Indicating and adjustment module PLICSCOM
Document ID:
36433
Indication
and adjustment
About this document
1.1
Function. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
4
1.2
Target group . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
4
1.3
Symbolism used. . . . . . . . . . . . . . . . . . . . . . . . . . . .
4
2
For your safety
2.1
Authorised personnel . . . . . . . . . . . . . . . . . . . . . . . .
5
2.2
Appropriate use . . . . . . . . . . . . . . . . . . . . . . . . . . . .
5
2.3
Warning about misuse . . . . . . . . . . . . . . . . . . . . . . .
5
2.4
General safety instructions . . . . . . . . . . . . . . . . . . . .
5
2.5
CE conformity . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
6
2.6
Fulfillment of NAMUR recommendations . . . . . . . . . .
6
2.7
Environmental instructions. . . . . . . . . . . . . . . . . . . . .
6
3
Product description
3.1
Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
7
3.2
Principle of operation . . . . . . . . . . . . . . . . . . . . . . . .
7
3.3
Packaging, transport and storage . . . . . . . . . . . . . . .
8
4
Prepare setup
4.1
Insert indicating and adjustment module. . . . . . . . . . .
9
4.2
Adjustment system . . . . . . . . . . . . . . . . . . . . . . . . . .
10
5
Set up - Sensors up to SW 3.99
5.1
Adjustment menu for all signal outputs. . . . . . . . . . . .
11
5.2
Adjustment menu 4 … 20 mA/HART . . . . . . . . . . . . .
18
5.3
Adjustment menu Profibus PA. . . . . . . . . . . . . . . . . .
21
6
Set up - Sensors from HW 2.0.0 and SW 4.0.0
6.1
Adjustment menu for all signal outputs. . . . . . . . . . . .
24
6.2
Adjustment menu 4 … 20 mA/HART . . . . . . . . . . . . .
30
7
Maintenance and fault rectification
7.1
Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
32
7.2
How to proceed in case of repair. . . . . . . . . . . . . . . .
32
8
Dismount
8.1
Dismounting steps . . . . . . . . . . . . . . . . . . . . . . . . . .
33
8.2
Disposal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
33
9
Supplement
9.1
Technical data . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
34
9.2
Dimensions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
35
2
Indicating and adjustment module PLICSCOM
Contents
36433-EN-091104
Safety instructions for Ex areas
Please note the Ex-specific safety information for installation and
operation in Ex areas. These safety instructions are part of the
operating instructions manual and come with the Ex-approved
instruments.
Indicating and adjustment module PLICSCOM
3
Contents
36433-EN-091104
1 About this document
1.1
Function
This operating instructions manual provides all the information you
need for mounting, connection and setup as well as important
instructions for maintenance and fault rectification. Please read this
information before putting the instrument into operation and keep this
manual accessible in the immediate vicinity of the device.
1.2
Target group
This operating instructions manual is directed to trained qualified
personnel. The contents of this manual should be made available to
these personnel and put into practice by them.
1.3
Symbolism used
Information, tip, note
This symbol indicates helpful additional information.
Caution: If this warning is ignored, faults or malfunctions can
result.
Warning: If this warning is ignored, injury to persons and/or serious
damage to the instrument can result.
Danger: If this warning is ignored, serious injury to persons and/or
destruction of the instrument can result.
Ex applications
This symbol indicates special instructions for Ex applications.
l
List
The dot set in front indicates a list with no implied sequence.
à
Action
This arrow indicates a single action.
1
Sequence
Numbers set in front indicate successive steps in a procedure.
4
Indicating and adjustment module PLICSCOM
1
About this document
36433-EN-091104
2 For your safety
2.1
Authorised personnel
All operations described in this operating instructions manual must be
carried out only by trained specialist personnel authorised by the plant
operator.
During work on and with the device the required personal protective
equipment must always be worn.
2.2
Appropriate use
The pluggable indicating and adjustment module is used for measured
value indication, adjustment and diagnoses with level and pressure
sensors.
You can find detailed information on the application range in chapter
"Product description".
Operational reliability is ensured only if the instrument is properly used
according to the specifications in the operating instructions manual as
well as possible supplementary instructions.
2.3
Warning about misuse
Inappropriate or incorrect use of the instrument can give rise to
application-specific hazards, e.g. vessel overfill or damage to system
components through incorrect mounting or adjustment.
2.4
General safety instructions
This is a state-of-the-art instrument that fulfils the prevailing regulations
and guidelines. The instrument must only be operated in a technically
flawless and reliable condition. The operator is responsible for the
trouble-free operation of the instrument.
During the entire duration of use, the user is obliged to determine the
compliance of the required occupational safety measures with the
current valid rules and regulations and also take note of new
regulations.
The safety instructions in this operating instructions manual, the
national installation standards as well as the valid safety regulations
and accident prevention rules must be observed by the user.
For safety and warranty reasons, any invasive work on the device
beyond that described in the operating instructions manual may be
carried out only by personnel authorised by the manufacturer. Arbitrary
conversions or modifications are explicitly forbidden.
The safety approval markings and safety tips on the device must be
also observed.
Indicating and adjustment module PLICSCOM
5
2
For your safety
36433-EN-091104
The device fulfills the legal requirements of the applicable EC
guidelines. By attaching the CE mark, VEGA provides a confirmation
of successful testing. You can find the CE conformity declaration in the
download area of www.vega.com.
