SICK FLOWSIC600-XT Service Manual

Specifications

Brand: SICK
Category: Measuring Instruments
Model: FLOWSIC600-XT
File Type: PDF
File Size: —
Pages: 68

Read Online

Read the SICK FLOWSIC600-XT Service Manual online using the built-in PDF viewer.

Text of the manual

FLOWSIC600-XT · Service Manual · 8019179/V0-2/2016-03 · © SICK Engineering GmbH

Document Information Described Product Product name: FLOWSIC600-XT Document ID Title: Service Manual FLOWSIC600-XT Part No.: 8019179 Document ID: 9238933 Version: 0-2 Release: 2016-03 Manufacturer SICK Engineering GmbH Bergener Ring 27 · D-01458 Ottendorf-Okrilla · Germany Phone: +49 35 20552410 Fax: +49 35 20552450 E-Mail: [email protected] Original document The English version 8019179 of this document is an original docu- ment from SICK Engineering GmbH. SICK Engineering GmbH assumes no liability for the correctness of an unauthorized translation. Please contact SICK Engineering GmbH or your local representa- tive in case of doubt. Legal Information Subject to change without notice. © SICK Engineering GmbH. All rights reserved. Glossary ATEX ATEX: Atmosphères Explosifs: Abbreviation for Euro- pean standards that govern safety in potentially explosive atmospheres CSA Canadian Standards Association (www.csa.ca) DC Direct Current HF High frequency, e.g. HF pulse IEC International Electronical Commission IECEx IEC system for certification in accordance with stan- dards for devices for use in potentially explosive atmospheres IPxy Ingress Protection: Degree of protection of a device in accordance with IEC/DIN EN 60529; x specifies the protection against contact and impurities, y pro- tection against moisture. MDR Manufacturer Data Record NAMUR Abbreviation for "Normen-Arbeitsgemeinschaft für Mess- und Regeltechnik in der chemischen Indu- strie", now "Interessengemeinschaft Automatisie- rungstechnik der Prozessindustrie" (www.namur.de) SPU Signal Processing Unit

FLOWSIC600-XT · Service Manual · 8019179/V0-2/2016-03 · © SICK Engineering GmbH

3 Warning Symbols Warning Levels / Signal Words HAZARD Risk or hazardous situation which will result in severe personal injury or death. WARNING Risk or hazardous situation which could result in severe personal injury or death. CAUTION Hazard or unsafe practice which could result in less severe or minor injuries. NOTICE Hazard which could result in property damage. Information Symbols IMMEDIATE HAZARD of severe injuries or death Hazard (general) Hazard by electrical voltage Hazard in potentially explosive atmospheres Hazard by explosive substances/mixtures Hazard by unhealthy substances Hazard by toxic substances Information on product condition relating to protection against explosions (general) Information on product characteristics related to Euro- pean Directive 94/9/EC (ATEX) Information on product characteristics related to explo- sion protection in accordance with the IECEx scheme. Important technical information for this product Important information on electric or electronic func- tions Nice to know Supplementary information Link referring to information at another place

FLOWSIC600-XT · Service Manual · 8019179/V0-2/2016-03 · © SICK Engineering GmbH

1 Important Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7 1.1 About this document . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8 1.2 For your safety . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9 1.2.1 Intended use of the device . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9 1.2.2 Authorized personnel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9 1.2.3 General safety instructions and protective measures . . . . . . . . . . . . . . . . . . . . . . . . 10 1.2.4 Hazards by hot, corrosive and explosive gases, or high pressure . . . . . . . . . . . . . . 10 1.2.5 Hazard through heavy loads . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11 2 Components . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13 2.1 Complete overview of the device family . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14 2.1.1 FLOWSIC600-XT . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14 2.1.2 FLOWSIC600-XT Forte . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15 2.1.3 FLOWSIC600-XT 2plex . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16 2.1.4 FLOWSIC600-XT Quatro . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17 2.2 SPU (Signal Processing Unit) overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18 2.2.1 Ex-d enclosure version . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18 2.2.2 Ex-de enclosure version . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19 2.2.3 Ex-i enclosure version . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20 3 Repairs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21 3.1 General information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22 3.2 Prerequisites for repair work . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22 3.3 Tools required . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23 3.4 Working on the meter body . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24 3.4.1 Replacing the sensors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24 3.4.1.1 Remove the sensors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24 3.4.1.2 Fit the new sensor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26 3.4.2 Replacing the integrated p/T sensor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28 3.4.2.1 Remove the p/T sensor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28 3.4.2.2 Fit the new p/T sensor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29 Contents

FLOWSIC600-XT · Service Manual · 8019179/V0-2/2016-03 · © SICK Engineering GmbH

5 3.5 Working on the SPU . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31 3.5.1 Replacing the display unit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31 3.5.1.1 Swivel the display unit downwards . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31 3.5.1.2 Remove the display unit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32 3.5.1.3 Fit the new display unit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33 3.5.1.4 Connect the display unit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34 3.5.1.5 Swivel the display unit upwards and screw tight . . . . . . . . . . . . . . . . . . . . . . . . . . 34 3.5.2 Replacing the display cover . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35 3.5.3 Replacing the backup battery . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36 3.5.3.1 Remove the backup battery . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36 3.5.3.2 Insert the new backup battery . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36 3.5.4 Replacing the RTC battery . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37 3.5.4.1 Replace the RTC battery . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37 3.5.5 Replacing the SPU circuit board . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38 3.5.5.1 Remove the electronics cover . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38 3.5.5.2 Remove the SPU circuit board . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39 3.5.5.3 Fit the new SPU circuit board . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40 3.5.5.4 Fit the electronics cover . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 41 3.5.6 Replacing the Ex-d I/O block . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42 3.5.6.1 Open the Ex-d terminal compartment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42 3.5.6.2 Remove the Ex-d I/O block . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42 3.5.6.3 Fit the new Ex-d I/O block . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44 3.5.6.4 Close the Ex-d terminal compartment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 45 3.5.7 Replacing the Ex-d I/O block fuse . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 46 3.5.8 Replacing the Ex-e connection block . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 47 3.5.8.1 Open the Ex-e terminal compartment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 47 3.5.8.2 Remove the connection blocks in the Ex-e terminal compartment. . . . . . . . . . . 48 3.5.8.3 Fit the new connection block in the Ex-e terminal compartment . . . . . . . . . . . . 48 3.5.8.4 Close the Ex-e terminal compartment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 49 3.5.9 Replacing the cable gland between the Ex-d and Ex-e terminal compartments . . 50 3.5.9.1 Remove the cable gland . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 50 3.5.9.2 Fit the Ex-e cable gland . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 51 3.5.10 Replacing the cable gland between the Ex-d terminal compartment and the Ex-i transducer electronics . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 52 3.5.10.1 Remove the cable gland . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 52 3.5.10.2 Fit the new cable gland . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 53 3.6 Parameter input . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 54 3.6.1 Configure sensor settings after transducer exchange . . . . . . . . . . . . . . . . . . . . . . . . 54 3.6.2 SOS Check . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 54 3.6.3 Configure Stacking . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 55 3.6.4 Configure Zerophase . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 55 3.6.5 Configure Pfix and Tfix . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 55 4 Accessories and Spare Parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 57 4.1 SPU Parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 58 4.2 Electronic boards . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 58 4.3 Transducers . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 59 4.4 Pressure/Temperature sensors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 59 4.5 O-rings for transducers . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 60 4.6 Electronics accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 60 4.7 Transducer covers . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 61

