Excell PH3 Operation & User’s Manual
Read Online
Read the Excell PH3 Operation & User’s Manual online using the built-in PDF viewer.
Text of the manual
EXCELL PRECISION CO., LTD.
PH3 IP 68 Waterproof
Check-weighing Indicator
User Manual
© EXCELL PRECISION CO., LTD. 2018. All rights reserved worldwide.
The information contained herein is the property of EXCELL PRECISION CO., LTD.
and is supplied without liability for errors or omissions. No part may be reproduced or
used except as authorised by contract or other written permission. The copyright and
the foregoing restriction on reproduction and use extend to all media in which the
information may be embodied.
EXCELL PRECISION CO., LTD.
TABLE OF CONTENTS
BEFORE USING THE SCALE ..................................................................................................... 3
PRECAUTIONS FOR USE .......................................................................................................... 3
PREPARING TO USE THE SCALE ............................................................................................. 4
SPECIAL NOTICE ....................................................................................................................... 6
CHAPTER 1 INTRODUCTION .................................................................................................... 7
1-1 FEATURES ............................................................................................................................ 7
1-2 SPECIFICATIONS ................................................................................................................... 7
1-3 SCALE APPEARANCE ............................................................................................................. 8
1-4 POWER SUPPLY .................................................................................................................... 8
1-5 EXPLANATION OF DISPLAY ..................................................................................................... 9
1-6 EXPLANATION OF KEYPAD .................................................................................................... 10
CHAPTER 2 BASIC OPERATING FUNCTIONS ....................................................................... 12
2-1 SIMPLE WEIGHT MODE ........................................................................................................ 14
2-2 LIMIT WEIGHT MODE ........................................................................................................... 14
2-3 TARGET WEIGHT MODE ....................................................................................................... 16
2-4 TARE FUNCTION .................................................................................................................. 18
2-5 CANCEL TARE FUNCTION ..................................................................................................... 19
2-6 PRE-TARE FUNCTION .......................................................................................................... 19
2-7 CANCEL PRE-TARE FUNCTION .............................................................................................. 20
2-8 GROSS/NET FUNCTION ........................................................................................................ 20
2-9 MEMORY STORE AND RECALL FUNCTIONS ............................................................................. 21
CHAPTER 3 EXTERAL CALIBRATION ..................................................................................... 22
3-1
GENERAL FUNCTION SETTINGS ............................................................... 23
3-1-1
Automatic Backlight Function Settings....................................... 25
3-1-2
Automatic Power-off Timer Settings ........................................... 26
3-1-3
HI / LO Weight Check Settings .................................................. 27
3-1-4
HI / LO Alarm Settings Function ................................................ 28
3-1-5
Analog and Digital Display Settings ........................................... 29
3-1-6
Weighing Speed Settings ........................................................... 30
3-1-7
Operation Mode Settings ........................................................... 31
3-1-8
Weight and Percentage Mode Settings ..................................... 32
3-1-9
Lock Function Settings .............................................................. 33
3-1-10
Relay Settings ........................................................................... 34
3-1-11
Auto Zero Tracking .................................................................... 35
3-1-12
Printer Key Accumulation Function ............................................ 36
3-1-13
Tare Cancellation Settings ......................................................... 37
3-1-14
Previous Zero Record Settings ................................................. 38
EXCELL PRECISION CO., LTD.
3-1-15
HI / LO / OK Positive / Negative Weight Settings ........................ 38
3-1-16
HI / LO / OK Tare Time Settings .................................................. 39
3-2
WEIGHT CALIBRATION .................................................................................. 40
3-2-1
Weight Calibration ........................................................................... 41
3-2-2
Restore to the Default setting .......................................................... 42
3-3
RS-232 BI-DIRECTIONAL FUNCTION SETTINGS .......................................... 43
3-3-1
Baud Rate Settings ................................................................... 45
3-3-2
Communication Protocol Settings ............................................. 46
3-3-3
Output Format Settings ............................................................. 47
3-3-4
Continuous Transmission Settings ............................................ 48
3-3-5
Selection of the Continuous Transmission Rate ........................ 49
3-3-6
Output Condition Settings ......................................................... 50
3-3-7
Auto Transmission at Zero ........................................................ 51
3-3-8
Reset of Auto Transmission ....................................................... 52
3-3-9
Condition of Manually Printing ................................................... 54
3-3-10
Real Time Clock (Hour) ............................................................. 55
3-3-11
Real Time Clock (Date) ............................................................. 56
3-3-12
RS-232 Serial Interface Settings Mode ..................................... 57
3-3-13
RS-232 General or Simple Output Weight Selection ................. 58
APPENDIX 1 RS-232 SERIAL INTERFACE ........................................................................... 59
APPENDIX 2 ERROR MESSAGES ........................................................................................ 68
APPENDIX 3 7-SEGMENT DISPLAY CHARACTERS ............................................................ 69
EXCELL PRECISION CO., LTD.
PRECAUTIONS FOR USE
1.
Please keep the indicator in a cool place. Do not store it at a high temperature.
2.
Please keep it clean and avoid hosting of cockroach and any other possible live creatures.
3.
Avoid objects impacting with the indicator. Do not drop loads onto the scale or subject the
platter to any physically strong shock.
4.
The load placed on the platter must not exceed the maximum weighing capacity of the
scale.
5.
The indicator is IP68 waterproof design. Only the cables with Φ3 to 5.5mm caliber can be
used, or the indicator’s waterproofness will be affected.
6.
Clean and wrap the scale into a clean bag for long time no use. A desiccant sachet may be
included to prevent any moisture built up.
7.
To avoid current leakage, please do not subject the batteries to excessive heat or attempt
to open the batteries.
8.
In order to maintain the re-chargeable battery in good condition it should be kept fully
charged whenever possible. If the scale is to be stored, the battery should be fully charged
before storage, and then re-charged at 3 month intervals.
9.
The number of times that the battery can be recharged will vary according to the conditions
of use. However it can be maximized by re-charging the battery frequently and by avoiding
conditions of total discharge. The battery cannot be overcharged.
10. Please operate or charge the scale in an open area. Do not squeeze the power cord to
avoid wire on fire.
11. If the scale is not working properly, please send it to our distributor for service.
12. Any suggestion for the product is warmly welcome.
BEFORE USING THE SCALE
Thank you for purchasing an EXCELL Electronic
Digital Indicator. In order to use the indicator
properly, please read this User Manual carefully
before use. If you have a problem concerning the
indicator, please contact your supplier.
EXCELL PRECISION CO., LTD.
PREPARING TO USE THE SCALE
1.
Locate the scale on a firm level surface free from vibrations for accurate weight readings.
2.
Avoid operating the scale in direct sunlight or drafts of any kind.
3.
If possible, avoid connecting the scale to AC power outlet sockets which are adjacent to
other appliances to minimise the possibility of interference affecting the performance of the
scale.
4.
Remove any weight that might be on the platter before the scale is switched on and avoid
leaving weight on the platter for long periods of time.
