SciCan statim 2000 G4 Operator's Manual
Text of the manual
95-113364 US EN R4. STATIM 2000/5000 G4 Operator’s Manual. Copyright 2016 SciCan Ltd. All rights reserved.
STATIM
® 2000/5000 G4
CASSETTE AUTOCLAVE
• Operator’s Manual
95-113364 US EN R4 STATIM 2000 G4 5000 G4 OpMan.indd 1
2017-01-18 11:01 AM
Table of Contents
1. Introduction .......................................................... 3
2. Important Information .......................................... 4
2.
Disclaimers
2.2 STATIM 2000 G4 - Unit Overview
2.3 STATIM 5000 G4 - Unit Overview
2.4 Touchscreen Overview
2.5 Setup Menu Overview
3. Setting Up your STATIM ...................................... 10
3.1
Positioning and Powering your Unit
3.2
Leveling your Unit
3.3
Connecting the Waste Bottle
3.4
Filling the Water Reservoir
3.5
Priming the Pump
3.6
Shipping the Unit
3.7
Setting the Time
3.8
Setting the Date
3.9
Setting the Language
3.10 Assigning Unit Identifier Number
3.11 Creating a User ID and PIN
3.12 Setting Up Process Enforced Usage
3.13 Changing the Touchscreen Display Themes
3.14 Adjusting the Screensaver Delay
3.15 Adjusting the Screen Contrast
3.16 Turning the Button Sound ON or OFF
3.17 Adjusting the Button Beep Volume
3.18 Setting up and Using your STATIM G4 Web Portal
3.19 Connecting to a Network
3.20 Connecting to a Wireless Network
3.21 Reserving an IP Address for your STATIM
4. Using Cassettes and Preparing Instruments .... 20
4.1 Using the STATIM 2000 G4 Cassette
4.2 Using the STATIM 5000 G4 Cassette
4.3 Preparing and Loading Instruments
4.4 Using Biological and Chemical Indicators
4.5 Instrument Weight Guide
5. Using your STATIM ............................................. 26
5.1 Preparing Unit for Use
5.2 STATIM 2000 G4 — Selecting a Cycle
5.3 STATIM 5000 G4 — Selecting a Cycle
5.4 STATIM 2000 / 5000 G4 — Running a Cycle
5.5 STATIM 2000 / 5000 G4 — Stopping a Cycle
6. Storing and Retrieving Cycle Information.......... 37
6.1 Retrieving Cycle Information Using the Touchscreen
6.2 Retrieving Cycle Information Using the USB Data Backup
6.3 Cycle Printout Overview
7. Retrieving a Remote Access Code .................... 39
8. Printing Cycle Information ................................. 40
8.1 Connecting to a Printer
8.2 Adjusting your Printer Settings
8.3 External Printers and Specifications
9. Maintaining your STATIM ................................... 41
9.1 Cleaning the Cassette
9.2 Cleaning the Water Reservoir Filter
9.3 Cleaning the Reservoir
9.4 Cleaning the Exterior Surfaces
9.5 Changing the Bacteria Retentive Filter and the Air Filter
9.6 Replacing the Cassette Seal
9.7 Maintaining Fluid Levels
9.8 Reading Water Quality
9.9 Using On-Screen Instructions
9.10 Preventative Maintenance Schedules
10. Troubleshooting your STATIM ........................... 47
11. Spare Parts .......................................................... 50
12. Warranty ............................................................ 51
13. Specifications ..................................................... 52
13.1 STATIM 2000 G4
13.2 STATIM 5000 G4
14. Legal Disclaimer and Privacy Policy .................. 54
STATIM Cassette Autoclave and STATIM are registered trademarks
and Your Infection Control Specialist, DriTecR and DriTecC logos are
trademarks of SciCan Ltd. All other trademarks referred to in this
manual are the property of their respective owners.
For all service and repair inquiries:
In Canada
1-800-870-7777
United States:
1-800-572-1211
Email:
[email protected]
SciCan Inc.
701 Technology Drive
Canonsburg, PA 15317 USA
Phone: (724) 820 1600
Fax: (724) 820 1479
Toll free: 1-800-572-1211
Manufactured by:
SciCan Ltd.
1440 Don Mills Road,
Toronto ON M3B 3P9
CANADA
Phone:
(416) 445-1600
Fax:
(416) 445-2727
Toll free:
1-800-667-7733
95-113364 US EN R4 STATIM 2000 G4 5000 G4 OpMan.indd 2
2017-01-18 11:01 AM
Congratulations on selecting the STATIM® G4 Cassette Autoclave. The STATIM is a compact, countertop
unit that is suitable for steam sterilization.
The details of installing, operating and maintaining your STATIM are all contained within this operator’s
manual. Please read these instructions before operating this unit and keep them for future reference.
Operational, maintenance and replacement instructions should be followed. Contents of this manual are
subject to change without notice to reflect changes and improvements to the STATIM product.
The STATIM is suitable for the sterilization of dental and medical instruments designed to withstand steam
sterilization. The STATIM has not been designed to sterilize liquids, cloth loads, bio-medical waste or
materials not compatible with steam sterilization. The processing of such loads may result in incomplete
sterilization and / or damage to the autoclave. For more information about instrument suitability for steam
sterilization, consult the instrument manufacturers’ reprocessing instructions.
STATIM 2000 G4
STATIM 5000 G4
95-113364 US EN R4 STATIM 2000 G4 5000 G4 OpMan.indd 3
2017-01-18 11:01 AM
2 Important Information
2.1 Disclaimers
Use only steam-process distilled water or VistaPure deionized water in your STATIM.*
Do not permit any person other than certified personnel to supply parts for, service or maintain your
STATIM. SciCan shall not be liable for incidental, special or consequential damages caused by any
maintenance or services performed on the STATIM by a third party, or for the use of equipment or parts
manufactured by a third party, including lost profits, any commercial loss, economic loss, or loss arising
from personal injury.
Never remove the cover of the unit and never insert objects through holes or openings in the cabinetry.
Doing so may damage the unit and/or pose a hazard to the operator.
All elements of this book are common to both STATIM 2000 G4 and STATIM 5000 G4 except when noted.
IMPORTANT: Follow local guidelines governing verification of the sterilization procedure.
Drying Performance
The STATIM 2000 G4 and 5000 G4 have been designed to provide a complete sterilization solution for
your unwrapped and wrapped instrument needs: Rapid sterilization balanced with rapid drying through
the use of SciCan's Dri-Tec Drying Technology.
To dry instruments, the STATIM 2000 G4 uses a combination of forced filtered air and convection heat.
The convection heat is derived from utilizing the remaining heat in the system after the sterilization phase
has been completed. The heat is then captured and released into the cassette to rapidly dry a properly
loaded cassette.
The STATIM 5000 G4 also uses forced filtered air and heat generated by the sterilization phase to dry the
instruments.
Please refer to Section 4.4 Preparing and Loading Instruments for instructions on the proper arrangement
of instruments in the cassette. By carefully following these directions on how to properly load the cassette
chamber, rapid drying of the load will be achieved.
* Although, SciCan recommends the use of steam-processed distilled water in the STATIM and Bravo autoclaves, SciCan has tested and
approved the deionized water from the VistaPure and found it to be acceptable for these units. The VistaPure is an advanced water system that
produces two grades of high quality water, including deionized water suitable for autoclaves, making it a viable alternative to steam distilled
water. Other water systems have not been tested or approved for use with SciCan sterilizers.
95-113364 US EN R4 STATIM 2000 G4 5000 G4 OpMan.indd 4
2017-01-18 11:01 AM
2 Important Information
2.2 STATIM 2000 G4 — Unit Overview
The following symbols may appear on the unit:
When you receive your STATIM 2000 G4, the items listed below will be included. If any of the items are
missing, contact your dealer immediately.
Cassette Tray and Lid
Instrument Rack
Waste Bottle
Bottle Lid Fitting
Tube Mounting Hardware
Operator’s Manual
Exhaust Tube
USB Memory Stick
Power Cord
Statmat
Caution: Hot Surface and/or
Hot Steam
Caution: Risk of electrical
shock. Disconnect supply
before servicing
Caution: Refer to user
manual for details
Steam distilled water or
VistaPure deionized water
(do not use tap water)
The following symbols appear
in the margins of this book.
A potential hazard
to the operator.
A situation that
may lead to a
mechanical failure.
Important
information.
1
12
11
10
9
8
7
6
5
4
3
2
LCD Touchscreen
Reservoir cap /
water filter
USB port
Power switch
Power cable port
Leveler feet
Ethernet port
Exhaust tube port
Compressor
Cassette
Biological filter
RS232 port
START
BUTTON
UNWRAPPED
CYCLE
AIR DRY ONLY
CYCLE
STOP
BUTTON
WRAPPED
CYCLE
RUBBER /
PLASTIC CYCLE
1
10
12
11
9
8
6
5
4
3
2
7
STATIM 2000/5000 G4 Operator’s Manual 95-113364 Rev. 2.0 Copyright 2014 SciCan Ltd. All rights reserved.
statim 2000/5000 G4
Cassette autoClave ™
• operator’s Manual
95-113364 US EN R4 STATIM 2000 G4 5000 G4 OpMan.indd 5
2017-01-18 11:01 AM
2 Important Information
2.3 STATIM 5000 G4 — Unit Overview
The following symbols appear on the unit:
When you receive your STATIM 5000 G4, the items listed below will be included. If any of the items are
missing, contact your dealer immediately.
Cassette Tray and Lid
Unwrapped
Instrument Rack
Waste Bottle
Bottle Lid Fitting
Tube Mounting Hardware
Statmat
Power Cord
Operator’s Manual
Exhaust Tube
USB Memory Stick
Wired Pouch Rack
The following symbols appear
in the margins of this book.
A potential hazard
to the operator.
A situation that
may lead to a
mechanical failure.
Important
information.
1
11
12
10
9
8
7
6
5
4
3
2
LCD Touchscreen
Reservoir cap /
water filter
USB port
Power switch
Power cable port
Leveler feet
Exhaust tube port
Biological filter
Cassette
RS232 port
Ethernet port
Air filter
Caution: Hot Surface and/or
Hot Steam
Caution: Risk of electrical
shock. Disconnect supply
before servicing
Caution: Refer to user
manual for details
Steam distilled water or
VistaPure deionized (do not
use tap water)
START
BUTTON
UNWRAPPED
CYCLE
HEAVY DUTY
UNWRAPPED
CYCLE
STOP
BUTTON
WRAPPED
CYCLE
RUBBER /
PLASTIC CYCLE
1
11
10
12
8
6
5
4
2
7
9
3
STATIM 2000/5000 G4 Operator’s Manual 95-113364 Rev. 2.0 Copyright 2014 SciCan Ltd. All rights reserved.
statim 2000/5000 G4
Cassette autoClave ™
• operator’s Manual
95-113364 US EN R4 STATIM 2000 G4 5000 G4 OpMan.indd 6
2017-01-18 11:01 AM
SELECT A CYCLE
Wrapped cycles
Unwrapped cycles
Rubber / Plastic cycles
Air Dry Only (2000) or
Heavy Duty Unwrapped
(5000)
Start
button
Display
Stop
button
Settings and unit
information
Cassette properly inserted
Water level
and quality status
USB storage device
connected/not connected
(turns green when active)
95-113364 US EN R4 STATIM 2000 G4 5000 G4 OpMan.indd 7
2017-01-18 11:01 AM
2.5 Setup Menu Overview
mm/dd/yyyy hh:mm:ss
SELECT A CYCLE
Time
1
2
3
4
5
6
7
8
9
CL
0
EN
--:--
24H
14:02
Date
--/--/--
DD/MM/YY
01/01/2011
1
2
3
4
5
6
7
8
9
CL
0
EN
On
Default
Off
DST On/Off
Yyyy/mm/dd xxxxx OK
Yyyy/mm/dd xxxxx OK
Yyyy/mm/dd xxxxx OK
Yyyy/mm/dd xxxxx OK
Yyyy/mm/dd xxxxx OK
Yyyy/mm/dd xxxxx (CFxx)
Yyyy/mm/dd xxxxx (CFxx)
Yyyy/mm/dd xxxxx (CFxx)
Yyyy/mm/dd xxxxx (CFxx)
Yyyy/mm/dd xxxxx (CFxx)
12
Default
24
12h/24h Setup
MM/DD/YYYY
Default DD/MM/YYYY
Date Format
French
Default English
Language
Unit No
1
2
3
4
5
6
7
8
9
CL
0
EN
000
xx.x uS
xx.x ppm
WATER LEVEL
Serial Printer
Default No Printer
Printer type
1200
Default 9600
Baud rate
Setup
Time
Date
DST On/Off
Setup
Time
Date
DST On/Off
Setup
Time
Date
DST On/Off
Setup
Time 12/24
Date Format
Recent Printouts
Setup
Time 12/24
Date Format
Recent Printouts
Setup
Time 12/24
Date Format
Recent Printouts
Setup
Unit No
Water Quality
Language
Setup
Unit No
Water Quality
Language
Setup
Unit No
Water Quality
Language
Setup
Printer Type
Baud Rate
Serial Number
Setup
Printer Type
Baud Rate
Serial Number
Setup
Printer Type
Baud Rate
Serial Number
CR
Default -
End of line CR/LF
Setup
Printer user ° char
Process Enforced
End of Line CR/LF
95-113364 US EN R4 STATIM 2000 G4 5000 G4 OpMan.indd 8
2017-01-18 11:01 AM
New PIN xxxx
OK
Upgrade Theme
Upgrade
PLEASE WAIT
Looking for Theme …
Upgrade Theme
Upgrade
P04_B091!