FLOWSIC600-XT · Service Manual · 8019179/V0-2/2016-03 · © SICK Engineering GmbH

4.8 Covers p/T sensor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 62 4.9 Accessories meter body . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 62 5 Annex . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 63 5.1 Component placement specification SPU (sensor cables) . . . . . . . . . . . . . . . . . . . . . . . 64

FLOWSIC600-XT · Service Manual · 8019179/V0-2/2016-03 · © SICK Engineering GmbH

7 Subject to change without notice FLOWSIC600-XT 1 Important Information About this document For your safety

FLOWSIC600-XT · Service Manual · 8019179/V0-2/2016-03 · © SICK Engineering GmbH

Important Information Subject to change without notice 1.1 About this document This Service Manual is part of the device documentation of the FLOWSIC600-XT with the further components: ● Operating Instructions (Part No. 8018846) ● Interface Description FLOWSIC600-XT This version is valid for: ● FLOWSIC600-XT with firmware as from firmware version V1.x.xx ● Operating and Parameter program FLOWgate as from version 1.x.x All service work that can be carried out by trained personnel is described. The valid Operat- ing Instructions contain all basic information on the measurement method, layout and function of the measuring system and its components as well as on using the device, there- fore additional information is only provided here where necessary to understand the func- tion. The Service Manual may therefore only be used together with the valid Operating Instruc- tions.

FLOWSIC600-XT · Service Manual · 8019179/V0-2/2016-03 · © SICK Engineering GmbH

9 Subject to change without notice 1.2 For your safety 1.2.1 Intended use of the device The measuring system FLOWSIC600-XT is used for measuring the actual volumetric flow rate of gases transported in pipelines. Apart from that, the FLOWSIC600-XT can also deter- mine the actual corrected volume and the sound velocity. The measuring system shall only be used as specified by the manufacturer and as set forth below. Special attention must be paid that: ● Usage is compliant with the technical data, specifications on allowable use as well as the requirements for assembly and connection, and ambient and operating conditions (to be found in the order documents, on the type plate and in approval documents, and the Operating Instructions). ● All measures required for value conservation, e.g. for maintenance and inspection, and/or transport and storage, are observed. ● The device is not subjected to any unallowed mechanical stress, e.g. through pig clean- ing. 1.2.2 Authorized personnel Persons responsible for safety must ensure the following: ● Any work on the measuring system may only be carried out by qualified persons and be approved by responsible skilled persons. ● Qualified persons are those who, based on their training, experience or instruction as well as their knowledge of relevant standards, regulations, accident prevention rules and plant conditions, and authorized by those responsible for persons and plants to carry out such work. It is decisive that these persons can recognize and avoid any possi- ble hazards in good time. Skilled persons are persons in accordance with DIN VDE 0105 or IEC 364, or directly comparable standards. ● These persons must have and demonstrate exact knowledge on hazards arising from operation, e.g. through hot, toxic, explosive gases or gases under pressure, gas-liquid mixtures or other media as well as adequate knowledge of the measuring system gained through training. ● It is important that, during all work on the measuring system, the persons have received training by “SICK Engineering GmbH”. ● In hazardous areas with potentially explosive atmospheres, wiring and installation shall only be carried out by staff trained according to EN60079-14 and according to national regulations.

FLOWSIC600-XT · Service Manual · 8019179/V0-2/2016-03 · © SICK Engineering GmbH

Important Information Subject to change without notice 1.2.3 General safety instructions and protective measures Improper use or handling can cause health or material damage. Please therefore read this Section thoroughly and observe this information during all tasks on the FLOWSIC600-XT as well as the warning and caution information in the individual Sections of this Service Man- ual. Basically applicable: ● Always comply with the statutory provisions and the associated technical rules and reg- ulations relevant to the equipment when preparing for and carrying out any work on the measuring system. Pay particular attention to potentially hazardous aspects of the equipment, such as pressurized piping and explosion protection zones. Always pay spe- cial attention to the relevant regulations. ● All work must be carried out in accordance with the local, system-specific conditions and with due consideration to operating hazards and specifications. ● The Operating Instructions belonging to the measuring system as well as system docu- mentation must be available on site. Always observe the information on the prevention of injuries and damage given therein. ● Depending on the particular hazard potential, an adequate number of suitable protec- tion devices and personal safety equipment must be available and used by the person- nel. 1.2.4 Hazards by hot, corrosive and explosive gases, or high pressure The FLOWSIC600-XT is fitted directly in the gas-carrying line. The operating company is responsible for safe operation and for complying with additional national and company- specific regulations. WARNING: Hazards by the gas in the system The following conditions can increase the risk: ● Toxic gas or gas dangerous to health ● Chemically aggressive gas ● Explosive gas ● High gas pressure ● High gas temperature ▸ In installations with an increased risk, the FLOWSIC600-XT may only be fitted and removed when the line is vented or when the installation is at a standstill. ▸ The same applies to repair and service work which involves opening any pressurized component or the explosion-proof SPU. Otherwise health or injury risks can possibly arise through escaping gas (e.g. poisoning, burns). WARNING: Hazards through leaks Operation in leaky condition is not allowed and potentially dangerous. ▸ Regularly check leak tightness of equipment.

FLOWSIC600-XT · Service Manual · 8019179/V0-2/2016-03 · © SICK Engineering GmbH

11 Subject to change without notice 1.2.5 Hazard through heavy loads The FLOWSIC600-XT measuring system must be attached securely to the carrying structure when being transported and installed. ▸ Only use lifting gear and equipment (e.g. lifting straps) suitable for the weight to be lifted. Maximum load information can be found on the type plate of the lifting gear. NOTICE: The lifting lugs are designed for transporting the measuring device only. Do not lift or transport the FLOWSIC600-XT with additional loads using these lugs.

