Handheld Odor Meter

OMX-ADM Operation Manual www.GlobalTestSupply.com Find Quality Products Online at: [email protected]

Table of Contents

1. Introduction............................................................................................................... 4 1-1. Supplied Components List ..................................................................................... 4 1-2. Warning ................................................................................................................... 5 1-3. Cautions .................................................................................................................. 6 1-4. Unit Dimension ...................................................................................................... 7 2. General Description .................................................................................................. 8 2-1. Features .................................................................................................................. 9 2-1-1. LCD Display ........................................................................................................ 9 2-1-2. Base Unit ........................................................................................................... 10 2-2. Accessories ............................................................................................................ 11 3. Measurement ........................................................................................................... 12 3-1. Real-time Sampling Mode .................................................................................... 12 3-1-1. Measurement Preparation ............................................................................... 12 3-1-2. Measurement .................................................................................................... 13 3-2. Memory Sampling Mode ...................................................................................... 13 3-2-1. Measurement Preparation ............................................................................... 13 3-2-2. Measurement .................................................................................................... 13 3-2-3. Memory Overflow .............................................................................................. 14 3-3. Peak-hold Function .............................................................................................. 15 3-4. Zero-point Function .............................................................................................. 15 4. Data Filing Software ................................................................................................ 16 4-1. How to Download Software.................................................................................. 16 4-2. How to Install Software ....................................................................................... 16 How to Install USD Drivers Software ......................................................................... 17 4-3. Preparation ........................................................................................................... 18 www.GlobalTestSupply.com Find Quality Products Online at: [email protected]

Open Software ...................................................................................................... 18

5. Specification ............................................................................................................. 19 6. Maintenance and Servicing ..................................................................................... 20 6-1. Replacement of “AA” Alkaline Batteries ............................................................ 20 6-2. Replacement of Button Battery ........................................................................... 20 6-3. Maintenance of Air Purifying Unit ..................................................................... 21 6-4. Replacement of Filter for Suction Nozzle ........................................................... 21 6-5. Troubleshooting .................................................................................................... 22 www.GlobalTestSupply.com Find Quality Products Online at: [email protected]

1-1. Supplied Components List

Thank you for purchasing of Shinyei Handheld Odor Meter. Handheld Odor Meter is packaged carefully and includes all the components necessary for full operation. Upon receipt, check the contents and find you received the items as below in good condition. The instrument kit contains: Name Q’ty OMX-ADM Base Unit 1 Suction Nozzle 1 AC Adaptor 1 ‘AA’ Alkaline Battery 4 USB Connection Cable 1 Air Purifying Unit 1 Activated Carbon (50cc/bag) 1 Replacement Filter for Air Purifying Unit 4 Replacement Filter for Suction Nozzle 5 Operation Manual 1 www.GlobalTestSupply.com Find Quality Products Online at: [email protected]

Do not use any adaptors except supplied original AC adaptor.

 Do not use in explosive atmosphere.  Do not use the device where combustible and/or explosive gasses exist. The device is not designed for explosion proof.  Do not dismantle the base unit. www.GlobalTestSupply.com Find Quality Products Online at: [email protected]

Do not install, use, or store the unit at a place where:

- Ambient temperature is not in the range of -10 to 40 degree in C. (14 to 104 degree in F) - Ambient humidity exceeds 90%R.H. - Condensation occurs due to rapid temperature change. - Corrosive and/or explosive gases exist. - Direct vibration and/or stream occur. - Mush dust, salty materials, and/or iron powders exist. - Static electrons, magnetic, and/or electronic noise occur. - Cool or warm air from conditioners influences directly. - The unit is exposed to direct sunlight. - Heating occurs due to heat radiation.  Do not use or store the unit at a place where the following gases exist: - Extremely concentrated odor. - Cigarette smoke and/or oil mist. - Silicon compound gas included in hair spray, adhesive, and silicon rubber etc. - High concentrated chloride, acidic substances, corrosive gas, and chlorofluorocarbon etc.  Exchange new batteries if battery status displays “Low.”  Do not use rechargeable batteries.  Do not use liquid such as benzene and/or thinner to clean the unit surface.  Do not block suction nozzle and/or exhaust nozzle. www.GlobalTestSupply.com Find Quality Products Online at: [email protected]

1-4. Unit Dimension

Unit: mm www.GlobalTestSupply.com Find Quality Products Online at: [email protected]

2. General Description

Handheld Odor Meter is the most popular simplified tool for odor analysis which enables to show the relative odor strength numerically by comparing odor gases and purified air. The reading shown on the display is calculated based on our original principle. With the installation of data filing software, measured data can be transferred to PC in easy steps. Odor Strength Simply turn on the unit and start measurement. Real-time relative odor strength can be measured continuously. Odor Intensity Odor Strength is converted into ammonia odor intensity by the scale of 2.5 to 5.0 (in 0.1 unit) based on odor concentration of measurement environment. NOTE Odor Intensity is determined based on Japanese Regulation that allows evaluating odor level by 6 scales. Peak-hold Maximum odor strength level can be retained. Memory Sampling Measurement Mode Up to 511 measuring data can be saved upon selecting sampling rate: 1, 2, 5, 10, 20, 60, 120 or 300 seconds. Data Filing Software By using original data filing software, measuring data from the unit can be transferred to PC. www.GlobalTestSupply.com Find Quality Products Online at: [email protected]