P04_B092!
P04_B093!
P04_B094!
Done!
2.5 Setup Menu Overview
Printer user ° char
1
2
3
4
5
6
7
8
9
CL
0
EN
000
Setup
Printer user ° char
Process Enforced
End of Line CR/LF
ON
Default OFF
Process Enforced
Setup
Printer user ° char
Process Enforced
End of Line CR/LF
PIN Setup
1
2
3
4
5
6
7
8
9
CL
0
EN
0000
Statim Silver
Default Statim NG
Themes
Upgrade Theme
Upgrade
Setup PIN
Setup
Themes
Screensaver
User
Themes
Change Theme
Upgrade Theme
Setup
Themes
Screensaver
User
3 min
Default 2 min
Screensaver
Setup
Themes
Screensaver
User
xx
Default 100
LCD Contrast
Setup
Network Setup
Button Beep
LCD Contrast
Setup
Network Setup
Button Beep
LCD Contrast
OFF
Default ON
Button Beep
Setup
Network Setup
Button Beep
LCD Contrast
80
Default 100
Beep Volume
Setup
Level
Beep Volume
Setup
Level
Beep Volume
95-113364 US EN R4 STATIM 2000 G4 5000 G4 OpMan.indd 9
2017-01-18 11:01 AM
3 Setting up your STATIM
3.1 Positioning and Powering your Unit.
Positioning your Unit
There are several factors that may affect the performance of your STATIM. Please review these factors,
and select a suitable location in which to install the unit.
•
Temperature and Humidity
Avoid installing your STATIM in direct sunlight or close to a heat source (e.g. vents or radiators). The
recommended operating temperatures are 15-25˚C (59˚F to 77˚F) with humidity of 25-70%.
•
Spacing
The vents and openings on the STATIM should remain uncovered and unobstructed. Leave a minimum
of 50 mm/2˝ between the top, sides and back of the unit and any wall or partition. For more detailed
information on clearances, see Section 13 Specifications.
•
Venting
The STATIM should be operated in a clean, dust free environment.
•
Work Surface
Your STATIM should be placed on a flat, level, water-resistant surface. Never install and operate the unit
on a sloped surface.
•
Electromagnetic Environment
Your STATIM has been tested and meets applicable standards for electromagnetic emissions. While
the unit does not emit any radiation, it may itself be affected by other equipment which does. We
recommend that your unit be kept away from potential sources of interference.
•
Electrical Requirements
To power your unit, use properly grounded and fused power sources with the same voltage rating as
indicated on the label at the back of your STATIM. Avoid multiple outlet receptacles. If using a surge
suppressor power bar, connect only one STATIM to it.
Powering your Unit
To power your STATIM, connect the power cord to the A.C. inlet receptacle at the back of the unit.
Ensure the power switch is in the OFF position and connect the unit to your power source.
95-113364 US EN R4 STATIM 2000 G4 5000 G4 OpMan.indd 10
2017-01-18 11:01 AM
3 Setting up your STATIM
3.2 Leveling your Unit
When placing your unit on a counter top, ensure it is stable and that all four feet are securely in contact
with the counter surface. This will prevent the unit from moving freely. Next, use the level indicator bubble
function in the settings menu to adjust the three leveler feet so that the unit drains properly. To access the
level indicator bubble from the main screen, follow these steps:
1.
2. Scroll to
and select.
3. Adjust the leveler feet to move the bubble. Position the bubble in the bottom right quadrant of the
target. This will ensure that the unit drains properly. Press STOP to exit and return to the cycle select
menu. When the unit is properly leveled, the bubble level will change from red to green.
Level
95-113364 US EN R4 STATIM 2000 G4 5000 G4 OpMan.indd 11
2017-01-18 11:01 AM
3 Setting up your STATIM
3.3 Connecting the Waste Bottle
The waste bottle 1 is used to collect the wastewater after it has been converted to steam and then
drained from the cassette. To connect the waste bottle to the STATIM, follow these steps (see Figure 4):
1. Insert the exhaust tube 2 into the fitting
on the back of the unit and pull gently to
confirm a tight fit.
2. Cut the tube to length and slide the waste
bottle fitting 4 into place.
3. Place the free end of the tube into the
hole in the lid of the waste bottle and
hand-tighten the fitting. Do not coil the
exhaust tube.
4. Unscrew the lid and copper condenser
coil 5 assembly from the waste bottle.
The lid and coil should come out together.
5. Fill the waste bottle with water to the
MIN line and replace the lid and copper
condenser assembly. Empty the waste
bottle often to avoid unpleasant odors and
discoloration of the contents. (A low-level
disinfectant, prepared according to the
manufacturer’s instructions, may be added
to the waste bottle to remedy this situation).
As a minimum, empty the waste bottle each
time you refill the reservoir.
6. Place the waste bottle near the unit. Store
the bottle below the unit. The tube can
be routed through a hole, (8 mm/0.3˝ in
diameter) in the countertop and secured
with the provided nylon clamps.
4
3
3
2
2
5
5
1
1
Figure 4
95-113364 US EN R4 STATIM 2000 G4 5000 G4 OpMan.indd 12
2017-01-18 11:01 AM
3 Setting up your STATIM
3.4 Filling the Water Reservoir
When filling the reservoir, ensure you only use steam-processed distilled water or VistaPure deionized
water containing less than 5 ppm total dissolved solids (having conductivity of less than 10 µS / cm).
The impurities and additives in other water sources will cause an error reading on the LCD.
To fill the reservoir, follow these steps (see Figure 5):
1. Remove the reservoir cap 1 .
2. Pour steam-processed distilled water or
VistaPure deionized water into the reservoir
until almost full (a maximum of 4L / 1 U.S. gal).
Use a funnel to avoid spillage.
3. Replace and secure the cap.
3.5 Priming the Pump
To prime the STATIM pump, follow these steps
(see Figure 6):
1. Move the unit to the edge of the work surface.
The front leveler feet should be approximately
12 mm/0.5˝ from the edge.
2. Lift the front left corner of the unit upward and
remove the drain tube 2 from the clip located on
the underside of the unit.
3. Pull the drain tube outward so the free end can be
positioned over a water container.
4. Fill the reservoir with steam-process distilled
water or VistaPure deionized water.
5. Remove the plug 3 from the end of the drain tube and allow water to drain from the tube into a
container for 30 seconds. When the water flows in a steady stream, replace the stopper.
6. Lift the front left corner of the unit upward and reinsert the tube into the clip on the underside of the
unit. Push the excess length of tubing back into the space provided.
Make sure the plug in the drain tube is secured.
After installation, and before sterilizing any instruments, run two Wrapped cycles. For further
instructions, see Section 5.1 Preparing the Unit for Use.
3.6 Shipping the Unit
Before you move the unit, you will need to drain the reservoir. To do so, follow these steps:
1. Place a water container below the unit.
2. Using the drain tube (see Section 3.5 Priming the Pump, Figure 6) empty the contents of the reservoir
into the water container.
3. Remove any remaining water from the reservoir with a non-linting, absorbent towel.
4. Screw-in the three leveler feet found underneath the unit.
5. Repack the unit in the original packing materials and include all accessories originally included with
the unit.
6. Specify heated and insured shipping.
1
Figure 5
Figure 6
3
2
95-113364 US EN R4 STATIM 2000 G4 5000 G4 OpMan.indd 13
2017-01-18 11:01 AM
3 Setting up your STATIM
3.7 Setting the Time
1.
2. Scroll to
Time
and select.
3. From the TIME screen, use the keypad to set the time. Press
EN to save and
to return to the
Setup menu.
4. To change your unit to display 12-hour time format (24-hour time format is the default setting), go
to the Setup menu and use
to scroll to TIME 12/24, select it and toggle to 12. Press
to
save and return to the Setup menu.
5. To activate daylight savings time (DST), which is recommended if you have connected your unit to a
network, go to the Set up menu and use
to scroll to DST ON/OFF and select. Use
to
toggle DST ON or OFF and press
to save and return to the Setup menu.
3.8 Setting the Date
1.
2. Scroll to
Date
and select.
3. From the DATE screen, use the keypad to set the date. Press
EN to save and
to return to the
Setup menu.
4. To change your format in which the date appears, return to the Setup menu and use
to scroll
to DATE FORMAT. Select it, and follow the prompts to have the date displayed in the desired format.
Press
to save and return to the Setup menu.
3.9 Setting the Language
The messages displayed by your STATIM can be presented in a number of different languages.
To change the current language, follow these steps:
1.
2. Scroll to
Language
and select.
3. From the Language screen, press
to scroll through the list of languages. When you
have found the desired language, press
to save your selection and return to the Setup menu.
95-113364 US EN R4 STATIM 2000 G4 5000 G4 OpMan.indd 14
2017-01-18 11:01 AM
3 Setting up your STATIM
3.10 Assigning Unit Identifier Number
1.
2. Scroll to
Unit No
and select.
3. Using the keypad, select a maximum of 3 digits to be used as the unit’s identifier number. Press
EN
to save and
to return to the Setup menu.
3.11 Creating a User ID and PIN
1.
2. Scroll to
User
and select.
3. From the SETUP PIN screen, you can assign up to four PINs. Select one of the User icons to assign
a PIN.
4. Using the keypad, assign a PIN of up to four digits and press
EN to save and
to move to
the confirmation screen.
5. If all of the information presented in the confirmation screen is correct, press OK to be returned to
the PIN USER screen. To make a correction, select the PIN User you want to change and repeat
the process described above.
3.12 Setting Up Process Enforced Usage
When process enforced usage is activated, users are required to enter a PIN both at the beginning and
at the end of a cycle. For Process Enforced usage to function, User IDs and PINs must first be assigned.
To set up User ID and PINs, refer to Section Creating a User ID and PIN. To activate Process Enforced
usage, follow these steps:
1.
2. Scroll to
Process Enforced and select.
3. Use
to toggle Process Enforced function ON or OFF. Press
to save your selection and
return to the Setup menu.
NOTE: Any user can stop a cycle and remove the cassette even with Process Enforced usage ON.
However, the cycle data will record that an unauthorized user has stopped the cycle and/or removed
the cassette.
Setup PIN
PIN Setup
1
2
3
4
5
6
7
8
9
CL
0
EN
0000
User x
PIN Updated
New PIN xxxx
OK
95-113364 US EN R4 STATIM 2000 G4 5000 G4 OpMan.indd 15
2017-01-18 11:01 AM
3 Setting up your STATIM
3.13 Changing the Touchscreen Display Themes
Your STATIM G4 touchscreen themes (i.e. the colours of the icons and background) can be changed to
one of the preset options or you can upload additional themes, as they become available from SciCan,
using the USB port. To change themes follow these steps:
1.
2. Scroll to
Theme
and select.
3. From here, you can either select
Change Theme for a menu of preloaded themes or
Upgrade Theme to access a new theme that can be loaded using the USB port.
4. In the
Change Theme screen,
use to scroll through your available options. As you scroll,
each theme will display on the touchscreen. Press
to select your theme and return to the Setup
menu.
5. To upgrade a theme available from SciCan, download the theme onto your computer’s desktop and
save the files onto a portable USB storage device. Insert the device into your STATIM’s USB port and,
from the UPGRADE THEME screen, press
Upgrade
.
5.1. The unit will load the files directly from the USB storage device. Do not remove the USB storage
device while files are loading (this could take as long as 10 minutes). When it is complete, the
screen will display the ‘Done’ message. This new theme will now be accessible from your THEMES
menu.
5.2. Press
to select this theme and return to the Setup screen.
3.14 Adjusting the Screensaver Delay
To change the length of time before your inactive touchscreen activates the screensaver,
follow these steps:
1.
2. Scroll to
Screensaver
and select.
3. Use
to scroll through you time options. When you have found the amount of time you require,
press it. Press
to save and return to the Setup menu.
3.15 Adjusting the Screen Contrast
The STATIM G4 touchscreens are calibrated for the lighting condition of most sterilization centres. Should
you need to adjust the contrast for your office, follow these steps:
1.
2. Scroll to
LCD Contrast
and select.
3. Use
to scroll through your contrast options. When you have found the contrast you require,
press it. Press
to save and return to the Setup menu.
95-113364 US EN R4 STATIM 2000 G4 5000 G4 OpMan.indd 16
2017-01-18 11:01 AM
3 Setting up your STATIM
3.16 Turning the Button Sound ON or OFF
The STATIM G4 is preset to beep when a button is pressed. If you would like to turn the button sound off,
follow these steps:
NOTE: Turning OFF the button sound does NOT turn off other alarms and cycle notification beeps.
1.
2. Scroll to
Beep ON/OFF
and select.
3. Use
to scroll through your ON or OFF options and select it by pressing it. Press
to save
and move back to the Setup menu.
3.17 Adjusting the Button Beep Volume
If you would like to adjust the beep volume, follow these steps:
1.
2. Scroll to
Beep Volume
and select.
3. Use
to scroll through the volume settings. Select the one you want by pressing it. Press
to save and move back to the Setup menu.