FLOWSIC600-XT · Service Manual · 8019179/V0-2/2016-03 · © SICK Engineering GmbH

Important Information Subject to change without notice

FLOWSIC600-XT · Service Manual · 8019179/V0-2/2016-03 · © SICK Engineering GmbH

13 Subject to change without notice FLOWSIC600-XT 2 Components Complete overview of the device family SPU (Signal Processing Unit) overview

FLOWSIC600-XT · Service Manual · 8019179/V0-2/2016-03 · © SICK Engineering GmbH

Components Subject to change without notice 2.1 Complete overview of the device family 2.1.1 FLOWSIC600-XT Fig. 1 Exploded view FLOWSIC600-XT 100 Meter body 300 Sensor 120 Meter body screws 320 Sensor assembly set 140 Pressure compensation element 340 Sensor cover 190 Meter body type plate 400 Pressure sensor 200 SPU 410 pT sensor set 240 SPU neck assembly set 420 pT sensor cover 250 SPU neck

FLOWSIC600-XT · Service Manual · 8019179/V0-2/2016-03 · © SICK Engineering GmbH

15 Subject to change without notice 2.1.2 FLOWSIC600-XT Forte Fig. 2 Exploded view FLOWSIC600-XT Forte 100 Meter body 300 Sensor 120 Meter body screws 320 Sensor assembly set 140 Pressure compensation element 340 Sensor cover 190 Meter body type plate 400 Pressure sensor 200 SPU 410 pT sensor set 240 SPU neck assembly set 420 pT sensor cover 250 SPU neck

FLOWSIC600-XT · Service Manual · 8019179/V0-2/2016-03 · © SICK Engineering GmbH

Components Subject to change without notice 2.1.3 FLOWSIC600-XT 2plex Fig. 3 Exploded view FLOWSIC600-XT 2plex 100 Meter body 300 Sensor 120 Meter body screws 320 Sensor assembly set 140 Pressure compensation element 340 Sensor cover 190 Meter body type plate 400 Pressure sensor 200 SPU 410 pT sensor set 240 SPU neck assembly set 420 pT sensor cover 250 SPU neck

FLOWSIC600-XT · Service Manual · 8019179/V0-2/2016-03 · © SICK Engineering GmbH

17 Subject to change without notice 2.1.4 FLOWSIC600-XT Quatro Fig. 4 Exploded view FLOWSIC600-XT Quatro 100 Meter body 300 Sensor 120 Meter body screws 320 Sensor assembly set 140 Pressure compensation element 340 Sensor cover 190 Meter body type plate 400 Pressure sensor 200 SPU 410 pT sensor set 240 SPU neck assembly set 420 pT sensor cover 250 SPU neck

FLOWSIC600-XT · Service Manual · 8019179/V0-2/2016-03 · © SICK Engineering GmbH

Components Subject to change without notice 2.2 SPU (Signal Processing Unit) overview The SPU enclosure of the FLOWSIC600-XT comprises a flameproof enclosure and an adjacent separate chamber. 2.2.1 Ex-d enclosure version Fig. 5 Exploded view Ex-d 1 Flameproof enclosure with I/O electronics 2 Ex-d terminal compartment cover 3 Ex-e terminal compartment 4 Ex-e terminal compartment cover 5 Ex-i transducer electronics with cover and backup battery 6 Display unit

FLOWSIC600-XT · Service Manual · 8019179/V0-2/2016-03 · © SICK Engineering GmbH

19 Subject to change without notice 2.2.2 Ex-de enclosure version Fig. 6 Exploded view Ex-de With Ex-e wiring, the Ex-d inputs and outputs run through a line duct to the Ex-e terminals in the Ex-e terminal compartment. The type of the line duct depends on the I/O configuration. 1 2 1 Flameproof enclosure with I/O electronics 2 Ex-d terminal compartment cover 3 Ex-e terminal compartment 4 Ex-e terminal compartment cover 5 Ex-i transducer electronics with cover and backup battery 6 Display unit 3 4 5 6

FLOWSIC600-XT · Service Manual · 8019179/V0-2/2016-03 · © SICK Engineering GmbH

Components Subject to change without notice 2.2.3 Ex-i enclosure version Fig. 7 Exploded view Ex-i 1 2 3 4 5 6 1 Flameproof enclosure 2 Ex-d terminal compartment cover 3 Ex-i terminal compartment 4 Ex-i terminal compartment cover 5 Ex-i transducer electronics with cover and backup battery 6 Display unit

FLOWSIC600-XT · Service Manual · 8019179/V0-2/2016-03 · © SICK Engineering GmbH

21 Subject to change without notice FLOWSIC600-XT 3 Repairs General information Tools required Working on the meter body Working on the SPU

FLOWSIC600-XT · Service Manual · 8019179/V0-2/2016-03 · © SICK Engineering GmbH

Repairs Subject to change without notice 3.1 General information The work described in the following may only be carried out by qualified personnel accord- ing to → p. 9, §1.2.2 and after training by the manufacturer. Repairs not authorized by the manufacturer result in the loss of possibly existing warranty claims. SICK recommends backing up all system parameters before component disassembly/ assembly or replacement. 3.2 Prerequisites for repair work ▸ Record all work carried out in a repair protocol. ▸ Send the repair protocol to SICK to update the Data books at the factory. WARNING: Hazards due to combustible gases or high pressure Gas under line pressure flows through the FLOWSIC600-XT when in operation. The pipeline at the installation site must be free from pressure and vented before replacing the ultrasonic transducers. Exception: Extraction tools, Handling in accordance with the Operating Instruc- tions for the extraction tool Before starting work: ▸ Ensure the pipeline is free from pressure and free from combustible gases. ▸ Purge the pipeline with inert gas if necessary. ▸ Observe the safety information in §1.1 and §3.1 of the Operating Instruc- tions. WARNING: Danger through power voltage ▸ Disconnect the measuring system from the power supply before starting work on the FLOWSIC600-XT! ▸ When using the FLOWSIC600-XT in an Ex zone, the operating voltage must be switched off in the safe zone (switching off in the Ex zone is only allowed with instrinsically safe power supply).

FLOWSIC600-XT · Service Manual · 8019179/V0-2/2016-03 · © SICK Engineering GmbH

23 Subject to change without notice 3.3 Tools required ● SW 3 Allen key ● SW 3 Allen key ● SW 5 Allen key ● SW 4 socket wrench ● Phillips screwdriver, size 1 ● Slot screwdriver, size 2 (blade width 3 mm) ● Slot screwdriver, size 3 (blade width 6 mm) ● Pointed pliers ● Snap ring pliers ● Sensor puller (Part No. 7041772) ● Coax connection tool (Part No. 4047938) ● SW 22 torque wrench, setting range up to 10 Nm) ● Adjustable wrench (8") ● Wrench, 19 mm

FLOWSIC600-XT · Service Manual · 8019179/V0-2/2016-03 · © SICK Engineering GmbH

Repairs Subject to change without notice 3.4 Working on the meter body 3.4.1 Replacing the sensors 3.4.1.1 Remove the sensors WARNING: Hazards due to combustible gases or high pressure The pipeline at the installation site must be free from pressure and vented before replacing the ultrasonic transducers. 1 Take the sensor cover off: Loosen the 4 screws (captive) with an SW 4 Allen key. 2 Pull the sensor cable carefully off the sensor by hand. Only apply even tension when pulling off. Important: Keep the sensor cable plug clean, make sure it does not come into contact with greasy sub- stances (e.g. assembly paste). Protect the plug against dirt as necessary, e.g. with insulating tape. Ensure the requirements for explosion protection are complied with 3 Loosen the retaining screws with a size 19 bit (ratchet or socket wrench).

