Pyxis SP-910 Operation Manual
Text of the manual
SP-910 Portable Water Analyzer
Operation Manual
Rev. A
Firmware version 1.0.170
Pyxis Lab, Inc 2016
All Rights Reserved
1 General Description....................................................................................... 6
1.1 Pyxis Major Features...............................................................................6
1.2 Unpackaging the Instrument.................................................................. 7
1.3 Standard Accessories..............................................................................8
1.4 Optional Accessories...............................................................................8
1.5 Sample Vial Compartment......................................................................8
1.6 Light Shield Cover................................................................................... 8
2 Start Pyxis SP-910.........................................................................................10
2.1 Battery Installation............................................................................... 10
2.2 Description of the Navigational Control Pad........................................10
2.3 Turning On Pyxis SP-910....................................................................... 11
2.4 Main Page............................................................................................. 11
2.5 Turning Off Pyxis SP-910.......................................................................12
2.6 Pyxis SP-910 Auto Power Off................................................................ 12
2.7 Auto LCD Power Saving.........................................................................12
3 PTSA Measurement..................................................................................... 13
3.1 PTSA Measurement.............................................................................. 13
3.2 PTSA calibration....................................................................................13
4 Fluorescein Measurement...........................................................................14
4.1 Fluorescein Measurement............................................................... 14
4.2 Fluorescein calibration......................................................................15
5 Colorimetric Measurement......................................................................... 17
5.1 Supported Methods..............................................................................17
5.2 Select a Method....................................................................................19
5.3 Single Timing Step Method...................................................................19
5.4 Single-Vial Procedure............................................................................20
5.5 Two-Vial Methods.................................................................................20
5.6 Multiple Timing Steps........................................................................... 21
5.7 Advanced Methods...............................................................................22
5.8 Method Setup and Calibration............................................................. 23
6 Turbidity Measurement............................................................................ 25
6.1 Operation..............................................................................................25
6.2 Turbidity Calibration............................................................................. 25
7 Absorbance Measurement...................................................................... 27
8 Bluetooth Interface................................................................................... 28
8.1 Install Software.....................................................................................28
8.2 Open SP-910 Bluetooth Interface.........................................................28
8.3 Scan and Connect................................................................................. 28
8.4 Device Configuration............................................................................ 29
Setup Device Name.................................................................................29
Setup Product......................................................................................... 29
Add User Defined Colorimetric Methods............................................... 29
9 Maintenance.............................................................................................. 30
10 Troubleshooting...................................................................................... 32
11 Appendix A.............................................................................................. 33
Figure 1 Insert the Sample Vial.........................................................................7
Figure 2 Open and Close the Light Shield Cover...............................................9
Figure 3 Replace Batteries.............................................................................. 10
Figure 4 Main Menu........................................................................................11
Figure 5 PTSA Measurement.......................................................................... 13
Figure 6 PTSA Calibration................................................................................14
Figure 7 Fluorescein Measurement................................................................15
Figure 8 Fluorescein Calibration..................................................................... 15
Figure 9 Method Selection..............................................................................19
Figure 10 Single Timer Method...................................................................... 20
Figure 11 Concentration as a Function of Time..............................................20
Figure 12 Multiple Timer Method.................................................................. 22
Figure 13 Method Form Selection..................................................................24
Figure 14 Method Unit Selection....................................................................24
Figure 15 Slope Calibration.............................................................................24
Figure 16 Turbidity Calibration-1....................................................................26
Figure 17 Turbidity Calibration-2....................................................................26
Figure 18 Absorbance Measurement............................................................. 27
Figure 19 Nebula Scans Bluetooth Device......................................................28
Figure 20 Change Device Name......................................................................29
Tables
Table 1 Feature Groups on Main Menu................................................................11
Table 2 List of Supported Colorimetric Methods..................................................17
Table 3 Wavelength of each channel....................................................................27
Table 4 Contact Information................................................................................. 32
Trademarks and Patents
Hach ® is a registered trademark of the Hach Company, Loveland, CO USA
Confidentiality
The information contained in this manual may be confidential and proprietary and is the property of
Pyxis Lab. Information disclosed herein shall not be used to manufacture, construct, orotherwise
reproduce the goods disclosed herein. The information disclosed herein shall not be disclosed
toothers or made public in any manner without the express written consent of Pyxis Lab.
Standard Limited Warranty
PyxisLab warrants its products for defects in materials and workmanship. PyxisLab will, at its option,
repair or replace instrument components that prove to be defective with new or remanufactured
components (i.e., equivalent to new). The warranty set forth isexclusive and no other warranty,
whether written or oral, is expressed or implied.
Warranty Term
The Pyxis warranty term is thirteen (13) months ex-works. In no event shall the standard limited
warranty coverage extend beyond thirteen (13) months from original shipment date.
Warranty Service
Damaged or dysfunctional instruments may be returned to Pyxis for repair or replacement. In some
instances, replacement instruments may be available for short duration loan or lease.
Pyxis warrants that any labor services provided shall conform to the reasonable standards of
technical competency and performance effective at the time of delivery. All service interventions are
to be reviewed and authorized as correct and complete at the completion of the service by a
customer representative, or designate. Pyxis warrants these services for 30 days after the
authorization and will correct any qualifying deficiency in labor provided that the labor service
deficiency is exactly related to the originating event. No other remedy, other than the provision of
labor services, may be applicable.
