Tektronix 7D11 Instruction Manual
Read Online
Read the Tektronix 7D11 Instruction Manual online using the built-in PDF viewer.
Text of the manual
Tektronix, Inc .
�
P.O . Box 500
�
Beaverton, Oregon 97005
070-1329-00
�
772
materials and workmanship
for one year . Any
questions with respect to the warranty should be taken up
with your TEKTRONIX
Field Engineer or representative .
All
requests for repairs and replacement parts should be
directed to the TEKTRONIX Field Office or representative
in your
area .
This
will
assure
you
the
fastest possible
service . Please include the instrument Type Number or Part
Number and Serial Number with
all requests for parts or
service .
Specifications and price change privileges reserved .
Copyright © 1972 by Tektronix, Inc ., Beaverton, Oregon.
Printed in the United States of America . All rights reserved .
Contents of this publication may not be reproduced in any
form without permission of Tektronix, Inc .
U.S.A . and foreign TEKTRONIX products covered by U .S .
and foreign patents and/or patents pending .
TEKTRONIX
is a registered trademark of Tektronix, In
TABLE OF CONTENTS
SECTION 1
SECTION 2
OPERATING INSTRUCTIONS
Page
7Q11 Features
1-1
Preliminary Information
Installation
1-1
Display
1-1
Controls and Connectors
1-3
General Operating Information
Signal Connection
1-3
Count Mode
1-4
Trigger Controls
1-4
Triggered Light
1-4
Trigger Coupling
1-4
Trigger Source
1-5
Trigger Slope/Level
1-5
Events Start Trigger
1-5
Delay Time or Events
1-6
Fine Delay
1-6
B Sweep Delay Mode
1-6
Output Signals
Front-Panel Output Signals
1-6
Output Signals to Mainframe
1-7
Operating Modes
Sweep Delay
1 -8
Echo Delay Time Mode
1
Applications
'
General
1-9
Delayed-Sweep Time Measurements
1-9
Displaying Complex Signals with Delayed Sweep
1-9
Pulse Jitter Measurements
1-10
SPECIFICATION
Introduction
2-1
Table 2-1, Electrical Characteristics
2-1
Table 2-2, Environmental Characteristics
2-2
Table 2-3, Physical Characteristics
2-2
The 71311 Digital Delay plug-in unit is designed for use
with TEKTRONIX 7000-Series Oscilloscope
mainframes
equipped
with a
readout
system . The
71311
uses
the
readout system to display the selected delay count on the
CRT. A delayed trigger output is generated by counting
time or events . Digital delay time to one second is read out
in
100-nanosecond
increments . Additional
analog
delay
time
from
zero
to
100
nanoseconds selected from
a
calibrated front-panel dial provides added resolution . An
"echo" time-delay mode provides a divide-by-two scaler to
read out the "one-way trip" time up to two seconds for
radar ranging and TDR applications. Delay time accuracy is
controlled by an internal crystal oscillator ; greater accuracy
can be obtained by the use of an external one-megahertz
standard . In the Count by Events mode, the CRT readout
displays the integer number of events from one to 101
events at count rates up to 50 megahertz .
The 71311 can be used to delay a 7B-series time-base unit
in either a runs-after or triggerable-after delay time mode .
Other 71311
features
include on-screen display of delay
interval
by
vertical
signal
or
display
blanking, trigger
pickoff from vertical amplifier unit, and blanking of all
zeros to the left of the most significant digit (Count by
Time mode only).
Installation
The 71311
is designed to operate in any plug-in compart-
ment of TEKTRONIX 7000-series mainframes. However,
certain modes of operation require the 71311 to be installed
in a specific compartment. The unit must be operated in a
horizontal compartment to trigger from a signal applied to
OPERATING INSTRUCTIONS
PRELIMINARY INFORMATION
Section 1-71311 Operators
a vertical amplifier unit . The 71311 must be operated in the
A Horizontal compartment to control the delay mode of a
time-base unit in the B Horizontal compartment. Operation
in a vertical compartment is necessary to view the Delay
Interval Pedestal without the use of external cables .
To install the 71311
into a plug-in compartment, push
the unit in until it is seated flush against the front panel of
the
mainframe.
To
remove,
pull
the
release
latch
to
disengage the 71311 . Continue to pull the release latch to
remove the unit from the mainframe.
Display
The 71311' readout display
is presented on the CRT of
the mainframe,
along with information encoded by the
other plug-in units. Digital delay time (in milliseconds) is
displayed in four to eight digits . All zeros to the left of the
most significant digit are blanked out. The + symbol to the
right of the digital display reminds the operator to add any
analog delay time selected by the FINE DELAY (ns) dial to
the delay time . The number of events being counted
is
presented in a seven-digit display.
The 71311 readout display appears on the CRT in a
location corresponding to the plug-in compartment used .