5.
All goods weighed should be placed in the centre of the platter for accurate weighing. The
overall dimensions of the goods being weighed should not exceed the dimension of the
platter.
6.
Once the scale has been powered on, it will go through a LCD display test and it is ready
for use when the display shows zero.
7.
The scale requires 15 to 20 minutes warm up before operation to ensure best accuracy.
8.
When the power bar indication turns to be low battery , it means the battery
should be recharged immediately.
9.
Introduction of Storage Battery
EXCELL PRECISION CO., LTD.
No burning
Caution Corrosion
Due to the storage battery adopt the advanced free-maintaining technique, customers need not
to replenish electrolyte.
The scale should be recharged every 3 months to prevent failure of the internal rechargeable
battery.
1. The battery should be charged for 8~10 hours.
2. The temperature of battery should below 45°C.
Maintaining
1. Please do not discharge with over-current when using the battery. Please charge the battery
after discharging current.
2. Please take down the battery when the scale is not used for a long time or break the
connection of cathode.
3. Do not short the battery terminals to check whether there is current. Please check whether
the connection point is firm to guarantee good connection.
4. The battery should be replaced by specialized person. No reverse-battery or the product
will be damaged.
a) Anode of battery should be connected with Anode of product battery (usually red cable)
b) Cathode of battery should be connected with Cathode of product battery (usually brown
cable or black cable)
c) See the following picture:
Safety warnings
1. The electrolyte of battery is caustic which causes metal, cotton, etc to corrode.
2. The hydrogen will be resolved when using or charging the battery and it will cause explosion
when approaches fire.
Brown cable(or black cable)
connected with Anode of
battery
Red cable connected with
Cathode of battery
Warning explosion Children faraway
EXCELL PRECISION CO., LTD.
SPECIAL NOTICE
In some certain conditions, the stainless case of indicator will be opened to set up the load
cell, connect the power cable, or replace the new rechargeable battery, please contact your
local service people to avoid any damage of waterproof.
Before opening the stainless case, please ensure the indicator is free of moisture, if
necessary, any liquid should be wiped off.
The assembling notice as below should be strictly followed to assure the waterproof
performance. We strongly suggest that these procedures should be operated by the technical
staff of your supplier.
ASSEMBLING NOTICE
Screw the case with a torque screwdriver of 12 kgf-cm in the following orders. Do not screw
tightly before all screws are in the positions.
Screwing orders:
EXCELL PRECISION CO., LTD.
CHAPTER 1 INTRODUCTION
1-1 Features
IP68 waterproof: moisture-proof, mist-proof, with IP68 waterproof and dustproof
Adjustable weighing speed: three weighing speed modes (fast, moderate, and slow)
supported.
Selectable units: kilogram (kg), gram (g), pound (lb), and ounce (oz) weighing units
HI, OK, LO indication: Large LCD display with HI, OK, LO indication and the signals for
relay output with max. current 8 A at 120VAC or 30VDC, 5A at 230VAC
High resolution: the maximum external resolution is 1 / 30 000, and the internal is 1 / 300
000. (Input voltage should be over 6 mV)
Practical functions: 10 preset settings, Percentage display, Simple / Limit / Target
operation modes, Pre-Tare function, Net / Gross switch function, Auto backlight, RTC
settings and Auto power-off timer etc.
Others: Auto calibration, Auto zero tracking, Low battery indication, Adjustable gravity value,
and Linearity adjustment.
Built-in Bi-directional RS-232 interface
1-2 Specifications
Analog Input and A/D Conversion
: Input Sensitivity 0.18 V/d or more
Input Signal Range
: -2 ~ +16 mV
Input Zero Range
: -1 ~ +5 mV
Load Cell Excitation
: 5 V DC
Load Cell Drive Capacity
: up to 4 load cells at 350 / load cell
Non-linearity
: 0.01% of full scale
A/D Resolution
: 1 000 000 counts (Max.)
EXCELL PRECISION CO., LTD.
1-3 Scale Appearance
1-4 Power Supply
Power Supply Selection
1. 6 V / 4.0 Ah Rechargeable battery
2. AC adaptor DC 12 V / 1A
Power Consumption
Approximately DC 19 mA (Indicator)
Approximately DC 38 mA (Indicator + Display backlight)
Approximately DC 70 ~ 80 mA (Indicator + Display backlight + Relay)
Low Battery Warning
When the power bar indication turns to be low battery , it means the battery should
be recharged immediately.
The package includes:
1. Indicator
x1
2. Power Cable x1
3. User Manual x1
When you unpack the
indicator and find any of the
above items are missing,
please contact your supplier.
256 mm
170 mm
90 mm
Power cable connector
Load cell connector
RS-232 connector
(For RS-232 and load cell, only the
cables with Ф 3 ~ 5.5mm caliber can
be used, or the indicator’s
waterproofness will be affected.)
EXCELL PRECISION CO., LTD.
1-5 Explanation of Display
LCD1: Analog indication
column with maximum
50 graduation
LCD2:Digital Display
column with maximum 6
digits
Keypad
HI, LO, OK status
indication
In the Simple mode, if the
weight is over the maximum
graduation, the “+” zone will
be flashing.
Weigh unit indication: kg,
g, lb, lb-oz, Taiwan-kg, %
Zero point indication
Stable reading indication
NET indication
Battery indication
GROSS indication
Current memory reading
(HI, LO, OK, RANGE)
data
LIMIT indication
TARGET indication
EXCELL PRECISION CO., LTD.
1-6 Explanation of Keypad
Press this key to switch the display on/off.
Press this key to obtain a zero reading, when the platter is empty (free of
load) and the display is not showing zero.
When the weight value is within the zero range, the zero function operates to
zero the scale or cancel the tare function.
Press this key to clear memory and delete data.
Press this key to perform action.
� Digit: 9
Press this key to switch between the “Net value” and the “Gross value”.
(This key is functional only in Tare mode.)
Digit: 6
Press this key to deduct the weight of container via entering the known
value, so that the scale readout shows the net weight of samples.
Digit: 3
Press this key to deduct the weight of container, so that the scale
readout shows the net weight of samples.
Press this key to recall memory.
EXCELL PRECISION CO., LTD.
Digit 8
Press this key to output the displayed values via the built-in interface to
a printer or a computer.
Digit 5
In Target weight mode, press this key to preset the allowable “Range”
value.
Digit 2
In Limit weight mode, press this key to input “Low” value, so that the
weighing result cannot be smaller than this value.
Press this key to store HI, LO, OK and Range settings or exit the
settings mode.
Digit 7
Press this key to switch among Simple weight mode, Limit weight
mode and Target weight mode.
Digit 4
In Target weight mode, press this key to input value which meets the OK
requirement.
Digit 1
In Limit weight mode, press this key to input “HI” value, so that the
weighing result cannot be greater than this value.
Digit 0
Press this key to switch the weight unit, the icons or arrows will indicate
the enable units as appropriate. The units available are dependent on
the exact scale model.