Repair components (parts and materials), but not consumables, provided in the course of a repair, or
purchased individually, are warranted for 90 days ex-works for materials and workmanship. In no
event will the incorporation of a warranted repair component into an instrument extend the whole
instrument’s warranty beyond its original term.
Shipping
A Repair Authorization Number (RA) must be obtained from the Technical Support (service@pyxis-
lab.com) before any product can be returned to the factory. Pyxis will pay freight chargesto ship
replacement or repaired products to the customer. The customer shall pay freight charges for
returning products to Pyxis. Any product returned to the factory without an RA number will be
returned to the customer.
1 General Description
Specification
Colorimeter Wavelength:
365, 420, 455, 525, 560, 570, and
630nm
Turbidity Excitation Wavelength:
White LED Infrared LED/90 degree
scattering
Fluorescence Excitation Wavelength:
365 nm LED 460nmLED
Fluorescence Emission Wavelength:
410 nm 520nm
Wavelength Accuracy:
±1 nm
Absorbance Reproducibility:
0.005 au in the range of 0 to 1.0 au
(3sigma)
Absorbance Linearity Range:
0 to 1.0 au
PTSA Fluorescence Reproducibility:
1 ppb PTSA (3 sigma)
PTSA Fluorescence Detection Limit:
1 ppb
Fluorescein Fluorescence Reproducibility:0.1ppb
Fluorescein Fluorescence Detection Limit:0.1ppb
Turbidity Reproducibility:
1 NTU (3 sigma)
Turbidity Detection Limit:
1 NTU
Battery:
4 AA alkaline
Typical Battery Life:
2 months
Display:
Graphical LCD 192x128 pixels,
visible under direct sunlight
Instrument Dimension:
L 265mm W 88mm H 62mm
Instrument Weight:
600 g without batteries
Storage Temperature Range:
0 to 140°F (-18 to 60°C)
Operation Temperature Range:
40 to 120 °F (4 to 49°C)
Humidity:
85% at 106 °F (41 °C)
Environmental:
IP67, dustproof and waterproof
Note:
1.
Specifications are subject to change without notice with Pyxis’ continuous
development.
1.1 Pyxis Major Features
The Pyxis SP-910 analyzer shown in Figure 1 is a combination of photometer and
fluorometer. It provides colorimetric measurements at 6 LED wavelengths, fluorometric
measurement of fluorescent tracer PTSA and fluorescein, and nephelometric turbidity
measurement using white LED and infrared LED as the excitation sources. Pyxis SP-
910 is pre-calibrated for colorimetric measurements of analyses common in industrial
water treatment and other water testing in the laboratory or in the field, such as
chlorine, phosphate, iron, and copper. Main features include:
Pyxis SP-910 is pre-calibrated for measuring PTSA (pyrenetetrasulfonic acid) in
the range of 0 to 300 ppb. The fluorescence PTSA measurement is automatically
compensated for sample color and turbidity interference.
Pyxis SP-910 is pre-calibrated for measuring fluorescein in the range of 0 to 20
ppb.
Pyxis SP-910 is pre-calibrated for measuring turbidity in the range of 0 to 500 NTU.
Automatically select the primary wavelength according to the method selected and
switches to the secondary wavelength to extend the primary measurement range.
Navigational Control Pad
LCD
Sample Vial Compartment
Light Shield Cover
Sample Vial
Figure 1 Insert the Sample Vial
Display a concentration-time profile curve during the last time period in a
colorimetric measurement. The user can terminate the timing process and take a
reading if the displayed concentration reaches a plateau before completing the
predefined time period.
The user can update the calibration parameter of any pre-calibrated colorimetric
method by testing a standard solution first and then following a setup procedure to
update the calibration parameters.
1.2 Unpackaging the Instrument
Remove the instrument and accessories from the shipping container and inspect each
item for any damage that may have occurred during shipping. Verify that all items
listed on the packing slip are included. If any items are missing or damaged, please
contact Pyxis Customer Service at [email protected].
1.3 Standard Accessories
Sample Vials – two 10 ml, round, 0.78 inch (20 mm) pathlength, glass vials,
which can be used for all measurements including turbidity and fluorescence
measurements.
16 mm adapter
4 AA alkaline batteries
Instrument Manual, also available from www.pyxis-lab.com
1.4 Optional Accessories
Silica desiccant gel pack
100 ppb PTSA standard in a 500 ml brown plastic bottle
20 ppb fluorescein standard in a 500 ml brown plastic bottle
1.5 Sample Vial Compartment
The sample vial compartment is shown in Figure 1, along with a 10-ml sample vial.
When the sample vial is inserted into the sample vial compartment, the triangular mark
on the sample vial should be aligned approximately with the 6 o’clock position of the
sample vial compartment or any position consistently.
The sample vial compartment should be kept clean. A small amount foreign material
could significantly affect turbidity and fluorescence measurement results. Use a soft
cloth or lint free paper tissue to clean sample vial compartment periodically. Remove
debris, scale, and deposit promptly.
1.6 Light Shield Cover
The light shield cover is shown in Figure 2. The light shield cover can be conveniently
slid between the open and closed positions. The light shield cover is held firmly at the
rest positions by permanent magnets.
The light shield cover should be in the closed position during storage, transportation,
and measurements, especially during the turbidity and fluorescence measurements.
When turned on, Pyxis SP-910 carries out self-diagnosis including checking the
performance of a variety of optical devices. The light shield door shall be at the closed
position to shield interference from ambient light during self-diagnosis.
Care should be taken to avoid water or debris being trapped in the track of the light
shield door.