3.18 Setting up and Using your STATIM G4 Web Portal
The STATIM G4 web portal is a direct connection to the STATIM on your local area network. It is
protected by your firewall and not accessible to outside users (unless they have a Remote Access
Code. For more information, see Section 7. Retrieving a Remote Access Code). The web portal provides
real-time cycle information and archived sterilization records unique to this unit. From here, you can
print reports, set up email notification and search cycle histories. To set up your web portal follow the
instructions available on the portal’s ‘HELP’ tab.
95-113364 US EN R4 STATIM 2000 G4 5000 G4 OpMan.indd 17
2017-01-18 11:01 AM
3 Setting up your STATIM
3.19 Connecting to a Network
The STATIM G4 has a 10/100Base-T Ethernet port located at the back of the unit. To connect your
STATIM to a network using a router, follow these steps:
1. Connect your network cable to the Ethernet port at the back of the unit. If your office uses a router,
the router should automatically assign the unit an IP address. You will know the IP address has been
assigned when the red X over the Network icon disappears.
NOTE: In some circumstances, where you do not have a router, for example when using Windows
Network Sharing, you may have to assign a dedicated or ‘static’ IP address. To assign a static IP
address, contact your local network administrator.
2
From the main screen, press the Network icon. The Network screen displays information about your
STATIM’s connectivity, including its IP address.
3. Type the IP address displayed on the touchscreen into the browser of any web enabled device to
access your unit’s web portal. You will know that an IP address has been assigned when the red X
over the Network icon disappears. When the Network icon is active (for example when sending email)
it will turn green.
NOTE: Use QR code if connecting to a mobile device.
NOTE: Connection time will vary depending on your network speed, and making an initial connection can
take longer.
3.20 Connecting to a Wireless Network
STATIM G4 models offer Wireless capability for connecting units to the G4 network, adding flexibility
to the Network Cable connection that is also available with STATIM G4 units. STATIM G4 WiFi offers a
simple set-up and secure connection to the G4 network.
1. From the main screen, press the Network Icon.
2. Select WiFi, choose
your network and enter
your password.
NOTE: Your connection
preference can be
changed at any time.
mm/dd/yyyy hh:mm:ss
SELECT A CYCLE
95-113364 US EN R4 STATIM 2000 G4 5000 G4 OpMan.indd 18
2017-01-18 11:01 AM
3 Setting up your STATIM
3.21 Reserving an IP Address for your STATIM
When your STATIM is connected to a router on your network, it is assigned a unique IP address. If
the connection between your STATIM and the router is lost (e.g. power failure, STATIM or router was
restarted), the re-established connection may be a different IP address. This may cause any previously
saved bookmarks or other saved links to become invalid.
To set a ‘permanent’ IP address (also known as establishing a DHCP/IP/Address Reservation), follow
these steps:
1. Select the Network icon From the STATIM G4 touchscreen and write down the numbers found for the
GATE (Router Local IP). Enter this GATE number in the address bar of your web browser to access
your router settings. (NOTE: You will need your network password to change the settings.)
NOTE: Accessing this screen will vary for each router, please consult your router’s User Manual or
contact your network administrator for more detailed instructions. SciCan currently recommends D-Link
DIR-615 Wireless N router.
2. Use the ‘DHCP/IP/Address Reservation’ function to assign your STATIM a ‘permanent’ IP address.
(NOTE: Each manufacturer may use different nomenclature; this setting may be called: HCP
Reservation, IP Reservation, or Address Reservation.)
3. You will have to select the appropriate device to apply this setting. Your unit’s default name will be:
‘statim_###’.
IMPORTANT: Please contact your network administrator for more detailed instructions.
mm/dd/yyyy hh:mm:ss
SELECT A CYCLE
95-113364 US EN R4 STATIM 2000 G4 5000 G4 OpMan.indd 19
2017-01-18 11:01 AM
4 Using Cassettes and Preparing Instruments
4.1 Using the STATIM 2000 G4 Cassette
When removing the cassette after a cycle, exercise caution as the metal areas will be hot and the
cassette may contain hot steam.
Opening the Cassette:
1. Hold the cassette handle with your thumbs facing inward on
the cassette latch.
2. Push downward on the cassette latch.
3. Raise the cassette lid upwards and disengage the hinge.
4. Rest the lid on its outer surface.
Closing the Cassette:
1. Align the hinge tab on the cassette lid with the hinge slot on the rear of the bottom tray.
2. As you begin to close the lid, the hinge tab and slot will engage.
•
Inserting the Cassette into the STATIM 2000 G4:
1. Place the end of the cassette into the unit.
2. Gently push inward until you hear a “click” sound and check the LCD icon for change
from
to
.
Never force the cassette into the STATIM as the interior components could be damaged.
NOTE: The main menu screen will display
if the cassette is not properly inserted in the unit.
Removing the Cassette:
1. Grasp the handle with two hands and pull away from the unit.
2. Pull the cassette clear of the unit and set down on a firm surface.
•
Disengaging the Cassette:
When not in use, the cassette should be disengaged. To disengage the cassette, grasp the handle
and pull the cassette out until there is a 15 mm to 20 mm (½ to ¾”) gap between the front of the
STATIM 2000 G4 and the cassette handle.
Figure 7
cassette latch
95-113364 US EN R4 STATIM 2000 G4 5000 G4 OpMan.indd 20
2017-01-18 11:01 AM
4 Using Cassettes and Preparing Instruments
4.2 Using the STATIM 5000 G4 Cassette
When removing the cassette after a cycle, exercise caution as the metal areas will be hot and the
cassette may contain hot steam.
Opening the Cassette:
1. Push the carry handle 2 into the open position.
2. Put your hands on either side of the cassette
handle 1 .
3. Insert your forefingers in the slots and place your
thumbs on the thumb pads.
4. Press down with your thumbs and pull up with your
forefingers until the lid opens.
5. Raise the cassette lid and disengage from the tray.
Rest the lid on its outer surface.
Closing the Cassette:
1. Align the hinge tab on the lid with the hinge slot on
the tray.
2. As you begin to close the lid, the hinge tab and slot
will engage.
3. Place the carry handle 2 into the closed position.
•
Inserting the Cassette into the STATIM 5000 G4:
1. Hold the cassette handle in one hand and the carry
handle in the other as shown in Figure 8.
2. Place the end of the cassette into the unit and drop
the carry handle into its closed position.
3. Gently push inward until you hear a “click” sound and
check the LCD icon for change from
to
.
Never force the cassette into the STATIM as the interior components could be damaged.
NOTE: The main menu screen will display
if the cassette is not properly inserted
in the unit.
n1
Figure 8
1
2
2
95-113364 US EN R4 STATIM 2000 G4 5000 G4 OpMan.indd 21
2017-01-18 11:01 AM
4 Using Cassettes and Preparing Instruments
Removing the Cassette:
1. Grasp the cassette handle with one hand and pull out from the unit.
2. As the cassette emerges from the unit, grasp the carry handle with your free hand and lift it upwards.
3. Pull the cassette clear from the unit and set down on a firm surface.
Disengaging the cassette
When not in use, the cassette should be disengaged. To disengage the cassette, grasp the handle
and pull the cassette out until there is a 15 mm to 20 mm (1/2 to 3/4”) gap between the front of the
STATIM 5000 G4 and the cassette handle.
4.3 Preparing and Loading Instruments
Before loading any instruments into the STATIM, consult the manufacturer’s reprocessing instructions.
Clean Instruments
Clean and rinse all instruments before loading them into the cassette. Disinfectant residues and solid
debris may inhibit sterilization and damage the instruments, the cassette, and the STATIM. Lubricated
instruments must be wiped thoroughly and any excess lubricant should be removed before loading.
Unwrapped Instruments
Arrange unwrapped instruments on the rack in the tray so that they do not touch one another. This
ensures that steam reaches all surfaces and will promote drying.
Instruments must not be stacked or piled in the cassettes, as this will impede the sterilization process.
95-113364 US EN R4 STATIM 2000 G4 5000 G4 OpMan.indd 22
2017-01-18 11:01 AM
4 Using Cassettes and Preparing Instruments
Wrapped Instruments (single wraps)
Place the instruments into single layer autoclave bags according to the manufacturer’s instructions.
Orient the instrument rack in the cassette to ensure that wrapped instruments rest approximately 6 mm
/ 0.25” above the cassette base. Place the wrapped instruments on the rack and arrange them to avoid
overlap. Ensure that all wrapped loads are dry before handling and /or storage to maintain sterility.
The use of cloth wraps in the STATIM is not recommended. SciCan recommends the use of paper /
paper and plastic / paper autoclave pouches. Pouches that have been cleared by the FDA for use
with the respective cycle time and temperature of the STATIM 2000 / 5000 G4 Cassette Autoclaves
must be used. The use of cloth wraps in the STATIM is not recommended. Loosely pack instruments
in the bags to allow steam penetration to all instrument surfaces. Care must be taken to ensure that
the combined weight of the loaded bags does not exceed 1.5 kg (3.3 lbs) for STATIM 5000 G4 or
1.0 kg (2.2 lbs) for STATIM 2000 G4
For STATIM 2000 G4:
For STATIM 5000 G4:
The rack will hold 12 autoclave bags.
Rubber and Plastic Instruments
The following materials can be sterilized in the STATIM:
•
Nylon
•
Polycarbonate (Lexan™)
•
Polypropylene, PTFE (Teflon™)
•
Acetal (Delrin™)
•
Polysulfone (Udel™)
•
Polyetherimide (Ultem™)
•
Silicone
•
Rubber
•
Polyester
When loading rubber and plastic instruments in the tray, leave a space between the instruments and
the cassette walls. This ensures that steam reaches all surfaces, and will promote drying.
95-113364 US EN R4 STATIM 2000 G4 5000 G4 OpMan.indd 23
2017-01-18 11:01 AM
4 Using Cassettes and Preparing Instruments
The following materials CANNOT be sterilized in the STATIM:
•
Polyethylene
•
ABS
•
Styrene
•
Cellulosics
•
PVC
•
Acrylic (Plexiglas™)
•
PPO (Noryl™)
•
Latex
•
Neoprene and similar materials.
Use of these materials may lead to instrument or equipment damage. If you are unsure of your
instrument’s material or construction, do not load into your STATIM until you have checked with the
instrument manufacturer.
All Instruments
The STATIM is NOT intended for sterilizing textiles, liquids or biomedical waste. Instruments will
remain sterile after a successful cycle until the cassette is disengaged from the unit. Unwrapped
instruments, once exposed to ambient or external conditions, cannot be maintained in a sterile state.
If sterile storage is desired, wrap the instruments to be sterilized in autoclave bags, according to
the instrument manufacturer’s instructions, and then allow the wrapped cycle to run until the air-dry
phase is complete.
Best Practice: Allow instruments (wrapped or unwrapped) to dry completely prior to handling.
Wrapped or pouched instruments must not touch each other to promote drying and enable effective
sterilization.
SciCan recommends the final user carefully choose the most appropriate sterilization cycle according
to the recommendations of their leading infection control authorities and local regulatory guidelines /
recommendations.
Routine Monitoring
Chemical process indicators suitable for steam sterilizers should be included in or on each package or
load being sterilized. In addition, the weekly use of biological indicators, which allow you to ascertain
whether the instruments have been exposed to sterilization condition, is recommended. Chemical and
biological indicators that have been cleared by the FDA, for use with the respective cycle time and
temperature of the STATIM 2000 / 5000 G4 Cassette Autoclaves must be used.
Note for Ophthalmology Use
In the field of ophthalmology, proper wrapping or pouching of surgical instruments will reduce the
exposure of instruments to any process residues during the sterilization cycle. Due to the highly sensitive
nature of some types of surgery (particularly in ophthalmology), SciCan recommends that all instruments
be routinely packaged or wrapped and processed through the wrapped cycle of the sterilizer.
This practice is the suggested approach for the majority of sterile surgical procedures and is referenced
in most leading infection control publications and guidelines.
95-113364 US EN R4 STATIM 2000 G4 5000 G4 OpMan.indd 24
2017-01-18 11:01 AM
Using Cassettes and Preparing Instruments
4.4 Using Biological and Chemical Indicators
For detailed instructions on how to handle, use and dispose of both the biological and chemical
indicators, please consult the product literature accompanying the indicators or contact the manufacturer
directly.
To use the indicators with the STATIM, follow these steps:
1. Place the appropriate biological indicator in the STATIM chamber.
2. Process the load in the sterilizer according to your usual practice Ensure that the message “Cycle
Complete” is displayed on the LCD after the cycle is finished.
3. Recover the biological and / or chemical indicator and process further according to the literature that
accompanied the indicator.
At the first indication of a potential sterilization failure:
1. Do not process any more instruments until favourable test results have been returned.
2. Ensure the correct indicator type was chosen.
3. Ensure the cassette was not overloaded. Consult the earlier portion of this section for proper loading
instructions.
4. If the results do not change, do not process any more instruments within the STATIM and contact
your SciCan dealer for further assistance.
It is recommended that the tests be conducted so that the incubation period occurs during a period of
planned downtime such as the last cycle before a weekend.
4.5 Instrument Weight Guide
NOTE: The above weights are to be used as reference only. For exact weights of your instruments,
consult the manufacturer’s specifications.