After power off, the scale will memorize the enable units. When the scale is
powered on again, it displays the previous enable units.
EXCELL PRECISION CO., LTD.
CHAPTER 2 BASIC OPERATING FUNCTIONS
PH3 has three weight modes: Simple, Limit and Target, press to
switch.
Simple weight mode
In this mode, the weight value can be displayed by means of analog and
digital. For example, when an object weighs 20 kg, the LCD2 shows the digits
20 and LCD1 has slight changes on its graduation.
Limit weight mode
This is a weight check mode, press “HI” and “LO” keys to input high and low
limit values, LCD1 displays if the weight on platter is greater than the high
limit, lower than the check value or equal to the OK range. When it meets the
OK range, the OK zone will blink. For example, High limit = 20, Low limit = 18,
then OK range is between 18 and 20.
Target weight mode
Use this mode to set a “Target” weight and an allowable range. When the
weight is within the range, it means the weight meets the target requirement.
For example, pour liquid into a container, press “OK” key to set target as 30
and press “Range” key to set an allowable range as ±2, when the weight
reaches the range between 28 and 32,the OK zone on LCD1 will blink.
EXCELL PRECISION CO., LTD.
Simple Weight Mode
� �
�
Target Weight Mode
�
Limit Weight Mode
EXCELL PRECISION CO., LTD.
2-1 Simple Weight Mode
Steps:
1.
Turn on the scale
Scale starts to count down, LCD2 displays ”------“
After reaching to Zero, LCD2 displays ”000000”.
2. If needed, select the weight unit.
Zero the scale, if the Zero point value is floating.
3. Put object on the platter
LCD1 (analog) and LCD2 (Digit) display the object’s weight
After the weight is stable, the STABLE icon appears at the left bottom
of LCD2.
2-2 Limit Weight Mode
Steps:
1. Press “MODE” key, select Limit Mode, the Limit icon is switched on.
If the weight is floating, press “ZERO” key to obtain a zero reading.
2. Press “HI” key, the screen displays the previous setting of High value.
.
kg
The previous setting is 30 kg.
EXCELL PRECISION CO., LTD.
3. Use “0” ~ “9” keys to input the High value. After inputting one digit, the cursor will
move to right. Press “ENTER” key when finishing. (For example 30 kg, press
0030 )
When input incorrectly, press “C” key to clear.
4. Press “LO” key, the screen displays the previous setting of Low value.
5. Use “0”~“9” keys to input the Low value, see step 3.
“High value” must be set before “Low value”.
6. Put object on the platter, the screen reflects HI / LO / OK status.
If the weight is between the high and low limit, the “OK” zone will blink.
If the weight is over the high limit, the “+” zone will blink.
If the weight is under the low limit, the “-” zone will blink.
.
kg
The previous setting is 28 kg.
to
EXCELL PRECISION CO., LTD.
2-3 Target Weight Mode
Steps:
1. Press “MODE” key, select Target Weight Mode, the Target icon is switched on.
Press ZERO key to zero the scale, if the weight is floating.
2. Set the target weight (There are two methods)
Method 1: input value via keypad
(1)
Press “OK” key, the screen displays the previous setting.
(2)
Use 0 ~ 9 keys to input the target value. After keying one digit, the cursor will
move to right. Press “ENTER” key when finishing.
Method 2: Input value via sampling weight
(1) Press “OK” key, the screen displays the previous setting.
.
kg
.
kg
The previous settings is 30 kg.
The previous setting is 30 kg.
to
EXCELL PRECISION CO., LTD.
(2)
Put the sample on the platter, press “ENTER” key. The screen displays “------”.
After the weight is stable, the sampling is finished.
3. Set the Range Value
(1)
Press “RANGE” key,the screen displays the previous setting.
(2)
Use “0” ~ “9” keys to input the target value. After keying one digit, the cursor
will move to right. Press “ENTER” key when finishing.
4. Put a container on the platter, the screen displays the container’s weight.
5. Press TARE key, the screen displays “------”. The “Net” icon is on, after the weight
returns to zero.
kg
.
kg
.
kg
.
kg
kg
Net
.
kg
2 kg
The weight of container is 2 kg.
20 kg
The previous setting is 20 kg.
The previous setting is 2 kg.
to
EXCELL PRECISION CO., LTD.
6. Put objects into the container and observe the changes of graduation on LCD1.
When the weight reaches the OK section, stop adding the objects and LCD2
displays the weight at OK status.
The Range value can be shown as percentage.
Refer to “3-1-8 Weight and Percentage Mode Settings” for details.
2-4 Tare Function
Steps:
1. When the platter is free of load and the display is not showing zero, to zero
the scale.
2.
Put a container on the platter
LCD2 displays the weight of container
After the weight is stable, the STABLE icon will blink.
3. To zero the weight of container, the screen displays the “Net” icon.
4. Put the goods in the container, the screen displays the net weight value of
the goods.
The tare function can be operated continually to the full weighing capacity of the scale.
To continue tare operation, please press key each time when you finishing adding
tare objects on or removing tare objects from the platter.
For OIML and NTEP approval models, the tare function is automatically cancelled when
instantaneous weight returns near zero point.
For Sri Lanka approval models, the zero key
is disabled when using tare function.
Tare and Pre-tare functions cannot be used simultaneously.
�
�
EXCELL PRECISION CO., LTD.
2-5 Cancel Tare Function
Steps:
1.
Move the full container from the platter.
2.
Zero the scale and Tare
The “Net” icon is switched off.
or Clear Tare
The “Net” icon is switched off.
2-6 Pre-Tare Function
Steps:
1. Zero the scale, if needed.
2. The screen displays the previous Pre-Tare value for two seconds
Enter Pre-Tare mode 00000P .
3. Use “0” ~ “9” keys to input the known weight of container, after that .
4. The PT icon on LCD2 will blink, which means the settings is finished.
5. Put the container on platter
The screen displays the weight is zero
Put the goods into the container
The screen displays the weight of goods
PT icon is on
LCD2 displays the Net value of goods.
Tare and Pre-tare functions cannot be used simultaneously.
�
�
�
�
�
�
EXCELL PRECISION CO., LTD.
2-7 Cancel Pre-Tare Function
Method 1
Steps:
1. Remove the full container from the platter
2. Zero the scale and Tare
The “PT” icon is switched off
Cancel the Pre-Tare.
Method 2
Steps:
1. The screen displays the previous Pre-Tare value for two seconds
Enter Pre-Tare mode 00000P
The screen displays 00000P
2. The “PT” icon is switched off Cancel the Pre-Tare.
2-8 Gross/Net Function
In GROSS or NET mode, the screen displays “Net”:
Steps:
1. Gross icon is switched on and “Net” icon is off, which means the
weight value on the display is the total amount of the tare value and net
value.
2. The screen displays Net value again. The “Net” icon is on, while the
“Gross” icon is off. Use this key to switch between “Net value” and “Gross value”.