2.6
Fulfillment of NAMUR recommendations
The requirements of the following NAMUR recommendations are
fulfilled:
l
NE 21 - "Electromagnetic Compatiblity Act of the process and
laboratory control technology"
The sensors as well as the corresponding indicating and adjustment
components are generally upward and downward compatible.
2.7
Environmental instructions
Protection of the environment is one of our most important duties. That
is why we have introduced an environment management system with
the goal of continuously improving company environmental protection.
The environment management system is certified according to DIN
EN ISO 14001.
Please help us fulfil this obligation by observing the environmental
instructions in this manual:
l
Chapter "Packaging, transport and storage"
l
Chapter "Disposal"
6
Indicating and adjustment module PLICSCOM
2
For your safety
36433-EN-091104
3 Product description
3.1
Configuration
This operating instructions manual applies to the following hardware
and software versions of the indicating and adjustment module:
l
Hardware from 1.0.0
l
Software from 1.0.0
The indicating/adjustment module consists of a display with full dot
matrix as well as four keys for adjustment.
A background lighting is optionally integrated and can be adjusted via
the adjustment menu.
The display can be optionally equipped with heating to ensure good
readability at low temperatures down to -40 °C (-40 °F).
The scope of delivery encompasses:
l
Indicating and adjustment module
l
Documentation
-
this operating instructions manual
-
Supplementary instructions manual 31708 "Heating for in-
dicating and adjustment module" (optional)
3.2
Principle of operation
The indicating and adjustment module is used for measured value
indication, adjustment, and diagnostics for the following VEGA plics®
sensors:
l
VEGAPULS series 60
l
VEGAFLEX series 60
l
VEGASON series 60
l
VEGABAR series 50 and 60
l
VEGACAL series 60
The indicating and adjustment module is mounted in the sensor
housing. After installation, the sensor as well as the indicating and
adjustment module are splash-proof, even without housing cover.
Another installation possibility is the external indicating and adjustment
unit VEGADIS 61. The operation of two indicating and adjustment
module in parallel in the sensor and in VEGADIS 61 however is not
supported.
The functional range of the indicating and adjustment module is
determined by the sensr and corresponds to the respective software
version of the sensor.
Power is supplied directly by the respective sensor or by VEGADIS 61.
An additional connection is not necessary.
Scope of the operating
instructions manual
Instrument versions
Scope of delivery
Application area
Functional range
Power supply
Indicating and adjustment module PLICSCOM
7
3
Product description
36433-EN-091104
The backlight is also powered by the sensor or via VEGADIS 61.
Prerequisite for this is a supply voltage at a certain level. The exact
voltage specifications can be found in the operating instructions
manual of the respective sensor.
The optional heating requires its own power supply. You can find
further details in the supplementary instructions manual "Heating for
indicating and adjustment module".
3.3
Packaging, transport and storage
Your instrument was protected by packaging during transport. Its
capacity to handle normal loads during transport is assured by a test
according to DIN EN 24180.
The packaging of standard instruments consists of environment-
friendly, recyclable cardboard. For special versions, PE foam or PE foil
is also used. Dispose of the packaging material via specialised
recycling companies.
Transport must be carried out under consideration of the notes on the
transport packaging. Nonobservance of these instructions can cause
damage to the device.
The delivery must be checked for completeness and possible transit
damage immediately at receipt. Ascertained transit damage or
concealed defects must be appropriately dealt with.
Up to the time of installation, the packages must be left closed and
stored according to the orientation and storage markings on the
outside.
Unless otherwise indicated, the packages must be stored only under
the following conditions:
l
Not in the open
l
Dry and dust free
l
Not exposed to corrosive media
l
Protected against solar radiation
l
Avoiding mechanical shock and vibration
l
Storage and transport temperature see chapter "Supplement -
Technical data - Ambient conditions"
l
Relative humidity 20 … 85 %
Packaging
Transport
Transport inspection
Storage
Storage and transport
temperature
8
Indicating and adjustment module PLICSCOM
3
Product description
36433-EN-091104
4 Prepare setup
4.1
Insert indicating and adjustment module
The indicating and adjustment module can be inserted into the sensor
and removed again at any time. It is not necessary to interrupt the
power supply.
Proceed as follows:
1
Unscrew the housing cover
2
Place the indicating and adjustment module in the desired position
on the electronics (you can choose any one of four different
positions - each displaced by 90°)
3
Press the indicating and adjustment module onto the electronics
and turn it to the right until it snaps in.
4
Screw housing cover with inspection window tightly back on
Removal is carried out in reverse order.
The indicating and adjustment module is powered by the sensor, an
additional connection is not necessary.
Fig. 1: Insert indicating and adjustment module
Note:
If you intend to retrofit the instrument with an indicating and adjustment
module for continuous measured value indication, a higher cover with
an inspection glass is required.
Mount/Dismount indica-
ting and adjustment mo-
dule
Indicating and adjustment module PLICSCOM
9
4
Prepare setup
36433-EN-091104
Fig. 2: Indicating and adjustment elements
1
LC display
2
Adjustment keys
l
[OK] key:
-
Move to the menu overview
-
Confirm selected menu
-
Edit parameter
-
Save value
l
[->] key:
-
Presentation change measured value
-
list entry
-
Select editing position
l
[+] key:
-
Change value of the parameter
l
[ESC] key:
-
interrupt input
-
Jump to next higher menu
The sensor is adjusted via the four keys of the indicating and
adjustment module. The LC display indicates the individual menu
items. The functions of the individual keys are shown in the above
illustration. Approx. 10 minutes after the last pressing of a key, an
automatic reset to measured value indication is triggered. Any values
not confirmed with [OK] will not be saved.