Instrument
Typical Instrument Weight
Scissors
30 g / 0.96 oz
Dental scalers
20 g / 0.64 oz
Forceps
15 g / 0.48 oz
Dental handpiece
40 to 60 g / 1.29 to 1.92 oz
Wrapped instrument rack
260 g / 8.35 oz
Unwrapped instrument rack
225 g / 7.23 oz
Suction cannula
10 g / 0.32 oz
Plastic mouth mirror
8 g / 0.25 oz
Impression tray
15 to 45 g / 0.48 to 1.45 oz
Plastic x-ray positioning ring
20 g / 0.64 oz
95-113364 US EN R4 STATIM 2000 G4 5000 G4 OpMan.indd 25
2017-01-18 11:01 AM
5 Using your STATIM
Before using your STATIM for the first time, make sure the reservoir is full and the pump is properly
primed. Refer to Sections Filling the Reservoir and Priming the Pump for detailed instructions.
5.1 Preparing Unit for Use
Once the unit is installed and before any instruments are sterilized, run two Wrapped Cycles (see Section
STATIM 2000 / 5000 G4 - Running a Cycle. Remove the cassette once it has cooled. Clean the top (lid)
and bottom (tray) sections using a soft cloth to wipe the inside surfaces and then rinse thoroughly with
tap water.
5.2 STATIM 2000 G4 — Selecting a Cycle
The STATIM 2000 G4 has three sterilization cycles, each designed to sterilize a specific type of
instrument. The instruments will remain sterile after a successful cycle until the cassette is removed from
the autoclave. At the end of each sterilization cycle, microbiologically filtered air-drying will commence
for 60 minutes by default. Air-drying can be interrupted at any time.
Unwrapped instruments, once exposed to ambient or external conditions, cannot be maintained in
a sterile state. If sterile storage is desired, wrap the instruments to be sterilized in autoclave bags
according to the instrument manufacturer’s instructions, and allow the wrapped cycle to run until the
air-dry phase is complete.
The types of instruments, sterilization requirements, and a graph depicting each cycle are described
over the next few pages. Consult the Instrument Weight Guide for information on how to make up an
appropriate load for the masses specified for individual cycles.
5.2.1. Unwrapped Cycle
(STATIM 2000 G4)
The Unwrapped Cycle is a general purpose
sterilization cycle used to sterilize up to 1.0 kg
(2.2 lbs) of solid metal instruments such as pliers,
burrs, scalers and forceps. Dental handpieces
may be sterilized in this cycle.
To select the Unwrapped Cycle, press the Unwrapped
Cycle button, then press the START button.
The sterilization temperature in the cassette is 135˚C (273˚F) and the holding time is 3.5 minutes.
See Section 4.4 Preparing and Loading Instruments before running this cycle.
A
B
C
D
E
F
A
Warming up
B
Conditioning
C
Pressurizing
D
Sterilizing
E
Venting
F
Air Drying
95-113364 US EN R4 STATIM 2000 G4 5000 G4 OpMan.indd 26
2017-01-18 11:01 AM
5 Using your STATIM
5.2.2. Wrapped Cycle
(STATIM 2000 G4)
The Wrapped Cycle is used to sterilize up to 1.0
kg (2.2 lbs) of solid and hollow metal instruments
which have been sealed in paper / paper, or paper
/ plastic autoclave bags. Dental handpieces may
be sterilized in this cycle.
To select the Wrapped Cycle, press the Wrapped
Cycle button, then press the START button.
The sterilization temperature in the cassette is 135˚C (273˚F) and the holding time is 10 minutes.
See Section 4.4 Preparing and Loading Instruments before running this cycle.
If wrapped instruments are intended for storage, the wraps must be dry when the cassette is removed
from the unit and opened.
5.2.3. Rubber and Plastics Cycle
(STATIM 2000 G4)
The Rubber and Plastics Cycle is used to
sterilize up to 0.4 kg (0.9 lbs) of solid unwrapped
instruments constructed of metal or the
materials listed in Section Preparing and Loading
Instruments.
To select the Rubber and Plastics Cycle, press the
Rubber and Plastics Cycle button, then press the
START button.
The sterilization temperature in the cassette is 121˚C (250˚F) and the holding time is 15 minutes.
A
B
C
D
E
F
A
Warming up
B
Conditioning
C
Pressurizing
D
Sterilizing
E
Venting
F
Air Drying
A
B
C
D
E
F
A
Warming up
B
Conditioning
C
Pressurizing
D
Sterilizing
E
Venting
F
Air Drying
95-113364 US EN R4 STATIM 2000 G4 5000 G4 OpMan.indd 27
2017-01-18 11:01 AM
5 Using your STATIM
5.2.4. Air Dry Only Cycle
(STATIM 2000 G4)
This is not a sterilization cycle.
The load is considered sterile after the successful completion of the sterilization phase of the cycle.
The Air Dry phase starts automatically after each sterilizing cycle and runs for 60 minutes. Air drying
may be interrupted by pressing the STOP button anytime after the sterilization phase of the cycle
is complete. To ensure that the contents of the cassette are dry, the cycle should run for the full 60
minutes. Dryness is important for unwrapped instruments for corrosion prevention. For wrapped
instruments, a dry wrap is required to maintain sterility.
If the STOP button is pressed during the air drying stage of the sterilization cycle, and the cassette
has not been removed from the autoclave, the Air Dry Only cycle may be used to promote further
drying. If the cassette has been removed from the autoclave, it may NOT be reinserted for the Air Dry
Only cycle. If the cassette contains wrapped instruments and the wraps are not dry when the cassette
is opened, the instruments must be handled in an aseptic manner for immediate use or resterilized.
To start, press the Air Dry Only cycle button, then press the START button. User will be prompted to
confirm cycle selection.
When started independently this cycle will run for 1 hour.
NOTE: Sterilized instruments should only be handled once they are dry. Drying times can vary depending
on the weight of load. If best practices are exercised (refer to Sections ‘Preparing and Loading
Instruments’ and ‘Maintenance’) and the load is less than the maximum capacity, instruments may be dry
in less than 60 minutes.
95-113364 US EN R4 STATIM 2000 G4 5000 G4 OpMan.indd 28
2017-01-18 11:01 AM
5 Using your STATIM
Before using your STATIM for the first time, make sure the reservoir is full and the pump is properly
primed. Refer to Sections Filling the Reservoir and Priming the Pump for detailed instructions.
5.3 STATIM 5000 G4 — Selecting a Cycle
The STATIM 5000 G4 has four sterilization cycles, each designed to sterilize a specific type of
instrument. It is important not to overload the chamber as this can inhibit steam access to all instrument
surfaces. The instruments will remain sterile after a successful cycle until the cassette is removed from
the autoclave housing. At the end of each sterilization cycle, microbiologically filtered air-drying will
commence for 60 minutes by default. Air-drying can be interrupted at any time.
Once unwrapped instruments are exposed to ambient or external conditions, they cannot be maintained
in a sterile state. If sterile storage is desired, wrap the instruments to be sterilized in autoclave bags
according to the instrument manufacturer’s instructions, and allow the wrapped cycle to run until the
air-dry phase is complete.
The types of instruments, sterilization requirements, and graphs depicting each cycle are described on
the next few pages. Consult the Instrument Weight Table in Section 4.6 for information on how to make
up an appropriate load for the masses specified for individual cycles.
5.3.1. Unwrapped Cycle
(STATIM 5000 G4)
The Unwrapped Cycle is used to sterilize light
loads (less than 0.5 kg or 1.1 lbs) of solid metal
instruments such as pliers, burrs, scalers, and
forceps.
To select the Unwrapped Cycle, press the Unwrapped
Cycle button, then press the START button.
The sterilization temperature in the cassette is 132˚ C (270˚ F) and the holding time is 3.5 minutes.
See Section 4.4 Preparing and Loading Instruments before running this cycle.
A
B
C
D
E
F
A
Warming up
B
Conditioning
C
Pressurizing
D
Sterilizing
E
Venting
F
Air Drying
95-113364 US EN R4 STATIM 2000 G4 5000 G4 OpMan.indd 29
2017-01-18 11:01 AM
5 Using your STATIM
5.3.2. Wrapped Cycle
(STATIM 5000 G4)
The Wrapped Cycle is used to sterilize up to
1.5 kg (3.3 lbs) of solid and hollow metal
instruments which have been sealed in paper /
paper, or paper / plastic autoclave bags. Dental
hand pieces may be sterilized in this cycle.
To select the Wrapped Cycle, press the Wrapped
Cycle button, then press the START button.
The sterilization temperature in the cassette is 132˚ C (270˚ F) and the holding time is 6 minutes.
See Section 4.4 Preparing and Loading Instruments before running this cycle.
If wrapped instruments are intended for storage, the wraps must be dry when the cassette is removed
from the unit and opened.
5.3.3. Rubber and Plastics Cycle
(STATIM 5000 G4)
The Rubber and Plastics Cycle is used to sterilize
up to 0.4 kg (0.9 lbs) of solid or hollow unwrapped
instruments constructed of metal or the materials
listed in Section 4.4 Preparing and Loading
Instruments.
To select the Rubber and Plastics Cycle, press the
Rubber and Plastics Cycle button, then press the
START button.
The sterilization temperature in the cassette is 121 ˚C (250 ˚C) and for the holding time is 35 minutes.
A
B
C
D
E
F
A
Warming up
B
Conditioning
C
Pressurizing
D
Sterilizing
E
Venting
F
Air Drying
A
B
C
D
E
F
A
Warming up
B
Conditioning
C
Pressurizing
D
Sterilizing
E
Venting
F
Air Drying
95-113364 US EN R4 STATIM 2000 G4 5000 G4 OpMan.indd 30
2017-01-18 11:01 AM
5 Using your STATIM
5.3.4. Heavy Duty Unwrapped Cycle
(STATIM 5000 G4)
The Heavy Duty Cycle is used to sterilize larger
loads of unwrapped metal instruments weighing
up to 1.5 kg (3.3 lbs). Dental hand pieces can be
sterilized in this cycle. Using the STATIM 5000
Extended Cassette (order no. 01-112509S) will
allow for the sterilization of rigid endoscopes in
this cycle.
To select the Heavy Duty Unwrapped Cycle, press the Heavy Duty Unwrapped Cycle button, then press
the START button.
The sterilization temperature in the cassette is 132˚ C (270˚ F) and the holding time is 6 minutes.
HEAVY DUTY UNWRAPPED
Press START
HEAVY DUTY UNWRAPPED
Press START
A
B
C
D
E
F
A
Warming up
B
Conditioning
C
Presurizing
D
Sterilizing
E
Venting
F
Air Drying
95-113364 US EN R4 STATIM 2000 G4 5000 G4 OpMan.indd 31
2017-01-18 11:01 AM
5 Using your STATIM
5.4 STATIM 2000 / 5000 G4 — Running a Cycle
To operate each cycle, follow these steps.
1. Turn the power switch at the back of the unit to ON.
At start up, the unit will display the main menu.
2. Press the appropriate cycle button on the touchscreen. The display will show the cycle name and
parameters.
3. Press the START icon.
NOTE: If Process Enforced usage is enabled, a PIN screen will appear after you have pressed START.
Enter your PIN to start the cycle.
mm/dd/yyyy hh:mm:ss
SELECT A CYCLE
95-113364 US EN R4 STATIM 2000 G4 5000 G4 OpMan.indd 32
2017-01-18 11:01 AM
5 Using your STATIM
If the unit is not properly levelled, a warning screen is displayed advising to level the unit to attain optimal
instrument dryness. Once the bubble level indicates a proper level, press Continue to start the cycle.
95-113364 US EN R4 STATIM 2000 G4 5000 G4 OpMan.indd 33
2017-01-18 11:01 AM
5 Using your STATIM
5.4 STATIM 2000 / 5000 G4 — Running a Cycle
When the cycle starts, the cycle parameters are displayed at the top of the screen. Below it is the current
phase. The unit’s cycle counter is displayed at right. A graph charts the progress of the cycle while
current cycle information is displayed at right. As the cycle is running, various sounds will be heard. This
is the normal functioning of the unit.
Example of sterilization cycle phases in STATIM 2000 G4:
95-113364 US EN R4 STATIM 2000 G4 5000 G4 OpMan.indd 34
2017-01-18 11:01 AM
5 Using your STATIM
5.4 STATIM 2000 / 5000 G4 — Running a Cycle
When the cycle is running, an intermittent buzzing sound will be heard as the pump injects water into the
steam generator. A random clicking sound will also be heard as the exhaust valve opens and closes.
Once the sterilization cycle is complete, a reminder tone will sound as the cycle moves into the Air Drying
phase.
The buzzing noise during the air drying stage is the compressor operating. The air drying phase of the
cycle may be interrupted at any time by pressing the STOP button. The display will then read:
To ensure that wrapped instruments are dry, allow the cycle to run to completion.
If the cassette contains wrapped instruments and the wraps are not dry when the cassette is open,
the instruments must be handled in an aseptic manner for immediate use or resterilized.
When the automatic 60-minute drying stage is finished the display reads:
By default, if the sterilization cycle is successful, the reminder tone sounds for 30 seconds or until the
STOP button is pressed, or the cassette is removed from the unit.
Be careful. The metal parts will be hot, and the cassette will contain hot steam. After the cassette is
removed from the unit, it should be opened to hasten drying of unwrapped instruments.
95-113364 US EN R4 STATIM 2000 G4 5000 G4 OpMan.indd 35
2017-01-18 11:01 AM
5 Using your STATIM
If the screen displays a CYCLE FAULT code or a NOT STERILE message, the cassette contents are not
sterile. See Section 10 Troubleshooting for more information.