Figure 2 Open and Close the Light Shield Cover
Warning
Magnetic sensitive devices, including but not limited to, credit cards, watched, hard
disks, should be keep at a distance of at least 2 inches from the Light Shield Door to
avoid possible damage and/or loss of information recorded.
2 Start Pyxis SP-910
2.1 Battery Installation
Pyxis SP-910 is powered by four AA-size alkaline batteries. Do not use rechargeable
nickel cadmium (NiCad) batteries or any AA-size lithium batteries. A set of batteries
typically lasts for two months. When the batteries capacity is low, Pyxis SP-910 will
prompt a LOW BATTERY warning. Replace all four batteries to resume operation of
Pyxis SP-910 after the battery warning.
The Pyxis SP-910 battery compartment, shown in Figure 3, is on the back side of the
instrument. Insert a small pad underneath the screen area to make the back surface
level when the instrument is turned upside down. Install batteries as followings:
1.
Remove the battery compartment cover by loosening four screws.
2.
Insert four batteries into the battery holder as shown in Figure 3. Make sure the
positive battery polarity marker (+) is aligned with the positive marker (+) on the
battery holder.
3.
Replace the battery compartment cover, making sure that the sealing O-ring is
lying flat on the battery holder and tighten the four screws.
Figure 3 Replace Batteries
2.2 Description of the Navigational Control Pad
The Pyxis SP-910 navigational control pad consists of five keys as shown in Figure 1.
The left, right, up, and down keys are navigational keys that are used to select an icon,
a button, or other items in various pages. The center key is the OK key. Press the OK
key on a selected item to launch the action associated with the selected item. The OK
key is also used to accept the current selection, like the return key in a computer
keyboard.
2.3 Turning On Pyxis SP-910
To turn on SP-910, press and hold on the OK key for 3 seconds, and release the OK
key when the LCD is lit.
After new batteries installation,Pyxis SP-910 will be automatically turned on, and its
LCD backlight will be on. You can navigate the main page menu and launch an
operation by pressing on an icon. If battery voltage is too low for the instrument to work
properly, Pyxis SP-910will show a low battery warning message for 5 seconds and
turn off automatically. If this happens, replace all four batteries.
2.4 Main Page
Pyxis SP-910provides intuitive icon menu-assisted user operations. On the main page,
eight major feature groups are illustrated as below:
Figure 4 Main Menu
A brief description of each feature group is given in Table 1.Detailed operation instructions
can be found in the following chapters.
Table 1 Feature Groups on Main Menu
No.
Title
Description
1
M-P
PTSA measurement
2
M-F
Fluorescein measurement
3
COLOR
Colorimetric measurement methods
4
TURB
Turbidity measurement
5
Power
Turn off SP-910
6
ABSB
Absorbance measurements
7
CAL
Calibration routines
8
SYS
System information
2.5 Turning Off Pyxis SP-910
Turn Pyxis SP-910 off by navigating to Power icon and press the OK key.
2.6 Pyxis SP-910 Auto Power Off
Pyxis SP-910 automatically turns itself off with no-key activity, except for during a
measurement.The auto power off time can be set. Pressing and holding the OK key for
3 seconds will wake up the instrument, and return to the original page if it has any
measurement data.
2.7 Auto LCD Power Saving
During a colorimetric method measurement, Pyxis SP-910 automatically turns LCD
backlight off with no-key activity. The auto LCD power time can be set.Pressing any
key will turn on the LCD backlight. Under normal ambient lighting condition, icons and
other contents shown on Pyxis SP-910LCD are readable without backlight being on.
3 PTSA Measurement
3.1 PTSA Measurement
1.
Fill the 10 ml sample vial with the test solution and tightly cap the sample vial.
2.
Place the sample vial into the sample vial compartment and slide the light shield
cover to the closed position.
3.
Press the M-P on the main page, then press the ok button,Pyxis SP-910 will start
to measure the PTSA concentration in the sample.
4.
PyxisSP-910 will display the PTSA concentration in ppb as PTSA.
Figure 5 PTSA Measurement
During the fluorescence measurement to determine the PTSA concentration, Pyxis
SP-910 checks the sample turbidity. If the sample turbidity value detected is greater
than 40 NTU, Pyxis SP-910 will display a warning.For best results, the sample should
be filtered if turbidity exceeds 40 NTU.
Sample color causes a lower PTSA concentration to be measured. Pyxis SP-910
automatically compensates for sample color. If the sample color is too intense, Pyxis
SP-910 will display a warning.
For best results, ensure that the sample vial is clean. Wipe off water on the outside
wall of the sample vial using a lint-free tissue paper. Fill the sample vial to the 10 ml
mark. If the sample contains air bubbles, tap the sample vial gently to remove the
bubbles before placing the sample vial to sample vial compartment.
3.2 PTSA calibration
Deionized water (DI) as the blank calibration solution and the 100 ppb PTSA calibration
standard solution are needed.
1.
Press the CAL on the main page, then choose the M-P and press the OK key to
launch the PTSA calibration page.
2.
Follow the message prompts, insert the DI blank into the sample vial compartment
and press the OK key.
Follow the message prompts, and insert the 100 ppb standard into the sample vial
compartment and press the OK key.
4.
Press the OK key to return to the main page.
Figure 6 PTSA Calibration
If calibration fails, the followings should be checked:
The DI blank is being contaminated.
The 100 ppb PTSA standard solution is decayed or being contaminated.
The light shield cover is not in the closing position.