The delay time or number of events will be displayed in the
top division of the CRT graticule. The delay-time measure-
ment unit (ms) will be displayed in the bottom division . It
is not necessary to select the 71311 with the mainframe
Vertical or Horizontal Mode switches to view the digital
display .
In
order to
view
the
Delay Interval
Pedestal
waveform,
selection
with the
Vertical
Mode switch
is
required .
CONTROLS AND CONNECTORS
The major controls and connectors for operation of the
71311
are located on the front panel of the unit . Two
controls located inside the unit for auxiliary functions are
described
in the General Operating Information section.
The front-panel controls and connectors are identified
in
Fig. 1-2; their functions are as follows:
COUNT MODE Switch
Selects
mode
of
operation
and
clock-signal
source for Time count mode .
B SWEEP DELAY MODE Switch
Selects the delay mode logic for time-base unit in
B Horizontal compartment of mainframe.
TRIGGER SLOPE/LEVEL Controls
Select slope and amplitude point of input signal
where the delay is initiated.
TRIG'D Indicator
Lights when a trigger is produced .
COUPLING Switch
Selects the method of coupling the input signal
to the Trigger circuit.
6O
SOURCE Switch
Selects Trigger input signal source .
EVENTS
START
TRIGGER
SLOPE/LEVEL
Controls
Select slope and amplitude point of input signal
where the Events Start count is initiated.
DELAY TIME OR EVENTS Control
Selects delay time or number of events counted.
Direction of rotation selects increase or decrease
in delay time or number of events .
FINE DELAY (ns) Dial
Selects analog delay time added to digital delay
time
selected by DELAY TIME OR EVENTS
control.
REV. JAN 1974
10
DLY'D TRIG OUT
Operating Instructions-71371 Operators
BNC
connector
for
Delayed
Trigger output
signal .
EXT 1 MHz IN or EVENTS START TRIG IN
BNC connector for input of
external
1 MHz
time-reference
signal
or
input of Events Start
Trigger input signal . Connector function is deter-
mined by setting of COUNT MODE switch .
DLY INTERVAL OUT
BNC connector for Delay Interval output signal .
EXT TRIG IN
BNC connector for external
input to Trigger
circuit
(selected by SOURCE
switch
in
the
external positions) .
GENERAL OPERATING INFORMATION
Signal Connection
RESET
Resets the DELAY TIME OR EVENTS to
0000001 (EFF SN B030000 and up) .
In general, probes offer the most convenient means of
connecting
signals to the 71311
external
trigger inputs.
TEKTRONIX
probes are shielded to prevent pickup of
electrostatic interference . A 10X attenuation probe offers a
high input impedance and allows the circuit under test to
perform very close to normal operating conditions . Also, a
10X probe attenuates the input signal ten times.
TEKTRONIX probes are designed to monitor the signal
source with minimum circuit loading. The use of a probe
will, however, limit the maximum trigger frequency range.
To obtain maximum trigger bandwidth when using probes,
select
a
probe
capable
of
compensating
the
input
capacitance ; observe the grounding considerations given in
the probe manual . The probe-to-connector adapters and the
bayonet-ground tip provide the best frequency response.
In high-frequency applications, requiring maximum over-
all bandwidth, use a coaxial cable terminated at both ends
in the characteristic impedance of the cable. To maintain
the high-frequency characteristics of the applied signal, use
high-quality
low-loss cable.
Resistive coaxial attenuators
can be used to minimize reflection if the applied signal has
suitable amplitude.
Operating Instructions-71311 Operators
High-level,
low-frequency
signals
can
be
connected
directly to the external trigger inputs with short, unshielded
leads.
When
this method
is
used,
establish
a common
ground between the 71311 and the associated equipment.
The common ground
provided by
line cords
is usually
inadequate .
If interference
is excessive with
unshielded
leads, use a coaxial cable or probe.
A signal
can also be routed to the 71311 through an
amplifier
unit
via the
internal
trigger
circuitry of the
mainframe (71311
installed in a horizontal compartment) .
This method of signal connection minimizes circuit loading,
especially when triggering a time-base unit in parallel with
the 71311 . (NOTE : Only external signals can be used with
the Events Start Trigger.)
The front-panel output signals, DLY'D TRIG OUT and
DLY INTERVAL OUT, should
be connected to
other
equipment with 50-ohm coaxial cables . The cables should
be terminated in 50 ohms to maintain the rise and falltime
characteristics of these signals.
Count Mode
General. Two basic count modes, Time and Events, can
be performed by the 71311,
as selected by the COUNT
MODE switch . The delay interval in both modes is selected
by the DELAY TIME OR EVENTS control and is read out
on the CRT.
TIME
INT
CLOCK. The
71311
counts
precise
increments of
time
after the receipt of
a
trigger . The
TRIGGER controls select and condition the signal to start
the time delay. Accuracy in this mode
is determined by an
internal, crystal-controlled oscillator .
TIME EXT 1 MHz. This count mode
is the same as
TIME
I NT CLOCK except the accuracy is derived from an
external, one-megahertz standard .