�
�
�
�
EXCELL PRECISION CO., LTD.
This function is only used in TARE and Pre-Tare modes.
All keys except for “On/Off” and “Net/Gross” do not work, while the Gross icon
is on.
2-9 Memory Store and Recall Functions
Store
When the settings of HI, LO, OK and Range are finished,
LCD2 displays “MEM S”, input any digits from “0” ~ “9”, such as “2”
When finishing, “M2” appears at the center of LCD1.
Recall
LCD2 displays “MEM R”, input any digits from “0” ~ “9”, such as “2”
When finishing, “M2” appears at the center of LCD1.
This indicator can store 10 pieces of data (from 0 to 9).
EXCELL PRECISION CO., LTD.
CHAPTER 3 EXTERAL CALIBRATION
In the weighing mode, press the TARE and ZERO keys at the same time to enter
the Advanced Function settings mode. The LCD2 shows
and
use MR to select the desired mode.
Overall workflow of the Advanced Function settings mode is as follows:
Press MR key
Press MR key
Press MR key
General Function Settings Mode
Weight Calibration
RS-232 Bi-directional Function Settings
Exit the Advanced Function Settings Mode
Refer to the following sections for the detailed operating
procedures of each function settings.
Press MR key
EXCELL PRECISION CO., LTD.
3-1
General Function Settings
There are 16 functions in the General Function settings mode from
to
.
C key
clear parameter
Enter key perform action
Refer to the following overview of
General Functions
Press Enter key
To exit the general
function settings, key in
“00” and press Enter
key.
EXCELL PRECISION CO., LTD.
Return to the Advanced Function Settings Mode
Automatic Backlight Function Settings
Automatic Power-off Timer Settings
HI / LO Weight Check Settings
HI / LO Alarm Settings Function
Analog and Digital Display Settings
Weighing Speed Settings
Operation Mode Settings
Weight and Percentage Mode Settings
Lock Function Settings
Relay Function Settings
Auto Zero Tracking Settings
Printer Key Accumulations Function
Tare Cancellation Settings
Previous Zero Record Settings
HI / LO / OK Positive / Negative Weight Settings
HI / LO / OK Tare Time Choose Settings
Overview of the General Functions
EXCELL PRECISION CO., LTD.
3-1-1
Automatic Backlight Function Settings
Select
in the General Function settings mode
to change the backlight function settings.
C key
clear parameter
Enter key perform action
Automatic backlight function
When the weight is over 10d, the display backlight will be on. After the weight is
stable for 10 seconds or when the scale returns to zero, the display backlight
switches off.
Parameter settings:
01 Turn on automatic backlight
00 Turn off automatic backlight
Factory default setting: 00
Press Enter key
Press Enter key
To exit the general function
settings, key in “00” and
press Enter key.
EXCELL PRECISION CO., LTD.
3-1-2
Automatic Power-off Timer Settings
Select
in the General Function settings mode
to change the automatic Power-off timer function settings.
C key
clear parameter
Enter key perform action
Automatic power-off function
When the weight on platter is less than 10d or keeps idle for the set time, the
scale will automatically power off.
Enter a number from 01 to 06 so as to
instruct the scale to be automatically
powered off after being idle 1 to 6 minutes
while 00 is the default which means the
auto power-off function is disable.
Factory default setting: 00
Press Enter key
Press Enter key
To exit the general
function settings, key in
“00” and press Enter
key.
EXCELL PRECISION CO., LTD.
3-1-3
HI / LO Weight Check Settings
Select
in the General Function settings mode
to set the HI / LO weight check function.
FNC 03 settings are available only when the scale is in Limit Weight Mode or Target
Weight Mode.
C key
clear parameter
Enter key perform action
Parameter settings:
00 Check the weight whatever it is stable
or not
01 Check the weight when it is stable
02 Whatever it is stable or not, check the
weight when the weight is over 10d.
03 Check the weight when it is stable
and over 10 d.
04 Do not check
Factory default setting: 00
Press Enter key
Press Enter key
To exit the general
function settings, key in
“00” and press Enter
key.
EXCELL PRECISION CO., LTD.
3-1-4
HI / LO Alarm Settings Function
Select
in the General Function settings mode
to set the HI / LO Alarm function.
FNC 04 settings are available only when the scale is in Limit Weight Mode or Target
Weight Mode.
C key
clear parameter
Enter key perform action
Parameter settings:
00 No Alarm
01 The Beeper sounds intermittently at LO
status
02 The Beeper sounds intermittently at OK
status
03 The Beeper sounds intermittently at LO
and OK status
04 The Beeper sounds intermittently at HI
status
05 The Beeper sounds intermittently at HI
and Low status
06 The Beeper sounds intermittently at HI
and OK status
07 The Beeper sounds intermittently at
High, Low and OK status
08 No Alarm
09 The Beeper sounds continually at LO
status
10 The Beeper sounds continually at OK
status
11 The Beeper sounds continually at LO
and OK status
12 The Beeper sounds continually at HI
status
13 The Beeper sounds continually at HI
and Low status
14 The Beeper sounds continually at HI
and OK status
15 The Beeper sounds continually at HI,
Low and OK status
Often used settings are “02” and “05”.
Factory default setting: 00
Press Enter key
Press Enter key
To exit the general
function settings, key in
“00” and press Enter
key.
EXCELL PRECISION CO., LTD.
3-1-5
Analog and Digital Display Settings
Select
in the General Function settings mode
to change the Analog and Digital display function settings.
C key
clear parameter
Enter key perform action
Parameter settings:
00 Both Analog and Digital displays
are enable
01 Only display Digital
Factory default setting: 00
Press Enter key
Press Enter key
To exit the general
function settings, key in
“00” and press Enter
key.
EXCELL PRECISION CO., LTD.
3-1-6
Weighing Speed Settings
Select
in the General Function settings mode
to set the weighing speed function.
C key
clear parameter
Enter key perform action
Parameter settings:
00 Slow reaction
01 Moderate reaction
02 Fast reaction
Factory default setting: 01
Press Enter key
Press Enter key
To exit the general
function settings, key in
“00” and press Enter
key.
EXCELL PRECISION CO., LTD.
3-1-7
Operation Mode Settings
Select
in the General Function settings mode
to change the operation mode function settings.
C key
clear parameter
Enter key perform action
Parameter settings:
00 Enter “Simple Weight Mode” after turn on
01 Enter “Limit Weight Mode” after turn on
02 Enter “Target Weight Mode” after turn on
03 Use previous setting mode after turn on
Factory default setting: 03
Press Enter key
Press Enter key
To exit the general
function settings, key in
“00” and press Enter
key.
EXCELL PRECISION CO., LTD.
3-1-8
Weight and Percentage Mode Settings
Select
in the General Function settings mode
to change the weight and percentage mode settings.
C key
clear parameter
Enter key perform action
Parameter settings:
00 Display in weight mode
01 Display in percentage mode
(Target mode only)
Factory default setting: 00
Press Enter key
Press Enter key
To exit the general
function settings, key in
“00” and press Enter
key.