5.5 STATIM 2000 / 5000 G4 — Stopping a Cycle
To stop a cycle press the STOP button. If the STOP button is pushed, the cassette is removed, or the unit
detects a problem while operating, the cycle will stop. Once a cycle has been stopped, the STOP button
must be pressed before another cycle can be started. The display reads any of the following messages:
If the display shows the message, CYCLE FAULT or NOT STERILE, the cassette contents are not sterile!
See Section 10 Troubleshooting for more information.
If the cassette contains wrapped instruments and the wraps are not dry when the cassette is open, the
instruments must be handled in an aseptic manner for immediate use and should not be stored.
or
95-113364 US EN R4 STATIM 2000 G4 5000 G4 OpMan.indd 36
2017-01-18 11:01 AM
6 Storing and Retrieving Cycle Information
The STATIM G4 has an internal Data Logger capable of storing all cycle data on every cycle, whether
successful or incomplete, for the lifetime of the unit. You can access this information through the
touchscreen, through the web portal, using a USB storage device or by attaching a printer.
6.1 Retrieving Cycle Information Using the Touchscreen
1. From the main menu, press
.
2. The unit will record the last five successful cycles and the last five incomplete cycles. If you select a
cycle from the list, it will display cycle information in a format similar to how it would be printed.
3. Use the arrow keys to scroll through and read.
NOTE: Regardless of whether you have a USB storage device attached to the unit or not, you can always
see the last five successful cycles and the last five incomplete cycles. Use the STATIM G4 web portal to
access all the cycle information stored on your STATIM from your computer. To connect your STATIM to
a network consult the separate manual titled STATIM 2000/5000 G4 – Setting Up and Using Your Web
Portal.
6.2 Retrieving Cycle Information Using the USB Data Back Up
The USB storage device can be used to transfer cycle information stored in the unit to a computer. Best
practice suggests this should be done once a week. To transfer data using the USB port, follow these
steps:
1. Plug the USB storage device into the USB port.
2. The STATIM keeps track of what data has already been transferred to the USB storage device and will
automatically load only new data.
3. When the activity light on the provided USB storage device stops blinking or the USB icon on the
LCD turns from a flashing green to a solid grey, remove the USB storage device and transfer the
information to your computer.
NOTE: If you select the USB storage device icon from the main menu, you will only be able to view the
last five complete cycles and the last five incomplete cycles. To view all the cycles stored on the USB
storage device, you must use your computer.
mm/dd/yyyy hh:mm:ss
SELECT A CYCLE
Yyyy/mm/dd xxxxx OK
Yyyy/mm/dd xxxxx OK
Yyyy/mm/dd xxxxx OK
Yyyy/mm/dd xxxxx OK
Yyyy/mm/dd xxxxx OK
Yyyy/mm/dd xxxxx (CFxx)
Yyyy/mm/dd xxxxx (CFxx)
Yyyy/mm/dd xxxxx (CFxx)
Yyyy/mm/dd xxxxx (CFxx)
Yyyy/mm/dd xxxxx (CFxx)
95-113364 US EN R4 STATIM 2000 G4 5000 G4 OpMan.indd 37
2017-01-18 11:01 AM
6 Storing and Retrieving Cycle Information
6.3 Cycle Printout Overview
Model: STATIM 2000 software: S203R604
STATIM 2000
S203R604
Unit Identifier: autoclave has been set up as number 000
UNIT #:
000
Water Quality: conductivity measurements
WATER QUALITY
0.5 µS /0.3 ppm
Cycle Counter: the number of cycles having been run on the unit = 2
CYCLE NUMBER
000002
Time / Date: 10:47 am 25th October 2007
10:47
10/25/2007
Cycle Name: UNWRAPPED
UNWRAPPED
Cycle Name cont’d - parameters: 135°C / 3.5 min.
135°C FOR 3.5 MIN.
Cycle Clock: starting at 0:00
CYCLE START
0:00
Warm up complete: start of the conditioning phase is 1:19 (see cycle
graph - ‘A’ phase complete, start of ‘B’ phase)
CONDITIONING
1:19
Start Time of Pressurization: 1:27 (start of ‘C’ phase)
PRESSURIZING
1:27
Start Time of Sterilization: 2:27 (start of ‘D’ phase)
STERILIZING
2:27
Temp. / Press. & Time at start of sterilization (‘D’ phase)
136.4ºC 220KPA
2:27
Temp. / Press. & Time printed at 30 second intervals during
sterilization. (“D” phase)
136.0ºC 219KPA
135.9ºC 222KPA
136.1ºC 222KPA
136.5ºC 225KPA
136.4ºC 225KPA
136.1ºC 221KPA
2:57
3:27
3:57
4:27
4:57
5:27
Temp. / Press. & Time of end of sterilization phase (end of ‘D’ phase)
136.1ºC 221KPA
5:57
Time Venting started: 5:57 (start of ‘E’ phase)
VENTING
5:57
Time Air Drying started: 6:42 (start of ‘F’ phase)
AIR DRYING
6:42
Cycle completion time: 22:42
CYCLE COMPLETE
22:42
95-113364 US EN R4 STATIM 2000 G4 5000 G4 OpMan.indd 38
2017-01-18 11:01 AM
7 Retrieving a Remote Access Code
To allow access to external persons to gain remote access to the STATIM G4, you will need to provide a
security token to the person requesting access. To obtain this code follow these steps:
1.
2. Scroll to
Remote Access and select.
3. Press
Enable
and wait a few seconds for a security token to be provided. Once displayed,
provide the token to the person requiring Remote Access.
NOTE: This token will be valid for only 2 hours, at which time the remote session will automatically
disconnect. Should you wish to end the session earlier, select
Disable
from the same menu
to disable the token.
95-113364 US EN R4 STATIM 2000 G4 5000 G4 OpMan.indd 39
2017-01-18 11:01 AM
Printers by SciCan
End of Line CR/LF
Serial Port Bitrate
Printer user ° char
Epson
TM-U220D (C31C515603)
CR/LF
9600
248 [0xF8]
Citizen
IDP-3110-40 RF 120B
CR
9600
N / A
Star Micro
SP212FD42-120
CR
9600
210 [0xd2]
Star Micro
SP216FD41-120
CR/LF
9600
210 [0xd2]
Star Micro
SP512MD42-R
CR/LF
9600
210 [0xd2]
8 Printing Cycle Information
The STATIM G4 is equipped with an RS232 serial port to allow you to connect it to an external printer.
(For a list of recommended printers, see the table below.)
8.1 Connecting to a Printer
To connect the printer, follow these steps:
1. Connect the external printer to the STATIM G4’s RS232 port using the serial printer cable supplied
with your printer.
2. Power on the printer.
3. On the STATIM G4, select
4. Scroll to
Printer Type
and select.
5. Use
to toggle to the Serial Printer option and select. Press
to save and move back to
the Setup menu.
8.2 Adjusting your Printer Settings
The STATIM G4 allows for several printer adjustments. You can access these settings from the user setup
menu (see instructions above). Use the table below or your printer’s operator manual to make the correct
adjustments to your printer’s
Baud Rate
,
End of Line CR/LF and
Printer User Char .
8.3 External Printers and Specifications
95-113364 US EN R4 STATIM 2000 G4 5000 G4 OpMan.indd 40
2017-01-18 11:01 AM
9 Maintaining your STATIM
9.1 Cleaning the Cassette
Keeping the STATIM cassette clean is good clinical practice and assists in the function of the unit. SciCan
recommends that the interior surface be cleaned at least once a week. Use dishwashing soap or a mild
detergent that does not contain chlorine. Scrub the inside of the cassette with a cleaning pad designed
for use with Teflon™ coated surfaces. After scouring, rinse thoroughly with water to remove all traces of
the detergent. Cleaning the inside of your cassette is very important if you regularly sterilize lubricated
instruments.
9.2 Cleaning the Water Reservoir Filter
The water reservoir filter should be cleaned at least once a week or when required. The filter can easily be
removed and cleaned by placing the filter upside down under running water to wash away the particles
until clean, and then placed back into the reservoir opening. If a replacement water reservoir filter is
required, order part number 01-109300S.
9.3 Cleaning the Reservoir
Check the inside of the reservoir for dirt or particles. To clean the reservoir, drain it, remove any visible
particles, rinse with steam process distilled water or VistaPure deionized water ONLY and drain it
again. Do not use chemicals or cleaning agents as these can cause damage to the unit .
9.4 Cleaning the Exterior Surfaces
Use a soft cloth moistened with soap and water to clean all exterior surfaces. Do not use harsh cleaning
chemicals or disinfectants.
95-113364 US EN R4 STATIM 2000 G4 5000 G4 OpMan.indd 41
2017-01-18 11:01 AM
9 Maintaining your STATIM
9.5 Changing the Bacteria Retentive Filter and the Air Filter
The filters should be replaced every six months or after 500 cycles to maintain an adequate supply of
clean air during the air drying cycle.
To change the bacteria retentive air filter on the STATIM 2000 G4 and 5000 G4, follow these steps:
1. Power the STATIM OFF.
2. Disconnect tube A 1 from the bacteria retentive filter 2 and
remove the filter from the filter bracket 3 . As you remove the
filter from the bracket, note the orientation of the arrow mark on
the filter.
3. When the filter is free of the bracket, carefully disconnect tube
B 4 from the filter.
4. Before installing the replacement bacteria retentive filter 2
(SciCan order no. 01-102119S) check that the arrow mark on the
filter matches the direction of the arrow on the bracket. Push the
left hand filter fitting into tube B 4 .
5. Gently press the replacement filter into the filter
bracket 3 . The arrow mark of the filter should be facing
out and pointing to the left.
6. Re-connect tube A 1 to the right hand filter fitting.
To change the air filter on the STATIM 2000 G4, follow
these steps:
1. Turn the power switch at the back of the unit OFF.
2. Remove and discard the old foam air filter 7 .
3. Install the new filter (SciCan part no. 01-100207S).
4. Secure the filter plate 6 to the back of the compressor using the screw 5 retained during the
disassembly procedure.
To change the air filter on the STATIM 5000 G4, follow these steps:
1. Unscrew the cylindrical air filter 8 in a counter-clockwise direction.
2. Discard the old filter.
3. Screw the new filter (SciCan part no. 01-101652S), finger tight only, into place.
Figure 11
5000 G4
1
8
4
2
3
Figure 10
2000 G4
5
7
6
1
4
2
3
95-113364 US EN R4 STATIM 2000 G4 5000 G4 OpMan.indd 42
2017-01-18 11:01 AM
9 Maintaining your STATIM
9.6 Replacing the Cassette Seal
To ensure optimum performance of your STATIM cassette autoclave, change the cassette seal every 500
cycles or every six months, whichever comes first. Replacement seals are available from SciCan (order
number 01-100028S for STATIM 2000 G4 and 01-101649S for STATIM 5000 G4).
To change the cassette seal, follow these steps:
Place the cassette lid and the new seal on a clean work surface. Examine
the position of the old seal in the cassette lid and arrange the new seal in
the same orientation, next to the lid.
Remove the old seal and discard. Clean any residue out of the
seal channel and flush out the channel with steam-distilled water
or VistaPure deionized water.
Lubricate the new seal with the liquid seal lubricant provided.
Insert the rounded edge of the seal under the round lip of
the lid. Align the holes in the new seal with the holes in the lid.
NOTE: At every corner and at the holes in the lid, two square nibs
should be visible. The nibs should fit flush with the lid’s outer surface.
Ensure the seal is completely inserted. Feel around
the periphery to ensure the seal is securely in place.
NOTE: During a cycle, steam may appear between the lid
and the tray. If this persists, remove the cassette and check
that the seal is correctly installed.
Be careful. The metal parts will be hot, and the cassette may contain hot steam.
Figure 13
95-113364 US EN R4 STATIM 2000 G4 5000 G4 OpMan.indd 43
2017-01-18 11:01 AM
9 Maintaining your STATIM
9.7 Maintaining Fluid Levels
1. The water reservoir level is continually monitored by your STATIM. If the reservoir is low, a red X will
appear on the water icon
of the cycle select screen. Press the icon to go to the next screen to
confirm that it is a water level issue and not a water quality issue.
2. If the reservoir is low, a
will appear next to WATER LEVEL.
3. To fill the reservoir, use steam-processed distilled water or VistaPure deionized water containing less
than 5 ppm total dissolved solids (having conductivity of less than 10 μS / cm). Remove the cap from
the top of the unit and fill the reservoir. We recommend using a funnel to minimize spills. Each time
you refill the reservoir, empty the waste bottle and refill with water to the MIN line. Empty the waste
bottle often to avoid unpleasant odors and discoloration of the contents. (A low-level chlorine-free
disinfectant, prepared according to the manufacturer’s instructions, may be added to the waste bottle
to remedy this situation).
9.8 Reading Water Quality
1. The water quality is continually monitored by your STATIM to ensure that the optimal quality of water
is being used. If the water quality is beyond the tolerances specified above, a red X will appear on the
water icon
of the cycle select screen. Press the icon to go to the next screen to confirm that it is a
water quality issue and not a water level issue.
2. If the water quality not suitable,
will appear next to micro S. and parts per million values.
3. Using the drain tube (see Section Priming the Pump, Figure 6) empty the contents of the reservoir
into the water container and replace with steam-process distilled water or VistaPure deionized water
containing less than 5 ppm total dissolved solids (having conductivity of less than 10 μS / cm).