Key functions
Adjustment system
10
Indicating and adjustment module PLICSCOM
4
Prepare setup
36433-EN-091104
5 Set up - Sensors up to SW 3.99
5.1
Adjustment menu for all signal outputs
With the parameter adjustment, the instrument is adapted to the
application conditions. The parameter adjustment is carried out via an
adjustment menu.
Information:
In this operating instructions, all general parameters are described.
Additional instrument-specific parameters are described in the
operating instructions of the respective sensor.
The following presentations are available in the measured value
display:
l
Level as digital value, sensor TAG
l
Level as digital value and bar graph, sensor TAG
l
Only with pressure transmitters: Level or pressure as digital value,
temperature value
With [->] you select the different presentations of the measured value.
You can reach the menu overview from any presentation with [OK].
With [ESC] you return from the menu overview to the measured value
display.
The main menu is divided into five areas with the following functions:
▶ Basic adjustment
Display
Diagnostics
Service
Info
To damp process-dependent measured value fluctuations, you have to
set an integration time of 0 … 999 s in this menu item.
Depending on the sensor type, the factory setting is 0 s or 1 s.
Damping
0 s
In this menu item you select the linearization curve:
l
linear
l
Horizontal cylindrical tank
l
Spherical tank
Measured value indica-
tion
Main menu
Basic adjustment/Dam-
ping
Basic adjustment/Li-
nearization curve
Indicating and adjustment module PLICSCOM
11
5
Set up - Sensors up to SW 3.99
36433-EN-091104
User programmable means: Switching on a linearization curve
programmed via PC and PACTware
A linearization is necessary for all vessels in which the vessel volume
does not increase linearly with the level - e. g. in a horizontal cylindrical
or spherical tank - and the indication or output of the volume is
required. Corresponding linearization curves are preprogrammed for
these vessels. They represent the correlation between the level
percentage and vessel volume. The linearisation applies to the
measured value indication and the current output. By activating the
appropriate curve, the volume percentage of the vessel is displayed
correctly. If the volume should not be displayed in percent but e.g. in l
or kg, a scaling can also be set in the menu item "Display".
Factory setting is linear.
Linearisation curve
linear
Caution:
Note the following, if the respective sensor is used as part of an overfill
protection system according to WHG:
If a linearisation curve is selected, the measuring signal is no longer
compulsorily linear proportional to the level. This must be taken into
consideration by the user, particularly when adjusting the switching
point on the level switch.
In the menu item "Sensor-TAG" you edit a 12-digit measurement loop
name. An unambiguous designation can hence be assigned to the
sensor, e.g. the measurement loop name or the tank or product
designation. In digital systems and in the documentation of larger
plants, a singular designation should be entered for exact identification
of individual measuring sites.
The available digits comprise:
l
Letters from A … Z
l
Numbers from 0 … 9
l
Special characters +, -, /, -
Factory setting is "Sensor".
Sensor-TAG
Sensor
Basic adjustment/Sen-
sor TAG
12
Indicating and adjustment module PLICSCOM
5
Set up - Sensors up to SW 3.99
36433-EN-091104
An integrated background lighting can be adjusted via the adjustment
menu. The function depends on the height of the supply voltage, see
operating instructions of the respective sensor.
In the default setting, the lightning is switched off.
Min. and max. measured values are saved in the sensor. The values
are displayed in the menu item "Peak values".
l
Min. and max. distance in m(d): Radar, guided microwave,
ultrasonic sensors
l
Min. and max. pressure: pressure transmitter1)
l
Min. and max. temperature: ultrasonic sensors, pressure trans-
mitters
Pointer
When non-contact level sensors are used, the measurement can be
influenced by the respective process conditions. In this menu item, the
measurement reliability of the level echo is displayed as dB value. The
measurement reliability equals signal strength minus noise. The higher
the value, the more reliable the measurement. With a functioning
measurement, the values are > 10 dB.
In this menu item, the device status is displayed. If the sensor detects
a fault, "OK" will be displayed. If a fault is detected, a flashing failure
message is outputted sensor-specifically, e.g. "E013". The fault is also
displayed in clear text, e.g. "No measurement value".
Information:
The fault message as well as the clear text indication are also carried
out in the measured value display.
Meas. reliability
Sensor status
1)
Pressure: -50 … +150 % of the nominal pressure range; temperature:
-50 … +150 °C.
Display/Lighting
Diagnosis/Pointer
Diagnosis/Measurement
reliability
Diagnosis/Device status
Indicating and adjustment module PLICSCOM
13
5
Set up - Sensors up to SW 3.99
36433-EN-091104
With ultrasonic and radar sensors as well as sensors with guided
microwave, the "Echo curve" represents the signal strength of the
echoes over the measuring range. The units of the signal strength are
"dB" (ultrasonic and radar) and "Volt" (guided microwave). The signal
strength enables the assessment of the quality of the measurement.
With ultrasonic and radar sensors, the "False echo curve" represents
the saved false echoes (see menu "Service") of the empty vessel with
signal strength in "dB" over the measuring range.
Up to 3000 measured values are recorded (depending on the sensor)
when starting a "Trend curve". Then the values can be displayed on a
time axis. The oldest measured values are always deleted.
In the menu item "Choose curve", the respective curve is selected.