The sample vial compartment is blocked with debris, water, or other materials.
The 100 ppb standard solution shall be stored in a brown or black opaque bottle.
Exposing the PTSA standard to light will cause the standard losing the PTSA
concentration. Many substances, such as quaternary amine cause a negative
interference. Many other substances such laundry detergents that contain optical
brightener will cause a significant positive interference.
4 Fluorescein Measurement
4.1 Fluorescein Measurement
1.
Fill the 10 ml sample vial with the test solution and tightly cap the sample vial.
2.
Place the sample vial into the sample vial compartment and slide the light
shieldcoverto the closed position.
3.
Press the M-F on the main page, then press the ok button,Pyxis SP-910 will start
to measure the fluorescein concentration in the sample.
4.
Pyxis SP-910 will display the fluorescein concentration in ppb as fluorescein.
Figure 7 Fluorescein Measurement
For best results, ensure that the sample vial is clean. Wipe off water on the outside
wall of the sample vial using a lint-free tissue paper. Fill the sample vial to the 10 ml
mark. If the sample contains air bubbles, tap the sample vial gently to remove the
bubbles before placing the sample vial to sample vial compartment.
4.2 Fluorescein calibration
Deionized water (DI) as the blank calibration solution and the 20 ppb fluorescein
calibration standard solution are needed.
1.
Press the CAL on the main page, then choose the M-F and press the OK key to
launch the fluorescein calibration page.
2.
Follow the message prompts, insert the DI blank into the sample vial compartment
and press the OK key.
3.
Follow the message prompts, and insert the 20 ppb standard into the sample vial
compartment and press the OK key.
4.
Press the OK key to return to the main page.
Figure 8 Fluorescein Calibration
If calibration fails, the followings should be checked:
The DI blank is being contaminated.
The 20 ppb fluorescein standard solution is decayed or being contaminated.
The light shield cover is not in the closing position.
The sample vial compartment is blocked with debris, water, or other materials.
The 20 ppb standard solution shall be stored in a brown or black opaque bottle.
Exposing the fluorescein standard to light will cause the standard losing the fluorescein
concentration. Many substances, such as quaternary amine cause a negative
interference. Many other substances such laundry detergents that contain optical
brightener will cause a significant positive interference.
5 Colorimetric Measurement
5.1 Supported Methods
A wide range of colorimetric methods is supported by the Pyxis SP-910 analyzer and
the number of them keeps increasing with continuous development of Pyxis. See
corresponding Hach® methods in Appendix A.
Table 2 List of Supported Colorimetric Methods
Abbreviated
Method
Name
Method Name
Description
Range
CL-F
F-Chlorine
Free chlorine, DPD method
2.2 ppm
CL-T
T-Chlorine
Total chlorine, DPD method
2.2 ppm
CL2H
CL2High
High range, DPD method
6.0ppm
CuBi
Cu_Bicinch
Bicinchoninate, EPA approved
for reporting wastewater
analysis
5.0 ppm
DEHA
DEHA
Method for N,N-
diethylhydroxylamine and other
oxygen scavengers
0.5 ppm
Ca
Ca
Calmagite method for calcium
4.00 ppm as
CaCO3
Mg
Mg
Calmagite method for
magnesium
4.00 ppm as
CaCO3
FePh
Fe_phenanth
Total iron using 1,10-
phenanthroline, USEPA
approved for reporting
wasterwater analysis
3.00 ppm
FeZi
FeZine
FerroZine method
1.300 ppm
FeTp
FeTptz
Total iron using TPTZ
1.80 ppm
MoHR
Mo_HighRange
High range molybdate
40.0 ppm
MoLR
Mo_LowRange
Low range molybdate using
ternary complex method
3.0 ppm
NO2H
NO2H
High range nitrite, ferrous
sulfate method
150 ppm as
NO2
NO2L
NO2L
Low range nitrite, diazotization
method, USEPA approved for
reporting wastewater and
drinking water analysis
0.350 ppm
as NO2
PMoV
OPO4-MoV
Reactive phosphate using
molybdovanadate method
30.0 ppm as
PO4
OPO4
OPO4
Reactive phosphate using
ascorbic acid molybdenum blue
method, USEPA accepted for
wastewater analysis
2.5 ppm as
PO4
OrgP
Phosphonate
UV digestion and ascorbic acid
reduction molybdenum blue
7.1 ppm as
PBTC
Reactive phosphate, amino
acid reduction method
30.0 ppm
ClO2
ClO2-DPD
DPD method, USEPA accepted
for reporting drinking water
analysis
5.00 ppm
ClO2D
ClO2Direct
Direct method for chlorine
dioxide
35.0 ppm
SiHR
SiHR
High range silica
75.0 ppm as
SiO2
SiLR
SiLR
Low range silica
1.60 ppm as
SiO2
AZOL
Azole
UV digestion for tolyltriazole
and benzotriazole
16.0 ppm
SO4
SO4
Barium sulfate turbdimetric
method
70.0 ppm
POLY
Polymer
Turbidimetric method for
anionic polymeric dispersant
13.0 ppm as
PAA
FeMo
FeMo
Total iron method for water
containing molybdate
1.80 ppm
Cr6
Cr6
1,5-Diphenylcarbohydrazide
method for chromium
hexavalent,USEPA accepted
for wastewater analyses
0.6ppm
CrT
CrTot
Alkaline hypobromite Oxidation
method for chromium total
0.6ppm
NH3S
NH3Sal
Salicylate method for
nitrogen,ammonia
0.5ppm
NH2C
NH2CL
Indophenol method for
chloramine mono
3.0ppm
N2H4
N2H4
p-Dimethylaminobenzaldehyde
method for hydrazine
0.5ppm
MnL
MnLow
Low range manganese
0.7ppm
MnH
MnHigh
High range manganese
20.0ppm
BLCH
Bleach
Direct Method measuring
sodium hypochlorite
concentration
AL
Alumi
Aluminon method for aluminum
0.8ppm
F
Floride
SPADNS method for fluoride
2.0ppm
CuL
CuPorp
Porphyrin Method for copper
0.2ppm
Zn
Zinc
Zincon method for zinc,USEPA
approved for wastewater
analysis
3.0ppm
S2-
Sulfide
Methylene blue method for
sulfide,USEPA accepted for
reporting wastewater analysis
0.7ppm
Pyridine-Pyrazalone method for
cyanide
0.2ppm
NO3M
NO3M
Middle range nitrate
5ppm as N
NO3H
NO3H
High range nitrate
30ppm as N
Ni
Ni
PAN method for nickel
1ppm
CYAN
CYAN
Turbidimetric method for
cyanuric acid
55ppm
pH
pH
Phenol red method for pH
8.5
5.2 Select a Method
Move the icon focus to the method icon Mthd using the navigational (left, right, up, or
down) keys. Press OK on the icon to launch the first method selection page. The
methods shown on the top row of the page are the most frequently selected methods.