9.9 Using On-Screen Instructions
The STATIM G4 on-screen help instruction can be viewed by following these steps:
1.
2. Scroll to
Instructions
and select.
3. Choose the Instructions you would like to view and select.
95-113364 US EN R4 STATIM 2000 G4 5000 G4 OpMan.indd 44
2017-01-18 11:01 AM
9 Maintaining your STATIM
9.10 Preventative Maintenance Schedules
To ensure trouble-free performance, both the operator and the dealer must follow a preventative
maintenance schedule.
NOTE: Please refer to your National, Regional, State or Safety laws for any additional reoccurring user
testing that may be required.
The schedules below describe the necessary actions.
Operator
Daily
Water Reservoir
•
Replace water as needed.
•
For opthalmic use, drain after each working day, leave empty, and refill at
the start of the next workday.
Waste Bottle
•
Empty the waste bottle every time you refill the reservoir.
•
Fill with tap water up to MIN line marking. You may also add some
chlorine-free disinfectant.
Unit
•
To ensure optimal performance of the STATIM autoclave, SciCan
recommends that a full sterilization cycle, which must include the full
Air Drying phase, be completed at the end of each day of use. This is
especially important when the unit is left idle over the weekend or for any
extended period of time.
Weekly
Cassette
•
Wash the interior of the cassette with dishwashing soap or a mild
detergent that does not contain chlorine.
•
Scrub the inside with a cleaning pad designed for use with Teflon™-coated
surfaces.
Biological and / or
Air Filter
•
Check the filter for dirt and moisture. Replace if dirty.
Call for service if wet.
Water Filter
•
Check the water reservoir filter every week and clean if necessary.
Replace only if necessary.
Every 6
months
Cassette Seal
•
Replace every 500 cycles or six months (whichever comes first), or
whenever necessary.
Biological and / or
Air Filter
•
Replace every 500 cycles or six months (whichever comes first).
Extended Period
of NON-USE
Unit
•
If the STATIM is not going to be in use for a week (or more), ensure that
a full sterilization cycle, which must include the full air drying phase, is
completed before the period of non-use. After the period of non-use and
before any instruments are sterilized, run three wrapped cycles. Remove
the cassette once it has cooled. Clean the top (lid) and bottom (tray)
sections using a soft cloth to wipe the inside surfaces and then rinse
thoroughly with tap water.
95-113364 US EN R4 STATIM 2000 G4 5000 G4 OpMan.indd 45
2017-01-18 11:01 AM
9 Maintaining your STATIM
Technician
Once a year
Cassette
•
Check the tray, lid and seal for damage. Replace if necessary.
Biological Filter
•
Inspect the biological filter for moisture.
Solenoid Valve
•
Inspect the valve and clean if dirty. Replace the plunger if defective.
Pump
•
Clean the filters, replace if dirty.
Check Valve
•
Remove the exhaust tube from the back of the unit during the air drying
phase. Check for air coming from the fitting.
•
Remove the air compressor tube from the check valve inlet while running
a cycle. Make sure no steam is leaking from the valve. Replace if there are
any leaks.
Water Reservoir
•
Check the reservoir for dirt. Clean and rinse with Steam distilled water or
VistaPure deionized water if necessary.
Calibration
•
Calibrate the unit.
95-113364 US EN R4 STATIM 2000 G4 5000 G4 OpMan.indd 46
2017-01-18 11:01 AM
10 Troubleshooting your STATIM
Problem
Solution
Unit displays:
Cycle interrupted — NOT STERILE,
Cycle aborted — NOT STERILE or
CYCLE FAULT messages.
Wait a few minutes and attempt another cycle before proceeding to the
next solution. Remove the cassette.
Be careful. The metal parts will be hot and the cassette will contain hot
steam.
Inspect the cassette to ensure that the holes in the back of the seal are
perfectly aligned, and that the flexible lip of the seal is completely free.
Check the exhaust tube for kinks or obstructions. If kinked, straighten
the tube. If the tube cannot be straightened, remove it from the push-in
fitting attached to the STATIM. Depress the collar on the fitting and, with
the other hand pull firmly on the tube. Once the tube is free of the fitting,
cut the damaged section of tubing away using a sharp instrument. Be
sure that you leave enough tube to reach the unit when you re-attach the
tube to the exhaust fitting. If the tube is too short to remove a section,
contact your SciCan dealer for a replacement.
Check that the STATIM has not inadvertently been exposed to any electrical
interference. Refer to Section 3.1 Positioning and Powering your Unit.
Try running another cycle. If the problem persists, record the cycle fault
message number and contact your dealer.
Excessive steam issuing from the front of
the machine.
Remove and reinsert the cassette. Attempt another cycle. Remove
and check the cassette seal for misalignment or damage. Replace the
seal if required.
Be careful as the metal parts will be hot and the cassette will contain
hot steam.
If the leak persists, turn the unit OFF, remove and unload the cassette
and contact your SciCan dealer.
Machine will not start and touchscreen
shows:
Press on the icon to confirm whether it is a water level problem or a
water quality problem.
If it is a water quality problem, you have likely used water which is not
steam-process distilled or is improperly distilled.
Empty the reservoir and refill with steam-process distilled water or
VistaPure deionized water containing less than 5 ppm total dissolved
solids (having conductivity of less than 10 µS / cm). If you have the
water conductivity meter, check the quality of the water before refilling
the reservoir. To empty the reservoir, see Section 3.6 Shipping the Unit.
If the level of the water in the reservoir is low, refill the reservoir. Refer
to the steps described in Section 3.4 Filling the Water Reservoir.
The printer does not work.
Make sure that the printer cable is connected securely with the
connector on the back of the STATIM. Make sure that the printer is
powered ON. Power unit OFF for 10 seconds and then power ON
again.
95-113364 US EN R4 STATIM 2000 G4 5000 G4 OpMan.indd 47
2017-01-18 11:01 AM
10 Troubleshooting your STATIM
Problem
Solution
Unit does not power ON.
Check that the unit is plugged into a properly grounded outlet and that
the power cord is firmly seated at the rear of the machine.
Try another circuit. Power unit OFF for 10 seconds and then power
ON again.
Check the condition of the line circuit breaker or fuse.
There is water under the machine.
Check that water was not spilled when refilling the reservoir. Make
sure the plug in the exhaust tube is secured. Remove and reinsert
the cassette. Attempt another cycle.
Be careful. The metal parts will be hot, and the cassette will contain
hot steam.
The cassette is leaking. If water drips from the underside of the unit
during operation, check the cassette seal for misalignment or damage
and replace the seal if required.
Be careful. The metal parts will be hot, and the cassette will contain
hot steam.
Attempt another cycle. If it still leaks attempt another cycle using a
different cassette if possible.
If the leak persists, turn the unit OFF, remove and unload the cassette,
unplug the unit, and call your dealer.
Instruments do not dry.
Best drying occurs when the cycle continues to completion. Allow the
cycle to finish. Make sure the instruments are loaded correctly in the
cassette. Refer to Section 4.4 Preparing and Loading Instruments.
Check the unit leveling.
Check air/biological filters and replace if dirty.
Examine the exhaust tube (tube to the waste bottle) for kinks. If
kinked, straighten the tube. If the tube cannot be straightened,
remove it from the push-in fitting attached to the STATIM. Depress
the collar on the fitting and with the other hand pull firmly on the tube.
Once the tube is free of the fitting, cut the damaged section of tubing
away using a sharp instrument. Be sure that you leave enough tube to
reach the unit when you re-attach the tube to the exhaust fitting.
If the tube is too short to remove a section, contact your SciCan
dealer for a replacement.
Make sure the compressor is working. To check, remove the exhaust
tube from the waste bottle. Start the Air Drying Only Cycle, and place
the free end into a glass of water. If there is not a strong, steady flow
of bubbles, the compressor is not functioning properly. Contact your
SciCan dealer.
95-113364 US EN R4 STATIM 2000 G4 5000 G4 OpMan.indd 48
2017-01-18 11:01 AM
10 Troubleshooting your STATIM
Problem
Solution
Time and date are incorrect.
The time and date have not been set. See Section 3 Setting up your STATIM,
for time and date instructions.
Touchscreen is blank/white.
Power was interrupted during a firmware upgrade. Power off the unit
and power it on again. It can take up to 6 minutes before the main
menu screen appears.
Touchscreen is blank/ dark.
Check power source.
USB storage device does not contain
the last print out.
Re-insert the USB storage device and wait for the data to copy over
again.
If problem persists, back up all the information you have on the USB
device and reformat it.
NOTE: You can always access all your unit’s cycle information
through the unit’s web portal.
Touchscreen shows:
An X over the connectivity icon means the unit is not connected to a
network. If it is supposed to be connected to a network and the X is
visible, it is because the unit is unable to acquire an IP address. To
resolve the issue, try some of the following:
•
Check that the router is functioning properly
•
Check the LAN cable (try a new cable if possible)
•
Make sure your router assigns IP addresses automatically.
Renew the IP address by following these steps:
1. Scroll through the setup menu to NETWORK SETUP and
select.
2. Select RENEW IP.
Unit is not sending emails.
Check email settings by using the TEST button on the unit’s web
portal. From the SETUP web page, select the TOOLS tab. Click
on TEST to check your router, unit, and Internet connections. If all
settings appear to be OK. Go to the unit’s touchscreen and renew
the IP address by following these steps:
1. Scroll through the setup menu to NETWORK SETUP and
select.
2. Select RENEW IP.
Not receiving emails from the unit
Check your spam filter. Be certain the unit has been identified as
an accepted email source.
95-113364 US EN R4 STATIM 2000 G4 5000 G4 OpMan.indd 49
2017-01-18 11:01 AM
11 Spare Parts
STATIM G4 SPARES
01-100028S
Cassette Seal (2000)
01-112409S
Cassette Lid (2000 G4)
01-112410S
Cassette Handle - Lid (2000 G4)
01-103945S
Rack-tray Unwrapped Instr. Kit
(2000)
01-101649S
Cassette Seal (5000)
01-112386S
Cassette Lid (5000 G4)
01-112387S
Cassette Handles Tray / Lid
(5000 G4)
01-112388S
Cassette Handle - Lid (5000 G4)
01-112511S
Cassette Lid (5000 Ext G4)
01-112512S
Cassette Handles Tray / Lid
(5000 Ext G4)
01-112513S
Cassette Handle - Lid
(5000 Ext G4)
01-101787S
Reservoir Cap and Filter
01-100812S
Condenser Bottle
01-100724S
Condenser Bottle w/o Condenser
01-100735S
Waste Water Bottle Fitting
01-100204S
Exhaust Tube
01-100207S
Compressor Filter (2000)
01-101652S
Air Compressor Filter (5000)
01-102119S
Filter Biological
01-109300S
Water Reservoir Filter Kit
01-104343S
Plug - Drain Tubing
01-100780S
Bumper
01-112408S
Cassette Tray (2000 G4)
01-112407S
Cassette Tray with Mesh Rack
(2000 G4)
01-106653
Mesh Rack - STATIM 2000
01-112385S
Cassette Tray (5000 G4)
01-112510S
Cassette Tray (5000 Ext G4)
01-101709S
Mesh Rack (STATIM 5000)
01-103935
Drying Plates (Qty 5)
(STATIM 5000)
01-103923
Condenser Additional Bottle
01-101647S
Power Cord North America
STATIM G4 ACCESSORIES
01-112406S
Cassette Complete (2000 G4)
01-114989S
Cassette Complete (5000 G4)
01-112509S
Cassette Complete (5000 Ext G4)
95-113364 US EN R4 STATIM 2000 G4 5000 G4 OpMan.indd 50
2017-01-18 11:01 AM
For a period of one year, SciCan guarantees that the STATIM 2000 / 5000 G4, when manufactured by
SciCan in new and unused condition, will not fail during normal service due to defects in material and
workmanship that are not due to apparent abuse, misuse, or accident.
The one year warranty will cover the performance of all components of the unit except consumables such
as the cassette seal, the compressor filter and the microbiological filter, provided that the product is being
used and maintained according to the description in the user’s manual.
In the event of failure due to such defects during this period of time, the exclusive remedies shall be repair
or replacement, at SciCan’s option and without charge, of any defected part(s) (except gasket), provided
SciCan is notified in writing within thirty (30) days of the date of such a failure and further provided that the
defective part(s) are returned to SciCan prepaid.
This warranty shall be considered to be validated, if the product is accompanied by the original purchase
invoice from the authorized SciCan dealer, and such invoice identifies the item by serial number and
clearly states the date of purchase. No other validation is acceptable. After one year, all SciCan’s
warranties and other duties with respect to the quality of the product shall be conclusively presumed to
have been satisfied, all liability therefore shall terminate, and no action or breach of any such warranty or
duty may thereafter be commenced against SciCan.
Any express warranty not provided hereon and any implied warranty or representation as to performance,
and any remedy for breach of contract which, but for this provision, might arise by implication, operation
of law, custom of trade or course of dealing, including any implied warranty of merchantability or of
fitness for particular purpose with respect to all and any products manufactured by SciCan is excluded
and disclaimed by SciCan. If you would like to learn more about SciCan products and features, visit our
website at www.scican.com.