Curve selection
Echo curve ▼
Information:
The trend recording is not activated when being shipped. It must be
started by the user via the menu item "Start trend curve".
A comparison of echo and false echo curve allows a more detailled
specification on the measurement reliability. The selected curve is
updated permanently. With the [OK] key, a submenu with zoom
functions is opened.
The following functions are available with "Echo and false echo
curve":
l
"X-Zoom": Zoom function for the meas. distance
l
"Y-Zoom": 1, 2, 5 and 10-fold signal magnification in "dB"
l
"Unzoom": Reset the presentation to the nominal measuring range
with single magnification
In the menu item "Trend curve" the following are available:
l
"X-Zoom": Resolution
-
1 minute
-
1 hour
-
1 day
l
"Stop/Start": Interrupt a recording or start a new recording
l
"Unzoom": Reset the resolution to minutes
As default setting, the recording pattern has 1 minute. With the
adjustment software PACTware, this pattern can be also set to 1 hour
or 1 day.
Diagnosis/Curve selec-
tion
Diagnosis/Curve pre-
sentation
14
Indicating and adjustment module PLICSCOM
5
Set up - Sensors up to SW 3.99
36433-EN-091104
In this menu item you simulate a user-defined level or pressure value
via the current output. This allows you to test the signal path, e.g.
through connected indicating instruments or the input card of the
control system.
The following simulation variables are available:
l
Percent
l
Current
l
Pressure (with pressure transmitters)
l
Distance (with radar and guided microwave)
With Profibus PA sensors, the selection of the simulated value is made
via the "Channel" in the menu "Basic adjustments".
How to start the simulation:
1
Push [OK]
2
Select the requested simulation variable with [->] and confirm with
[OK]
3
Set the requested numerical value with [+] and [->].
4
Push [OK]
The simulation is now running, with 4 … 20 mA/HART a current is
outputted and with Profibus PA or Foundation Fieldbus a digital value.
How to interrupt the simulation:
à
Push [ESC]
Information:
The simulation is terminated automatically 10 minutes after the last key
has been pushed.
Simulation
Start simulation?
With the reset function, modified values are reset. Three subfunctions
are available:
l
Basic adjustment
-
Reset the values modified with the indicating and adjustment
module to the sensor-specific basic setting
l
Factory setting
Service/Simulation
Service/Reset
Indicating and adjustment module PLICSCOM
15
5
Set up - Sensors up to SW 3.99
36433-EN-091104
As basic adjustment, but also reset of special parameters to
the default values2)
l
Peak values measured value and temperature3)
-
Reset of the min./max. values of pressure, level and temper-
ature to the current values
Information:
Because the reset values are nearly sensor-specifc, they are listed in
the operating instructions manual of the respective sensor.
Reset
Select reset?
In this menu item you select the internal arithmetic unit of the sensor.
With radar, guided microwave and ultrasonic sensors this is m(d) or
ft(d).
For pressure transmitters more comprehensive units are available.
They are described in the operating instructions manual of the
respective sensor in the menu "Basic adjustments".
Unit of measurement
m(d)
The sensor is already set to the ordered national language. In this
menu item you can change the language. The following languages are
available, e.g. in software version 3.50:
l
Deutsch
l
English
l
Français
l
Espanõl
l
Pycckuu
l
Japanese
l
Italiano
l
Netherlands
l
Japanese
l
Chinese
2)
Special parameters are parameters which are set customer-specifically on
the service level with the adjustment software PACTware.
3)
Temperature only with pressure transmitters and ultrasonic sensors.
Service/Adjustment unit
Service/Language
16
Indicating and adjustment module PLICSCOM
5
Set up - Sensors up to SW 3.99
36433-EN-091104
With this function
l
Load parameter adjustment data from the sensor into the
indicating and adjustment module
l
Write parameter adjustment data from the indicating and adjust-
ment module into the sensor
The data are permanently saved in an EEPROM memory in the
indicating and adjustment module and remain there even in case of
power failure. From there, they can be written in one or several
sensors or kept as backup for a probable sensor exchange.
Kind and volume of the copied data depend on the respective sensor.
Information:
Before writing the data into the sensor, it is checked if the data fit the
sensor. If data do not fit, a fault signal is triggered or the function is
blocked. When writing data into the sensor, you will see from which
instrument type the data originate and which TAG-no. this sensor had.
The following items are checked:
l
Software version
l
WHG approval
l
SIL activated
l
Measuring principle
l
Radar C-band/K-band
l
Radar measuring range < 30 m or > 30 m
l
Signal output
l
Pressure measuring range
Copy sensor data
Copy sensor data?
In this menu item, the PIN is activated/deactivated permanently.
Entering a 4-digit PIN protects the sensor data against unauthorized
access and unintentional modifications. If the PIN is activated
permanently, it can be deactivated temporarily (i.e. for approx. 60 min.)
in any menu item. The instrument is delivered with the PIN set to 0000.
Service/Copy sensor
data
Service/PIN
Indicating and adjustment module PLICSCOM
17
5
Set up - Sensors up to SW 3.99
36433-EN-091104
Only the following functions are permitted with activated PIN:
l
Select menu items and show data
l
Read data from the sensor into the indicating/adjustment module.
In this menu item the most important sensor information can be
displayed:
l
Instrument type
l
Serial number: 8-digit number, e.g. 12345678
Instrument type
Serial number
12345678
l
Date of manufacture: Date of the factory calibration, e.g. 10.