Figure 9 Method Selection
The followings are the operations associated with this page:
1.
Use the navigational keys and the OK key to select and launch a method.
2.
Press the arrow icon at the end of the favorite methods row to return to the main
page.Press the arrow icon at the lower right corner of the page to display the
second method selection page if the device is loaded with more than 23 methods.
Note: Methods shown in the method selection pages include Hach© equivalent
methods and Pyxis specific advanced methods. The table in Appendix A provides a
brief description of Pyxis method names and their corresponding Hach® program
number. Hach® reagents for 10 ml sample can be used for the test.
5.3 Single Timing Step Method
Most of colorimetric methods have only one timing step. As an example, in the DPD
freechlorine method, it takes three minutes for the DPD powder reagent to completely
react with chlorine in the water sample. The DPD free chlorine method has a single
three-minute timing step. Figure 10 shows the main page of a method with a single
timing step.
Figure 10 Single Timer Method
5.4 Single-Vial Procedure
1.
Place the sample vial filled with the water sample in the Pyxis SP-910 sample vial
compartment and press the ZERO button. Pyxis SP-910will display the page
shown in Figure 11.
2.
Take the sample vial out and add the reagent to the sample vial.
3.
Place sample vial back into the sample vial compartment and press the timer
button TIMER. Pyxis SP-910 will start to monitor the reaction between the reagent
and the species you want to measure in the water sample. The concentration is
shown in the chart as a function of time (Figure11).
4.
When the timer reaches the preset time and the reaction is complete, the value of
concentration will be shown on the top right corner of the page.
5.
The rate of the reaction is often faster than the standard pre-set time, which will
become apparent from the concentration-time plot. You can press the STOP
button to stop the timer and terminate the timing step. The last read concentration
value will be displayed on the top right corner of the page after you terminate the
timing step.
Figure 11 Concentration as a Function of Time
5.5 Two-Vial Methods
Some colorimetric methods require using two vials. The water sample is added to two
identical vials. One vial is being used to zero the colorimeter, referred as to the
prepared blank. A reagent is added to the other vial, referred as to the prepared
sample. The absorbance value is determined from the prepared sample.
If the method requires two or more reagents, the prepared blank could be the resulting
solution after one or more reagents have been added to the sample.
The following procedure is typical for two-vial methods:
1.
Place the prepared blank into the Pyxis SP-910 sample vial compartment and
press the ZERO button to zero the instrument.
2.
Place the prepared sample into the Pyxis SP-910 sample vial compartment and
press the TIMER button to start the method timer.
3.
When the timing step is completed, the measured concentration will be displayed
on the top of the page. The timing step could be terminated earlier by pressing
STOP button.
4.
Optionally, Pyxis SP-910can be re-zeroed using the prepared blank after the
timing step is completed or terminated. The blank reading will be subtracted from
the measured concentration value, and the displayed concentration value on the
top-right corner will be updated. This step is optional. It is only necessary if the
prepared blank changes its color during the timing period.
5.
Optionally, the prepared sample vial can be put back and read again by pressing
the READ button if the blank is re-zeroed after the timing step is completed or
terminated. A new concentration value based on the last absorbance value
measured will be calculated and displayed.
5.6 Multiple Timing Steps
Some colorimetric methods have two or three timing steps. Pyxis SP-910 shows a
count-down timer for the timing steps before the last timing step (Figure 12). During
these timing steps, one or more reagents are added to the sample, or operations such
as swirling the vial to mix the reagent and the sample are being performed. These
methods usually use one vial for the prepared blank and the other for the prepared
sample.
Figure 12 Multiple Timer Method
In order to show the concentration-time curve as shown in Figure 9during the last
timing step, Pyxis SP-910must be zeroed using the prepared blank before the last
timing step. Thus, the last timer button will not be selectable until Pyxis SP-910 has
been zeroed using the prepared blank. Multi-timing step Hach® methods require
zeroing the colorimeter using the prepared blank after the last timing step is completed.