EVENTS . The
71311
counts
events,
periodic
or
aperiodic, at count rates to 50 megahertz. The EVENTS
START TRIGGER
provides
a means of discriminating
between the event that starts the delay and the events to be
counted.
The
events
to
be
counted are
selected and
conditioned by the TRIGGER controls.
REV. JAN 1974
Trigger Controls
The input signal may have a wide variety of shapes and
amplitudes, many of which are unsuitable as delay-initiating
triggers. For this reason, these signals are first applied to a
trigger
circuit where
they
are
converted
to
pulses of
uniform amplitude and shape. This makes
it possible to
start the delay with
a pulse that
has
a constant
size,
eliminating variations of the delay circuit operation caused
by changing input signals. The TRIGGER controls provide
a means
to
select
the
signal
source,
filter unwanted
frequencies, and start the delay at any voltage
level on
either slope of the waveform .
Triggered Light
The TRIG'D light provides a convenient indication of
the Trigger circuit condition. If the TRIGGER controls are
correctly set and an adequate signal is applied, the TRIG'D
light is on .
If the TRIG'D light is off, no delay interval is
started. The cause might be an incorrectly set TRIGGER
control, low signal
amplitude, or a signal repetition rate
outside the usable frequency range. This feature can be
used as a general indication of correct triggering when there
is no display on the CRT. The Delay Interval Pedestal and
Z-Axis Blanking
displays
also
aid
in
obtaining
correct
TRIGGER control
settings. See the discussion of these
features under Output Signals to Mainframe for further
information .
NOTE
When the 71311 is used in the EVENTS count mode,
the EVENTS START TRIGGER affects the output
of the
Trigger
circuit but has no
effect on the
TRIG'D light.
Trigger Coupling
The TRIGGER pushbuttons located below the COUP-
LING title select the method in which the input signal is
connected to
the Trigger circuit . Each
position permits
selection or rejection of various frequency components of
the signal used to trigger the delay start.
AC .
In this position of the COUPLING switch, the DC
component of
the input signal
is blocked. Signals with
low-frequency
components
below
about 30
hertz
are
attenuated .
In general, AC COUPLING
can be used for
most applications. However, if the signal contains unwanted
frequency components or if the delay is to be triggered at a
low repetition rate or DC level, one of the other switch
positions will provide better results .
The triggering point in the AC position depends upon
the average voltage level of the input signal .
If the input
signal occurs randomly, the average voltage level will vary,
causing the triggering point to vary also . This shift of the
triggering point may be enough
so
it
is
impossible to
maintain
a
stable
delay
start ;
in
such
cases,
use DC
coupling .
AC
LF
REJ. In
this
position, DC
is
rejected
and
low-frequency input signals below about 30 kilohertz are
attenuated . Therefore, the delay
is triggered only by the
higher-frequency components of the input signal . The AC
LF REJ position is particularly useful for providing stable
triggering
if
the
input
signal
contains
line-frequency
components.
AC HF REJ. This COUPLING switch position passes
all low-frequency signals between about 30 hertz and 50
kilohertz. DC is rejected and signals above 50 kilohertz are
attenuated . This position is useful to trigger the delay from
the low-frequency components of a complex waveform .
DC . The DC
position
can be used to
provide stable
triggering from low-frequency or low-repetition-rate signals
which would be attenuated in other modes. It can also be
used to trigger the delay when the input signal reaches a DC
level selected by the setting of the SLOPE/LEVEL control.
When triggering from the internal source, the setting of the
vertical
unit position control(s) affects the DC triggering
point.
Trigger Source
The TRIGGER pushbuttons located below the SOURCE
title select the source of the signal connected to the Trigger
circuit.
INT. In this position, the input signal
is derived from
the associated vertical unit . Therefore, the 71311 must be
installed in a horizontal compartment to use the internal
source .
Further
selection of the
internal signal may be
provided
by
the
vertical
unit and mainframe;
see the
instruction
manuals
for
these instruments
for further
information .
LINE .
In this SOURCE switch position, a sample of the
power-line voltage from the mainframe is connected to the
Trigger circuit. Line triggering
is useful when the input
signal
is time related (multiple or submultiple) to the line
frequency.
It is also useful for providing stable triggering
from a line-frequency component in a complex waveform .
(K
REV. JAN 1974
EXT. A
signal
connected
to
the
EXT TRIG
IN
connector can be
used to trigger the delay in the EXT
position of the SOURCE switch . An external signal can be
used to provide a trigger when the internal signal amplitude
is too low.
EXT =10. Operation
in
this
position
is the same as
described for EXT except the external signal is attenuated
10 times. Attenuation of high-amplitude signals is desirable
to extend the range of the LEVEL control .