January 2008
l
Software version: Edition of the sensor software, e.g. 3.50
Date of manufacture
10. January 2008
Software version
3.50
l
Date of last change using PC: Date of the last change of sensor
parameters via PC, e.g. 10. January 2008
Last change using PC
10. January 2008
l
Sensor details, e.g. approval, process fitting, seal, measuring cell,
measuring range, electronics, housing, cable entry, plug, cable
length etc.
Sensor characteristics
Display now?
5.2
Adjustment menu 4 … 20 mA/HART
In the menu item "Display" you can define how the measured value
should be presented on the display.
Info/Info
Display/Display
18
Indicating and adjustment module PLICSCOM
5
Set up - Sensors up to SW 3.99
36433-EN-091104
The following values can be displayed:
l
Height
l
Pressure (only with pressure transmitters)
l
Distance (only with radar, guided microwave, ultrasonics)
l
Current
l
Scaled
l
Percent
l
Lin. percent
l
Temperature (only with pressure transmitters).
The selection "scaled" opens the menu items "Display unit" and
"Scaling". In "Display unit" there are the following options:
l
Height
l
Mass
l
Flow
l
Volume
l
Without unit
Depending on selection, the different units are in turn available.
In the menu item "Scaling", the requested numerical value with
decimal point is entered for 0 % and 100 % of the measured value.
There is the following relation between the indication value in the menu
"Display" and the adjustment unit in the menu "Basic adjustment":
l
With radar, guided microwave and ultrasonics, displayed value
"Distance" means: presentation of the measured value in the
selected adjustment unit, e.g. m(d)
l
With pressure, displayed value "Pressure" or "Height" means:
presentation of the measured value in the selected adjustment
unit, e.g. bar or m.
Displayed value
Scaled ▼
Display unit
Volume ▼
l ▼
Scaling
0 % = 0.0 l
100 % = 100.0 l
In the menu item "Current output" you determine the behaviour of the
current output during operation and in case of failure. The following
options are available:
Service/Current output
Indicating and adjustment module PLICSCOM
19
5
Set up - Sensors up to SW 3.99
36433-EN-091104
4 … 20 mA
20 … 4 mA
Failure mode4)
Hold value
20.5 mA
22 mA
< 3.6 mA
Min. current5)
3.8 mA
4 mA
Max. current6)
20 mA
20.5 mA
The values in bold font represent the data of the factory setting.
In HART multidrop mode, the current is constantly 4 mA. This value
does not change even in case of failure.
Current output
Output mode: 4-20 mA ▼
Failure mode: 22 mA ▼
Min. current 3.8 mA ▼
The functional safety is already activated Ex factory for instruments
with SIL qualification. For instruments Ex factory without SIL
qualification, the functional safety must be activated by the user for
applications according to SIL via the indicating and adjustment
module. The SIL factory setting cannot be deactivated by the user.
The activation of SIL has the following impact:
l
In the menu item "Failure mode" under "Current output", the
parameters "Hold value" and "20.5 mA" are blocked
l
In the menu item "HART mode", the function "Multidrop" is blocked
Note:
For such applications, it is absolutely necessary to take note of "Safety
Manual".
HART offers standard and multidrop mode.
The mode standard with the fixed address 0 means output of the
measured value as 4 … 20 mA signal.
4)
Value of the current output in case of failure, e.g. if no valid measured value
is delivered.
5)
This value is not underrun during operation.
6)
This value is not exceeded during operation.
Service/Functional safe-
ty (SIL)
Service/HART mode
20
Indicating and adjustment module PLICSCOM
5
Set up - Sensors up to SW 3.99
36433-EN-091104
In Multidrop mode, up to 15 sensors can be operated on one two-wire
cable. An address between 1 and 15 must be assigned to each
sensor.7)
In this menu item you determine the HART mode and enter the
address for multidrop.
HART mode
Standard
Address 0
The default setting is standard with address 0.
5.3
Adjustment menu Profibus PA
Level and pressure sensors operate as slaves on the Profibus PA. To
be identified as a bus participant, each sensor must have a unique
address. Each instrument is delivered with address 126. With this
address, it can at first be connected to an existing bus. However, the
address must be changed. This can be done in this menu item.
Sensor address
126
The channel is the input selector switch for function block (FB) of the
sensor. Within the function block, additional scalings (Out-Scale) are
carried out. In this menu item, the value fir the function block is
selected:
l
SV1 (Secondary Value 1):
-
Percent with radar, guided microwave and ultrasonic sensors
-
Pressure or height with pressure transmitters
l
SV2 (Secondary Value 2):
-
Distance with radar, guided microwave and ultrasonic sensors
-
Percent with pressure transmitters
l
PV (Primary Value):
-
Linearised percentage value
Channel
PV lin. value
7)
The 4 … 20 mA signal of the HART sensor is switched off. The sensor
consumes a constant current of 4 mA. The measuring signal is transmitted
exclusively as digital HART signal.
Basic adjustment/Sen-
sor address
Channel
Indicating and adjustment module PLICSCOM
21
5
Set up - Sensors up to SW 3.99
36433-EN-091104
Radar, guided microwave and ultrasonic sensors deliver the following
measured values:
l
SV1 (Secondary Value 1): Percentage value after the adjustment
l
SV2 (Secondary Value 2): Distance value before the adjustment
l
PV (Primary Value): Linearised percentage value
l
PA-Out (value after passing the function block): PA output
A pressure transmitter delivers the following measured values:
l
SV1 (Secondary Value 1): Pressure or height value before
adjustment
l
SV2 (Secondary Value 2): Percentage value after the adjustment
l
PV (Primary Value): Linearised percentage value
l
PA-Out (value after passing the function block): PA output
l
Temperature
In the menu item "Display" you can define which measured value
should be presented on the display.