EXCELL PRECISION CO., LTD.
3-1-9
Lock Function Settings
Select
in the General Function settings mode
to change the Lock function settings.
C key
clear parameter
Enter key perform action
Parameter settings:
00 All keys act normally
01 Only ON/OFF ZERO TARE
are enable.
Factory default setting: 00
Press Enter key
Press Enter key
To exit the general
function settings, key in
“00” and press Enter
key.
EXCELL PRECISION CO., LTD.
3-1-10
Relay Settings
Select
in the General Function settings mode
to
change the relay function settings.
C key
clear parameter
Enter key perform action
Parameter settings:
00 Default open
01 Default short
Factory default setting: 00
Press Enter key
Press Enter key
To exit the general
function settings, key in
“00” and press Enter
key.
EXCELL PRECISION CO., LTD.
3-1-11
Auto Zero Tracking
Select
in the General Function settings mode
to set auto zero tracking.
Only for none-approved model
C key
clear parameter
Enter key perform action
Parameter settings:
00 1/4 d
01 1/2 d
02 1 d
03 2 d
04 4 d
Factory default setting: 00
Press Enter key
Press Enter key
To exit the general
function settings, key in
“00” and press Enter
key.
EXCELL PRECISION CO., LTD.
3-1-12
Printer Key Accumulation Function
Select
in the General Function settings mode
to
set printer key accumulation function.
This function is enabled only when the parameter in
is set
as “11”.
When this function is enable, the Enter key functions as C key and has up to 99
accumulations.
C key
clear parameter
Enter key perform action
Parameter settings:
00 Turn off the accumulation function
01 Turn on the accumulation function
Factory default setting: 01
Press Enter key
Press Enter key
To exit the general
function settings, key in
“00” and press Enter
key.
EXCELL PRECISION CO., LTD.
3-1-13
Tare Cancellation Settings
Select
in the General Function settings mode
to
change the tare cancellation settings.
FNC 13 settings are not available for OIML and NTEP approval models.
C key
clear parameter
Enter key perform action
Parameter settings:
00 Use TARE key to call off settings
01 TARE value cancelled automatically
by moving objects from the platter
Factory default setting: 00
Press Enter key
Press Enter key
To exit the general
function settings, key in
“00” and press Enter
key.
EXCELL PRECISION CO., LTD.
3-1-14
Previous Zero Record Settings
Available in non-approved model only
3-1-15
HI / LO / OK Positive / Negative Weight Settings
FNC 15 settings can only be available under “Limit Weight Mode” and “Target Weight
Mode.”
FNC 15 settings can be arranged in groups to work with FNC 03 and FNC 04.
If FNC 03 setting is “00” and FNC 15 setting is “01” or “03”,immediate tare function is
available.
C key
clear parameter
Enter key perform action
C key
clear parameter
Enter key perform action
Zero key perform zero record
Press Enter key
Press Enter key
Parameter settings:
00 To abort previous zero recording
01 To enable previous zero recording
Factory default setting: 00
Press Enter key
Parameter settings:
00 HI / LO / OK positive weight display
01 HI / LO / OK positive weight display
and then tare
02 HI / LO / OK negative weight display
03 HI / LO / OK negative weight display
and then tare
Factory default setting: 00
EXCELL PRECISION CO., LTD.
Press Enter key
3-1-16
HI / LO / OK Tare Time Settings
FNC 16 settings can be available only when FNC 15 has been set to “01”.
If FNC 03 is set to “00”, immediate tare function is available whatever setting of FNC 16 is.
C key
clear parameter
Enter key perform action
Zero key perform zero record
Press Enter key
Parameter settings:
00 To tare after being stable for 1/2 s
01 To tare after being stable for 1 s
02 To tare after being stable for 2 s
03 To tare after being stable for 3 s
Factory default setting: 00
EXCELL PRECISION CO., LTD.
3-2
Weight Calibration
There are 2 functions in the Weight Calibration settings mode from
to
.
The function is disable when the parameter in
is set as “01” or “03”
(for approval model).
C key
clear parameter
Enter key perform action
Exit the Weight Calibration Mode
Calibration
Restore to the Default setting
Press Enter key
Input parameter
from 00 ~ 02
To exit the weight
calibration, key in “00”
and press Enter key.
EXCELL PRECISION CO., LTD.
The scale is
calculating the
internal value of
the zero point.
After stabilization,
the LCD shows
the calibration
weight value.
Put the platter on
the scale and
ensure there is
free of load. Then
press Enter key.
kg
kg
Place the displayed weight on
the platter and press Enter
key. The scale calculates the
full capacity internal value and
after stabilization, the buzzer
beeps 3 times. Remove the
weight from the
platter and the
procedure is
completed.
kg
Press MR key to calculate
the internal value of zero point.
Press MR key to quit
SPAN calculating mode
3-2-1
Weight Calibration
Select
in the General Function settings mode
to set
weight calibration.
To exit the weight calibration,
key in “00” and press
Enter key.
Weight calibration conditions:
The calibration procedure will not be carried out if the difference between the internal and
external ratio of external calibration weight is > 3 of that of initial factory calibration.
The internal and external ratio of the initial factory calibration -3 < The internal and
external ratio of external weight calibration < The internal and external ratio of the
initial factory calibration +3
.
.
C key
clear parameter
Enter key perform action
Press Enter key
The LCD displays the calibration weight
value (the maximum capacity of 1st unit) and
the unit. The digit on the left side keeps
flashing.
Revise the calibration weight value if needed.
About 2 seconds later
About 2 seconds later
EXCELL PRECISION CO., LTD.
3-2-2
Restore to the Default setting
Select
in the General Function settings mode
to
restore the default setting.
C key
clear parameter
Enter key perform action
Press Enter key
Key in parameter
from 00 to 01
Press Enter key
To exit the weight
calibration,
key in “00” and press
Enter key.
Parameter settings:
00 Exit the settings
01 Restore to the factory
default setting
EXCELL PRECISION CO., LTD.
3-3
RS-232 Bi-directional Function Settings
There are 14 functions in the Bi-directional function settings mode from
to
.
C key
clear parameter
Enter key perform action
To exit RS-232 Bi-directional
settings mode, key in “00”
and press Enter key.
Press Enter key
Key in parameter from 00 to 14
Refer to the following overview of
RS-232 Bi-directional Settings
EXCELL PRECISION CO., LTD.
Exit the RS-232 Bi-directional Settings Mode
Baud Rate Settings
Communication Protocol Settings
Output Format Settings
Continuous Transmission Settings
Selection of Continuous Transmission Rate
Output Condition Settings
Auto Transmission at Zero
Reset of Auto Transmission
Condition of Manually Printing
Real Time Clock Settings (Hour)
Real Time Clock Settings (Date)
RS-232 Serial Interface Settings Mode
RS-232 General or Simple Output Weight Selection
The Date Format Settings
Overview of RS-232 Bi-directional
Settings
EXCELL PRECISION CO., LTD.