FLOWSIC600-XT · Service Manual · 8019179/V0-2/2016-03 · © SICK Engineering GmbH

25 Subject to change without notice 4 Loosen the lock washer with a small flat-blade screwdriver and take out. 5 Screw the sensor puller onto the sen- sor socket. Screw on carefully (not too tight, it is sufficient when the sensor puller is steady). 6 Turn the knurled nut until the sensor can be pulled out of the sensor sup- port. 7 Unscrew the sensor from the sensor puller

FLOWSIC600-XT · Service Manual · 8019179/V0-2/2016-03 · © SICK Engineering GmbH

Repairs Subject to change without notice 3.4.1.2 Fit the new sensor 1 Unpack the new sensor and check for transport damage. Do not use the sensor when any damage can be seen at the sensor or the O-ring seal. 2 Screw the new sensor onto the assembly tool. 3 Grease the O-ring with lubricant. Important: – Do not let lubricant into the area with media contact. – Step 3 does not apply to oxygen applications, do not use lubri- cant! 4 Insert the sensor into the sensor support until the resistance of the O- ring seal can be felt. 5 Fix the sensor carefully with one light axial circular movement. Important: The O-ring can be easily damaged. Only position once and do not tilt! 6 Unscrew the assembly tool from the sensor. If the sensor turns as well, press the assembly tool slightly from the top and carefully continue to screw. 7 Refit the lock washer. Make sure the washer does not tilt otherwise the retaining screw cannot be completely screwed in. 8 Screw the retaining screw in by hand 9 Then tighten with torque 20 Nm. Check the position of the manufac- turer's marking: – The retaining screw has been screwed in completely when the markings are aligned. – If the markings cannot be aligned, remove the retaining screw again and check the lock washer seating. 10 After sensor replacement, perform a leak tightness check. Apply leak detection spray to the sensors and slowly increase the pressure in the device.

FLOWSIC600-XT · Service Manual · 8019179/V0-2/2016-03 · © SICK Engineering GmbH

27 Subject to change without notice 11 Push the sensor cable into the con- nection tool 12 Plug the sensor plug onto the sensor connection according to the cable assignment in §5.1 (→ p. 64). The plug must engage audibly. 13 Make a tension test: The plug is engaged correctly when the cable cannot be pulled out with a slight pull. 14 Refit the sensor cover: – First position the cover side with the cable gland. – Make sure the cable is not crimped under the cover edge: No gap should be visible between the cover and meter body. 15 First slightly tighten the 4 screws with an SW 4 Allen key. 16 Then tighten crosswise by hand (5 Nm).

FLOWSIC600-XT · Service Manual · 8019179/V0-2/2016-03 · © SICK Engineering GmbH

Repairs Subject to change without notice 3.4.2 Replacing the integrated p/T sensor 3.4.2.1 Remove the p/T sensor 1 Take the sensor cover off: Loosen the 4 screws (loss-secure) with an SW 4 Allen key. 2 Carefully pull the cable out (locking lug upwards) 3 Then disconnect the plug. Make sure that the cables are not pulled out the plug. 4 Push the sensor cable with plug into the socket wrench 5 Position the socket wrench and turn the PT sensor out. 6 Take the PT sensor out of the meter body.

FLOWSIC600-XT · Service Manual · 8019179/V0-2/2016-03 · © SICK Engineering GmbH

29 Subject to change without notice 3.4.2.2 Fit the new p/T sensor 1 Unpack the new sensor and check for transport damage. Do not use the sensor when any damage can be seen at the sensor or the seal. 2 Position the p/T sensor in the SW 22 socket wrench (push the cable in first). 3 Grease the thread (separating agent that does not rub on stainless steel). 4 Insert the sensor as straight as pos- sible and screw tight (tightening torque: 20 Nm). 5 After sensor replacement, perform a leak tightness check. Apply leak detection spray to the sensors and slowly increase the pressure in the device. 6 Connect the plug connection again. 7 Fold the sensor cable as small as possible so that it is not crimped under the sensor cap.

FLOWSIC600-XT · Service Manual · 8019179/V0-2/2016-03 · © SICK Engineering GmbH

Repairs Subject to change without notice 8 Position the sensor cover from the top downwards. 9 First slightly tighten the 4 screws. 10 Then tighten crosswise hand-tight (5 Nm).

FLOWSIC600-XT · Service Manual · 8019179/V0-2/2016-03 · © SICK Engineering GmbH

31 Subject to change without notice 3.5 Working on the SPU 3.5.1 Replacing the display unit 3.5.1.1 Swivel the display unit downwards 1 Open the display cover. 2 Loosen the 4 screws on the display unit with an SW 3 Allen key. 3 Carefully swivel the display unit downwards.

FLOWSIC600-XT · Service Manual · 8019179/V0-2/2016-03 · © SICK Engineering GmbH

Repairs Subject to change without notice 3.5.1.2 Remove the display unit 1 Disconnect the connection cable for external power supply and backup battery. 2 Disconnect the connection cables of the display unit from the electronics unit: To do this, carefully lift the plugs with a small flat-blade screwdriver. Disconnect the plugs carefully with even tension on the cables. 3 Loosen the retaining screws on the swivel joint of the electronics unit with a Phillips screwdriver. When swiveled, the display unit and display cover are pressed against the swivel joint when in the lowest posi- tion. This obviates any additional pro- tection against the unit dropping down during the work step.

FLOWSIC600-XT · Service Manual · 8019179/V0-2/2016-03 · © SICK Engineering GmbH

33 Subject to change without notice 3.5.1.3 Fit the new display unit 4 Swivel the display unit and display cover to a vertical position and remove upwards out of the swivel holder. 1 Unpack the new display unit and check for transport damage. Do not use the display unit when any damage can be seen. 2 Position the display unit vertically in the swivel holder and swivel down. 3 Position the retaining screws and small plates on the swivel joint of the electronics and screw tight. Check the small retaining plates are positioned correctly (straight side towards the hinge).