Pyxis SP-910can optionally be re-zeroed using the prepared blank after the last timing
step. The blank value measured will be subtracted from the concentration value
measured at the end of the last timing step. Optionally, the READ button could be
pressed to read the prepare sample again.
The following procedure is typical for methods having two-timing steps:
1.
Press the TIMER1 button to start the first timer. Complete the necessary
operations to prepare the blank and the sample.
2.
Place the prepared blank into the Pyxis SP-910 sample vial compartment
and press the ZERO button.
3.
Place the prepared sample into the Pyxis SP-910 sample vial compartment
and press the TIMER2 button to start the second timer. Pyxis SP-910 will
display the measured concentration as a function of time as shown in Figure
9.
4.
When the timing step is completed, the measured concentration will be
displayed on the top right of the screen. The timing step could be terminated
earlier by pressing STOP button.
5.
Optionally, Pyxis SP-910can be re-zeroed using the prepared blank after the
timing step is completed or terminated. The blank reading will be subtracted
from the measured concentration value, and the displayed concentration
value on the top-right corner will be updated. This step is optional. It is only
necessary if the prepared blank changes its color during the timing period.
5.7 Advanced Methods
Pyxis SP-910provides 6 LED wavelengths and can measure absorbance values at
multiple LED wavelengths. Consequently, Pyxis SP-910can provide many predefined
advanced methods that traditionally require complex and often expensive lab testing
procedures.
Low range, direct reading chlorine dioxide, 0 to 35.0 ppm
The maximum absorption bank of aqueous chlorine dioxide is around 360nm. Pyxis
SP-910has a 365nm UV LED and can be used to directly measure chlorine dioxide. It
offers a much lower detection limit (0.2 ppm) than direct methods available from other
portable colorimeters having only light sources in the visible range.
Select ClO2D in the method selection page and carry out the following steps to
measure chlorine dioxide:
1.
Place a vial filled with deionized water into the vial compartment and press the
ZERO button to zero Pyxis SP-910.
2.
Discard the deionized water and fill the same vial with the sample. Place the
vial into the vial compartment and press READ button to read. The measured
chlorine dioxide concentration will be displayed in the top of the method page.
Turbdimetric Anionic Polymer Method
1.
Add polymer reagent 1 to 10 ml sample and inverse the sample vial 5 times to
mix the reagent with the sample. Place the sample via to the sample vial
compartment.
2.
Press on ZERO.
3.
Add polymer reagent 2 and press on TMR1 to start the five minutes timer.
4.
Gently inverse the sample via for 10 times and place the sample vial to the
sample vial compartment.
5.
Polymer concentration will be measured and displayed when the five minute
timer is reached. The polymer concentration is shown as ppm PAA (polyacrylic
acid) equivalent.
5.8 Method Setup and Calibration
Press the SETUP button in the method result page to launch the method setup and
calibration page.
5.8.1
Set up the method parameters
Press the FORM button to select a concentration form from the list of forms
that are available for this specific method (Figure 13).
Press the UNIT button to select a concentration unit among the list of ppb,
ppm, mg/L, and ug/L (Fihure 14).
Figure 13 Method Form Selection
Figure 14 Method Unit Selection
5.8.2
Slope Calibration
If the method has been calibrated priorto shipping, there is no need to
calibrate unless a calibration check indicates that the method needs a
calibration. The following steps are used to calibration a method:
1.
Use a calibration standard of known concentration. Follow the steps
required by the method and note the value reported by SP-910.
2.
If the measured value differs from the known standard value,Press
theCONFG button to launch the method configuration page.
3.
Press the slope calibration button SlpCal. A numeric keyboard will be
displayed.
4.
Enter the concentration value and press the OK key on the enter key in
the numeric keyboard to return to the configuration page.
5.
Press the EXIT button. Press the OK key to accept the calibration or other
key to cancel the calibration.
Figure 15 Slope Calibration
For best results, the concentration of the standard solution should be less than
the maximum concentration for the method (table 2) and greater than the half
of the maximum concentration. For example to calibrate total chlorine, the
chlorine concentration in the standard solution should be between 1.1 and 2.2
ppm.
The corresponding calibration parameters will be updated and saved in the
memory as the working calibration parameter set. Note that this set of
calibration parameters are not the same as the default set. You can use
Default button to copy the default calibration parameters to the working set.
5.8.3
Reagent Blank Calibration
Some methods have a non-zero intercept value in the calibration equation. For
these methods, a proper non-zero intercept value is pre-loaded in SP-910prior
to shipping. The following steps are used to carry out a reagent blank
calibration:
1.
Follow the normal steps to carry out a measurement on a deionized water
sample.
2.
Press the CONFG button to launch the method configuration page.
3.
Press the reagent blank calibration button BLKC
4.
Press the OK key to save when exiting from the configuration page or
press other keys to cancel.
5.8.4
Resume to Default Calibration Parameters
Pressing the Default button will copy the default calibration intercept and
slope to the working intercept and slope, respectively. If the default calibration
parameters were created prior to shipping, this button action is to restore the
working calibration parameters to the original factory loaded calibration
parameters.
6
Turbidity Measurement
6.1 Operation
Follow the following steps to measure turbidity:
1.
Fill the 10 ml sample vial to above the 10 ml mark.
2.
Insert the sample vial to the sample vial compartment.
3.
Slide the light shield cover to the closed position.
4.
Press the Turb on the main page, then press the OK key, Pyxis SP-910 will
start to measure the turbidity in the sample.
6.2 Turbidity Calibration
1.