Trigger Slope/Level
The TRIGGER SLOPE/LEVEL controls determine the
slope and voltage level of the input signal where the Trigger
circuit responds . Generally, the best point on a waveform
for triggering the delay
is where the slope
is steep, and
therefore usually free of noise . Assuming a sine-wave input
waveform, the steepest slope occurs at the zero-crossing
point . This
is the point selected for triggering when the
LEVEL control
is set to 0 (center) . A more positive or
negative point on the waveform is selected as the LEVEL
control
is rotated clockwise or counterclockwise
respec-
tively from 0 (towards + or - symbols on panel) .
Before
setting
the
TRIGGER
LEVEL,
the
desired
SLOPE, MODE, COUPLING,
and SOURCE should be
selected . Then adjust the LEVEL control so the delay starts
from the desired point.
Events Start Trigger
Operating Instructions-71311 Operators
The Events Start Trigger is used in the EVENTS count
mode to differentiate between the event that starts the
delay and the events being counted.
The EVENTS START TRIG IN connector provides the
input to the events-start
signal . The EVENTS START
TRIGGER SLOPE and LEVEL controls select the ampli-
tude point and slope on the input signal where the delay is
triggered.
Generally, the best point on a waveform for triggering
the delay is where the slope is steep, and therefore usually
free of noise, Assuming a sine-wave input waveform, the
steepest slope occurs at the zero-crossing point. This is the
point selected for triggering when the LEVEL control is set
to 0 (center) . A more positive or negative point on the
waveform
is selected
as
the LEVEL control
is rotated
clockwise or counterclockwise respectively from 0 (towards
+ or - symbols on panel) .
Operating Instructions-7D11 Operators
Delay Time or Events
The DELAY TIME OR EVENTS control selects the
digital delay time in the TIME count mode, and the number
of events counted in the EVENTS count mode . The delay
time
in milliseconds, or integer number of events, selected
is displayed on the CRT readout.
This
control
is a spring-return-to-center
control that
increases or decreases the count at which a delayed pulse
will occur. The direction of rotation determines whether
the count is increased or decreased . The rate at which the
count
increments
is
determined
by the
magnitude of
rotation . After either extreme of the range is reached, the
next count starts from the other end of the range . For
example, if the delay time is increased above 1000.0000 ms
(one second), the count will go to 0.0001 ms . Conversely, if
the delay time is decreased past 0.0001 ms, the count will
go to 1000.0000 ms.
Fine Delay
The FINE DELAY (ns) dial selects analog delay time
from zero to 100 nanoseconds in the TIME count mode .
This
one-turn
control provides added resolution to the
digital
delay
time
selected
by the DELAY TIME OR
EVENTS control . The delay time selected by the FINE
DELAY (ns) dial
is read from the calibrated knob as the
analog delay time
is not read out on the CRT. Each minor
division on the dial represents two nanoseconds.
B Sweep Delay Mode
The B SWEEP DELAY MODE switch permits the 7D11,
under
specific conditions, to select the delay mode of a
compatible time-base unit . To use this feature, the 7D11 is
installed in the A Horizontal compartment and the time-
base in the B Horizontal compartment of a four-plug-in
mainframe. With this arrangement, the time-base unit can
be controlled through the mainframe interface. Some dual
time-base units are not compatible with this feature ; see the
time-base unit instruction manual for further information .
INDEPENDENT. The 7D11
and
the time-base
unit
operate independently .
B STARTS AFTER DELAY. The time-base unit pro-
duces a sweep immediately following the selected delay
interval . This
provides
the same mode of
operation as
triggering
the time-base
unit
with
the
delayed
trigger
output .
B TRIGGERABLE AFTER DELAY. The time-base unit
produces a sweep after the first trigger pulse
is received
following the selected delay interval . This mode of opera-
tion provides a stable display of a signal having time jitter .
Precision time measurements cannot be made in this mode
because the actual delay time is only partially dependent on
the delay interval of the 7D11 .
OUTPUT SIGNALS
Front-Panel Output Signals
General . The Delay
Interval and Delayed Trigger out-
puts are available at the front-panel DLY INTERVAL OUT
and
DLY'D TRIG OUT connectors
respectively . These
outputs can be used to control other equipment during or
immediately following the delay interval . To maintain the
rise and falltime characteristics of these signals, connection
to other equipment should be made with 50-ohm coaxial
cable ; the output of the cable should be terminated in 50
ohms .
DLY INTERVAL OUT. This output is a positive-going,
rectangular waveform coincident with the generated delay
interval .
In time mode DLY INTERVAL OUT is approxi-
mately 20-30 ns shorter than indicated delay time because
of internal propagation delays and trigger recognition time .
In the event mode, the DLY INTERVAL OUT is within
30 ns of actual delay, usually C 10 ns .
DLY'D TRIG
OUT.
This
signal
is
generated
as
a
positive-going rectangular pulse coincident with the end of
the delay interval .
The front-panel output signals are shown
in
Fig
1-3,
along with the input signal . The input signal, Fig. 1-3(A), is
comprised of one- and ten-microsecond time markers. The
7D11
is
set
for
a 0.0038-millisecond delay time
after
triggering on the ten-microsecond markers. The resultant
Delay Interval and Delayed Trigger outputs are shown in
Fig.