FLOWSIC600-XT · Service Manual · 8019179/V0-2/2016-03 · © SICK Engineering GmbH

Repairs Subject to change without notice 3.5.1.4 Connect the display unit 3.5.1.5 Swivel the display unit upwards and screw tight 1 Connect the display cables. 2 Connect the connection for external power supply to the “BAT1/Ext. Power” connection. 3 Insert the backup battery and con- nect to the “BAT2” connection. Make sure the external power supply is active before connecting the backup battery. Otherwise the backup battery is immediately active. 1 Swivel the display unit upwards. 2 Tighten the 4 screws on the display unit with an SW 4 Allen key (5 Nm).

FLOWSIC600-XT · Service Manual · 8019179/V0-2/2016-03 · © SICK Engineering GmbH

35 Subject to change without notice 3.5.2 Replacing the display cover Prerequisites ▸ The display unit has been removed: – Swivel the display unit downwards, → p. 31, §3.5.1.1. – Remove the display unit, → p. 32, §3.5.1.2. Replace the display cover 1 Loosen the pins on the display cover. 2 Take the display cover out of the swivel joint. 3 Unpack the new display cover and check for transport damage. Do not use the display cover when any dam- age can be seen. 4 Position the display cover in the swivel joint. 5 Fix the new display cover with the pins. 6 Refit the display unit, → p. 33, §3.5.1.3.

FLOWSIC600-XT · Service Manual · 8019179/V0-2/2016-03 · © SICK Engineering GmbH

Repairs Subject to change without notice 3.5.3 Replacing the backup battery Prerequisites: ▸ The display unit is swiveled down: – Swivel the display unit downwards, → p. 31, §3.5.1.1. 3.5.3.1 Remove the backup battery 3.5.3.2 Insert the new backup battery 1 Disconnect the backup battery con- nection. 2 Take the backup battery out of the holder. 1 Unpack the new backup battery and check for transport damage. 2 Do not use the backup battery when any damage can be seen. 3 Make sure the external power supply is active before connecting the backup battery. Otherwise the backup battery is immediately active. 4 Insert the backup battery in the holder and connect to the “BAT2” connection. 5 Swivel the display unit back up and screw tight, → p. 34, §3.5.1.5.

FLOWSIC600-XT · Service Manual · 8019179/V0-2/2016-03 · © SICK Engineering GmbH

37 Subject to change without notice 3.5.4 Replacing the RTC battery Prerequisites ▸ The display unit is swiveled down: – Swivel the display unit downwards, → p. 31, §3.5.1.1. 3.5.4.1 Replace the RTC battery 1 Loosen the battery cover screw with a Phillips screwdriven. 2 Remove the battery cover. 1 Take the RTC battery out of the holder 2 Insert the new battery. 3 Refit the battery cover. 4 Swivel the display unit back up and screw tight, → p. 34, §3.5.1.5.

FLOWSIC600-XT · Service Manual · 8019179/V0-2/2016-03 · © SICK Engineering GmbH

Repairs Subject to change without notice 3.5.5 Replacing the SPU circuit board Prerequisites ▸ The display unit is swiveled down: – Swivel the display unit downwards, → p. 31, §3.5.1.1. ▸ The connection cables of the display unit and the connections for external power supply and backup battery are disconnected: – Remove the display unit, → p. 32, §3.5.1.2, step 1 and 2 3.5.5.1 Remove the electronics cover 1 Loosen the 4 screws of the electron- ics cover with a T20 Torx key. 2 Take the electronics cover off.

FLOWSIC600-XT · Service Manual · 8019179/V0-2/2016-03 · © SICK Engineering GmbH

39 Subject to change without notice 3.5.5.2 Remove the SPU circuit board 1 Loosen the 5 screws of the SPU cir- cuit board with a T10 Torx key. 2 Carefully tilt the circuit board to the front. Take care that the power connec- tions on the front side of the circuit board do not impact against the enclosure and get damaged. 3 Pull the circuit board out until the connections of the interface plugs on the top edge of the circuit board can be accessed. 4 Disconnect the connections of the interface plugs. 5 Pull the circuit board out further so that the transducer connections on the rear side of the circuit board can be accessed. 6 Carefully loosen the transducer con- nections with a flat-blade screw- driver. Too much force when loosening the connecting plugs can damage the cable connections or the cables. Replace cables when any damage can be seen. 7 Carefully remove the SPU circuit board. – Take care not to damage the con- nections on the front side of the circuit board.

FLOWSIC600-XT · Service Manual · 8019179/V0-2/2016-03 · © SICK Engineering GmbH

Repairs Subject to change without notice 3.5.5.3 Fit the new SPU circuit board 1 Unpack the new SPU circuit board and check for transport damage. Do not use the SPU circuit board when any damage can be seen. 2 Carefully insert the new SPU circuit board. Take care that the power connec- tions on the front side of the circuit board do not impact against the enclosure and get damaged during all work steps. 3 Connect the transducer cables on the rear side of the circuit board according the wir- ing diagrams in §5.1 (→ p. 64). 4 Connect the interface plugs on the rear side of the circuit board accord- ing to the markings on the circuit board and the plugs. 5 Insert the circuit board. 6 Screw the 5 retaining screws tight (1 Nm) with T10 Torx key.

FLOWSIC600-XT · Service Manual · 8019179/V0-2/2016-03 · © SICK Engineering GmbH

41 Subject to change without notice 3.5.5.4 Fit the electronics cover 1 Position the electronics cover. 2 Lead the cable of the external power supply through the recess in the cover. Make sure the cable is not crimped. 3 Tighten the 4 screws of the electron- ics cover with a T20 Torx key. 4 Connect the display unit again and close. → p. 34, §3.5.1.4.

FLOWSIC600-XT · Service Manual · 8019179/V0-2/2016-03 · © SICK Engineering GmbH

Repairs Subject to change without notice 3.5.6 Replacing the Ex-d I/O block 3.5.6.1 Open the Ex-d terminal compartment 3.5.6.2 Remove the Ex-d I/O block 1 Loosen the securing screw on the Ex- d terminal compartment cover with an SW 10 Allen key. 2 Unscrew the clamping compartment cover. A wrench can be applied in the grooves, for example, to loosen the compartment cover. 1 Disconnect terminal connections in use on the Ex-d electronics block. 2 Loosen the 3 screw fittings of the I/O block with a Phillips screwdriver.

FLOWSIC600-XT · Service Manual · 8019179/V0-2/2016-03 · © SICK Engineering GmbH

43 Subject to change without notice 3 Pull the I/O block out with the tab provided. – Take care not to damage the cable gland in the Ex-e terminal compartment. – Carefully push the cables upwards and move the I/O block past the cables.

FLOWSIC600-XT · Service Manual · 8019179/V0-2/2016-03 · © SICK Engineering GmbH

Repairs Subject to change without notice 3.5.6.3 Fit the new Ex-d I/O block 1 Unpack the new I/O block and check for transport damage. Do not use the I/O block when any damage can be seen. 2 Carefully guide the I/O block past the Ex-e cables and plug into the main board. Take care not to bend any pins. 3 Insert the 3 screws and screw tight. Take care that the cutting ring washers do not drop down. 4 Connect the electrics as described in §3.4 “Electrical Installation” in the FLOWSIC600-XT Operating Instructions.

