ABB 8241 8240 Series Analyzers
OUT OF SERVICE
ALARM 1
CAL
ALARM 2
HOLD
OUT OF SAMPLE
Phosphate EIL 8242
Clarification of an instruction or additional information.
Information.
Further reference for more detailed information or
technical details.
Although Warning hazards are related to personal injury, and Caution hazards are associated with equipment or property damage,
it must be understood that operation of damaged equipment could, under certain operational conditions, result in degraded
process system performance leading to personal injury o
INTRODUCTION ......................................................... 2
2
PREPARATION ........................................................... 2
2.1
Company Standard Settings ............................ 2
3
INSTALLATION ........................................................... 3
3.1
Serial Communication
Adaptors for Personal Computers ................... 3
3.1.1
Five-wire Configuration ...................... 3
ABB 8241 This Supplement must be read in conjunction with the
Instruction Manual supplied with the instrument:
Instrument
Manual Reference
•
Model 8241
Colorimetric Silica Monitor
–
IM/8241
•
Model 8242
Colorimetric Phosphate Monitor
–
IM/8242
For repeatable and reliable serial communication to take place
between a master (host computer) and slaves (instruments) it
is essential that the two conditions detailed in this section
Fig. 4.1 Schematic Diagram
Computer Terminal or
Host Computer
Chain of 8240 Monitors
(Connected)
Last 8240
in Chain
Termination Resistor
(on 8240's receive input)
Observe the limitations outlined in the installation information
of the relevant Instruction Manual. The maximum serial data
transmission line length for both RS422 and RS485 systems is
1200m.
3.1
Serial Communication
Adaptors for Personal Computers
An RS
ABB 8241 Fig. 4.2 Serial Connections
…4.1
Serial Connections
Motherboard
See Figs. 2.4 & 2.5 of the
relevant Instruction Manual
3
2
1
6
5
4
Output Board
Output Board
1 – Tx+
2 – Rx+
3 – Common (0V)
4 – Tx–
5 – Rx–
6 – No connection
A – Five-wire Connections
B – Three-wire Connections
3-wire
5-wire
IC32
IC31
IC33
IC34
C – Three- and Five-wire Links
6
7
1
12
1 – Tx+/Rx+
2 – Not Connected
3 – Common (0V)
4 �
ABB 8241 Fig. 5.1 Termination Resistor Links
For all aspects other than serial data transmission the
transmitter is set up as shown in the relevant Instruction
Manual. Unless otherwise requested, the instrument is
despatched with a transmission rate of 9600 baud and
transmission line termination resistors linked-out. If the
resistors are to be linked-in (see Fig. 5.1) carry out the
following Section.
5.1
Termination Resistors – Fig. 5.1
For long t
Fig. 6.1 Location of Serial Data Interface Page
The general programming procedure is as detailed in the relevant Instruction Manual, but with an additional Set Up Serial Interface
frame in the Set Up Instrument page.
6.1 Serial Interface Page
Refer to Section 6.3 (Set Up Instrument) in the relevant Instruction Manual.
CLOCK SETUP. PAGE 3.2
Date (DD/MM/YY) = xx/xx/xx
Time (HH:MM) = xx:xx
SET UP INSTRUMENT
Read Coil Status
Read Holding Register
Force Single Coil
Preset Single Register
Loopback Diagnostic Test
Preset Multiple Registers
Read up to 16 consecutive discrete (boolean) points from a specific point. The 8240
returns zeros for points which do not contain defined data and NAKs* any request for
point numbers greater that 100.
Up to 8 consecutive registers from a specific starting register. The 8240 returns zeros
from registers which do not contain de