent #1. Cap and mix. 8. Use 1.0 mL pipet (0354) to add 1.0 mL of *Ammonia Nitrogen Reagent #2. 9. Cap and invert to mix. Wait 5 minutes for full color development. 10. Insert tube into chamber. Close lid. Select Scan Sample. Record result as AMMONIA-NITROGEN. See manual for calculations. *WARNING: Reagents marked with an * are considered to be potential health hazards. To view or print a Safety Data Sheet (SDS) for these reagents go to www.lamotte.com. Search for the four digit reagent code number listed on the reagent label, in the contents list or in the test procedures. Omit any letter that follows or precedes the four digit code number. For example, if the code is 4450WT-H, search 4450. To obtain a printed copy, contact LaMotte by email, phone or fax. Emergency information for all LaMotte reagents is available from Chem-Tel (US, 1-800-255-3924) (International, call collect, 813- 248-0585). DC1500 AMMONIA Quick Guide

Calibrate meter with ammonia nitrogen standards. Use a distilled or

deionized water blank and one ammonia nitrogen standard of known concentration. PROCEDURE 1. Press and hold until colorimeter turns on. 2. Press ENTER to select Testing Menu. 3. Scroll to and select 008 Ammonia-N HR from menu. 4. Rinse tube (0290) three times with the standard. Fill the tube to the 10 mL line. Cap tube. Dry tube. 5. Insert tube. Close lid. Select Scan Blank. 6. Remove tube. Add 8 drops of Ammonia Nitrogen Reagent #1. Cap and mix. 7. Use 1.0 mL pipet (0354) to add 1.0 mL of *Ammonia Nitrogen Reagent #2. 8. Cap and invert to mix. Wait 5 minutes for full color development. 9. Insert tube. Close lid. Select Scan Sample. 10. Press to scroll to Calibrate. 11. Press ENTER to select Calibrate. 12. Press or to scroll to the concentration of the standard. 13. Press ENTER to save the value. Press and ENTER at any time to return to the default value. For additional information see the DC1500 Ammonia manual. 802 Washington Ave • Chestertown • Maryland • 21620 • USA 410-778-3100 • 800-344-3100 • www.lamotte.com 3241-QG 10.14.16 DC1500 AMMONIA Quick Guide

Other Measuring Instruments Manuals