FLOWSIC600-XT · Service Manual · 8019179/V0-2/2016-03 · © SICK Engineering GmbH

45 Subject to change without notice 3.5.6.4 Close the Ex-d terminal compartment 1 Screw the terminal compartment cover back on. 2 Insert the securing screw on the Ex-d terminal compartment cover and screw tight with an SW 10 Allen key. Do not operate the device without the securing screw!

FLOWSIC600-XT · Service Manual · 8019179/V0-2/2016-03 · © SICK Engineering GmbH

Repairs Subject to change without notice 3.5.7 Replacing the Ex-d I/O block fuse Prerequisites ▸ The Ex-d terminal compartment is open: – Open the Ex-d terminal compartment, → p. 42, §3.5.6.1 Replace the fuse 1 Remove the fuse. 2 Unpack the new fuse and check for transport damage. Do not use the fuse unit when any damage can be seen. 3 Insert the new fuse. Take care not to bend the pins. 4 Make sure the fuse has a tight seat. 5 Close the Ex-d terminal compartment again → p. 45, §3.5.6.4.

FLOWSIC600-XT · Service Manual · 8019179/V0-2/2016-03 · © SICK Engineering GmbH

47 Subject to change without notice 3.5.8 Replacing the Ex-e connection block 3.5.8.1 Open the Ex-e terminal compartment 1 Loosen the 3 screws (captive) of the Ex-e terminal compartment cover with an SW 4 Allen key. 2 Put the terminal compartment cover in the holder provided.

FLOWSIC600-XT · Service Manual · 8019179/V0-2/2016-03 · © SICK Engineering GmbH

Repairs Subject to change without notice 3.5.8.2 Remove the connection blocks in the Ex-e terminal compartment. 3.5.8.3 Fit the new connection block in the Ex-e terminal compartment 1 Disconnect the connections in the Ex-e terminal compartment. 2 Loosen the screws of the front bracket of the connection block with a T10 Torx key and remove the bracket. Loosen the screws of the rear bracket. 3 Carefully take the Ex-e connection block out. 1 Unpack the new connection block and check for transport damage. Do not use the connection block when any damage can be seen. 2 Carefully position the new Ex-e con- nection block. 3 Fit the second bracket and fix with the screws (1 Nm). 4 Connect the electrics as described in §3.4 “Electrical Installation” in the FLOWSIC600-XT Operating Instructions.

FLOWSIC600-XT · Service Manual · 8019179/V0-2/2016-03 · © SICK Engineering GmbH

49 Subject to change without notice 3.5.8.4 Close the Ex-e terminal compartment 1 Position the cover on the Ex-e termi- nal compartment. 2 Tighten the 3 screws (captive) of the terminal compartment cover with an SW 4 Allen key (5 Nm).

FLOWSIC600-XT · Service Manual · 8019179/V0-2/2016-03 · © SICK Engineering GmbH

Repairs Subject to change without notice 3.5.9 Replacing the cable gland between the Ex-d and Ex-e terminal compartments Prerequisites ▸ Ex-e terminal compartment is open – Open the Ex-e terminal compartment, → p. 47, §3.5.8.1 ▸ The Ex-d I/O block has been removed: – Open the Ex-d terminal compartment, → p. 42, §3.5.6.1 – Remove the Ex-d I/O block, → p. 42, §3.5.6.2 3.5.9.1 Remove the cable gland 1 Disconnect the terminals in the Ex-e terminal compartment; these can be spring-loaded or screw terminals (Figure shows an example). To create space for the repair work, the connection blocks can be removed, → p. 48, §3.5.8.2. 2 Loosen the snap ring in the Ex-e ter- minal compartment with the snap ring pliers. Hold the cable gland tight when loos- ening the snap ring so that it does not slip down. 3 Remove the cable gland downwards through the Ex-d terminal compart- ment.

FLOWSIC600-XT · Service Manual · 8019179/V0-2/2016-03 · © SICK Engineering GmbH

51 Subject to change without notice 3.5.9.2 Fit the Ex-e cable gland 1 Unpack the new cable gland and check for transport damage. Do not use the cable gland when any damage can be seen. 2 Guide the cable gland through the Ex-d terminal compartment. 3 Fit the snap ring. 4 Make sure the snap ring engages properly in the groove. 5 Connect the electrics as described in §3.4 “Electrical Installation” in the FLOWSIC600-XT Operating Instructions. 6 Close the Ex-e terminal compartment again, → p. 49, §3.5.8.4.

FLOWSIC600-XT · Service Manual · 8019179/V0-2/2016-03 · © SICK Engineering GmbH

Repairs Subject to change without notice 3.5.10 Replacing the cable gland between the Ex-d terminal compartment and the Ex-i transducer electronics Prerequisites ▸ The SPU circuit board has been removed: – Swivel the display unit downwards, → p. 31, §3.5.1.1. – Remove the electronics cover, → p. 38, §3.5.5.1. – Remove the SPU circuit board, → p. 39, §3.5.5.2 ▸ The Ex-d I/O block has been removed: – Open the Ex-d terminal compartment, → p. 42, §3.5.6.1. – Remove the Ex-d I/O block, → p. 42, §3.5.6.2. 3.5.10.1 Remove the cable gland 1 Loosen the 3 long screws of the main board. 2 Loosen the snap ring of the cable gland with the snap ring pliers. 3 Remove the main board with cable gland.

FLOWSIC600-XT · Service Manual · 8019179/V0-2/2016-03 · © SICK Engineering GmbH

53 Subject to change without notice 3.5.10.2 Fit the new cable gland 1 Unpack the new cable gland and check for transport damage. Do not use the cable gland when any damage can be seen. 2 Position the cable gland. Use a small screwdriver as aid when necessary. 3 Fix the main board with the 3 long screws. 4 Fit the snap ring. 5 Make sure the snap ring engages properly in the groove. 6 Refit the Ex-d I/O block,→ p. 44, §3.5.6.3. 7 Refit the SPU circuit board, → p. 40, §3.5.5.3.

