Extech Instruments IR300UV Operation & User’s Manual

Specifications

Brand: Extech Instruments
Category: Thermometer
Model: IR300UV (Operation & User’s Manual)
File Type: PDF
File Size: —
Pages: 10

Read Online

Read the Extech Instruments IR300UV Operation & User’s Manual online using the built-in PDF viewer.

Text of the manual

Visit us at www.TestEquipmentDepot.com

This  symbol,  adjacent  to  another  symbol  or

terminal, indicates the user must refer to the manual for further information Warnings o Do not directly or indirectly point the laser at the eyes of a person or an animal o Inspect  for  damage  or  for  any  shortage  of  parts  or accessories before use o Replace the batteries immediately after the battery indicator flashes o Do not use the thermometer near explosive gases, steam, or dust o Note that an object with high reflectivity will normally cause the measured temperature value to read much lower than the actual temperature o Use the device only as described in this User Guide Cautions To avoid thermometer damage, please avoid the following hazards: o EMF from welding equipment or electro‐induction heaters o Static electricity o Thermal  shock  caused  by  large  or  abrupt  environmental  temperature  changes;  wait  30  minutes  to allow the thermometer to stabilize to new environmental conditions o Do not use this device in excessively high temperature environments

EXTECH INSTRUMENTS 3                                              IR300UV‐en‐GB_V1.0   1/16 Description Meter Description 1. LCD Display 2. Multi‐Switch control 3. Battery Compartment 4. Laser pointer 5. White/Blue LEDs 6. IR sensor 7. Lens cover control 8. Measurement Scan Trigger Display Description Laser pointer active icon Hold mode (displayed temperature freezes with trigger release) Maximum/Minimum/Average temperature icons Battery capacity indicator White LEDs active Temperature units (°F /°C) Temperature Alarm icons UV LEDs active Emissivity setting

The  Settings  switch  has  five  positions  to  select  a  parameter,  turn  a

parameter ON or OFF, and to adjust parameter values.  Press and release the Measurement Trigger.  Press  the   button  left  to  enter  the  Setting  Mode.  The parameter selected will appear in the display.  Press the  button right to advance to the next parameter.  Press the up  or down  buttons to change a parameter status or adjust a parameter value.  Press the  button to store parameter changes and advance to the next parameter.  Press the  button left to exit the Settings mode and return to Measurement mode. Note: If no button is pressed within the APO time the meter will shut down. It may be helpful to extend the APO time when making adjustments in the Settings mode.

EXTECH INSTRUMENTS 5                                              IR300UV‐en‐GB_V1.0   1/16 Setting Mode Detail 1. Flashlight Setting Enter the Setting mode. The  Flashlight icon flashes. Press up  for ON or OFF. Press  to store the setting 2. UV light Setting Enter the Setting mode. Advance until the UV icon flashes. Press up  for ON or OFF. Press  to store the setting 3. Temperature Units Setting Enter the Setting mode. Advance until the °F or °C icon flashes. Press up  to change the units. Press  to store the setting 4. Emissivity Setting Enter the Setting mode. Advance until the EMS icon flashes. Press up  or down  to adjust the setting. Press  to store the setting 5. Laser Pointer Setting Enter the Setting mode. Advance until the  icon flashes. Press up  for ON or OFF. Press  to store the setting 6. Max Min Avg Max‐Min Setting Enter the Setting mode. Advance until the MAX – MIN AVG icon flashes. Press up  for ON or OFF. Press  to store the setting 7. Alarm HI Setting Enter the Setting mode. Advance until the Alarm HI icon flashes. Press up  or down  for ON or OFF. Press  to store the setting If ON was selected, Press up  or down  to adjust the limit. Press  to store the setting 8. Alarm LO Setting Enter the Setting mode. Advance until the Alarm LO icon flashes. Press up  or down  for ON or OFF. Press  to store the setting If ON was selected, Press up  or down  to adjust the limit. Press  to store the setting 9. Backlight Setting Enter the Setting mode. Advance until the LItE icon flashes. Press up  or down  for ON or OFF. Press  to store the setting 10. Auto Power OFF Setting Enter the Setting mode. Advance until the APO icon flashes. Press up  or down  to adjust the setting. Press  to store the setting

The  laser  pointer  circle  identifies  the  area  being

measured. The size of the area is determined by the distance to spot ratio. Objects outside the circle are not measured. The spot measured should always be larger than the circle for accurate measurements. MAX, MIN, AVERAGE and MAX‐MIN In the MAX MIN AVG mode the meter will record the Maximum value scanned, Minimum value scanned, the Average of the scanned values and the Difference between the Max and Min value. 1. Turn Max Min ON in the settings mode 2. Press and hold the trigger to scan an area. Max, Min, Avg or Max‐Min will be displayed in the lower right display. Press the up  button to select the parameter to be displayed. The values will be continuously updated for the length of the scan.

