ABB 8241 8240 Series Analyzers

OUT OF SERVICE ALARM 1 CAL ALARM 2 HOLD OUT OF SAMPLE Phosphate EIL 8242

Clarification of an instruction or additional information.

Information. Further reference for more detailed information or technical details. Although Warning hazards are related to personal injury, and Caution hazards are associated with equipment or property damage, it must be understood that operation of damaged equipment could, under certain operational conditions, result in degraded process system performance leading to personal injury o

INTRODUCTION ......................................................... 2

2 PREPARATION ........................................................... 2 2.1 Company Standard Settings ............................ 2 3 INSTALLATION ........................................................... 3 3.1 Serial Communication Adaptors for Personal Computers ................... 3 3.1.1 Five-wire Configuration ...................... 3

ABB 8241 This Supplement must be read in conjunction with the

Instruction Manual supplied with the instrument: Instrument Manual Reference • Model 8241 Colorimetric Silica Monitor – IM/8241 • Model 8242 Colorimetric Phosphate Monitor – IM/8242 For repeatable and reliable serial communication to take place between a master (host computer) and slaves (instruments) it is essential that the two conditions detailed in this section

Fig. 4.1 Schematic Diagram

Computer Terminal or Host Computer Chain of 8240 Monitors (Connected) Last 8240 in Chain Termination Resistor (on 8240's receive input) Observe the limitations outlined in the installation information of the relevant Instruction Manual. The maximum serial data transmission line length for both RS422 and RS485 systems is 1200m. 3.1 Serial Communication Adaptors for Personal Computers An RS

ABB 8241 Fig. 4.2 Serial Connections

…4.1 Serial Connections Motherboard See Figs. 2.4 & 2.5 of the relevant Instruction Manual 3 2 1 6 5 4 Output Board Output Board 1 – Tx+ 2 – Rx+ 3 – Common (0V) 4 – Tx– 5 – Rx– 6 – No connection A – Five-wire Connections B – Three-wire Connections 3-wire 5-wire IC32 IC31 IC33 IC34 C – Three- and Five-wire Links 6 7 1 12 1 – Tx+/Rx+ 2 – Not Connected 3 – Common (0V) 4 �

ABB 8241 Fig. 5.1 Termination Resistor Links

For all aspects other than serial data transmission the transmitter is set up as shown in the relevant Instruction Manual. Unless otherwise requested, the instrument is despatched with a transmission rate of 9600 baud and transmission line termination resistors linked-out. If the resistors are to be linked-in (see Fig. 5.1) carry out the following Section. 5.1 Termination Resistors – Fig. 5.1 For long t

Fig. 6.1 Location of Serial Data Interface Page

The general programming procedure is as detailed in the relevant Instruction Manual, but with an additional Set Up Serial Interface frame in the Set Up Instrument page. 6.1 Serial Interface Page Refer to Section 6.3 (Set Up Instrument) in the relevant Instruction Manual. CLOCK SETUP. PAGE 3.2 Date (DD/MM/YY) = xx/xx/xx Time (HH:MM) = xx:xx SET UP INSTRUMENT

Read Coil Status

Read Holding Register Force Single Coil Preset Single Register Loopback Diagnostic Test Preset Multiple Registers Read up to 16 consecutive discrete (boolean) points from a specific point. The 8240 returns zeros for points which do not contain defined data and NAKs* any request for point numbers greater that 100. Up to 8 consecutive registers from a specific starting register. The 8240 returns zeros from registers which do not contain de