3-3-1
Baud Rate Settings
Select
in the RS-232 Bi-directional settings mode
to set the Baud Rate.
C key
clear parameter
Enter key perform action
LCD displays the last setting.
Factory default value: 04
Parameter settings:
00 600
(bits/s)
01 1 200
(bits/s)
02 2 400 (bits/s)
03 4 800 (bits/s)
04 9 600 (bits/s)
05 19 200 (bits/s)
06 38 400 (bits/s)
Press Enter key
Press Enter key
To exit the RS-232
Bi-directional settings,
key in “00” and press
Enter key.
EXCELL PRECISION CO., LTD.
3-3-2
Communication Protocol Settings
Select
in the RS-232 Bi-directional settings mode
to set the communication protocol.
Relevant information please refer to Appendix 1
C key
clear parameter
Enter key perform action
LCD displays the previous setting.
Factory default setting: 00
Parameter settings:
00 N 8 1
(none parity check, 8 data bits, 1 stop bit)
01 E 7 1
(even parity check, 7 data bits, 1 stop bit)
02 O 7 1
(odd parity check, 7 data bits, 1 stop bit)
Press Enter key
Press Enter key
To exit the RS-232
Bi-directional settings,
key in “00” and press
Enter key.
EXCELL PRECISION CO., LTD.
3-3-3
Output Format Settings
Select
in the RS-232 Bi-directional settings mode
to set the output format.
When
and
are both set to 01 all options of
are
allowed.
When settings
or
is set to 00, only the options 00, 01, 02, 09,
10 of
are allowed.
With parameter setting of “11”, please then enter
to set the RS-232 serial interface.
Relevant information please refer to
Appendix 1.
C key
clear parameter
Enter key perform action
LCD shows the last setting.
Factory default setting: 11
Parameter settings:
00 Same data as the scale in general
format
01 Gross weight in general format
02 Net weight in general format
03 Same data as the scale in simple
format
04 Gross data in simple format
05 Net data in simple format
06 HI / LO /OK status + Same data as
the scale in simple format
07 HI / LO / OK status + Simple gross
weight in simple format
08 HI/LO/OK status + Simple net weight
in simple format
09 Tare value in general format
10 Pre-Tare value in general format
11 RS-232 serial format
Press Enter key
Press Enter key
To exit the RS-232
Bi-directional settings,
key in “00” and press
Enter key.
EXCELL PRECISION CO., LTD.
3-3-4
Continuous Transmission Settings
Select
in the RS-232 Bi-directional settings mode
to set the continuous transmission.
Relevant information please refer to Appendix 1.
C key
clear parameter
Enter key perform action
LCD shows the last setting.
Factory default setting: 00
Parameter settings:
00 Command mode
01 Continuous transmission mode
02 Auto transmission mode
03 Manually printing format mode
04 Shut down RS-232
Press Enter key
Press Enter key
To exit the RS-232
Bi-directional settings,
key in “00” and press
Enter key.
EXCELL PRECISION CO., LTD.
3-3-5
Selection of the Continuous Transmission Rate
Select
in the RS-232 Bi-directional settings mode
to set the continuous transmission rate.
C key
clear parameter
Enter key perform action
LCD shows the last setting.
Factory default setting: 02
Parameter settings:
00 1 times/s
01 2 times/s
02 4 times/s
03 8 times/s
04 16 times/s
05 More than 16 times/s
Press Enter key
Press Enter key
To exit the RS-232
Bi-directional settings,
key in “00” and press
Enter key.
EXCELL PRECISION CO., LTD.
3-3-6
Output Condition Settings
Select
in the RS-232 Bi-directional settings mode
to set the output condition.
For approval models,
settings are available only when setting
is set to 00, 01, 02, 09, or 10.
C key
clear parameter
Enter key perform action
Press Enter key
Press Enter key
LCD shows the last setting.
Factory default setting: 01
Parameter settings:
00 Output all the time
01 Stable weight output only
(No output when under / overload or
unstable)
EXCELL PRECISION CO., LTD.
3-3-7
Auto Transmission at Zero
Select
in the RS-232 Bi-directional settings mode
to set the auto transmission at zero.
C key
clear parameter
Enter key perform action
LCD shows the last setting.
Default setting: 05
Parameter settings:
Please enter a number from 00
to 99 to set its value from 00d to
99d accordingly.
Press Enter key
Press Enter key
To exit the RS-232
Bi-directional settings,
key in “00” and press
Enter key.
EXCELL PRECISION CO., LTD.
3-3-8
Reset of Auto Transmission
Select
in the RS-232 Bi-directional settings mode
to reset the auto transmission.
C key
clear parameter
Enter key perform action
LCD shows the last setting.
Default setting: 00
Parameter settings:
Please enter number from 00
to 99 to set its value from 00d
to 99d accordingly.
Press Enter key
Press Enter key
To exit the RS-232
Bi-directional settings,
key in “00” and press
Enter key.
EXCELL PRECISION CO., LTD.
RS1 07 & RS1 08 Settings Instruction
1. Under auto transmission mode, if RS1 07 and RS1 08 are all set at “00”, no matter what
weight it is, no any data of the weight will be transmitted out.
2. Under auto transmission mode, if RS1 07 is set between “00 ~ 99”, and RS1 08 is set at
“00”, when the weight is over or under the zero range that set in RS1 07, after being
stable, the data of weight will be transmitted out; if the weight is back to the zero range
(RS1 07,) and then becomes higher or lower than the range again, after being stable, the
data of weight will be transmitted out (the format is set in RS1 03.)
e.g.: When RS1 07 is set at “05”, RS1 08 is set at “00”, and RS1 04 is set at “02”, if the weight
is less than 5g, no data of weight will be transmitted out.
Put on a weight of 6g, after being stable ST, NT, +00000006g
If take off the weight of 6g, weight column displays “0g”
No transmission (because it’s in the zero range)
Put on a weight of 50g, after being stable ST, NT, +0000050g
If take off the weight “48g,” weight column displays “2g”
No transmission (because it’s in the zero range)
Add the weight to 30g, after being stable ST, NT, +0000030g
And so on…
3. Under auto transmission mode, if RS1 07 is set at “00” and RS1 08 is set between “00 ~
99”, as long as the weight is higher than “0”, after being stable, the data of weight will be
transmitted out. And then, the weight becomes the benchmark, as long as the weight
changes higher or lower than the range of RS1 08, after being stable, the data of weight
will be transmitted out. To sum up, on condition that the weight changes and after being
stable, the data of weight will be transmitted out (the format is set in RS1 03.)
4. Under auto transmission mode, if RS1 07 is set between “00 ~ 99” and RS1 08 is set
between “00 ~ 99”, as long as the weight is higher or lower than the zero range set in
RS1 07, after being stable, the data of weight will be transmitted out. To sum up, on
condition that the weight changes, not including “0”, after being stable, the data of weight
will be transmitted out (the format is set in RS1 03.