1-3(B)
and
(C), respectively . The characteristics of
these output signals are given in the Specification section .
Fig.
1-3. Display showing time relationship of :
(A) Input signal
to the front-panel,
(B) Delay Interval, and (C) Delayed Trigger
outputs.
REV B, SEP 1978
Output Signals to Mainframe
General. Signal outputs are provided to the mainframe
via
the
interface
connector.
The
following
discussion
describes these signals and the operating conditions neces-
sary for their use .
Delay-Interval
Pedestal . This output
provides an
on-
screen display of the approximate delay interval . To view
the pedestal
display, the 71311
must
be
installed
in a
vertical
plug-in
compartment
and
be
selected
by
the
mainframe
Vertical
Mode
switch . The position of this
display is fixed near the vertical center of the graticule area .
The Delay-Interval Pedestal display is shown in Fig. 1-4 (A) .
The input signal, shown in Fig. 1-4 (B), is comprised of one-
and ten-microsecond time markers. The 71311
is set to
trigger on the ten-microsecond markers, and to generate a
0.0038-millisecond delay time .
Delayed Trigger. The Delayed Trigger output provides an
internal
Delayed Trigger source for a time-base unit . A
time-base
unit
can
be triggered from the Delay Trigger
when the 71311
is in a vertical compartment. To use this
output, the 71311
must
be selected by the appropriate
trigger source switch (mainframe) .
Z-Axis Blanking . Z-axis blanking provides an on-screen
display
of
the
approximate
delay
interval .
This
is
ac-
complished
by blanking out the CRT display during the
delay
interval . Z-axis blanking can be obtained with the
71311
installed
in any plug-in compartment . The Z-axis
blanking display is selected by a slide switch located inside
the unit (on the left side ; see Fig. 1-6) .
Fig.
1-4. Waveform display of : (A) Delay Interval
Pedestal ;
(B)
input signal .
REV. JAN 1974
Operating Instructions-71311 Operators
A. 71311 triggered incorrectly.
B.7D11 triggered correctly .
Fig.
1-5. Use of Delay-Interval
Pedestal display to indicate :
(A)
Incorrect triggering, and; (B) correct triggering .
Fig. 1-6. Location of Z-Axis Blanking Display switch (on left side
of unit). The switch position towards the rear of the unit selects
Z-axis blanking .
Operating Instructions-7D11 Operators
Sweep Delay
NOTE
At faster sweep speeds (100 nsldiv or faster) care
must be taken when interpreting CRT display because
relative propagation
delays through the 7D11 and
vertical amplifier plug-ins are not the same.
This
appears as a relative time shift between delay interval
pedestal or Z-axis blanking generated by the 7D 11
and the signal(s) viewed through a vertical amplifier
on the CRT. Changing the TRIG SOURCE between
INT and EXT or EXT - 10 will vary this apparent
time shift due to differences in propagation delays of
the signal path .
OPERATING MODES
The 7D11 can be used to delay the start of a sweep for a
selected time interval
following the receipt of a trigger .
Low-jitter sweep delay can be used for accurate time, jitter,
and stability measurements . Sweep delay can also be used
to select a portion of a complex signal for display. A sweep
is delayed by triggering the sweep from the Delayed Trigger
output of the 7D11, rather than from the signal to be
displayed . Several methods of coupling the Delayed Trigger
to the sweep are possible, depending on the application .
These methods are described in the following discussions .
B Sweep Delay Mode Switch . The sweep produced by a
time-base unit can be controlled and delayed by a 7D11 via
the mainframe interface and the B SWEEP DELAY MODE
switch . To use this mode of sweep delay, the 7D11 must be
installed in the A Horizontal compartment and the time-
base unit in the B Horizontal compartment of a four-plug-in
mainframe. For further information, see B Sweep Delay
Mode .
NOTE
The
logic
levels provided to
the 7D11 from
the
mainframe are designed to control a time-base unit
delaying sweep.
For this reason, the 7D11 might
become locked out (no output) when the setting of
either the B-Sweep unit Time/Division switch or the
B SWEEP DELAY MODE switch is changed. If this
occurs, a delayed sweep will not be produced. To
reset the 7D11, set the B SWEEP DELAY MODE
switch
first
to INDEPENDENT,
then
select
the
desired delay mode.
REV. JAN 1974
Internal Trigger. The sweep produced by a time-base
unit
in
a
horizontal
compartment
can
be
internally
triggered from a 7D11 in a vertical compartment. To use
this sweep delay mode, the 7D11 must be selected by the
mainframe
trigger
source
switch .
Delaying
a time-base
sweep from the internal source can be used with the units
installed in either a three- or four-plug-in mainframe.
External Trigger Source . A sweep can be delayed by
external triggering from the DLY'D TRIG OUT connector .