Displayed value
PA-Out
Profibus transmits two values cyclically. The first value is determined in
the menu item "Channel". The selection of the additional cyclical value
is made in the menu item "Additional PA value".
The following values are available with radar, guided microwave and
ultrasonic sensors:
l
SV1 (Secondary Value 1): Percentage value after the adjustment
l
SV2 (Secondary Value 2): Distance value before the adjustment
l
PV (Primary Value): Linearised percentage value
With pressure transmitters the following values are available:
l
SV1 (Secondary Value 1): Pressure or height value before
adjustment
l
SV2 (Secondary Value 2): Percentage value after the adjustment
l
PV (Primary Value): Linearised percentage value
Additional PA value
Display/Display
Service/Additional PA
value
22
Indicating and adjustment module PLICSCOM
5
Set up - Sensors up to SW 3.99
36433-EN-091104
Here, you determine the unit and scaling for PA-Out. These settings
also apply to the values displayed on the indicating and adjustment
module if in the menu item "Displayed value" PA-Out was selected.
The following displayed values are available in "Out-Scale unit":
l
Pressure (only with pressure transmitters)
l
Height
l
Mass
l
Flow
l
Volume
l
Others (no unit, %, mA)
In the menu item "PV-Out-Scale", the requested numerical value with
decimal point is entered for 0 % and 100 % of the measured value.
Out-Scale-Unit
Service/Determine Out-
Scale
Indicating and adjustment module PLICSCOM
23
5
Set up - Sensors up to SW 3.99
36433-EN-091104
6 Set up - Sensors from HW 2.0.0 and SW 4.0.0
6.1
Adjustment menu for all signal outputs
With the parameter adjustment, the instrument is adapted to the
application conditions. The parameter adjustment is carried out via an
adjustment menu.
Information:
In this operating instructions, all general parameters are described.
Further, instrument-specific parameters are described in the operating
instructions of the respective sensor.
The following presentations are available in the measured value
display:
l
Level as digital value, sensor TAG
l
Level as digital value and bar graph, sensor TAG
With [->] you select the different presentations of the measured value.
You can reach the menu overview from any presentation with [OK].
With [ESC] you return from the menu overview to the measured value
display.
The main menu is divided into five areas with the following functions:
Setup: Contains settings for medium, application, vessel, adjustment,
signal output as well as device TAGs
Display: Contains language changeover, settings to the indicated
measured value as well as for lighting
Diagnosis: Contains information on pointer, reliability, echo curve and
simulation
Additional settings: Contains false signal suppression, linearization
curve, sensor length, reset, adjustment unit, SIL function, HART mode
Info: Shows instrument type, serial number, software version, last
change, sensor characteristics
Measured value indica-
tion
Main menu
24
Indicating and adjustment module PLICSCOM
6
Set up - Sensors from HW 2.0.0 and SW 4.0.0
36433-EN-091104
In the menu item "Sensor-TAG" you edit a 12-digit measurement loop
name. An unambiguous designation can hence be assigned to the
sensor, e.g. the measurement loop name or the tank or product
designation. In digital systems and in the documentation of larger
plants, a singular designation should be entered for exact identification
of individual measuring sites.
The available digits comprise:
l
Letters from A … Z
l
Numbers from 0 … 9
l
Special characters +, -, /, -
To damp process-dependent measured value fluctuations, you have to
set an integration time of 0 … 999 s in this menu item.
Depending on the sensor type, the factory setting is 0 s or 1 s.
In the delivery status, the sensor is set to the ordered national
language.
This menu item enables the setting of another national language.
The optionally integrated background lighting can be adjusted via the
adjustment menu. The function depends on the height of the supply
voltage, see operating instructions of the respective sensor.
The lighting is switched off in the delivery status.
Setup/Measurement
loop name
Setup/Damping
Display/Language
Display/Lighting
Indicating and adjustment module PLICSCOM
25
6
Set up - Sensors from HW 2.0.0 and SW 4.0.0
36433-EN-091104
In this menu item, the device status is displayed.
Min. and max. measured values are saved in the sensor. The values
are displayed in the menu item "Peak values".
In this menu item, the electronics temperature is displayed.
In this menu item you simulate measured values via the current output.
This allows the signal path to be tested, e.g. via connected indicating
instruments or the input card of the processing system.
Information:
With Profibus PA sensors, the selection of the simulated value is made
via the "Channel" in the menu "Basic adjustments".
How to start the simulation:
1
Push [OK]
2
Select the requested simulation variable with [->] and confirm with
[OK]
3
Set the requested numerical value with [+] and [->].
4
Push [OK]
The simulation is now running, with 4 … 20 mA/HART a current is
outputted and with Profibus PA or Foundation Fieldbus a digital value.
Diagnosis/Device status
Diagnosis/Pointer
Diagnosis/Electronics
temperature
Diagnosis/Simulation
26
Indicating and adjustment module PLICSCOM
6
Set up - Sensors from HW 2.0.0 and SW 4.0.0
36433-EN-091104
How to interrupt the simulation:
à
Push [ESC]
Information:
The simulation is terminated automatically 10 minutes after the last key
has been pushed.