EXCELL PRECISION CO., LTD.
3-3-9
Condition of Manually Printing
Select
in the RS-232 Bi-directional settings mode
to set the condition of manually printing.
For approval models,
settings are available only when settings
is set to 00, 01, 02, 09, or 10.
Before
settings, ensure that you set
to 03, then press the
PRINT key.
C key
clear parameter
Enter key perform action
LCD shows the last setting.
Parameter settings:
00 Transmit immediately
01 Transmit when stable
Factory default setting: 01
Press Enter key
Press Enter key
EXCELL PRECISION CO., LTD.
3-3-10
Real Time Clock (Hour)
Select
in the RS-232 Bi-directional settings mode
to set the real time clock for time.
MR key move the flashing letter to left
Enter key perform action
LCD shows the current time (24 hours)
Press Enter key
Use“0” ~ “9” keys to key
in “hour”, “minute”
and “second”.
Press Enter key
To exit the RS-232
Bi-directional settings,
key in “00” and press
Enter key.
EXCELL PRECISION CO., LTD.
3-3-11
Real Time Clock (Date)
Select
in the RS-232 Bi-directional settings mode
to set the real time clock for date.
MR key move the flashing letter to left
Enter key perform action
LCD shows the current date (A.D.)
Press Enter key
Use “0” ~ “9” keys to
input “YY”, “MM”
and “DD”.
Press Enter key
To exit the RS-232
Bi-directional settings,
key in “00” and press
Enter key.
EXCELL PRECISION CO., LTD.
3-3-12
RS-232 Serial Interface Settings Mode
This function is enabled only when the parameter in
is set
as “11”.
Select
in the RS-232 Bi-directional settings mode
to set the RS-232 serial interface.
This
settings are available for most models. These settings may be
unavailable for some specific models without bi-directional RS-232 serial interface
component.
More information please refer to Appendix 1.
Enter key perform action
Press Enter key
Start to receive
Press Enter key
To exit the RS-232
Bi-directional settings,
key in “00” and press
Enter key.
EXCELL PRECISION CO., LTD.
Input parameters “00”、“01”
Input parameters “00 ~ 02”
3-3-13
RS-232 General or Simple Output Weight Selection
Please refer to RS-232 output data format at pages 64 ~ 65.
3-3-14
The Date Format Settings
The date order printed will be in accordance with RS1 14.
C key
clear parameter
Enter key perform action
C key
clear parameter
Enter key perform action
Press Enter key
Parameters instruction
00 6 digits
01 7 digits
default value: 00
Press Enter key
The display shows the last setting.
Parameter setting:
00 YY.MM.DD
01 DD.MM.YY (EUROPE)
02 MM.DD.YY (AMERICA)
Default setting : 01 (EUROPE)
Press Enter key
Press Enter key
EXCELL PRECISION CO., LTD.
Appendix 1 RS-232 Serial Interface
RS-232 Command format:
(1) The Host (PC) writes the data to the EEPROM of Slave (Scale).
(2) The Host (PC) reads the data in the EEPROM of Slave (Scale), and the Slave
(Scale) has the data feedback to the host (PC).
Content planning
1.
W
,
,
,
CR LF
2.
R
,
CR LF
Write above two commands to the Host (PC), the Slave (Scale) has the following format
feedback to the Host (PC).
L
e
n
g
t
h
:
CR LF
A
S
C
I
I
:
CR LF
S
y
m
b
o
l
:
n CR LF
Write
EEPROM
( 00 ~ 58 )
Length of
storage
( ≤15 )
1 2 3 ………………. n
ASCII code n = length of storage
Read
EEPROM
( 00 ~ 58 )
Length of
storage
1 2 3 ……… … n
Store ASCII code n = length of storage
1 2 3 ………………. n
Store ASCII symbol n = length of storage
EXCELL PRECISION CO., LTD.
Code
First EEPROM
address
Item
Final EEPROM
address
00
00
Gross
01
01
02
Net
03
02
04
Tare
05
03
06
Pre-Tare
07
04
08
Target Weight
09
05
10
Range
11
06
12
High Limit
13
07
14
Low Limit
15
08
16
Percent (%)
17
09
18
EEPROM address
Weight units
24
19
First unit EEPROM address
20
Second unit EEPROM address
21
Third unit EEPROM address
22
Forth unit EEPROM address
23
Fifth unit EEROM address
10
25
Time
26
11
27
Date
28
12
29
EEPROM address
HI、LO、OK status
33
30
LO EEPROM address
31
OK EEPROM address
32
HI EEPROM address
13
34
EEPROM address
Stable (Unstable) sign
37
35
Stable sigh EEPROM address
36
Unstable sign EEPROM
address
14
38
Accumulations quantity
39
15
40
Accumulations total weight
41
16
42
Check the x weight
43
17
44
User’s definition 1
18
45
User’s definition 2
19
46
User’s definition 3
20
47
User’s definition 4
21
48
User’s definition 5
22
49
User’s definition 6
23
50
User’s definition 7
24
51
User’s definition 8
25
52
User’s definition 9
26
53
User’s definition 10
27
54
User’s definition 11
28
55
User’s definition 12
29
56
User’s definition 13
30
57
User’s definition 14
31
58
User’s definition 15
EXCELL PRECISION CO., LTD.
Format
1.
W
,
,
,
CR LF
2.
R
,
CR LF
Write above two commands to the Host (PC), the Slave (Scale) has the following format
feedback to the Host (PC).
L
e
n
g
t
h
:
CR LF
I
t
e
m
:
CR LF
Write
EEPROM
( 59 ~ 63 )
Code
numbers
( < 15 )
1 2 3 …………………. n
code n =number
Read
EEPROM
( 59 ~ 63 )
Code numbers
1 2 3 ………………. n
code n = number
EXCELL PRECISION CO., LTD.
EEPROM address
Format
Transmission
59
Simple Format
Press “Print” key in
Simple Weight mode
60
Limit Format
Press “Print” key in
Limit Weight mode
61
Target Format
Press “Print” key in
Target Weight mode
62
Transmission of accumulation
clear
Press “Enter” key
63
Control code
Transmit one time
when turn on the scale
Code function
Code+80 Do not print the first and final letters
Code+40 Display zero by blank
Code+B0 Do not print the first and final letters and display zero by blank
Example:
Tare Code 03
03+80 = 83
Do not print the first and final letters of Tare
03+40 = 43
Display zero by blank, such as 000.00kg
03+B0 = B3
Do not print the first and final letters, and display zero
when the content of Tare value is blank.
EXCELL PRECISION CO., LTD.