This method can be used with any triggered sweep.
Echo Delay Time Mode
The Echo delay time mode provides a CRT readout
of the "one-way-trip" time, or one-half of the generated
delay time . This mode of operation is selected for use by an
internal switch (on left side of unit, see Fig. 1-7) . In the
Echo mode, the delay time is selected by the DELAY TIME
OR EVENTS control
in 200-nanosecond increments . An
insertion delay of about 160 nanoseconds in this mode
requires
adding
analog
delay time
to
the
first delay
increment to obtain a 200-nanosecond delay interval . This
can be accomplished by displaying the DELAY INTERVAL
OUT and setting the FINE DELAY dial for a total delay
interval of 200 nanoseconds as measured on the graticule.
Fig. 1-7. Location of Normal-Echo Delay Time Mode switch (on
left side of unit) . Set the switch towards the front of the unit to
select Normal Mode .
The following information describes the procedures and
techniques for making measurements using a 71311 Digital
Delay unit . These applications are not described in detail,
since each application must be adapted to the requirements
of the individual measurement. This instrument can also be
used for many applications which are not described in this
manual . Contact your local TEKTRONIX Field Office or
representative
for assistance in making specific measure-
ments with this instrument .
Delayed-Sweep Time Measurements
The 71311 can be used to delay the start of a sweep to
make
accurate
time measurements . The following pro-
cedure
describes one method of determining
the time
difference between two pulses displayed on the same trace.
This
application
can
also
be
used
to
measure
time
difference between two signal sources or to measure the
time duration of a single pulse . Other methods of delayed-
sweep time measurements
with the 71311
are described
under General Operating Information .
1 . Within the 71311, set the Z-Axis switch to Z to select
the Z-Axis Blanking display.
2. Install the 71311, time-base unit, and amplifier unit in
the A Horizontal, B Horizontal, and vertical compartments
respectively of a four-plug-in mainframe.
3. Connect the signal to be measured to the amplifier
unit.
4. Set both the 71311 and time-base unit to trigger from
the signal source .
5. Set the 71311 for TIME INT CLOCK operation in the
INDEPENDENT mode .
6 . Set the time-base unit sweep rate to view several
cycles of the applied signal .
7. Set the 71311 TRIGGER controls to trigger the delay
start. The TRIG'D indicator will
light when a trigger is
being produced .
APPLICATIONS
Operating Instructions-71311 Operators
8. Turn the DELAY TIME OR EVENTS control clock-
wise
to produce a blanked display. Set this control to
position the blanked portion to the left of the first pulse .
9. Set the 71311 B SWEEP DELAY MODE switch to B
STARTS AFTER DELAY.
10 . Set the time-base unit for a sweep rate 100 times
faster than its original sweep rate .
11 . Set the DELAY TIME OR EVENTS and FINE
DELAY controls
to move the first pulse to the center
vertical line . Note the total delay time read out on the CRT
and the FINE DELAY dial .
12 . Set the DELAY TIME OR EVENTS and FINE
DELAY controls to position the second pulse to the same
point. Again note the total delay time .
13 . Subtract
the
first delay time
reading from
the
second . This gives the time interval between pulses .
Displaying Complex Signals With Delayed Sweep
Complex signals often consist of a number of individual
events of differing amplitudes . Since the time-base trigger
circuits are sensitive only to changes in signal amplitude, a
stable display can normally be obtained only when the
sweep
is
triggered by
the event(s)
having
the
greatest
amplitude.
However,
this may not produce the desired
display of a lower amplitude event which follows the
triggering event. The 71311 provides a means of delaying the
start of the sweep by a selected amount following the event
which triggers the 71311 . Then, the part of the waveform
which
contains
the
information
of
interest
can
be
displayed . The following procedure describes one method
of using the 71311 to display complex signals .
1 . Within the 71311, set the Z-Axis switch to Z to select
the Z-Axis Blanking display.
2. Install the 71311, time-base unit, and amplifier unit in
the A Horizontal, B Horizontal, and vertical compartments
respectively of a four-plug-in mainframe.
3. Connect the signal to be measured to the amplifier
unit .
Operating Instructions-71311 Operators
4 . Set the 71311 and the time-base unit to trigger from
the signal source .
5
.
Set the 71311 for TIME INT CLOCK operation in the
INDEPENDENT mode .
Fig. 1-8.
(A) The circled portion of this display cannot be viewed
at
a faster sweep rate ;
(B) Display blanked during delay interval ;
(C) Area of interest displayed at faster sweep rate .
1-10
6 . Set the time-base unit
to view several cycles of the
applied signal .
7 . Set the 71311 TRIGGER controls to trigger the delay
start . The TRIG'D
indicator
will
light when
a trigger
is
being produced .
8 . Turn the DELAY TIME OR EVENTS control clock-
wise to blank that portion of the display preceding the area
of interest . The blanked portion of the display represents
the delay interval .