Fill the 10 ml sample vial to above 10 ml mark with the deionized water.
2.
Insert the sample vial to the sample vial compartment.
Slide the light shield cover to the closed position.
4.
Press the CAL on the main page, then choose the Turb calibration and press
the ok button to launch the Turb calibration page.(Figure 16)
5.
Press the OK key to measure the deionized water
6.
Fill the 10 ml sample vial to above 10 ml mark with the 100 NTU formazin
standard. Insert the sample vial to the sample vial compartment.
7.
Press the OK key to measure the 100 NTU standard. Low range turbidity
calibration is successful
8.
Press the OK key to continue high range turbidity calibration. If high range
turbidity calibration not required,press any keys to exit. (Figure 17)
9.
Fill the 10 ml sample vial to above 10 ml mark with the 400 NTU formazin
standard. Insert the sample vial to the sample vial compartment.
10. Press the OK key to measure the 400 NTU standard. High range turbidity
calibration is successful,
11. Press any keys to exit.
Figure 16 Turbidity Calibration-1
Figure 17 Turbidity Calibration-2
The following steps are used to measure the absorbance values of a sample:
1.
Press the ABS to launch the absorbance measurement page.
2.
Place a vial filled with the blank sample in the sample vial compartment. Press
the ZERO button to zero the method.
3.
Place a vial filled with the sample in the sample vial compartment. Press the
READ button to read absorbance.The absorbance values of first 6
wavelengths (Table 3) will be shown. Press the READ button again to show
the absorbance values of the last three wavelengths.
Figure 18 Absorbance Measurement
Press EXIT to return to the main page. Timing function for absorbance measurement may
not be available for some models.
Table 3 Wavelength of each channel
Channel
Wavelength (nm)
1
560
2
570
3
470
4
not used
5
455
6
525
7
365
8
630
9
420
Note that the absorbance values measured with Pyxis SP-910 is generally smaller than
those measured with a spectrophotometer equipped with a monochromatic light source or
detector. The Pyxis SP-910 absorbance values should, however, linearly correlate with the
absorbance values measured with the spectrophotometer. Thus for any colorimetric
system, Pyxis SP-910 absorbance follows Lambert-Beer law.
8
Bluetooth Interface
SP-910 equipped with a Bluetooth interface, which allows a user to connect to SP-910
with a computer or a smartphone to do the following tasks:
Configure device
Add user defined colorimetric methods
Upgrade device firmware
Also with the Bluetooth interface, user can calibrate an inline fluorometer directly from
SP-910 in the field. Below sections describe how to connect and communicate with
your SP-910 via a computer and Pyxis Nebula software.
8.1 Install Software
Download Nebula software by clicking this link, unzip and install Nebula software,
Bluetooth adapter driver will be installed as well. Plugin the Bluetooth adapter shipped
along with your SP-910 device, open Nebula software.
8.2 Open SP-910 Bluetooth Interface
Bluetooth function is normally switched off in order to reduce power consumption, to
open Bluetooth interface, select SYS in the main menu and click BTLE in the next
screen.
8.3 Scan and Connect
Open Nebular software, select the right COM port, depends on hardware variance,
actual COM port may not be same as below picture shows. Then click Scan button to
scan Bluetooth enabled device, SP-910 device will appear in device list. Choose the
SP-910 device by clicking device name, and click Connect button to connect to it.
Figure 19 Nebula Scans Bluetooth Device
8.4 Device Configuration
Setup Device Name
When connected, the device information will be uploaded and displayed, only the
device name is allowed to change as Fig 21 shows
Figure 20 Change Device Name
Setup Product
Your product name product/PTSA ratio can be setup according to Fig 22
Figure 21 Setup Product
Add User Defined Colorimetric Methods
User can edit new method parameters in Nebula software and download to SP-
910via Bluetooth interface according to Fig 23.
“Add”: add one empty method entry in user defined method list
“Delete”: delete one method from user defined method list
“Download”: send new method to SP-910
“Upload”: read user defined methods from SP-910
“Load/Save methods” read/write methods to local file
Fig 22 Add User Defined Colorimetric Method
9
Maintenance
Use a soft cloth or lint free paper tissue to clean the sample vial compartment
periodically. Remove debris, scale, and deposit promptly.
Although Pyxis SP-910 is protected from water damage, it is a good practice to avoid
water entering the sample vial compartment and becoming trapping underneath the
navigational control pad. Deposits left behind when the water is evaporated could
affect Pyxis performance.
Pyxis SP-910should be stored in the temperature range of 0 to 140°F (-18 to 60°C)
and relative humidity less than 85% at 106 °F (41 °C). Do not leave the Pyxis SP-910
in a parked vehicle. The temperature inside a parked vehicle can reach above 150 °F
in summer and -20 °F in winter. Exposing the Pyxis SP-910 to extreme temperature or
humidity will cause a gradual decay in performance of fluorescence measurements
and require more frequent calibrations.
During storage and transportation, do not leave a sample vial in the sample vial
compartment. Close the lid of the sample vial compartment during storage and
transportation.
Replace batteries when the Pyxis SP-910 displays a warning message indicating LOW
BATTERY voltage. Remove batteries from Pyxis SP-910 battery compartment if Pyxis
SP-910 is going to be placed in storage for a long period time.
When the Pyxis SP-910 is shipped, a desiccant pack is included in the desiccant
compartment underneath the cover of the battery compartment. It is recommended
that a new desiccant pack is replaced each time the batteries are replaced.