In this menu item, you select the measured variable of the system.
A linearization is necessary for all vessels in which the vessel volume
does not increase linearly with the level - e. g. in a horizontal cylindrical
or spherical tank - and the indication or output of the volume is
required. Corresponding linearization curves are preprogrammed for
these vessels. They represent the correlation between the level
percentage and vessel volume. The linearisation applies to the
measured value indication and the current output. By activating the
appropriate curve, the volume percentage of the vessel is displayed
correctly. If the volume should not be displayed in percent but e.g. in l
or kg, a scaling can be also set in the menu item "Display".
Caution:
Note the following, if the respective sensor is used as part of an overfill
protection system according to WHG:
If a linearisation curve is selected, the measuring signal is no longer
compulsorily linear proportional to the level. This must be taken into
consideration by the user, particularly when adjusting the switching
point on the level switch.
Additional settings/De-
vice settings
Further settings/Linear-
ization
Indicating and adjustment module PLICSCOM
27
6
Set up - Sensors from HW 2.0.0 and SW 4.0.0
36433-EN-091104
In this menu item, the PIN is activated/deactivated permanently.
Entering a 4-digit PIN protects the sensor data against unauthorized
access and unintentional modifications. If the PIN is activated
permanently, it can be deactivated temporarily (i.e. for approx. 60 min.)
in any menu item. The instrument is delivered with the PIN set to 0000.
Caution:
With active PIN, the adjustment via PACTware/DTM as well as other
systems is also blocked.
Only the following functions are permitted with activated PIN:
l
Select menu items and show data
l
Read data from the sensor into the indicating/adjustment module.
In delivery status, the PIN is "0000".
The instrument has an internal clock.
With this function the sensor settings are copied. The following
functions are available:
l
Store data from the sensor into the indicating/adjustment module.
l
Store data from the indicating and adjustment module in the
sensor
The following data or settings for adjustment of the indicating and
adjustment module are saved:
l
All data of the menu "Setup" and "Display"
l
In the menu "Additional settings" the items "Distance unit,
temperature unit and linearization"
l
The values of the user programmable linearization curve
Additional settings/PIN
Further settings/Date
and time
Further settings/Copy
sensor settings
28
Indicating and adjustment module PLICSCOM
6
Set up - Sensors from HW 2.0.0 and SW 4.0.0
36433-EN-091104
The copied data are permanently saved in an EEPROM memory in the
indicating and adjustment module and remain there even in case of
power failure. From there, they can be written into one or more sensors
or kept as backup for a possible sensor exchange.
Kind and volume of the copied data depend on the respective sensor.
Note:
Before the data are stored in the sensor, a check is carried out to
determine if the data fit the sensor. If the data do not fit, a fault signal is
triggered or the function is blocked. When data are being written into
the sensor, the display shows which instrument type the data originate
from and which TAG-no. this sensor had.
In this menu, you read out the instrument name and the instrument
serial number:
In this menu item, the hardware and software version of the sensor is
displayed.
In this menu item, the date of factory calibration of the sensor as well
as the date of the last change of sensor parameters are displayed via
the indicating and adjustment module or via the PC.
Info/Instrument type
Info/Sensor version
Info/Calibration date
Indicating and adjustment module PLICSCOM
29
6
Set up - Sensors from HW 2.0.0 and SW 4.0.0
36433-EN-091104
In this menu item, the features of the sensor such as approval, process
fitting, seal, measuring range, electronics, housing and others are
displayed.
6.2
Adjustment menu 4 … 20 mA/HART
In the menu item "Current output mode" you determine the output
characteristics and the fault mode of the current output.
In the menu item "Current output Min./Max.", you determine the
reaction of the current output during operation and in case of failure.
In this menu item, you define how the measured value is displayed.
There is the following relationship between the indication value in the
menu "Display" and the adjustment unit in the menu "Device settings":
Sensor characteristics
Setup/Current output
mode
Setup/Current output
Min./Max.
Display/Displayed value
30
Indicating and adjustment module PLICSCOM
6
Set up - Sensors from HW 2.0.0 and SW 4.0.0
36433-EN-091104
With radar, guided microwave and ultrasonic sensors, displayed
value "Distance" means: presentation of the measured value in the
selected adjustment unit, e.g. m(d)
l
With pressure transmitters, displayed value "Pressure" or "Height"
means: presentation of the measured value in the selected
adjustment unit, e.g. bar or m.
In this menu item, you define the scaling of the measured value on the
display.
In the menu item "Scaling", the requested numerical value with
decimal point is entered for 0 % and 100 % of the measured value.
In this menu item, you define the scaling of the measured value on the
display.
The sensor offers the HART modes standard and Multidrop. In this
menu item you determine the HART modes and enter the address with
Multidrop.
The mode standard with the fixed address 0 means output of the
measured value as 4 … 20 mA signal.
In Multidrop mode, up to 63 sensors can be operated on one two-wire
cable. An address between 1 and 126 must be assigned to each
sensor.8)
The default setting is standard with address 0.
8)
The 4 … 20 mA signal of the HART sensor is switched off. The sensor
consumes a constant current of 4 mA. The measuring signal is transmitted
exclusively as digital HART signal.
Display/Scaling size
Display/Scaling
Further settings/HART
mode
Indicating and adjustment module PLICSCOM
31
6
Set up - Sensors from HW 2.0.0 and SW 4.0.0
36433-EN-091104
7 Maintenance and fault rectification
7.1
Maintenance
When the device is used correctly, no maintenance is required in
normal operation.