Command Mode
Command Format A
Host
Command
Slave
Command
MZ
Zero
SM
Manually transmit
MT
Tare
SO
Command mode
MG
Gross weight
UA
Switch to the first unit
MN
Net weight
UB
Switch to the second unit
MM
Switch operation mode
UC
Switch to the third unit
CT
Clear Tare value
UD
Switch to the forth unit
CP
Clear Pre-Tare value
UE
Switch to the fifth unit
SA
Auto transmit
%
Stop continuous transmission
and enter the command mode
SC
Continuous transmission
Note : UA ~ UE settings are dependent on the model of the scale
Command Format B
Host
Command
Slave
Data
RW
Read current weight
RH
Read Gross (simple)
RG
Read Gross weight
RI
Read Net (simple)
RN
Read Net weight
RJ
Read comparison situation +
current display of weight (simple)
RT
Read Tare
RK
Read comparison situation +
Gross (simple)
RB
Read current display of weight
(simple)
RL
Read comparison situation + Net
(simple)
RE
Read Pre-Tare
Note : ◇
1 add % before the command to read continuously
◇
2 add # before the command to transmit a stable value
Read Weight Comparison Settings Value
RS : group (00 ~ 09) : settings item
LO
Read LO settings value
RG
Read Range settings value
HI
Read HI settings value
PT
Read Pre-Tare settings value
OK
Read Target settings value
Note: No group number for PT.
EX:
RSPT<CR><LF>
Read Pre-Tare settings value
ANS:
RSPTXXXXXX<CR><LF>
EX:
RS03RG
Read the 3rd group of Range
settings value
ANS:
RS03RGXXXXXX
EXCELL PRECISION CO., LTD.
Command Format C
Host
Command+ Data
Slave
Command+ Data
Write Weight Comparison Settings Value
WSXXXXXX
: group (00 ~ 09) : settings item XXXXXX: Settings value
LO
Write LO settings value
HI
Write HI settings value
OK
Write Target settings value
RG
Write Range settings value
PT
Write Pre-Tare settings value
EX:
WSPTXXXXXX<CR><LF>
Write Pre-Tare settings value
ANS:
WSPTXXXXXX<CR><LF>
EX:
WS03RG XXXXXX<CR><LF>
Write Range settings value
ANS:
WS03RG XXXXXX<CR><LF>
The above 3 command formats are all Bi-directional RS-232 commands.
Error messages on Slave (Scale):
E1: Wrong command
E2: Command format error (Wrong parameters)
E3: Command not recognized
EXCELL PRECISION CO., LTD.
Output Data Format
1. Seven digits (without including one decimal point)
Weight format
Gross
S T
,
G S
,
+ 0
1
2
3
4
5
6 7 SP SP o
z
CR LF
Net
S T
,
N T
,
+ 1
.
2
3
.
4
5 6
t
l
.
g
Tare
S T
,
T R
,
+ 0
1
2
.
3
4
5 6 SP SP k
g
Plus OL
O L
,
G S
,
+ SP SP SP SP SP SP SP SP SP SP SP SP
Minus OL O L
,
G S
,
- SP SP SP SP SP SP SP SP SP SP SP SP
Unstable
U S
,
G S
,
+ 0
1
2
3
4
.
5 6 SP SP l
b
Pre-Tare
S T
,
P T
,
+ 0
1
2
3
4
5
6 7 SP SP SP g
Simple format
Gross
+
0
1
2
3
4
5
6
7
CR LF
Net
+
1
.
2
3
.
4
5
6
Tare
+
0
1
2
.
3
4
5
6
Plus OL
+ SP SP SP SP SP SP SP SP
Minus OL - SP SP SP SP SP SP SP SP
Comparison status + Simple format
Byte0 Byte1 Byte2 +/-
1
.
2
3
.
4
5
6
CR
LF
Byte0 : HI 30H / 31H
Byte1 : OK 30H / 31H
Byte2 : LO 30H / 31H
EXCELL PRECISION CO., LTD.
2. Six digits (without including one decimal point )
General weight format
Gross
weight
S
T
,
G
S
,
+
1
2
3
4
5
6
7 SP SP o
z
CR LF
Net weight S
T
,
N
T
,
+
.
2
3
.
4
5
6
t
l
.
g
Tare
S
T
,
T
R
,
+
1
2
.
3
4
5
6 SP SP k
g
Positive
overload
O
L
,
G
S
,
+ SP SP SP SP SP SP SP SP SP SP SP
Negative
overload
O
L
,
G
S
,
- SP SP SP SP SP SP SP SP SP SP SP
Unstable
U
S
,
G
S
,
+
1
2
3
4
.
5
6 SP SP l
b
Pre-tare
S
T
,
P
T
,
+
1
2
3
4
5
6
7 SP SP SP g
Simple format
Gross
weight
+
1
2
3
4
5
6
7
CR LF
Net
weight
+
.
2
3
.
4
5
6
Tare
+
1
2
.
3
4
5
6
Positive
overload
+ SP SP SP SP SP SP SP
Negative
overload
- SP SP SP SP SP SP SP
Comparative condition + Simple format
Byte0 Byte1 Byte2 +/-
.
2
3
.
4
5
6
CR
LF
Byte0 : HI 30H / 31H
Byte1 : OK 30H / 31H
Byte2 : LO 30H / 31H
EXCELL PRECISION CO., LTD.
Serial Data Transfer / Receive Format
LSB MSB
1
0
n , 8 , 1
S
8-bit data
STOP
1
0
e , 7 , 1
S
7-bit data
P STOP
1
0
o , 7 , 1
S
7-bit data
P STOP
Notes:
S : Start bit
STOP: Stop bit
P : Parity bit
RS-232 Cable Connection Illustrations
Open the housing to connect RS-232 cable with RS-232 connector locating at
the right-bottom of the main board, take the most normal connection way of 9PIN
and 25PIN as an example, please refer to below illustration:
PC
PIN
Function
Female 9 PINS
PIN
Function
PH3
2
T x D
1
SG
3
R x D
2
T x D
5
SG
3
R x D
Printer
PIN
Function
Male 25 PINS
PIN
Function
PH3
2
R x D
1
SG
3
T x D
2
T x D
7
SG
3
R x D
If take the other ways, please recognize the signals and follow the above rules.
After finishing, please lock the housing tightly according to the tighten or loosen
orders and screwing method which attached described in the Special Notice at the
beginning of this manual.
EXCELL PRECISION CO., LTD.
Appendix 2 Error Messages
EEPROM does not function.
Zero value* after power on is too high.
Zero value* after power on is too low.
The internal value is unstable.
The platter is slanted (i.e. not level).
A/D IC is over the resolution.
When the above error messages occur, please contact your
supplier for professional help immediately.
The weight on the platter is over the capacity + 9 division.
When the above error message occurs, please take away some
of your objects on the platter.
*For Sri Lanka approval and non-approval models, zero value inspection range and
return-to-zero range is less than or equal to 3%.
EXCELL PRECISION CO., LTD.
Appendix 3 7-Segment Display Characters
Number
Display
Letter
Display
Letter
Display
0
A
N
1
B
O
2
C
P
3
D
Q
4
E
R
5
F
S
6
G
T
7
H
U
8
I
V
9
J
W
K
X
L
Y
°C
M
Z