Text of Extech Instruments 407750 Operation & User’s Manual
Model 407750 Version 2.1 February 2004
2
Warranty
EXTECH INSTRUMENTS CORPORATION warrants this instrument to be free of defects in parts and
workmanship for one year from date of shipment (a six month limited warranty applies on sensors and
cables). If it should become necessary to return the instrument for service during or beyond the warranty
period, contact the Customer Service Department at (781) 890-7440 ext. 210 for authorization or visit our
website
Model 407750 Version 2.1 February 2004 3
Meter Description
1.
LCD Display
2.
Microphone
3.
ON/OFF key
4.
REC (Record) key
5.
MAXHLD (Max Hold) key
6.
C/A Weighting Select key
7.
BA (Background Absorber) key
8.
F/S Fast / Slow Response select key
9.
DOWN
10. Backlit (LCD backlighting) key
11. UPPER
12. AC adaptor jack
13. Calibration screw adjust
14. AC analog output jack
15. DC analog output jack
16. RS-232 output jack
Note: The Battery Compartment and the
Model 407750 Version 2.1 February 2004 4
Operation
Quick Start
1. Power the meter by pressing the ON/OFF key.
2. The meter’s LCD will count down to zero (99.9, 88.8, 77.7, etc.) and then begin
measuring sound levels. If the LCD does not switch on after pressing the ON/KEY
check the 9V battery.
3. Point the microphone toward the source of the sound level to be measured and view
the reading on the meter’s LCD.
‘A’ and ‘C’ Frequency Weighting
Sele
Model 407750 Version 2.1 February 2004 5
BA (Background Noise Absorber) Mode
The Background Noise Absorber allows the user to accurately measure equipment noise
by “eliminating” background noise. The Sound Level Meter first stores the background
noise as a reference level. From there, when a sound is measured, the display will show
the sound level measurement minus the background noise. To operate the meter in BA
mode, follow these steps:
1. Power the
Model 407750 Version 2.1 February 2004 6
Analog Outputs
The meter includes an AC and a DC analog output. These outputs are proportional to the
displayed sound level and are ideal for use with chart recorders and dataloggers. The AC
output is 0.707V rms full scale and the DC output is 10mV per dB. The labeled 3.5mm
output mini-plugs are located on the bottom of the instrument.
RS-232 Output
The meter includes an RS-232 PC interface jack. This PC interface a
Model 407750 Version 2.1 February 2004 7
Reference Information
Frequency Weighting Characteristics
Frequency (Hz)
A Weighting
C Weighting
Tolerance (IEC 651 Type 2)
31.5
-39.4dB
-3dB
±3dB
63
-26.2dB
-0.8dB
±2dB
125
-16.1dB
-0.2dB
±1.5dB
250
-8.6dB
0dB
±1.5dB
500
-3.2dB
0dB
±1.5dB
1 k
0dB
0dB
±1.5dB
2 k
+1.2dB
-0.2dB
±2dB
4 k
+1dB
-0.8dB
±3dB
8 k
-1.1dB
-3dB
±5dB
Typical A-Weighted Sound Levels
50HP Siren (100')
130
120
110
100
90
80
70
60
50
40
30
Model 407750 Version 2.1 February 2004 8
�
Support Hotline (781) 890-7440
Tech support: Ext. 200; Email: [email protected]
Repair/Returns: Ext. 210; Email: [email protected]
Website: www.extech.com
Battery Replacement
When the low battery message appears on the LCD, the 9V battery has fallen to a critically
low voltage level and should be replaced as soon as possible. The battery compartment
cover resides at the rear of the meter. Remove the rear batter