7.2
How to proceed in case of repair
If a repair is necessary, please proceed as follows:
You can download a return form (23 KB) from our Internet homepage
www.vega.com under: "Downloads - Forms and certificates - Repair
form".
By doing this you help us carry out the repair quickly and without
having to call back for needed information.
l
Print and fill out one form per instrument
l
Clean the instrument and pack it damage-proof
l
Attach the completed form and, if need be, also a safety data
sheet outside on the packaging
l
Please ask the agency serving you for the address of your return
shipment. You can find the respective agency on our website
www.vega.com.
32
Indicating and adjustment module PLICSCOM
7
Maintenance and fault rectification
36433-EN-091104
Before dismounting, be aware of dangerous process conditions such
as e.g. pressure in the vessel, high temperatures, corrosive or toxic
products etc.
Take note of chapters "Mounting" and "Connecting to power supply"
and carry out the listed steps in reverse order.
8.2
Disposal
The instrument consists of materials which can be recycled by
specialised recycling companies. We use recyclable materials and
have designed the electronics to be easily separable.
WEEE directive 2002/96/EG
This instrument is not subject to the WEEE directive 2002/96/EG and
the respective national laws. Pass the instrument directly on to a
specialised recycling company and do not use the municipal collecting
points. These may be used only for privately used products according
to the WEEE directive.
Correct disposal avoids negative effects to persons and environment
and ensures recycling of useful raw materials.
Materials: see chapter "Technical data"
If you have no possibility to dispose of the old instrument
professionally, please contact us concerning return and disposal.
Indicating and adjustment module PLICSCOM
33
8
Dismount
36433-EN-091104
approx. 150 g (0.33 lbs)
Ambient conditions
Ambient temperature
-15 … +70 °C (+5 … +158 °F)
Storage and transport temperature
-40 … +80 °C (-40 … +176 °F)
Indicating and adjustment module
Voltage supply and data transmission
through the sensor
Indication
LC display in dot matrix
Adjustment elements
4 keys
Protection rating
-
unassembled
IP 20
-
mounted into the sensor without cover
IP 40
Materials
-
Housing
ABS
-
Inspection window
Polyester foil
Display light
Power supply
through the sensor, voltage range see sensor
operating instructions manual
34
Indicating and adjustment module PLICSCOM
9
Supplement
36433-EN-091104
Fig. 3: Dimensions PLICSCOM
Indicating and adjustment module PLICSCOM
35
9
Supplement
36433-EN-091104
Industrial property rights
VEGA product lines are global protected by industrial property rights.
Further information see http://www.vega.com.
Only in U.S.A.: Further information see patent label at the sensor
housing.
VEGA Produktfamilien sind weltweit geschützt durch gewerbliche
Schutzrechte.
Nähere Informationen unter http://www.vega.com.
Les lignes de produits VEGA sont globalement protégées par des
droits de propriété intellectuelle. Pour plus d'informations, on pourra
se référer au site http://www.vega.com.
VEGA lineas de productos están protegidas por los derechos en el
campo de la propiedad industrial. Para mayor información revise la
pagina web http://www.vega.com.
Линии продукции фирмы ВЕГА защищаются по всему миру
правами на интеллектуальную собственность. Дальнейшую
информацию смотрите на сайте http://www.vega.com.
VEGA系列产品在全球享有知识产权保护。
进一步信息请参见网站。
9.4
Trademark
All the brands as well as trade and company names used are property
of their lawful proprietor/originator.
36
Indicating and adjustment module PLICSCOM
9
Supplement
36433-EN-091104
Additional PA value
- Primary Value 22
- Secondary Value 22
Adjustment system 10
Application area 7
C
Channel with Profibus PA 21
Copy sensor data 17
Copy sensor settings 28
Current output 19
Current output Min./Max. 30
D
Damping 11, 25
Date of manufacture 29
Date/Time 28
Display 18
Display with Profibus PA 22
Displayed value 30
E
Echo curve 14
Electronics temperature 26
F
Functional range 7
Functional safety (SIL) 20
H
HART mode
- Multidrop 20, 31
- Standard 20, 31
I
Info 18
Instrument type 29
Instrument versions 7
K
Key function 10
L
Language 16, 25
Lighting 13, 25
Linearisation 27
Linearisation curve 11
M
Main menu 11, 24
Meas. reliability 13
Measured value indication 11, 24
Measured value presentation 18
Measured variable of the system 27
Measurement loop name 25
O
Out-Scale with Profibus PA 23
P
PIN 17, 28
Pointer 13, 26
Power supply 7
R
Reset 15
S
Scaling measured value for display 31
Scaling of the measured value for display 31
Sensor address Profibus PA 21
Sensor characteristics 30
Sensor status 13, 26
Sensor version 29
Sensor-TAG 12
Simulation 15, 26
U
Unit of measurement 16
Indicating and adjustment module PLICSCOM
37
Index
36433-EN-091104
VEGA Grieshaber KG
Am Hohenstein 113
77761 Schiltach
Germany
Phone +49 7836 50-0
Fax +49 7836 50-201
E-mail: [email protected]
www.vega.com
Printing date:
ISO 9001�
All statements concerning scope of delivery, application,
practical use and operating conditions of the sensors and
processing systems correspond to the information avail-
able at the time of printing.
© VEGA Grieshaber KG, Schiltach/Germany 2009
Subject to change without prior notice