9 . Set
the
B
SWEEP
DELAY MODE
switch
to
B
STARTS AFTER DELAY . Set the time-base to a sweep
rate which displays the area of interest .
Example . Fig .
1-8(A)
shows
a
complex waveform
as
displayed on the CRT . The circled portion of the waveform
cannot be viewed in any greater detail because the sweep is
triggered by the larger amplitude pulses at the start of the
display ; a faster sweep rate moves this area of the waveform
off the viewing area . Fig . 1-8(B) shows the display blanked
out during the delay interval (delay interval is area between
the arrows) . Fig . 1-8(C) shows the sweep triggered after the
delay at a faster sweep rate to display the area of interest .
Pulse Jitter Measurements
In
some
applications,
it
is necessary
to
measure the
amount of jitter on the leading edge
of a pulse, or jitter
between pulses .
Use the following procedure :
1 .
Install an amplifier unit and the 71311
in the vertical
compartments of a mainframe . Install a time-base unit in a
horizontal compartment .
2 .
Set the mainframe
for a dual-trace vertical display
mode .
3 .
Connect the signal to the amplifier unit and to the
71311 EXTTRIG IN connector .
4 . Set the time-base unit to trigger on the applied signal
at a sweep rate which displays the complete waveform .
A
5 . Set the 7D11 for TIME INT CLOCK count mode . Set
the TRIGGER
SOURCE
to
EXT .
Set
the
TRIGGER
LEVEL control for a triggered condition
as indicated by
the TRIG'D indicator light .
6 .
Set the DELAY TIME OR EVENTS control to select
a delay
interval which
ends
just before
the pulse to be
measured .
7 . Connect the
DLY'D TRIG OUT to the time-base
external trigger input . Set the time-base to trigger from the
external source
.
8 . Set the time-base unit for a sweep rate which displays
just the pulse or leading edge to be measured .
9 .
Pulse jitter is shown by horizontal movement of the
pulse (take into account inherent jitter of 7D11) . Measure
the amount of horizontal movement directly from the CRT
graticule .
Operating Instructions-7D11 Operators
Example . Assume
the
horizontal
movement
is
0.2
division at a 200-nanosecond sweep rate (see Fig . 1-9) . The
amount of pulse jitter is then 0.2 X 200 nanoseconds = 40
nanoseconds .
Fig . 1-9 . Measuring pulse jitter .
DOrr
41 1 V2
&
I
~
I
.
I
t
~
I
Jitter
The electrical characteristics listed in Table 2-1 are valid over the stated environmental range for instruments calibrated at
an ambient temperature of +20°C to +30°C, and after a 20-minute warmup unless otherwise stated .
TIME DELAY
Digital Delay-Selected by front-panel DELAY TIME OR EVENTS
control .
Range
is
100
nanoseconds
to
one
second
in
100-nanosecond increments in Normal mode ; 200 nanoseconds
to two seconds in 200-nanosecond increments in Echo mode .
Normal/Echo modes are selected by internal Mode switch .
Analog Delay-Selected by front-panel FINE DELAY (ns) control .
Range
is continuously
variable from
zero
to
at
least 100
nanoseconds; accuracy
is within two nanoseconds of indicated
delay time; resolution is one nanosecond or less .
Jitter with Internal Clock-2 .2 nanoseconds or (delay time X 10-7 ),
whichever is greater .
Insertion Delay in Normal Mode-Zero within two nanoseconds.
Insertion
Delay
in
Echo
Mode-Less
than
160
nanoseconds,
adjustable to 200 nanoseconds within four nanoseconds with
FINE DELAY dial .
Recycle Time-Not greater than 575 nanoseconds.
Time Base-500-megahertz oscillator, phase locked to internal or
external clock, as selected by COUNT MODE switch . Internal
clock (TIME INT CLOCK) is a five-megahertz crystal oscillator .
Accuracy is within 0.5 part per million from 0°C to +50° C ;
long-term drift is one part or less in 107 per month .
Oi7
External clock (TIME EXT
1 MHz) requires a one-megahertz
signal applied to EXT 1 MHz IN connector . Input requirements
are : Frequency, one megahertz within 1% to insure phase lock of
500-megahertz oscillator; amplitude, at least 400 millivolts RMS
sinewave
or
pulse with 10% to 90% duty cycle . AC input
coupling . Input impedance
is 50 ohms t2% . Maximum input
voltage
is 5 volts RMS (DC + peak AC must not exceed 500
volts) .
TRIGGERING
Source-Selected by front-panel SOURCE switch .
INT-Internal from associated
vertical unit (71311
installed in
horizontal compartment) .
LINE-Internal from power source (AC) .
SPECIFICATION
TABLE 2-1
ELECTRICAL
1 Triggering frequency range is also affected by the bandwidth limits of the vertical plug-in unit used .
REV . JAN 1974
TRIGGERING (cont)
Section 2-71311 Operators
EXT-External from front-panel EXT TRIG IN connector .