Note  that  measurements  should  normally  be  made  further  than  60cm  (24  inches)  from  the

target. The meter can measure from further distances but the measurement may be affected by external sources of light. In addition, the spot size may be so large that it encompasses surface areas not intended to be measured. 2.5cm @ 30cm 5cm @ 60cm 10cm @ 120cm 1” @ 12” 2” @ 24” 4” @ 48” It is necessary to ensure that the size of the target is larger than the spot size. The smaller the target, the closer the distance should be.

EXTECH INSTRUMENTS 8                                              IR300UV‐en‐GB_V1.0   1/16 Emissivity Emissivity represents the energy reflectivity of a material. Most organic materials and painted or oxidized surfaces have an emissivity of approximately 0.95. If possible, masking tape or flat black paint should be applied to cover the measured surface. Wait a period of time to allow the tape or paint to reach thermal equilibrium with the surface of the covered object. Measure the temperature of the surface covered with tape or paint only after equilibrium has been achieved. High Low Limits The HI / LO limits provide an alarm when a preset temperature is exceeded. Lens Cover The lens cover protects the lens from dust when the meter is not in use. Always close the cover when storing the meter. Measurement Notes 1. The object under test should be larger than the spot (target) size calculated by the field of view diagram. 2. If the surface of the object under test is covered with frost, oil, grime, etc., clean before taking measurements. 3. If an object's surface is highly reflective apply masking tape or flat black paint to the surface before measuring. 4. The meter may not make accurate measurements through transparent surfaces such as glass. 5. Steam, dust, smoke, etc. can obscure measurements. 6. The meter compensates for deviations in ambient temperature. It can, however, take up to 30 minutes for the meter to adjust to extremely wide ambient temperature changes. 7. To find a hot spot, aim the meter outside the area of interest then scan across (in an up and down motion) until the hot spot is located. 8. When the meter is held at the optimum distance from the target, the single laser dot will appear at the top of the circular laser target.  As long as the single laser dot is within the circular laser target, the meter will make an accurate measurement.

EXTECH INSTRUMENTS 9                                              IR300UV‐en‐GB_V1.0   1/16 Maintenance Cleaning To clean the lens use compressed air to clear dust and other particles, then carefully clean with a wet cotton swab. The cotton swab should be moistened with clean water. To clean the meter housing, wipe with a damp, soft cloth. Do not use solvents or abrasives. Do not immerse the IR300UV in water or other liquid. Replacing Batteries When the battery icon indicates a low battery, or when the meter doesn’t switch ON, replace the batteries. The battery compartment lid is located on the rear of the handle. Remove the one Philips head screw and lift the lid off. Replace the three 1.5V AA batteries following correct polarity and then replace and secure the battery compartment lid. Battery Safety Notes: Please dispose of batteries responsibly; never dispose of batteries in a fire, batteries may explode or leak. If the meter is not to be used for 60 days or more, remove the battery and store separately. Do not mix battery types or freshness levels; please use batteries of the same type and of the same freshness level. Never dispose of used batteries or rechargeable batteries in household waste. As consumers, users are legally required to take used batteries to appropriate collection sites, the retail store where the batteries were purchased, or wherever batteries are sold. Disposal: Do not dispose of this instrument in household waste. The user is obligated to take end‐of‐life devices to a designated collection point for the disposal of electrical and electronic equipment. Specifications Infrared Thermometer Specifications Range / Resolution ‐30~500oC (‐22~932oF) Accuracy  0°C:   ±1.5°C or 1.5% of reading (whichever is greater)  32°F:   ±3°F or 1.5% of reading (whichever is greater)  ‐10°C to 0°C:   ±2°C 14° to 32°F:   ±4°F Note: Accuracy is specified for the ambient temperature range: 21~25°C (70~77°F) Response Time 500ms (95% of reading) Repeatability (% of reading) ±8% of reading or ±1.0°C (1.8°F) whichever is greater Emissivity Adjustable from 0.10 to 1.00 Temp. Coefficient 0.1°C/°C or ±0.1%/°C of reading (whichever is greater) Field of View D/S = Approx. 12:1 ratio (D = distance, S = spot) Laser power Less than 1mW Spectral response 8 to 14 microns