FLOWSIC600-XT · Service Manual · 8019179/V0-2/2016-03 · © SICK Engineering GmbH

Repairs Subject to change without notice 3.6 Parameter input 3.6.1 Configure sensor settings after transducer exchange In order to configure settings after exchanging a transducer pair, follow these steps: 1 Open a FLOWSIC600-XT meter using the Device Manager with user level "Service" or higher. 2 Navigate to "Service"/"Register Tree". 3 For each exchanged transducer select the appropriate branch underneath "Flow/Path plane X/Path X-Y" and modify the following values according to the "Data Sheet Trans- ducer Pair": – Register "Sensor Serial X-Y Position". – Register "Sensor Length X-Y Position". – Register "TimeOffset X-Y". – Register "TimeDifference". – Register "DiffractionTime". 4 Confirm each modification by clicking "Write". 5 Repeat each above step for the each exchanged transducer. 6 Disable configuration mode, if required. 7 If required, generate a new Parameter Protocol" in "Parameter Modification". 3.6.2 SOS Check In order to compare theoretical and measured SOS, follow these steps For a detailed description, please refer to " FLOWgate Software Manual": 1 Open a FLOWSIC600-XT meter using the device manager with user level "Service" or higher. 2 Navigate to “Diagnosis/SOS Calculator”. 3 Select the “Gas Composition” and click "Apply" 4 Calculate the theoretical speed of sound. 5 Open the trend chart. – The theoretical speed of sound is visualized as a horizontal red line. – The measured speed of sound for each path is plotted along the timeline in different colors.

FLOWSIC600-XT · Service Manual · 8019179/V0-2/2016-03 · © SICK Engineering GmbH

55 Subject to change without notice 3.6.3 Configure Stacking In order to configure Stacking of signals, please follow these steps: 1 Open a FLOWSIC600-XT meter using the device manager with user level "Service" or higher. 2 Navigate to "Diagnostics"/"Signal View". 3 Ensure, that configuration mode is set on meter. 4 In the section "Parameters" select the appropriate stacking size 1, 2, 4, 8, or 16. 5 Click "Write to device" in the toolbar. 6 Disable configuration mode, if required. 3.6.4 Configure Zerophase In order to configure or adjust the zerophase of a path, follow these steps: 1 Open a FLOWSIC600-XT meter using the Device Manager with user level "Service" or higher. 2 Navigate to "Diagnostics"/"Signal View". 3 Ensure, that configuration mode is set on meter. 4 Select a path to configure. 5 Identify the current zero phase error inside the panel "Signal Values" identified by "0- Error" 6 Inside Panel "Parameter", please modify the parameter "Phaseshift" by adding the 0- Error to the current value of "Phase Shift" 7 Click "Write to device" in the toolbar. 8 Repeat steps 4-7 for each required path and direction. 9 Disable configuration mode, if required. 3.6.5 Configure Pfix and Tfix In order to configure settings for Pfix or Tfix, follow these steps: 1 Open a FLOWSIC600-XT meter using the Device Manager with user level "Service" or higher. 2 Navigate to "Parameter Modification"/"Pressure/Temperature Sensors". 3 Ensure, that configuration mode is set on meter. 4 For Pfix value, please modify "Replacement and fix value" inside panel "Pressure value source settings". 5 For Tfix value, please modify "Replacement and fix value" inside panel "Temperature value source settings". 6 Click "Write to device" in the toolbar. 7 Disable configuration mode, if required.

FLOWSIC600-XT · Service Manual · 8019179/V0-2/2016-03 · © SICK Engineering GmbH

Repairs Subject to change without notice

Accessories and Spare Parts

FLOWSIC600-XT · Service Manual · 8019179/V0-2/2016-03 · © SICK Engineering GmbH 57 Subject to change without notice FLOWSIC600-XT 4 Accessories and Spare Parts SPU Parts Electronic boards Transducers Pressure/Temperature sensors O-rings for transducers Electronics accessories Transducer covers Covers p/T sensor Accessories meter body

FLOWSIC600-XT · Service Manual · 8019179/V0-2/2016-03 · © SICK Engineering GmbH

Accessories and Spare Parts Subject to change without notice 4.1 SPU Parts 4.2 Electronic boards Order number Description 2085275 Ex-d SPU housing; metric M25; incl. Backplane and Cable feedthrough 2085333 Ex-d SPU housing; imperial 3/4" NPT; incl. Backplane and Cable feedthrough 2085282 Ex-e SPU housing; metric M25; incl. Backplane and Cable feedthrough 2085332 Ex-e SPU housing; imperial 3/4" NPT; incl. Backplane and Cable feedthrough 2085283 Ex-i SPU housing; metric M25; incl. Backplane and Cable feedthrough 2085331 Ex-i SPU housing; imperial 3/4" NPT; incl. Backplane and Cable feedthrough 2085284 Terminal compartment Ex-d 2085285 Terminal compartment cover Ex-e /Ex-i 2085287 Display flap 2085286 Display unit with flap 2085290 Electronik cover 2085288 Display connection cable 2085289 Keder 2085300 Cable feedthrough SPU -- I/O 2085301 Cable feedthrough Ex-d -- Ex-e 2085303 Screw type terminal Ex-e 2085304 Spring type terminal Ex-e 2085320 Electronics neck; single; screw in 2085321 Electronics neck; single; screw in; extended 2085322 Electronics neck; Y-type; screw in; extended 2085323 Electronics neck; single 2085324 Electronics neck; single; extended 2085325 Electronics neck; Y-type Order number Description 2085291 Electronic board; 200kHz; IIA; 8path 2085292 Electronic board; 200kHz; IIC; 8path 2085293 Electronic board; 200kHz; IIC; 4path 2085294 Electronic board; 300kHz; IIA; 8path 2085295 Electronic board; 135kHz; IIA; 8path 2085296 Electronic board; 135kHz; IIC; 8path 2085297 Electronic board; 135kHz; IIA; 8path; HS 2085298 Electronic board; 80kHz; IIA; 8path 2085315 I/O unit with Ethernet 2085305 I/O unit with Ethernet 2085302 Fuse I/O unit