EXT
-
10-External
divide
by
ten
(10X
input
signal
attenuation) .
Coupling-Selected by front-panel COUPLING switch .
AC-Capacitive coupling .
AC LF REJ-Capacitive coupling, low-frequency reject .
AC HF REJ-Capacitive coupling, high-frequency reject .
DC-Direct coupling .
Polarity-Selected
by
front-panel
TRIGGER
LEVEL/SLOPE
control . Time delay can be triggered from positive-going or
negative-going portion of trigger signal .
Internal Trigger Sensitivity )
AC-0 .3
division
of
deflection,
minimum,
30 hertz to
10
megahertz ; increasing to 1 .0 division at 50 megahertz .
AC LF REJ-0.3 division of deflection, minimum, 30 kilohertz
to 10 megahertz ; increasing to 1 .0 division at 50 megahertz . Will
not trigger on sinewaves of three divisions or less below 120
hertz .
AC HF REJ-0.3 division of deflection, minimum, 30 hertz to
50 kilohertz .
DC-0.3 division of deflection, minimum, DC to 10 megahertz;
increasing to 1 .0 division at 50 megahertz .
Slope-0.5
volt/nanosecond or
greater
for
specified
jitter and
accuracy ; slower risetime signals will trigger the instrument with
reduced accuracy and increased jitter .
External Trigger Sensitivity (SOURCE switch set to EXT ; triggering
signal requirements increased 10 times for EXT - 10 position) .
AC-150
millivolts,
minimum,
30
hertz to
10 megahertz;
increasing to 500 millivolts at 50 megahertz .
AC LF REJ-150
millivolts, minimum, 150 kilohertz to 10
megahertz ; increasing to 500 millivolts at 50 megahertz .
AC HF REJ-150 millivolts, minimum, 30 hertz to 50 kilohertz .
DC-150 millivolts, minimum, DC to 10 megahertz ; increasing to
500 millivolts at 50 megahertz .
External Trigger Input
EVENTS DELAY
Delay Range-One to 107 events.
Delay Increment-One event .
Insertion Delay-35 nanoseconds t five nanoseconds.
Recycle Time-500 nanoseconds or less .
Maximum Events Frequency-50 megahertz or greater .
Events Start Triggering
2-2
TRIGGERING (cont)
Maximum Input Voltage-500 volts (DC + peak AC), 500 volts
peak-to-peak AC at 1 kHz or less .
Input Resistance-One megohm within 2% .
Input Capacitance-20 picofarads within two picofarads.
LEVEL Range-At least +1 .75 volts to -1.75 volts in EXT, at
least +17.5 volts to -17 .5 volts in EXT - 10 .
Source-External
only
from
EVENTS
START
TRIG
IN
connector .
Coupling-DC .
Maximum Input Voltage-150 volts (DC + peak AC) .
Sensitivity-40 millivolts, minimum, 30 hertz to four megahertz ;
increasing to 100 millivolts, four megahertz to 20 megahertz ;
increasing to 250 millivolts, 20 megahertz to 50 megahertz .
Input Resistance-Approximately one megohm
TABLE 2-2
ENVIRONMENTAL CHARACTERISTICS
Refer to the Specification for the associated oscilloscope .
TABLE 2-1 (cont)
REV . JAN 1974
EVENTS DELAY (cont)
Input Capacitance-Approximately 20 picofarads
LEVEL Flange-At least + three volts to - three volts.
OUTPUT SIGNALS
Delayed
Trigger-Positive-going,
rectangular
pulse
available
at
front-panel DLY'D TRIG OUT connector .
Amplitude-Two volts or greater into open circuit; one volt or
greater into 50-ohm load .
Risetime with 50-Ohm Load-Two nanoseconds or less.
Falltime with 50-Ohm Load-Five nanoseconds or less .
Pulse Width-200 to 250 nanoseconds .
Delay
Interval-Positive-going, rectangular waveform
available
at
front-panel DLY INTERVAL OUT connector .
Amplitude-Two volts or greater into open circuit ; one volt or
greater into 50-ohm load .
Rise and Falltime with 50-Ohm Load-Five nanoseconds or less.
Accuracy-Delay interval pulse width is equal to the generated
Events Delay within 30 nanoseconds and equal to the generated
Time Delay less 20 to 30 nanoseconds .
Relative Timing of Delayed Trigger and Delay Interval Output
Signals-Leading edge of Delayed Trigger pulse is coincident with
falling edge of Delay Interval pulse within two nanoseconds .
Pedestal Display-CRT analog display of the delay interval, available
when 71311
is used in vertical compartment . (Pedestal Display
pulse ends coincident with Delayed Trigger start .)
De-Intensified
Display-Signal
to dim CRT display during delay
interval . (De-Intensified Display selected by internal switch .)
TABLE 2-3
PHYSICAL
Size
�
Fits all 7000-series plug-in compartments.
Weight
�
2 Pounds 11 Ounces (1219 grams)