95-113364 US EN R4 STATIM 2000 G4 5000 G4 OpMan.indd 51
2017-01-18 11:01 AM
13.1 STATIM 2000 G4
Machine Dimensions:
Length:
500 mm (19.6”)
Width:
414 mm (16.3”)
Height:
157 mm (6.2”)
Cassette Size (External):
Length:
410 mm (16”) (includes handles)
Width:
195 mm (7.7”)
Height:
40 mm (1.6”)
Cassette Size (Internal):
Length:
280 mm (11”)
Width:
180 mm (7.1”)
Height:
35 mm (1.4”)
Sterilization Chamber Volume:
1.8 L (61 fl. oz.) U.S.
Reservoir Volume:
4.0 L (140 fl. oz.) U.S.
Weight (Without water):
21 kg (46 lbs)
Clearance required:
Top:
50 mm (1.9”)
Sides:
50 mm (1.9”)
Back:
50 mm (1.9”)
Front:
480 mm (18.9”)
Maximum Steam Temperature:
138˚C (280˚F)
Maximum Operating Pressure:
341kP abs (49.5 psia)
Electrical Rating* (+/- 10%):
100 V, 50 / 60 Hz, 11A
110 V, 50 / 60 Hz, 11A
220 - 240 V, 50 / 60 Hz, 6 A
*see serial number label for requirements specific to your unit.
Ethernet Port:
10/100 Base-T
USB Port:
USB 2.0
Protection Class:
I
Protection:
covered (indoor use only)
Ambient Operating Temperature and Humidity:
15˚C to 25˚C (59˚F to 77˚F)
and 25% to 70%
Max. Altitude:
Up to 2000 meters (6600 ft)
Installation Category:
1
95-113364 US EN R4 STATIM 2000 G4 5000 G4 OpMan.indd 52
2017-01-18 11:01 AM
13.2 STATIM 5000 G4
Machine Dimensions:
Length:
600 mm (23.6”)
Width:
414 mm (16.3”)
Height:
190 mm (7.5”)
Cassette Size (External):
Length:
495 mm (19.5”) (includes handles)
Width:
195 mm (7.7”)
Height:
80 mm (3.2”)
Extended Cassette Size (External): Length:
565 mm (22.2”) (includes handles)
Width:
195 mm (7.7”)
Height:
80 mm (3.2”)
Casette Size (Internal):
Length:
380 mm (15”)
Width:
180 mm (7.1”)
Height:
75 mm (3”)
Extended Section (Internal):
Length:
110 mm (4.3”)
Width:
130 mm (5.1”)
Height:
28 mm (1.1”)
Sterilization Chamber Volume:
5.1 L (170 fl. oz.) U.S.
Extended Sterilization Chamber Volume:
5.5 L (186 fl. oz.) U.S.
Reservoir Volume:
4.0 L (140 fl. oz.) U.S.
Weight (Without water):
33 kg (73 lbs)
Clearance required:
Top:
50 mm (1.9”)
Sides:
50 mm (1.9”)
Back:
50 mm (1.9”)
Front:
570 mm (22.4”)
Maximum Steam Temperature:
138˚C (280˚F)
Maximum Operating Pressure:
341kP abs (49.5 psia)
Electrical Rating* (+/- 10%):
100 V, 50 / 60 Hz, 11A
110 V, 50 / 60 Hz, 11A
220 - 240 V, 50 / 60 Hz, 6 A
*see serial number label for requirements specific to your unit.
Ethernet Port:
10/100 Base-T
USB Port:
USB 2.0
Protection Class:
I
Protection:
covered (indoor use only)
Ambient Operating Temperature and Humidity:
15˚C to 25˚C (59˚F to 77˚F)
and 25% to 70%
Max. Altitude:
Up to 2000 meters (6600 ft)
Installation Category:
1
95-113364 US EN R4 STATIM 2000 G4 5000 G4 OpMan.indd 53
2017-01-18 11:01 AM
SCICAN SOFTWARE PRODUCT LICENSE
This SciCan Software Product Agreement is made as of the date of delivery (“Effective Date”) to Customer of
the equipment containing the SciCan Software Product (the “SciCan Equipment”) by and between SciCan Ltd.,
1440 Don Mills Road, 2nd Floor, Toronto, Canada, M3B 3P9 (“SciCan”) and the purchaser or lessee of the SciCan
Equipment and each of its end users (collectively, “Customer”). “SciCan Software Product” means all SciCan
proprietary software contained in the SciCan Equipment.
This SciCan Software Product License constitutes the entire agreement (“Agreement”) between SciCan and the
Customer concerning Customer’s use of the SciCan Software Product. No purchase order which purports to modify
or supplement this Agreement will add to or vary the terms of this Agreement even if signed or initialed by SciCan.
ARTICLE 1
INTERPRETATION
1.1
Definitions
(a) “Affiliate” will mean any affiliated entity, which Controls, is Controlled by, or is under common Control with
Customer.
(b) “Confidential Information” will mean non-public, commercially sensitive information of each of the parties and,
in the case of SciCan, the SciCan Software Product, Updates, Documentation, and all SciCan information which
is marked as confidential or proprietary at the time of disclosure.
(c) “Control” will mean the possession, directly or indirectly, of the power to direct or cause the direction of the
management and operating policies of an entity through the ownership of voting securities (at least fifty-one
percent (51%) of its voting or equity securities), contract, voting trust, or otherwise.
(d) “Documentation” will mean the user manuals relating to the use of the SciCan Software Product and the SciCan
Equipment delivered with the SciCan Equipment.
(e) “SciCan Licensors” will mean third parties which have granted SciCan distribution rights with respect to their
software.
(f) “Updates” will mean modifications made by SciCan to the SciCan Software Product that SciCan generally makes
available at no additional charge to its customers who are current subscribers for support services and who are
current in payment of support services fees, if any. Updates become part of the SciCan Software Product for
purposes of this Agreement.
ARTICLE 2
LICENSE
2.1
License Grant
Subject to the terms and conditions of this Agreement, SciCan grants Customer a perpetual, fully paid, non-
exclusive, and non-transferable license to use the SciCan Software Product, solely at Customer’s site, solely in
connection with the operation of the SciCan Equipment for Customer’s internal business purposes.
ARTICLE 3
LICENSE RESTRICTIONS
3.1
Restrictions
Except as expressly authorized herein, Customer will not engage in or permit any:
(a) copying or modification of the SciCan Software Product or Documentation;
(b) reverse engineering, decompilation, translation, disassembly, or discovery of the source code of all or any portion
of the SciCan Software Product;
95-113364 US EN R4 STATIM 2000 G4 5000 G4 OpMan.indd 54
2017-01-18 11:01 AM
(c) distribution, disclosure, marketing, rental, leasing or service bureau use or transfer to any third party of the
SciCan Software Product or the Documentation, except as part and parcel of the sale of the SciCan Equipment
containing the SciCan Software Product;
(d) disclosure of the results of SciCan Equipment or SciCan Software Product performance benchmarks to any third
party without SciCan’s prior written consent; or
(e) disclosure of any source code (if any) provided hereunder to any third party.
ARTICLE 4
UPDATES
4.1
Updates
(a) Provided Customer has opened an account and provided all requested information to SciCan, and, if applicable,
paid the associated fees for Updates, SciCan will provide Updates for the SciCan Software Product in
accordance with SciCan’s Update policies and procedures of general application. Prior to discontinuing Updates
for the SciCan Software Product, SciCan will provide Customer with at least six (6) months advance notice.
Customer will allow SciCan to use remote access software to aid in the resolution of problems or questions.
Update fees, if any, will be billed on an annual basis payable in advance.
(b) SciCan will have no obligation to provide Updates or assistance if Customer fails to make any required payment
or otherwise elects to discontinue Update services. In order to reinstate or renew support services, Customer
must first pay SciCan the then current annual Updates services fee and all past unpaid Updates services fees
and agree that all past Updates will be loaded on the SciCan Equipment.
(c) SciCan will have no obligation to provide Updates for any (i) altered, damaged or modified SciCan Equipment or
SciCan Software Product, (ii) SciCan Software Product that is not the then-current or previous sequential release,
(iii) SciCan Software Product problems caused by Customer’s negligence or other causes beyond the control
of SciCan, or (iv) a failure that cannot be reproduced at SciCan’s facility or via remote access to the Customer’s
facility.
ARTICLE 5
OWNERSHIP
5.1
Ownership
SciCan retains all right, title and interest in the SciCan Software Product, Updates, and Documentation and any
copies thereof. Except as otherwise expressly granted in this Agreement, no license, right or interest in any SciCan
trade mark, copyright, trade name or service mark is granted hereunder.
ARTICLE 6
PATENT AND COPYRIGHT INDEMNITY
6.1
SciCan Indemnity
SciCan will defend and indemnify Customer for all costs (including reasonable attorney fees) arising from a claim
that SciCan Software Product furnished and used within the scope of this Agreement infringes a registered
copyright or patent provided that:
(a) Customer notifies SciCan in writing within thirty (30) days of the claim;
(b) SciCan has sole control of the defense and all related settlement negotiations, and
(c) Customer provides SciCan with the assistance, information, and authority necessary to perform the above.
Reasonable expenses incurred by Customer in providing such assistance will be reimbursed by SciCan.
95-113364 US EN R4 STATIM 2000 G4 5000 G4 OpMan.indd 55
2017-01-18 11:01 AM
SciCan will have no liability for any claim of infringement based on:
(a) use of a superseded or modified release of the SciCan Software Product (except for such alterations or
modifications which have been made by SciCan or under SciCan's direction) if such infringement would have
been avoided by the use of a current unaltered release of the SciCan Software Product; or
(b) the combination, operation, or use of the SciCan Software Product with hardware, programs or data not
furnished or otherwise approved by SciCan if such infringement would have been avoided by the use of the
SciCan Software Product without such hardware, programs or data.
6.3
SciCan’s Obligation
In the event the SciCan Software Product is held or believed by SciCan to infringe, or Customer’s use of the SciCan
Software Product is enjoined, SciCan will have the option, at its expense, to:
(a) modify the SciCan Software Product to be non-infringing; or
(b) obtain for Customer a license to continue using the SciCan Software Product; or
(c) substitute the SciCan Software Product with other software reasonably suitable to operate the SciCan
Equipment; or
(d) if none of the foregoing remedies are commercially feasible, terminate the license for the infringing SciCan
Software Product and refund the price of the affected SciCan Equipment, prorated over a five-year term
calculated from the Effective Date.
6.4
Entire Liability for Infringement
This Article 6 states SciCan's entire liability for infringement or misappropriation of intellectual property rights.
ARTICLE 7
WARRANTY
7.1
Warranty
SciCan warrants that it has title to and/or the authority to grant licenses of the SciCan Software Product. Customer’s
exclusive remedy with respect to breach of this provision will be pursuant to Article 6 (Patent and Copyright
Indemnity).
7.2
Functionality
SciCan warrants for a period of ninety (90) days from the Effective Date that the SciCan Software Product, unless
modified by Customer and provided that all Updates have been installed, will perform, in all material aspects, the
functions described in the Documentation when operated on the related SciCan Equipment.
7.3
Services
SciCan reserves the right to charge Customer for services performed by SciCan in connection with reported failures
which are later determined to be caused by operator error, untrained users, site electrical malfunction, software or
hardware not supplied or recommended by SciCan or by alterations or additions to the SciCan Equipment or the
SciCan Software Product other than by way of Updates or by persons other than SciCan employees or consultants.
7.4
DISCLAIMER
THE WARRANTIES ABOVE ARE EXCLUSIVE AND IN LIEU OF ALL OTHER WARRANTIES, WHETHER EXPRESS OR
IMPLIED, INCLUDING THE IMPLIED WARRANTIES OF MERCHANTABLE QUALITY, MERCHANTABILITY, NON-
INFRINGEMENT AND FITNESS FOR A PARTICULAR PURPOSE.
95-113364 US EN R4 STATIM 2000 G4 5000 G4 OpMan.indd 56
2017-01-18 11:01 AM
LIMITATION OF LIABILITY
8.1
LIMITED LIABILITY
EXCEPT AS PROVIDED IN Article 6 (PATENT AND COPYRIGHT INDEMNITY), SCICAN’S LIABILITY FOR DAMAGES
UNDER THIS AGREEMENT WILL IN NO EVENT EXCEED THE AMOUNT PAID BY LICENSEE TO SCICAN FOR THE
SCICAN EQUIPMENT IN CONNECTION WITH WHICH THE CLAIM AROSE. IN NO EVENT WILL SCICAN BE LIABLE
FOR INDIRECT, INCIDENTAL, SPECIAL OR CONSEQUENTIAL DAMAGES, INCLUDING BUT NOT LIMITED TO
LOST DATA OR LOST PROFITS, HOWEVER ARISING, EVEN IF IT HAS BEEN ADVISED OF THE POSSIBILITY OF
SUCH DAMAGES. IN NO EVENT WILL SCICAN LICENSORS BE LIABLE FOR ANY DIRECT, SPECIAL, INDIRECT,
CONSEQUENTIAL, INCIDENTAL OR EXEMPLARY DAMAGES ARISING OUT OF THIS AGREEMENT, REGARDLESS
OF THE FORM OF ACTION WHETHER BASED IN CONTRACT, TORT, OR ANY LEGAL THEORY. THE PARTIES
AGREE TO THE ALLOCATION OF LIABILITY RISK WHICH IS SET FORTH IN THIS SECTION 8.1.