2-1-1. LCD Display

No. Name Function ① Ready Blinks for about 10min for initializing when the unit is turned on. Start measuring when it stops blinking. ② Odor Strength Displays odor strength in numeric value from 0 to 999. Start blinking when overflowed. ③ Calendar Function Upper: Year-Month-Date Lower: Hour-Time ④ Peak Hold Retains Maximum odor strength. ⑤ Battery Status ■■■■ Full ■■■□ Good – Measureable for 3 to 4 hours approx. ■■□□ Moderate – Measureable for 1 to 2 hours approx. ■□□□ Low – Change battery immediately. □□□□ No battery. ⑥ USB Shows USB Cable is connected. ⑦ Odor Intensity Displays odor intensity of ammonia gas concentration based by the scale of 2.5 to 5.0 (in 0.1 unit). Hold ENTER for approx. 3 seconds to turn ON or OFF. ⑧ Data No. Shows number of saved data. ⑨ Memory Displays remaining memory capacity from 0 to 100%. ⑩ Sampling Rate Select sampling rate. NOTE “RT” refers to Real-time Sampling that measures Odor Strength and Odor Intensity every 1 second continuously. www.GlobalTestSupply.com Find Quality Products Online at: [email protected]

2-1-2. Base Unit

No. Name Function ①  Hold 3 seconds for “peak-hold function”. (See 3-3.)  Hold and turn on the device at the same time to set calendar. ②  Select sampling rate. (See 3-2.)  Set calendar. ③  Use for “zero-point function". (See 3-4.)  Set calendar. ④ Confirm sampling rate and/or “zero-point” function. ❶ Power Switch Turn on/off the unit. ❷ AC Adaptor Port Connect AC adaptor if necessary. ❸ Suction Nozzle Takes sampling air and/or object gases. ❹ Exhaust Nozzle Exhaust sampling air and/or object gases. ❺ USB Port Connect attached USB cable for PC connection. ❻ Battery Box Open/Close to replace batteries. ❼ Hook Pin Use for straps. Calendar Settings 1. Hold and turn on the device at the same time to set calendar. 2. When “Year” is blinking, change to the arbitrary number. 3. Press to confirm the settings. 4. Repeat 2. And 3. to set month, date, hour, and minute. www.GlobalTestSupply.com Find Quality Products Online at: [email protected]

Connect to the unit during the measurement if necessary.

Air Purifying Unit Attach to the unit for “zero-point function”. (See 3-4.) Activated Carbon Exchange activated carbon when deodorization using air purifying unit is not effective. Replacement Filter for Air Purifying Unit Replace filter when deodorization using air purifying unit is not effective. Replacement Filter for Suction Nozzle Replace filter when gets dirt. USB Connection Cable Use to transfer measuring data from the unit to PC. www.GlobalTestSupply.com Find Quality Products Online at: [email protected]

3-1. Real-time Sampling Mode

3-1-1. Measurement Preparation 1. Place filter to suction nozzle. 2. Attach suction nozzle to the unit before measurement. 3. Use “AA” alkaline batteries or AC adaptor. or 4. Turn on the unit at clean ambient. It takes about 10 min for the unit initializing and start measuring. NOTE The unit sets “zero” point when the initializing is done. NOTE To get the zero point stable, initializing at clean environment is recommended. NOTE To get “Zero” point forcibly, use air purifying unit. (See 3-4.) www.GlobalTestSupply.com Find Quality Products Online at: [email protected]

1. Turn on the unit and wait for about 10 minutes until “Ready” stop blinking and the

initializing is done. 2. Start measuring. 3. Use data filing software if necessary. 4. Turn off the unit when odor strength shows close to “0”. 3-2. Memory Sampling Mode 3-2-1. Measurement Preparation Refer to Page. 11. 3-2-2. Measurement 1. Turn on the unit and wait for about 10 minutes until “Ready” stop blinking and the initializing is done. 2. Press until sampling rate on LCD display starts blinking. 3. Select sampling rate by pressing. 4. Press to finish selecting sampling rate. 5. Press (one time) and (twice) to save data. www.GlobalTestSupply.com Find Quality Products Online at: [email protected]

3-2-3. Memory Overflow

Memory overflow occurs when the internal memory is full. If there is no space remained, calendar on LCD display starts blinking and data number becomes “0”. To delete saved memory, press and at the same time while the unit is turned off. Keep pressing and and turn on the device. NOTE Saved memory in the unit will not be deleted automatically when it is transferred to PC. www.GlobalTestSupply.com Find Quality Products Online at: [email protected]