Accessories and Spare Parts

FLOWSIC600-XT · Service Manual · 8019179/V0-2/2016-03 · © SICK Engineering GmbH 59 Subject to change without notice 4.3 Transducers 4.4 Pressure/Temperature sensors 4.5 O-rings for transducers 4.6 Electronics accessories Order number Description 2080032 Transducer pair, type 210 2080027 Transducer pair, type 218 7041787 Transducer pair, type S2 (200 kHz, 100 bar, 120 C, titanium, P18) 2039997 Transducer pair, type 22 (200 kHz, 250 bar, 120 C, titanium, P18) 2039441 Transducer pair, type 26 (200 kHz, 250 bar, 120 C, titanium, P10) 2032538 Transducer pair, type 28 (135 kHz, 250 bar, 180 C, titanium, P18) 7042607 Transducer pair, type S6 (200 kHz, 100 bar, 120 C, titanium, P10) 2034125 Transducer pair, type S8 (135 kHz, 100 bar, 180 C, titanium, P18) 7042604 Transducer pair, type S3 (135 kHz, 16 bar, 280 C, titanium, P14) On request Transducer pair, type S5 (300 kHz, 16 bar, 280 C, titanium, P10) 7042605 Transducer pair, type S4 (135 kHz, 100 bar, 180 C, titanium, P14) 2040391 Transducer pair, type 46 (200 kHz, 450 bar, 120 C, titanium, P14) 2040392 Transducer pair, type S7 (135 kHz, 16 bar, 180 C, titanium, P18) 2032592 Transducer pair, type M6 (200 kHz, 10 bar, 120 C, titanium/ +layer, P10) 2040393 Transducer pair, type N4 (135 kHz, 40 bar, 180 C, NIROSTA, P14) 2040394 Transducer pair, type L8 (135 kHz, 100 bar, 280 C, titanium, P18) Order number Description 2083110 p/T sensor I (20 bar) 2083118 p/T sensor II (110 bar) 2083119 p/T sensor III (160 bar) Order number Description 2085270 O-ring Set Natural gas 15,0 * 2,0 VITON LT170-TT 2085271 O-ring Set Oxygen 2085272 O-ring Set Hydrogen 2085273 O-ring Set Carbon dioxide 2085274 O-ring Set Natural gas 7,5 * 1,5 VITON LT170-TT Order number Description 2085326 Tool kit 2085327 Tamper proof protection set 2085328 Set cable connection I/O unit 6050602 Interface converter USB/IR 2085329 Set cable glands / blind plugs Ex-e 2085330 Backup battery unit 2085319 RTC backup battery On request Set of screws for Fl600-XT SPU housing

FLOWSIC600-XT · Service Manual · 8019179/V0-2/2016-03 · © SICK Engineering GmbH

Accessories and Spare Parts Subject to change without notice 4.7 Transducer covers Order number Description 2083170 Transducer cover, Al, 3" 2083172 Transducer cover, Al, 4" 2085207 Transducer cover, Al, 4"; exchangeable version 2085202 Transducer cover, Al, 6"-A 2085201 Transducer cover, Al, 6"-B 2085203 Transducer cover, Al, 8"-A 2085204 Transducer cover, Al, 8"-B 2085205 Transducer cover, Al, 10"-12"-A 2085206 Transducer cover, Al, 10"-12"-B 2085208 Transducer cover, Al, 14"-A 2085209 Transducer cover, Al, 14"-B 2085214 Transducer cover, Al, 16"-A 2085215 Transducer cover, Al, 16"-B 2085216 Transducer cover, Al, 18"-A 2085217 Transducer cover, Al, 18"-B 2085218 Transducer cover, Al, 20"-A 2085219 Transducer cover, Al, 20"-B 2085220 Transducer cover, Al, 22"-A 2085221 Transducer cover, Al, 22"-B 2085222 Transducer cover, Al, 24"-A 2085223 Transducer cover, Al, 24"-B 2083171 Transducer cover, SS, 3" 2083174 Transducer cover, SS, 4" 2085224 Transducer cover, SS, 4"; exchangeable version 2085225 Transducer cover, SS, 6"-A 2085226 Transducer cover, SS, 6"-B 2085227 Transducer cover, SS, 8"-A 2085228 Transducer cover, SS, 8"-B 2085229 Transducer cover, SS, 10"-12"-A 2085230 Transducer cover, SS, 10"-12"-B 2085231 Transducer cover, SS, 14"-A 2085232 Transducer cover, SS, 14"-B 2085233 Transducer cover, SS, 16"-A 2085234 Transducer cover, SS, 16"-B 2085235 Transducer cover, SS, 18"-A 2085236 Transducer cover, SS, 18"-B 2085237 Transducer cover, SS, 20"-A 2085238 Transducer cover, SS, 20"-B 2085239 Transducer cover, SS, 22"-A 2085240 Transducer cover, SS, 22"-B 2085241 Transducer cover, SS, 24"-A 2085242 Transducer cover, SS, 24"-B

Accessories and Spare Parts

FLOWSIC600-XT · Service Manual · 8019179/V0-2/2016-03 · © SICK Engineering GmbH 61 Subject to change without notice 4.8 Covers p/T sensor 4.9 Accessories meter body Order number Description 2083181 p-/T transducer cover 4"- 6"; aluminum 2083179 p-/T transducer cover 8"-16"; aluminum 2083182 p-/T transducer cover 4"- 6"; stainless steel 2083180 p-/T transducer cover 8"-16"; stainless steel Order number Description On request Meterbody parts for transport and storage On request Set of screws for Fl600-XT meter body

FLOWSIC600-XT · Service Manual · 8019179/V0-2/2016-03 · © SICK Engineering GmbH

Accessories and Spare Parts Subject to change without notice

FLOWSIC600-XT · Service Manual · 8019179/V0-2/2016-03 · © SICK Engineering GmbH

63 Subject to change without notice FLOWSIC600-XT 5 Annex Component placement specification SPU (sensor cables)

FLOWSIC600-XT · Service Manual · 8019179/V0-2/2016-03 · © SICK Engineering GmbH

Annex Subject to change without notice 5.1 Component placement specification SPU (sensor cables) Bild 8 4-Path devices: FLOWSIC600-XT (Single), FLOWSIC600-XT 2plex and Quatro, measuring system 1 ������������������ p&T 1 2 3 4 6 5 7 8 4 2 3 1 5 6 7 8 ��� ��� ��� ��� ��� ��� ��� ���

FLOWSIC600-XT · Service Manual · 8019179/V0-2/2016-03 · © SICK Engineering GmbH

65 Subject to change without notice Bild 9 FLOWSIC600-XT 2plex, measuring system 2 (1-Path) ������������������ p&T 1 5 1 5 ��� ���

FLOWSIC600-XT · Service Manual · 8019179/V0-2/2016-03 · © SICK Engineering GmbH

Annex Subject to change without notice Bild 10 FLOWSIC600-XT Quatro, measuring system 2 ������������������ p&T 1 2 3 4 6 5 7 8 4 2 3 1 5 6 7 8 ��� ��� ��� ��� ��� ��� ��� ���

FLOWSIC600-XT · Service Manual · 8019179/V0-2/2016-03 · © SICK Engineering GmbH

67 Subject to change without notice Bild 11 FLOWSIC600-XT Forte, measuring system 1 and 2 � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � ������������������� � ������������������� ���� �� �� �� �� �� �� �� �� �� �� �� �� �� �� �� �� �� �� �� �� �� �� �� �� �� �� �� �� �� � �� �� ��� ��� ��� ��� ��� ��� ��� ��� ��� ���� ���� ���� ���� ���� ���� ���

United Arab Emirates

�������������������������� �������������������� USA/Mexico ���������������������� � ��������������������������� ��������������������� Vietnam �������������������� ����������������������������������� 8019179/V0-2/2016-03 SICK AG | Waldkirch | Germany | www.sick.com

Other Measuring Instruments Manuals