The Pyxis SP-910 will prompt a warning message if it detects an abnormal condition or
operation. On screen prompts direct the user to take appropriate corrective actions in
most cases.
If an unspecific error occurs or Pyxis SP-910 cannot be turned on, reboot the
instrument by taking a battery out of the battery holder and re-install the battery.
If the Pyxis SP-910 has been idle for more than two months and cannot be turned on,
replace all four batteries with four new AA alkaline batteries.
A diagnostics page can be launched by press the SYS icon in the main page. The
software version and its associated hash code can be found in the diagnosis page.
Contact Pyxis professionals at [email protected] and provide with following
information to ensure high quality technical support.
Table 4 Contact Information
Items
Note
Contact Name
Phone
Email
Customer Name
Product Number (P/N)
Can be found on the product label on back of
product
Serial Number (S/N)
Can be found on the product label on back of
product
Firmware version
Can be found in diagnosis page
Problem Description
Capture warning message if applicable
11 Appendix A.
Pyxis Method and Hach® Method Number (PRMP) Cross Reference
Abbreviated
Method
Name
Method Name
Corresponding Hach © method
Hach Methed
Number
CL-F
F-Chlorine
Chlorine, Free, DPD, PRMP 9
8021
CL-T
T-Chlorine
Chlorine, Totoal, DPD, PRMP 9
8167
Cl2H
Cl2High
High Range DPD Chlorine, No
sample change needed
10070
CuBi
Cu_Bicinch
Copper, Bicinchoninate, PRMP 20
8506
DEHA
DEHA
DEHA, Iron Reduction Method for
Oxygen Scavengers, PRMP 25
8140
Ca
Ca
Calcium: Calmagite Colorimetric
Method, PRMP 29
8030
Mg
Mg
Magnesium: Calmagite
Colorimetric Method, PRMP 30
8030
FePh
Fe_phenanth
Iron, 1,10 phenanthroline, PRMP
33
8008
FeZi
FeZine
Iron, FerroZine, PRMP 37
8147
FeTp
FeTptz
Iron, TPTZ, PRMP 39
8112
MoHR
Mo_HighRange
Molybdenum, High Range,
Mercaptoacetic Acid, PRMP 44
8036
MoLR
Mo_LowRange
Molybdenum, Low Range, Ternary
Complex, PRMP 47
8169
NO2H
NO2H
Nitrite, High Range, Ferrous
Sulfate, PRMP 59
8153
NO2L
NO2L
Nitrite, Low Range, Diazotization,
PRMP 60
8507
PMoV
OPO4-MoV
Phosphorus, Reactive,
Molybdovanadate, GRMP 77
8114
OPO4
OPO4
Phosphorus, Reactive,
Orthophosphate Ascorbic Acid,
GRMP 79
8048
OrgP
Phosphonate
Phosphonates, Persulfate UV
Oxidation, PRMP 80
8007
PAmi
OPO4-Amino
Phosphorus, Reactive, Amino
Acid, GRMP 85
8178
ClO2
ClO2-DPD
Chlorine Dioxide, DPD, PRMP 112
10126
ClO2D
ClO2Direct
Chlorine Dioxide, Direct
Reading,PRMP7
8345
SiHR
SiHR
Silica, High Range,
Silicomolybdate, PRGM 89
8185
SiLR
SiLR
Silica, Low Range, Heteropoly
Blue, PRMP 90
8186
AZOL
Azole
Benzotriazole, UV Photolysis,
PRMP 3
8079
Sulfate. PRMP 91
8051
POLY
Polymer
Turbidimetric method for anionic
polymers
N/A
FeMo
FeMo
Iron, for cooling water with
molybdenum-based treatment,
PRMP 38
8365
Cr6
Cr6
Hexavalent chromium, 1,5-
Diphenylcarbohydrzaide Method,
PRMP 13
8023
CrT
CrTot
ChromiumtotalAlkaline
Hypobromite Oxidation Method,
PRMP15
8024
NH3S
NH3Sal
Ammonia Salicylate Method,
PRMP 64
8155
NH2C
NH2Cl
Indophenol Method for
MonoChloramine, PRMP 110
10171
N2H4
N2H4
P-Dimethylaminobenzaldehyde
Method for Hydrazine, PRMP 31
8141
MnL
MnLow
Low Range Manganese PAN
Method, PRMP 43
8149
MnH
MnHigh
High Range Manganese,
Periodate Oxidation Method,
PRMP 41
8034
BLCH
Bleach
Direct Method measuring sodium
hypochlorite concentration
N/A
Al
Alumi
Aluminon Method for Aluminum,
PRMP 1
8012
F
Floride
SPADNS 2 Method for Fluoride,
PRMP 27
8029
CuL
CuPorp
Porphyrin Method for Copper,
PRMP 22
8143
Zn
Zinc
Zincon Method for Zinc, PRMP 97
8009
S2-
Sulfide
Methylene Blue Method for
Sulfide, PRMP 93
8131
CN
Cyanide
Pyridine-Pyrazalone Method for
Cyanide, PRMP 23
8027
NO3M
NO3M
Middle range nitrate,PRMP54
8171
NO3H
NO3H
High range nitrate,PRMP51
8039
Ni
Ni
PAN method for nickel,PRMP48
8150
CYAN
CYAN
Turbidimetric method for cyanuric
acid ,PRMP 24
8139
pH
pH
Phenol red method for pH,PRMP
75
10076
Hach ® is a registered trademark of the Hach Company, Loveland, CO USA