3-3. Peak-hold Function

Press for over 3 seconds during Real-time sampling. By using this function, the highest value of odor strength keeps being displayed until the unit measures higher value. NOTE Press for over 3 seconds again to cancel the function. NOTE The function is not displayed on the saved data. 3-4. Zero-point Function By using this function, the unit sets the baseline accurately when it is turned on. The unit sets the baseline at the place where the initialization is done. After 10 minutes initializing, the unit will display Odor Strength “0” (baseline) automatically regardless of the measurement environment. If you need to set the baseline accurately, connect Air Purifying Unit to purify the air forcibly. 1. Press until calendar on LCD display “0” and starts blinking. 2. Press to proceed, or press to cancel. NOTE If you set zero-point forcibly under dirty environment, odor strength may keep displaying “0”. In this case, please set zero-point again under clean environment. NOTE Zero-point function cannot be performed in memory sampling mode. www.GlobalTestSupply.com Find Quality Products Online at: [email protected]

How to Install USB Drivers Software

Install USB Drivers software before connecting USB cable to PC. 1. Download the latest software from the link as below. https://www.silabs.com/products/mcu/Pages/USBtoUARTBridgeVCPDrivers.aspx 2. Install the software. NOTE If PC CPU is 32bit, select “CP210xVCPInstaler_x86.exe”. If CPU is 64bit, select “CP210xVCPInstaler_x64.exe”. NOTE Warning message appears if incorrect file is selected. 3. Confirm Device Manager displays “Silicon Labs CP210x USB to UART Bridge (COM)” after the installation. 4. Confirm assigned COM number. www.GlobalTestSupply.com Find Quality Products Online at: [email protected]

Connect the unit to PC by USB connection cable.

4-4. Open Software 1. Double-click shortcut icon on desktop, or open the file from program files. 2. Confirm main screen appears. 3. Read “Handheld Odor Meter Data Filing Software Manual” for details. www.GlobalTestSupply.com Find Quality Products Online at: [email protected]

Semiconductor gas sensors

Object Gas Hydrogen sulfide, methyl mercaptan, ammonia, etc. Sampling Method Continuous sampling by built-in pump Detection Range 0 ppm to 25 ppm approx. (ammonia concentration) Display Digital LCD display Odor Strength (0 to 999) Odor Intensity (2.5 to 5.0) Response Time Ammonia (2ppm): Reaches 90% responsiveness within 20 seconds Ammonia (10ppm): Reaches 90% responsiveness within 20 seconds Power Source 4 “AA” alkaline batteries or AC adaptor Consumed Power 1,100mW approx. with “AA” alkaline batteries Storage Temp. Range 0 to 40 degree in C (No dew condensation) Dimensions 74(W) x 167.50(L) x 35(D) Memory 32,732 data or 511 files www.GlobalTestSupply.com Find Quality Products Online at: [email protected]

6. Maintenance and Servicing

6-1. Replacement of “AA” Alkaline Batteries Replace “AA” alkaline batteries when battery status on LCD display becomes “Low”. 6-2. Replacement of Button Battery The unit uses a button battery (model: CR2032) for memory back up. When to replace: - Pasts 5 years approx. after purchasing. - Confirms when calendar or time on LCD display is incorrect. 1. Turn off the unit and remove “AA” alkaline batteries. NOTE Do not connect AC adaptor. 2. Push the round part of button battery firmly until the left part of button battery moves up. 3. Take out the battery. NOTE Do not use metal tools or easily broken stools like screwdrivers or pencils. 4. Insert the new battery. www.GlobalTestSupply.com Find Quality Products Online at: [email protected]

6-3. Maintenance of Air Purifying Unit

Replace activated carbon and filters in Air Purifying Unit to prevent gases absorbed in the unit. 1. Twist the Air Purifying Unit to open. 2. Take out top wire gauze, top filter, activated carbon, and bottom filter. NOTE Do not take out bottom wire gauze. 3. Replace bottom filter and fill in activated carbon up to the line. NOTE Do not exceed the line when replace activated carbon. 4. Replace top filter and top wire gauze. 6-4. Replacement of Filter for Suction Nozzle Twist Suction Nozzle and replace filter if necessary. Top Top Wire Gauze Top Filter Arrow Label Activated Carbon Bottom Filter Bottom Wire Gauze Bottom Filter Suction Nozzle Do not exceed the line. www.GlobalTestSupply.com Find Quality Products Online at: [email protected]

If you find:

Check: ① The unit does not operate. Check battery status or connect AC adaptor. Check battery polarity. ② Odor Strength does not change. Get the unit closer to object. Connect Air Purifying Unit to clean the absorbed gas in the unit. Clean nozzle with alcohol and make sure to dry it up. ③ Odor Strength on LCD display keeps blinking. Odor concentration is too high and overflowed. Try to lower the gas concentration. ④ Time on LCD display keeps blinking. Memory is overflowed. Delete data. ⑤ Sampling rate cannot be selected. Memory is overflowed. Delete data. ⑥ Data # becomes “0”. Memory is overflowed. Delete data. ⑦ “SErr” is shown on LCD display. Sensor malfunction. Contact us if the message keeps being displayed. ⑧ “PErr” is shown on LCD display. Built-in pump malfunction. Contact us if the message keeps being displayed. www.GlobalTestSupply.com Find Quality Products Online at: [email protected]