ARTICLE 9
CONFIDENTIALITY
9.1
Maintain Confidentiality
By virtue of the account with SciCan that the Customer will open for purposes of registering the SciCan Equipment
and obtaining Updates, SciCan will obtain and possess Confidential Information and personal information relating
to the Customer. Customer information that SciCan obtains does not include local area network (“LAN”) topology
or information about other devices connected to the LAN. Personal information that SciCan obtains will include the
names of persons to whom emails may be sent by SciCan respecting the functioning of the SciCan Equipment and
the Updates. Each of the parties acknowledges to the other that by virtue of their licensing and Update services
relationship it may have access to Confidential Information of the other party. The parties agree, both during the
term of this Agreement and after termination, to hold each other's Confidential Information in confidence. The
SciCan Software Product shall be treated as confidential in perpetuity. The parties agree not to make each other's
Confidential Information available in any form to any third party (other than those of its employees or consultants
under nondisclosure obligations) or to use each other's Confidential Information for any purpose other than as
contemplated by this Agreement. Each party will take commercially reasonable steps to ensure that Confidential
Information is not disclosed or distributed by its employees or consultants in violation of the provisions of this Article
9. The parties agree that the terms and conditions of this Agreement are considered confidential.
9.2
Exception
Notwithstanding any provision contained in this Agreement, neither party will be required to maintain in confidence
any of the following information:
(a) information which, at the time of disclosure to the receiving party, is in the public domain;
(b) information which, after disclosure, becomes part of the public domain, except by breach of this Agreement;
(c) information which was in the receiving party’s possession at the time of disclosure, and which was not acquired,
directly or indirectly, from the disclosing party;
(d) information which the receiving party can demonstrate resulted from its own research and development,
independent of disclosure from the disclosing party;
(e) information which the receiving party receives from third parties, provided such information was not obtained by
such third parties from the disclosing party on a confidential basis; or
(f) information which is produced in compliance with applicable law or a court order, provided the other party is
given reasonable notice of such law or order and an opportunity to attempt to preclude or limit such production.
95-113364 US EN R4 STATIM 2000 G4 5000 G4 OpMan.indd 57
2017-01-18 11:01 AM
Law and Venue
This Agreement will be governed and construed under the laws of the Province of Ontario and the applicable federal
laws of Canada. In no event will this Agreement be governed by the United Nations Convention on Contracts for the
International Sale of Goods.
10.2
Notices
All notices will be in writing and sent by first class mail, overnight courier, or transmitted by facsimile and confirmed by
mailing, to the addresses indicated on the first page of this Agreement, or such other address as either party may indicate
by at least ten (10) days prior written notice to the other party. Notices to SciCan will be sent to [email protected].
Notice will be deemed to have been given upon personal delivery (in the case of overnight courier or facsimile) or five (5)
business days after being sent by first class mail or on the next business day if sent by facsimile.
10.3
Assignment
Customer may not assign this Agreement (by operation of law or otherwise) or sublicense the SciCan Software
Product without the prior written consent of SciCan. Customer may, however, sell or otherwise dispose of the
SciCan Equipment with the SciCan Software Product loaded onto its internal operating system. Customer
acknowledges that no Upgrades will be available for any equipment that is sold or disposed of unless
the purchaser or recipient opens an account for Upgrades with SciCan and pays the appropriate fees.
Any prohibited assignment or sublicense of the SciCan Software Product will be null and void. The foregoing
notwithstanding, upon written notice to SciCan, Customer may assign, or otherwise transfer this Agreement to an
Affiliate of Customer, provided such Affiliate agrees with SciCan to be bound by the terms and conditions of this
Agreement.
10.4
Legal Costs
If any legal action, including arbitration, is required in order to enforce or interpret any of the provisions of this
Agreement, the prevailing party in such action will recover all reasonable costs and expenses, including attorney's
fees, incurred in connection therewith.
10.5
Extraordinary Relief
Each party acknowledges that any breach of its obligations with respect to the proprietary rights of the other party
or such party's licensors may cause such other party irreparable injury for which there may be inadequate remedies
at law and that such other party and its licensors will be entitled to injunctive relief, in addition to all other remedies
available to it.
10.6
Headings
The article and section headings herein are provided for convenience only and have no substantive effect on the
construction of this Agreement.
10.7
Force Majeure
Neither party will be liable for any failure to perform due to causes beyond its reasonable control.
10.8
Severability
If any provision of this Agreement is held to be unenforceable, the parties will substitute for the affected provision an
enforceable provision, which approximates the intent and economic effect of the affected provision.
10.9
Non-Waiver
The failure by a party to exercise any right hereunder will not operate as a waiver of such party’s right to exercise
such right or any other right in the future.
10.10
Amendment
This Agreement may be amended only by a written document executed by a duly authorized representative of each
of the parties.
10.11
Exclusive Agreement
This Agreement replaces and supersedes any prior verbal understandings, written communications or representations.
95-113364 US EN R4 STATIM 2000 G4 5000 G4 OpMan.indd 58
2017-01-18 11:01 AM
1. Read first — regulatory information
Read this document before using your sterilizer. This sterilizer complies with the radio frequency, safety standards
and regulations of the countries that have approved its importation. Contact SciCan for the latest list of approved
countries. Install and use your sterilizer according to the following instructions.
IMPORTANT NOTE: To comply with FCC* & IC RF** exposure compliance requirements, the antenna used for this
transmitter must be installed to provide a separation distance of at least 20cm from all persons and must not be
co-located or operating in conjunction with any other antenna or transmitter.
*FCC (Federal Communications Commission)
**IC RF (Industry Canada Radiofrequency)
2. STATIM™ Wireless adapter
The STATIM Sterilizer includes a Wi-Fi IEEE 802.11b,g,n module that allows functions that were previously available
only when using a wired interface.
The wireless adapter supports connection to IEEE 802.11b,g,n networks, WPA™ Personal and WPA2™ Personal
(EAP* Types: EAP-TLS, EAP-TTLS/MSCHAPv2, PEAPv0/EAP-MSCHAPv2, PEAPv1/EAP-GTC ,EAP-FAST). The
WiFi module uses the frequency range 2412-2462MHz and has a maximum 0.111W RF Output power.
*Extensible Authentication Protocol
While the STATIM unit is connected to a WiFi network, the security of the connection depends on the configuration
of the Wireless infrastructure (router or access point).
Securing your Wi-Fi® connections is an important element of securing your personal data. A Wi-Fi network using
WPA2™ provides both security (you can control who connects) and privacy (the transmissions cannot be read by
others) for communications as they travel across your network. For maximum security, your network should include
only devices with the latest in security technology – Wi-Fi Protected Access® 2 (WPA2). Wi-Fi CERTIFIED™ devices
implement WPA2. See more at: http://www.wi-fi.org/discover-wi-fi/security#sthash.tk28zkHJ.dpuf
Most access points, routers, and gateways are shipped with a default network name (SSID), and administrative
credentials (username and password) to make configuration as simple as possible. These default settings should
be changed as soon as you set up your network. - See more at: http://www.wi-fi.org/discover-wi-fi/security#sthash.
tk28zkHJ.dpuf
It’s also important to consider employing other measures to secure your communications after they travel beyond
your Wi-Fi network. See more at: http://www.wi-fi.org/discover-wi-fi/security#sthash.tk28zkHJ.dpuf.
Tips on securing a new network
• Change the network name (SSID) from the default name
• Change the administrative credentials (username and password) that control the configuration settings of
your Access Point/Router/Gateway
• Enable WPA2-Personal (aka WPA2-PSK) with AES encryption
• Create a network passphrase that meets recommended guidelines
• Enable WPA2 security features on your client device and enter the passphrase for your network
See more at: http://www.wi-fi.org/discover-wi-fi/security#sthash.tk28zkHJ.dpuf
95-113364 US EN R4 STATIM 2000 G4 5000 G4 OpMan.indd 59
2017-01-18 11:01 AM
Checking security on an existing network
When you add a new device to your Wi-Fi network, it’s a great time to make sure you’re taking advantage of the
highest level of security. Take the opportunity to ensure your network is configured for WPA2.
If your network was set up some time ago, or a service provider (e.g. consultant or cable provider) configured your
network, it may be worth checking that it’s configured for the highest level of security. If your network is configured
for an older generation of security (WEP or WPA), Wi-Fi Alliance® recommends you move to WPA2. WPA2 has been
required on all Wi-Fi CERTIFIED products since 2006 – the vast majority of Wi-Fi CERTIFIED devices in service
today are capable of WPA2.
2.2.
Passphrase quality & lifespan
A secure network passphrase greatly enhances network security, so it is important to select an effective
passphrase. In general, increasing length, complexity and randomness all improve the quality of a passphrase. Wi-Fi
Alliance recommends that a passphrase is at least eight characters long, and includes a mixture of upper and lower
case letters and symbols. A passphrase should not contain a word found in a dictionary and should not include
personal information (identification number, name, address, etc).
Periodically changing the passphrase on your network also increases security.
See more at: http://www.wi-fi.org/discover-wi-fi/security#sthash.tk28zkHJ.dpuf
2.3.
Wireless Setup
The STATIM unit allows for Wireless or Wired connections, but only one type at a time.
2.3.1.
Selecting between the Wired and Wireless networks
2.3.2.
Connecting to a Wireless network
mm/dd/yyyy hh:mm:ss
SELECT A CYCLE
mm/dd/yyyy hh:mm:ss
SELECT A CYCLE
95-113364 US EN R4 STATIM 2000 G4 5000 G4 OpMan.indd 60
2017-01-18 11:01 AM
USA — Federal Communications Commission (FCC)
3.1.
Approved wireless devices
This section presents the FCC ID and model number of the wireless device
3.2.
Preinstalled wireless LAN adapter
FCC ID: YOPGS2011MIE (Model: GS2011MIE)
3.3.
FCC ID location
3.3.1. On the rear side of your STATIM Sterilizer, you will find an indicator label of the format “Contains FCC
ID YOPGS2011MIE”, where YOPGS2011MIErepresents the FCC ID that corresponds to your preinstalled
Wireless LAN module.
3.4.
FCC RF Exposure compliance
The total radiated energy from the Main antenna connected to the Wireless Card conforms to the FCC limit of the
SAR (Specific Absorption Rate) requirement regarding 47 CFR Part 2 section 1093, when the sterilizer was tested.
The transmission antenna for the Wireless Card are located in the front fascia.
3.5.
Radio Frequency interference requirements
The device has been tested and found to comply with the limits for a Class B digital device pursuant to FCC Part 15
Subpart B.
Due to differences in channel allocation, if you cannot connect using the WIFI device, it may be because these
channels are not available in your region or due to interference. If this occurs, the Ethernet connection should be
used.
4.
Canada — Industry Canada (IC)
4.1.
Approved wireless devices
This section presents the IC Certification and model number of each wireless device.
4.2.
Preinstalled wireless LAN adapter
IC:9154A-GS2011MIE (Model GS2011MIE)
4.3.
Low power license-exempt radio communication devices (RSS-210)
Operation is subject to the following two conditions:
1. This device may not cause interference.
2. This device must accept any interference, including interference that may cause undesired operation of the
device. The transmitter devices have been designed to operate with the antennas integrated in the sterilizer, and
having a maximum gain of within 3 dBi.
4.4.
Exposure of humans to RF fields (RSS-102)
The STATIM unit employs low gain integral antenna that do not emit RF field in excess of Health Canada limits for
the general population; consult Safety Code 6, obtainable from Health Canada's Web site at www.hc-sc.gc.ca/
The radiated energy from the antennas connected to the wireless adapters conforms to the IC limit of the RF
exposure requirement regarding IC RSS-102, Issue 2 clause 4.1
95-113364 US EN R4 STATIM 2000 G4 5000 G4 OpMan.indd 61
2017-01-18 11:01 AM
Electronic emission notices
5.1.
Federal Communications Commission (FCC) Declaration of Conformity
NOTE: The wireless adapters (Model: GS2011MIE) underwent certification process for the FCC Part 15 Subpart B
compliance under the respective FCC ID number.
This equipment has been tested and found to comply with the limits for a Class B digital device, pursuant to Part
15 of the FCC Rules. These limits are designed to provide reasonable protection against harmful interference in a
residential installation. This equipment generates, uses, and can radiate radio frequency energy and, if not installed
and used in accordance with the instructions, may cause harmful interference to radio communications. However,
there is no guarantee that interference will not occur in a particular installation. If this equipment does cause harmful
interference to radio or television reception, which can be determined by turning the equipment off and on, the user
is encouraged to try to correct the interference by one or more of the following measures:
Reorient or relocate the receiving antenna.
• Increase the separation between the equipment and receiver.
• Connect the equipment into an outlet on a circuit different from that to which the receiver is connected.
• Consult an authorized dealer or service representative for help.
SciCan LTD. is not responsible for any radio or television interference caused by unauthorized changes or
modifications to this equipment. Unauthorized changes or modifications could void the user's authority to operate
the equipment. This device complies with Part 15 of the FCC Rules. Operation is subject to the following two
conditions: (1) this device may not cause harmful interference, and (2) this device must accept any interference
received, including interference that may cause undesired operation.
Responsible Party:
SciCan LTD.
1440 Don Mills Road
Toronto, Ontario, Canada
M3B 3P9
Telephone: 1.800.667.7733
5.2.
Industry Canada Class B Emission Compliance Statement
This Class B digital apparatus complies with Canadian ICES-003.
95-113364 US EN R4 STATIM 2000 G4 5000 G4 OpMan.indd 62
2017-01-18 11:01 AM